Home
        Manuals - OMEGA Engineering
         Contents
1.                                                                                                                                                                   Figure 9  Connections for Measuring Pressure    3 4 1 Zeroing with Absolute Pressure Modules     To zero  adjust the calibrator to read a known pressure  such as  barometric pressure     To adjust the calibrator  follow these steps   1  Enter the pressure zeroing menu     2  Select  ZERO d     SET REFERENCE ABOVE  will  appear  Enter the pressure using the keypad     3  The calibrator stores the Barometric zero offset in non   volatile memory     The zero offset is stored for one absolute pressure module at a  time  If anew absolute module is connected this process must  be repeated     4  Using Source Modes  Lower Display     The calibrator can generate calibrated signals for testing and  calibrating process instruments  It can source voltages  currents   resistances  frequencies  pulses  and the electrical output of RTD  and thermocouple temperature sensors     4 1 Setting 096 and 10096 Output Parameters  To set the 0  and 100  points  use the following steps     1  Select the lower display  LOWER  from Main Menu  and choose  the desired primary parameter     2  Select output  OUT  from the input output control  and enter the  desired value  For example select  VOLTS OUT      3  Enter 5V with the keypad and press Enter     Press any one of the four cursor control arrows to display the  setpoint control menu     5  
2.                                                                                                          RTD Type Range    C  Uncertainty  Miniumum Maximum 1 year    C    PT385  10 ohm  200  80 0 78    80 0 0 82   0 100 0 84   100 300 0 95   300 400 0 99   400 630 1 09   630 800 1 19   PT385  50 ohm  200  80 0 18    80 0 0 25   0 100 0 21   100 300 0 26   300 400 0 29   400 630 0 34   630 800 0 39   PT385  100 ohm  200  80 0 10    80 100 0 13   100 300 0 18   300 400 0 20   400 630 0 25   630 800 0 29   PT3926  100 ohm  200  80 0 10    80 0 0 11   0 100 0 13   100 300 0 17   300 400 0 19   400 630 0 24   PT3916  100 ohm  200  190 0 08    190  80 0 10    80 0 0 11   0 100 0 13   100 260 0 17   260 300 0 17   300 400 0 19   400 630 0 25                100                                                                                        RTD Type Range    C  Uncertainty  Miniumum Maximum 1 year    C    PT385  200 ohm  200  80 0 40   80 0 0 42   0 100 0 45   100 260 0 45   260 300 0 52   300 400 0 53   400 630 0 66   PT385  500 ohm  200  80 0 18   80 0 0 19   0 100 0 21   100 260 0 25   260 300 0 26   300 400 0 29   400 630 0 34   PT385  1000 ohm  200  80 0 10   80 0 0 12   0 100 0 14   100 260 0 17   260 300 0 18   300 400 0 19   400 630 0 25   NI120  80 260 0 06  Cu10  100 260 0 82  Cu50  180 199 0 20  Cu100  180 199 0 13  YSI 400 15 50 0 05                   Read Accuracy is based on 4 wire input  For 3 wire input add   0 059 assuming all three    RTD leads are matched    
3.                 Figure 20  Calibrating an I P Device    6 3 Calibrating a Transmitter    To calibrate a transmitter both the upper and the lower displays will  be used  one for measuring and the second a source  This section  covers all but the pressure transmitters  A thermocouple temperature    27    28    transmitter is used in this example   The following steps show how to calibrate a temperature transmitter     1     From the Main Menu select upper display  and choose current  loop  mA LOOP     Switch to lower display from the Main Menu  and select  TC   from the primary parameters  Choose output  OUT  from the  input output control  and select TC type    Set the 0   and 100   span points using the keypad and the  0  and 100  keys  refer to Setting O   and 100   Parameters  section     Connect the calibrator to the transmitter as shown in Figure 21     Test transmitter at 0  25  50  75  100   using the 25   step  function  2596 key      Adjust the transmitter a necessary    To calibrate a different transmitter  follow the above steps with the  exception of choosing TC on the lower display  Replace TC with the  correct parameter for the transmitter                                                                 Figure 21  Calibrating a Transmitter    6 4 Calibrating a Pressure Transmitter  To calibrate a pressure transmitter  use these steps     1     Select upper display from the Main Menu  and choose current   mA LOOP  from the primary parameters  Return to Main Menu   
4.             Figure 11  Connections for Simulating a Transmitter    19    20       mA IN  25 00          0 000 mA    mA 2W SIM    4 000 mA           uer            4 5 Sourcing volts  To source volts follow these steps              1  Select lower display from the Main Menu     2  Choose  VOLTS  from the primary parameters  Switch to  input output control and select output  OUT      3  Connect the leads for the voltage source terminals  as shown in    Figure 12   4  Enter the voltage using the keypad              Figure 12  Connections for Sourcing Voltage and Frequency       mA LOOP  25 00     VOLTS  0 000 mA  6 000 V    MENU   LIGHT                  4 6 Sourcing frequency  To source a signal use these steps     1  Switch to the lower display and select frequency from the  primary parameters     2  Select output  and than choose the frequency units     3  Connect the leads to the frequency output terminals as shown in    Figure 12   4  Enter the desired frequency using the keypad     5  To change the amplitude  select  FREQ LEVEL  from frequency    level menu   6  Enter the amplitude     4 7 Sourcing a pulse train    The calibrator can produce a pulse train with an adjustable number  of pulses at a desired frequency  For example  setting the frequency  to 60Hz and the number of pulses to 60 would produce 60 pulses for  a period of 1 second  To source a pulse  use the same connection as  for frequency  and proceed as follows     1     Switch to the lower display and select pul
5.          SAVE             Save As Found data  and time stamp    Y          Validate test results        YES   NO                   Tolerance field selection          Y    SELECT   DONE            Tolerance field data entry          DONE              Show test results        YES                  EMT        View test results       DONE                 Save AF results as AL        YES                   Document Menus        Y       56    Adjust Instrument        YES                                                                                                                                                                                                                       Downloaded Tag As  Found Test  v  Either display                    39 Pressure Zero Menu  pressure mode    Yes mE     DONE  No  pl    Dal  v  Pulse mode  Nec AF data collection     pulse trigger  TRIGGER     READY  No Y  AF data collection     pulse stop  STOP      v  Manual Input   ve    AF data collection     manual entry  NEXT                               1                    Y Y  Manual input entry K  DONE                  Y  y     AF data collection     save test point  No Last data point   I                       SAVE                               1              Yes Y  v K             Save As Found test  data and time stamp                v  L                            Tag ID field selection          Y       SELECT   NEXT   SAVE             Tag field data entry          58       DONE               
6.       Kee eb  cicli     OK LO    4 x    mg       Figure 17  Connection Using Current Loop       mA LOOP  25 00          0 000 mA    T  0 001 mA            LIGHT            24    5 2 1 HART  Resistor Selection    The CL310 can be set up so that the 250 ohm resistor required for  HART  configuration devices resides inside the CL310  Enabling  the CL310 s internal 250 ohm resistor eliminates the need to  manually add a series resistor during a HART  calibration process   NOTE  When the CL310 s internal 250 resistor is enabled  maximum  load driving capability drops from 1000 ohms  2 20mA to 750 ohms   20mA     Enable Disable Procedure    1  Remove the battery cover and remove the 2 screws that are  at the top of the case     2  Remove the 2 screws on the bottom or lower portion of the  case     3  Gently remove the top half of the case from the bottom   4  Figure 10a  shows the location of the HART   jumpers     5 3 Measuring Pressure    Note  The CL310 is compatible with OMEGA Pressure Modules  The  accessory PCL PMA is required for pressure measurement     Note  Pressure is not read from modules with frequency or pulse train  mode enabled     Warning   To avoid a violent release of pressure in a pressurized system  shut off    the valve and slowly bleed off the pressure before you attach the  pressure module to the pressure line     Caution    To avoid mechanically damaging the pressure module  never apply  more than 10 ft lb  of torque between the pressure module fittings  or
7.       OME          YEAR C          WARRANTY       GA       Shop online at          omega com       e mail  info omega com  For latest product manuals   www  omegamanual info                                 Wen   8 fee  E3S      fen  ES      a  EE BEN  x             CL310    Multi Function Calibrator          OMEGA    omega com info omega com    Servicing North America   U S A  Omega Engineering  Inc   Headquarters  Toll Free  1 800 826 6342  USA  amp  Canada only   Customer Service  1 800 622 2378  USA  amp  Canada only   Engineering Service  1 800 872 9436  USA  amp  Canada only   Tel   203  359 1660 Fax   203  359 7700  e mail  info omega com    For Other Locations Visit omega com worldwide    The information contained in this document is believed to be correct  but OMEGA accepts no liability for any errors it contains  and reserves  the right to alter specifications without notice     1  Introduction    The OMEGA CL310 Multifunction Process Calibrator is a handheld   battery operated instrument that measures and sources electrical  and physical parameters  The calibrator has the following features  and functions     e A dual display  The upper display is used for the measurement of  volts  current  and pressure  The lower display can be used to  measure volts  current  pressure  resistance temperature detectors   RTDs   thermocouples  frequency  and resistance  and to source  pulse trains    e A thermocouple  TC  input output terminal with automatic  reference junction tem
8.     101    11  Maintenance   Warranty    11 1 Replacing Batteries    Replace batteries as soon as the battery indicator turns on to avoid  false measurements  If the batteries discharge too deeply the CL310  will automatically shut down to avoid battery leakage     Note  Use only AA size alkaline batteries or optional rechargeable  batteries     Warning  This product has non rechargeable alkaline batteries installed     Prior to connecting any AC adapter or battery charger  these  batteries must be removed and or replaced with rechargeable  batteries such as Nickel Cadmium or Nickel Metal Hydride types     Failure to remove the batteries will result in damage to the  calibrator and voids the warranty     11 2 Cleaning the Unit    Warning    To avoid personal injury or damage to the calibrator  use only the  specified replacement parts and do not allow water into the case     Caution    To avoid damaging the plastic lens and case  do not use solvents or  abrasive cleansers     Clean the calibrator with a soft cloth dampened with water or water  and mild soap     11 3 Service Center Calibration or Repair    Only qualified service personnel should perform calibration  repairs   or servicing not covered in this manual  If the calibrator fails  check  the batteries first  and replace them if needed     Verify that the calibrator is being operated as explained in this  manual  If the calibrator is faulty  send a description of the failure with  the calibrator  Be sure to pack the c
9.    12345678    LOCAL    Restores the calibrator to local operation if it was in remote mode   Also clears LOCKOUT if the unit was in lockout mode     LOWER MEAS    Sets the lower display to measure mode  The command is followed  by any of the parameters except for pulse and mA sim  which are  source only  Enter DCI for mA  DCV for volts  TC for thermocouple   RTD for RTD  FREQUENCY for frequency  and PRESSURE for  pressure  For example     LOWER MEAS DCV sets the lower display mode to VOLTS IN    L PRES UNIT    Sets the unit for measuring pressure on the lower display  Add the  unit after the command  The available pressure units and their syntax    45    46    are shown in the Table 7   Parameter Units    For example     L PRES UNIT KPAL sets the pressure unit to kiloPascals    OUT    Sets the output of the calibrator  This command may be used to  output mA  volts  frequency  temperature  and ohms  Frequency  output  which is set by the command FREQ TYPE  is either  continuous or pulse  The calibrator is automatically set to source  mode when OUT is entered  A number and its unit must follow the  command  See Table 7   Parameter Units  for a list of available units   For example     OUT 10 MA sets the calibrator to mA OUT mode and sets the  output to 10mA     OUT     Returns the output of the calibrator  Using the above example  OUT   Would return     1 000000E 02  A    PRES     Returns the model and serial number of the attached pressure unit   Returns NONE if no pressure u
10.    Use the numeric keys  followed by ENTER  to enter a user value for  generation  or press the READY function key to proceed to the next  menu where the UP and DOWN function keys can be used to scroll  through the As Found data points     Press the TRIG function key to generate the pulse train  The lower  line changes to the following while the pulse train is being generated     STOP        Press the STOP function key to abort the pulse train before it is  finished  When the pulse train is complete  or is aborted  the lower  line changes back to its original state                 Press the READY function key to change the lower line to the  prompts for scrolling through the As Found test points  or  proceeding to As Left data collection     UP   DOWN   AS LEFT    Press the UP or DOWN function key to cycle through the As Found  test points     Press the AS LEFT function key to proceed directly to As Left data  collection     Press the LEFT ARROW key to return to the trigger menu and  generate a new pulse train                 77    78    8 4 Downloaded Tag As Found Test    8 4 1 Tag Selection    At the main menu  press the MENU function key  followed by the  MORE function key  followed by the DOCUMENT function key to  display the first level document mode menu        TEST REVIEW EXIT    Press the TEST function key to display the test selection menu     AS FOUND   AS LEFT   EXIT    Press the AS FOUND function key to display the As Found Tag  Selection menu which shows all unca
11.   FREQ LEVEL  with the above example would return   5 000000E  00  V    FREQ TYPE    When in frequency mode  sets the calibrator to output a continuous  wave  Frequency Out   or a pulse train  To set the calibrator to  continuous wave enter CONT after the command  To set the  calibrator to pulse enter PULSE after the command  For example     FREQ TYPE CONT will set the calibrator to FREQ OUT     Note  This command does not put the calibrator in frequency mode   Use the OUT command to put the calibrator in frequency mode     FREQ TYPE     Tells whether calibrator is sourcing a pulse or a continuous wave   The command will return CONT if the calibrator is in FREQ OUT  mode  and PULSE if the calibrator is in PULSE mode     FREQ UNIT  Sets the unit for frequency  There are three ranges of frequencies for    frequency and pulse modes  CPM  cycles per minute   Hz  and kHz   Use this command to select the right range  For example     FREQ UNIT HZ sets the frequency to Hz range    FREQ UNIT     Returns the frequency unit currently being used by the frequency  and pulse modes     FUNC     Returns the current mode of the upper and lower displays  For  example if the calibrator is set to volts on the upper display  and  pressure on the lower display  FUNC  Would return     DCV  PRESSURE    GET CLOCK    Returns the current calibrator time and date as  yyyy mm dd  hh mm ss   for example     2006 03 25 19 02 56    GET SN  Returns the calibrator serial number of up to 10 digits  for example
12.   SAVE                   Save Tag ID data                                           changes  v  Validating Yes  results  m Show test results   YES   NO    No  v  View test results  DONE                          Save AF results as AL        YES                Document Menus                3   i       Adjust Instrument              YES                                m                   D                                                                                                                                                                                                          y  Either display in  ve    Pressure Zero Menu  pressure mode  ZERO     DONE  No  2     y  Pulse mode  Vas Adjust     pulse trigger  TRIGGER     READY  No i  Adjust     pulse stop  STOP       gt  lt   y  Adjust     monitor step enter  UP   DOWN   AS LEFT  E  As Left Menu  As Left tag selection  SELECT     EXIT    Ly  Home Menu  y  Already has AL  ves Overwrite AL results   7  results  YES   NO            pre  y  Either display in wes com Pressure Zero Menu  pressure mode  ZERO     DONE             No  4    4                      y  F             59       60    No                                                                                                                                                                                                          v  Pulse mode  Ves AL data collection     pulse trigger  TRIGGER     READY  No     AL data collection     pulse stop  STOP      Ya  y  Manu
13.   Select lower display  and choose  PRESSURE  from the primary  parameters     Connect the calibrator to the transmitter and the pressure  module as in Figure 22     Zero the pressure module     Test the transmitter at O   and 100   of the span  and adjust as  necessary                                      1                                                                                                                                          PRESSURE    MODULE                                                                                                                                                    HAND PUMP       Figure 22  Calibrating a Pressure Transmitter    7  Remote Operation    The calibrator can be remotely controlled using a PC terminal  or by   a computer program running the calibrator in an automated system    It uses an RS232 serial port connection for remote operation  With   this connection the user can write programs on the PC  with   Windows   languages like Visual Basic  to operate the calibrator  or use  a Windows terminal  such as Hyper Terminal  to enter single  commands  Typical RS232 remote configurations are shown in   Figure 23     COM PORT       f  EOMEGNK       1                                                   Figure 23  Calibrator to Computer Connection    29    7 1 Setting up the RS 232 Port for Remote Control    Note  The RS 232 connection cable should not exceed 15m unless  the load capacitance measured at connection points is less than 
14.   Use the numeric  keys followed by ENTER to change the value  The year must be  2006 to 2100  the month 1 to 12  the day 1 to the number of days in  the specified month  the hour 0 to 23  the minute O to 59  and the  second 0 to 59  If the day is made invalid by changing to a month  with fewer days  the day value will automatically be changed to the  maximum number of days valid for the new month     Press the SAVE function key to set the clock and return to the home  menu     Press the HOME key to exit the data entry screen without changing  the clock              9  CL310 SW Utility Program    The CL310 SW  Upload  Save  View  Utility program is a Microsoft  Windows based application which aids in the download of tag test  configurations and the retrieval of test results from the CL310  calibrator  and organizing its storage on your PC     9 1 System Requirements  The minimum system requirements are as follows     Windows XB  Windows 2000  Windows Vista  or later    500 KB of free hard drive space for the application  85      Sufficient extra disk space for storage of tag reports    9 2 Installation    Insert the Video Tutorial CD into your PC s drive  A menu will be  automatically displayed     Choose the menu option  Install Omega CL310 SW Utility Software    Follow the on screen prompts to complete the installation     As an alternative  use Explorer to browse to the  Omega CL310 SW   directory on the CD and run SETUPEXE  Follow the on screen  prompts to install the app
15.   between the fittings an the body of the module     To avoid damaging the pressure module from overpressure  never  apply pressure above the rated maximum printed on the module     To avoid damaging the pressure module from corrosion  use it only  with specified materials  Refer to the pressure module documentation  for material compatibility     To measure pressure  follow these steps   1  Connect the pressure module to the calibrator as shown in Figure  18     The calibrator can measure pressure on both the upper and the  lower display  This makes it possible to measure pressure in two  different units at the same time    25    6     26    Switch to either upper or lower display from the Main Menu   Select  PRESSURE  from the primary parameters   Select the desired measuring unit     Zero the pressure module  The zero function on the calibrator can  be found in the pressure zeroing menu     af ON          VALVE    Qu 9                      PRESSURE  MODULE                                      1                                                                                                                                                                           sovact MEASURE                               Cy O    We        O                Figure 18  Measuring Pressure Transmitter    Note  On high pressure modules engineering units normally  associated with low pressure ranges such as  inH O  cmH O  etc  are  not valid selections  Selecting one of these units with a high  
16.  2500pF     Serial parameter values   9600 baud  8 data bits  1 stop bit  no parity  Xon Xoff  EOL  End of Line  character or CR  Carriage Return  or both    To set up remote operation of the calibrator on the Windows Hyper  Terminal  connected to a COM port on the PC as in Figure 23  use  the following procedure     1  Start Hyper Terminal  located in Accessories Communications of  the Windows Start menu     2  Select New Connection     3  For Name enter CL310  Select the serial port that the unit is  connected to     4  Enter the above information for port settings     5  Select ASCII setup from File Properties Settings and mark these  choices     Echo typed characters locally  Wrap lines that exceed terminal width  6  Select Ok        To see if the port works enter  IDN   This command will return  information on the unit     7 2 Changing Between Remote and Local Operation    There are three modes of operation of the calibrator  Local  Remote   and Remote with Lockout  Local mode is the default mode   Commands may be entered using the keypad on the unit or using a  computer  In Remote mode the keypad is disabled  and commands  may only be entered using a computer  but choosing  GO TO  LOCAL  from the menu on the calibrator display will restore keypad  operation  In Remote with Lockout  the keypad can not be used at  all  To switch modes proceed as follows     1  To enable Remote mode  type in the serial command REMOTE at  the computer terminal     2  To enable Remote with L
17.  DONE       Initially the span tolerance is set to 0 25   and the input and output  values are set to the corresponding minimum and maximum test  results values  As the test results values will only approximate the  actual instrument range  they will need to be modified to exact range  values for the validation to work properly  Use the UP DOWN  LEFT   and RIGHT arrow keys to select a value to be modified  and press  SELECT to display the editing prompt described below  Once all  values have been modified  press DONE to continue on to the overall  test status display     When a value is selected for modification  a prompt similar to the  following one for error tolerance is displayed        Enter tolerance as      Span   0 25    DONE        The upper part of the display indicates the value being modified  and  the lower part indicates the current value  Enter the new value using   the numeric keys  followed by ENTER  Press the DONE function key  to return to the current settings display              The overall test status display provides the option of displaying  details for each test point     PASSED       Display test  details           YES NO    The overall test status is displayed on the top line  If all points  passed  PASSED is displayed  If one or more points have failed   FAILED is displayed     Press the NO function key to bypass detail results display and  proceed to the test conclusion stage     Press the YES function key to display the test details  The content 
18.  Format    Click on this item to upload calibration reports in text format  A  checkmark indicates that this format is selected for all further  uploads  The presently selected format is also displayed on the  status line at the bottom of the screen     Upload in CSV Format    Click on this item to upload calibration reports in CSV   spreadsheet  format  A checkmark indicates that this format is  selected for all further uploads  The presently selected format is  also displayed on the status line at the bottom of the screen     Calibrator Identity  Display the identity of the connected calibrator     89    9 6 Erase Menu    All Tags    Erase all tag configurations from the calibrator  Both list boxes  are cleared after the erasure  Use the Configure menu to select  the communications port     Selected Tags    Erase selected tag configurations from the calibrator  Before  using this function  upload a list of the tag configurations  contained in the calibrator into the Tag ID s list box using the Tag  ID s function on the Upload menu  In the Tag ID s list box   highlight the tags which are to be erased  and then select this  function to erase them  The Reports list box is cleared  and the  Tag ID s list box is updated after the erasure  Use the Configure  menu to select the communications port     Multiple individual tags are selected by holding down the Ctrl key  and clicking on them one at a time  Multiple contiguous tags are  selected either by clicking on the first and dra
19.  Terminals    3 3 2 Using Resistance Temperature Detectors  RTDs     The supported types of RTDs are shown in Section 10   Specifications  RTDs are characterized by their 0     resistance  RO   The calibrator accepts two  three  and four wire inputs  with four wire  input being the most accurate    To use the RTD option  apply the following steps    1  Switch to lower display  LOWER  from Main Menu     2  Select RTD from the primary parameters  Select  IN  from  input output control     3  Choose 2  3  or 4 wire connection  2W  3W  4W    4 wire  allows for the most precise measurement     4  Select RTD type from the sensor types   5  Attach RTD leads as shown in Figure 8     3W     0 0O  0 Or  ill O                                      mA IN  25 00         0 000 mA  RTD IN 4W P10 385    25 0_      MENT       _             Figure 8  Measuring Temperature with RTD Connection  Resistance can also be measured using this function  To do so   use the above procedure and choose OHMS from the sensor  types  This option can be used along with a table to measure an  RTD which is not programmed into the calibrator     3 4 Measuring Pressure  Note  The CL310 is compatible with OMEGA Pressure Modules  The  accessory PCL PMA is required for pressure measurement   Note  Pressure is not read from modules with frequency or pulse  train mode enabled   Note  On high pressure modules engineering units normally  associated with low pressure ranges such as             cmH O  etc  are  not valid se
20.  Warning    To avoid possible electric shock or personal injury       Do not apply more than the rated voltage  See specifications for  supported ranges     Follow all equipment safety procedures     Never touch the probe to a voltage source when the test leads are  plugged into the current terminals     Do not use the calibrator if it is damaged  Before you use the  calibrator  inspect the case  Look for cracks or missing plastic  Pay  particular attention to the insulation surrounding the connectors     Select the proper function and range for your measurement     Make sure the battery cover is closed and latched before you  operate the calibrator     Remove test leads from the calibrator before you open the battery  door     Inspect the test leads for damaged insulation or exposed metal   Check test leads continuity  Replace damaged test leads before  you use the calibrator     e When using the probes  keep your fingers away from the probe  contacts  Keep your fingers behind the finger guards on the  probes    Connect the common test lead before you connect the live test    lead  When you disconnect test leads  disconnect the live test lead  first     Do not use the calibrator if it operates abnormally  Protection may  be impaired  When in doubt  have the calibrator serviced     Do not operate the calibrator around explosive gas  vapor  or dust       When using a pressure module  make sure the process pressure  line is shut off and depressurized before you connect it or  d
21.  a Cold   Junction Compensation  CJC  function  Normally this function should    13       be ON and the actual temperature of mA IN  25 00   the thermocouple will be measured     With CJC OFF  the calibrator will 0   000 mA    measure the difference between the TC IN CJC OFF B  thermocouple at the junction and at its     TC input terminal  0 L C       Note  CJC off mode should only be          LIGHT          used when calibration is being done  using an external ice bath     To use the thermocouple to measure temperature  follow these  steps     1  Attach the thermocouple leads to the TC miniplug  and insert the  plug into the input output of the calibrator  as in Figure 7     Note  For best accuracy wait 2 to 5 minutes for the temperature  between the miniplug and the calibrator to stabilize before any  measurements are taken     2  Switch to lower display from Main Menu     3  Select TC from the primary parameters  Choose  IN  in the  input output control  and than the thermocouple type from the  sensor types  The temperature unit may also be changed from  Celsius to Fahrenheit     The calibrator can also measure the mV of a Thermocouple  which  can be used along with a table in case the corresponding TC type is  not supported by the calibrator  To do so  proceed as above and  choose mV from sensor types     TC PLUG       Note  The TC wire  used must match the  thermocouple type  being calibrated                             Figure 7  Measuring Temperature Using Thermocouple
22.  above  CPRT_COEFA  Would return     3 908000E 03    CPRT_COEFB    This command is used for entering a custom RTD into the calibrator   The numeric value entered after the command will be set as the  second coefficient of the polynomial used by the custom RTD     For example   CPRT COEFB  5 8019E 07 enters  5 8019e 7 as coefficient B     CPRT COEFB     Returns the number  which was entered for the first coefficient of the  polynomial used in the custom RTD  Using the example above   CPRT COEFB  Would return      5 801900E 07    CPRT COEFC    This command is used for entering a custom RTD into the calibrator   The numeric value entered after the command will be set as the first  coefficient of the polynomial used by the custom RTD     For example   CPRT COEFC  5 8019E 12 enters  5 8019e 12 as coefficient A     CPRT  COEFC     Returns the number which was entered for the first coefficient of the  polynomial used in the custom RTD  Using the example above  CPRT COEFC  Would return      5 801900E 12    CPRT MIN T    Sets the minimum temperature of the custom RTD range  The  temperature value must be entered with a degrees label  CEL for  Celsius and FAR for Fahrenheit     For example     CPRT MIN T  260 CEL enters  260  C as the minimum  temperature     CPRT MIN T   Returns the value entered for minimum temperature in the range for    a custom RTD  Note that the Calibrator always returns numbers in  scientific notation  The above example would return      2 600000E  02  CEL    CP
23.  all of the parameters must be  supplied  The instrument input range  the difference between  low and high  must be at least 0 0001     Tag Test Points Tab    Enter the number of test points and the instrument input test  point values     Click the Insert button to insert a new value into the middle of the  list of test points  The specified test point value is shifted up one  position  as are all of the values following it  and the specified  test point value is set to zero  The 21st test point is lost     Click the Delete button to remove a value from the middle of the  list of test points  The specified test point value is lost  and all of  the values following it are shifted down one position  The 21st  test point is set to zero     Click the Generate button to generate a set of test points in a  predefined scheme  If tolerance testing is enabled the zero and  span for generation are set to the instrument input low and high  from the tolerance tab  otherwise they are set to the minimum  and maximum test point values  The number of points for  generation is set to the present number of points  Alter the  generation parameters as desired  select a scheme  and click  OK to generate a new set     9 10 Group Configuration Details    The display contains two lists  The leftmost contains all available  tags which are not presently members of the group  The rightmost  contains the present group members  All tags which are members    95    96    of the group must reside in the sam
24.  and lower display  For example  if the upper display is measuring  5mA  and the lower display is measuring 10V  then VAL  will return     5 000000E 03  A  1 000000E 01  V    ZERO_MEAS    Zeroes the attached pressure module  Enter the zeroing value in PSI  after the command when zeroing an absolute pressure module     ZERO MEAS     Returns the zero offset or the reference value for absolute pressure  modules     8  Document Mode    8 1 Introduction    Document mode allows an instrumentation technician to create  repeatable calibration tests for up to 50 tags while in the field  to  download predefined calibration tests from a PC database for up to  50 tags before going to the field  or some combination of both types  up to a maximum of 50 tags  Each test may consist of 1 to 21 user  selectable test points  and may be repeated as many times as  necessary to complete adjustment and calibration of the tag  Two  sets of test results are stored for each tag  an initial As Found set  stored during the first test performed  and a final As Left set stored  during the last test performed  If only a single test is required to  confirm a tag s calibration  both sets of test results will be the same     When creating a test in the field  the technician selects the calibrator  input and output to be used for the test before the As Found test is  started  These include the standard measure and source types of the  calibrator  plus two special types  manual input and manual output   Manual 
25.  and operation of this display is identical to that described in the  section on viewing test results  When viewing is complete  press the  DONE function key to proceed to the test conclusion stage        8 2 5 Test Conclusion    A prompt is displayed asking whether the As Found test results are  to be saved for the As Left test results  and the testing for this tag is  to be ended     If the test has been successful and no further testing is required   press the YES function key to return to the first level document mode  menu     If further testing is required  press the NO function key and a prompt  is displayed asking whether an adjustment step is required before  proceeding to the first As Left test     If an adjustment is required  press the YES function key  otherwise  press the NO function key to proceed directly to As Left data  collection     8 3 Adjustment    During the adjustment process  the output values entered in the As  Found test can be generated by stepping up or down through the  test points  In addition  if the output type is one generated by the  calibrator  i e  not pressure or manual  a user entered value  which  may be different from the test points  may be generated     75    76    If pressure is selected on the upper or lower display  the pressure  zero prompt is displayed first and operates as described under As  Found Setup     The prompt displayed during the test depends on the input and  output types selected  Any combination which includes puls
26.  before  the Event Status Register was read     CME    Command Error  Set to 1 when the calibrator receives an  invalid command  Entering an unsupported RTD type may  cause such an error     EXE    Execution Error  Set to 1 when the calibrator runs into an  error while executing is last command  A parameter that  has too significant figures may cause such an error     DDE    Device dependent Error  Set to 1 when  for example  the  output of the calibrator is overloaded     QYE  Query Error     OPC    Operation Complete  Set to 1 when the calibrator has  finished executing all commands before the command   OPC was entered     Error Queue    If an error occurs due to invalid input or buffer overflow  its error code  is sent to the error queue  The error code can be read from the  queue with the command FAULT   The error queue holds 15 error  codes  When it is empty  FAULT  returns 0  The error queue is  cleared when power is reset or when the clear command  CLS is    entered     Input Buffer    Calibrator stores all received data in the input buffer  The buffer holds  250 characters  The characters are processed on a first in  first out    basis     7 4 Remote Commands and Error Codes    The following tables list all commands  and their descriptions  that  are accepted by the calibrator     Table 5  Common Commands                            Command Description    CLS  CLS  Clear status   Clears the ESR  the error queue  and the  RQS bit in the status byte  Terminates pending O
27.  downloaded tags have status Loaded     If the download is terminated with the Cancel button  any tags  already downloaded remain in the calibrator and must be erased  separately if needed     Once the download is complete  the download status display  remains on the screen for review  Click the OK button to dismiss  it     9 8 Configure Menu    Tag    Initially CL310 SW is set to store tags and groups in the root  folder  Omega CL310 SW  within the  My Documents  folder  If  you wish to store them into another folder  navigate to the  alternate folder before saving the first tag or group  From then  on  until a further alternate folder is chosen  that folder will be the  default for storage     New    Create a new tag configuration  A dialog screen with the tabs  Identification  Instrument Input  Instrument Output  Tolerance   and Test Points is displayed for entry of the configuration data   This dialog screen is described in detail in section 9 9 below   When complete  the configuration is saved into a user named file  with the fixed extension UTG     Edit   Edit an existing tag configuration  After a tag configuration file is  selected  a dialog screen with the tabs Identification  Instrument  Input  Instrument Output  Tolerance  and Test Points is displayed  for modification of the configuration data  This dialog screen is  described in detail in section 9 9 below  When complete  the  configuration is saved into a user named file with the fixed  extension UTG  This may b
28.  gt  al                            5  1 2 3  0          MEASURE   SOURCE MEASURE       ET    v                                              O                   S GG    Figure 1  Input Output Terminals    Table 1  Input and Output Terminals                               No    Name Description  1 2   Measure Isolated V  Input terminals for measuring current  voltage   mA terminals and supplying loop power   3 TC input output Terminal for measuring  or simulating thermo   couples  Accepts miniature polarized thermo   couple plugs with flat in line blades spaced 7 9  mm  0 312 in  center to center   4 5   Source Measure Terminals for sourcing and measuring voltage   V RTD 2W  Hz  frequency  pulse train  and RTDs  6 7   Source Measure Terminals for sourcing and measuring current   mA terminals  3W 4W   and performing  RTD measurements with 3 wire or 4 wire  setups   8 Pressure module Connects calibrator to a pressure module for  connector pressure measurements   9 Serial port Connects calibrator to a PC for uploading data  or remote control or to a serial printer for print   ing calibration certificates        Figure 2 shows the location of the keys on the calibrator  Table 2 lists  the functions of each key                OMEGA    cL310                                           FA F2          o                         O0          O los                                                    w  zmazm       MEASURE   SOURCE  MEASURE                   Figure 2  Keypad    Table 2  Key Fun
29.  input has settled  the technician presses a function  key to prompt for manual input entry if any  to save both the As  Found output and input values  and finally to step to the next point     51    52    Once the last point has been tested  the As Found test automatically  ends  the real time clock time stamp is saved  and the technician is  prompted for changes to the tag identification data  any field but tag  number which can not be changed   If test validation was selected   the technician has the option of displaying the As Found test results     After either type of As Found test is ended  the technician has the  option of saving the As Found results into the As Left storage and  ending the tag calibration  of entering an adjustment mode to  monitor physical adjustments to the tag prior to starting an As Left  test  or of directly entering the As Left test     During adjustment mode  the technician can step up and down the  list of test points and the calibrator automatically generates the  corresponding output in the case of a calibrator source type  or  displays the value of a manual output type  In this way any changes  to the original test points resulting from physical adjustments to the  instrument may be evaluated  Alternatively  the technician can enter  output values different from the test points and monitor  corresponding input values  When the adjustment is complete  a  function key is pressed to enter the As Left test     During the As Left test the calib
30.  shown  in the upper right corner of the display  Note that the point number is  the next point to be tested and data has not been saved for it yet     If the calibrator output is pressure  adjust the external pressure  source until the displayed output value matches the desired test  value  For all other output types  enter the output value to be  generated using the numeric keys  followed by ENTER     Once the displayed calibrator input value has settled  press the  SAVE function key to save the data and step to the next point  The  point number in the upper right increases to show that the test data  is saved     Once the desired number of points have been tested  press the  DONE function key to end the data collection stage and proceed to  the entry of test identification data     If for any reason the data collection is to be terminated without    saving any data  press either the HOME key or the ABORT function  key  A prompt is displayed requesting confirmation that the test data  accumulated so far is to be abandoned  Press the NO function key  to return to the data collection process at the point where it was  interrupted  Press the YES function key to abandon the data  collected     Pulse Train Type Prompt    When the input output combination includes a pulse train output  the  following is displayed for each test point        VOLTS IN AF 1  PULSE OUT 1 0 VpP    60 0 Hz    TRIG     READY             The current test type  AF for As Found  and point number are shown  
31.  the temperature unit that is currently used for measuring and  sourcing RTD and TC     TRIG    Starts and stops the pulse train when the calibrator is in pulse mode   The parameters of the pulse train are set by commands PULSE          and FREQ LEVEL  Entering TRIG initializes the train  Entering the  command while the pulse train is running stops it     TRIG     Returns TRIGGERED if the pulse train is running  and returns  UNTRIGGERED when the pulse train is not running  Returns NONE  when the calibrator is not in pulse mode     TSENS TYPE    Sets the temperature sensor type to thermocouple  or to RTD for  temperature measurement  After the command add TC for  thermocouple  or RTD for RTDs  For example     TSENS TYPE TC sets the sensor type to thermocouple  49    50    TSENS TYPE     Returns the type of sensor that is currently set to measure  temperature  either TC or RTD     UPPER MEAS    Sets the measuring mode for the upper display  After the command  enter DCI for mA  DCI LOOP for mA with loop power  DCV for volts   and PRESSURE for pressure  For example     UPPER MEAS DCV sets the upper display to measure volts    U PRES UNIT    Sets the unit for measuring pressure on the upper display  Add the  unit after the command  The available pressure units and their syntax  are shown in Table 7   Parameter Units   For example     U PRES UNIT MMHG sets the pressure unit to millimeters of  mercury at 0  C    VAL     Returns the value of any measurement taking place on the upper 
32.  through the  remaining pages of data  cycling from the last page to the first page  and vice versa  The second page contains the remaining  identification fields and the validation error tolerance  The third page  contains the instrument input setup  The fourth page contains the  instrument output setup  The fifth and subsequent pages display the  test results  On these pages only  use the LEFT ARROW and RIGHT  ARROW keys to switch between the As Found results and the As Left  results                 Note that the input output orientation of the data on these pages is  with respect to the instrument under test  This is in contrast to the    orientation during testing  which was with respect to the calibrator     Press the DONE function key to return to the tag selection menu   Press the EXIT function key to return to the review selection menu     8 7 Printing Test Results    At the main menu  press the MENU function key  followed by the  MORE function key  followed by the DOCUMENT function key to  display the first level document mode menu        TEST REVIEW EXIT    Press the REVIEW function key to display the review selection menu        VIEW PRINT   CLEAR    Press the PRINT function key to display the tag selection prompt     FT 384  PI 884                      SELECT     EXIT    Tag numbers are displayed on multiple pages  six tags to a page   Press the UP ARROW and DOWN ARROW keys to scroll the tag  selector highlight one tag at a time  scrolling from page to page   cycl
33.  to the product  relative to the product     OMEGA s policy is to make running changes  not model changes  whenever an improvement is possible  This affords our  customers the latest in technology and engineering    OMEGA is a registered trademark of OMEGA ENGINEERING  INC       Copyright 2017 OMEGA ENGINEERING  INC  All rights reserved  This document may not be copied  photocopied   reproduced  translated  or reduced to any electronic medium or machine readable form  in whole or in part  without the prior  written consent of OMEGA ENGINEERING  INC     103    Where Do   Find Everything   Need for  Process Measurement and Control   OMEGA   Of Course    Shop online at omega com       TEMPERATURE     4 Thermocouple  RTD  amp  Thermistor Probes  Connectors   Panels  amp  Assemblies    A Wire  Thermocouple  RTD  amp  Thermistor    A Calibrators  amp  Ice Point References     Recorders  Controllers  amp  Process Monitors   IA Infrared Pyrometers    PRESSURE  STRAIN AND FORCE   4 Transducers  amp  Strain Gages      Load Cells  amp  Pressure Gages      Displacement Transducers    A Instrumentation  amp  Accessories    FLOW LEVEL    4 Rotameters  Gas Mass Flowmeters  amp  Flow Computers   A Air Velocity Indicators     Turbine  Paddlewheel Systems    4 Totalizers  amp  Batch Controllers    pH CONDUCTIVITY       pH Electrodes  Testers  amp  Accessories    A Benchtop Laboratory Meters    A Controllers  Calibrators  Simulators  amp  Pumps   4 Industrial pH  amp  Conductivity Equipment    
34. Cs  Frequency response is                    lt   5ms   Table 15  Thermocouple Measurement Source  Range Accuracy     of reading   floor   Read  mV   10 000mV   75 000mV  0 02    10uV  Source  mV   10 000mvV   75 000mV 0 02    10uV                Maximum current output in voltage ranges is 1mA with an output impedance of  lt   12       98             Table 16  Thermocouple Read and Source  errors in  C                                                                                                     Range    C  Accuracy    C    TC Type Minimum Maximum CJC OFF CJC ON  J  210 0  150 0 0 4 0 6   150 0 1200 0 0 2 0 4  K  200 0  100 0 0 5 0 7   100 0 600 0 0 2 0 4  600 0 1000 0 0 3 0 5  1000 0 1372 0 0 4 0 6  T  250 0  200 0 1 5 1 7   200 0 0 0 0 5 0 7  0 0 400 0 0 2 0 4  E  250 0  200 0 1 0 1 2   200 0  100 0 0 3 0 5   100 0 1000 0 0 2 0 4  R1 0 0 200 0 1 7 1 9  200 0 1767 0 1 0 1 2  51 0 0 200 0 1 7 1 9  200 0 1767 0 1 1 1 3  B1 600 0 800 0 1 5 1 7  800 0 1000 0 1 2 1 4  1000 0 1820 0 1 0 1 2   C 0 0 1000 0 0 5 0 7  1000 0 2316 0 1 5 1 7  XK  200 0 800 0 0 2 0 4  BP 0 0 800 0 1 9 2 1  800 0 2500 0 0 6 0 8  L  200 0 900 0 0 2 0 4  U  200 0 0 0 0 4 0 6  0 0 600 0 0 2 0 4  N  200 0  100 0 0 8 1 0   100 0 1300 0 0 3 0 5                   CJC error outside of 23   5  C is 0 05  C   C       1  For thermocouple measure mode on B  R and S  round the specification up or    down accordingly as there is no resolution past the decimal point        99    Table 17  RTD Read and Source       
35. DATA ACQUISITION    IA Communications Based Acquisition Systems   A Data Logging Systems     Wireless Sensors  Transmitters   amp  Receivers     Signal Conditioners      Data Acquisition Software    HEATERS      Heating Cable      Cartridge  amp  Strip Heaters  14 Immersion  amp  Band Heaters   A Flexible Heaters    A Laboratory Heaters    ENVIRONMENTAL  MONITORING AND CONTROL       Metering  amp  Control Instrumentation    4 Refractometers      Pumps  amp  Tubing    LA Air  Soil  amp  Water Monitors    4 Industrial Water  amp  Wastewater Treatment      pH  Conductivity  amp  Dissolved Oxygen Instruments       M4590 0217    
36. DC mA Measurement Source       Range Accuracy   96 of reading   floor     Read  Isolated  Upper Display  0 000mA   24 000mA 0 015    2uA  Read  non Isolated  Lower Display   0 000mA   24 000mA 0 015    2uA  Source 0 000mA   24 000mA 0 015    2uA    Maximum load on mA source is 10000  Voltage input range on simulate mode 5V   30V                          97    Table 12  Frequency Measurement Source       Range    Accuracy   96 of reading   floor                          Read 2 0CPM   600 0CPM 0 05    0 1CPM  1 0Hz   1000 0Hz 0 05    0 1Hz  1 00KHz   10 00KHz 0 05    0 01KHz  Source 2 0CPM   600 0CPM 0 05   1 0Hz   1000 0Hz 0 05   1 00KHz   10 00KHz 0 250        assure zero crossing     Input voltage amplitude range on frequency is 1V to 20V zero based square wave  only Output amplitude is adjustable from 1V to 20V  and is a square wave with 50  duty  cycle For output frequency  a slight negative offset of approximately  0 1V is present to       Table 13  Resistance Measurement                                        Range Accuracy     of reading   floor   Ohms low 0 000   400 09 0 015    0 039  Ohms high 401 00   4000 02 0 015    0 39  Table 14  Resistance Source  Range Excitation Current Accuracy   96 of reading   floor   Ohms low 5 00   400 00 0 1mA   0 5mA 0 015    0 12  5 0Q   400 00 0 5mA   3mA 0 015    0 03     Ohms high 4000   15000 0 05mA   0 8mA 0 015    0 3Q  15000   40000 0 05mA   0 4mA 0 015    0 39             Note  Unit is compatible with smart transmitters and PL
37. Data Collection    If pressure is selected on the upper or lower display  the pressure  zero prompt is displayed first  and operates as described under As  Found Setup     As Left data collection is identical to As Found data collection except  for four minor differences     1  The AF in the upper right corner of the display is changed to AL  to indicate that this is an As Left test     2  The test outputs can not be changed  they will be the same as  the As Found values  This ensures that the As Found and As  Left test results line up when being compared later  For pressure  and manual output types  the As Found value is displayed for the  technician to set it on the external source  For all other output  types  the calibrator automatically generates the As Found value     3  Once the SAVE function key is pressed for the last test point  the  process automatically proceeds to the test conclusion stage     4  AREPEAT function key is available for restarting the As Left test  at the first point     8 5 3 Test Conclusion    If test validation was selected  a display showing the overall pass fail  status of the test is displayed with a prompt asking whether details  should be displayed  Press the YES function key to display the  details  or the NO function key to skip the details  In either case the  adjustment prompt is displayed next    A prompt is displayed asking whether further adjustment and testing  is required  Press the YES function key to proceed to the adjustment  men
38. ER   and   MORE      Choosing  UPPER  calls up the parameter selection menu for the  upper display  Choosing  LOWER  calls up the parameter selection  menu for the lower display   MORE  enters the next menu level     UPPER   LOWER   MORE    The Document Mode selection menu is next  Its options are   DOCUMENT    NEXT   and  DONE   Choosing  DOCUMENT   enters the document mode menu system described in section 8 of  this manual   NEXT  proceeds to the next menu level  and  DONE   returns to the home menu     DOCUMENT   NEXT   DONE    The Auto Function Menu is the next menu if you are in source mode   Its options are  AUTO FUNC    NEXT  and  DONE    AUTO FUNC   allows you to adjust the Automatic Output Function parameters    NEXT  proceeds to the next menu level and  DONE  returns to the  home menu  See Section 4 2 Using the Automatic Output Functions     AUTO FUNC   NEXT   DONE    The contrast menu is usually the next menu level  Its options are   CONTRAST    NEXT   and  DONE   The  CONTRAST  option is  used to adjust contrast   NEXT  proceeds to the auto off main menu   and  DONE  returns to home menu  Contrast is adjusted using the  arrow options  which are available after choosing  CONTRAST      CONTRAST  NEXT   DONE    NOTE  The CL310 calibrator offers a wide range contrast adjustment  feature to accommodate operation in extreme temperatures                                                                 In certain cases making large changes in contrast may render the  displ
39. Hold down the Function Key that corresponds to 096  F1   096  will flash and the setpoint is stored     6  Repeat these steps  entering 20V and holding the Function Key  that corresponds to 100             7  Use the 25926 key to cycle 5 V and 20 V in 2596 increments     4 1 1 Stepping the current output  To use the 25  function with mA output  follow these steps     1  Select the lower display from the Main Menu  and choose  mA     2  Use the 2596 key to cycle between 4 mA and 20 mA in 25    intervals     4 2 Using the Automatic Output Functions    There are two automatic output functions  step and ramp  The  selected function can be turned on and off using the Output Home  Menu  The Automatic Output Function parameters can be set in the  Auto Function Menu     17    Parameters include    1  Which auto function will be available  Step or Ramp     2  The Auto Function Time  time between steps for step and time to  get from over one limit to the next for ramp     The limits for the ramp and step functions are set to the 096 and  10096 values  See Section 4 1 Setting 096 and 10096 Output  Parameters  Steps are in 25  increments from the 0  value to the  10096 value     4 3 Sourcing mA    To source a current  follow these steps     1  From the Main Menu select lower display  LOWER   Choose   mA  from the primary parameters     2  Switch to input output control  and select output  OUT    3  Connect the leads to the mA terminals  as shown in Figure 10   4  Enter the desired curren
40. RT MAX T    Sets the maximum temperature of the custom RTD range  The  temperature value must be entered with a degrees label  CEL for  Celsius and FAR for Fahrenheit     For example     CPRT MAX T 0 0 CEL enters 0 0  C as the maximum  temperature     CPRT MIN T     Returns the value entered for minimum temperature in the range for  a custom RTD  The above example would return     0 000000E 00  CEL    CPRT_RO    Sets the 0   resistance  RO  in the custom RTD  The value must be  entered with a units label  Refer to the Parameter Units table for  assistance     For example   CPRT_RO 100 OHM sets RO to 100 ohms   CPRT RO     Returns the value for the resistance in custom RTD  The above  example would return     1 000000E  02  OHM    43    44    FAULT     Returns the error code number of an error that has occurred  The  command may be entered when the previous command did not do  what it was meant to do     For example  if a value for current output is entered that is bigger  than the supported range  0 24mA  FAULT  Would return     103 which is the code number for an entry over range     Refer to the Error Codes table for more information on error code  numbers     FREQ LEVEL    Sets the amplitude of the wave used in the Frequency Out and Pulse  modes  The range for amplitude entered may be found in the  Specifications section     For example   FREQ LEVEL 5 V sets the amplitude at 5Vpp     FREQ LEVEL     Returns the amplitude of the wave used in Frequency Out and Pulse  modes   
41. RTD_WIRE 4W sets the connection to 4 wire    RTD_WIRE   Returns the number of wires used in the RTD connection     SET_CLOCK yyyy mm dd hh nn ss    Sets the calibrator date and time  where yyyy is the four digit year   2006 to 2100   mm is the two digit month  1 to 12   dd is the two  digit day  1 to the number of days in the specified month   hh is the  two digit hour in 24 hour format  0 to 23   nn is the two digit minute   0 to 59   and ss is the two digit second  0 to 59   Note that all two  digit values must have a leading zero if they have a value less than  10  Returns   Complete if successful  otherwise an error message  enclosed in angle brackets  For example     SET CLOCK 2006 03 20 09 16 33  47    48    SIM    Sets the output for current simulation  This command also switches  the calibrator into mA simulation mode  A number and a unit must be  entered after the command  For example     SIM 5 MA sets the current simulation at 5 mA    SIM     Returns the output of the current simulation  With the example  above  the output would be     5 000000E 08  A    TAG CLEAR tag    Clears test data for the specified tag from the calibrator  Once the  test data is cleared  it can not be recovered  Returns   Complete   if  successful  otherwise an error message enclosed in angle brackets   For example     TAG CLEAR FT 8567    TAG CLEAR ALL    Clears test data for all tags from the calibrator  Once the test data is  cleared  it can not be recovered  Returns   Complete   if succes
42. T   EXIT          p          Menu                   Downloaded Tag  As Found Test                As Found Setup Menu          UPPER   LOWER   DONE                     Parameter Selection          SELECT   NEXT   DONE             Document Mode Menu Tree    54     gt                                Frequency Out Pulse Frequency Level Menu  or Pulse Mode    Y  S DONE           1                   Pulse Mode   No             i Yes       Pulse Parameter Menu       FREQ      COUNTS   DONE       Ya                   Upper des is    Upper Units Menu       manual type                DONE                     Lower Units Menu       Lower display is mmt    DONE                   manual      Yes    Either display in    Pressure Zero Menu       pressure mode           ZERO        DONE          No                Pulse mode  AF data collection     pulse trigger  TRIGGER     READY                No i    AF data collection     pulse stop  STOP                                Manual Input or AF data collection     manual entry                Yes       Output    NEXT   DONE                 1              l                                                   Manual output entry  Yes  DONE               Manual input entry  Yes             DONE                                   AF data collection     save test point       SAVE   DONE              LEFT KEY             turp rn                            Tag ID field selection       SELECT              SAVE          Tag field data entry          DONE     
43. TER     If the calibrator input is not manual  wait for the displayed calibrator  input value to settle and then press the NEXT function key to prompt  for the manual entry s  as described in the following subsection     If the calibrator input is manual  wait for the external device reading  to settle and then press the NEXT function key to prompt for the  manual entry s  as described in the following subsection     Upon return from the manual entry prompts s   the NEXT function  key label changes to SAVE  If the data displayed is correct  press the  SAVE function key to save the data and step to the next point  The  point number in the upper right increases to show that the test data  is saved  If the manual data must be corrected before saving it  press  the LEFT ARROW key to change the SAVE function key back to  NEXT  and press it to repeat the manual entry s      Once the desired number of points have been tested  press the  DONE function key to end the data collection stage and proceed to  the entry of test identification data     If for any reason the data collection is to be terminated without  saving any data  press either the HOME key or the ABORT function  key  A prompt is displayed requesting confirmation that the test data  accumulated so far is to be abandoned  Press the NO function key  to return to the data collection process at the point where it was  interrupted  Press the YES function key to abandon the data  collected     Manual Data Entry Prompt    A prom
44. The byte is returned in decimal format   For example     If 40 is returned  bits 5 and 3 are enabled      STB    Returns the status byte in decimal form from the Serial Poll Status  Byte  For example     If 72 is returned  bits 6 and 3 are enabled      WAI    Prevents further remote commands from being executed until all  previous commands are executed  For example     OUT 10 MA    WAI   OUT 5 V will out put 10mA and wait until  output settles  than volts command will be processed     7 5 2 Calibrator Commands    CAL START    Puts the calibrator in calibration mode  The main display will say  CALIBRATION MODE and a calibration menu will be displayed on  the terminal     CJC STATE    Turns Cold Junction Compensation  CJC  on or off  when the  calibrator is in thermocouple  TC  mode  The command is used by  adding ON or OFF after it    For example     CJC STATE OFF turns CJC off     CJC STATE     Tells whether the Cold Junction Compensation in thermocouple  mode is turned on or turned off  The calibrator returns OFF if CJC is  off  and ON if CJC is on     CPRT COEFA   This command is used for entering a custom RTD into the calibrator    The numeric value entered after the command will be set as the first  41    42    coefficient of the polynomial used by the custom RTD   For example   CPRT COEFA 3 908E 03 enters 3 908e 3 as coefficient A     CPRT_COEFA     Returns the number which was entered for the first coefficient of the  polynomial used in the custom RTD  Using the example
45. UPPER MEAS Sets the measuring mode for the upper display   U PRES UNIT Sets the upper pressure unit  VAL  Returns the measured values  ZERO MEAS Zeros the pressure module  ZERO MEAS  Returns the zero offset of the pressure module  Table 7  Parameter units  Units Meaning  MA milliamps of current  MV Voltage in millivolts  V Voltage in volts  CPM Frequency in cycles per minute  Hz Frequency in Hertz  KHz Frequency in kiloHertz  Ohms Resistance in Ohms  Cel Temperature in Celsius  Far Temperature in Fahrenheit  Psi Pressure in pounds per square inch  InH204C Pressure in inches of water at 4  C  InH2020C Pressure in inches of water at 20  C  CmH204C Pressure in centimeters of water at 4  C  CmH2020C Pressure in centimeters of water at 20  C  Bar Pressure in bars  Mbar Pressure in millibars  KPal Pressure in kiloPascals  InHg Pressure in inches of mercury at 0  C  MmHg Pressure in millimeters of mercury at 0  C  Kg cm2 Pressure in kilograms per square centimeter          Table 8  Error codes                Error Number Error Description  100 A non numeric entry was received where it should be a  numeric entry  101 Too many significant digits entered  102 Invalid units or parameter value received  103 Entry is above the upper limit of the allowable range          38             Error Number Error Description                   104 Entry is below the lower limit of the allowable range  105 A required command parameter was missing   106 An invalid pressure unit was received   107 An i
46. V  DONE        Enter the new peak to peak voltage using the numeric keys followed  by the ENTER key  or if satisfied with the current voltage press the  DONE function key to bypass changing it  When satisfied with the  voltage value  press the DONE function key to continue              Optional Setup   Pulse Train Output on Lower Display    If pulse train output is selected on the lower display  a prompt for  peak to peak voltage is displayed  followed by a prompt for a fixed  frequency or a fixed number of pulse counts  The peak to peak  voltage prompt is as follows        Enter peak to peak    voltage  1 0 V  DONE        Enter the new peak to peak voltage using the numeric keys followed  by the ENTER key  or if satisfied with the current voltage press the  DONE function key to bypass changing it  When satisfied with the  voltage value  press the DONE function key to continue     Pulse train has two variable parameters  frequency and pulse count   Since the calibrator can only store one variable parameter per test   one of these must be set to a fixed value for all test points  while the  other is changed  A second prompt allows the fixed parameter to be  selected  and a value to be entered for it  In the sample below   frequency is currently selected as the fixed parameter with a value of  2 0 Hz                 Enter fixed pulse  frequency    2 0 Hz    FREQ   COUNTS   DONE             Press the FREQ function key or the COUNTS function key to select  the fixed parameter  Th
47. al Input   ve P AL data collection     manual entry  NEXT                               1                   Y     Manual input entry F  DONE               py  v    AL data collection     save test point  Last data point   I                SAVE                               1              Yes Y  v F  Save As Left data  and time stamp  v  Validating Yes  results           Show test results   YES   NO          v  View test results  DONE      pre  y       AL complete  adjustment required           YES   NO            Y      D Document Menus                               Review Menu                     Review Function Select             VIEW   PRINT   CLEAR                T       View Tag Select       SELECT     EXIT                                                      Print Tag Select                                                                                                 SELECT   EXIT  View Tag    DONE   EXIT    Review Menu  G  Y    v  Clear Tag Select  SELECT   CLEAR ALL   EXIT     y v  Confirm Tag Clear Confirm Clear All Tags  YES NO YES   NO    Y Y  G Review Menu                                  Final Confirm Clear All Tags          YES   NO                           61                   Function Selection Menu       TEST         REVIEW   EXIT          Y          Review Function Select          VIEW      PRINT       CLEAR          Test Selection Menu          AS FOUND   ASLEFT   EXIT             Download       New       Tag Source                      Configure Upper a
48. alibrator securely  using the  original shipping container if it is available     11 4 Replacement Parts  amp  Accessories    102    PCL PMA OMEGA Pressure Module Adapter  DPG4000 C32   RS232 Cable with Miniature Plug  PCL422 TL 48  Stackable Testleads  CL20 Kit 2 Thermocouple Test Lead Kit    WARRANTY DISCLAIMER    OMEGA ENGINEERING  INC  warrants this unit to be free of defects in materials and workmanship for a  period of 13 months from date of purchase  OMEGA s WARRANTY adds an additional one  1  month grace  period to the normal one  1  year product warranty to cover handling and shipping time  This ensures  that OMEGA s customers receive maximum coverage on each product     If the unit malfunctions  it must be returned to the factory for evaluation  OMEGA s Customer Service  Department will issue an Authorized Return  AR  number immediately upon phone or written request   Upon examination by OMEGA  if the unit is found to be defective  it will be repaired or replaced at no  charge  OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser   including but not limited to mishandling  improper interfacing  operation outside of design limits   improper repair  or unauthorized modification  This WARRANTY is VOID if the unit shows evidence of  having been tampered with or shows evidence of having been damaged as a result of excessive corrosion   or current  heat  moisture or vibration  improper specification  misapplication  misuse or other operati
49. an RTD  follow these steps     1  Select lower display from the Main Menu  and choose  RTD   from the primary parameters    2  Choose output  OUT  from the input output control  and select  RTD type from the sensor types     3  Connectthe calibrator to the instrument being tested  as in  Figure 14     4  Enter the temperature or resistance using the keypad          MS Use                     Os test leads for    y U        shared connections                                                                     22 Figure 15  Using a 3  or 4 wire Connection for RTDs          mA LOOP    RTD OUT     25 00     ___Q 000 mA    P500 385    10 0      C       MENU   LIGHT            Note  The calibrator simulates a 2 wire RTD  To connect 3  or 4 wire  transmitter  use stacking cables  as shown in Figure 15     4 9 1 Custom RTD    A custom curve fit PRT may be entered into the calibrator for  sourcing and measuring  To do so follow these steps     1  Switch to lower display  Select RTD and set sensor type to  CUSTOM     2  Enter the RTD custom setup main menu  and select  SET    CUSTOM      3  Using the keypad  enter the values that the calibrator  prompts for  minimum temperature  maximum  temperature  RO  and the values for each of the  temperature coefficients     The custom function uses the Calendar Van Dusen equation for  outputting and measuring custom RTDs  The coefficient C is only  used for temperatures below 0  C  Only A and B coefficients are  needed for the range above 0  C  
50. as described under viewing test results  Press the  SELECT function key to erase the tag  An additional prompt is  displayed to confirm the request  Press the YES function key to erase  the tag  or the NO function key to abort the request  To erase  additional tags  highlight each in turn and press the SELECT function  key  When all tags have been erased  press the EXIT function key to  return to the review selection menu     8 9 Setting the Date and Time    At the main menu  press the MENU function key  followed by the  MORE function key  followed by the NEXT function multiple times to  get to the clock selection menu     CLOCK NEXT   DONE    Press the CLOCK function key to display the clock data entry screen        84       2006 03 28 15 21 14    Enter year    2006    BACK NEXT   SAVE    Initially the date and time values displayed on the top line are the  precise clock setting when the screen was first displayed  After that  they are static values representing any data changes made  The  clock is not updated until the SAVE function key is pressed  For the  most accurate setting it is recommended that all values but the  seconds be entered first  then enter a seconds value slightly ahead  of the current time  and finally press the SAVE function key exactly  when the entered seconds is reached     Press the NEXT and BACK function keys to switch between the  individual date and time fields  The upper part of the display will  change to reflect the current value for that field
51. ay difficult to read under normal conditions  If this occurs and  the display is too dim or dark to read  proceed with the following    process to set the contrast back to a default setting   1  Turn on the unit while holding down the  HOME   key     2  Hold the key down for a count of 10 seconds to restore contrast  default settings     If the display is so dim that you cannot tell if the unit is on or off  use  the backlight key to determine if the power is on or off     The auto off main menu contains the options  AUTO OFF    NEXT    and  DONE      The  AUTO OFF  option is used to turn the automatic shutoff on and  off  and to set the amount of time the unit needs to stay dormant  before it shuts off   NEXT  proceeds to the clock menu  and  DONE   returns to the home menu     AUTO OFF   NEXT   DONE    The Clock menu is the next menu displayed in the  MORE  menu  sequence  Its options are  CLOCK    NEXT   and  DONE   Choose   CLOCK  to set the calibrator date and time as described in section 8  of this manual   NEXT  to proceed to the terminal emulation menu   and  DONE  to return to the home menu     CLOCK   NEXT   DONE    The Terminal menu is the last menu displayed after choosing   MORE  in the main menu  Its options are  TERMINAL    NEXT   and   DONE   Choose  TERMINAL  to enter terminal emulation mode   Select  NEXT  or  DONE  to return to the home menu     TERMINAL  NEXT   DONE    When the lower display is in the frequency or pulse mode  the  frequency level menu is a
52. be used in or with any nuclear installation or  activity  medical application  used on humans  or misused in any way  OMEGA assumes no responsibility  as set forth in our basic WARRANTYDISCLAIMER language  and  additionally  purchaser will indemnify  OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the  Product s  in such a manner     RETURN REQUESTS INQUIRIES    Direct all warranty and repair requests inquiries to the OMEGA Customer Service Department  BEFORE  RETURNING ANY PRODUCT S  TO OMEGA  PURCHASER MUST OBTAIN AN AUTHORIZED RETURN  AR   NUMBER FROM OMEGA S CUSTOMER SERVICE DEPARTMENT  IN ORDER TO AVOID PROCESSING  DELAYS   The assigned AR number should then be marked on the outside of the return package and on any  correspondence     The purchaser is responsible for shipping charges  freight  insurance and proper packaging to prevent  breakage in transit     FOR WARRANTY RETURNS  please have the FOR NON WARRANTY REPAIRS  consult  following information available BEFORE contacting OMEGA for current repair charges  Have  OMEGA  the following information available BEFORE  1  Purchase Order number under which the product contacting OMEGA    was PURCHASED  1  Purchase Order number to cover the COST  2  Model and serial number of the product under of the repair    warranty  and 2  Model and serial number of the product  and  3  Repair instructions and or specific problems 3  Repair instructions and or specific problems   relative
53. calibrator into  the Reports list box  Before using this function  upload a list of  the tag configurations contained in the calibrator using the Tag  ID s function above  In the Tag ID s list box  highlight the tags  which are to be uploaded  and then select this function to upload  them  The Reports list box is cleared before the upload  Use  the Configure menu to select the communications port     Only reports for calibrated tags may be uploaded  Downloaded  tags which have not yet been calibrated are skipped after a  warning message is displayed     Multiple individual tags are selected by holding down the Ctrl key  and clicking on them one at a time  Multiple contiguous tags are  selected either by clicking on the first and dragging to the last  or  by clicking on the first and holding down the Shift key while  clicking on the last     When multiple tags are uploaded into the Reports list box they  are separated from each other by a line of dashes  If the tags  are saved into separate files  the dashes are removed     The reports may be uploaded in text format suitable for direct  printing and use with word processors  or in CSV  comma  separated variables  format suitable for importing into  spreadsheets  The format is selected below     Your company name can be inserted at the top of each  uploaded report  Use the Configure menu to enter a name to be  inserted into all calibration reports  Nothing is inserted if this  configuration item is left blank     Upload in Text
54. ctions                      No    Name Function  1 Function Keys Used to operate the menu bar at the bottom of  F1  F2  F8 the calibrator display  F1 is used for selecting  options in the left box  F2 for the center box   and F3 for the right box   Home Returns to home menu on the menu bar   Power Turns calibrator on and off   Cursor Control Key Left and right arrow keys are used to select  which decade to be changed in output value   Up and down arrow keys are used to increase   decrease  or ramp output value   5 Numeric Keypad Allows user to enter Numeric values              2 1 Main Display     7          mA IN    M  Ero IN 4W    Qum MENU   LIGHT             PME              Figure 3  Display    The display of the calibrator  shown in Figure 3  is divided into three  main sections  the upper display  the lower display  and the menu    bar     The upper display is used for measuring dc voltage  dc current with  and without loop power  and pressure     The lower display can be used for both measuring and sourcing     The menu bar is used to setup both the upper and the lower display  to perform the desired function     Table 3 explains the different parts of the display     Table 3  Display Functions       No     Name    Primary Parameters    Description    Determine what parameter is going to be meas   ured or sourced  The available options for the  upper display are VOLTS IN  PRESSURE  mA  IN  and mA LOOP The available options for the  lower display are  VOLTS  TC  thermoc
55. d  Otherwise pressing DONE  proceeds directly to the As Found data collection prompt     Optional Setup   Manual Input or Output on Upper Display    If either manual type is selected on the upper display  the upper  manual unit description prompt is displayed        Upper Units     7ABC 8DEF 9GHI  4JKL 5MNO 6PQR  ISTU 2VWX 3YZ     0          DONE              Up to 5 characters may be entered for a unit description using a  telephone style entry method  The four lower lines provide a  mapping of the allowed alphanumeric characters to the numeric  keypad keys  To enter a character press the corresponding numeric  key multiple times to cycle through the mapped characters until the  desired character is displayed  For example  to enter E  press the 8  key 3 times  If the next character is on a different numeric key   simply press that key to advance the cursor and display the first  mapped character  If the next character is on the same key  press  ENTER to advance the cursor before pressing the same key  To  insert a space  press ENTER twice  To erase characters press CE   Press the DONE function key when the unit description is complete     Optional Setup   Manual Input or Output on Lower Display    If either manual type is selected on the lower display  the lower  manual unit description prompt is displayed  Except for the title on  the first line  it looks like and operates identically to the upper display  version described in the previous paragraph     Optional Setup   Pre
56. dded after the main menu  The options  available in this menu are  FREQ LEVEL    NEXT   and  DONE   The   FREQ LEVEL  option is used to adjust the amplitude of the wave    NEXT  is used to access the contrast main menu  and  DONE   returns to the home menu     FREQ LEVEL  NEXT  DONE    When the calibrator is in RTD CUSTOM mode  the RTD custom  setup menu  is inserted after the main menu  Options  SET  CUSTOM    NEXT   and  DONE  are available   SET CUSTOM  is  used to enter a custom PRT into the calibrator  Refer to Section 4 1                                   8a for instructions   NEXT  is used to enter the contrast main menu   and  DONE  to return to the home menu     SET CUSTOM   NEXT   DONE    The pressure zeroing main menu is the final variation to choosing   MORE  in the main menu  It has the options  ZERO    used to zero  pressure   NEXT  and  DONE   which have the same function as  above  Refer to the Section 5 3 for instructions on zeroing                    ZERO      NEXT   DONE    The parameter selection menu is called up when  UPPER  or   LOWER  is selected from the main menu  It contains the following  options   SELECT    NEXT   and  DONE   When the display is  selected  a parameter will start to flash  Use the  SELECT  option to  change the parameter  and the  NEXT  option to switch to another  variable   DONE  returns to the home menu and enables the  selected mode        SELECT   NEXT   DONE             2 3 Cursor control   Setpoint control    The output valu
57. e can be controlled by the four cursor control arrows  on the keypad  By pressing one of the arrows a cursor will be added  to the display under the last digit of the output value  The left and  right arrow keys are used to select which decade to be changed in  the output value  The up and down arrow keys are used to increase   decrease  or ramp the output value     The menu bar will change to the setpoint menu with the touch of any  one of the four arrow keys     0  25  100     The three function keys are associated with 0  25  and 100  values   respectively  0 and 100  values can be stored by entering a value  and then holding down the corresponding function key  The 2596 key  will then automatically step through the 25  values        11    12                                                                                                                                                                                        Home Menu  MENU   LIGHT   RAMP  Selection Menu  UPPER   LOWER   MORE  Parameter Selection Document Mode Selection  SELECT   NEXT   DONE DOCUMENT   NEXT   DONE      Home Menu Document Menus  AA  Source Mode  Lus T Auto Function Menu  es  AUTO FUNC   NEXT   DONE              M  v      v  Frequency Out    Frequency Level Menu  or Pulse mode    Yes    FREQLEVEL   NEXT   DONE              M  v      v  RTD                                RTD Custom Menu  mode  Yes  d sercustom   NEXT   DONE              M  N  Y  v  Either display                   J   Pressure Z
58. e directory as the group     To add tags to the group  highlight them in the available list and click  the right arrow  To add a single tag  double click it     To remove tags from the group  highlight them in the group list and  click the left arrow  To remove a single tag  double click it     Multiple individual tags are selected by holding down the Cir  key  and clicking on them one at a time  Multiple contiguous tags are  selected either by clicking on the first and dragging to the last  or by  clicking on the first and holding down the Shift key while clicking on  the last     10  Specifications  All measurements apply at 23  C   5  C  unless specified otherwise     Outside of this range the stability of the measurements is    0 005 of reading   C     Table 9  General Specifications       Operating Temperature    10  C to 50  C       Storage Temperature    20     to 70  C       Power 4 X AA batteries  Alkaline or optional rechargeable       Low battery warning Yes       Serial Communications   Yes  ASCII  CE   EMC EN50082 1  1992 and EN55022  1994 Class B  Safety CSA C22 2 No  1010 1  1992                Table 10  DC Voltage Measurement Source                Range Accuracy     of reading   floor   Read  Isolated Upper Display  0 000V  30 000V 0 015    2mV  Read  non lsolated Lower Display    0 000V   20 000V 0 015    2mV  Source 0 000V   20 000V 0 015    2mV             Maximum current output in voltage ranges is 3mA with an output impedance of  lt   1Q        Table 11  
59. e the original file name or a new one     Extract from Report    Extract a tag configuration from a previously uploaded and  saved report file  If the selected file contains multiple tags  a list  is displayed for selection of the single tag to be extracted  Once  the tag is selected  or directly if the report file contained a single  91    tag  the configuration data is extracted and presented in a dialog  screen with the tabs Identification  Instrument Input  Instrument  Output  Tolerance  and Test Points for modification of the  configuration data  This dialog screen is described in detail in  section 9 9 below  When complete  the configuration is saved  into a user named file with the fixed extension UTG     Display Print  Display a tag configuration  and optionally print it     Group    Initially CL310 SW is set to store tags and groups in the root  folder  Omega CL310 SW  within the  My Documents  folder  If  you wish to store them into another folder  navigate to the  alternate folder before saving the first tag or group  From then  on  until a further alternate folder is chosen  that folder will be the  default for storage     New    Create a new group of tags  Once the first tag for the group is  selected with a standard file dialog  a configuration screen is  displayed for modifying the contents of the group  This dialog  screen is described in detail in section 9 10 below  All tags  which are members of the group must reside in the same  directory as the group  Whe
60. e train  output has its own unique prompt sequence  All other combinations  are covered by a generic prompt     Generic Type Prompt    A prompt similar to the following is displayed for each test point  when the output is not pulse train     mA IN        4 000         VOLTS OUT    1 000 V   UP   DOWN   ASLEFT    The PT in the upper right corner of the display indicates that this is  the adjustment display  If an As Found data point is being  generated  its number is displayed  If a user entered value is being  generated  two dashes are displayed     Press the UP or DOWN function key to cycle through the As Found  test points  If the output type is not pressure or manual  the  calibrator will generate it  If the output type is pressure or manual  the  output value is displayed so that it can be set on the external source              If the output type is not pressure or manual  any value different from  the test points can be entered using the numeric keys  followed by  ENTER     Press the AS LEFT function key to proceed directly to As Left data  collection     Pulse Train Type Prompt    When the output is pulse train  the following is displayed for each  test point        VOLTS IN PT 1    TTA    PULSE OUT 1 0 VpP    60 0 Hz    TRIG     READY    The PT in the upper right corner of the display indicates that this is  the adjustment display  If an As Found data point is selected  its  number is displayed  If a user entered value is selected  two dashes  are displayed           
61. e upper display will change to reflect the  current value for that parameter  Enter the new parameter value  using the numeric keys followed by the ENTER key  or if satisfied  with the current parameter selection and value press the DONE  function key to bypass changing it  When satisfied with the  parameter selection and value  press the DONE function key to  continue     67    68    8 2 2 Data Collection    The test for each point consists of setting the output value  waiting  for the input to settle  and saving the data  This process is repeated  until the desired number of test points are recorded  For ease of  analysis  it is suggested that the test points be evenly spaced across  the test range and that they be entered in ascending order   descending order  ascending followed by descending  or  descending followed by ascending     The prompt displayed during the test depends on the input and  output types selected  Any combination which includes pulse train  output has its own unique prompt sequence  as does any  combination which includes a manual input  manual output  or both   All other combinations are covered by a generic prompt     Generic Type Prompt    A prompt similar to the following is displayed for each test point  when the input output combination does not include pulse train  output  manual input  or manual output     mA IN AF 1    4 000 mA    VOLTS OUT    1 000 V    SAVE DONE   ABORT             The current test type  AF for As Found  and point number are
62. ed at a  later time  the tag must be selected first     8 5 1 Tag Selection    At the main menu  press the MENU function key  followed by the  MORE function key  followed by the DOCUMENT function key to  display the first level document mode menu     79    80       TEST REVIEW EXIT    Press the TEST function key to display the test selection menu     AS FOUND   AS LEFT   EXIT    Press the AS LEFT function key to display the tag selection prompt  which shows all previously calibrated tags available for recalibration                       FT 384  PI 884    SELECT     EXIT    Tag numbers are displayed on multiple pages  six tags to a page   Press the UP ARROW and DOWN ARROW keys to scroll the tag  selector highlight one tag at a time  scrolling from page to page   cycling from top to bottom of the list and vice versa  Use the LEFT  ARROW and RIGHT ARROW keys to scroll the tag selector highlight  one page at a time  cycling from top to bottom of the list and vice  versa        Press the EXIT function key to return to the main menu without  selecting a tag     When the desired tag number is highlighted  press the SELECT  function key     If the there has been no previous As Left test proceed directly to data  collection     If there has been at least one previous As Left test  a prompt  requesting confirmation to overwrite existing As Left data is  displayed  Press the YES function key to continue to data collection   Press the NO function key to return to tag selection     8 5 2 
63. ep              8 2 4 Results Validation    After the identification data has been saved  a prompt is displayed  asking whether test results are to be validated or not  If validation is  selected  each test point is assigned a pass or fail status on the test  results displays and reports  and validation is automatically  performed at the end of each As Left test  If validation is not selected   no status is displayed on the test results displays and reports  and  no As Left validation is performed     Validation is based on a linear relationship between calibrator output  and input  Zero and span values are entered for both input and  output  along with a   span error tolerance  The zero and span end  points are used to calculate expected calibrator input values  corresponding to the actual calibrator output values  The expected  input values are then compared to the actual input values and  pass fail status determined by the   span error tolerance     73    74    If the instrument s output input relationship is not linear  test results  validation should not be selected    Press the YES function key to validate the results  Press the NO  function key to bypass validation and proceed to the test conclusion  stage    When validation is selected  the following prompt is displayed  showing the current settings for output zero and span  input zero  and span  and the 96 span error tolerance              Output  V   1 000 5 000  Input  mA   4 000 20 000    Span Tol   0 25  SELECT  
64. ero Menu  Pressure mode              ZERO   NEXT   DONE  No  Contrast Menu  CONTRAST   NEXT   DONE  y  Auto Off Menu  AUTO OFF   NEXT   DONE                      9Y  y  Clock Menu  CLOCK   NEXT   DONE k  y  Terminal Menu  TERMINAL   NEXT   DONE                         J  9Y  Ya  y  Home Menu                                                                      Figure 4  The Menu Tree    3  Using Measure Modes  Lower Display     3 1 Measuring volts and frequency    Electrical parameters volts and frequency can be measured using  the lower display  To make the desired measurements  follow these  steps     1  Switch to the lower display  LOWER  from Main Menu   2  Select the desired parameter for measurement   3  Connect leads as shown in Figure 5           VOLTS IN  0 000 V    9 000 V           uer         VOLTS IN                      Figure 5   Measuring Volts and Frequency with Input Output Terminals    3 2 Measuring mA  To measure mA follow these steps   1  Switch to lower display and select mA   2  Make sure the input output control is set to IN   3  Connect leads as shown in Figure 6        mA IN  25 00     0 000            5000 mA            uem                Figure 6  Measuring mA with Input Output Terminals  3 3 Measuring Temperature    3 3 1 Using Thermocouples    The calibrator supports the following thermocouple types  B  C  E  J   K  L  N  R  S  T  U       and XK  The characteristics of all the types are  described in Specifications section  The calibrator also has
65. evels     The top level of the menu is the home menu  It can be accessed  anytime by pressing the HOME key  There are three variations of the  home menu  the input home menu  the output home menu  and the  pulse home menu     In the input home menu the only active options are  MENU  and   LIGHT   The  MENU  option is used to enter the next level of the  menu bar  the main menu  Press the corresponding function key  F1   to enter the main menu  The  LIGHT  option is used to turn on the  LCD back light  Press the corresponding function key  F2  to turn on  the back light     MENU   LIGHT      In the output home menu there are three active options   MENU     LIGHT  and  STEP  or  RAMP   The first two options work the same  as in the input home menu  The third option is selectable in the Auto  Function Menu and is used to turn on and off the selected auto  function  See Section 4 2  Using the Automatic Output Functions   Also leaving this menu or pressing the Home button will stop the  auto functions                    MENU   LIGHT   STEP    The pulse home menu also has three active options   MENU     TRIG   and  COUNTS   The  TRIG  and  COUNTS  options are used                for pulse simulation  The function of these options is explained in  Section 4 2 6  Sourcing a Pulse      MENU   TRIG   COUNTS    The next level of the menu bar is the main menu  The levels under  the main menu depend on what mode the calibrator is in     The main menu has three active options  UPPER    LOW
66. fication data entry  stage     3  AREPEAT function key is available for restarting the test at the  first point     8 4 3 Identification Data Entry    At this step the technician is given the opportunity to update any of  the downloaded tag identification fields  except for the tag number  which must remain the same in order for the PC program to  synchronize the test results with its database during upload   Otherwise this process is identical to that described under New Tag  As Found identification data entry  When the SAVE function key is  pressed  the process automatically proceeds to the test conclusion  stage     8 4 4 Test Conclusion    If test validation was selected  a display showing the overall pass fail  status of the test is displayed with a prompt asking whether details  should be displayed  Press the YES function key to display the  details  or the NO function key to skip the details  In either case the  adjustment prompt is displayed next     A prompt is displayed asking whether further adjustment and testing  is required  Press the YES function key to proceed to the adjustment  menu  or the NO function key to proceed to the first level document  mode menu     8 5 As Left Test    An As Left test can immediately follow an As Found test or a previous  As Left test  or it can be initiated at a later time for any tag stored in  memory  If it immediately follows another test  no tag selection step  is required and data collection starts right away  If it is initiat
67. figure menu contains functions to enter and modify tag  configurations  to enter and modify tag groups  to set the optional  company name to be added to all uploaded calibration reports  and    to select the communications port to which the calibrator is  connected     The Help menu contains the application help  support  and program  information     9 4 File Menu    Initially CL310 SW is set to store reports in the root folder  Omega  CL310 SW  within the  My Documents  folder  If you wish to store  them into another folder  navigate to the alternate folder before  saving the first report  From then on  until a further alternate folder is  chosen  that folder will be the default for storage     Open  Open an existing calibration report file and load the contents into  the Reports list box  Use the Files of type selection in the file  open dialog to specify the format  TXT or CSV  of the files to be  displayed for opening     Save Reports as Single File    Save the tag calibration reports contained in the Reports list box  into single user named file within a user selected root directory   If the reports were uploaded from a calibrator  the file extension  is TXT or CSV depending on the report format selection on the  Upload menu at the time the reports were uploaded  If the  reports were loaded from a file  the file extension is set the same  as that of the source file     Save Reports as Separate Files    Save the tag calibration reports contained in the Reports list box  i
68. gging to the last  or  by clicking on the first and holding down the Shift key while  clicking on the last     9 7 Download Menu    Tag    Select and download a single tag to a calibrator  The tag ID  must be unique within the calibrator or the tag will not be  downloaded  Use the Configure menu to select the  communications port     Once the download begins  click the Cancel button in the  displayed status box to terminate the download     Group    Select a group from which to select multiple tags for download  to the calibrator  Highlight all of the tags in the list which are to  be downloaded  then click OK  Multiple individual tags are  selected by holding down the Ctrl key and clicking on them one  atatime  Multiple contiguous tags are selected either by  clicking on the first and dragging to the last  or by clicking on the  first and holding down the Shift key while clicking on the last   Any tag ID which is not unique within the calibrator will not be  downloaded  Use the Configure menu to select the  communications port     During download a status display shows the progress  and  provides a Cancel button for terminating the download  As the    download proceeds  a right angle bracket indicates the tag  presently being processed  and the status is displayed as  Loading  All tags not yet downloaded have status Pending  Ifa  tag can not be downloaded for some reason  an error message  is displayed explaining the problem and the tag status shows  FAILED  All successfully
69. he LEFT  ARROW key to change back to the TRIG READY prompts and repeat  the test     Once the desired number of points have been tested  press the  DONE function key to end the data collection stage and proceed to  the entry of test identification data     If for any reason the data collection is to be terminated without  saving any data  press either the HOME key or the ABORT function  key  A prompt is displayed requesting confirmation that the test data  accumulated so far is to be abandoned  Press the NO function key  to return to the data collection process at the point where it was  interrupted  Press the YES function key to abandon the data  collected        Manual Input Output Type Prompt    When the input output combination includes a manual input  manual  output  or both  a prompt similar to the following is displayed for  each test point     MANUAL IN AF 1            TONS    VOLTS OUT  1 000 V  NEXT DONE   ABORT    The current test type  AF for As Found  and point number are shown  in the upper right corner of the display  Note that the point number is  the next point to be tested and data has not been saved for it yet              If the calibrator output is pressure  adjust the external pressure  source until the displayed output value matches the desired test  value  If the calibrator output is manual  adjust the external source to    the desired test value  For all other output types  enter the output  value to be generated using the numeric keys  followed by EN
70. in the upper right corner of the display  Note that the point number is  the next point to be tested and data has not been saved for it yet     Enter the output value to be generated using the numeric keys   followed by ENTER  Press the TRIG function key to generate the  pulse train  The lower line changes to the following while the pulse  train is being generated     STOP        Press the STOP function key to abort the pulse train before it is  finished  When the pulse train is complete  or is aborted  the lower  line changes back to its original state     Once the calibrator input value has settled  and the results are  acceptable  press the READY function key to proceed to the prompts  for saving the data        If the calibrator input is manual  the lower line changes to the  following prompts              NEXT DONE   ABORT    Press the NEXT function key to enter the manual input value as  described in the subsection  Manual Data Entry Prompt  below  See    69       70    the following paragraphs for a description of the DONE and ABORT  function keys     If the calibrator is not manual  or after manual data has been entered   the lower line changes to the following prompts              SAVE DONE   ABORT    If the data displayed is correct  press the SAVE function key to save  the data and step to the next point  The point number in the upper  right increases to show that the test data is saved  If the pulse train  or manual data must be corrected before saving it  press t
71. ing from top to bottom of the list and vice versa  Use the LEFT  ARROW and RIGHT ARROW keys to scroll the tag selector highlight  one page at a time  cycling from top to bottom of the list and vice  versa     Connect a serial interface printer to the serial port of the CL310  turn  the power on and ensure that it is warmed up and ready to print     When the desired tag number is highlighted  press the SELECT  function key to print the tag  To print additional tags  highlight each  in turn and press the SELECT function key    When all tags have been printed  press the EXIT function key to  return to the review selection menu        83    8 8 Clearing Test Results    At the main menu  press the MENU function key  followed by the  MORE function key  followed by the DOCUMENT function key to  display the first level document mode menu     TEST REVIEW EXIT    Press the REVIEW function key to display the review selection menu                    VIEW PRINT   CLEAR    Press the CLEAR function key to display the tag selection prompt  and clear menu options     SELECT   CLEAR       EXIT    To erase all tags from the calibrator memory  press the CLEAR ALL  function key  Additional prompts are displayed to confirm the  request  Press the YES function key in response to every prompt to  proceed with clearing all tags  To abort the clear request  press the  NO function key in response to any prompt                 To erase a single tag from the calibrator memory  highlight the  desired tag 
72. input and output extend the versatility of the calibrator by  providing a means for the technician to key in data from auxiliary  measurement or source equipment  or from field devices that can  not be physically connected to an instrument  such as the dial  position of a panel meter  The calibrator output points are entered  one at a time during the As Found test  The point values may be  entered in any order  but this is usually done as a sequence of  ascending values  descending values  ascending followed by  descending  or vice versa  After each output is generated  the  technician waits for the input to settle  Once settled  he presses a  function key to prompt for manual entries if any  to save both the  output and input values  and finally to step to the next point  Once all  required test points have been entered  a different function key is  pressed to end the As Found test by prompting for the tag  identification data and the test validation limit  and automatically  adding the real time clock time stamp  If test validation was  selected  the technician has the option of displaying the As Found  test results     To perform a downloaded test  the technician selects the tag to be  tested from a list on the display for the As Found test  The calibrator  cycles through the test points one at a time automatically generating  the As Found  downloaded  output value in the case of a calibrator  source type  or displaying the As Found value of a manual output  type  Once the
73. isconnect it from the pressure module     Disconnect test leads before changing to another measure or  source function       When servicing the calibrator  use only specified replacement  parts   To avoid false readings  which could lead to possible electric shock    or personal injury  replace the battery as soon as the battery  indicator appears       To avoid a violent release of pressure in a pressurized system  shut  off the valve and slowly bleed off the pressure before you attach  the pressure module to the pressure line     A Caution    To avoid possible damage to calibrator or to equipment under test     e Use the proper jacks  function  and range for your measurement or  sourcing application       To avoid mechanically damaging the pressure module  never apply  more than 10 ft lb  of torque between the pressure module fittings  or  between the fittings an the body of the module       To avoid damaging the pressure module from overpressure  never  apply pressure above the rated maximum printed on the module       To avoid damaging the pressure module from corrosion  use it only  with specified materials  Refer to the pressure module  documentation for material compatibility     2  Calibrator Interface    Figure 1 shows the location of the input and output connections on  the calibrator  while Table 1 describes their use              oH         755     LE OMEGA       1                                                                 lo      i 4    gt   zmazm   amp g   
74. lections  Selecting one of these units with a high  pressure module attached will cause the display to read           Warning     To avoid a violent release of pressure in a pressurized system  shut  off the valve and slowly bleed off the pressure before you attach the  pressure module to the pressure line     15    Caution    To avoid mechanically damaging the pressure module  never apply  more than 10 ft lb  of torque between the pressure module fittings  or  between the fittings an the body of the module     To avoid damaging the pressure module from overpressure  never  apply pressure above the rated maximum printed on the module     To avoid damaging the pressure module from corrosion  use it only  with specified materials  Refer to the pressure module  documentation for material compatibility     To measure pressure  follow these steps     1  Connect the pressure module to the calibrator as shown in  Figure 9     The calibrator can measure pressure on both the upper and the  lower display  This makes it possible to measure pressure in two  different units at the same time     Switch to either upper or lower display from the Main Menu   Select  PRESSURE  from the primary parameters   Select the desired measuring unit     Zero the pressure module  The zero function on the calibrator  can be found in the pressure zeroing menu     ak                  VALVE       OE OMEGA          10             PRESSURE  MODULE                                                                
75. librated downloaded tags                       U TIC 221 5  U LI 22    NEW SELECT   EXIT             Tag numbers are displayed on multiple pages  six tags to a page   The letter U indicates an uncalibrated downloaded tag  Press the UP  ARROW and DOWN ARROW keys to scroll the tag selector highlight  one tag at a time  scrolling from page to page  cycling from top to  bottom of the list and vice versa  Use the LEFT ARROW and RIGHT  ARROW keys to scroll the tag selector highlight one page at a time   cycling from top to bottom of the list and vice versa     Press the EXIT function key to return to the main menu without  selecting a tag     When the desired tag number is highlighted  press the SELECT  function key to proceed to data collection     8 4 2 Data Collection    If pressure is selected on the upper or lower display  the pressure  zero prompt is displayed first  and operates as described under New  Tag As Found Setup     Downloaded Tag As Found data collection is identical to New Tag As  Found data collection except for three minor differences     1  The test outputs can not be changed  they will be the predefined  values that were downloaded  For pressure and manual output  types  the downloaded value is displayed for the technician to  set it on the external source  For all other output types  the  calibrator automatically generates the As Found value     2  Once the SAVE function key is pressed for the last test point  the  process automatically proceeds to the identi
76. lication     9 3 Overview  Application features   enter and save tag test configurations for repeated use    collect tag configurations into related groups for ease of  download  saving the groups for repeated use    download individual tag configurations to a calibrator   select multiple tags from a group for download to a calibrator  upload all  or selected  tag calibration reports from a calibrator  save uploaded tag calibration reports in text or spreadsheet  format   erase all  or selected  tag configurations from a calibrator   print tag calibration reports  saved tag test configurations  and  saved groups   confirm the identity of the calibrator connected to the computer  The File menu contains functions to save uploaded calibration  reports to disk using three different methods  to print calibration  reports  and to open previously saved calibration reports for viewing   Once a previously saved file containing multiple tags is opened  it    can be resaved using one of the methods where the tags are saved  in individual files     The Upload menu contains functions to list all tag configurations in  the calibrator  to upload calibration reports for all or selected tags   and to display the identity of the connected calibrator     The Erase menu contains functions to erase all or selected tag  configurations from the calibrator     The Download menu contains functions to download individual tag  configurations  or selected tags from a group  to the calibrator     The Con
77. mmand  For example     TRIG   WAI    Triggers the pulse train  Once the pulse train has been triggered   the calibrator can proceed to the next command     Sequential Commands    Commands that are executed immediately after the are entered   This type includes most of the commands     7 3 2 Character Processing  The data entered into the calibrator is processed as follows       ASCII characters are discarded if their decimal equivalent is less  than 32  space   except 10  LF  and 13  CR        Data is taken as 7 bit ASCII    The most significant data bit is ignored     Upper or lower case is acceptable     7 3 3 Response Data Types  The data returned by the calibrator can be divided into four types   Integer    For most computers and controllers they are decimal numbers  ranging from  32768 to 32768  For example      ESE 140   ESE  returns 140    Floating    Numbers that have up to 15 significant figures and exponents   For example     CPRT_COEFA  returns 3 908000E 03    Character Response Data  CRD   Data returned as keywords  For example   RTD_TYPE  returns PT385_10    Indefinite ASCII  IAD   Any ASCII characters followed by a terminator  For example    IDN  returns MARTEL  ASC300  250  1 00    7 3 4 Calibrator Status    Status registers  enable registers  and queues provide status  information on the calibrator  Each status register and queue has a  summary bit in the Serial Poll Status Byte  Enable registers generate  summary bits in the Serial Poll Status Byte  The follo
78. n complete  the group is saved into  a user named file with the fixed extension UGP     Edit   Edit an existing group of tags  After a group file is selected  a  configuration screen is displayed for modifying the contents of  the group  This dialog screen is described in detail in section  9 10 below  All tags which are members of the group must  reside in the same directory as the group  When complete  the  group is saved into a user named file with the fixed extension  UGP This may be the original file name or a new one     Display Print  Display the members of a group of tags  and optionally print it     Company Name    Display a dialog for entering and changing the company name  which is inserted at the top of each tag report  The maximum  length is 50 characters  If nothing is entered for this item   nothing is inserted at the top of the tag reports     COM Port    The communications port is selected by clicking on one of the  following menu items  The selected port is indicated by the  menu item with the checkmark  and is also displayed on the  status line at the bottom of the screen     Use COM1  Use COM2  Use COM3  Use COM4  Use COM5  Use COM6  Use COM7  Use COM8    9 9 Tag Configuration Details    Tag Identification Tab    The Tag ID field is required and must not contain a space as its  first character  The remaining fields are optional  All fields are a  maximum of 16 characters in length     Tag Instrument Input Tab    The Type field contains a drop down list pre
79. n the field           1 000 V    VOLTS IN   aN SS  VOLTS OUT   UPPER   LOWER       DONE       Press the UPPER or LOWER function key to display the respective    type and parameter selection menu     63    64    VOLTS IN              10 999 V    VOLTS OUT    1 000 V  SELECT   NEXT DONE    These two menus operate the same as the normal upper and lower  type and parameter selection menus described earlier in this manual   except that they have two extra type selections  MANUAL IN and  MANUAL OUT  The currently selected parameter field blinks   indicated by bold italicized text above   Press the SELECT function  key to cycle through the possible selections for that field  Press the  NEXT function key to move to the next parameter field  When all  parameters are set for the selected display  press the DONE function  key to return to the As Found setup menu              One display must be selected as a calibrator input type  and the  other must be selected as a calibrator output type  Pressure can not  be selected simultaneously on both displays as the calibrator is only  capable of handling one pressure at a time  Pressure takes its input  or output orientation as the opposite of the orientation selected on  the other display     Once both the upper and lower displays are properly selected  press  the DONE function key on the As Found setup menu to proceed with  the setup     At this point some display type selections  as described below   require optional setup data to be entere
80. nd  Lower Displays       Y       Collect As Found Test  Data       Y       Enter Tag Identification  Data       Y       Optionally  Enter Tag  Validation Data  Validate  Tag  and Display Results       v       Select Tag for Testing       v       Data    Collect As Found Test       y       Optionally  Modify Some  Tag Identification Data                   Optionally  Display  Validation Results                M    v       As Left Test Required        No    Yes    v       Y       Select Tag for Testing          Collect As Left Test Data       Y       Optionally  Display  Validation Results       Y       As Left Test Complete   adjustment required           Y Yes                Adjustment Required        Yes    Y          62    Adjust Tag             No    Adjust Tag                      No       8 2 New Tag As Found Test    8 2 1 Setup    At the main menu  press the MENU function key  followed by the  MORE function key  followed by the DOCUMENT function key to  display the first level document mode menu           TEST    REVIEW       EXIT       Press the TEST function key to display the test selection menu           AS FOUND   AS LEFT   EXIT          Press the AS FOUND function key to display the As Found Tag  Selection menu which shows all uncalibrated downloaded tags           U TIC 221 5  U LI 22  NEW SELECT       EXIT       The letter U indicates an uncalibrated downloaded tag  Press the  NEW function key to display the As Found Tag setup menu for    configuring a new tag i
81. ng  conditions outside of OMEGA s control  Components in which wear is not warranted  include but are not  limited to contact points  fuses  and triacs     OMEGA is pleased to offer suggestions on the use of its various products  However   OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for  any damages that result from the use of its products in accordance with information provided  by OMEGA  either verbal or written  OMEGA warrants only that the parts manufactured by the  company will be as specified and free of defects  OMEGA MAKES NO OTHER WARRANTIES OR  REPRESENTATIONS OF ANY KIND WHATSOEVER  EXPRESSED OR IMPLIED  EXCEPT THAT OF  TITLE  AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY  AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED  LIMITATION OF  LIABILITY  The remedies of purchaser set forth herein are exclusive  and the total liability of  OMEGA with respect to this order  whether based on contract  warranty  negligence   indemnification  strict liability or otherwise  shall not exceed the purchase price of the  component upon which liability is based  In no event shall OMEGA be liable for  consequential  incidental or special damages     CONDITIONS  Equipment sold by OMEGA is not intended to be used  nor shall it be used   1  as a  Basic  Component  under 10 CFR 21  NRC   used in or with any nuclear installation or activity  or  2  in medical  applications or used on humans  Should any Product s  
82. nit is attached  For example     PRES  Will return MARTEL 001PNS 3 0    PRES UNIT     Returns the pressure units of both the upper and the lower display   For example if the unit on the upper display is bars  and on the  lower it is psi  the command will return     BAR  PSI    PULSE CNT    Sets the number of pulses the calibrator will produce when it is  triggered while in pulse mode  For example     PULSE        3000 will set the number of pulses to 3000     PULSE CNT     Returns the number of pulses in the pulse train  Using the above  example  the returned value would be     3000    REMOTE    Puts the calibrator in remote mode  From the remote mode the user  can still use the keypad to get back to local unless the command  LOCKOUT was entered before REMOTE  Than the keypad is totally  locked out  and the user has to send the LOCAL command to get  back to local operation     RTD TYPE    Sets the RTD type  The following is a list of RTD types they way they  should be entered after the command     PT385 10        85 50  PT385 100        85 200   PT385 500  PT385 1000   PT392 100  PTJIS 100   Ni120  Cu10  Cu50  Cu100  YSI 400   OHMS  CUSTOM    For example            TYPE PT385_10 sets RTD type to Pt385 10    RTD_TYPE   Returns the RTD type     RTD_WIRE   Sets the number of wires used for connection in measuring RTDs   The calibrator measures RTDs using 2   3   and 4 wire connections   After the command  enter 2W for 2  wire  3W for 3 wire  and 4W for 4   wire  For example     
83. nto separate files within a user selected root directory  The files  are named using the tag identifier from each report  plus the  current computer date and time  If the reports were uploaded  from a calibrator  the file extension is TXT or CSV depending on  the report format selection on the Upload menu at the time the  reports were uploaded  If the reports were loaded from a file  the  file extension is set the same as that of the source file     Save Reports in Separate Tag Directories    Save the tag calibration reports contained in the Reports list box  into separate files in separate subdirectories within a user  selected root directory  The files are named using the tag  identifier from each report plus the current computer date and  time  The subdirectories are named using the tag identifier from  each report  If the tag subdirectory already exists within the root    87    directory  the file is saved there  If the tag subdirectory does not  exist within the root directory  the tag subdirectory is created and  the file is saved there  If the reports were uploaded from a  calibrator  the file extension is TXT or CSV depending on the  report format selection on the Upload menu at the time the  reports were uploaded  If the reports were loaded from a file  the  file extension is set the same as that of the source file     Print  Print the tag calibration reports contained in the Reports list box     Print Setup  Select the font size used for all printing functions     E
84. nvalid        STATE was received   108 An invalid TSENS TYPE was received   109 Pressure module not connected       110    An unknown command was received       111    An invalid RTD or TC parameter value was received       112    The serial input buffer overflowed       113    Too many entries in the command line       114    The serial output buffer overflowed       115    Output is overloaded       116    Calibrator not in pulse train mode when TRIG was received          117    An invalid FREQ TYPE was received          7 5 Entering Commands    Commands for the calibrator may be entered in upper or lower case   There is at least one space required between the command and  parameter  all other spaces are optional  Almost all commands for  the calibrator are sequential  any overlapped commands will be  indicated as such  This section will briefly explain each of the  commands and describe their general use  which will include any  parameters that may be entered with the command as well as what  the output of the command is     7 5 1 Common Commands   CLS    Clears the ESR  the error queue and the RQS bit  Also terminates all  pending operations  When writing programs  use before each  procedure to avoid buffer overflow      ESE    Loads a byte into the Event Status Enable register  The command is  entered with a decimal number that  when converted to binary   enables the right bits in the Event Status Register  For example      ESE 133    When 133 is converted to binary it i
85. ockout  type in REMOTE and LOCKOUT  30 in either order     3  To switch back to local operation enter LOCAL at the terminal  This  command also turns off LOCKOUT if it was on  For more  information on commands refer to the Remote Commands  section     7 3 Using Commands    7 3 1 Command types    Refer to the Section on Remote Commands for all available  commands     The calibrator may be controlled using commands and queries  All  commands may be entered using upper or lower case  The  commands are divided into the following categories     Calibrator Commands  Only the calibrator uses these commands  For example  LOWER MEAS DCV  tells the calibrator to measure voltage on the lower display     Common Commands    Standard commands used by most devices  These commands  always begin with an      For example     IDN   tells the calibrator to return its identification     Query Commands    Commands that ask for information  They always end with a       For example     FUNC   Returns the current modes of the upper and lower displays     Compound Commands    Commands that contain more than one command on one line   For example     LOWER MEAS RTD  RTD TYPE CU10    Sets the calibrator to measure RTD in the lower display and sets  RTD type to Cu 10     31    32    Overlapped Commands    Commands that require more time to execute than normal  The  command  WAI can be used after the overlapped command to  tell the calibrator to wait until the command finishes before  executing the next co
86. on standard cold junction option  enter the text string for it as  documented in the user manual addendum accompanying your  calibrator     The Amplitude field appears only for the Frequency and Pulse  types  Enter the peak to peak amplitude to be generated     The Fixed Type and Fixed Value fields appear only for the Pulse  type  Select counts or frequency as the fixed type and enter the  value to be used     Tag Instrument Output Tab    The Type field contains a drop down list presenting all of the  available options for a standard calibrator  Select the type  corresponding to the instrument to be calibrated  If your  calibrator contains special software implementing a  non standard type  enter the text string for it as documented in  the user manual addendum accompanying your calibrator     The Units field contains a drop down list presenting all of the  standard units for the type selected above  Select the units  corresponding to the instrument to be calibrated  The Manual  type always requires user defined units which may be up to 5  characters in length  If your calibrator contains special software  implementing a non standard unit  enter the text string for it as  documented in the user manual addendum accompanying your  calibrator     The Curve field appears only for the Thermocouple and RTD  types  It contains a drop down list presenting all of the available  options for a standard calibrator  Select the curve corresponding  to the instrument to be calibrated  If yo
87. ouple    RTD  FREQ  frequency   PULSE  PRESSURE   mA  and mA 2W SIM        Input Output control    Switches the lower display between input mode   read   and output mode  source         Additional Settings    Available only for TC  thermocouple   and RTD  measurements  For TC this setting turns the  CJC  Cold Junction Connection  on and off  For  RTD measure  RTD IN   this setting sets the  number of wires used in the measurement  2   wire  3 wire  or 4 wire           Span Indicator          Available only for mA and mA LOOP Shows  where in the preset span the measured value  falls  Fixed for mA at 4  0   and 20  100             5 Units Shows what unit the measurement or source  value is in  Available options are for RTD and  TC    C or   F   and for FREQ and PULSE  CPM   Hz  or KHz        6 Sensor Types Allow for measurements to be made for different  types of RTDs and TCs  All types are shown in  the Specifications  Also  displays the amplitude  of the pulse and frequency source  and pres   sure units           7 Numeric Displays Display the numeric values of the signal being  measured  or sourced  An  OL  reading indi   cates an out of range or overload condition              2 2 Menu Bar    The parameters on the display are controlled by the menu bar  which  is located at the bottom of the LCD  The function keys  F1  F2  and  F3  are used to navigate through all the levels and choices of the  menu bar  Refer to the menu tree for a clarification on the layout of all  the l
88. peration  Complete commands    ESE Loads a byte into the Event Status Enable register     ESE  Returns the contents of the Event Status Enable register     ESR  Returns the contents of the Event Status register and clears the  register     IDN  Identification query  Returns the manufacturer  model number   and firmware revision level of the Calibrator     OPC Enables setting of bit O  OPC for  Operation Complete   in the  Event Status Register to 1 when all pending device operations  are complete     OPC  Returns a 1 after all pending operations are complete  This       command causes program execution to pause until all opera   tions are complete        35        RST    Resets the state of the instrument to the power up state  This  command holds off execution of subsequent commands until it  is complete                                                                                                   SRE Loads a byte into the Service Request Enable register    SRE  Returns the byte from the Service Request Enable register    STB  Returns the status byte    WAI Prevents further remote commands from being executed until  all previous remote commands have been executed   Table 6  Calibrator Commands  Command Description  CAL START Puts the calibrator in calibration mode  CJC STATE Turns CJC on or off   CJC STATE  Returns the state of the CJC  CPRT COEFA Sets the custom RTD coefficient A  CPRT COEFA  Returns the custom RTD coefficient A  CPRT COEFB Sets the custom RTD coefficien
89. perature compensation       Five setpoints in each range for increasing decreasing output      An interactive menu     Complete RS232 interface for remote control       Isolated read back for transmitter calibration    e Documenting capability for up to 50 tags     1 1 Customer Service    OMEGA Engineering    Tel   203  359 1660  Fax   203  359 7900    www omega com  email  info omega com    1 2 Standard Equipment    Check to see if your calibrator is complete  It should include   CL310 Calibrator  Instruction Manual  Test Leads  Rubber Boot   RS 232 cable  USB adapter cable  Calibration Certificate  software CD     1 3 Safety information    Symbols Used    The following table lists the International Electrical Symbols  Some or  all of these symbols may be used on the instrument or in this  manual     Symbol Description       AC  Alternating Current     AC DC    Hee    Battery    CE Complies with European Union Directives           MA    DC    Double Insulated   Electric Shock   Fuse   PE Ground   Hot Surface  Burn Hazard    Read the User s Manual  Important Information   Off    On     O gt  i enl    The following definitions apply to the terms    Warning    and    Caution        e    Warning    identifies conditions and actions that may pose hazards  to the user     e    Caution    identifies conditions and actions that may damage the  instrument being used     Use the calibrator only as specified in this manual  otherwise injury  and damage to the calibrator may occur     A
90. pressure module attached will cause the display to read            Using the Upper and the Lower Display for  Calibration and Testing    6 1 Testing an Input or Indicating Device    To test and calibrate actuators  recording  and indicating devices  using the source functions  follow these steps     1  Select the lower display and choose the correct primary  parameter     2  Switch to  OUT  in the input output control     3  Connect the leads to the instrument and the calibrator as shown  in Figure 19     1 mA max  current       qu                      Figure 19  Connections for Testing an Output Device    6 2 Calibrating an I P Device    The following steps show how to calibrate a device that controls   pressure    1  Select upper display from the Main Menu  and select pressure  from the primary parameters    2  Switch to lower display from the Main Menu  and select current  source  mA out  from the primary parameters     3  Connect the calibrator to the device as shown in Figure 20  The  calibrator will simulate the transmitter current and measure the  output pressure     4  Enter a current using the keypad        M            OMEGA          1                                                                                                                                           ois        1531  eee     TYN  El Pv  C ll       e el kl fe  OO   mazm  g                                                                                                                         i f  
91. pt similar to the following is displayed for entry of manual  inputs and outputs  When both input and output are manual  the  output prompt is displayed first        AF 1    Enter manual input  for Upper Display               TONS    DONE   BACK      Enter the manual value using the numeric keys  followed by ENTER   If the value is correct  press the DONE function key to continue  The  BACK function key may be pressed to return to the previous prompt  display  71             72    8 2 3 Identification Data Entry    When the DONE function key is pressed at the data collection stage   the first page of the following two page identification data display is  shown  Initially all data fields are blank  but after data is entered the  current value of each field is shown to the right of its description        TAG   MFG  MDL   SER     SELECT   NEXT SAVE          TECH   TEMP  HUMD  OTHR    SELECT  NEXT SAVE    These two pages provide space to enter eight phrases of up to 16  characters each to describe the test that has been performed  The  first seven phrases are predefined to be Tag Number  Instrument  Manufacturer  Model Number  Instrument Serial Number  Technician  Name or ID number  Temperature  and Humidity  The last phase   Other is for client use  There is no restriction on the data that may be  entered except for Tag Number which must have a non blank first  character and must be unique among all tags currently stored in the  calibrator              Press the NEXT function ke
92. rator cycles through the test points  one at a time automatically generating the As Found output value in  the case of a calibrator source type  or displaying the As Found  value of a manual output type  Once the input has settled  the  technician presses a function key to prompt for manual input entry if  any  to save both the As Left output and input values  and finally to  step to the next point  Once the last point has been tested  the As  Left test automatically ends and  if test validation was selected  the  technician has the option of displaying the As Left test results   Finally  the technician has the option of ending the tag calibration  of  entering an adjustment mode to monitor physical adjustments to the  tag prior to starting a further As Left test  or of directly entering a  further As Left test     Additional As Left tests may be performed after tag calibration has  ended as long as the original tag data has not been cleared from  storage  A menu item is available for selecting any tag and starting  another As Left test    Review functions are available to view test results on the calibrator  display    Individual or all tag test results may be cleared from storage using  menu functions     Document Menus             Function Selection Menu          TEST   REVIEW   EXIT          Review Menu          Test Selection Menu          AS FOUND   AS LEFT   EXIT       p          Menu             As Left Menu             As Found Tag Selection Menu          NEW   SELEC
93. s 10000101  Bits 7  2  and 0  will be enabled     39    40     ESE     Returns the contents of the Event Status Enable register  The value  returned is a decimal  For example  if the register has the following  settings     10000101 than the value returned will be 133      ESR     Returns the contents of the Event Status Register in decimal form   For example     If the ESR contains 10111001   ESR  will return 185      IDN   Returns the manufacturer  model number  and firmware revision of  the Calibrator  For example    IDN  will return OMEGA CL310 0 2 00   OPC    Enables the Operation Complete setting in the ESR  This setting  makes it possible to check if an operations is complete after it has  been initialized     For example this operation could be used with the command TRIG      OPC   Returns 1 when all operations are complete  and causes program    execution to pause until all the operations are complete  For  example     TRIG    OPC  will return a 1 when the pulse train initiated by  TRIG is complete      RST    Resets the state of calibrator to the power up state  All subsequent  commands are held off until the execution of the command is  complete      SRE    Loads a byte into the Service Request Enable register  A decimal  number must be entered  which when converted to binary   corresponds to the correct settings     For example      SRE 8 enters the binary number 00001000 to the SRE  This  enables bit 3  Bit 6 is not used      SRE     Returns a byte from the SRE  
94. se from the primary  parameters     Choose the desired unit and enter the frequency using the  keypad    Select the  COUNTS  function from the home menu to enter the  number of pulses  Use  TRIG  to start and stop the signal     The amplitude of the pulse can be adjusted in the same manner  as for frequency           mA LOOP  25 00  mA IN  25 00   0 000      0 000 mA  PULSE 5 0 Vpp FREQ OUT 5 0 Vpp    9 0 kHz 4 0 kHz  MENU   TRIG   COUNTS MENU   LIGHT                            TC PLUG                                                                     SURE   SOURCE MEASURE  A Y Q d v  Y                            Note       wire used must match  the thermocouple type being calibrated     Figure 13  Connections for Outputting Thermocouples    4 8 Sourcing Thermocouples  To source a thermocouple use the following steps     1     Connect the thermocouple leads to the appropriate polarized TC  miniplug  and insert the plug into the TC terminals on the  calibrator  as shown in Figure 13     Select lower display from the Main Menu  and choose  thermocouple  TC  from the primary parameters     Choose output  OUT  from the input output control   21    4  Select the desired thermocouple type from the sensor types   5  Enter the temperature using the keypad     mA IN  25 00         U 000         TC OUT CJC OFF    100 0            ight T _                                     MEASURE SOURCE MEASURE          Figure 14  Connections for Outputting RTDs    4 9 Sourcing Ohms RTDs  To source 
95. senting all of the  available options for a standard calibrator  Select the type  corresponding to the instrument to be calibrated  If your  calibrator contains special software implementing a nonstandard  type  enter the text string for it as documented in the user  manual addendum accompanying your calibrator     The Units field contains a drop down list presenting all of the  standard units for the type selected above  Select the units  corresponding to the instrument to be calibrated  The Manual  type always requires user defined units which may be up to 5  characters in length  If your calibrator contains special software  implementing a nonstandard unit  enter the text string for it as  documented in the user manual addendum accompanying your  calibrator     The Curve field appears only for the Thermocouple and RTD  types  It contains a drop down list presenting all of the available  options for a standard calibrator  Select the curve corresponding  to the instrument to be calibrated  If your calibrator contains  special software implementing a nonstandard curve  enter the  text string for it as documented in the user manual addendum    accompanying your calibrator        94    The CJ field appears only for the Thermocouple type  It contains  a drop down list presenting all of the available options for a  standard calibrator  Select the cold junction option  corresponding to the instrument to be calibrated  If your  calibrator contains special software implementing a  n
96. sful   otherwise an error message enclosed in angle brackets     TAG DNLD    See the separate Tag Download and Upload Procedures manual for  detailed instructions on using this command to download tag  calibration test configuration data from a PC to the calibrator     TAG UPLD    See the separate Tag Download and Upload Procedures manual for  detailed instructions on using this command to upload tag calibration  test results to a PC from the calibrator     TAGS     Sends a list of all tag ID s in the calibrator to a PC  Returns a list of  tag ID s  one per line  followed by    Complete   if successful   otherwise an error message enclosed in angle brackets  Each line  contains three data items separated from each other by spaces   First  the position of the tag in the calibrator  an integer from 1 to 50     Second  the status of the tag  a single character  U  for downloaded  uncalibrated or  C  for calibrated  Third  the tag number string of up  to 16 characters     TC TYPE    Sets the type of the thermocouple  All available types are shown in  the TC Types table in Section 8   Specifications   For example     TC TYPE B sets thermocouple type to B    TC TYPE   Returns the type of thermocouple the calibrator is set to     TEMP UNIT    Sets the temperature unit for sourcing and measuring RTD and TC   Add CEL after the command for Celsius  and FAR for Fahrenheit  For  example     TEMP UNIT CEL sets the temperature to be measured or  sourced to Celsius     TEMP UNIT     Returns
97. so coefficient C should be set  to 0  The RO is the resistance of the probe at 0  C  The  coefficients for PT385  PT3926  and PT3616 are shown in Table 4                      below   Table 4  RTD Coefficients   RTD Range C  RO Coefficient A Coefficient B Coefficient C  PT385  260   0 100 3 9083x10 3  5 775  10 7  4 183x10 12  PT385 0   630 100 3 9083x10 3  5 775  10 7      PT3926 Below 0 100 3 9848x10 3  5 87x10 7  4x10 12  PT3926  Above0 100 3 9848x10 3  5 87  10 7      PT3916 Below 0 100  3 9692x10 3  5 8495x10 7  4 2325x10 12  PT3916           0 100  3 9692x10 3   5 8495x10 7        23    5  Using Isolated Measure Modes  Upper Display     5 1 Measuring volts and mA    Use the following steps to measure the voltage or current output of a  transmitter     1  Select the upper display from the Main Menu     2  Select the desired primary parameter to be measured  Connect  the leads to the isolated inputs of the calibrator  as in Figure 16              Figure 16  Isolated Input Connection    5 2 Measuring current with loop power    To test a 2 wire  loop powered transmitter that is disconnected from  wiring  use the loop power function  This function activates a 24V  supply in series with the current measuring circuit  To use this option  proceed as follows     1  Select  mA LOOP  as primary parameter in the upper display     2  Connect the calibrator to transmitter current loop terminals  as  shown in Figure 17                                                                    
98. ssure on Upper or Lower Display    If pressure is selected on the upper or lower display  a pressure zero  prompt is displayed  Note that this prompt is displayed again before  each adjustment or subsequent test  If the installed module is not an  absolute pressure module  the following prompt is displayed     0  54 in H20     20  C  Zero pressure reading    ZERO     DONE    The value displayed on the top line is the current pressure reading  adjusted with the current zero value  Press the ZERO function key to  zero the pressure  or if satisfied with the current zero press the  DONE function key to bypass changing it  When satisfied with the  pressure zero  press the DONE function key to continue                 If the installed module is an absolute pressure module  the following  prompt is displayed     65    66          54 in H20     20  C   Enter reference  pressure              DONE    The value displayed on the top line is the current pressure reading  adjusted with the current reference pressure  Enter the new reference  pressure using the numeric keys followed by the ENTER key  or if  satisfied with the current reference value press the DONE function  key to bypass changing it  When satisfied with the pressure  reference  press the DONE function key to continue              Optional Setup   Frequency Output on Lower Display    If frequency output is selected on the lower display  a prompt for  peak to peak voltage is displayed        Enter peak to peak    voltage  1 0
99. t B  CPRT COEFB  Returns the custom RTD coefficient B  CPRT COEFC Sets the custom RTD coefficient C  CPRT COEFC  Returns the custom RTD coefficient C  CPRT MIN T Sets the custom RTD minimum temperature  CPRT MIN T  Returns the custom RTD minimum temperature  CPRT MAX T Sets the custom RTD maximum temperature  CPRT MAX T  Returns the custom RTD maximum temperature  CPRT RO Sets the custom RTD RO resistance  CPRT RO  Returns the custom RTD RO resistance  FAULT  Returns the error code of an error that has occurred  FREQ LEVEL Sets the frequency and pulse amplitude  FREQ LEVEL  Returns the frequency and pulse amplitude  FREQ TYPE Set the frequency output to continuous  frequency  or  pulse   FREQ TYPE  Returns frequency output type  continuous or pulse  FREQ UNIT Sets the unit for frequency and pulse  FREQ UNIT  Returns the unit for frequency and pulse  FUNC  Returns the current mode of the upper and lower dis   play  GET CLOCK Returns the calibrator date and time setting  GET SN Returns the calibrator serial number    36             Command    Description                                                                                                 LOCAL Returns user to manual operation of the calibrator   LOCKOUT Locks out the keypad of the calibrator  and allows for  remote operation only   LOWER MEAS Sets the mode for measuring on the lower display    L PRES UNIT Sets the pressure unit on the lower display   OUT Sets the output of the calibrator   OUT  Returns the output of 
100. t using the keypad    1000 ohms max              Figure 10  Connections for Sourcing Current       mA LOOP  25 00      ___0 000 mA    mA OUT  9 000 mA  MENU   LIGHT    4 3 1 HART   Resistor Selection    The CL310 can be set up so that the 250 ohm resistor required for  HART   configuration devices resides inside the CL310  Enabling  the CL310 s internal 250 ohm resistor eliminates the need to  manually add a series resistor during a HART   calibration process     NOTE  When the CL310 s internal 250 resistor is enabled  maximum  load driving capability drops from 1000 ohms   20mA to 750 ohms   20mA              Enable Disable Procedure    1  Remove the battery cover and remove the 2 screws that are  at the top of the case     2  Remove the 2 screws on the bottom or lower portion of the  case     3  Gently remove the top half of the case from the bottom   4  Figure 10a  shows the location of the HART    jumpers                 HART  Resistor                                                                                              D  oo 4B    circuit board                      Figure 10a     4 4 Simulating a Transmitter    To have the calibrator supply a variable test current to a loop in place  of a transmitter  follow these steps     1  Select lower display from the Main Menu     2  Choose mA simulation from the primary parameters  mA 2W  SIM   and enter the desired current     3  Connect the 24V loop as shown in Figure 11        LOOP  POWER  SUPPLY    30 VDC  MAX          
101. the calibrator   PRES  Returns the model and serial number of the attached  pressure module   PRES UNIT  Returns the pressure unit for the upper and lower dis   play   PULSE CNT Sets the number of pulses for the pulse train   PULSE CNT  Returns the number of pulses in the pulse train   REMOTE Puts the calibrator in remote mode   RTD TYPE Sets the RTD type   RTD TYPE  Returns the RTD type   RTD WIRE Sets the number of wires used by the RTD mode    RTD WIRE  Returns the wire number setting used in the RTD mode   SET CLOCK Set calibrator clock   SIM Sets the output for mA simulation   SIM  Returns the output of the mA simulation   TAG CLEAR Clear test data for a specified tag   TAG CLEAR ALL Clear test data for all tags   TAG DNLD Download tag calibration test configuration data from a  PC to the calibrator   TAG UPLD Upload tag calibration test results to a PC from the cali   brator   TAGS  Upload a list of all tag iD s to PC   TC TYPE Sets the thermocouple type   TC TYPE  Returns the thermocouple type   TEMP UNIT Sets input output temperature unit for RTD and TC   TEMP UNIT  Returns the temperature unit for RTD and TC   TRIG Starts and stops the pulse train in pulse mode   TRIG  Returns TRIGGERED when a pulse train is on  Returns  UNTRIGGERED when the pulse train is off    TSENS TYPE Sets temperature sensor type    TSENS TYPE  Returns temperature sensor type          37                                                                                        Command Description  
102. u  or the NO function key to proceed to the first level document  mode menu     8 6 Viewing Test Results    At the main menu  press the MENU function key  followed by the  MORE function key  followed by the DOCUMENT function key to  display the first level document mode menu        TEST REVIEW EXIT       Press the REVIEW function key to display the review selection menu        VIEW PRINT   CLEAR    Press the VIEW function key to display the tag selection prompt              81    82       FT 384    U TIC 221 5  U LI 22  PI 884          SELECT     EXIT    The letter U indicates an uncalibrated downloaded tag  the  remainder have been calibrated  Tag numbers are displayed on  multiple pages  six tags to a page  Press the UP ARROW and  DOWN ARROW keys to scroll the tag selector highlight one tag at a  time  scrolling from page to page  cycling from top to bottom of the  list and vice versa  Use the LEFT ARROW and RIGHT ARROW keys  to scroll the tag selector highlight one page at a time  cycling from  top to bottom of the list and vice versa     Press the EXIT function key to return to the review selection menu  without selecting a tag   When the desired tag number is highlighted  press the SELECT    function key to display the first page of data showing the tag number   calibration date and time  and the first few identification fields     TAG  FT 384    2006 03 28 15 21 14  MFG MARTEL   MDL  F100   SER  123456    DONE     EXIT    Press the UP ARROW and DOWN ARROW keys to scroll
103. ur calibrator contains  special software implementing a non standard curve  enter the  text string for it as documented in the user manual addendum  accompanying your calibrator     The CJ field appears only for the Thermocouple type  It contains  a drop down list presenting all of the available options for a  standard calibrator  Select the cold junction option  corresponding to the instrument to be calibrated  If your  calibrator contains special software implementing a  non standard cold junction option  enter the text string for it as  documented in the user manual addendum accompanying your  calibrator     The Wires field appears only for the RTD type  It contains a drop  down list presenting all of the available options for a standard  calibrator  Select the connection option corresponding to the  instrument to be calibrated  If your calibrator contains special  software implementing a non standard connection option  enter  the text string for it as documented in the user manual  addendum accompanying your calibrator     Tag Tolerance Tab    Tolerance testing is based on linear percent of range error using  the parameters provided  If enabled  at the end of a calibration  each test point pair is assigned a pass or fail status based on the  actual percent of span error calculated using the low and high  values provided for instrument input and output     When tolerance testing is disabled  none of the parameters need  to be supplied     When tolerance testing is enabled 
104. wing is a list of  registers and queues along with their function     Serial Poll Status Byte  STB   The STB is sent when the calibrator responds to the  STB     command  Figure 24 demonstrates how it functions  Cleared  when power is reset     Service Request Enable Register  SRE     Enables or disables the bits of the STB  Cleared when power is reset   Setting bits to O disables them in the STB  Setting the bits to 1  enables them  Bit assignments for the SRE and the STB are shown  below           MSS   ESB 0 EAV 0 0 0                                  MSS    Master Summary Status  Set to 1 when ESB or EAV are 1   enabled   Read using the  STB  command     ESB  Set to 1 when at least one bit in ESR is 1     EAV    Error Available  An error has been entered into the error  queue  and may be read using the Fault  command     33    34    Event Status Register  ESR     A two byte register  in which the lower bits represent conditions of  the Calibrator  Cleared when read and when power is reset     Event Status Enable Register  ESE     Enables and disables bits in the ESR  Setting a bit to 1 enables the  corresponding bit in the ESR  and setting it to O disables the  corresponding bit  Cleared at power reset  Bit assignments for the  ESR and the ESE respectively are shown below                                               15 14 13 12 11 10 9 8   0 0 0 0 0 0 0 0   7 6 5 4 3 2 1 0  PON 0          EXE   DDE   QYE 0 OPC       PON    Power On  Set to 1 if power was turned on and off
105. xit  Exit the application    9 5 Upload Menu    Tag ID s  Upload a list of all tag configurations in the calibrator  and  display it in the Tag ID s list box  Each line contains the physical  position of the tag in the calibrator  a C if the tag has been  calibrated or a U if it has not been calibrated  and the tag ID   Both list boxes are cleared before the upload  Use the Configure  menu to select the communications port     All Tag Reports    Upload all tag calibration reports from the calibrator into the  Reports list box  The list box is cleared before the upload  Use  the Configure menu to select the communications port     Only reports for calibrated tags may be uploaded  Downloaded  tags which have not yet been calibrated are skipped after a  warning message is displayed     When multiple tags are uploaded into the Reports list box they  are separated from each other by a line of dashes  If the tags  are saved into separate files  the dashes are removed     The reports may be uploaded in text format suitable for direct  printing and use with word processors  or in CSV  comma  separated variables  format suitable for importing into  spreadsheets  The format is selected below     Your company name can be inserted at the top of each  uploaded report  Use the Configure menu to enter a name to be  inserted into all calibration reports  Nothing is inserted if this  configuration item is left blank     Selected Tag Reports    Upload selected tag calibration reports from the 
106. y to switch between the two pages     Press the SAVE function key to save the data and proceed to the  results validation step     Press the UP ARROW and DOWN ARROW keys to move the field  selector highlight from one field to the next  To enter data into the  highlighted field  press the SELECT function key and a prompt  similar to the following is displayed showing the field name and  current value on the top line     TAG  FT 384             8DEF 9GHI  4JKL 5MNO 6PQR  1STU 2VWX qvz     02970        DONE NEXT SAVE    Up to 16 characters may be entered for the field using a telephone  style entry method  The four lower lines provide a mapping of the  allowed alphanumeric characters to the numeric keypad keys  To  enter a character press the corresponding numeric key multiple  times to cycle through the mapped characters until the desired  character is displayed  For example  to enter E  press the 8 key 3  times  If the next character is on a different numeric key  simply  press that key to advance the cursor and display the first mapped  character  If the next character is on the same key  press ENTER to  advance the cursor before pressing the same key  To insert a space   press ENTER twice  To erase characters press CE     Press the DONE function key to return to the two page identification  data display    Press the NEXT function key to display and edit the next data field in  sequence     Press the SAVE function key to save the data and proceed to the  results validation st
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
GOCLEVER QUANTUM 785 8GB Black    移動科 - 大竹市役所  Sony Cyber-shot DSC-W35 User's Manual  Casinha Pit Stop    Copyright © All rights reserved. 
   Failed to retrieve file