Home
        series - Jenway
         Contents
1.     European or  American  using the Up Down keys  The Left Right arrow keys allow  forward and backward movement along the row of digits  Press the  STO key to accept the setting     13 950 050 REV A 04 03    3 3  DO  MODE    This gives a direct readout of the  DO  of the sample under test as either   oxygen saturation or    air saturation  If a cal value  lt 25 0  is entered the resolution will automatically increase to 0 1        Oxygen Saturation   this will give a reading of 20 9  in air where 20 9  is the amount of  oxygen as a percentage of oxygen in air and will give a reading directly as a percentage of oxygen in  the sample  All measurements in this range are given to a resolution of 0 1        Air Saturation   this will directly assign a percentage dissolved oxygen to air saturated water  and further readings are then given as a percentage of this air saturated water sample  All  measurements in this range are given to a resolution of 1      DO  Calibration    If DO  measurements are to be performed in mg l the current barometric pressure should be  entered into the instrument via the Calibration Setup    menu        Calibration  D 02 100 0   Manual Pressure bar  1 013   Manual Salinity afl  00  Setup  oboe  Sates  Ree    Calibration  DO  allows the setting of  DO  value for calibration between the limits of 0 1 to  199 9 using the Up Down arrow keys     Manual Pressure  bar  allows the setting of the current barometric pressure between the limits of  0 1 to 6 0bar
2.    3 TXD 3 TXD   4 DTR 4 DTR   5 GND 5 GND   6 DSR 6 DSR   7RTS ee 7RTS   8 CTS 8 CTS   9 9    NOTE  Interface Cable  Order Code  013 203  is required     30 950 050 REV A 04 03    Interfacing  continued     The RS232 communications parameters on the computer or printer need to be set to match those of  the Model 9500  as detailed below     1200 Baud 9600 Baud  7 Data Bits 8 data bits  Odd Parity OR No parity  1 Stop Bit 1 stop bit    Setting of these options is detailed in Section 3 2 8     The Model 9500 supports both hardware  CTS RTS  flow control and software XON XOFF flow con   trol     Pressing the PRINT key outputs from the RS232 interface     Sending an ASCII    D    to the 9500 causes a printout of the current displayed reading plus sample  number     Sending an ASCII    C    causes a printout of the last calibration parameters     Sending an ASCII  P  causes a printout of the stored readings     6 3 Keypad Emulation    Keypad remote control using RS232 interface     7   Instrument On   Standby   Escape  1   Calibrate   Memory Clear   9   Print   8   Up Arrow   2   Down Arrow   4   Left Arrow   6   Right Arrow   30r5   Enter   Store    31 950 050 REV A 04 03    6 4 Printing    A 32 column serial printer  037 701  is available for use with the Model 9500   There are two methods of connecting the serial printer to the Model 9500     a  IrDA   the IrDA interface is a line of sight  wireless communication protocol  The IrDA sensor   located on the front left hand corner
3.   Fit membrane    Replace membrane    Fill or replace membrane   Replace probe     Clean the tip or replace the membrane   Replace membrane     Refill or replace membrane     Replace membrane     Replace probe   Use fresh solution and re calibrate     Reset salinity values   Recalibrate with fresh solution   Replace solution    Replace standards    Reset barometric pressure     Line up units or alternatively use  supplied RS232 connector    The feed light on the printer will flash if  the unit requires paper    Connect ac power supply     If the above does not answer your query try the FAQ section on the www Jenway com Website     33    950 050 REV A 04 03    HEALTH  amp  SAFETY   PRODUCT  Potassium Chloride Solution  PHYSICAL DATA   Description  Colourless solution   Specific Gravity  1 0   Solubility in water  miscible in all    proportions    HEALTH HAZARD   May be harmful if ingested in quantity  causing nausea  vomiting and diarrhoea   May irritate eyes     FIRST AID   Eyes Irrigate thoroughly with water  If discomfort persists OBTAIN MEDICAL ATTENTION    Lungs Remove from exposure    Skin Wash off thoroughly with soap and water    Mouth Wash out mouth thoroughly with water  In severe cases OBTAIN MEDICAL  ATTENTION     PRODUCT  Sodium Sulphite Anhydrous  PHYSICAL DATA    Description  White Powder  Specific Gravity  2 63  Solubility in water  very soluble    HEALTH HAZARD   If ingested in quantity can cause gastric irritation  colic  diarrhoea  central  nervous system de
4.   e g  the dissolved oxygen  c1  day  0  3  was not less  than or equal to dissolved oxygen  C1  day  0      dissolved oxygen  C2  day  n   This does not  prevent the answer being displayed  but does warrant a warning as it is a condition of the BSI  standards EN 1889 1 1998  EN 1889 2 1998    The same explanation applies to the following warning message        The following BOD condition was not met    C1 C2  lt    2 C1  3   Press any key to continue       ANON KR mg l    AN The following BOD condition    was not met     e1 c2  lt    2 1  3      Press any key to continue            Warning Messages     General Messages        Oxygen consumed over n days  is greater than 1 5 mg l  Press any key to continue        Rei       q Oxygen consumed over  n days  is greater  than 1 5 mg l  Press  any key to continue     B  B  S  D       The above warning message alerts the user to a possible contamination within the blank samples   It is a condition of the BSI standard that over n days the oxygen within the blank should not exceed  1 5 mg l     22 950 050 REV A 04 03       Overwrite existing blank STO Key to continue ESC Key to cancel      amp      Overwrite existing sample STO Key to continue ESC Key to cancel      There is a blank sample reading already stored  the BOD time limit of 24 hours has not expired     therefore the user can overwrite the existing blank sample measurement   If 24 hours has expired the appropriate message will be shown to indicate this        BOD Batch and S
5.   they are unable to go back and repeat any day  0  measurement   unless they delete the batch  Deleting the batch erases any results stored in that batch from  sample 1 to 10     When displaying the BOD answer  if there is only one blank reading day  0  or day  n  but not both   then the BOD answer is calculated without blank criteria    Overwrite   You can overwrite a day  0  or day  n  reading providing 24 hours hasn t elapsed from any initial    reading in said batch  If you overwrite a day  n  reading the BOD result is updated straight away  with the new answer if there is one     20 950 050 REV A 04 03    BOD Calculation    Using the two BSI Standards for BOD calculation  with without blank solution   EN 1889 1 1998 and  EN 1889 2 1998  There are three conditions that need to be met for a successful BOD answer     C1   is the dissolved oxygen of the sample on day  0     C2   is the dissolved oxygen of the sample on day  n     C3   is the dissolved oxygen of the blank sample on day  0    C4   is the dissolved oxygen of the blank sample on day  n    Ve   is the volume sample within the test solution    Vt   is the total volume of the test solution     Condition 1   with blank    During the course of  n  days the dissolved oxygen measurement shouldn t fall by more than 1 5  mg l     Condition 2 3     with out blank   BOD answer is evaluated providing the following conditions are met     C1 3   lt    C1   C2   lt     2 C1  3  broken down into two parts     C1 3   lt    C1  
6.  Alarm High  mg l Alarm Low    3 2 5 G L P Setup        Exit  Cal Reminder    Cal Reminder Interval    Cal Reminder Audible Alarm    User ID    Alarm Outputs Disabled  Audible Alarm Warning Disabled   DO2 Alarm High 199 0   DO2 Alarm Low 000 0  mg l Alarm High 19 99  mg l Alarm Low 00 00       enables the user to exit the Setup menu and return to the main   set up screen    can be enabled or disabled by using the Up Down keys which toggle  between the two settings    can be enabled or disabled by using the Up Down keys which toggle  between the two settings    allows the user to set the high alarm limit up to 199 0   allows the user to set the low alarm limit down to 000 0   allows the user to set the high alarm limit up to 19 99   allows the user to set the low alarm limit down to 00 00    Cal Reminder Disabled  Cal Reminder Interval Hr  024  Cal Reminder Audible Alarm Disabled  User ID 0000  Batch ID 000    Security Setup          enables the user to exit the Setup menu and return to the main   set up screen    can be enabled or disabled by using the Up Down keys which toggle  between the two settings    can be set within the limits of 001 to 999 hours by using the Up Down  keys  The Left Right arrow keys allow forward and backward  movement along the row of digits  Press the STO key to accept the  value    can be enabled or disabled by using the Up Down keys which toggle  between the two settings    up to a 4 digit code can be set using the Up Down keys  The Left   Right a
7.  C2  and    C1   C2   lt     2 C1  3    Failure of any of these conditions doesn   t prevent an answer being displayed  but a warning  message is displayed when viewing the results of the particular sample whose conditions failed   BOD Calculation   with blank   BOD     C1   C2    Vt     Ve  Vt   C3   C4    Vt Ve    BOD Calculation   without blank    BOD    C1     C2     General Points    Batches 1 to 20 work independently  allowing up to 20 batches to be running at the same time   Blank time elapsed is shown on the BOD printout   When the second blank reading is stored  sample ID is set to 1  and the cursor moves to the  sample square    When changing to a different batch number  sample ID reverts back to 1 and sample type reverts  back to blank    The maximum number of days shown on the results screen and measuring screen after a  complete BOD measurement is 99    BOD Batch is a separate set up parameter to Batch ID in G L P set up  the two are treated  independently of each other    Cal reminders may be enabled  but do not function while using the BOD mode    Results can be printed from the BOD measuring screen  and also the result screen    21 950 050 REV A 04 03    Warning Messages     BOD Conditions not met        The following BOD condition was not met    C1 3  lt   C1 C2   Press any key to continue       CY QB m        i  The following bod condition  2 was not met   ife1 3  lt   e1 c2      Press any key to continue    i       The first part of the BOD condition failed
8.  Optional Accessories  Optional Accessories  Spares    Section 6 Interfacing  Analogue  RS232  Keypad Emulation  Printing  Alarm Outputs    Section 7 Troubleshooting  Troubleshooting    Health  amp  Safety    EC Declaration of Conformity    a ae  hu    2 1  2 2  2 3  2 4  2 5    5 1  5 2    6 1  6 2  6 3  6 4  6 5    7 1    950 050 REV A 04 03    Section 1    1 1 Instrument Description    Model 9500  Dissolved Oxygen Meter  Operating Manual    Introduction    The 9500 is a fully specified DO  Temperature meter that includes full support for the B O D  5 day  test protocol  Up to 20 B O D  tests  each with up to 10 samples  can be stored  Powerful data logging  capabilities are included  with the ability to store up to 250 DO  readings either manually  at timed    intervals or alarm events     1 2 Specification    DO    Ranges   Resolution  Accuracy   Pressure correction  Salinity correction    Temperature  Range  Resolution  Accuracy    ATC Range  Manual Temp  Comp  Range    Calibration    Outputs    Clock    GLP    Display  Language option  Power   Size   Weight    0 to 199    0 to 25 0    0 to 19 99mg   1    0 1    0 01mg l    2  within 10  C of calibration temperature  Manual   Manual     10 to  60  C  14 to 140  F   0 1  C  1  F    0 5 C   1 F     0 to 60  C  32 to 140 F   0 to 60  C  32 to 140  F     Automatic    Analogue   RS232 serial  amp  IrDA printer interface   Alarm   24 hours  hrs min sec or day of month month year  leap year corrected  European and American form
9.  of data screens are available  The current screen and total  number of screens are shown in the bottom right hand corner of the screen     000 O 19mag l  25 0   C 03 50 40 20706703  000  0 19mgil   25 0   C 03 50 43 20706703  000   0 00mg l   25 0   C 04 15 03 23106703  000   0 00mgil   25 0   C 04 15 05 23106703       Other screens of data can be accessed using the Up Down arrow keys     The CLR key will delete data according to the set mode     If the PROMPT BEFORE DELETING warning is set then a second CLR is required to delete  results     The PRINT key will print all stored data     26 950 050 REV A 04 03    3 7 GLP Functions    A variety of GLP functions are available via the GLP Setup menu     Cal Reminder Disabled  Cal Reminder Interval Hr  024  Cal Reminder Audible Alarm Disabled  User ID 0000  Batch ID 000    Security Setup      Setup       A reminder that calibration is due can be set via the Cal Reminder option  Once this option is set a  valid calibration is required to use the 9500 after the time limit has elapsed  This option is  DISABLED by default    The interval time is set in hours on this menu     NOTE  A calibration should be performed immediately after the setting of this value to  reset the clock  This should be done as soon as set up is complete     The on screen reminder can be accompanied by an audible alarm  This is set using the audible  alarm setting on this menu     NOTE  The use of the Calibration Reminder feature will prevent the user from  per
10.  of the printer  should be in line with the Ir window on the side of  the 9500  The Ir icon on the symbol display indicates whether the units are attempting to connect   single icon flashing  or connected  two icons      b  Connect serial cable supplied with the printer to the 9 way socket located on the rear panel of the  instrument     To intiate a print out of data press the print key   When the first print is performed a header section will be printed showing     Instrument name   Time and Date   Spacing for entry of Operator  amp  User ID  Operator ID number   Most recent calibration information    This will be followed by results data dependent on mode selected  Details will also be given on  temperature  Time and date of the stored readings will be displayed     An asterisk     indicates that manual temperature compensation is being used    Each reading will be identified by a batch number    A calibration will reset the printout and the header information will be re printed    To obtain a print out of stored readings  enter the RESULTS MODE and press the print key  A print  out of all filled memory locations will then be generated     6 5 Alarm Outputs    The 9500 provides two alarm outputs  These can be accessed using the two 4mm connectors on  the rear of the 9500  They are open collector outputs  To set the alarm limts at which these are  activated  please refer to section 3 2 4    To use the alarm outputs  they should be ENABLED in the Alarms Setup screen see sect
11.  the Up Down keys which toggle between the  two settings     1 950 050 REV A 04 03    Data Log Interval  Memory Full    Prompt Before Deleting    Clr Key    3 2 8 Printer Setup        can be set between 00 00 01 and 23 59 59    gives the user to select the Stop  cease storing results and not to  overwrite existing stored information  or Overwrite  overwrite existing  results  options when the memory is full  Selection is made via the  Up Down keys which toggle between the two settings    can be enabled or disabled by using the Up Down keys which toggle  between the two settings    toggles between Deletes Results Before  Deletes All Results and  Deletes Results Since  Selection is made via the Up Down keys  which toggle between the four settings     Select the Printer Setup sub menu by highlighting the option and pressing the STO key  The follow     ing menu will be shown     Exit    Printer Interface    Serial Printer Baudrate    Printer Interface Serial    Serial Printer Baudrate 9600       enables the user to exit the Setup menu and return to the main  instrument display    toggles between Infrared and Serial  Selection of the preferred option  can be made using the Up Down keys which toggle between the two  settings    toggles between 9600 and 1200  Selection of the preferred option can  be made using the Up Down keys which toggle between the two  settings   Refer Section 6 2      12 950 050 REV A 04 03    3 2 9 Endpoint Detection Set up        Exit  Endpoint detection  En
12.  using the Up Down keys     Manual Salinity  g l  allows the setting of salinity values between the limits of 0 0 to 75 0     Prepare a zero oxygen solution by mixing 2gms of sodium sulphite in 100mls of water  Allow to  stand for a few minutes prior to use     To perform a calibration select the DO  measurement mode  Immerse the probe in the prepared  zero solution and stir gently for approximately two minutes  Press the CAL CLR key  Once the  reading has stabilised  if the endpoint detection is enabled the endpoint symbol will be displayed      The display will update to show     0  D 0 2    E IN Calibration 0 K   25 0 sa  Cal  0 0 D02  y j    00 13 27           14 950 050 REV A 04 03    Remove the probe from the zero solution     Fill a suitable container with a sample of clean water  Hold the probe so that the membrane is close  to  but not touching the surface of the water  approx  1cm above      Press the CAL CLR key  Once the reading has stabilised  if the endpoint detection is enabled the  endpoint symbol will be displayed   The display will update to show     meson       15 950 050 REV A 04 03    3 4 B O D  Biochemical Oxygen Demand     B O D  Biochemical Oxygen Demand  measurements are important in water treatment plants   laboratories and industrial waste water facilities for water quality testing and pollution control     The B O D test is a measure of the amount of oxygen that is consumed by bacteria as they  decompose the organic components of waste  Dissolved o
13.  when selected  SETUP  MODE  DO   mg l or  B O D    STATUS or RESULTS  If a double headed arrow symbol is present this indicates that  the mode can be changed to an alternative option    7  Real time clock   will display either date or time    8  The following symbols will appear along the display in  DO  and mg l modes    Padlock   Set up parameters security locked  Notepad symbol   data logging to internal memory    IrDA status  Alarm indication   an Up arrow refers to Hi alarm   a Down arrow refers to Low alarm     4 950 050 REV A 04 03    2 4 Keypad    2 4 1 Keypad layout    1  ESC    2  CAL  CLR    3  Print key  4  Up Arrow  Down Arrow  Left Arrow  Right Arrow    STO             Q          used to switch the instrument on and to place into standby mode  only if power  supply lead remains connected to the instrument   Also used to escape exit a  mode     used to select and perform a calibration sequence  This key is also used to clear  readings from Memory     used to initiate a print    used for adjustment during set up  to scroll results and to toggle between modes   used for adjustment during set up  to scroll results and to toggle between modes   used for adjustment during set up and to move between mode tags    used for adjustment during set up and to move between mode tags     used to accept an entered value in set up mode and to instigate a stored reading   This key can also be used as a CAL key during calibration     5 950 050 REV A 04 03    2 5 Inputs Outputs    
14. 037 901  903 311    Power Supplies    021 030  021 031  021 032  021 033    5 2 Spares    522 023  522 019    B O D  Kit   Zero Powder  Membrane Kit  amp  KCI  Dust Cover    IrDA Serial printer supplied with roll of thermal paper  serial connection lead  power  supply and power connection lead  UK    Paper Roll   Serial communication software  3 2  disk    Interface Cable Kit    Bench stirrer  battery operated   Extended length electrode rod  recommended for use with bench stirrer     U K  230V Power Supply  European 230V Power Supply  U S  115V Power Supply  230V leaded Power Supply    Replacement DO  probe  Membrane kit  amp  KCI    29 950 050 REV A 04 03    Section 6    Interfacing  6 1 Analogue  All units are provided with 2 x 4mm sockets  marked as ANALOG OUT  on the rear panel  An analogue  output voltage of 1mV per least significant digit is available from these sockets   6 2 RS232    The Bi directional RS232 interface is available on the rear panel 9 way D type connector   The connections are as follows     DCD 1   LINKED TO DTR AND DSR  RXD 2   INPUT TO 9500   TXD 3   OUTPUT FROM 9500   DTR 4   LINKED TO DCD AND DSR  GND 5   DSR 6   LINKED TO DCD AND DTR  RTS 7   OUTPUT FROM 9500   CTS 8   INPUT TO 9500    Suggested interconnections are detailed below     9500 IBM PC XT  25 way    D      1 DCD DCD 8   2 RXD RXD 3   3 TXD TXD 2   4 DTR DTR 20   5 GND GND7   6 DSR DSR 6   7 RTS RTS 4   8CTS            oTSS   9   9500 IBM PC XT  9 way    D      1 DCD 1 DCD   2 RXD 2 RXD
15. Fig 2 5 1 Rear panel layout       Alar OF ain    Dissolver  RS 252  Oxygen Analog OP           1  Dissolved Oxygen 7 pin DIN socket for connection of DO  probe     2  Alarm Output 2 x 4mm sockets  Open collector alarm outputs   Red for Hi   Black for Low     3  Printer Socket 9 way socket for RS232 connection     4  Power In AC 9V I P socket  2 1 x 5 5mm socket allowing the power supply to be  connected to the instrument     5  Analog Out 2x4mm sockets  Analogue output     6 950 050 REV A 04 03    Section 3    Operation    3 1 Good Practice Guidelines    1     Ensure the membrane is kept wet at all times  When not in use the probe should be stored  with the tip in a beaker of deionised water  For longer periods  overnight  the protective sheath  should be fitted  with the sponge insert soaked in distilled water  This will prevent the  electrolyte fill solution from drying out due to loss of water through the porous membrane  For  extended periods store dry     The probe is fitted with a temperature compensating element  This is housed underneath the  silver ring on the probe body  Always ensure the probe is immersed in solution to a depth  suitable to cover this ring  Allow sufficient time for these to respond if measuring samples  with varying temperatures or where sample temperature is significantly different to ambient  temperature     Ensure the sample is moving across the membrane at a speed greater than 15cm min to  avoid oxygen starvation at the membrane  If the flow r
16. Model 9500  Dissolved Oxygen Meter  Operating Manual    950 050 REV A 04 03    SAFETY    Please read this information carefully prior to installing or using this equipment     1  The unit described in this manual is designed to be operated only by trained personnel  Any  adjustments  maintenance and repair must be carried out as defined in this manual  by a person  qualified to be aware of the hazards involved     2  Itis essential that both operating and service personnel employ a safe system of work  in addition to  the detailed instructions specified in this manual     3  References should always be made to the Health  amp  Safety data supplied with any chemicals used   Generally accepted laboratory procedures for safe handling of chemicals should be employed     4  If it is suspected that safety protection has been impaired in any way  the unit must be made    inoperative and secured against any intended operation  The fault condition should immediately be  reported to the appropriate servicing authority     950 050 REV A 04 03    Model 9500  Dissolved Oxygen Meter  Operating Manual    Contents    Section 1 Introduction  Instrument Description  Instrument Specification    Section 2 Installation  Unpacking  Installation  Displays  Keypad  Inputs Outputs    Section 3 Operation  Good Practice Guidelines  Set Up Parameters   DO   Mode  B O D   Status Page  Results Storage  amp  Display  GLP Functions    Section 4 Maintenance  General  Probe Care  amp  Maintenance    Section 5
17. ample ID     Value being adjusted is above upper limit        This message will be shown during BOD Batch and Sample ID set up if the value selected is above    the upper limit of 10 and 20 respectively     23 950 050 REV A 04 03    3 5 Status Page  The Status page displays the current calibration information     If no valid calibration data is stored  e g  after a reset or a failed calibration  a warning screen will be  shown     The status page will show the calibration data in the order it was carried out  Date  time   temperature  and measurement        24 950 050 REV A 04 03    3 6 Results storage and display    The Model 9500 has a variety of options relating to the storage of data     Data Log Event Manual  Data Log To Memory  Data Log Interval 00 01 00  Memory Full Stop  Prompt Before Deleting Disabled  Clr Key Disabled    Printer Setup       Setup  nails J States J Rem     To initiate data logging the Data Log Event setting should be ENABLED    By default this option is MANUAL    The settings available are  Manual  Timed Interval  Endpoint Detection  Timed After Endpoint  Alarm  Set  Alarm Clear  Alarm Set  amp  Clear and Disabled  Press the Up Down arrows to cycle through  them     Manual   logs results on pressing the STO key    Timed Interval   logs on time and interval set by Data Log Interval in Hr Min Sec   Endpoint Detection   logs data when the endpoint is detected    Timed After Endpoint   logs at the Data Log Interval after the endpoint    Alarm Set   data 
18. ate is insufficient then the sample  should be stirred  e g  either by a gentle stirring motion with the probe or with a magnetic  stirrer      When using the probe in liquors  sludges or polymers a coating may be deposited on the  membrane  causing low response or drifting  This can be reduced by rinsing the probe in  deionised water after each test     Ensure the probe is rinsed in deionised water after each test     When measuring in mg l  ppm  results are pressure dependent and determinations carried  out atpressures other than 760mm Hg will need to be compensated     When measuring in mg l  ppm  results obtained from saline samples will need to be adjusted  for salinity     Ensure the probe is polarised prior to use  refer Operation   It is essential to perform this  procedure after replacement of the membrane or probe     When replacing the membrane ensure no air bubbles are trapped in the electrolyte fill solution  and that the membrane is not creased or damaged after fitting  If the probe response is  sluggish  or the readout is unstable after membrane replacement  clean the anode and  cathode  refer Section 4 2 Maintenance      7 950 050 REV A 04 03    3 2 Set Up Parameters    The following section details the set up modes available to the user on the main set up menu  screen        3 2 1 Instrument Setup    This option will allow the following parameters to be set        Exit allows the user to exit this menu    Language enables the selection of the appropriate lang
19. ats   Calibration reminder interval  1 to 999 hours    Alarm outputs  open collector and audible    Security code protected user data   Back lit 1 8 VGA monochrome LCD   English  French  German  Italian  Spanish  Portuguese  9V power supply   210x250x55mm   850g    1 950 050 REV A 04 03    Section 2  Installation  2 1 Unpacking  Remove the Model 9500 from the packaging and ensure the following items are included     Model 9500 Dissolved Oxygen Meter   Dissolved Oxygen probe with ATC  522 008    3 membranes  amp  KCI solution  522 019    Zero salts  983 030    Adaptor bush   Electrode holder   Power Supply  as specified at time of ordering the product   Condensed operating instructions  950 051    Operating Manual  950 050     OCOANOARWNH  gt     The electrode stand requires minimal assembly  refer to the diagram below   Any shortages or damage should be reported immediately to the manufacturer or your local  distributor     2 2 Installation    The Model 9500 is supplied ready to use  Connect the dissolved oxygen probe to the rear panel DIN  socket     The electrode stand requires minimal assembly  refer to the diagram below      Fig  2 2 1 Electrode Holder Assembly          2 950 050 REV A 04 03    Preparation for use    The probe is delivered with the membrane module detached  Prior to use the membrane module  must be fitted to the probe body    A protective cap is also supplied  This should be fitted to the probe when not in use  The sponge  contained within the cap should 
20. be wetted with deionised water prior to fitting  This covers the  membrane and keeps it wet  thus preventing the electrolyte filling solution from drying out     1     2     Remove the protective cap from the probe     Take a membrane module and hold in the vertical position  Fill with the electrolyte solution   5  KCI  supplied with the instrument     Whilst still holding the module upright  screw the probe slowly down onto the thread  allowing  the excess electrolyte to escape through the screw thread  Ensure no air bubbles are present  and that the membrane is not creased     Connect the probe to the 9500 and turn on the instrument     The complete probe and meter should be allowed to polarise for 30 minutes prior to  calibration     3 950 050 REV A 04 03    2 3 Displays       OST   29 0     ra bd   15 28 25  mails     ee    1  Primary display     4  digit  Provides direct readout in  DO    mg l or BOD of samples and  standards                    2  Mode annunciators     shows selected measurement mode   DO    mg l or BOD     3  Secondary display     31  digit display  Provides direct readout of automatic or manual  temperature  Also displays B O D  set up information when working in B O D  mode     4    Mode annunciators     indicates temperature in   C or   F and whether the measurements are  manually temperature compensated  MAN symbol      5  Endpoint symbol     this symbol is displayed when an endpoint has been detected    6  Mode tags     Each mode tag is highlighted
21. dpoint Audible Alarm    Endpoint Stability  Sec     3 2 10 Clock Set up        Endpoint Detection Enabled  Endpoint Audible Alarm Disabled  Endpoint Stability Sec  005       enables the user to exit the Set Up menu and return to the main  instrument display    can be enabled or disabled by using the Up Down keys which toggle  between the two settings    can be enabled or disabled by using the Up Down keys which toggle  between the two settings    can be set within the limits of 001 to 999 seconds     Select the Clock Setup sub menu by highlighting the option and pressing the STO key  The  following screens will be shown     Disnlay Time Display Time  IN WARNING Date Format DDIMMIYVY   i Adjusting time or date    may delete all stored results Time 13 29 42       Date 29 05 03       Exit    Display    Date Format    Time    Date    enables the user to exit the Setup menu and return to the main  instrument display    toggles between Time and Date  Selection of the preferred option  can be made using the Up Down keys which toggle between the two  settings    toggles between European  DD MM YY  and American  MM DD YY   formats  Selection of the preferred option can be made using the Up   Down keys which toggle between the two settings    allows time to be set  hrs min sec  using the Up Down   keys  The Left Right arrow keys allow forward and backward  movement along the row of digits  Press the STO key to accept the  setting    allows date to be set  in previously selected format 
22. e BOD process begins with the preparation of the samples and blanks  if required    Sample volume and total volume are both required to be entered by the user for every blank   sample     Exit  Exit  Sample Volume Vejml 0050 Sample Total Volume Vt ml 0150    Total Volume   t ml 0250 Total V 0250 0250     Setup           Once the Blank measurement is taken  the cursor will automatically move down to the sample  square  ready for the user to input the first sample measurement  If the user wishes to repeat the  blank measurement  the sample type must be changed to blank in the BOD set up menu  providing  24 hours has not elapsed from the previous measurement   The user can continue to input  samples and there respective volumes  until they have finished the batch  The batch does not  need to consist of 10 samples  the user can finish entering the day  0  samples at any time  If the  user forgets the blank  providing 24 hours has not passed since the first sample reading on day  0   the user can select blank for sample type and continue to store the blank day  0  reading     Once 24 hours has passed from the oldest day  0  reading  sample blank   the cursor will  automatically move to the right side to continue to store day  n  readings  After each day  n   reading is stored the user is given the chance to move onto the next sample number  If they do not   the BOD answer is shown on the measuring screen  minus endpoint icon   Once the user is on  the right hand side of the screen
23. forming measurements with the 9500 until a valid calibration is carried out     The user and batch ID can be used to identify sets of samples and a specific user  This information  is printed when data is output to the printer  The batch number is also stored in the results memory     3 7 1 Security    To control access to set up options and data manipulation functions a security code can be set  using this menu  When enabled a password is required to make any changes to the set up menus     When the code is ENABLED a user can measure a sample  log data and calibrate the unit  but  cannot change settings within the set up screens until a valid password is entered     27 950 050 REV A 04 03    Section 4  Maintenance  4 1 General    The Model 9500 is designed to give optimum performance with minimum maintenance  It is only  necessary to keep the external surfaces clean and free from dust  To give added protection when  the unit is not in use the unit should be disconnected from the mains supply and covered with the  optional dust cover  060 406   For longer term storage or re shipment it is recommended that the  unit be returned to the original packing case     4 2 Probe Care and Maintenance    Although the oxygen probe is supplied in a clean and tested condition it may  after some time   become sluggish or erratic due to contamination of the gold cathode  silver anode or membrane     a  Membrane Replacement    1  Hold the probe in a vertical position and carefully unscrew the 
24. ion 3 2 4  The alarm outputs will remain active until the alarm condition is no longer evident or the alarm limits  are reset in the Alarm Set up menu     For the Hi alarm output please use the Red 4mm phone connector   For the Low alarm output please use the Black 4mm Phono connector     For further information please contact your local distributor or the manufacturer     32 950 050 REV A 04 03    Section 7    Troubleshooting and functional checks    7 1 Troubleshooting  Fault    No display    Erratic display    Unable to calibrate  to 100     Permanently reading  zero    Unable to calibrate    Possible Cause    Check power supply    Check power supply    Calibrating to previously  set value    Membrane not fitted  Membrane damaged  Membrane not filled with KCI  Probe defective    Sensor tip may be tarnished  Membrane damaged    Erratic slow response Air bubbles inside the    Measurement errors    Will not print    membrane cap   Coating of membrane due to  use in oils or slurries  strong  solvents  acids or alkalis   Probe defective    Zero solution will absorb  oxygen if left open   Incorrect salinity   Calibration error  Contaminated solution  Incorrect standards   Incorrect barometric pressure    IrDA Connection broken  Paper out    Battery flat    Action    Check that correct 9V ac power supply  is connected and switched on     Unit must be used with supplied 9V  acpower supply  Usage of other units will  cause the 9500 not to operate     Manually reset value to 100    
25. k readings     To view a BOD answer straight away  after recording a day  n  reading  The user can choose no  when prompted to increment sample number  To move to the next sample for the day  n  reading  the user can change the sample number in the set up menu     Measuring Screen   Incubation Days    The Incubation period is only shown as 5 or 7 days  depending on the time elapsed between the  two readings  There is a 4 hour window either side of 5 days  120 hours   whereby if the time  elapsed is within this window  the number 5 is shown underneath BOD  the same applies for 7  days respectively   If the time elapsed is outside those boundaries  nothing is shown below BOD     In any given batch  if the time period since the first reading stored is greater than 24 hours the    flashing cursor will automatically move to the right highlighting the blank or sample square  depending on the selection of sample type within the BOD set up screen     16 950 050 REV A 04 03    Results Screen    The result screen shows all results for the current selected batch  If there is no saved result or  time elapsed data then         is displayed in the results place        Batch  amp  Sample ID stay the same unless the user increments the sample number when storing a  result or edits the sample or batch ID within the BOD set up menu     The sample and solution volume can be changed for every sample  the divisible is shown on the  results screen  The value shown in the set up screen for the samp
26. le and solution volume is that of  current reading  and doesn   t relate to previous values used for previous readings  The exact values  used for the sample and solution volume for each sample are shown on the BOD result printout        Results Screen   Deleting    Selection within the set up menu allows the user to delete one  current  or all batches  By deleting  all this resets the sample and batch ID to 1  By choosing to delete one batch  current   batch ID  number stays the same  but the sample ID number reverts back to 1        17 950 050 REV A 04 03    BOD Set Up Parameters    The following section details the set up modes available to the user on the BOD set up menu  screen        BOD Batch ID    This option allows the user to enter a 2 digit batch ID number from 01 to 20  Enter the value using  the keypad  Press the STO key to accept the value        Sample ID    This option allows the user to enter a 2 digit sample ID number from 01 to 10  Enter the value using  the keypad  Press the STO key to accept the value        18 950 050 REV A 04 03    Sample Type    This option toggles between Sample and Blank           Cir Key    This option toggles between Delete All Batches and Delete Current Batch           Volumes       Sample value  Ve  ml allows the user to set the required sample Ve   Total volume  Vt  allows the user to set the required total volume Vt           Clock Setup       Refer Section 3 2 10 Set Up Parameters    19 950 050 REV A 04 03    BOD Process    Th
27. logs when the alarm set point is reached    Alarm Clear   data logs when the alarm is cleared    Alarm Set  amp  Clear   data logs at the alarm set point and when the alarm condition clears   Disabled   no data logging is possible when this option is selected     The 9500 can log to either the memory or the printer  The memory can hold 250 data points which  can be accessed via the Results screen  plus 20 BOD batches  10 samples per batch      When the memory is full there are two options available regarding any additional logging  The  default is STOP     Stop   when the memory is full no further data logging can occur until some  locations are deleted     Overwrite overwrites data from the earlier storage point   To prevent the accidental erasure of data the  Prompt Before Deleting  option can be either Enabled  or Disabled     The function of the CLR key can also be set in this menu  The options define the use of the delete  key     25 950 050 REV A 04 03    Disabled   no manual deleting of results is possible     Deletes Results Before   deletes all results before the selected data point   Delete Results Since   deletes all results since the selected data points   Deletes All Results   deletes all stored results     3 12 1 Accessing Stored Results    To access results which have been stored use the Right arrow key to select results    The most recently stored results will be displayed on the screen  If more results are stored than  can fit on a single screen then a series
28. membrane module    2  Take the new membrane module and holding it in a vertical position  fill with O  electrolyte   5  KCl     3  Still holding the module in a vertical position  screw the probe slowly down onto the thread   allowing excess electrolyte to escape through the screw thread  Ensure no air bubbles are  present and the membrane is not creased     b  Cathode and Anode Cleaning    The gold cathode tip can be re polished using a fine abrasive     crocus paper     material  Lay the  abrasive sheet on a flat surface  hold the probe in a vertical position and gently polish by moving the  tip over the sheet in a circular motion     A toothbrush dipped in diluted ammonia solution will remove any deposits from the silver anode   Rinse with deionised water prior to re assembly  If the probe is not to be used for 24 hours  store  with the protective sheath fitted to prevent the electrolyte from drying out due to evaporation through  the membrane  which is porous to water vapour as well as oxygen  If the probe is disconnected  from the unit or anew membrane has been fitted  it will be necessary to allow the probe to polarise  before stable readings can be obtained    Polarisation will normally be achieved within 30 minutes     28 950 050 REV A 04 03    Section 5    Optional Accessories    5 1 Optional Accessories    The following list of items are available for use with the Model 9500     552 050  983 030  522 019  060 406    037 701    037 702  050 002  037 801    Stirrers  
29. pression and death  due to liberation of sulphur dioxide  Irritating to skin  eyes  and respiratory system  Used in controlled quantities as a food preservative and antioxidant     FIRST AID   Eyes Irrigate thoroughly with water for at least 10 minutes  OBTAIN MEDICAL ATTENTION    Lungs Remove from exposure  rest and keep warm  In severe cases OBTAIN MEDICAL  ATTENTION    Skin Wash off thoroughly with water  Remove contaminated clothing and wash before re   use  In severe cases OBTAIN MEDICAL ATTENTION    Mouth Wash out mouth thoroughly with water and give plenty to drink  In severe cases    OBTAIN MEDICAL ATTENTION     34 950 050 REV A 04 03    EC Declaration of Conformity    JENWAY Model 9500 Dissolved Oxygen Meter complies with the following European Standards   EN 50081 1 1992 Electromagnetic compatibility   Generic emission standard    EN 50082 1 1992 Electromagnetic compatibility   Generic immunity standard  Performance crite   rion B     EN 61010 1 2001 Safety requirements for electrical equipment for measurement  control and  laboratory use    Following the provision of     EMC Directive   89 336 EEC and Low Voltage Directive   73 23 EEC    Martyn J  Fall   Managing Director  Jenway   Gransmore Green  Felsted  Dunmow   Essex  CM6 3LB  England    35 950 050 REV A 04 03    
30. rrow keys allow forward and backward movement along the  row of digits  Press the STO key to accept the value    10 950 050 REV A 04 03    Batch ID up to a 3 digit code can be set by using the Up Down keys  The Left   Right arrow keys allow forward and backward movement along the  row of digits  Press the STO key to accept the value    3 2 6 Security Setup        Select the Security Setup sub menu buy highlighting the option and pressing the STO key  The  following menu will be shown     Data Entry Security Disabled    Security Code 660       Setup  moe J Status   Results    Exit enables the user to exit the Set Up menu and return to the previous  set up screen    Data Entry Security can be enabled or disabled by using the Up Down keys which toggle  between the two settings    Security Code up to a 3 digit code can be set by using the Up Down keys  The Left     Right arrow keys allow forward and backward movement along the  row of digits  Press the STO key to accept the value     3 2 7 Data Logging Set up        Data Log Ewent Manual  Data Log To Memory  Data Log Interval 00 01 00  Memory Full Stop  Prompt Before Deleting Disabled  Cir Key Disabled    Printer Setup       Setup       Exit enables the user to exit the Set Up menu and return to the main  set up screen    Data Log Event can be enabled or disabled by using the Up Down keys which toggle  between the two settings    Data Log To allows the data to be sent to Memory or to the external Printer     Selection is made via
31. uage     English  French   German  Italian  Spanish or Portuguese  Use the Up Down keys to scroll  through the language options    LCD Brightness     enter the value using the keypad  Press the STO key to accept the value        8 950 050 REV A 04 03    3 2 2 Temperature Setup       Select the Temperature Setup sub menu by highlighting the option and pressing the STO key  The  following menu will be shown     Exit  Temp Units    c  Manual Temperaturel   C   025 0       Temp units allows selection of the preferred unit of measurement  either   C or   F  using  the Up Down keys which toggle between the two units     Manual temperature allows the manual temperature value to be set  Press the STO key to  accept the value     3 2 3 Calibration Set up        This option allows the following parameters to be set     Calibration  D 02 100 0  Manual Pressure bar  1 013  Manual Salinity  gfl  00       Exit enables the user to exit the Setup menu and return to the main set up  screen   Calibration  DO  allows the setting of  DO  value for calibration between the limits of 0 1 to    199 9 using the Up Down arrow keys     Manual Pressure  bar  allows the setting of the current barometric pressure between the limits of  0 1 to 6 0bar using the Up Down keys     Manual Salinity  g l  allows the setting of salinity values between the limits of 0 0 to 75 0     9 950 050 REV A 04 03    3 2 4 Alarms Setup       Exit   Alarm Outputs   Audible Alarm Warning   DO  Alarm High   DO  Alarm Low    mg l
32. xygen measurements are made at the  beginning and end of a five day incubation period for the standard five day B O D test     The dissolved oxygen probe membrane is not affected by intefering ions  the pH of the sample   sludge  clays or inorganic matter  This is due to the electrolytic cell being separated from the  sample by a selectively permeable membrane through which only gases can pass  The potential  within the electrolytic cell is adjusted so that only the oxygen is reduced at the cathode  giving rise  to a current flow proportional to the partial pressure of oxygen in the sample     B O D  Measuring Screen    The main BOD screen serves 2 purposes    1  to display the measuring screen and current sample number selected    2  to display the BOD answer if a measurement has been taken    Note  In this mode the padlock and IrDA icon are not displayed  even when enabled       12 58      BatchNo   1 SampleNo   1  Blank  0   12 77 Blank  nj  0 19  Sample 0   12 77 Sample nj  0 19  Day n    Day 0   17 days 22 hrs         lf there is a BOD answer  it is displayed on the measuring screen  To identify when the measuring  screen is displaying a BOD answer  BOD is shown under the units of measure  Below the BOD  caption is the incubation period in days providing the BOD conditions have been met     Shown at the bottom of the results screen     Day  n      Day  0     relates to the time elapsed for day  n   and day  0  of the current sample reading  it does not relate to the blan
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Bogo LifeStyle 10QC 16GB Black  取扱説明書 - ヒロセ電機  取扱説明書 - NTTドコモ  Unity™ for Vari-Flex  USER`S MANUAL  brp 取扱説明書  USER MANUAL - Silver Pines Sound & Light  Sony SDM-HS73 17" Flat Panel LCD Monitor      Copyright © All rights reserved. 
   Failed to retrieve file