Home
        User Manual - Temperature & Process Instruments, Inc.
         Contents
1.                  Parameter Selection  SELECT   NEXT   DONE                  Home Menu    MENU   LIGHT                   If calibrator is in  Source mode    Auto Function Menu    AUTO FUNC   NEXT   DONE                           If calibrator is in If calibrator is in  RTD CUSTOM    mode    If calibrator is in    Frequency Out    or Pulse mode Pressure mode             Frequency Level  FREQ LEVEL  NEXT   DONE    RTD Custom Menu    SET CUSTOM   NEXT    DONE                   Pressure Zeroini  ZERO    NEXT   DONE                      Home Menu Home Menu Home Menu  MENU   LIGHT MENU   LIGHT MENU   LIGHT                Contrast Menu    CONTRAST  NEXT   DONE          Auto Off Menu    AUTO OFF    NEXT    DONE                   Home Menu    MENU   LIGHT             Figure 4  The Menu Tree    11          3  Using Measure Modes    3 1 Measuring volts and frequency    The calibrator can measure volts or frequency on these jacks  To make the desired  measurements  follow these steps     1  Select the  CONFIG  option from the main menu   2  Select the desired parameter for measurement   3  Connectleads as shown in Figure 5     VOLTS IN              0 001 V  Menu  uem               Figure 5  Measuring Volts and Frequency with Input Output Terminals             3 2 Measuring mA  To measure mA follow these steps   1  Select the  CONFIG  option from the main menu   2  Choose  mA  from the primary parameters   3  Make sure the input output control is set to IN   4  Connectleads as shown in Fig
2.     0 000 V    MENU   LIGHT   STEP             4 6 Sourcing frequency  To source a signal use these steps     1     oa FON    Select the  CONFIG  option from the main menu and select frequency from the primary  parameters     Select output  and than choose the frequency units   Connect the leads to the frequency output terminals as shown in Figure 15   Enter the desired frequency using the keypad   To change the amplitude  select  FREQ LEVEL  from frequency level menu   Enter the amplitude   19    4 7 Sourcing a pulse train    The calibrator can produce a pulse train with an adjustable number of pulses at a desired  frequency  For example  setting the frequency to 60Hz and the number of pulses to 60 would  produce 60 pulses for a period of 1 second  To source a pulse  use the same connection as  for frequency  and proceed as follows     1     Select the  CONFIG  option from the main menu and select pulse from the primary  parameters        2  Choose the desired unit and enter the frequency using the keypad   3  Select the  COUNTS  function from the home menu to enter the number of pulses  Use   TRIG  to start and stop the signal   4  The amplitude of the pulse can be adjusted in the same manner as for frequency   PULSE 5 0 Vpp FREQ OUT 5 0 Vpp  MENU TRIG COUNTS MENU   LIGHT   STEP                   TC PLUG                                                                Note  TC wire used must match  the thermocouple type being calibrated     Figure 15  Connections for Outputti
3.    Floating  Numbers that have up to 15 significant figures and exponents  For example   CPRT_COEFA  returns 3 908000E 03    Character Response Data  CRD   Data returned as keywords  For example   RTD TYPE  returns PT385 10    Indefinite ASCII  IAD   Any ASCII characters followed by a terminator  For example    IDN  returns MARTEL  MC1010  250  1 00    25    26    5 3 4 Calibrator Status    Status registers  enable registers  and queues provide status information on the calibrator   Each status register and queue has a summary bit in the Serial Poll Status Byte  Enable  registers generate summary bits in the Serial Poll Status Byte  The following is a list of  registers and queues along with their function     Serial Poll Status Byte  STB     The STB is sent when the calibrator responds to the  STB  command  Figure 24  demonstrates how it functions  Cleared when power is reset     Service Request Enable Register  SRE     Enables or disables the bits of the STB  Cleared when power is reset  Setting bits to 0  disables them in the STB  Setting the bits to 1 enables them  Bit assignments for the SRE and  the STB are shown below           0 MSS   ESB 0 EAV 0 0 0                                  MSS    Master Summary Status  Set to 1 when ESB or EAV are 1  enabled   Read using  the  STB  command     ESB  Set to 1 when at least one bit in ESR is 1     EAV    Error Available  An error has been entered into the error queue  and may be read  using the Fault  command     Event Status R
4.    MARTEL    LECTRONICS    MC1010    Reference Manual       1  Introduction aii cc ed ee is Ce ee Ee EORR OX AAA 3    ded Custome SORViGO    ust oir Lu EAE La Haut al etal aig  at aos 3  1 2 Standard Equipment          0    ehh hh 3  1 8 Safety information  A EEan aaa 4  2  Calibrator Interface                neuuunnnanann nan nan nan nn nn nan nn nn 6  2 1 Main Display  sssmassas a Em ar a t aa 7  VNB E kh a t a aa aa ann 00 8  2 3 Cursor control   Setpoint control morros Ro RR Ra RR ama 11  3  Using Measure Modes                  n n nnnununannn nn na nan nn nn nan nn 12  3 1 Measuring volts and frequency                                  e eeeeeeee 12  3 2 Meas  ring MA       iiusuuiuduiusuiaraciad oh R3 AR ROG sm DOR RC E CERE A EC ROY A 3H 12  3 8 Measuring mA with Loop Power    o o ooooooocc nauan aa aaa 13  3 4 Measuring Temperature      o o o o oooooooo leer 13  3 5 Measuring Pressure  ooo A ninn 15  4  Using Source Modes                    unuununnannan naa nan n nan ann na 16  4 1 Setting 0  and 10096 Output Parameters       oooooccccconcccn 16  4 2 Using the Automatic Output Functions                          n uu  rennuna 17  4 3  Sourcing MA  Cama Ropa do dora d ed eA ded eee Se VRE RED RRS REX  2 17  4 4 Simulating a Transmitter    ees 18  4 5 Sourcing VONS icc 00 or he Pa quaa paa dede dada Que NN RETE ERE RR RARE E 19  4 6 Sourcing fr  quency vss ss scu usu a A AO UU RR RE EOS RO RARE S ERE EXER 19  4 7 SOUICING A pulse tralh AAA e e XR AA NUN DER 20  4 8 Sourc
5.   Choose 2  3  or 4 wire connection  2W  3W  4W    4 wire allows for the most precise  measurement     Select RTD type from the sensor types     Attach RTD leads as shown in Figure 9                    RTD IN 4W P100 385  100 00 Q             0 0     C  meno Tucan T                              Figure 9  Measuring Temperature with RTD Connection    Resistance can also be measured using this function  To do so  use the above procedure  and choose OHMS from the sensor types  This option can be used along with a table to  measure an RTD which is not programmed into the calibrator     3 5 Measuring Pressure    Note  The MC1010 is compatible with BETA Calibrator Pressure Modules  The accessory  BPPA 100 is required for pressure measurement     Note  Pressure is not read from modules with frequency or pulse train mode enabled   Note  On high pressure modules engineering units normally associated with low pressure  ranges such as  inH O  cmH O  etc  are not valid selections  Selecting one of this units with a  high pressure module attached will cause the display to read           Warning   To avoid a violent release of pressure in a pressurized system  shut off the valve and slowly  bleed off the pressure before you attach the pressure module to the pressure line    Caution    To avoid mechanically damaging the pressure module  never apply more than 10 ft lb  of  torque between the pressure module fittings  or between the fittings an the body of the module     To avoid damaging the 
6.  10uV       Maximum current output in voltage ranges is 1mA with an output impedance of  lt   12       Table 16  Thermocouple Read and Source  errors in  C                                            Range    C  Accuracy    C   TC Type Minimum Maximum CJC OFF CJC ON  J  210 0  150 0 0 4 0 6   150 0 1200 0 0 2 0 4  K  200 0  100 0 0 5 0 7   100 0 600 0 0 2 0 4  600 0 1000 0 0 3 0 5  1000 0 1372 0 0 4 0 6                                                                                                 Range    C  Accuracy    C   TC Type Minimum Maximum CJC OFF CJC ON   T  250 0  200 0 1 5 1 7   200 0 0 0 0 5 0 7   0 0 400 0 0 2 0 4   E  250 0  200 0 1 0 1 2   200 0  100 0 0 3 0 5    100 0 1000 0 0 2 0 4   R  0 0 200 0 1 7 1 9  200 0 1767 0 1 0 1 2   S 0 0 200 0 1 7 1 9  200 0 1767 0 1 1 3   B  600 0 800 0 1 5 1 7  800 0 1000 0 1 2 1 4   1000 0 1820 0 1 0 1 2   C 0 0 1000 0 0 5 0 7  1000 0 2316 0 1 5 1 7   XK  200 0 800 0 0 2 0 4  BP 0 0 800 0 1 9 2 1  800 0 2500 0 0 6 0 8   L  200 0 900 0 0 2 0 4   200 0 0 0 0 4 0 6   0 0 600 0 0 2 0 4   N  200 0  100 0 0 8 1 0   100 0 1300 0 0 3 0 5   CJC error outside of 23   5  C is 0 05  C   C          1  For thermocouple measure mode on B  R and S  round the specification up or down accordingly as there  is no resolution past the decimal point     Table 17  RTD Read and Source                                        RTD Type Range    C  Accuracy    C   Minimum Maximum Measure 4W Source   Measure 3W  PT385  10 ohm  200 0  80 0 1 3 2 5   80 0 0 
7.  RTD OUT       4 9 1 Custom RTD    A custom curve fit PRT may be entered into the calibrator for sourcing and measuring  To  do so follow these steps     1  Select the  CONFIG  option from the main menu  Select RTD and set sensor type  to CUSTOM     2  Enter the RTD custom setup main menu  and select  SET CUSTOM      3  Using the keypad  enter the values that the calibrator prompts for  minimum  temperature  maximum temperature  RO  and the values for each of the  temperature coefficients     The custom function uses the Calendar Van Dusen equation for outputting and measuring  custom RTDs  The coefficient C is only used for temperatures below 0  C  Only A and B  coefficients are needed for the range above 0  C  so coefficient C should be set to 0  The  RO is the resistance of the probe at 0  C  The coefficients for PT385  PT3926  and PT3616  are shown in Table 4 below     Table 4  RTD Coefficients                            RTD Range   C  RO Coefficient A Coefficient B Coefficient C  PT385  260   0 100 3 9083x10 3  5 775x10 7  4 183x10 12  PT385 0   630 100 3 9083x10 3  5 775x10 7       PT3926 Below 0 100 3 9848x10 3  5 87x10 7  4x10 12  PT3926 Above 0 100 3 9848x10 3  5 87x10 7       PT3916 Below 0 100 3 9692x10 3  5 8495x10 7  4 2325x10 12  PT3916 Above 0 100 3 9692x10 3  5 8495x10 7                          5  Remote Operation    22    The calibrator can be remotely controlled using a PC terminal  or by a computer program  running the calibrator in an automated system  It
8.  on the unit while holding down the    HOME    key   2  Hold the key down for a count of 10 seconds to restore contrast default settings     If the display is so dim that you cannot tell if the unit is on or off  use the backlight key to  determine if the power is on or off     CONTRAST  NEXT   DONE       The auto off main menu contains the options  AUTO OFF    NEXT   and  DONE      The  AUTO OFF  option is used to turn the automatic shutoff on and off and set the amount of  time the unit needs to stay dormant to shut off   NEXT  and  DONE  both return to home menu     AUTO OFF   NEXT   DONE    When the calibrator is in the frequency or pulse mode  the frequency level menu is added  after the main menu  The options available in this menu are  FREQ LEVEL    NEXT   and   DONE   The  FREQ LEVEL  option is used to adjust the amplitude of the wave   NEXT  is  used to access the contrast main menu  and  DONE  returns to the home menu     FREQ LEVEL  NEXT  DONE    When the calibrator is in RTD CUSTOM mode  the RTD custom setup menu  is inserted after  the main menu  Options  SET CUSTOM    NEXT   and  DONE  are available   SET CUSTOM   is used to enter a custom PRT into the calibrator  Refer to Section 4 1 8a for instructions    NEXT  is used to enter the contrast main menu  and  DONE  to return to the home menu     SET CUSTOM  NEXT   DONE    The pressure zeroing main menu is the final variation to choosing  MORE  in the main menu     It has the options  ZERO     used to zero pressur
9.  uses an RS 232 serial port connection for  remote operation  With this connection the user can write programs on the PC  with Windows  languages like Visual Basic to operate the calibrator  or use a Windows terminal  such as       Hyper Terminal  to enter single commands  Typical RS 232 remote configurations are shown  in Figure 18     COM PORT USB PORT    e GE o  o                     G   G   G     a    O    0000  0000  rs   29                                OO  OO    Optional RS 232 or USB Cable   See section 7 4        Figure 18  Calibrator to Computer Connection    5 1 Setting up the RS 232 Port for Remote Control    Note  The RS 232 connection cable should not exceed 15m unless the load capacitance  measured at connection points is less than 2500pF     Serial parameter values     9600 baud   8 data bits   1 stop bit   no parity   Xon Xoff   EOL  End of Line  character or CR  Carriage Return  or both    To set up remote operation of the calibrator on the Windows Hyper Terminal  connected to a  COM port on the PC as in Figure 23  use the following procedure     1     a FON    Start Hyper Terminal  located in Accessories Communications of the Windows Start  menu     Select New Connection   For Name enter MC1010  Select the serial port that the unit is connected to   Enter the above information for port settings   Select ASCII setup from File Properties Settings and mark these choices   Echo typed characters locally  Wrap lines that exceed terminal width  Select Ok  To see 
10.  zero offset or the reference value for absolute pressure modules     39    6  Specifications    One year at 23  C   5  C  unless specified otherwise  Outside of this range the stability of the  measurements is   0 003  of FS   C     Table 9  General Specifications       Operating Temperature     10  C to 50  C       Storage Temperature     20  C to 60  C       Power    4 X AA batteries  Alkaline or optional rechargeable       Battery Life    Approximately 20 hours       Serial Communications    ASCII  RS232 or USB depending on the communications cable used       CE   EMC    EN61326  2006  ANNEX B Criterion A       Safety    CSA C22 2 No  1010 1  1992       Size  with holster     8 5  x 4    x 2     21 6cm x 10 2cm x 5 1cm        Weight       Approximately 280z   0 8kg        Table 10  DC Voltage Measurement Source                Range Accuracy    of reading   floor   Read 0 000V   20 000V 0 015    2mV  Source 0 000V   20 000V 0 015    2mV             Maximum current output in voltage ranges is 3mA with an output impedance of  lt   1           Table 11  DC mA Measurement Source          Range Accuracy    of reading   floor   Read 0 000mA   24 000mA 0 015    2uA  Source   0 000mA 24 000mA   0 015    2uA          Maximum load on mA source is 10002  Voltage input range on simulate mode 5V   30V        For units configured to support 50mA the table below shows the specification           Range Accuracy    of reading   floor   Read 0 000mA   52 000mA 0 02    4uA  Source 0 000mA   
11. 0 0 2 0 2  100 0 260 0 0 2 0 2  260 0 300 0 0 3 0 3  300 0 400 0 0 3 0 3  400 0 630 0 0 4 0 4  NI120  80 0 260 0 0 1 0 2  Cu10  100 0 260 0 1 3 2 5  Cu50  180 0 200 0 0 3 0 5  Cu100  180 0 200 0 0 1 0 3  YSI 400 15 0 50 0 0 1 0 1                7  Maintenance   Warranty    7 1 Replacing Batteries    Replace batteries as soon as the battery indicator turns on to avoid false measurements  If  the batteries discharge too deeply the MC1010 will automatically shut down to avoid battery    leakage     Note  Use only AA size alkaline batteries or optional rechargeable batteries     Warning    This product has non rechargeable alkaline batteries installed     Prior to connecting any AC adapter or battery charger  these batteries must be removed  and or replaced with rechargeable batteries such as Nickel Cadmium or Nickel Metal Hydride    types     Failure to remove the batteries will result in damage to the calibrator and voids the warranty     7 2 Cleaning the Unit    Warning    To avoid personal injury or damage to the calibrator  use only the specified replacement parts  and do not allow water into the case     Caution    To avoid damaging the plastic lens and case  do not use solvents or abrasive cleansers     Clean the calibrator with a soft cloth dampened with water or water and mild soap     44    7 3 Service Center Calibration or Repair    Only qualified service personnel should perform calibration  repairs  or servicing not covered  in this manual  If the calibrator fails  c
12. 0 1 3 2 5  0 0 100 0 1 4 2 7  100 0 300 0 1 5 3 0  300 0 400 0 1 6 3 1  400 0 630 0 1 8 3 3  630 0 800 0 1 9 3 6          42                                                                                                                               RTD Type Range    C  Accuracy    C   Minimum Maximum Measure 4W Source   Measure 3W  PT385  50 ohm  200 0  80 0 0 3 0 5   80 0 0 0 0 4 0 7  0 0 100 0 0 4 0 7  100 0 300 0 0 4 0 7  300 0 400 0 0 5 0 7  400 0 630 0 0 5 0 8  630 0 800 0 0 6 0 9  PT385  100 ohm  200 0  80 0 0 1 0 3   80 0 0 0 0 2 0 4  0 0 100 0 0 2 0 4  100 0 300 0 0 2 0 4  300 0 400 0 0 3 0 5  400 0 630 0 0 3 0 5  630 0 800 0 0 4 0 6  PT3926  100 ohm  200 0  80 0 0 1 0 2   80 0 0 0 0 2 0 3  0 0 100 0 0 2 0 3  100 0 300 0 0 2 0 3  300 0 400 0 0 3 0 4  400 0 630 0 0 3 0 5  PT3916  100 ohm  200 0  80 0 0 1 0 2   80 0 0 0 0 2 0 3  0 0 100 0 0 2 0 3  100 0 260 0 0 2 0 3  260 0 300 0 0 3 0 4  300 0 400 0 0 3 0 4  400 0 630 0 0 3 0 5  PT385  200 ohm  200 0  80 0 0 6 0 6   80 0 0 0 0 7 0 7  0 0 100 0 0 7 0 8  100 0 260 0 0 7 0 8  260 0 300 0 0 7 0 8  300 0 400 0 0 9 0 9  400 0 630 0 0 9 1 0  PT385  500 ohm  200 0  80 0 0 2 0 3   80 0 0 0 0 3 0 3  0 0 100 0 0 3 0 3                43                                                                RTD Type Range    C  Accuracy    C   Minimum Maximum Measure 4W Source   Measure 3W  100 0 260 0 0 4 0 4  260 0 300 0 0 4 0 4  300 0 400 0 0 4 0 5  400 0 630 0 0 5 0 5  PT385  1000 ohm  200 0  80 0 0 2 0 2   80 0 0 0 0 2 0 2  0 0 100 
13. 3  Bit 6 is not  used      SRE   Returns a byte from the SRE  The byte is returned in decimal format  For example   If 40 is returned  bits 5 and 3 are enabled      STB  Returns the status byte in decimal form from the Serial Poll Status Byte  For example   If 72 is returned  bits 6 and 3 are enabled      WAI    Prevents further remote commands from being executed until all previous commands are  executed  For example     OUT 10 MA    WAI   OUT 5 V will out put 10mA and wait until output settles  than volts  command will be processed     5 5 2 Calibrator Commands    CAL START    Puts the calibrator in calibration mode  The main display will say CALIBRATION MODE and a  calibration menu will be displayed on the terminal     CJC STATE    Turns Cold Junction Compensation  CJC  on or off  when the calibrator is in thermocouple   TC  mode  The command is used by adding ON or OFF after it     For example   CJC STATE OFF turns CJC off     CJC STATE     Tells whether the Cold Junction Compensation in thermocouple mode is turned on or turned  off  The calibrator returns OFF if CJC is off  and ON if CJC is on     CPRT COEFA    This command is used for entering a custom RTD into the calibrator  The numeric value  entered after the command will be set as the first coefficient of the polynomial used by the  custom RTD     For example   CPRT COEFA 3 908E 03 enters 3 908e 3 as coefficient A     CPRT COEFA     Returns the number which was entered for the first coefficient of the polynomial us
14. 52 000mA 0 02    4uA             Maximum load on mA source is 400  at 52mA  Voltage input range on simulate mode 5V   60V           40             Table 12  Frequency Measurement Source                            Range Accuracy    of reading   floor   Read 2 0CPM   600 0CPM 0 05    0 1CPM  1 0Hz   1000 0Hz 0 05    0 1Hz  1 00KHz   10 00KHz 0 05    0 01KHz  Source 2 0CPM   600 0CPM 0 05   1 0Hz   1000 0Hz 0 05   1 00KHz   10 00KHz 0 250        Input voltage amplitude range on frequency is 1V to 20V zero based square wave only Output amplitude is  adjustable from 1V to 20V  and is a square wave with 50  duty cycle For output frequency  a slight negative  offset of approximately  0 1V is present to assure zero crossing        Table 13  Resistance Measurement                   Range Accuracy    of reading   floor   Ohms low 0 0092   400 092 0 025    0 052  Ohms high 401 0Q   4000 00 0 025    0 59       Table 14  Resistance Source                      Range Excitation Current Accuracy    of reading   floor   Ohms low 5 0Q   400 092 0 1mA   0 5mA 0 025    0 12  5 0Q   400 09 0 5mA   3mA 0 02596   0 05     Ohms high 4002   15002 0 05mA   0 8mA 0 025    0 5Q  1500Q   4000    0 05mA   0 4mA 0 025    0 5Q       Note  Unit is compatible with smart transmitters and PLCs  Frequency response is  lt   5ms        Table 15  Thermocouple Measurement Source                Range Accuracy    of reading   floor   Read  mV   10 000mV   75 000mV 0 02    10uV  Source  mV   10 000mV   75 000mV 0 02   
15. and off using the Output Home Menu  The Automatic Output Function parameters  can be set in the Auto Function Menu    Parameters include     1  Which auto function will be available  Step or Ramp      2  The Auto Function Time  time between steps for step and time to get from over one limit  to the next for ramp     3  Step Size  the step size can be set to either 1096 or 25     The limits for the ramp and step functions are set to the 0  and 100  values  See Section  4 1 Setting 0  and 100  Output Parameters  Steps are in either 10  or 25  increments from  the 0  value to the 100  value     4 3 Sourcing mA  To source a current  follow these steps     1  Select the  CONFIG  option from the main menu  Choose  mA  from the primary  parameters     2  Switch to input output control  and select output  OUT    3  Connect the leads to the mA terminals  as shown in Figure 11   4  Enter the desired current using the keypad     1000 ohms max     MEASURE   SOURCE              Figure 11  Connections for Sourcing Current    mA OUT  25 00         0 000 mA  MENU uem                       17    4 3 1 HART   Resistor Selection    The MC1010 can be set up so that the 250 ohm resistor required for Hart    configuration  devices resides inside the MC1010  Enabling the MC1010 s internal 250 ohm resistor  eliminates the need to manually add a series resistor during a Hart    calibration process     NOTE  When the MC1010 s internal 250 resistor is enabled  maximum load driving capability  drops fro
16. e   NEXT  and  DONE   which have the  same function as above  Refer to the Section 5 3 for instructions on zeroing                                            ZERO    NEXT   DONE    The parameter selection menu is called up when  CONFIG  is selected from the main menu   It contains the following options   SELECT    NEXT   and  DONE   When the display is  selected  a parameter will start to flash  Use the  SELECT  option to change the parameter   and the  NEXT  option to switch to another variable   DONE  returns to the home menu and  enables the selected mode                 SELECT   NEXT   DONE             2 3 Cursor control   Setpoint control    The output value can be controlled by the four cursor control arrows on the keypad  By  pressing one of the arrows a cursor will be added to the display under the last digit of the  output value  The left and right arrow keys are used to select which decade to be changed in  the output value  The up and down arrow keys are used to increase  decrease  or ramp the    output value     The menu bar will change to the setpoint menu with the touch of any one of the four arrow    keys           0  25  100           The three function keys are associated with 0  25  and 100  values  respectively  0 and 100   values can be stored by entering a value and then holding down the corresponding function  key  The 25  key will then automatically step through the 25  values        Home Menu    MENU   LIGHT    Selection Menu  CONFIG MORE              
17. e options is  explained in Section 4 2 6  Sourcing a Pulse      MENU   TRIG   COUNTS    The next level of the menu bar is the main menu  The levels under the main menu depend on  what mode the calibrator is in     The main menu has two active options  CONFIG  and  MORE      Choosing  CONFIG  calls up the parameter selection menu   MORE  enters the next menu  level                                CONFIG MORE    The Auto Function Menu is the next menu if you are in source mode  Its options are  AUTO  FUNC    NEXT  and  DONE    AUTO FUNC  allows you to adjust the Automatic Output  Function parameters   NEXT  proceeds to the next menu level and  DONE  returns to the  home menu  See Section 4 2 Using the Automatic Output Functions     AUTO FUNC   NEXT   DONE    The contrast menu is usually the next menu level  Its options are  CONTRAST    NEXT   and   DONE   The  CONTRAST  option is used to adjust contrast   NEXT  proceeds to the auto off  main menu  and  DONE  returns to home menu  Contrast is adjusted using the arrow options   which are available after choosing  CONTRAST      NOTE  The MC1010 calibrator offers a wide range contrast adjustment feature to  accommodate operation in extreme temperatures                 In certain cases making large changes in contrast may render the display difficult to read  under normal conditions  If this occurs and the display is too dim or dark to read  proceed  with the following process to set the contrast back to a default setting     1  Turn
18. ed in the  custom RTD  Using the example above CPRT COEFA  Would return     3 908000E 03    CPRT COEFB    This command is used for entering a custom RTD into the calibrator  The numeric value  entered after the command will be set as the second coefficient of the polynomial used by the  custom RTD     For example   CPRT COEFB  5 8019E 07 enters  5 8019e 7 as coefficient B     CPRT COEFB     Returns the number  which was entered for the first coefficient of the polynomial used in the  custom RTD  Using the example above  CPRT COEFB  Would return      5 801900E 07    33    34    CPRT COEFC    This command is used for entering a custom RTD into the calibrator  The numeric value  entered after the command will be set as the first coefficient of the polynomial used by the  custom RTD     For example   CPRT COEFC  5 8019E 12 enters  5 8019e 12 as coefficient A   CPRT COEFC     Returns the number which was entered for the first coefficient of the polynomial used in the  custom RTD  Using the example above CPRT COEFC  Would return      5 801900E 12    CPRT MIN T    Sets the minimum temperature of the custom RTD range  The temperature value must be  entered with a degrees label  CEL for Celsius and FAR for Fahrenheit     For example   CPRT MIN T  260 CEL enters  260 C as the minimum temperature     CPRT MIN T     Returns the value entered for minimum temperature in the range for a custom RTD  Note that  the Calibrator always returns numbers in scientific notation  The above example wo
19. egister  ESR     A two byte register  in which the lower bits represent conditions of the Calibrator  Cleared  when read and when power is reset     Event Status Enable Register  ESE     Enables and disables bits in the ESR  Setting a bit to 1 enables the corresponding bit in the  ESR  and setting it to 0 disables the corresponding bit  Cleared at power reset  Bit  assignments for the ESR and the ESE respectively are shown below        15 14 13 12 11 10 9 8                                  0 0 0 0 0 0 0 0                                           T 6 5 4 3 2 1 0  PON 0 CME  EXE   DDE   QYE 0 OPC  PON  Power On  Set to 1 if power was turned on and off before the Event Status Register  was read   CME    Command Error  Set to 1 when the calibrator receives an invalid command   Entering an unsupported RTD type may cause such an error     EXE    Execution Error  Set to 1 when the calibrator runs into an error while executing is  last command  A parameter that has too significant figures may cause such an  error     DDE    Device dependent Error  Set to 1 when  for example  the output of the calibrator is  overloaded     QYE  Query Error     OPC    Operation Complete  Set to 1 when the calibrator has finished executing all  commands before the command  OPC was entered     Error Queue    If an error occurs due to invalid input or buffer overflow  its error code is sent to the error  queue  The error code can be read from the queue with the command FAULT   The error  queue holds 15 e
20. he output would be   5 000000E 03  A    TEMP_UNIT    Sets the temperature unit for sourcing and measuring RTD and TC  Add CEL after the  command for Celsius  and FAR for Fahrenheit  For example     TEMP_UNIT CEL sets the temperature to be measured or sourced to Celsius     TEMP_UNIT   Returns the temperature unit that is currently used for measuring and sourcing RTD and TC     TRIG    Starts and stops the pulse train when the calibrator is in pulse mode  The parameters of the  pulse train are set by commands PULSE_CNT  and FREQ_LEVEL  Entering TRIG initializes  the train  Entering the command while the pulse train is running stops it     TRIG     Returns TRIGGERED if the pulse train is running  and returns UNTRIGGERED when the pulse  train is not running  Returns NONE when the calibrator is not in pulse mode     TSENS_TYPE    Sets the temperature sensor type to thermocouple  or to RTD for temperature measurement   After the command add TC for thermocouple  or RTD for RTDs  For example     TSENS TYPE TC sets the sensor type to thermocouple    TSENS_TYPE   Returns the type of sensor that is currently set to measure temperature  either TC or RTD     VAL     Returns the value of any measurement taking place on the display  For example  if the display  is measuring 10V  then VAL  will return     1 000000E  01  V    ZERO MEAS    Zeroes the attached pressure module  Enter the zeroing value in PSI after the command  when zeroing an absolute pressure module     ZERO MEAS   Returns the
21. heck the batteries first  and replace them if needed     Verify that the calibrator is being operated as explained in this manual  If the calibrator is  faulty  send a description of the failure with the calibrator  Be sure to pack the calibrator  securely  using the original shipping container if it is available  To return the product to Martel  go to www martelcalibrators com and click on the Customer Service link  Download the RMA  form and follow the instructions  Calibration procedures and other related documents can  also be found on the website     7 4 Replacement Parts  amp  Accessories    700MA Fluke Module Adapter  BPPA 100 BETA Pressure Module Adapter  80029 JKTE T C Wire Kit  Mini Plugs  80036 RSNB T C Wire Kit  Mini Plugs  CC572 Carrying Case  C232SJ RS232 Cable with Miniature Plug  1919896 USB RS232 Converter Cable with Miniature Plug  5353050 48  Stackable Testleads  1001 9V Battery Adapter Charger  7 5 Warranty    Martel Electronics Corporation warrants all products against material defects and  workmanship for a period of twelve  12  months after the date of shipment  Problems or  defects that arise from misuse or abuse of the instrument are not covered  If any product is to  be returned  a    Return Material Authorization    form can be obtained from our website  www martelcorp com under customer service  You can also call 1 800 821 0023 to have a  form faxed  Martel will not be responsible for damage as a result of poor return packaging   Out of warranty re
22. if the port works enter  IDN   This command will return information on the unit     23    5 2 Changing Between Remote and Local Operation    There are three modes of operation of the calibrator  Local  Remote  and Remote with  Lockout  Local mode is the default mode  Commands may be entered using the keypad on  the unit or using a computer  In Remote mode the keypad is disabled  and commands may  only be entered using a computer  but choosing  GO TO LOCAL  from the menu on the  calibrator display will restore keypad operation  In Remote with Lockout  the keypad can not  be used at all  To switch modes proceed as follows     1  To enable Remote mode  type in the serial command REMOTE at the computer terminal   2  To enable Remote with Lockout  type in REMOTE and LOCKOUT in either order     3  To switch back to local operation enter LOCAL at the terminal  This command also turns off  LOCKOUT if it was on  For more information on commands refer to the Remote Commands  section     5 3 Using Commands    5 3 1 Command types  Refer to the Section on Remote Commands for all available commands     The calibrator may be controlled using commands and queries  All commands may be  entered using upper or lower case  The commands are divided into the following categories     Calibrator Commands  Only the calibrator uses these commands  For example  MEAS DCV  tells the calibrator to measure voltage     Common Commands    Standard commands used by most devices  These commands always begin wi
23. ime  If a new absolute  module is connected this process must be repeated        PRESSURE            psi    MENU   LIGHT               4  Using Source Modes    The calibrator can generate calibrated signals for testing and calibrating process instruments   It can source voltages  currents  resistances  frequencies  pulses  and the electrical output of  RTD and thermocouple temperature sensors     4 1 Setting 096 and 10096 Output Parameters  To set the 0  and 100  points  use the following steps     1     T     Select the  CONFIG  option from the main menu  and choose the desired primary  parameter     Select output  OUT  from the input output control  and enter the desired value  For  example select  VOLTS OUT      Enter 5V with the keypad and press Enter   Press any one of the four cursor control arrows to display the setpoint control menu     Hold down the Function Key that corresponds to 0   F1   0  will flash and the setpoint is  stored     Repeat these steps  entering 20V and holding the Function Key that corresponds to 10096   F3      Use the  F2  key to step in either 2596 or 1096 increments     4 1 1 Stepping the current output    To use the 25  function with mA output  follow these steps   1  Select the  CONFIG  option from the main menu  and choose mA   2  Use the 2596 key to cycle between 4 mA and 20 mA in 25 96 intervals     4 2 Using the Automatic Output Functions    There are two automatic output functions  step and ramp  The selected function can be  turned on 
24. ing Themocouples    sisssusus s A e 20  4 9 Sourcing OHMSIRIDS  los maal aa aaa pa S PRU A A Eee 21  5  Remote Operation                eens 22  5 1 Setting up the RS 232 Port for Remote Control               liliis 23  5 2 Changing Between Remote and Local Operation              o oooooooooooooo   24  5 3 Using Commands coco 24  5 4 Remote Commands and Error Codes         0    cette 27  5 5 Entering Commands       0 00 tee 31  6  Specifications                 nnusnsnnunnnanunn nan nan nan n nn ann nn na 40  7  Maintenance   Warranty                  nununnnnn na na nan nunnana nan nn 44  7 1 Replacing Batteries    sisse A ua PERRET ES 44  7 2 Cl  aning the Umit 2    a ra UR aa ua mu a a d 44  7 3 Service Center Calibration or Repair             iilii 44  7 4 Replacement Parts 8 Accessories      45    5 Warranty  AS 45    1  Introduction    The Martel MC1010 Multifunction Process Calibrator is a handheld  battery operated   instrument that measures and sources electrical and physical parameters  The calibrator has   the following features and functions      A graphics display  The display can be used to source and measure volts  current   pressure  resistance temperature detectors  RTDs   thermocouples  frequency  and  resistance  and to source pulse trains      Athermocouple  TC  input output terminal with automatic reference junction temperature  compensation       Setpoints in each range for quickly increasing decreasing output    An interactive menu    Complete serial in
25. ipment safety procedures       Never touch the probe to a voltage source when the test leads are plugged into the current  terminals       Do not use the calibrator if it is damaged  Before you use the calibrator  inspect the case   Look for cracks or missing plastic  Pay particular attention to the insulation surrounding the  connectors       Select the proper function and range for your measurement     Make sure the battery cover is closed and latched before you operate the calibrator     Remove test leads from the calibrator before you open the battery door       Inspect the test leads for damaged insulation or exposed metal  Check test leads continuity   Replace damaged test leads before you use the calibrator       When using the probes  keep your fingers away from the probe contacts  Keep your fingers  behind the finger guards on the probes       Connect the common test lead before you connect the live test lead  When you disconnect  test leads  disconnect the live test lead first       Do not use the calibrator if it operates abnormally  Protection may be impaired  When in  doubt  have the calibrator serviced       Do not operate the calibrator around explosive gas  vapor  or dust       When using a pressure module  make sure the process pressure line is shut off and  depressurized before you connect it or disconnect it from the pressure module       Disconnect test leads before changing to another measure or source function     When servicing the calibrator  use only 
26. ister  For  example      ESE 133  When 133 is converted to binary it is 10000101  Bits 7  2  and 0 will be enabled      ESE   Returns the contents of the Event Status Enable register  The value returned is a decimal  For  example  if the register has the following settings     10000101 than the value returned will be 133      ESR   Returns the contents of the Event Status Register in decimal form  For example   If the ESR contains 10111001   ESR  will return 185     31    32     IDN     Returns the manufacturer  model number  and firmware revision of the Calibrator  For  example      IDN  will return MARTEL  MC1010  250  1 00   OPC    Enables the Operation Complete setting in the ESR  This setting makes it possible to check if  an operations is complete after it has been initialized     For example this operation could be used with the command TRIG      OPC   Returns 1 when all operations are complete  and causes program execution to pause until all  the operations are complete  For example    TRIG    OPC  will return a 1 when the pulse train initiated by TRIG is complete      RST    Resets the state of calibrator to the power up state  All subsequent commands are held off  until the execution of the command is complete      SRE    Loads a byte into the Service Request Enable register  A decimal number must be entered   which when converted to binary  corresponds to the correct settings     For example      SRE 8 enters the binary number 00001000 to the SRE  This enables bit 
27. lse modes   FREQ LEVEL  with the above example would return   5 000000E 00  V    FREQ TYPE    When in frequency mode  sets the calibrator to output a continuous wave  Frequency Out   or  a pulse train  To set the calibrator to continuous wave enter CONT after the command  To set  the calibrator to pulse enter PULSE after the command  For example     FREQ TYPE CONT will set the calibrator to FREQ OUT     Note  This command does not put the calibrator in frequency mode  Use the OUT command  to put the calibrator in frequency mode     FREQ TYPE     Tells whether calibrator is sourcing a pulse or a continuous wave  The command will return  CONT if the calibrator is in FREQ OUT mode  and PULSE if the calibrator is in PULSE mode     FREQ UNIT    Sets the unit for frequency  There are three ranges of frequencies for frequency and pulse  modes  CPM  cycles per minute   Hz  and kHz  Use this command to select the right range   For example     FREQ UNIT HZ sets the frequency to Hz range    35    36    FREQ UNIT   Returns the frequency unit currently being used by the frequency and pulse modes     FUNC     Returns the current mode of the display  For example if the calibrator is set to volts FUNC   Would return     DCV    LOCAL    Restores the calibrator to local operation if it was in remote mode  Also clears LOCKOUT if the  unit was in lockout mode     MEAS    Sets the display to measure mode  The command is followed by any of the parameters  except for pulse and mA sim  which are sou
28. m 1000 ohms  2 20mA to 750 ohms  20mA     Enable Disable Procedure  1  Remove the battery cover and remove the 2 screws that are at the top of the case   2  Remove the 2 screws on the bottom or lower portion of the case   3  Gently remove the top half of the case from the bottom   4  Figure 12  shows the location of the Hart   jumpers           HART  Resistor                                                                                                          D  o0 i    circuit board                Figure 12     4 4 Simulating a Transmitter    To have the calibrator supply a variable test current to a loop in place of a transmitter  follow  these steps     1  Select the  CONFIG  option from the main menu     2  Choose mA simulation from the primary parameters  mA 2W SIM   and enter the desired  current     3  Connect the 24V loop as shown in Figure 13              LOOP  POWER  SUPPLY    E    30 VDC  MAX    MEASURE   SOURCE                    Figure 13  Connections for Simulating a Transmitter       mA 2W SIM 0 00     4 000 mA    MENU   LIGHT   STEP             4 5 Sourcing volts  To source volts follow these steps     1   2     Select the  CONFIG  option from the main menu     Choose  VOLTS  from the primary parameters  Switch to input output control and select  output  OUT      Connect the leads for the voltage source terminals  as shown in Figure 14   Enter the voltage using the keypad              Figure 14  Connections for Sourcing Voltage and Frequency    VOLTS OUT   
29. ng Thermocouples    4 8 Sourcing Thermocouples  To source a thermocouple use the following steps     1     20    Connect the thermocouple leads to the appropriate polarized TC miniplug  and insert the  plug into the TC terminals on the calibrator  as shown in Figure 15     Select the  CONFIG  option from the main menu  and choose thermocouple  TC  from  the primary parameters     Choose output  OUT  from the input output control   Select the desired thermocouple type from the sensor types   Enter the temperature using the keypad        TC OUT TYPE E  CJC ON CJ 25 5   C    0 0   C    MENU   LIGHT   STEP                                           Figure 16  Connections for Outputting RTDs    4 9 Sourcing Ohms RTDs  To source an RTD  follow these steps   1  Select the  CONFIG  option from the main menu  and choose  RTD  from the primary  parameters   2  Choose output  OUT  from the input output control  and select RTD type from the sensor  types   3  Connect the calibrator to the instrument being tested  as in Figure 16     4  Enter the temperature or resistance using the keypad     Use stackable  test leads for  shared connections    MEASURE   SOURCE MEASURE   SOURCE                      4W                                                 NS    Figure 17  Using a 3  or 4 wire Connection for RTDs  21       P200 385  200 0 2    0 0   C  MENU   LIGHT      Note  The calibrator simulates a 2 wire RTD  To connect 3  or 4 wire transmitter  use stacking  cables  as shown in Figure 17    
30. nu Bar    The parameters on the display are controlled by the menu bar  which is located at the bottom  of the LCD  The function keys  F1  F2  and F3  are used to navigate through all the levels and  choices of the menu bar  Refer to the menu tree for a clarification on the layout of all the    levels     The top level of the menu is the home menu  It can be accessed anytime by pressing the  HOME key  There are three variations of the home menu  the input home menu  the output  home menu  and the pulse home menu     In the input home menu the only active options are  MENU  and  LIGHT   The  MENU  option  is used to enter the next level of the menu bar  the main menu  Press the corresponding  function key  F1  to enter the main menu  The  LIGHT  option is used to turn on the LCD  back light  Press the corresponding function key  F2  to turn on the back light     MENU   LIGHT         In the output home menu there are three active options   MENU    LIGHT  and  STEP  or   RAMP   The first two options work the same as in the input home menu  The third option is  selectable in the Auto Function Menu and is used to turn on and off the selected auto  function  See Section 4 2  Using the Automatic Output Functions  Also leaving this menu or  pressing the Home button will stop the auto functions     MENU    LIGHT   STEP    The pulse home menu also has three active options   MENU    TRIG   and  COUNTS   The   TRIG  and  COUNTS  options are used for pulse simulation  The function of thes
31. pairs and recalibration will be subject to specific charges  Under no  circumstances will Martel Electronics be liable for any device or circumstance beyond the  value of the product     45    MARTEL    LECTRONICS    www martelcorp com  e mail  sales martelcorp com  Tel   603  434 1433 800 821 0023  Fax   603  434 1653    Martel Electronics    3 Corporate Park Dr   Derry  NH 03038    0219999  RevF 5 13    
32. plete    SRE Loads a byte into the Service Reguest Enable register    SRE  Returns the byte from the Service Reguest Enable register    STB  Returns the status byte    WAI Prevents further remote commands from being executed until all previous remote  commands have been executed   Table 6  Calibrator Commands  Command Description  CAL START Puts the calibrator in calibration mode  CJC STATE Turns CJC on or off   CJC STATE  Returns the state of the CJC  CPRT COEFA Sets the custom RTD coefficient A  CPRT COEFA    Returns the custom RTD coefficient A  CPRT COEFB Sets the custom RTD coefficient B  CPRT COEFB    Returns the custom RTD coefficient B  CPRT COEFC Sets the custom RTD coefficient C  CPRT COEFC    Returns the custom RTD coefficient C  CPRT MIN T Sets the custom RTD minimum temperature  CPRT MIN T  Returns the custom RTD minimum temperature  CPRT MAX T Sets the custom RTD maximum temperature  CPRT MAX T    Returns the custom RTD maximum temperature  CPRT RO Sets the custom RTD RO resistance  CPRT RO  Returns the custom RTD RO resistance  FAULT  Returns the error code of an error that has occurred  FREQ LEVEL Sets the frequency and pulse amplitude       FREQ LEVEL        Returns the frequency and pulse amplitude       28                                                                                                          Command Description  FREQ TYPE Set the frequency output to continuous  frequency  or pulse   FREQ TYPE  Returns frequency output type  continuous or p
33. pressure module from overpressure  never apply pressure above the  rated maximum printed on the module     To avoid damaging the pressure module from corrosion  use it only with specified materials   Refer to the pressure module documentation for material compatibility     To measure pressure  follow these steps    1  Connect the pressure module to the calibrator as shown in Figure 10    Note  Make sure the calibrator is on before you plug in the pressure module   Select the  CONFIG  option from the main menu    Select  PRESSURE  from the primary parameters    Select the desired measuring unit     Zero the pressure module  The zero function on the calibrator can be found in the  pressure zeroing menu     aR wD           ES  VALVE                 E  ch PRESSURE  7 MODULE                                                             9   d                                                                          0909   OOOO   OOOO  g                         OO  OO          Figure 10  Connections for Measuring Pressure    15    3 4 1 Zeroing with Absolute Pressure Modules     To zero  adjust the calibrator to read a known pressure  such as barometric pressure    To adjust the calibrator  follow these steps    1  Enter the pressure zeroing menu    2  Select  ZERO       SET REFERENCE ABOVE  will appear  Enter the pressure  using the keypad    3  The calibrator stores the Barometric zero offset in non volatile memory     The zero offset is stored for one absolute pressure module at a t
34. rature  follow these steps    1  Attach the thermocouple leads to the TC miniplug  and insert the plug into the  input output of the calibrator  as in Figure 8    Note  For best accuracy wait 2 to 5 minutes for the temperature between the miniplug and   the calibrator to stabilize before any measurements are taken    2  Select the  CONFIG  option from the main menu     3  Select TC from the primary parameters  Choose  IN  in the input output control  and than  the thermocouple type from the sensor types  The temperature unit may also be changed  from Celsius to Fahrenheit              13    The calibrator can also measure the mV of a Thermocouple  which can be used along with a  table in case the corresponding TC type is not supported by the calibrator  To do so  proceed  as above and choose mV from sensor types     TC PLUG          Note  The TC wire  used must match the  thermocouple type  being calibrated     MEASURE   SOURCE                Figure 8  Measuring Temperature Using Thermocouple Terminals    3 4 2 Using Resistance Temperature Detectors  RTDs     The supported types of RTDs are shown in Section 8  Specifications  RTDs are  characterized by their 0  C resistance  RO  The calibrator accepts two  three  and four wire  inputs  with four wire input being the most accurate     To use the RTD option  apply the following steps     1   2   3     Select the  CONFIG  option from the main menu   Select RTD from the primary parameters  Select  IN  from input output control   
35. rce only  Enter DCI for mA  DCV for volts  TC for  thermocouple  RTD for RTD  FREQUENCY for frequency  and PRESSURE for pressure  For  example     MEAS DCV sets the display mode to VOLTS IN    PRES UNIT    Sets the unit for measuring pressure on the display  Add the unit after the command  The  available pressure units and their syntax are shown in the Table 7   Parameter Units    For example     PRES UNIT KPAL sets the pressure unit to kiloPascals    OUT    Sets the output of the calibrator  This command may be used to output mA  volts  frequency   temperature  and ohms  Frequency output  which is set by the command FREQ TYPE  is  either continuous or pulse  The calibrator is automatically set to source mode when OUT is  entered  A number and its unit must follow the command  See Table 7   Parameter Units  for a  list of available units  For example     OUT 10 MA sets the calibrator to mA OUT mode and sets the output to 10mA     OUT   Returns the output of the calibrator  Using the above example  OUT  Would return   1 000000E 02  A    PRES     Returns the model and serial number of the attached pressure unit  Returns NONE if no  pressure unit is attached  For example     PRES  Will return MARTEL 001PNS 3 0    PRES UNIT     Returns the pressure units  For example if the unit on the display is psi the command will  return     PSI    PULSE CNT    Sets the number of pulses the calibrator will produce when it is triggered while in pulse mode   For example     PULSE CNT 3000 will 
36. rror codes  When it is empty  FAULT  returns 0  The error queue is cleared  when power is reset or when the clear command  CLS is entered     Input Buffer    Calibrator stores all received data in the input buffer  The buffer holds 250 characters  The  characters are processed on a first in  first out basis     5 4 Remote Commands and Error Codes    The following tables list all commands  and their descriptions  that are accepted by the  calibrator     27    Table 5  Common Commands                                                                                                    Command Description   CLS  CLS  Clear status   Clears the ESR  the error queue  and the RQS bit in the status byte   Terminates pending Operation Complete commands   ESE Loads a byte into the Event Status Enable register    ESE  Returns the contents of the Event Status Enable register    ESR  Returns the contents of the Event Status register and clears the register    IDN  Identification guery  Returns the manufacturer  model number  and firmware revision level  of the Calibrator    OPC Enables setting of bit 0  OPC for  Operation Complete   in the Event Status Register to 1  when all pending device operations are complete    OPC  Returns a 1 after all pending operations are complete  This command causes program  execution to pause until all operations are complete    RST Resets the state of the instrument to the power up state  This command holds off  execution of subseguent commands until it is com
37. set the number of pulses to 3000     PULSE CNT     Returns the number of pulses in the pulse train  Using the above example  the returned value  would be     3000    REMOTE    Puts the calibrator in remote mode  From the remote mode the user can still use the keypad  to get back to local unless the command LOCKOUT was entered before REMOTE  Than the  keypad is totally locked out  and the user has to send the LOCAL command to get back to  local operation     RTD TYPE    Sets the RTD type  The following is a list of RTD types they way they should be entered after  the command     PT385 10  PT385 50  PT385 100  PT385 200  PT385 500   PT385 1000  PT392 100  PTJIS 100  Ni120  Cu10  Cu50   Cu100   YSI 400  OHMS  CUSTOM    For example     RTD_TYPE PT385_10 sets RTD type to Pt385 10    RTD_TYPE   Returns the RTD type     RTD_WIRE    Sets the number of wires used for connection in measuring RTDs  The calibrator measures  RTDs using 2   3   and 4 wire connections  After the command  enter 2W for 2  wire  3W for 3   wire  and 4W for 4 wire  For example     RTD_WIRE 4W sets the connection to 4 wire    37    38    RTD WIRE   Returns the number of wires used in the RTD connection     SIM    Sets the output for current simulation  This command also switches the calibrator into mA  simulation mode  A number and a unit must be entered after the command  For example     SIM 5 MA sets the current simulation at 5 mA    SIM   Returns the output of the current simulation  With the example above  t
38. specified replacement parts       To avoid false readings  which could lead to possible electric shock or personal injury   replace the battery as soon as the battery indicator appears       To avoid a violent release of pressure in a pressurized system  shut off the valve and slowly  bleed off the pressure before you attach the pressure module to the pressure line     A Caution    To avoid possible damage to calibrator or to equipment under test     Use the proper jacks  function  and range for your measurement or sourcing application       To avoid mechanically damaging the pressure module  never apply more than 10 ft Ib  of  torque between the pressure module fittings  or between the fittings an the body of the module       To avoid damaging the pressure module from overpressure  never apply pressure above  the rated maximum printed on the module       To avoid damaging the pressure module from corrosion  use it only with specified materials   Refer to the pressure module documentation for material compatibility     2  Calibrator Interface    Figure 1 shows the location of the input and output connections on the calibrator  while Table    1 describes their use        9   BETA                    208  0  000    OOONEB  OOO  1000                                           Figure 1  Input Output Terminals    Table 1  Input and Output Terminals       No     Name       TC input output    Description    Terminal for measuring  or simulating thermocouples  Accepts  miniature polari
39. ter was missing  106 An invalid pressure unit was received  107 An invalid CJC STATE was received  108 An invalid TSENS TYPE was received       30                                           Error Number Error Description  109 Pressure module not connected  110 An unknown command was received  111 An invalid RTD or TC parameter value was received  112 The serial input buffer overflowed  113 Too many entries in the command line  114 The serial output buffer overflowed  115 Output is overloaded  116 Calibrator not in pulse train mode when TRIG was received  117 An invalid FREQ TYPE was received       5 5 Entering Commands    Commands for the calibrator may be entered in upper or lower case  There is at least one  space required between the command and parameter  all other spaces are optional  Almost  all commands for the calibrator are sequential  any overlapped commands will be indicated  as such  This section will briefly explain each of the commands and describe their general  use  which will include any parameters that may be entered with the command as well as  what the output of the command is     5 5 1 Common Commands   CLS    Clears the ESR  the error queue and the RQS bit  Also terminates all pending operations   When writing programs  use before each procedure to avoid buffer overflow      ESE    Loads a byte into the Event Status Enable register  The command is entered with a decimal  number that  when converted to binary  enables the right bits in the Event Status Reg
40. terface for remote control    1 1 Customer Service    Corporate Office   www martelcorp com    e mail  sales martelcorp com  Tel   603  434 1433 800 821 0023 Fax   603  434 1653  Martel Electronics  3 Corporate Park Dr   Derry  NH 03038    1 2 Standard Equipment  Check to see if your calibrator is complete  It should include   MC1010 Calibrator  Instruction Manual  Test Leads  Rubber Boot  NIST Certificate    1 3 Safety information    Symbols Used    The following table lists the International Electrical Symbols  Some or all of these symbols  may be used on the instrument or in this manual           Symbol Description     AC  Alternating Current     AC DC  Battery    ON  AN    CE Complies with European Union Directives    DC    Double Insulated   Electric Shock   Fuse   PE Ground   Hot Surface  Burn Hazard    Read the User s Manual  Important Information     Off     O gt   gt  i enl    On    Canadian Standards Association    o  E  Yd  c  a    The following definitions apply to the terms    Warning    and    Caution           Warning    identifies conditions and actions that may pose hazards to the user      Caution  identifies conditions and actions that may damage the instrument being used     Use the calibrator only as specified in this manual  otherwise injury and damage to the  4 calibrator may occur     AN Warning    To avoid possible electric shock or personal injury     Do not apply more than the rated voltage  See specifications for supported ranges     Follow all equ
41. th an      For  example     IDN   tells the calibrator to return its identification     Query Commands  Commands that ask for information  They always end with a      For example   FUNC   Returns the current modes of the calibrator     Compound Commands  Commands that contain more than one command on one line  For example   MEAS RTD  RTD TYPE CU10  Sets the calibrator to measure RTD and sets RTD type to Cu 10     Overlapped Commands    Commands that require more time to execute than normal  The command  WAI can be used  after the overlapped command to tell the calibrator to wait until the command finishes before  executing the next command  For example     TRIG   WAI    Triggers the pulse train  Once the pulse train has been triggered  the calibrator can proceed  to the next command     Sequential Commands    Commands that are executed immediately after the are entered  This type includes most of  the commands     5 3 2 Character Processing  The data entered into the calibrator is processed as follows       ASCII characters are discarded if their decimal equivalent is less than 32  space   except 10   LF  and 13  CR        Data is taken as 7 bit ASCII    The most significant data bit is ignored     Upper or lower case is acceptable     5 3 3 Response Data Types  The data returned by the calibrator can be divided into four types     Integer    For most computers and controllers they are decimal numbers ranging from  32768 to 32768   For example      ESE 140   ESE  returns 140 
42. uld return      2 600000E  02  CEL    CPRT MAX T    Sets the maximum temperature of the custom RTD range  The temperature value must be  entered with a degrees label  CEL for Celsius and FAR for Fahrenheit     For example   CPRT MAX T 0 0 CEL enters 0 0  C as the maximum temperature     CPRT MIN T     Returns the value entered for minimum temperature in the range for a custom RTD  The  above example would return     0 000000E  00  CEL    CPRT  RO    Sets the 0   resistance  RO  in the custom RTD  The value must be entered with a units label   Refer to the Parameter Units table for assistance     For example   CPRT RO 100 OHM sets RO to 100 ohms   CPRT RO     Returns the value for the resistance in custom RTD  The above example would return   1 000000E  02  OHM  FAULT     Returns the error code number of an error that has occurred  The command may be entered  when the previous command did not do what it was meant to do     For example  if a value for current output is entered that is bigger than the supported range   0 24mA  FAULT  Would return     103 which is the code number for an entry over range   Refer to the Error Codes table for more information on error code numbers     FREQ LEVEL    Sets the amplitude of the wave used in the Frequency Out and Pulse modes  The range for  amplitude entered may be found in the Specifications section     For example   FREQ LEVEL 5 V sets the amplitude at 5Vpp     FREQ LEVEL   Returns the amplitude of the wave used in Frequency Out and Pu
43. ulse  FREQ UNIT Sets the unit for frequency and pulse  FREQ UNIT  Returns the unit for frequency and pulse  FUNC  Returns the current mode of the calibrator  LOCAL Returns user to manual operation of the calibrator  LOCKOUT Locks out the keypad of the calibrator  and allows for remote operation only  MEAS Sets the mode for measuring of the calibrator  PRES UNIT Sets the pressure unit of the calibrator  OUT Sets the output of the calibrator  OUT  Returns the output of the calibrator  PRES  Returns the model and serial number of the attached pressure module  PRES UNIT  Returns the pressure unit for the calibrator  PULSE CNT Sets the number of pulses for the pulse train  PULSE CNT  Returns the number of pulses in the pulse train  REMOTE Puts the calibrator in remote mode  RTD TYPE Sets the RTD type  RTD TYPE  Returns the RTD type  RTD WIRE Sets the number of wires used by the RTD mode   RTD WIRE  Returns the wire number setting used in the RTD mode  SIM Sets the output for mA simulation  SIM  Returns the output of the mA simulation  TEMP UNIT Sets input output temperature unit for RTD and TC  TEMP UNIT  Returns the temperature unit for RTD and TC  TRIG Starts and stops the pulse train in pulse mode  TRIG  Returns TRIGGERED when a pulse train is on  Returns UNTRIGGERED when the pulse  train is off   TSENS TYPE Sets temperature sensor type   TSENS TYPE    Returns temperature sensor type  VAL  Returns the measured values  ZERO MEAS Zeros the pressure module  ZERO MEAS  Returns the 
44. ure 3  Display             The display of the calibrator is shown in Figure 3     The menu bar is used to setup the display to perform the desired function   Table 3 explains the different parts of the display     Table 3  Display Functions       No     Name       Primary Parameters    Description    Determine what parameter is going to be measured or  sourced    The available options are  VOLTS  TC  thermocouple   RTD   FREQ  frequency   PULSE  PRESSURE  mA  and mA 2W SIM        Input Output control    Switches between input mode  read   and output mode   source         Additional Settings    Available only for TC  thermocouple   and RTD measurements   For TC this setting turns the CUC  Cold Junction Connection   on and off    For RTD measure  RTD IN   this setting sets the number of  wires used in the measurement  2 wire  3 wire  or 4 wire        Units    Shows what unit the measurement or source value is in   Available options are for RTD and TC    C or   F   and for FREQ  and PULSE  CPM  Hz  or KHz        Sensor Types    Allow for measurements to be made for different types of RTDs  and TCs  All types are shown in the Specifications  Also  dis   plays the amplitude of the pulse and frequency source  and  pressure units        Numeric Displays    Display the numeric values of the signal being measured  or  sourced  An    Ol    reading indicates an out of range or overload  condition              Secondary Valve    Shows the base unit Q for RTD or mV for TC           2 2 Me
45. ure 6     mA IN  25 00           0 000 mA      MENU T   LIGHT              Figure 6  Measuring mA with Input Output Terminals    3 3 Measuring mA with Loop Power    To test a 2 wire  loop powered transmitter that is disconnected from wiring  use the loop  power function  This function activates a 24V supply in series with the current measuring  circuit  To use this option proceed as follows     To measure mA with loop power follow these steps   1  Use the  CONFIG  option from the main menu to select mA IN LOOP  2  Connect leads as shown in Figure 7                                                                    bodia  Ex   E MEASURE   SOURCE  aja cen 9 Q   E     eo   ee                    Figure 7  Connection Using Current Loop  3 4 Measuring Temperature    3 4 1 Using Thermocouples       The calibrator supports the following  thermocouple types  B  C  E  J  K  L  N  R  S  T  U  TC IN VRE  BP and XK  The characteristics of all the types are   CJC ON GJ 24 9    G    described in Specifications section  The calibrator  also has a Cold Junction Compensation  CJC   function  Normally this function should be ON and    the actual temperature of the thermocouple will be 5 0 0 3 5  measured  With CJC OFF  the calibrator will L   measure the difference between the thermocouple       at the junction and at its TC input terminal  MENU LIGHT    Note  CJC off mode should only be used when calibration is being done using an external  ice bath     To use the thermocouple to measure tempe
46. zed thermocouple plugs with flat in line blades  spaced 7 9 mm  0 312 in  center to center                       2 3 Source Measure Terminals for sourcing and measuring voltage  frequency   V RTD 2W  Hz  pulse train  and RTDs  4 5 Source Measure Terminals for sourcing and measuring current  and performing  mA terminals  3W 4W RTD measurements with 3 wire or 4 wire setups   6 Pressure module connector   Connects calibrator to a pressure module for pressure meas   urements   7 Serial port Connects calibrator to a PC for remote control           Figure 2 shows the location of the keys on the calibrator  Table 2 lists the functions of each  key                       O                                                                M       Figure 2  Keypad    Table 2  Key Functions       No  Name Function       1 Function Keys F1  F2  F3 Used to operate the menu bar at the bottom of the calibrator  display  F1 is used for selecting options in the left box  F2 for  the center box  and F3 for the right box              2 Home Returns to home menu on the menu bar   3 Power Turns calibrator on and off   Cursor Control Key Left and right arrow keys are used to select which decade to    be changed in output value  Up and down arrow keys are used  to increase  decrease  or ramp output value                 5 Numeric Keypad Allows user to enter Numeric values        2 1 Main Display        3   ja _      _RTD IN 4W P100 385  Gy   100 00 91     7     a O  0 0 6       2    MENU   LIGHT    Fig
47. zero offset of the pressure module          29    Table 7  Parameter units                                                                                     Units Meaning  MA milliamps of current  MV Voltage in millivolts  V Voltage in volts  CPM Frequency in cycles per minute  Hz Frequency in Hertz  KHz Frequency in kiloHertz  Ohms Resistance in Ohms  Cel Temperature in Celsius  Far Temperature in Fahrenheit  PSI Pressure in pounds per square inch  INH204C Pressure in inches of water at 4  C  INH2020C Pressure in inches of water at 20  C  INH2O60F Pressure in inches of water at 60 F  CMH204C Pressure in centimeters of water at 4  C  CMH2020C Pressure in centimeters of water at 20  C  BAR Pressure in bars  MBAR Pressure in millibars  MPAL Pressure in MegaPascals  KPAL Pressure in kiloPascals  INHG Pressure in inches of mercury at 0  C  MMHG Pressure in millimeters of mercury at 0  C  KG CM2 Pressure in kilograms per square centimeter  MMH204C Pressure in millimeters of water at 4  C  MMH2020C Pressure in millimeters of water at 20  C  FTH2O60F    Pressure in feet of water at 60  F    Table 8  Error codes                                     Error Number Error Description  100 A non numeric entry was received where it should be a numeric entry  101 Too many significant digits entered  102 Invalid units or parameter value received  103 Entry is above the upper limit of the allowable range  104 Entry is below the lower limit of the allowable range  105 A required command parame
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
American Power Conversion SURT005 User's Manual  ORFX100 – Starball General Information: Orion  取扱説明書  DRSA-1200 User Manual - Digi-Cool  Quand votre Caf se mobilise pour le logement  17027278_1014 DKP310 NLO.book  Electrolux 115 D Ventilation Hood User Manual  Betriebsanleitung EC04 - k-tech  Télécharger  RM-PICC Salvo Compiler Reference Manual – HI    Copyright © All rights reserved. 
   Failed to retrieve file