         Contents
1.      Accurate maintenance increases the service life as well and protects against  premature rotor failures     Corrosion and damages resulting from insufficient maintenance could not be  subject of claims lodged against the manufacturer        One can use all standard disinfectants  The centrifuges and accessories are constructed from  various materials and one should to take into account possible variety of materials  During  sterilization by means of steam one should to consider temperature resistance of individual  materials     6 6 Sterilization    sterilization   radiation B  radiation y        formalin    ethylene oxide  ethanol          may be used     cannot be used  x Laboratory vessels have to be exactly cleaned and rinsed with the distilled water before the sterilization in the    autoclave  It is always necessary to remove closures from containers   1  The frequent steam sterilization reduces mechanical durability  PC test tubes may become useless     2  Except PVC hoses which are resistant to the steam sterilization in the temperature 121 C    19    6  Centrifuging    Chemical resistance of plastic    vV           5 o  op 9  55   d  5                     o        aldehydes  cyclic alcohols  cyclic  hydrocarbons  haloid  hydrocarbons      e   of Las   oe         o        e           Lasel  o                             0   oe                                           oe       ojoj of      o  o  o     Permanent action of the substance does not cause damage through
2.    3500 or  L  1000 rpm  MPW 54s     e set up time of centrifuging in the range 130 min or hold time mode     13    5  Operating and design  5 6 Safety features    Cover lock    The centrifuge can be started only with properly closed cover  While  the cover can be  opened only after stopping the rotor  In case of emergency opening of the cover during operation   the centrifuge will be immediately switched off and the rotor will brake till complete stopping     Rest state inspection    Opening of the centrifuge   s cover is possible only with the rotor in the state of rest     EMERGENCY COVER RELEASE    In case of e g  mains failure it is possible to open cover manually  At first  one must be sure  that rotor is not in the move  use inspection glass   On the right hand side of the casing    there is a hole  Insert emergency opening key  17162  into the hole     It is not allowed to use emergency cover release when the rotor is running        14    6  Centrifuging  6  Centrifuging    Power switching ON OFF is carried out with master switch situated on the side wall of the centrifuge   All settings on the centrifuge are done by means of the control panel     6 1 Control panel   The control panel  equipped with membrane keyboard  placed on the front casing serves the  purpose of controlling centrifuge operation  STATUS SPEED display show information about state of  centrifuge  left side  and speed mode  right side   TIME display show information about time remain  to the end of c
3.    centrifuge No   MPW 54  user manual   Cat  No   20054  EN       MPW    MED  INSTRUMENTS    2014 12 09    USER MANUAL          LABORATORY CENTRIFUGE    MPW 54        READ IT BEFORE USE         MPW MED  INSTRUMENTS  46 Boremlowska Street  04 347 Warsaw Poland  tel   48 22 610 81 07  service   fax  48 22 610 55 36      IVD    mpw mpw pl  www mpw pl       Technical specification  WARNING SIGNS AND HAZARD ICONS    WARNING   Warning of potential injury or health risk     DANGER     Risk of electric shock with potential for severe injury or death as a consequence     DANGER     Biohazard with potential for risk to health or death as a consequence     DANGER     Risk of explosion with potential for severe injury or death as a consequence        This manual was prepared with special care  MPW MED  INSTRUMENTS may change the manual at any  time and without notice because of improvements and inaccuracies of current information     Technical specification    Content  1 TECHNICAL SPECIFICATION gege Ee 4  2  OR Le WR      DEE 5       SAFEN NOTE ee D  E e Malte DEE D  e SAEN ETAN E EE 6  LOADING THE ROTOR EE 7  SAEED A A D A OT         a EEE A eene E E    E          8  3 5 SAFETY  Oh KEEN 8  3 EXIVTAINSEENANCE CONDITIONS a a OS Dir 9  SU SAFELY PRECAUTIONS EE 9  4  INSTALLATION deeg 11  A 1 THE CONTENTSOF THE PACKAGE EE 11  Ph NEO Ne NEE 11  5 OPERATING AND DESIGN                                                    ENEE Ee EE 12  5 1 DESCRIPTION  OF CENTRIFUGE TE t 2 2 02 0        Age 12 
4.  30 days  The    very good We  IP material is able to be resistant through years     Continuous action of the substance causes insignificant and partly reversible damage  good to limited through the period of 7 30 days  e g  puffing up  softening  reduced mechanical  durability  discolouring      The material should not have the continuous contact with the substance  The  limited immediate occurrence of damage is possible  e g  the loss of mechanical durability   deformation  discolouring  bursting  dissolving         Plastics                                     polymethylpentene                   20    6  Centrifuging    DANGER     For centrifuging infectious materials it is necessary to use hermetically closed buckets  in    order to prevent they migration into the centrifuge        Rotors  buckets and round carriers can be sterilized in autoclave with temperature 121     124  C and  pressure 215 kPa during 20 min  In the centrifuge  disinfectants and cleaning agents generally used in  medical care should be used  e g  Aerodesina 2000  Lysoformin 3000  Melseptol  Melsept SF   Sanepidex  Cutasept         Additional accessories can be sterilized using autoclave depending on material that they are made of   See table STERILIZATION     User is responsible for proper disinfections of the centrifuge  if some dangerous material  was spilled inside or outside of the centrifuge  During the above mentioned works one    must wear safety gloves        6 7 Chemical resistance    One can
5.  Directive 2004 108 WE  EMC     Standard PN EN ISO 15223 1 2012    SPOLDZIELNIA PRACY  Wojciel ojszewski            Warsaw  46 Boremlowska Street  Quality policy in line with ISO 9001 2008         CT  PREZESZ im  MPW MED  INSTRUMENTS           MPW MED  INSTRUMENTS   SPOLDZIELNIA PRACY  w Warszawie    Certifying authority       DECLARATION OF CONFORMITY      Warsaw  13 11 2014       DECLARATION OF DECONTAMINATION    In order to protect our employees please fill out the declaration of decontamination    completely before sending centrifuge to the manufacturer  repair      1  Device                   So ETC Ce GG    2  Description of decontamination     see user Manual     3  Decontamination carried out by                 qj              J  kRkKIJk lqll Qo nnnm    4  Date and signature RE    DECLARATION OF DECONTAMINATION    In order to protect our employees please fill out the declaration of decontamination    completely before sending centrifuge to the manufacturer  return      5  Device    est bd SE      serial No      6  Description of decontamination     see user Manual     7  Decontamination carried out by                      8  Dateandsisnature               ooneacnicii i at e Env vet                                       
6.  ENEE          ce PE                                                            12  5 3 ROTORAND  ACCESSORIES INSTAEEATIOB 30          eite uera Eyed Fue ame                                     ee b                        13  5  Be hei fl een DEEN 13  5 5 SET       PARAIVIETERS EE 13  OAE EE a ee ENEE 14  Ge CE CHILE  LEE EE EE EE 15  b  eeh al E EE 15  op          Tele 16  6 3 CENTRIFUGING FINISH EE 16  6 4  SIGNALLING STATES OF CENTRIFUG                            a          vr Dav Quad ove ga                         17  6 5 MAINTENANCE OF CENTRIFUGING ELEMENTS                                     sese              18  E H sic  HP SERIO PNEU RR RR 19  EERSTEN 21  7  GUARANTEE  SERVICE EE 22  g DISPOSAL EE 23  EK  MANUFACTURER    SINFO                                                      EE 24  ENEE ele ee 25      ADDITIONAL ACCESSORIES       DECLARATION OF CONFORMITY     DECLARATION OF DECONTAMINATION  REPAIR     DECLARATION OF DECONTAMINATION  RETURN     Technical specification    1 Technical specification       MPW MED  INSTRUMENTS  SP    DZIELNIA PRACY   Boremlowska 46 Street  04 347 Warsaw   typ wes  running time 01 00   30 00    min   s   step 1min    short time operation mode     SHORT no    continuous operation mode     HOLD yes    electromagnetic compatibility according to PN EN 55011    ambient conditions PN EN 61010 1  pkt 1 4 1   set up site indors only    ambient temperature 2     40        humidity  maximum relative humidity   lt  80     excess voltage category 
7.  use all standard disinfectants  The centrifuges and accessories are constructed from various  materials and one should to take into account possible variety of materials  During sterilization by  means of steam one should to consider temperature resistance of individual materials     DANGER     For centrifuging infectious materials it is necessary to use hermetically closed buckets   in order to prevent they migration into the centrifuge     User is responsible for proper disinfections of the centrifuge  if some dangerous  material was spilled inside or outside of the centrifuge  During the above mentioned  works one must wear Safety gloves        21    Guarantee  service    7  Guarantee  service    Manufacturer grants to the Buyer the guarantee on conditions specified in the Guarantee  Certificate  Buyer forfeits the right to guarantee repair when using the device inconsistently with the  User manual provisions  when damage results from the User   s fault    Repairs should be carried out in authorized service workshops  granted with the MPW  Certificate    The centrifuge shall be sent to repair after decontaminating disinfections  Information about  authorized service workshops could be obtained from the Manufacturer     22    Disposal    8  Disposal    When you are disposing the device  the respective statutory rules must be  observed     Pursuant to guideline 2002 96 EC  WEEE   all devices supplied after August  13  2005  may not be disposed as part of domestic waste     
8.  used  for spinning in this centrifuge  The use of equipment made by other manufacturers should  be consulted with the manufacturer of the centrifuge     It is not allowed to lift or shift the centrifuge during operation  and rest on it     It is nor allowed to stay in the safety zone within 30 cm distance around the  centrifuge neither leave within this zone some things  e g  glass vessels     It is not allowed to put any objects on the centrifuge     COVER OPENING      It isn t allowed to open the cover manually in emergency procedure when rotor is  still turning     ROTORS    It is not allowed to use the rotors and round carriers with signs of corrosion or  other mechanical defects     It is not allowed to centrifuge highly corrosive substances which may cause  material impairment and lower mechanical properties of rotor and round carriers     It is not allowed to use rotors and accessories not admitted by the manufacturer   Let to use commercial glass and plastic test tubes  which are destined to  centrifuging in this laboratory centrifuge  One should absolutely not use poor  quality elements  Cracking of glass vessels and test tubes could result in dangerous  vibration of the centrifuge     It is not allowed to carry out centrifugation with the rotor caps taken off or not  driven tight        10    Instalation  4 instalation  Open the package  Remove the box containing the accessories  Take out centrifuge from the    container  Keep the box and packing materials in case 
9. 14082  13 530   13 30 13 3530 13 3030  Brak dna        3120   5800            26                5    Q xL  mm  12x100 16x113   16x108 12x75  Catalogue number 15119 15048 15053 15118    6 a           N P ei     e   EU  d          D  t     D  1 1          Angle rotor  30    11145 with  buckets 13080    Time of decelaration  5     Capacity  ml  BL BL 19          Q xL  mm  11 7x95 12x75 12x85 16x108  Catalogue number 15054 15120 15419 15053    i           Angle rotor  30    11145 with j      buckets 13081    14082 14082  ack of bottom   3120  5800           27    ANNEXES    ADDITIONAL ACCESSORIES   INNE OTHER    Przetwornica napiecia 300W   Power inverter 300W  12DC  gt  230 AC   e g     umozliwiajaca zasilanie pradem statym possibility of supply from the cigarette  o napi  ciu 12 V  m in  z zapalniczki lighter socket     samochodowej     Only original MPW buckets can be used  One should use tubes which  dimensions and durability is proper Using of tubes from other  companies should be consulted with manufacturer of centrifuge  In  the centrifuge  disinfectants and cleaning agents generally used in  medical care should be used  e g  Aerodesina 2000  Lysoformin 3000   Melseptol  Melsept SF  Sanepidex  Cutasept F         28       MED INSTRUMENTS          Product Laboratory centrifuge         Model       MPW 54    Product classification on the basis of   Non classified to list A or B and not for    PN EN ISO 13485 2012 AC 2013 03 PN EN 61010 2   PN EN 13612 2  PN EN 61326 2 6 2     
10. PN EN 61010 1  pollution degree PN EN 61010 1    safety area 300 mm    dimensions         height       width  W   depth  D   height with open lid Hoc     56 dB 56 dB  weight of centrifuge 230V  weight of centrifuge 230V    There is a possibility of using power inverter 12DC 230AC  look  10 ANNEXES  gt  ADDITIONAL  ACCESSORIES   gt  OTHER        Application    2 Application    The MPW 54 is a table top laboratory centrifuge for in vitro diagnostic  IVD   Device is used for  separation samples taken from people s  animal   s and plant   s components of different densities   under the influence of the centrifugal force  to provide information about their biological state  Its  construction ensures easy operation  safe work and wide range of applications at laboratories  engaged in routine medical analyses  biochemical research works etc  This centrifuge is not biotight  and therefore during centrifugation of preparations requiring biotightness one has to use closed and  sealed containers and rotors  In the centrifuge  it is prohibited to centrifuge caustic  inflammable and  explosive preparations     Safety notes    3 Safety notes    3 1 Personnel    3 2 Guarantee    Laboratory centrifuge can be operated by laboratory personnel after getting  acquainted with user manual     User manual shall be always held near the centrifuge     The centrifuge can not be misused        Guarantee period amounts to minimum 24 months  unless otherwise specified in  the purchase documents      The 
11. The device belongs to 8th group  medical devices  and is categorized in  business to business field     The icon of the crossed out trash can shows that the device may not be  disposed as part of domestic waste  The waste disposal guidelines of the  individual EC countries might vary  If necessary  contact your supplier        23    Manufacturer   s info    9  Manufacturer   s info    MPW MED  INSTRUMENTS 22 610 56 67 sales department  22 610 81 07 service  22 610 55 36 fax    Boremlowska 46 Street http    www mpw pl  04 347 Warsaw e mail  mpw mpw pl       DISTRIBUTOR     24                5    10  ANNEXES    Capacity  ml  15 10    Nalgene conical with ca  Tube glass 8 bottom H glass    P x L  mm  12x100 16x113   17x120   16x108 12x75    Catalogue number 15119 15053 15118    EE 2 e      wile 4     d     Angle rotor  30    11101 with    buckets 13080    14082    Capacity  ml       6   5 J  5     10       J    Q xL  mm  11 7x95 12x75 12x85 16x108  Catalogue number 15054 15120 15419 15053    LI EL LE JL    14082 14082  13 530   13 330 13 330 13 330  ack of bottom   3120  5800  23    Time of decelaration  s                  Angle rotor  30   11101 with      buckets 13081       25    ANNEXES    Weeer conical  Tube glass   bottom                12  100    Catalogue number 15119 15048    Angle rotor  30    11144 with d  buckets 13080    14082  lack of bottom  4    Time of decelaration  s     Capacity  ml           mm  11 7x95        Angle rotor  30    11144 with    buckets 13081    
12. a tube can be filled only  partially  or limit the speed of a centrifuge  which shall be determined from the formula     1 2    n admissible   Nmax X            density   5   Nmax  Max speed  rpm   y cm       3 7 Safety precautions    For safety reasons  inspections of the centrifuge carried out by the authorized service at least  once a year after the period of warranty  The reason for more frequent inspections could be  corrosion inducing environment  Examinations should end with issuing report of validation that  checks on the technical state of the laboratory centrifuge  It is being recommended to establish  document where every repairs and reviews are being registered  Both these documents should be  stored in the place of use of the centrifuge     Safety notes    INSPECTION PROCEDURES CARRIED OUT BY THE OPERATOR    Operator has to pay special attention to the fact that the centrifuge parts of key  importance due to safety reasons are not damaged  This remark is specifically important  as for   Centrifuge accessories and especially structural changes  corrosion  preliminary  cracks  abrasion of metal parts     Screw joints   Inspection of the rotor assembly   Inspection of bioseals of the buckets if such are used     Control of execution of the guarantee yearly technical inspection of the centrifuge     Only the manufacturer specified holders  included in the equipment list  as well as  centrifuge capillaries  which diameter  length and durability are suitable  should be
13. entrifuging         x    STATUS SPEED SPEED       MPW       START STOP COVER    OOO    4  Control panel    Buttons application     EN START Start of centrifuging  Oo  STOP Cancel of centrifuging    COVER Open the lid  TIME Set up runtime  in 1 minute step   L H SPEED Change speed mode       15    6  Centrifuging    6 2 HOLD mode    continuous mode  centrifuging until  pressing STOP     HOLD mode    In order to run the HOLD mode  set the  time value to           6 3 Centrifuging finish         WAYS OF FINISH THE CENTRIFUGING         After reaching the preset working time centrifuging is stopped       Before the expiry of the set time  you can interrupt the cycle by pressing  the STOP       16    6 4 Signalling states of centrifuge    SIATUSISPEED    STATUS SPEED    STATUS SPEED    S ATUSISPEED    17    6  Centrifuging    Cover is locked    Cover is open    Rotor is running  speed is increasing or  constant      mark turn counter clockwise    Rotor is running  breaking      mark turn  clockwise    Rotor is running  time is counting     dot  next to time indication is blinking    Cycle cancelled or end of centrifuging     a  dot next to the time indication stops  flashing  after braking the rotor  four  short beeps are issued       6  Centrifuging      Centrifuging in high rotational speed      Centrifuging in low rotational speed       Centrifuge has memory for remembering the last settings even after turn off and turn on the  centrifuge    6 5 Maintenance of centrifuging eleme
14. ll test tubes outside the centrifuge     Put on or screw the caps on vessels and rotors  if applicable       n case of centrifuging in an angle rotor  test tubes  buckets  have to be filled properly in    order to prevent from pouring fluids during centrifuging     Centrifuge will tolerate small weight differences occurring during loading of rotors   However it is recommended to equalize vessels loads as much as possible in order to    ensure minimal vibrations during operation  When the centrifuge is started with large  imbalance  the unbalance control system will switch off the drive system and error signal  will be transmitted  On the monitoring panel  error message will be displayed          Threaded parts shall be lubricated with the petroleum jelly       For replacement of the rotor one shall unscrew clamping and then grab the rotor with both  hands at opposite sides  taking it away from drive shaft by pulling it up     5 4 Control device    The microprocessor control unit of the centrifuge ensures broad possibilities of providing   realisation and reading of work parameters     5 5 Set up parameters    Data setting and read out system forms hermetically closed keyboard with distinctly accessible  operation points  Easily readable display signaling time to end of centrifuging and set mode of speed   It make easy for the operator to the registration status of the device   Control system provides possibility of    e set up speed mode      5800 or  L  3500 rpm  MPW 54     
15. ntrifugation  Rotation chamber was made from plastic  Rotors and containers are from aluminum   reductive inserts from the polypropylene     5 2 Control elements            d 1  Control panel    E         2  Power switch  3 3  Hole for emergency lid  5 opening  a 4  Cover  1 5  Glass inspection    1  General view    1  Motor axle       2  Rotor  3  Rotor lid    IN    4  Complete clamp 1    1  Mains socket    2  Fuse socket       3  Back of the centrifuge    2  Angle rotor assembly 12    5  Operating and design    5 3 Rotor and accessories installation      Connect the centrifuge to the mains  master switch on left side of the centrifuge        Open the cover of the centrifuge by pressing the COVER key  Prior to putting the rotor in   one has to check if the rotating chamber is free of impurities  e g  such as dust  glass  splinters  residues of fluids that must be taken away       One shall fit the rotor on the motor shaft driving it home on the cone     Fitting the rotor too shallow will result in lack of identification of the rotor after start of    the centrifuge  displaying the error message and stopping the centrifuge          Screw in the bolt for fixing the rotor  clockwise  and screw it tightly home with the supplied  spanner for the rotor       In case of rotors designed with the cover they must not be used without it  Rotor covers  must be closed exactly  Rotor covers ensure smaller drags of the rotors  proper setting of the  test tubes and airtight sealing       Fi
16. nts       n order to increase durability of gaskets and threaded places  it shall be  lubricated with petroleum jelly  cat  No 17201         Cleaning of the accessories    In order to ensure safe operation one shall carry out in regular way periodical  maintenance of the accessories     Rotors  buckets and round carriers have to withstand steady high stresses  originating from the centrifugal force  Chemical reactions as well as corrosion   combination of variable pressure and chemical reactions  can cause  corrosion or destruction of metals  Hard to observe surface cracks increase  gradually and weaken material without visible symptoms     In case of observation of surface damage  crevice or other change  as well as  the corrosion  the given part  rotor  bucket  etc   shall be immediately  replaced     In order to prevent corrosion one has to clean regularly the rotor with the  fastening bolt  buckets and round carriers     Cleaning of the accessories shall be carried out outside of the centrifuge once  every week or still better after each use     Then  those parts shall be dried using soft fabric or in the chamber drier at ca   50         Especially prone to the corrosion are parts made of aluminium  For cleaning  them one should use neutral agent of pH value 6 8     It is forbidden to use alkaline agent of pH  gt  8     In this way  the useful service life of the device is substantially increased and       18    6  Centrifuging    susceptibility to corrosion is diminished
17. of service shipping    4 1 The contents of the package    Quantity    Cat No    pcs                 Centrifuge MPW 54  Complete clamp    17168  Spanner for the rotor F       17099L    Key for emergency                            1117162   Powercord230V 15V     J 1 f 17866 17867       Fuse WTAT2A250V     17859  Petroleum jelly 20m     0       User manual        2004EN O    4 2 Location       The centrifuge shall not be located near source of heat and shall not  be subjected to direct sunlight     The table for the centrifuge shall be stable and shall have flat levelled  table top     It is necessary to ensure a safety zone of the minimum 30cm round the  centrifuge from every direction     At the change of the place from cold to warm one  condensation of  water will occur inside the centrifuge     It is important then that sufficient time be provided for drying the  centrifuge prior to starting the centrifuge again  min  4 hours         11    5  Operating and design  5  Operating and design    5 1 Description of centrifuge    New generation of MPW MED  INSTRUMENTS laboratory centrifuges is provided with state of   the art microprocessor control systems  very durable and quiet asynchronous brushless motors and  accessories consistent with requirements of the present day user     The centrifuge has rigid self supporting structure  Housing was made of ABS  Cover is fixed on  steel axles of hinges and from the front it is locked with electromagnetic lock blocking opening during  ce
18. otes    3 4 Current protection    The centrifuge is equipped with thermal current protection  Fuse is  situated in the plug in socket unit at back wall of the centrifuge    Supply voltage given on the rating plate has to be consistent  with local supply voltage  MPW MED  INSTRUMENTS laboratory  centrifuges are 151 safety class devices and they are provided  with the three core cable with the plug resistant to dynamic    loadings     Mains socket shall be provided with the safety pin  It is  recommended to install emergency cut out that shall be located  far from the centrifuge  near the exit or beyond the room     Before switching on  check whether the centrifuge is  connected to power supply correctly        3 5 Safety hints    ROTORS MAINTENANCE    Use only accessories in good condition       Protect equipment against corrosion using accurate preventive  HAZARDOUS MATERIALS  Infectious materials could be processed in closed buckets only     It is not allowed to subject to centrifugation toxic or infectious materials  with damaged leak proof seals of the rotor or test tube  Proper  disinfection procedures have to be carried out when dangerous  substances contaminated the centrifuge or its accessories     EXPLOSIVE AND COMBUSTIBLE MATERIALS    It is not allowed to centrifuge explosive and inflammable materials     It is not allowed to centrifuge substances prone to reacting in result of  supplying high energy during centrifugation  The centrifuge can not be  operated in explo
19. service life of the centrifuge specified by the manufacturer amounts to 10  years     After termination of guarantee period it is necessary to carry out yearly technical  inspections of the centrifuge     Manufacturer reserves the right to make technical changes in manufactured  products     Maximum period of storage of not used centrifuge amounts to 1 year  After this  period  a service authorized by manufacturer should carry out technical  inspection of the centrifuge        Safety notes    3 3 Loading the rotor    Fix the rotor on the motor axis firmly   Avoid unbalance     Load opposite buckets with the same accessories     Centrifugation of the test tubes of different sizes     There is a possibility to centrifuge test tubes of different sizes  however  it is  absolutely necessary in such cases that opposite buckets and round carriers be the  same     CORRECT INCORRECT    1115 necessary to insert test tubes symmetrically on the opposite sides      It is necessary to insert buckets symmetrically on the opposite sides       UWAGA    If the centrifuge will run with a large unbalance it will occur dangerous vibrations      Buckets 13080 and 13081 are part of common balanced kit and marked by one number   One should load rotor with buckets from the same kit     FILLING TUBES      Fill test tubes outside the centrifuge       Please pay special attention to the quality and proper thickness of the glass test  tubes walls  Those shall be test tubes for centrifuges        Safety n
20. sion endangered areas    It is not allowed to centrifuge materials capable of generating inflammable  or explosive mixtures when subjected to air        Safety notes    3 6 Maintenance conditions    START UP    Prior to switching the centrifuge on  one shall read all sections of this instruction  carefully in order to ensure smooth operation and avoid damages of this device or  its accessories     In order to protect the centrifuge against unbalance  fill in the test tubes up to the  same weight     TRANSPORTATION      Centrifuge must not be transported with the rotor mounted on the shaft     GENERAL HINTS    One must use original rotors  test tubes and spare parts only     In case of faulty operation of the centrifuge one shall ask for assistance of service  of MPW MED  INSTRUMENTS company or its authorized representatives     It is not allowed to switch the centrifuge on if it is not installed properly or rotor is  not fitted correctly     CENTRIFUGES SUBSTANCES      It is not allowed to exceed load limit set by the manufacturer  Rotors are intended  for fluids of average homogeneous density equal to 1 2 g cm  or smaller when  centrifugation is carried out at maximum or submaximum speed     In order to avoid overloading of the rotor  please observe the maximum load  which is  stored on each rotor  The maximum permissible load is achieved when all the tubes are  filled with a liquid of 1 2 g cm  density    If the density of the centrifuging liquid is greater than 1 2 g cm   
    