        ART-M1 Magnetostrictive Ultrasonic Scaler
         Contents
1.     Do not immerse the ART M1 in water or liquid  If the ART MI has water damage   return the machine to Bonart for servicing    Do not modify the ART M1  Modifications will invalidate the warranty on the    machine as well as invalidate safety codes and endanger the patient and operator     Section III  Precautions    3 A  Precautions for all Ultrasonic Scaler Units and Systems    Ensure sufficient water flow to the scaler tip during use to cool the hand piece and  insert    Take precaution as all ultrasonic scalers produce aerosols that may transmit  contagious diseases    Use inserts manufactured by Bonart for optimal performance and results    Keep the unit away from intense heat  Excessive heat may damage the electronic  components    Switch the water valve off when the ART M1 is not in use    Avoid treating patients with pacemakers due to the magnetic field produced by the  ART M1  The magnetic field may cause monitoring devices to malfunction and    provide inaccurate readings     3 B  Precautions for Ultrasonic Prophylaxis Procedures     Scaling tips  inserts  will wear with use  Inserts with 2mm of wear will lose  approximately 50  of their scaling efficiency  In general  it is recommended that  inserts be discarded and replaced after 2mm of wear to maintain optimal efficiency  and avoid breakage      Avoid direct contact of inserts with the patient s lips  cheeks and tongue      Do not reuse scaling tips  inserts  that are damaged  bent  or reshaped  Discard  immedi
2.   Push the insert firmly into the hand piece  If installed properly  the insert band should  be flush against the top of the hand piece  To remove the insert  pull the insert at the  grip  Be careful not to cut your hand while removing the insert     4  Check the water tube connection to the water supply source and main unit     Section VI  General Description and Information of Parts             Figure 2  ART MI Outlook diagram    1  Main Unit    The main unit generates power and produces a signal that is passed to the hand piece     The hand piece is then powered up and vibrates the installed insert      14     2     Hand piece    The hand piece is a housing for scaling inserts  Inserts are interchangeable and can  be installed with a firm push  To remove the insert  pull from the grip  but be  careful not to puncture or cut yourself  Aside from  Bonart inserts   Dentsply Cavitron and Hu Friedy inserts are also compatible with the ART M1     When the hand piece is powered  the hand piece will begin to heat due to the power  generated by the main unit  To cool the temperature of the hand piece  a stream of  water is released into the hand piece and through the tip  This stream of water is    also used to wash away tooth calculus and tartar     CAUTION   Make sure that new insert is firmly seated  The insert band    should be flush against the top of the hand piece    3     4     CAUTION  Do not depress the footswitch when there is no insert installed   Doing so will cause th
3.  indicators or chemical indicators for    ensuring the efficiency of the sterilization cycle are recommended     NOTE   Routine spore testing will determine the sterile conditions of your office     NOTE   The Sterilization Assurance Level  SAL  of steam autoclave should be 10    According to ISO 13683 Sterilization of health Care products      10     WARNING   High room temperature conditions  improper dilutions  or extended    immersion time in chemical sterilizers can result in damage to the  plastic and other materials of the insert Tips     CAUTION   The use of a dry heat oven  incompatible chemical vapor type    sterilizers or quaternary ammonium compounds must be avoided as  damage can result to the plastic and elastomeric materials     Main Unit   Carefully wipe the Main Unit with absolute alcohol or an Acrylonitrile Butadiene Styrene   ABS  plastic compatible disinfectant to prevent damage to the plastic components of the  casing    Section V  Installation    5 A  General Information    If the installation of your ART M1 system is performed by an unauthorized distributor of    Bonart products  check to see that the requirements below are followed     5 B  Water Line Requirements    e Incoming water supply line pressure to the Scaler must be 25 psi  172 kPa minimum   to 60 psi  414Kpa maximum   If your dental water supply line pressure is above 60  psi  install a water regulator on the water supply line to your ART M1 unit    e After the above installation requirement
4.  is fulfilled  thoroughly flush the water prior  to connecting to the scaler    e Please use the manual shut off valve on the dental water system supply line to    shut off water when the office is unoccupied     5 C  Electrical Requirements    Refer to Section X  Specifications    zd    5 D  Unpacking the Unit    Carefully unpack your ART MI unit and verify that all components and accessories are    included         Figure 1  All components and accessories                            Item Quantity   Item Quantity   1  Main Unit 1  2  Foot Control   Foot Switch 1   3  AC Power Cord Set  1  4  PU Water Tubing 1   5  User Manual  amp  Literature Packet 1   6   ART IF 50 Streamline Universal Plastic Handle Internal Flow Insert  Optional  ART P 10 Straight Angle Universal Metal Handle External Flow Insert  Optional   7  Water Hose  Quick Disconnect Assembly  Optional          NOTE      While unpacking  check your ART M1 unit for any damages  If any damages are found   please contact your dealer immediately  Enter the unit serial number on your warranty card    and mail it within 14 days from the date of purchase     5 E  Power Cord   Power Connection    e Always make sure that the power switch is set to the OFF position before performing    the following tasks   o Plug the detachable AC cord into the back of the unit   o Plug the 3 prong plug into a grounded outlet       42         A N Safety Instructions    A  Grounding   Prior to connecting accessories to the unit  check that th
5.  procedures  WARNING   Alerts users of danger that may cause severe injury when proper    procedures are not followed    e SYMBOLS         e     Compliance with MDD 93 42 EEC    0434    BF equipment    Grounding terminal    Attention  Please read instructions     gt    B    Warnings    Important        Only trained and qualified doctors  dentist  and hygienists should operate  the ART M1   Patients with pacemakers cannot be treated with the ART M1    v wv    Do not immerse the ART MI in water     gt  If patient or operator is pregnant  consult a physician prior to performing  dental scaling for safety purposes    The ART M1 requires a shielded AC power cord       Plug the power cord into a well grounded outlet     v    AN Attention to users with cardiac pacemakers     Patients with pacemakers should avoid treatment with the ART M1  Electronic  appliances including razors  hair dryers  microwave ovens  TV receptors  and other  medical equipment  such as the ART MI  may interfere with the performance of    pacemakers   For more information on this subject  please refer to the following articles        Advances in Cardiac Pacemaker   The New York Academy of Sciences  Vol  167   Article 2  pp  515 1075       Electromagnetic Radiation Interference with Cardiac Pacemaker   U  S     Department of Health  Education and Welfare       The Individual with a Pacemaker in the Dental Environment   Journal of the  American Dental Association  Vol  91  No  6  pp  1224 1229    Please Read Bef
6.  see figure below         4  During treatment  try to keep the angle between the surface of the tooth and scaling  insert at 15 degrees  If the patient experiences any discomfort during treatment  follow    the suggestions below     e Increase the hand piece movement speed on the surface when treating sensitive areas   e Treat less sensitive area first  and then return to sensitive areas     e Ifthe problem persists  reduce the power output intensity of the hand piece     Section VIII  System Maintenance  Daily start up    1  Open the water supply valve    2  Turn the ART MI on using the main power switch  Verify that the LED power indicator   LED  is lit    3  Set the power control knob to full power    4  Turn the water control knob to the maximum setting    5  Hold the empty hand piece upright over a sink or drain    6  Depress the footswitch and flush the water line for at least 2 minutes    7 Place a Bonart scaling insert  or a compatible insert  into the hand piece and adjust the    water level as desired     NOTE  If no water is detected  check the water supply  NOTE  If the power LED is not lit  contact a local authorized agent      18     Preparations for treating each patient  1  Remove  clean  and sterilize all used ultrasonic inserts   2  Make sure the scaling insert tip has been sterilized before treating the next patient     3  Clean and disinfect the surface of the power cord  hand piece  and cable assembly     Daily shutdown  1  Use the power switch to turn th
7. 230V  5   50 60Hz 92VA  24 5KHz  5   for 25KHz Insert   25 60 PSIG 172 414KPa   24 5cm L  x 22 5cm W  x 8 5cm H   4 Kg include hand piece    250 cm   250 cm    0   60    10    90   non condensing     0   60    10    90   non condensing   860 1060 hPa    Please follow county and state regulations for disposal of the ART M1      21     
8. ART M1 Magnetostrictive Ultrasonic Scaler    Instruction Manual    ias Bonart Co   Ltd  Jorgen Kruuse A S  4F 11  No 3 Wuquan Ist Rd   Xinzhuang Havretoften 4  DK 5550 Langeskov   Dist   New Taipei City 24892  Taiwan  R O C  Denmark  Tel  886 2 22983980  Fax 886 2 22983981 C       w 0434    Email  sales bonartmed com       COPYRIGHTO 2013 BONART CO   LTD     BONART Co   Ltd  owns all rights to this publication  This publication shall be used solely as  a reference for operation  maintenance  and repair on BONART equipment  No part of this    document may be reproduced or distributed in any form for any other purpose    In the event of inadvertent or deliberate publication  BONART intends to enforce its rights to  this manual under the copyright law  Reproduction of the materials presented in this manual  without the express written permission of BONART Co   Ltd  is prohibited    The contents of this manual are subject to change without prior notice     PROPERTY OF BONART CO   LTD   ALL RIGHTS RESERVED    ART is a registered trademark of BONART Co   Ltd   amp  its affiliates     Document No  RD 02 01 02 Rev 1 1  07 12   Printed in Taiwan  R O C     Operator Safety    Please read this manual before operating unit     The ART M1 magnetostrictive ultrasonic scaler should be operated  maintained and repaired    by qualified and properly trained personnel   e NOTE  CAUTION  AND WARNING STATEMENTS    NOTE   Provides tips and advice  CAUTION  Provides correct operating or maintenance
9. ars  Do not repeatedly tap or  hold the footswitch down  The empty hand piece will melt     Lubricate the rubber O ring on the insert with water before placing it into the hand    piece  To install the insert  firmly push the insert into the hand piece     Hold the hand piece over a sink or drain with an insert installed  Verify that the water  is reaching the tip     Check your Bonart scaling inserts for wear and replace as needed     Use the foot control to control the water flow to the hand piece  Hold down the foot  control to allow water to flow through  Release the foot control to stop the water    flow     Adjust the water flow using the water control knob to the desired level of water flow  and hand piece temperature  Increase the water flow by turning the water control  knob clockwise  Decrease the water flow by turning the water control knob    counter clockwise     NOTE  Increased water flow results in a cooler hand piece temperature and vice  versa    CAUTION   During operation  a continuous flow of water is required to keep the  hand piece cool       17     Z B  Patient Comfort Considerations      Position the patient comfortably on the patient chair  Adjust the chair angle and position    to access the patient s oral cavity with ease and comfort   2  Put the saliva ejector into the patient s mouth to remove excess saliva and debris     3  Rotate the patient s head gently so that the oral cavity can be easily accessed without    causing discomfort to the patient 
10. ately      Use the long axis of the scaling tip  insert  to wipe accretions from the tooth  Do not    gouge the tooth with the tip     3 C  Precautions for hand piece  In order for the system s hand piece to function consistently and effectively   Accumulated air pressure from the handle  or hand piece  must be released every  10 to 15 minutes on every procedure by doing the following steps   1  Put the insert into the hand piece and make sure it is fully seated   2  Hold the hand piece vertically with the insert tip on top  depress the foot Switch   control      3  Let the water spray out for a minute or so  then start scaling procedure     Section IV  Infection Control    4 A  General Infection Control Recommendations     As with all dental procedures  use standard protective equipment such as face masks  eyewear   face shields  gloves and protective gowns    To ensure safety to the operator and patient  carefully follow the Infection Control Procedures  detailed in Section XII     4 B  Water Supply Recommendations    All dental water supply systems should conform to applicable Centers for Disease Control  and Prevention  CDC  and the American Dental Association  ADA  standards  These  standards should apply to flushing  chemical flushing and general infection control    procedures     4 C  Cleaning and Sterilization     Hand piece    Prior to cleaning  remove the insert from the hand piece  Next  drain any water left in  the hand piece to clear the hand piece of any residu
11. e ART M1 off   2  Close the water supply valve     Daily Maintenance    The suggestions below will help maximize the lifespan of the scaler   A  The ART MI should be used in a well ventilated room    B  Handle the ART MI with care when moving it    C  Before leaving the operation room  make sure that the ART M1 is turned off and the  water faucet is tightly closed    D  If you find that the power output to the hand piece is insufficient  the insert is probably  worn out  Inserts should typically be replaced after six months or when worn out     E  If service is needed  please contact a local authorized agent for technical support     Section IX  Troubleshooting    While Bonart offers repair service  the following trouble shooting procedures can help you    save time   9 A Troubleshooting Guide  Basic procedures     1  Check all lines and connections to and from the system  A A loose plug or connection    is the most common cause of problems   2  Check the setting of all control knobs   System will not operate     LED power indicator is not lit  1  Check that the power control switch is in the ON position  and that the detachable  power cord is fully seated in the receptacle on back of the system   2  Check that the three prong power plug is fully seated in the AC receptacle on the back  of the ART M1  and that the voltage is supported       19      3  Check that the power plug is plugged into the wall outlet     LED power indicator is lit  1  Check that the footswitch connect
12. e accumulated during treatment   Then  clean the surface of the hand piece with antiseptic soap or solution and rinse  thoroughly with water  Afterwards  wipe or spray a chemical disinfectant on the hand  piece  Please try to use a disinfectant that has a 1 6 dilute solution  ie Sporicidin    Disinfectant  to avoid damaging the hand piece     After disinfecting the hand piece  a sterile insert may be re inserted into the hand piece    in preparation for the next patient     At the end of the day  remove the insert from the hand piece and scrub both  the hand  piece and cable  with antiseptic soap  Rinse thoroughly with water     NOTE   Cleaning the hand piece after every patient is advised     WARNING   Do not dip the hand piece and extension cable directly into the  sterilized fluid  Doing so will cause the ART M1 to malfunction     WARNING   Do not leave the disinfectant on the surface of the hand piece and  extension cable any longer than instructed by the manufacturer of  the disinfectant     Insert    Saliva  blood  and other debris will leave a residue on the insert after each use   Therefore  it is necessary to clean inserts after each use  Cleaning can be done  manually by scrubbing the insert with a brush  water  and detergent  Be sure to rinse  thoroughly after scrubbing  Let the insert dry  Then  place the insert in a bag and  autoclave at 275  F  135  C  for 15 minutes  or as recommended by the manufacturer of  your sterilization unit  to sterilize the insert  Bio
13. e hand piece to melt     NOTE   The ART M1 is a stand alone unit and is now only available in 25Khz     Please do not use 30Khz inserts with the hand piece     Footswitch    The footswitch is designed to allow the operator to use two hands while scaling     Main power switch  ON OFF     The main power switch turns the ART M1 unit on and off  When the power is  switched on  the LED power indicator on the main unit should be lit  When the    power is switched off  the LED power indicator should be unlit     AN Make sure the footswitch is in the OFF position before turning  on the ART M1      15     5  Turbo switch    The turbo switch allows you to switch between turbo and scaling mode  Pushing  the button once during scaling mode will trigger turbo mode and light the LED mode  indicator  Under turbo mode  power output increases to make cleaning easier  To  return to scaling mode  push the turbo switch button again  The LED mode    indicator should diminish   6  Water control    The water control knob adjusts the amount of water flowing through the hand piece   To increase the water flow  turn the knob clockwise  To decrease the water flow   turn the knob counter clockwise  Keep in the mind that the water level should be  adjusted not only for cleaning tartar but also for cooling the hand piece during use   Increasing the water level will cool the hand piece faster     WARNING   Do not turn the water knob counter clockwise more than three full  turns  Doing so will damage the in
14. e main unit is plugged into a  grounded wall outlet     B  Main voltage range and fuse   Prior to plugging the AC adapter into the power outlet  check that the voltage is  supported     5 F  Foot Control Cable Assembly Connection  Align the pins of the foot control plug with the receptacle on the back of the device and push  in until the plug is firmly seated     5 G  Water Supply Line Connection    Push the blue water tube  hose  into the stainless steel receptacle until the hose cannot be  pushed any further  Then tighten the screw     Connect the quick connect to the water supply line  The ART MI should be prepackaged  with a male quick connect  If a female quick connect is necessary  please contact a Bonart  distributor to purchase the part    Inspect all connections for leaks    To remove the water line from the ART M1 scaler  first  turn off the water supply or  disconnect the water supply line  Then loosen the screw on the water tube from the  receptacle of the unit and gently pull the water hose out    5 H  Hand piece Cable Assembly Connection   The ART M1 hand piece  attached to the main unit  is designed only for inserts with a    frequency of 25khz  The ART MI is not available in 30khz so please do not try using a  30khz insert with the ART MI     219    5 I  Assembling the Scaler Unit    1  Switch the power of the ART M1 off  The power indicator should not be lit if the  machine is off    2  Plug the power cord of the main unit into a grounded AC power outlet    3
15. minum plates to become more familiar with the    scaling tips before performing on patients     Table of Contents       Copyright  C      7   Quisih pO h utu ias  Operator Safety ee rere CU Rae erty ade  Warnings  shpat Sadun ee a Swed ev ds  Prefage i Rear wR ell be a ee  Contents           RNE NR Peer Tar  section Indications Tor Use    gt  deduce eot eei evel  Section II  Contraindications and Warnings                                 Section II  Precautions     eei eere e erede ees  Section IV  Infection Control                2525 IA  Section V  Installation     quest eecpees dote v Et REESE YE RE SER te s   pXaEEER     Section VI  General Description and Information of Parts                   Section VIL Techniques         j  weit tasse rex vea   t veta  Section VIII  System Maintenance                                           Section IX  Troubleshooting                  22252eeemI  Section X Specifications   a vL       eD o                                         Section XL Disposal                             desee o eed as    Section I  Indications for Use    Ultrasonic procedures     Removal of calculus and plaque during dental prophylaxis   General supra and sub gingival scaling applications   Periodontal debridement for all types of periodontal diseases     Endodontic procedures     Section II  Contraindications and Warnings    Do not use the ART MI for amalgam restorative dental procedures    Do not use the ART MI if the patient or operator is wearing a pacemaker
16. or is fully seated in the foot control receptacle on    the back of the system     System operates   No water reaches the tip     Ensure that the water control knob is turned on or adjusted to allow water to flow   2  Check that water supply control valve s   dental office water supply  is open     3  Check that the external water tube is properly connected     9 B Technical Support and Repairs  For technical support and assistance call 1 888 526 6278 or 626 600 5330 Monday  through Friday  9 00 AM to 5 00 PM  Pacific Time   For other areas  contact your local    Bonart representative     X  Specifications    Classification       Protection against electric shock   Class I       Degree of protection against electric shock   Type BF      Mode of operation   Continuous  e    Medical device directive   Ha    Standards Followed  18013485  Medical Devices  Quality Management Systems  Requirements for  regulatory purposes  93 42 EEC  MDD CE Directive  EN60601 1    Medical Electrical Equipment  General Requirement for Safety  EN60601 1 2  Medical Electrical Equipment  Electromagnetic Compatibility     Requirements and Tests      20      Specifications  e Power supply    e Working frequency   e Water supply   e Dimension   e Weight   e Hand piece Cable   e Footswitch Cable  Operation environment  e Temperature   e Relative Humidity  Transportation and Storage conditions  e Temperature   e Relative Humidity   e Atmospheric pressure    Section XI  Disposal    115V  5     50 60Hz 92VA  
17. ore Operating         Make sure the device is connected to a grounded outlet   Failure to meet this requirement may severely damage the    user and equipment         Place the device on a flat and stable surface  Placing the  device on a tilted or unstable surface may degrade    performance and also cause damage to the device         Do not modify this device  Modifications to the device  will invalidate the safety codes and warranty while    endangering the patient and operator         Do not place heavy objects on top of the power cord and    keep the power cord away from intense heat         Unplug the device and call Bonart for service and  instructions if you observe any abnormalities while    operating the device        Preface    About the ART M1 Magnetostrictive Ultrasonic Scaler    The ART MI Magnetic Ultrasonic Scaler is manufactured by Bonart Co   Ltd   an ISO  13485 2003 company  The ART MI is also CE 0434 certified in compliance with the  applicable requirements of the Council Directive 93 42 EEC  The reviewing council also  classifies the ART M1 in accordance with EN60601 1 and ENO0601 1 2 safety    requirements   How it works    The ART M1 consists of two parts  The main electronic system and the scaling hand piece   The main electronic system utilizes two closed loops  One loop provides automatic  tuning to match the resonance of the insert  The other loop will control the tip stroke based    on various working conditions     The scaling hand piece utilizes a 
18. pulsing magnetic field  This magnetic field is applied to  a metal  stack  which flexes and moves an installed scaling tip  insert  approximately  25 000 30 000  cycles per second in an elliptical pattern based on the amount of power  used  Thus  more power results in greater tip displacement  which allows for more robust  scaling  However  the vibrating metal stack generates heat and requires adequate  amounts of fluid to cool the hand piece and tissues during treatment     When the two parts of the ART MI are placed together  the properties of the main unit and  scaling hand piece enable the ART M1 to    power away  the heaviest calculus deposits   In addition  tips may be rotated 360 degrees within the hand piece to relieve cord tension  and alignment  As a result  the ART MI provides maximum comfort to the patient and    ease of use to the operator     Technical Support    For technical support and repairs in the U S   call 1 888 526 6278 or 626 600 5330   Monday through Friday  9 00 AM to 5 00 PM  Pacific Time   For other areas  contact your    local Bonart representative   Supplies  amp  Replacement Parts    To order supplies or replacement parts in the U S   please contact one of our authorized    distributors listed on our website  www bonartmed com     CAUTION   Since the scaling tip  insert  is made of stainless steel  avoid direct contact  with teeth  Direct contact will cause damage to teeth enamel  For optimum performance   we suggest practicing on a model or alu
19. ternal water line needle     WARNING   If needed  gently adjust the water control knob to prevent any  serious damages to the water valve     When used for the first time  turn the knob clockwise to allow   N water to flow through  The knob is set to the off position  when assembled by the manufacturer     7  Power control    The power control adjusts the power intensity to the desired level based on the  cleaning procedure  When turned clockwise  power output increases  When    turned counter clockwise  power output decreases   8  Water in Connector    The water in connector is the contact point for the water tube and the main water  supply line  Use either a male or female quick connect to connect the water tube  to the water supply  The ART MI comes with a male quick connect attached to  the water tube     9  AC Line Socket Seat    The AC line socket seat is the connection point between the power cord and the    main unit      16     Section VII  Techniques    Z A  Performing Ultrasonic Scaling Procedures    l     Use purified or distilled water to prevent infection when patients experience tissue    laceration during treatment  Keep the power cord tidy to avoid tripping and other accidents     Position the footswitch in an easily accessible spot for the user  Keep the footswitch    cord tidy to avoid tripping and other accidents     Hold the empty hand piece in an up right position when testing to see if the scaler is  working properly  Tap the footswitch until water appe
    