Home

Model 43i - User Manual

image

Contents

1. 7 28 Figure 7 14 Replacing the Flow Transducer ooococcocicinoncnnincncnncnrncnonncnrnnono 7 31 Figure 7 15 Replacing the Heater Assembly oo cece ce ceteeeteteeseetenenees 7 34 Figure 7 16 Replacing the Kicker Assembly Standard Kicker 7 35 Figure 7 17 Replacing the Kicker Assembly Optional Heated Kicker 7 36 Figure 7 18 Replacing the Thermistor 7 37 Figure 7 19 Replacing the Input Board ooocccocicicoccccinnococinrnononocnrocornnnononoso 7 39 Figure 7 20 Replacing the I O Expansion Board Optional 7 41 Figure 7 21 Rear Panel Board CONNECTOFS oocicconicccnnoconicccnnncononocnnnncnnnoos 7 41 Figure 7 22 Replacing the Measurement Interface Board 7 44 Figure 7 23 Replacing the Front Panel Board and the LCD Module 7 45 Figure 8 1 Hardware COMPONENTS oocoocicococococinccncncononnncnconononocnnnnrnonocnrnncnnnoos 8 2 Figure 9 1 Internal Permeation Span Source Flow Diagram 9 2 Figure 9 2 Cal Oven Therm Resistor Sereen 9 5 Figure 9 3 Cal Gas Therm Bath Screen oooooociciociciccincooincoconcnncnconononocnroncnnnnos 9 6 Figure 9 4 Cal Gas Therm Resistor SCTeeN oocooocinciciocincnccnnocinonncnnnornnonocnnnnos 9 7 Figure 9 5 Rack Mount Option Assembly o oooocoonicccnionnconinccnincnnoninninooosoo 9 12 Figure 9 6 Bench MOUNtING ocooconinocononn
2. The Flow Sensor Cal screen appears Note If Service Mode is not displayed refer to Accessing the Service Mode in this chapter then return to the beginning of this step A 10 At the Flow Sensor Cal screen press to select Zero The Calibrate Flow Zero screen appears Wait at least 10 seconds for the zero reading to stabilize then press to save the zero flow value Reconnect the pump cable to the AC PUMP connector on the measurement interface board Connect a calibrated flow sensor at the SAMPLE bulkhead on the rear panel Press to return to the Flow Sensor Cal screen At the Flow Sensor Cal screen press to select Span The Calibrate Flow Span screen appears Wait at least 10 seconds for the reading to stabilize use gt and _ J _4_ to enter the flow sensor reading and press to save the value Thermo Fisher Scientific Heater Assembly Replacement Thermo Fisher Scientific A Servicing Heater Assembly Replacement 11 Install the cover Use the following procedure to replace the heater assembly Figure 7 15 Equipment Required Heater assembly Heat sink grease Philips screwdriver Allen wrench 5 32 inch Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to
3. C 5 0x03 0x04 Read Holding Registers Read Input Registers C 5 0x05 Force Write Single Collie lis C 7 MODBUS Commands Supported i cmioiaridatnd inicia C 8 Geysitech Bayern Hessen Protocol s sssssssssessesseeseessessessesseeseeneenss D 1 Serial Communication Parameters usas D 1 TCP Communication Parameters esla il D 2 ii A ANGAR D 2 Abbreviations Usd Os D 2 Basic Command eds D 2 Block Checksum lt BCO gt 5 As D 3 Geysttech Commandos D 3 Model 43 Instruction Manual xix Contents xx Model 43 7 Instruction Manual Instrument Control Command ST ccccccccssccsseessssscssessssecesssseess D 3 Data Sampling Data Query Command DA eccoonicninocnicnonncnnnconnannns D 4 Measurements reported in response to DA command cee D 6 Single Range Mode iii rd D 6 Diral Auto Range Modernas ina D 6 Operating and Error Stat e iia D 7 Thermo Fisher Scientific Thermo Fisher Scientific Figures Figure 1 1 Model 431 Flow SchematiC ooooicoonicconinconincnninconinccnrnccnrnccnrnncnrnoos 1 3 Figure 2 1 Remove the Packing Material oooooonicnocoicicicinocincinincnnnncnrnnnnnoooso 2 2 Figure 2 2 Removing the Shipping Screws ooconcccicccncnccninconinccnrnccnrnccornncnonoss 2 2 Figure 2 3 Rear Panel ooooocococoiciccccccccnncncoconnocnccncnrnnnonorocorornnnonorornnnonocorannanooso 2 4 Figure 2 4 Atmospheric Dump Bypass Plumbing ccconininicncnininnonooo oo 2 5 Figure 2 5
4. Figure 3 7 Pin Out of Rear Panel Connector in Auto Range Mode Table 3 4 Default Analog Outputs in Auto Range Mode Channel Connector Pin 1 0 Terminal Pin Description 1 14 1 SO Analog Output 2 33 3 SO Status Output half scale high range zero scale low range 3 15 5 None 4 34 7 None 5 17 9 None 6 36 11 None Ground 16 18 19 35 37 2 4 6 8 10 12 Signal Ground Note All channels are user definable If any customization has been made to the analog output configuration the default selections my not apply A Gas Units The Gas Units screen defines how the SO concentration readings are expressed Gas units of parts per billion ppb or micrograms per cubic meter g m are available The pg m gas concentration is calculated using a standard pressure of 760 mmHg and a standard temperature of 20 s 3 12 Model 43i Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific S0 Range Operation Range Menu When switching the selected units from ppb or ppm to pg m or mg m the analog ranges all default to the highest range in that mode For example when switching from mg m to ppm all the ranges default to 10 Therefore whenever you change units you should also check the range settings In the Main Menu choose Range gt Gas Units Note If the units change from ppb or ppm to g m or mg m or vice versa the instrument should be re calibrated particularly if the user s
5. 1 From the Main Menu press _4_ to scroll to Service press gt to scroll to Analog Input Calibration and press e The Analog Input Cal screen displays Note If Service Mode is not displayed refer to Accessing the Service Mode in this chapter then return to the beginning of this step A 2 At the Analog Input Cal screen press _4_ to scroll to a channel and be 3 With the cursor at Calibrate Zero press e The screen displays the input voltage for the selected channel 4 Make sure that nothing is connected to the channel input pins and press to calibrate the input voltage on the selected channel to zero volts Model 43i Instruction Manual 7 25 Servicing Analog Input Calibration 7 26 Calibrating the Input Channels to Full Scale Model 43i Instruction Manual The screen displays 0 00 V as the voltage setting 5 Press Cm gt Cm to return to the Analog Input Cal screen and repeat Steps 2 through 4 to calibrate other input channels to zero as necessary 6 Continue with the Calibrating the Input Channels to Full Scale procedure that follows Use the following procedure to calibrate the input channels to full scale by applying a known voltage to the channels Equipment Required DC voltage source greater than O volts and less than 10 volts 1 Connect the known DC voltage source to the input channel 1 8 to be calibrated Figure 7 12 shows the analog output pins and Table 7 4 i
6. The calculation begins by flashing the lamp 10 times per second After the lamp flash a sample and hold circuit on the input board is used to sample the PMT output The raw accumulated counts are scaled according to the gain setting of the input board Next the uncorrected values are determined according to a unique averaging algorithm which minimizes errors resulting from rapidly changing gas concentrations This algorithm results in values which are stored in RAM in a circular buffer that holds all the ten second data from the previous five minutes This data is averaged over the selected time interval which can be any multiple of ten between 1 and 300 seconds The background values which are corrected for temperature are subtracted from their respective averages The reading is corrected by the stored span factor and by the temperature factor The front panel display serial and Ethernet data ports and analog outputs are the means of communicating the results of the above calculations The front panel display presents the SO concentrations The display is updated every 1 10 seconds depending on the averaging time The analog output ranges are user selectable via software The analog outputs are defaulted based on the measurement range The defaults are Model 43i Instruction Manual 8 5 System Description Electronics Electronics Motherboard External Connectors Internal Connectors 8 6 Model 43i Instruction Manual
7. 10 through 2 from left to right In the last data byte the status of outputs 15 11 is shown as the byte value 0x0A or binary 0000 1010 Output 15 is in the fifth bit position from the left and output 11 is the LSB of this byte The four remaining high order bits are zero filled Read holding input registers reads the measurement data from the instrument Issuing either of these function codes will generate the same response These functions read the contents of one or more contiguous registers These registers are 16 bits each and are organized as shown below All of the values are reported as 32 bit IEEE standard 754 floating point format This uses 2 sequential registers least significant 16 bits first The request specifies the starting register address and the number of registers Registers are addressed starting at zero Therefore registers numbered 1 16 are addressed as 0 15 The register data in the response message are packed as two bytes per register with the binary contents right justified within each byte For each register the first byte contains the high order bits and the second contains the low order bits The status of outputs 2 10 is shown as the byte value 0xCD or binary 1100 1101 Output 10 is the MSB of this byte and output 2 is the LSB By convention bits within a byte are shown with the MSB to the left and the LSB to the right Thus the outputs in the first byte are 10 through 2 from left to right I
8. 5V Ground Measurement Frequency Output Sample Hold NC GAIN A GAIN B HVPS MJ9 YN 0 0 A OO N HV Power Supply Voltage Adjust Ground HV Power Supply On Off Ground HV Power Supply Voltage Monitor Ground Ground PRES MJ10 gt a WO N Pressure Sensor Input Ground 15V 15V FLOW MJ11 on A WO N Flow Sensor Input Ground 15V 15V Ground LED MJ14 LED Drive LED Supply AMB TEMP MJ15 Ambient Temperature Thermistor Model 431 Instruction Manual 6 17 Troubleshooting Connector Pin Descriptions 6 18 Model 43i Instruction Manual Connector Label Reference Designator Pin Signal Description Ground BENCH HEATER NTJ1 FP WO N Bench Temperature Input Ground AC HOT Bench Heater AC CONVERTER MJ17 gt o oO A 0 N Converter Temperature Input Ground Converter Heater On Off Ground 15V 15V 24V IN 24V Ground AC PUMP PJ5 os w N AC HOT AC NEUT AC Ground FAN PJ4 24V Ground AC IN PJ3 y w N AC HOT AC NEUT AC Ground AC CONVERTER PJ12 AC HOT AC NEUT AC Ground Z S SOL PJ7 24V Zero Span Solenoid Control SAMPLE SOL 24V Sample Solenoid Control CONV SOL 24V Spare 1 Solenoid Control SPAN2 SOL 24V Spare 2 Solenoid Control PERM OVEN THERM Perm Oven Gas Ther
9. will be read and inserted in the text field of the line This is followed by a possible string enclosed in quotes This is used to place a string into the value field The value source which is the item or word number in the DATA ERec response appears next This is followed by an optional bitfield designator The datum identified by the value source can be printed as a string s hexadecimal x decimal d or floating point or binary b number Typically bitfield extractions are only done for decimal or hexadecimal numbers Floating point numbers can be followed with an optional precision specifier which will be used as an argument to printf s f format e g a field of 4 would be translated into the printf command of 3f Alternately the special character causes an indirection on the precision specifier which now becomes a field number can precede the precision specifier this This is useful when formatting for example numbers which have varying precision depending on the mode of the instrument Binary numbers can also have an optional precision specifier which is used to determine how many bits to print For example the specifier b4 will print the lowest four bits of the parsed number There are serious restrictions on where an s field may appear currently sources 1 and 2 must be s and no others may be s The value source is followed by optional alarm information indicated by a commerci
10. Operation The commands can be sent in either uppercase or lowercase characters Each command must begin with the proper instrument identification number ASCII character The command in the example that follows begins with the ASCII character code 171 decimal which directs the command to the Model 43 and is terminated by a carriage return CR ASCII character code 13 decimal If an incorrect command is sent a bad command message will be received The example that follows sends the incorrect command set unit ppm instead of the correct command set gas unit ppm Send set unit ppm Receive set unit ppm bad cmd The save and set save params commands stores parameters in FLASH It is important that each time instrument parameters are changed that this command be sent If changes are not saved they will be lost in the event of a power failure Streaming data is sent out the serial port or the Ethernet port on a user defined periodic basis Streaming data over Ethernet is only generated when a connection is made on TCP port 9881 When interfacing to an instrument via C link commands always enter the concentration values in ppb or 1g m units For example to set a background value to 20 ppm enter 20000 ppb as the value for the set background command Table B 1 lists the 437 C Link protocol commands The interface will respond to the associated command strings Thermo Fisher Scientif
11. Operation Instrument Controls Menu Descriptor The Descriptor screen allows the user to enter the descriptor for the selected analog input channel The descriptor is used in datalogging and streaming data to report what data is being sent out The descriptor may be from 1 to 3 characters in length and defaults to IN1to IN8 user input channel number In the Main Menu choose Instrument Controls gt I O Configuration gt Analog Input Config gt Select Channel gt Descriptor EUs EEE TE Sanat Bone nr Units The Units screen allows the user to enter the units for the selected analog input channel The units are displayed on the diagnostic screen and in datalogging and streaming data The units may be from 1 to 3 characters in length and defaults to V volts In the Main Menu choose Instrument Controls gt I O Configuration gt Analog Input Config gt Select Channel gt Units Thermo Fisher Scientific Model 43i Instruction Manual 3 49 Operation Instrument Controls Menu Decimal Places The Decimal Places screen allows the user to select how many digits are displayed to the right of the decimal from 0 to 6 with a default of 2 In the Main Menu choose Instrument Controls gt I O Configuration gt Analog Input Config gt Select Channel gt Decimal Places Number of Table Points The Number of Table Points screen allows the user to select how many points are used in the analog input conversion table Th
12. 102646 00 Cable DB37F to Open End Six Feet optional 102659 00 Cable DB25M to Open End Six Feet optional 6219 Cable RS 232 optional 102888 00 Terminal Board PCB Assembly DB37F standard with all instruments 102891 00 Terminal Board PCB Assembly DB37M standard with all instruments 103084 00 Terminal Board PCB Assembly DB25M included with optional 1 0 Expansion Board in all instruments Model 43i Instruction Manual 7 5 Servicing Removing the Measurement Bench and Lowering the Partition Panel Flash Intensity Assy Reaction Chamber Assy gt P Mirror Housing Assy PMT Assy Chamber Inlet Assy Flash Lamp Housing Flow Transducer Permeation Oven Assy Opt Pressure Transducer Thermistor Measurement Interface Board LCD Displays y IZ Front Panel Board p 3 Ae gt Removing the Measurement Bench and Lowering the Partition Panel 7 6 Model 43i Instruction Manual Lf z F 2 o I Flash Trigger Assy E i a ey OA S y Kicker Assy s HAR ge at gt Fan e 5 p y tos D y le wa KS ZA Pump le ES 5 Solenoid Valves Opt An e 7 g 1 0 Expansion Board Opt els mo p i Qe eq NI Digital Output Board E 3 Power Supply HVPS Input Board Assy Optional Transformer Mother Board not shown Figure 7 2 Model 43i Component Layout The measurement bench can be removed and the pa
13. Min and Max Internal Temperature Limits Chamber Temperature Min and Max Chamber Temperature Limits 3 64 Model 43i Instruction Manual The Minimum Internal Temperature alarm limit screen is used to change the minimum internal temperature alarm limit The minimum and maximum internal temperature screens function the same way In the Main Menu choose Alarms gt Internal Temp gt Min or Max The Chamber Temperature screen displays the current chamber temperature and sets the minimum and maximum alarm limits The chamber temperature is regulated at 45 0 degrees C Acceptable alarm limits range from 43 to 47 C If the chamber temperature reading goes beyond either the minimum or maximum limit an alarm is activated The word alarm appears in the Run screen and in the Main Menu In the Main Menu choose Alarms gt Chamber Temp The Minimum Chamber Temperature alarm limit screen is used to change the minimum chamber temperature alarm limit The minimum and maximum internal temperature screens function the same way In the Main Menu choose Alarms gt Chamber Temp gt Min or Max Thermo Fisher Scientific Operation Alarms Menu The Pressure screen displays the current reaction chamber pressure reading Pressure oh ao ie and sets the minimum and maximum alarm limits Acceptable alarm limits range from 400 to 1000 mmHg If the pressure reading goes beyond either the minimum or maximum limit an alarm is activated The
14. The command terminator determines the presence or absence of lt BCC gt If a command is terminated by lt ETX gt then the next two characters are the checksum if the command is terminated with lt CR gt no checksum is attached The block checksum is represented by two characters which represent a 2 digit hex number 1byte e g 1 byte 0xAB hex checksum will be represented by the two characters A amp B The checksum is referred to as lt BCC gt throughout this document The following commands are supported by the Geysitech protocol e Instrument Control Command ST e Data Sampling Data Query Command DA There are three control commands supported by the Geysitech protocol This lt control commands is a single letter which triggers an action in the instrument These commands are active only when service mode is inactive and the zero span option is present Command N switches the instrument gas mode to Zero mode Command K switches the instrument gas mode to Span mode Command M switches the instrument gas mode to Sample mode The following are the different acceptable formats of the ST command lt STX gt ST lt address gt lt control command gt lt ETX gt lt BCC gt OR lt STX gt ST lt address gt lt control command gt lt CR gt OR lt STX gt ST lt address gt lt SP gt lt control command gt lt CR gt OR lt STX gt ST lt address gt lt SP gt lt control command gt lt ETX gt lt BCC g
15. The input board accepts the current signal from the PMT and converts it to a voltage which is scaled by a factor of approximately 1 10 or 100 depending on the full scale range of the SO channel The scaled voltage signal is converted to a frequency and sent to the microprocessor The digital output board connects to the motherboard and provides solenoid driver outputs and relay contact outputs to a connector located on the rear panel of the instrument Ten relay contacts normally open with power off are provided which are electrically isolated from each other Eight solenoid driver outputs open collector are provided along with a corresponding 24VDC supply pin on the connector The front panel connector board interfaces between the motherboard and the front panel mounted function key panel and Graphics display It serves as central location to tie the three connectors required for the function key panel the graphics display control lines and the graphics display backlight to a single ribbon cable extending back to the motherboard This board also includes signal buffers for the graphics display control signals and a high voltage power supply for the graphics display backlight Thermo Fisher Scientific Flash Trigger Board Flash Intensity Board 1 0 Expansion Board Optional 1 0 Components Analog Voltage Outputs Thermo Fisher Scientific System Description 1 0 Components The flash trigger board is located in the bas
16. This can be set via the front panel The Instrument Address is represented by lt address gt in the examples throughout this document Note Device IDs 128 through 247 are not supported because of limitations imposed by the C Link protocol A The following is a list of abbreviations used in this document lt CR gt is abbreviation for Carriage Return ASCII code 0x0D lt STX gt is abbreviation for Start of Text ASCII code 0x02 lt ETX gt is abbreviation for End of Text ASCII code 0x03 lt SP gt is abbreviation for space ASCII code 0x20 The following is the basic structure of a Geysitech command lt STX gt Command text lt ETX gt lt BCC gt OR lt STX gt Command text lt CR gt Each Command is framed by control characters lt STX gt at the start and terminated with either lt ETX gt or lt CR gt If a command is terminated with lt ETX gt then additional two characters lt BCC gt is attached after lt ETX gt this is the block checksum Thermo Fisher Scientific Block Checksum lt BCC gt Geysitech Commands Instrument Control Command Thermo Fisher Scientific ST Geysitech Bayern Hessen Protocol Block Checksum lt BCC gt The block checksum is calculated beginning with a seed value of 00000000 binary 0x00 and bitwise exclusive ORing with each of the characters of the command string or response including the framing characters lt STX gt and lt ETX gt The checksum works as an error check
17. at this point pressing indicates that these are proposed changes as opposed to implemented changes To change the selected record format and erase record log file data see Commit Content that follows In the Main Menu choose Instrument Controls gt Datalogging Settings gt Select Content gt Select Field gt Other Measurements Analog Inputs The Analog Inputs screen allows the user to select the output signal none or analog inputs 1 8 that is tied to the selected field item The selected item is shown by lt after it Note that at this point pressing indicates that these are proposed changes as opposed to implemented changes To change the selected record format and erase record log file data see Commit Content that follows In the Main Menu choose Instrument Controls gt Datalogging Settings gt Select Content gt Select Field gt Analog Inputs 3 28 Model 43i Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu Commit Content The Commit Content screen is used to save the selected output signal that is tied to the selected field item If no changes have been made NO CHANGES TO RECORD LIST will appear For more information about selecting the analog output signal group choices see Commit Content in this chapter In the Main Menu choose Instrument Controls gt Datalogging Settings gt Select Content Reset to Default Content The Reset to Default Con
18. calculated by dividing the data values by the full scale range for each of the three parameters and then multiplying each result by the user selected output range Negative concentrations can be represented as long as they are within 5 of full scale The zero and span values may be set by the user to any desired value All electronics operate from a universal switching supply which is capable of auto sensing the input voltage and working over the entire operating range The internal pump and heaters all operate on 110VAC An optional transformer is required if operating on the 210 250VAC or 90 110VAC ranges An on off switch controls all power to the instrument and is accessible on the front panel The motherboard contains the main processor power supplies a sub P processor and serves as the communication hub for the instrument The motherboard receives operator inputs from the front panel mounted function key panel and or over I O connections on the rear panel and sends commands to the other boards to control the functions of the instrument and to collect measurement and diagnostic information The motherboard outputs instrument status and measurement data to the front panel mounted graphics display and to the rear panel I O The motherboard also contains I O circuitry and the associated connector to monitor external digital status lines and to output analog voltages that represent the measurement data Connectors located on
19. er Returns a brief description of the main operating conditions in B 17 the format specified in the commands erec Returns a brief description of the main operating conditions in B 18 the format specified in the command erec format Reports sets erec format ASCII or binary B 19 erec layout Reports current layout of erec data B 20 flags Reports 8 hexadecimal digits or flags that represent the B 11 status of the ozonator PMT gas mode and alarms flow Reports current measured sample flow in LPM B 9 format Reports sets current reply termination format B 37 gas mode Reports current mode of sample zero or span B 30 gas unit Reports sets current gas units B 31 he help Simulates pressing help pushbutton B 26 high avg time Reports sets high range averaging time B 8 high range Reports selects current SO high range B 28 high so2 Reports SO concentration calculated with high range B 8 coefficients high so2 coef Reports sets high range SO coefficients B 25 high sp conc Reports sets high span concentration B 25 host name Reports sets host name string B 37 instr name Reports instrument name B 37 instrument id Reports sets instrument id B 38 internal temp Reports current internal instrument temperature B 9 isc iscreen Retrieves framebuffer data used for the display B 26 lamp set Sets the flash lamp ON or OFF B 33 lamp setpoint Reports sets flash lamp setpoint B 15 lamp status Reports sets flash lamp status on or off B 33 lamp int Reports the current
20. for data that is sent out in response to the erec lrec srec and related commands The example that follows shows a typical response For details on how to interpret the strings see Record Layout Definition later in this appendix Send lrec layout Receive lrec layout s s lx f f f f f f t D L TERETE so2 pmtv lmpv ain4 ain5 lmpi lrec mem size srec mem size These commands report the long and short records that can be stored with the current settings and the number of blocks reserved for long and short records To calculate the number of short records per block add 2 to the number of records and then divide by the number of blocks The example that follows shows that 7 blocks were reserved for long records and the maximum number of long records that can be stored in memory is 1426 Send lrec mem size Receive lrec mem size 1426 recs 7 blocks lrec per srec per These commands report the long and short records logging period The example that follows shows that the short record logging period is 5 minutes Send srec per Receive srec per 5 min Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Datalogging set srec per value set srec per value value 1151151301601 These commands set the long and short records logging period to value in minutes The example that follows sets the long record logging period to 15 minutes Send set lrec per 15 Receive set lrec per 15
21. gt Bit 8 7 6 5 4 3 2 1 gt Hex value 80 40 20 10 08 04 02 01 MSB LSB Operating status Service Mode On 0 0 0 0 0 10 0 l1 Maintenance Local 0 10 0 10 0 10 1 0 Zero gas On 0 10 0 10 0 1 0 10 Span gas On 0 10 0 10 1 0 0 10 Gas Unit Indication ppb OR ppm 10 10 0 1 o 0 0 0 Test LED On 0 0 1 10 0 0 0 0 Flash Lamp Off 0 1 0 0 0 0 0 0 Not used 1 10 0 0 0 0 0 0 Table D 2 Error Status for Model 437 D7 D6 D5 D4 D3 D2 D1 DO gt Bit 8 7 6 5 4 3 4 gt Hex value 80 40 20 10 08 04 02 01 MSB LSB Error status Internal Temperature Alarm 0 0 0 0 0 0 0 1 Reaction Chamber Temperature o 0 0 0 0 10 1 0 Alarm Lamp Intensity Alarm 0 10 0 10 0 1 0 10 Lamp Voltage Alarm 0 10 0 10 1 0 0 10 Pressure Alarm 0 0 0 1 0 0 0 10 Sample Flow Alarm o 10 1 10 o 10 0 0 Permeation Gas Temperature 0 1 0 0 0 10 o 0 Alarm Not used 1 0 0 0 0 0 0 0 Thermo Fisher Scientific Model 431 Instruction Manual D 7
22. set analog iout range 4 1 ok Table B 10 Analog Current Output Range Values Range Output Range 1 0 20 mA 2 4 20 mA 0 cannot be set to this but may report Undefined analog vin channel This command retrieves the analog voltage input channel data both the calculated value and the actual voltage In the example that follows the calculated value of channel 1 is 75 325 degrees F volts are 2 796 This command responds with feature not enabled if the I O expansion board is not detected Send analog vin 1 Receive analog vin 1 75 325 2 796 analog vout range channel This command reports the analog voltage output channel range where channel is between 1 and 6 inclusive according to Table B 11 Send analog vout range 2 Receive analog vout range 2 3 Model 43 Instruction Manual B 39 C Link Protocol Commands 1 0 Configuration B 40 Model 43 Instruction Manual set analog vout range channel range This command sets analog voltage output channel to the range where channel is between 1 and 6 inclusive and range is set according to Table B 11 The example that follows sets channel 2 to the 0 10 V range Send set analog vout range 2 3 Receive set analog vout range 2 3 ok Table B 11 Analog Voltage Output Range Values Range Output Range 1 0 1 V 2 0 100 mV 3 0 10 V 4 0 5 V 0 cannot be set to this but may report Undefined dig in This command reports the status of the digital
23. to scroll to Analog Out Calibration and press The Analog Output Cal screen displays Note If Service Mode is not displayed refer to Accessing the Service Mode in this chapter then return to the beginning of this step A 3 At the Analog Output Cal screen press _4_ to scroll to the voltage channel or current channel corresponding to the rear panel terminal pins where the meter is connected then press 4 4 With the cursor at Calibrate Zero press e The Analog Output Cal line displays Zero Thermo Fisher Scientific Analog Input Calibration Calibrating the Input Channels to Zero Volts Thermo Fisher Scientific Servicing Analog Input Calibration Note When calibrating the analog output always calibrate zero first and then calibrate full scale A 5 Use until the meter reads the value shown in the Set Output To line then press to save the value 6 Press Cm to return to the previous screen 7 Press to select Calibrate Full Scale 8 Use until the meter reads the value shown in the Set Output To line then press to save the value Use the following procedures to calibrate the analog inputs after replacing the optional I O expansion board These procedures include selecting analog input channels calibrating them to zero volts and then calibrating them to full scale using a known voltage source Use the following procedure to calibrate the input channels to zero volts
24. 1 0 Terminal Board Views ococoociccccioconicccnnnconinccnrnccnrnccccornncnoos 2 6 Figure 2 6 D O Terminal Board ViewS oooooociccociconinconnnconinccnrnccnrnccnrnccncnoos 2 7 Figure 2 7 25 Pin Terminal Board ViQWS c ccccceccccscetesesesteteseseeteteseseseenees 2 8 Figure 3 1 Front Panel Display ooonocooconnccnnnonocncnrnnoncnonrnnonnororonoonoooso 3 2 Figure 3 2 Front Panel PUSADUTLONS o oocioonoccicicincncnnnnononncnrnononocnrnrnononononoso 3 3 Figure 3 3 Flowchart of Menu Driven Software oocioocinconincnninccnincnrnccnonoso 3 5 Figure 3 4 Pinout of Rear Panel Connector in Single Range Mode 3 8 Figure 3 5 Pinout of Rear Panel Connector in Dual Range Mode 3 10 Figure 3 6 Analog Output in Autorange MOdE ocoociconincnnncinocninccoinooos 3 11 Figure 3 7 Pin Out of Rear Panel Connector in Auto Range Mode 3 12 Figure 4 1 Cylinder Gas Dilution SySteM oooociconcicconincnonnconinccnnorncnnrncnornnnnoos 4 3 Figure 4 2 Permeation Tube System o oocicoocicccciococinccnnnccnrnccnrnncnrnccnrnncnroncnronos 4 4 Figure 5 1 Inspecting and Replacing the Capillary 5 4 Figure 5 2 Capillary ASSEMDly ooooooooocicionicconiococnocononccnrncononccnrnononornccnrnnnooos 5 4 Figure 5 3 Inspecting and Cleaning the Fan 5 5 Figure 5 4 Rebuilding the PUMP ooocococicicocicononoconon
25. 2 Bytes 1 Byte 1 Byte 0x03 or 0x04 0x0000 to maximum allowed by instrument 1 to maximum allowed by instrument 0x03 or 0x04 2 x N N N or N 1 Function code 0x80 01 IIlegal Function 02 IIlegal Address 03 Illegal Data 04 Slave Device Failure Here is an example of a request and response to read registers 10 13 Request Field Name Function Starting Address Hi Starting Address Lo No of Registers Hi No of Registers Lo Response Field Name Hex 0x03 0x00 0x09 0x00 0x04 Hex Thermo Fisher Scientific 0x05 Force Write Single Coil Thermo Fisher Scientific MODBUS Protocol Function Codes Function 0x03 Byte Count 0x06 Register value Hi 10 0x02 Register value Lo 10 0x2B Register value Hi 11 0x00 Register value Lo 11 0x00 Register value Hi 12 0x00 Register value Lo 12 0x64 Register value Hi 13 0x00 Register value Lo 13 0x64 The contents of register 10 are shown as the two byte values of 0x02 0x2B Then contents of registers 11 13 are 0x00 0x00 0x00 0x64 and 0x00 0x64 respectively The force write single coil function simulates the activation of the digital inputs in the instrument which triggers the respective action This function code is used to set a single action to either ON or OFF The request specifies the address of the action to be forced Actions are addressed starting at zero Therefore action number 1 is addressed as 0 The requested ON OFF state is s
26. 21 Troubleshooting Connector Pin Descriptions Connector Label Reference Designator Pin Signal Description 8 RS485 to Motherboard Table 6 9 Digital Output Board Connector Pin Descriptions Connector Label Reference Designation Pin Signal Description MOTHER BD J1 5V 24V 24V Ground Ground Ground SPI Reset SPI Input SPI Output SPI Board Select SPI Clock Relay 1 Contact a O O N on A O N e 8 O DIGITAL OUTPUTS J2 Relay 2 Contact a Relay 3 Contact a Relay 4 Contact a Relay 5 Contact a Relay 6 Contact a Relay 7 Contact a Relay 8 Contact a Oo aoa N on A O N Relay 9 Contact a gt o Relay 10 Contact a NC Solenoid Drive Output 1 gt gt Solenoid Drive Output 2 Solenoid Drive Output 3 Solenoid Drive Output 4 Solenoid Drive Output 5 Solenoid Drive Output 6 Solenoid Drive Output 7 gt gt gt aS gt gt gt gt co loo N o a au N Solenoid Drive Output 8 6 22 Model 43i Instruction Manual Thermo Fisher Scientific Connector Label Reference Designation Pin 20 21 22 23 24 25 26 21 28 29 30 31 32 33 34 35 36 37 Troubleshooting Connector Pin Descriptions Signal Description Relay 1 Contact b Relay 2 Contact b Relay 3 Contact b Relay 4 Contact b Relay 5 Contact b Relay 6 Contact b Relay 7 Contact b Relay 8 Contact b Relay 9 Contact b Relay 10 Contact b 24V 24V 24V 24V 24V 24V 24V 24V Table 6 10 Input
27. 3 53 Service Moder ae irra E ARUN Ure a ROTE 3 53 Dat Time A a R i 3 54 A iiA Rent er A er are Tee 3 54 Diagnostic Meninas ena a A 3 55 AAA da ieie a 3 55 Voltages fa cased cad illes eden EE sheds 3 56 Motherboard Voltages mor deco i 3 56 Interface Board Voltages de din Sans ee 3 56 VO Board Voltages viii 3 57 A esis onian aa i a Aie E ERT 3 57 A R A 3 57 Sample Flor A ah cee A E et ate 3 58 lamp Its aaa 3 58 peal S pati Tests oe 3 58 Analog Input Readings 1 daa dida 3 59 Analog Input Voltage 3 59 Disttal PUE aaa sita 3 60 Model 437 Instruction Manual xiii Contents xiv Model 43 Instruction Manual TRG ay SEAL CS nc atertcetcaesh cy ite cal sects ous a ESNS 3 60 Test Analog Outputs isisisi iineccleidccapaeteetindidaete Aiea 3 61 S t Analog Output did id 3 61 Instrument Configuration bs 3 62 Contact Info nao al in 3 62 Alarms Menu a A R Rat 3 62 Internal per ot AA 3 63 Min and Max Internal Temperature LiMitS oconnccnnnnnncnnnninnnnnoso 3 64 Chamber Temperature iia ii 3 64 Min and Max Chamber Temperature Limits 0 0 cesses 3 64 DESEA 3 65 Min and Max Pressure Erin cayt ners stceecs tobe Nace 3 65 Sample Pl da cd esta eta eee ee pus lacado 3 65 Min and Max Sample Flow Limits eiii 3 66 Lamp TiN A NI 3 66 Min and Max Lamp Intensity Limits comida 3 66 limp NY AAA A laa de eel ee 3 67 Min and Max Lamp Voltage Limits ici ciicis 3 67 Zero and Span Check rris oe e da ERNE 3 67 Max Zero and Span OiEtaadaiandti it id 3 68 Zer
28. 9999999 to 99999999 The default table is a two point table with point 1 0 00 V 000 0 U and point 2 10 00 V 10 0 U In the Main Menu choose Instrument Controls gt I O Configuration gt Analog Input Config gt Select Table Point gt User Value Thermo Fisher Scientific Model 43i Instruction Manual 3 51 Operation Instrument Controls Menu Temperature Compensation Pressure Compensation 3 52 Model 43i Instruction Manual Temperature compensation provides compensation for any changes to the instrument s output signal due to internal instrument temperature variations The effects of internal instrument temperature changes on the analyzer s subsystems and output have been empirically determined This empirical data is used to compensate for any changes in temperature This compensation can be used for special applications or when operating the instrument outside the recommended temperature range When temperature compensation is on the first line of the display shows the current internal instrument temperature measured by a thermistor on the Interface board When temperature compensation is off the first line of the display shows the factory standard temperature of 30 C In the Main Menu choose Instrument Controls gt Temperature Compensation Pressure compensation provides compensation for any changes to the instrument s output signal due to reaction chamber pressure variations The effects of reaction cham
29. A Handle all printed circuit boards by the edges A Do not remove the LCD panel or frame from the LCD module A The polarizing plate is very fragile handle it carefully a Do not wipe the polarizing plate with a dry cloth it may easily scratch the plate A Do not use alcohol acetone MEK or other Ketone based or aromatic solvents to clean the LCD module use a soft cloth moistened with a naphtha cleaning solvent A Do not place the LCD module near organic solvents or corrosive gases A Do not shake or jolt the LCD module A gt 2 4 Ze Figure 7 1 Properly Grounded Antistatic Wrist Strap The firmware can be updated by the user in the field via the serial port or over the Ethernet This includes both the main processor firmware and the firmware in all low level processors Refer to the Port manual for the firmware update procedure If the Service Menu is not displayed on the Main Menu use the following procedure to display it 1 At the Main Menu press _ Y to scroll to Instrument Controls gt press 1 gt Y J to scroll to Service Mode gt and press Ce The Service Mode screen appears Thermo Fisher Scientific Replacement Thermo Fisher Scientific Parts List Servicing Replacement Parts List 2 Press to toggle the Service Mode to ON 3 Press gt Jto return to the Main Menu 4 Return to the procedure Table 7 1 lists the replacement parts for the Model 431 major subass
30. Analog Outputs y a MEE iF Mid Set Analog Outputs The Set Analog Outputs screen contains three choices Set to full scale set to zero or reset to normal Full scale sets the analog outputs to the full scale voltage zero sets the analog outputs to 0 volts and normal operation The example that follows shows the selected output state ALL is set to full scale In the Main Menu choose Diagnostics gt Test Analog Outputs gt ALL Voltage Channel 1 6 or Current Channel 1 6 PE dra Pi Tae Dl itt EN EEE TE Sanat Bone PME Fad A woe EE fee EE EE Thermo Fisher Scientific Model 43i Instruction Manual 3 61 Operation Alarms Menu Instrument Configuration Contact Information Alarms Menu 3 62 Model 43i Instruction Manual The Instrument Configuration screen displays information on the hardware configuration of the instrument In the Main Menu choose Diagnostics gt Instrument Configuration Note If the analyzer is in service mode pressing on the item will toggle it yes or no with the exception of dilution ratio and auto calibration which may only be enabled at the factory A The Contact Information screen displays the customer service information In the Main Menu choose Diagnostics gt Contact Information The alarms menu displays a list of items that are monitored by the analyzer If the item being monitored goes outside the lower or upper limit the status of that item will go fr
31. Cho 0 C1 C2 C2 C2 C2 C3 c3 C3 C3 C1 C2 and C3 are custom ranges For more information about custom ranges see Set Custom Ranges that follows Table 3 6 Extended Ranges 3 14 Model 43i Instruction Manual ppb ppm pg m mg m 500 05 2000 2 10001 5000 5 2000 2 10000 10 5000 5 20000 20 10000 10 50000 50 20000 20 100000 100 50000 50 200000 200 100000 100 250000 250 C1 Ch C C1 C2 C2 C2 C2 Thermo Fisher Scientific Operation Range Menu ppb ppm pg m mg m C3 Cs 6 C3 Set Custom Ranges The Set Custom Ranges menu lists three custom ranges C1 C2 and C3 Custom ranges are user defined ranges Any value between 50 ppb 0 05 ppm and 10000 ppb 10 ppm can be specified as a range In the pg m mg m mode any value between 200 pg m 0 2 mg m and 25000 ug m 25 mg m can be specified as a range In the extended range mode any value between 500 ppb 0 5 ppm and 100000 ppb 100 ppm can be specified as a range In the pg m mg m mode any value between 2000 g m 2 mg m and 250000 g m 250 mg m can be specified as a range In the Main Menu choose Range gt Set Custom Ranges CH Cie hi Custom Ranges The Custom Ranges screen is used to define the custom ranges The display shows the current custom range The next line of the display is used to set the range To use the custom full scale range be sure to select it Custom range 1 2 or 3 in the SO Range sc
32. Coeffcjent anri et E e E RE 3 18 Reset User Calibration Detail a 3 19 Calibration Mia 3 19 Calibrate SOx Background leidos 3 19 Calibrate SOs Corriente os 3 20 Model 43 Instruction Manual xi Contents xii Model 437 Instruction Manual Teeter Spans Check Mercado tina 3 20 A e aa EA ETa 3 21 Period Holt nidos 3 21 Zero Span Purge Duration Minutes sccsssssordssstsesesasidecvosterenes sees 3 21 Zero Span AVETASING TAME dui a A ia 3 22 E A 3 22 Instrument Controls Menta Ole 3 23 A cae cincc ee ceca cece ce at OTA auc EE EEA bendy 3 23 Datalogging Settings eiii 3 23 Select SREC ERE Girres enii td ass 3 24 o O REA 3 24 Number GF Records titted ala aca cA a N 3 25 Wate aid me iio 3 25 Record Display nina edi ere edness 3 26 PLASC lOO sooni ti a a 3 26 Select Contentiuoin inicia 3 26 Choose ARTE 3 27 C ONCENt ATIONS eiiieaen ia gad ibeeria asii 3 27 Other Measurements eea i a eel beaten ee 3 28 Analog ptas a 3 28 Commit Co ibe nt civic ado ii 3 29 Reset to Default Content atada 3 29 Configure Datalogic dida 3 30 Logging Penod Mitre sr a 3 30 Memory Allocation Percent cesta dae 3 30 Data Veter iii 3 31 ns a AAA 3 31 Baud Ratio dai 3 31 A Otte Rael E ened or nen eas RN 3 32 A A cca Aas telat eae et a nt 3 32 Data Bits eresien isiin AA AA 3 32 PA SAA 3 33 Stop BUS eta 3 33 Instrument ID tana ninas 3 33 Communication Protocol rallies 3 34 Streaming Data Contatori 3 34 Streaming Data Interval snai O le oat 3 35 Choose
33. Default Output AsSignMBNt ocococicoonicconincconnconinccnrnconrnccnrnccnnoos B 41 Table C 1 Read Registers for 4Bl ooocicoociccociccocnoconinccnnocononocnnoncnnoncnrnrcnrnrcnnnos C 8 Table C 2 Write Coils Td iaa tato C 9 Table C 3 Read Coils adi C 10 Table D 1 Operating Status for Model 437 oocociccociccccnococicccnnacininicnnnncnnonoo D 7 Table D 2 Error Status for Model 43 0 0 ccccccccsceseesecsecesteseeseeseeseetesneseeseens D 7 Thermo Fisher Scientific Thermo Fisher Scientific Chapter 1 Introduction The Model 437 Pulsed Fluorescence Analyzer combines proven detection technology easy to use menu driven software and advanced diagnostics to offer unsurpassed flexibility and reliability The Model 437 has the following features 320 x 240 graphics display Menu driven software Field programmable ranges User selectable single dual auto range modes Multiple user defined analog outputs Analog input options High sensitivity Fast response time Linearity through all ranges Internal sample pump Totally self contained Insensitive to changes in flow and ambient temperature User selectable digital input output capabilities Standard communications features include RS232 485 and Ethernet C Link MODBUS Geysitech Bayern Hessen streaming data and NTP Network Time Protocol protocols Simultaneous connections from different locations over Ethernet For details of the analyzer s principle of operation
34. Item Sigival incita db 3 35 Goncentrationsaanr e a e e EE a A a 3 35 Other Measurement sirios ds 3 36 Analog O See el ee tere eae 3 36 RS 232 RS 485 Selection adas 3 37 TEPAP Settings caia ia 3 37 Use DHC Presin naaa eaa e e 3 38 IP Address iii da 3 38 Na O 3 39 Thermo Fisher Scientific Thermo Fisher Scientific Contents GALE WAY nadaanan aa a a E 3 39 LONE 3 39 Network Time Protocol Severa iria a 3 40 VO Configuration 3 40 Output Relay Settings ros 3 41 A seas Gis AA Masts 3 41 T strument State ar ctenten ieee aia aiaeGeiasdedl waded 3 42 IA A Ns 3 42 otis lata ds 3 43 Digital Input Settings end obsididteliriin tai 3 43 Logic tela en R nadaa pt 3 44 Jostrument Actio eisenai neina ai a 3 44 Analog Output Configuration Select Channel eee 3 44 Allow Over Under Ranger 3 45 Analog Output Configuration Select Action oooonnninninmmmm 3 45 Select Ramee ninia ida 3 46 Minimum and Maximum Value iS 3 46 Choose Signal To Updated 3 47 Analog Input Configuration e sssessesssesssssssrsrsesrsesesesesesesesesese 3 48 DS iesst ree ierg ea EEA EE EE ia 3 49 iS 3 49 Decimal Plica il atras ps 3 50 Number of Table Pot ans ii i 3 50 A O ity on Renee eer 3 51 NOMS re sedate hms ale tne ER ES E Oe watt gl Pate att ea ase 3 51 User Valencina 3 51 Temperature om pensatie Mires cossesaecasvodd tanbveroc arose otsssanenteaiontsn 3 52 Pressure Compensation scriata a ai 3 52 Sereen Contiastaraa reesi rroi osi oE E E A OKEN e aai eari
35. Level Connection Diagrams on page 6 10 e Connector Pin Descriptions on page 6 12 e Service Locations on page 6 24 Read the safety precautions in the Preface and the Servicing chapter before performing any actions listed in this chapter The troubleshooting guides presented in this chapter are designed to help isolate and identify instrument problems Table 6 1 Troubleshooting Power Up Failures Table 6 2 Troubleshooting Calibration Failures and Table 6 3 Troubleshooting Measurement Failures provide general troubleshooting information and Model 43i Instruction Manual 6 1 Troubleshooting Troubleshooting Guides 6 2 Model 43i Instruction Manual indicate the checks that you should perform if you experience an instrument problem Table 6 4 Troubleshooting Alarm Messages lists all the alarm messages you may see on the display and provides recommendations about how to resolve the alarm condition Table 6 1 Troubleshooting Power Up Failures Malfunction Does not start The backlight on the display does not come on and the pump motor is not running Possible Cause No power or wrong power configuration Main fuse is blown or missing Bad switch or wiring connection Action Check the line to confirm that power is available and that it matches the voltage and frequency configuration of the instrument Unplug the power cord open the fuse drawer on the back panel and check the fuses vi
36. System Description Hardware Mirror Housing Assy Flash Intensity Assy PMT Assy Chamber Inlet Assy Flash Lamp Housing f fo Reaction Chamber Assy AAL I Flash Trigger Assy oy Kicker Assy Fan Flow Transducer Permeation Oven Assy Opt Pressure Transducer Pump Thermistor a Solenoid Valves Opt as 1 0 Expansion Board Opt Measurement Interface Board LCD Display gy 4 lt Digital Output Board Front Panel sade UE T g i Power Supply HVPS Input Board Assy Optional Transformer Mother Board not shown Figure 8 1 Hardware Components Hydrocarbon Kicker The hydrocarbon kicker removes hydrocarbons from the gas stream while leaving the SO concentration unaffected It operates on a selective permeation principle using differential pressure to force hydrocarbon molecules to pass through the tube wall The differential pressure is created across the tube wall as sample gas passes through a capillary tube which reduces its pressure The sample gas is then fed to the shell side of the hydrocarbon kicker Heated Hydrocarbon The optional heated hydrocarbon kicker is used in applications requiring Kicker Optional more efficiency than the standard kicker The heated kicker operates at the same temperature as the optical bench and requires no additional power 8 2 Model 43i Instruction Manual Thermo Fisher Scientific Optics Flash Lamp Condensing Lens Mirror A
37. and present only if auto calibration is installed A In the Main Menu choose Calibration gt Zero Span Check 3 20 Model 43i Instruction Manual Thermo Fisher Scientific Operation Calibration Menu mph Ca EUs EEE Eut Beee Next Time The Next Time screen is used to set the initial date and time of the zero span check Once the initial zero span check is performed the date and of the next zero span check is calculated and displayed In the Main Menu choose Calibration gt Zero Span Check gt Next Time Eat Pike ONE F Period Hours The Zero Span Period Hours screen defines the period or interval between zero span checks Periods between 0 and 999 hours are acceptable To turn the zero span check off set the period to 0 In the Main Menu choose Calibration gt Zero Span Check gt Period Hours Zero Span Purge The Zero Duration Minutes screen defines how long zero air is sampled by Duration Minutes the instrument The span and purge duration screens look and functions Thermo Fisher Scientific Model 43i Instruction Manual 3 21 Operation Calibration Menu Zero Span Averaging Time Zero Span Ratio 3 22 Model 43i Instruction Manual the same way as the zero duration screen and are used to set how long the span gas and sample gas are sampled by the instrument Durations between 0 and 60 minutes are acceptable Each time a zero span check occurs the zero check is done first followed by the span check T
38. and product specifications see the following topics Principle of Operation on page 1 2 describes the analyzer s operating principles Specifications on page 1 3 provides a list of the analyzer s P pag P y performance specifications Model 43i Instruction Manual 1 1 Introduction Principle of Operation Principle of Operation 1 2 Model 43i Instruction Manual Thermo Fisher Scientific is pleased to supply this pulsed fluorescence SO analyzer We are committed to the manufacture of instruments exhibiting high standards of quality performance and workmanship Service personnel are available for assistance with any questions or problems that may arise in the use of this analyzer For more information on servicing see the Servicing chapter The Model 43 operates on the principle that SO molecules absorb ultraviolet UV light and become excited at one wavelength then decay to a lower energy state emitting UV light at a different wavelength Specifically SO hv gt SO gt SO ho The sample is drawn into the Model 43 i through the SAMPLE bulkhead as shown in Figure 1 1 The sample flows through a hydrocarbon kicker which removes hydrocarbons from the sample by forcing the hydrocarbon molecules to permeate through the tube wall The SO2 molecules pass through the hydrocarbon kicker unaffected The sample then flows into the fluorescence chamber where pulsating UV light excites the
39. ani la 8 3 Bandpass Fla to tido slo eee necedad aia 8 4 Photomultpler Tube tia a old cado sa Lee ase 8 4 o A ad 8 4 Flow Sentir da 8 4 Pressure Arana ucer a alos 8 4 Capillary viii o 8 4 Vacum PIM a 8 4 e isa es de oie in Dido 8 4 Instrument o O 8 4 Menttorinesienal Susie 8 5 Measurement Calculations init 8 5 Output Communication ct sees E eats 8 5 A dh cect chao aceasta aes ceca eth aaa 8 6 Motherboard A ela ic Oe a a e a Ae ba Oo ead 8 6 External Conor aire etc EIT Per Aver eee 8 6 Internal Connectors denia 8 6 Measurement Interface Board usina ede dial lslscten 8 7 Measurement Interface Board ConnectoFs ceceseseeeseseeeeneeeeeees 8 7 Flow Sensor Assembly A ah eau Manat beutauesmmealtanue Stet 8 7 Pressure Sensor Assembly 4 sis ds coon eat ea re od SG 8 7 Bench A O a 8 8 PMT Power Supply Asser bl ycescitieccossctecnesssstecchave toadieienateced een 8 8 Diagnostic LED ie 8 8 ii A asd don hace h onau wet paie EASA 8 8 Digital OQutp t A e A 8 8 Front Panel Connector Dodo ia iS 8 8 Model 43 Instruction Manual xvii Contents xviii Chapter 9 Appendix A Appendix B Model 43 Instruction Manual Flash Frigger DOM senean ngaga ainn a 8 9 Flash Intensity Board aa 8 9 I O Expansion Board Optional cintia re 8 9 VO Components nike dies 8 9 Analog Voltage Outputs a seeds 8 9 Analog Current Outputs Optional fc ciccccscssecssesiscetssseveseseencesocest 8 10 Analog Voltage Inputs Optional de 8 10 Digital Relay Outpu
40. atmospheric bypass a From the Run screen press to display the Main Menu b From the Main Menu select Diagnostics gt gt Flow to display the Sample Flow screen c Make sure that the reported sample flow is less than the actual zero air flow 3 Press gt to monitor the zero air reading and wait for the reading to stabilize 4 Record the measured SO value as the zero drift since the last adjustment If the zero has changed by more than 0 015 ppm it is recommended that a new calibration be performed Model 43i Instruction Manual 4 11 Calibration Zero Span Check 4 12 Model 43i Instruction Manual Connect a source of vented span gas to the SAMPLE bulkhead The span gas should be approximately 80 of the full scale range To ensure that the span gas is being measured at atmospheric pressure check that the flow is approximately 0 5 LPM or more with the excess flowing out the atmospheric bypass Press _ gt to monitor the span gas reading and wait for the reading to stabilize Record the difference between the measured SO value and the actual SO span concentration used This is the span drift since the last adjustment If the calibration has changed by more than 10 a new calibration should be performed Thermo Fisher Scientific Safety Precautions Thermo Fisher Scientific A A Chapter 5 Preventive Maintenance This chapter describes the periodic maintenance procedures that
41. concentration alarm warning value B 14 so2 analog out range Reports analog current output range per channel B 39 analog vin Retrieves analog voltage input data per channel B 39 analog vout range Reports analog voltage output range per channel B 39 avg time Reports sets averaging time B 8 baud Reports sets current baud rate B 36 Model 431 Instruction Manual B 3 C Link Protocol Commands Commands List B 4 Model 43 Instruction Manual Command Description Page cal high so2 coef Sets auto calibrates high range SO coefficient B 24 cal low so2 coef Sets auto calibrates low range SO coefficient B 24 cal pres Sets current measured pressure as pressure during calibration B 26 for pressure compensation cal so2 bkg Sets auto calibrates SO background B 24 cal so2 coef Sets auto calibrates SO coefficient B 24 clr lrecs Clears away only long records that have been saved B 16 clr records Clears away all logging records that have been saved B 16 clr srecs Clears away only short records that have been saved B 16 contrast Reports sets current screen contrast B 32 conv set temp Reports sets temperature setpoint for SO converter B 32 Model 4507 only conv temp Reports current SO converter temperature Model 450 only B 9 copy Irec to sp Sets copies current Irec selection into the scratch pad B 22 copy sp to Irec Sets copies current selections in scratch pad i
42. connector d PMT power cable from the HVPS e LED cable from the LED connector on the measurement interface board f Flasher assembly from the FLASH TRIG connector on the measurement interface board g Green ground wire from the floor plate 3 Disconnect the optical bench plumbing 4 Using a 5 32 inch Allen wrench remove the four optical bench retaining screws and lift the optical bench off the floor plate 5 Replace the optical bench by following the previous steps in reverse order 6 Calibrate the instrument Refer to the Calibration chapter in this manual Retaining Screws 4 Figure 7 6 Replacing the Optical Bench 7 12 Model 43i Instruction Manual Thermo Fisher Scientific Servicing Cleaning the Mirrors The mirrors located in the optical bench do not come in contact with the Cleaning the Mirrors sample gas and DO NOT need cleaning Read the Equipment Damage warning that follows f Equipment Damage DO NOT attempt to clean the mirrors in the optical bench These mirrors do not come in contact with the sample gas and should not be cleaned Cleaning the mirrors can damage the mirrors A Flash Lamp Use the following procedure to replace the flash lamp Figure 7 7 Re p lacement Equipment Required Flash lamp Philips screwdriver f Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handlin
43. days for all other products the Warranty Period Seller agrees during the Warranty Period to repair or replace at Seller s option defective Products so as to cause the same to operate in substantial conformance with said published specifications provided that a Buyer shall promptly notify Seller in writing upon the discovery of any defect which notice shall include the product model and serial number if applicable and details of the warranty claim b after Seller s review Seller will provide Buyer with service data and or a Return Material Authorization RMA which may include biohazard decontamination procedures and other product specific handling instructions and c then if applicable Buyer may return the defective Products to Seller with all costs prepaid by Buyer Replacement parts may be new or refurbished at the election of Seller All replaced parts shall become the property of Seller Shipment to Buyer of repaired or replacement Products shall be made in accordance with the Delivery provisions of the Seller s Terms and Conditions of Sale Consumables including but not limited to lamps fuses batteries bulbs and other such expendable items are expressly excluded from the warranty under this warranty Notwithstanding the foregoing Products supplied by Seller that are obtained by Seller from an original manufacturer or third party supplier are not warranted by Seller but Seller agrees to assign to Buyer any war
44. fan filter Figure 5 3 Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Turn the instrument OFF and unplug the power cord 2 Remove the fan guard from the fan 3 Flush the filter with warm water and let dry a clean oil free purge will help the drying process or blow the filters clean with compressed air 4 Re install the filter and fan guard Mounting Screws 4 Filter Fan Guard A e Figure 5 3 Inspecting and Cleaning the Fan If a sample pre filter is used it should be inspected regularly for excessive dust and particulates which will restrict flow Replace the sample filter as necessary Model 43i Instruction Manual 5 5 Preventive Maintenance Lamp Voltage Check Lamp Voltage Check Leak Test 5 6 Model 43i Instruction Manual The instrument is equipped with a lamp voltage control circuit which automatically corrects for degradation of the flash lamp However after several years of use the lamp may have degraded to the point that is being driven with the maximum voltage that the power supply can deliver Use the following procedure to check the lamp v
45. flow rate Ensure all devices are properly calibrated and that all flows are corrected to 25 C and 1 atm FLOWMETER FLOW CONTROLLER FLOW PERMEATION CONTROLLER FLOWMETER DEVICE MIXING CHAMBER A A EXTRA OUTLETS CAPPED WHER NOTIN YSE TO INLET OF ANALYZER UNDER CALIBRATION Figure 4 2 Permeation Tube System The SO output level is calculated from PxK SOz lovr T Where SOz our SO output concentration in ppm P permeation rate in pg min Fr total flow rate of gas after mixing chamber Fp Fp in liters minute K SO2 0 382 constant for SO permeant Commercial Commercial permeation systems such as the Thermo Scientific Model 146 Permeation Systems Multigas Calibration System are available for this requirement Refer to the instrument instruction manual for more information 4 4 Model 43i Instruction Manual Thermo Fisher Scientific Multi Point Calibration Thermo Fisher Scientific Calibration Multi Point Calibration Regulations typically require zero span calibration when the instrument is newly installed moved repaired interrupted for more than a few days or when span or zero shift by more than 15 All gas must be supplied to the instrument at atmospheric pressure It may be necessary to employ an atmospheric bypass plumbing arrangement to accomplish this Refer to Figure 2 4 in the Installation chapter for an example of an atmospheric bypass p
46. from the top assembly by twisting and gently pulling the glass away from the top Note Keep the glass clean when handling it A Place the permeation tube s in the glass chamber Attach the glass chamber to the top assembly by gently pushing the two together with a slight twisting motion Replace the glass chamber assembly into the oven until the top of the assembly is flush or slightly below the top of the oven Equipment Damage Do not use tools to tighten the knurled screw in the following step A Thermo Fisher Scientific Computation of Concentrations Oven Installation and Configuration Thermo Fisher Scientific Optional Equipment Internal Permeation Span Source 7 Tighten the knurled screw finger tight Do not use tools to tighten 8 Tighten the white plastic retaining collar 9 Replace the oven cover being careful to place the tubing and wiring into the cover slot The computation of SO output level is shown in the following information Note that is assumed that all devices are properly calibrated and that all flows are corrected to 25 C and 1 atm Permeation Tube R K o Output ppm Where R permeation rate in ng min Qo flow rate of gas scc min into the charcoal scrubber during span mode K constant for the specific permeant 24 45 MW MW molecular weight K SO 0 382 Use the following procedure to install and configure the permeation oven 1 Physically install the
47. good unit if possible Identify the defective component and replace Check analog cable shielding and grounding Try to localize source of noise by comparing analog signal to data collected through RS 232 or Ethernet Poor linearity Problem with calibrator Problem with input board range switching Verify accuracy of the multipoint calibration system with an independent flow meter Go to the Input Board Test screen Service menu and step through each range while the instrument samples a known stable Thermo Fisher Scientific Malfunction Possible Cause Leak in sample probe line Troubleshooting Troubleshooting Guides Action source of S02 Stay on the Input Board Test screen and while holding instrument on the lowest gain step the calibrator through all SO levels Manually plot signal vs concentration to verify linearity Check for variable dilution Excessive response time Averaging time is not set correctly Instrument is not drawing in sample at normal flow rate SO is being absorbed and or released by dirt in the tubing or filters of the sampling system or inside the instrument Go to Averaging Time Main Menu and verify setting Check sample Flow and Pressure readings on the Diagnostics screen Use an independent flow meter to check flows at the sample inlet and exhaust bulkheads they should match Perform a leak test as described in the Preventive Ma
48. gt TCP IP Settings gt NTP Server The Ol meh A A The I O Configuration menu deals with configuration of the analyzer s I O system The analog input configuration is displayed only if the I O expansion board option is installed Note The digital outputs may take up to one second after the assigned state occurs to show up on the outputs A Note Analog Input Config is only displayed if the optional I O Expansion Board is installed A In the Main Menu choose Instrument Controls gt I O Configuration Thermo Fisher Scientific Thermo Fisher Scientific Operation Instrument Controls Menu Ea EE Py Tame al Cilia MARTE mid Lario Md Output Relay Settings The Output Relay Settings menu displays a list of the 10 digital output relays available and allows the user to select the instrument parameter or logic state to change for the relay selected In the Main Menu choose Instrument Controls gt I O Configuration gt Output Relay Settings gt 1 10 Logic State The Logic State screen is used to change the I O relay to either normally open or normally closed Press to toggle and set the logic state between open and closed Ea Cir Pern pai piini Liai Pa Mid EUs EEE TE oa nr Model 43i Instruction Manual 3 41 Operation Instrument Controls Menu Instrument State The Instrument State submenu allows the user to select the instrument state that is tied to the selected relay output A submenu lis
49. gt lt BCC gt lt STX gt DA lt ETX gt lt BCC gt The data query string is valid and will be answered with data transmission only if the command starts with lt STX gt which is followed by the characters DA and the lt address gt if present matches the Instrument Address and the command is terminated with either lt CR gt with no checksum or lt ETX gt followed by the correct checksum lt BCC gt D 4 Model 43 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Geysitech Bayern Hessen Protocol Geysitech Commands Sample Data Reply String in response to Data Query Command DA In response to a valid data query command DA the instrument responds in the following format lt STX gt MD02 lt SP gt lt address gt lt SP gt lt measured valuel gt lt SP gt lt status gt lt SP gt lt SFKT gt lt SP gt lt address 1 gt lt SP gt lt measured value2 gt lt SP gt lt status gt lt SP gt lt SFKT gt lt ETX gt lt BCC gt The response uses the same command terminators as used by the received command i e if the received command was terminated with a lt CR gt the response is terminated with lt CR gt and if the command was terminated with a lt ETX gt lt BCC gt the response is terminated with lt ETX gt and the computed checksum lt BCC gt The 02 after the MD indicates that two measurements are present in the reply string a 03 for three measurements and so on this will also determine the length of the r
50. high level processor that also controls the front panel user interface The low level Thermo Fisher Scientific Monitoring Signals Measurement Calculations Output Communication Thermo Fisher Scientific System Description Software processors all run common firmware that is bundled with the high level firmware and loaded on power up if a different version is detected Each board has a specific address that is used to identify to the firmware what functions are supported on that board This address is also used for the communications between the low level processors and the high level processor Every tenth of a second the frequency counters analog I O and digital I O are read and written to by the low level processor The counters are accumulated over the past second and the analog inputs are averaged over that second The high level processor polls the low level processors once per second to exchange the measurement and control data Signals are gathered from the low level processors once per second and then processed by the high level processor to produce the final measurement values The one second accumulated counts are accumulated and reported for the user specified averaging time If this averaging time is greater than ten seconds the measurement is reported every 10 seconds The one second average of the other analog inputs are reported directly no additional signal conditioning is performed by the high level processor
51. in diameter Temperature range 20 30 C Physical dimensions Contained inside the Model 437 Power requirements 120 VAC 50 60 Hz 50 watts in addition to the standard Model 43i Weight Approximately five pounds in addition to standard Model 437 Model 43i Instruction Manual 1 5 Introduction Specifications 1 6 Model 43i Instruction Manual Thermo Fisher Scientific Lifting A Unpacking and Thermo Fisher Scientific Inspection Chapter 2 Installation Installation of the Model 431 includes lifting the instrument unpacking and inspection connecting sample zero span and exhaust lines and attaching the analog outputs to a recording device The installation should always be followed by instrument calibration as described in the Calibration chapter of this manual This chapter provides the following recommendations and procedures for installing the instrument e Lifting on page 2 1 e Unpacking and Inspection on page 2 1 e Setup on page 2 3 e Connecting External Devices on page 2 5 e Startup on page 2 9 When lifting the instrument use a procedure appropriate to lifting a heavy object such as bending at the knees while keeping your back straight and upright Grasp the instrument at the bottom in the front and at the rear of the unit Although one person can lift the unit it is desirable to have two persons lifting one by grasping the bottom in the front and the other by gras
52. in PDA A a B 2 Commands Lts e Ra B 2 Ea MEMES Sa B 8 Alarms A ates road ae B 11 Thermo Fisher Scientific Thermo Fisher Scientific Contents O ties ish stetaadatetadsads Supe nian cist caatacdatese dicate B 15 Datalopoine ia ti Hae eee B 16 A E ay cael Sex dacs ss uae E E ST B 24 Keys Display csin erini esisi sirinin a AAEE B 26 Measurement Configuration oO ooo B 28 Hardware COn tE ai E ORR R B 32 Communications Configuration ti ada B 35 1OConto raton dera B 39 Record Layout Definition ii ri B 43 Format Specifier for ASCII ResponseS mcionionocononmarconosenaccncnrenos B 43 Format Specifier for Binary Responses ensiantdais B 44 Format Specifier for Front Panel Layout s sescecsesesceeesorteres B 44 Oc nomen ee rrr RED oR NT ra We Ste ni OUR any Meera ATS ois MoO PERE ea B 45 Value String iia dain B 45 Valie Sour E ANS B 45 Alarm Information A ei aie alice ea or et coat B 45 Translation able DSi B 45 Selection Laia ist B 46 Button Descritor os B 46 A easel aid eaat ane eai e dC ted ed A B 46 MOD BUS Protocol C 1 Serial Communication Parameters 3206 2264s ences eatin hae C 1 TCP Communication Parametros C 2 Application Data Unit Deo C 2 Slave Address anana Seed ce et ae a ret Nan N ETA C 2 MBAP eden EE C 2 Function Codes atacan C 3 Dataene al ae Atel Mun aN C 3 Error Checa C 3 A eas oak ae td Set aa eo pat C 3 0x01 0x02 Read Coils Read Inputs ovina C 3 0x03 0x04 Read Holding Registers Read Input Registers
53. in a Thermo non ElA rack includes mounting slides and retrofit front panel rack mounting handles Model 43i Instruction Manual 9 11 Optional Equipment Mounting Options Figure 9 5 Rack Mount Option Assembly 9 12 Model 43i Instruction Manual Thermo Fisher Scientific Optional Equipment Mounting Options 16 75 REF Figure 9 6 Bench Mounting Thermo Fisher Scientific Model 43i Instruction Manual 9 13 Optional Equipment Mounting Options 16 75 Figure 9 7 EIA Rack Mounting 9 14 Model 43i Instruction Manual Thermo Fisher Scientific Optional Equipment Mounting Options 16 75 Figure 9 8 Retrofit Rack Mounting Thermo Fisher Scientific Model 43i Instruction Manual 9 15 Optional Equipment Mounting Options 9 16 Model 43i Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Appendix A Warranty Seller warrants that the Products will operate or perform substantially in conformance with Seller s published specifications and be free from defects in material and workmanship when subjected to normal proper and intended usage by properly trained personnel for the period of time set forth in the product documentation published specifications or package inserts Ifa period of time is not specified in Seller s product documentation published specifications or package inserts the warranty period shall be one 1 year from the date of shipment to Buyer for equipment and ninety 90
54. in the Permeation Tube Oven Calibration procedure described previously Model 43i Instruction Manual 9 7 Optional Equipment Internal Permeation Span Source Determining Release Rate by Transfer Standard 9 8 Model 43i Instruction Manual 10 Insert the permeation tube carefully Do not touch with fingers Turn on the instrument Wait 48 72 hours for the permeation tube to stabilize Carefully remove the permeation tube from the oven and weigh to an accuracy of 0 1mg Perform this measurement as quickly as possible Replace the permeation tube into the oven of the instrument Repeat Steps 5 and 6 after two weeks Compute the weight loss of the permeation tube from the values determined in Steps 5 through 7 Repeat Steps 5 through 8 until the weight loss has been determined to a precision of 1 2 For the most accurate work use the permeation tube in the same oven that was used to determine the weight loss of permeation tube Use the following procedure to determine the release rate by transfer standard Make sure the oven has been calibrated as described in the Permeation Tube Oven Calibration procedure described previously Also make sure that the Transfer Standard has been properly calibrated Determine the permeation rate for the permeation tube in the Transfer Standard or install a certified permeation tube Allow the permeation tubes in both the 437 and the Transfer Standard to stabilize for at least
55. individual voltage channels 1 6 and Thermo Fisher Scientific Operation Instrument Controls Menu individual current channels 1 6 if the I O expansion board option is installed In the Main Menu choose Instrument Controls gt I O Configuration gt Analog Output Config ERED CH Py GE Ebel PPP LES Pied Bai PTT PLWH Allow Over Under Range The Allow Over Under Range screen is used to select whether or not the analog outputs are allowed to exceed the maximum selected value of 100 mV 1 V 5 V 10 V or 20 mA or the minimum selected value of 0 V 0 mA or 4 mA By default this parameter is set to on and 5 over and under range is allowed for all analog output channels e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Analog Output Config gt Allow Over Under Range a Cdra MT CH ciinii BMRB EE Saeed Besse Pr Aad PRO E Analog Output Configuration The Analog Output Configuration menu displays a list of the analog Select Action output configuration choices from which the user selects the parameter to adjust for the selected output channel Configuration choices include selecting range setting minimum maximum values and choosing the signal to output e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Analog Output Config gt All Voltage Channels All Current Channels Voltage Channel 1 6 or Current Channel 1 6 Thermo Fisher Scientific Model 43i Inst
56. integer representing degrees C The example that follows sets the converter temperature to 50 1 C Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Hardware Configuration Send set conv set temp 50 1 Receive set conv set temp 50 1 ok date This command reports the current date The example that follows reports the date as December 1 2004 Send date Receive date 12 01 04 set date mm dd yy mm month dd day yy year This command sets the date of the analyzer s internal clock The example that follows sets the date to December 1 2004 Send set date 12 01 04 Receive set date 12 01 04 ok set default params This command sets all the parameters to their default values This does not affect the factory calibrated parameters Send set default params Receive set default params ok lamp status This command reports the status of the flash lamp on or off The example that follows reports the flash lamp is on Send lamp status Receive lamp status on set lamp onoff These commands set the flash lamp on or off The example that follows turns the flash lamp off Send set lamp off Receive set lamp off ok pmt status This command reports the status of the PMT on or off The example that follows reports that the PMT is on Model 43 Instruction Manual B 33 C Link Protocol Commands Hardware Configuration B 34 Model 43 Instruction Manual Send pmt status Receive pmt
57. lamp intensity B 16 lamp voltage Reports flash lamp voltage B 16 layout ack Disables stale layout layout changed indicator B 38 le left Simulates pressing left pushbutton B 26 led status led Reports sets optical test LED status on or off B 16 list din Lists current selection for digital input B 17 list dout Lists current selection for digital output B 17 Model 43 Instruction Manual B 5 C Link Protocol Commands Commands List B 6 Model 43 Instruction Manual Command Description Page list Irec Lists current selection Irec logging data B 17 list sp Lists current selection in the scratchpad list B 17 list srec Lists current selection srec logging data B 17 list stream Lists current selection streaming data output B 17 list var aout Reports list of analog output index numbers and variables B 42 list var din Reports list of digital input index numbers and variables B 42 list var dout Reports list of digital output index numbers and variables B 42 low avg time Reports sets low averaging time B 8 low range Reports sets current SO low range B 28 low so2 Reports SO concentration calculated with low range B 9 coefficients low so2 coef Reports sets low range SO coefficient B 25 low sp conc Reports sets low span concentration B 25 Ir Outputs long records in the format specified in the command B 17 Irec Outputs long records B 18 Irec f
58. line 3 Connect a suitable recording device to the rear panel connector For detailed information about connecting to the instrument refer to Connecting External Devices on page 2 5 External Device Connection Components on page 7 4 Model 43i Instruction Manual 2 3 Installation Setup Terminal Block and Cable Kits on page 9 9 Instrument Controls gt I O Configuration in the Operation chapter For detailed information about troubleshooting a connection refer to Analog Output Testing on page 7 21 4 Plug the instrument into an outlet of the appropriate voltage and frequency WARNING The Model 43 is supplied with a three wire grounding cord Under no circumstances should this grounding system be defeated A EXTERNAL ACCESSORY 95 RS 232 485 8 lt j ETHERNET POWER FAIL 2 RELAY DIGITAL INPUTS ANALOG VOLTAGE OUTPUTS 2 4 Model 43i Instruction Manual O DIGITAL SIT OUTPUTS SPAN ZERO PON vo no NN wg AR EXPANSION E3 ki AN a O y NO y A WT xY EXHAUST SAMPLE Figure 2 3 Rear Panel Thermo Fisher Scientific A Connecting External Devices Terminal Board PCB Assemblies 1 0 Terminal Board Thermo Fisher Scientific Installation Connecting External Devices Vent to Exhaust Line at Instrument Bulkhead Atmospheric Pressure s a Input Gas Line Figure 2 4 Atmo
59. membranes Listen for the rapid clicking of the flash lamp Check the flash lamp voltage and intensity Check the PMT voltage and run the optical span test A good optical span test indicates that the PMT is OK and that the problem is more likely the flash lamp Zero or Span will not stabilize Flow rate of the diluted span mix is inadequate Instrument is not drawing in span gas SO is being absorbed Check by pass or atmospheric pressure vent to verify that the zero air system is providing more flow than the instrument is drawing Check sample Flow and Pressure readings on the Diagnostics screen Use an independent flow meter to check flows at the sample inlet and exhaust bulkheads they should match Perform a leak test as described in the Preventive Maintenance chapter Replace any lines made of vinyl or Thermo Fisher Scientific Thermo Fisher Scientific Malfunction Possible Cause and released by dirt in the tubing or filters of the calibration system or contamination inside the instrument Averaging time is not Troubleshooting Troubleshooting Guides Action other plastics with fresh Teflon or stainless steel Replace Teflon filter membranes that look dirty Remove any filters that are not Teflon membranes Check the Averaging Time in Main set correctly Menu If too high the unit will be slow to stabilize If too low the signal may appear noisy PMT high volt
60. not at earth ground A 1 Turn instrument OFF unplug the power cord and remove the cover 2 Remove the I O expansion board optional if used See the I O Expansion Board Replacement procedure in this chapter 3 Disconnect the digital output board ribbon cable from the motherboard 4 Using the nut driver remove the two standoffs securing the board to the rear panel Figure 7 19 5 Pop off the digital output board from the mounting studs and remove the board 6 To install the digital output board follow previous steps in reverse Use the following procedure to replace the motherboard Figure 7 18 Equipment Required Motherboard Philips screwdriver Nut driver 3 16 inch Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Turn instrument OFF unplug the power cord and remove the cover 2 Remove the I O expansion board optional if used See the I O Expansion Board Replacement procedure in this chapter Thermo Fisher Scientific Measurement Interface Board Replacement Thermo Fisher Scientific A Servicing 3 Remove the digital out
61. panel Remove two screws from the left side of the case viewed from front Remove one screw from the bottom rear of the case Remove one screw from the top front of the partition panel While holding the case securely loosen the captive screw at the rear of the measurement bench and pull the measurement bench from the rear of the case Remove the screw at the top rear of the partition panel securing the top of partition panel to the measurement bench and lower the panel being careful not to put excessive tension on the cables Replace the measurement bench by following previous steps in reverse Use the following procedure to replace the fuse Equipment Required Replacement fuses refer to the Replacement Parts List in this chapter Turn instrument OFF and unplug the power cord Remove fuse drawer located on the AC power connector If either fuse is blown replace both fuses Thermo Fisher Scientific Pump Replacement Thermo Fisher Scientific Servicing Pump Replacement 4 Insert fuse drawer and reconnect power cord Use the following procedure to replace the pump Figure 7 4 To rebuild the pump see Pump Rebuilding in the Preventive Maintenance chapter Equipment Required Pump Philips screwdriver Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal componen
62. qualified service personnel only A The Model 437 is supplied with a three wire grounding cord Under no circumstances should this grounding system be defeated 4 Equipment Damage Do not attempt to lift the analyzer by the cover or other external fittings 4 Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged viii Model 437 Instruction Manual Thermo Fisher Scientific FCC Compliance WEEE Symbol Thermo Fisher Scientific About This Manual FCC Compliance Alert Description the chassis is not at earth ground 4 This adjustment should only be performed by an instrument service technician A Handle all printed circuit boards by the edges only A Do not remove the panel or frame from the LCD module A The LCD module polarizing plate is very fragile handle it carefully A Do not wipe the LCD module polarizing plate with a dry cloth it may easily scratch the plate 4 Do not use Ketonics solvent or aromatic solvent to clean the LCD module use a soft cloth moistened with a naphtha cleaning solvent 4 Do not place the LCD module near organic solvents or corrosive gases amp Do not shake or jolt the LCD module A Changes or modifications
63. reports the outputs of the 6 or 12 Digital to Analog converters according to Table B 12 The example that follows shows that the D A 1 is 97 7 full scale Send dtoa 1 Receive dtoa 1 97 7 Note All channel ranges are user definable Ifany customization has been made to the analog output configuration the default selections may not apply Table B 12 Default Output Assignment DtoA Function Single Range Dual Range Autorange 1 Voltage Output Low SO Low SO High Low SO 2 Voltage Output Low S0 High SO Range Status 3 Voltage Output Not Assigned Not Assigned Not Assigned 4 Voltage Output Not Assigned Not Assigned Not Assigned 5 Voltage Output Not Assigned Not Assigned Not Assigned 6 Voltage Output Not Assigned Not Assigned Not Assigned 7 Current Output Low SO Low SO High Low SO 8 Current Output Low SO High SO Range Status 9 Current Output Not Assigned Not Assigned Not Assigned 10 Current Output Not Assigned Not Assigned Not Assigned 11 Current Output Not Assigned Not Assigned Not Assigned Model 43 Instruction Manual C Link Protocol Commands 1 0 Configuration B 42 Model 43 Instruction Manual DtoA Function Single Range DualRange Autorange 12 Current Output Not Assigned Not Assigned Not Assigned list var aout list var dout list var din These commands report the list of index numbers and the variables associated with that index number available for selection in the curr
64. scattered light from dust in the optical bench If so carefully clean the optical bench Disconnect the input board from the interface board by unplugging ribbon cable labelled INPUT The instrument reading should drop to zero or to a negative value Instrument appears to zero but there is weak or Span cylinder empty Check the source pressure Model 43i Instruction Manual 6 3 Troubleshooting Troubleshooting Guides 6 4 Model 43i Instruction Manual Malfunction Possible Cause Action no response to span gas Calibration system failure Flow rate of the diluted span mix is inadequate Instrument is not drawing in span gas SO is being absorbed by tubing filters or dirt in the calibration system Flash lamp has failed PMT or input board has failed Check solenoids or other hardware to be sure that span gas Is being delivered Check by pass or atmospheric vent to verify that the zero air system Is providing more flow than the instrument draws Check sample Flow and Pressure readings on the Diagnostics screen Use an independent flow meter to check flows at the sample inlet and exhaust bulkheads they should match Perform a leak test as described in the Preventive Maintenance chapter Replace any lines made of vinyl or other plastics with fresh Teflon or stainless steel Replace Teflon filter membranes that look dirty Remove any filters that are not Teflon
65. screen 8 At the Pressure Sensor Cal screen press to select Span The Calibrate Pressure Span screen appears 9 Wait at least 10 seconds for the ambient reading to stabilize use e Jand to enter the known barometric pressure and press _ to save the pressure value 10 Reconnect the instrument tubing to the pressure transducer 11 Install the cover Thermo Fisher Scientific Model 431 Instruction Manual 7 29 Servicing Flow Transd ucer Use the following procedure to replace the flow transducer Figure 7 14 Replacement Equipment Required Flow transducer Philips screwdriver Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Turn instrument OFF unplug the power cord and remove the cover 2 Disconnect the plumbing connections from the flow transducer Note the plumbing connections to facilitate reconnection 3 Disconnect the flow transducer cable from the FLOW connector on the measurement interface board 4 Remove the two screws securing the flow transducer to the kicker bracket and remove the flow transducer 5 Install the new flow
66. selected item is shown by lt after it In dual or auto range mode HI or LO is displayed to indicate high or low range concentrations In the Main Menu choose Instrument Controls gt Communication Settings gt Streaming Data Config gt Select Item gt Other Measurements Analog Inputs The Analog Inputs screen allows the user to select the analog input signal none or analog inputs 1 8 that is tied to the selected streaming data item The selected item is shown by lt after it In the Main Menu choose Instrument Controls gt Communication Settings gt Streaming Data Config gt Select Item gt Analog Inputs 3 36 Model 43i Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu Ea rir ETOES cn Lea EUs EEE o Bone nr E dd RS 232 RS 485 Selection The RS 232 RS 485 Selection screen allows the user to choose between the RS 232 or RS 485 specification for serial communication 1 Equipment Damage Disconnect the serial cable before changing RS 232 and RS 485 selection to prevent damage to any equipment currently connected to the analyzer A In the Main Menu choose Instrument Controls gt Communication Settings gt RS 232 RS 485 Selection a a ee Ea my O pH mpi eo ee Lira E A pH piri PPLE Pil BAL PRI Pihke fini i PLETE i a hai TCP IP Settings The TCP IP Settings menu is used for defining TCP IP settings Note The instrument power must be cycled afte
67. standard temperature is different from 20 C A display warning will appear that ranges will be defaulted and calibration parameters reset A The SO Range screen defines the concentration range of the analog outputs For example a SO range of 0 50 ppb restricts the analog output to concentrations between 0 and 50 ppb The display shows the current SO range The next line of the display is used to change the range The range screen is similar for the single dual and autorange modes The only difference between the screens are the words High or Low displayed to indicate which range is displayed The example that follows shows the SO range screen in single mode For more information about the dual and autorange modes see Single Range Mode Dual Range Mode and Autorange Mode earlier in this chapter Table 3 5 lists the available standard ranges Table 3 6 lists the extended ranges When switching from standard to extended ranges the PMT voltage must be readjusted For more information about readjusting the PMT voltage see the Servicing chapter In the Main Menu choose Range gt SO Range Model 43i Instruction Manual 3 13 Operation Range Menu POE E TEE E uee Beane Table 3 5 Standard Ranges ppb ppm pg m mg m 50 0 05 200 02 100 010 50 05 200 0 20 1000 1 500 0 50 2000 2 1000 1 5000 5 2000 2 10000 10 5000 5 20000 20 10000 10 25000 25 C1
68. sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Turn instrument OFF unplug the power cord and remove the cover 2 Disconnect the plumbing connections from the flow transducer the optical bench and the rear bulkhead Note the plumbing connections to facilitate reconnection 3 Cut the ties securing the kicker assembly to the kicker bracket and remove the kicker Thermo Fisher Scientific Servicing Heated Kicker Replacement 4 Install the new kicker following the previous steps in reverse Kicker Ties Kicker Bracket Flow Transducer Figure 7 16 Replacing the Kicker Assembly Standard Kicker Heated Kicker Use the following procedure to replace the heated kicker assembly Figure 1 17 Replacement Equipment Required Heated kicker assembly optional Allen wrench 5 32 inch amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A Equipment Damage Some internal components can be damaged by small 1 Turn instrument OFF unplug the power cord and remove the cover 2 Using a 5 32 inch Allen wrench remove the four optical bench ret
69. the power supply to the bracket and remove the power supply 6 To install the power supply follow the previous steps in reverse 7 Recalibrate the instrument Refer to the calibration procedures in the Calibration chapter PMT Voltage Use the following procedure to adjust the PMT voltage after replacing the z PMT High Voltage Power Supply or switching from standard to extended Adj ustment ranges or vice versa Note The service procedures in this manual are restricted to qualified service representatives A 1 Select the SO range Refer to Range Menu in the Operation chapter Thermo Fisher Scientific Model 431 Instruction Manual 7 19 Servicing DC Power Supply Replacement 7 20 DC Power S upply Replacement Model 43i Instruction Manual A 2 Reset the SO BKG calibration factors to 0 0 Refer to Calibration Factors Menu in the Operation chapter 3 Reset the SO COEF to 1 000 4 Set the Averaging Time to 10 seconds Refer to Averaging Time in the Operation chapter 5 Connect the calibration gas and allow the instrument to sample calibration gas until the reading stabilizes 6 From the Main Menu press _ to scroll to Service gt press gt to scroll to PMT Supply Settings gt and press The PMT Supply Settings screen appears Note If Service Mode is not displayed refer to Accessing the Service Mode in this chapter then return to
70. then the first logged data record that is displayed is the first record after this time If set to one minute logging this would be at 20 JUN 2005 10 01 Thermo Fisher Scientific Model 43i Instruction Manual 3 25 Operation Instrument Controls Menu DH pikari Ce a E SO a EA Record Display The Record Display screen read only displays the selected records E a i a TE Erase Log The Erase Log is used to erase all saved data for the selected record type not all short records and long records In the Main Menu choose Instrument Controls gt Datalogging Settings gt Erase Log CREED RGN PTEN i CH ciinii E PUHE Eut Beee ETET hei Bast i i y iNi Lab LBA PF EH girs AS EHH i i et Select Content The Select Content submenu displays a list of 32 record fields to use and a submenu list of the analog output signal group choices to choose from Choices are Concentrations Other Measurements and Analog Inputs if the I O expansion board is installed This is a temporary list of items for the selected record type that must be committed via the datalogging menu before the changes will apply Note that committing any changes to this list will erase all currently logged data as the format of the stored data is changed 3 26 Model 43i Instruction Manual Thermo Fisher Scientific Choose Item Type Thermo Fisher Scientific Concentrations Operation Instrument Controls Menu In the Main Menu
71. to this unit not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment Note This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense amp The following symbol and description identify the WEEE marking used on the instrument and in the associated documentation Symbol Description Marking of electrical and electronic equipment which applies to electrical and electronic equipment falling under the Directive 2002 96 EC WEEE and the equipment that has been put on the market after 13 August 2005 4 Model 437 Instruction Manual ix About This Manual Where to Get Help Where to Get Help Service is available from exclusive distributors worldwide Contact one of the phone numbers below for product support and technical information or visit us on the web at www thermo com aqi 1 86
72. to view and set the permeation oven thermistor resistance The low and high point screens function the same way In the Main Menu choose Service gt Perm Oven Settings gt Factory Cal Oven Therm gt Low Point Set Defaults The Set Defaults screen allows the user to reset the configuration values to Extended Ranges 3 84 Model 43i Instruction Manual factory defaults In the Main Menu choose Service gt Perm Oven Settings gt Factory Cal Oven Therm gt Set Defaults The Extended Ranges screen is used to turn the extended ranges feature ON and OFF The extended ranges is visible only when the instrument is in service mode For more information on the service mode see Service Mode described previously in this chaper Thermo Fisher Scientific Dilution Ratio Display Pixel Test Thermo Fisher Scientific Restore User Defaults Operation Service Menu The Dilution Ratio screen allows the user to view and set the dilution ratio Acceptable values are 1 500 1 The default is 1 1 When this value is set the dilution ratio is applied to all concentration measurements This screen is only accessible if the dilution ratio option is installed In the Main Menu choose Service gt Dilution Ratio The Display Pixel Test is used to test the LCD display The display pixel test screen is visible only when the instrument is in service mode For more information on the service mode see Service Mode ear
73. touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Turn instrument OFF unplug the power cord and remove the cover 2 Disconnect heaters from the AC BENCH cable 3 Using the 5 32 inch wrench remove the four screws securing the optical bench to the floor plate Figure 7 6 4 Lift the optical bench from the floor plate to gain access to the heater assembly 5 Remove two retaining screws and washers from each heater and remove the heaters Figure 7 15 6 Apply heat sink grease to the new heaters as appropriate 7 Fasten each heater with the two retaining screws and washers 8 Secure the optical bench to the floor plate with the four screws 9 Connect the heaters to the AC BENCH cable Model 431 Instruction Manual 7 33 Servicing Kicker Replacement Retaining Screws 2 each heater Kicker Replacement A 7 34 Model 43i Instruction Manual 10 Replace the cover and plug in the power cord Heaters Figure 7 15 Replacing the Heater Assembly Use the following procedure to replace the kicker assembly Figure 7 16 Equipment Required Kicker assembly Cable ties Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be
74. translated before they are compared to the selection list options L Indicates a button which pops up a selection list without any translation The output value is number of the selected option T Indicates a button which pops up a selection list with output translation The number of the option selected is used as an index into the translation table to generate an output string N Indicates a button which only sends the subsequent command to the instrument No user prompting happens Some examples n is the C syntax for an end of line character Concentrations n This is a single text only line An This is a single blank line NO 3s n Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Record Layout Definition This is a line which appears slightly indented The text field is NO the value is taken from the third element of the data response and interpreted as a string NO 18sBd ddd set no coef s n This is a line which also appears slightly indented The next field is also NO but the value is taken from the eighteenth element of the data response again interpreted as a string A button appears on this line which when pressed pops up an input dialog which will state Please enter a new value for NO using a d ddd format The string entered by the user is used to construct the output command If the user enters for example 1 234 the constructed command will be set
75. two averaging times high and low The averaging time screen functions the same way in the single dual and autorange modes The following averaging times are available 1 2 5 10 20 30 60 90 120 180 240 and 300 seconds In the Main Menu choose Averaging Time Calibration factors are used to correct the SO concentration readings that the instrument generates using its own internal calibration data The Calibration Factors menu displays the calibration factors The screens that follow show the calibration factors menu in single mode and dual autorange modes The only differences between the screens are the words HI and LO to indicate which range is displayed Normally the instrument is calibrated automatically using the commands listed in the Calibration menu described later in this chapter However the instrument can also be calibrated manually using the Calibration Factors menu Thermo Fisher Scientific S02 Background Thermo Fisher Scientific Operation Calibration Factors Menu To manually calibrate the instrument see SO Background and SO Span Coefficient that follows for more information In the Main Menu choose Calibration Factors co ees aran a ALTERAR ni hat h POES e Pi EE EE Eio a E La EEY id hab is E The SO Background correction is determined during zero calibration The SO background is the amount of signal read by the analyzer while sampling zero air Alt
76. using lower case letters to easily distinguish them from the operating screens To exit a help screen press Jor to return to the previous screen or to return to the Run screen e JH Up Down e gt Left Right The four arrow pushbuttons e Ce and En move the cursor up down left and right or change values and states in specific screens Enter The is used to select a menu item accept set save a change and or toggle on off functions Model 43i Instruction Manual 3 3 Operation Software Overview Soft Keys Software Overview 3 4 Model 43i Instruction Manual The Soft Keys are multi functional keys that use part of the display to identify their function at any moment The function of the soft keys allows immediate access to the menu structure and most often used menus and screens They are located directly underneath the display and as the keys functions change this is indicated by user defined labels in the lower part of the display so that the user knows what the keys are to be used for To change a soft key place the menu cursor gt on the item of the selected menu or screen you wish to set Press gt followed by the selected soft key within 1 second of pressing the right arrow key The edit soft key prompt will be displayed for configuration for the new label Note Not all menu items may be assigned to soft keys If a particular menu or screen item cannot be assigned t
77. valves from the run screen The lower left hand corner of the Run screen indicates which mode is active ZERO SPAN or SAMPLE 2 To ensure that the zero air is being measured at atmospheric pressure check that the zero air flow reported by the analyzer is approximately 0 5 LPM Actual zero air flow should be 0 8 LPM or more with the excess leaving out the atmospheric bypass a From the Run screen press Cm to display the Main Menu b From the Main Menu choose Diagnostics gt Flow to display the Sample Flow screen c Make sure that the reported sample flow is less than the actual zero air flow 3 Press to monitor the zero air reading and wait for the reading to stabilize 4 Press and choose Calibration gt SO Background to display the SO Background screen 5 When the reading stabilizes press to set the SO reading to zero The display flashes the message SAVING and the SO reading is added to the background correction Record the stable zero air reading as Zso2 Thermo Fisher Scientific Thermo Fisher Scientific 10 11 12 13 Calibration Multipoint Calibration in Dual Auto Range Mode Press to return to the Calibration menu Connect a source of calibration gas to the SAMPLE bulkhead The calibration gas should be approximately 80 of the high full scale range To ensure that the calibration gas is being measured at atmospheric pressure check that the flow repo
78. word ALARM appears in the Run screen and in the Main Menu In the Main Menu choose Alarms gt Pressure mph aa EN EEE Eek Boes ETOT Fad pri ii Min and Max Pressure The Minimum Pressure alarm limit screen is used to change the minimum Limits pressure alarm limit The minimum and maximum pressure screens function the same way In the Main Menu choose Alarms gt Pressure gt Min or Max Sample Flow The Sample Flow screen displays the current sample flow reading and sets the minimum and maximum alarm limits Acceptable alarm limits range from 0 to 2 5 LPM If the sample flow reading goes beyond either the minimum or maximum limit an alarm is activated The word ALARM appears in the Run screen and in the Main Menu In the Main Menu choose Alarms gt Sample Flow Thermo Fisher Scientific Model 43i Instruction Manual 3 65 Operation Alarms Menu Min and Max Sample Flow Limits Lamp Intensity Min and Max Lamp Intensity Limits 3 66 Model 43 Instruction Manual The Minimum Sample Flow alarm limit screen is used to change the minimum sample flow alarm limit The minimum and maximum sample flow screens function the same way In the Main Menu choose Alarms gt Sample Flow gt Min or Max The Lamp Intensity screen displays the current lamp intensity reading and sets the minimum and maximum alarm limits Acceptable alarm limits range from 20 to 100 percent If the lamp intensity reading
79. 0 000 Smpfl 0 440 Pres 715 478 avgt 10 lo avgt 10 SO2bkg 6 795 502 coef 1 000 lo S02 coef 1 000 SO2range 10000 000 lo SO2range 10000 000 Lmpi 92 000 lrec srec lrec xxxx yy SEEC XXXX YY lrec aa bb 00 pp qq yy srec aa bb 00 pp qq yy xxxx the number of past records yy the number of records to return 1 to 10 aa hours 01 to 24 bb minutes 01 to 59 oo month 01 to 12 pp day 01 to 31 qq year These commands output long or short records The output format is determined by the set Irec format and set srec format commands The logging time is determined by the set lrec per and set srec per commands In the following example there are 740 long records currently stored in memory When the command lrec 100 5 is sent the instrument counts back 100 records from the last record collected record 740 and then Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Datalogging returns 5 records 640 641 642 643 and 644 For details on how to decode the flag fields within these records see the flags command Send lrec 5 Receive lrec 100 5 18 09 05 12 03 flags 5c000044 S02 9658E 5 low S02 8920E 5 Intt 28 6 Rctt 25 6 Pgast 0 00 Pres 720 3 Smpfl 0 489 PMt V 808 LmpV 449 Lmpi 4740E 1 18 10 05 12 03 flags 5c000044 S02 6923E 5 low S02 6429E 5 Intt 28 6 Rctt 25 6 Pgast 0 00 Pres 720 3 Smpfl 0 489 PMt V 808 LmpV 449 Lmpi 4740E 1 18 11 05 12 03 flags 5c0000
80. 0 HMO PU l a a a a a a There is an optional single digit d which may follow any of the numeric fields which indicates that after the field has been parsed out the resulting value is to be divided by 104d Thus the 16 bit field OxFFC6 would be interpreted with the format specifier n3 as the number 0 058 The subsequent lines in the ERec Layout response describe the appearance of the full panel The full instrument panel as it appears on the screen has two columns of lines Each line is composed of three major components 1 a text field 2 a value field and 3 a button None of these three components is required The text field contains statically displayed text The value field displays values which are parsed out of the response to a DATA ERec command It also displays though background changes alarm status The button when pressed triggers input from either a dialog box or a selection list There are five kinds of buttons B I L T and N Each line in the layout string corresponds to one line on the display The layout string describes each of the three major fields as well as translation mechanisms and corresponding commands Thermo Fisher Scientific Text Value String Value Source Alarm Information Translation Table Thermo Fisher Scientific C Link Protocol Commands Record Layout Definition The first field in the layout string is the text It is delimited by a The string up to the first
81. 04 sabeyop Indul Bojeuy w qd synejag 129 1SN jesey s uey wojsng jos 1e9 dway 19949 019Z sHulpeay induj Bojeuy Apo S91A19S os 07 189 jaog 205 07 abuey os 07 1531 peog Indu abeyoA dweT 1581 ueds jeondo ISENuOg U3919S OS IH 189 209 70S H ew Bay 07 abuey Os H 189 mols ysuaju due ysuajuj dwe7 uonesuedwioy said yg ogies Byg os oui SAY IH sun sen 189 aanssaid MO 4 9jdwes moj4 uonesuadwog dwa J909 9S spo obuey 21NSSIId ainsseld 6yuog oll py a6e110A Ld dua saquieyd sdwe sBuijes wwog x99y9 ueds o1ez synejaq 189 Jasn Jesey sebuey wojsng Jos 398 UySe 4 jenu dua jeusaquy sobeyon sbuijes BubBojejeg 3909 oge 4209 OS afuey os py Beyo ysela papaq suey UOISI9A WesGO1g dwe7 ysel yg os 180 Byg os au Bay suun sen S 01 UO S10 9 oul 8913S suuely sonsoubeig sui aa uonesques uon EN ae o bulb ae obuey nuay uen uaas9g uny uSeJ9s 1591 1195 u 19S dn 1amodg Figure 3 3 Flowchart of Menu Driven Software 3 5 Model 43i Instruction Manual Thermo Fisher Scientific Operation Software Overview Power up Screen The Power up screen is displayed when power is applied to the Model 432 Run Screen This screen is displayed while the internal components are warming up and diagnostic checks are being performed Thermo SCIENTIFIC iSeries Please wait while booting The Run screen displays the SO concentration The statu
82. 10 LMA GMT 11 DLE GMT 12 EDT GMT 5 4 CDT GMT 6 5 MDT GMT 7 6 and PDT GMT 8 7 In the Main Menu choose Instrument Controls gt Timezone Thermo Fisher Scientific Operation Diagnostics Menu Diagnostics Menu The Diagnostics menu provides access to diagnostic information and functions This menu is useful when troubleshooting the instrument The analog input readings and analog input voltages are only displayed if the I O expansion board option is installed In the Main Menu choose Instrument Controls gt Diagnostics Program Versions The Program Versions screen read only shows the version numbers of the 8 y programs installed Prior to contacting the factory with any questions regarding the instrument please note the product model program version numbers In the Main Menu choose Diagnostics gt Program Versions Thermo Fisher Scientific Model 43i Instruction Manual 3 55 Operation Diagnostics Menu Voltages The Voltages menu displays the current diagnostic voltage readings This screen enables the power supply to be quickly read for low or fluctuating voltages without having to use a voltage meter The I O board item is only displayed if the I O expansion board is installed In the Main Menu choose Diagnostics gt Voltages ERED CH Py Timms Lid bal E E CH CDE Pim Toes ria Motherboard Voltages The Motherboard screen read only is used to display the current voltage reading
83. 3 Pin da 3 Pini 24 DC POWER SUPPLY REAR PANEL AC 24VDC AC INPUT 3 Cond Pl 2 Pin me FRONT PANEL gt POWER SW E ORO a PJG i TRANSFORMER wae 3 Pin AAA AN E OOVAC or 240VAC oats 1 CPTIONS I Figure 6 1 Board Level Connection Diagram Common Electronics 6 10 Model 43i Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Troubleshooting Board Level Connection Diagrams TRIGGER PAK 3 Pin FLASH 4 Pin INTENSITY TO COMMON DIAGNOSTIC LED ELECTRONICS 2 Pin INPUT BOARD Coax E 5 10 Pin J 2 Cond o OO PM 7 Pin 2 Cond PRESSURE 4 Pin FLOW MEASUREMENT 5 Pin INTERFACE BOARD LL AMBIENT TEMP THERM 2 Pin RS RR a aa E RR 1 _ GAS THERM i DATA 2Pin PERM OVEN L_ HEATER 15V 1W optional 3 Pin POJ amp OVEN THERM I Mei os ee au au nan A _ ZERO SPAN 2 Pin L SAMPLE SOL f CONVERTER SOL Not 2 Pin Ll span 2 sor 2 Pin 24V 2 Pin PAN 2 Pin Pa CONVERTER 6 Pin NEEN OE AC CONNECTORS e 3 Pin Not Used 3 i PJ BENCH HEATER AND 3 Pin f 4 Pin AC i EET PUMP Figure 6 2 Board Level Connection Diagram Measurement System Model 431 Instruction Manual 6 11 Troubleshooting Connector Pin Descriptions Connector Pin The connector pin descriptions in Table 6 5 through Table 6 12 can be used along with the board level connection diagrams to troubleshoot Desc riptions board level faults Table 6 5 Motherboard Connector Pin Descriptions C
84. 4 Analog Input Calibration ici iii 7 25 Calibrating the Input Channels to Zero Volts osses 7 25 Calibrating the Input Channels to Full Scale 7 26 Pressure Transducer Assembly Replacement csseseseeseeeereeeeeees 7 27 Pressure Transducer Calibre 7 28 Flow Transducer Replacement 2 idas 7 30 Flow Transducer alibration ts ios vesenifenvoas seasteduteseaset adi een doayngtadaal beeen 7 31 Heater Assembly Replacement socia tia 7 33 Kicker Replacement ratio 7 34 Heated Kicker Replace e de it Ghee 7 35 Thermistor Replace ote 7 36 Ambient Temperature Calibrati n aa cen tedace Ree 7 37 Thermo Fisher Scientific Thermo Fisher Scientific Contents Input Board Replacement etnia land 7 39 I O Expansion Board Optional Replacement sssssseeeseseeeeeeeeeeeeeese 7 40 Digital Output Board Replacement aii 7 41 Motherboard Replacement iii 7 42 Measurement Interface Board Replacement somos 7 43 Front Panel Board Replicante aia 7 44 LCD Module Replacements o at aa pcs 7 45 Service LOCO aaa 7 46 System Description sscsscsscssessssssssecsssssessessesesssessesesassaessessneesessesaesaeeatees 8 1 Hardwarea nee bec a tat aa tant are N ee E a 8 1 Hydrocarbon Kicker ad 8 2 Heated Hydrocarbon Kicker Optional ooocconnicnnnnnonnnnenencno 8 2 OPUS A A AA 8 3 Pasan ps o adria ode 8 3 Condensing Lens vda acu rd teat shes a ios 8 3 Mitror Assembly A O a earache Aca 8 3 A O 8 3 Flash Lamp Trigger Assembly icons it 8 3 Reaction Chambers
85. 44 S02 8814E 5 low S02 9488E 5 Intt 28 6 Rctt 25 6 Pgast 0 00 Pres 720 3 Smpfl 0 489 PMt V 808 LmpV 449 Lmpi 4740E 1 18 12 05 12 03 flags 5c000044 SO2 7313E 5 low S02 6997E 5 Intt 28 6 Rctt 25 6 Pgast 0 00 Pres 720 3 Smpfl 0 489 PMt V 808 LmpV 449 Lmpi 4740E 1 18 13 05 12 03 flags 5c000044 S02 8168E 5 low S02 8237E 5 Intt 28 6 Rctt 25 6 Pgast 0 00 Pres 720 3 Smpfl 0 489 PMt V 808 LmpV 449 Lmpi 4740E 1 erec format Irec format srec format These commands report the output format for long and short records and dynamic data in various formats such as ASCII without text ASCII with text or binary The example that follows shows the output format for long records is ASCII with text according to Table B 4 Send lrec format Receive lrec format 01 set erec format format set lrec format format set srec format format These commands set the output format for long and short records and dynamic data according to Table B 4 The example that follows sets the long record output format to ASCII with text Send Receive set lrec format 1 set lrec format 1 ok Model 43 Instruction Manual B 19 C Link Protocol Commands Datalogging B 20 Model 43 Instruction Manual Table B 4 Record Output Formats Format Output Format 0 ASCII no text 1 ASCII with text 2 binary data erec layout Irec layout srec layout These commands reports the layout string indicating the data formats
86. 48 hours Carefully calibrate the 437 using the Transfer Standard The output of the Transfer Standard should be connected to the SAMPLE bulkhead on the rear panel of the 432 Thermo Fisher Scientific Heated Hydrocarbon Kicker Teflon Particulate Filter 1 0 Expansion Board Assembly 25 Pin Terminal Board Assembly Terminal Block and Cable Kits Thermo Fisher Scientific Optional Equipment Heated Hydrocarbon Kicker 5 Switch the calibrated 437 into the span mode 6 Measure the flow rate into the ZERO bulkhead on the rear panel of the 431 Be sure that the charcoal scrubber is connected Note the flow and measured SO concentration 7 From the flow and measured concentration compute the permeation tube release rate The optional heated hydrocarbon kicker is used in applications requiring more efficiency than the standard kicker The heated kicker operates at the same temperature as the optical bench and requires no additional power A 5 10 micron pore size two inch diameter Teflon element is available for the Model 43i This filter should be installed just prior to the SAMPLE bulkhead When using a filter all calibrations and span checks must be performed through the filter The 1 O expansion board provides six analog current output channels 0 20 mA or 4 20 mA and eight analog voltage inputs 0 10V The DB25 connector on the rear panel provides the interface for these inputs and outputs The 25 pin terminal
87. 5 2 1 1 5 2 2 10 3 5 20 4 10 50 5 20 100 6 50 200 7 100 250 8 C1 C1 9 C2 C2 10 C3 C3 custom range range 1 2 3 This command reports the user defined value of custom range 1 2 or 3 The example that follows reports that custom range 1 is defined to 5 50 Send custom 1 Receive custom 1 550E 2 ppm set custom range range value set custom 1 range value set custom 2 range value set custom 3 range value These commands are used to set the maximum concentration for any of the three custom ranges 1 2 or 3 to range value where value is a floating point number representing concentration in ppm or mg m The example that follows sets the custom 1 range to 55 5 ppm Send set custom 1 range 55500 Receive set custom 1 range 55500 ok Model 43 Instruction Manual B 29 C Link Protocol Commands Measurement Configuration B 30 Model 43 Instruction Manual range mode This command reports the current range mode Send range mode Receive range mode single set range mode mode This command sets the current range mode to single dual or auto The example that follows sets the range mode to single Send set range mode single Receive set range mode single ok gas mode This command reports the current mode of sample zero or span The example that follows reports that the gas mode is sample Send gas mode Receive gas mode sample set sample set sample gas This command sets the zero span valves to the sample
88. 6 282 0430 Toll Free 1 508 520 0430 International x Model 43 Instruction Manual Thermo Fisher Scientific Chapter 1 Chapter 2 Chapter 3 Thermo Fisher Scientific Contents PENCE eA CMC ENGIN ica 1 1 Principle OL Operators colon waa eae tahoe 1 2 SPECICA TIONS dd lista 1 3 Italia 2 1 A ta oe PRES od E Ee EE PIED OTST ODE EO POT 2 1 Unpacking and Inspection its 2 1 A O A 2 3 Connecting External Devices oriol raid 2 5 Terminal Board PCB Assemblies cc cssccccecssecsncecasecseeesi cesscnaa anna 2 5 I O Terminal Bondad 2 5 D O Terminal Dori in a 2 7 25 Pip Terminal Board nassib 2 8 SUIT UES cst as sonnei A ode hang secemunianeres 2 9 TI sass se incu che vst dhs cv onesies wean Wenn aban cued even AAN 3 1 Displayen inai a aA i AA ATEA ich ae dein 3 2 P shbuttons a ia SANOS 3 3 Soft Keys A 3 4 Software Overview sesessssssssessssesessesressssessessessesessestessssessesseseesesseseese 3 4 Power up SCLCe Ths tes das tes casucsee aid 3 6 R Scheer isan eee ee erate ae ay eE oda AT ENNS 3 6 A etiri t be Nvahwect hasta ean di E E aE 3 7 Range MED A e 3 8 Single Range Modera cd 3 8 Dual Range Mode de Mit aa aed a 3 9 Autorange Modet raran cues a uit halo 3 10 Gases dia a E a Us 3 12 SO RE eds 3 13 SEE CUSCO RANEES cintia 3 15 Custom Ranges ainia id 3 15 AVERAGING DIME ii 3 16 Calibration Factors Menu s sssssessssessssessseesesesssststseesesesesessssesesreeesese 3 16 SOZ Background sirena in an i i a be EAN I 3 17 SOS pani
89. 8 2 RS 485 DB Connector Pin Configuration DB9Pin Function 2 receive 8 receive 7 transmit 3 transmit 5 ground An RJ45 connector is used for the 10Mbs Ethernet connection supporting TCP IP communications via standard IPV4 addressing The IP address may be configured for static addressing or dynamic addressing set using a DHCP server Any serial port protocols may be accessed over Ethernet in addition to the serial port The external accessory connector is not used in the Model 431 analyzer This port is used in other models to communicate with smart external devices that may be mounted hundreds of feet from the analyzer using an RS 485 electrical interface Thermo Fisher Scientific Chapter 9 Optional Equipment Internal Ze ro S pan With the zero span assembly option a source of span gas is connected to dS le Val the SPAN port and a source of zero air is connected to the ZERO port an ample ValvesS zoo and span gas should be supplied at atmospheric pressure It may be necessary to use an atmospheric dump bypass plumbing arrangement to accomplish this See the Installation and Operation chapters for more information Internal Permeation The Internal Permeation Span Source option is designed to provide a simple span check This option is intended as a quick convenient check to Span Source be used between zero and span calibrations for determining instrument malfunction or drift Because this o
90. 810 V led status led These commands report the status of the optical test LED on or off The example that follows reports that the optical test LED is on Send lamp status Receive lamp status on set led onoff These commands set the optical test LED on or off The example that follows turns the optical test LED off Send set lamp off Receive set lamp off ok clr records This command will clear all long and short records that have been saved Send clear records Receive clear records ok set clr Irecs set clr srecs These commands will clear only the long records or only the short records that have been saved The example that follows clears short records Send set clr srecs Receive set clr srecs ok Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Datalogging list din list dout These commands report the current selection for the digital outputs in the format Output no Index number variable name active state The active state for digital outputs is open or closed The active state for digital inputs is high or low Send list dout Receive list dout output index variable state 1 28 CONC ALARM closed 2 2 LOCAL REMOTE open 3 4 UNITS closed 4 11 HI S02 CONC open 7 7 SAMPLE MODE open 8 8 S02 MODE open list Irec list srec list stream list sp These commands report the list of current selections for long record logging data short record logging data streaming da
91. ANY PERSON OR ENTITY OTHER THAN SELLER WITHOUT SELLER S PRIOR WRITTEN APPROVAL OR ANY USE OF REPLACEMENT PARTS NOT SUPPLIED BY SELLER SHALL IMMEDIATELY VOID AND CANCEL ALL WARRANTIES WITH RESPECT TO THE AFFECTED PRODUCTS THE OBLIGATIONS CREATED BY THIS WARRANTY STATEMENT TO REPAIR OR REPLACE A DEFECTIVE PRODUCT SHALL BE THE SOLE REMEDY OF BUYER IN THE EVENT OF A DEFECTIVE PRODUCT EXCEPT AS EXPRESSLY PROVIDED IN THIS WARRANTY STATEMENT SELLER DISCLAIMS ALL OTHER WARRANTIES WHETHER EXPRESS OR IMPLIED ORAL OR WRITTEN WITH RESPECT TO THE PRODUCTS INCLUDING WITHOUT LIMITATION ALL IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE SELLER DOES NOT WARRANT THAT THE PRODUCTS ARE ERROR FREE OR WILL ACCOMPLISH ANY PARTICULAR RESULT A 2 Model 43 Instruction Manual Thermo Fisher Scientific Instrument Identification Thermo Fisher Scientific Number Appendix B C Link Protocol Commands This appendix provides a description of the C Link protocol commands that can be used to remotely control a Model 43 analyzer using a host device such as a PC or a datalogger C Link protocol may be used over RS 232 RS 485 or Ethernet C Link functions can be accessed over Ethernet using TCP IP port 9880 Streaming data may be accessed over Ethernet using TCP IP port 9881 e Instrument Identification Number on page B 1 e Commands on page B 2 e Measurements on page B 8 e Alarms on page B 11
92. APPED WHEN NOT IN USE TO INLET OF ANALYZER UNDER CALIBRATION Figure 4 1 Cylinder Gas Dilution System The exact SO concentration is calculated from SO x F so SO ligor f sro 0 Fp Fso Where SOz our diluted SO concentration at the output manifold ppm SOz srp concentration of the undiluted SO standard ppm Esoz flow rate of the SO standard corrected to 25 C and 760 mm Hg Fp flow rate of the dilution air corrected to 25 C and 760 mm Hg Commercial precision dilution systems are available which reliably and accurately dilute a high concentration gas mixture to provide a reliable span gas A high concentration 50 ppm of SO in air is precisely diluted to the concentration range required The Thermo Scientific Model 146 Multigas Calibration System is one such system for precision dilution Permeation tube systems which precisely maintain a set temperature to within 0 1 C and hold a zero air flow rate to within 0 5 can be used for generation of span gas The flow rate of the permeation system must be at least 0 5 LPM for proper operation Model 43i Instruction Manual 4 3 Calibration Permeation Tube System A permeation tube system shown in Figure 4 2 can be constructed All connections between components in the system should be made with glass Teflon or other non reactive material The air flow controllers should be capable of maintaining a constant air flow within 2 of the required
93. Board Connector Pin Descriptions Connector Label Reference Designator Pin Signal Description PMT IN J1 1 2 PMT Input Ground INTF BD J2 oOo o N on A UO N gt Thermo Fisher Scientific 15V Ground 15V 5V Ground Measurement Frequency Output Sample Hold NC Gain A Gain B Model 431 Instruction Manual 6 23 Troubleshooting Service Locations Table 6 11 Flash Trigger Pack Pin Descriptions Connector Label Reference Designator Pin Signal Description FLASH TRIG TO J01 1 Flash Voltage 2 Flash Trigger 3 Ground Table 6 12 Flash Intensity Assembly Pin Descriptions Connector Label Reference Designator Pin Signal Description FLASH INT 1 15V 2 15V 3 Ground 4 Lamp Intensity Se rvice For additional assistance service is available from exclusive distributors x worldwide Contact one of the phone numbers below for product support Locations and technical information or visit us on the web at www thermo com aqi 1 866 282 0430 Toll Free 1 508 520 0430 International 6 24 Model 43i Instruction Manual Thermo Fisher Scientific Safety Precautions Thermo Fisher Scientific Chapter 7 Servicing This chapter explains how to replace the Model 43i subassemblies It assumes that a subassembly has been identified as defective and needs to be replaced For fault location information refer to the Preventive Maintenance chapter and the Troubleshooting chapter in this manual The
94. C flow This command reports the current sample flow The example that follows reports that the current sample flow is 0 503 liters minute Send flow Receive flow 0 503 1 m internal temp This command reports the current internal instrument temperature The first reading is the temperature being used in instrument calculations The second temperature is the actual temperature being measured If temperature compensation is on then both temperature readings are the same If temperature compensation is off a temperature of 30 C is used as the default temperature even though the actual internal temperature is 27 2 C The example that follows shows that temperature compensation is on and that the internal temperature is 27 2 C Send internal temp Receive internal temp 27 2 deg C perm gas temp This command reports the current permeation gas temperature The example that follows reports that the permeation gas temperature is 45 C Send perm gas temp Receive perm gas temp 45 deg C Model 431 Instruction Manual B 9 C Link Protocol Commands Measurements B 10 Model 43 Instruction Manual pmt voltage This command reports the current PMT voltage The example that follows reports that the current PMT voltage is 510 volts Send pmt voltage Receive pmt voltage 510 volts pres This command reports the current reaction chamber pressure The first pressure reading is the pressure reading being used in instrument calc
95. DHCP see Use DHCP in this chapter In the Main Menu choose Instrument Controls gt Communication Settings gt TCP IP Settings gt Netmask Gateway The Default Gateway screen is used to edit the gateway address The default gateway can only be changed when DHCP is OFF For more information on DHCP see Use DHCP in this chapter Any traffic to addresses that are not on the local subnet will be routed through this address In the Main Menu choose Instrument Controls gt Communication Settings gt TCP IP Settings gt Gateway Cu gu Bs ETE EE oa Pia EE EE ERED CHa o E Host Name The host name screen is used to edit the host name When DHCP is enabled this name is reported to the DHCP server In the Main Menu choose Instrument Controls gt Communication Settings gt TCP IP Settings gt Host Name Thermo Fisher Scientific Model 43i Instruction Manual 3 39 Operation Instrument Controls Menu 3 40 Network Time Protocol Server 1 0 Configuration Model 43i Instruction Manual Eo EPOPEE Road Booe The Network Time Protocol NTP Server screen is used to edit the IP address of the NTP server An NTP server may be used to periodically synchronize the instrument s real time clock with a standard More information about the NTP servers and a list of public servers may be found at http www ntp org In the Main Menu choose Instrument Controls gt Communication Settings
96. Defaults ina 3 83 Factory Calibrate Oven Tbetiiistor ss sisescccscsasenisesarsevavanedsesetanceoes 3 83 Lowand High Points nensis sei 3 84 Set Defaults ana R A NR N 3 84 Extended Ran e 3 84 A O EE Calctatacare con sdensgesncaucevessbedetateaee 3 85 Display Pixel Testi 3 85 Restore User Del di ica 3 85 Password 3 86 Set Password os itatis A IO 3 86 Lock Instrument coda cote ieee tea a 2h E E 3 87 Change Passwords sa a a e 3 87 Re mov PassWord ada da 3 88 Unlock Instrument ao 3 88 Calibr ti i asaan aa aa aaan eaa aas 4 1 Zero Gas Genta i n 4 1 Commercial Heatless Diaria 4 1 Absorbing Column 4 2 Calibration Gas Generation iS 4 2 Cylinder Gas Dil tio m ora 4 2 Commercial Precision Dilution SysteMS cooononionconocnnnnoncnnanacnnnnncnns 4 3 Permeation Tube SySte mi gsic tots sateacdajadsutadghestidesac noted saawdaarart habs hectanees 4 3 Commercial Permeation Systems and iidtaiadoada s 4 4 Multi Point Catral id 4 5 Multipoint Calibration in Dual Auto Range Mode eee 4 7 Zeol Span Check AA II a E 4 10 Preventive Maintenance caccsmsmsaresesrarissrasarasinssrosarsnraneroniassasarosacssrenaes 5 1 Sly Pecan ONS oiiire ria E On 5 1 Replacement Parets dot ice 5 2 Cleanino thie Outside Cases esa este ee crete Leva a 5 2 Visual Inspection and Cleaning alada 5 2 Cleaning the Mirror ata 5 2 Capillary Inspection and Replacement 2 ccses0e0 42 accceee teh eceaieieneiies 5 3 Fan Filter Inspection and Clin ss 5 5 Sample Particulate
97. Federal Regulations Designated Equivalent Method Number EQSA 0486 060 EPA Designation Date April 10 1986 amended for Model 43i The Model 43i meets EPA designation requirements when operated as follows Range 50 to 1000 ppb Averaging Time 10 to 300 seconds Temperature Range 20 to 30 C Line Voltage 90 to 110 Vac 50 60 Hertz 105 to 125 Vac 50 60 Hertz 210 to 250 Vac 50 60 Hertz Pressure Compensation ON or OFF Temperature Compensation ON or OFF Flow Rate 0 5 to 1 0 LPM RS 232 RS 485 Interface With or without the following options Teflon Particulate Filter Rack Mounts High Flow Rate Internal Zero Span and Sample Valves Internal Permeation Span Source 1 0 Expansion Board Thermo Fisher Scientific Model 43 Instruction Manual Thermo Fisher Scientific About This Manual This manual provides information about installing operating maintaining and servicing the Model 43i It also contains important alerts to ensure safe operation and prevent equipment damage The manual is organized into the following chapters and appendixes to provide direct access to specific operation and service information Chapter 1 Introduction provides an overview of the product features describes the principle of operation and lists the specifications Chapter 2 Installation describes how to unpack setup and start up the analyzer Chapter 3 Operation descr
98. Filter peca 5 5 Lamp Wo tapeo erat o rain e ses 5 6 Model 437 Instruction Manual xv Contents xvi Chapter 6 Chapter 7 Model 43 Instruction Manual Real Tesouro ct gc 5 6 Pump Rebeca ia iiiod datar 5 7 Troublesh 0tind da 6 1 Salety Pr cautions arar o id ES 6 1 Troubleshooting Guides edad 6 1 Board Level Connection Diagrams cinc 6 10 Connector Pin Descriptions ne da e E 6 12 Service Loca ida 6 24 A E E E A E 7 1 Safety Precautions dias 7 1 Firmware Updates A hain e a R E ae 7 2 Accessing the Service Modes conidios 7 2 Replacement Parts ist o e aa 7 3 Cable Lita bea 7 4 External Device Connection Components occccccocccnnnnoncnnnonanonnnnonannnnnos 7 4 Removing the Measurement Bench and Lowering the Partition Panel 7 6 Fuse Replacement xi west seteci A iscajacs Rak etal eee LAs tie tet aut meee 7 8 Pimp Rep Ace rnc leo act II O Pace ateas es 7 9 Fan Replacement rioan i aa sie a aTa 7 10 Optical Bench Replacement iii tii 7 11 Cleaning the Witt Gre id ai dls 7 13 Flash Lamp Replacement vivia atea 7 13 Flash Lamp Voltage Adjustment 2274 Ges Sie len essen eve ieee 7 14 Flash Trigger Assembly Replacement acc 7 15 Flash Intensity Assembly Replacement ctas da 7 16 Photomultiplier Tube Replacement ancla 7 17 PMT High Voltage Power Supply Replacement ocoociconnnnnnonmmm 7 18 PMT Voltage Adjustment ici onde Ada 7 19 DC Power Supply Replacement aisla dins leads 7 20 Analog Output Test a e AOS 7 21 Analog Output Calor do iaa 7 2
99. Fisher Scientific Model 43 Instruction Manual B 25 C Link Protocol Commands Keys Display B 26 Keys Display Model 43 Instruction Manual set sp conc value set high sp conc value set low sp conc value These commands set the span concentrations to user defined values to value where value is a floating point representation of the span concentration in current selected units The example that follows sets the span concentration to 80 ppm in the single range mode Send set sp conc 80000 Receive set sp conc 80000 ok pres cal This command reports the pressure recorded at the time of calibration The example that follows shows that the pressure at calibration is 85 5 mmHg Send pres cal Receive pres cal 85 5 mmHg set pres cal This command automatically sets the current pressure as the calibration pressure The example that follows successfully sets the calibration pressure to 120 5 mmHg Send set pres cal 120 5 Receive set pres cal 120 5 ok set cal pres This command automatically sets the current pressure as the calibration pressure The example that follows successfully sets the calibration pressure Send set cal pres Receive set cal pres ok push button button do down en enter he help le left me menu ri right ru run up 1 2 3 4 These commands simulates pressing the front panel pushbuttons The numbers represent the front panel soft keys from left to right Send push enter Rec
100. HE TE Sanat Bone nr Analog Input The Analog Input Voltages screen read only displays the 8 raw analo 8 inp 8 y play 8 Voltages voltage readings if the 1 O expansion board option is installed In the Main Menu choose Diagnostics gt Analog Input Voltages Enek prun ERER EN EEE Eut Boes nr Thermo Fisher Scientific Model 43i Instruction Manual 3 59 Operation Diagnostics Menu Digital Inputs The Digital Inputs screen read only displays the state of the 16 digital inputs Pull ups are provided on all the inputs so if nothing is connected they will read 1 ifan input is brought to ground it will read 0 In the Main Menu choose Diagnostics gt Digital Inputs is EHO tiers Relay States The Relay States screen displays the state of the 10 digital outputs and allows toggling of the state to either on 1 or off 0 The relays are restored to their original states upon exiting this screen In the Main Menu choose Diagnostics gt Relay States Press to toggle the relay state open and closed EHO Calor 3 60 Model 43i Instruction Manual Thermo Fisher Scientific Operation Diagnostics Menu Test Analog Outputs The Test Analog Outputs menu contains a number of digital to analog converter DAC calibration items Channel choices include all analog outputs 6 voltage channels and 6 current channels if the I O expansion board option is installed In the Main Menu choose Diagnostics gt Test
101. Meaningful data should not be collected when the instrument is in the service mode In dual or auto range modes HI or LO multi point calibration is displayed to indicate the calibration of the high or low concentrations In the Main Menu choose Service The Flash Voltage Adjustment screen allows the user to manually adjust the flasher supply voltage The flash voltage adjustment screen is visible only when the instrument is in service mode For more information on the service mode see Service Mode earlier in this chapter Note This adjustment should only be performed by an instrument service technician A In the Main Menu choose Service gt Flash Voltage Adjustment Thermo Fisher Scientific Operation Service Menu Ea Mn mate mri NaC nd Le Initial Flash The Initial Flash Reference allows the user to view and set the initial flash reference The initial flash reference screen is visible only when the instrument is in service mode For more information on the service mode see Service Mode earlier in this chapter Reference Note This adjustment should only be performed by an instrument service technician A In the Main Menu choose Service gt Initial Flash Reference PMT Voltage The PMT Voltage Adjustment screen is used to manually adjust the PMT supply voltage The PMT voltage adjustment is visible only when the Adjustment instrument is in service mode For more information on the
102. Mode earlier in the chapter Note This adjustment should only be performed by an instrument service technician A Note Wait at least 30 seconds for the reading to stabilize before saving the value A In the Main Menu choose Service gt Flow Calibration Thermo Fisher Scientific Calibrate Flow Zero Calibrate Flow Span Thermo Fisher Scientific Operation Service Menu Ph dm mT Pi AG LS The Calibrate Flow Zero screen calibrates the flow sensor at zero flow Note The pump must be disconnected before performing the zero calibration A Note Wait at least 30 seconds for the reading to stabilize before saving the value A In the Main Menu choose Service gt Flow Calibration gt Zero Rohde Lira P PP Pi AG Lt The Calibrate Flow Span screen allows the user to view and set the flow sensor calibrate span point Note An independent flow sensor is required to read the flow then the operator enters the flow value on this screen to perform the calibration A Note Wait at least 30 seconds for the reading to stabilize before saving the value A In the Main Menu choose Service gt Flow Calibration gt Span Model 43i Instruction Manual 3 75 Operation Service Menu Restore Default Flow The Restore Default Flow Calibration screen allows the user to reset the Calibration configuration values to factory defaults Note Wait at least 30 seconds for the reading to stabilize before saving the va
103. Model 43 Instruction Manual Pulsed Fluorescence SO Analyzer Part Number 101589 00 14Jan2008 CE Thermo 2007 Thermo Fisher Scientific Inc All rights reserved Specifications terms and pricing are subject to change Not all products are available in all countries Please consult your local sales representative for details Thermo Fisher Scientific Air Quality Instruments 27 Forge Parkway Franklin MA 02038 1 508 520 0430 www thermo com aqi Thermo Fisher Scientific WEEE Compliance This product is required to comply with the European Union s Waste Electrical amp Electronic Equipment WEEE Directive 2002 96 EC It is marked with the following symbol Thermo Fisher Scientific has contracted with one or more recycling disposal companies in each EU Member State and this product should be disposed of or recycled through them Further information on Thermo Fisher Scientific s compliance with these Directives the recyclers in your country and information on Thermo Fisher Scientific products which may assist the detection of substances subject to the RoHS Directive are available at www thermo com WEEERoHS Model 43 Instruction Manual Equivalent Method Designation The Thermo Scientific Model 431 is designated by the United States Environmental Protection Agency USEPA as an Equivalent Method for SO2 as set forth in the U S Environmental Protection Agency regulations at Title 40 Part 53 of the U S Code of
104. O expansion board cable from the EXPANSION BD connector on the motherboard 3 Remove the two standoffs holding the I O expansion board connector to the rear panel Figure 7 21 4 Pop off the board from the mounting studs and remove the board 5 To install the I O expansion board follow previous steps in reverse 6 Calibrate the analog current outputs and analog voltage inputs per the procedure earlier in this chapter Thermo Fisher Scientific Servicing Motherboard Support Bracket Mounting Studs Motherboard Digital Output Board 1 0 Expansion Board Opt Standoffs 2 1 0 Expansion Board Motherboard Digital Output Board Figure 7 21 Rear Panel Board Connectors Digital Outp ut Board Use the following procedure to replace the digital output board Figure 7 18 and Figure 7 19 Replacement Equipment Required Digital output board Nut driver 3 16 inch amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis f Equipment Damage Some internal components can be damaged by small Thermo Fisher Scientific Model 43i Instruction Manual 7 41 Servicing Motherboard Replacement Motherboard Replacement 7 42 Model 431 Instruction Manual A before touching any internal components When the instrument is unplugged the chassis is
105. Possible Cause Detection circuit failure Instrument is not properly calibrated Input board malfunction Signal cable failure PMT failure Flash lamp assembly failure Action other problems Check all external plumbing and the source of the sample to verify that the SO is not being adsorbed by the sampling system Lines carrying SO must be made from clean Teflon or stainless steel Go to Diagnostics menu and run the optical span test to verify PMT and associated electronics Go to the Calibration Factors menu and verify that the SO Background and SO Coefficient are set appropriately Go to Service Menu and select Input Board Test to verify A D signal on each range While viewing the Input Board Test screen set the gain to 100 and then disconnect the signal cable from the input board The signal should drop from a value above 1000 to near zero Check the PMT voltage Service menu Check the lamp voltage Service menu Span calibration coefficient outside acceptable limits of 0 5 2 0 Bad span gas System leak Insufficient calibrator flow Verify quality of span gas Perform leak test Verify calibrator is providing a flow of at least 0 8 LPM Excessive noise or spikes on analog outputs Defective or low sensitivity PMT Defective input board or BNC connection Noise pick up by recorder or data logger Check PMT voltage and run an optical span test Replace PMT with known
106. SO molecules The condensing lens focuses the pulsating UV light into the mirror assembly The mirror assembly contains four selective mirrors that reflect only the wavelengths which excite SO molecules As the excited SO2 molecules decay to lower energy states they emit UV light that is proportional to the SO2 concentration The bandpass filter allows only the wavelengths emitted by the excited SO2 molecules to reach the photomultiplier tube PMT The PMT detects the UV light emission from the decaying SO2 molecules The photodetector located at the back of the fluorescence chamber continuously monitors the pulsating UV light source and is connected to a circuit that compensates for fluctuations in the UV light As the sample leaves the optical chamber it passes through a flow sensor a capillary and the shell side of the hydrocarbon kicker The sample then flows to the pump and is exhausted out the EXHAUST bulkhead of the analyzer The Model 437 outputs the SO concentration to the front panel display and the analog outputs and also makes the data available over the serial or Ethernet connection Thermo Fisher Scientific Introduction Specifications 1 l z Pass Electronics a A s Flashlamp F E id Te LIIN D A yo E ne SA Photodetector wr TON A AFAN gt Hydrocarbon y Kick
107. Scientific Operation Instrument Controls Menu Streaming Data Interval The Streaming Data Interval screen is used to adjust the streaming data interval The following interval times are available 1 2 5 10 20 30 60 90 120 180 240 and 300 seconds In the Main Menu choose Instrument Controls gt Communication Settings gt Streaming Data Config gt Streaming Data Interval Choose Item Signal The Choose Signal screen displays a submenu list of the analog output signal group choices Group choices are Concentrations Other Measurements and Analog Inputs if the I O expansion board is installed In the Main Menu choose Instrument Controls gt Communication Settings gt Streaming Data Config gt Item 1 8 Concentrations The Concentrations screen allows the user to select the output signal that is tied to the selected streaming data item The selected item is shown by lt after it In dual or auto range mode HI or LO is displayed to indicate high or low concentrations Range Status is also visible in auto range mode In the Main Menu choose Instrument Controls gt Communication Settings gt Streaming Data Config gt Select Item gt Concentrations Thermo Fisher Scientific Model 43i Instruction Manual 3 35 Operation Instrument Controls Menu Other Measurements The Other Measurements screen allows the user to select the output signal that is tied to the selected streaming data item The
108. Send set alarm sample flow max 1 Receive set alarm sample flow max 1 ok alarm trig conc so2 This command reports the SO concentration alarm trigger action for minimum alarm current setting to either floor or ceiling The example that follows shows the SO concentration minimum alarm trigger set to ceiling according to Table B 3 Send alarm trig conc so2 Receive alarm trig conc so2 1 These commands set the SO concentration alarm minimum value where value is set to either floor or ceiling according to Table B 3 The example that follows sets the SO concentration minimum alarm trigger to ceiling Send set alarm trig conc so2 1 Receive set alarm trig conc so2 1 ok Table B 3 Alarm Trigger Values Value Alarm Trigger 00 Floor 01 Ceiling Thermo Fisher Scientific Diagnostics Thermo Fisher Scientific C Link Protocol Commands Diagnostics age int This command reports the current of the reference channel AGC circuit The example that follows reports that the current AGC intensity is 90 percent Send agc int Receive agc int 90 diag volt mb This command reports the diagnostic voltage measurements on the motherboard The sequence of voltages is Positive 24 positive 15 positive 5 positive 3 3 and negative 3 3 Each voltage value is separated by a space Send diag volt mb Receive diag volt mb 24 1 14 9 4 9 3 2 3 2 diag volt mib This command reports the diagnostic voltage measurements on the mea
109. T USED 40013840014 HI S02 CONC Dual Auto Range mode Function code 0x80 01 IIlegal Function 02 Illegal Address Thermo Fisher Scientific Thermo Fisher Scientific MODBUS Protocol MODBUS Commands Supported Register Number Variable 40015840016 NOT USED 40017840018 NOT USED 400198 40020 RANGE S02 Dual Auto Range mode 400218 40022 INTERNAL TEMP 40023840024 CHAMBER TEMP 40025840026 NOT USED 40027 amp 40028 PERM OVEN GAS TEMPERATURE 400298 40030 PERM OVEN HEATER TEMPERATURE 400318 40032 CHAMBER PRESSURE 400338 40034 SAMPLE FLOW 400358 40036 PMT VOLTS 40037 amp 40038 FLASH VOLTS 400398 40040 FLASH REF 40041 amp 40042 ANALOG IN 1 1 0 Expansion Board installed 40043840044 ANALOG IN 2 1 0 Expansion Board installed 400458 40046 ANALOG IN 3 I O Expansion Board installed 40047 amp 40048 ANALOG IN 4 1 0 Expansion Board installed 400498 40050 ANALOG IN 5 1 0 Expansion Board installed 40051 amp 40052 ANALOG IN 6 1 0 Expansion Board installed 40053 amp 40054 ANALOG IN 7 1 0 Expansion Board installed 40055840056 ANALOG IN 8 1 0 Expansion Board installed Table C 2 Write Coils for 43 CoilNumber Action Triggered 101 ZERO MODE 102 SPAN MODE 103 S02 MODE 104 CS MODE 105 SET BACKGROUND 106 CAL TO LOW SPAN 107 AOUTS TO ZERO 108 AOUTS TO FS 109 CAL TO HIGH SPAN Model 43 Instruction Manual C 9 MODB
110. US Protocol MODBUS Commands Supported C 10 Model 43 Instruction Manual Table C 3 Read Coils for 43 Coil Number Status 1 AUTORANGE 2 LOCAL REMOTE 3 SERVICE 4 UNITS 0 ppm or ppb 1 mg m or g m 5 ZERO MODE 6 SPAN MODE 7 SAMPLE MODE 8 NOT USED 9 NOT USED 10 GEN ALARM 11 S02 CONC MAX ALARM 12 S02 CONC MIN ALARM 13 NOT USED 14 NOT USED 15 NOT USED 16 NOT USED 17 INTERNAL TEMP ALARM 18 CHAMB TEMP ALARM 19 CONVERTER TEMP 20 PERM GAS TEMP 21 PRESSURE ALARM 22 SAMPLE FLOW ALARM 23 FLASH REF ALARM 24 FLASH VOLTAGE ALARM 25 MOTHERBOARD STATUS ALARM 26 MEASUREMENT INTERFACE BOARD STATUS ALARM 27 1 0 EXP BD STATUS ALARM 28 CONC ALARM 29 PURGE MODE 30 ZERO CHK CAL ALARM Zero Span Option 31 SPAN CHK CAL ALARM Zero Span Option Thermo Fisher Scientific Serial Communication Parameters Thermo Fisher Scientific Appendix D Geysitech Bayern Hessen Protocol This appendix provides a description of the Geysitech Bayern Hessen or BH Protocol Interface and is supported both over RS 232 485 as well as TCP IP over Ethernet The Geysitech Commands that are implemented are explained in detail in this document The Geysitech protocol support for the Series enables the user to perform the functions of reading the various concentrations and to trigger the instrument to be in sample zero span mode if valid
111. Water Bath procedure Alternatively you can perform the Setting Perm Oven Temperature procedure and then continue with the Setting Temperature with Known Resistance procedure In either case use the following procedure for setting the perm oven temperature 1 Unplug POJ3 from the measurement interface board Place a 4 369 KQ resistor across pins 3 and 4 on the board 2 From the Permeation Oven menu in the Service menu select Cal Oven Thermistor Figure 9 2 Figure 9 2 Cal Oven Therm Resistor Screen Thermo Fisher Scientific Model 43i Instruction Manual 9 5 Optional Equipment Internal Permeation Span Source Setting Temperature with Water Bath 9 6 Model 43i Instruction Manual Enter the exact value of the attached resistor and press to save the offset Press Cm to return to the Permeation Oven menu Remove the resistor and re attach the permeation oven to POJ3 Use the following procedure for setting the measure temperature with water bath l Remove the thermistor from the permeation tube oven but leave the thermistor connected to the measurement interface board Insert the thermistor into the water bath next to an NIST traceable thermometer if necessary use an extension cable to reach Turn on the power to the water bath Using an NIST traceable thermometer with a resolution of 0 01 C adjust the water bath to 45 C From the Permeation Oven menu in the Service menu selec
112. ables color codes for pins 1 37 are for 37 pin cables A Table 9 1 Cable Options Description Cable Length DB37M to open end Six feet DB37F to open end Six feet DB25M to open end Six feet RS 232 Table 9 2 Color Codes for 25 Pin and 37 Pin Cables Pin Color Pin Color 1 BLACK 20 RED BLACK 2 BROWN 21 ORANGE BLACK 3 RED 22 YELLOW BLACK 4 ORANGE 23 GREEN BLACK 5 YELLOW 24 GRAY BLACK 6 GREEN 25 PINK BLACK 7 BLUE End color codes for 25 pin cables continue for 37 pin cables 8 VIOLET 26 PINK GREEN 9 GRAY 27 PIND RED 19 WHITE 28 PINK VIOLET 11 PINK 29 LIGHT BLUE Thermo Fisher Scientific Mounting Options Thermo Fisher Scientific Optional Equipment Mounting Options Pin Color Pin Color 12 LIGHT GREEN 30 LIGHT BLUE BROWN 13 BLACK WHITE 31 LIGHT BLUE RED 14 BROWN WHITE 32 LIGHT BLUE VIOLET 15 RED WHITE 33 LIGHT BLUE BLACK 16 ORANGE WHITE 34 GRAY GREEN 17 GREEN WHITE 35 GRAY RED 18 BLUE WHITE 36 GRAY VIOLET 19 VIOLET WHITE 37 LIGHT GREEN BLACK The analyzer can be installed in the configuration described in Table 9 3 and shown in Figure 9 5 through Figure 9 8 Table 9 3 Mounting Options Mounting Type Description Bench Positioned on bench includes mounting feet and front panel side trim handles ElA rack Mounted in an ElA style rack includes mounting slides and front panel ElA rack mounting handles Retrofit rack Mounted
113. age power Check the PMT high voltage power supply failure Flasher lamp supply voltage This voltage should be about 600 volts violet wire is positive Replace with known good lamp to see if the lamp is the problem Table 6 3 Troubleshooting Measurement Failures Malfunction Reduced response or no response to sample gas with alarm s indicated Possible Cause Undefined electronic failure or pump failure Instrument is not drawing in sample as expected Action Check alarm screens and the diagnostic voltage screen to localize fault Check the response to known span gas Run an optical span test Check sample Flow and Pressure readings on the Diagnostics screen Use an independent flow meter to check flows at the sample inlet and exhaust bulkheads they should match Perform a leak test as described in the Preventive Maintenance chapter Reduced response or no response to sample gas with no alarms indicated Instrument is not drawing in sample as expected Check sample Flow and Pressure readings on the Diagnostics screen Use an independent flow meter to check flows at the sample inlet and exhaust bulkheads they should match Perform a leak test as described in the Preventive Maintenance chapter Check the external plumbing for leaks or Model 43i Instruction Manual 6 5 Troubleshooting Troubleshooting Guides 6 6 Model 43i Instruction Manual Malfunction
114. aily As experience is gained with the instrument the frequency of these checks can be adjusted accordingly Thermo Fisher Scientific Thermo Fisher Scientific Calibration Zero Span Check The span gas concentration used in the span check should be between 70 and 90 of the full scale range The zero and span drift should be measured and recorded prior to making any adjustments All gas must be supplied to the instrument at atmospheric pressure It will be necessary to employ an atmospheric bypass plumbing arrangement to accomplish this If a filter is used all gas must enter the instrument through the filter For detailed information about the menu parameters and the icons used in these procedures see the Operation chapter Use the following procedure to check the zero span 1 Ifthe instrument is equipped with the optional zero span and sample valves connect the zero and span gas to the rear panel bulkheads labeled ZERO and SPAN otherwise connect a source of vented zero air to the SAMPLE bulkhead Note Do not use nitrogen for zero air A gt is used to activate the zero and span valves The lower left hand corner of the Run screen indicates which mode is active ZERO SPAN or SAMPLE 2 To ensure that the zero air is being measured at atmospheric pressure check that the flow reported by the analyzer is approximately 0 5 LPM Actual zero air flow should be 0 8 LPM or more with the excess flowing out the
115. aining screws Figure 7 17 Thermo Fisher Scientific Model 43i Instruction Manual 7 35 Servicing Thermistor Replacement Thermistor Replacement 7 36 Model 43i Instruction Manual 3 Disconnect the plumbing connections as appropriate to gain enough clearance to lift the optical bench off the heated kicker Carefully note the plumbing connections to facilitate reconnection 4 Lift the heated kicker off the rubber mounts 5 Install the new heated kicker assembly 6 Reposition the optical bench on top of the heated kicker insert and tighten the four optical bench retaining screws and reconnect any plumbing connections by following the previous steps in reverse order 7 Calibrate the instrument Refer to the Calibration chapter in this manual Retaining Screws 4 pe a Figure 7 17 Replacing the Kicker Assembly Optional Heated Kicker Use the following procedure to replace the ambient temperature thermistor Figure 7 18 Equipment Required Thermistor Thermo Fisher Scientific A Thermistor Ambient Temperature Calibration Thermo Fisher Scientific A Servicing Ambient Temperature Calibration Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before tou
116. aintenance Pump Rebuilding 7 Ifthe instrument passes the leak test but still has low flow check the capillary for blockage 8 Ifthe flow drops to zero during the flow check but the pressure is above 250 mm Hg the pump may need to be rebuilt Pump Rebuilding Use the following procedure to rebuild the pump Figure 5 4 To replace the pump see Pump Replacement in the Servicing chapter Equipment Required Flatblade screwdriver Pump rebuild kit flapper valve and diaphragm amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A f Equipment Damage Some internal components can be damaged by small 1 Turn instrument OFF unplug the power cord and remove the cover 2 Loosen the fittings and remove both lines going to the pump 3 Remove the four screws from the top plate remove top plate flapper valve and the bottom plate Figure 5 4 4 Remove the screw the securing the diaphragm to piston and remove diaphragm 5 Assemble the pump by following the previous steps in reverse make sure the Teflon white side of the diaphragm is facing up and that the flapper valves cover the holes of the top and bottom plate Thermo Fisher Scie
117. al at sign with a source indicator and a starting bit indicator All alarm information is presumed to be two bits long low and high The bitfield extraction is performed on the integer part of the source Typical alarm information would appear as 6 4 Then there appears an optional translation table within braces This is a string of words separated by spaces An example translation table would be Code_0 Code_1 Code_2 Code_3 The value once extracted is used as a zero based index into the translation table to determine the string to display Model 43 Instruction Manual B 45 C Link Protocol Commands Record Layout Definition Selection Table Button Designator Examples B 46 Model 43 Instruction Manual Then there appears an optional selection table within parentheses This is a string of numbers separated by spaces 0 1 The selection table lists the translation table entries which the user may select from when setting the parameter This is not necessarily the same as the entries which may be displayed Then there appears an optional button designator This will be one of B Jz L i pA or N B Indicates a button which pops up an input dialog prompting the user for a new value using the designated input format The input format is specified from the B through the subsequent semicolon I Indicates a button which pops up a selection list with input translation That is the values read are
118. alarm is triggered when the concentration goes above the minimum value If the SO concentration goes beyond either the minimum or maximum limit an alarm is activated The word ALARM appears in the Run screen and in the Main Menu In the Main Menu choose Alarms gt SO Concentration Min and Max S0 The Minimum SO Concentration alarm limit screen is used to change the Concentration Limits minimum SO concentration alarm limit The minimum and maximum SO concentration alarm limit screens function the same way In the Main Menu choose Alarms gt SO Concentration gt Min or Max Min Trigger The Minimum Trigger screen allows the user to view and set the SO concentration alarm trigger type to either floor or ceiling The minimum alarm may be programmed as a floor trigger alarm is triggered when the concentration falls below the minimum value or a ceiling trigger alarm is triggered when the concentration goes above the minimum value In the Main Menu choose Alarms gt SO Concentration gt Min Trigger Thermo Fisher Scientific Model 43i Instruction Manual 3 69 Operation Service Menu Service Menu Flash Voltage Adjustment 3 70 Model 43i Instruction Manual The Service menu appears only when the instrument is in the service mode To put the instrument into the service mode In the Main Menu choose Instrument Controls gt Service Mode Advanced diagnostic functions are included in the service mode
119. and reports whether temperature compensation is on or off The example that follows shows the temperature compensation is off Send temp comp Receive temp comp off set temp comp onoff These commands turn the temperature compensation on or off The example that follows turns temperature compensation off Send set temp comp off Receive set temp comp off ok Thermo Fisher Scientific Model 437 Instruction Manual B 31 C Link Protocol Commands Hardware Configuration Hardware Configuration B 32 Model 43 Instruction Manual contrast This command reports the screen s level of contrast The example that follows shows the screen contrast is 50 according to Table B 8 Send contrast Receive contrast 5 50 set contrast level This command sets the screen s level of contrast according to Table B 8 The example that follows sets the contrast level to 50 Send set contrast 5 Receive set contrast 5 ok Table B 8 Contrast Level Level Contrast Level 0 0 10 20 30 40 50 60 70 80 90 100 a O O0O ws os oa AIAN o conv set temp This command reports the temperature that the SO converter is set to The example that follows reports that the converter temperature is set to 50 1 C Send conv set temp Receive conv set temp 50 1 deg C set conv set temp value This command sets the temperature that the SO converter is set to value where value is an
120. anel pushbuttons and menu driven software Thermo Fisher Scientific Display on page 3 2 describes the LCD graphics display Pushbuttons on page 3 3 describes the various front panel pushbuttons and the expected key actions for each Software Overview on page 3 4 describes the menu driven software and submenus Range Menu on page 3 8 describes the gas units SO range and custom ranges Averaging Time on page 3 16 describes the averaging period applied to SO measurements Calibration Factors Menu on page 3 16 describes the calibration factors used to correct SO measurement readings Calibration Menu on page 3 19 describes manual and automatic calibration of zero and span Instrument Controls Menu on page 3 23 describes the instrument hardware control and configuration Diagnostics Menu on page 3 55 describes the diagnostic information and functions Alarms Menu on page 3 62 describes a list of items that are monitored by the analyzer Service Menu on page 3 70 describes service related menu items Password on page 3 86 describes how to enter change a password and lock unlock the analyzer Model 43i Instruction Manual 3 1 Operation Display Display The 320 x 240 graphics liquid crystal display LCD shows the sample concentrations instrument parameters instrument controls help and error messages Some menus contain more items than can be d
121. ant to note the averaging time when calibrating The longer the averaging time the more accurate the calibration will be To be most accurate use the 300 second averaging time For more information about calibration see Chapter 4 Calibration In the Main Menu choose Calibration gt Cal SO Background Thermo Fisher Scientific Model 43i Instruction Manual 3 19 Operation Calibration Menu a Cie hi Calibrate SO The Calibrate SO Coefficient screen is used to adjust the SO coefficient Coefficient and enter the span concentration The display shows the current SO2 concentration reading and the current SO range The next line of the display is where the SO calibration gas concentration is entered The SO span coefficient is calculated stored and used to correct the current SO reading For more information about calibration see Chapter 4 Calibration In dual or auto range modes HIGH or LOW is displayed to indicate the calibration of the high or low coefficient In the Main Menu choose Calibration gt Cal SO Coefficient a e Zero Span Check The Zero Span Check menu is available with the zero span valve option It Menu is used to program the instrument to perform fully automated zero and span check or adjustments Total Duration Hr is the sum of zero span and purge duration minutes Note Zero and Span Calibration Reset are toggle items that change between yes or no when selected
122. ard by lifting it up and off the slotted bottom support 4 Replace the front panel board by following previous steps in reverse LCD Module LCD Module Screws 4 Front Panel Board Bottom Support Figure 7 23 Replacing the Front Panel Board and the LCD Module Use the following procedure to replace the LCD module Figure 7 23 Equipment Required LCD module Philips screwdriver CAUTION If the LCD panel breaks do not to let the liquid crystal contact your skin or clothes If the liquid crystal contacts your skin or clothes wash it off immediately using soap and water A Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must Model 43i Instruction Manual 7 45 Servicing Service Locations be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A Do not remove the panel or frame from the module A The polarizing plate is very fragile handle it carefully A Do not wipe the polarizing plate with a dry cloth it may easily scratch the plate A Do not use alcohol acetone MEK or other Ketone based or aromatic solvents to clean the module use a soft cloth moistened with a naphtha cleaning solvent A Do not place the module n
123. are also supported Model 43i Instruction Manual 8 9 System Description 1 0 Components Analog Current Outputs Optional Analog Voltage Inputs Optional Digital Relay Outputs Digital Inputs 8 10 Model 43i Instruction Manual The analog outputs may be assigned to any measurement or diagnostic channel with a user defined range in the units of the selected parameter The voltage outputs are independent of the current outputs The optional I O Expansion board includes six isolated current outputs These are software configured for any one of the following ranges while maintaining a minimum resolution of 11 bits e 0 20 mA e 4 20mA The user can calibrate each analog output zero and span point through firmware At least 5 of full scale over and under range are also supported The analog outputs may be assigned to any measurement or diagnostic channel with a user defined range in the units of the selected parameter The current outputs are independent of the voltage outputs The current outputs are isolated from the instrument power and ground but they share a common return line Isolated GND The optional I O expansion board includes eight analog voltage inputs These inputs are used to gather measurement data from third party devices such as meteorological equipment The user may assign a label unit and a voltage to user defined unit conversion table up to 16 points All voltage inputs have a resolution of 12 bits ove
124. asurement interface board 3 Remove the bench see Optical Bench Replacement in this chapter 4 Loosen the retaining screw on the top of the lamp housing and pull out the trigger assembly and lamp Model 431 Instruction Manual 7 15 Servicing Flash Intensity Assembly Replacement Flash Intensity Assembly Replacement A 7 16 Model 43i Instruction Manual Remove the lamp from the trigger assembly by pulling straight out and insert the lamp into the new trigger assembly Insert the new flash trigger assembly into the lamp housing tighten the retaining screw and reconnect the flash trigger cable Re install the bench Use the following procedure to replace the flash intensity assembly Equipment Required Flash intensity assembly Phillips screwdriver small Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A Turn instrument OFE unplug the power cord and remove the cover Disconnect the flash intensity cable from the FLASH INT connector on the measurement interface board Remove the three screws securing the flash intensity assembly to the reaction chamber Figu
125. ber pressure changes on the analyzer s subsystems and output have been empirically determined This empirical data is used to compensate for any change in reaction chamber pressure When pressure compensation is on the display represents the current pressure in the fluorescence chamber When pressure compensation is off the display shows the factory standard pressure of 750 mmHg In the Main Menu choose Instrument Controls gt Pressure Compensation Thermo Fisher Scientific Screen Contrast Thermo Fisher Scientific Service Mode Operation Instrument Controls Menu The Screen Contrast screen is used to change the contrast of the display Values between 0 and 100 in increments of 5 are available Changing the screen contrast may be necessary if the instrument is operated at extreme temperatures In the Main Menu choose Instrument Controls gt Screen Contrast Notes The optimal contrast will change with changes in temperature A The optimal contrast will change from one LCD screen to another If the LCD screen is replaced the contrast may need to be reset A If the display contrast is not optimal but the content on the screen is visible select Instrument Controls gt Screen Contrast and adjust the screen contrast If the content on the screen is not visible use the set contrast 10 C Link command to set screen contrast to mid range then optimize the contrast See Contrast Levels in the C Link Protoco
126. ble B 5 Model 43 Instruction Manual B 23 C Link Protocol Commands Calibration Calibration B 24 Model 43 Instruction Manual Send stream time Receive stream time O set stream time value This command enables value where value is to attach or disable time stamp to streaming data string according to Table B 5 The example that follows attaches a time stamp to streaming data Send set stream time O Receive set stream time O ok Table B 5 Stream Time Values Value Stream Time 00 Attaches time stamp to streaming data string 01 Disables time stamp to streaming data string set cal so2 coef set cal high so2 coef set cal low so2 coef These commands will auto calibrate SO coefficients based on SO span gas concentrations The high and low commands are only available in dual and auto range mode If the mode is incorrect the instrument responds with can t wrong settings The example that follows shows a successful auto calibration of the low SO coefficient Send set cal low so2 coef Receive set cal low so2 coef ok set cal so2 bkg This command will auto calibrate the SO background The example that follows shows a successful auto calibration of the SO background Send set cal so2 bkg Receive set cal so2 bkg ok so2 coef coef so2 high so2 coef low so2 coef These commands report SO coefficients in single range mode or the high and low range coefficients in dual or auto range mode If the mod
127. bly Replacement Thermo Fisher Scientific A Servicing Flash Trigger Assembly Replacement 1 From the Main Menu press _ to scroll to Service gt press gt to scroll to Flash Voltage Adjustment gt and press The Set Flash Voltage Adjustment Man screen appears Note If Service Mode is not displayed refer to Accessing the Service Mode in this chapter then return to the beginning of this step A 2 At the Set Flash Voltage Adjustment Main screen use Cer until the SUPPLY voltage is 1000V 3 Press to store the value 4 Perform a PMT Voltage Adjustment and recalibrate the instrument Refer to PMT Voltage Adjustment in this chapter and the calibration procedures in the Calibration chapter Use the following procedure to replace the flash trigger assembly Figure 7 7 Equipment Required Flash trigger assembly Philips screwdriver Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Turn instrument OFF unplug the power cord and remove the cover 2 Disconnect the flash trigger cable from the FLASH TRIG connector on the me
128. board assembly is included with the optional I O expansion board Refer to Terminal Board PCB Assemblies in the Installation chapter for information on attaching the cable to the connector board For associated part numbers refer to the Servicing chapter The optional terminal block and cable kits provide a convenient way to connect devices to the instrument These kits break out the signals on the rear panel connector to individual numbered terminals Two types of terminal block and cable kits are available One kit is for the DB37 connectors and can be used for either the analog output connector or the relay output connector The other kit is for the DB25 connector and can be used for the optional I O expansion board For associated part numbers refer to the Servicing chapter Each kit consists of e one six foot cable e one terminal block Model 43i Instruction Manual 9 9 Optional Equipment Cables Cables 9 10 Model 43i Instruction Manual e one snap track Note Supporting all of the connections on units with the optional I O expansion board requires e two DB37 kits e one DB25 kit Table 9 1 identifies the optional individual cables that are available for the instrument and Table 9 2 provides the cable color codes For associated part numbers refer to the Servicing chapter Note Table 9 2 provides the color coding for both 25 pin cables and 37 pin cables Color codes for pins 1 25 are for 25 pin c
129. bration Factors menu or to return to the Run screen Press to actually set the SO reading to 0 ppb and store the new background correction of 25 7 ppb In the Main Menu choose Calibration Factors gt SO Bkg Model 43i Instruction Manual 3 17 Operation Calibration Factors Menu s0 Span Coefficient The SO span coefficient is usually calculated by the instrument processor during calibration The span coefficients are used to correct the SO readings and normally has a value near 1 000 The SO Span Coefficient screen allow the SO span coefficient to be manually changed while sampling span gas of known concentration The display shows the current SO concentration reading The next line of the display shows the SO span coefficient that is stored in memory and is being used to correct the SO concentration Notice that as the span coefficient value is changed the current SO concentration reading on the above line also changes However no real changes are made to the value stored in memory until is pressed Only proposed changes as indicated by a question mark prompt are displayed until is pressed In dual or autorange modes HIGH or LOW is displayed to indicate the calibration of the high or low coefficient The example that follows shows the coefficient screen in dual or auto range mode Note The concentration value will show ERROR if the measured concentration is not a valid span value either higher tha
130. ce a reading equivalent to less than 0 015 ppm S02 Possible Cause Zero air system is faulty needs new SO scrubbers or requires maintenance Zero air flow rate is inadequate Instrument is not drawing in span gas Span gas containing SO NO ora hydrocarbon is contaminating system Internal or external lines filters and other sample handling equipment are contaminated or dirty Hydrocarbon kicker has failed High scattered light Input board failure Action Test against an ultra zero cylinder from a reputable scientific gas supplier or check effect of a new chromatography grade activated charcoal scrubber installed at the instrument inlet Check by pass or atmospheric pressure vent to verify that the zero air system is providing more flow than the instrument is drawing Check sample Flow and Pressure readings on the Diagnostics screen Use an independent flow meter to check flows at the sample inlet and exhaust bulkheads they should match Perform a leak test as described in the Preventive Maintenance chapter Verify that span gases connected to the calibration system are shut off and leak tight Replace inlet filter if installed and as much of the tubing as possible Replace the hydrocarbon kicker Go to Instrument Controls select Flash Lamp and toggle to OFF If the previously high signal drops to zero or less when the flash lamp is off the problem may be caused by
131. ces 25 Pin Terminal Board The 25 pin terminal board is included with the optional I O Expansion Board See Detail A NE See Detail B Assembled Connector Figure 2 7 25 Pin Terminal Board Views Detail A Table 2 3 25 Pin Terminal Board Pin Descriptions Detail B Pin Signal Description Pin Signal Description 1 lOut1 13 Analog_In1 2 Isolated ground 14 Analog_In2 3 lOut2 15 Analog_In3 4 Isolated ground 16 GNDD 5 lOut3 17 Analog_In4 6 Isolated ground 18 Analog_In5 7 lOut4 19 Analog_In6 8 Isolated ground 20 GNDD 9 lOut5 21 Analog_In7 10 Isolated ground 22 Analog_In8 11 lOut6 23 GNDD 12 Isolated ground 24 GNDD 2 8 Model 43i Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Installation Startup Sta rtup Use the following procedure when starting the instrument l Turn the power ON Allow 30 minutes for the instrument to stabilize Set instrument parameters such as operating ranges and averaging times to appropriate settings For more information about instrument parameters see the Operation chapter Before beginning actual monitoring perform a multipoint calibration as described in the Calibration chapter Model 43i Instruction Manual 2 9 Installation Startup 2 10 Model 43i Instruction Manual Thermo Fisher Scientific Chapter 3 Operation This chapter describes the front panel display front p
132. ces and the line is terminated by a n the normal line separator character Valid fields are s parse a string d parse a decimal number Model 43 Instruction Manual B 43 C Link Protocol Commands Record Layout Definition Format Specifier for Binary Responses Format Specifier for Front Panel Layout B 44 Model 43i Instruction Manual ld parse a long 32 bit decimal number f parse a floating point number X parse a hexadecimal number lx parse a long 32 bit hex number ignore the field Note Signed versus unsigned for the integer values does not matter it is handled automatically A The second line of the Layout response is the binary parameter list for parsing the fields from a binary response Parameters MUST be separated by spaces and the line is terminated by a n Valid fields are 1 parse a time specifier 2 bytes parse a date specifier 3 bytes ignore one 8 bit character 1 byte parse a 24 bit floating point number 3 bytes n x parse a 24 bit floating point number 3 bytes N x parse a 32 bit floating point number 4 bytes parse an 8 bit signed number 1 byte parse an 8 bit unsigned number 1 byte parse a 16 bit signed number 2 bytes parse a 16 bit unsigned number 2 bytes parse a 24 bit signed number 3 bytes parse a 24 bit unsigned number 3 bytes parse a 32 bit signed number 4 bytes parse a 32 bit unsigned number 4 bytes Tmresesp 250
133. ching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Turn instrument OFF unplug the power cord and remove the cover 2 Squeeze the thermistor latch and pull the thermistor from the AMB TEMP connector on the measurement interface board 3 Snap the new thermistor into the AMB TEMP connector Figure 7 18 Replacing the Thermistor Use the following procedure to calibrate the ambient internal temperature for the instrument Equipment Required Calibrated thermometer or 10K 1 resistor WARNING The service procedures in this manual are restricted to qualified representatives 4 Model 43i Instruction Manual 7 37 Servicing Ambient Temperature Calibration 7 38 Model 43i Instruction Manual If the equipment is operated in a manner not specified by the manufacturer the protection provided by the equipment may be impaired A Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Remove the instrument cover 2 Tape the thermistor plugged into the measurement interface board to a calibrated thermometer Note Since the thermi
134. choose Instrument Controls gt Datalogging Settings gt Select Content WH mpini Pin Paid The Choose Item Type submenu is a list of the type of data that can be logged for the current field Choices are Concentrations Other Measurements and Analog Inputs if the I O expansion board is installed In the Main Menu choose Instrument Controls gt Datalogging Settings gt Select Content gt Field 1 32 ERED Cdra MT Dl Cie Lo a EN PUHE o Bone nr Note The ANALOG INPUTS item is only displayed if the I O Expansion Board option is installed A The Concentrations screen allows the user to select the output signal that is tied to the selected field item The selected item is shown by lt after it Note that at this point pressing indicates that these are proposed changes as opposed to implemented changes To change the selected record format and erase record log file data see Commit Content that follows Range Status is visible only in auto range mode In the Main Menu choose Instrument Controls gt Datalogging Settings gt Select Content gt Select Field gt Concentrations Model 43i Instruction Manual 3 27 Operation Instrument Controls Menu Other Measurements The Other Measurements screen allows the user to select the output signal that is tied to the selected field item The selected item is shown by lt after it Items displayed are determined by the options installed Note that
135. conicncnnnonnornnononononoso 3 48 Table 6 1 Troubleshooting Power Up FallureS oococociniconcinininoncnnincnnnocoso 6 2 Table 6 2 Troubleshooting Calibration Fallures oooconociciocicniccinnncnnnnccnnoss 6 3 Table 6 3 Troubleshooting Measurement Failures oooococoocicininonnincnicncono 6 5 Table 6 4 Troubleshooting Alarm Messages oococonocoocinoconcnnnncnccncnrnnnnnooos 6 8 Table 6 5 Motherboard Connector Pin DescriptiONS o oooo 6 12 Table 6 6 Measurement Interface Board Connector Pin Descriptions 6 16 Table 6 7 Front Panel Board Connector Pin Diagram o o o o 6 19 Table 6 8 0 Expansion Board Optional Connector Pin Descriptions 6 21 Table 6 9 Digital Output Board Connector Pin Descriptions 6 22 Table 6 10 Input Board Connector Pin Descriptions ooocinoininonninicnco 6 23 Table 6 11 Flash Trigger Pack Pin Descriptions oococociciconononncncnncncnnocnnnnoo 6 24 Table 6 12 Flash Intensity Assembly Pin Descriptions 6 24 Table 7 1 Replacement PartS ooociccoccocicocccccnnnnnconcnrnnnrnnnocornrnnnonororannnnorarannonos 7 3 Table 7 2 Model 431 Cables cccceccccssesseseeseesecsessestsseeseeseenestesteseeseeneeee 7 4 Table 7 3 External Device Connection Component cooocinconicccnnncininicninccnnoss 7 4 Table 7 4 Analog Output Channels and Rear Pa
136. del 43i Instruction Manual 4 5 Calibration Multi Point Calibration 4 6 Model 43i Instruction Manual 10 0 5 LPM Actual zero air flow should be 0 8 LPM or more with the excess flowing out the atmospheric bypass a From the Run screen press _M_ to display the Main Menu b From the Main Menu choose Diagnostics gt Flow to display the Sample Flow screen c Make sure that the reported sample flow is less than the actual zero air flow Monitor the zero air reading and wait for the reading to stabilize From the Main Menu choose Calibration to display the Calibration menu At the Calibration menu press to display the SO Background screen after 10 minutes of zero air and when the reading is stable press to set the SO reading to zero The display flashes the message SAVING and the SO reading is added to the background correction Record the stable zero air reading as Zso2 Press _M_ to return to the Calibration menu Connect a source of calibration gas to the SAMPLE bulkhead The calibration gas concentration should be approximately 80 of the full scale range To ensure that the calibration gas is being measured at atmospheric pressure check that the sample flow in the analyzer is approximately 0 5 LPM Actual sample air flow should be 0 8 LPM or more with the excess flowing out the atmospheric bypass Sample the vented calibration gas and wait for the instrument reading to stabilize Pres
137. dentifies the associated channels 2 From the Main Menu press 4 to scroll to Service press gt to scroll to Analog Input Calibration and press The Analog Input Cal screen displays input channels 1 8 3 At the Analog Input Cal screen press _ to scroll to the channel selected in Step 1 and press 4 Press to scroll to Calibrate Full Scale and press Ce The screen displays the current input voltage for the selected channel 5 Use e f and Ce to enter the source voltage and press to calibrate the input voltage for the selected channel to the source voltage Thermo Fisher Scientific Servicing 6 Press gt to return to the input channels display and repeat Steps 3 5 to calibrate other input channels to the source voltage as necessary Pressure Transducer Use the following procedure to replace the pressure transducer assembly Assembly Replacement Thermo Fisher Scientific Figure 7 13 Equipment Required Pressure transducer assembly Philips screwdriver Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A Turn instr
138. dr dns 192 168 1 1 ok addr gw This command reports the default TCP IP gateway address Send addr gw Receive addr gw 192 168 1 1 set addr gw address This command sets the default gateway address where address consists of gt four numbers ranging from 0 255 inclusive separated by Send set addr gw 192 168 1 1 Receive set addr gw 192 168 1 1 ok addr ip This command reports the IP address of the analyzer Send addr ip Receive addr ip 192 168 1 200 set addr ip address This command sets the analyzer s IP address where address consists of four 9 numbers ranging from 0 255 inclusive separated by Send set addr ip 192 168 1 200 Receive set addr ip 192 168 1 200 ok addr nm This command reports the IP netmask Send addr nm Receive addr nm 255 255 255 0 Model 43 Instruction Manual B 35 C Link Protocol Commands Communications Configuration B 36 Model 43 Instruction Manual set addr nm address This command sets the nm address where address consists of four numbers 9 ranging from 0 255 inclusive separated by Send set addr nm 255 255 255 0 Receive set addr nm 255 255 255 0 ok baud This command reports the current baud rate for the serial port RS232 RS485 The example that follows reports that the current baud rate is 9600 baud Send baud Receive baud 9600 set baud rate rate 1200 2400 4800 9600 19200 38400 57600 115200 T
139. e Diagnostics on page B 15 e Datalogging on page B 16 e Calibration on on page B 24 e Keys Display on page B 26 e Measurement Configuration on page B 28 e Hardware Configuration on page B 32 e Communications Configuration on page B 35 e I O Configuration on page B 39 e I O Configuration on page B 39 Each command sent to the analyzer must begin with the American Standard Code for Information Interchange ASCII symbol or byte value equivalent of the instrument s identification number plus 128 For example if the instrument ID is set to 25 then each command must begin with the ACSII character code 153 decimal The analyzer ignores any command that does not begin with its instrument identification number If the instrument ID is set to 0 then this byte is not required For more information on changing Instrument ID see Chapter 3 Operation Model 437 Instruction Manual B 1 C Link Protocol Commands Commands Commands Accessing Streaming Data Entering Units in PPB Commands List B 2 Model 43 Instruction Manual The analyzer must be in the remote mode in order to change instrument parameters via remote However the command set mode remote can be sent to the analyzer to put it in the remote mode Report commands commands that don t begin with set can be issued either in the remote or local mode For information on changing modes see Chapter 3
140. e Appendix C MODBUS Protocol provides a description of the MODBUS Protocol Interface and is supported both over RS 232 485 RTU protocol as well as TCP IP over Ethernet e Appendix D Geysitech Bayern Hessen Protocol provides a description of the Geysitech Bayern Hessen or BH Protocol Interface and is supported both over RS 232 485 as well as TCP IP over Ethernet Safety and Equipment This manual contains important information to alert you to potential safety azards and risks of equipment damage Refer to the followin es o Damage Alerts hazards and risks of equip damage Refi he following types of alerts you may see in this manual Safety and Equipment Damage Alert Descriptions Alert Description DANGER A hazard is present that will result in death or serious personal injury if the warning is ignored A serious personal injury if the warning is ignored 4 CAUTION The hazard or unsafe practice could result in minor to moderate personal injury if the warning is ignored 4 f WARNING A hazard is present or an unsafe practice can result in Equipment Damage The hazard or unsafe practice could result in property A N damage if the warning is ignored A Safety and Equipment Damage Alerts in this Manual Alert Description WARNING If the equipment is operated in a manner not specified by the manufacturer the protection provided by the equipment may be impaired 4 The service procedures in this manual are restricted to
141. e chapter Notes This adjustment should only be performed by an instrument service technician A Current channels are visible only if the I O expansion board is installed a In the Main Menu choose Service gt Analog Output Calibration gt Voltage Channel 1 6 or Current Channel 1 6 Thermo Fisher Scientific Model 43i Instruction Manual 3 77 Operation Service Menu y me o opH OE Bd tPA PEPE Analog Output Calibrate Zero Analog Output Calibrate Full Scale 3 78 Model 43i Instruction Manual The Analog Output Calibrate Zero screen allows the user to calibrate the zero state of the selected analog output The operator must connect a meter to the output and adjust the output until it reads 0 0 V on the meter In the Main Menu choose Service gt Analog Output Calibration gt Selected Channel gt Calibrate Zero The Analog Output Calibrate Full Scale screen allows the user to calibrate the full scale state of the selected analog output The operator must connect a meter to the output and adjust output until it reads the value shown in the set output to field In the Main Menu choose Service gt Analog Output Calibration gt Selected Channel gt Calibrate Full Scale Thermo Fisher Scientific Analog Input Calibration Operation Service Menu i E El Cie EE haa nr j WPi The Analog Input Calibration menu is a selection of analog input to calibrate and allows the user to select the calibrati
142. e instrument is in service mode For more information on the service mode see Service Mode earlier in the chapter Note This adjustment should only be performed by an instrument service technician A In the Main Menu choose Service gt Perm Oven Settings Thermo Fisher Scientific Operation Service Menu Calibrate Gas Thermistor The Calibrate Gas Thermistor submenu is used for calibrating the permeation oven gas thermistor using a water bath or known resistance In the Main Menu choose Service gt Perm Oven Settings gt Cal Gas Thermistor gt Water Bath or Known Resistor Water Bath The Calibrate Gas Thermistor Water Bath screen allows the user to view and set the permeation oven gas temperature to a known temperature value Lira My Py AG Ls Resistor The Calibrate Gas Thermistor Resistance screen allows the user to view and set the permeation oven gas thermistor resistance to a known resistor value Thermo Fisher Scientific Model 43i Instruction Manual 3 81 Operation Service Menu Calibrate Oven Thermistor Permeation Oven Setpoint Factory Calibrate Gas Thermistor 3 82 Model 43i Instruction Manual The Calibrate Oven Thermistor menu is used to view and set the permeation oven thermistor resistance to a known resistor value In the Main Menu choose Service gt Perm Oven Settings gt Cal Oven Thermistor The Permeation Oven Setpoint screen is used to select the permea
143. e instrument uses linear interpolation between the points in this table to determine what the reading value is based on the analog input voltage Each point in the table consists of an analog input voltage value 0 10 5 V and a corresponding reading value Only two points are necessary for linear inputs however a larger number of points may be used to approximate non linear inputs The points range from 2 to 10 with a default of 2 In the Main Menu choose Instrument Controls gt I O Configuration gt Analog Input Config gt Select Channel gt Table Points Hai Thess Minter E ee ee miii 3 50 Model 43i Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu Table Point The Table Point submenu allows the user to set up an individual table point In the Main Menu choose Instrument Controls gt I O Configuration gt Analog Input Config gt Table Point 1 10 Volts The Volts screen allows the user to set the input voltage for the selected table point in the conversion table from 0 00 to 10 50 The default table is a two point table with point 1 0 00 V 000 0 U and point 2 10 00 V 10 0 U In the Main Menu choose Instrument Controls gt I O Configuration gt Analog Input Config gt Select Table Point gt Volts User Value The User Value screen allows the user to set the output value for the corresponding input voltage for the selected table point in the conversion table from
144. e is incorrect the instrument responds with can t wrong settings The example that follows reports that the SO coefficient is 1 200 Thermo Fisher Scientific C Link Protocol Commands Calibration Send so2 coef Receive so2 coef 1 200 set so2 coef value set coef so2 value set high so2 coef value set low so2 coef value These commands set the SO coefficients to user defined values to value where value is a floating point representation of the coefficient The example that follows sets the SO coefficient to 1 200 Send set so2 coef 1 200 Receive set so2 coef 1 200 ok so2 bkg This command reports the current SO backgrounds The example that follows reports that the SO background is 21 4 ppb Send so2 bkg Receive so2 bkg 21 4 ppb set so2 bkg value This command is used to set SO backgrounds to user defined values to value where value is a floating point representation of the background in current selected units The example that follows sets the SO2 background to 21 4 ppb Send set so2 bkg 21 4 Receive set so2 bkg 21 4 ok sp conc high sp conc low sp conc These commands report span concentration in single range mode or the high and low span concentrations in dual or auto range mode If the mode is incorrect the instrument responds with can t wrong settings The example that follows reports the span gas concentration in single range mode Send sp conc Receive sp conc 1000 Thermo
145. e of the flash lamp assembly It receives high voltage and the trigger signal from the measurement interface board and uses a small transformer to produce a short high voltage pulse to fire the flash lamp The flash intensity board amplifies the lamp intensity signal detected by the photodetector which is used to control the lamp voltage The I O expansion board connects to the motherboard and adds the capability to input external analog voltage inputs and to output analog currents via a connector located on the rear panel of the instrument It contains local power supplies a DC DC isolator supply a sub processor and analog circuits Eight analog voltage inputs are provided with an input voltage range of OV tol0VDC Six current outputs are provided with a normal operating range of 0 to 20 mA External I O is driven from a generic bus that is capable of controlling the following devices e Analog output voltage and current e Analog input voltage e Digital output TTL levels e Digital input TTL levels Note The instrument has spare solenoid valve drivers and I O support for future expansion A The instrument provides six analog voltage outputs Each may be software configured for any one of the following ranges while maintaining a minimum resolution of 12 bits e 0 100mV e 0 1V e 0 5V e 0 10V The user can calibrate each analog output zero and span point through firmware At least 5 of full scale over and under range
146. e two screws at the top of the measurement interface board Pop off the measurement interface board from the two bottom mounting studs and remove the board 4 To install the measurement interface board follow previous steps in reverse 5 Re install the measurement bench Refer to Removing the Measurement Bench and Lowering the Partition Panel in this chapter Mounting Studs 2 Mounting Screws 2 Figure 7 22 Replacing the Measurement Interface Board Front Panel Board Use the following procedure to replace the front panel board Figure 7 23 Replacement Paper Required Eront panel board amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Equipment Damage Some internal components can be damaged by small 7 44 Model 43i Instruction Manual Thermo Fisher Scientific Mounting Studs 2 Ribbon Cables 3 F LCD Module Replacement A A Thermo Fisher Scientific Servicing LCD Module Replacement 1 Turn instrument OFF unplug the power cord and remove the cover 2 Remove the three ribbon cables and the two wire connector from the front panel board 3 Pop off the board from the two top mounting studs and remove the bo
147. ear organic solvents or corrosive gases A Do not shake or jolt the module A 1 Turn instrument OFF unplug the power cord and remove the cover 2 Disconnect the ribbon cable and the two wire connector from the front panel board 3 Remove the four screws at the corners of the LCD module 4 Slide the LCD module out towards the top of the instrument 5 Replace the LCD module by following previous steps in reverse Service Locations For additional assistance service is available from exclusive distributors worldwide Contact one of the phone numbers below for product support and technical information or visit us on the web at www thermo com aqi 1 866 282 0430 Toll Free 1 508 520 0430 International 7 46 Model 43i Instruction Manual Thermo Fisher Scientific Chapter 8 System Description This chapter describes the function and location of the system components provides an overview of the software structure and includes a description of the system electronics and input output connections and functions as follows Hardwa re Model 43i hardware components Figure 8 1 include e Hydrocarbon kicker e Heated hydrocarbon kicker optional e Optics e Flash lamp e Condensing lens e Mirror assembly e Light baffle e Trigger pack e Reaction chamber e Bandpass filter e Photomultiplier tube e Photodetector e Flow sensor e Pressure transducer e Capillary e Vacuum Pump Thermo Fisher Scientific Model 43i Instruction Manual 8 1
148. ed Internal Temp User set alarm min value User set alarm max value Chamber Temp User set alarm min value User set alarm max value Perm Oven Gas Temp Perm Oven Gas alarm min value Perm Oven Gas alarm max value Perm Oven Heater Perm Oven Heater alarm min Perm Oven Heater alarm max Temp value value Pressure User set alarm min value User set alarm max value Sample Flow User set alarm min value User set alarm max value PMT Volts 700 volts 1100 volts Flash Voltage User set alarm min value User set alarm max value Flash Reference User set alarm min value User set alarm max value Everything Else 0 units 10 units Choose Signal To Output The Choose Signal To Output screen displays a submenu list of the analog output signal group choices Group choices are Concentrations Other Measurements and Analog Inputs if the I O expansion board is installed This allows the user to select the output signal to the selected output channel In dual or auto range mode HI or LO is displayed to indicate high or low concentrations Range SO2 is visible only in auto range mode The Concentrations screen is shown below See Table 3 8 that follows for a list of items for each signal group choice In the Main Menu choose Instrument Controls gt I O Configuration gt Analog Output Config gt Select Channel gt Choose Signal To Output PR Ea Cir Py Tit Le EUs EEE TE haa nr Thermo Fisher Scienti
149. ed relay s B 42 ri right Simulates pressing right pushbutton B 26 ru run Simulates pressing run pushbutton B 26 sample Sets zero span valves to sample mode B 30 sample gas save Stores parameters in FLASH B 33 save params Stores parameters in FLASH B 33 sc screen C series legacy command that reports a generic response Use B 27 iscreen instead so2 Reports current SO concentration B 9 so2 bkg Reports sets current SO background B 25 so2 coef Reports sets current SO coefficient B 25 sp conc Reports sets span concentration B 25 sp field Reports sets item number and name in scratch pad list B 22 span Sets zero span valves to span mode B 30 span gas sr Reports last short record stored B 17 srec Reports maximum number of short records B 18 srec format Reports sets output format for short records ASCII or binary B 19 srec layout Reports current layout of short record data B 20 srec mem size Reports maximum number of short records B 20 srec per Reports sets short record logging period B 20 stream per Reports sets current set time interval for streaming data B 23 stream time Reports sets a time stamp to streaming data or not B 23 temp comp Reports sets temperature compensation on or off B 31 time Reports sets current time 24 hour time B 34 up Simulates pressing up pushbutton B 26 zero Sets zero span valves to zero mode B 30 zero gas Model 431 Instruction Manual B 7 C Link Protocol Commands Measurements Measurements B 8 Model 43 In
150. edure to setup the instrument 1 Connect the sample line to the SAMPLE bulkhead on the rear panel Figure 2 3 Ensure that the sample line is not contaminated by dirty wet or incompatible materials All tubing should be constructed of FEP Teflon 316 stainless steel borosilicate glass or similar tubing with an OD of 1 4 inch and a minimum ID of 1 8 inch The length of the tubing should be less than 10 feet Note Gas must be delivered to the instrument free of particulates It may be necessary to use the Teflon particulate filter as described in Teflon Particulate Filter on page 9 9 A If the sample may contain particulates larger than 5 microns it should be filtered before introducing it to the instrument Use a filter such as Teflon that does not interact with SO in the sample If a sample filter is used all calibrations and span checks must be performed through the filter The filter element should be replaced regularly to prevent the absorption of SO by trapped material on the filter A Note Gas must be delivered to the instrument at atmospheric pressure It may be necessary to use an atmospheric bypass plumbing arrangement as shown in Figure 2 4 if gas pressure is greater than atmospheric pressure A 2 Connect the EXHAUST bulkhead to a suitable vent The exhaust line should be 1 4 inch OD with a minimum ID of 1 8 inch The length of the exhaust line should be less than 10 feet Verify that there is no restriction in this
151. eive push enter ok Thermo Fisher Scientific C Link Protocol Commands Keys Display isc iscreen This command retrieves the framebuffer data used for the display on the iSeries instrument It is 19200 bytes in size 2 bits per pixel 4 pixels per byte arranged as 320 by 240 characters The data is sent in RLE encoded form to save time in transmission It is sent as a type 5 binary c_link response with no checksum The RLE encoding consists of a 0 followed by an 8 bit count of consecutive OxFF bytes The following c code will expand the incoming data Void unpackDisplay void far tdib unsigned char far rlescreen int i j k unsigned char far sc4bpp sc2bpp screen ptr ptr screen unsigned char far malloc 19200 RLE decode the screen for i 0 i lt 19200 4 ptr screen lt 19200 i ptr rlescreen i if rlescreen i 0 unsigned char rlecount unsigned char rlescreen i while rlecount ptr 0 tlecount else if rlescreen i Oxff unsigned char rlecount unsigned char rlescreen i while rlecount ptr oxff rlecount To convert this data into a BMP for use with windows it needs to be turned into a 4BPP as that is the smallest windows can display Also note that BMP files are upside down relative to this data i e the top display line is the last line in the BMP sc screen This command is meant for backwa
152. emblies Refer to Figure 7 2 to identify the component location Table 7 1 Replacement Parts Part Number Description 100480 00 Front Panel Pushbutton Board 101491 07 Processor Board 100533 00 Motherboard Assembly 100539 00 Digital Output Board 100542 00 1 0 Expansion Board Optional 102340 00 Front Panel Connector Board 102496 00 Front Panel Display 101399 00 Transformer Step Down 220 240V Optional 101863 00 Transformer 100V Step up Transformer 100862 00 Measurement Interface Board Assembly 100883 00 Input Board Assembly 8774 Trigger Board Assembly 101023 00 Pressure Transducer Assembly 102055 00 Flow Transducer 8666 Lamp Flashtube 8884 Flash Intensity Photocell Assembly 8746 Kicker Assembly Standard and Heated optional 8868 Photomultiplier Tube PMT 101024 00 PMT High Voltage Power Supply 100727 00 PMT Base Socket Assembly 01426 00 Pump 110VAC w Plate and Fittings 8606 Pump Repair Kit for 101426 00 01055 00 AC Receptacle Assembly 4510 Fuse 250VAC 3 0 Amp SlowBlow for 100VAC and 110VAC models Model 43i Instruction Manual 7 3 Servicing Cable List Cable List External Device Connection Components 7 4 Model 43i Instruction Manual Part Number Description 14007 Fuse 250VAC 1 60 Amp SlowBlow for 220 240VAC models 101681 00 Power Supply Assembly 24VDC w Base Plate and Screws 101688 00 Ambient Te
153. emp alarm max 45 Receive set internal temp alarm max 45 ok alarm pressure min alarm pressure max These commands report the pressure alarm minimum and maximum value current settings The example that follows reports that the pressure alarm minimum value is 205 mmHg Send pressure alarm min Receive pressure alarm min 205 mmHg set alarm pressure min value set alarm pressure max value These commands set the pressure alarm minimum and maximum values to value where value is a floating point number representing pressure alarm limits in millimeters of mercury The example that follows sets the pressure alarm maximum value to 215 mmHg Send set alarm pressure max 215 Receive set alarm pressure max 215 ok Thermo Fisher Scientific Model 43 Instruction Manual B 13 C Link Protocol Commands Alarms B 14 Model 43 Instruction Manual alarm sample flow min alarm sample flow max These commands report the sample flow alarm minimum and maximum value current settings The example that follows reports that the sample flow alarm minimum value is 2 LPM Send alarm sample flow min Receive alarm sample flow min 2 1 min set alarm sample flow min value set alarm sample flow max value These commands set the sample flow alarm minimum and maximum values to value where value is a floating point number representing sample flow alarm limits in liters per minute The example that follows sets the sample flow alarm maximum value to 1 LPM
154. ent mode determined by single dual auto gas mode for analog output digital output and digital inputs The index number is used to insert the variable in a field location in a list using set sp field index The example that follows reports the list of analog output index numbers and variables Send list var aout Receive list var aout index variable O none 1 so2 11 intt 12 rctt 16 pres 17 smplfl 18 pmtv 19 Impv 20 lmpi 21 aint 22 ain2 23 ain3 24 ain4 25 ains 26 ain6 27 ain7 28 ain8 relay stat This command reports the current relay logic normally open or normally closed if all the relays are set to same state that is all open or all closed The example that follows shows that the status when all the relays logic is set to normally open Send relay stat Receive relay stat open Note If individual relays have been assigned different logic then the response would be a 4 digit hexadecimal string with the least significant byte LSB being relay no 1 A Thermo Fisher Scientific Record Layout Definition Format Specifier for ASCII Responses Thermo Fisher Scientific C Link Protocol Commands Record Layout Definition For example Receive relay stat 0x0001 indicates relay no 1 is set to normally open logic all others are normally closed Receive relay stat 0x0005 indicates relay no 1 and 3 are set to be normally open logic all others are normally closed set
155. eplacement Parts Cleaning the Outside Case Visual Inspection and Cleaning Cleaning the Mirrors 5 2 Model 43i Instruction Manual Equipment Damage Do not attempt to lift the instrument by the cover or other external fittings A Equipment Damage Do not use solvents or other cleaning products to clean the outside case A Equipment Damage DO NOT attempt to clean the mirrors in the optical bench These mirrors do not come in contact with the sample gas and should not be cleaned Cleaning the mirrors can damage the mirrors A See the Servicing chapter for a list of replacement parts Clean the outside case using a damp cloth being careful not to damage the labels on the case Equipment Damage Do not use solvents or other cleaning products to clean the outside case A The instrument should be inspected occasionally for obvious visible defects such as loose connectors loose fittings cracked or clogged Teflon lines and excessive dust or dirt accumulation Dust and dirt can accumulate in the instrument and can cause overheating or component failure Dirt on the components prevents efficient heat dissipation and may provide conducting paths for electricity The best way to clean the inside of the instrument is to first carefully vacuum all accessible areas and then blow away the remaining dust with low pressure compressed air Use a soft paint brush or cloth to remove stubborn dirt The mirrors located in the optical bench do
156. eply string lt address gt is the Instrument Address Each subsequent measurement attached to the response will have the lt address X gt where X keeps incrementing by 1 for each measurement included lt measured value gt is the concentration value in currently selected gas units represented as exponential representation with 4 characters mantissa and 2 characters exponent each with sign Mantissa sign and 4 digits The decimal point is assumed to be after the first digit and is not transmitted Exponent sign and 2 digits Example 5384000 0 is represented as 5384 06 0 04567 is represented as 4567 02 lt status gt is formed by lt operating status gt and lt error status gt and separated by a space i e lt operating status gt lt SP gt lt error status gt Each of the two lt operating status gt and lt error status gt are formed by two characters each representing a 2 digit hex number which is one byte 8 Bits operation status and one byte 8 Bits error status These two bytes contain the information about the main operating conditions of the instrument at that instant For details on how to interpret the status bytes refer to Table D 1 and Table D 2 below lt SFKT gt is the space provided for future use for special function it currently contains a string of ten 0 s i e lt 0000000000 gt Model 437 Instruction Manual D 5 Geysitech Bayern Hessen Protocol Geysitech Commands Measurement
157. er Clow Optical Char r Sensor gt O gt gt Pressure Figure 1 1 Model 43i Flow Schematic Specifications Preset ranges Table 1 1 Model 43 Specifications 0 0 05 0 1 0 2 0 5 1 2 5 and 10 ppm 0 02 0 5 1 2 5 10 20 and 25 mg m Extended Ranges 0 0 5 1 2 5 10 20 50 and 100 ppm 0 2 5 10 20 50 100 200 and 250 mg m Custom ranges 0 0 05 to 10 ppm 0 0 5 to 100 ppm in extended range 0 0 2 to 25 mg m 0 2 to 250 mg m in extended range Zero noise 1 0 ppm RMS 10 second averaging time 0 5 ppm RMS 60 second averaging time 0 25 ppm RMS 300 second averaging time Lower detectable limit 2 ppb 10 second averaging time 1 ppb 60 second averaging time 0 5 ppb 300 second averaging time Zero drift 24 hour lt 1 ppb Span drift 1 full scale per week Response time 80 sec 10 second averaging time 110 sec 60 second averaging time 320 sec 300 second averaging time Thermo Fisher Scientific Model 43i Instruction Manual 1 3 Introduction Specifications 1 4 Model 43i Instruction Manual Linearity 1 of full scale Sample flow rate 0 5 Ipm standard 1 0 Ipm optional Interferences EPA levels Less than lower detectable limit except for the following NO lt 3 ppb M Xylene lt 1ppb H20 lt 3 of reading Operating 15 35 C may be safely operated over the range of 0 45 C te
158. erated This is the instrument calibration curve All future measurements should be interpreted using this curve Regulations typically require multipoint calibration when the instrument is newly installed moved repaired interrupted for more than a few days or when span or zero shift by more than 15 All gas must be supplied to the instrument at atmospheric pressure It may be necessary to employ an atmospheric bypass plumbing arrangement to accomplish this If a filter is used all gas must enter the instrument through the filter For detailed information about the menu parameters and the icons used in these procedures see the Operation chapter Model 43i Instruction Manual 4 7 Calibration Multipoint Calibration in Dual Auto Range Mode 4 8 Model 43i Instruction Manual Note The calibration and calibration check duration times should be long enough to account for the transition purge process when switching from sample to zero and from zero to span This transition time is the time required to purge the existing air A Use the following procedure to perform a multipoint calibration in dual autorange mode 1 Ifthe instrument is equipped with the optional zero span and sample valves connect the zero and span gas to the rear panel bulkheads labeled ZERO and SPAN otherwise connect a source of vented zero air to the SAMPLE bulkhead Note Do not use nitrogen for zero air A gt is used to activate the zero and span
159. ermination Formats Format Reply Termination 00 lt CR gt 01 lt NL gt sum xxxx lt CR gt where xxxx 4 hexadecimal digits that represent the sum of all the characters bytes in the message host name This command reports the host name string Send host name Receive host name analyzero1 set host name string This command sets the host name string where string is 1 13 alphanumeric characters Send set host name analyzer01 Receive set host name analyzer01 ok instr name This command reports the instrument name Send instr name Receive instr name S02 Analyzer S02 Analyzer Model 43 Instruction Manual B 37 C Link Protocol Commands Communications Configuration B 38 Model 43 Instruction Manual instrument id This command reports the instrument id Send instrument id Receive instrument id 12 set instrument id value This command sets the instrument id to value where value is a decimal number between 0 and 127 inclusive Note Sending this command via RS 232 or RS 485 will require the host to use the new id for subsequent commands amp Send set instrument id 12 Receive set instrument id 12 ok mode This command reports what operating mode the instrument is in local service or remote The example that follows shows that the instrument is in the remote mode Send mode Receive mode remote set mode local set mode remote These commands set the instrument to local or remote mode T
160. et The zero and span check screens are visible only if the zero span check option is enabled and function the same way In the Main Menu choose Alarms gt Zero or Span Check Thermo Fisher Scientific Model 43i Instruction Manual 3 67 Operation Alarms Menu Max Zero and Span Offset Zero and Span Auto Calibration S0 Concentration 3 68 Model 43i Instruction Manual EHO Calor The Max Zero Check Offset screen is used to change the maximum zero check offset The maximum zero and span offset screens function the same way In the Main Menu choose Alarms gt Zero or Span Check gt Max Offset DH mri PEDER i Priv ig The Zero and Span Check screen allows the user to view the status of the most recent zero span checks and set the maximum check offsets An alarm will be triggered if a zero or span check indicates drift that exceeds the offset value The zero and span check screens are visible only if the zero span check option is enabled Their functions are similar In the Main Menu choose Alarms gt Zero or Span Autocal CPE Basen Cp gies The SO Concentration screen displays the current SO concentration and sets the maximum alarm limit Acceptable alarm limits range from 0 to 10000 ppm The minimum alarm may be programmed as a floor trigger Thermo Fisher Scientific Operation Alarms Menu alarm is triggered when the concentration falls below the minimum value or a ceiling trigger
161. f two or more of the same instruments are connected to one computer Valid Instrument ID numbers are from 0 to 127 The Model 43 has a default Instrument ID of 43 For more information about the Instrument ID see Appendix B C Link Protocol Commands or Appendix C MODBUS Protocol In the Main Menu choose Instrument Controls gt Communication Settings gt Instrument ID ENORME EN EUH o Bene Thermo Fisher Scientific Model 43i Instruction Manual 3 33 Operation Instrument Controls Menu Communication Protocol The Communication Protocol screen is used to change the instrument communication protocol for serial communications In the Main Menu choose Instrument Controls gt Communication Settings gt Communication Protocol Streaming Data The Streaming Data Configuration menu is used to allow for configuration Configuration of the 8 streaming data output items streaming interval current data format and current timestamp setting The Choose Item Signal submenu displays a list of the analog output signal group choices to choose from Choices are Concentrations Other Measurements and Analog Inputs if the I O expansion board option is installed Note Add Labels Prepend Timestamp and Add Flags are toggle items that change between yes or no when selected A In the Main Menu choose Instrument Controls gt Communication Settings gt Streaming Data Config 3 34 Model 43i Instruction Manual Thermo Fisher
162. fic Model 43i Instruction Manual 3 47 Operation Instrument Controls Menu Table 3 8 Signal Types Group Choices a Cie hi A Pida Concentrations Other Analog Measurements Inputs None None None S02 Internal Temperature Analog Input 1 1 0 Expansion single auto range only Board installed LO S02 Chamber Temperature Analog Input2 1 0 Expansion dual range only Board installed HI S02 Perm Oven Gas only Analog Input3 1 0 Expansion dual auto range only if perm oven installed Board installed Range Status auto range only Perm Oven Heater only if perm oven installed Analog Input 4 1 0 Expansion Board installed Chamber Pressure Analog Input 5 1 0 Expansion Board installed Sample Flow Analog Input6 1 0 Expansion Board installed PMT Volts Analog Input 7 1 0 Expansion Board installed Flash Volts Analog Input8 1 0 Expansion Board installed Flash Ref The Analog Input Configuration menu displays a list of the 8 analog input channels available for configuration This screen is only displayed if the I O expansion board option is installed Configuration includes entering descriptor units decimal places and choice of 2 10 table points and corresponding number of points selected Analog Input Configuration In the Main Menu choose Instrument Controls gt I O Configuration gt Analog Input Config 3 48 Model 43i Instruction Manual Thermo Fisher Scientific
163. for broadcast MODBUS commands is not supported Device IDs 128 through 247 i e 0x80 hex to OxF7 hex are not supported because of limitations imposed by C Link A MBAP Header In MODBUS over TCP IP a MODBUS Application Protocol Header MBAP is used to identify the message This header consists of the following components Transaction Identifier 2 Bytes 0x0000 to OxFFFF Passed back in response Protocol Identifier 2 Bytes 0x00 MODBUS protocol Length 2 Bytes 0x0000 to OxFFFF Number of following bytes Unit Identifier 1 Byte 0x00 to OxFF Passed back in response A Slave address is not required in MODBUS over TCP IP because the higher level protocols include device addressing The unit identifier is not used by the instrument C 2 Model 43 Instruction Manual Thermo Fisher Scientific Function Code Data Error Check Function Codes 0x01 0x02 Read Coils Thermo Fisher Scientific Read Inputs MODBUS Protocol Function Codes The function code is a single byte in length The following function codes are supported by the instrument Read Coils 0x01 Read Inputs 0x02 Read Holding Registers 0x03 Read Input Registers i 0x04 Force Write Single Coil 0x05 Read Exception Status 0x07 If a function code is received that is not in this list and invalid function exception is returned The data field varies depending on the function For more description of these data fields see Function Codes below I
164. for that instrument This is achieved by using the supported Geysitech commands listed below For details of the Model 437 Geysitech Protocol specification see the following topics Serial Communication Parameters on page D 1 TCP Communication Parameters on page D 2 Instrument Address on page D 2 Abbreviations Used on page D 2 Basic Command Structure on page D 2 Block Checksum on page D 3 Geysitech Commands on page D 3 The following are the communication parameters that are used to configure the serial port of the Series to support Geysitech protocol Number of Data bits 8 Number of Stop bits 1 Parity None Data rate from 1200 115200 Baud 9600 is default Model 437 Instruction Manual D 1 Geysitech Bayern Hessen Protocol TCP Communication Parameters TCP Communication Parameters Instrument Address Abbreviations Used Basic Command Structure D 2 Model 43 Instruction Manual iSeries Instruments support the Geysitech TCP protocol over TCP IP The register definition is the same as for the serial interface TCP connection port for Geysitech 9882 The Geysitech instrument address has a value between 0 and 127 and is represented by 3 digit ASCII number with leading zeros or leading spaces if required e g Instrument address of 1 is represented as 001 or lt SP gt lt SP gt 1 The instrument Address is the same as the Instrument ID used for C Link and MODBUS commands
165. g 32 data items and a full srec containing 32 items once each minute for a week gt 20 000 total records If logging is limited to the minimum content of date time concentration and error flags the analyzer can store data once each minute for four months gt 190 000 records In the Main Menu choose Instrument Controls gt Datalogging Settings Select SREC LREC The Select SREC LREC is used to select short record or long record format for other operations in this menu In the Main Menu choose Instrument Controls gt Datalogging Settings gt Select SREC LREC View Logged Data The View Logged Data screen is used to select the start point to view the logged data by number or records or date and time 3 24 Model 43i Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu In the Main Menu choose Instrument Controls gt Datalogging Settings gt Select SREC or LREC gt View Logged Data ts EUS EEE E Suai baene PME UY fad iat ah PUE Si aet FE hae PEE F Number of Records The Number of Records screen is used to select the starting record to display the number of records back to view ER Ea Cdra MT CH Ci EN EEE TE haa PME Aad Bast ade 5 ii The Record Display screen read only displays the selected records Date and Time The Date and Time screen is used to set a start date and time for which to view logged data For example if 20 JUN 2005 10 00 is entered
166. g any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Turn instrument OFF unplug the power cord and remove the cover 2 Disconnect the flash trigger cable from the FLASH TRIG connector the measurement interface board 3 Remove the bench see Optical Bench Replacement in this chapter 4 Loosen the retaining screw on the top of the lamp housing and pull out the socket assembly and lamp Thermo Fisher Scientific Model 43i Instruction Manual 7 13 Servicing Flash Lamp Voltage Adjustment Flash Lamp Voltage Adjustment 7 14 Model 43i Instruction Manual Flash Trigger Assy Socket Assy f Retaining Screw Figure 7 7 Replacing the Flash Lamp and Flash Trigger Assembly 5 Remove the old lamp from the socket assembly by pulling straight out and insert the new lamp 6 Insert the socket assembly into the lamp housing tighten the retaining screw and reconnect the flash trigger cable 7 Re install the bench 8 Adjust the lamp voltage Refer to the Flash Lamp Voltage Adjustment procedure that follows Use the following procedure to adjust the flash lamp voltage ote The service procedures in this manual are restricted to qualified service representatives 4 Thermo Fisher Scientific Flash Trigger Assem
167. goes beyond either the minimum or maximum limit an alarm is activated The word ALARM appears in the Run screen and in the Main Menu In the Main Menu choose Alarms gt Lamp Intensity The Minimum Lamp Intensity alarm limit screen is used to change the minimum lamp intensity alarm limit The minimum and maximum lamp intensity screens function the same way In the Main Menu choose Alarms gt Lamp Intensity gt Min or Max Thermo Fisher Scientific Operation Alarms Menu ENORME Bs EEE Eset Boes Lamp Voltage The Lamp Voltage screen displays the current lamp voltage and sets the minimum and maximum alarm limits Acceptable alarm limits range from 500 to 1200 volts If the lamp voltage goes beyond either the minimum or maximum limit an alarm is activated The word ALARM appears in the Run screen and in the Main Menu In the Main Menu choose Alarms gt Lamp Voltage Ea Car Py Tite mb Drini Le EN EEE TE haa PME Fad CP dd Min and Max Lamp The Minimum Lamp Voltage alarm limit screen is used to change the Voltage Limits minimum lamp voltage alarm limit The minimum and maximum lamp voltage screens function the same way In the Main Menu choose Alarms gt Lamp Voltage gt Min or Max ty i I E DHI Py cag ee mid Li il rrini iL miii Zero and Span The Zero Span Check screen allows the user to view the status of the most Check cent zt check and set the maximum zero check offs
168. gure Datalogging gt Memory Allocation t E ERE Pi Ba EHO Calor Thermo Fisher Scientific Operation Instrument Controls Menu Data Treatment The Data Treatment screen is used to select the data type for the selected record type whether the data should be averaged over the interval the minimum or maximum measured during the interval or the current value last value measured Data treatment doesn t apply to all data just to the concentration measurement All other data points log the current value at the end of the interval In the Main Menu choose Instrument Controls gt Datalogging Settings gt Configure Datalogging gt Data Treatment Ea Cad Ene EMO Mri wail nd A Lr i Communication The communication Settings menu is used with communications control and configuration Settings 5 In the Main Menu choose Instrument Controls gt Communication Settings E E Cdra PTH Dl itt Bet LEP Eesi my Lid E afl EN EEE TE haa PME Aad A EE feo PEE aE Baud Rate The Baud Rate screen is used to set the RS 232 RS 485 interface baud rate Baud rates of 1200 2400 4800 and 9600 19200 38400 57600 and 115200 are available In the Main Menu choose Instrument Controls gt Communication Settings gt Baud Rate Thermo Fisher Scientific Model 43i Instruction Manual 3 31 Operation Instrument Controls Menu tse ERE p CH ENMENSZ AE bli Lid bal CH CDE Pida baa nit A Ha Serial Setting
169. he example that follows sets the instrument to the local mode Send set mode local Receive set mode local ok program no This command reports the analyzer s model information and program version number which will be dependant on the current version Send program no Receive program no iSeries 43i 01 01 10 003 set layout ack This command disables the stale layout layout change indicator that is attached to each response if the layout has changed Send set layout ack Receive set layout ack ok Thermo Fisher Scientific 1 0 Configuration Thermo Fisher Scientific C Link Protocol Commands 1 0 Configuration analog iout range channel This command reports the analog current output range setting for channels where channel must be between 1 and 6 inclusive The example that follows reports current output channel 4 to the 4 20 mA range according to Table B 10 This command responds with feature not enabled if the I O expansion board is not detected Send analog iout range 4 Receive analog iout range 4 2 set analog iout range channel range This command sets analog current output channel to the channel range where channel is between 1 and 6 inclusive and range is set according to Table B 10 The example that follows sets current output channel 4 to the 0 20 mA range This command responds with feature not enabled if the I O expansion board is not detected Send set analog iout range 4 1 Receive
170. he key assignment screen will not come up upon entering right arrow soft key combinations All items under the Service menu including the menu itself cannot be assigned soft keys A The Model 43 utilizes the menu driven software as illustrated by the flowchart in Figure 3 3 The Power Up screen shown at the top of the flowchart is displayed each time the instrument is turned on This screen is displayed while the instrument is warming up and performing self checks After the warm up period the Run screen is automatically displayed The Run screen is the normal operating screen It displays the SO2 concentration depending on operating mode From the Run screen the Main Menu can be displayed by pressing _ _ The Main Menu contains a list of submenus Each submenu contains related instrument settings This chapter describes each submenu and screen in detail Refer to the appropriate sections for more information Thermo Fisher Scientific Operation Software Overview piomsseg 13 UQ puomssed SAQUIIY Psomsseg abueyo juaun43su 490 un 49077 Piomssed 18S 195f ynejeq a1ojsey 1591 9XId Aejdsiq oney uonnig sefuey pepua xy sues uA Weg 189 indy Bojeuy 189 dino Bojeuy snyeis Pa dx3 O I snje1s pg Jul snes a 23u09 os e9omy ueds e90ny 013Z 199249 ueds ju ye U0D 6yuoy jusunsul sindino Bojeuy s l sajels ejoy syndy eu61g piomssed JUO ebuey oyny 10 eng 1
171. hese commands report the converter temperature alarm minimum and maximum value current settings The example that follows reports that the converter temperature alarm minimum value is 205 0 C Send alarm converter temp min Receive alarm converter temp min 205 0 deg C set alarm converter temp min value set alarm converter temp max value These commands set the converter temperature alarm minimum and maximum values to value where value is a floating point number representing converter temperature alarm limits in degrees C The example that follows sets the converter temperature alarm maximum value to 215 0 C Send set alarm converter temp max 215 Receive set alarm converter temp max 215 ok Thermo Fisher Scientific C Link Protocol Commands Alarms alarm internal temp min alarm internal temp max These commands report the internal temperature alarm minimum and maximum value current settings The example that follows reports that the internal temperature alarm minimum value is 15 0 C Send internal temp alarm min Receive internal temp alarm min 15 0 deg C set internal temp alarm min value set internal temp alarm max value These commands set the internal temperature alarm minimum and maximum values to value where value is a floating point number representing internal temperature alarm limits in degrees C The example that follows sets the internal temperature alarm maximum value to 45 0 C Send set internal t
172. his command sets the instrument baud rate The example that follows sets the instrument s baud rate to 9600 Note After the command is sent the baud rate of the sending device must be changed to agree with the instrument A Send set baud 9600 Receive set baud 9600 ok dhcp This command reports the current state of use of DHCP on or off DHCP is used to assign an IP address to the analyzer automatically The example that follows shows that DHCP is on Send dhcp Receive dhcp on set dhcp onoff These commands enables and disables the DHCP service by either on or off Changes to this parameter will only take effect when the analyzer is powered up The example that follows sets the DHCP service on Note When DHCP is set to on the user supplied addr gw addr dns addr ip and addr nm parameters are not used A Send set dhcp on Receive set dhcp on ok Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Communications Configuration format This command reports the current reply termination format The example that follows shows that the reply format is 00 which means reply with no checksum according to Table B 9 Send format Receive format 00 set format format This command sets the reply termination format where format is set according to Table B 9 The example that follows sets the reply termination format to checksum Send set format 01 Receive set format 01 ok Table B 9 Reply T
173. hough the background is expressed in terms of concentration the background signal is actually the combination of electrical noise and scattered light Before the analyzer sets the SO reading to zero it stores this value as the SO background correction The SO Background screen is used to perform a manual adjustment of the instruments zero background As such the instrument should sample zero air until stable readings are obtained The display shows the current SO reading This reading is the SO2 background signal The next line of the display shows the SO background correction that is stored in memory and is being used to correct the SO reading That is the SO background correction is subtracted from the SO reading In the example that follows the analyzer is reading 25 7 ppb of SO while sampling zero air The SO background correction is 0 0 ppb That is the analyzer is not applying a zero background correction The question mark is used as a prompt to change the background correction In this case the background correction must be increased to 25 7 ppb in order for the SO reading to be at 0 ppb To set the SO reading in the example that follows to zero use _ to increment the SO background correction to 25 7 ppb As the SO background correction is increased the SO2 concentration is decreased At this point however no real changes have been made To escape this screen without saving any changes press to return to the Cali
174. ibes the front panel display the front panel pushbuttons and the menu driven software Chapter 4 Calibration provides the procedures for calibrating the analyzer and describes the required equipment Chapter 5 Preventive Maintenance provides maintenance procedures to ensure reliable and consistent instrument operation Chapter 6 Troubleshooting presents guidelines for diagnosing analyzer failures isolating faults and includes recommended actions for restoring proper operation Chapter 7 Servicing presents safety alerts for technicians working on the analyzer step by step instructions for repairing and replacing components and a replacement parts list It also includes contact information for product support and technical information Chapter 8 System Description describes the function and location of the system components provides an overview of the software structure and includes a description of the system electronics and input output connections Chapter 9 Optional Equipment describes the optional equipment that can be used with this analyzer Appendix A Warranty is a copy of the warranty statement Appendix B C Link Protocol Commands provides a description of the C Link protocol commands that can be used to remotely control an analyzer using a host device such as a PC or datalogger Model 43 Instruction Manual vii About This Manual Safety and Equipment Damage Alerts
175. ic Thermo Fisher Scientific C Link Protocol Commands Commands List Table B 1 C Link Protocol Commands Command Description Page addr dns Reports sets dns address B 35 addr gw Reports sets default gateway address B 35 addr ip Reports sets IP address B 35 addr nm Reports sets netmask address B 35 agc int Reports current AGC intensity B 15 alarm chamber Reports sets chamber temperature alarm maximum value B 11 temp max alarm chamber Reports sets chamber temperature alarm minimum value B 11 temp min alarm conc so2 Reports sets current SO concentration alarm maximum value B 12 max alarm conc so2 Reports sets current SO concentration alarm minimum value B 12 min alarm converter Reports sets SO converter temperature alarm maximum value B 12 temp max Model 4507 only alarm converter Reports sets SOzconverter temperature alarm minimum value B 12 temp min Model 450 only alarm internal Reports sets internal temperature alarm maximum value B 13 temp max alarm internal Reports sets internal temperature alarm minimum value B 13 temp min alarm pressure Reports sets pressure alarm maximum value B 13 max alarm pressure Reports sets pressure alarm minimum value B 13 min alarm sample flow Reports sets sample flow alarm maximum value B 14 max alarm sample flow Reports sets sample flow alarm minimum value B 14 min alarm trig conc Reports sets current SO
176. ignal LD3_5V LCD Signal LCD_ONOFF_5V LCD Signal LFLM_5V LCD Signal NC LLP_5V LCD Signal LCLK_5V LCD Signal 5V Ground 25V LCD Bias Voltage Ground KEYBOARD J3 0 N OQO wo A WO N Keypad Row 1 Input Keypad Row 2 Input Keypad Row 3 Input Keypad Row 4 Input Keypad Col 1 Select Keypad Col 2 Select Keypad Col 3 Select Keypad Col 4 Select LCD BACKLIGHT J4 5V Supply NC Ground Thermo Fisher Scientific Thermo Fisher Scientific Troubleshooting Connector Pin Descriptions Table 6 8 0 Expansion Board Optional Connector Pin Descriptions Connector Label EXPANSION 1 0 Reference Designator J1 Pin Signal Description 1 OO a N on A WW N gt gt o gt gt gt gt gt gt gt gt co loo N o a au N N N N N N WN 0d Aa Wo N O Analog Voltage Input 1 Analog Voltage Input 2 Analog Voltage Input 3 Ground Analog Voltage Input 4 Analog Voltage Input 5 Analog Voltage Input 6 Ground Analog Voltage Input 7 Analog Voltage Input 8 Ground NC Current Output Return Ground Current Output 1 Current Output Return Current Output 2 Current Output Return Current Output 3 Current Output Return Current Output 4 Current Output Return Current Output 5 Current Output Return Current Output 6 MOTHER BD J2 YN 0 mo Aa WO N 5V 24V 24V Ground Ground Ground RS485 to Motherboard Model 431 Instruction Manual 6
177. imit value If not select proper range Thermo Fisher Scientific Thermo Fisher Scientific Troubleshooting Troubleshooting Guides Alarm Possible Cause Action Message Concentration low Check user defined low set point be sure the min trigger is set as desired Alarm Zero Instrument out of Recalibrate instrument Check calibration Alarm Span Check Alarm Zero Instrument out of Check gas supply Perform manual calibration Autocal calibration Alarm Span Autocal Alarm Internal cables not Check that all internal cables are connected Motherboard connected properly properly Recycle AC power to instrument If still Status alarming change board Alarm Board is defective Interface Status Alarm 1 0 Exp Status Model 43i Instruction Manual 6 9 Troubleshooting Board Level Connection Diagrams Board Level Figure 6 1 and Figure 6 2 are board level connection diagrams for the common electronics and measurement system These illustrations can be Connection Diagrams used along with the connector pin descriptions in Table 6 5 through Table 6 12 to troubleshoot board level faults VOEXPANSION BOARD EXPANSION I O J1 25 Pin TOMEASUREVENT SYSTEM DIGTALOUIPUT BOARD DIGITAL OUTPUTS 37 Pin J2 RS232 RS485 FRONT ee LCD 9 Pin in j PANEL DISPLAY BOARD RS232 RS485 y pin Pn R 10 BASE T B e KEY PANEL 8 Pin 8 Pin EXPANSION RS485 13 Pin SPARE DATA ANALOG OUTPUTS DIGITAL INPUTS 37 pin 1PJ8
178. inputs as a 4 digit hexadecimal string with the most significant bit MSB being input 16 Send dig in Receive dig in Oxff7f din channel This command reports the action assigned to input channel and the corresponding active state The example that follows reports the input 1 to be assigned an index number 3 corresponding to action of SO2 mode with the active state being high Send din 1 Receive din 1 3 S02 MODE high set din channel index state This command assigns digital input channel 1 16 to activate the action indicated by index 1 35 when the input transitions to the designated state high or low Use list din var command to obtain the list of supported index values and corresponding actions Send set din 5 9 high Receive set din 1 9 high ok Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands 1 0 Configuration dout channel This command reports the index number and output variable and the active state assigned to output channel The example that follows reports the input 4 to be assigned an index number 11 corresponding to general alarm with the active state being open Send dout 4 Receive dout 4 11 GEN ALARM open set dout channel index state This command assigns digital output channel to be assigned to the action associated with index and assigns it an active state of state open or closed Send set dout 4 11 open Receive set dout 4 11 open ok dtoa channel This
179. intenance chapter Replace any lines made of vinyl or other plastics with fresh Teflon or stainless steel Replace Teflon filter membranes that look dirty Remove any filters that are not Teflon membranes Analog signal doesn t match expected value Software has not been configured Analog output goes above full scale value or below zero Recorder is drawing down output Verify that the selected analog output has been properly configured to match the data system By default a 5 over and under range on the analog outputs is provided If this is not desirable due to system restrictions it may be turned off in the INSTRUMENT CONTROLS gt 1 0 CONFIGURATION gt ANALOG OUTPUT CONFIG screens Verify that the recorder or data logger input impedance meets minimum requirements Flow meter fluctuations Thermo Fisher Scientific Dirty pump diaphragm Capillary blocked Clogged Teflon line Clean or replace pump diaphragm Clean or replace capillary Inspect all sample lines Model 43i Instruction Manual 6 7 Troubleshooting Troubleshooting Guides 6 8 Model 43i Instruction Manual Table 6 4 Troubleshooting Alarm Messages Alarm Message Alarm Internal Temp Possible Cause Instrument overheating Action Replace fan if not operating properly Clean or replace foam filter refer to the Preventive Maintenance chapter in this manual Check 10K thermistor on measurement in
180. isplayed at one time For these menus use Jand 4 to move the cursor up and down to each item ia NED Figure 3 1 Front Panel Display CAUTION If the LCD panel breaks do not to let the liquid crystal contact your skin or clothes If the liquid crystal contacts your skin or clothes wash it off immediately using soap and water A 3 2 Model 43i Instruction Manual Thermo Fisher Scientific Operation Pushbuttons Pushbuttons The Pushbuttons allow the user to traverse the various screens menus Table 3 1 lists the front panel pushbuttons and their functions Thermo Fisher Scientific Figure 3 2 Front Panel Pushbuttons Table 3 1 Front Panel Pushbuttons Key Name E Soft Keys Function The E soft keys are used to provide shortcuts that allow the user to jump to user selectable menu screens For more information on processing soft keys see Soft Keys Cr Run The Cr is used to display the Run screen The Run screen normally displays the SO concentration Cm Menu The Cm is used to display the Main Menu when in the Run screen or back up one level in the menu system For more information about the Main Menu see Main Menu later in this chapter 2 Help The 2 is context sensitive that is it provides additional information about the screen that is being displayed Press 2 for a brief explanation about the current screen or menu Help messages are displayed
181. l Commands Appendix section of this document for more information on this command A The Service Mode screen is used to turn the service mode on or off The service mode locks out any remote actions and includes parameters and functions that are useful when making adjustments or diagnosing the Model 437 For more information about the service mode see Service Menu later in this chapter Model 43i Instruction Manual 3 53 Operation Instrument Controls Menu Note The service mode should be turned off when finished as it prevents remote operation A In the Main Menu choose Instrument Controls gt Service Mode Date Time The Date Time screen and allows the user to view and change the system date and time 24 hour format The internal clock is powered by its own battery when instrument power is off In the Main Menu choose Instrument Controls gt Date Time Timezone The Timezone screen is used to set the timezone for the NTP time server 3 54 Model 43i Instruction Manual This should be set to the timezone that the instrument is located in If the exact timezone is not shown in the list it may be entered via the CLINK TZ command see Appendix B The selections are UTC GMT EST GMT 5 CST GMT 6 MST GMT 7 PST GMT 8 YST GMT 9 HST GMT 10 NST GMT 11 DLW GMT 12 CET GMT 1 EET GMT 2 BST GMT 3 DLT GMT 4 ECH GMT 5 FOX GMT 6 GLF GMT 7 CCT GMT 8 JST GMT 9 GST GMT
182. l pressure by subtracting 0 027 mm Hg per foot of altitude A Do not try to calibrate the pressure transducer unless the pressure is known accurately A Equipment Required Vacuum pump Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A Thermo Fisher Scientific Servicing Pressure Transducer Calibration 1 Remove the cover 2 Disconnect the tubing from the pressure transducer and connect a vacuum pump known to produce a vacuum less than 1 mm Hg 3 From the Main Menu press to scroll to Service gt press gt to scroll to Pressure Calibration gt and press The Pressure Sensor Cal menu appears Note If Service Mode is not displayed refer to Accessing the Service Mode in this chapter then return to the beginning of this step A 4 At the Pressure Sensor Cal screen press to select Zero The Calibrate Pressure Zero screen appears 5 Wait at least 10 seconds for the zero reading to stabilize then press to save the zero pressure value 6 Disconnect the pump from the pressure transducer 7 Press to return to the Pressure Sensor Cal
183. lecules onto the mirror assembly Model 43i Instruction Manual 8 3 System Description Bandpass Filter Photomultiplier Tube Photodetector Flow Sensor Pressure Transducer Capillary Vacuum Pump Software Instrument Control 8 4 Model 43i Instruction Manual The bandpass filter restricts the light reaching the photomultiplier tube to the SO fluorescence wavelengths The PMT converts optical energy from the reaction to an electrical signal This signal is sent to the input board which transmits it to the processor The photodetector is located in the fluorescence chamber and continuously monitors the pulsating UV flash lamp This photodetector is connected to a circuit that automatically compensates for fluctuations in flash lamp output The flow sensor is used for measuring the flow of sample gas in the measurement system The pressure transducer or pressure sensor measures the reaction chamber pressure The capillary along with the pump is used to control flow in the sample line The vacuum pump draws the reacted gases out of the reaction chamber The processor software tasks are organized into four areas e Instrument Control e Monitoring Signals e Measurement Calculations e Output Communication Low level embedded processors are used to control the various functions on the boards such as analog and digital I O and heater control These processors are controlled over a serial interface with a single
184. lier in the chapter In the Main Menu choose Service gt Display Pixel Test The Restore User Defaults screen is used to reset the user calibration and configuration values to factory defaults The restore default user is visible only when the instrument is in service mode For more information on the service mode see Service Mode earlier in the chapter In the Main Menu choose Service gt Restore User Defaults Model 43i Instruction Manual 3 85 Operation Password Set Password 3 86 Model 43i Instruction Manual DH Guided PIL The Password menu allows the user to configure password protection If the instrument is locked none of the settings may be changed via the front panel user interface The items visible under the password menu are determined by the instrument s password status In the Main Menu choose Set Password DH pikari E EA The Set Password screen is used to set the password to unlock the front panel The set password is shown if the instrument is unlocked and the password is set In the Main Menu choose Password gt Password Ehhe EE EE Thermo Fisher Scientific Operation Password Lock Instrument The Lock Instrument screen is used to lock the instrument s front panel so users can not change any settings from the front panel The lock instrument screen is shown if the instrument is unlocked and the password is set If the instrument keyb
185. lue A In the Main Menu choose Service gt Flow Calibration gt Set Defaults Input Board Test The Input Board Test screen is used to manually adjust the input board gain The input board test is visible only when the instrument is in service mode For more information on the service mode see Service Mode earlier in the chapter Note This adjustment should only be performed by an instrument service technician A In the Main Menu choose Service gt Input Board Test 3 76 Model 43i Instruction Manual Thermo Fisher Scientific Operation Service Menu Temperature The Temperature calibration screen allows the user to view and set the Calibration ambient temperature sensor calibration The temperature calibration is visible only when the instrument is in service mode For more information on the service mode see Service Mode earlier in the chapter Note This adjustment should only be performed by an instrument service technician A Note Wait at least 30 seconds for the reading to stabilize before saving the value A In the Main Menu choose Service gt Temperature Calibration Analog Output The Analog Output Calibration menu is a selection of analog output to Calibration calibrate and allows the user to select the calibration action zero or span The analog output calibration is visible only when the instrument is in service mode For more information on the service mode see Service Mode earlier in th
186. lumbing arrangement If a filter is used all gas must enter the instrument through the filter For detailed information about the menu parameters and the icons used in these procedures see the Operation chapter Note The calibration and calibration check duration times should be long enough to account for the transition purge process when switching from sample to zero and from zero to span This transition time is the time required to purge the existing air A Depending on the plumbing configuration and the instrument data from approximately the first minute of a zero calibration or check should be disregarded because of residual sample air Also data from approximately the first minute of a span calibration or check should be disregarded because the span is mixing with the residual zero air A Use the following procedure to perform a zero span calibration 1 Ifthe instrument is equipped with the optional zero span and sample valves connect the zero and span gas to the rear panel bulkheads labeled ZERO and SPAN otherwise connect a source of vented zero air to the SAMPLE bulkhead Note Do not use nitrogen for zero air A gt is used to activate the zero and span valves The lower left hand corner of the Run screen indicates which mode is active ZERO SPAN or SAMPLE 2 To ensure that the zero air is being measured at atmospheric pressure check that the zero air flow reported by the analyzer is approximately Mo
187. m 1200 to 19 200 baud The user can also set the data bits parity and stop bits The following protocols are supported e C Link e Streaming Data e Modbus Slave e Geysitech Bayern Hessen The Streaming Data protocol transmits user selected measurement data via the serial port in real time for capture by a serial printer data logger or PC RS 232 Connection A null modem crossed cable is required when connecting the analyzer to an IBM Compatible PC However a straight cable one to one may be required when connecting the analyzer to other remote devices As a general rule when the connector of the host remote device is female a straight cable is required and when the connector is male a null modem cable is required Data Format 1200 2400 4800 9600 19200 38400 57600 or 115200 BAUD 8 data bits 1 stop bit no parity All responses are terminated with a carriage return hex 0D Refer to Table 8 1 for the DB9 connector pin configuration Thermo Fisher Scientific Model 43i Instruction Manual 8 11 System Description 1 0 Components RS 485 Connection Ethernet Connection External Accessory Connector 8 12 Model 43i Instruction Manual Table 8 1 RS 232 DB Connector Pin Configurations DB9Pin Function 2 RX 3 TX 7 RTS 8 CTS 5 Ground The instrument uses a four wire RS 485 configuration with automatic flow control SD Refer to Table 8 2 for the DB9 connector pin configuration Table
188. mistor Ground Thermo Fisher Scientific Thermo Fisher Scientific Troubleshooting Connector Pin Descriptions Connector Label Reference Designator Pin Signal Description PERM OVEN POJ3 1 Perm Oven Heater On Off 2 15V Power 3 Perm Oven Heater Thermistor 4 Ground Table 6 7 Front Panel Board Connector Pin Diagram Connector Label MOTHER BOARD Reference Designator Pin J1 1 Oo a N O o A N gt gt o gt gt gt gt gt gt gt gt co loo N o a au N N N N N N N N N YN OO RA UO N O Signal Description Ground Ground LCLK LCD Signal Ground Ground LLP LCD Signal LFLM LCD Signal LD4 LCD Signal LDO LCD Signal LD5 LCD Signal LD1 LCD Signal LD6 LCD Signal LD2 LCD Signal LD7 LCD Signal LD3 LCD Signal LCD Bias Voltage 5V Ground Ground LCD_ONOFF LCD Signal Keypad Row 2 Input Keypad Row 1 Input Keypad Row 4 Input Keypad Row 3 Input Keypad Col 2 Select Keypad Col 1 Select Keypad Col 4 Select Model 43i Instruction Manual 6 19 Troubleshooting Connector Pin Descriptions 6 20 Model 43i Instruction Manual Connector Label Reference Designator Pin 28 29 30 31 32 33 34 Signal Description Keypad Col 3 Select Ground Ground Ground Ground 24V 24V LCD DATA J2 ay oOo a nN oon FP WW DN gt gt gt gt w N LDO_5V LCD Signal LD1_5V LCD Signal LD2_5V LCD S
189. mode The example that follows sets the instrument to sample mode that is the instrument is reading the sample gas Send set sample Receive set sample ok set zero set zero gas This command sets the zero span valves to the zero mode The example that follows sets the instrument to zero mode that is the instrument is reading the sample gas Send set zero Receive set zero ok set span set span gas This command sets the zero span valves to the span mode The example that follows sets the instrument to span mode that is the instrument is sampling span gas Send set span Receive set span ok Thermo Fisher Scientific C Link Protocol Commands Measurement Configuration gas unit This command reports the current gas units ppb or 1g m The example reports that the gas unit is set to ppm Send gas unit Receive gas unit ppm set gas unit unit ppm mg m This command sets the gas units to ppm or mg m The example that follows sets the gas units to mg m Send set gas unit mg m3 Receive set gas unit mg m3 ok pres comp This command reports whether pressure compensation is on or off The example that follows shows that pressure compensation is on Send pres comp Receive pres comp on set pres comp onoff These commands turn the pressure compensation on or off The example that follows turns pressure compensation off Send set pres comp off Receive set pres comp off ok temp comp This comm
190. mperature Power requirements 100 VAC O 50 60 Hz 115 VAC 50 60 Hz 220 240 VAC 50 60 Hz 165 watts Physical 16 75 W X 8 62 H X 23 D dimensions Weight Approximately 48 Ibs Analog outputs 6 voltage outputs 0 100 mV 1 5 10 V user selectable 5 of full scale over under range user selectable 12 bit resolution measurement output user selectable per channel Digital outputs 1 power fail relay Form C 10 digital relays Form A user selectable alarm output relay logic 100 mA 200 VDC Digital inputs 16 digital inputs user select programmable TTL level pulled high Serial Ports 1 RS 232 or RS 485 with two connectors baud rate 1200 115200 data bits parity and stop bits protocols C Link MODBUS Geysitech Bayern Hessen and streaming data all user selectable Ethernet connection RJ45 connector for 10 Mbs Ethernet connection static or dynamic TCP IP addressing In non condensing environments Performance specifications based on operation within 20 30 C range Thermo Fisher Scientific Thermo Fisher Scientific Introduction Specifications Table 1 2 Model 43i Optional Permeation Oven Specifications Temperature control Single Point 45 C Temperature 0 1 C stability Warm up time 1 hour permeation device can take 24 to 48 hours to stabilize Carrier gas flow 70 scc min Chamber size Accepts permeation tubes up to 9 cm in total length 1 cm
191. mperature Thermistor with Connector 100907 00 Fan 24VDC 8630 Filter Guard Assembly w foam 8919 Capillary 0 013 inch ID 0 5 LPM flow rate standard 4800 Capillary O ring 4119 Capillary 0 008 inch ID Perm Oven Option Table 7 2 describes the Model 43i cables See the Troubleshooting chapter for associated connection diagrams and board connector pin descriptions Table 7 2 Model 43i Cables PartNumber Description 101036 00 DC Power Supply 24V Output 01037 00 115VAC Supply to Measurement Interface Board 01048 00 RS 485 Data 01038 00 Power Switch to Motherboard 01364 00 DC Power Supply Status Monitor 01054 00 Motherboard to Front Panel Board 101035 00 DC Power Supply AC Input 101033 00 AC from Receptacle 01377 00 AC to Power Switch 01355 00 Signal Output Ribbon 01695 00 Permeation Oven 01055 00 Main AC Receptacle Assembly 01267 00 Fan Power Cable Table 7 3 lists the standard and optional cables and components used for connecting external devices such as PCs and data loggers to an Series instrument Table 7 3 External Device Connection Components Thermo Fisher Scientific Thermo Fisher Scientific Servicing External Device Connection Components Part Number Description 102562 00 Terminal Block and Cable Kit DB25 optional 102556 00 Terminal Block and Cable Kit DB37 optional 102645 00 Cable DB37M to Open End Six Feet optional
192. n read only displays the current sample flow rate The flows is measured by an internal flow sensor In the Main Menu choose Diagnostics gt Flow OL oer Lamp Intensity The Lamp Intensity screen read only displays the current lamp intensity in Hertz Normally the intensity would read 90 but this number will decrease over time as the lamp degrades In the Main Menu choose Diagnostics gt Lamp Intensity Pa E mh rh POPENEE Paria rita Li irii EH Cie hi Optical Span Test The Optical Span Test screen is used to turn the optical span test LED on or off and displays the SO2 concentration reading Within the fluorescence chamber is a light emitting diode LED which may be used to simulate a particular concentration of SO It is designed to provide a quick and easy way of checking the optics and electronics for span drift or other problems The difference between when the LED is ON and OFF must be greater than 50 ppb In the Main Menu choose Diagnostics gt Optical Span Test 3 58 Model 43i Instruction Manual Thermo Fisher Scientific Operation Diagnostics Menu righ ieee rh igs A py ppe Led EN EEE TE Beee nr A EE fee PEE EE Analog Input The Analog Input Readings screen read only displays the 8 current user Readings scaled analog readings if the I O expansion board option is installed In the Main Menu choose Diagnostics gt Analog Input Readings PRE Car ire pui Drini Lii EN PU
193. n MODBUS over Serial an error check is included in the message This is not necessary in MODBUS over TCP IP because the higher level protocols ensure error free transmission The error check is a two byte 16 bit CRC value This section describes the various function codes that are supported by the Model 432 Read Coils Inputs reads the status of the digital outputs relays in the instrument Issuing either of these function codes will generate the same response These requests specify the starting address i e the address of the first output specified and the number of outputs The outputs are addressed starting at zero Therefore outputs numbered 1 16 are addressed as 0 15 The outputs in the response message are packed as one per bit of the data field Status is indicated as 1 Active on and 0 Inactive off The LSB of the first data byte contains the output addressed in the query The other outputs follow toward the high order end of this byte and from low order to high order in subsequent bytes If the returned output quantity is not a multiple of eight the remaining bits in the final data byte will be padded with zeros toward the high order end of the byte The Byte Count field specifies the quantity of complete bytes of data Model 43 Instruction Manual C 3 MODBUS Protocol Function Codes C 4 Model 43 Instruction Manual Note The values reported may not reflect the state of the actual relays in the instrument a
194. n Signal Description 1 Analog1 13 Power_Fail_NC 2 Analog ground 14 Power_Fail_COM 3 Analog2 15 Power_Fail_NO 4 Analog ground 16 TTL_Input1 5 Analog3 17 TTL_Input2 6 Analog ground 18 TTL_Input3 7 Analog4 19 TTL_Input4 8 Analog ground 20 Digital ground 9 Analog5 21 TTL_Input5 10 Analog ground 22 TTL_Input6 11 Analog6 23 TTL_Input7 12 Analog ground 24 Digital ground Thermo Fisher Scientific Installation Connecting External Devices D O Terminal Board Figure 2 6 shows the recommended method for attaching the cable to the terminal board using the included tie down and spacer Table 2 2 identifies the connector pins and associated signals See Detail A See Detail B a Detail A Detail B d Assembled Connector Figure 2 6 D O Terminal Board Views Table 2 2 D O Terminal Board Pin Descriptions Pin Signal Description Pin Signal Description 1 Relay1_ContactA 13 Relay7_ContactA 2 Relay1_ContactB 14 Relay7_ContactB 3 Relay2_ContactA 15 Relay8_ContactA 4 Relay2_ContactB 16 Relay8_ContactB 5 Relay3_ContactA 17 Relay9_ContactA 6 Relay3_ContactB 18 Relay9_ContactB 7 Relay4_ContactA 19 Relay10_ContactA 8 Relay4_ContactB 20 Relay10_ContactB 9 Relay5_ContactA 21 Not Used 10 Relay5_ContactB 22 24V 11 Relay6_ContactA 23 Not Used 12 Relay6_ContactB 24 24V Thermo Fisher Scientific Model 43i Instruction Manual Installation Connecting External Devi
195. n the last data byte the status of outputs 15 11 is shown as the byte value 0x0A or binary 0000 1010 Output 15 is in the fifth bit position from the left and output 11 is the LSB of this byte The four remaining high order bits are zero filled Read holding input registers reads the measurement data from the instrument Issuing either of these function codes will generate the same response These functions read the contents of one or more contiguous registers Model 43 Instruction Manual C 5 MODBUS Protocol Function Codes C 6 Model 43 Instruction Manual These registers are 16 bits each and are organized as shown below All of the values are reported as 32 bit IEEE standard 754 floating point format This uses 2 sequential registers least significant 16 bits first The request specifies the starting register address and the number of registers Registers are addressed starting at zero Therefore registers numbered 1 16 are addressed as 0 15 The register data in the response message are packed as two bytes per register with the binary contents right justified within each byte For each register the first byte contains the high order bits and the second contains the low order bits Request Function code Starting Address Quantity of Registers Response Function code Byte count Register value N Quantity of Registers Error Response Function code Exception code 1 Byte 2 Bytes 2 Bytes 1 Byte 1 Byte N x
196. n the selected range or 0 or lower A In the Main Menu choose Calibration Factors gt Hi SO Coef 3 18 Model 43i Instruction Manual Thermo Fisher Scientific Operation Calibration Menu Reset User Calibration The Reset User Calibration Default screen allows the user to reset the Default calibration configuration values to factory defaults In the Main Menu choose Calibration Factors gt Reset User Cal Defaults KH cid ee Pi TTT oH tithe E Pi EEE i Oc ca dde ba DF Beene PEE EE Calibration The Calibration menu is used to calibrate zero and span The calibration Menu menu is similar for the single dual and autorange mode as shown below The dual and autorange modes have two SO span factors HI and LO This allows each range to be calibrated separately This is necessary if the two ranges used are not close to one another For example a low SO2 range of 500 ppb and a high SO range of 10 ppm For more information about calibration see the Calibration chapter In the Main Menu choose Calibration Cai CA mt BREE E usok Besse Lie i Sel Haat is T Cp bebe EAR Ea Hy perm CH ciinii Pii iini i EEE E unk Bases nr Beet Ba 5 PU Eh DEE EE Calibrate S02 The Calibrate SO Background screen is used to adjust the SO Background background or perform a zero calibration Before performing a zero calibration ensure the analyzer samples zero air for at least 5 minutes It is import
197. nel Pin Connections 7 23 Table 7 5 Analog Input Channels and Rear Panel Pin Connections 7 24 Table 8 1 RS 232 DB Connector Pin Configurations o io o o 8 12 Table 8 2 RS 485 DB Connector Pin Configurati0N onocinonninio o 8 12 Table 9 1 Cable Opticas 9 10 Table 9 2 Color Codes for 25 Pin and 37 Pin Cables 9 10 Table 9 3 Mounting Options a a te ese a es ak pete tee 9 11 Table B 1 C Link Protocol Commands cece ceseeeetestesteseeeeeetesteseesee B 3 Model 43 Instruction Manual xxiii Tables xxiv Model 437 Instruction Manual Tabl o B 2 Averaging Meson tas B 8 Table B 3 Alarm Trigger Values ooooiococoncicicoccininoncnncncnononocnrnnnonononororonoooos B 14 Table B 4 Record Output FormatS ooocicinnicioninconinoninnnnnnononocnnnncncnncnnnncnnnnos B 20 Table B 5 Stream Time Values oooocococicicociononiocinconnnonornornocnncoroororoornonoos B 24 Table B 6 Standard A ee B 28 Table B 7 Extended Ranges cccccccccccccecesesescetesesestetesesesteteseseetetessseeeeees B 29 Table B B Contrast sra As B 32 Table B 9 Reply Termination Formats ocociniconcocicincncnnnocnnnnnnncornnnnnncoronnonos B 37 Table B 10 Analog Current Output Range Values oocoooiincininincicncono B 39 Table B 11 Analog Voltage Output Range Values oooocoininoinincinicicninao B 40 Table B 12
198. ng x 0 250 inch OD x 0 013 inch ID and O ring Figure 5 2 Refer to Replacement Parts in the Servicing chapter 5 Check capillary for particulate deposits Clean or replace as necessary 6 Inspect O ring for cuts or abrasion and replace as necessary 7 Replace capillary in the holder making sure the O ring is around the capillary before inserting it into the body 8 Finger tighten the capillary nut enough to ensure a tight seal Model 43i Instruction Manual 5 3 Preventive Maintenance Capillary Inspection and Replacement 5 4 Model 43i Instruction Manual 9 Reinstall the cover 10 Connect the power cord and turn the instrument ON Capillary Nut Capillary Holder Figure 5 1 Inspecting and Replacing the Capillary Union Elbow rae ria Capillary Nut xK Capillary Adapter Capillary ahs ee O Ring re Capillary Holder Union 4 Figure 5 2 Capillary Assembly Thermo Fisher Scientific Fan Filter Inspection and Cleaning A Sample Particulate Filter Inspection Thermo Fisher Scientific Preventive Maintenance Fan Filter Inspection and Cleaning Under normal use the filter over the instrument fan located on the rear panel of the instrument should be cleaned and reconditioned at six month intervals If the instrument is operated in excessively dirty surroundings this procedure should be instituted on a more frequent schedule Use the following procedure to inspect and clean the
199. nnniconononnnnocorononnocoronnonororannonorornnnnnos 9 13 Figure 9 7 EIA Rack Mounting csi ni 9 14 Figure 9 8 Retrofit Rack Mounting 9 15 Figur B 1 Had Scs nue eee eee eae eel ae B 11 Thermo Fisher Scientific Thermo Fisher Scientific Tables Table 1 1 Model 43 SpecifiCatiONS oooconicicnicinininninonincnnoncnncnononornrnncnnnnos 1 3 Table 1 2 Model 43i Optional Permeation Oven Specifications 1 5 Table 2 1 O Terminal Board Pin DescriptiONS oionoininninincincnnonininncnnnnso 2 6 Table 2 2 D O Terminal Board Pin DescriptiONS onocinccnincinnnccnnccnnoss 2 7 Table 2 3 25 Pin Terminal Board Pin DescriptiONS o ocociciconccninincnncnrncnnnnnooos 2 8 Table 3 1 Front Panel Pushbuttons 0 0 cece cece cece cece eeeeeeeseeeeseeeeeeeeeseeeseeneees 3 3 Table 3 2 Default Analog Outputs in Single Range Mode 3 9 Table 3 3 Default Analog Outputs in Dual Range Mode 3 10 Table 3 4 Default Analog Outputs in Auto Range Mode 3 12 Table 3 5 Standard Ranges ccccccccccccscsscsssscssestssesessesesesesseseseseseeeeeeees 3 14 Table 3 6 Extended Ranges ooococoococicococonnnncnccncnnnononnrnnnonocorannonororannonororanos 3 14 Table 3 7 Analog Output Zero to Full Scale Values 3 47 Table 3 8 Signal Types Group Choices oooconcocinccncocinno
200. no coef 1 234 NO 21f Code_0 Code 1 Code 2 Code 3 Code 4 Code 5 Code 6 Code 7 Code 8 Code 9 Code 10 Code 11 Lset range no d n This is a line which appears slightly indented the title is again NO and the value the twenty first element of the data response interpreted as a floating point number There is a no translation button which creates a selection list of twelve Code nn options The number of the user selection is used to create the output command Mode 6 12 13x local remote service service 0 1 Tset mode s n This is a line which has a title of Mode and value taken from the sixth field of the data response There is a bitfield extraction of bits 12 through 13 from the source the value type is not important here because the value is being translated to an output string Once the bits have been extracted they are shifted down to the bit zero position Thus the possible values of this example will be 0 through 3 The translation list shows the words which correspond to each input value the zeroth value appearing first 0 gt local 1 gt remote etc The selection list shows that only the first two values in this case are to be shown to the user when the button is pressed The T button indicates full translation input code to string and user selection number to output string xC This is a line that starts a new column the xC or L Comp 6 11x off on Tset temp comp s n Model 43 Instruction Man
201. not come in contact with the sample gas and DO NOT need cleaning Read the Equipment Damage warning that follows Equipment Damage DO NOT attempt to clean the mirrors in the optical bench These mirrors do not come in contact with the sample gas and should not be cleaned Cleaning the mirrors can damage the mirrors A Thermo Fisher Scientific Capillary Inspection and Replacement Thermo Fisher Scientific A Preventive Maintenance Capillary Inspection and Replacement The capillary normally only requires inspection when instrument performance indicates that there may be a flow problem A decreased sample flow may be an indication that a capillary is blocked Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A Use the following procedure to inspect and replace the capillary 1 Turn the instrument OFF and unplug the power cord 2 Remove the instrument cover 3 Locate the capillary holder Figure 5 1 and unscrew the knurled capillary nut Note The capillary used with the optional heated kicker is located adjacent to the flow transducer A 4 Remove the glass capillary 1 5 inch lo
202. nt 6 Perform a photomultiplier tube calibration See PMT Voltage Adjustment in this chapter PMT Housing Cover Retaining Screws 3 PMT Cover Figure 7 9 Replacing the PMT PMTH igh Vo Itage Use the following procedure to replace the PMT high voltage power supply Figure 7 10 Power Supply Equipment Required Replacement PMT high voltage power supply Philips screwdriver amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A f Equipment Damage Some internal components can be damaged by small 1 Turn instrument OFF unplug the power cord and remove the cover 2 Disconnect the two PMT high voltage supply cables 3 Loosen the two retaining screws securing the power supply bracket to the floor plate and slide the power supply towards the rear slightly and lift it off the base screws 7 18 Model 43i Instruction Manual Thermo Fisher Scientific Servicing PMT Voltage Adjustment Assembly Bracket Screws 2 Input Box Assembly Figure 7 10 Replacing the PMT High Voltage Power Supply HVPS 4 Loosen two screws on the input box assembly and lift the input box assembly off the bracket 5 Remove the four screws securing
203. ntific Model 43i Instruction Manual 5 7 Preventive Maintenance Pump Rebuilding Top Plate Figure 5 4 Rebuilding the Pump 6 Check that the flow reads approximately 0 5 LPM when power is turned ON 7 Perform a leak test as described in this chapter 5 8 Model 43i Instruction Manual Thermo Fisher Scientific Safety Precautions Troubleshooting Thermo Fisher Scientific Guides Chapter 6 Troubleshooting This instrument has been designed to achieve a high level of reliability In the rare event of problems or failure the troubleshooting guidelines board level connection diagrams connector pin descriptions and testing procedures presented in this chapter should be helpful in isolating and identifying problems For additional fault location information refer to the Preventive Maintenance chapter in this manual The service mode described in the Operation chapter includes parameters and functions that are useful when making adjustments or diagnosing problems The Technical Support Department at Thermo Fisher Scientific can also be consulted in the event of problems See Service Locations at the end of this chapter for contact information In any correspondence with the factory please note both the serial number and program number of the instrument This chapter provides the following troubleshooting information e Safety Precautions on page 6 1 e Troubleshooting Guides on page 6 1 e Board
204. nto Irec list B 22 copy sp to srec Sets copies current selections in scratch pad into srec list B 22 copy sp to stream Sets copies current selections in scratch pad into stream list B 22 copy srec to sp Sets copies current srec selection into the scratch pad B 22 copy stream to sp Sets copies current streaming data selection into the scratch B 22 pad custom Reports sets defined custom range concentration B 29 data treatment Reports sets the current selection of data treatment for B 22 Irec srec concentrations in the Irecs or srecs date Reports sets current date B 33 default params Sets parameters to default values B 33 dhcp Reports sets state of use of DHCP B 36 diag volt ob Reports diagnostic voltage level for I O expansion board B 15 diag volt mb Reports diagnostic voltage level for motherboard B 15 diag volt mib Reports diagnostic voltage level for measurement interface B 15 board dig in Reports status of the digital inputs B 40 din Reports sets digital input channel and active state B 40 do down Simulates pressing down pushbutton B 26 dout Reports sets digital output channel and active state B 41 dtoa Reports outputs of the digital to analog converters per channel B 41 Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Commands List Command Description Page en enter Simulates pressing enter pushbutton B 26
205. o and Span Auto Calibration vincia latas 3 68 AAA a a a iSo ioi Oa EEOSE a S 3 68 Min and Max SO Concentration Limits oocoocncononocncnninoncnnnnnnnnns 3 69 Min Tigger UA A ae R T ERE der ie 3 69 Service Me iS 3 70 Flash Voltage Adjustment iii di 3 70 Initial Flash Reference rias Dni its 3 71 PMT Voltage Adjustment 3 71 Range Mode Selec ot edo ae 3 72 Pressure Calibration ii dit 3 72 Calibrate Pressure Zero ii ti E 3 73 Calibrate Pressure Pan nd oca 3 73 Restore Default Pressure Calibration ooononcicnonnonononnonreres 3 74 Flow Calibrations andinas 3 74 Calibrate Flow Zero cite seca seste hasene lane Oc aE GNR 3 75 Calibrate Mow Spas ie a ee ed ead he 3 75 Restore Default Flow Calibration a14 cise aceite atta 3 76 Input Board Testunai n pit eacteabhaeess 3 76 Temperature Calibrtion ad nia 3 77 Analog Output Calibration iii 3 77 Analog Output Calibrate Lo o 3 78 Analog Output Calibrate FulleSeale anciana 3 78 Analog Input Calibrations ii iS is 3 79 Analog Input Calibration Zero ici sedi tit 3 79 Analog Input Calibrate Full Scale viii iaicins 3 80 Permeation Oven Settings AA A Matias Cesescananty 3 80 Thermo Fisher Scientific Thermo Fisher Scientific Contents Calibrate Gas Thermistor usa utili aaiicitiasds 3 81 Water Barradas tects 3 81 A A A santa aR E plat e aia 3 81 Calibrate Oven Thermistor caidas 3 82 Permeation Oven Sepia iros 3 82 Factory Calibrate Gas Therion aan 3 82 Lowand High Potts io oer o dos 3 83 Set
206. o perform just a zero check set the span and purge duration screens to 0 off The same applies to perform just a span or purge check In the Main Menu choose Calibration gt Zero Span Check gt Zero Span or Purge Duration Min ERE ES CHA Pi Tam DH Cil BME EE umk Bone ae Ae Haat ake EE m owe AS EE EE The Zero Span Averaging Time screen allows the user to adjust the zero span averaging time The zero span averaging time is used by the analyzer only when performing an automatic zero or span check or calibration The analyzer s averaging time is used for all other functions The following averaging times are available 1 2 5 10 20 30 60 90 120 180 240 and 300 seconds In the Main Menu choose Calibration gt Zero Span Check gt Zero Span Avg Sec The Zero Span Ratio screen is used to adjust the ratio of zeros to spans For example if this value is set to 1 a span check will follow every zero check If this value is set to 3 there will be two zero checks between each zero span check This value may be set from 1 to 10 with 1 as default In the Main Menu choose Calibration gt Zero Span Check gt Zero Span Ratio Thermo Fisher Scientific Operation Ea E Py Tame Ol Cilia MARTE nd Lario Md Instrument The Instrument Controls menu contains a number of items The software Controls Menu controls listed in this menu enable control of the listed instrument functions In the Main Menu choose In
207. oard is locked via the front panel using Password gt Lock Instrument the instrument reports being in Remote mode In this mode the keypad is locked data can be viewed but not changed using the front panel interface and the remote Set commands are active If the instrument keyboard is unlocked via the front panel using Password gt Unlock Instrument the instrument reports being in Local mode the front panel interface is unlocked and data can be changed from the front panel Refer to the C Link Protocol Commands appendix for detailed information about mode allow mode and power up mode commands In the Main Menu choose Password gt Lock Instrument ER E EUs EEE TE oa Change Password The Change Password is used to set or change the password used to unlock the instrument s front panel The change password screen is shown if the instrument is unlocked In the Main Menu choose Password gt Change Password Thermo Fisher Scientific Model 43i Instruction Manual 3 87 Operation Password Remove Password The Remove Password screen is used to erase the current password and disable password protection The remove password screen is shown if the instrument is unlocked and the password set In the Main Menu choose Password gt Remove Password Unlock Instrument The Unlock Instrument screen is used to enter the password to unlock the front panel The unlock instrument is sho
208. oduced by measuring the pressure difference across a precision orifice This unit is used for measuring the flow of sample gas in the measurement system Pressure Sensor Assembly The pressure sensor assembly consists of a board containing an instrumentation amplifier and a pressure transducer with a gas input fitting The pressure transducer output is produced by measuring the Thermo Fisher Scientific Model 43i Instruction Manual 8 7 System Description Electronics Bench Heater PMT Power Supply Assembly Diagnostic LED Input Board Digital Output Board Front Panel Connector Board 8 8 Model 43i Instruction Manual pressure difference between the sample gas pressure and ambient air pressure The fluorescence chamber temperature is measured with a thermistor The voltage across the thermistor is fed to the main processor for use in calculating and displaying the reaction chamber temperature The voltage across the thermistor is also compared to a set point voltage and used to control that the reaction chamber heaters to maintain a constant temperature The PMT power supply produces high voltage to operate the photomultiplier tube used in the measurement system The output voltage is adjustable from approximately 600 to 1200 volts under software control The diagnostic LED is used to provide an alternate light source to the PMT to determine if the PMT is operating when the operating condition of the flash lamp is unknown
209. ok no of lrec no of srec These commands report the number of long and short records stored in the long and short records memory The example that follows shows that 50 long records have been stored in the memory Send no of lrec Receive no of lrec 50 recs malloc Irec malloc srec These commands report the currently set memory allocation for long and short records in percent of total memory Send malloc lrec Receive malloc lrec 10 set malloc lrec value set malloc srec value value 0 to 100 These commands set the percent of memory space allocated for long and short records to value where value is a floating point number representing percent The example that follows sets the memory allocation for long records to 10 Note Issuing these commands will clear all the logging data memory All the existing records should be retrieved using appropriate commands if required A Send set malloc lrec 10 Receive set malloc lrec 10 ok Model 43 Instruction Manual B 21 C Link Protocol Commands Datalogging B 22 Model 43 Instruction Manual data treatment lrec data treatment srec These commands report the current selection of data treatment for concentrations in the long records lrecs or short records srecs The example that follows reports the data treatment for concentrations in lrec is minimum Send data treatment lrec Receive data treatment lrec min set data treatment lrec string set data treatment srec s
210. oltage 1 Press J to display the Main Menu 2 From the Main Menu select Diagnostics gt Voltages gt Interface Board and press to display the Interface Board Voltages screen 3 Check the flash supply voltage If this voltage is at 1200 V it is necessary to either replace the lamp or adjust the lamp voltage control circuit For detailed information about this screen refer to the Operation chapter For more information about replacing the lamp or adjusting the lamp voltage control circuit see the Servicing chapter A normal flow rate is approximately 0 5 LPM If the flow rate is less than 0 35 LPM use the following procedure to perform a leak test 1 Block the SAMPLE bulkhead on the rear panel with a leak tight cap 2 Wait two minutes 3 Press to display the Main Menu 4 Press 4 to move the cursor to Diagnostics and press to display the Diagnostics menu 5 Press to move the cursor to Flow and press to display the Sample Flow screen The flow reading should indicate zero flow 6 Press to display the Diagnostics Menu then press Ct Jio move the cursor to Pressure and press to display the Pressure screen The pressure reading should be less than 180 mm Hg If not check to see that all fittings are tight and that none of the input lines are cracked or broken For detailed information about these screens refer to the Operation chapter Thermo Fisher Scientific Preventive M
211. om OK to either LOW or HIGH respectively If the alarm is not a level alarm the status will go from OK to FAIL The number of alarms detected is displayed to indicate how many alarms have occurred If no alarms are detected the number zero is displayed To see the actual reading of an item and its minimum and maximum limits move the cursor to the item and press Ce Thermo Fisher Scientific Operation Alarms Menu Items displayed are determined by the options displayed Zero Autocal or Span Autocal are visible only if Zero Cal Reset or Span Cal Reset are turned ON Otherwise Zero Check or Span Check are displayed The motherboard status interface board status and I O expansion board status if installed indicates that the power supplies are working and connections are successful There are no setting screens for these alarms In the Main Menu choose Alarms Internal The internal temperature screen displays the current internal temperature Temperature and sets the minimum and maximum alarm limits Acceptable alarm limits range from 8 to 47 C If the internal temperature reading goes beyond either the minimum or maximum alarm limit an alarm is activated The word ALARM appears in the Run screen and in the Main Menu In the Main Menu choose Alarms gt Internal Temp PESEE Lira rm PA Py AG Lt Thermo Fisher Scientific Model 43i Instruction Manual 3 63 Operation Alarms Menu
212. on action zero or span The analog input calibration is visible only when the instrument is in service mode For more information on the service mode see Service Mode earlier in the chapter Note This screen is present only when the I O expansion board is installed A Note This adjustment should only be performed by an instrument service technician A In the Main Menu choose Service gt Analog Input Calibration gt Input Channel 1 8 y Piro Analog Input Calibration Thermo Fisher Scientific Zero The Analog Input Calibrate Zero screen allows the user to calibrate the zero state of the selected analog input In the Main Menu choose Service gt Analog Input Calibration gt Select Channel gt Calibrate Zero Hook up a voltage source of 0 V to the analog input channel Model 43i Instruction Manual 3 79 Operation Service Menu Analog Input Calibrate Full Scale Permeation Oven Settings 3 80 Model 43i Instruction Manual DH ind A A The Analog Input Calibration Full Scale screen allows the user to calibrate the full scale state of the selected analog input In the Main Menu choose Service gt Analog Input Calibration gt Select Channel gt Calibrate Full Scale Hook up a voltage source of 10 V to the analog input channel The Permeation Oven Settings menu is used for setting up and calibrating the permeation oven option The permeation oven settings menu is visible only when th
213. onnector Label Reference Designator Pin Signal Description INTF DATA J1 1 Ground 2 RS485 to Interface Board 3 RS485 to Interface Board Ethernet Output 10 BASE T J2 Ethernet Output Ethernet Input NC NC Ethernet Input NC NC 5V 24V 24V 0 N OQO wo A WO N EXPANSION BD J3 Ground Ground Ground RS485 to Expansion Board RS485 to Expansion Board 5V 24V 24V 0 N O wo A OQ N SPARE DATA J4 Ground Ground Ground RS485 to Spare Board RS485 to Spare Board o N oon A WO N 1 0 J5 Power Fail Relay N C Contact 6 12 Model 43i Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Connector Label Reference Designator Pin oOo a N oon A O gt s gt gt gt gt gt gt gt gt co loo N o a gt au N O 0 0 0 wo U U N NNN N NNN N NH Oo om F WON O O O N O o A WO N Q Troubleshooting Connector Pin Descriptions Signal Description Ground TTL Input 1 TTL Input 2 Ground TTL Input 5 TTL Input 7 TTL Input 8 TTL Input 10 Ground TTL Input 13 TTL Input 15 Ground Analog Voltage Output 1 Analog Voltage Output 3 Ground Analog Voltage Output 5 Ground Ground Power Fail Relay COM Power Fail Relay N O Contact Ground TTL Input 3 TTL Input 4 TTL Input 6 Ground TTL Input 9 TTL Input 11 TTL Input 12 TTL Input 14 TTL Input 16 Ground Analog Voltage Output 2 Analog Voltage Ou
214. ormat Reports sets output format for long records ASCII or binary B 19 Irec layout Reports current layout of Irec data B 20 Irec mem size Reports maximum number of long records that can be stored B 20 Irec per Reports sets long record logging period B 20 malloc Irec Reports sets memory allocation for long records B 21 malloc srec Reports sets memory allocation for short records B 21 me menu Simulates pressing menu pushbutton B 26 mode Reports operating mode in local service or remote B 38 no of Irec Reports sets number of long records stored in memory B 21 no of srec Reports sets number of short records stored in memory B 21 perm gas temp Reports current perm gas temperature B 9 pmt status Reports sets PMT on or off B 33 pmt voltage Reports current PMT voltage B 10 pres Reports current reaction chamber pressure B 10 pres cal Reports sets pressure used for calibration B 26 pres comp Reports sets pressure compensation on or off B 31 program no Reports analyzer program number B 38 push Simulates pressing a key on the front panel B 26 range Reports sets current SO range B 28 Thermo Fisher Scientific C Link Protocol Commands Commands List Thermo Fisher Scientific Command Description Page range mode Reports sets current range mode B 29 react temp Reports current reaction chamber temperature B 10 relay status Reports sets relay logic status to for the designat
215. pecified by a constant in the request data field A value of OxFFOO requests the action to be ON A value of 0x0000 requests it to be OFF All other values are illegal and will not affect the output The normal response is an echo of the request returned after the state has been written Request Function code 1 Byte 0x05 Starting Address 2 Bytes 0x0000 to maximum allowed by instrument Output Value 2 Bytes 0x0000 or OxFFOO Response Function code 1 Byte 0x05 Starting Address 2 Bytes 0x0000 to maximum allowed by instrument Output Value 2 Bytes 0x0000 or OxFFOO Model 43 Instruction Manual C 7 MODBUS Protocol MODBUS Commands Supported MODBUS Commands Supported C 8 Model 43 Instruction Manual Error Response Function code Exception code 1 Byte 1 Byte 03 Illegal Data 04 Slave Device Failure Here is an example of a request to write Coil 5 ON Request Field Name Function Dutput Address Hi Output Address Lo Output Value Hi Output Value Lo Response Field Name Function Output Address Hi Output Address Lo Output Value Hi Output Value Lo Tables C1 through C3 list the MODBUS commands supported for the Model 432 Hex 05 00 05 FF 00 Hex 05 00 05 FF 00 Table C 1 Read Registers for 43 Register Number Variable 40001840002 S02 CONC 10003840004 NOT USED 40005840006 NOT USED 40007840008 LO S02 CONC Dual Auto Range mode 40009840010 NOT USED 40011840012 NO
216. permeation oven and valves into the instrument and connect the cables and plumbing 2 From the Perm Oven Settings menu in the Service menu select Perm Oven Selection then select 45 C 3 Remove the thermistor from POJ1 on the measurement interface board 4 Connect a resistance of approximately 3 8 KQ across pins 1 and 2 of POJ1 Model 43i Instruction Manual 9 3 Optional Equipment Internal Permeation Span Source 9 4 Model 43i Instruction Manual 10 11 12 13 14 15 16 17 18 Go to the Factory Cal Gas Therm menu from the Perm Oven Settings menu in the Service menu Select Low Point enter the exact value of the resistor and press to calibrate the low resistance point Press Cm to return to the Factory Cal Gas Therm menu Connect a resistance of approximately 5 KQ across pins 1 and 2 of POJ1 Go to the High Point screen enter the exact value of the resistor and press to calibrate the high resistance point Press Cm to return to the Factory Cal Gas Therm menu Remove resistor from POJ1 and re attach the gas thermistor Connect a resistance of approximately 3 8 KQ across pins 3 and 4 of POJ3 Go to the Factory Cal Oven Therm menu from the Perm Oven Settings menu in the Service menu Select Low Point enter the exact value of the resistor and press to calibrate the low resistance point Press Cu to return to the Factory Cal Oven Therm menu Connect a resistance of appro
217. ping the bottom in the rear Equipment Damage Do not attempt to lift the instrument by the cover or other external fittings A The Model 431 is shipped complete in one container If there is obvious damage to the shipping container when you receive the instrument notify the carrier immediately and hold for inspection The carrier is responsible for any damage incurred during shipment Use the following procedure to unpack and inspect the instrument 1 Remove the instrument from the shipping container and set it on a table or bench that allows easy access to both the front and rear Model 43i Instruction Manual 2 1 Installation Unpacking and Inspection 2 Remove the cover to expose the internal components 3 Remove the packing material Figure 2 1 Remove Packing 2 pieces Remove Packing 2 pieces Units without Optional I O Board Units with Optional 1 0 Board Figure 2 1 Remove the Packing Material 4 Remove the three shipping screws from the pump Figure 2 2 Shipping Screws 3 Figure 2 2 Removing the Shipping Screws 5 Check for possible damage during shipment 2 2 Model 43i Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Setup Installation Setup 6 Check that all connectors and circuit boards are firmly attached 7 Re install the cover 8 Remove any protective plastic material from the case exterior Use the following proc
218. ption does not precisely control dilution gas flow it should not be used as a basis for instrument zero and span adjustments calibration updates or adjustment of ambient data Whenever there is an indication of possible instrument drift or malfunction a full zero and multipoint calibration Level 1 should be performed prior to corrective action For further information on zero span and calibration of air pollution monitors refer to Section 2 0 9 of the US EPA s Quality Assurance Handbook for Air Pollution Measurement Systems Volume ID Figure 9 1 shows how this option is integrated with the Model 431 components Energizing the sample valve V1 shuts off the sample flow and permits the flow of zero air for analysis When valves V1 and V2 are energized the flow of zero air mixes with air containing SO from the permeation oven This mode of operation provides a single point span check Thermo Fisher Scientific Model 43i Instruction Manual 9 1 Optional Equipment Internal Permeation Span Source Permeation Tube Installation 9 2 Model 43i Instruction Manual A Figure 9 1 Internal Permeation Span Source Flow Diagram Use the following procedure to install the permeation tube Remove the oven cover Remove the glass chamber assembly by loosening the white plastic retaining collar loosening not removing the knurled screw and gently pulling the assembly upward Completely remove the oven Separate the glass chamber
219. put board See the Digital Output Board Replacement procedure in this chapter 4 Unplug all connectors from the motherboard Note connector locations to facilitate reconnection 5 Using the nut driver remove the standoffs securing the board to the rear panel Figure 7 19 6 Pop off the motherboard from motherboard support bracket and remove the motherboard 7 To install the motherboard follow previous steps in reverse 8 Calibrate the analog voltage outputs Refer to Analog Output Calibration in this chapter Use the following procedure to replace the measurement interface board Figure 7 22 Equipment Required Measurement interface board Philips screwdriver Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Refer to Removing the Measurement Bench and Lowering the Partition Panel in this chapter to lower the partition panel then proceed to the next step below 2 Unplug all connectors Note the locations of the connectors to facilitate reconnection Model 431 Instruction Manual 7 43 Servicing Front Panel Board Replacement 3 Unscrew th
220. r the range of 0 to 10 volts The instrument includes one power fail relay on motherboard and ten digital output relays on the digital output board These are reed relays rated for at least 500 mA 200VDC The power fail relay is Form C both normally opened and normally closed contacts All other relays are Form A normally opened contacts and are used to provide alarm status and mode information from the analyzer as well as remote control to other devices such as for controlling valves during calibration The user may select what information is sent out each relay and whether the active state is opened or closed Sixteen digital inputs are available which may be programmed to signal instrument modes and special conditions including e Zero Gas Mode e Span Gas Mode Thermo Fisher Scientific System Description 1 0 Components e Set Background e Cal to low span e Cal to high span e Analog outputs to zero e Analog outputs to full scale The actual use of these inputs will vary based on analyzer configuration The digital inputs are TTL level compatible and are pulled up within the analyzer The active state can be user defined in firmware Serial Ports Two serial ports allow daisy chaining so that multiple analyzers may be linked using one PC serial port The standard bi directional serial interface can be configured for either RS 232 or RS 485 The serial baud rate is user selectable in firmware for standard speeds fro
221. r this parameter has been changed for the change to take effect A In the Main Menu choose Instrument Controls gt Communication Settings gt TCP IP Settings Thermo Fisher Scientific Model 43i Instruction Manual 3 37 Operation Instrument Controls Menu CARTED iat MOTOR gp gypi ENTRE I oak bonoe PET Sal be d EE 3 PL dd Use DHCP The Use DHCP screen is used to specify whether to use DHCP or not When DHCP is enabled the network dynamically provides an IP address for the instrument The instrument s power must be cycled for a change to this parameter to take affect In the Main Menu choose Instrument Controls gt Communication Settings gt TCP IP Settings gt Use DCHP Champ fier T ES AAA A i E Cara Tce cn ier E ir PA ar fae a Phe a l a R i IP Address The IP Address screen is used to edit the IP address The IP address can only be changed when DHCP is OFF For more information on DHCP see Use DHCP above In the Main Menu choose Instrument Controls gt Communication Settings gt TCP IP Settings gt IP Address P T p DH pirri at i PERE PUMP 3 38 Model 43i Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu Netmask The Netmask screen is used to edit the netmask The netmask is used to determine the subnet for which the instrument can directly communicate to other devices on The netmask can only be changed when DHCP is OFF For more information on
222. racket Input Board Input Board Screws 4 Input Box Screws 2 Power Supply Bracket Screws 2 Input Box Figure 7 19 Replacing the Input Board 3 Loosen the two screws securing the input board assembly to the bracket slide the assembly up and remove from bracket 4 Install the input board assembly by following the previous steps in reverse Thermo Fisher Scientific Model 43i Instruction Manual 7 39 Servicing 1 0 Expansion Board Optional Replacement I O Expansion Board Optional Replacement 7 40 Model 43i Instruction Manual Use the following procedure to replace the optional I O expansion board Figure 7 20 Note After replacing the optional I O expansion board calibrate the current outputs and the analog voltage inputs See the Analog Output Calibration procedure and the Analog Input Calibration procedure in this chapter A Equipment Required I O expansion board Nut driver 3 16 inch Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Turn instrument OFF unplug the power cord and remove the cover 2 Unplug the I
223. rameters on page C 2 e TCP Communication Parameters on page C 2 e Application Data Unit Definition on page C 2 e Function Codes on page C 3 e MODBUS Commands Supported on page C 8 Additional information on the MODBUS protocol can be obtained at http www modbus org References are from MODBUS Application Protocol Specification V1 1a MODBUS IDA June 4 2004 The following are the communication parameters that are used to configure the serial port of the Series to support MODBUS RTU protocol Number of Data bits 8 Number of Stop bits 1 Parity None Data rate from 1200 115200 Baud 9600 is default Model 43 Instruction Manual C 1 MODBUS Protocol TCP Communication Parameters iSeries Instruments support the MODBUS TCP protocol The register definition is the same as for the serial interface Pa rameters TCP connection port for MODBUS 502 TCP Communication Application Data Here are the MODBUS ADU Application Data Unit formats over serial i age d TCP IP Unit Definition Serial Slave Address Function Code Data Error Check TCP IP MBAP Header Function Code Data Slave Address The MODBUS slave address is a single byte in length This is the same as the instrument ID used for C Link commands and can be between 1 and 127 decimal i e 0x01 hex to 0x7F hex This address is only used for MODBUS RTU over serial connections Note Device ID 0 used
224. ranty rights in such Product that Seller may have from the original manufacturer or third party supplier to the extent such assignment is allowed by such original manufacturer or third party supplier In no event shall Seller have any obligation to make repairs replacements or corrections required in whole or in part as the result of i normal wear and tear ii accident disaster or event of force majeure iii misuse fault or negligence of or by Buyer iv use of the Products in a manner for which Model 437 Instruction Manual A 1 Warranty Mounting Options they were not designed v causes external to the Products such as but not limited to power failure or electrical power surges vi improper storage and handling of the Products or vii use of the Products in combination with equipment or software not supplied by Seller If Seller determines that Products for which Buyer has requested warranty services are not covered by the warranty hereunder Buyer shall pay or reimburse Seller for all costs of investigating and responding to such request at Seller s then prevailing time and materials rates If Seller provides repair services or replacement parts that are not covered by the warranty provided in this warranty Buyer shall pay Seller therefor at Seller s then prevailing time and materials rates ANY INSTALLATION MAINTENANCE REPAIR SERVICE RELOCATION OR ALTERATION TO OR OF OR OTHER TAMPERING WITH THE PRODUCTS PERFORMED BY
225. rd compatibility on the C series Screen information is reported using the Screen command above Thermo Fisher Scientific Model 43 Instruction Manual B 27 C Link Protocol Commands Measurement Configuration B 28 Measurement Configuration Model 43 Instruction Manual Send screen Receive screen This is an I series Instrument Screen Information not Available range range so2 high range low range These commands report SO range in single range mode or the high and low ranges in dual or auto range mode If the mode is incorrect the instrument responds with can t wrong settings The example that follows reports that the SO full scale range to 50 ppm according to Table B 6 and Table B 7 Send range Receive range 6 5000E 2 ppm set range Selection set high range Selection set low range Selection These commands select the SO full scale ranges according to Table B 6 and Table B 7 The example that follows sets the SO full scale range to 100 ppm Send set range 3 Receive set range 3 ok Table B 6 Standard Ranges Selection ppm mg m 0 0 05 0 2 1 0 1 0 5 2 0 2 1 3 0 5 2 4 1 5 5 2 10 6 5 20 7 10 25 8 C1 C1 9 C2 C2 Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Measurement Configuration Selection ppm mg m 10 C3 C3 Table B 7 Extended Ranges Selection ppm mg m 0 0
226. re 7 8 Install the new flash intensity assembly by following this procedure in reverse Thermo Fisher Scientific Photomultiplier Tube Replacement Thermo Fisher Scientific A Servicing Photomultiplier Tube Replacement Retaining Screws 3 Flash Intensity Assembly Figure 7 8 Replacing the Flash Intensity Assembly Use the following procedure to replace the photomultiplier tube Equipment Required Photomultiplier tube Philips screwdriver Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Turn analyzer OFF unplug the power cord and remove the cover 2 Disconnect the high voltage cable from the PMT power supply cable connector and unplug the BNC cable from the input board connector 3 Remove the three retaining screws holding the PMT cover to the PMT housing and pull back the cover to access the two PMT base retaining screws 4 Pull PMT and PMT base from the PMT housing by twisting it slightly back and forth 5 To install a new PMT follow previous steps in reverse Model 43i Instruction Manual 7 17 Servicing PMT High Voltage Power Supply Replaceme
227. re frequently depending on local conditions A calibration gas system capable of providing accurate levels of SO calibration gas between zero and 80 of the full scale range is required The calibration system must provide a flow rate of at least 0 5 LPM for an instrument with the standard flow instruments with higher flow rates will require a higher minimum calibration system flow rate All calibration gas should be derived from local or working standards such as cylinders of compressed gas or permeation devices that are certified as traceable to an NIST primary standard A cylinder gas dilution system shown in Figure 4 1 can be constructed All connections between components in the system should be made with glass Teflon or other non reactive material The air flow controller should be capable of maintaining a constant air flow within 2 of the required flow rate The SO flow controller should be capable of maintaining constant SO flows within 2 of the required flow rate Ensure both flow controllers are properly calibrated The pressure regulator for the standard SO cylinder must have a non reactive diaphragm and internal parts and a suitable delivery pressure Thermo Fisher Scientific Commercial Precision Dilution Systems Permeation Tube Thermo Fisher Scientific System Calibration Permeation Tube System MIXING CHAMBER PRESSURE REGULATOR OUTPUT MANIFOLD EXTRA OUTLETS C
228. reen For more information about selecting ranges see SO Range described previously In the Main Menu choose Range gt Set Custom Ranges gt Custom range 1 2 or 3 Thermo Fisher Scientific Model 43i Instruction Manual 3 15 Operation Averaging Time Averaging Time Calibration Factors Menu 3 16 Model 43 Instruction Manual The Averaging Time defines a time period 1 to 300 seconds during which SO measurements are taken The average concentration of the readings are calculated for that time period The front panel display and analog outputs are updated every 10 seconds for averaging times between 10 and 300 seconds For averaging times of 1 2 and 5 seconds the front panel display and analog outputs are updated every second An averaging time of 10 seconds for example means that the average concentration of the last 10 seconds will be output at each update An averaging time of 300 seconds means that the moving average concentration of the last 300 seconds will be output at each update Therefore the lower the averaging time the faster the front panel display and analog outputs respond to concentration changes Longer averaging times are typically used to smooth output data The Averaging Time screen for the single range mode is shown below In the dual and auto range modes an averaging time menu is displayed before the averaging time screen This additional menu is needed because the dual and autorange modes have
229. relay open set relay open value set relay closed set relay closed value These commands set the relay logic to normally open or closed for relay number value where value is the relay between 1 and 16 The example that follows sets the relay no 1 logic to normally open Note If the command is sent without an appended relay number then all the relays are assigned the set logic of normally open closed A Send set relay open 1 Receive set relay open 1 ok The Erec Lrec Srec layouts contain the following e A format specifier for parsing ASCII responses e A format specifier for parsing binary responses In addition to these the Erec Layout contains e A format specifier for producing the front panel displays In operation values are read in using either the ASCII or binary format specifiers and converted to uniform internal representations 32 bit floats or 32 bit integers These values are converted into text for display on the screen using the format specifier for the front panel display Normally the specifier used to parse a particular datum from the input stream will be strongly related to the specifier used to display it e g all of the floating point inputs will be displayed with an f output specifier and all of the integer inputs will be displayed with a d specifier The first line of the Layout response is the scanf like parameter list for parsing the fields from an ASCII ERec response Parameters are separated by spa
230. rnnnnnncnrnnoonorornnnonororonoooos 5 8 Figure 6 1 Board Level Connection Diagram Common Electronics 6 10 Figure 6 2 Board Level Connection Diagram Measurement System 6 11 Figure 7 1 Properly Grounded Antistatic Wrist Strap o oonocoiciccnic 7 2 Figure 7 2 Model 43i Component Lay0Ut oooocicoocincconnconinccnnnccnonccnrornconrnononos 7 6 Figure 7 3 Removing the Measurement Bench and Lowering the Partition E Raita me A A A A 7 7 Figure 7 4 Replacing the PUMP cececesessesesesescetesesestsesteesesesteseseseees 7 10 Figure 7 5 Replacing the Fan ccc ceceescscesesesesceeseseseseeceeseseeceseseseaes 7 11 Figure 7 6 Replacing the Optical Bench ooooocicincincncincnnincnninccnrnonnnnoos 7 12 Figure 7 7 Replacing the Flash Lamp and Flash Trigger Assembly 7 14 Figure 7 8 Replacing the Flash Intensity Assembly 1 17 Figure 7 9 Replacing the PMT ou ccc cece cececesesescetesesesteteseesesesteeseseees 7 18 Figure 7 10 Replacing the PMT High Voltage Power Supply HVPS 7 19 Figure 7 11 Replacing the DC Power Supply ocooninoonininninocoioooooo 7 21 Figure 7 12 Rear Panel Analog Input and Output Pins 7 23 Model 437 Instruction Manual xxi Figures xxii Model 43 Instruction Manual Figure 7 13 Replacing the Pressure Transducer Assembly
231. rted by the analyzer is approximately 0 5 LPM Actual sample flow should be 0 8 LPM or more with the excess flowing out the atmospheric bypass Sample the calibration gas and wait for the instrument reading to stabilize Use _ to move the cursor to Calibrate HI SO2 and press to display the Calibrate HI SO screen The first line of the display shows the current SO reading The second line of the display is where you enter the high calibration gas concentration To enter the high calibration gas concentration use to move the cursor left and right Use J to increment and decrement the digit Press 4_ to calibrate the instrument to the high SO2 calibration gas The display flashes the message SAVING and the corrected SO reading is displayed Press Cm to return to the Calibration menu If using a chart recorder the SO recorder response will equal SO lour x100 URL Z soz Where URL upper range limit of the instrument s operating range Zso2 instrument response to zero air full scale Generate five SO concentrations equally spaced between zero and the concentration above Model 43i Instruction Manual 4 9 Calibration Zero Span Check Zero Span Check 4 10 Model 43i Instruction Manual 14 15 16 17 18 19 20 2t 22 Record instrument reading for each concentration after allowing time for both gas generation system and instrumen
232. rtition panel can be lowered to improve access to connectors and components Refer to the following steps when a procedure requires lowering the partition panel Figure 7 3 Thermo Fisher Scientific Servicing Removing the Measurement Bench and Lowering the Partition Panel Pass Through REMOVE SCREWS crore AAN SLIDE BENCH OUT Partition Panel Loosen Captive Screw Na gt Remove Screws Remove Screw LOWER PANEL _ ESA ES Remove Hidden Screw Figure 7 3 Removing the Measurement Bench and Lowering the Partition Panel Equipment Required Philips screwdriver Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Turn instrument OFF and unplug the power cord Thermo Fisher Scientific Model 43i Instruction Manual 7 7 Servicing Fuse Replacement Fuse Replacement 7 8 Model 43i Instruction Manual 10 11 If the instrument is mounted in a rack remove it from the rack Remove the cover Disconnect the plumbing connections at the rear of the measurement bench Disconnect the connectors that pass through the center of the partition
233. ruction Manual 3 45 Operation Instrument Controls Menu E Pri Select Range The Select Range screen is used to select the hardware range for the selected analog output channel Possible ranges for the voltage outputs are 0 100 mV 0 1 0 5 and 0 10 V Possible ranges for the current outputs are 0 20 mA and 4 20 mA e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Analog Output Config gt select Channel gt Select Range Minimum and Maximum Value The Minimum Value screen is used to edit the zero 0 to full scale 100 value in percentages for the selected analog output channel See Table 3 7 for a list of choices In dual or auto range mode HI or LO is displayed to indicate high or low concentrations Range Status is visible only in auto range mode The minimum and maximum output value screens function the same way The example that follows shows the set minimum value screen In the Main Menu choose Instrument Controls gt IO Configuration gt Analog Output Config gt Select Channel gt Set Minimum or Maximum Value 3 46 Model 43i Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu Table 3 7 Analog Output Zero to Full Scale Values Output Zero Value Full Scale 100 Value S02 Zero 0 Range Setting LO S02 Zero 0 Range Setting HI S02 Zero 0 Range Setting Range S02 Changing the setting for this output is not recommend
234. s J to move the cursor to Cal SO Coefficient and press to display the Calibrate SO screen The first line of the display shows the current SO reading The SPAN CONC line of the display is where you enter the calibration gas concentration Thermo Fisher Scientific Multipoint Calibration in Dual Auto Range Thermo Fisher Scientific Mode Calibration Multipoint Calibration in Dual Auto Range Mode 11 To enter the calibration gas concentration use to move the cursor left and right Use to increment and decrement the digit Press to calibrate the instrument to the SO2 calibration gas The display flashes the message SAVING and the corrected SO2 reading is displayed 12 Press the _ pushbutton to return to the Run screen If using a chart recorder hooked up to an analog output channel the SO2 recorder response will equal SO Where URL upper range limit of the analyzer s operating range Zso analyzer response to zero air fullscale Note The minimum and maximum analog output configuration values should be set to the default settings minimum 0 maximum 100 a 13 For multi point calibration generate five SO2 concentrations equally spaced between zero and the concentration above Record instrument reading for each concentration after allowing time for both gas generation system and instrument to stabilize Plot a graph of instrument readings against the SO2 concentrations gen
235. s Concentration Title Bar Gas Type Gas Units Status Bar Status Bar Icons Programmable Soft Keys The Main Menu contains a number of submenus Instrument parameters and features are divided into these submenus according to their function The concentration appears above the main menu and submenus in every screen The SERVICE menu is visible only when the instrument is in service mode For more information on the service mode see Service Mode later in this chapter e Use t and to move the cursor up and down e Press to select a choice e Press to return to the top of the Main Menu or gt to return to the Run screen Model 43i Instruction Manual 3 7 Operation Range Menu Range Menu Single Range Mode 3 8 Model 43i Instruction Manual The Range menu allows the operator to select the gas units SO ranges and to set the custom ranges The screens that follow show the range menu in single mode and dual autorange modes The only difference between the screens are the words HI and LO to indicate which range is displayed For more information about the single dual and auto range modes see Single Range Mode Dual Range Mode and Autorange Mode In the Main Menu choose Range if EH Cush i fed coos OS EDH pirri MAMA In the single range mode there is one range one averaging time and one span coefficient By default two SO analog o
236. s The Serial Settings menu is used for serial communications control and configuration In the Main Menu choose Instrument Controls gt Communication Settings gt Serial Settings a Cie hi Baud Rate The Baud Rate screen is used to set the RS 232 RS485 interface baud rate Baud rates of 1200 2400 4800 9600 19200 38400 57600 and 115200 are available The analyzer s default baud rate is set to 9600 to provide backwards compatibility with the older C series analyzers Data Bits The Data Bits Screen is used to set the number of serial data bits Selections of 7 or 8 are available defaults to 8 e PT ee A Pei A A EE EE 3 32 Model 43i Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu Parity The Parity screen is used to select the parity bit for the serial port Selections of NONE EVEN or ODD are available defaults to NONE ty ERED CH EN EEE TE hos ETET Fad Stop Bits The Stop Bits screen is used to select the number of stop bits for the serial port Selections of 1 and 2 are available defaults to 1 ty e Erap prup Lira PY ps i pued PA a in bab i Saal cows EE fee PEE AEE Instrument ID The Instrument ID screen allows the operator to edit the instrument ID The ID is used to identify the instrument when using the C Link or MODBUS protocols to control the instrument or collect data It may be necessary to edit the ID number i
237. s bar displays the time 24 hour format the password lock icon service wrench icon alarm bell icon and optional zero span sample solenoid valve status if installed Status Bar Icons The password lock icon indicates that no parameter changes can be made from gt B A lt The alarm bell icon indicates that an alarm is active the front panel 3 6 Model 43i Instruction Manual The service wrench icon indicates that the instrument is in the service mode The word SAMPLE on the bottom left of the display indicates the analyzer has the span zero valve option and is in SAMPLE mode Other modes appear in the same area of the display as ZERO or SPAN For more information about the optional solenoid valves see Chapter 9 Optional Equipment When operating in dual or auto range mode two sets of coefficients are used to calculate the CO High and Low concentrations Also two averaging times are used one for each range The title bar indicates which range concentrations are displayed The words LOW RANGE CONCENTRATION on the top of the display indicates that the low concentration is displayed In dual range mode pressing the Ce Jand arrows will toggle between high and low concentrations The example that follows shows the Run screen in single range mode Thermo Fisher Scientific Thermo Fisher Scientific Main Menu Operation Software Overview Ga
238. s on the motherboard In the Main Menu choose Diagnostics gt Voltages gt Motherboard Voltages EHO Calor Interface Board Voltages The Interface Board screen read only is used to display the current voltage readings on the interface board In the Main Menu choose Diagnostics gt Voltages gt Interface Board Voltages 3 56 Model 43i Instruction Manual Thermo Fisher Scientific Operation Diagnostics Menu ERED CH BEBE E usek Besse Ler et 1 0 Board Voltages The I O Board screen read only is used to display the current voltage readings on the I O expansion board This menu is only accessible if the I O expansion board is installed In the Main Menu choose Diagnostics gt Voltages gt I O Board Voltages Temperatures The Temperatures screen read only displays the current internal instrument temperature and chamber temperature The internal temperature is the air temperature measured by a sensor located on the interface board In the Main Menu choose Diagnostics gt Temperatures aa Ph dm me OEE OE Bed PPP LES Pilla fm PLP PLE Pressure The Pressure screen read only displays the current optical bench pressure The pressure is measured by a pressure transducer in line with the optical bench Thermo Fisher Scientific Model 43i Instruction Manual 3 57 Operation Diagnostics Menu In the Main Menu choose Diagnostics gt Pressure Sample Flow The Sample Flow scree
239. s reported in response to DA command Single Range Mode Dual Auto Range Mode D 6 Model 43 Instruction Manual Example Geysitech Protocol with transmission of three concentrations Instrument ID is 1 Operation Status is 03 Error Status is 04 Data Query String lt STX gt DA lt CR gt Reply String lt STX gt MDO3 lt SP gt 001 lt SP gt 2578 01 lt SP gt 03 lt SP gt 04 lt SP gt 0000000000 lt SP gt 002 lt SP gt 1 T i Address First Concentration E format 25 78 Address 1 5681 00 lt SP gt 03 lt SP gt 04 lt SP gt 0000000000 lt SP gt 003 lt SP gt 1175 01 lt SP gt 03 lt SP gt 04 lt SP T 1 T Second Concentration 5 681 Address 2 Third Concentration 11 75 0000000000 lt SP gt lt CR gt The attached concentrations are in the selected gas units The measurements that are attached to the response if not valid in a particular mode then a value of 0 0 will be reported The following measurements reported in response to DA command are for the Model 43i The 1 measurement reported in single range mode includes e SO The 2 measurements reported in dual or auto range modes include e low SO Thermo Fisher Scientific Geysitech Bayern Hessen Protocol Geysitech Commands Operating and Error Status See Table D 1 for operating status and Table D 2 for error status for the Model 437 Table D 1 Operating Status for Model 437 D7 D6 D5 D4 D3 D2 D1 DO
240. s sample lines and particulate filters which are used in normal monitoring should also be used during calibration Finally it is recommended that the recording devices and outputs used during normal monitoring be calibrated prior to the instrument calibration and that they be used during the calibration or the zero span check An SO gt free lt 0 0005 ppm air supply is required for the proper calibration and checkout of the instrument There are several methods that are acceptable to generate this zero gas Commercial heatless air dryers filled with a mixed bed of activated charcoal and a 13X molecular sieve have been found effective in removing SO from compressed air The use of this type of zero gas system is recommended when minimum maintenance is of prime importance This system requires a source of compressed air Refer to the manufacturer s recommendations for installation of such a system Model 43i Instruction Manual 4 1 Calibration Calibration Gas Generation Absorbing Column Calibration Gas Generation Cylinder Gas Dilution 4 2 Model 43i Instruction Manual An absorbing column packed with activated charcoal is acceptable for scrubbing SO from ambient air Ambient air is forced through a laboratory gas absorption column packed with the charcoal and the SO is removed to acceptable levels lt 0 0005 ppm The charcoal should be changed at a minimum of every six months It may be necessary to change the charcoal mo
241. s the user may program these outputs for either active closed or open amp Request Function code 1 Byte 0x01 or 0x02 Starting Address 2 Bytes 0x0000 to maximum allowed by instrument Quantity of outputs 2 Bytes 1 to maximum allowed by instrument Unit Identifier 1 Byte 0x00 to OxFF Passed back in response Response Function code 1 Byte 0x01 or 0x02 Byte count 1 Byte N Output Status n Byte N NorN 1 N Quantity of Outputs 8 if the remainder not equal to zero then N N 1 Error Response Function code 1 Byte 0x01 or 0x02 Exception code 1 Byte 01 IIlegal Function 02 Illegal Address 03 Illegal Data 04 Slave Device Failure Here is an example of a request and response to read outputs 2 15 Request Field Name Hex Function 0x01 Starting Address Hi 0x00 Starting Address Lo 0x02 Quantity of Outputs Hi 0x00 Quantity of Outputs Lo Ox0D Response Field Name Hex Function 0x01 Thermo Fisher Scientific 0x03 0x04 Read Holding Registers Read Input Registers 0x03 0x04 Read Holding Registers Read Input Registers Thermo Fisher Scientific MODBUS Protocol Function Codes Byte Count 0x03 Output Status 2 10 OxCD Output Status 11 15 Ox0A The status of outputs 2 10 is shown as the byte value 0xCD or binary 1100 1101 Output 10 is the MSB of this byte and output 2 is the LSB By convention bits within a byte are shown with the MSB to the left and the LSB to the right Thus the outputs in the first byte are
242. service mode see Service Mode earlier in this chapter Note This adjustment should only be performed by an instrument service technician A In the Main Menu choose Service gt PMT Voltage Adjustment Thermo Fisher Scientific Model 43i Instruction Manual 3 71 Operation Service Menu Range Mode Select The Range Mode Select screen is used to switch between the various range modes single dual and autorange In the Main Menu choose Service gt Range Mode Select Pressure Calibration The Pressure Calibration menu is used to calibrate the pressure sensor to zero span or restore factory default values The pressure calibration is visible only when the instrument is in service mode For more information on the service mode see Service Mode earlier in this chapter The pressure sensor s zero counts and span slope are displayed on the menu Note This adjustment should only be performed by an instrument service technician A Note Wait at least 30 seconds for the reading to stabilize before saving the value A In the Main Menu choose Service gt Pressure Calibration 3 72 Model 43i Instruction Manual Thermo Fisher Scientific Calibrate Pressure Zero Calibrate Pressure Span Thermo Fisher Scientific Operation Service Menu The Calibrate Pressure Zero screen calibrates the pressure sensor at zero pressure Note A vacuum pump must be connected to the pressure sensor before pump P performing
243. service mode described in the Operation chapter includes P P parameters and functions that are useful when making adjustments or diagnosing problems For additional service assistance see Service Locations at the end of this chapter Read the safety precautions before beginning any procedures in this chapter WARNING The service procedures in this manual are restricted to qualified representatives 4 If the equipment is operated in a manner not specified by the manufacturer the protection provided by the equipment may be impaired A CAUTION Carefully observe the instructions in each procedure A Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A Model 43i Instruction Manual 7 1 Servicing Firmware Updates Firmware Updates Accessing the Service Mode 7 2 Model 43i Instruction Manual DO NOT point the photomultiplier tube at a light source This can permanently damage the tube A DO NOT attempt to clean the mirrors in the optical bench These mirrors do not come in contact with the sample gas and should not be cleaned Cleaning the mirrors can damage the mirrors
244. set to index number 3 which is for the variable SO concentration Send sp field 1 Receive sp field 1 1 so2 set sp field number value number 1 32 is the maximum number of fields in long and short record lists number 1 18 is for streaming data lists This command sets the scratch pad field number item number in scratch pad list to value where value is the index number of a variable in the analog out variable list Available variables and their corresponding index numbers may be obtained using the command list var aout The set sp field command is used to create a list of variables which can then be transferred into the long record short record or streaming data lists using the set copy sp to lrec set copy sp to srec or set copy sp to stream commands Send set sp field 1 34 Receive set sp field 1 34 ok stream per This command reports the currently set time interval in seconds for streaming data Send stream per Receive stream per 10 set stream per number value number value 1 2 5 10 20 30 60 90 120 180 240 300 This command sets the time interval between two consecutive streaming data strings to number value in seconds The example that follows sets the number value to 10 seconds Send set stream per 10 Receive set stream per 10 ok stream time This command reports if the streaming data string will have a time stamp attached to it or not according to Ta
245. should be performed on the instrument to ensure proper operation Since usage and environmental conditions vary greatly you should inspect the components frequently until an appropriate maintenance schedule is determined This chapter includes the following maintenance information and replacement procedures e Safety Precautions on page 5 1 e Replacement Parts on page 5 2 e Cleaning the Outside Case on page 5 2 e Visual Inspection and Cleaning on page 5 2 e Cleaning the Mirrors on page 5 2 e Fan Filter Inspection and Cleaning on page 5 5 e Sample Particulate Filter Inspection on page 5 5 e Lamp Voltage Check on page 5 6 e Leak Test on page on page 5 6 e Pump Rebuilding on page 5 7 Read the safety precautions before beginning any procedures in this chapter WARNING If the equipment is operated in a manner not specified by the manufacturer the protection provided by the equipment may be impaired A Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A Model 43i Instruction Manual 5 1 Preventive Maintenance Replacement Parts R
246. span coefficients so that each range can be calibrated separately This is necessary if the two ranges are not close to one another For example the low SO range is set to 0 50 ppb and the high SO range is set to 0 10000 ppb By default in the dual range mode the analog outputs are arranged on the rear panel connector as shown in Figure 3 5 See Table 3 3 for channels and pin connections Dual range mode may be selected from the Range Mode Select in the Service Menu later in this chapter Thermo Fisher Scientific Model 43i Instruction Manual 3 9 Operation Range Menu 000000000000000000 0000000000000000000 e Figure 3 5 Pinout of Rear Panel Connector in Dual Range Mode Table 3 3 Default Analog Outputs in Dual Range Mode Channel Connector Pin 1 0 Terminal Pin Description 1 14 1 SO High Range 2 33 3 SO Low Range 3 5 5 None 4 34 7 None 5 7 9 None 6 36 11 None Ground 16 18 19 35 37 2 4 6 8 10 12 Signal Ground Note All channels are user definable If any customization has been made to the analog output configuration the default selections my not apply A Autorange Mode The autorange mode switches the SO analog outputs between high and 3 10 Model 43i Instruction Manual low ranges depending on the concentration level The high and low ranges are defined in the Range menu For example suppose the low range is set to 50 ppb and the high range is se
247. spheric Dump Bypass Plumbing Several components are available for connecting external devices to Series instruments These connection options include e Individual terminal board PCB assemblies e Terminal block and cable kits optional e Individual cables optional For detailed information on the optional connection components refer to the Optional Equipment chapter For associated part numbers refer to the Servicing chapter The following terminal board PCB assemblies are available for iSeries instruments e I O terminal board PCB assembly 37 pin standard e D O terminal board PCB assembly 37 pin standard e 25 pin terminal board PCB assembly included with optional I O Expansion Board Figure 2 5 shows the recommended method for attaching the cable to the terminal board using the included tie down and spacer Table 2 1 identifies the connector pins and associated signals Note All of the I O available in the instrument are not brought out on this terminal board if more I O is desired an alternative means of connection is required A Model 43i Instruction Manual 2 5 Installation Connecting External Devices See Detail A 2 6 Model 43i Instruction Manual See Detail B Assembled Connector Figure 2 5 I O Terminal Board Views Detail A Detail B Table 2 1 I O Terminal Board Pin Descriptions Pin Signal Description Pi
248. ssembly Light Baffle Flash Lamp Trigger Assembly Reaction Chamber Thermo Fisher Scientific System Description Hardware The optics section provides the light source for the fluorescence reaction and optimizes the reaction with a system of lenses and mirrors It includes a flash lamp condensing lens mirror assembly and light baffle The flash lamp provides the ultraviolet light source that causes the fluorescence reaction in the SO molecules The condensing lens focuses light from the flash lamp into the mirror assembly A set of eight mirrors selectively reflects only those wavelengths used in exciting SO molecules This reflective filtering causes the radiation reaching the detection chamber to be more intense and more stable throughout the lifetime of the instrument The circular baffle helps keep stray light from entering the detection volume The flash lamp trigger assembly pulses the UV flash lamp at a rate of 10 times per second for improved signal to noise ratio and long term stability The lamp is operated in the pulsed mode for six major reasons e Long life e High optical intensity improved signal to noise ratio e Small size e Low power requirements less than 1 watt e Long term stability e Chopped signal processing no dark current drift In the reaction chamber pulsating light from the flash lamp excites the SO molecules A condenser lens collects and focuses light from fluorescing SO mo
249. st one second for the action to be activated A Note Not all of the I O available in the instrument is brought out on the supplied terminal board if more I O is desired an alternative means of connection is required A In the Main Menu choose Instrument Controls gt I O Configuration gt Digital Input Settings Thermo Fisher Scientific Model 43i Instruction Manual 3 43 Operation Instrument Controls Menu Logic State Instrument Action Analog Output Configuration Select Channel 3 44 Model 43i Instruction Manual The Logic State screen is used to change the I O relay to either normally open or normally closed The default state is open which indicates that a relay connected between the digital input pin and ground is normally open and closes to trigger the digital input action If nothing is connected to the digital input pin the state should be left at open to prevent the action from being triggered Press to toggle and set the logic state open or closed a e The Instrument Action screen allows the user to select the instrument state that is tied to the selected digital input In the Main Menu choose Instrument Controls gt 1 O Configuration gt Digital Input Settings gt Select Relay gt Instrument State il mid The Analog Output Configuration menu displays a list of the analog output channels available for configuration Channel choices include all voltage channels all current channels
250. status on set pmt onoff These commands set the PMT on or off The example that follows turns the PMT off Send set pmt off Receive set pmt off ok save set save params This command stores all current parameters in FLASH memory It is important that each time instrument parameters are changed that this command be sent If changes are not saved they will be lost in the event of a power failure The example that follows saves the parameters to FLASH memory Send set save params Receive set save params ok time This command reports the current time 24 hour time The example that follows reports that the internal time is 2 15 30 pm Send time Receive time 14 15 30 set time hh mm ss hh hours mm minutes ss seconds This command sets the internal clock 24 hour time The example that follows sets the internal time to 2 15 pm Note If seconds are omitted the seconds default to 00 A Send set time 14 15 Receive set time 14 15 ok Thermo Fisher Scientific Communications Configuration Thermo Fisher Scientific C Link Protocol Commands Communications Configuration addr dns This command reports the TCP IP address for the domain name server Send addr dns Receive addr dns 192 168 1 1 set addr dns address This command sets the dns address where address consists of four numbers gt ranging from 0 255 inclusive separated by Send set addr dns 192 168 1 1 Receive set ad
251. stors are interchangeable to an accuracy of 0 2 C and have a value of 10K ohms at 25 C an alternate procedure is to connect an accurately known 10K resistor to the thermistor input AMB TEMP on the measurement interface board and enter the temperature reading A A 1 C change corresponds to a 5 change in resistance thus this g P g alternative procedure can be quite accurate as a check however it clearly is not NIST traceable A 3 From the Main Menu press _ to scroll to Service gt press gt to scroll to Temperature Calibration gt and press 4 The Calibrate Ambient Temperature screen appears Note If Service Mode is not displayed refer to Accessing the Service Mode in this chapter then return to the beginning of this step A 4 Waitat least 10 seconds for the ambient reading to stabilize use e Jand Cer Js to enter the known temperature and press to save the temperature value 5 Install the cover Thermo Fisher Scientific Servicing Input Board Replacement Use the following procedure to replace the input board Figure 7 19 Input Board gP p Replacement Esuipmen Required Input board Philips screwdriver 1 Disconnect the BNC signal cable and the ribbon cable 2 Loosen the two screws securing the bracket holding the input board to the floor plate slide the bracket towards the optical bench slightly and lift it off the base screws Power Supply B
252. struction Manual avg time high avg time low avg time These commands report the averaging time in seconds when operating in single range or averaging time used with the high and low ranges when operating in dual or auto range mode The example that follows shows that the averaging time is 300 seconds according to Table B 2 Send avg time Receive avg time 11 300 sec set avg time selection set high avg time selection set low avg time selection These commands set the averaging time high and low averaging times according to Table B 2 The example that follows sets the low range averaging time to 120 seconds Send set low avg time 8 Receive set low avg time 8 ok Table B 2 Averaging Times Selection Averaging Time seconds 0 1 seconds 2 5 10 20 30 oJ oino ay SI wt rp o N 5 o Y gt ow o o Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Measurements so2 high so2 low so2 These commands report the measured SO concentration when operating in single range or high and low SO when operating in dual or auto range mode The example that follows shows that the SO2 concentration is 40 Send S02 Receive S02 0040E 0 ppm conv temp This command reports the current SO converter temperature The example that follows reports that the current converter temperature is 45 C Send conv temp Receive conv temp 45 0 deg
253. strument Controls y as Lal Ebel Flash Lamp The Flash Lamp screen is used to turn the flash lamp on or off The flash lamp must be off when using the optical span test LED For more information about the optical span test LED see Optical Span Test later in this chapter In the Main Menu choose Instrument Controls gt Flash Lamp A pruni Lira el Pa rite in BA i PERI Piim Hi i Datalogging Settings The Series instruments include a built in data logging capability as a standard feature The operator is allowed to create two different types of records which for historical reasons are named lrecs and srecs Each record Thermo Fisher Scientific Model 43i Instruction Manual 3 23 Operation Instrument Controls Menu can contain up to 32 different fields or data items and records can be created at user defined intervals ranging from 1 to 60 minutes Record generation is tied to the instrument s real time clock For example if the logging period for srecs is set to 30 minutes a new srec will be generated on every hour and every half hour 10 00 10 30 11 00 Lrecs and srecs can be interleaved For example an srec containing just the current concentration level could be generated every five minutes while an lrec containing a full set of diagnostic data could be generated once every hour The analyzer s computer system includes three megabytes of flash memory which is enough to store a full lrec containin
254. sually or with a multimeter Unplug the power cord disconnect the switch and check operation with a multimeter Display does not come on Pump is running DC power supply failure DC power distribution failure Display failure Check the green LED on the back edge of the power supply If the LED is off the supply has failed Check surface mount LEDs labelled 24V PWR on the motherboard and the interface board If lit power is OK If possible check instrument function through RS 232 or Ethernet Contact Thermo Fisher Scientific Service Department Power comes on and the display functions but the pump is not running AC power is not reaching the pump Pump is jammed due to a new or stiff diaphragm Pump bearings have failed Locate the three pin connector on the interface board and use a voltmeter to check AC voltage across the two black wires should read 110 120V even on 220V or 100V instruments Carefully rotate the pump fan to get past sticking point Disconnect AC power and remove the gas lines from the pump head then try to rotate the pump fan If it is jammed or noisy the motor bearings may have failed Thermo Fisher Scientific Thermo Fisher Scientific Troubleshooting Troubleshooting Guides Table 6 2 Troubleshooting Calibration Failures Malfunction Cannot zero instrument or there is a high background signal when sampling zero air Zero air should produ
255. surement interface board The sequence of voltages is Positive 24 positive 15 negative 15 positive 5 and positive 3 3 Each voltage value is separated by a space Send diag volt mib Receive diag volt mib 24 1 14 9 14 9 4 9 3 2 diag volt iob This command reports the diagnostic voltage measurements on the I O expansion board The sequence of voltages is Positive 24 positive 5 positive 3 3 and negative 3 3 Each voltage value is separated by a space Send diag volt iob Receive diag volt iob 24 1 4 9 3 2 3 2 lamp setpoint This command reports the set lamp setpoint The example that follows reports the lamp setpoint is 1000 counts Send lamp setpoint Receive lamp setpoint 1000 counts set lamp setpoint value value O lt value lt 4095 These commands set the lamp setpoint in counts The example that follows sets the lamp setpoint to 100 counts Model 43 Instruction Manual B 15 C Link Protocol Commands Datalogging Datalogging B 16 Model 43 Instruction Manual Send set lamp setpoint 100 Receive set lamp setpoint 100 ok lamp int This command reports the current flash lamp intensity The example that follows reports that the current flash lamp intensity is 3867 Hz Send lamp int Receive lamp int 3867 Hz lamp voltage This command reports the current flash lamp voltage The example that follows reports that the current flash lamp voltage is 810 volts Send lamp voltage Receive lamp voltage
256. t Model 431 Instruction Manual D 3 Geysitech Bayern Hessen Protocol Geysitech Commands The lt address gt is optional which means it can be left out completely The lt address gt if present must match the Instrument Address Additional space can be present after the lt address gt If the received command does not satisfy the above formats or if the lt address gt does not match the Instrument Address the command is ignored This is a sample command to switch the instrument to zero mode instrument id 5 lt STX gt STO05 lt SP gt N lt CR gt Data Sampling Data This command DA initiates a data transfer from the instrument The Query Command instrument responds with measurement data which depends on the range DA mode and is listed in Measurements reported in response to DA command below The command structure for a data query command is as follows lt STX gt DA lt address gt lt ETX gt lt BCC gt The lt address gt is optional which means it can be left out completely The lt address gt if present must match the Instrument Address Additional space can be present after the lt address gt If the lt address gt is left out then no space is allowed in the query string A command with no address is also a valid command The following are the different acceptable formats of the DA command with Instrument Address 5 lt STX gt DA lt CR gt lt STX gt DA005 lt CR gt lt STX gt DA lt SP gt lt SP gt 5 lt ETX
257. t Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Turn instrument OFF unplug the power cord and remove the cover 2 Disconnect the pump power cable from the AC PUMP connector on the measurement interface board 3 Remove both lines from the pump 4 Loosen the four captive screws holding the pump bracket to the shock mounts and remove the pump assembly Model 43i Instruction Manual 7 9 Servicing Fan Replacement Pump Bracket Captive Screws 4 O 3 _ Floor Plate Pump Mounting Screws 2 Bottom Figure 7 4 Replacing the Pump 5 Install the new pump by following the previous steps in reverse 6 Perform a leak test as described in the Preventive Maintenance chapter Fan Replaceme nt Use the following procedure to replace the fan Figure 7 5 Equipment Required Fan Philips screwdriver amounts of static electricity A properly ground antistatic wrist strap must 1 Equipment Damage Some internal components can be damaged by small be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Turn instrument OFF unplug
258. t Cal Gas Thermistor gt Water Bath Enter the thermistor temperature from the thermometer and press Figure 9 3 Figure 9 3 Cal Gas Therm Bath Screen Remove the thermistor from the water bath dry and replace the permeation tube oven Make sure the charcoal scrubber is connected to the ZERO bulkhead on the rear panel Thermo Fisher Scientific Setting Temperature with Known Resistance Determining Permeation Rate by Weight Loss Thermo Fisher Scientific Optional Equipment Internal Permeation Span Source 6 Wait for the permeation gas temperature reading to stabilize Use the following procedure to set the gas temperature with an accurate known resistance Remove the thermistor from POJ1 on the measurement interface board Connect a 4 369 KQ resistor across pins 1 and 2 of POJ1 use a resistance substitution box and an accurate meter if necessary From the Permeation Oven menu select Cal Gas Thermistor then select known resistor Figure 9 4 Figure 9 4 Cal Gas Therm Resistor Screen Enter the exact value of the attached resistor and press to save the offset Remove the resistor and reconnect the gas thermistor Make sure the charcoal scrubber is connected to the ZERO bulkhead on the rear panel Wait for the permeation gas temperature reading to stabilize Use the following procedure to determine the permeation rate by weight loss Make sure the oven has been calibrated as described
259. t to 100 ppb Figure 3 6 Sample concentrations below 50 ppb are output based on low range selection and sample concentrations above 50 ppb are output based on high range selection When the low range is active the status output is at 0 volts When the high range is active the status output is at half of full scale Thermo Fisher Scientific Operation Range Menu When the high range is active the concentration must drop to 95 of the low SO range for the low range to become active In addition each SO analog output has a span coefficient There are two span coefficients so that each range can be calibrated separately This is necessary if the two ranges are not close to one another For example the low SO range is set to 0 50 ppb and the high SO range is set to 0 10000 ppb 100 ppb SO2 Concentration 50 ppb 0 ppb 10V SO2 Analog Output 5V d N ov 10V 100 ppb range SO2 Status Output 5V 50 ppb 50 ppb range range OV Figure 3 6 Analog Output in Autorange Mode By default in the autorange mode the analog outputs are arranged on the rear panel connector as shown in Figure 3 7 See Table 3 4 for channels and pin connections Autorange mode may be selected from Range Mode Select in the Service Menu later in this chapter Thermo Fisher Scientific Model 43i Instruction Manual 3 11 Operation Range Menu 000000000000000000 0000000000000000000 e Si
260. t to stabilize Plot a graph of instrument readings against the SO concentrations generated for the high range Connect a source of calibration gas to the SAMPLE bulkhead The calibration gas should be approximately 80 of the low full scale range Use to move the cursor to Calibrate LO SO and press to display the Calibrate LO SO screen The first line of the display shows the current SO reading The second line of the display is where you enter the low calibration gas concentration To enter the low calibration gas concentration use Ce gt to move the cursor left and right Use to increment and decrement the digit Press 4 to calibrate the instrument to the low SO calibration gas The display flashes the message SAVING and the corrected SO reading is displayed Press _M_ to return to the Calibration menu Generate five SO concentrations equally spaced between zero and the concentration above Record instrument reading for each concentration after allowing time for both gas generation system and instrument to stabilize Plot a graph of instrument readings against the SO concentrations generated for the low range This is the instrument calibration curve All future measurements should be interpreted using this curve The zero span check procedure is normally performed any time a quick check of the accuracy of the instrument is required Normally zero and span are checked d
261. ta output or the scratch pad sp list The example that follows shows the list for streaming data output Send list stream Receive list stream field index variable x x time 1 1 s02 2 18 pmtv er xy Ir xy sr xy x 0 1 Reply termination format see set format format command y 0 1 2 Output format see set erec lrec srec format format command These commands report the last long or short records stored and dynamic data In the example that follows the command requests a long record with no checksum in ASCII format with text For details on how to decode the flag fields within these records see the flags command Model 43 Instruction Manual B 17 C Link Protocol Commands Datalogging B 18 Model 43 Instruction Manual Send 1r01 Receive 1r01 18 09 05 12 03 flags 5c000044 SO2 9658E 5 low S02 8920E 5 Intt 28 6 Rctt 25 6 Pgast 0 00 Pres 720 3 Smpfl 0 489 PMt V 808 LmpV 449 Lmpi 4740E 1 erec This command returns a brief description of the main operating conditions at the time the command is issued i e dynamic data The example that follows shows a typical response The format is defined by the current settings of format and erec format commands For details on how to decode the flag fields within these records see the flags command Send erec Receive erec 09 45 04 07 05 flags 5C040000 so2 0 001 4 loso2 0 001 4 pmtv 595 331 FlshV 797 227 Intt 29 263 Rctt 44 938 PGast
262. tent screen is used to reset all of the datalogging field items to default values for the selected record type For more information about selecting the content of logged data fields see Select Content described previously In the Main Menu choose Instrument Controls gt Datalogging Settings gt Reset to Default Content Thermo Fisher Scientific Model 43i Instruction Manual 3 29 Operation Instrument Controls Menu Configure Datalogging Logging Period Min Memory Allocation Percent 3 30 Model 43i Instruction Manual The Configure Datalogging menu deals with datalogging configuration for the currently selected record type In the Main Menu choose Instrument Controls gt Datalogging Settings gt Configure Datalogging EHO gies The Logging Period Min screen is used to select the logging period in minutes for the record format srec or lrec List of choices include off 1 5 15 30 and 60 minutes default In the Main Menu choose Instrument Controls gt Datalogging Settings gt Configure Datalogging gt Logging Period Min EH mind MAMA The Memory Allocation Percent screen is used to select the percentage of each record type for both srecs and lrecs Percentages between 0 and 100 are available in increments of 10 Changing this value results in log erasure for both short records and long records In the Main Menu choose Instrument Controls gt Datalogging Settings gt Confi
263. terface board replace if bad Alarm Chamber Temp Chamber temperature below set point Heaters have failed Check 10K thermistor replace if bad Check temperature control board to insure the LEDs are coming on If not temperature control board could be defective Check connector pins for continuity Alarm Pressure High pressure indication Check plumbing for leaks Check the pump for a tear in the diaphragm replace with pump repair kit if necessary Refer to the Preventive Maintenance chapter in this manual Check that capillary is properly installed and 0 rings are in good shape Replace if necessary Alarm Flow Flow low Flow high Check sample capillary for blockage Replace as necessary If using sample particulate filter make sure it is not blocked Disconnect sample particulate filter from the sample bulkhead if flow increases replace the filter Perform leak test to check for internal leaks When delivering zero air or gas to the instrument use an atmospheric dump Alarm Intensity Low lamp is failing Check that the lamp and trigger pak are securely fastened Replace lamp Alarm Lamp Low voltage lt 750V Replace the lamp power supply voltage lamp power supply has failed High voltage gt 1200V Replace the flash lamp flash lamp failed Alarm SO Concentration has Check to insure range corresponds with expected Conc exceeded range l
264. the beginning of this step A 7 At the PMT Supply Settings screen press to select Manual PMT Adjustment The Set PMT Voltage Manual screen appears 8 At the Set PMT Voltage Manual screen use Ce t increment decrement the counts until the instrument displays the calibration gas concentration value Use the following procedure to replace the DC power supply Figure 7 11 Equipment Required DC power supply Philips screwdriver Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A Thermo Fisher Scientific Analog Output Testing Thermo Fisher Scientific Servicing Analog Output Testing 1 Turn instrument OFF unplug the power cord and remove the cover 2 Disconnect all the power supply electrical connections Note connector locations to facilitate re connection 3 Loosen the captive screw securing the power supply to the chassis plate and lift out the power supply 4 To install the DC power supply follow the previous steps in reverse Power Supply Captive Screw Figure 7 11 Replacing the DC Power Supply The analog outputs should be tested if the concentra
265. the motherboard include External connectors include e External Accessory e RS 232 485 Communications two connectors e Ethernet Communications e I O connector with Power Fail Relay 16 Digital Inputs and 6 Analog Voltage Outputs Internal connectors include e Function key panel and Display Thermo Fisher Scientific System Description Electronics e Measurement Interface Board Data e I O Expansion Board Data e Digital Output Board e AC distribution Measurement The measurement interface board serves as a central connection area for all Interface Board measurement electronics in the instrument It contains power supplies and interface circuitry for sensors and control devices in the measurement system It sends status data to the motherboard and receives control signals from the motherboard Measurement Interface Connectors located on the measurement interface board include Board Connectors e Data communication with the motherboard e 24V and 120VAC power supply inputs e Fan and solenoid outputs e 120VAC outputs for the pump and temperature control e Flow and pressure sensors e Ambient temperature sensor e Bench heater e PMT high voltage supply e Input board e Flash trigger e Flash intensity e Diagnostic LED Flow Sensor Assembly The flow sensor assembly consists of a board containing an instrumentation amplifier and a flow transducer with input and output gas fittings The flow transducer output is pr
266. the power cord and remove the cover 2 Remove the fan guard from the fan and remove the filter 3 Pull the power connector off the fan 7 10 Model 43i Instruction Manual Thermo Fisher Scientific Optical Bench Replacement Thermo Fisher Scientific Servicing Optical Bench Replacement 4 Remove the four fan mounting screws and remove the fan 5 Install a new fan following the previous steps in reverse Mounting Screws 4 Filter Fan Guard Figure 7 5 Replacing the Fan Use the following procedure to replace the optical bench Figure 7 6 Equipment Required Allen wrench 5 32 inch Philips screwdriver Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Turn instrument OFF unplug the power cord and remove the cover 2 Disconnect the electrical cables from the optical bench a Flash intensity assembly cable from the FLASH INT connector on the measurement interface board b Heater cable from AC BENCH connector on the measurement interface board Model 43i Instruction Manual 7 11 Servicing Optical Bench Replacement c PMT BNC cable from the input board
267. the zero calibration A Note Wait at least 30 seconds for the reading to stabilize before saving the value A In the Main Menu choose Service gt Pressure Calibration gt Zero EN EEE o Bone nr The Calibrate Pressure Span screen allows the user to view and set the pressure sensor calibration span point Note The plumbing going to the pressure sensor should be disconnected so the sensor is reading ambient pressure before performing the span calibration The operator should use an independent barometer to measure the ambient pressure and enter the value on this screen before calibrating A Note Wait at least 30 seconds for the reading to stabilize before saving the value A In the Main Menu choose Service gt Pressure Calibration gt Span Model 43i Instruction Manual 3 73 Operation Service Menu Restore Default Pressure Calibration The Restore Default Pressure Calibration screen allows the user to reset the configuration values to factory defaults Note Wait at least 30 seconds for the reading to stabilize before saving the value A In the Main Menu choose Service gt Pressure Calibration gt Set Defaults Flow Calibration The Flow Calibration menu is used to calibrate the flow sensor to zero 3 74 Model 43i Instruction Manual span or restore factory default values The flow calibration is visible only when the instrument is in service mode For more information on the service mode see Service
268. tion oven state to not present or the desired temperature choice of 30 35 and 45 G In the Main Menu choose Service gt Perm Oven Settings gt Perm Oven Selection The Factory Calibrate Gas Thermistor submenu is used for calibrating the permeation oven gas thermistor to either low point high point or set defaults In the Main Menu choose Service gt Perm Oven Settings gt Factory Cal Gas Therm Thermo Fisher Scientific Low and High Points Set Defaults Factory Calibrate Oven Thermo Fisher Scientific Thermistor Operation Service Menu The Calibrate Gas Thermistor Low Point screen allows the user to view and set the permeation oven thermistor resistance The low and high point screens function the same way In the Main Menu choose Service gt Perm Oven Settings gt Factory Cal Gas Therm gt Low Point The Set Defaults screen allows the user to reset the configuration values to factory defaults In the Main Menu choose Service gt Perm Oven Settings gt Factory Cal Gas Therm gt Set Defaults The Factory Calibrate Oven Thermistor submenu is used for calibrating the permeation oven heater thermistor to either low point high point or set defaults In the Main Menu choose Service gt Perm Oven Settings gt Factory Cal Oven Therm Model 43i Instruction Manual 3 83 Operation Service Menu Low and High Points The Calibrate Oven Thermistor Low Point screen allows the user
269. tion value on the front panel display disagrees with the analog outputs To check the analog outputs you connect a meter to an analog output channel voltage or current and compare the meter reading with the output value displayed on the Test Analog Outputs screen Equipment Required Multimeter Use the following procedure to test the analog outputs 1 Connect a meter to the channel to be tested Figure 7 12 shows the analog output pins and Table 7 4 identifies the associated channels 2 From the Main Menu press to scroll to Diagnostics press gt J to scroll to Test Analog Outputs and press Model 431 Instruction Manual 7 21 Servicing Analog Output Testing 7 22 Model 431 Instruction Manual The Test Analog Outputs screen displays Press _ 4 to scroll to the desired channel corresponding to the rear panel terminal pins where the meter is connected and press Ce The Set Analog Outputs screen displays Press _ to set the output to zero The Output Set To line displays Zero Check that the meter is displaying a zero value If the meter reading differs by more than one percent the analog outputs should be adjusted Refer to the Analog Output Calibration procedure that follows Press to set the output to full scale The Output Set To line displays Full Scale Check that the meter is displaying the full scale value If the meter reading differs by more than one percent the analog outp
270. tput 4 Ground Analog Voltage Output 6 Model 43i Instruction Manual 6 13 Troubleshooting Connector Pin Descriptions 6 14 Model 43i Instruction Manual Connector Label Reference Designator Pin 3 7 Signal Description Ground SER EN J7 Serial Enable Jumper 3 3V 24V IN 24V Ground DIGITAL OUTPUT oOo a Nn on FP WW DN i 5V 24V 24V Ground Ground Ground SPI Reset SPI Input SPI Output SPI Board Select SPI Clock EXT RS485 J15 OO a N on A WW N gt gt gt gt gt al w N RS485 to Rear Panel RS485 to Rear Panel 5V 5V 5V Ground Ground Ground NC NC 24V 24V 24V 24V 24V 24V MONITOR J17 24V Power Monitor Ground FRONT PANEL BD J18 Ground Thermo Fisher Scientific Thermo Fisher Scientific Troubleshooting Connector Pin Descriptions Connector Label Reference Designator Pin Signal Description 2 Ground 3 LCLK LCD Signal 4 Ground 5 Ground 6 LLP LCD Signal 7 LFLM LCD Signal 8 LD4 LCD Signal 9 LDO LCD Signal 10 LD5 LCD Signal 11 LD1 LCD Signal 12 LD6 LCD Signal 13 LD2 LCD Signal 14 LD7 LCD Signal 15 LD3 LCD Signal 16 LCD Bias Voltage 17 5 18 Ground 19 Ground 20 LCD_ONOFF LCD Signal 21 Keypad Row 2 Input 22 Keypad Row 1 Input 23 Keypad Row 4 Input 24 Keypad Row 3 Input 25 Keypad Col 2 Select 26 Keypad Col 1 Select 27 Ke
271. transducer following the previous steps in reverse 6 Calibrate the flow transducer Refer to the Flow Transducer Calibration procedure that follows 7 30 Model 43i Instruction Manual Thermo Fisher Scientific Flow Transducer Calibration Thermo Fisher Scientific Servicing Flow Transducer Calibration Figure 7 14 Replacing the Flow Transducer Use the following procedure to calibrate the flow transducer Equipment Required Calibrated flow sensor WARNING The service procedures in this manual are restricted to qualified representatives 4 If the equipment is operated in a manner not specified by the manufacturer the protection provided by the equipment may be impaired A Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly ground antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch the instrument chassis before touching any internal components When the instrument is unplugged the chassis is not at earth ground A 1 Remove the cover Model 431 Instruction Manual 7 31 Servicing Flow Transducer Calibration 7 32 Model 43i Instruction Manual 2 3 Disconnect the pump cable from AC PUMP connector on the measurement interface board From the Main Menu press _ to scroll to Service gt press gt to scroll to Flow Calibration gt and press
272. tring string cur avg min max These commands set the data treatment to string where string is current average minimum or maximum for the concentration values recorded in the long records lrecs or short records srecs The example that follows sets the data treatment for concentrations in lrec to minimum Send set data treatment lrec min Receive set data treatment lrec min ok set copy sp to lrec set copy sp to srec set copy sp to stream These commands copy the current selections in scratch pad sp into the long record short record or streaming data list The example that follows copies the current list in scratch pad into the long records list Send set copy sp to lrec Receive set copy sp to lrec ok set copy lrec to sp set copy srec to sp set copy stream to sp These commands copy the current contents of the long record short record or streaming data list into the scratch pad sp These commands are useful in easy modification of current long record short record or streaming data lists The example that follows copies the current list of long records into the scratch pad Send set copy lrec to sp Receive set copy lrec to sp ok sp field number This command reports the variable number and name stored at index in the scratch pad list The example that follows shows that the field 1 in the Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Datalogging scratch pad is
273. ts signal types of either alarm and non alarm to choose from In the Main Menu choose Instrument Controls gt I O Configuration gt Output Relay Settings gt Select Relay gt Instrument State CH CH Erer pi prip CH Py Timms Lid bal EME EP Saad Bone BME TY Saad Alarms The Alarms status screen allows the user to select the alarm status for the selected relay output The selected item is shown by lt after it In the Main Menu choose Instrument Controls gt I O Configuration gt Output Relay Settings gt Select Relay gt Instrument State gt Alarms Note The I O BD STATUS alarm is only present if the I O expansion board is installed ZERO CHK CAL and SPAN CHK CAL are only present if autozero span check is enabled A DH iy A E 3 42 Model 43i Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu Non Alarm The Non Alarm status screen allows the user to select the non alarm status for the selected relay output The selected item is shown by lt after it In the Main Menu choose Instrument Controls gt I O Configuration gt Output Relay Settings gt Relay 1 10 gt Instrument State gt Non Alarm Digital Input Settings The Digital Input Settings menu displays a list of the 16 digital inputs available and allows the user to select the instrument parameter or logic state to change for the relay selected Note The digital inputs must be asserted for at lea
274. ts suisse i ea 8 10 Digital Input ii 8 10 Serial Por ind gis iaaa ai kanian EER raais 8 11 RS 232 GO mm ect Om in as 8 11 RS 485 Connection nena nnn aid 8 12 Ethernet o A a a eraa a aoit 8 12 External Accessory Connector 8 12 Internal Zero Span and Sample Valves iii aii tii 9 1 Optional Equipment sscsscssssssssscsssssssssssessssseseessessssassatesssenessssesaseateases 9 1 Internal Permeation Span Source 22s soccsuncodt acts acetic escact warcran ce eects 9 1 Permeation Tube Installation pisito nce dees 9 2 Computation of Concentrations sisi 9 3 Oven Installation and Contrato sio 9 3 Permeation Tube Oven Cabrio 9 5 Setting Perm Oven Tenperitlle io dia 9 5 Setting Temperature with Water Bath 9 6 Setting Temperature with Known Resistance ssec 9 7 Determining Permeation Rate by Weight LoSS cooncnnncnmmmm 9 7 Determining Release Rate by Transfer Standard coccion 9 8 Heated Hydrocarbon cer ondaa acta 9 9 Tefl n Particulate lt ea ea 9 9 I O Expansion Board Assembly taria bi 9 9 25 Pin Terminal Board Assembly iii ic 9 9 Terminal Block and Cable Kits coi seceateeceetteactd ei eaetae coe 9 9 EI EEA E O as 9 10 Mounting Opt siii ii ae 9 11 Waai a a a A 1 C Link Protocol CommManda ssssssessssssssessseesssessseessseessseessseessseessneesess B 1 Instrument Identification Number aiii iia B 1 Commands Fossa dba apne ef oa a a a wae Doles eda enti EEE A eieaa B 2 Accessing Streaming Data is B 2 Enteting Units
275. ual B 47 C Link Protocol Commands Record Layout Definition This shows that the bitfield end the second part of a bitfield specification is optional The bitfield will be one bit long starting in this case at the eleventh bit Background 7 8Bd ddd set 03 bkg s n This shows the use of indirect precision specifiers for floating point displays The background value is taken from the 7th element and the precision specifier is taken from the 8th If the asterisk were not present it would indicate instead that 8 digits after the decimal point should be displayed B 48 Model 43 Instruction Manual Thermo Fisher Scientific Serial Communication Thermo Fisher Scientific Parameters Appendix C MODBUS Protocol This appendix provides a description of the MODBUS Protocol Interface and is supported both over RS 232 485 RTU protocol as well as TCP IP over Ethernet The MODBUS Commands that are implemented are explained in detail in this document The MODBUS protocol support for the iSeries enables the user to perform the functions of reading the various concentrations and other analog values or variables read the status of the digital outputs of the analyzer and to trigger or simulate the activation of a digital input to the instrument This is achieved by using the supported MODBUS commands listed below For details of the Model 43 MODBUS Protocol specification see the following topics e Serial Communication Pa
276. ulations The second pressure is the actual pressure reading being measured If pressure compensation is on then both pressure readings are the same If pressure compensation is off a pressure of 760 mmHg is used as default pressure even though the actual pressure is 753 4 mmHg The example that follows shows that the actual reaction chamber pressure is 753 4 mmHg Send pres Receive pres 753 4 mmHg react temp This command reports the current reaction chamber temperature The example that follows reports that the current reaction chamber temperature is 45 2 C Send react temp Receive react temp 45 2 deg C flags This reports 8 hexadecimal digits or flags that represent the status of the flash lamp LED pressure and temperature compensation status gas units gas mode and alarms To decode the flags each hexadecimal digit is converted to binary as shown in the Table B 1 It is the binary digits that define the status of each parameter In the example that follows the instrument is reporting that the flash lamp is on and the SO high concentration alarm is activated Send flags Receive flags 40028000 Thermo Fisher Scientific C Link Protocol Commands Alarms Figure B 1 Flags Alarms alarm chamber temp min alarm chamber temp max These commands report the chamber temperature alarm minimum and maximum value current settings The example that follows reports that the chamber temperature alarm minimum
277. ument OFF unplug the power cord and remove the cover Disconnect plumbing from the pressure transducer assembly Note the plumbing connections to facilitate reconnection Disconnect the pressure transducer cable from the measurement interface board Loosen the two pressure transducer assembly retaining screws and remove the pressure transducer assembly by sliding it down then taking it out To install the pressure transducer assembly follow the previous steps in reverse Calibrate the pressure transducer Refer to the Pressure Transducer Calibration procedure that follows Model 431 Instruction Manual 7 27 Servicing Pressure Transducer Calibration Pressure Transducer 7 28 Calibration Model 43i Instruction Manual Retaining Screws 2 Pressure Transducer Assy Figure 7 13 Replacing the Pressure Transducer Assembly Use the following procedure to calibrate the pressure transducer Notes An error in the zero setting of the pressure transducer does not introduce a measurable error in the output concentration reading Therefore if only a barometer is available and not a vacuum pump adjust only the span setting A A rough check of the pressure accuracy can be made by obtaining the current barometric pressure from the local weather station or airport and comparing it to the pressure reading However since these pressures are usually corrected to sea level it may be necessary to correct the reading to loca
278. utputs are arranged on the rear panel connector as shown in Figure 3 4 See Table 3 2 for channels and pin connections Single range mode may be selected from the Range Mode Select in the Service Menu later in this chapter 000000000000000000 0000000000000000000 37 O Y Figure 3 4 Pinout of Rear Panel Connector in Single Range Mode Thermo Fisher Scientific Operation Range Menu Table 3 2 Default Analog Outputs in Single Range Mode Channel Connector Pin 1 0 Terminal Pin Description 1 14 1 SO Analog Output 2 33 3 SO Analog Output 3 5 5 None 4 34 7 None 5 17 9 None 6 36 11 None Ground 16 18 19 35 37 2 4 6 8 10 12 Signal Ground Note All channels are user definable Ifany customization has been made to the analog output configuration the default selections my not apply A Dual Range Mode In the dual range mode there are two independent analog outputs These are labelled simply as the High Range and the Low Range Each channel has its own analog output range averaging time and span coefficient This enables the sample concentration reading to be sent to the analog outputs at two different ranges For example the low SO analog output can be set to output concentrations from 0 to 50 ppb and the high SO analog output set to output concentrations from 0 to 100 ppb In addition each SO analog output has a span coefficient There are two
279. uts should be adjusted Refer to the Analog Output Calibration procedure that follows Press to reset the analog outputs to normal 8 p Thermo Fisher Scientific Analog Voltage Outputs o ms Servicing Analog Output Testing Current Outputs Analog Voltage Inputs Figure 7 12 Rear Panel Analog Input and Output Pins Table 7 4 Analog Output Channels and Rear Panel Pin Connections Voltage Pin Current Pin Channel Channel 1 14 1 15 2 33 2 17 3 15 3 19 4 34 4 21 5 17 5 23 6 36 6 25 Ground 16 18 19 35 37 Current Output Return 16 18 20 22 24 Thermo Fisher Scientific Model 431 Instruction Manual 7 23 Servicing Analog Output Calibration 7 24 Model 43i Instruction Manual Table 7 5 Analog Input Channels and Rear Panel Pin Connections Input Channel Pin 1 o INIo on AIUI N oI INI omy on wi rm o Ground 4 8 11 Use the following procedure to calibrate the analog outputs ifa meter reading in the Analog Output Testing procedure differed by more than one percent or after replacing the motherboard or optional I O expansion board Equipment Required Multimeter 1 Connect a meter to the channel to be adjusted and set to voltage or current as appropriate Figure 7 12 shows the analog output pins and Table 7 4 identifies the associated channels 2 From the Main Menu press _4_ to scroll to Service press gt 4
280. value is 35 0 C Send alarm chamber temp min Receive alarm chamber temp min 35 0 deg C set alarm chamber temp min value set alarm chamber temp max value These commands set the chamber temperature alarm minimum and maximum values to value where value is a floating point number representing chamber temperature alarm limits in degrees C The example that follows sets the chamber temperature alarm maximum value to 55 0 Es Send set alarm chamber temp max 55 0 Receive set alarm chamber temp max 55 0 ok Thermo Fisher Scientific Model 437 Instruction Manual B 11 C Link Protocol Commands Alarms B 12 Model 43 Instruction Manual alarm conc so2 min alarm conc so2 max These commands report the SO concentration alarm minimum and maximum values current setting The example that follows reports that the SO concentration minimum is 5 2 ppb Send alarm conc so2 min Receive alarm conc so2 min 5 2 ppb set alarm conc so2 min value set alarm conc so2 max value These commands set the SO concentration alarm minimum and maximum values to value where value is a floating point representation of the concentration alarm limits Values must be in the ppb or pg m Refer to Entering Units in PPB in this chapter The example that follows sets the SO concentration alarm maximum value to 80 ppm Send set alarm conc so2 max 80000 Receive set alarm conc so2 max 80000 ok alarm converter temp min alarm converter temp max T
281. wn if the instrument is locked In the Main Menu choose Password gt Unlock Instrument mi Cua aa nid 3 88 Model 43i Instruction Manual Thermo Fisher Scientific Zero Gas Generation Commercial Heatless Air Dryers Thermo Fisher Scientific Chapter 4 Calibration The Model 43i requires initial and periodic calibration according to the procedures described in this chapter A quality control plan that allows the frequency and number of calibration points to be modified depending on calibration and zero span check data should be implemented Such a quality control program is essential to ascertain the accuracy and reliability of the air quality data collected The data compiled for such a program might include items such as dates of calibration atmospheric conditions control settings and other pertinent data For more detailed quality assurance guidelines see the Quality Assurance Handbook for Air Pollution Measurement Systems published by the U S EPA Research Triangle Park NC 27711 There are a number of conditions which should be met prior to a calibration or a zero span check First the instrument should have at least 30 minutes to warm up and stabilize Second the range used during the calibration or zero span check should be the same as that used during normal monitoring Third all operational adjustments to the instrument should be completed prior to calibration Fourth all parts of the gas flow system such a
282. ximately 5 KQ across pins 3 and 4 of POJ3 Go to the High Point screen enter the exact value of the resistor and press to calibrate the high resistance point Press Cm to return to the Factory Cal Oven Therm menu Remove resistor from POJ3 and re attach the permeation oven Perform a thermistor calibration using the Permeation Tube Oven Calibration procedure that follows Thermo Fisher Scientific Optional Equipment Internal Permeation Span Source Permeation Tube There are two general approaches that can be used to calibrate the permeation tube oven The first is to calibrate the temperature indicator Oven Calibration very accurately to better than 0 02 C and to use a permeation tube whose weight loss has been previously determined at that temperature Note An error of approximately 0 1 C corresponds to an error of 1 in release rate A The second approach is to note that the thermistors used to measure temperature are interchangeable to better than 0 2 C Thus a 1 resistor of the proper value 4 369 KQ for 45 C can be used to set the span on the measurement interface board The release rate for the permeation tube is then determined by weight loss in the actual oven being used Setting Perm Oven You can use either of the two calibration methods presented here One Temperature method involves performing the Setting Perm Oven Temperature procedure and then continuing with the Setting Temperature with
283. ypad Col 4 Select 28 Keypad Col 3 Select 29 Ground 30 Ground 31 Ground 32 Ground 33 24V 34 24V RS232 RS485 A PTA 1 NC Model 431 Instruction Manual 6 15 Troubleshooting Connector Pin Descriptions Connector Label Reference Designator Pin Signal Description Serial Port 1 RX RS485 IN Serial Port 1 TX RS485 OUT NC Ground NC Serial Port 1 RTS RS485 OUT Serial Port 1 CTS RS485 IN NC NC Serial Port 2 RX RS485 IN Serial Port 2 TX RS485 OUT NC Ground NC Serial Port 2 RTS RS485 OUT Serial Port 2 CTS RS485 IN NC AC HOT AC NEUT AC Ground AC HOT AC NEUT AC Ground AC HOT 2 AC NEUT 3 AC Ground oOo a nN oon FP WW NY RS232 RS485 B P1 B OO a N on A WW N AC IN o E 0 N AC 24VPWR PJ2 w N AC INTF BD PJ3 p Table 6 6 Measurement Interface Board Connector Pin Descriptions Connector Label Reference Designator Pin Signal Description DATA IPJ8 1 Ground RS485 from Motherboard 2 3 RS485 from Motherboard FLASH TRIG FJ1 Flash Voltage 2 Flash Trigger 6 16 Model 43i Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Connector Label Reference Designator Pin 3 Troubleshooting Connector Pin Descriptions Signal Description Ground FLASH INT FJ2 lt FP WO N 15V 15V Ground Lamp Intensity INPUT BD MJ8 oOo ao N own A UO N gt o 15V Ground 15V

Download Pdf Manuals

image

Related Search

Related Contents

AWH-1010RB  User`s Manual  User manual (GB) PAGE 3-4 Bedienungsanleitung (D  LG 55EA8800 Energy Guide : Free Download, Borrow, and Streaming : Internet Archive  User`s Guide  Philips Fidelio SoundAvia  FI VM209.indd  PRODUITS COMPLEMENTAIRES  MANUAL PULSCALC PROGRAM  データシート  

Copyright © All rights reserved.
Failed to retrieve file