Home

Hart Scientific 9101 USER

image

Contents

1. Set up Place the 9101 zero point dry well upright on a flat stable surface The area around the instrument must be clear to allow adequate air circulation For best results operate the dry well in room of constant temperature between 18 to 25 C 64 to 77 F free from drafts Stable ambient temperatures allow stable well gradients and less drift Turn on the power to the calibrator by toggling the power switch on The fan should begin quietly blowing air through the instrument The amber LED will light when the unit is energized The green LED will light when the unit is ready to use The unit is not at the correct temperature until the green LED is illuminated Power Connect the power cord provided into the instrument and plug the other end into a power socket of specified voltage Normally this will be 115 VAC 10 50 60 Hz If specified upon ordering the instrument may optionally be configured to operate at 230 VAC 10 5 General Operation 5 2 5 3 General Operation Set up Place the 9101 zero point dry well upright on a flat stable surface The area around the instrument must be clear to allow adequate air circulation For best results operate the dry well in room of constant temperature between 18 to 25 C 64 to 77 F free from drafts Stable ambient temperatures allow stable well gradients and less drift Cooling Turn on power by pressing the switch on the top of the instrument to the on position Notice th
2. Scientific Inc 799 E Utah Valley Drive American Fork UT 84003 9775 USA Phone 1 801 763 1600 Telefax 1 801 763 1010 E mail support hartscientific com Fluke Nederland B V Customer Support Services Science Park Eindhoven 5108 9101 User s Guide 5692 EC Son NETHERLANDS Phone 31 402 675300 Telefax 31 402 675321 E mail ServiceDesk fluke nl Fluke Int l Corporation Service Center Instrimpex Room 2301 Sciteck Tower 22 Jianguomenwai Dajie Chao Yang District Beijing 100004 PRC CHINA Phone 86 10 6 512 3436 Telefax 86 10 6 512 3437 E mail xingye han fluke com cn Fluke South East Asia Pte Ltd Fluke ASEAN Regional Office Service Center 83 Clemenceau Avenue 15 15 06 Ue Square 239920 SINGAPORE Phone 65 737 2922 Telefax 65 737 5155 E mail antng singa fluke com When contacting these Service Centers for support please have the follow ing information available Model Number Serial Number Voltage Complete description of the problem 2 Introduction 2 Introduction The Hart model 9101 zero point dry well is a precision instrument which maintains a constant temperature at O C It can be used to provide a stable zero reference for thermocouples or for calibrating temperature probes The 9101 has three wells so that multiple probes may be calibrated and com pared at once The 9101 uses solid state electronics to control the temperature of an alumi num block The block temp
3. any time to agree more closely with a particular standard thermometer To calibrate the instrument insert the standard probe into the center well as discussed in Section5 3 and allow the temperature to stabilize At the back of the 9101 is a small hole covered by a calibration void if removed sticker which allows access to a trimmer potentiometer used to set the temperature Insert a small adjustment tool or screwdriver through the hole and slowly turn the trimmer until the standard measures 0 0 C Allow a few minutes between adjustments to allow time for the temperature to change and stabilize 15 8 Maintenance Maintenance The calibration instrument has been designed with the utmost care Ease of operation and simplicity of maintenance have been a central theme in the product development Therefore with proper care the instrument should re quire very little maintenance Avoid operating the instrument in an oily wet dirty or dusty environment If the outside of the instrument becomes soiled it may be wiped clean with a damp cloth and mild detergent DO NOT use harsh chemicals on the surface which may damage the paint It is important to keep the well of the calibrator clean and clear of any foreign matter DO NOT use fluid to clean out the well The dry well calibrator should be handled with care Avoid knocking or dropping the calibrator DO NOT slam the probe stems into the well This type of action can cause
4. into the calibrator causing damage to the calibra tor or binding and damage to your probe Comparison Calibration Comparison calibration involves testing a probe against a similar reference probe The advantage to this method is that better accuracy can be achieved since errors due to dry well inaccuracy stem effect and drift can be reduced After inserting the probes to be calibrated allow sufficient time for the probes to settle and the temperature of the dry well to stabilize It is best if both the reference probe and the probe under test are the same size and construction Using probes with different lengths diameters and materials will have different stem effects causing an unknown temperature difference All dry wells have horizontal and vertical gradients that change with temperature This is an unknown variable which can be factored out if probes are the same type length diameter and material Probes should be inserted to the same depth in the well The following procedure can be used to calibrate a probe against a refer ence while eliminating error due to temperature gradients between wells e Place the reference probe in one well e Place the probe to be calibrated the unit under test UUT in another well e With the reference inserted into one well and the probe under test in serted into a second well make measurements of each e Swap the locations of the reference probe and probe under test Allow plenty o
5. the instrument may be impaired The following definitions apply to the terms Warning and Caution Warning identifies conditions and actions that may pose hazards to the user Caution identifies conditions and actions that may damage the in strument being used 1 2 1 A Warnings To avoid personal injury follow these guidelines Before initial use or after transport or after storage in humid or semi humid environments or anytime the instrument has not been en ergized for more than 10 days the instrument needs to be energized for a dry out period of 2 hours before it can be assumed to meet all of the safety requirements of the IEC 1010 1 If the product is wet or has been in a wet environment take necessary measures to remove mois ture prior to applying power such as storage in a low humidity temper ature chamber operating at 50 degree centigrade for 4 hours or more Use only a grounded AC mains supply of the appropriate voltage to power the instrument The dry well requires a maximum of 1 amp 50 60 Hz 115 230 VAC 10 DO NOT connect this unit to a non grounded non polarized outlet DO use a ground fault interrupt device HIGH VOLTAGE is used in the operation of this equipment SEVERE INJURY or DEATH may result if personnel fail to observe safety pre cautions Before working inside the equipment turn power off and dis connect power cord DO NOT use this unit in environments other than t
6. 115 VAC 10 1 A or 230 VAC 10 0 5 A 50 60 Hz 125 W Dimensions 12 25 H x 8 5 W x 5 75 D 311 x 216 x 146 mm Environmental Conditions Although the instrument has been designed for optimum durability and trouble free operation it must be handled with care The instrument should not be operated in an excessively dusty or dirty environment Maintenance and cleaning recommendations can be found in the Maintenance Section of this manual The instrument operates safely under the following conditions e temperature range 5 50 C 41 122 F e ambient relative humidity maximum 80 for temperature lt 31 C de creasing linearly to 50 at 40 C e pressure 75kPa 106kPa e mains voltage within 10 of nominal e vibrations in the calibration environment should be minimized e altitude does not effect the performance or safety of the unit 9101 User s Guide 3 3 Warranty Hart Scientific Inc Hart warrants this product to be free from defects in material and workmanship under normal use and service for a period as stated in our current product catalog from the date of shipment This war ranty extends only to the original purchaser and shall not apply to any product which in Hart s sole opinion has been subject to misuse alteration abuse or abnormal conditions of operation or handling Software is warranted to operate in accordance with its programmed in structions on appropriate Hart products It is not warran
7. 2 Calibration of Multiple Probes o 14 6 2 Dry Well Characteristics 14 6 2 1 Vertical Gradient lt c cos 3a e dG aie oe a a ea eS 14 6 2 2 Stabilization and Accuracy o o a 14 7 Calibration Procedure o 15 8 Maintenance 1 5666 eae wh ewe secan 17 9 Troubleshooting cocer a ee A 19 1 Before You Start 1 1 Before You Start Symbols Used Table 1 lists the International Electrical Symbols Some or all of these sym bols may be used on the instrument or in this manual Table 1 International Electrical Symbols Symbol Description NU AC Alternating Current PT AC DC Battery Mm Mm O gt l ei CE Complies with European Union Directives DC Double Insulated Electric Shock Fuse PE Ground Hot Surface Burn Hazard Read the User s Manual Important Information Off On 9101 User s Guide Symbol Description G Canadian Standards Association us OVERVOLTAGE Installation CATEGORY II Pollution Degree 2 per CATI IEC1010 1 refers to the level of Impulse Withstand Voltage protection provided Equipment of OVERVOLTAGE CATEGORY II is energy con suming equipment to be supplied from the fixed installation Exam ples include household office and laboratory appliances C TIC Australian EMC Mark 1 2 Safety Information Use this instrument only as specified in this manual Otherwise the protec tion provided by
8. a shock to the sensor If a hazardous material is spilt on or inside the equipment the user is responsible for taking the appropriate decontamination steps as out lined by the national safety council with respect to the material If the mains supply cord becomes damaged replace it with a cord with the appropriate gauge wire for the current of the instrument If there are any questions call a Hart Scientific Authorized Service Center for more information Before using any cleaning or decontamination method except those recommended by Hart users should check with a Hart Scientific Au thorized Service Center to be sure that the proposed method will not damage the equipment If the instrument is used in a manner not in accordance with the equipment design the operation of the dry well may be impaired or safety hazards may arise 17 9 Troubleshooting 9 Troubleshooting If the amber LED on the top of the unit is off e Check that there is power to the unit If there is power to the unit con tact a Hart Scientific Authorized Service Center for further assistance 19
9. at the amber POWER indicator light turns on and the fan begins circulating air The dry well will begin to cool the well until it reaches 0 C and stabilizes Typically it takes 15 to 30 minutes for the block to reach O and stabilize longer at higher ambient temperatures When the block has reached O and stabilized the green READY indicator light will turn on and the instrument is ready for use Measurement When the instrument is ready insert the probe into one of the three well holes at the top of the instrument The probe should be inserted the full depth of the well since the temperature is most accurate at the bottom The probe must make good contact with the bottom of the well A little heat conduction grease on the probe may help It also helps to apply a small constant sideways and downwards force on the probe After inserting the probe allow approximately 10 minutes for the temperature to stabilize At this point the probe may be measured for accuracy or used as a stable zero reference 6 Test Probe Calibration 6 1 6 1 1 Test Probe Calibration For optimum accuracy and stability allow the calibrator to warm up for 10 minutes after power up and then allow adequate stabilization time after reaching the set point temperature After completing calibration allow the block to cool before switching the power off Calibration Methods Never introduce any foreign material into the probe hole of the insert Fluids etc can leak
10. erature is monitored with a high sensitivity ther mistor sensor The controlling circuit adjusts the current to thermoelectric modules which cool the block maintaining a stable temperature Since the current to the cooling modules is constant DC there is no problem with in duced EMF noise in the user s probes or thermocouples due to switching Figure 1 9101 Zero Point Calibrator 9101 User s Guide transients of the cooling circuits Because the temperature control is done with solid state electronics the 9101 is rugged and stable The zero point dry well will not be damaged by subfreezing temperatures during shipping or storage The 9101 is calibrated at the factory for NIST traceable accuracy It may also be easily calibrated in the field to the user s own standard 3 Specifications and Environmental Conditions 3 3 1 3 2 Specifications and Environmental Conditions Specifications Temperature 0 C Accuracy 0 02 C typical 0 05 C maximum 18 C 25 C ambient Stability 0 005 C Temperature 0 005 C C amb Coefficient Stabilization Approximately 30 minutes the ready lamp indicates stable control at Time 0 C Operable 5 to 27 C 40 to 80 F temperature range Accuracy 18 to 25 C 64 to 77 F temperature range Cooling time 15 to 30 min Depending on ambient Well 3 wells 2 outside wells 25 dia X 6 deep center well 280 dia x 6 deep Power
11. f time for thermal settling e Make another set of measurements of the reference probe and the probe under test e Average the two measurements of the reference probe Average the two measurements of the probe under test Averaging the two mea surements in this way eliminates error due to temperature gradients between the two wells 13 9101 User s Guide 14 6 1 2 6 2 6 2 1 6 2 2 e Compare the averaged measurement of the probe under test with the averaged measurement of the reference probe e For best results repeat the test several times at the same temperature and at different temperatures This method can be used with different types of probes but the user must determine the uncertainty of the measurement Calibration of Multiple Probes Fully loading the calibrator with probes increases the time required for the temperature to stabilize after inserting the probes Be sure that the tempera ture has stabilized before starting the calibration Multiple probes may be calibrated simultaneously using either the direct or comparison calibration method Stem effect will cause less error in the com parison calibration method than with the direct calibration method Dry Well Characteristics Understanding the thermal characteristics of the dry well calibrator can help you achieve the best accuracy and efficiency possible Vertical Gradient There is a temperature gradient vertically in the test well The hea
12. hose listed in the user s manual 1 Before You Start 1 2 2 1 3 e DO NOT use this unit for any application other than calibration work e Follow all safety guidelines listed in the user s manual e Calibration Equipment should only be used by Trained Personnel A Cautions To avoid possible damage to the instrument follow these guidelines e Operate the instrument in room temperatures between 5 50 C 41 122 F Allow sufficient air circulation by leaving at least 6 inches of space between the instrument and nearby objects Overhead clear ance needs to allow for safe and easy insertion and removal of probes for calibration e The dry well is a precision instrument Although it has been designed for optimum durability and trouble free operation it must be handled with care The instrument should not be operated in excessively wet oily dusty or dirty environments It is important to keep the wells of the instrument clean and clear of any foreign matter Do not operate near flammable materials e DO NOT use fluids to clean out the well e Jfa mains supply power fluctuation occurs immediately turn off the in strument Power bumps from brown outs and black outs could damage the instrument Wait until the power has stabilized before re energiz ing the instrument Hart Scientific Authorized Service Centers Please contact one of the following authorized Service Centers to coordinate service on your Hart product Hart
13. meet SCIENTIFIC A Fluke Company 9101 Zero Point Dry Well User Manual Rev 4A0801 Hart Scientific LLC 799 E Utah Valley Drive American Fork UT 84003 9775 USA Phone 1 801 763 1600 Telefax 1 801 763 1010 E mail support hartscientific com www hartscientific com Subject to change without notice e Copyright 1991 Printed in USA Rev 4A0801 Table of Contents 1 Before You Start 1 1 1 symbols Used gt o e a m e o ceros 1 1 2 Safety Informati0D a 2 1 2 1 WW ATIINIGS La eh eh etc de aes eae cer tee Gap eee oo she Ga deca cE es 2 1 2 2 GaUtlOnS z a ee Sk ale eek ee A ae ek kee eS 3 13 Hart Scientific Authorized Service Centers 3 2 MITOGUCHON ss s ca eon e a oe we ee a 5 3 Specifications and Environmental Conditions 7 34 SpeciiCatOns ice va ee oe ee ee EEO T AA 7 3 2 Environmental Conditions 7 3 3 Walranty s ra 0884 a ee eee ee ede ba ee a 8 4 Quick Start ewe eo a a a es eh A 9 41 Unpacking 2 6045 86485454 8 65 e O a e 9 A a cca MEG A Eee ee ea a a r 9 43 POWGr sa sr a a 9 5 General Operation 2 2 02 ee eee es 11 Dil DEUDA once eo ee oe ee ae 11 dudo COOMO s naska paa ee Ee Ge ee AP ee oS 11 5 3 Measurement 11 6 Test Probe Calibration 05858685 13 6 1 Calibration Methods 004 13 6 1 1 Comparison Calibration o o 13 6 1
14. ted to be error free Hart s obligation under this warranty is limited to repair or replacement of a product which is returned to Hart within the warranty period and is deter mined upon examination by Hart to be defective If Hart determines that the defect or malfunction has been caused by misuse alteration abuse or abnormal conditions or operation or handling Hart will repair the product and bill the purchaser for the reasonable cost of repair To exercise this warranty the purchaser must forward the product after call ing or writing a Hart Scientific Authorized Service Center see Section 1 3 for authorization The Service Centers assume NO risk for in transit damage THE FOREGOING WARRANTY IS PURCHASER S SOLE AND EXCLUSIVE REM EDY AND IS IN LIEU OF ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OR MECHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE OR USE HART SHALL NOT BE LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSS WHETHER IN CONTRACT TORT OR OTHERWISE 4 Quick Start 4 2 4 3 Quick Start Unpacking Unpack the dry well carefully and inspect it for any damage that may have occurred during shipment If there is shipping damage notify the carrier immediately Verify that the following components are present e 9101 Dry well e 2130 Well tube sizing kit e Power Cord e User s Guide e Report of Calibration e Calibration Label
15. ter has been applied to the block in such a way as to compensate for nominal heat losses out of the top of the dry well and minimize vertical temperature gra dients However actual heat losses will vary depending on the number and types of probes inserted into the calibrator and the block temperature For best results insert probes the full depth of well Stabilization and Accuracy The stabilization time of the dry well calibrator will depend on the condi tions and temperatures involved Typically the dry well will be stable to 0 1 C within 5 minutes of reaching the set point temperature as indicated by the display Ultimate stability will be achieved 10 to 20 minutes after reaching the set temperature Inserting a cold probe into a well will require another period of stabilization depending on the magnitude of the disturbance and the required accuracy For example inserting a 4 inch diameter room probe at temperature into a sleeve at 300 C will take approximately 5 minutes to be within 0 1 C of its set point and will take 10 minutes to achieve maximum stability Decreasing the time required for the calibration process can be accom plished by knowing how soon to make the measurement It is recom mended that typical measurements be made at the desired temperatures with the desired test probes to establish these times 7 Calibration Procedure Calibration Procedure The user may easily calibrate the 9101 zero point dry well at

Download Pdf Manuals

image

Related Search

Related Contents

CURVAMATIC-2  Philips LED solutions 129466000KX1    取扱説明書 NS-31SDR 録画機能付低照度カメラ Ver.1307  1 - Jvc  vRanger Pro  n°180 - Hérault  偏芯座付ボルト36、PZ偏芯座付ボルト36 取扱説明書  A135.009 - Bartscher GmbH  DATA Count S-25  

Copyright © All rights reserved.
Failed to retrieve file