1.    TE Heartsine      samaritan  PAD    USERS MANUAL          H017 001 030 5          TABLE OF CONTENT    Available PAD Configuration  CPR Guideline Comparsion    aA    Introduction  Sudden Cardiac Arrest  Heart Rhythm  Detecting Fibrillation  Intended Users  Training    Your Samaritan   PAD  Unpacking Your PAD   PAD Features   Please Note    OOOO a oO O oO oA Ol    Preparing for Use  Installation   Where to Store Your PAD  Self Test   Replace the Pad Pak  Maintenance   Status Indicator   Soft Carry Case  Cleaning    O O O CO CO CO Y N N    How to Use the PAD  Pre Defibrillation Action  CPR Fundamentals  CPR Guidelines    ek ee A a  2000    Icons used in this manual  Performing Defibrillation 12    Step by Step Introduction 12 15 ES NOTE  After Using the PAD 16    Troubleshooting 16 A WARNING    Warnings 17  Incident Notification 17 e    VOICE PROMPT    Glossary 18    Abbreviations 18  Product List 19  gt  TO DO             www heartsine com 3       AVAILABLE SAMARITAN   PAD CONFIGURATION    HeartSine Technologies provides you with a fully configureable samaritan   PAD system to allow you to comply with your chosen SCA treatment  protocol  Our current devices can be configured to be compliant with the 2000 or 2005 version of the AHA ERC guidelines on Cardiopulmonary  Resuscitation  CPR  and Emergency Cardiovascular Care  ECC     You should have been trained in the appropriate version of the AHA ERC guidelines and the use of your device configuration    Contact HeartSine or you
2.   online at www heartsine com or can be requested on CD  USA      If you have had any occasion to use your samaritan   PAD or if  you require any further information on the samaritan   PAD  its  accessories or any other products please contact us     ABBREVIATIONS    CPR  Cardiopulmonary resuscitation    CPR D  Cardiopulmonary resuscitation Defibrillation    SCA  Sudden cardiac arrest    VF  Ventriculary Fibrillation    BLS  Basic Life Support    ACLS  Advanced Cardiac Life Support          PRODUCT LIST  SELECTION     PAD DEVICES  PAD BAS 05  PAD DAT 05  PAD FIR 05  PAD TRN 05  DISPOSABLES  PAD PAK 01    DATA MANAGEMENT  amp  ACCESSORIES    PAD ACC 01  PAD ACC 02  PAD ACC 03  PAD ACC 04  PAD ACC 09  PAD ACC 16  TRANSPORT  amp  CASES  PAD BAG 01  PAD BAG 02  PAD BAG 03  PAD BAG 04  PAD CAB 01  PAD CAB 02  PAD CAB 03    TRAINING SYSTEMS 8  LITERATURE  TRN SYS 05  TNR ACC 01  TNR ACC 02  TNR ACC 03    Order Fax   44 2890939401    PAD User Manual    PAD Basic Life Saver System  AHA ERC 2005 Guidelines   PAD with Data Management Package  AHA ERC 2005 Guidelines  PAD First Responder  AHA ERC 2005 Guidelines   PAD with Trainer Kit  AHA ERC 2005 Guidelines    Pad Pak    ADULT Cartridge    USB Cable with SAVER    Software CD ROM  amp  License   USB Data Download Cable   SAVER    Software  SAVER Software CD ROM  amp  License   Triangular Wall Sign  AED Wall Sign  Includes Triangular Wall Mount   PAD Prep Kit   Resuscitation Bag with O  Connection and Adult Mask    Soft Carrying Case   
3.  defibril   lation therapy is in process  Defibrillation energy can cause  injury     As long as the defibrillator is used according to the directions   and no one isin contact with the patient when the Shock button is  pressed  there is no risk of harm to the rescuer or bystanders   The samaritan   PAD cannot deliver a shock unless the  electrodes are applied to someone whose heart is in need of a  shock        Ee See Warnings and Precautions for more details        13       AFTER USING THE PAD    After using your samaritan   PAD HeartSine Technologies  recommend you perform the following actions     1  Use Saver    software to download information  about the therapy performed and store  appropriately   If you do not have Saver  software please contact your dealer who can  arrange for the incident to be downloaded         2  Remove the used Pad Pak    from your  samaritan   PAD and dispose of in a suitable  manner   For recommended disposal methods  please refer to section Pad Pak    disposal     3  Check the exterior of the samaritan   PAD for  cracks or other signs of damage  Contact your  distributor or HeartSine Technologies  immediately if any damage is found     4  Check the exterior of the samaritan   PAD for  dirt or contamination  If needed clean device  with approved cleaning products     5  Check supplies  accessories and spares for    Lise  Sits   damage or expiration  Replace immediately if  2008 er    any damage or expiration is found  Contact  Lot No   i  143 
4.  is passed through  the heart  firstly in one direction and then in another     Defibrillation Pads  Defibrillation pads are the electrodes that are connected to the  patient s chest in order to administer therapy     Electromagnetic Interference  Electromagnetic interference is radio interference that may  Cause erroneous operation of electronic equipment     Impedance Measurement  Impedance measurement is a check that is performed to  check the integrity of PAD patient contact     samaritan   PAD   The PAD is a semi automatic device used for the delivery of  external defibrillation therapy to resuscitate victims of SCA   who are unresponsive  are not breathing  or without life signs     Saver      Saver    is software that can be used in conjunction with the  PAD and a USB cable  It can retrieve and view information  about therapy delivered using the samaritan   PAD  Also   Saver    software can be used to configure the PAD     SCOPE       SCOPE    stands for Self Compensating Output Pulse Envelope    Waveform  This is the biphasic technology developed by  HeartSine that is incorporated into the samaritan   PAD     18    Sinus Rhythm   Sinus Rhythm is the normal electrical rhythm by which the heart  muscle contracts and expands to create blood flow around the  body     Ventricular Fibrillation  Is a life threatening heart rhythm that is treatable with the therapy  using the samaritan   PAD     MORE INFORMATION  A copy of the detailed samaritan   PAD user manual is available
5.  location   a flat surface     Wall cases differ in some countries  Ask HeartSine or  your authorised distributor     Push Pad Pak    into the opening and  listen for the    click    sound to ensure it WHERE TO STORE YOUR PAD  is properly inserted   The PAD should be kept in a convenient central area  Place  Once the Pad Pak    is installed properly it near a telephone so that the rescuer can call Emergency  neo ada MA begiga Medical Services and retrieve the PAD without wasting time        Some important points to remember when storing      gt  Store the PAD in a suitable location for easy  INSTALL 2 CHECK DEVICE is working optimally  access     b gt  PUSH the ON Button  Ensure you can  gt  Do not lock the location where PAD is being 7    hear the voice prompts  placed     e CALL for medical assistance l l   gt  Do store the PAD in a clean and dry environment   Remove clothing from patients    chest to expose bare skin Make all necessary arrangements to ensure that the device     gt    is accessible at all times  Inform any possible users of the  Switch OFF by pressing the Button location of the samaritan   PAD     www heartsine com T7       PREPARING YOUR PAD FOR USE    SELF TEST    The samaritan   PAD includes an automatic self test which is  performed on a weekly basis  The self test program will run  automatically and requires no user interaction     Upon completion of self test the PAD will emit a    beep     The  self test program will test your PAD and ascertain if it
6.  of pad positioning  instructions  as indicated on the labelling and in  training  is essential  Care must be taken to ensure  pads are adhered to the patients    skin properly  Air  pockets between the adhesive pad and skin must be  eliminated  Failure in pad adhesion may hinder  effectiveness of therapy or cause excessive skin  burns to the patient if a therapeutic shock is applied    A Reddening of the skin may appear after use  this is  normal     Periodic checks of this device must be undertaken to     ensure among other things that the samaritan   PAD  is not damaged in any way     The Pad Pak    is a single use item and must be  replaced after each use or if pouch that seals  defibrillation pads has been broken compromised in  any way  If damage is suspected the Pad Pak    must  be replaced immediately     INCIDENT NOTIFICATION  As a user of an automated external defibrillator it is essential that  you inform HeartSine of any incident where your PAD is sus     pected to have caused a death  serious injury or illness  If you  have any suspicions that this is the case inform HeartSine directly  or through your authorised HeartSine dealer              17       GLOSSARY    Pad Pak      The Pad Pak    is a tray that fits into the samaritan   PAD  This  pack contains the defibrillation electrodes and the battery that  supplies power to the samaritan   PAD  Pull the green tab to  access the defibrillation pads     Biphasic Shock  A biphasic shock is an electrical current that
7.  safeguard against interference you  must operate the PAD at least 2 m    6 feet  away from all RF devices and  other susceptible equipment  Alter   natively switch off equipment  affected by or causing the electro   magnetic interference     Pull green tab to remove Pad       7    Grip the SECOND GREEN Tab of Pad Pak    and PULL         STEP 5    Remove clothing to expose the patient s chest  If the patient  has an excessively hairy chest  shave the area where the  electrodes are about to be applied     e Peel pads from liner  Peel the electrodes from the  liner        Apply pads to patient s bare  chest as shown in picture        and apply them to the bare chest  as illustrated on the electrodes     Press pads firmly to patient   s  bare skin       PERFORMING DEFIBRILLATION    Place the electrodes on the patients chest as indicated below   Sternum and Apex electrode pads are clearly identified on the  respective electrodes        Press the electrodes firmly to the  patient s bare chest to ensure proper contact is made     Placement of the pads is critical  Strict  observance of pad positioning instruc   tions  as indicated on the labelling and  in training  is essential  Failure in pad  adhesion may hinder effectiveness of  therapy or cause excessive skin burns   A to the patient if a therapeutic shock is  applied     STEP 6    When the electrode pads are attached correctly to the patient  you will hear the audio prompt     i Assessing heart rhythm    id Do not touch the pa
8.  use     The samaritan   PAD has the capability to deliver  therapeutic electrical shocks  The electrical shock can  cause serious harm to either operators or bystanders   Caution must be taken to ensure that neither the   A operators nor bystanders touch the patient when a  shock is to be delivered     To safeguard against interference you must operate   the samaritan PAD 2 meters  6 feet  away from all RF   devices and other susceptible equipment    Alternatively switch off equipment affected by or  A causing Electromagnetic Interference     The samaritan   PAD has been designed to work on  unconscious  non responding patients  If the patient is   A responsive or conscious do not use the samaritan    PAD to provide treatment     Touching the patient during the analysis phase of   treatment can cause interference with the diagnostic   process  Avoid contact with the patient while analysis   is being carried out  The device will instruct you when  A it is safe to touch the patient     It has been determined that the samaritan   PAD is   safe to use in conjunction with oxygen mask delivery   systems  However  due to the danger of explosion it is   strongly advised that the samaritan   PAD should not   be used in the vicinity of explosive gases  This   includes flammable anaesthetics or concentrated  A oxygen     www heartsine com    The samaritan PAD must only be used on patients  over 8 years old     Proper placement of the samaritan   PAD pads is  critical  Strict observance
9.  with abrasive materials  cleaners or    Do not clean the samaritan   PAD  A solvents           HOW TO USE THE PAD    The HeartSine samaritan   PAD is designed for the treatment of  sudden cardiac arrest  SCA   It should only be used to treat  someone who may be a victim of a SCA and is     B   unresponsive to stimulus   B   not breathing   B   exhibiting no signs of life     If the person is unresponsive but you are unsure that they have  suffered from a SCA begin CPR  When appropriate apply the  defibrillator and follow the audible instructions     The PAD will only administer a shock if   it is needed  A voice prompt will tell you  A when to press the shock button to   administer defibrillation therapy     The samaritan   PAD should not be  A used on someone who is responsive  when shaken or breathing normally     PRE DEFIBRILLATION ACTION  Prior to using a PAD it is advised to perform the following checks  and actions in order to prepare the patient     Remove clothing to expose bare chest     If excessively hairy shave hair from areas to which  defibrillation pads are to be applied     Ensure that the patient chest is dry  If necessary dry  chest area          gt  Ensure no rescuer or bystander are in contact with  the patient while the PAD is assessing the patients   heart rhythm or while defibrillation shock is being   applied            CPR FUNDAMENTALS    The following is a brief overview of the fundamentals of CPR for lay  rescuers as advised by both the American He
10. Hard Carrying Case  Pelican Yellow Hard Case with Foam Insert    First Responder Rucksack  PAD CODRA Rucksack  Red  Light Version   Professional Rescuer Rucksack  PAD CODRA EMC    Rescue Cabinet with Alarm   Basic   Wall Bracket   samaritan   PAD   Vehicle Mount  USA UK     PAD Training System  AHA ERC 2000 2005 Guidelines  Replacement Battery Charger   Trainer Pads  set of 10    Trainer Pads  set of 25     OR contact your authorised HeartSine Distributor 19       Worldwide Headquarters  Manufactured by     HeartSine Technologies Inc HeartSine Technologies Ltd  105 Terry Drive Canberra House  Newtown 203 Airport Road West  PA 18940   USA Belfast BT3 9ED   Northern Ireland  Tel  1 215 860 8100 Tel   44  0 28 9093 9400  Fax  1 215 860 8192 Fax   44  0 28 9093 9401       y    Authorised Distributor    supportiheartsine com       
11. are  ECC  and or the use of an AED      specifically the use of the samaritan   PAD   Many training organi       zations offer a course that combines CPR and defibrillator  CPR D        use  Course lengths and detail covered vary but most basic courses      can be completed in one day  For information on how to get training    et your local authorised distributor or HeartSine Technologies     directly     SEA Y      TRAINING      SCA is a condition requiring immediate emergency medical interven       tion  This intervention  due to the nature of the condition  can be per       formed prior to seeking the advice of a physician  In order to properly      diagnose this condition HeartSine recommends that all potential     users of the samaritan   PAD    as a minimum  are fully trained in      cardiopulmonary resuscitation  CPR   basic life support  BLS  and in      particular the use of an Automated External Defibrillator  specifically      the PAD  It is also recommended that this training be kept up to date      by means of regular refresher courses as and when recommended     by your training provider  If potential users of the PAD are not trained      in these techniques contact your HeartSine dealer or HeartSine     directly either of whom can arrange for training to be provided      Alternatively contact your local government health department for     information on certified training organisations in your region     a nnannnaunnunuununununnnnannnnnnunnnaunununnnnnnnnnnnum
12. art Association  AHA   and the European Resuscitation Council  ERC  in their 2005  published Guidelines for CPR and ECC  This is intended only as a  quick reference for trained CPR providers  HeartSine recommend  that all potential users of the samaritan   PAD are trained  by a  competent training organisation in both CPR and PAD use  prior to  placing of a samaritan   PAD into service           CPR GUIDELINES 2005    CALL 911 112 get AED    HA  Alternate with second person after one cycle or 2 minutes 2 MINUTES  5 CYCLES    www heartsine com    HeartSine PAD 300    Quick Card  User Instruction  PERSON NOT RESPONSIVE  samaritan  FAD  NO SIGNS OF LIFE     Adress person  amp  shake on shoulder     Engage other people to help you  and alternate CPR     fa    a  N A    OPEN THE AIRWAY      CHECK FOR BREATHING     2 breaths  if necessary    PERFORM CPR           Until an AED is available  OR arrival of emergency physician    If AED AVAILABLE switch ON  amp  follow instructions    YES    DELIVERSHOCK m    CONTINUE CPR FOR    2 MINUTES  5 CYCLES  Ga  a  lt a  lt  lt   Goa eh    No    CONTINUE CPR FOR    ANALYSIS  SHOCK DECISION       11       PERFORMING DEFIBRILLATION    Call for medical assistance        STEP 2 Lay the PAD on a flat surface      gt  Press ON OFF Button and open  the GREEN Tab of the soft carry  case    The samaritan   PAD is activated and you  hear the audio prompts     ag CALL for medical assistance    Remove clothing from patients  chest to expose bare skin    To
13. ator  lf the Pad Pak    has been  inserted correctly  Status Indicator flashes  approximately every 5 seconds       If necessary inform relevant safety officer or person  responsible for maintenance of the PAD       Update the relevant records to show the date that the  replacement Pad Pak    was placed into service     MAINTENANCE    HeartSine recommends users perform regular maintenance  checks  A suggested maintenance check would be     1  Check the Status Indicator  If the Status Indicator is not  flashing a problem has been detected  Refer to the  troubleshooting section of this manual     2  Check the expiration date of the Pad Pak    currently  lise   inserted into the PAD  If the Pad Pak    has  Until    exceeded its use by date  remove it from the   PAD and replace with a new Pad Pak      _ Lot No    Contact your local authorised HeartSine  distributor for replacements     3  Check supplies  accessories and spares for damage or  expiration  Replace any accessories found to be  damaged or that have exceeded their expiration date        4  Check the exterior of the samaritan   PAD for cracks  or other signs of damage  Contact your authorised  HeartSine distributor if any damage is found     5  Check that trained responders are aware of the PADs  location and that it is easily accessible for those  Responders at all times     6  Ensure all trained responders have up to date training  for both CPR and AED use  For recommended  retraining intervals please consult the orga
14. ators is 150J 200J  Give second and subsequent shocks at 150J or greater  The 2000 guidelines  recommended up to 3 shocks followed by 1 minute of CPR 15 compressions to 1 ventilation   Further details can be found on the AHAand ERC websites     The samaritan   PAD manufactured in 2006 can be factory set to perform to the AHA ERC 2005 orthe AHA ERC 2000 guidelines as required by users        INTRODUCTION    The HeartSine samaritan   PAD is an automatic external defibrillator   AED  used for the fast delivery of defibrillation electric shock therapy  to resuscitate victims of sudden cardiac arrest  SCA        SUDDEN CARDIAC ARREST  SCA      Sudden cardiac arrest is a condition in which the heart suddenly     stops pumping effectively due to a malfunction of the heart s elec      trical system  Often victims of SCA have no prior warning signs or     symptoms  SCA can also occur in people with previously diagnosed    heart conditions  Survival for an SCA victim depends on immediate    cardio pulmonary resuscitation  CPR   The use of an external def      ibrillator within the first few minutes of collapse can greatly improve    the patients  chances of survival  Heart attack and SCA are not the      same  though sometimes a heart attack can lead to a SCA  If you are        experiencing symptoms of a heart attack  pain  pressure  shortness    of breath  squeezing feeling in chest or elsewhere in the body  seek    emergency medical attention immediately      HEARTRAYTHM Ccccccccccct
15. cctccnten tse l      The normal electrical rhythm by which the heart muscle contracts to      create blood flow around the body is known as Normal Sinus Rhythm         NSR   Ventricular Fibrillation  VF  caused by chaotic electrical sig     nals in the heart is often the cause of SCA  but an electrical shock    can be administered to re establish normal sinus rhythm  This treat       ment is called defibrillation  The samaritan   PAD is a device design           ed to automatically detect ventricular fibrillation  VF  and perform  defibrillation on victims of sudden cardiac arrest     A        DETECTING FIBRILLATION    The electrical rhythm by which the heart muscle contracts can be      detected and used for medical diagnosis and the resulting reading is      called an Electrocardiogram  ECG   The PAD has been designed to        analyse a patients ECG in order to detect ventricular fibrillation  VF       in the heart  If ventricular fibrillation  VF  is detected the samaritan          PAD will deliver a carefully engineered electrical shock designed to    stop the chaotic electrical activity experienced within the heart     muscle during SCA  This may allow the victim s heart to return to a    normal sinus rhythm     a     n  an n n snn nun unnn nuunuunununununun nunun NuNNENENE   S    SAA A S      INTENDED USERS    The samaritan   PAD is intended for use by those who have been      user receives training in cardiopulmonary resuscitation  CPR       emergency cardiovascular c
16. e samaritan   PAD will only administer a  shock if it is needed     rd It   s safe to touch the  patient    Check for Airway  Check for Breathing    Check for Circulation       If needed begin CPR    4  e  4  4       If needed begin CPR     The PAD will continue in a 3 escalating  energy shock protocol with a one minute CPR    interval       The energy and CPR configuration of your  device can be seen by reading the label  situated on the front membrane  e g     Follow the voice prompts until the  emergency services arrive     TNA A A 2      AHA ERC 2005 configuration    El It s safe to touch the  patient          E Begin CPR       Begin CPR immediately       The PAD will remain in CPR mode for 2 minutes  After 2    minutes of CPR you will hear the following audible prompt        if Stop CPR         The PAD will then return to Step 6  Ensure no one is in contact        with the patient and proceed as before     NOTE PERFORMING CPR   When performing CPR watch and listen to the PAD  The    OK to  touch indicator    will flash  The PAD may emit 100 beeps per  minute as a guide to CPR when configured to do so    100 is the recommended rate to perform compressions under  AHA ERC 2005 guidelines     Your HeartSine dealer will have trained you in the  particular SCA treatment protocol you have  chosen  In all cases follow the audio and visual  prompts given by your PAD     www heartsine com       NOTE USER AND BYSTANDER SAFETY   Do not touch the patient while the PAD is analysing or
17. nisation  or body used to provide the Training     STATUS INDICATOR    The samaritan   PAD includes a status indicator  This is an  indicator which will flash approximately once every five seconds   When itis flashing itis an indication that the PAD is ready for use     If this indicator is not flashing then there is a problem with your  samaritan   PAD  If this is the case please refer to the trouble   shooting section for further guidance and fault finding advice     This is an indicator which will flash   once  every five seconds  When it is flashing it is  an indication that the PAD is ready for use     SAMARITAN   PAD SOFT CARRY CASE    The samaritan   PAD and samaritan   PAD soft carry case have  been designed to allow the rescuer to use the samaritan   PAD  without having to open the carry case     FRONT BACK    Clear window for     quick start card     www heartsine com    A clear plastic cover protects the samaritan   PAD while  allowing the rescuer to operate the unit  If your PAD is stored in  the soft carry case it is not necessary to remove it from the case  to operate it     Pull GREEN Tab  to remove electrodes     CLEANING YOUR PAD    To clean the samaritan   PAD wipe the device with a soft cloth  that has been dampened by one of the following       gt  Soapy water     gt  Isopropyl alcohol  70  solution    Do not immerse any part of the PAD  in water or any type of fluid  Contact    with fluids may seriously damage the  device or cause fire or shock hazard    
18. nu     A The samaritan   PAD must be used  by a person trained in its use     Included with this packaging are training materials to help familiarize  people trained in the use of AEDs with the operation of the PAD     There is limited published data on defibrillators  used by minimally trained rescuers in the home   It is not clear what safety issues may occur from  defibrillator use in such an environment        Check with local government health  department for Information about any    requirements associated with ownership  and use of a defibrillator in the region  where it is to be used             trained in its operation  It is highly recommended that any potential                              A A E   www heartsine com 5       YOUR SAMARITANO PAD    UN PACKING YOUR SAMARITANO PAD    Open the outer box  remove the  PAD and all accessories     Fill out the Warranty Registration card and send  to HeartSine Technologies     Read this User Manual   Review the product training materials  USA    Ensure all potential users are suitably trained     Place the HeartSine samaritan   PAD into service     PLEASE NOTE    HeartSine Technologies or the authorised distributor are not  obliged to carry out service repairs under warranty if     1  Unauthorized modifications have been made to the device   2  Non standard components are used   3  The user has not used the device in accordance with the    indications for use or the instructions provided in this manual     4  The serial numbe
19. r authorised HeartSine distributor for further information     AHA ERC 2000 AHA ERC 2005    Configuration  Label    AHA ERC 2000  100J 150 J 200J    3 Shocks k Sr An Hmm a     Ar 1 Shock  60 sec CPR Pause     e   120 sec CPR Pause  150J 150J 200J     150J 150J 200J    CPR GUIDELINES 2005 VERSUS 2000   In December 2005 both the American Heart Association  AHA  and the European Resuscitation Council  ERC  in collaboration with the International Liaison  Committee on Resuscitation  ILCOR  released new guidelines for Cardiopulmonary Resuscitation  CPR  and Emergency Cardiovascular Care  ECC   These  2005 guidelines superseded the previous guidelines released in 2000 and contained several changes to the recommended procedures for both CPR and  automatic external defibrillation  While the AHA ERC 2005 guidelines are based on the latest research and represent what is believed to be best practice by  both the AHA and ERC both organisations state that the 2000 guidelines continue to offer safe therapy for patients suffering from Sudden Cardiac Arrest  SCA      Main changes in defibrillation therapy from 2000 to 2005    Treat ventricular fibrillation pulseless ventricular tachycardia  VF VT  with a single shock  followed by immediate resumption of CPR  2 ventilations and 30  compressions   Do not reassess the rhythm or feel for a pulse  After 2 min of CPR    5 cycles of 30 2   check the rhythm and give another shock  if indicated    The recommended initial energy for biphasic defibrill
20. r of the apparatus is removed  defaced   misused or altered     5  The device  electrodes or batteries are stored or used  operationally outside of environmental specifications     6  Pad Pak    packaging is not returned     Any claims made under warranty must be directed via your  dealer from whom the device was originally purchased  The  dealer  before carrying out service under warranty  requires  evidence of purchase           The product must be used in accordance with the user manual  and for the purpose for which it was intended  If you have a    query please contact support HeartSine com for assistance     SAMARITAN   PAD FEATURES    LED Status    Data Port Indicator    Action a    Arrows    Pad Pak     Placement  Indicator    Configura   tion Label    Do not  Touch  Indicator  Speaker  y  ON OFF  PAD Pak TM Button  Cartdrige   expiration date     7 year PAD warranty  3 year PAD PAK    shelf time after manufacturing date    Data Port    Saver Software  amp  USB Cable             PREPARING YOUR SAMARITANO PAD FOR USE          PAD PAK    INSTALLATION INSTALL 3  The Pad Pak    includes the battery and defibrillation electrode    pads in one product  This allows for a single expiration date  for  easy consumable management  Place the samaritan   PAD into its Soft    Carry Case  For alternative transport cases  ask HeartSine or your distributor     Remove Pad Pak    from its packaging  INSTALL 4    Place the PAD and the Pad Pak    on    Put into a wall case or safe visible
21. s basic  functions are running  Self test is not able to determine if the  battery and defibrillation pads currently inserted in samaritan  PAD are within their use by date     You must remember to check the expira   tion date on the Pad Pak     regularly     Check that the PAD status indicator can be  seen easily  Ensure that it is flashing  approximately once every 5 seconds     WHEN TO REPLACE THE PAD PAK       The battery lifetime in use can be up to 6 hours monitoring or  30 shocks or a combination of both  A Pad Pak    in the stand  by mode  inserted into the PAD  has a shelf life indicated by the  expiration date  typically 3 years from manufacture   Replace   ment of the battery and defibrillation pack must be carried out if     the expiry date of the Pad Pak    has    gt  been exceeded  If the Pad Pak    has been used  is a    gt  single use item  it must be replaced with a  new Pad Pak    cartdrige     ilf the status indicator on the PAD is not flashing you may need  to replace the Pad Pak        Far diagnosis of the reason for status indicator is not flashing  please refer to the troubleshooting section of this manual        HOW TO REPLACE A PAD PAK      1  Take the replacement Pad Pak    from its protective  bag       Remove the old Pad Pak    which is to be replaced     3  Slide the replacement Pad Pak    into the slot on  the underside of the samaritan   PAD until it clicks       Push the Pad Pak    firmly to ensure it is fully  inserted       Check status indic
22. tient     a Stand clear of patient     The do not touch indicator  below  on  the PAD will be illuminated     www heartsine com       A    NOTE    e    Touching the patient during the  analysis phase of treatment can cause  interference with the diagnostic  process  Avoid contact with the patient  while analysis is being carried out  The  device will instruct you when it is safe  to touch the patient     Follow audio guidance  Do not touch patient or allow  any others to touch the patient while the PAD is  analyzing  After completion of analysis the PAD will  advise you of treatment recommended  Care must be  taken to keep the patient still  A moving patient can  lead to incorrect  delayed or less effective diagnosis  and therapy           13       PERFORMING DEFIBRILLATION    TASA A      AHA ERC 2000 configuration         If the patient requires a therapeutic shock the PAD will start to  charge  In such a scenario you will hearthe following prompts     E Shock advised    rg Stand clear of patient    The samaritan   PAD delivers electrical  shocks which can cause serious harm  to operators and bystanders  Caution  must be taken to ensure no one is in  contact with the patient when a shock       When the PAD has charged to the required level you will hear the    The shock button will be flashing   K Press Shock button now       Ensure no one is touching the patient   When you are certain that no one is  touching the patient press the shock button  to deliver the therapy     Th
23. your local HeartSine approved dealer     6  Install a new Pad Pak     Before installing the  new PadPak    check that its expiration date  has not been exceeded     16     8     hi       7  After installation of the new Pad Pak     Check  the Status Indicator  If the Status Indicator is not  flashing refer to the troubleshooting section of  this manual  If the problem persists  contact  HeartSine or your local approved dealer for  technical support     8  Turn on the PAD and verify that the PAD  operates in the correct manner i e  audible  prompt    Call for medical assistance    can be  heard  Turn off the PAD     9  Contacting HeartSine after use  At HeartSine  we like to hear from our customers whenever  they have any occasion to use any of our pro   ducts  even if therapy is not delivered as part  of the incident  This information is vital to the  continued development and constant improve   ment we strive for in the treatment of sudden  cardiac arrest     TROUBLESHOOTING    Check expiry date of your Pad Pak     Change the Pad Pak     if expiry date has been exceeded      gt        If the status indicator is still not flashing  or    a warning message is heard when samaritan Pad  is turned on  or    If for any reason  you have suspicions that your  PAD is not working correctly contact your authorised  HeartSine dealer or HeartSine directly for support     support heartsine com         WARNINGS AND PRECAUTIONS    The samaritan   PAD must be used by a person trained  in its
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
  User`s Manual  Peerless Alpha 2 x 1.4mm², 100m  Whirlpool GX900QPPS1 User's Manual  Nokia 3200 Bluetooth Headset User Manual  ML1800 Se ML1800 Series User Manual 1800  Easy-WESTERN Super  日立ルームエアコン 取扱説明書  Aide-mémoire en cas de dégâts de tempète  Honeywell 88065 User's Manual    Copyright © All rights reserved. 