Home
GE Druck PACE5000 / PACE6000 Control Modules
Contents
1. Replacement Parts Use only the replacement parts listed in Table 4 2 WARNING TURN OFF THE SOURCE PRESSURE AND CAREFULLY VENT THE PRESSURE LINES BEFORE DISCONNECTING THE PRESSURE LINES FOR MAINTENANCE PROCEED WITH CARE ISOLATE THE INSTRUMENT POWER SUPPLIES BEFORE REPLACING PARTS WITH POWER APPLIED THE INSTRUMENT CONTAINS LETHAL VOLTAGES Table 4 2 Replacement Parts List fe Fuse T2 0A 250V HRC PACE5000 Fuse T5 0A 250V HRC PACE6000 IO FILTER KIT Kit filter CMX XXXxX T Module pressure control refer to data sheet Fuse Replacement Fig 4 1 Replace the fuse when detailed in Section 5 Testing and Fault Finding Set the power switch to OFF Isolate the power supply to the instrument and disconnect the IEC power supply connector 1 Note To gain access to rack mounted instruments it may be necessary to partially or completely withdraw the instrument Isolate pneumatic supplies and de pressurise all pressure supply inlet and outlet lines Remove the fuse carrier 2 from the power supply input socket assembly and replace the fuse cartridge 3 Note Fit the correct type of fuse detailed in Table 4 2 Refit the fuse carrier 2 in the power supply inlet socket assembly Refit and reconnect rack mounted units as detailed in Section 2 Installation Switch on the power supply and set the power supply switch to ON The instrument should now be operational If the fuse blows immediately on switch on contact
2. The PACE instrument contains selectable pressure units and user defined units Use the conversion factors to calculate a user defined unit from the table above Refer to the data sheets SDS0001 or SDS0008 for the list of selectable pressure units K0443 Issue No 3 iv CONTENTS Section Title Page 1 SAA EE 1 1 1 1 a EE 1 1 2 MS rre 2 1 2 1 Beie Le a q 10 EE 2 1 2 2 Packaging for Storage and Tronsportotton 2 1 23 Preparation tere EE 2 1 2 4 RSU leger te 2 2 2 5 ROCE MOUNTON U rt pst OENE 2 7 2 6 Begleet te 2 8 3 SE 3 1 3 1 Preparados a O 3 1 3 2 See ee Ne 3 2 3 3 PCOS US MOS E 3 3 3 4 A E 3 5 3 5 Operation and Example PIOCCQUICS usos ios 3 8 3 6 Elo El EE 3 12 Sel Barometric Re erence O PUN ib la celos 3 13 3 8 UN A A 3 14 59 MS CIPI Cbs UCM APP O ono A a eletece 3 15 K0443 Issue No 3 V 4 MAINTENANC ao bo 4 1 Jet zege ter Lee EE 4 2 MS SDE COM cess eee ee Need ke 4 3 Se Lo LE 4 4 a A 4 5 o e A o PA 4 6 SA ER 4 7 BLU foto yd e a lt Gl ese Rattner O 4 8 Fall iC 21g 218 2 Xe 0 o ge meen EA 4 9 Beie ENER E 5 TESTING AND PAULI PINIDIIN G adios ove Jet ee rei Die le EE EE Elia elle ele ue TeSt EE 55 A II 5 4 Approved SCV ICSAC e Su etorri leeis 6 REFERENCEAND SPECIFICA TEE 6 1 ASIN a o 6 2 Opera leste et Mens a Os 6 3 A O E O 6 4 Medsure PP e oi E OO cate onen 6 5 O ena 6 6 EE oe S EE o AA E 6 7 SUPE SON od 6 8 EE K0443 Issue No 3 vi 4 1 4 1 4 1 4 1 4 1 dial 4 2 4 2 4 4 4 5 al
3. Systems without a negative supply Without a negative supply vacuum pump release the positive pressure from the system to atmosphere through the negative supply port The release from the negative port may be piped away to a place where this discharge causes no disturbance or hazard Alternatively a diffuser may be fitted to the negative supply port to diffuse air flow The need for a negative supply for optimum performance General Supply pressures at least 110 of range but less than the MWP there must be a difference of 10 of full scale between the supply pressure and the maximum output pressure When operating at positive or negative full scale there must be a pressure difference between supply and output to cause a gas flow Operating near atmospheric pressure or below Any controller operating near atmospheric pressure or below requires a vacuum pump or other negative supply connected to the negative supply port for optimum performance Without a vacuum supply as the output pressure approaches atmospheric pressure the differential pressure approaches zero resulting in a reduced flow to the output Reduced flow causes an increase in the time to control to atmosphere especially with large user volumes and an increased overshoot at low pressures see figures 2 3 2 4 and 2 5 Vacuum Pump Each PACE Control Module has a vacuum sensor so it is normal practice to connect a vacuum pump to the ve supply port The flow rate of the vac
4. Task Selecting Task enables a set of pre determined functions and software enabled optional functions Units Select the new units from the list of pressure measurement units Special units can also be defined see supervisor set up Global set up See paragraph 6 6 Set up zero Mode off auto timed Interval 00 00 00 Isolation status isolated non isolated K0443 Issue No 3 6 9 6 Reference and Specification 6 5 Control Set up Vent Select Vent to reduce the system pressure to near atmospheric pressure Use this feature to reduce system pressure to a safe value before disconnecting the Unit Under Test Use vent set up to adjust the slew rate of venting Note The vent key can be selected in the control set up menu or programmed as an on screen selection in the status area from the global set up display status area menu Nudge Sets the incremental resolution of the nudge control for trimming the set point digits Set point Limits Defines the limits of pressure that can be entered as a set point useful for protecting sensitive UUT Slew rate Sets how the controller achieves a set point Linear Controller changing pressure linearly to set point at a rate set by user Note This function should be used for UUT that have hysteresis errors Rate Either maximum rate or at a rate value set by user Overshoot Fast changing pressure as fast as possible may go beyond set point overshoot No overshoot changing pre
5. To connect the power supply to the instrument proceed as follows e Insert the IEC power connector 1 into the power supply connector at the rear of the instrument e Set the ON OFF switch to ON e Check that the front panel display shows the power up sequence K0443 Issue No 3 2 8 PACE Pressure Controller User Manual 1 IEC power connector 2 Fuse carrier 5 Fuse Figure 2 6 Electrical Connections Pressure Control Module Input and Output Connectors 24V DC Output O 100mA maximum Using a 4 way connector pin 24 Vdc pin O Vdc This facility can energise external equipment An integral self resetting fuse protects this output Logic switch Input Using a 4 way connector Input Output This facility can be used to trigger the instrument from a pressure switch contact during the Pressure Switch Task see Section 3 4 Connections are not polarised and can be connected either way Integral opto isolators protect this input circuit K0443 Issue No 3 2 9 2 Installation Communication Connections Connect the appropriate connectors into the rear panel communications ports and if appropriate secure with the captive screws Note The RS232 and IEEE 488 interfaces are both enabled at power up Set the required parameters in Supervisor Setup communications menu see Section 3 8 SUPPLYA OUTPU AVENT REF A A sl SE 8 eg E p CAL LABEL mos av NOMINAL CAL LABEL A N a 100mA MAX Logic A INPUT
6. Touch screen areas 4 The instrument is now ready for use K0443 Issue No 3 3 2 PACE5000 Model serial no Module serial no xxxxxxx Module Full Scale xxxxxxx PACE Pressure Controller User Manual 3 3 Measure mode PACE5000 aa Measured Pressure mbarg ES 198 03 7 ve Full Scale ve Full Scale 1000 00 3500 00 Control sl w l 0 00 S _ 789 346 346 5llelalelsl PACE6000 Le 7 la lo lo lM 70 0000 bar g FS 1 Measured Pressure bar g 0 0009 ZS ve Full Scale 70 a J ve ve Full Scale 1 0000 Scale 1 0000 Control 73 por 5 NIESEN Le z Le lo loll 1 Pressure measurement of current selected sensor in current selected pressure measurement units 2 Current enabled functions 3 Control measure selection 4 Nudge up changed in control set up 5 Current set point value change with numeric keys 6 Nudge down changed in control set up 7 Status area changed in global set up Display Icons gt lt Auto zero Control mode with overshoot Percentage KA Reference level difference Control mode with no y E S Bo Timed zero A gas head correction overshoot A Al Control mode active Filter pressure reading tare Tare enabled Y Control mode passive Linear rate 0 Zero 4 Maximum rate GC Control mode gauge Sil Mr K0443 Issue No 3 3 3 3 Operation Measure set up e ES E current pressure reading to ze
7. command 2 Always use the vent function before disconnecting pressure equipment from the outlet port g Press the ese key to return to measure mode The screen display changes as follows The current pressure reading changes from blue green to black If enabled the effort meter indicates the controller at rest K0443 Issue No 3 3 10 PACE Pressure Controller User Manual Divider Select and set up the divider task by pressing Divider from the task screen The divider menu defines high set point low set point and then divides the span into a number of equal test points min 2 max 25 Controlled Pressure mbar g Divider menu structure 789 473 Divider or Select required units Rate etc in A y y ai the set up menus When Divider is Divider Setpoint 6 20 S Lel San gt then entered from the Task menu a these test point pressures and Divider H Divide as set up press contra tO start number of test points can be set By entering control mode enables the Divider sequence of test pressures and controlled at the selected rate Example High set point 2 bar Low set point O bar Number of points 5 Test pressures 0 0 5 1 1 5 and 2 bar D ivid e r pa ra m ete rs Divider Parameters High Setpoint Get 3 00000 A High Setpoint 3 00000 bar High Set point pam int eae Low set point d st ca 8 Number of points de a 0001 90 e Y ve Full Scale Lo 3
8. dl Sal 5 2 3 5 6 1 6 1 6 4 6 6 6 9 6 10 6 11 6 12 6 15 6 8 Calibration stas 6 9 Communications Instrument EN ara 6 10 EEN 6 11 Return Goods Material Procedure AS 6 12 Beie oleo A K0443 Issue No 3 vil 6 28 6 29 6 29 6 29 6 30 intentionally left blank K0443 Issue No 3 viii PACE Pressure Controller User Manual 1 1 1 Description Introduction The PACE5000 single channel and PACE6000 single dual channel Pressure Automated Calibration Equipment measures and controls pneumatic pressures and displays on a touch screen the pressure measurement and controller status The touch screen enables selections and settings in both measure and control modes The instrument can be operated remotely through communication interfaces Figure 1 1 PACE5000 General view Figure 1 2 PACE6000 General view The rear of the instrument houses all the electrical and pneumatic output and input connections The electrical connections provide an a c power supply serial and parallel communication interfaces d c output and logic input and output The system pneumatic controller module contains a positive and negative pressure supply port an output port vent port and reference port K0443 Issue No 3 1 1 1 Description The instrument can be mounted in a standard 19 inch rack system rack mount option CAL LABEL de PART f IWER 1007A MAX P F 4 PUN eov Hz Abed E FU 260 80 80
9. 2 PACE Pressure Controller User Manual 5 4 Approved Service Agents For the list of service centres visit our web site www gesensinginspection com K0443 Issue No 3 5 Testing and Fault Finding intentionally left blank K0443 Issue No 3 5 4 PACE Pressure Controller User Manual 6 Reference and Specification 6 1 Installation notes The PACE instrument pressure controller calibrator requires an independent pressure supply and set of connections with the exception of the reference connection this provides a reference to atmosphere for gauge sensors and barometric sensors The instrument must have the correct supply pressure and a suitable supply medium see data sheet specification The supply gas density and type does not affect the accuracy of pressure measurement assuming that the UUT is at the same level height as the controller or gas head correction is accurately set Gas supply For normal operation the instrument requires a positive supply of at least 110 of range but less than the MWP with a gas regulator For absolute operation negative gauge operation or if the installation requires a fast response around atmospheric pressure a vacuum source must be connected to the negative supply recommended configurations shown in figures 2 3 2 4 and 2 5 To achieve control performance the source pressure must be maintained at 10 full scale above the required set point Controller performance is maintained during
10. 24 V d c maximum 30 V d c wrt chassis 1A resistive maximum Volts free schematic K0443 Issue No 3 6 27 6 Reference and Specification 6 9 Calibration The calibration menu provides facilities for programming settings for maintenance as follows Note A PIN protects the Calibration menu against unauthorised use Each instrument on delivery contains the factory set PIN 4321 To continue protecting the supervisor set up menu the PIN should be changed as soon as possible ve source zero ve source zero sensor correction e Selects the range for a three point calibration routine valve correction source PDCR correction three point calibration of both source pressure sensors screen calibration Time amp Date e Sets instrument clock and calendar Change PIN e Changes the Calibration PIN Enter the existing PIN the new DIN and confirmation of the new PIN Should this new PIN become lost it can only be reset by returning the instrument to a GE service centre Option enable process Use the following process to enable the test program option on a PACE instrument Example e Touch the top Measure area of the screen Select Global Setup Serial e Select Calibration nUmMBer o Enter a Calibration PIN of 1234 Enter new option key xxxxxxxxxx 10 digits After entry of this key PACE confirms the options have been enabled K0443 Issue No 3 6 28 PACE Pressure Controller User Manual 6 10 Commu
11. 30 V maximum Start Controls pressure at a fast slew rate to Setpoint 0 100000 Se A Setpoint 0 00 mbar Stow Rate 100 00 mbertes Start Duo Time 00 00 0 ee 50ms Setpoint 0 100000 bar Slew Rate 1 000000 barfsecon Test Setpoint 1 0 00 mbar Dwell Time 00 00 00 Test Setpoint 2 0 00 mbar Switch de bounce time 50 ms Slew Rate 100 00 mbar second Test Bs d Time 00 00 00 ni 1 CN a set point just below expected switch End Be 9 mn operating point De bounce time can Ad be d for sl test sl tes or slower test slew rates 67 819 lo H Controls pressure between two set points and if required can repeat loop n times and auto reverse can be selected End Controls pressure to a safe condition to disconnect switch under test Procedure Select the switch test parameters in the switch test menu including the start pressure end pressure and test rate of change Slower rates give more accurate results The de bounce time can be set between 0 and 200 ms After the test the display shows the pressures at which the contacts open and close and the switch hysteresis the difference between the two switching pressures Before disconnecting the switch under test press Release pressure to release any residual pressure Pressure module TORN Outlet Vent O O O Logic input DC Output 24V max 24V OV Note This switch test procedure can be repeated to exercis
12. 500000 E I I I I I I I I I I I I I I I I I I I I I I I I 2 l l l l l l l l l l l l l l l l l l l l l l l l l 1 l l I I I EE EA I I 2 3 4 5 N P pressureN Number of pointsR range between low set point 1 and high set point 5 controlled pressure to each set point K0443 Issue No 3 3 11 3 Operation Preset The Preset function is similar to the Divider function except that using this menu individual set point values can be defined for each of 25 set points _1501 027 The set up function displays Preset a preset number pressing the o a y a soft key for that number lt SS e assigns a pressure value to the key After setting all the 25 preset pressures enter control mode and then press a soft key to change to the pressure assigned to that key Preset as set u p Preset Pressure 1 GENEE SS 1121314 15 in izle lolol Preset Pressure 1 0 000000 b 0 Value _ Preset and controlled at the E En 3 selected rate lA 1501 027 Ft SSeS Vin mm 250 150 SE 50 0 A 50 0 P P pressure T time K0443 Issue No 3 3 12 PACE Pressure Controller User Manual 3 6 Global Set up Selections Global set up selections provide access to the instrume
13. SETZ0A SUPPLYA CAL LABEL O PART ATIC CONTROL MODULE el OUTPUT AVENT REF i DC OUTPUT o p ipaa O ONTROL MODULE O z DRUCK LE6 OFH UK Figure 1 3 PACE5000 Rear view AVENT REF DC OUTPUT EE IDOS 24V NOMINAL CAL LABEL 100MA MAX INPUT AN PART DAUCK LES OFH UK OPTION 2 OPTION 1 Figure 1 4 PACE6000 Rear view N CONSULT HANDBOOK OPTION 2 ETHERNET A VENT REF DRUCK LE6 OFH UK ETHERNET uss i ATA USB A Options available for the PACE5000 and PACE6000 are detailed in datasheet SDSO001 PACE5000 and SDSO0008 PACE6000 Further information and notes on applications are available in section 6 of this manual and on the GE Sensing amp Inspection web site at www gesensinginspection com K0443 Issue No 3 PACE Pressure Controller User Manual 2 Installation 2 1 Packaging On receipt of the PACE5000 or PACE6000 check the contents of the packaging against the following list Packaging List i PACE5000 or PACE6000 Pressure Controller li Cable power supply iii User guide and CD UD 0001 containing the full documentation suite iv Pneumatic Control Module blanking plate keep this plate for future use CAUTIONS 1 AFTER REMOVING A CONTROL MODULE USE A BLANKING PLATE TO KEEP THE FLOW OF COOLING AIR 2 AFTER UNPACKING AN INSTRUMENT THAT HAS BEEN IN COLD CONDITIONS ALLOW TIME TO STABILISE AND ANY CONDENSATION TO EVAPORATE 2 2 Packaging for Storage or
14. T E 20 0 DRUCK LES OFH UK OPTION 2 OPTION 1 e de Figure 2 7 Communication Connectors 1 RS232 2 Canbus option 3 IEEE488 4 USB B 5 USB A 6 Ethernet option RS232 Interface When using the RS232 interface a cable must be connected directly from the instrument to a suitable port on the computer in a point to point link The pin connections for the 9 pin D type RS232 connector and the relationship between the instrument and the RS232 control signals together with device interconnection interface is shown in Table 2 1 The instrument is configured as Data Circuit Terminating Equipment DCE K0443 Issue No 3 2 10 PACE Pressure Controller User Manual Instrument Control Line Computer Printer Instrument Connector Signal RS232 Connector Type Function 9 way Direction Terminology D type Pin No TxD O P pom 2 foe gt ICAA ICI Leni mell ee EEN O A EE Gett eeh Pulled RLSD high DCD internally Not connected Pulled l DSR AE DCE Ready internally Equipment Connector Cable AAA Table 2 1 RS232 Connections Handshaking connections Software handshaking use TXD RXD and GND Hardware handshaking use TXD RXD GND CTS RTS and DTR K0443 Issue No 3 2 11 2 Installation IEEE 488 Interface This interface complies with IEEE 488 standard The IEEE 488 parallel interface connects a computer controller to one or more PACE instrument and possibly other ins
15. details of other system components t High pressure gas exhaust depending on pressure range Optimum controller transient response and minimum time to set point may be degraded if either the pneumatic supply or vacuum system has restricted flow Installing a reservoir volume which has larger capacity than the load volume located in close proximity to the controller supply ports can improve the controller response Optional vacuum system kit To protect the control module for ranges above 100 bar from over pressure a suitable protection device such as a relief valve or bursting disc must be fitted to limit the applied supply pressure to below the MWP Optional differential connection kit K0443 Issue No 3 2 5 2 Installation SUPPLY OUTLET VENT REF E le lt gt SE EN 12 Figure 2 4 Pneumatic Connections with negative gauge pressure generator Key to Figure 2 4 1 4 5 8 10 12 Pressure source d Conditioner 3 Filter Regulate to between 110 full scale and MWP Diffuser 6 Unit under test 7 Optional reservoir T Protection device 9 Optional differential connection a atmosphere Vacuum generator f 11 Source pressure regulated compressed air supply Exhaust to atmosphere 13 Check valve f Notes Refer to Section 6 Reference and Specification for details of other system components t High pressure gas exhaust depending on pressure range Optimum controller transi
16. edit an existing program press Edit Enter a name for the program using the text editor keys to select letter range to select the letter within the range Use the right arrow key to move to the next character position o Press enter on completion e The display shows the line instruction with insert and delete selections Pressing insert changes the display to a list of the available programming commands Table 3 1 Test Program Commands BEEP Beep on off RATE_VALUE Specifies controller rate Breaks to this point when stop count BREAK command or STOP icon is selected RESOLUTION Sets display resolution then executing code to the end CONTROL Selects Control mode SETPOINT Allows set point to be entered COUNT ES EECHER SETTLING_FAST Used to specify overshoot requirements DWELL Specifies dwell time seconds SETTLING NO Used to specify no overshoot GOTO Used to set up a loop Enter program STOP COUNT Used to stop loop program after a number of line number to go to loops IN_LIMITS in_Limits band setting full scale TEXT Waits within In_Limits for this time IN_LIMITS_TIME period before setting a valid In_Limits UNITS Selects required display units condition IP_LOGIC sce acral tic se ste EES VENT Instructs instrument to vent contacts as a halt condition MEASURE Selects Measure mode VFC OFF Sets VFC allocation test program off PAUSE ue EES VFC ON Sets VFC allocation test program on input Resume RAN
17. slow variations in source pressure down to a source pressure of 20 full scale range Supply conditioning equipment Supplies should be provided with an isolation valve and any other necessary conditioning equipment Important note To protect the control module for ranges above 100 bar from over pressure a suitable protection device such PA as a relief valve or bursting disc must be fitted to limit Maximum Working Pressure Exceeded E the applied supply pressure to below the MWP Reduce ve Source Pressure within M I AN MWP Limits then select OK the instrument may indicate this condition by Jl OK uo displaying this message Maximum Working Pressure 0 00 a If the measured ve source pressure exceeds the limits below then display a persistent message box as shown Clear this message by selecting OK Measured Pressure mbarg Should the applied supply pressure exceed the MWP K0443 Issue No 3 6 1 6 Reference and Specification Supply contamination Some supplies may need the removal of water oil or particulate contamination Any water in the compressed gas supply will be in vapour form Le non condensing and must be removed using a mist filter Oil must be completely removed as this causes a rapid deterioration of the control valve performance The compressed gas supply must not contain particulates and must be removed using a particulate filter Do not use a compressed gas supply containing corrosive material
18. the manufacturer or Service Agent K0443 Issue No 3 4 2 PACE Pressure Controller User Manual 1 IEC connector 2 Fuse carrier 3 Fuse Figure 4 1 Power Supply Fuse Replacement K0443 Issue No 3 4 3 4 Maintenance 4 8 Filter replacement Fig 4 2 When necessary change the filters in the pressure module Note To gain access to rack mounted instruments it will be necessary to completely withdraw the instrument from the rack Procedure e Switch off instrument Depressurize the system and isolate the pneumatic supplies Switch off and then disconnect the electrical power supply Disconnect the pneumatic pipes to the pressure module Unscrew the four captive Phillips cross head screws securing the pressure module in the instrument case Remove the pressure module to gain access to the filters Using the 5 mm hexagonal key unscrew the filter retainer 1 in the bore of each of the pressure connection Remove the five filters 2 if necessary invert the pressure module so that the filters fall out FULL SCALE a PNEUMATIC CONTROL MODULE gie Wbdo dE LEILA SEO PART O 2 4 1 Retainer 2 Filter Figure 4 2 Pressure module filters Insert a new filter in the bore of each of the pressure connection Using the 5 mm hexagonal key screw in each filter retainer hand tight Locate the pressure module in the instrument case and secure with the four captive Phillips
19. GE Sensing amp Inspection Technologies Pressure Automated Calibration Equipment User manual K0443 PACES5000 PACE6000 The General Electric Company All rights reserved Introduction This technical manual provides operating instructions for the PACE Pneumatic Pressure Controllers The features shown and described in this manual may not be available on some models Safety The manufacturer has designed this equipment to be safe when operated using the procedures detailed in this manual Do not use this equipment for any other purpose than that stated This publication contains operating and safety instructions that must be followed to ensure sate operation and to maintain the equipment in a safe condition The safety instructions are either warnings or cautions issued to protect the user and the equipment from injury or damage Use suitably qualified technicians and good engineering practice for all procedures in this publication Pressure Do not apply pressures greater than the maximum working pressure to the equipment Toxic Materials There are no known toxic materials used in construction of this equipment Maintenance The equipment must be maintained using the procedures in this publication Further manufacturer s procedures should be carried out by authorized service agents or the manufacturer s service departments Technical Advice For technical advice contact the manufacturer A qualifie
20. GE Specifies instrument range WAIT_IN_ LIMITS Waits until pressure is within the limits RATE_MAX Sets controller rate to maximum ZERO Output zeroed To select a command use the command on the display and press insert key to write it into the program Place the UNITS RATE SETTLING and RESOLUTION commands at the start of the program this protects pressure sensitive UUT When selected certain commands require a value or selection to be set e g RANGE RATE TEXT the display shows a screen prompt for the appropriate setting K0443 Issue No 3 6 18 PACE Pressure Controller User Manual Example Program Note Changes to instrument settings made in a test program remain valid only for the test program The instrument reverts to the pre test settings on completion Step Command Argument Action START Program start 1 UNITS mbar Select units mbar 2 RATE 100 Select rate 100 mbar min 5 IN LIMITS TIME 10 00 00 10 secs 4 IN LIMITS Set In Limits Band 5 RESOLUTION 5 Display resolution 5 digits 6 SETTLING No overshoot 7 TEXT Operator instruction e g Connect UUT 8 ZERO 9 SET POINT 400 Set point 400 mbar 10 CONTROL Controller ON 11 WAIT IN LIMITS Wait for In Limits Condition 12 BEEP Beep on approx 1 sec Beep off 13 MEASURE Switch to Measure controller off 14 DWELL 30 Wait for 30 sec 00 00 30 15 SET POINT 800 Set point 800 mbar 16 CONTROL Controller on 17 WAIT IN LIMITS Wait for In Limits Condition 18 BEEP ON Beep on appr
21. ISSVHO cl GND ve 8010 91 ZOIC S 9010 Pl an mie SOIC Cl Jul MIC void v old LOld 3719VN3 310038 NJH ZL NOLLN311W NIV LL LSINOJH 30 4438 OS OL YV319 30V4431NI OJI 6 ASILNAGI YO GN3 103 4314320 VIVO ON OYAN VIVO HOS AGVSY LON GSYN GNVA VIVA Ava or W JINV4 SISSVHO AND AO c O D O wnwxew 08 N SSIYOAOQV O 00 SI W a I SN N w L gt gt z SSIYAQV P 88r 3331 M L SSIYUOV Z w gt N E m J J O x 2 Installation intentionally left blank K0443 Issue No 3 2 14 PACE Pressure Controller User Manual 3 3 1 Operation Preparation Make sure the electrical cables and pneumatic pipes comply with the installation requirements in Section 2 Carry out the following before use e If necessary carry out the maintenance task detailed in Section 4 For bench top single instrument operation carry out the following de Make sure the instrument power supply switch on the rear panel is set to OFF 2 Connect the instrument to the electrical supply make sure the supply includes a connection to a protective earth 3 Inspect the pneumatic hoses for damage ingress of dirt and moisture Before use the instrument should be tested This section contains quick reference charts detailing all the available functions The functions shown and described may not be available on some models Review and become familiar with the who
22. K0443 Issue No 3 2 7 2 Installation e Refer to the electrical connections below before finally fitting the instrument into the equipment rack e Temporarily locate and screw in the two spigots to each side of the equipment rack e Locate and slide the instrument into the rack locating the instrument on the spigots e Secure the instrument in the equipment rack with two of the screws and washers supplied e Remove the two spigots and replace with the remaining two screws and washers supplied 2 6 Electrical connections WARNINGS 1 THE GROUND LEAD OF THE INSTRUMENT MUST BE CONNECTED TO THE AC SUPPLY PROTECTIVE SAFETY GROUND 2 ISOLATE THE POWER SUPPLY BEFORE MAKING ANY ELECTRICAL CONNECTIONS TO THE REAR PANEL General The instrument must be connected to the correct electrical power supply as stated on a label next to the power connector Also see section 6 Reference and Specification Make sure that the power supply is off before connecting the power cable Requirements for rack mounted instruments e Install an isolator in the power supply circuit The power supply connector and switch on the rear panel of the instrument will not be accessible e Set the power supply isolator to OFF Connect the power supply and set the power supply switch to ON before sliding the instrument into the rack e Set the power supply isolator to ON Check that the front panel display shows the power up sequence Connecting Figure 2 6
23. Start 1 499547 up Leak rate Li minute IN E Test Pressure 2 Start 0 053995 8 Test Pressure 2 End 0 054107 Leak Rate 2 0 0000056 sec TP1 l l l l l l lk l i l l l l l l Control Measure I l Dwell i Dwell l l l l l l l l l l l l End l l l l l l 0 l l l l l l P l l l l l l l l l l l l l l l l l l l l lk l l l l l I l l TP 21 Control Measure i Dwell Dwell l l l l l l l l l l l l T Ik leak P pressure T time seconds TP test pressure K0443 Issue No 3 6 15 6 Reference and Specification Switch Test option This function automates the testing of pressure switch devices Connect the pressure port of the switch to be tested to the output port Connect the switch contacts in series with the 24V dc output and the Logic Input Note The volt free logic input connections require a switching potential 24V max to be applied If necessary this can be an external d c Switch test Switch test I Switch test Measured Pressure mbar g B ie 1499 42 E Test Setpoint 1 1000 000 Release pressure Stop S e Test Setpoint 2 0 000 Slew Rate 100 000 ep Dwell Time 00 00 20 d OOO Repeat 2 Switch Test Parameter Selection Example of setting a parameter Switch Test Start Parameters Module 1 Switch Test Parameter Selection source Common mode must be kept within
24. Task To control pressure in the task proceed as follows a Select the required units of pressure measurement from the measure set up menu o Press the status area and enter control set up select the required slew rate The display changes to show the type of slew rate selected Select the required vent slew rate in vent set up Caution Use the vent set up to prevent damage to rate sensitive equipment connected to this controller The vent slew rate setting is independent of the controller slew rate settings Return to the task screen In basic task use the numeric keys to enter a set point c Press the a key to start controlling pressure d The screen display changes as follows The current pressure reading changes from black to blue If enabled the effort meter indicates the amount of work done effort by the controller e When the controller achieves the selected pressure set point the screen display changes as follows The current pressure reading changes from blue to green If enabled the effort meter shows the controller effort to keep the pressure at the set point f On completion of testing select control set up and select Vent to reduce the system pressure to near atmospheric pressure This feature should be used to reduce system pressure to a safe value before disconnecting the Unit Under Test Notes L The vent valve opens and remains open until a key press or receipt of a communications
25. Transportation To store the instrument or to return the instrument for calibration or repair carry out the following procedures 1 Pack the instrument as detailed in section 6 8 Reference and Specification 2 Return the instrument for calibration or repair complete the return goods procedure detailed in section 6 8 Reference and Specification S This procedure also applies to the pressure control module as a separate item 2 3 Preparation for Use The instrument can be used as a e Free standing instrument positioned on a horizontal surface e Rack mounted in a standard 19 inch rack using the rack mount accessory option kit refer to section 2 5 For free standing instruments use the two front feet on the base to elevate the instrument to provide a better viewing angle Note The cooling air outlet on the underside of the instrument must not be obstructed Allow a free flow of air around the instrument especially at high ambient temperatures K0443 Issue No 3 2 1 2 Installation 2 4 Pneumatic connections WARNINGS TURN OFF THE SOURCE PRESSURE AND CAREFULLY VENT THE PRESSURE LINES BEFORE DISCONNECTING OR CONNECTING THE PRESSURE LINES PROCEED WITH CARE ONLY USE EQUIPMENT WITH THE CORRECT PRESSURE RATING BEFORE APPLYING PRESSURE EXAMINE ALL FITTINGS AND EQUIPMENT FOR DAMAGE REPLACE ALL DAMAGED FITTINGS AND EQUIPMENT DO NOT USE ANY DAMAGED FITTINGS AND EQUIPMENT Connection adaptor Input supply G 1 8 suppl
26. When connecting the output ports of two pressure modules together make sure both are either e below 70 bar OR e between 100 to 210 bar To comply with the Pressure Equipment Directive PED do not mix these categories Supply equipment Pneumatic supplies should have isolation valves and where necessary conditioning equipment The positive pressure supply should be regulated to between 110 of the full scale pressure range and MWP stated on the control module To protect the control module for ranges above 100 bar from over pressure a suitable protection device such as a relief valve or bursting disc must be fitted to limit the applied supply pressure to below the MWP On instruments without a negative supply the positive pressure discharges from the system to atmosphere through the negative supply port Fit the diffuser to the negative port to diffuse airflow During system pressure vent operations the pressure discharges from the system to atmosphere through the vent port Fit a diffuser to the vent port to diffuse airflow K0443 Issue No 3 2 3 2 Installation Pneumatic Connection Examples Figures 2 2 and 2 4 These examples show a single channel connection detail using supply equipment described above SUPPLY OUTLET VENT REF E Q OL 2 O Figure 2 2 Pneumatic Connections without vacuum supply Key to Figure 2 2 1 Pressure source 2 Conditioner E Filter 4 Regulate to between 110 full scale
27. ability or precision The instrument uses the in limits flag when performing a control task such as Leak Test or Switch Test Note In remote control the control computer can be used to interrogate the in limits register to confirm the controller has achieved set point Alarms One or two pressure alarms can be set An alarm triggers when the pressure exceeds the high alarm or falls below the low alarm A buzzer sounds when the alarm triggers and the alarm symbol bell appears on the display Comms Selects a communication port and parameters for simultaneous operation of both the RS232 and the IEEE 488 interfaces The user can select appropriate settings for communicating with the control computer PC and the required command protocol Refer to publication KO472 SCPI Remote Communications Manual or K0469 Heritage Communications Manual RS232 Located on the rear panel an external RS232 connection requires Connector 9 way D female wired as per Table 2 1 Communications RS232 point to point only DPI 520 daisy chain is not supported Baud Rate power up default 9600 no parity amp handshake xon xoff Baud rates selectable 2400 4800 9600 19k2 38k4 57k6 115k2 Parity None Odd amp Even Flow control None Hardware amp xon xoff Protocols SCPI DPI 520 DPI 500 New data up date rate refer to data sheet Selectable through the user interface Built in EMC Filtering and Transient voltage protect
28. al option The aeronautical option is a specialised application of the PACE instrument Special application note The PACE instrument must be set up very carefully so that the aeronautical pressures applied do not exceed maximum pressure values and rates of change Leak testing Cautions 1 Do not exceed the maximum pressures stated in the appropriate Component Maintenance Manual for the unit under test 2 Carefully de pressurize all pipes to atmospheric pressure before disconnecting and connecting to the unit under test Before testing an aeronautical component carry out a leak test This task sets the test pressure dwell time at the test pressure and the leak test time At the start of the test the instrument applies a test pressure to the user system A dwell time allows the user system to stabilise Aeronautical testing The aeronautical task enables control and measurement of altitude and airspeed in aeronautical units such as feet and metres altitude and knots mph km h airspeed This task utilises dual pressure displays to show the parameter and the rate of change of Altitude Airspeed Mach and Airspeed with Mach number The aeronautical task enables the testing and calibration checking of aeronautical indicators and system components by controlling and displaying values and rates in aeronautical units When using a single instrument the pressure supply must be changed when changing from Altitude to Airspeed Modu
29. and MWP 5 Diffuser 6 Unit under test 7 Optional reservoir T 8 Protection device 9 Optional differential connection q atmosphere Notes Refer to Section 6 Reference and Specification for details of other system components i High pressure gas exhaust depending on pressure range t Optimum controller transient response and minimum time to set point may be degraded if either the pneumatic supply or vacuum system has restricted flow Installing a reservoir volume which has larger capacity than the load volume located in close proximity to the controller supply ports can improve the controller response To protect the control module for ranges above 100 bar from over pressure a suitable protection device such as a relief valve or bursting disc must be fitted to limit the applied supply pressure to below the MWP Optional differential connection kit K0443 Issue No 3 2 4 PACE Pressure Controller User Manual SUPPLY OUTLET VENT REF Ei E TO OO O Q Figure 2 3 Pneumatic Connections with vacuum supply Key to Figure 2 3 1 4 5 8 10 13 Pressure source 2 Conditioner 3 Filter Regulate to between 110 full scale and MWP Diffuser 6 Unit under test 7 Optional reservoir t Protection device 9 Optional differential connection Oil mist trap 11 Vacuum source 12 Normally open electrical release valve Check valve q atmosphere Notes Refer to Section 6 Reference and Specification for
30. cross head screws K0443 Issue No 3 4 4 PACE Pressure Controller User Manual 4 9 Pressure module replacement Fig 4 3 WARNING TURN OFF THE SOURCE PRESSURE AND VENT THE PRESSURE LINES BEFORE DISCONNECTING OR CONNECTING THE PRESSURE LINES PROCEED WITH CARE example instrument 1 Pressure module 2 Instrumentcase 3 Blanking plate Figure 4 3 Pressure module Procedure Note For rack mounted units isolate the electrical power and pneumatic supplies Partially withdraw the unit from the rack to gain access to the electrical and pneumatic connections Removal e Switch off instrument Depressurize the system and isolate the pneumatic supplies Switch off and then disconnect the electrical power supply Disconnect the pneumatic pipes to the pressure module Unscrew the four captive Phillips cross head screws securing the pressure module 1 in the instrument case 2 Note If necessary temporarily fit a blanking plate supplied to protect the internal components from ingress of foreign matter Fitting e Fit a fully compatible pressure module 1 into the instrument case 2 Note See below for additional fitting instructions for the PACE 6000 Secure the pressure module 1 with the four captive Phillips cross head screws Reconnect the pneumatic supplies to the instrument use appropriate sealing for all pneumatic connections K0443 Issue No 3 4 5 4 Maintenance Note PACE 6000 Caut
31. d technician must have the necessary technical knowledge documentation special test equipment and tools to carry out the required work on this equipment K0443 Issue No 3 i Abbreviations The following abbreviations are used in this manual the abbreviations are the same in the singular and plural abs a C ALT BSP CAS CSK d c DPI e g etc Fig ft g Hg HTS Hz IAS Le IEC IEEE 488 in kg kts kn LCD m mA max mbar min mm mV MWP No NPT PACE Para PDCR PED psi PTX K0443 Issue No 3 Absolute Alternating current Altitude British pipe thread Calculated airspeed Countersunk Direct current Digital Pressure Instrument For example And so on Figure Foot Gauge Mercury High tensile steel Hertz Indicated airspeed That is International Electrotechnical Commission Institute of Electrical and Electronic Engineers standard 488 data Inch Kilogram knot Liquid crystal display Metre Milliampere Maximum Millibar Minute or minimum Millimetre Millivolts Maximum working pressure Number National Pipe Thread Pressure Automated Calibration Equipment Paragraph Pressure transducer Pressure Equipment Directive Pounds per square inch Pressue transmitter ROC RS232 SCPI UUT V VFC Ve Ve G Rate of Climb Serial communications data standard Standard Commands for Programmable Instruments Unit under test Volts Volts free contact Positive Negative Degrees C
32. e the switch unit Logic input is opto isolated An external supply 5V to 24V can be used provided its common mode voltage remains with 30 Vmax with respect to chassis A Pressure Switch TITITTTTT Example Switch Test Connections K0443 Issue No 3 6 16 PACE Pressure Controller User Manual Test Program option The test program task provides a facility for writing and executing test procedures Selecting a test program from the Task menu displays all the task programs currently stored together with the ability to write new ones Test program Starts J 0 999770 Bun f test program 0 999770 z ve Full Scale Sto 3 500000 z Ae Select a test program single step from list Test Programs La le Lo Lo Le lele 8 My Prog AL Filename On a e le la le Lt lo y ran yal Lo Lo Lola lete le Ela ls ll lala le ala la a a A Edit K0443 Issue No 3 6 17 6 Reference and Specification Program To start a test program select the test programs listed on the screen the step function moves down the list of available test programs To start a test program press run When the program starts a stop legend replaces the run legend Press the stop key at any time to stop the test program To write a test program press Set up and select New Note A new test always contains a START and END command first and last line instructions these lines cannot be deleted To
33. e lower range then the higher range controller will control to this set point When the controller is in limits then the range is changed to the lower ranged controller and this then controls to the entered set point Module 2 Module 1 SUPPLY OUTLET VENT REF SUPPLY OUTLET VENT REF rear view 09 9990 0 d O O OI ofPACE6000 instrument Pressure output Pressure regulator Pressure source Pressure regulator K0443 Issue No 3 3 5 3 Operation 3 4 Control Mode In measure mode press opd and the instrument changes to control mode Press Term and the instrument stops controlling pressure and changes to measure mode Controlled Pressure mbar g 499 ous 7 SC Zeg Controlled Pressure mbarg QQ A 2 1 1499 995 x lt cor y ee ku 1500 000 _ Key to display 1 Current measured pressure value in limits green out of limits blue 2 In active control mode Pressure reading filter ON Head pressure value applied Control mode with overshoot Tare enabled Press to switch between controlled pressure and measured pressure Nudge up Set point press and the display changes to numeric keys Nudge down 5 4 5 6 7 Notes a Active mode control active except in measure mode Status area shows effort meter set in global set up press to enter control set up Passive mode when the controller achieves inlimits condition measure mode is automatically selected Gau
34. elsius Associated publications Symbols K0447 PACE 5000 6000 User Guide and Safety Instructions K0450 PACE Series Calibration Manual K0476 Pressure Control Module User Guide and Safety Instructions K0472 Remote Communications Manual K0469 Heritage Communications Manual Instrument Emulation The equipment contains the following symbols to identify hazards CE A 5 This equipment meets the requirements of all relevant European safety directives The equipment carries the CE mark This symbol on the instrument indicates that the user should refer to the user manual This symbol on the instrument indicates do not throw away in domestic bin hazardous material dispose correctly in accordance with local regulations K0443 Issue No 3 iii Pressure units and conversion factors Pressure units Factor hPa Pressure units Factor hPa mbar 1 0 CH OO 20 C 0 978903642 bar 1000 0 mH 0 O 20 C 97 8903642 Pa N m 0 01 0 0980665 kPa 10 0 1 333223684 MPa atm 1013 25 mmHg 0 C psi 68 94757293 cmHg 0 C 0 4788025898 mHg 0 C 2 4908891 inHg 0 C 2 486413 mmH gt 0 O 4 C 2 487641558 CMH 70 O 4 C 0 980665 ftH20 O 4 C 29 8906692 MH 50 O 4 C 98 0665 ftH20 20 C 29 836983 mmH gt 20 O 20 C 0 097890364 ftH20 O 60 F 29 8516987 Unit Conversion To convert FROM pressure VALUE 1 in pressure UNITS 1 TO pressure VALUE 2 in pressure UNITS 2 calculate as follows VALUE 2 VALUE 1 x FACTOR 1 FACTOR 2 Note
35. en in aeronautical option K0443 Issue No 3 3 8 PACE Pressure Controller User Manual 3 5 Operation and Example Procedures Introduction e Before operation the instrument must be connected to the correct electrical and pneumatic supplies as detailed in Section 2 Installation Switch the instrument ON and after a short time the display shows measured pressure mode except when regulator mode is selected and to the task set before the power off Measure and Control Modes The instrument operates in two modes Measure or Control In Measure mode the instrument works as a precision pressure indicator and shows the pressure measured at the output port In Control mode the instrument works as a precision pressure controller and shows the controlled pressure measured at the output port Pressing Task enables various pre determined functions Task T Enable Stop Set up Task W Leak Test Switch Test Start program T Test program b nin ul D Al S Gl u p Test Program Preset Step L Select program SCH Enable Divider Po Set up Enable Preset Set up select a preset and store new values Selecting a task Option refer Press exits the menu and to 6 8 changes the display to exit to the task selected The display shows the task screen When selected e g Basic the screen changes to show the selected task K0443 Issue No 3 3 9 3 Operation
36. ent response and minimum time to set point may be degraded if either the pneumatic supply or vacuum system has restricted flow Installing a reservoir volume which has larger capacity than the load volume located in close proximity to the controller supply ports can improve the controller response Optional negative gauge generator kit To protect the control module for ranges above 100 bar from over pressure a suitable protection device such as a relief valve or bursting disc must be fitted to limit the applied supply pressure to below the MWP Optional differential connection kit K0443 Issue No 3 2 6 PACE Pressure Controller User Manual 2 5 Rack mount option Figure 2 5 General There must be enough space at the rear of the instrument for all the cables and pipes The length of the cables and pipes must allow for the removal and fitment of the instrument The cooling air of the instrument must not be obstructed Allow a free flow of air through the equipment rack and around the instrument especially at high ambient temperatures Example instrument Figure 2 5 Rack mounting Procedure e Locate and secure the back brackets to the rear of the equipment rack e Unscrew and remove the four countersunk screws from each of the instrument side panels e Locate the two brackets on each side of the instrument and secure with the four countersunk screws e Support the instrument and connect the cables and pipes
37. ent value in Pascals Supervisor Restore Last Settings Un restore previous defaults Y N K0443 Issue No 3 3 Operation 3 9 Instrument Status The control set up menu provides access to the status of the instrument DREES Pressing the escape key stores Gam settings and returns to status aga in measure or control set up v BL Instrument Status gau a CACICA Instrument ME SS LA HU Instrument Main code Instrument OS Build Software build Instrument Boot ROM Module 1 Main Code Module 1 Boot ROM Module 2 Main Code Module 2 Boot ROM Analogue O P Main Code Hardware build Additions to the standard instrument Calibration Analogue O P Boot Code History Valve correction tele Zero VFC Boot Code Software Message c icati IEEE 488 current settings see 6 7 Communications RS232 current settings see 6 7 Current Set up Zero In Limits Current set up High Alarm Low Alarm l l Slew Rate U Operation time days hours minutes Power Up se log control time days hours minutes PACE5000 Instrument Status More next page y Instrument Status rc o a K0443 Issue No 3 3 16 PACE Pressure Controller User Manual Software Software history in the status menu provides read only information on the current software in
38. es from ICAO Standard Atmosphere 1964 Reference Pressure Select the required reference pressure this can be either the barometric pressure from the instrument s internal barometric sensor or any numeric value e g 1013 25 mbar Go to ground Returns the instrument and any unit under test UUT connected to it safely to ground pressure at a controlled timed rate K0443 Issue No 3 6 23 6 Reference and Specification In the aeronautical mode the display goes to the last selected parameter altitude airspeed or Mach Altitude K0443 Issue No 3 6 24 PACE Pressure Controller User Manual Airspeed Mach K0443 Issue No 3 6 25 6 Reference and Specification Analogue Output Option The analogue output option provides a selectable output of voltage or current 0 to 10V Off Selects Analogue O P range 30V maximum with On off Condition Of respect to chassis Update rate of Analogue O P option from the control module Update Rate 10 readings sec Analogue O P bandwidth 0 5 x update rate Hz K0443 Issue No 3 6 26 PACE Pressure Controller User Manual Volts free Contact Option The Volts free Contact option provides a selectable output of voltage or current Contact 1 Not Set Each selection has three VFC ox Volts free connections Selection of event trigger trigger V conditions Selection of trigger Divider conditions OV sv
39. fitted as standard Additional communication types are options TBA Current set up read only data Fault history read only data Software history read only data Use log read only data Vent set up Use the vent set up to prevent damage to rate sensitive equipment connected to this controller The vent slew rate setting is independent of the controller slew rate settings 6 6 Global set up Supervisor set up PIN protected menu Calibration PIN protected menu Save recall user set up Save user set up Recall user set up Display Resolution Backlight Audio volume Status area K0443 Issue No 3 6 11 6 Reference and Specification 6 7 Supervisor Set up The Supervisor menu provides facilities for programming settings These are usually made during installation as follows Important Note A PIN protects the Supervisor menu against unauthorised use Each instrument on delivery contains the factory set PIN 0268 To continue protecting the Supervisor set up menu the PIN should be changed as soon as possible Protective Vent The protective vent can be enabled or disabled and causes the discharge of pressure at a controlled rate if the measured pressure exceeds 110 full scale This protects the pressure sensor from over range In Limits A tolerance value can be set at the set point When the controller achieves the set point the instrument controls within this set tolerance value It does not affect controller st
40. g pressure leaking Apply control valve In measure mode with output port sealed the Decreasing pressure leaking Release control valve pressure continues to increase or decrease Confirm by isolating supplies Contact approved service agent Display pressure reading in red Over range use vent de pressurize O OS Idle time out enabled but timeout period setting too short at a new pressure Blocked vent port Instrument will not zero does not achieve set point Check for blockage Contact approved service agent for repair Instrument controlling to set point no pneumatic Blocked isolation valve output Contact approved service agent for repair Leaking isolation valve Erratic or inaccurate zero Reference port restrictor not fitted Contact approved service agent for repair System internal leak Carry out leak test at full scale pressure Contact approved service agent for repair Reference port restrictor not fitted Increased gas consumption Unstable control at set point or does not achieve set point Table 5 1 Fault Location Note When the controlled pressure stays within tolerance band the pressure at the outlet is within the limits set in Set up Supervisor In Limits If the controller status indicator is outside the tolerance band then this could indicate either a leak in the system or that the supply pressure differs from the pressure for which the contro valves have been characterised K0443 Issue No 3 5
41. ge mode when the controller achieves zero gauge inlimits condition measure mode is automatically selected and the zero valve opens K0443 Issue No 3 3 6 PACE Pressure Controller User Manual Controlling to a new set point To change the set point value touch the set point area of the screen and the display shows the numeric keys Set the new set point value Setpoint mbar A GAR 789 346 4 2 3 4 5 L 6 7 L8 Lo Lo IH If necessary use the key to remove the last digit in the set point value display To save the new set point touch the set point area of the screen This returns the display to the measured pressure screen and showing the new set point Touch the escape Jl to leave the numeric setting unchanged To control pressure to the new set point value press the opd key The display shows the pressure value changing as the instrument controls to the new set point at the set rate of change When changing from measured to controlled mode the displayed pressure digits change from black measured pressure to blue controlled pressure out of limits to green controlled pressure in limits If enabled the effort meter shows the effort the controller exerts to achieve the set point The status area can be changed to various displays showing pressure and controller performance Effort T v Y Effort meter Note In normal controlled pressure conditions the effort meter stays w
42. ion K0443 Issue No 3 6 12 PACE Pressure Controller User Manual IEEE 488 Located on the rear panel an external IEEE 488 connection requires Connector 24 way D female wired as IEEE 488 standard Communications IEEE 488 GPIB Default Address 16 Protocols SCPI DPI 520 DPI 500 New data up date rate refer to data sheet Built in EMC filtering and transient voltage protection Timeout Presets the times for automatically changing from control to measure mode Note Controller timeouts can save supply gas extending contro valve life and minimising acoustic noise Idle Timeout starts when the controller achieves the set point after the set time if new set points are not entered the timer will time out and return to measure mode Gas Head Correction Corrects pressure output for the height difference between instrument and UUT e for UUT positioned higher than the reference level of the PACE instrument enter a positive height REF LEVEI correction for UUT positioned lower than the reference level of the PACE instrument enter a negative height correction When calibrating the PACE instrument disable the gas head correction and correct the actual applied pressures for height K0443 Issue No 3 6 13 6 Reference and Specification Power Up With Normal selected the instrument powers up in Measure mode With power up set point selected the instrument powers up in Control mode at a pre selec
43. ions Reconnect the electrical power supply connector For rack mounted units set the power switch to on locate and secure the instrument in the rack Apply the pneumatic pressure and or vacuum supplies switch on the electrical power supply Switch on the instrument and carry out a serviceability test detailed in section 5 1 AFTER REMOVING A CONTROL MODULE USE A BLANKING PLATE TO KEEP THE FLOW OF COOLING AIR 2 TWO CHANNEL OPERATION COMBINING THE OUTPUTS OF TWO PRESSURE CONTROL MODULES CAN BE DONE WHEN BOTH PRESSURE MODULES ARE BELOW 70 BAR OR BETWEEN 100 TO 210 BAR DAMAGE OCCURS IF ONE PRESSURE MODULE IS BELOW 70 BAR AND THE OTHER IS 100 TO 210 BAR Note When installing two pressure modules the module with the higher pressure range must be located in the right hand side Module 1 above the power supply connector If using two pressure modules of the same pressure range the module with the lower serial number must be located in the right hand side Module 1 above MWP bar FULL SCALE bar the power supply connector PNEUMATIC CONTROL MODULE Ay al e llo Me e tbh Pere ruck i o When connecting the output ports of two pressure modules together make sure both are either e below 70 bar Serial number OR e between 100 to 210 bar To comply with the Pressure Equipment Directive PED do not mix these categories K0443 Issue No 3 PACE Pressure Controller User Manual 5 5 1 5 2 Testing and Fau
44. ithin the band green If the status indicator moves outside the band there may be a leak into or out of the system Controlling to ambient zero pressure Use the numeric key display and set the new set point value to ambient or zero gauge pressure When the display shows the new set point value press the opd key The display shows the pressure value changing as the instrument controls to the new set point at the set rate of change When the display shows ambient or zero pressure press the Te key to switch off the controller and return to measure mode K0443 Issue No 3 3 7 3 Operation Control Setup D wiel ki iS Vent Nudge Setpoint Limits Slew Rate Resolution ai e vent z Ves No v E Ge Li Nudge resolution So Scholar El nudge resolution 1st to 7th digit U Press to exit pper n i Set point limits Lower set up Enable disable Linear Enter units second el Maximum rate Toggle between fast no overshoot Control set up DO Active Continuous control to set point Control mode Passive Controls outside user defined band Gauge Controls from zero gauge to set point Global set up See 3 6 Status see 3 9 Module 1 Vent Setup Vent Rate 0 1000 Vent Rate 0 1000 barfsecond Vent set up A press l to exit set up v D Dress l to exit set up Replaced by GO to Ground wh
45. ld be restricted Using a reference port restrictor short term ambient pressure variations can be prevented from affecting controller performance The controller and UUT references should be connected together using the optional differential connection kit to provide a common reference to atmosphere K0443 Issue No 3 6 5 6 Reference and Specification 6 3 Icons Display Icons in Set up Menus unction unction EC E E ES unction Aero mode Airspeed range Altitude range Ai Y Auto range Audio volume R Ven Change supervisor Canbus PIN Calibration history Communications Contrast Correction analogue output d Control mode Copy Correction SCM Correction sensor i o Es Current set up ET Date amp time Correction source oe Correction valve sensor rH 4 P Diagnostic Canbus Diagnostic a Diagnostic analogue OP barometric option Diagnostic control Diagnostic Diagnostic general sensor controller Diagnostic source sensor oo volt Diagnostics Display Diagnostic vacuum sensor Diagnostic RS232 K0443 Issue No 3 6 6 PACE Pressure Controller User Manual unction Icon unction unction Instrument Gas head pressure Hardware build s Passive mode Power up Pressure filter Instrument alias accuracy name e Question La Range Lk Restore to as does ue SE shipped settings le Pr
46. le 2 Module 1 SUPPLY OUTLET VENT REF SUPPLY OUTLET VENT REF OO 0 0 0 OOOOO viewed from rear of instrument Module 2 Module 1 SUPPLY OUTLET VENT REF SUPPLY OUTLET VENT REF EK 000 O E 00000 viewed from rear of instrument Open to Atmosphere K0443 Issue No 3 6 22 PACE Pressure Controller User Manual Example of Altitude and Airspeed Testing This example shows how two channel PACE instruments can be used to generate altitude and airspeed Cautions 1 Before testing set the rates of change for both Pitot and Static to a safe value A high rate of change can damage sensitive aeronautical components Refer to the appropriate Component Maintenance Manual for the unit under test 2 In this example configuration negative airspeed can be generated this can damage an airspeed indicator To prevent negative airspeed apply the static pressure before the pitot pressure for increasing and decreasing airspeed values Static Altitude Pitot rear view Module 2 Module 1 of PACE SUPPLY OUTLET VENT REF SUPPLY OUTLET VENT REF O LL Pressure source Pressure Pressure regulator regulator Units The units can be either the aeronautical or pressure units At any time the units can be changed between pressure and pressure converted to aeronautical units The display shows the outlet pressure converted to Altitude CAS or Mach using BS 2G 199 1984 conversions and assuming standard atmospheric conditions Based on tabl
47. le procedure before starting a process on a component or system K0443 Issue No 3 3 1 3 Operation 3 2 Power up sequence The following sequences of operation show the instrument in measure or control mode Note The following sequence is an example the values and selections displayed depend on the range s and options enabled in the instrument To contro pressure the outlet port must be connected to a UUT or a blanking plug fitted The UUT must be of the correct pressure rating or the instrument set to limit the set point value to a safe pressure Set the power supply to ON and the power up example display routine starts 1 The display screen first shows 2 After a short time the display shows the start of the power up sequence the instrument carries out a self test If the test finds a fault the display shows an error refer to Section 5 Fault Finding and Testing 3 After a successful self test sequence the system enables the touch screen and changes to measure mode The touch screen shows the measured pressure in the parameters selected in measure set up Note The PACE 6000 shows a single display default this is the left hand pressure control module Change to dual display in Global Set up Display Druck at Heart Measured Pressure mbarg H 1 198 03 5 ee Westen a SS a sli am 0 00 om E 4 1 Measure set up 2 PO Aa 3 Nudge up 4 Set point entry 5 Status touch to enter control set up 6 Nudge down
48. ller is entered and Control is then selected then the range is changed to the higher ranged controller and this then controls to the entered set point Controller Off Decreasing Set point With both controllers in Measure mode if a set point within the range of the higher ranged controller is entered and Control is then selected the higher ranged controller controls to the entered set point With both controllers in Measure mode if a set point above the range of the lower ranged controller is entered and Control is then selected then the higher ranged controller will control to this set point When the controller is in limits then the range is changed to the lower ranged controller and this then controls to the entered set point Controller On Increasing Set point With the lower ranged controller in Control mode if a set point within the range of the lower ranged controller is entered then the lower ranged controller controls to the entered set point If the set point is increased to above the lower range but still within the higher range then the lower range controller switches off and that the higher range controller switches on and controls to the entered set point Controller On Decreasing Set point With the higher ranged controller in Control mode if a set point within the range of the higher ranged controller is entered then the higher ranged controller controls to the entered set point If the set point is decreased to within th
49. lt Finding Introduction The PACE instrument contains a built in self test and diagnosis system that continuously monitors the performance of the unit At power up the system performs a self test This section details the standard serviceability test A fault finding table lists possible faults the probable cause and the procedures to rectify the fault Standard Serviceability Test The following procedure shows if the unit is serviceable and checks functions and facilities of the PACE instrument Procedure Connect the instrument as detailed in Section 2 Installation Connect a UUT or fit blanks to the output port After a successful power up select measure set up and proceed as follows a Select the required units of pressure measurement from the measure set up menu bo Press the status area to enter control set up and select the required slew rate and vent rate c Press set point and using the numeric keys set a value within the pressure range of the instrument ld Check that the screen display shows the selected units of pressure measurement the selected type of slew rate and the set point Press the contra key to start le The screen display changes as follows The measured pressure digits change from black to blue and shows the pressure value changing towards the set point If enabled the effort meter indicates the work effort of the controller f When the controller achieves the selected pressure
50. n of logic input of control module Vent and ve FS enables vent selection and shows full scale pressure in current selected units Zero and ve FS enables zero selection and shows full scale pressure in current selected units Rate indicates slew rate set Barometric displays barometric pressure Tare indicates tare condition and value Status area examples Measured Pressure mbarg 198 03 ve Full Scale ve Full Scale Full scale 1000 00 3500 00 Controlled Pressure barg O SC e In Limits meter Time to In Limits ee S e 4499 Ee ZS Effort meter pre ee Vent and ve FS ve Full Scale mei 3 50000 Control age Y ve Full Scale Zero and ve FS 0 See so Measured Pressure mbar g A 288 35 za Barometric arometric esla E lt 500 00 El 3 7 Barometric Reference Option When installed this option allows absolute or gauge pressure range selection To obtain absolute pressure the instrument uses a summation of gauge pressure and barometric pressure measured by the barometric sensor Refer to Section 6 8 and the data sheets for the performance of barometric reference and precision of absolute ranges E PACE Pressure Controller User Manual 3 8 Supervisor set up Supervisor st 4 S r Calibr upervisol Setu ll ll hl More next page y Normal _Powerup j enter four digit code 0268 Protecti
51. nd the pump When the pump supply is switched off the valve opens allowing atmospheric pressure to enter the pump directly rather than through the pipe to the instrument Note Without this arrangement oil may progressively move up the supply pipe and into the instrument Pump Performance Recommended for ranges above 2 bar 30 psi gauge positive full scale 1 When installing a vacuum supply protect the vacuum pump against the discharge of positive pressure by the controller into the vacuum pump This may result in reducing vacuum pump performance 2 Use a check valve in the negative supply to vent excess pressure to atmosphere if the vacuum pressure rises above atmospheric pressure The check valve should be installed on the instrument side of a volume which is approximately equal to the system volume The volume slows any rapid pressure rise giving the vacuum pump time to reduce the pressure Note A wide bore vacuum pipe can have enough volume and used with a check valve could provide the necessary overpressure protection K0443 Issue No 3 6 4 PACE Pressure Controller User Manual Venting Either a zero or vent operation uses the vent port Vent The system gas at the output pressure can be released from the vent port Unrestricted gas flow occurs in this operation Recommended Use a controlled method to reduce the system pressure at a controlled rate to near atmospheric pressure then select vent Zero During a zero opera
52. ng software enabled options can be selected from the task menu Leak Test option This task applies one or two test pressures to either an external system to find any leaks in a system connected to the instrument or an internal leak check This task sets the test pressure control dwell time at the test pressure and the leak test time measure dwell time Measured Pressure barg 1 499854 z Test Pressure 1 1 500000 Test Pressure 2 0 054000 End Pressure 0 000000 Control Dwell 00 00 20 Measure Dwell 00 00 20 Leak Rate fsecond At the start of the test Leak test the instrument applies a test pressure to the user system A control dwell time allows the user system to thermally stabilise The instrument changes to measured mode and then records the pressure change during measure dwell time On completion the display shows the leak rate results with leak rates per second or per minute in the current pressure units DOC Example set up parameter Leak Test Parameters Measure Dwell 00 00 20 A Test Pressure 1 1500 000 mbar T Test pressure2 Se Wim 00 00 20 End pressure l1l2 3l4 5 Control dwell time le z 8 lo Lo EL Example leak rate results Measure dwell time Measured Pressure bar g 0 000392 0 Vu er go be 0 N N c s 0 selected In measure set Leak rate second Test Pressure 1
53. nications Instrument Emulation Refer to PACE Communications Manual Instrument Emulation KO469 6 11 Specification Refer to the PACE 5000 datasheet or PACE 6000 datasheet for details Note The data sheet SDS 0001 or SDS 0008 is contained in the CD shipped with the product 6 12 Return Goods Material Procedure Should the unit require calibration or become unserviceable it can be returned to the nearest GE Service Centre listed at gesensinginspection com Please contact our Service Department either by phone fax or E mail to obtain a Return Goods Authorisation RGA or in the USA Return Material Authorization RMA providing the following information Product i e PACE instrument Serial number Details of defect work to be undertaken Calibration traceability requirements Operating conditions Safety Precautions You must also tell us if the product has been in contact with anything hazardous or toxic and the relevant COSHH or in the USA MSDS references and precautions to be taken when handling Important notice Service or calibration by unauthorized sources will affect the warranty and may not guarantee further performance K0443 Issue No 3 6 29 6 Reference and Specification 6 13 Packaging Procedure 1 The instrument should be at zero ambient pressure Set the power switch to off Shut off the pneumatic pressure and vacuum supplies to the instrument Switch off and isolate the electrical power supply to the instrume
54. nt Remove the instrument from the equipment rack to access the rear panel Disconnect the power supply cable and the pneumatic supply hose assemblies Stow the power supply cable in the packaging below Remove any pressure adaptors diffusers and restrictors 2 If available use the original packing material When using packing materials other than the original proceed as follows Fit protection to all the ports to prevent ingress of moisture and dirt Note Use the original red plastic plugs or low tack masking tape e Wrap unit in polyethylene sheeting e Select a double wall cardboard container Inside dimensions must be at least 15 cm 6 greater than the equipment The carton must meet test strength requirements of gt 125 kg 275 lbs e Protect all sides with shock absorbing material to prevent equipment movement within the container Seal carton with approved sealing tape Mark carton FRAGILE on all sides top and bottom of shipping container Environment The following conditions apply for both shipping and storage Temperature range 20 to 70 C LA to 158 F K0443 Issue No 3 6 30 The General Electric Company All rights reserved
55. nt s settings for both measure and control modes This set up menu provides PIN protected access to the supervisor set up and calibration Pressing Global Set up from measure or control set up menu changes the touch screen display to show five selections Supervisor Set up Calibration Save Recall User Set up and Display Global Setup Supervisor Calibration Setup 5 Display 1 Selections 2 Escape key Global set up Gis Supervisor Setup mm Supervisor set up set up Calibration Save recall user Save recall user set up Save recall user set up Display Save Recall User Setup Lo Global Setup E Calibration Save Recall User Setup Un Recall current user set up Save current user set up Resolution Measure display resolution Back light brightness Timer time out setting Audio volume of full volume Status area Dual single display PACE 6000 only K0443 Issue No 3 Status area settings Enables the user to view an operating condition or parameter of the instrument Full scale pressure in current selected units of the pressure range Source positive and negative source pressure values in current selected units Effort meter indicates controller effort In Limits meter indicates controller in limits condition and time to in limits Module logic P indicates status conditio
56. onally left blank K0443 Issue No 3 3 18 PACE Pressure Controller User Manual 4 4 1 4 2 4 3 4 4 4 5 Maintenance Introduction This section contains the routine maintenance and procedures to replace components detailed in Section 5 Testing and Fault Finding and listed in Table 4 2 Table 4 1 Maintenance Tasks Visual Inspection Daily before use week Calibration 12 months t Pressure Module filters Determined by usage Replace Pressure Module Pressure controller operating hours may change depends on usage e g rack mounted bench top and environment e g humidity dust t may change depends on the required accuracy Visual Inspection Inspect the external of the instrument and associated equipment for obvious signs of damage and dirt If necessary clean the instrument as detailed below Cleaning Caution Donotuse solvents for cleaning Clean the instrument every week Clean the front panel with a damp lint free cloth and mild detergent Test Carry out a standard serviceability test refer to section 5 2 Calibration The pressure control module or the complete instrument should be withdrawn from service and returned to the manufacturer or calibration facility refer to section 6 14 To find the date of the last calibration press Measure or Control set up Status Calibration history Do not use an instrument with out of date calibration K0443 Issue No 3 4 1 4 Maintenance 4 6 4 7
57. otective vent Recall user set up E K0443 Issue No 3 6 7 6 Reference and Specification unction E ee m unction con Restore settings 2 a Save as shipped A x fv Set point disable Select range enable limit Slew rate linear Slew rate max rate Save recall user set up settings Set point limits Set point higher limit Set serial number Software build Software upgrade histon Software upgrade Switch test Test program Test program copy Test program SC K0443 Issue No 3 6 8 PACE Pressure Controller User Manual con unction unction con unction AN Vent time out Vent Yes No es i Vent set up Y output A 6 4 Measure Set up Pressure zero During use the instrument pressure sensor can show small zero shifts caused by time and temperature changes Regular zeroing increases measuring precision Process Selects display processing features that change the reading as follows Pressure can be expressed as a percentage of full scale or as a percentage of a specified span Filter The reading can be filtered by a custom low pass filter or the filter can be disabled default disabled The controller works at a speed independent of the filter time constant Tare A specific tare value can be selected or the current displayed pressure reading can be captured as the tare value The display shows the selected tare value in the pressure window
58. ox 1sec Beep off 19 MEASURE Switch to Measure controller off 20 TEXT Operator instruction e g Wait for beep record pressure 21 DWELL 30 Wait for 30 sec 22 BEEP Beep on approx 1 sec Beep off 23 TEXT Operator instruction e g Min pressure allowed 785 mbar 24 PAUSE WAIT for operator input to touch single step 25 VENT Vent END Program end K0443 Issue No 3 6 19 6 Reference and Specification Programming Loops To program a loop use the GOTO command Include the COUNT command in the loop for counting the number of loop cycles Note The test program commands do not include tests for conditional jumps to stop a test program from looping Stop must be selected by the operator Example of programming a loop START 1 UNITS mbar 2 RATE_VALUE 100 3 RESOLUTION 5 4 IN LIMITS 5 IN LIMITS TIME 6 SETTLING_N_ O zero 7 TEXT 8 ZERO 9 SET POINT 400 10 CONTROL 11 WAIT IN LIMITS 12 BEEP ON 13 MEASURE 14 DWELL 50 15 SET POINT 800 16 CONTROL 17 WAIT IN LIMITS 18 BEEP ON 19 MEASURE 20 COUNT 21 VENT 22 GOTO 9 END K0443 Issue No 3 Program start Select units mbar Select rate 100mbar min Display resolution 5 digits Sets the In limits band 10 00 00 10 secs No overshoot Operator instruction eg Connect UUT Performs a sensor zero Set point 400 mbar Controller ON Wait for In limits condition Beep on approx 1sec Beep off Switch to measure controller off Wait 30 sec Set point 800 mba
59. r Controller on Wait for In limits condition Beep on approx 1sec Beep off Switch to measure controller off Increment loop counter Vent Loop back to program line 9 Program end PACE Pressure Controller User Manual Barometric Reference Option The barometric reference option measures the barometric pressure at the reference port When installed this option allows absolute or gauge pressure range selection To obtain absolute pressure the instrument uses a Summation of gauge pressure and barometric pressure measured by the barometric sensor a1n osqb opnasg uondo 21119410109 000939Yd 9 NSSIIJ OJIZ IINJOSQY sqo 1090 aJnssaud aduaJajas abnob i d X sqo JOQWOZOL 6 oquo 6 soquuoss SGD JOQWOZOT 01 SQD JOQUIOS6 IBUDY IMJOSQY Opnasd 6 sOQqUuQ F SBUDY afnpe sqb JOQWQS9 INSS Jd DUJ WOING SqD 1Dq2 0 SQD JOGO IBUDY ajnjosqy Opnasd 5 DIA00TF 6 ISAGTE 6 10QTF SBUDY afnpe uodo IJBWOING YUM INPON O1JUOD ADAT SD JOQTT 01 Sqd 10q0 aBUDY aynjosqy opnasd 5 JOQOT 0 6 JOQT 3BUDY 3BNDO uodo DWOING YUM ANPON JO1 UO0D JOG OT SQD JOQUIOSET 0 sqo JOQUIOSI IBUDY IMJOSQY OpNasd 6 Joquos sF SBUDy 3BNDO uodo IIJAWOING YUM ANPON O1JUOD JOQUIOSS uodo DILJAWOING YUM ANPON O1 U0D JOQUIOZ Refer to SDS 0001 or SDS 0008 for the performance of barometric reference and precision of absolute ranges 6 21 K0443 Issue No 3 6 Reference and Specification Aeronautic
60. ro offset stored for current range Pressure zero Range allows selection of available pressure ranges 1 Measure Setup at Process Filter pressure reading 4 _ ieee Tare SS Task tg Y LA D Units select from list of available pressure measurement units y i Global set u see 3 6 Pressing stores settings i WI and exits set up ENEE Press to exit set up Measure Setu Timed p Set up zero H Auto 2 Off AE Zero T Status bat k Pressing 5 stores settings Pressing wor returns to the first page and exits set up Absolute range selection available when press to exit set up barometric option installed Auto range only available with two channel instruments PACE6000 Measure Setup y 2 gt BB ta Pressure Zero Range Process Task y E T d Units Global Setup Zero Setup Analogue O P Setup pem VFC Teip Seng H K0443 Issue No 3 3 4 PACE Pressure Controller User Manual AUTO RANGE only available with two channel instruments Controller Off Increasing Set point With both controllers in Measure mode if a set point within the range of the lower ranged controller is entered and Control is then selected the lower ranged controller controls to the entered set point With both controllers in Measure mode if a set point above the range of the lower ranged contro
61. set point the screen display changes as follows The colour of the displayed pressure value changes from blue to green indicating that the controller is within the in limits tolerance If enabled the effort meter shows the controller effort to keep the pressure at the set point lg Select vent and the pressure reduces to atmospheric pressure at a controlled rate vent slew h The test is completed when the controller is at atmospheric pressure Notes 1 The vent opens and remains open until OK is pressed 2 Always use the vent function before disconnecting pressure equipment from the outlet port j The instrument automatically returns to measure mode The colour of the displayed pressure value changes to black After a successful serviceability test the instrument is ready for use K0443 Issue No 3 5 1 5 Testing and Fault Finding 5 3 Fault Finding Check the fault conditions and solutions listed in the following table before contacting gesensinginspection com or a recommended Service Agent Check rear panel switch set to on Power supply connected display not lit Check fuse and if necessary replace Check electrical power supply fuse or circuit breaker l 24 Y DC output intermittent Internal self resetting fuse operating Reduce load current to specified value Instruments functions but does not reach all set Check pneumatic supplies for correct pressures points Check system for leaks Increasin
62. ssure at an exponential rate but remaining within limits Control mode Selects one of three modes Active Control In this mode the controller continuously maintains the set point compensating for small pressure leaks and thermal affects Passive Control In this mode the user can define a band either side of the set point the default band equals to the instrument s precision When the controlled pressure enters this band the controller automatically shuts off Subsequently if the measured pressure exits the band the controller automatically re establishes the pressure without instability the controlled pressure re enters the band Note If passive mode is in use with a leak free and thermally stable system then the control stability contribution can be discounted from the uncertainty calculation Zero Gauge Control In this mode the controller switches off once stable at zero gauge and the zero valve opened Entering a new set point causes the zero valve to close and the controller starts to control to the new set point K0443 Issue No 3 6 10 PACE Pressure Controller User Manual Global set up See para 6 6 not a PIN protected menu Status The display shows e Instrument status model module control sensor ve source sensor and ve source sensor e Software build read only data Hardware build read only data Calibration history read only data Zero history read only data Communications IEEE 488 and RS232 are
63. ted set point value Note When using power up set point be sure to disable the Idle Timeout feature Lock Tasks Individual tasks Allows any combination of individual tasks to be disabled Note Restricts operation of the instrument to specific tasks or functions ideal for production procedures All Disables all tasks Change PIN Changes the Supervisor PIN enter the existing PIN then the new PIN and confirmation of the new PIN Note Confirmation of the new PIN permanently replaces the old PIN Record this new PIN and keep this record ona safe place Should this new PIN become lost it can only be reset by returning the instrument to a GE service centre User defined units Permits the user to define a set of units Following the on screen prompts special units may be set by selecting a Pascal multiplier and assigning a five character name Instrument alias name Permits the user to define a 20 character alias name for the instrument The instrument returns this name through the communications interfaces Language Operation in any of the following languages can be selected English default French German Italian Portuguese Spanish Russian Chinese Japanese Further languages can be added Restore as shipped settings Restores instrument settings to factory default Note Does not affect PIN settings K0443 Issue No 3 6 14 PACE Pressure Controller User Manual 6 8 Options Options in the Task Menu The followi
64. the instrument Software history Instrument Main Code read only Instrument OS Build read only Instrument Boot ROM read only m Module 1 Main Code read only __ Module 1 Boot ROM read only Analogue O P Main Code read only L More next page y VFC Main Code read only L Analogue O P Boot Code read only Software History E fo Instrument OS Build Instrument Main Code Boot ROM Module Main Module Boot Analogue O P pl Code ROM Main Code Example Module Main Code DK0364 v00 00 55 installed 10 06 08 09 33 21 DK0364 v00 00 54 installed 06 06 08 14 47 06 DKO364 v00 00 50 installed 27 05 08 09 30 54 DKO364 v00 00 49 installed 22 05 08 08 36 32 DKO364 v00 00 48 installed 20 05 08 11 17 31 DKO364 v00 00 46 installed 20 05 08 10 29 14 DKO364 v00 00 48 installed 19 05 08 14 32 31 DKO364 v00 00 46 installed 16 05 08 14 43 11 DKO364 v00 00 44 installed 14 05 08 16 30 28 DKO364 v00 00 43 installed 14 05 08 10 19 22 Software History VFC Main VFC Boot Code ROM Analogue O P Boot ROM Pressing More returns to the first page ____ VFC Boot Code read only press TI to exit set up Module 2 Main ROM read only Module 2 Boot Code read only K0443 Issue No 3 3 Operation intenti
65. tion only the internal volume of the instrument vents to atmosphere Recommended Do not obstruct the vent port To diffuse gas exhaust a diffuser may be fitted to the vent port Output port The output port provides the controlled test pressure to the unit under test UUT Reference Port The reference port provides the negative pressure to the gauge sensor and to the barometric reference option Gauge sensors use this port identified as REF For gauge sensors without a barometric reference small pressures can be applied refer to the MWP stated on the rear panel of the control module All other pressure measurement requires the port to be opened to atmosphere When in gauge mode the instrument shows and controls the pressure difference between the reference port and the output port Note This is not a true differential operation as there is no true differential calibration of the sensor The transducer of the barometric reference option senses atmospheric pressure via the reference port when enabled the port MUST be open to atmosphere The reference connection should be actively used differential connection option for precision low pressure measurement The instrument controls pressure relative to the pressure at the reference port An atmospheric pressure change causes the controller to adjust the pressure and appears at the pressure output as apparent instability To keep a stable controlled pressure the reference port shou
66. truments Up to 30 instruments can be connected through a high speed data bus to the computer controller Note The length of each IEEE 488 cable must be less than 3 metres to comply with the EMC requirements see Section 6 Reference and Specification Single Unit Installation Figure 2 8 Connect an IEEE 488 connector cable assembly to the rear panel IEEE 488 connector of the instrument e Connect the other end of the connector cable assembly to the IEEE 488 connector on the controller computer e Change the IEEE 488 communication parameters as described in the Supervisor set up refer to Section 6 7 Multiple Unit Installation Figure 2 8 To install multiple units use stacking plugs to link from first instrument to the second instrument Proceed as follows e Connect a pair of IEEE 488 stacking connectors to the rear panel IEEE 488 connector of the instrument 1 Connector to rear panel of first instrument 2 Connector from controller computer 3 Connector to rear panel of second instrument e Connect the other end of one of the connectors to the IEEE 488 connector on the controller computer and the other connector into the next instrument e Repeat this procedure for all the instruments in the system e Use the Supervisor set up communications menu on each instrument to set up the required communication parameters refer to Section 3 8 K0443 Issue No 3 2 12 PACE Pressure Controller User Manual AWVYs S
67. uum pump does not matter but generally higher pump flow improves PACE control performance However low pressure ranges lt 700mbar require vacuum regulation or the use of the negative gauge pressure generator lO NEG G GEN 1 option Conclusion Use a vacuum supply for Absolute ranges Negative gauge ranges A vacuum supply improves Time to reduce system pressure at pressures below 2 bar 30 psi full scale Control stability near atmospheric pressure Overshoot at low pressures To improve performance at or near gauge zero K0443 Issue No 3 6 2 PACE Pressure Controller User Manual Table 6 1 Air Density Values Values of air density kg m 3 for air of relative humidity 50 and containing 0 04 carbon dioxide by volume Air temperature C Note that 100 kPa 1 bar SC SES SES SES SES K0443 Issue No 3 6 3 6 Reference and Specification 6 2 Operational Requirements Special Note A contaminated UUT must have additional in line filters connected between the output port and the UUT to prevent contamination of the instrument Negative or Vacuum Supply The negative supply for absolute control does not need to be regulated Any variation between this and absolute zero will affect instrument operation if controlling at low absolute pressures Oil Contamination Precautions must be taken against oil transfer to the instrument Recommended 1 A normally open venting solenoid connected to atmosphere a
68. ve Vent Vent Enable disable In Limits select High source pressure Low source pressure Marins Enable disable high pressure Enable disable low pressure Communications MX Idle timeout Time out i L enablefdicable Gas Ges head correction correction Enable Ges head correction E A Test port medium Pressing the escape key stores settings and returns to global set up Gas Head Correction 4 iJ Supervisor Protective Vent In limits value full scale time period time to stay in limits before triggering indication IEEE 488 MO Correction parameters Condition enable disable Head in or m Local gravity m s Supervisor Barometric mbar Reference port medium Change PIN Temperature gas Humidity gas Power Up E Lock Tasks gi User Defined Units Mm Instrument Alias Name Language ee air or nitrogen Rotor Power on set point Lock Lock unlock individual tasks Lock tasks Lock unlock all tasks Change PIN PIN User User defined units units Instrument Instrument alias name name Language next page gt More Restore last _ Restore last settings Enter new supervisor PIN _ User defined units User defined units user unit 1 Enter new PIN Re enter new PIN User defined units enter the name of the defined units enter the equival
69. y G 1 8 Output G 1 8 Vent G 1 8 Reference G 1 8 Refer to the data sheets SDS0001 or SDS0008 for a complete range of adaptors Pressure supply Figure 2 1 1 2 5 1 The pressure supply must be clean dry gas nitrogen or air and at the correct pressure refer to the specification Section 6 Ensure the user systems can be isolated and vented Connect pressure and vacuum supplies to the SUPPLY and SUPPLY connection ports Connect the Unit Under Test UUT to the required outlet connection port QS CONSULT HANDBOOK OPTION 2 ETHERNET PNEUMA CONTROL MO a WT CAN 2 NATI CONTROL MODULE el i E j a USB B AN FITO PIP 44440 Tee DRUCK LES OFH UK 2 Q 7 Example instrument 2D BSP connector 2 Bonded seal Note For instruments with NPT connections use adequate pressure sealing Figure 2 1 Pneumatic Connections K0443 Issue No 3 2 2 PACE Pressure Controller User Manual Installation The instrument requires a positive pressure supply instruments operating in an absolute range or negative pressure range require a vacuum supply A vacuum supply should be used for a fast response for instruments operating near atmospheric pressure For dual channel operation two independent pressure and vacuum supplies can be used Important notes When using two pressure modules make sure that the modules are compatible refer to Section 4 9 All connections must comply with the Pressure Equipment Directive PED
Download Pdf Manuals
Related Search
Related Contents
NEO-SOLAR - Motive Srl Siemens Webcam TP177B User's Manual Acorn multiscan monitor User Guide Phonix S9070GPR mobile phone case HP 冬のリーフレット外面 Scanner digital 3000 DV Plus IC Portal – Clerk User Manual (IS Claim Module) ABUS TVCC40530 surveillance camera B.Tech EIE Syllabus - Narayana Engineering College, Nellore Copyright © All rights reserved.
Failed to retrieve file