Centura Shoulder CPM User Manual

image

Contents

1. Replace the fuse s of the connector with fuses of the same type and calibre 2 fuses T 750 mA 250V 6 3 x 32 KINETEC order 4610007434 If the display still does not indicate any information contact your nearest KINETEC technician If after switching on your KINETEC e Your KINETEC does not work and the display indicates 50 STOP 25 115 Press START again e Your KINETEC still does not function Contact your nearest KINETEC technician e Your KINETEC does not function and the display indicates SERVICE D1 angle measurement function failure ou SERVICE D2 no movement ou SERVICE D3 motor rotation failure ou SERVICE D4 abnormal consumption Contact your nearest KINETEC technician if the same message is displayed after having switched the device off then on and started it by pressing START CLEANING Before conducting any cleaning operation SWITCH OFF the unit and disconnect the power supply In order to ensure optimal hygiene you are advised to clean the machine for each new patient Cleaning should be carried out in the environmental conditions specified in the Technical Specifications section below Use a DISINFECTANT product alcohol free or lt 5 alcohol solution in spray plastic cases and metal components In order to ensure optimal hygiene you are advised to clean the covers for each new patient All the consumables enable hazard free disposal Recommendations to obtain a m
2. of the movement is immediately press limit 20 BYPASS 20 160 recorded Continue the session with FLEX EXT er a a the new movement limits 20 RUN 50 160 9 movement Comments e The values shown in the Display column are examples They actually depend on the stored programs The movement speed is fixed and cannot be changed default factory setting 2 Centura 18 29 EN INSTRUCTIONS FOR USE HOW TO USE THE PADS COMPONENTS The following pads and straps labeled at left are included in the pad kit a elbow pad b wrist pad c hand strap d biceps pad e Abduction adduction e Flexion extension Hand strap and wrist pad attachment e Rotation e Synchronized abduction adduction with rotation SETTING UP THE MACHINE Each set up requires only some of the pads as follows e Set up for abduction adduction requires pads a b and c e Set up for rotation requires a b and c e Set up for synchronized abduction adduction with rotation requires a b and c e Set up for flexion extension requires b c and d Attach the pads and straps to the CPM machine as follows a Elbow pad Position it in the elbow support with the seam distal and the straps and D rings toward the elbow support Thread the straps and D rings through the slots in the elbow support b Wrist pad Position it on the wrist support with the strap and D ring toward the wrist support Thread the strap and D ring through the slot
3. OF THE HAND CONTROL RESET TIME FUNCTION This function allows one to read the running time since the last resetting of the counter Check if the locking switch is To stop the unit FLEX EXT in the following position 50 STOP 15 100 i Press simultaneously on the REST TIME 125H Pape since cat 2 keys Reset Limit low resetting To reset the counter press REST TIME 125H The counter is now reset the key Reset is done Or After 5 seconds the reset FLEX EXT function switches off and the running time remains in the 50 STOP 15 100 memory How to choose a movement aw amp The movement can only be changed when the machine is stopped and the locking switch is in the unlocked position on or a You can select a movement by pressing the appropriate button The LED is on When a movement is first selected the system returns to the original parameters of the movement default settings Default settings hor abd abd add flex rot q Z 3 i abd add rot 0 Abduction Horizontal Abduction Flexion Extension Rotation Rotation Abduction e Lower limit 30 e Upper limit 60 e Speed 2 e Load 6 e Extension pause 0 e Flexion pause 0 e Timer 0 Possible values for each parameter Horizontal Abduction Flexion Extension Rotation Abduction Rotation abduction 60 to 85 30 to 85 e Lower limit 20 to 155 20 to 175 30 to 105
4. amp ASSEMBLING THE PARTS Loosen the knob 1 and slide the motor Assemble the abduction motor 3 and Use the color code to assemble the support 2 to the right or the left tighten the screw 4 rotation motor red for right blue for left Plug in the hand control Plug in the motor The assembly is secure when you hear a click Plug in the motor Use the color code to assemble the elbow Use the color code to assemble the forearm The KINETEC Centura is shown assembled splint 6 red for right blue for left The splint 7 Tighten the knob 8 for a left shoulder mobilisation assembly is secure when you hear a click Centura 20 29 EN INSTRUCTIONS FOR USE PATIENT SET UP Make sure the straps are clean Put the unit in the position that is the most comfortable for the patient Position the patient in the chair in a Slide the arm supports toward the Adjust the lengths comfortable position and supportin atient and put the arm in the a arm p pp g p P b forearm the affected arm supports Secure pads ADJUSTING THE SHOULDER JOINT AXIS c Vertical adjustment e Loosen the 2 knobs 1 e With the handle 2 adjust the height of the entire mechanism e Tighten the 2 knobs 1 d Side to side adjustment e Loosen the knob 3 Slide the entire mechanism e Tighten the knob 3 e Scapula plan choice e Loosen the knob 4 e Rotate the arm support e Tighten the knob 4 f S
5. assigned to a specific connector Connect the power supply cable 1 1 IMPORTANT Check that the electrical socket is in good condition and is suitable for the splint power supply cord The latter complies with current standards and has a grounding socket The plug may be connected to any standard socket The socket must however have a grounded pin To connect the power supply only use the original cable supplied with the machine Check that the cables remain free around the device so that they do not get damaged Check that the machine is not damaged in particular the protective housings Watch out for the position of the motor cables they must remain clear around the machine Wrong position Good position STARTING THE UNIT Switch on the unit 2 The indicator light 3 lights up the machine carries out an auto diagnostic and then the display shows Your KINETEC Centura is ready to be used FLEX EXT 30 STOP 15 130 CHANGING THE DISPLAY LANGUAGE Press the following buttons simultaneously SPeed and force followed AN or Z to change the display language Press the o button to confirm limit then switch the machine off and then on again to apply the changed display language Available languages English French German Italian and Spanish Centura 7 29 EN USE OF THE HAND CONTROL LOCKING THE HAND CONTROL SETTING f 3 A This function allows the physician to authorise or block access to ch
6. original cable supplied with the machine Check that the cables remain free around the device so that they do not get damaged Warning Before using this machine always check that the machine is not damaged in particular the protective housings Warning In case of electromagnetic interference with other devices move the device Warning Please do not touch the moving parts while the unit is running pinching risk Warning Modifying the machine in any way is strictly forbidden Warning Before each use always check that the red knobs under the chair have been tightened see page 6 Warning Before each use always check that the various locking knobs have been tightened see pages 3 and 4 Centura 1 29 EN DESCRIPTION DEFINITION The KINETEC Centura is a upper extremity PASSIVE mobilisation device enabling the following movements e Abduction Adduction from 20 to 160 in an exterior rotation position e Abduction Adduction from 20 to 160 with the elbow in a flexed position e Internal rotation 60 to external rotation 90 e Combined movements Abduction Adduction from 20 to 160 and external Rotation from 30 to 90 e Flexion Extension from 20 to 180 e Horizontal Abduction from 30 to 110 available on Centura 5 e Indications e Total shoulder replacement e Repeated dislocation of the humerus e Rotator cuff injury e Upper humerus fractures e Scapula fractures e Acromioplasty e Capsu
7. page 2 Device of type B class Speeds from 50 to 140 min IP 20 Patient height from 1 40 to 2m Fuse T 750mA 250V 6 3x32mm 4T to 6 7 KINETEC order 4610007434 Acoustic pressure lt 70dB SYMBOLS USED TYPE B device ne A protection against electric shocks r Lower limit imit Caution consult the accompanying Upper limit documents limit BH STOP power off a x Flexion movement O ON power on te G Rotation movement TYPE B device protection against electric shocks a F ro a Q a Abduction movement start Start movement Combined movement abd add rot hor abd Stop movement Ww Horirontal abduction movement program Program access A Hand control locked speed Speed T Hand control unlocked timer Timer o Hand control half locked Switch on LED and defect signal force Force when the LED blinks Pause Pause ma Alternative current Contains electric and electronic Increase components not to throw in the dustbins of household refuse 10 C Temperature Limit during storage or Z Decrease ac transport t WARRANTY Right way up when box is stored 4 Keep dry during storage or transport The KINETEC warranty is strictly limited to the replacement free of charge or repair in the plant of the component or components found to be defective KINETEC guarantees its joint passive mobilization systems for 1 ye
8. settings are fixed PARTS NEEDED e Chair e Abduction or flexion splint e Hand control Arih T centura E29 RVD Se Loosen the knob 1 and slide the motor Assemble the abduction motor 3 and Use the color code to assemble the support 2 to the right or the left tighten the screw 4 forearm support red for right blue for left Plug in the hand control Plug in the motor The assembly is secure when you hear a click E Sl l Sie Use the color code to assemble the forearm Position arm splint 8 splint 7 Tighten the knob 6 The assembly is secure when you hear a click The KINETEC Centura is shown assembled for left shoulder mobilisation Centura 22 29 EN INSTRUCTIONS FOR USE PATIENT SET UP Make sure the straps are clean Put the unit in the position that is the most comfortable for the patient Position the patient in the chair in a Slide the arm supports toward the Adjust the lengths comfortable position and supporting patient and put the arm in the a arm the affected arm supports b forearm Secure pads ADJUSTING THE SHOULDER JOINT AXIS c Vertical adjustment e Loosen the 2 knobs 1 e With the handle 2 adjust the height of the entire mechanism e Tighten the 2 knobs 1 d Side to side adjustment e Loosen the knob 3 Slide the entire mechanism e Tighten the knob 3 e Scapula plan choice e Loosen the knob 4 e Rotate the arm suppo
9. 12 Abduction motor 13 Shoulder depth sliding lock Centura 3 29 EN DESCRIPTION ABDUCTION ASSOCIATED WITH ROTATION SPLINT 14 Centura 18 14 Arm length setting lock 15 90 elbow splint 16 90 elbow splint lock 17 Rotation motor 18 Elbow splint support lock 19 Forearm length setting lock 20 Forearm slider 21 Right left bean swivel lock 22 Forearm splint 18 Elbow splint support lock 23 Swiveling splint support 24 Elbow flexion setting lock 25 Arm splint 26 Arm splint lock 27 Forearm length setting lock 28 Horizontal abduction column 14 Arm length setting lock 18 Elbow splint support lock 29 Elbow support older 30 90 elbow splint lock 19 Forearm length setting lock 20 Forearm slider 4 29 EN DESCRIPTION DESCRIPTION ELECTRICAL 1 Supply cable connector switch 2 Fuse 3 ON OFF switch 4 Hand control lock switch 5 Defect or power light 6 Motor or hand control connectors 7 Hand control location for transport 8 Hand control Q 9 Liquid crystal display 3 10 Flexion extension an 59 11 Abduction adduction t centura 12 Rotation pu an N 13 Abduction adduction synchronized with rotation 14 Lower limits setting 15 Upper limits setting 16 Increase decrease 17 START 18 STOP 19 FORCE 20 SPEED 21 PA
10. AL 100 150 movement To set the pain level when FLEX EXT The new lower value limit ap immediately L 25 MANUAL 25 150 ee is The angle display changes with current movement To start the session with FLEX EXT the new movement limits 25 RUN 30 150 Specific rules for synchronized movement e You can only change the upper limit of the movement through the manual mode and only beyond the synchronization point Comments e The values shown in the Display column are examples They actually depend on the stored programs The movement speed is fixed and cannot be changed default factory setting 2 Centura 17 29 EN USE OF THE HAND CONTROL HOW TO DEFINE THE UPPER AND LOWER MOVEMENT LIMITS e During the session The BY PASS MODE is a way to record the pain threshold of a patient during a session The angle display changes with current movement The unit is runnin Check if the locking switch 3 FLEX EXT is in the following position 25 RUN 30 150 or To select the BY PASS FLEX EXT Wheunibexceedsithe MODE 25 BYPASS 160 150 recorded upper limit continuous press To set the new pain level FLEX EXT The new upper value limit when reached of the movement is immediately press limit 25 BYPASS 160 160 recorded To select the BY PASS FLEX EXT ara toane mode for lower limits 25 BYPASS 20 160 continuous press movement o set the new pain leve e new lower value limit T h in level FLEX EXT Th lue limi when reached
11. Centura Centura 5 Before use please read this document EN Kinetec reserves the right to effect technical modifications 467896351 05 2011 S rie 5 0459 Notice Originale User manual Xkinetec kinetec MONS REST Rolyan Sammons Preston Rolyan wi A Patterson Medical Company PA I ERSON W68 N158 Evergreen Blvd MEDICAL Cedarburg WI 53012 Phone 800 228 3693 Fax 262 387 8748 SUMMARY Select your language SEE PAGE 7 Warning and Safety instructions Warning The physician physiotherapist defines the protocol and ensures that it is correctly implemented adjustments session time and frequency of use Warning Run a cycle with the device unloaded before installing the patient on the machine Warning For optimum safety always give the hand control to the patient before starting the system The patient must know the start stop reverse function on the hand control see page 8 Warning To avoid the parameters being changed lock the machine s hand control before giving it to the patient Warning Danger risk of explosion Do not use the machine with anaesthetic gas or in an environment that is rich in oxygen Warning To avoid all risks of electric shock the machine should only be connected to a power supply that has protective grounded pin Warning Before using this machine always check that the electrical socket is in good condition and is suitable for the splint power supply cord Only use the
12. EX EXT The value change at the RUN 10 60 speed of the movement Comments e The values show in the Display column are examples They actually depend on the stored programs e The current movement parameters can be changed while using that program but no data will be stored in the original program See the programming mode p 14 to modify programs Centura 12 29 EN USE OF THE HAND CONTROL READING THE VALUES OF A PROGRAM EXAMPLE SPEED program Check if the locking switch is in the following position ROTATION or To stop the unit STOP 15 90 To access the program PROGRAM 1 The program number mode ROTATION 90 blinks To change the program if PROGRAM 3 The new program number necessary ROTATION 30 blinks PROGRAM 3 Displaying the speed To read the speed value SPEED 1 val After 15 seconds or after PROGRAM 3 pressing on another key 0 ROTATION 30 To exit and validate the ROTATION aea a ae selected program 0 STOP 15 60 program 3 Start the unit ROTATION The value change at the 0 RUN 20 60 speed of the movement Comments e The values showed in the Display column are examples They actually depend on the stored programs The current movement parameters can be changed while using that program but no data will be stored in the original program See the programming mode p 14 to modify programs Centura 13 29 EN USE OF THE HAND CONTROL HOW TO MODIFY PROGRAMS PROGRAM MODE prog
13. USE 22 TIMER 23 PROGRAM access 24 Horizontal abduction Centura 5 29 EN ASSEMBLY amp TRANSPORT BASE ASSEMBLY Remove all the components from their packaging Place the chair 1 on the floor we recommend that two people carry out this operation Take care not to trap your fingers Straighten up the back of the chair 2 The tilting axis 3 must always be in good condition Never use the splint without it Lock the back of the chair by tightening the red knobs 4 Before using this machine we recommend that you always check that these knobs have been tightened Place the Abduction motor 5 on right or left depending on the limb be moved The other components to be used depend on the selected movement see page 20 to 25 UNIT TRANSPORT For easy transport of the unit it features 2 wheels 6 anda handle 7 Place the arm support as close as possible to the chair to limit the overall dimensions and help balance the unit Place your foot as indicated to balance the unit You can adjust the height of the handle with knob 8 Centura 6 29 USE OF THE HAND CONTROL ELECTRICAL CONNECTION SAFETY FIRST KINETEC Centura is a type B class device Before connecting the device to the power supply check that the mains voltage matches that shown on the plate 100 240 V 50 60 Hz below switch ON 2 The connectors can be plugged in any one of the sockets the sockets are not
14. anging the movement parameters 4 a The switch 4 has 3 positions N Fully locked position gt The only functions that are possible are reading the operational settings and operating the START STOP function Fully unlocked position All the operational settings can be changed Semi locked position The program can still be changed and the lower and upper movement limits can still be modified the START STOP function is still accessible Double locking soft lockin Simultaneously press the A and VZ keys to lock the hand control the display shows LOCK To unlock the hand control simultaneously press the A and Z keys the display shows UNLOCK You cannot change the parameters if you try the display shows LOCK SOFT We recommend that you lock the hand control when you give it to the patient START STOP REVERSE FUNCTION As with all KINETEC systems KINETEC Centura is equipped with a START STOP REVERSE function Press the stop key of the hand control The movement stops The display reads STOP Press the key of the hand control The movement starts start in the opposite direction and the display reads RUN Caution For optimum safety always give the hand control to the patient before starting the system Procedure to stop the machine To stop the machine s movement Press the i key To switch power off press the ON OFF switch see page 7 Centura 8 29 EN USE
15. ar against all defects of manufacture from the date of purchase by the consumer KINETEC is the only organization able to assess the application of the warranty to its systems The warranty will be considered null and void if the device has been used abnormally or under conditions of use other than those indicated in the user s manual The warranty will also be considered null and void in the event of deterioration or an accident due to negligence inappropriate surveillance or inappropriate maintenance or due to transformation of the equipment or an attempt to repair the equipment Centura 28 29 EN PRODUCT INFORMATION GUIDANCE AND MANUFACTURER S DECLARATION Electromagnetic emissions The KINETEC Centura is intended for use in the electromagnetic environment specified below The customer or the user of the KINETEC Centura should ensure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance The KINETEC Centura uses RF energy only for its Radio electric Frequency emissions Group 1 internal function Therefore its RF emissions are very low CISPR 11 and are not likely to cause any interference in nearby electronic equipment Radio electric Frequency emissions Class B CISPR 11 The KINETEC Centura is suitable for use in all Harmonic emissions Class A establishments including domestic establishments and those IEC 61000 3 2 directly connected t
16. aximum hygiene of the pads Sterilization of the pads if necessary Sterilizer at 134 C during 18 minutes e Desinfection of the pads Washing at 30 C with use of a disinfection solution during the rising cycle e Example of product which can be used Solution Baclinge at 0 125 or Souplanios at 0 125 from ANIOS Laboratory A complete list of distributors in your country is available on request ELIMINATION AND RECYCLING a Packing Packing must be separated from the components plastic and paper cardboard and given to the specific sites from recycling b Kinetec patient pad kit To clean with a product of disinfection then to give it to the specific sites of recycling c Unit It contains electronic components cables aluminium steel and plastic parts When the splint is not operational any more to dismount and separate in groups from materials and to give them to correct unit of recycling or to turn over the machine to Kinetec for destruction Centura 27 29 EN PRODUCT INFORMATION TECHNICAL SPECIFICATIONS Product Weight 22 Kg 48lb Splint dimensions Environment Storage transport conditions Temperature 40 to 70 C 40 to 160 F Relative humidity up to 90 Operating conditions Room temperature 10 to 40 C 50 to 105 F Relative humidity up to 80 Electricity Power supply 100 240 V Frequency 50 60 Hz 56x100x76cm 22 x39 x30 Power consumption 50 VA Angular limits see
17. extern extern Upper limit 25 to 160 25 to1g0 090 25 t0 90 55 to 110 extern extern e Speed 1 to 5 from 50 to 140 per minute e Load 1to6 0 to 900 seconds 15 minutes 0 to 900 seconds 15 minutes No time 00H00 to 24H00 e Extension pause e Flexion pause e Timer Available on Centura 5 or in option Centura 9 29 EN USE OF THE HAND CONTROL om crc aI HOW TO ADJUST THE PARAMETERS OF SINGLE MOVEMENTS Check if the locking switch Te see sne wink ROTATION is in the following STOP 0 60 position oe ABD ADD To choose the movement STOP 90 90 The display shows the new movement selected Or Bee and the default settings of STO 0 60 the upper and lower limits FLEX EXT of this movement i 30 STOP 89 100 The display shows not available Or for Centura 5 30 aan 60 j You dont have the horizontai i To display the lower limit FLEX EXT of the movement u 30 EXT 89 100 The value blinks To change the lower limit FLEX EXT merens bink if necessary 50 EXT 89 100 R FLEX EXT While the value blinks 50 FLEX 89 100 PREE To validate the new value i RTN he AN 9r key A l EA 3 to change if necessary eseanae FLEX EXT 7 SPEED 2 FLEX EXT Loads Or to display pause FLEX EXT The pause value in upper PAUSE HIGH 0S limit of movement blinks To change pause value in upper limit of movement if necessary FLEX EXT The new pause value in upper limit of moveme
18. hoice of Horizontal abduction hor abd Press wy and adjust your parameters see page 10 Or select a program see page 12 Centura 25 29 EN ACCESSORIES Horizontal Abduction Kit Except for Centura 5 Part number to order 4670020020 Head and shoulder rests Part number to order 4670024551 Scales kit Part number to order 4670019726 Transport bag Part number to order 4640001464 CEM kit elbow module E an Part number to order 4621007002 gt P Centura 26 29 EN PRODUCT INFORMATION MAINTENANCE After 2 000 hours of operation KINETEC Centura requires a few lubrication and maintenance operations lubrication of the joints pointer stops and ball screws The need for maintenance is indicated by display of the message SERVICE TIME Mx when the system is switched on Despite that warning you can continue to use your KINETEC by pressing START but you should contact your nearest KINETEC technician to have the maintenance operations conducted as soon as possible When the system is no longer in operating order kindly return it to us with the accessories for destruction TROUBLE SHOOTING A spare parts list and technical catalogue are available to you on request from your KINETEC distributor If after connecting the power supply cable to the power supply and switching on KINETEC Centura e The display does not indicate any information Check that the electrical socket is live using another device
19. houlder depth adjustment e Loosen the knob 5 Slide the entire mechanism e Tighten the knob 5 e Adjust the arm rest STARTING THE UNIT Adjustments for ABDUCTION ADDUCTION with fixed EXTERNAL ROTATION 1 Adjustment of the rotation position Press q and find the right position with the MANUAL MODE see page 17 dy abd add 2 Choice of the abduction adduction motion Press and adjust your parameters see page 10 3 Or select a program see page 12 Adjustments for ABDUCTION ADDUCTION with ASSOCIATED ROTATION 1 Choice of the abduction adduction associated with rotation motion Press and adjust your parameters see page 11 abd add rot 2 Or selecta program see page 12 Adjustments for ROTATION abdiadd 1 Adjustment of the abduction position Press a and find the right position with the MANUAL MODE see page 17 rot 2 Choice of the rotation motion Press they and adjust your parameters see page 11 3 Or select a program see page 12 Centura 21 29 EN INSTRUCTIONS FOR USE STANDARD SETTINGS FOR A MOVEMENT USING ONE MOTOR Possible movement awaa Adjustments for ABDUCTION ADDUCTION with fixed ELBOW EXTENSION FLEXION Vv The KINETEC Centura provides motion from 20 to 160 of abduction During this motion the elbow flexion settings are fixed fe Adjustments for EXTENSION FLEXION R The KINETEC Centura provides motion from 20 to 180 of flexion During this motion the elbow flexion
20. lotomy e Arthrolysis e Synovectomy for Rheumatoid Arthritis e All type of shoulder styffness joint e Clinical Benefits e Breaks the cycle of trauma inflammation and the loss of range of motion e Prevents joint stiffness e Speeds the recovery of post operative range of motion e Maintains the quality of the articular surface e Reduces pain and oedema e Promotes articular cartilage healing e Reduces hospitalization time e Reduces the need for pain medication e Contraindications e Unstable fractures e Spastic paralyses e Uncontrolled infection e The machine are not adapted for patients height more 2m 6 7 or under 1 4m 4 7 Centura 2 29 EN DESCRIPTION COMPLIANCE The KINETEC Centura complies with the standards of Directive 93 42 EEC and bears the CE mark The KINETEC Centura complies with the standards in force IEC 601 1 2 concerning the electromagnetic compatibility of medical devices and IEC 60601 1 concerning electrical safety The KINETEC Centura meets the requirements of the Machinery Directive No 2006 42 EC DESCRIPTION COMPONENTS KINETEC Centura consists of the following components 1 Chair 2 Frame 3 Wheels 4 Transport handle 5 Arm rest knobs 6 Arm rest 7 Hand control support 8 Locking of the right left sliding 9 Locking of the up down sliding 10 Chair locking knob 11 Locking of the abduction motor
21. misee pade 12 Centura 15 29 EN USE OF THE HAND CONTROL Comments e When a program has been deleted the display shows PROGRAM 11 EMPTY e The values shown in the Display column are examples They actually depend on the stored programs Program table Timer lower limit Upper limit G speed force pause pause timer limit Pause on Pause on upper limit Load Speed Lower limit limit hor abd abd add A abd add rot Movement type a ey SK INTO ST TOTO TM Oo oO POE 10 11 12 13 14 15 16 Centura 16 29 EN USE OF THE HAND CONTROL HOW TO DEFINE THE UPPER AND LOWER MOVEMENT LIMITS At the start of a session The MANUAL MODE is a way to set within the tolerance of a patient at the beginning of a session Proceed as below KINETEC CENTURA V x x Check if the locking switch is in the following Please wait position Switch the unit on Movement verif To select the MANUAL MODE for upper limits by continually holding or pressure on the key Movement verif Switch the unit on Motor M1 M2 FLEX EXT 30 STOP 89 100 To select the MANUAL The unit is moving to the MODE for upper limits by FLEX EXT upper limit of the continually holding 30 MANUAL 89 100 movement pressure on the key continuous press TA FLEXIEXT een y 30 MANUAL 150 150 press recorded To select the manual FLEX EXT ae tohe mode for lower limits E ioe 30 MANU
22. n from 50 to 90 30 60 100 Abdcution i 50 90 Rotation Comments e Speed load pauses and timer are the same for both of the movement components The setting is the same as for a single movement e Pauses can be set at the lower and or the upper limits of the abduction movement e You will have successive displays of abduction movement limits or associated rotation movement by repeatedly pressing the synchronized movement button e You cannot change the settings while the machine is running e BY PASS Mode is not available in synchronization mode Centura 11 29 EN USE OF THE HAND CONTROL USING PROGRAMS program The KINETEC Centura allows you to store up to 16 programs including the type of movement ROM speed load pauses and timer The original parameter values of the program are empty These values can be modified and recorded at any time see How to enter a program p 14 To select a program Check if the locking switch is in the following position f ROTATION or To stop the unit stop 0 STOP 0 60 To access the program PROGRAM 1 The program number mode OTS EMPTY blinks To change the program if PROGRAM 3 The new program number necessary FLEX EXT 60 blinks To exit and validate the FLEX EXT ae cia kath i Selected proglar STOP 0 60 ave been recorded in program 3 To exit without validation ROTATION Back to the starting of selected program STOP 0 60 parameters Start the unit FL
23. nt PAUSE HIGH 10S blinks To validate and display the pause value in lower limit of movement FLEX EXT The pause value in lower PAUSE LOW 0S limit of movement blinks value BKUS INNO eve lover link Gt movement movement if necessary PAUSE LOW 15S blinks To validate the new value press another key FLEX EXT The unit is ready to start or wait more than 3 seconds The display shows the 50 STOP 89 100 with the new parameters selected movement Centura 10 29 EN USE OF THE HAND CONTROL HOW TO SET THE SYNCHRONIZED MOVEMENT PARAMETERS abd add rot Check if the locking switch To stop the unit FLEX EXT is in the following 50 STOP 15 100 position TA The indication ABD blinks and the display To select the combined SYNC ABD ROT shows the values for the movement 30 STOP 15 100 abduction movement abd add rot To change it proceed as for a single movement The indication ROT blinks and the display To press a second time on heed SYNC ABD ROT shows the values for the the key 0 STOP 15 60 rotation movement abd add rot To change it proceed as for a single movement Synchronization rules e The degress of rotation are lower than or equal to the degrees of abduction 1 of abduction means 1 of rotation e When the degrees of rotation are lower than the degrees of abduction the synchronization applies to the upper degrees of the movement Example abduction from 30 to 100 rotatio
24. o the public low voltage power supply Voltage fluctuations Flicker emissions A network that supplies buildings used for domestic purposes IEC 61000 3 3 Complies Electromagnetic immunity The KINETEC Centura is intended for use in the electromagnetic environment specified below The customer or the user of the KINETEC Centura should ensure that it is used in such an environment Immunity test E Compliance level Electromagnetic environment guidance 2 kV 4 kV 6 kV 2 kV Electrostatic discharge contact contact At the interruption there is a reset of the Electrical fast transient for power supply lines for power supply lines ESD KINETEC Centura After turning on push IEC 61000 4 2 2 KV 4 kV 8 kV 2 kV 4 kV 8kV START to begin the session alr air 2 kV 2 kV Mains power quality should be that of a typical 0 5 kV 1 kV 2 kV between line and earth 0 5 kV 1 kV 2 kV between line and earth Oe aa 1 kV 1 kV commercial or hospital environment for input output for input output 0 5 kV 1 kV 0 5 kV 1 kV Surge petweerlines petweenilines Mains power quality should be that of a typical IEC 61000 4 5 commercial or hospital environment Voltage interruptions CEI 61000 4 11 lt 5 Ur gt 95 dip in Ur for 5 seconds lt 5 Ur gt 95 dip in U7 for 5 seconds At the interruption there is a reset of the KINETEC Centura After turning on pu
25. ram To switch off the unit Check if the locking switch is in the following position J To press the two keys at the same time to switch the unit on Then KINETEC CENTURA V x x PROGRAM 3 EMPTY Welcome text during 3 seconds The program number blinks To change the program if necessary PROGRAM 10 30 SYNC ABD 100 The new program number blinks To choose the movement PROGRAM 10 30 ABD ADD 90 Or PROGRAM 10 0 ROTATION 60 Or PROGRAM 10 30 FLEX EXT 100 PROGRAM 10 Or 30 SYNC ABD 100 The display indicates the selected movement the program number blinks again Or for Centura 5 PROGRAM 10 30 ABD HOR 60 The display shows not available if you don t have the horizontal abduction module To display the lower limit of the movement PROGRAM 10 SYNC ABD 100 The value blinks To change the lower limit of the movement if necessary PROGRAM 10 SYNC ABD 100 The new value blinks To validate the new value press another key PROGRAM 10 SYNC ABD 100 PROGRAM 10 TIMER 00H01MIN PROGRAM 10 SPEED 1 PROGRAM 10 LOAD The value blinks Press the A or rey to change if necessary Or display pauses PROGRAM 10 PAUSE HIGH 0S The pause value in upper limit of movement blinks To change pause value in upper limit of movement if necessary Centura PROGRAM 10 PAUSE HIGH 15S 14 29 The new pause
26. rt e Tighten the knob 4 f Shoulder depth adjustment e Loosen the knob 5 Slide the entire mechanism e Tighten the knob 5 e Adjust the arm rest STARTING THE UNIT Adjustments for ABDUCTION ADDUCTION with fixed ELBOW EXTENSION FLEXION 1 Adjustment of the elbow flexion position Unscrew the knob 6 and adjust the flexion as appropriate 2 Choice of the abduction adduction motion Press 244 and adjust your parameters see page 10 3 Or select a program see page 12 Adjustments for EXTENSION FLEXION with fixed ELBOW EXTENSION FLEXION 1 Adjustment of the elbow flexion position Unscrew the knob 6 and adjust the flexion as appropriate flex 2 Choice of flexion extension motion Press amp and adjust your parameters see page 10 w 3 Or select a program see page 12 Centura 23 29 EN INSTRUCTIONS FOR USE ADJUSTMENTS FOR HORIZONTAL ABDUCTION Possible movement forabd Horizontal Abduction Available on Centura 5 or in option Contact your nearest KINETEC distributor The KINETEC Centura provides motion from 30 to 110 of horizontal abduction PARTS NEEDED e Chair e Horizontal abduction splint e Hand control os 2 ASSEMBLING THE PART Assemble the horizontal abduction column to the unit Assemble the motor M1 to the column You need to see the Use the colour code to turn the column in the right position Fasten the red point on the M1 part for a right side set up bl
27. s in the wrist support c Hand strap Turn the hand support so the rod is up With the hook material on the strap facing up thread the D ring end of the strap under the rod d Biceps pad Position it on the biceps support with the strap and D ring toward the biceps support Thread the strap and D ring through the slots in the biceps support After patient has been placed in the supports secure straps by threading them through the D rings and securing the hook closures Be sure the straps are snug but not so tight that they impair circulation The pads are intended for single patient use only Pads for Kinetec Centura Shoulder CPM Machine 5315 0130 Centura 19 29 EN INSTRUCTIONS FOR USE STANDARD SETTINGS FOR A MOVEMENT USING BOTH MOTORS Possible movement awaa Adjustments for ABDUCTION ADDUCTION with fixed EXTERNAL ROTATION Va The KINETEC Centura provides motion from 20 to 160 of abduction The rotation position can be adjusted between 60 and 90 Adjustments for ABDUCTION ADDUCTION with ASSOCIATED ROTATION The KINETEC Centura provides motion from 20 to 160 of abduction associated with 120 of rotation in maximum abd add rot Adjustments for ROTATION motion roe The KINETEC Centura provides motion from 60 of internal rotation to 90 of external rotation During this motion the abduction settings are fixed PARTS NEEDED e Chair e Abduction rotation splint e Hand control HD w
28. sh START to begin the session Voltage dips and voltage variations on power supply input lines CEI 61000 4 11 lt 5 Ur gt 95 dip in Ur for 0 5 cycle 40 Ur 60 dip in Ur for 5 cycles 70 Ur 30 dip in U for 25 cycles lt 5 Ur gt 95 dip in Ur for 0 5 cycle 40 Ur 60 dip in Ur for 5 cycles 70 Ur 30 dip in U for 25 cycles Mains power quality should be that of a typical commercial or hospital environment Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3A m 3A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment NOTE UT is the a c mains voltage prior to application of the test level Centura 29 29 EN Xkinetec MONS RRolyan Sammons Preston Rolyan A Patterson Medical Company W68 N158 Evergreen Blvd Cedarburg WI 53012 Phone 800 228 3693 Fax 262 387 8748
29. ue for left screw Several positions are possible to adjust the plan of the motion horizontal to more or less inclined Insert the elbow support holder into the extremity of the M1 part The required colour point red for right and blue for left needs to Insert the elbow splint into the elbow support holder respecting be positioned internally For safety reason this part cannot be the colour code removed from the lower part and must be so lifted when removed Centura 24 29 EN INSTRUCTIONS FOR USE PATIENT SET UP Make sure the straps are clean Put the unit in the position that is the most comfortable for the patient Position the patient in the chair in a Slide the arm supports toward the Adjust the lengths comfortable position and supporting patient and put the arm in the a arm the affected arm supports b forearm Secure pads ADJUSTING THE SHOULDER JOINT AXIS c Vertical adjustment e Loosen the 2 knobs 1 e With the handle 2 adjust the height of the entire mechanism e Tighten the 2 knobs 1 d Side to side adjustment e Loosen the knob 3 Slide the entire mechanism Tighten the knob 3 e Shoulder depth adjustment e Loosen the knob 4 Slide the entire mechanism e Tighten the knob 4 f Plan of the motion from horizontal to inclined e Loosen the knob 5 e Rotate the arm support e Tighten the knob 5 e Adjust the arm rest STARTING THE UNIT e C
30. value in upper limit of movement blinks EN USE OF THE HAND CONTROL HOW TO MODIFY PROGRAMS PROGRAM MODE continued program To validate and display PROGRAM 10 The pause value in lower the pause lower limit of rats h vement PAUSE LOW 0S limit of movement blinks To change the pause PROGRAM 10 The new pause value in lower limit of movement if lower limit of movement necessary PAUSE LOW 10S blinks The program number To validate and display of blinks and the display the combined rotation 0 eee indicates the rotation setting values combined with abduction To change the lower limit G PROGRAM 10 The value blinks of the movement 0 SYNCROT 60 To change the lower limit PROGRAM 10 The new value blinks if necessary 50 SYNC ROT 60 To validate and display PROGRAM 10 the upper limit of the The value blinks a 50 SYNC ROT 60 The new value blinks To change the upper limit PROGRAM 10 see page 11 for more if necessary 50 SYNC ROT 75 information about combined movement To record the program 10 program Se PROGRAM 10 The program 10 has been SAVING recorded and the display Then indicates the next program PROGRAM 11 so you can change EMPTY another program PROGRAM 10 The program 10 has been OR CLEARING recorded and the display indicates the next program To cancel the program PROGRAM 11 so you can change EMPTY another program ee Pade mae KINETEC To use the modified Peun i CENTURA V x x progra