Home

OMICRON CPC100 Primary Test System Specifications

image

Contents

1. OMICRON World Leader in Innovative Power System Testing Solutions PRIMARY HT ES TIN G L a DAD syn ny wis j i n i A nn ai ily I JN Company Profile CUSTOMER AND COMPANY PROFILE About our customers subjects covered in other catalogs Typical customers of our CP product line which represents our products for primary testing are engineers and technicians working mainly in commissioning maintenance testing and test fields of e Power transformers e Current transformers e Voltage transformers e Circuit breakers and other electrical power system equipment in utilities and industry Our test equipment is also used for automated measuring of resistances contact resistances winding resistances grounding resistances cable impedances power factor dissapation factor and capacitance but also for single phase testing of primary and secondary protective relays I gt V gt V lt or frequency relays OMICRON electronics is an international company providing innovative solutions for primary and secondary testing Combining innovation leading edge technology and creative software solutions OMICRON s sales have earned world leader status for OMICRON within this niche market With sales in more than 100 countries offices in Europe the United States and Asia and a worldwide network of distributors and represen
2. 6 7 INTERFACE ELEMENTS 6 A or 130 V AC Output Current Output 6 A DC Current Measuring Input AC or DC Voltage Measuring Input 300 V AC Low Level Voltage Measuring Input 3 V AC Voltage Measuring Input V DC 2 wire resistance measurement Binary input for potential free contacts or voltages up to 300 V DC See page 7 for details on outputs and inputs Corot 8 Safety key lock If locked the quantities which are currently put out are frozen The unit does not accept any commands except for an Emergency stop 9 Signal Lights Green light indicates a safe operation whereas red light indicates an operation with hazardous voltage and or current levels at the outputs 10 Emergency stop button 11 Keys for quick selection of your application 12 Keys for quick selection of your desired view 13 LCD monitor 14 Soft touch keys changing according to the selected application 15 Keys for selecting stacked test cards 16 Numerical keyboard 17 Advanced jog dial handwheel with click Enter function The handwheel allows for navigation within test cards within other views or across views and it also allows for the entering of values In entering mode the wheel s adaptive acceleration function will increase decrease the input value in ever bigger steps if the wheel is turned fast Slow turning will increase decrease the value in ever smaller steps 18 Up down keys for navigation and entering
3. 70 C 4 158 F Rel humidity 5 95 non condensing Operating temperature Storage temperature Humidity range Shock IEC68 2 27 operating 15 g 11 ms half sine Vibration IEC68 2 6 operating 10 150 Hz 2g EMC EN 50081 2 EN 55011 EN 61000 3 2 FCC Subpart B of Part 15 Class A EN 50082 2 IEC 61000 4 2 3 4 8 CE conform 89 336 EEC Safety EN 61010 1 EN 60950 EN 50191 IEC 61010 1 Produced and tested in an EN ISO 9001 certified company Weight 29 kg 64 Ibs robust case with cover 468 x 394 x 233 mm 18 6 x 15 5 x 9 2 W x H x D with cover without handles Dimensions CPC 100 Applications APPLICATIONS The software described in the Software section runs in the embedded processor within the CPC 100 and can be operated with front panel control No additional PC is required Quick Manual Control Quick Testing almost any function using direct manual control of the current and voltage outputs and measurements This function allows any manipulation of the selected output range of the CPC 100 All output terminals available on the CPC 100 can be selected and controlled Measuring two quantities in two columns such as AC or DC voltages in amplitude and phase angle AC or DC currents in amplitude and angle or the frequency of any selected output is easily done by selecting the desired quantities Quick also automatically performs calculations derived from the quantities describ
4. A oa queria piai Ly Transferred measurement results gt CPC 100 Report Test Preparation REPORTS PREPARING TESTS OFFLINE B CPC100 E Substation Alach HLJ Substation Kaus GHI Feeda ED Feda 2 L CTY Limi T CPC Explorer The CPC Explorer is a PC software shipped with the CPC 100 The unit provides internet connectivity through an Ethernet connection both in a network and from a directly connected PC From within CPC Explorer the following is possible EE CTI LS e Reporting E CTZ LU mmal e Downloading data for archiving and further processing on the PC E C22 e Uploading data for re using pre prepared templates e Easy software upgrading e Service and maintenance remote trouble shooting Automatic saving of all results or test parameters All test results are automatically saved and stored with the test This data can be organized according to the user s needs A Microsoft Windows Explorer like display provides access to stored data and allows for easy management and organization of test data File view Also once tests have been defined satisfactorily they can be saved to serve as templates for new tests This allows easy reproduction of any test at any time no matter how complex Once the brainwork has been done execution at different locations or different points in time become a matter of pressing a button Gat creme fil L
5. CT CT Ratio Burden and Excitation D Ue et bocae PP Cineni clang liee tram EET Isc aoa pi GR Mrusrp Aa POOtEOs foaak Posay K E Merse buden ETRato CTRiaste CTiuden CTEnctstion 4 liati Eina sec ror oe E hiiio Bunei E GRR VA Cot gt Ea TRiato CTiuden CTEacitation ale aad NASRA Vmax 00 Aue S HH a mhz Henane Canc RE oe i Dea Reni ns Ta Save Ae ihe zj Tepla iknee 107m Vkree Eiee fee Assessed nia ii NRE erent eee reer rea lo r oer go Burden CT P1 P2 Burden Burden Note for input current measuring the direct current input or the current probe can be used CT Ratio Burden Tests ratio polarity and burden with direct injection to CT primary current input and measuring of secondary output After entering primary secondary and test current and pressing the Start button the test card measures e Secondary current with magnitude and angle CT angle error e Ratio with error in percent e Polarity on the CT terminals e Connected burden in VA and power factor cos Duration of the test 3s including automatic reporting Output up to 800 A 2000 A AC Input up to 10 A AC 3 V or 300 V via current clamp CT Burden Measures connected CT bu
6. Impedance Measuring 17 Reports Preparing Tests Offline 70100 aces E E E T TD1 Tangent Delta Power Factor Testing 12 CP TD1 Technical Data CPC 100 with CP TD1 CU1 Line amp Ground Impedance Testing 19 CP CU1 Technical Data CP CU1 and CP GB1 Other Accessories Sy saaevudeuseckssedwageassacnsvai E O 20 Oil Test Cell Compensating Reactor 2 Polarity Checker Current Booster 23 Package Options caent Tanstonner Tesina Ata AA 24 CT Analyzer CT1 Technical Data suber Feauena sponse Ane Se een ee 25 FRAnalyzer FA1 Technical Data info obmmeter cP son nec nceeneccescencsoneccenenenenesaeneneresasnssnssssesesessesey 26 CPM 500 CPM Remote Control Accessory 27 Contact Addresses ISO 9001 certified CPC too OVERVIEW Multifunctional Primary Test System for Substation Commissioning and Maintenance This world wide unique system allows for automated testing of power transformers CTs VTs resistance testing and more Providing up to 800 A or 2000 A with current booster and 2000 V with a frequency range of 10 400 Hz it also comes complete with an integrated PC Its software routines test a wide range of substation equipment and automatically create customizable reports The compact design 29 kg 64 Ibs and the innovative software save testing time and minimize transportation costs Analog voltages and currents can be measured with very high precision Its Q meters offer ranges from uQ to kQ
7. 0 10 0 10 0 20 0 05 0 05 0 10 30 V 0 10 0 10 0 20 0 05 0 05 0 10 V1 AC 500 kQ 3V 0 20 0 10 0 20 0 10 0 05 0 10 300 mV 0 30 0 10 0 20 0 15 0 05 0 10 BIV 0 05 015 0 20 0 03 0 08 010 v2 Ae 10 M 300 mv 015 015 020 0 08 0 0 0 109 30 mV 020 0 50 030 010 0 25 0 159 10V 0 05 0 15 0 03 0 08 1V 0 05 0 15 0 03 0 08 VDC 500 kQ 100 mV 0 10 0 20 0 05 0 10 10mV 0 10 0 30 0 05 0 15 e Automatic range switching Galvanically separated potential groups AC DC V1 amp V2 V DC AC frequency range 15 400 Hz Protection of AC DC input 10 A FF fuse Binary input for dry contacts or voltages up to 300 V DC see 7 on page 5 Toggling with potential free contacts or voltages of up to 300 V gt 100 kQ Trigger criteria Input impedance Response time 1 ms Q meter DC Generator Output Current outputs Range Amplitude tmax Vmax Powelmax f 0 800 A 5S 6 0 V 4800 VA 15 400 Hz 800 A AC 0 400 A 8 min 6 4 V 2560 VA 15 400 Hz 0 200 A gt 2i 0 5 V 1300 VA 15 400 Hz 6 A 0 6 A gt 2h 55V 330 VA 15 400 Hz 3 A 0 3 A gt 2h 110 V 330 VA 15 400 Hz 0 400 A 2 min 6 5V 2600 W DC 400 A DC
8. 2006 20 F Tower 2 The Gateway Harbour City Kowloon Hong Kong S A R Phone 852 2634 0377 Fax 852 2634 0390 info asia omicron at www omicron at North and South America OMICRON electronics Corp USA 12 Greenway Plaza Suite 1510 a Houston TX 77046 USA Phone 1 713 830 4660 1 800 OMICRON Fax 1 713 830 4661 info omicronusa com www omicronusa com OMICRON V6 2 1 0 ENU Subject to change without notice
9. LBS e POWERFUL UP TO 500 A DC e TRUE DC CURRENT e AUTOMATIC MEASURING e UP TO 100 RESULTS CAN BE SAVED e CPM 500 REMOTE CONTROL e SELECTABLE USER INTERFACE LANGUAGE Technical Data Power input 90 250V 50 60Hz Test current range 10 500ADC Resistance range 1 uQ 200 mQ Max resolution 0 1 uQ Accuracy 0 2 1 LSD Operating temperature 10 C 50 C 14 F 122 F Weight 7 5 kg 16 5 Ibs Dimensions 198 x 255 x 380 mm 78 0 x 100 4 x 149 6 in W x H x D without handle Standard test cable set e Current cables 2 x 6m 35mm with battery clips e Voltage sense cable set 2 x 6m with crocodile clips e Mains power cable e Ground cable Further accessories Visit the OMICRON website at www omicron at or Www omicronusa com 2 E V Sense CPM 500 Worldwide CONTACT ADDRESSES North and South America Europe Africa Middle East Asia Pacific OMICRON electronics Corp USA OMICRON electronics GmbH OMICRON electronics Asia Ltd 12 Greenway Plaza Suite 1510 Oberes Ried 1 Suite 2006 20 F Tower 2 Houston TX 77046 USA The Gateway Harbour City A 6833 Klaus Austria Phone 1 713 830 4660 or 1 800 OMICRON Phone 43 5523 507 0 Kowloon Hong Kong S A R Fax 1 713 830 4661 Fax 43 5523 507 999 Phone 852 2634 0377 Email info omicronusa com Email info omicron at Fax 852 2634 0390 Email info asia omicron at Pa lt a gN OMICRON offices OMICRON sales partner
10. OMICRON has a worldwide customer base which continues to grow Direct customer contact is essential in developing and building lasting relationships around the world To achieve this OMICRON has an extensive network of more than 50 representatives distributors and local offices To identify the contact for your area please visit Contact Us on our website at www omicron at or Www omicronusa com OMICRON is an international company providing innovative power system testing solutions With sales in more than 120 countries offices in Europe the North America and Asia and a worldwide network of sales partners OMICRON has established a reputation as a supplier of leading edge technology OMICRON s automated primary and secondary testing capabilities are important benefits in light of the changing market conditions resulting in restructured organizations required to do more with less Services in the area of consulting commissioning testing and training make the product range complete Specialization and visionary leadership allows OMICRON to continue with revolutionizing developments for its solutions to meet the customer needs of the 21st century OMICRON Sales Service Centers Europe Middle East Africa OMICRON electronics GmbH Oberes Ried 1 A 6833 Klaus Austria Phone 43 5523 507 0 Fax 43 5523 507 999 info omicron at www omicron at Asia Pacific OMICRON electronics Asia Ltd Suite
11. Resistance uQ Measurement Measures test objects with very low resistance such as contacts of circuit breakers and buswire connectors A 4 263 ma After connecting the test object 4 wires entering the test current and pressing the Start button the test card measures e the resistance of the test object The test current can be selected from 0 400 A Duration of the test 5 s including automatic reporting Output up to 400 A DC Input up to 10 V DC Ground Resistance Measures ground resistance or soil resistivity TATancheck FiGiround Comment iiet fio O00 ELIN aar The ground resistance of substations can be measured with two IAMS Eim Fence auxiliary electrodes which are put into the ground To avoid Y Paes a F F BD a influences from the mains current and its harmonics this EEN aE Des measurement should preferably be carried out with 128 Hz The i ee Ene Rents aay aiiai high output power of the CPC 100 ensures a high ratio of the E i oo signal level to the noise level Thanks to the signal processor technology the measuring procedure is very selective and disturbances by ground currents are reduced to a minimum For large grounding systems the auxiliary electrode for the current injection is replaced by a second grounding system which is connected to the current output of the CPC 100 via an approx Temrin No 10xa Assessed n
12. Seven successive automatic measurements on predefined levels are possible Further the signals can be generated with a repeat function so that the sequence can be run in an endless loop up to 100 results can be recorded Example Overcurrent relay with auto reclosure function State 1 wait for the circuit breaker CB to open Output 400 A until the trigger condition Overload occurs The measurement table displays relay time the CB opening time 290 ms State 2 wait for the CB to close short dead time Output 50 A until the Overload condition that startet state 2 clears The measurement table displays short dead time CB closing time 477 ms State 3 wait for the CB to open Like state 1 State 4 wait for the CB to close long dead time The measurement table displays long dead time CB closing time 3 1910 s Ramping The functionality of the Ramping test card includes the automatic measurement of the pick up and drop off values of overcurrent relays Also CB contact resistances can be measured with ramp functions to avoid induced voltages in CT windings Up to five ramps can be defined with detailed results being available for each CPC 100 Applications CU1 POWER FACTOR IMPEDANCE MEASURING Insulation Diagnosis Tangent Delta Power Factor Test The condition of the insulation is an essential aspect for the operational reliability of electrical power transformers generators and other
13. high voltage equipment CPC 100 CP TD1 provide laboratory precision for capacitance and dissipation power factor measure ments in the field Quantities measured include e Capacitance Cp e Dissipation factor tan 6 tangent delta e Power factor cos e Power active reactive apparent e Impedance absolute value phase inductivity resistance Q Tg jadin EDI au a Gate Primary Measurements Line amp Ground Impedances k factors mutual coupling Accurate data of the primary lines feeders and grounding are critical for operational reliability of electrical power i where modeling programs are used for design or when protective settings are applied inst Cad CPC 100 CP CU1 coupling unit provides the ability to safely and me GS E T N Deste accurately measure parameters of overhead lines power cables and 700 S00 Overload nwa son 500 5000 Overload n a mow ffe panne grids This system can determine 400 0 0 Overioad na 30000 me 6000 Overload nls i aro n oe TETE a e ee E Tr eN a Fyre Too De bie Hee Line Impedances amp k factors of ite AP ke Fong e ele overhead lines or power cables e Mutual coupling of parallel lines e Coupling of power lines to signal cables e Ground Impedances of large substations e Step amp touch voltages ei we itp i E fw bik i ja EER ee a5 Pokies MiquencE impedance Z i
14. of the test 5s including automatic reporting Output up to 130 V AC Input up to 10 A AC and 300 V AC Electronic Voltage Transformer Measures ratio and polarity of non conventional electronic VTs After entering the primary voltage the secondary voltage and the test voltage and pressing the Start button the test card e measures the low level secondary voltage e calculates the actual ratio the deviation and the polarity Duration of the test 5s including automatic reporting Output up to 2 kV Input up to 3 V or 300 V AC Voltage Withstand Test The voltage withstand test is described in the CT section page 10 Polarity Checker The Polarity Checker option can be found in the Accessories section page 21 CPC 100 Applications POWER TRANSFORMER VTBuden VTE lectoracs ji THRs Vinh s i vp iA Vase iGo ERTE M u naie ESS Vier POY M0 prim festi pA E 14 i 18k 31456 S ES ima Me Lk H 15 n JIR C 1 3 We 15 E 2 pHeRI T a oc We Tene SOT WoE ma T mao T 1e p fe Tap F maas Dev A mel Fipple Siope a een Reat Paci to Tta Power Transformer Ratio Tap Changer Test Transformer Ratio per Tap Measures ratio and excitation current per tap For this test a test voltage of up to 2 kV is injected on the transformer high voltage side This voltage is measured
15. shown on page 22 e Transport case similar model as for CPC 100 shown on page 22 but with different inner shell CPC too Accessories Cables clamps and plugs supplied with CPC 100 Standard Package High current cable set 800 A 70 mm2 High voltage cable set Connection clamps for high Grounding cable GR YE 1 x 6 m 2000 V screened 0 5 mm2 voltage for connection with 6 mm2 with connection clamp standard 2x6m_ VEHK0612 standard 2x6m_ VEHK0613 banana plugs 4 mm VEHKO0615 alternative 2x9m _ VEHKO617 alternative 2 x 10 m VEHK0618 VEHZ0610 Measurement cable set 2 5 mm2 Low voltage adapter low voltage plug Power cord with VEHK0623 VEHS0610 standard 6x6m VEHK0614 for V2 AC low voltage level input 0 3 V open ends VII connector ZA connector BS connector alternative 6x 10 m VEHK0619 VEHK0621 VEHK0616 VEHK0620 VEHK0624 standard alternative alternative alternative Crocodile clamps for connection of banana plugs 4 x 4 mmVEHZ0620 Current Clamp Active ac and dc current probe with voltage output USB memory stick 64 MB Article Nr VEHZ0666 2 measuring ranges 10 Aand 80A symbol picture Frequency range dc to 10 kHz Accuracy error lt 2 for currents up to 40 A and frequencies up to 1 kHz e Phase error lt 0 5 at 50 60 Hz Length 230 mm 9 1 Article Nr VEHZ4000 Transport cases OMICRON offers sturdy transport cases with hard form interiors Recommended for heavy transport stres
16. the CP TD1 to measure the Permittivity and Tangent Delta Power Factor of insulation liquids e g transformer oil The three electrode design with guard allows precise measurement especially of small losses The circular electrodes are constructed from rigid stainless steel and require a sample of 1 2 to 2 liters 41 to 68 fl oz Electrical connection to the Test Cell is made using the standard cables provided with the CP TD1 Technical Data Cell Type Three terminal guarded Cell Gap Spacing 11 mm 2 31 nominal Sample Volume 1 2 liters min to 2 liters max 41 to 68 fl oz Cell Capacitance 65 pF nominal in air Maximum Test Voltage 12 kV rms Voltage Operating Range 2 5 to 12 kV for a stress of 200 to 1100 V mm 7 87 to 43 31 V inch Connectors 6 mm 0 24 sockets for high voltage cable 4 mm 0 15 sockets for measuring cables Dimensions 22 x25 cm 8 7 x 10 2 diameter x height Weight approx 6 kg 13 2 Ibs Article Nr VEHZ0601 compensationg Reactor CP CR500 The CP CR500 allows Power Factor Tangent Delta measurements of large motors and generators In combination with the multifunctional Primary Test Set CPC 100 and the Tangent Delta Power Factor unit CP TD1 one CP CR500 compensates capacities of up to 500 nF To compensate even larger capacities two devices can be connected in parallel This allows compensation of up to 1000 nF Benefits e Portable high power solution due to size and
17. turns amp open windings technica Data FRAnalyzer FA1 Frequency range Source Output Output impedance Connector Wave form Amplitude 10 Hz to 20 MHz 50 Q BNC Sine signal 0 01 to 1 Vrms into 50 Q Inputs Reference CH1 Measurement CH2 Impedance Impedance high Connector Input sensitivity Input sensitivity Input attenuator Input attenuator Receiver bandwidth Dynamic range Dynamic range Gain error Phase error 500 1 MQ 60 pF BNC 1V rms full scale 100 mV rms full scale 20 dB O dB 10 dB 20 dB 30 dB 40 dB 10 Hz 30 Hz 100 Hz 300 Hz 1 kHz 3 kHz gt 90 dB gt 100 dB at 10 Hz receiver bandwidth lt 0 1 dB calibrated lt 0 5 calibrated using the network analyzer toolkit software PC Software for e operation of the FRAnalyzer e diagnosis of the measuring results e database handling e vector network analyzer toolkit e reporting PC requirements Interface USB 1 1 PC operating system Windows 2000 or Windows XP Processor Pentium 500 MHz or higher Memory 256 MB RAM or higher Drive CD ROM Mechanical data supply voltage Weight lt 2 kg without measuring cables Dimensions w x h x d 26x 5x 26 5 cm 10 25 x 2 x 10 5 inch DC power supply DC 10 V to 24 V 10 W For detailed information on the FRAnalyzer please see our product information sheet FRAnalyzer Sweep Frequency Response Analyzer for Power Transformer Winding Diagnosis M
18. weight e Flexible by using different numbers of units and or by changing the test frequency 35 85 Hz Technical Data with one CP CR500 Maximum Test Voltage 12 kV rms 50 Hz Inductors 2x 40H Current Compensation 2x1A Capacity Compensation 2 x 250 nF 50 Hz 2 x 500 nF 35 Hz Dimensions 468 x 394 x 233 mm 18 6 x 15 5 x 9 2 Weight 33 kg 73 Ibs oa 3 gf For detailed information on these products please see our product Article Nr VEHZ0602 single unit information sheet 12 kV Oil Test Cell CP TC 12 and VEHZ0603 set of two units Compensationg Reactor CP CR 500 CPC 100 Accessories POLARITY CHECKER CURRENT BOOSTER Polarity Checker Polarity check test card CPOL polarity checker Current Booster CP CB2 The Current Booster CP CB2 connected to CPC 100 Polarity Checker Replacement for Battery Checking Method Checks a series of test points for correct wiring Just inject a the special continuous test signal at one point with the CPC 100 and check the polarity at all terminals with CPOL as shown in fig 1 getting a clear indication whether the polarity is OK green LED or not red LED This procedure is much faster than the conventional method and can easily be performed by a single person Duration of the test depending on the number of test points 3 5 s per test point The Polarity Checker in use Current Booster CP CB2 Tests applications requiring up
19. with a new filename CPC too Accessories CP TD1 TANGENT DELTA amp POWER FACTOR The CP TD1 capacitance and tangent delta power factor measuring instrument Is an accessory unit to the CPC 100 completing its use for power transformer testing with the ultimate insulation diagnosis solution Controlled by the CPC 100 the CPC 100 CP TD1 combination provides fully automated testing and reporting capabilities for the comprehensive testing of transformer parameters within one portable system The application of innovative measurement techniques and the use of high precision components in the CP TD1 bring laboratory precision with a rugged design into the field of insulation condition testing The CP TD1 also offers new test methods such as testing with frequency sweeps A custom built trolley allows for practical handling on and off site along with easy and quick breakdown into portable components Quantities measured include e Capacitance Cp e Dissipation factor tan 6 tangent delta e Power factor cos e Power active reactive apparent e Impedance absolute value phase inductivity resistance Q Technical Data CP TD1 with CPC 100 The CP TD1 is connected via interfaces to the CPC 100 and thus does CP TD1 with transport trolley not need further control elements
20. 0 300 A 3 min 0 5 V 1950 W DC 0 200 A SU 6 5 V 1300 W DC 6 A DC 0 6 A gt 2h 60 V 360 W DC 2000 A AC with an optional Current Booster Please see page 21 for details Voltage outputs Range Amplitude tmax Imax Powermax f 0 2 kV 1 min 1 25A 2 5 kVA 15 400 Hz 2 kV AC Da2 kV 2h 0 5 A 1 0 kVA 15 400 Hz 0 1 kV 1 min 2 5A 2 5 kVA 15 400 Hz 1 kV AC 0 1 kV sAN 1 0 A 1 0 kVA 15 400 Hz 0 0 5 KV 1 min 5 0 A 2 5 kVA 15 400 Hz 500 V AC 0 0 5 V gt 2 li 2 0 A 1 0 kVA 15 400 Hz 130 V AC 0 130 V gt 2h 3 0A 390 VA 15 400 Hz Internal measurement of outputs Guaranteed accuracy Typical accuracy Output Range Amplitude Phase Amplitude Phase Reading Full Scale Full Scale Reading Full Scale Full Scale 800 A AC 0 20 0 20 0 20 0 10 0 10 0 10 400 A DC 0 40 0 10 0 20 0 05 2000 V 0 10 0 10 O20 0 05 0 05 omo 1000 V 0 10 0 10 080k 0 05 0 05 015S 2 kV AC 500 V 0 10 0 10 0 40 0 05 0 05 O20 5A 0 40 0 10 020k 0 20 0 05 0 10 50 mA 0 10 0 10 020k 0 05 0 05 0 10 Guaranteed values valid over one year within 23 C 5 C 73 F 10 F in the frequency range of 45 65 Hz or DC Accuracy values indicate that the error is smaller than Value read Reading error Full Scale of the range Full Scale Error With mains voltage 230 V with 2 x 6 m high current cable at 23 C 5 C 73 F 10 F ambient temperatur
21. EG Automatic reporting of all performed tests As all test results are automatically saved and stored with any test the results can be reported from a connected PC at any time via the CPC Explorer PC software Since the test data is stored in XML format a standard format for exchanging data and the report is provided in HTML format all Internet capabilities can be utilized including embedded graphics Conversely uploading data for archiving and further processing on the PC is supported Exporting to standard desktop applications such as text processors e g Microsoft Word databases or spreadsheet programs is also possible as XML format is used Reporting archived results is possible in exactly the same way as directly from the CPC 100 Adaptations of the report are possible by creating any custom data format using the XML data a CPC Editor Reduce on site testing time to a minimum by pa Leg n pa preparing tests offline on a standard PC a CPC Editor allows for setting up single test cards or entire tests with several test cards with all test Saad vied a specific settings This is done on a standard PC independent from the CPC 100 o ALLEI G These test files XML can be loaded to from a CPC 100 using CPC Explorer ai ee Once run they also contain the test results and assessments and can be viewed with CPC Editor Saved tests can be used as templates or procedures just clear results and save
22. High voltage output Dissipation factor DF tan V Tay at f Hz Range Resolution Accuracy Conditions 300 mA gt 2 min 0 0 i Bau 70 HZ 0 12 kV AC 15 400 oeae a 100 mA gt 60 min capacitive 0 005 lt 8 mA O 100 5 digits error lt 0 5 reading i Voltage current measurement 0 10000 0 02 Range Resolution Accuracy 12000 V AC 1V error lt 0 3 reading 1V Power factor cos gt AC 5 digits error lt 0 5 reading Range Resolution Accuracy Conditions 7 8 mA AC error lt 0 3 reading 100 nA 0 10 n error lt 0 1 reading 15 70 He capacitive 0 005 lt 8 mA Capacitance Cp equivalent parallel circuit o o i p p eq p 0 100 5 digits error lt 0 5 a reacing i Range Resolution Accuracy Conditions 0 02 error lt 0 05 readin _ 01 i j lt 8mA Representation of the following values is also possible 1 pF 3 uF 6 digits U I p e Power active reactive apparent error lt 0 2 reading gt 8 mA e Impedance absolute value phase inductivity resistance Q Nominal voltage CPC 100 1 x 100 240 VAC 50 60 Hz 16 A Signals below 45 Hz with reduced values possible Capacitive linear loads Operating temperature 10 55 C 14 131 F Transport and storage 20 70 C 4 158 F Relative humidity 5 95 non condensing CPC 100 Accessories CP CUT LINE amp GROUND IMPEDANCE The multifunctional primary test system CPC 100 in combi
23. NS OT EIE Des wig FS O34 IE Feni Polaaly for f Save fe Templates Choose test card i Winding Resistance Voltage Withstand Test and CT Ratio V Burden A Burden Burden Winding Resistance Measures CT winding resistance After entering the test current and pressing the Start button the test card e displays the deviation of the measurement over time during the period of charging the winding e automatically performs a discharging of the winding after saving the measurement e measures the DC voltage e measures the resistance e optionally compensates the temperature behavior of copper where the applied temperature compensation calculates the resistance for working temperature Duration of the test depending on the charging time After the charging period the user creates the report by pressing Save Results Output up to 6 A DC or 100 A DC 6 5 V Input up to 10 V DC Voltage Withstand Tests the voltage withstand capability of the insulation between primary and secondary winding or ground and secondary winding After entering the test voltage and the duration and pressing the Start button the test card e determines the leakage current flowing through the insulation The current threshold for maximum leakage current can be entered The CPC 100 wil
24. P2360 durden 1 272644 corm OE i v T too0v B i hw Delie Cad p Eemi Eae Renagre Canc yac aw 0 fevered La Faia JONG OAs HOAs Lew Fendi Polanly or Save fie Templates Assessed nfs VT Ratio Polarity and Burden Electronic VT VT a shielded cable transformer with twisted wires Burden a electronic voltage electronic protection relay with low V input VT Ratio and Polarity Measures the capacitive or inductive VT ratio and polarity After entering the primary voltage secondary voltage and test voltage and pressing the Start button the test card e measures amplitude and phase of the voltage on the transformer s secondary side e calculates the actual ratio the deviation and the polarity Duration of the test 5 s including automatic reporting Output up to 2 kV Input up to 300 V AC VT Burden Measures the connected secondary burden of the VT After entering the nominal secondary voltage and the test voltage and pressing the Start button the test card measures e the connected secondary burden with voltage injection on the VT s secondary side e the connected secondary burden in VA and the power factor cos including the secondary current and the angle between voltage and current Duration
25. PC 100 Nominal ac spark over voltage lt 1000 V rms Current Output Ranges Short circuit proof up to 30 kA for 100 ms Current range Compliance voltage Mechanical Data Dimensions x h 200 x 190 mm 7 9 x 7 5 inch ee ne one ne Weight 6 8 kg 13 2 Ib including grounding cable 0 20 Arms 250 V rms Weeaccarine O 50 Arms 100 V rms Special accessories for step touch and ground impedance on request 0 100 Arms 50 V rms Overall System with CPC 100 Measuring Transformers Output Power VT 600 V 30 V class 0 1 5000 VA cos lt 1 0 for 8 s 230 V ac mains voltage CT 100A 25A ee 5000 VA cos lt 0 4 for 8s 115 V ac mains voltage Mechanical Data Accuracy Protection IP 20 P C Dimensions w x h x d 450 x 220 x 220 mm 17 7 x 8 7 x 8 7 inch b d ane Ta ae ange Weight 28 5 kg 63 Ib 0 03 ss 02 Q Oes DaT 100 A Environmental Conditions 0 2 2Q 0 5 0 3 100 A Operating temperature 10 55 C 14 131 F 2 52 03 SOA Transport amp storage 2 Temperature 20 0 C 4 158 F gt 29 Q 0 3 20 A Relative humidity 5 95 non condensing 25 300Q9 03 10 10 A Safety EN 61010 1 Prepared for IEEE 510 EN 50191 VDE 104 EN 50110 1 VDE 105 Part 100 LAPG 1710 6 NASA At an ambient temperature of 23 C 5 C 73 F 10 F CPC too Accessories 12 KV OIL TEST CELL COMPENSATING REACTOR 12 kV Oil Test Cell CP TC12 The CP TC12 Is used in conjunction with the CPC100 and
26. ach 2 A compensation w inductors paralleled 500 nF CP CR500 Compensating Reactor set of two CP CR 500 compensation w units paralleled 1000 nF CP CU1 amp CP GB1 includes CP CU1 coupling unit w accessories for line impedance k factor mutual coupling ground impedance step and touch voltage and single cable coupling measurements Includes CP GB1 grounding box VEHZ0672 Customized packages with individual combinations are available on request CPC 100 Standard VE000611 CPC 100 Enhanced VE000621 S S N S YAS S NS NY NSAN NYS S S 0O SHO SHEOL S S SPS ENN A NYSS S OSHO SOLS ANNY NY SENSIS CP TD1 TanDelta Test Set VE000640 CP Transformer Test Set VE000645 N SNAN NS SNI SIS SSS S S Current Transformer Testing CT1 CT Analyzer OMICRON s CT Analyzer is a unique lightweight multifunctional instrument designed to meet the highest standards for performing excitation ratio polarity and winding resistance tests on current transformers CTs as well as burden impedance measurement The equipment provides automatic testing and calibration for all types of low leakage flux current transformers both on site in the power system as well in the controlled environment of CT and switchgear manufacturers The CT Analyzer allows an automatic assessment of test results clearly indicating whether the parameters of the CT under test match its specification I
27. d test 2 kV Rogowski coils Low power CT s CP VT test cards ratio burden voltage withstand test 2 kV electronic voltage transformers CP Transformer test cards winding resistance tap changer check ratio voltage withstand test 2 kV CP Resistance test cards contact resistance uQ mQ winding resistance uQ kQ CP Ramping test card programmable ramping generator and determination of thresholds CP Sequencer test card sequencer test card for testing with different states CP GR Ground resistance test option includes testing software hardware accessory VEHZ0660 CPOL polarity checking for CT VT wiring including software hardware accessory VEHZ0650 CP TD1 test card capacitance and dissipation power factor measurements CPC Editor Software for offline test preparation on a PC Standard 2 x 6 m 70 mm2 800 A Alternative 2 x 9 m 70 mm 800 A High voltage cable set 2 kV screened Standard 2 x 6 m 0 5 mm2 High current cable set Alternative 2 x 10 m 0 5 mm2 Standard 6 x 6 m 2 5 mm2 1 x 0 5 m 2 5 mm2 Alternative 6 x 10 m 2 5 mm2 1 x 0 5 m 2 5 mm2 High voltage clamps connection with banana plugs 4 mm 2 red 2 black Measurement cable set Crocodile clamps for connection of the banana plugs 4 mm 2 red 2 black Grounding cable GR YE 1 x 6 m 6 mm with connection clamp Ethernet PC connection cable 3 m Low voltage adapter 4 mm banana to low voltage plug USB Memory stick 64 MB User
28. e Signals below 50 Hz or above 60 Hz with reduced values possible 3 Output can be synchronized with mains 4 The input output is protected with surge arrestors between the pins and against protective earth In case of energies above a few hundred Joule the lightning arrestors apply a permanent short circuit to the input output gt Signals below 50 Hz or above 200 Hz with reduced values possible 98 of all units have an accuracy better than specified as Typical 7 Input is galvanically separated from all other inputs V1 and V2 are galvanically coupled but separated from all other inputs There are power restrictions for mains voltages below 190 V AC Fuse protected Error of reading lt than value Accuracy full scale Mode Connection Range Current Guaranteed Typical 0 5 VQ s 12 5 MO 4 wire 400 ADC 400 A 0 85 0 45 10 UQ TD 4 wire 6 A DC 6A 0 60 0 35 100 V s 10 Q 4 wire 6 A DC 1A 0 40 0 25 0 2 0 20 kQ 2 wire VDCin lt 5mA 1 00 0 20 0 50 0 1 Q General Display 1 4 VGA greyscale LCD display USB Memory USB 2 0 Guaranteed communication of USB sticks supplied by OMICRON Power Supply i Single phase nominal Single phase permissible 100 VAC 240 V AC 16 A 85 VAC 264 V AC LN or LL Frequency nominal 50 60 Hz Power consumption lt 3500 VA lt 7000 VA for short time lt 10 sec Connection IEC320 C20 Environmental conditions 10 55 C 14 131 F 20
29. e actual ratio e calculates the deviation from the nominal ratio Duration of the test 5 s including automatic reporting Output up to 800 A 2000 A with Current Booster CP CB2 Input up to 3 V AC built in CTFistcv ET Tages je Wed ST e add P1 burden CT Ratio Low Power Measures the ratio for CTs with low power principle output voltage is proportional to primary current shielded E l with twisted wires electronic protection relay iom 00A Y sec pon ites moa hi A Aus Delete Fi Liba Sal biia a ten SE yac Ew Doo i Maud no Rai Guama MTSE After entering primary current secondary voltage and the test current and pressing the Start button the test card e measures the amplitude of the injected current e measures the low power output voltage and phase angle e calculates the actual ratio e calculates the deviation from the nominal ratio Ascessed n a Duration of the test 5 s including automatic reporting Output up to 800 A 2000 A with Current Booster CP CB2 Input up to 3 V AC Polarity Checker The Polarity Checker option can be found in the Accessories section page 21 CPC 100 Applications VOLTAGE TRANSFORMER VT voim IDOOODY Tara ah Gower joa Manual np ae Cee ROE een Eel i Cunent clamp pen i t B Aus lsc ema
30. ed above such as S P Q Z R R X Re Ls Re Cp ratio 1 ratio 5 AV and Al and displays them in the third column ee 200 Om noe 595 92 Any relevant measured value can easily be stored by pressing Keep Result Thus all important results 000 53am DDE 1 000e are arranged and displayed in a table and can be further investigated at any time The entire testing ra R a 4 session can be saved as a file CPC 100 front panel displaying a CT ratio measurement A trigger functionality allows for triggering to a certain event such as reaching a threshold value or the switching of a contact The trigger can be binary or analog or it can be an overload of the unit For each trigger event both the current output value and the delay time are measured which allows the testing of pick up and drop off values of primary relays Complex impedances of transformers inductors or capacitors can be measured The amplifier technology used enables a wide range of testing frequencies different from the mains frequency The built in digital filters allow for sensitive and selective measuring without interference problems With this concept a lot of tests can be performed with small test signals instead of using big and heavy conventional test equipment with high test signals at mains frequency Reporting is done automatically Measurement of the complex impedances of a transformer CPC 100 Applications CURRENT TRANSFORMER
31. ia m T js substation ground 3 auxiliary line For this application the use of CP CU1 see page 19 i a is recommended to obtain higher output power For measuring the soil resistivity four auxiliary electrodes are used Duration of the test 5 s including automatic reporting Output up to 6 A AC Input up to 3 V or 300 V AC and 10 A AC Winding Resistance The winding resistance test is described in the CT section page 10 CPC 100 Applications AUTOMATIC TEST PROCEDURES Sequencer and Ramping With general application cards some simple testing of single phase relays is possible such as the determination of trip times of I gt V gt V lt or frequency relays With Sequencer and Ramping the user can create her his own automatic test procedures Sequencer Typical Applications e Testing of auto reclosure cycles with primary current injection up to 2000 A opening closing of the circuit breaker contact is detected internally by the OMICRON hardware so no extra wiring is required e Measurement of the opening closing time of circuit breakers e Testing of primary overcurrent relays e Testing of low voltage circuit breakers with protection functions etc e Automatic testing with different amplitudes e g CT ratio at 0 05 In 0 2 In 0 5 In 1 In and 1 2 In The user can define consecutive states and the transition between states which can be initiated by a timeout a trigger or a combination of the two
32. icroz Ohmmeter Description The CPM 500 is a completely electronic device that can be used for all Micro Ohmmeter applications from 10 to 500 A like resistances of circuit breakers disconnecting switches high current busbar joints cable splices welding joints or grounding connections The robust design makes the unit ideal for its use in high voltage substations and industrial environments The straightforward operation saves time both in training and actual testing After the test current has been set the automatic test procedure can be started by pressing the Q button The test result is displayed as R V I The CPM 500 uses advanced switch mode technology and generates true DC current with a stabilized output independent of the mains voltage Automatic ramps eliminate magnetic transients Up to 100 measurement results can be saved in the internal memory All measurements are marked with the testing time and date A remote control unit with built in sense connection allows starting and measuring the resistance directly at the test object It indicates when the test is running and if the result has qualified as an acceptable test or not CPM remote control for starting stopping of tests Tests with multiple measurements can be started directly from the remote control Connection to CPM 500 6 pole XLR CPM remote control Current and sense cables are connected using a 4 wire method e LIGHTWEIGHT 7 5 KG 16 5
33. internally with high precision The voltage amplitude and phase angle on the low level voltage winding is measured back via the measuring input The ratio is calculated automatically The excitation current in amplitude and phase angle is also measured and reported Duration of the test 5 s per tap including automatic reporting Output up to 2 kV Input up to 300 V Resistance and Continuity of OLTC per Tap Measures winding resistance per tap and detects interruptions of on load tap changer OLTC diverter switches The voltage drop at the winding resistance is measured with a sense line The resistance value of each tap can easily be saved to a table containing all taps An automatic temperature compensation is possible Interruptions of the current because of a faulty diverter can be detected Duration of the test depends on the inductivity of the winding inductance Due to the high output voltage of up to 65 V testing time is reduced Output up to 6 A DC 65 V or up to 100 A DC 6 5 V Input up to 10 V DC and 10 A DC Voltage Withstand Test The voltage withstand test is described in the CT section page 10 Winding Resistance The winding resistance test is described in the CT section page 10 Power Factor See CPC 100 Accessories CP TD1 tangent delta power factor measuring page 18 CPC 100 Applications RESISTANCE Resistance Q measurement Ground Resistance SSS a a a
34. l automatically switch off if the maximum leakage current is exceeded Duration of the test can be set by the user the test report will be created after the test automatically Output up to 2 kV AC CT Ratio V Tests ratio and polarity with direct injection of the voltage to the CT secondary input common method for bushing CTs After entering primary and secondary current and the test voltage and pressing the Start button the test card e measures the actual ratio e calculates the deviation of amplitude and phase angle of the voltage of the CT primary side Duration of the test 3s including automatic reporting Output up to 130 V Input up to 3 V and 300 V AC OMICRON CT Ratio Rogowski CT Ratio Low Power shield of cable connected to Rogowski coil maja ipm AOA vie OOM iron OOS hue Paai C Ues ssl bogshe jv ec perm C Man inp CT Ratio Rogowski Measures the ratio for CTs with Rogowski coil principle induced voltage is proportional to the conductor current differentiated with respect to time electronic protection relay with integrator Auto After entering primary current secondary voltage test current nominal frequency and pressing the Start button the test card e measures the amplitude of the injected current e measures the Rogowski coil s output voltage and phase angle e calculates th
35. manual Reference manual includded on the CPC Explorer CD ROM CP SA1 Surge Arrestor Box for 100 A winding resistance measurement from firmware v1 4 gt TH3631 Temperature Humidity Measurement Unit for humidity ambient amp surface temperature measurement Power Cord VII connector for most of EU M East VEHK0616 ZA connector for South Africa Namibia and India VEHK0620 BS connector for UK Hong Kong etc VEHK0624 C 19 3 x 1 5 mm 2 5 m with open ends VEHK0621 Accessories Low voltage plug spare plug for low voltage input 0 3 V CP Amplifier test card for using the CPC 100 like an amplifier Current Clamp 10 A 80 A with voltage output Polarity Tester hardware only CPOL polarity tester cover batteries Ground resistance accessory set 4 electrodes 50 m cable reel red 100 m cable reel black CP CB2 Current Booster up to 2000 A includes connection cable to CPC VEHK0611 ground cable w clamp VEHK0615 high current cable set w plugs amp clamps VEHK0610 including transport case with wheels VEHPOO71 CP TD1 TanDelta Power Factor Upgrade includes CP TD1 12 kV hardware w ref capacitor TanDelta test card TD1 Accessories cable drums w HV amp Measure cables 20 m Trolley elFC card if needed VEHZ0646 CP TC12 12kV Test Cell for measuring permittivity and TanDelta Power Factor of insulation liquids like tranformer oil CP CR500 Compensating Reactor that includes two inductors rated 40 H e
36. nation with the coupling unit CP CU1 is the world wide unique measurement system for e Line impedances and k factors of overhead lines or power cables e Mutual coupling of parallel lines e Coupling of power lines into signal cables e Ground impedances of large substations e Step and touch voltages The test system overcomes the problem of power system frequency interference that has previously made it necessary to use extremely large high power equipment to carry out these measurements The use of switched mode amplifiers and frequency shift techniques facilitates accurate measurements with compact portable equipment As measurements on high voltage lines need special safety measures the CP CU1 coupling unit is used to couple the inputs and outputs of the CPC 100 to the transmission line safely The grounding box CP GB1 is used to make the connection from the CP CU1 to the Multifunctional primary test system CPC 100 in combination with the overhead line and can be located at a safe distance from the operator coupling unit CP CUT and the grounding box CP GBI The grounding box allows the discharge of currents up to 30 kA once its voltage limit is exceeded Technical Data CP CU1 with CPC 100 Technical Data CP GB1 For detailed information on the CP CU1 please see our product Electrical Data information brochure CP CU1 Extension to the C
37. rden load with direct injection of secondary current with disconnected CT After entering the secondary nominal current and the test current and pressing the Start button the test card measures e Secondary voltage in magnitude and angle e Connected burden in VA and power factor cos Duration of the test 3s including automatic reporting Output up to 6 A AC Input up to 10 A AC 3 V via current clamp and 300 V CT Excitation Tests the excitation curve The necessary wiring is only two leads from the voltage output to the open secondary wiring of the CT After entering the current and voltage limits and pressing the Start button the test card will automatically record the CT excitation curve according to IEC 60044 1 ANSI 45 or ANSI 30 standards and the kneepoint will be automatically calculated After the test the core is demagnetized The test is done using a regulated voltage source Duration of the test 30 s including automatic reporting with recorded excitation curve and calculated kneepoint voltage Output up to 2000 V AC CT CONTINUED Riess eT mo Time cos itenat 0000 Dew TE wi Tempersiiae comoaesahon for Cu idiomas Tarim nema F view enone 04 SO He E eer al onl AL 4 vi Ture A00 Asses sed nia Rogale Add ipm 200A liee 500A eas vet 5000 S00 He A Aue Dese a Cad Veen Bomoy Goer Vow My o Mero Lo Raic PO
38. read with a standard PC and the CPC explorer Standards Safety EN60950 and EN61010 Mechanical Data Weight 8 kg 17 4 Ib without accessories Dimensions w x h x d 360 x 285 x 145 mm 9 2 x 7 2 x 3 7 in CT Analyzer The unique analyzing solution for current transformers European Patent EP 1 398 644 A1 The CT Analyzer performs an automatic parameter search and CT test using the Name plate guesser function For standard current transformers it allows the user to do a test with minimal training and no knowledge of the CT to be tested The extremely small and lightweight lt 8 kg 17 Ib hardware is particularly beneficial for on site testing and the clear user interface utilizing clear electronically implemented test cards identified by labeled tabs makes the device very easy to use Ordering Information VE000650 CT Analyzer standard package CT Analyzer hardware Compact Flash card 128 MB VEHZ0654 USB compact flash card reader VEHZ0655 power cord reference manual and CPC Explorer PC Software Accessories set for CT Analyzer Coax cable set with banana plugs VEHK0651 2 connection clamps for primary side VEHZ0652 Set of grounding cable VEHK0615 Crocodile clamp set VEHZ0656 6 flexible terminal adapters Carry bag for CT Analyzer VEHP0018 CT Analyzer standard package with Accessories CT Analyzer standard package VE000650 Accessories set for CT Analyzer VEHK0650 Transport case
39. s or for shipping Carry bag for CPC 100 For Current Booster Case model For CPC 100 aaa auxiliaries CP CB2 f CPC 100 CP CB2 connection cable Capacity Cable sets current clamps manual power cord high current cable set Dimensions 700 x 450 x 500 mm 400 x 250 x 450 mm 700 x 450 x 360 mm LxHxD 27 6 x 17 7 x 19 7 15 7 x 9 8 x 17 7 27 7 x 17 7 x 14 2 Weight 10 8 kg 23 8 Ib 24 kg 52 8 Ib 9 0 kg 19 8 Ib CPC 100 Package Options PACKAGE OPTIONS CPC 100 Package Overview Item No VESM0610 VESM0615 VESM0620 VESM0625 VESM0630 VESM0635 VESM0640 VESM0645 VESM0665 VESM0670 VEHK0612 VEHK0617 VEHK0613 VEHK0618 VEHK0614 VEHK0619 VEHZ0610 VEHZ0620 VEHK0615 VEHK0622 VEHK0623 VEHZ0666 VESD0601 VESD0600 VEHZ0665 VEHZ0644 VEHS0610 VESM0660 VEHZ4000 VEHZ0650 VEHZ0660 VEHZ0630 VE000641 VEHZ0601 VEHZ0602 VEHZ0603 VEHZ0671 Description CPC 100 Multifunctional Primary Test System including hardware and built in software for testing of the CT s VT s power transformers resistance etc The basic system includes e CPC 100 multifunctional test set hardware 110 230 V 50 60 Hz with AC 800 A AC 2000 V DC 400 A e QUICK test card manual control of the test set e Transport case with wheels for CPC 100 VEHPO061 e Carry bag for CPC 100 accessories VEHPO069 1 Packages CP CT test cards ratio V ratio I excitation curve burden winding resistance voltage withstan
40. t generation saves time by eliminating manual report writing accelerates project completion e Print reports via external PC SAFETY FEATURES e Emergency shut off button e Safety key lock block unauthorized use e Overvoltage protection on all inputs and outputs PREPARED FOR THE FUTURE e Able to test unconventional equipment like Rogowski Coils or current sensors e Unit accessible in a network or with direct PC connection via standard internet protocols e Maximum flexibility for future enhancements by using DSPs for signal generation and switched mode amplifier technique POWER TRANSFORMER e Ratio e Winding resistance e Tap changer testing e Excitation current e Short circuit impedance measurement e Leakage reactance RESISTANCE TESTING e Contact resistance uQ e Winding resistance uQ kQ e Ground resistance e Measuring of complex impedances winding impedances cable impedances etc PROTECTION RELAYS Single phase testing of primary and secondary relays I gt V gt V lt or frequency relais ADDITIONAL APPLICATIONS e TANGENT DELTA POWER FACTOR TEST Add the CP TD1 accessory for the most flexible insulation diagnosis system in the industry e LINE amp GROUND IMPEDANCES k FACTOR MUTUAL COUPLING more Add the CP CU1 accessory for the safest most accurate system for measuring primary power system parameters CPC 100 Technical Data 1 2 3 4 5
41. t is also applicable to the testing of CTs according to IEC 60044 6 TPS TPX TPY and TPZ With the CT Analyzer the testing of current transformers can be carried out to an extremely high level of accuracy The accuracy level 0 02 1 min makes the CT Analyzer the ideal tool for calibration and verification not only for protection CTs but also for class 0 2 CTs for metering Using the CT Analyzer provides a wide range of test and measurement possibilities e Burden measurement e CT Winding resistance measurement e CT Excitation characteristic recording e CT transient behavior measurement e CT ratio measurement with consideration of connected burden CT phase and polarity measurement Determination of accuracy limiting factor instrument security factor secondary time constant remanence factor knee point voltage current class saturated and non saturated inductance Assessment according to defined standards IEC60044 1 IECo0044 6 IEEE C57 13 1993 Technical Data CT Analyzer CT1 Generator amplifier section Output current 0 5 Arms 15 A peak Output voltage 0 120 V Output power 0 350 VA 1500 VA peak Ratio accuracy For O VA up to rated power ratio 1 2000 0 02 ratio 2000 5000 0 03 ratio 5000 10000 0 05 Phase measurement Resolution 0 1 min Accuracy 1 min for cos 0 8 1 Data transfer Compact flash card to store test results and for transfer of data to a PC Data can be
42. tatives OMICRON has truly established its reputation as a supplier of the highest quality The automated testing and documentation capabilities of OMICRON testing solutions are important benefits in light of the changing market conditions resulting in restructured organizations required to do more with less Today OMICRON s products revolve around a testing concept which provides the solutions to many challenges created by these competitive trends in the marketplace This integration of lightweight and reliable hardware with flexible and user friendly software is referred to as the OMICRON Test Universe Services in the area of consulting commissioning relay testing and training make OMICRON s product range complete Specialization in power system testing along with visionary leadership allows OMICRON to continue with innovative developments for its testing solutions to meet the customer needs of the 21st century SECONDARY TESTING CM line catalog OMICRON s product range in the area of secondary testing For a detailed list on literature currently available please refer to www omicron at support literature or Wwww omicronusa com support literature Table of Contents TABLE OF CONTENTS itunctiona mary Test Stem CPC 10 g A Overen 5 Interface Elements 7 Technical Data 8 Applications 9 Current Transformer CT 12 Voltage Transformer VT 13 Power Transformer 14 Resistance 15 Automatic Test Procedures 16 Power Factor
43. to 2000 A The output current of the CPC 100 can be increased to up to 2000 A by an electronically controlled current booster The CP CB2 can be connected close to the busbar using short high current leads and to the CPC 100 with a long control cable Current outputs Range Amplitude tmax Vmax Powermax f 0 1000 A 255 4 90 V 4900 VA 15 400 Hz 1000 A AC 0 500 A 30 min 5 00 V 2500 VA 15 400 Hz 2000 A 0 2000 A 255s 2 45 V 4900 VA 15 400 Hz Internal measurement of outputs Guaranteed accuracy Typical accuracy Output Amplitude Phase Amplitude Phase Reading Full Scale Full Scale Reading Full Scale Full Scale 2000 A AC 0 25 0 25 050 0 13 0 13 0 25 1000 A AC 0 25 0 25 0 50 0 13 0 13 0 259 Dimensions 186 x 166 x 220 mm 7 3 x 6 5 x 8 7 without handle Weight 16 0 kg 35 3 Ibs With mains voltage 230 V at 23 C 5 C 73 F 10 F ambient temperature Signals below 50 Hz or above 60 Hz with reduced values possible Outputs in series 4 Outputs in parallel i High current cable set for CP CB2 Connection cable to CPC 100 4x 1 5 m with 95 mm with plugs and clamps for CP CB2 1 x 0 6 m with 95 mm for serialization of outputs 20 m 3x 2 5 mm2 VEHK0610 VEHK0611 Includes in option e CP CB2 Current Booster e Connection cable to CPC 100 e High current cable set for CP CB2 e Grounding cable same as for CPC 100
44. to allow for a wide variety of applications Testing of unconventional equipment such as Rogowski Coils or current sensors completes the spectrum Additional aplications can be added with the use of flexible accessories to include tangent delta power factor testing and primary measurements of power system equipment LIGHTWEIGHT CURRENT TRANSFORMER CT 29 kg 64 Ibs automatic testing of eLess than one fourth the weight Ratio burden and polarity of equivalent conventional equipment e Phase and magnitude error Lower transportation costs saves money e Excitation curve e Easier handling reduces manpower needs Winding resistance eLow weight reduces injury potential e Secondary burden e Dielectric withstand voltage 2kV ac ONE TEST UNIT FOR MULTIPLE USES eCT circuit continuity e Replaces need for multiple test sets e g high current and voltage sources uQ meter e Saves time by eliminating need for multiple training VOLTAGE TRANSFORMER VT e User friendly interface reduces training time to one day automatic testing of e Ratio and polarity AUTOMATED TESTING AND REPORTING Phase and magnitude error e Prepare test plans ahead of time in office saves time in field testing phase e Secondary burden eSet up test once push button to start saves testing time e Dielectric withstand voltage 2kV ac e Results automatically saved saves time and reduces errors by eVT circuit continuity eliminating manual recording of data e Automatic repor
45. values 19 Test start stop button CPC too Technical Data INTERFACE ELEMENTS CONTINUED Plug to connect external functions e external test start stop push button e external I O signal lights 2 Serial interface for devices like CP TD1 3 Connection of CPC 100 to a network or direct connection to a PC s network connector 4 Connection of a USB memory stick 1 Grounding terminal 2 High AC Voltage Output 2 kV AC 1 A 5 A AC 3 Ext BOOSTER for the connection of the booster options 4 High DC current output 400 A DC 5 High AC current output 800 A AC 6 Mains power supply 1 phase 85 V 264 V AC 7 Overcurrent protection 8 POWER O I CPC 100 Technica TECHNICAL DATA The output is either voltage or current and is automatically selected by the software or manually by the user Current and voltage outputs are overload and short circuit proof and protected against over temperature Inputs Measuring inputs Guaranteed accuracy Typical accuracy Input Imped Range Amplitude Phase Amplitude Phase Reading Full Scale Full Scale Reading Full Scale Full Scale 10 AAC 010 0 10 020 0 05 0 05 010 TAAG 010 010 030 0 05 0 05 0 159 IACGDC lt 0 19 10 A DO 0 05 0 15 0 03 0 08 1ADE 0 05 015 0 03 0 08 300V
46. with wheels VEHK0650 VE000652 VEHPOO068 For detailed information on the CT Analyzer please see our product information brochure CT Analyzer The Revolution in Current Transformer Testing Sweep Frequency Response Analyzer FAT FRAnalyzer The FRAnalyzer detects defects in power transformer windings and faults in the magnetic core After the transport of transformers to site and following through faults with high currents the frequency response test should be made to ensure that the windings were not damaged Unique features of FRAnalyzer e High reliability of results due to the applied measurement method and innovative connection technique e Extremely small size through the use of leading edge technology e Ease of use because of creative software e Portability resulting from battery operation For high frequencies the equivalent circuit diagram of a transformer winding can be drawn as a complex network of resistors inductors and capacitors The frequency response of this network is unique It is like a fingerprint Winding deformation causes one or more of the capacitors in this network to be changed To identify these changes which can be quite small the new fingerprint is compared to that taken in a healthy condition Deviations are an indicator of e Coil deformation axial amp radial e Faulty core grounds e Partial winding collapse e Hoop buckling e Broken or loosened clamps e Shorted