Contents

1. Ifcylinder is not full after second attempt to fill contact the provider Cylinder Fillport Cover Connector Fillport Cover FIGURE 4 4 Fillport Covers Part No 1100873 28 HomeFill 1 Compressor SECTION 4 COMPRESSOR OPERATION Turning the Compressor On NOTE For this procedure refer to FIGURE 4 5 on page 30 1 Make sure the concentrator is On Refer to the concentrator Owner s Manual CAUTION The concentrator flow rate to the patient MUST be set at maximum 3 L min or less for Platinum 5 XL and 5 L min for Platinum 10 Otherwise the oxygen level to the compressor will be lt 90 and the compressor will not fill the cylinder 2 Set the concentrator flow rate to the patient to 3 L min or less for Platinum 5 XL and 5 L min for Platinum 10 Refer to the concentrator Owner s Manual 3 Make sure the cylinder is connected to the compressor Refer to Connecting the Cylinder to the Compressor on page 23 4 Push the power switch on the control panel to the On 1 position 5 Examine the control panel The following sequence of events should occur A 0 3 minutes The O Below Normal YELLOW light will be lit while the compressor warms up B After three minutes The FILLING GREEN light will be on while the cylinder is filling C The FULL GREEN light will be on when the cylinder is finished filling Proceed to STEP 6 NOTE The O BELOW NORMAL
2. Make sure the pressure test on cylinder has not expired Instruct the user on the safe operation of the concentrator and review all warnings Instruct the user on the safe operation of the II compressor and review all warnings Leave a full cylinder of oxygen with the user Leave a copy of all warnings and filling instructions with the user HomefFill 1 Compressor 6 Part No 1100873 HOMEFILL Il INFORMATION HOMEFILL Il INFORMATION HomefFill is an accessory device to the oxygen concentrator that enables the patient to refill custom portable oxygen cylinders for personal use NOTICE The HomeFill Il compressor is for use with specially designed Invacare Platinum 5 10 XL HomeFill 11 compatible oxygen concentrators The oxygen concentrator can be set to the following maximum flowrates when operating with HomeFill Il compressor Platinum 5 and XL 3 L min or less Platinum 10 5 L min or less To determine if your oxygen concentrator is HomeFill compatible examine the lower right corner of the rear of the oxygen concentrator for an outlet fitting Refer to the illustration below TOP VIEW Outlet Fitting Power Cord Concentrator HomeFill 1 Compressor 7 Part No 1100873 SECTION 1 SAFETY INSTRUCTIONS SECTION I IMPORTANT SAFETY INSTRUCTIONS A WARNING SECTION IMPORTANT SAFETY INSTRUCTIONS contains important informatio
3. Double Insulated Product ETL Certified Complying with UL73 and 01097 US ETL Certified Complying with CAN CSA 22 2 No 68 Canada For household use only TYPICAL PRODUCT PARAMETERS COMPRESSOR Operating temperature 50 to 95 F 10 to 35 20 to 60 non condensing humidity Storage temperature 10 to 150 F 23 to 65 15 to 95 humidity Input Pressure Required 14 21 psi Input Flow Required 2 L min Input O Required gt 90 O Width 20 4 inches Height 15 inches Depth 16 inches Weight 33 Ibs Shipping Weight Part No 1100873 39 Ibs 5 HomeFill 1 Compressor LABEL LOCATION Keep all connections free of oil and grease to avoid violent ignition See Owner s Manual before cleaning Pm 1075220 Rev c 1004 DO NOT REMOVE UNIT COVER This unit contains moving and high pressure components Manufacturer Service Only rs wiu Keep all connections free of oil and grease See Owner s Manual before cleaning 2 Cylinders filled on this unit must be labeled Contents of 22 this cylinder are the product of an oxygen concentrator DEALER INFORMATION NOTE Invacare recommends leaving a full cylinder of oxygen with the patient after setting up the HomeFill II compressor Before leaving a HomeFill Compressor with a user the following checklist MUST be completed Q Q
4. INITIAL SETUP Transporting the Compressor A WARNING NEVER transport a compressor with a cylinder connected to or stored on the compressor otherwise injury or damage can occur CAUTION When transporting the compressor be careful not to bump the compressor or the connectors into obstacles Otherwise damage to the compressor may occur NOTE For this procedure refer to FIGURE 2 1 1 There are two methods to safely transport the compressor A One hand Grasp the hand grip at the bottom rear of the compressor Hand Grip Hand Grip Two hands Grasp the hand grips on the Hand Grip sides of compressor FIGURE 2 1 Transporting the Compressor Installing the Power Cord NOTE For this procedure refer to FIGURE 2 2 on page 14 1 Ensure the compressor is hard flat stable surface such as a table or the Ready Rack 2 Make sure the power switch is in the Off O position 3 Plug the power cord into the electric connector in the back of the compressor HomeFill 1 Compressor 13 Part No 1100873 SECTION 2 INITIAL SETUP 4 Plug the opposite end of the power cord into the wall outlet Wall Outlet Compressor REAR Electrical Connection Power Cord NOTE This product has a polarized plug one blade is wider than the other As a safety feature this plug will fit in a polarized ou
5. PORE S GREEN DOTS ARE VISIBLE FIGURE 4 Resetting Connector Fillport 5 Grasp the cylinder regulator assembly in the area behind the cylinder gauge FIGURE 4 2 6 Position the cylinder in the compressor cradle FIGURE 4 2 7 Align the cylinder fillport with the connector fillport FIGURE 4 2 Part No 1100873 24 HomeFill 1 Compressor SECTION 4 COMPRESSOR OPERATION 8 Pull UP on the outer ring sleeve of the connector fillport while pushing DOWN on the cylinder regulator assembly to couple the cylinder fillport into the connector fillport FIGURE 4 2 NOTE The cylinder is properly connected when an audible click is heard HomeFill 1 Compressor 25 Part No 1100873 SECTION 4 COMPRESSOR OPERATION Compressor Cylinder Regulator Assembly Cylinder Fillport Connector Fillport Outer Ring Sleeve Cylinder Gauge PULL UP Regulator Regulator Cylinder Flow Dial Cylinder Outlet Cannula Gauge Connection Cylinder Fillport Power Switch Connector Fillport Compressor Grade Outer Ring Sleeve NOTE Fillport covers not shown for clarity FIGURE 4 2 Connecting the Cylinder to the Compressor Part No 1100873 26 HomeFill 1 Compressor SECTION 4 COMPRESSOR OPERATION Disconnecting the Cylinder from the Compressor A WARNING NEVER use tools of any kind to connect disconnect the cylinder and the compressor Otherwise severe injury
6. SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGE OR LIMITATION OF HOW LONG AN IMPLIED WARRANTY LASTS SO THE ABOVE EXCLUSION AND LIMITATION MAY NOT APPLY TO YOU THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE PROVINCIAL LAWS AND REQUIREMENTS Invacare Corporation www invacare com USA Invacare the Medallion Design One Invacare Way Yes You Can are registered Elyria Ohio USA trademarks of Invacare Corporation 44036 2125 Platinum and HomeFill are trademarks Yes you can 800 333 6900 of Invacare Corporation 2006 Invacare Corporation Technical Services 800 832 4707 Part No 1100873 Rev G 11 01 06
7. the plug not the cord 2 Remove the filter 3 Clean the filter with a vacuum cleaner or wash in warm soapy water and rinse thoroughly gt Dry the filter thoroughly before reinstallation Filter ie Sa EGY FIGURE 7 1 Cleaning the Compressor Filter HomeFill 1 Compressor 37 Part No 1100873 LIMITED WARRANTY Cabinet A WARNING Push the power switch to the Off O position and unplug the compressor before cleaning To avoid electrical shock DO NOT remove cabinet 1 Clean the cabinet with a mild household cleaner and non abrasive cloth or sponge LIMITED WARRANTY NOTE THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4 1975 This warranty is extended only to original purchaser who purchases this product when new and unused from Invacare Corporation or a dealer This warranty is not extended to any other person or entity and is not transferable or assignable to any subsequent purchaser or owner Coverage under this warranty will end upon any such subsequent sale or other transfer of title to any other person This warranty gives you specific legal rights and you may also have other legal rights which may vary from state to state Invacare warrants the HomeFill II Compressor to be free from defects in materials and workmanship for a period of one 1 year from date of purchase from Invacare or a dealer with a copy of the selle
8. 1 Compressor 3 Part No 1100873 SPECIAL NOTES SPECIAL NOTES Check all parts for shipping damage In case of damage DO NOT use Contact Invacare for further instruction Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage Refer to the following table for definitions of the signal words SIGNAL WORD MEANING DANGER Danger indicates an imminently hazardous situation which if not avoided will result in death or serious injury WARNING Warning indicates a potentially hazardous situation which if not avoided could result in death or serious injury CAUTION Caution indicates a potentially hazardous situation which if not avoided may result in property damage ACCESSORIES There are many different types of humidifiers oxygen tubing cannulas and masks that can be used with this device You should contact your local home care provider for recommendations on which of these devices will be best for you They should also give you advice on the proper usage maintenance and cleaning WARNING Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products HomeFill 1 Compressor 4 Part No 1100873 TYPICAL PRODUCT PARAMETERS Regulatory Listing
9. DO NOT REMOVE THIS PRODUCT LABEL Lake Mary Blvd Sanford FL 32773 Not for Commercial Resale 1 800 832 4707 Part Number 1125090 Rev 11 04 FIGURE Cylinder Prefill Inspection HomeFill 1 Compressor 17 Part No 1100873 SECTION 3 CHECKLIST AND CYLINDER INSPECTION The Hydrostatic Testing Date Aluminum cylinders MUST undergo testing every five years A WARNING DO NOT fill cylinders that have not been tested in the past five years Otherwise serious personal injury may result Contact your dealer for replacement External Examination 1 Examine the outside of the cylinder for the following conditions and replace the cylinder if they exist e Dents or dings e Arc burns Oil or grease e Any other signs of damage that might cause a cylinder to be unacceptable or unsafe for use Examine the cylinder for evidence of fire or thermal damage Evidence includes charring or blistering of the paint or other protective coating or heat sensitive indicator If fire or thermal damage is found replace the cylinder 2 Inspect the cylinder regulator assembly for the following and replace if found Debris oil or grease e Noticeable signs of damage Signs of corrosion inside the valve Signs of excessive heat or fire damage Part No 1100873 18 HomeFill 1 Compressor YIM 9599 REGISTER YOUR PRODUCT The benefits of registering 2 3 Safeguard your investmen
10. N OR ATTEMPT TO SERVICE UNLESS SERVICE IS PROVIDED BY AN INVACARE CERTIFIED TECHNICIAN otherwise this will void any and all warranties If within such warranty periods any such product shall be proven to Invacare s satisfaction to be defective such product shall be repaired or replaced at 5 option This warranty only applies to the labor for repairs performed by the Invacare Service Department or Invacare dealers It does not apply to the labor performed by the purchaser or user This warranty does not include normal wear and tear failure to adhere to the product instructions or shipping charges incurred in replacement part installation or repair of any such product Invacare s sole obligation and your exclusive remedy under this warranty shall be limited to such repair or replacement Routine maintenance items such as filters are excluded from this warranty For warranty service please contact Invacare s service department at the toll free number on the back page during normal working hours Upon receiving notice of an alleged defect in a product Invacare will issue a serialized return authorization It shall be the responsibility of the purchaser at the purchaser s cost to return the unit pack the unit a manner to avoid shipping damage and to ship the unit to either Invacare s plant or service center as specified by Invacare in advance Defective units must be returned for warranty inspection using the serial numb
11. Operator s Manual HomeFill Compressor User Before using the product read this manual and save for future reference Dealer This manual must be given to the end user For more information regarding Invacare products parts and services please visit www invacare com Yes you WARNING A WARNING DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL IF YOU ARE UNABLE TO UNDERSTAND THE DANGERS WARNINGS CAUTIONS AND INSTRUCTIONS CONTACT INVACARE TECHNICAL SERVICES BEFORE ATTEMPTING TO USE THIS EQUIPMENT OTHERWISE SERIOUS INJURY OR PROPERTY DAMAGE MAY RESULT NOTICE The information contained in this document is subject to change without notice A WARNING Users MUST NOT SMOKE while using this device Keep all matches lighted cigarettes or other sources of ignition out of the room in which this product is located Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air Failure to observe this warning can result in severe fire property damage and cause physical injury or DEATH NOTE Updated versions of this manual are available on www invacare com HomeFill 1 Compressor 2 Part No 1100873 TABLE OF CONTENTS TABLE OF CONTENTS SPECIAL NOTES TYPICAL PRODUCT D PARAMETERS ccccccccc
12. Technical Service If the compressor has a damaged cord or plug if it is not working properly if it has been dropped or damaged or submersed in water call a qualified technician for examination and repair Part No 1100873 10 HomeFill 1 Compressor SECTION I IMPORTANT SAFETY INSTRUCTIONS If any leakage of the oxygen cylinder is detected DO NOT attempt to use the cylinder Turn the flow selector knob on the regulator to Off If leakage persists place the cylinder outdoors and notify your Home Health Care Provider or Service Representative of this condition The HomefFill is equipped with a high pressure relief valve to ensure the user s safety When activated this safety feature will make an extremely loud noise If this noise occurs turn the unit Off and contact your Home Health Care Provider or Service Representative Unplug unit from outlet when not in use and before cleaning This appliance is provided with double insulation Use only identical replacement parts Refer to Servicing of Double Insulated Products on page 11 Servicing of Double Insulated Products In a double insulated product two systems of insulation are provided instead of grounding No grounding means is provided on a double insulated product nor is a means for grounding to be added to the product Servicing a double insulated product requires extreme care and knowledge of the system and is to be done only by qualified service personnel Replace
13. YELLOW light may come on if the oxygen level from the concentrator has not yet reached or drops below 90 The compressor will begin or resume filling when the HomeFill 1 Compressor 29 Part No 100873 SECTION 4 COMPRESSOR OPERATION oxygen level reaches 90 If light does not go Off within 10 minutes refer to Troubleshooting Cylinder Fill times on page 34 NOTE If the ATTENTION RED light and audible alarm are on Refer to Troubleshooting Cylinder Fill times on page 34 6 Remove the full cylinder Refer to Disconnecting the Cylinder from the Compressor on page 27 7 Push the power switch to the Off O position Control Panel Attention RED light and Audible Alarm Filling O GREEN Light INVACARE HOMEFILL II ATTENTION O2 BELOW ot fl FILLING FULL DANGER RISK OF ELECTRICAL SHOCK DO NOT remove cover refer servicing to qualified service personnel A WARNING RISK OF FIRE NO sources of ignition out of the roduct is located a away fom area ekal ing delivered Textiles oil and other are easily and will burn with great intensity in oxygen enriched air CAUTION Federal law restricts this device to sale by or the order of a physician or any other ees te law of the state in which he order the of this devi OPERATING INSTRUCTIONS See Operator s Manual gt YELLOW lig
14. and or damage may occur NOTE For this procedure refer to FIGURE 4 3 1 Press the compressor power switch to the Off O position 2 Grasp the cylinder regulator assembly in the area behind the cylinder gauge 3 With the other hand grasp the outer ring sleeve of the connector fillport and push DOWN 4 Liftup on the cylinder regulator assembly to remove from the connector fillport Cylinder Regulator 24 eS Assembl PUSH DOWN Be SY Cylinder ter Ri Gauge Sindee ing Compressor Ky Fillport Sleeve lt Q Connector Fillport FIGURE 4 3 Disconnecting the Cylinder from the Compressor HomeFill 1 Compressor 27 Part No 1100873 SECTION 4 COMPRESSOR OPERATION 5 When the cylinder fillport is disconnected from the connector fillport release the outer ring sleeve of the connector fillport and use two hands to remove the cylinder from the compressor cradle A WARNING The fillport cover on the connector fillport and the cylinder fillport MUST be replaced after filling and whenever not in use If either of the fillport covers are missing contact your healthcare provider before using 6 Place the fillport covers onto the connector fillport and cylinder fillport 7 Examine the cylinder gauge to ensure cylinder is full needle pointing into GREEN area If the cylinder is not full perform the following e Repeat Connecting the Cylinder to the Compressor on page 23
15. available for a particular medical condition IJ Yes No If yes please list any condition s here and we will send you information by email and or mail about any available Invacare products that may help treat care for or manage such condition s Would you like to receive updated information via email or regular mail about the Invacare home medical products sold by Invacare s dealers Yes No What would you like to see on the Invacare website Would you like to be part of future online surveys for Invacare products Yes No User s Year of birth If at any time you wish not to receive future mailings from us please contact us at Invacare Corporation CRM Department 39400 Taylor Parkway Elyria OH 44035 or fax to 877 619 7996 and we will remove you from our mailing list TO FIND MORE INFORMATION ABOUT OUR PRODUCTS VISIT WWW INVACARE COM SALVLS A3LINN AHLNI AYVSSAOAN ADVLSOd ON 9 86 9 0 HO VIHAA AVMMYVd YOTAVL OOV6E INSINLYVd4sd Wad JYVOVANI AASSAYdQGV 34 TIM ADVLSOd HO VIHATA 961 ON LIWHAd JIV SSV19 LSul4 TVN SSANISNG SECTION 4 COMPRESSOR OPERATION SECTION 4 COMPRESSOR OPERATION Connecting Disconnecting the Cylinder to from the Compressor Connecting the Cylinder to the Compressor A WARNING NEVER use tools of any kind to connect disconnect the cylinder and the compressor Otherwise severe injury a
16. ccccccccccccccccccccccccccccsccccccocs D LABEL LOCATION cccccccccccccccccccccccccccccccccsccssess DEALER INFORMATION ccccccccccccccccccccccccccccees HOMEFILL INFORMATION cccccccccccccccccccccccce 7 SECTION IMPORTANT SAFETY INSTRUCTIONS ccccccccccccccccccccccccccccccccccccscssssee Servicing of Double Insulated Products SECTION 2 INITIAL SETUP 0000000000000000000000000000 12 Initial Set Up of 12 SECTION 3 CHECKLIST AND CYLINDER INSPECTION 66 1 Compressor Operation 16 Cylinder Prefill 16 SECTION 4 COMPRESSOR OPERATION 23 Connecting Disconnecting the Cylinder to from the COMPRESSOR gt SS 23 Turning the Compressor 29 Setting the Flow Knob to the Prescribed Setting 31 SECTION 5 INDICATOR LIGHTS 32 SECTION 6 TROUBLESHOOTING CYLINDER FILL TIMES cccccccccccccccccccccccccccccccccccccccccccccccccs SF SECTION 7 MAINTENANCE 37 Cleaning the Compressor Filter 37 38 LIMITED WARRANTY 38 HomeFill
17. er as identification within thirty 30 days of return authorization date Do not return products to our factory without prior consent C O D shipments will be refused please prepay shipping charges LIMITATIONS AND EXCLUSIONS THE FOREGOING WARRANTY SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE ACCIDENT IMPROPER OPERATION MAINTENANCE OR STORAGE SOOT OR SMOKE FILLED ENVIRONMENTS OR OTHER THAN NORMAL APPLICATION USE OR SERVICE OR TO PRODUCTS MODIFIED WITHOUT INVACARE CORPORATION S EXPRESS WRITTEN CONSENT HomeFill 1 Compressor 39 Part No 1100873 INCLUDING BUT NOT LIMITED TO MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS OR TO PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE CORPORATION OR TO PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND INVACARE CORPORATION S CONTROL THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER WARRANTIES WHATSOEVER WHETHER EXPRESS OR IMPLIED INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER
18. gh concentrations of oxygen over a long duration without monitoring its effect on oxygen content of arterial blood may be harmful Oxygen Cylinders filled by the Venture HomeFill shall be used for personal use only Not to be filled for resale or use by professional users This cylinder is to be used with the Invacare Venture HomeFill system This cylinder contains 93 3 oxygen produced by an oxygen concentrator using the pressure swing adsorption process DO NOT handle cylinder or use contents until you are professionally trained including emergency procedures Use in accordance with Venture HomeFill Operator s Manual Open valve slowly Close valve after each use and when empty Secure cylinder during storage and use No smoking in cylinder area Keep away from heat flame and spark Keep out of reach of children DO NOT drop Transfilling of this gas is performed by Venture HomeFill Only Cylinders that have been refinished and or exposed to elevated temperatures must be hydrostatically tested before filling Cylinder temperature should not exceed 125 F 52 C Aluminum cylinders subject to the action of fire or heated to temperatures in excess of 350 F must be withdrawn from service and condemned by trained personnel DO NOT use caustic paint strippers Valve and safety relief devices must be removed and replaced by trained and authorized personnel DO NOT alter or change valve assembly Invacare Gorporation
19. ht GREEN Light Power Switch FIGURE 4 5 Turning the Compressor On Part No 1100873 30 HomeFill 1 Compressor SECTION 4 COMPRESSOR OPERATION Setting the Flow Knob to the Prescribed Setting NOTE For this procedure refer to FIGURE 4 6 1 Attach the nasal cannula to the outlet fitting of the cylinder 2 Turn the flow knob to the L min setting prescribed by your physician or therapist A WARNING As changing the L min setting on the flow knob will affect the dose of oxygen delivered DO NOT readjust the L min setting unless directed by your physician or therapist 3 To turn the flow of oxygen Off turn the flow knob counter clockwise to the Off position Outlet Fitting Flow Knob Prescribed Setting FIGURE 4 6 Setting the Flow Knob to the Prescribed Setting HomeFill 1 Compressor 3I Part No 1100873 SECTION 5 INDICATOR LIGHTS SECTION 5 INDICATOR LIGHTS INDICATOR LIGHT COLOR None COMPRESSOR STATUS Compressor is Off EFFECT flowing to user if concentrator is on O BELOW NORMAL YELLOW Compressor is On and warming up 0 3 minutes after power is On Cylinder filling not started flowing to user if concentrator is On from concentrator is lt 90 any time after three minute warmup Cylinder filling stops will restart when O gt 90 flowing to user if concentrator is On FILLING GREEN Com
20. ing oils suntan oils or sunblocks and similar oily products If you use these types of products Invacare suggests you wash your hands prior to using the HomefFill unit If the bottle fitting should come in contact with such substances it should be cleaned off with a damp cloth prior to connecting it to the compressor NEVER oil or lubricate the compressor coupler or bottle fittings NEVER use any penetrating oil or lubricant such as WD 40 or in One If connection difficulty is experienced verify that the bottle fitting is being inserted straight down into the compressor coupler If continued difficulty is experienced contact your supplier for assistance Before moving or repositioning either the compressor or concentrator ALWAYS push the power switch to the Off O position disconnect the AC power cord and the interconnect hose between the compressor and concentrator Failure to do so may result in damage to the unit or personal injury Children should ALWAYS be supervised around the HomefFill unit When used with the HomefFill table similar elevated support surfaces a child could pull over any such table causing the unit to fall Failure to observe this warning may result in personal injury or product damage Use extreme care when handling and filling an oxygen cylinder Full oxygen cylinders are under pressure and can become a projectile if dropped or mishandled NEVER transport a compressor with a cylinder connected to or
21. ment parts for a double insulated product MUST be identical to the parts they replace A double insulated product is marked with the words DOUBLE INSULATION or DOUBLE INSULATED The symbol El is also able to be marked on the product CAUTION DO NOT unplug the power cord by pulling on the cord To unplug grasp the plug not the cord otherwise damage may occur SAVE THESE INSTRUCTIONS HomeFill 1 Compressor lI Part No 1100873 SECTION 2 INITIAL SETUP SECTION 2 INITIAL SETUP NOTE Dealers Only Refer to Dealer Information on page 6 before performing this procedure Initial Set Up of Compressor Compressor Initial Setup Checklist Before operating the HomefFill Compressor for the first time complete the following checklist Q Choose a location for the compressor Q Transport the compressor to the desired location Q Connect the power cord to the compressor Q Connect the compressor to the concentrator Choosing a location The compressor should be at least three inches away from walls draperies or furniture to ensure sufficient air flow Avoid placing the compressor next to heaters radiators or hot air registers DO NOT use in a closet The compressor should sit on a hard flat stable surface such as a table or the Ready Rack A WARNING NEVER use near any type of flame or flammable explosive substances vapors or atmosphere Part No 1100873 12 HomeFill 1 Compressor SECTION 2
22. n for the safe operation and use of this product DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as Owner s Manuals Service Manuals or Instruction Sheets supplied with this product or optional equipment If you are unable to understand the Warnings Cautions or Instructions contact a health care professional dealer or technical personnel before attempting to use this equipment otherwise injury or damage may occur CAUTION Caution Federal law restricts this device to sale by or on order of a physician or any other practitioner licensed by the law of the State in which he she practices to use or order the use of this device When using the HomefFill basic precautions should ALWAYS be followed including the following A WARNING TO REDUCE THE RISK OF BURNS ELECTROCUTION FIRE OR INJURY TO PERSONS NEVER use near any type of flame or flammable explosive substances vapors or atmosphere Part No 1100873 8 HomeFill 1 Compressor SECTION I IMPORTANT SAFETY INSTRUCTIONS DO NOT allow oil from your hands or other sources to come into contact with the compressor coupling or bottle fittings These substances may become flammable in the presence of oxygen and cause injury Avoid touching the bottle fitting whenever possible Example of items to avoid are hand creams or lotions cook
23. nance on page 37 5 If performing STEPS 1 4 does not cause any indicator lights to come on contact your home health care provider or Invacare for service RED Light ON I Check the cylinder gauge to see if it is FULL Cylinder full Remove the cylinder and turn the unit Off Perform the Cylinder Prefill Inspection on page 16 with a cylinder that is NOT FULL Cylinder not full Proceed to STEP 2 2 Check that the cylinder is connected properly Turn the compressor Off Remove the cylinder and reinstall Turn the compressor On 3 If cylinder is not full and is connected properly the compressor may have internal failure Contact your home health care provider or Invacare Compressor cycles ON and Off or O BELOW NORMAL YELLOW and FILLING GREEN lights are alternating or BELOW NORMAL YELLOW light goes On and Off Part No 1100873 1 Check the concentrator liter flow setting is at 3 L min or less If necessary adjust the liter flow setting refer to the concentrator owner s manual Wait at least 20 minutes for the GREEN light to appear If it does not appear proceed to STEP 2 2 The concentrator may need service contact your home health care provider or Invacare 34 HomeFill 1 Compressor SECTION 6 TROUBLESHOOTING CYLINDER FILL TIMES TROUBLESHOOTING PROBLEM SOLUTION O BELOW NORMAL YELLOW light stays On Ensure the concentrator has warmed up for at least 20 minu
24. nd or damage may occur DO NOT drop oxygen cylinders Use two hands when handling transporting oxygen cylinders Otherwise injury or damage may occur DO NOT use a liquid leak detector to test for leaks NOTE For this procedure refer to FIGURE 4 1 on page 24 and FIGURE 4 2 on page 26 1 Examine the cylinder gauge If the cylinder pressure is less than 1500 psig proceed to STEP 2 FIGURE 4 2 NOTE If the cylinder pressure is greater than 1500 psig DO NOT attempt to top off this cylinder It may not fill 2 Set the regulator flow dial on the cylinder to Off FIGURE 4 2 3 Remove the cylinder and connector fillport covers if present HomeFill 1 Compressor 23 Part No 1100873 SECTION 4 COMPRESSOR OPERATION CAUTION DO NOT connect the outlet cannula connection to the HomeFill Il compressor otherwise the conserving cylinder will not work properly 4 Momentarily push DOWN on the outer ring sleeve of the connector fillport until GREEN dots are visible to reset the connector FIGURE 4 1 NOTE If the outer ring sleeve is in the UP position GREEN dots not visible the connector fillport will not be able to accept the cylinder fillport Pushing DOWN momentarily will reset the connector fillport GREEN dots visible to accept the cylinder fillport PUSH Outer Ring Sleeve DOWN Connector Fillport GREEN Dots BEFORE COUPLING Peg CYLINDER PS PUSH DOWN W ON SLEEVE
25. pressor is On and cylinder is filling O from concentrator is gt 90 any time after three minute warmup Cylinder is filling O flowing to user if concentrator is On FULL GREEN Part No 1100873 Compressor is On Cylinder is finished filling full remove cylinder FULL light is On in gt 10 mins or Cylinder is not attached properly check connection FULL light is On in lt 10 mins 32 Cylinder filling stops Cannot fill another cylinder until compressor is turned Off and On again flowing to user if concentrator is On HomeFill 1 Compressor SECTION 5 INDICATOR LIGHTS INDICATOR LIGHT COMPRESSOR EFFECT COLOR STATUS ATTENTION RED Compressor is On Compressor does not Audible Alarm Cylinder NOT filling begin to fill or stops Causes filling Full cylinder attached Unit must be turned Off when unit turned On and On before replace with empty filling can occur cylinder or flowing to user if Internal failure concentrator is On contact dealer HomeFill 1 Compressor 33 Part No 1100873 SECTION 6 TROUBLESHOOTING CYLINDER FILL TIMES SECTION 6 TROUBLESHOOTING CYLINDER FILL TIMES TROUBLESHOOTING PROBLEM SOLUTION No indicator lights on when compressor turned on Turn compressor Off 2 Check that power cord is properly plugged into compressor and wall outlet 3 Turn the compressor On 4 Clean air filters Refer to Mainte
26. r s invoice required for coverage under this warranty Invacare warrants the HomeFill internal wear components piston seals and rings bearings drive belts check valves pressure relief components and electric motor for a period of three 3 years from the date of purchase from Invacare or a dealer with a copy of the seller s invoice required for coverage under this warranty or 5 000 hours whichever comes first Invacare warrants the HomeFill control panel components electrical switches reset switch and indicator lights for a period of three 3 years from the date of purchase from Invacare or a dealer with a copy of the seller s invoice required for coverage under this warranty or 5 000 hours whichever comes first Part No 1100873 38 HomeFill 1 Compressor For Model Nos IOH2M6 and IOH2M9 Invacare warrants the HomeFill regulator NOT including the cylinder to be free from defects in materials and workmanship for a period of six 6 years from date of purchase from Invacare or a dealer with a copy of the seller s invoice required for coverage under this warranty For Model Nos HF2PCL4 HF2PCL6 and 2 Invacare warrants the HomeFill regulator NOT including the cylinder to be free from defects in materials and workmanship for a period of two 2 years from date of purchase from Invacare or a dealer with a copy of the seller s invoice required for coverage under this warranty DO NOT OPE
27. r is used to fill a cylinder complete the following checklist Q Ensure the concentrator has been On for at least thirty minutes Refer to the concentrator Owner s Manual Perform the prefill inspection on the cylinder Connect the cylinder to the compressor Push the compressor power switch to the on 1 position Examine the indicator lights on the control panel Disconnect and remove the full cylinder Push the compressor power switch to the Off O position If filling another cylinder repeat this checklist OOUOUWUD Cylinder Prefill Inspection A WARNING ONLY use cylinders that have the Invacare HomeFill 11 connection and the label All cylinders MUST be inspected before attempting to fill Otherwise injury or damage may occur NOTE For this procedure refer to FIGURE 3 1 on page 17 Part No 1100873 16 HomeFill 1 Compressor SECTION 3 CHECKLIST AND CYLINDER INSPECTION A DANGER Improper filling or abuse of this cylinder or failure to follow these safety instructions may cause serious bodily injury or death High pressure oxidizing gas vigorously accelerates combustion DO NOT alter or modify cylinder or related components DO NOT use Oil or Grease Keep cylinder cylinder valve all equipment and connections free of oil and grease to avoid violent ignition resulting in serious bodily injury or death For medical applications Use only as directed by a licensed practitioner Uninterrupted use of hi
28. stored on the compressor otherwise injury or damage can occur HomeFill 1 Compressor 9 Part No 1100873 SECTION IMPORTANT SAFETY INSTRUCTIONS DO NOT store filled oxygen cylinders near a furnace hot water tank in the trunk of a car or other high temperature area Storage in areas such as this can result in bursting of the cylinder or fire The use of this device is limited to the oxygen patient Cylinders MUST be used ONLY by the oxygen patient and are not to be distributed to any other individual for any purpose Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products DO NOT use outdoors The maximum patient flow while filling cylinders with the 200 is 3 L min for Platinum 5 XL 5 L min for Platinum 10 Patients with a higher prescribed flow should not transfill while receiving supplemental oxygen treatment from the Platinum Concentrator NEVER block the air openings of the product or place it soft surface such as a bed or couch where the air opening may be blocked Keep the openings free from lint hair and the like DO NOT disassemble The unit is provided with double insulation and has no serviceable parts Refer to Servicing of Double Insulated Products on page 11 If any service is required contact Invacare
29. t Ensure long term maintenance and servicing of your purchase Receive updates with product information maintenance tips and industry news Invacare can contact you or your provider if servicing is needed on your product It will enable Invacare to improve product designs based on your input and needs Register ONLINE at www invacare com Or Complete form and mail Any registration information you submit will be used by Invacare Corporation only and protected as required by applicable laws and regulations PRODUCT REGISTRATION FORM NAME ADDRESS CITY STATE PROVINCE ZIP POSTAL CODE EMAIL PHONE NO INVACARE MODEL NO SERIAL NO PURCHASED FROM DATE OF PURCHASE Method of purchase check all that apply LI Medicare Insurance Medicaid Other 2 This product was purchased for use by check one Self Parent Spouse Ly Other 3 Product was purchased for use at d Home Facility Ly Other 4 purchased an Invacare product because LI Price Features list features 5 Who referred you to Invacare products check all that apply 1 Doctor Therapist Friend Relative Dealer Provider Other Li Advertisement circle one TV Radio Magazine Newspaper LI No Referral 6 What additional features if any would you like to see on this product 10 Would you like information sent to you about Invacare products that may be
30. tes If has not been 20 minutes turn the compressor Off until the concentrator warms up If the YELLOW light still remains on proceed to STEP 2 2 Inspect the connection between the compres sor and the concentrator for damage or pinching If damaged replace If pinched straighten tubing Turn the compressor Off for at least 10 seconds Turn the compressor If the YELLOW light still remains on after 10 minutes proceed to STEP 3 3 The concentrator may need service contact your home health care provider or Invacare Fill times exceed maximum as recommended fill times OR GREEN filling light stays On Ensure the cylinder regulator is set to Off 2 Check the recommended fill times shown in Cylinder Fill Times on page 36 3 Check that the cylinder is connected properly Turn the compressor Off Remove the cylinder and reinstall Turn the compressor ON 4 Check the compressor concentrator line for proper connection and or damage leaks or kinks 5 Check the concentrator liter flow setting is at three L min or less If necessary adjust the liter flow setting refer to the concentrator owner s manual If the GREEN filling light does not go Off the compressor and or concentrator may need service Contact your home health care provider or Invacare Excessive noise when compressor is On Call home health care provider or Invacare for service Hissing sound from Regu lator
31. tlet only one way If the plug does not fit fully in the outlet reverse the plug If it still does not fit contact a qualified electrician DO NOT attempt to defeat this safety feature FIGURE 2 2 Installing the Power Cord Connecting the Compressor to the Concentrator NOTE For this procedure refer to FIGURE 2 3 on page 15 1 Insert one end of the interconnect hose into the outlet fitting on the back of the concentrator Detail A 2 Insert the opposite end of the interconnect hose into the inlet fitting on the side of the compressor Detail B NOTE The metal tabs on the concentrator outlet fitting and the compressor inlet fitting will pop out with an audible click when the end of the interconnect hose is properly installed Detail 3 Loop any excess interconnect hose and secure to the back of the concentrator with the fastening straps Part No 1100873 14 HomeFill 1 Compressor SECTION 2 INITIAL SETUP DETAIL A DETAIL Insert End of T Interconnect Interconnect Hose into inlew Gust Fitting Hose Inlet Outlet Fitting DETAIL B Compressor aS A L CONCENTRATOR TOP VIEW Outlet Fitting Part No 1100873 HomeFill 1 Compressor SECTION 3 CHECKLIST AND CYLINDER INSPECTION SECTION 3 CHECKLIST AND CYLINDER INSPECTION Compressor Operation Checklist Each time the HomeFill II Compresso
32. when flow selector is Off HomeFill 1 Compressor Turn flow selector Off 2 If hissing sounds continue call home health care provider or Invacare 35 Part No 1100873 SECTION 6 TROUBLESHOOTING CYLINDER FILL TIMES TROUBLESHOOTING PROBLEM SOLUTION Regulator is On but O is not being delivered Refill cylinder 2 Check that cannula is straight and not pinched If damaged replace 3 Call home health care provider or Invacare for service CYLINDER FILL TIMES Cylinder Concentrator Flow Rate to Patient Filling Platinum 5 up to 3 L min and Platinum 10 up to 5 Times L min HF2PC9 HF2PCL4 HF2PCL6 NORMAL 2 hr 20 min 60 min hr 25 min MAXIMUM 2 hr 45 min hr 20 min hr 50 min NOTE All filling times are approximate and may vary on environmental conditions Part No 1100873 36 HomeFill 1 Compressor SECTION 7 MAINTENANCE SECTION 7 MAINTENANCE Cleaning the Compressor Filter A WARNING Push the power switch to the Off O position and unplug the compressor before cleaning DO NOT operate the compressor without the filter installed CAUTION DO NOT unplug by pulling on cord To unplug grasp the plug not the cord otherwise damage may occur NOTE For this procedure refer to FIGURE 7 1 NOTE All compressors have one filter on the back of the cabinet NOTE Perform this procedure at least once a week 1 Unplug the compressor Grasp