Home
pin/pinless deep sensing moisture meter with spherical sensor and
Contents
1. you to check its calibration manually at any time e A display icon indicates when the meter s 9V battery is low on charge KEY FEATURES e Two measurement modes Pin conductivity based and pinless based on the fringing field effect e Spherical sensor gauges moisture level up to 4 in below a surface e Conventional test pins at end of 46 in remote cable e Pin measurement accuracy of 3 e Displays WME and relative readings simultaneously on two displays 0 to 99 9 count jumbo LCD with 0 9 in 23mm digits and 40 dot tri color LED bar graph e Manual or automatic calibration checking e Data hold display lock or freeze function e User programmable beeper function backlight function and auto power off time e Low battery warning e Includes two pairs of replacement test pins e Spare pins storage compartment WHAT S IN THE CASE The MMD950 comes in a custom molded plastic case along with a remote pin type probe and cable a 9V battery two pairs of replacement pins and this user s manual PRODUCT OVERVIEW The figure shows all of the controls indicators and physical features on the front bottom top back and right side of the MMD950 Familiarize yourself with their positions and functions before moving on to the Setup Instructions Spherical sensor Measurement mode indicator REL shown 1 Indicates beeper is enabled Indicates battery is weak Jumbo LCD display
2. ES Specialty Tools amp Instruments GENERAL TOOLS amp INSTRUMENTS 80 White Street New York NY 10013 3567 PHONE 212 431 6100 FAX 212 431 6499 TOLL FREE 800 697 8665 e mail sales generaltools com www generaltools com MMD950 User s Manual Specifications subject to change without notice 2011 GENERAL TOOLS amp INSTRUMENTS NOTICE WE ARE NOT RESPONSIBLE FOR TYPOGRAPHICAL ERRORS MAN MMD950 7 20 2011
3. Five function Mode button Four function button Battery spare pins compartment cover on back of meter Tri color 40 LED analog bar graph Q Indicates display is locked frozen Jack for remote probe Remote probe with 10mm test pins shown with protective cap off SETUP INSTRUCTIONS INSTALL BATTERY To open the battery compartment turn the meter over and lift the tab on the bottom of the battery compartment cover Remove the cover and set it aside Then plug the included 9V battery into the wired socket inside the compartment The terminals of the battery and the socket mate in only one way with the smaller male terminal plugging into the larger female terminal Close the battery compartment by replacing its cover and snapping it shut OPERATING INSTRUCTIONS To power on the meter press the button and hold it for at least three seconds To power off the meter follow the same instruction You can use the meter immediately to measure moisture levels using the factory set defaults for beeper and backlight operation and auto power off activation time By default e The beeper is enabled and beeps faster the higher the readout above 17 WME in pin measurement mode 17 in REL mode e The auto power off function is disabled e The backlight is off Raa tPteaenm mroowow rp To change the default settings for the beeper backlight or auto power function follow the Advanced Setup Instructions beginning on p 9 A
4. If the beeper has been disabled to re enable it you must power off the meter and then change the beeper s default setting using the Advanced Setup instructions in the next section To hold a measurement press the button briefly The display will freeze with the held value along with a O lock icon at upper right To unlock the display press the button briefly again ADVANCED SETUP INSTRUCTIONS In Setup mode you can change the default settings for the meter s beeper and backlight and the auto power off and calibration functions To enter Setup mode power off the meter by pressing the button and holding it for at least three seconds Then power the meter back on in a special way by simultaneously pressing the and Mode buttons and holding them for at least three seconds Doing this will cause the beeper to sound the backlight to come on and the word SEtUP to scroll across the LCD After a few seconds the LCD will read out 0 0 In this format the left digit is the Option No and the right digit is the current setting for that option For example 0 0 is shorthand for Option 0 Setting 0 The table below details the five function options available to you in Setup mode Function Action Default 0 Loads factory default settings Settings Loads user s default settings Source Backlight comes on for 30 seconds Beeper Programs beeper to beep faster the higher the moisture leve
5. PIN PINLESS DEEP SENSING MOISTURE METER WITH SPHERICAL SENSOR AND REMOTE PROBE USER S MANUAL MMD950 Please read this manual carefully and thoroughly before using this product TABLE OF CONTENTS Introduction 254 26 fede ease ed ee a 3 5 Key Features 00000 cee ee 5 What s in the Case 00 00sec eee eres 5 Product Overview c siescccewaeveedueecawees 6 7 Setup Instructions 000ee ee eee 7 Install Battery 0 002 ccs 7 Operating Instructions 200000es 7 12 Measuring Moisture Levels 5 8 9 Advanced Setup Instructions 9 11 Checking Calibration 55 11 12 Specifications 00 cee eee eee eee 12 Maintenance TipS 20000 cence eee 13 Warranty Information 000c eee anes 13 Return for Repair Policy 22000eeeee 14 INTRODUCTION Thank you for purchasing General Tools amp Instruments MMD950 Pin Pinless Deep Sensing Moisture Meter with Spherical Sensor and Remote Probe Please read this user s manual carefully and thoroughly before using the meter The MMD950 is designed for use in woodworking water damage restoration building construction and home renovation Examples include e Checking for moisture on or below the surface of carpets and sub flooring e Measuring the moisture content of wood drywall or concrete before painting wallpapering sealing or treating e Locating water leaks abo
6. ces 0 or 1 for Options 0 3 and 4 0 1 2 or 3 for Options 1 and 2 When the setting you want appears as the right digit on the display press the button to save it Each time you save a setting the display advances to the next Option To leave a setting unchanged press the button to advance to the next Option Once a setting has been chosen for Option 4 the meter will automatically exit Setup mode and enter Measurement mode You cannot access the Options out of sequence For example if you have disabled the beeper by pressing the button with 1 3 on the display you cannot change the setting without powering off the meter and re entering Setup mode The one exception to this rule is the setting of the display backlight You can turn the backlight on or off by pressing the Mode button and holding it for at least three seconds CHECKING CALIBRATION Option 4 gives you two choices for checking the meter s calibration Setting 0 lets you check the calibration manually whenever the meter is powered on Setting 1 automatically checks the calibration of the meter each time it is powered on To check the meter s calibration manually follow the Advanced Setup Instructions in the previous section to choose Setting 0 for Option 4 Next be sure the meter is operating in REL mode Then press and hold the and Mode buttons at the same time making sure that no part of your hands and fingers is within 4 in of the sphere This will ca
7. e internal or surface moisture level of a material enter REL mode by pressing the Mode button as many times as necessary until REL appears on the top line of the LCD To establish a baseline for relative readings press the spherical sensor against a material known to be completely dry This should produce a reading of 0 on both the LCD and LED bar graphs Then press the spherical sensor against one or more points on the target surface and compare the readings Some measurement tips 1 The spherical sensor s accuracy is not affected by its angle relative to the target surface Press the sphere against the material at the angle that makes the LCD easiest to read 2 Never use force to drive the test pins into a hard surface 3 Measurements of wood are skewed by two variables ambient humidity and the density of the wood species The best way to compensate for the effect of these variables is to develop your own moisture level curves based on your experience working with different species of wood on a day to day basis To scan a material for an area of peak wetness in either measurement mode WME or REL first make sure that the beeper is enabled indicated by the ses icon on the top line of the LCD If the icon is present scan the material while paying attention to the beeper s frequency The faster the beeper beeps the wetter an area is Near the high end of the meter s measurement range the beeping sounds almost constant
8. istinction between absolute and relative readings is important to understand The moisture level measurements made by the remote probe s test pins are displayed in units of WME Wood Moisture Equivalent simultaneously on a 0 99 9 count digital LCD and a 40 LED analog bar graph These pin measurements are accurate within 3 of the reading plus 5 digits By contrast readings made by the spherical sensor are simultaneously displayed with no units on the LCD and on a second scale called REL of the bar graph Although the readings from the spherical sensor have no accuracy specification they are nonetheless useful for quickly comparing the moisture levels of materials or the wetness of different areas of the same material Two possible uses for the relative readings are 1 Determining whether two pieces of wood to be joined have roughly the same moisture content which means they will dry out at the same rate without warping 2 Locating the source of a water leak above a ceiling by comparing the REL readings at various points on it If the ceiling is level the point with the highest REL reading is below the source of the leak The biggest advantage of the spherical sensor is its ability to sense moisture up to 4 in below or behind the surface of a material The remote probe s pins can also sense moisture just below the surface of some materials However for hard materials like wood or concrete measurements made by pins mostly represe
9. l above 17 WME 17 in REL mode Programs beeper to beep faster the higher the moisture level Programs beeper to sound when switching between measurement modes 3 Disables beeper Removes te icon from display Auto power 0 Disables auto power off function off 1 Sets auto power off interval to 3 minutes 2 Sets auto power off interval to 5 minutes 3 Sets auto power off interval to 10 minutes Backlight 0 Disables display backlight 1 Enables display backlight Calibration 0 Enables manual calibration checking by user Check Programs meter to automatically check its calibration each time it is powered on The factory default setting for all five functions including the source of the default settings is 0 To change any or all settings to your choice s you must first change the setting of Option 0 from 0 to 1 Opportunities to change the settings of the other four functions from their factory defaults are then made available in order from Option 0 to Option 4 As each Option is presented you can change its setting or leave the existing setting unchanged In either case you must confirm the setting to allow the sequence to continue It s important to understand that in Setup mode you have 30 seconds to confirm the setting of each Option If you take longer than 30 seconds to confirm any setting the MMD950 will automatically exit Setup mode and enter Measurement mode To change any setting use the Mode button to cycle through the choi
10. nt surface moisture content because 1 Moisture close to the surface has a greater effect on a reading than moisture deep below it and 2 The 10mm 0 4 in long pins on the remote probe are too short to be driven deep into a hard material For softer materials like soil paper or powders the pin measurements are more likely to reflect the average moisture level of the material between its surface and the penetration depth of the pins usually far less than 0 4 in Other features of the MMD950 include the following e Any reading WME or REL can be held by pushing a front panel button Doing so called locking or freezing the reading makes it possible to make a measurement in a dark place hold it and display it later in a better lighted area e The 40 LED analog bar graph groups LEDs of three different colors green yellow and red into low medium and high moisture bands Each readout Y WME or REL can be accompanied by a beeper that can be programmed to sound in different ways in response to different moisture levels or operations The correlations make it easy to use sight and or sound to pinpoint areas of peak or relative wetness The beeper can be disabled without affecting measurements e An auto power off function that can be triggered by three five or 10 minutes of inactivity e The meter can be set up to check its calibration automatically each time it is powered on Alternatively it can be programmed to allow
11. or replace this instrument if after examination the company determines it to be defective in material or workmanship This warranty does not apply to damages that General determines to be from an attempted repair by non authorized personnel or misuse alterations normal wear and tear or accidental damage The defective unit must be returned to General Tools amp Instruments or to a General authorized service center freight prepaid and insured Acceptance of the exclusive repair and replacement remedies described herein is a condition of the contract for purchase of this product In no event shall General be liable for any incidental special consequential or punitive damages or for any cost attorneys fees expenses or losses alleged to be a consequence of any damage due to failure of or defect in any product including but not limited to any claims for loss of profits RETURN FOR REPAIR POLICY Every effort has been made to provide you with a reliable product of superior quality However in the event your instrument requires repair please contact our Customer Service to obtain an RGA Return Goods Authorization number before forwarding the unit via prepaid freight to the attention of our Service Center at this address General Tools amp Instruments 80 White Street New York NY 10013 212 431 6100 Remember to include a copy of your proof of purchase your return address and your phone number and or e mail address NOT
12. ta minimum General recommends changing the default for the auto power off function to enabled If you leave the auto power off function disabled you run the risk of discharging the meter s battery if you forget to power the meter off when you are finished using it MEASURING MOISTURE LEVELS Surface moisture levels should be measured by the test pins of the remote probe if the surface can tolerate pinholes If it cannot use the meter s spherical sensor Internal moisture levels can be gauged by the spherical sensor for comparison purposes To use the remote probe remove the small rubber yellow plug from the jack on the right side of the meter callout K of the figure Then insert the white plug at the end of the cable into the jack To make a surface moisture level measurement press the Mode button as many times as necessary until WME appears on the top line of the LCD Then press the test pins against the target material The material s surface moisture level will be displayed as a percentage on both the LCD and the LED bar graph The bar graph uses LEDs of three different colors green yellow and red to indicate different moisture level ranges low medium and high respectively Readings below 17 are considered low readings between 17 and 30 medium and readings above 30 high If the beeper is enabled with the factory default setting it will beep faster the higher the moisture level rises above 17 To gauge th
13. unt with 0 9 in 23mm high digits LCD Resolution 0 1 Bar Graph Composition 40 LEDs of 3 colors green 0 to17 yellow 17 to 30 and red gt 30 Bar Graph Resolution 1 LED 2 5 Auto Power Off After 3 5 or 10 minutes user selectable with 0 reading Low Battery Warning Level lt 6 5VDC Current Consumption lt 70mADC Operating Temperature 32 to 122 F 0 to 50 C lt 80 relative humidity Dimensions 9 x 2 75 x 1 625 in 229 x 70 x 41mm Weight 8 oz 249g Power Source 9V battery included MAINTENANCE TIPS When the mm icon appears at the upper left of the LCD it s time to replace the 9V battery that powers the meter although measurements will remain valid for several hours after the icon first appears To replace the battery follow the Setup Instructions on p 7 When the tips of the remote probe s 10mm test pins show signs of wear replace them with a pair of 10mm or 15mm pins included in the case Remove the battery when storing the meter for an extended period of time Never drop or disassemble the meter or immerse it in water WARRANTY INFORMATION General Tools amp Instruments General s MMD950 Pin Pinless Deep Sensing Moisture Meter with Spherical Sensor and Remote Probe is warranted to the original purchaser to be free from defects in material and workmanship for a period of one year Subject to certain restrictions General will repair
14. use CAL to flash twice on the display and the beeper to sound twice if it is enabled If the meter is in Calibration the readout will change to 0 from its previous value and remain at 0 for 10 seconds After 10 seconds the display will resume showing real time readings in Measurement mode To program the meter to automatically check its calibration each time it is powered on follow the Advanced Setup Instructions in the previous section to choose Setting 1 for Option 4 That is all you need to do To confirm that the automatic calibration check function is enabled enter REL mode and power the meter off Then power it on and note that the sequence of initial displays includes two flashes of the letters CAL and two soundings of the beeper if it is enabled If the meter is in calibration the display will show O for 10 seconds and then begin displaying REL measurements If either calibration check produces a non zero reading for 10 seconds and the meter is still under warranty call General s Customer Service Department at 212 431 6100 to arrange to return the meter for service or replacement SPECIFICATIONS Measurement Ranges 0 to 99 9 WME and 0 to 99 REL Measurement Accuracy 3 of reading 5 digits in pin mode Length of Test Pins 10mm 0 4 in on remote probe 10mm and 15mm spares Length of Remote Probe Cable 46 in 1 168m Measurement Depth pinless mode 4 in 102mm LCD Range Size 99 9 co
15. ve ceilings below floors or behind walls e Selecting dry lumber The meter senses the moisture level of a material using either of two techniques 1 Inferring the material s electrical conductivity from the current flow induced between a pair of steel pins placed on or into the material The wetter a material the higher its conductivity The MMD950 comes with a remote probe consisting of a pair of replaceable stainless steel test pins at the end of a 46 in long cable 2 Measuring the change in the material s capacitance produced by the slight spreading of an electric field generated by the meter When the meter is powered on the spherical sensor at its top generates a three dimensional electromagnetic wave at radio frequency that extends about 4 in 100mm from the meter When the spherical sensor is placed against a wet material the increased capacitance of the material distorts the wave s electric field to an extent that can be measured This distortion called the fringing field effect is proportional to the material s moisture level WHICH SENSOR SHOULD YOU USE It depends on the application The spherical sensor is the only choice for estimating the surface or internal moisture level of materials like finished wood paint or wallpaper that cannot be marred by pinholes If accurate readings are important the remote pin type probe is the only option because the spherical sensor can only provide relative readings The d
Download Pdf Manuals