Operation Manual/1MB
Contents
1. Type a new file name at Name In this example the file name is Instant interrupt for 1s g3 Untitled Sequence Builder Iof xi File View Instrument Help e Wizard List f E EX Abnormality Simulation Wizard a Abrupt Phase Shift wizard for Single Phase Cancel Name Instant interrupt for 1s Location Create data from E Program FilessKikusuiNQuick Wave Sequencers Browse lt Note gt W The application supports the Windows 95 long file name format Since the report print output truncates any file names longer than 80 characters use 80 or fewer characters E Enter numbers in half size characters Click OK Enter test conditions Page 13 Quick Wave Sequencer User s Guide 3 Enter test conditions Abnormality Simulation Wizard Step 1 of 2 Volage VE fmm Frequency Hz 50 Intervals s D Simulation D ata No Type Test Voltagefv TestTime ms Repe Not Used Not Used Not Used Nat I lsed BR WMH Edit Loop times i Set Voltage Frequency Interval and Loop Input the steady voltage effective value of the test plan 100 for Voltage frequency 50 for Frequency time of instantaneous interruption interval 20 for Interval and 4 for Loop Double click Number 1 You can perform the same action by positioning the cursor at Number 1 and clicking the Edit button The Data Input screen appears
2. Select Dip interruption from the Type page Abnormality Simulation mode Simulation Wizard No 1 x Type edit Abnormality Simulation mode C Not Used R O A C Voltage Variat io Cancel Apply Help Page 14 Quick Wave Sequencer User s Guide Enter the simulation data for Dip interrupt in the Edit page Input the test plan Dip Interrupt voltage OV for Dip instant voltage an Dip interruption start phase angle of 90 for T1 Start phase and a Dip interruption period 1000 ms for the instant interruption period The Repeat should be one Note When Dip interruption end phase angle exceeds approximately 280 Dip interruption occurs at an interval of 1 cycle For details see page 40 Simulation Wizard No 1 x Type Edit Dip Interruption Voltage Vrms fo T1 Start phase 20 T2 Dip Interruption period ms fi Q0 Repeat times fi E Cancel Apply Help After entering the test data click OK Abnormality Simulation Wizard Step 1 of 2 Volage V 100 Frequency Hz 50 Intervals s 20 Simulation D ata HE Lim Hesse Tet Tiels Peps e Dip Interruption 0 0 1000 Not Used Not Used Nat I sed Edit Loop times Cancel Help Click the Next button Page 15 Quick Wave Sequencer User s Guide Abnormality Simulation Wizard Step 2 of 2 Voltage V 100 0 Freque
3. 200 Range Before or now sequence execution settings m After sequence execution settings Voltage V 00 Voltage V fi oo Erequency Hz po Frequency Hz 50 Wavebank o 2d Set Wavebank D m Sequence Start Step End Step Loop 4 Sequence execution Page 47 Quick Wave Sequencer User s Guide Settings made in the Running Information are valid when the Run dialog box is opened And waveform bank 0 are set as initial settings for the Run dialog in the PCR L Even when the Run dialog box is opened you can click the set button to enter Before or now sequence execution settings Operation settings in the PCR L Operation settings This output button turns PCR L output on or off 200V Range This button switches between 100V 200V for the PCR L a change that can be made only when PCR L output is off For a voltage less than 152 5V select the 100V range by unchecking the box For a voltage 152 5V or greater select the 200V range by checking the box When the Execute dialog box opens confirm the PCR L setting and check the voltage for setting before test execution in Running Information The range is automatically switched to the 200V Range if necessary Before or now sequence execution settings Voltage Set PCR L output voltage in the range 0 0 305 0V If the set value is 152 5V or greater click the check box for the 200V Range confirming that PCR L output is off Set the
4. 5 Note The voltage variation test consumes up to five sequence steps When Repeat is set to n up to 5 n steps are consumed Enter all test conditions and click OK to return to the Step 1 of 2 screen Then click the Next button to go to the Step 2 of 2 screen Here you can confirm the test conditions just entered Abnormality Simulation Wizard Step 2 of 2 Voltage V 100 0 Frequency Hz 50 00 Interval s 10 0 Simulation No Loop times 1 r Simulation Data Type Dip Interruption Dip Interruption Voltage Vrms 0 0 Ti Statphase 0 T2 Dip Interruption period ms 10 Repeat times 1 tea n After confirming test conditions click the Next button to display the Finish screen Clicking the Finish button closes the Wizard and waveform bank and sequence step information is developed to Wave Bank Server and Sequence Builder Page 55 Quick Wave Sequencer User s Guide Abrupt phase shift wizard for single phase Select Abrupt Phase Shift Wizard for Single Phase in the Sequence Wizard to activate the Abrupt Phase Shift Wizard for Single Phase Abrupt Phase Shift Wizard for Single Phase can abruptly shift the phase and create a discontinuous sine waveform The abrupt phase shift test is achieved by switching a waveform bank in which the phase shifted waveform using the applied waveform of the harmonic composite waveform is set in advance Select Abrupt Phase Shift Wizard for
5. Quick Wave Sequencer User s Guide Changing the 1 point data The waveform pattern is complete To patch the waveform click the right mouse button at the appropriate waveform point A popup menu appears enabling selection of Patch and a mark at the waveform and a dialog box are displayed Untitled Easy Power Waveformer QW B x File View Shadow Instruments Help Cea 1 ve Patchy Pointing Address fis OK waveform Data 2485 al Cancel lt lt Left lt lt gt gt Right gt gt Fast Move Ready x2 111 y 2 3472 NUM Z In the dialog box above Point Address 114 indicates that the current waveform position is 114 Similarly Waveform Data 3485 indicates that the current data value at that position is 3485 Make the appropriate changes to the waveform data As a general rule settings range from 0 to 4095 with bottom data assigned zero and top data assigned 4095 Saving waveform data You can save created waveforms to disk in the 12 bits x 1024 word format Select the File Save As menu to display a dialog box for selecting a file name Name the file the default extension is EPW Page 32 Quick Wave Sequencer User s Guide Reading existing data A saved waveform data file can be opened and read Select the File Open menu to display a dialog box Double click on the desired file Generating user defined waveforms Since the waveform pattern that can be used as the user de
6. 0 0 305 0V Start phase Set the phase angle for the point at which instantaneous interruption starts For example enter 90 when instantaneous interruption takes place at the peak AC current operating resolution is 18 at 50Hz Dip interruption period ms Set the time for which instantaneous interruption continues in the range 1 9999ms Repeat times Set number of test executions in the range 1 5 times Note The Dip interruption test consumes up to four sequence steps When Repeat is set to n up to 4 n steps are consumed Setting voltage variation test data Simulation Wizard No 1 x Type Edit Variation Voltage Vrms T1 Slope time ms T2 Variation period ms T3 Slope time2 ms Repeat times TA Cancel Help To set test conditions for voltage fluctuations select Voltage Variation from Abnormality Simulation mode and click the Edit tab Variation voltage Vrms Set the fluctuating test voltage in the range 0 0 305 0V Slope time 1 ms Set the duration over which voltage rises or falls to the fluctuating voltage in Page 54 Quick Wave Sequencer User s Guide the range 1 99990ms Variation period ms Set the duration for which voltage fluctuates in the range 1 9999ms Slope time 2 ms Set the duration over which voltage recovers in the range 1 99990ms Repeat times Set the number of test executions in the range 1
7. F1 4 Executing the test Switch on power for the PCR L and open the test execute screen on the Sequence Builder Instrument Run menu Page 17 Quick Wave Sequencer User s Guide Run To PCR1000L ersion 2 04 x Operation settings 200 Range Before or now sequence execution settings 4 gt After sequence execution settings Voltage V fi 00 Voltage V fi 00 Frequency Hz o Frequency Hz 50 W avebank 0 Set W avebank D Sequence Start Step End Step 4 Loop 4 Sequence execution Close Click the Sequence execution button The following confirmation dialog appears Sequence Builder B Do you wish to send to PCR L Wave Bank Data and Sequence steps Data lt Note gt This dialog is displayed only if the sequence or waveform data is not transferred It does not appear if the data is already transferred Click the Yes button The Dip interruption test executes the data transfer Sequence Time Will be h min s mik Execution Running When the test is completed the displayed message Running changes to Sequence End and the Stop button becomes a Close button Click Close to close the Running dialog Page 18 Quick Wave Sequencer User s Guide lt Descreption gt The forecast sequence running time is the total of all sequence running times When sequence wave sync is on actual running time is longer than the forecast sequenc
8. PCR L output voltage and frequency before the test is performed Voltage When the Execute dialog opens set PCR L output voltage in the range Page 45 Quick Wave Sequencer User s Guide 0 0 305 0V When set in the 100V range this setting automatically switches to the 200V range for a voltage 152 5V or greater When the PCR L current range differs from the range to be set you must shut off PCR L output to change the range Otherwise the Execute dialog box will not open Frequency When the Run dialog box is opened set the frequency to be set in the PCR L in the range 1 00 999 9Hz After Sequence settings Enter values for PCR L output voltage and frequency following test execution Voltage Set the voltage to be set in PCR L on test completion in the range 0 0 305 0V When a voltage before test execution is set to 152 5V or greater the set voltage following testing switches to 152 5V or greater Frequency Following testing set the frequency in the PCR L in the range 1 00 999 9 Hz Sequence settings A sequence test can be performed when sequence step data includes at least 2 steps Start step Set the step with which the sequence starts When testing data created by the Wizard this step is generally 1 The setting range from 1 to the end step End step Set the step with which sequence ends When testing data created by the Wizard the number of steps is the number that has already be
9. Single Phase The Step 1 of 1 screen appears Note The abrupt phase shift test is imitatively conducted by the sequence function with PCR L bank switching The test involves the following actions which can be considered the internal abrupt phase shift 1 The output phase at the start of the test begins with the abrupt shift phase 2 When the test is finished the phase switches to phase 0 since the waveform bank switches to waveform bank 0 3 Ina repeat phase test the same phenomenon seen in 1 occurs Setting abrupt phase shift test data Abrupt Phase Shift Wizard for Single Phase Step 1 of 1 T1 Before Abrupt phase shift time s P1 Abrupt phase shift phase P2 Abrupt phase shift phase T2 After Abrupt phase shift time s Voltage V Frequency Hz Set abrupt phase shift test data in this screen Before Abrupt phase shift time s Set the time before abrupt phase shift occurs in the range 0 001 9999s Operating time is rounded up to an integer multiple of 1 cycle Page 56 Quick Wave Sequencer User s Guide Abrupt phase shift phase Set the waveform phase when abrupt shift occurs in the range 0 359 In the displayed image graph this is approximately 180 Abrupt phase shift phase Set the waveform phase in which phase shifts abruptly in the range 0 359 In the displayed image graph this is approximately 0 After Abrupt phase shift time s Set the t
10. fi Set Voltage Frequency Interval and Loop Enter test plan steady voltage effective value 100 for Voltage frequency 50 for Frequency time of fluctuating interval 20 for Interval and 4 for Loop Double click Number 1 This performs the same action as positioning the cursor at Number 1 and clicking Edit The Data Input screen appears Select Voltage Variation from Type page Abnormality Simulation mode Set the simulation data for voltage variation in the Edit page Input fluctuating voltage 40V for Voltage Variation front slope time 1000 ms for Slope Time 1 fluctuation sustained time 1000 ms for Variation period and post slope time 1000 ms for Slope Time 2 The Repeat should be one Cancel Apply Help Enter test data and click OK Inspect the test data and click the Next button Click the Finish button to complete entry of test conditions When you close the Wizard a bank waveform and the sequence steps are created in Wave Bank Server and Sequence Builder 4 Execution of test In Sequence Builder open the Instrument Run menu Click the Sequence execution button to transfer the set data The fluctuating voltage test is executed Page 20 Quick Wave Sequencer User s Guide Chapter 3 Wave Bank Server waveform generation In previous chapters we performed tests by transmitting the sine waveform in Wave Bank Server to the PCR L Wave Bank Server creates the sine harmonic composite and distort
11. from sine wave harmonic composite waveforms distortion waveforms triangle waves and square waves as well as waveforms acquired from the oscilloscope The software transfers generated waveforms to the waveform bank of the PCR L The software is activated by activating Sequence Builder For novice Windows operators Quick Wave Sequence runs on Windows The basic operations such as windows menus and buttons are the same as those of other Windows applications Those who are unfamiliar with the basic operations of Windows should refer to the Microsoft Windows instruction manual Page 5 Quick Wave Sequencer User s Guide Features Quick Wave Sequence has the following features 1 The software supports single phase systems Using the PCR L waveform bank and the sequence function a highly flexible simulation test can be performed on the AC power source Dip interrupt test Voltage variation test Abrupt phase shift test supporting single phase systems only Output of harmonic composite waveform Waveform obtained from the oscilloscope is output from the PCR L The Wizard automatically creates sequence steps for Dip interruption voltage variation or abrupt phase shift tests with no need for complex settings for the simple test The combination of the waveform bank of the PCR L and the sequence function enables execution of tests of the composite combination of the sine distortion and harmonic composite
12. n 1 Voltage 100 0 V Voltage 10 0 V Frequency 50 00 Hz Frequency 50 00 Hz Time 15ms Time 1s Wave Bank 0 Wave Bank 1 When wave Sync for sequence step n 1 is disabled unchecked et Waveform of sequence stepn Waveform of sequence step n 1 When wave Sync for sequence step n 1 is enabled checked Waveform of sequence step n Waveform of sequence step n 1 When settings are complete click OK Note The time of waveform of sequence step n gets longer than the actual set time when wave Sync is enabled Sequence Builder adjusts synchronization by using this time difference however 3 5 ms are required to adjust it When the time is set to 18 ms for example waveform of sequence step n come to 2 cycles 40 ms If it becomes an issue contact your Kikusui agent Page 40 Quick Wave Sequencer User s Guide Simple printing Sequence Builder can print created sequence step data Data is sent to the printer entered in the File Print menu Select the File Print Preview menu to see a preview of the output Page 41 Quick Wave Sequencer User s Guide Chapter 5 Performing the Test You can perform various tests by operating the PCR L according to waveform bank or sequence step information But before performing the test you must set GPIB GPIB setting This application communicates with external equipment through the GPIB by device name Set the GPIB environment before using the GPIB function To ins
13. to changes from the frequency set in one step before to that set in this step If this item is not checked the frequency is immediately changed to that set in this step In this case first ramp value is disabled Time Set the step duration Enter values for hours minutes and seconds in the range 0 001 s 999h59m59 999s Note Depending on PCR L function specifications this value may differ by about 1 ms from the actual output time To operate the system in cycles set a value several ms shorter than cycle time and click the checkbox enabling Wave Sync for the next sequence step Wave Bank Enter a waveform bank number from 0 to 13 The waveform set by Wave Bank Server appears in this step Bank 0 indicates sine waveform fixing Output Set output in this step To enable output check this item Trigger Check this item to output a trigger signal the SEQ TRIG OUT terminal on the back panel from the PCR L at the start of this step Wave Sync In the PCR L sequence the set time is not synchronized with the output waveform Depending on the time set for the step discontinuous waveforms may be output when steps switched Checking Waveform Synchronization starts this step operation after waveform banks are switched in the previous step Check this item for phase control Otherwise leave unchecked Page 39 Quick Wave Sequencer User s Guide ka LC Waveform of sequence step n Waveform of sequence step
14. waveforms Page 6 Quick Wave Sequencer User s Guide System requirements System requirements for the Quick Wave Sequencer Personal computer IBM PC AT compatible machine with Processor i486 66 or faster At least 16MB memory CD ROM drive Hard disk installation requires a minimum of 20 MB of free disk space Microsoft mouse or compatible pointing device VGA 800x600 display adapter capable of 256 color depth and color monitor Printer compatible with Microsoft Windows Operating system Microsoft Windows 95 NT 2000 Professional XP Professional and Internet Explorer 4 01 or later Microsoft Windows NT Workstation 4 0 which Service pack3 is installed in GPIB card for personal computer GPIB board compatible with NI 488 2M made by National Instruments Corp AT GPIB TNT PCMCIA GPIB or PCI GPIB VISA 2 6 or later GPIB cable A 24 pin cable with connector conforming to ANSI IEEE std 488 1 1987 AC Power Supply Single phase system PCR L series To use this application the PCR L the main unit must have ROM Ver2 04 or later IB11 PCR L GPIB interface board for PCR L Equipment under test EUT EUT is able to operate with power that the PCR L series supplies Page 7 Quick Wave Sequencer User s Guide Install Before using Quick Wave Sequencer you must install the software on your PC hard disk Installation is performed by the setup program SETUP EXE found on the program diskette To install Setup 1 Laun
15. waveforms containing harmonic components For the harmonic component you can set the phase and any level from primary to 50th order To select a harmonic composite waveform select the waveform bank number from No 1 to No 13 and select Edit Change Category Harmonics menu You can also position the cursor at a waveform bank number and right click to select Harmonics from the popup menu Harmonics sets data after the category is changed Double click the waveform bank number or select Edit Edit Wave Bank Harmonics Bank 1 Harmonics Bank O M Harmonics Level o v rz Distortion Factor zz 0 0 Convert into Vrms Yims Set Voltage 100v 100 0 115v 115 0 Cancel Harmonic level Set level and phase for the harmonic component of each order The default level is 100 for the primary order and 0 for all other orders The default phase is 0 for all orders To edit harmonic component data double click an order number in the Harmonics Level list box Harmonic Order 2 Ed Level E Phase fo oaee Level Set the level of an order number Possible settings range from 0 to 100 Phase Set the phase for an order number Possible settings range from 0 to Page 25 Quick Wave Sequencer User s Guide 359 When you ve finished setting level and phase click OK Distortion Factor The following indicates a theoretical distortion factor of the waveform
16. Creating new data cette ia ee ec t e ee Ee Pe eie 36 Reading existing data ce RAO etie RAV AW tert to de eR euren 36 Datastorage RM C CR 36 Sequence step insertion generation sesssssesseseeeeer ener nnns 37 Sequence Step CUL iie TE RE ERR OE TH ORE RS nae cae Re e ER ae ean SE E ERE CHR EUER DG 37 SEQUENCE gqu M 37 SEQUENCE Step editing eet etian etat etti tebtet ums 38 Description of sequence parameters essssssssseseeeee ener nnne enn 38 Simple printing mero oie etie ee ete e TRE ee e De He E 41 Chapter 5 Performing the Test iuter Leite en Le de edida Use de b due 42 GPIB Setting 5 4 e denn Sie Ae rece es 42 Procedure for performing the test ssssssssseseeseeeeereeeene eene 44 Setting Running Information sssssssssseseseseeeeeeeenennee nennen nnne nennen nnne nns 45 Test executlol e rr RO e Rr p de lg He i e e RO TR ERU 47 Transmission of waveform bank eene ener reet nre 50 Transmission of sequence step ener ener nnne nnne enne 51 Chapter o Sequence WIZAard re aH ele i et p ae EE E idi tae te iba 52 Simulation Wizard To oda T ona aod abeo teer tbt tti ben er 52 Abrupt phase shift wizard for single plhase ccsccesccsseesseesceeeeeeeeeeeeeeeeseenseceseeseceeeneeenes 56 Appendix 1 Error Me SSaQ S 0 ccccccceceeeceeeeceeeeeeeceecaeceeeeeeesecacaaeeeeeeeeeseccaaeseeeeesetenniaaeees 58 Application error messages nennen nennen nennen err
17. ElectronicTest Instruments amp PowerSupplies Software for AC Power Supply SDO4 PCH L E Ver 1 0 Software Quick Wave Sequencer User s Manual jor Dip interrupt variation oltaee Part No Z1 001 882 IB001675 Dec 2006 Quick Wave Sequencer User s Guide Use of This User s Manual Please read through and understand this Users Manual before operating the product After reading always keep the manual nearby so that you may refer to it as needed When moving the product to another location be sure to bring the manual as well If you find any incorrectly arranged or missing pages in this manual they will be replaced If the manual it gets lost or soiled a new copy can be provided for a fee In either case please contact Kikusui distributor agent and provide the Kikusui Part No given on the cover This manual has been prepared with the utmost care however if you have any questions or note any errors or omissions please contact Kikusui distributor agent Trademarks Microsoft and Windows are the registered trademarks of Microsoft Corporation in the U S and other countries IBM and PC AT are the registered trademarks of International Business Machines Corporation National Instruments NI 488 NI 488 2M AT GPIB TNT PCMCIA GPIB and PCI GPIB are the trademarks or registered trademarks of National Instruments Corporation Other company names article names and product names in this manual are generally the tr
18. Sequencer User s Guide Sequence step editing To edit sequence steps position the cursor over the sequence step number and select or double click Edit Edit You can also position the cursor over the step number and right click to select Edit from the popup menu Sequence parameters that are required for a sequence are modified in the edit dialog box i3 Instant interrupt for 1s Sequence Builder CE x File MESES View Tools Instrument Help Di Cut Ctrl 3 B Copy Ctrl C Paste trl Hy 50 00 OFF 0 0 0025 oO N Inset m 50 00 OFF 0 0 0960 0 E dit Enter 50 00 OFF 0 D 0 020 0 o Lo RN pem 50 00 OFF 0 0D 20000 0 gt Edit Sequence step NUM 7 Description of sequence parameters Sequence Step 2 xi Voltage V fo Voltage Ramp Frequency Hz 50 Frequency Ramp Time h min s FC p NN Wave Bank p v Output Wave Sync Cancel Sequence parameter items Voltage Set a value for output voltage in the range 0 0 305 0 V Page 38 Quick Wave Sequencer User s Guide Voltage Ramp It takes the time set in this step to changes from the voltage set in one step before to that set in this step If this item is not checked the voltage is immediately changed to that set in this step In this case first ramp value is disabled Frequency Set a value for frequency output in the range 1 00 999 9 Hz Frequency Ramp It takes the time set in this step
19. ademarks or registered trademarks of their respective companies Reproduction and reprinting of this manual whole or partially without our permission is prohibited Both product specifications and manual contents are subject to change without notice Copyright c 1993 2006 Kikusui Electronics Corp All rights reserved Printed in Japan Page 2 Quick Wave Sequencer User s Guide Chapter 1 Introductio i a a a a a a eene te sten innert nennen nn 5 OVeLVI6 Winston ein e ee OL e Ro HE HUE RED ed ee ERE ee es 5 qM E 6 System requirements o ari tbe a e t e ve atia tti s 7 Install gosse ct et EN a dr e 8 PLECAULI ONS c eh 9 Chapter 2 Introduction to Testing ssssssseeeee memes 12 GPIB setting csset e ee e eee ci een e et 12 The Dipimterruption test a thee ede Sees eek Sood cee e REN TI E ee Ri LR ud 12 Executing the voltage variation test essere ener nnne 19 Chapter 3 Wave Bank Server waveform generation sse 21 Waveform generation categories sess nentes 21 Sine Waveltotth ecc dee eR Uem ig ep isl teet Re edt 22 P ak clipped waveforms eee o E ERR basse iecit ide 23 Harmonic composite waveform esses eene ener enne nennen nn nennen 25 User detined Wavetormi ne e ede ee e t es 27 Simple printing ecce e Ree e eee e ee eter eee de 35 Chapter 4 Sequence Bullder zer EHE HR RE Da eei npe a dale Da Pas 36
20. always connect OUTPUT terminal N to terminal G If OUTPUT terminals G and N at the rear of the PCR L are not connected while an oscilloscope is used to observe PCR L output a potential difference is produced between the PC or PCR L and oscilloscope cables possibly resulting in electric shock or equipment damage You also need to connect the oscilloscope GROUND to OUTPUT terminal G at the rear of the PCR L Caution Stopping the test To stop the test click the Stop button in the execute dialog In an emergency in which the EUT is damaged switch off power to the PCR L When a GPIB error occurs click OK button to exit Caution DC component This application software is capable of outputting triangle and square waves These waveforms are set so that the AC power source contains the DC component When AC current containing a DC component is input overcurrent may occur in some EUTs setting off an alarm Immediately stop testing when this occurs lt Caution gt Bank14 is used with this software After you used this software do not output Bank14 by using a remote controller or GPIB However you can output Bank14 when you do not let this software run after having set Bank14 anew Page 9 Quick Wave Sequencer User s Guide For trouble free operation Close all other applications before running this software If your PC hard disk has 5MB or less of free space data may not be written properly to disk Before running t
21. are added one by one The maximum step number is 100 Sequence step cut pi Instant interrupt for 1s Sequence Builder E dit File Click the cursor on any sequence step number m then click SHIFT sequence step number n This selects all sequence step numbers from m to n Select the Edit Cut menu or enter CTRL X from the keyboard to delete these sequence step numbers from m to n The sequence step data is moved to the clipboard To restore this data move the cursor to the desired insertion point of sequence number and select Edit Paste menu or enter CTRL V from the keyboard The clipboard contains only the most recent copying or cutting data if any material was copied or cut in other work the contents of the clipboard are cleared PII E3 View Tools Instrument Help O a Copy Ctrl C Stef Paste Cisy Frequency FRamp Hour Minute Second BankNo 50 00 OFF 0 0 0 025 0 E U iet sooo OFF 0 0 oss 0 EX ER Enter 50 00 OFF 0 D 0 020 n L Running Info 50 00 OFF 0 0 20 000 0 H Cut the selection and put it on the Clipboard NUM A Sequence step copy As with sequence step cut select any sequence step number and select Edit Copy menu or enter CTRL C from the keyboard This moves the selected data to the clipboard Position the cursor at the sequence step number you want to copy and select Edit Paste or enter CTRL V from the keyboard Page 37 Quick Wave
22. ce steps Create a sequence step in Sequence Builder Setting output voltage and frequency Open Edit Running Info menu in Sequence Builder and set all items Transfer bank waveform and sequence data Transfer the waveform bank data to the PCR L by selecting Instrument Send menu in Wave Bank Server Transfer sequence data to the PCR L by selecting Instrument Send menu in Sequence Builder Performing the Test Page 44 Quick Wave Sequencer User s Guide Open the Execute screen by selecting Instrument Run menu in Sequence Builder and execute the test Click the Sequence execution button When output must be turned on before performing the sequence test click the Output button for the operation settings PCR L is enabled or disabled Setting Running Information Running Information sets voltage and frequency in advance before and after sequence operation Open the dialog box by selecting Edit Running Info menu in Sequence Builder W Instant interrupt for 1s Sequence Builder i OF XI File B View Tools Instrument Help D Sut Str 829 Spy tHe Haste ER Insert Insert E Edt Enter Running Info Running Information Running Info Ed Before Sequence Settings After Sequence settings Voltage V Voltage V 100 Frequency Hz po Frequency Hz 50 Sequence Settings Start Step f Loop Cancel Before Sequence settings Enter values for
23. ch Windows close all other applications 2 Insert program disk into a CD ROM drive 3 To install Quick Wave Sequencer follow the instructions on the screen Note In the operating environment that Windows Installer is not installed in a reboot may be required The default location for the Quick Wave Sequencer files is C Program Files Kikusui NQuick Wave Sequencer You can modify default setting to change the destination directory Program Folder specifies the folder name for the application The default name is Kikusui Quick Wave Sequencer End of installation When installation is completed Windows creates the folder for Kikusui Quick Wave Sequencer that is accessible from the Windows Start menu f Kikusui Quick Wave Sequencer E3 File Edit View Help Sequence Wavebank Builder Help Server Help Sequence Builder 658 bytes Ui 1 object s selected Page 8 Quick Wave Sequencer User s Guide Precautions For correct and safe use of this software read and make sure you understand the following descreptions Safety Warning Connecting the EUT to the PCR L AC power supply The test involves high voltages and large AC currents Take extra care in connecting the EUT to the PCR L power supply since improper connections can damage equipment and lead to fire or explosion For connection directions refer to the PCR L instruction manual Warning For PCR L AC power supply
24. data Creating new data To create new data select File New The sequence step is cleared with file name untitled Reading existing data A saved data file can be reopened for test execution To read data select the File Open menu to display the file name selection dialog box Choose an existing file that was previously saved with extension WSL When the Sequence Builder file is read a Wave Bank Server waveform generation application file with extension WBL is read simultaneously Data storage Sequence Builder can store set data on a disk including set data for executing sequences and waveform generation application files To save data select File Save menu or File Save As menu Either displays the dialog box for file name selection Name the file appropriately The Sequence Builder extension is WSL A Wave Bank Server file is saved under the same name at the same time but with an WBL extension Note The Sequence Builder and Wave Bank Server files are the only paired files that have different extensions Changing the file name for one changes the file name for both When moved or copied the two files must reside in the same directory Page 36 Quick Wave Sequencer User s Guide Sequence step insertion generation Click the specified step number and select the Edit Insert menu or the INSERT key A new step is inserted just before the specified step number and step numbers following this step
25. e nennen 58 Page 3 Quick Wave Sequencer User s Guide GPIB Error 06883268 og teet pt iO e RU oe ee Rie eee es 58 Appendix 2 Application Specifications ssssssssssseeeeeeee nennen 59 Operating environment sss eere enne trennen trennen enr enne nn nennen 59 Sequence Builder inicia e ee e Hed C He e P LR YT EE de 60 Wave Bank Servet zen ento terae orco d ge Fa eue esee eas 60 Page 4 Quick Wave Sequencer User s Guide Chapter 1 Introduction Overview Software Quick Wave Sequencer SD04 PCR L E hereinafter referred to as Quick Wave Sequencer is application software that uses the sequence functions of our company s AC power supply PCR L series hereinafter referred to as PCR L to perform AC power source simulation Quick Wave Sequencer consists of the following two pieces of software which are activated and operated simultaneously 1 Sequence Builder SeqBuild exe This software generates the sequence parameters of the PCR L such as output voltage frequency and time for the waveform bank that was created and transferred from Wave Bank Server and transfers those parameters to the PCR L The software executes the transferred sequence and conducts voltage variation tests and Dip interruption tests The software contains wizard functions that automatically set the input of sequence parameters 2 Wave Bank Server WBanksvr exe The software creates AC voltage waveforms
26. e running time lt Note gt To halt sequence execution click the Stop button in the Running dialog In an emergency in which the unit being tested is damaged switch off power to the PCR L This will produce an GPIB error Click OK button to exit The same test can be repeated using the Sequence execution button in the Run dialog Executing the voltage variation test This section describes the steps required to conduct a voltage variation test using the Sequence Wizard 1 Make the test plan Steady voltage Single phase 100 0V Frequency 50 00 Hz Front slope time 1000 ms 1s Fluctuating voltage 40 0 V Fluctuation sustain time 1000 ms 1s Postslope time 1000 ms 1s Fluctuation interval 20s Number of repeats 4 repeats les 1s 1 91 20 _ _ TITULI TTELELLLEELLLLLLPPCTETTPPPEPPEPETPPPPEPEEFEPPPTTTPELLLLLECLLLLU 2 Open Sequence Builder and select Sequence Wizard When you open Sequence Builder Wave Bank Server is opened automatically In Sequence Builder click Tools Sequence Wizard and select the Abnormality Simulation Wizard from the Wizards List Type at file name Page 19 Quick Wave Sequencer User s Guide at Name Click OK 3 Enter test conditions Simulation Wizard No 1 x Type Edit Variation Voltage Vrms 40 T1 Slope time ms 1000 T2 Variation period ms 1000 T3 Slope time2 me 1000 Repeat times
27. e the waveform bank is used for sequence step data Data is sent continuously from Bank 1 to Bank 13 Transmission of the waveform bank data can be cancelled but if data has not been sent to the PCR L testing waveforms will be different from the set data If waveform data is not transmitted to the PCR L when the test is conducted from the menu of the Sequence Builder Instrument Run waveform bank data is transmitted with sequence steps pul LESSE CS ETT for 1s Wave Bank Server iof x File Edit View Instrument Help BankNo Categories Applets l 0 SIN 1 SIN 2 SIN 3 SIN 4 SIN ib a Sending Normal SIN Wave 8 SIN g SIN Sending to Wave Bank 3 mau Note The following are times for each bank at which a user defined waveform is transferred When the user defined waveform has been transferred to PCR L data transfer of that bank is skipped at waveform transmission Waveform data is cleared once power to the PCR L is shut off PCR L s ROM version 2 04 Single phase system Approximately 45 seconds Page 50 Quick Wave Sequencer User s Guide Transmission of sequence step Perform transmission of sequence step data to the PCR L from the Sequence Builder Instrument Send menu pi Instant interrupt for 1s Sequence Builder Too OFF 50 00 0 0 OFF 50 00 OFF D 0 020 0 OFF 50 00 OFF n 0 Send To PCR1000L Version 2 04 x Sequence Step 3 Send Sequence
28. eform appears in the main window Page 28 Quick Wave Sequencer User s Guide Untitled Easy Power Waveformer QW Iof x File View Shadow Instruments Help ojala v Ready NUM 4 Shadowing the unused area Easy Power Waveformer converts waveform data into a user defined waveform acceptable to the PCR L The format of user defined waveform data is 12 bits x 1024 words This is the only acceptable size and format Generally speaking not all 1024 words captured from the oscilloscope will fit in the full scale rectangle In many cases the 1024 word buffer contains data longer than the 1 period to be repeated and the signal level of the captured waveform does not fit You must shadow the unused area using left right upper and lower shadows The effective area left unshadowed is accepted as one waveform cycle Upper Shadow oo J Valid Area 1 ee 3 Lower Shadow Left Shadow Right Shadow To shadow the waveform point and click the right mouse button at the boundary of the area to be shadowed A popup menu prompts you to select a shadow type Selecting left displays a dialog box Page 29 Quick Wave Sequencer User s Guide Untitled Easy Power Waveformer QW File View Shadow Instruments Help pojela 1 m x Pointing Address 5 OK Waveform D ata Cancel lt lt Left lt lt i Fast Move In the above dialog Point Address 25 indicates that the current po
29. en set The setting range is from the start step to the number of set steps Loop Set the number of operation repeats from start to end step When testing data created by the Wizard this value is already set The setting range is from 1 to 99999 99999 sets for infinite repetitions lt Note gt In the PCR which SUBCPU version 2 07 installed in 99999 is not infinite repetitions but 99998 repetitions The version number of SUBCPU is indicated at the upper right of the indicator frequency indication section when turning on the PCR If you want to update SUBCPU in your PCR contact your Kikusui agent Page 46 Quick Wave Sequencer User s Guide Test execution Run the test from the Sequence Builder Instrument Run menu If sequence steps are edited but sequence data is not transmitted to the PCR L before the Execute dialog box appears transmit waveform bank and sequence data to the PCR L You can cancel transmission waveform bank and sequence data Following the first test after power on or modification of the last setting perform the operation by transmitting data Neither waveform bank or sequence data is saved to PCR L nonvolatile memory The waveform bank data cannot be transmitted unless PCR L output is turned off R3 Instant interrupt for 1s Sequence Builder CE Xx File Edit View Tools Mi oleja amp a j Sequence Run NUM A Run To PCR1000L Version 2 04 Ea Operation settings
30. equence Wizard GPIB setting Since Quick Wave Sequencer communicates with the PCR L through GPIB testing requires the setup described below Install the PC GPIB card and the driver software For information on installing the GPIB card refer to the instruction manual from National Instrument Corp Set the PCR L GPIB address to 5 The GPIB address for Quick Wave Sequencer is preset to 5 lt Descreption gt Set the PCR L GPIB address by pressing the SHIFT F GPIB keys from the PCR L panel In GPIB address setting mode enter 5 Then restart the PCR L to validate the GPIB address just entered Properly connect the designated GPIB cables The Dip interruption test This section gives a step by step description of an Dip interruption test using the Sequence Wizard 1 Make a test plan Steady voltage Single phase 100 0V Frequency 50 00 Hz Instant interrupt voltage 0 0 V Instant interrupt start phase angle 90 Instant interrupt duration 1000 ms 1s Instant interrupt interval time 20 s Number of repeats 4 repeats Page 12 Quick Wave Sequencer User s Guide Kl 208 HH s 4 amp 2 s YAMAMURA ai O 2 Open Sequence Builder and open Sequence Wizard When you open Sequence Builder Wave Bank Server is opened automatically Click the Sequence Builder Tools Sequence Wizard and select Abnormality Simulation Wizard from the Wizards List
31. fined waveform was created exit from Easy Power Waveformer on the File Exit menu Closing the application displays a message box Do you wish to update waveform data No waveform is created when you click No To generate the waveform created by Easy Power Waveformer as a user defined waveform click Yes Clicking Cancel returns you to the original screen Generating triangle waves Generate triangle waves by selecting Triangle Waveformer from the Applets of the user defined waveform In the Applets double click Triangle Waveformer or select Triangle Waveformer with the cursor and click the Import button The system displays a screen for creating triangle waves Triangle Waveformer Ea m Duty Ratio A so B o Close To generate enter a setting in the range 0 100 for A This automatically calculates the factor for B Once you ve entered a setting for A click the Close button The Do you wish to update waveform data message box appears Page 33 Quick Wave Sequencer User s Guide Triangle Waveformer x N Do you wish to update waveform data No waveform is created when you click No To go back and set a different value for A click Cancel Clicking Yes generates a triangle wave in the user defined waveform User Defined Bank 3 Eg Easy Power Waveformer Square W aveformer Triangle eformer Inspect the created waveform and click OK Generating square waveforms Square wave
32. he software check the remaining capacity of the hard disk Before running the software initialize PCR L Set the GPIB address for the PCR L to 5 the GPIB address for Quick Wave Sequencer is preset to 5 lt Descreption gt To initialize the PCR L press the SHIFT 6 RESET keys and SHIFT ENTER keys in sequence from the PCR L panel Initialization takes several seconds Set the GPIB address for the PCR L by pressing the SHIFT F GPIB keys from the PCR L panel In GPIB address setting mode enter 5 Then restart the PCR L The validates the GPIB address just entered Page 10 Quick Wave Sequencer User s Guide General Items for Software Use Voltage settings in the application are based on the effective value of the sine wave Voltage set values for another waveform may have different meanings For set voltage value V voltage 2 J 2 V p p is output A phase angle set by the Simulation Wizard can be entered in 1 Since the PCR L sequence function needs to be specified in 1 ms the value at a frequency of 50 Hz is set in 360 20 ms 18 You can enter up to the 50th order for harmonic composite waveforms The actual output waveform depends on frequency characteristics of the PCR L Page 11 Quick Wave Sequencer User s Guide Chapter 2 Introduction to Testing This chapter describes a simple method for performing the abnormality simulation Dip interruption and voltage Variations test using the S
33. ime after which the abrupt phase shift waveform is output in the range 0 001 9999s Operating time is rounded up to an integer multiple of 1 cycle Voltage V Set output voltage in the range 0 0 305 0V Frequency Hz Set output frequency in the range 1 00 999 9Hz Enter all test conditions and click the Next button to go to the Finish screen Click Finish to exit Wizard Waveform bank and sequence step information is developed to Wave Bank Server and Sequence Builder Page 57 Quick Wave Sequencer User s Guide Appendix 1 Error messages Application error messages You are using earlier version of PCR L unit ROM version 2 04 or later is required PCR L ROM version is older than version 2 04 Contact us GPIB error messages Unable to open instrument in device DEVx Unable to close instrument in device DEVx Access to GPIB failed in the initial stage The device name has a specific name such as DEV1 A problem has occurred For example the GPIB driver is not correctly installed or the appropriate device name has not been defined in the GPIB driver environment setting Different model access in device DEVx A wrong instrument has been connected through GPIB An instrument other than the PCR L may have been connected Error writing to instrument from file in device DEVx Error reading from instrument to file in device DEVx A communication error occurred while writing or reading data to and from the instru
34. ing user defined waveforms Selecting Applet enables the generation of each waveform Easy Power Waveformer Creates artificial waveforms obtained from the oscilloscope Square Waveformer Creates square waves Triangle Waveformer Creates triangle waves After confirming setting and display data click OK Obtaining data from the oscilloscope Selecting Easy Power Waveformer from the Applets of the user defined waveform enables generation of artificial waveforms obtained from the Page 27 Quick Wave Sequencer User s Guide oscilloscope Below is a procedure for creating artificial waveforms Obtain waveform Shadow the unused area Enlarge the waveform Change the 1 point data Obtaining a waveform The user defined waveform created by Easy Power Waveformer is digitized Obtaining the waveform is the first step toward creating a user defined waveform We use a digital oscilloscope connected to the GPIB interface The following oscilloscopes are compatible with this application Kikusui COM3000 series Kikusui COM7000A series Kikusui COR5500 series LeCroy 9350A series Tektronix TDS320 series For the names of specific models in the series refer to Appendix Application Specifications at the end of this volume With any of these oscilloscopes you can read waveforms through the GPIB interface Select Instrument Read from the Digital Scope menu select the CH1 or other channel and click Go The wav
35. ion waveforms triangle and square waves and artificial waveforms obtained from the oscilloscope It transmits the created waveforms to the PCR L waveform bank enabling use as the waveform setting for Sequence Builder In Wave Bank Server waveform generation creates the waveform without the attributes of voltage time or frequency Waveform generation categories Waveform generation has four categories e Sine waveform A combination of sine waves is used to achieve instant interrupt and voltage fluctuations Peakclipped waveform The distorted sine waveform is created by specifying the crest factor Harmonic composite waveform This category creates a waveform that includes harmonic components Harmonic components can be set from primary to 50th order You can also set the level and phase of each order User defined waveform This category covers triangle and square waves and artificial waveforms obtained from the oscilloscope Page 21 Quick Wave Sequencer User s Guide Sine waveform This is the default Wave Bank Server category In New Creation waveform bank numbers from 0 to 13 are all sine waveforms Bank No 0 of the waveform may not be changed to any category other than sine waveforms To select sine waveforms in Category position the cursor at a waveform number and select Edit Change Category SIN menu You can also select SIN from the right click popup menu In Sin you can select Category but may not
36. ment through GPIB Check that the GPIB cable is securely connected and that PCR L power is on Confirm that the I O Time Out of the GPIB driver software of National Instruments is set to 10s initial value You are using earlier version of PCR L unit ROM version 2 04 or later is required The ROM version of the PCR L is older than version 2 04 Page 58 Quick Wave Sequencer User s Guide Appendix 2 Application Specifications Operating environment Personal Computer IBM PC AT compatible with the following Processor of i486 66 or better Memory of 16MB or greater CD ROM drive Hard disk at least 20MB of free space for installation Microsoft mouse or compatible pointing device Display adapter of at least VGA 800x600 and 256 color depth and color monitor Printer compatible with Microsoft Windows Operating system Microsoft Windows 95 NT 2000 Professional XP Professional and Internet Explorer 4 01 or later Microsoft Windows NT Workstation 4 0 which Service pack3 is installed in GPIB card for PCs GPIB board compatible with NI 488 2M made by National Instruments Corp AT GPIB TNT PCMCIA GPIB or PCI GPIB VISA 2 6 or later GPIB cable Cable with a 24 pin connector conforming to ANSI IEEE std 488 1 1987 Power supply AC power Single phase system PCR L series This application requires a PCR L ROM version of 2 04 or later IB11 PCR L PCR L GPIB interface board Equipment Under Test The equipmen
37. ncy Hz 50 00 Interval s 20 0 imulation No i Simulation No Loop times 4 m Simulation Data Type Dip Interruption Dip Interruption Voltage Vrms 0 0 T1 Start phase 90 T2 Dip Interruption period ms 1000 Repeat times 1 coc e After confirming the current test conditions click the Next button Abnormality Simulation Wizard Finish You are finishing the wizard Used sequence steps 4 Generated files Instant interrupt for 1s WSL Instant interrupt for 1s w BL Cancel Help lt Descreption gt Used sequence steps Number of steps created in Sequence Builder Files to be created Instant interrupt for one second WSL Data file for Sequence Builder Instant interrupt for one second WBL Data file for Wave Bank Server Click on the Finish button to finish setting of the test conditions using Wizard When you exist from Wizard the bank waveform and sequence steps are created in Wave Bank Server and Sequence Builder Page 16 Quick Wave Sequencer User s Guide a Instant intrrupt for 1s Wave Bank Server e x File Edit View Instrument Help alaj le a Ja eS L IN l Calpe pum NA 6 sent LINZ i KZ 8 Tam N L kA 9 A LN 10 Huemugt Em 11 EI 12 Ek 13 Ready pi Instant intrrupt for 1s Sequence Builder of x File Edit View Tools Instrument Help ojeju eral z iere I el xj For Help press
38. pect and change GPIB settings specify a device name by selecting Instrument Options in Sequence Builder Note that the GPIB address is not specified but the GPIB device name is The GPIB device name is set by the GPIB board device driver At installation the device name is set to DEV5 By default the National Instrument GPIB driver provides device numbers from DEV1 to DEV16 GPIB addresses for corresponding devices are set from 1 to 16 thus a GPIB address of PCR L must be set to 5 to connect to the DEV5 device However this does not necessarily hold if GPIB driver environment settings are modified Caution After changing the GPIB device name restart the application to reflect the changed GPIB device name to Wave Bank Server Set I O Time Out for the National Instruments GPIB driver software to 10s initial value Setting it to 1s or less can produce an GPIB error Note To check for errors run the diagnostic program attached to the GPIB board This application supports only GPIB 0 board control Take the necessary precautions when using more than one GPIB board Page 42 Quick Wave Sequencer User s Guide Pile Es pi Untitled Sequence Builder File Edi View Tools marr CENA z _ Digital si E Bun a Options GPIB Device Name DEVS Cancel Change the options for this application NUM A For more information on the National Instruments GPIB refer to the in
39. s are generated by selecting Square Waveformer from the Applets of the user defined waveform Double click Square Waveformer in the Applets or select Square Waveformer with the cursor and click the Import button The system displays a screen for creating square waves Square Waveformer x Page 34 Quick Wave Sequencer User s Guide To generate enter a setting in the range O 100 for A This automatically calculates the factor for B Once you ve entered a setting for A click the Close button The Do you wish to update waveform data message box appears Square Waveformer x N Do you wish to update waveform data i No Cancel No waveform is created when you click No To go back and set a different value for A click Cancel Clicking Yes generates a triangle wave in the user defined waveform User Defined Bank 3 x Applets Easy Power W aveformer Square W aveformer Triangle w aveformer Import OK Cancel Inspect the created waveform and click OK Simple printing The system can print a waveform category waveform bank No 0 No 13 created by Wave Bank Server in a simple waveform image format To print select File Print To preview the printed image select File Print Preview menu Page 35 Quick Wave Sequencer User s Guide Chapter 4 Sequence Builder In addition to sequence execution and PCR L output control Sequence Builder can edit create and transfer sequence
40. set any other data p Untitled Wave Bank Server iof x File MSIE View Instrument Help amp Copy Ctrl C Haste tl Hy nnlate Change Category EX Edit Wave Bank Enter Peak Clip Hannanics p 2 SIN UserWefned CEA 3 SIN 4 SIN 5 SIN 6 SIN 7 SIN Kal 8 SIN 9 SIN 10 SIN 11 SIN 12 SIN 13 SIN SIN NM 2 Page 22 Quick Wave Sequencer User s Guide Peak clipped waveforms In creating a sine waveform with distorted peak you can specify the degree of waveform distortion To select a peak clipped waveform select a waveform bank number from No 1 No 13 and select Edit Change Category Peak Clip menu You can also position the cursor at the appropriate waveform bank number and right click to select Peak Clip from the popup menu Peak clipped Waveform sets data after categories are changed Double click the waveform at which the category has now been changed or select Edit Edit Wave Bank Peak Clip Bank 1 Ed Crest Factor 1 35 P OK Cancel Peak clipped waveform in single phase systems Page 23 Quick Wave Sequencer User s Guide Crest factor The Crest factor is used to specify the degree of waveform distortion 1 10 1 40 for single phase systems To display a preview of the waveform click the View button Clicking OK completes entry Page 24 Quick Wave Sequencer User s Guide Harmonic composite waveform This category creates
41. settings following sequence execution When time management is required for the last sequence create a sequence step to output Voltage Frequency and Waveform bank of After sequence execution settings and add this step to the last step Sequence The sequence test can be executed for two or more steps of sequence step data Start step Set the step to start the sequence To test data created by the Wizard this value is usually set to 1 in the range from 1 to the end step End step Set the step to end the sequence To test data created by the Wizard this value has been set to the number of steps set in the range from the start step to the number of steps set Loop Set the number of repeats from start to end step To test data created by the Wizard this value has been set in the range from 1 to 99999 repeats The value 99999 indicates an infinite number of repeats lt Note gt In the PCR which SUBCPU version 2 07 installed in 99999 is not infinite repetitions but 99998 repetitions The version number of SUBCPU is indicated at the upper right of the indicator frequency indication section when turning on the PCR If you want to update SUBCPU in your PCR contact your Kikusui agent Page 49 Quick Wave Sequencer User s Guide Transmission of waveform bank Perform waveform transmission to the PCR L on the Wave Bank Server Instrument Send menu Wave Bank Server waveform bank data is required sinc
42. sition of the left shadow boundary is 25 Similarly Waveform Data 1584 indicates that the data value of that position is currently 1584 As a general rule bottom data is assigned zero and top data is assigned 4095 Set other shadows on the waveform in the same way Clicking on the left or right button enables you to move between the boundary positions of the left and right shadows Similarly clicking on the Up or Down button enables you to move between the boundary positions of the upper and lower shadows These actions move a shadow position by one digit To move the shadow quickly check the Fast Move check box before performing Page 30 Quick Wave Sequencer User s Guide Enlarging the waveform Untitled Easy Power Waveformer QW Em E File View Shadow Instruments Help olsa l me ve 100 200 300 400 Ready 42365 y 1644 NUM 7 Assume shadow setting is complete and that the illustrated waveform has been obtained Since a PCR L user defined waveform must be 12 bits x 1024 words the above waveform must be enlarged to fit the size requested by the PCR L Select Shadow Expand waveform from the menu bar The waveform data is enlarged to the full size Untitled Easy Power Waveformer QW Oy x File View Shadow Instruments Help nisja 15 vej Q bBRb 4 oO bE 41 o L 0 x 70 y 2 3349 NUM 4 Page 31
43. step to the Page 51 Quick Wave Sequencer User s Guide Chapter 6 Sequence Wizard Within Quick Wave Sequencer the Sequence Wizard function can be used to automatically create waveform bank and sequence steps for voltage tests Dip interruption and voltage variation and abrupt phase shift tests which are general AC tests Click the Sequence Builder Tools Sequence Wizard menu to open Sequence Wizard Sequence Wizard consists of two subfunctions Simulation Wizard which executes voltage testing and Abrupt Phase Shift Wizard for single phase systems On the Sequence Wizard screen enter the file name to which waveform bank and sequence step information developed by the Wizard is saved Simulation Wizard Select Abnormality Simulation Wizard in Sequence Wizard to start Simulation Wizard Within one test Simulation Wizard can create compositions for up to four types of voltage tests Dip interruption and voltage variation The simulation test is executed in the order No 1 No 2 No 3 and No 4 Selecting the Simulation Wizard opens the Step 1 of 2 screen Abnormality Simulation Wizard Step 1 of 2 Voltage V jim Frequency Hz 50 Intervals s fio Simulation Data No Type Test Voltagefv TestTime ms Repe 4 Not Used Not Used Not Used Not I sed a Ww N oft Edit Loop times f Cancel Help Set Voltage Frequency Interval and Loop in this screen Vol
44. struction manual lt Descreption gt Set the PCR L GPIB address from the PCR L panel with the SHIFT F GPIB keys Enter 5 in GPIB address setting mode then restart the PCR L to validate the new GPIB address Page 43 Quick Wave Sequencer User s Guide Procedure for performing the test Use one of the following procedures depending on the specific test For a test to output a steady AC waveform The test continuously outputs sine harmonic composite and user defined 1 Creating bank waveforms Set the waveform to output to any bank number in Wave Bank Server The sine waveform is set to bank number 0 Creation and transfer of bank waveforms Transmit waveform bank data to the PCR L by selecting Instrument Send menu in Wave Bank Server Performing the Test Open the Sequence Builder Instrument Run menu and set Voltage Frequency and Wave Bank that ware set Before or now sequence execution settings Click the set button then the Output button for the operation settings PCR L output is enabled or disabled For a test to output AC waveforms that change over time The test outputs AC waveforms for which output voltage frequency and waveform change as a function of time Sequence Wizard automatically performs steps 1 to 3 below 1 Creating bank waveforms Set the waveform to be tested at any bank number in the Wave Bank Server A simple sine waveform is set to bank number 0 Creating sequen
45. t being tested must operate with the PCR L series Page 59 Quick Wave Sequencer User s Guide Sequence Builder Edit sequence steps in the following range Item Numeric Remarks NEN LEN EN NEG CNN No of sequence steps Upto100 Voltage range setting 0 0v 3050V Cd Voltage ramp setting 1 Enabed Cid Frequency setting range 100Hz 9999Hz O Oo re ee eee Frequency ramp setting Enabled Sequence setting time 0 000s 999h59m59 999s Whole sequence duration must be 0 001s or greater Waveform bank 07 13 For Bank 0 sine wave is fixed specification Output control ON OFF Pp Sequence trigger setting Enabled Waveform synchronizing Enabled function Functions for data storage and Yes WSL Sequence Builder file reading WBL Wave Bank Server file The above two files are saved and read simultaneously 1 Setting resolution and setting accuracy are shown in the following J Sefingresoton setting accuracy Wave Bank Server Numerical value contents No of waveform banks 14 0 13 0 Sine wave is fixed Selectable categories Sine waveform Peak clipped waveform Harmonic composite waveform User defined waveform Aplet in user defined Triangle waveform waveform Square waveform Waveform obtained from Data storage function is oscilloscope provided Compatible oscilloscopes COM3101 3051 Kikusui COM3000 series COM7203 02A Kikusui COM7000 series COR5502 01 61 41 21 Kikus
46. tage Set PCR L output voltage before and after testing in the range 0 0 305 0V Page 52 Quick Wave Sequencer User s Guide Frequency Set PCR L output frequency before and after testing in the range 1 00 999 9Hz Interval Set the time interval between tests in the range 1 0 999 9s Loop Set the number of voltage test repeats Dip interruption and voltage variation for the whole system in the range 1 99999 repeats The value 99999 sets for an infinite number of repeats Set the data Double click the specified number or position the cursor at the Edit button and click to display the Simulation Data Input screen Select Dip interruption or Voltage Variation from abnormality simulation mode and click the Edit tab to enter data Not Used cancels data for that number Simulation Wizard No 1 Ed Type edit Abnormality Simulation mode Voltage Variation Cancel Apply Help Setting Dip Interruption Test Data Simulation Wizard No 1 x Type Edit Dip Interruption Voltage Vrms fm T1 Start phase o T2 Dip Interruption period ms fi 0 Repeat times fi Cancel Apply Help Page 53 Quick Wave Sequencer User s Guide Select Dip Interruption from abnormality simulation mode and click the Edit tab This sets Dip interruption test conditions Dip Interruption voltage Vrms Set the testing voltage for instantaneous interruption in the range
47. ui COR5500 series 9350A 9354A Lecroy 9350A series TDS310 320 350 Textronics TDS series Page 60 Quick Wave Sequencer SD04 PCR L User s Manual
48. voltage value Clicking the set button sets the value in the PCR L along with frequency and waveform bank number Frequency Set the PCR L output frequency in the range 1 00 999 9Hz Clicking the Set button sets the value in the PCR L along with voltage and waveform bank number Wavebank Set a value for output bank number in the PCR L When the dialog box is opened the default value is 0 When the waveform bank is created by Wave Bank Server and transmitted to the PCR L you can output the bank stored in the PCR L bank in the range 0 13 Clicking the set button sets the number in the PCR L along with voltage and frequency After sequence execution settings Voltage Set the voltage for the PCR L on test completion in the range 0 0 305 0 V When the set value is 152 5V or greater click the check box for the 200V Range confirming that PCR L output is off Then set voltage Frequency Page 48 Quick Wave Sequencer User s Guide Set the frequency to be set in the PCR L on test completion in the range 1 00 999 9Hz e Wavebank Set an output bank number to be output to the PCR L on test completion When the dialog box is opened the default is 0 When the waveform bank is created by Wave Bank Server and transmitted to the PCR L you can output the bank stored in the PCR L bank in the range 0 13 Note It takes at least 10 ms for the PCR L to execute Voltage Frequency and Wavebank of After sequence execution
49. with a composed higher harmonic Equation L oad Eve OD be F y 5 100 DF Distortion Factor L level value at the nth order Convert into Vrms This indicates the set value for the voltage used to output the generated harmonic composite waveform at the effective value When this set value for voltage is input as voltage during sequence execution the harmonic composite waveform can be output at the designated effective value The set values for voltage for the 200V and 230V effective values are the doubled set values for voltage of the 100V and 115V effective values After confirming settings and display data click OK Page 26 Quick Wave Sequencer User s Guide User defined waveform This item uses a subcategory called Applet for simple settings for creating triangle and square waves and the artificial waveform obtained from the oscilloscope To select user defined waveform by category select a waveform bank number from No 1 to No 13 and select Edit Change Category User Defined menu You can also position the cursor at a waveform bank number and right click to select User Defined from the popup menu User define waveform sets data after the category is changed Double click the waveform bank number or select the Edit Edit Wave Bank menu User Defined Bank 3 x Applets Square W aveformer Triangle W aveformer Import Cancel Applets This is a subclassification for creat