        Model: HIT-1B USER`S MANUAL
         Contents
1.                                                                          3   3  Theory of Operation                                                       5  3 1 Functional Blocks                                                        5  3 1 1 Preamplifier                                                         5  3 1 2 Microcontroller                                                         3 1 3 Loop Driver                                                         6  3 1 4 Communications Interface                                        6   3 2 System Response Time                                                  6   4                                                      1                           7  4 1 Configuration Commands                                               8  42 Monitoring Diagnostic Commands                                  10  4 3 Default Configuration                                                  11  5  Installation                                                                  13  5 1 Typical Connections                                                   13  5 2 Communications Connections                                        16  5 3 Wiring                                                                    16         297          This page intentionally left blank        HP 297       Introduction 1    1  INTRODUCTION    The HIT 1  Hoffer Intelligent Transmitter  is a loop powered intelligent  flow rate transmitter  The transmitter is compatible with all
2.             OPTION      GENERAL PURPOSE   2 6 L X 2 6 H X 2 6 W MINIMUM MOUNTING SPACE      D  2  LONG DIN RAIL MOUNT SINGLE UNIT   UP TO 20 HIT 1B UNITS CAN BE MOUNTED ON A SINGLE RAIL   ADD 2  PER UNIT      3 0  MEETS CLASS 1  DIV  1 AND 2  GROUPS C  D  CLASS II  GROUPS E  F  G  NEMA 4X WITH       RING  CERTIFIED CSA  UL  BODY KILARK  GECCT 3  STOCK  200 0945  FLAT COVER  STOCK  200 0533  KILARK GECBC     3B O  MEETS CLASS L DIV  1 AND 2  GROUPS A  B  C  D  CLASS 1  ZONES 1 AND 2  GROUPS IIB   H2        CLASS     DIV  1 AND 2  GROUPS E  F  G  CLASS     NEMA 3  4  7  B  C  D   9  E  F  G   CERTIFIED CENELEC  CSA  UL  FM  BODY KILARK  HKB  STOCK  200 0406  FLAT COVER  STOCK  200 0773  KILARK  HKB B                       MEETS CLASS L DIV  1 AND 2  GROUPS A  B  C  D  CLASS 1  ZONES 1 AND 2  GROUPS IIB   H2        CLASS     DIV  1 AND 2  GROUPS E  F  G  CLASS     NEMA 3  4  7  B  C  D   9  E  F  G   CERTIFIED ATEX  CENELEC  CSA  UL  FM  BODY COVER STOCK  200 2006    LINEARIZED ANALOG OUTPUT  MODEL        1         8              OPTION   7  4      20      UP      20 POINTS   ACCURACY    0 02  OF FULL SCALE   INPUT POWER  MODEL        1            C       OPTION   D  8 TO 24 VDC LOOP POWERED    AC  115 220 VAC 50 60 HZ EXTERNAL POWER  USE MODEL ACC39B  SPECIAL FEATURES  MODEL                          D   OPTION   CE  CE MARK REQUIRED FOR EUROPE    SP  ANY SPECIAL FEATURES THAT ARE NOT COVERED IN THE    MODEL NUMBER USE A WRITTEN DESCRIPTION OF THE  SP     HP 297    Specification
3.    Flow Calculation Method  Sets the flow rate  calculation method to Average K Factor or  Linearization Table  Frequency vs K Factor Table   Set  to    0    for Average or    1    for Linearization     Average K Factor  Set to the average K factor from  the meter calibration data  This value is used in the flow  rate calculation when the FC command is set to     Average        Number of Points  Number of Frequency K factor  pairs in the Linearization table  Maximum 20     Frequencies 01 20  Frequency values for the  Linearization table  Maximum 20  Minimum number  specified by NP command     K Factors 01 20  K factor values for the Linearization  table  Maximum 20  Minimum number specified by NP  command     Configuration 9    The following table lists the various configuration commands  both  read and write  available for the HIT 1 transmitter along with the  response from the HIT 1 for each command                                   Description Read Write Response  Command Command  Tag Number DN lt CR gt  DN  lt value gt  lt CR gt    TAG NUM      value     20mA Maximum AF lt CR gt  AF  lt value gt  lt CR gt    20 mA Flow      value     Flowrate  Max Sample Time      lt      gt    NB  lt value gt  lt CR gt    MAX M TIME       value     Correction Factor CF lt CR gt  CF  lt value gt  lt CR gt    CORR FACT      value     Flowrate Units      lt      gt    FM  lt value gt  lt CR gt    FLOW UNITS    where value is 0  1       value     Where value is  2  or 3  SEC  MIN  HR  or  DA
4.  00  F08 500 00  F09 600 00  F10 700 00  F11 800 00  F12 900 00  F13 1000 0  F14 1100 0  F15 1200 0  F16 1300 0  F17 1400 0  F18 1500 0  F19 1600 0  F20 1700 0     01 100 00  K02 100 00         100 00  K04 100 00  K05 100 00  K06 100 00  K07 100 00  K08 100 00       HP 297    12 Configuration    HP 297                                     FIELD Value  KO09 100 00  K10 100 00       100 00  K12 100 00  K13 100 00  K14 100 00  K15 100 00  K16 100 00    17 100 00    18 100 00    19 100 00  K20 100 00          Installation 13    5  INSTALLATION    The HIT 1B has an input conditioning circuit which accepts signals  from most frequency or pulse producing flowmeters     The input will interface directly to         Turbine Flowmeters       Reed Switches       Logic Signals       Open Collector Outputs    The input is protected for over voltage up to 50 volts     The signal conditioning circuit is configured for the selected signal type  by appropriately setting the input selection DIP switches found on the  HIT 1B signal input board  see following for switch settings   When  interfacing with a MAG coil or dry contact  reed switch  signal the  preamp input sensitivity is adjustable by setting the DIP switches S1 3  and S1 4 according to the following table        22       off off  40mV p p on off  100mV p p off on  160mV p p on on       The transmitters are shipped from the factory with the sensitivity set to  the maximum setting  22mV     p  unless otherwise noted     5 1 Typical Conne
5.  Hoffer  turbine flowmeters and the H O G  series positive displacement  flowmeters     The transmitter can accept a low level frequency signal  a pulse signal   or a contact closure on the input and provides a linearized 4 20 mA  current output which is proportional to the flow rate     The HIT 1 is fully configurable with K factors  frequencies  and  timebase being configured via the RS232 communications port  The  configuration and monitoring can be performed using any PC based  communications program  e g   HyperTerminal  or ASCII terminal     The transmitter is housed in an aluminum enclosure and can be  mounted directly on a flowmeter or wall mounted using an optional  NEMA 4X or EX enclosure  An optional DIN rail mounting bracket is  also available for mounting on a standard DIN rail     This transmitter is designed to conform to the EMC Directive of the  Council of European Communities 89 336 EEC and the following  standards     Generic Emission Standard EN 50081 1  Residential  Commercial  amp  Light Industry Environment     Generic Emission Standard EN 5008 1 2  Industrial Environment     Generic Immunity Standard EN 50082 1  Residential  Commercial  amp  Light Industry Environment     Generic Immunity Standard EN 50082 2  Industrial Environment     HP 297    2 Introduction    11 Model Number Designation    MODEL HIT 1B          B           0      ENCLOSURE STYLE  LINEARIZED ANALOG OUTPUT  INPUT POWER   SPECIAL FEATURES          ENCLOSURE STYLE  MODEL HIT 1B        
6. ERTY  AND INJURY TO OR DEATH OF ANY PERSONS  HFC NEITHER ASSUMES  NOR AUTHORIZES ANY PERSON TO ASSUME FOR IT ANY OTHER LIABILITY IN  CONNECTION WITH THE SALE OR USE OF HFC S GOODS  AND THERE ARE NO ORAL  AGREEMENTS OR WARRANTIES COLLATERAL TO OR AFFECTING THE AGREEMENT   PURCHASER S SOLE AND EXCLUSIVE REMEDY IS THE REPAIR AND OR REPLACEMENT OF  NONCONFORMING GOODS AS PROVIDED IN THE PRECEDING PARAGRAPHS         SHALL         BE LIABLE FOR ANY OTHER DAMAGES WHATSOEVER INCLUDING INDIRECT   INCIDENTAL  OR CONSEQUENTIAL DAMAGES              HFC 9907 A    LIMITED WARRANTY POLICY FOR HOFFER FLOW CONTROLS    HOFFER FLOW CONTROLS  INC    HFC   warrants HFC s Precision Series and API  Series of turbine flowmeters to be free from defects in material and workmanship under  normal use and service  only if such goods have been properly selected for the service  intended  properly installed and properly operated and maintained as described in the  turbine flowmeter manual  Reference  turbine flowmeter manual  for specific details  This  warranty shall extend for a period of five  5  years from the date of shipment to the  original purchaser and covers the Precision Series and API Series of flowmeters supplied  with their standard hybrid ceramic ball bearings only  All other HFC products carry a one   1  year warranty  This warranty is extended only to the original purchaser   Purchaser     Purchaser s sole and exclusive remedy is the repair and or replacement of nonconforming  goods as pro
7. Hoffer Intelligent Transmitter  Model  HIT 1B    USER   S MANUAL    HP 297  October 2005    HOFFER FLOW CONTROLS  INC   The Turbine Flowmeter Company  107 Kitty Hawk Lane  P O  Box 2145  Elizabeth City  NC 27906 2145  800 628 4584 252 331 1997 FAX 252 331 2886  www hofferflow com E mail  info hofferflow com       NOTICE       Hoffer Flow Controls  Inc  makes no warranty of any kind with regard to this  material  including  but not limited to  the implied warranties of merchantability  and fitness for a particular purpose     This manual has been provided as an aid in installing  connecting  calibrating   operating  and servicing this unit  Every precaution for accuracy has been taken in  the preparation of this manual  however  Hoffer Flow Controls  Inc  neither  assumes responsibility for any omissions or errors that may appear nor assumes  liability for any damages that result from the use of the products in accordance  with information contained in the manual     HOFFER FLOW CONTROLS  policy is to provide a user manual for each item  supplied  Therefore  all applicable user manuals should be examined before  attempting to install or otherwise connect a number of related subsystems     During installation  care must be taken to select the correct interconnecting wiring  drawing  The choice of an incorrect connection drawing may result in damage to the  system and or one of the components     Please review the complete model number of each item to be connected and locate  th
8. Y   Flow Calculation FC lt CR gt  FC  lt value gt  lt CR gt                         Method where value                value      1  Where value is AVG or  LIN   Average K Factor AK lt CR gt    AK  lt value gt  lt CR gt _  AVG KFAC      value     Number of Points in NP lt CR gt  NP  lt value gt  lt CR gt    NUM PTS      value     Frequency vs  K Factor table  Frequency 1 20 F   lt CR gt    Fi    lt value gt  lt CR gt _  FREQ        value     K Factor 1 20 K   lt CR gt    K    lt value gt  lt CR gt    K FACT         value       For example                 To set the maximum flow rate for the 20mA output to 821 gal min     Send from the terminal     AF 821 lt CR gt                      responds     20 mA Flow  821     Send from the terminal     FM 1 lt CR gt                        responds     FLOW UNITS   MIN       HP 297    10 Configuration    4 2    Command    Monitoring Diagnostic Commands    DA    OM    OI    OF    MO    AA    Description    Dump All  Sends all the configuration settings to the  attached terminal     Output Maximum  Sets the output current to 20mA   After one minute resets to the calculated output current     Output Minimum  Sets the output current to 4mA   After one minute resets to the calculated output current     Output Frequency  Sets the output current to the  calculated current   Default mode     Medium Output  Sets the output current to 12mA   After one minute resets to the calculated output current     Continuous Display  Continuously sends the cal
9. ctions    The following figures illustrate the typical input and output connections  for the various pickup types     HP 297    14 Installation                                                                                                                                                                                                                                        MAG Coil Input           LJ bs      gt  4 20mA        8 24Vdc  D  4 20mA      O  CONTACT         MAG    SIG    s  COIL     N  SIG     Switch Settings q R   sw sw2 N  4                     LOAD  1    1 4  TTL Redi Pulse Pick up  QV  ere       4 20mA             4 20mA                                White  C  SIG    a       Black  B        2   WI SIG        REDI PULSE  COIL     R               N  x LOAD  TTL Pulse    51    Switch Settings                J                                    297                Installation          8 24Vde                      Reed Switch Input            gt  4 20mA             4 20              CONTACT  REED  SWITCH    SIG         SIG       Switch Settings    SW1    SW2                          s                     INPUT    RS232                   15                   LOAD             HP 297    16 Installation    5 2 Communications Connections    Communications with the HIT 1B using a terminal or a PC using the  Windows configuration program requires the use of the HOFFER  HIT 903 Communications Cable  The communication circuits of the  HIT 1B are externally powered and the po
10. culated  frequency and flow rate values to the attached terminal     The following table lists the various monitoring diagnostic commands  available for the HIT 1 transmitter along with the response from the  HIT 1 for each command     Description Read Response    Command       Continuously display AA lt CR gt                       Frequency and Flowrate RXXXXX XX       Dump Configuration DA lt CR gt    Dump settings to    terminal        Set Output to maximum      lt      gt    Output   20 mA       Set Output to minimum OI lt CR gt  Output   4 mA       Set Output to Frequency OF lt CR gt  Output   freq     scale    scale       Set Output to 12 mA      lt      gt    Output   12 mA             To terminate or interrupt any Monitoring Diagnostic command send  another command other than the one currently in use     HP 297    Configuration 11    43 Default Configuration    The HIT 1 is configured by the factory before shipment  If the unit is  purchased with a Hoffer Flowmeter or if calibration data and  configuration information is supplied  the unit is configured as  appropriate  If the unit is shipped without specific configuration  information  the unit is configured with the following default values                                                                                                                       FIELD Value  DN 00000000  AF 500 00  NB 1 00  CF 1 00  FM 0  FC 0  AK 100 00  NP 20    01 10 00  202 50 00  F03 90 00  F04 100 00  F05 200 00  F06 300 00  F07 400
11. e appropriate manual s  and or drawing s   Identify all model numbers exactly  before making any connections  A number of options and accessories may be added  to the main instrument  which are not shown on the basic user wiring  Consult the  appropriate option or accessory user manual before connecting it to the system  In  many cases  a system wiring drawing is available and may be requested from  Hoffer Flow Controls     This document contains proprietary information  which is protected by  copyright  All rights are reserved  No part of this document may be photocopied   reproduced  or translated to another language without the prior written consent  of Hoffer Flow Controls  Inc     HOFFER FLOW CONTROLS    policy is to make running changes  not model  changes  whenever an improvement is possible  This affords our customers the  latest in technology and engineering  The information contained in this  document is subject to change without notice     THIS WARRANTY IS EXPRESSLY IN LIEU      ALL OTHER WARRANTIES  EXPRESSED OR IMPLIED  INCLUDING ANY IMPLIED WARRANTY OF  MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE  HFC SHALL  NOT BE LIABLE FOR ANY LOSS OR DAMAGE RESULTING  DIRECTLY OR INDIRECTLY   FROM THE USE OR LOSS OF USE OF THE GOODS  WITHOUT LIMITING THE  GENERALITY OF THE FOREGOING  THIS EXCLUSION FROM LIABILITY EMBRACES THE  PURCHASER S EXPENSES FOR DOWNTIME OR FOR MAKING UP DOWNTIME  DAMAGES  FOR WHICH THE PURCHASER MAY BE LIABLE TO OTHER PERSONS  DAMAGES TO  PROP
12. nctional Blocks    3 1 1 Preamplifier    The Preamplifier accepts the input from the flowmeter  The  Preamplifier applies amplification  low pass filtering  and wave   shaping to the input signal  The wave shaping function converts  the signal into a square wave before sending it to the  Microcontroller     3 1 2 Microcontroller    The Microcontroller accepts the square wave output of the  preamplifier and performs all of the calculations that are required  to control the Loop Driver  After measuring the frequency of the  square wave  the Microcontroller uses the following equations to  compute the flow rate and current     Jrequency              flowrate    Kfactor    HP 297    6 Theory of Operation    Where   Kfactor   Is dependent on the Flow Calculation Method setting and  is either the Average K Factor or the Linearized K   Factor from the Frequency   K Factor table     FM   Is the Flow rate Units setting of 0  1       2  Where    0    is  for Seconds     1    is for Minutes  and    2    is for Hours   CF   Is the Correction Factor setting   flowrate    current   4mA     16mAx    Where   AF   Is the 20 mA maximum Flow rate value     If the calculated flowrate is greater than the AF setting  the  current will be set to 24mA to signal an    Over range    condition   After calculating the current  the Microcontroller digitally sends  the current information to the Loop Driver     3 1 3 Loop Driver    The loop driver uses the digital information sent to it by the  Microcontr
13. ods will be disposed of in HFC s  discretion     HFC does not warrant the product to meet the requirements of any safety code of any  state  municipality  or other jurisdiction  and Purchaser assumes all risk and liability  whatsoever resulting from the use thereof  whether used singlely or in combination with  other machines or apparatus     This warranty shall not apply to any HFC product or parts thereof  which have been  repaired outside HFC s factory or altered in any way  or have been subject to misuse   negligence  or accident       have not been operated in accordance with HFC s printed  instructions or have been operated under conditions more severe than  or otherwise  exceeding  those set forth in the specifications        FOR NON WARRANTY REPAIRS OR CALIBRATIONS  consult HOFFER FLOW CONTROLS  for current repair calibration charges  Have the following information available BEFORE  contacting HOFFER FLOW CONTROLS     1  P O  number to cover the COST of the repair calibration     2  Model and serial number of the product  and  3  Repair instructions and or specific problems relative to the product           HFC 9907 A    CONTENTS    1                                                                                    1  1 1 Model Number Designation                                            2   2  Specifications                                  1                        3  2 1 General                                                                     3  2 2            
14. oller to set the current of the loop  The Loop Driver  supplies power to both the Preamplifier and Microcontroller     3 1 4   Communications Interface    The Communications Interface provides an RS232C port to the  Microcontroller  The power for the Communication Interface is  provided by the external terminal device  The Communications  Interface is used to configure and trouble shoot the transmitter     3 2 System Response Time    The output current response time to flow rate changes depends on the  frequency of the input signal  The response time varies from 0 25  seconds to the Maximum Sample Time MST  setting  The MST  setting is adjustable from 1 to 10 seconds  If the input frequency is  greater then 4 Hz  the output is updated every 0 25 seconds regardless  of the MST setting  When the flow rate drops to    0     the output drops  to 4mA after the time set by MST  If the input frequency is less than 4  Hz  the output update time is equal to the period of the input signal   1              1 second  2Hz   0 5 second      The MST setting should be  set to a number higher than the period of the lowest input frequency     HP 297    Configuration 7    4  CONFIGURATION    The HIT 1 is configured using either a PC running Terminal Emulation  software or a dumb ASCII terminal  The PC terminal RS232C serial  port configuration must be set as follows     Baud rate  2400    Data bits  8  Parity  none  Stop bits  1    Handshaking  none    The transmitter configuration is set and o
15. r displayed by sending a series  of character commands via the communications port to the unit     There are two types of commands  READ and WRITE  READ  commands   lt command gt  lt CR gt   upload data from the unit to the  terminal and WRITE commands   lt command gt   lt value gt  lt CR gt    download data from the terminal to the unit  Each command is  terminated with a Carriage Return  CR   If an invalid command is sent   the unit will respond with the message    Invalid Command     or will  return the complete command list  The complete command list is  returned anytime a single Carriage Return  CR  character is sent to the  unit     The following sections discuss the various commands available for the         1          297    8 Configuration    4 1    Command    HP 297    Configuration Commands    DN    AF    NB    CF    FM    FC    AK    NP    FHH    K      Description    Tag Number  User defined ID number  Maximum  999999     20mA Maximum Flowrate  Flow rate for 20mA  output     Maximum Sample Time  This is the maximum time   in seconds  that the HIT 1 samples the input frequency  before returning zero  Output   4 mA   Default value    1 second  Maximum value   10 seconds  Recommend  changing only for minimum input frequencies of less  than 1 0 Hz     Correction Factor  User defined factor to multiply the  flow rate calculation  Default value   1     Flowrate Units  Sets the flow rate time units     0    for  SEC     1    for MIN     2    for HR  or    3    for DAY  
16. s 3    2  SPECIFICATIONS    2 1    2 2    General    Input     Frequency Range   DC power    Load Resistance   Output     Output Resolution     Accuracy     Temperature Drift     Linearity   Update Time   Linearization   Timebase     Scale    Communications   Physical   Temperature     Packaging     Dimensions for  General Purpose     AC or DC signal 10       to 50         Magnetic coil  Pulse  Open Collector  Dry  Contact     0 2 to 3500 Hz    8 to 24 Vdc  loop powered    650 O   24             Linearized    4 20 mA proportional to flow  rate     Over Range    capability  24 mA    16 bit  0 2 WA    0 02  of full scale   20  C    40 ppm   C     0 05  of full scale    0 25 to 10 seconds    Twenty point K factor table    Rate can be configured in units per second   minute  hour  or day    The 20mA point is configurable    RS232 for configuration and monitoring     Operating   40  C to 80  C    General Purpose  Explosion proof   amp  DIN  rail    1 47   38 mm  wide x 2 55   65 mm  long x  1 90   49 mm  tall    mounting footprint 2 535     65 mm  x 2 550      65 mm      HP 297    4 Specifications          This page intentionally left blank        HP 297       Theory of Operation 5    3  THEORY OF OPERATION    The HIT 1 consists of four functional blocks  These functional blocks  are the Preamplifier  Microcontroller  Loop Driver          Communications Interface     Hoffer s Intelligent Transmitter HIT1    POWER                        Flowmeter input Microcontroller        3 1 Fu
17. vided in the following paragraphs     In the event Purchaser believes the Hoffer product is defective  the product must be  returned to HFC  transportation prepaid by Purchaser  within the appropriate warranty  period relative to the product  If HFC s inspection determines that the workmanship or  materials are defective and the required maintenance has been performed and  has been  properly installed and operated  the product will be either repaired or replaced  at HFC s  sole determination  free of additional charge  and the goods will be returned  transportation  paid by HFC  using a transportation method selected by HFC     Prior to returning the product to HFC  Purchaser must obtain a Returned Material  Authorization  RMA  Number from HFC s Customer Service Department within 30 days  after discovery of a purported breach of warranty  but not later than the warranty period   otherwise  such claims shall be deemed waived  See the Return Requests inquiries Section  of this manual     If HFC s inspection reveals the Hoffer product to be free of defects in material and  workmanship or such inspection reveals the goods were improperly used  improperly  installed  and or improperly selected for service intended  HFC will notify the purchaser in  writing and will deliver the goods back to Purchaser upon receipt of Purchaser s written  instructions and agreement to pay the cost of transportation  If Purchaser does not respond  within thirty  30  days after notice from HFC  the go
18. wer for these circuits is  supplied by the RS232 serial port of the computer terminal  The serial  port configuration must be set to the following     Baud rate  2400    Data bits  8  Parity  none  Stop bits  1    Handshaking  none                                                                                        DB9  cil C X         NS           52716 DSR    1            2        mn V RTS       3          r    4     amp       CTS    5         4   O     6       19         SIG        5       Hoffer HIT 903 Communications Cable    5 3 Wiring    When connecting the HIT 1  it is good practice to use shielded cable   The shield should be connected to earth ground near the instrument   The other end of the shield should not be connected     In order to comply with the requirements for Electromagnetic  Compatibility  as per EMC Directive 89 336 EEC of the Council of  European Community  this wiring practice is mandatory     HP 297    
    