GKM and GKM-2 User Guide
Contents
1. Solutions May be due to dirty glass or glass chips on the surface of the instru ment This surface should be brushed clean with a small paint brush after each break to clean away any glass chips which may have been left from the previous break This problem could also be due to poor quality glass which has internal stresses Use quality Ultramicrotomy Glass see Ordering Information 14 SECTION 7 MAINTENANCE Your Glass Knife Maker requires very little maintenance Glass chips may however fall into the scoring shaft groove and impair smooth action of the scoring shaft To clean remove the screws holding the front and rear positioners then remove the four screws which secure the slotted top plate beneath the positioners Remove this plate The scoring mechanism is now easily accessible for cleaning Unscrew the scoring shaft and remove the scoring wheel holder Clean all parts with acetone and relubricate with a light grease Reassemble Since the micrometers remain in the same position the positioners should return to their previous position The scoring wheel can be replaced when necessary by this procedure Simply push the scoring wheel axle thru its hole and insert a new scoring wheel 15 SECTION 8 REFERENCES 1 Griffiths G Simons K Warren G and Tokuyasu K T 1983 Immunoelectron microscopy using thin frozen sections application to studies of the intracellular transport of Semliki Forest virus spike glycoprotein
2. By noting the readout numbers as you start using the GKM 2 you will soon have consistently high quality repeatable breaks With the GKM model it is necessary to estimate the pressure applied SECTION 2 CLEANING THE GLASS STRIPS The glass strips should be carefully cleaned with a good quality mild detergent such as used for cleaning fine crystal glassware Wear rubber gloves and be careful handling the glass to avoid cutting your hands with the sharp edges Use a soft sponge to clean off any residue Rinse with distilled water and dry with a good quality lint free laboratory tissue SECTION 3 PREPARING 25MM GLASS SQUARES WARNING Always wear protective eyewear while breaking glass with this or any other glass knife maker Small chips of glass can fly into your eyes while breaking glass and cause serious eye injury Do not rub your eyes while making glass knives as glass chips can easily be transferred accidentally to your eyes Also be very carefull to clean up glass chips frequently during knife making to prevent injury To produce straight controlled balanced breaks it is necessary to apply pressure equally to both sides of the score and to have the weight of the glass equal on both sides Therefore the glass strips should always be broken exactly in the middle of each piece To begin making glass squares 1 Carefully examine the labeled GKM illustration near the front of the this manual to familarize yourself with t
3. THE GLASS KNIVES Examine the shelves on both glass pieces to determine if the correct shelf width was formed for each opposing knife edge If the shelves were too wide or if the break occurred too close to the corner or even right across the corner or the wrong side then the front or back positioners or both may need to be adjusted Adjust only one positioner at a time Determine which corner was most out of alignment and take note of the setting on the micrometer scale for that positioner Decide which way you want to move the notch in order to shift the score mark relative to the corner of the glass square To move the positioner to the right rotate the micrometer knob clockwise You should be able to observe the movement of the micrometer shaft and the positioner to determine that you are moving the positioner in the direction desired The micrometer is calibrated in millimeters and the smallest graduations are 0 01 mm 1 Oum Very precise and highly reproducible adjustments are easily made This makes it possible to carefully monitor the effects of moving the positioner and also to return to previous settings if desired To start with move the positioner only a small amount as the shelf will shift by a somewhat greater amount Tighten the corresponding clamping knob Now break another glass knife by repeating the above steps Examine the shelves of the new knives you have just broken and determine if you have adjusted the position
4. GKM and GKM 2 Glass Knife Maker Instruction Manual April 2010 Products by Boeckeler Boeckeler Instruments Inc 4650 S Butterfield Drive Tucson AZ 85714 3403 520 745 0001 800 552 2262 Fax 520 745 0004 www rmcproducts com www boeckeler com info boeckeler com Part No 0209 Thank You for purchasing a GKM or GKM 2 Boeckeler Instruments Product Warranty Boeckeler Instruments Inc products are warranted to be free from defects in materials and workmanship for a period of one year from the date of delivery Boeckeler Instruments Inc will repair or replace and return free of charge any part which is returned to its factory within said period transportation prepaid by user and which is found upon inspection to have been defective in materials and workmanship This warranty does not include normal wear from use nor does it apply to any instrument or part which has been altered by anyone other than a Boeckeler Instruments Inc employee The warranty does not apply to any instrument which has been damaged through accident negligence failure to follow operating instructions the use of electric currents or circuits other than those specified on the plate affixed to the instrument use beyond the specified capacity of the instrument misuse or abuse Boeckeler Instruments Inc reserves the right to change alter modify or improve any of its instruments without any obligation whatsoever to make corresponding ch
5. anges to any instrument previously sold or shipped The foregoing obligations are in lieu of all other obligations and liabilities including negligence and all warranties of merchantability or otherwise expressed or implied in fact or by law and state our entire and exclusive liability and buyer s exclusive remedy for any claim or damages in connection with the sale or furnishing of goods or parts their design suitability for use installation or operation Boeckeler Instruments Inc will in no event be liable for any special or consequential damages whatsoever and our liability under no circumstances will exceed the contract price for the goods for which liability is claimed Should you need service for your unit please contact Boeckeler Instruments through any of the means listed on the cover of this manual We will be happy to help you The model and serial number of each unit is registered at the factory when the unit is shipped Please have these numbers handy when you call us or include them in any written correspondence Before You Start Usine Your Grass Knife MAKER The information contained in this manual has been verified and validated and is believed to be accurate for the intended use of the instrument If the instrument or procedures are used for purposes different than the capabilities specified herein confirmation of their validity and suitability should be obtained from an appropriate product specialist at Boeckeler Instrum
6. e GKM 2 Hand tighten the connector 2 Connect the other end of the db 9 serial cable to the serial port B on the back of the digital readout 3 Connect the power cord to the back of the digital readout C 4 Connect the other end of the power cord to an appropriate wall plug MICROMETER INSTALLATION INSTRUCTIONS The micrometers identical for both the GKM and GKM 2 are shipped in separate containers Unpack and install them as follows 1 Insert the micrometer through the hole in the micrometer mount 2 Push it forward until it stops 3 Tighten the set screw Found on the back of the micrometer holder Micrometero Set Screw Scalen Hole Figure 3 View of the GKM GKM 2 from the front left side SECTION 1 DESCRIPTION OF THE INSTRUMENT Glass knives for Ultramicrotomy and Cryoultramicrotomy are made from special process glass strips which may be purchased from Boeckeler Instruments in three thicknesses 6mm 8mm and 12mm The strips are supplied in nominal lengths of 200mm and 400mm The glass strips should be first cleaned with detergent rinsed and then dried The strips are then placed on the top of the instrument scored and then broken so as to yield 25mm glass squares The squares are then scored diagonally to just slightly to the sides of opposite corners and then broken so as to produce two glass triangles with 45 angle cutting edges This instrument is designed to produce glass knives by an adaptation of th
7. e new Balanced Break method of Griffiths et al With this method glass squares are first produced from glass strips by breaking the strip in half and each succesive strip in half until glass squares are produced In this manner each break is as nearly per pendicular as possible Using older methods with other instruments the squares were broken off the end of a longer strip This resulted in unequal pressure being applied on each side of the score due to the greater weight of the longer side of the strip Consequently the edges were not always perpendicular To make knives the squares are scored nearly exactly diagonally such that the score lines are pointing only slightly to the right of one corner and to the left of the opposite corner This method produces knives with an included angle as close to 45 as possible This gives knives of maximum sharpness If the score is pointing too far to the left or right of the opposite corners the final break will go to the nearest edge rather than towards the corner and the real included angle will be significantly greater than 45 This would produce a knife edge which is too blunt especially for cryoultramicrotomy The digital readout included with the GKM 2 makes it much easier to get consistently high quality glass knives from all your breaks The readout indicates the load Mechanical engineering term meaning force or weight on a structure on the glass strip or square at any particular time
8. eadout number The readout typically indicates around 60 for the correct load Some initial testing will help you find the best load for your instrument Once you discover it you will be able to make consistent high quality breaks on all your glass strips 11 After the glass breaks turn the clamping lever clockwise to raise the breaking anvil 12 Pull back on the front positioner retract knob with one hand and secure it with the catch carefully rotating the two halves of the glass square with the other hand until they can be slid out from under the breaking arm The actual knife edge should be about 1 mm or less away from the corner of the original glass square This distance can be evaluated by examining the opposing edge of the opposite piece of glass The opposing edge has a shelf rather than a knife edge since only one knife edge can be formed at each corner The width of this shelf should be carefully evaluated Knives for sectioning plastic resins should have shelves about 0 5 to 1 mm wide 13 Two usable knives should be formed from each glass square Note For cryoultramicrotomy knives the shelf should be only 0 2 mm or less since the smaller the shelf is the more closely the knife edge approaches a true 45 angle This will produce a knife which is maximally sharp for cryoultramicrotomy of frozen tissue but such a knife might dull too quickly for use with plastic resin embedded specimens SECTION 5 EVALUATING
9. ents Otherwise Boeckeler Instruments does not guarantee results and assumes no obligation or liability This publication is not a license to operate under nor a recommendation to infinge upon any process patents NOTES CAUTIONS and WARNINGS within the text of this manual are used to emphasize important and critical instructions WARNING A warning informs you of a hazard or an unsafe practice that could result in personal injury or effect the operators health CAUTION A caution informs the operator of an unsafe practice that could result in the damage of the equipment NOTE Notes are used to highlight essential information aR iii and Glass Strip 50 mm stop Breaking Anvil flipped back pu HOO MMSSE 200 mm stop Rear Positioner Adjustment Knob Breaking Lever Front Positioner fe Maker Adjustment Knob _ Glas Kni Scoring Shaft Score Selector A _ ae Clamping Lever Figure 1 Diagram of the GKM Glass Knife Maker ATTACHING THE DicitaAL READOUT GKM 2 Onty NOTE The GKM model glass knife maker does not have a digital readout If you have a GKM simply ignore all references in this manual to the digital readout The glass knife maker unit functions identically in both models To attach the digital readout to the GKM 2 Figure 2 Digital Readout Connections for the GKM 2 1 Connect one end of the db 9 serial cable included to the serial port A on the back of th
10. er in the right direction If you are close to the desired shelf width you may want to adjust the positioner for the opposite corner to improve its shelf width Continue breaking knives until you have both positioners adjusted to produce two good knife edges from each glass square Record the micrometer scale settings for future reference particularly if there are multiple users of this instrument Examine both the knife edge and the counterpiece shelf from each corner Be careful not to damage the opposite corner as its knife edge may also be usable Ideally the knife edge and its counterpiece edge should be both straight and perpendicular to the edge of the glass The usable knife edge is generally in the middle of the knife edge or towards the top of the curved stress mark As mentioned above the shelf on the counterpiece should be less than 0 2mm wide if the knives are to be used for cryoultramicrotomy If it is larger the break fracture line tends to wander away from the direction of the score and goes towards the nearest edge rather than towards the corner This can result in a knife edge whose real included angle at the very edge may be as much as 10 or more greater than 45 Such a knife will be relatively blunt and will not cut the thinnest ultrathin sections It would also be prone to cause section compression 13 SECTION 6 TROUBLESHOOTING Problem Slanted or curved breaks edges not perpendicular to the flat surface
11. ery nearly perpendicular and have minimal stress lines Position the glass square so the edges are oriented to become the side where the knife edge will be formed and the manufacturer s scored edges face upward Pull back on the front positioner retract knob and flip over the catch located to the left of this knob Insert the glass square The far corner of the glass square should be placed in the notched space of the rear positioner The corner nearest the operator should be held in position by the notch on the front positioner Flip back the catch and release the spring loaded retract knob to secure the glass square between the notches of the front and rear positioners This gives a positive and reproducible alignment of the squares Lower the breaking anvil over the glass by rotating the clamping lever counterclockwise The digital readout numbers will begin to rise on the GKM 2 Pull the scoring shaft towards you with a smooth motion to produce an even score 11 You may pre stress the glass by lifting the breaking lever see Section 3 6 if you need a deeper score for thicker glass 10 Slowly lift on the breaking lever until you see the score begin to spread upward through the glass forming a fracture Stop and let the glass break by itself This produces a break by the slow break method which is generally regarded as producing the best knives If you are using the GKM 2 make a note of the digital r
12. he control locations Be sure that the scoring shaft is positioned all the way in This places the scoring wheel behind the glass strip Select the proper score Rotate the score selector so that the number 1 lines up with the score index Take a freshly cleaned glass strip and place it on the top of the instrument such that its right edge stops against the 200mm or 100mm stop depending upon the length of the strip The stop should be flipped forward The manufacturer s score marks should face upward Be sure that the back edge is firmly against the rear positioner so that the score will be made exactly at a 90 angle Hold in postion with your hand 6 10 Tighten the glass strip in position by rotating the clamping lever counterclockwise until firm contact is made On the GKM 2 the numbers on the readout will begin to rise slightly indicating the load being placed on the glass strip When breaking 8mm or 12mm glass you may wish to produce a deeper score To do this raise the breaking lever to a point approximately 10 mm above the top surface of the instrument to pre stress the glass Pull the scoring shaft forward with a smooth even stroke Leave in the forward position Should the glass break at this point the glass was pre stressed too hard Apply pressure to the anvil by slowly moving the breaking lever upward As slow breaks often yield straighter more stress free breaks move the lever up o
13. nly until resistance is felt and then watch until the score begins to spread upward through the glass and the glass breaks This will require some practice until you have acquired the right feel However even faster breaks will often yield good edges Once the glass breaks lower the breaking lever and lift up the breaking arm from the glass by rotating the clamping lever clockwise Pull the glass pieces away carefully to avoid damaging the freshly broken edge 11 Push the scoring shaft inwards to prepare for the next break You now have two equal length strips Place one of the strips aside on a clean cloth until needed Always keep the manufacturer s score sideup NOTE The top surface of the instrument should be brushed off with the paint brush to remove all glass chips after each glass breaking operation Failure to remove all glass chips before positioning the glass for the next break could result in uneven pressure and a poor quality break fracture edge Do not brush glass chips into the scoring shaft slot 12 Flip the 100mm or 50mm stop down 13 Place the glass strip on the instrument so that the right edge is against the stop Firmly position the glass strip against the back positioner and repeat steps 5 through 10 14 Continue breaking the glass strips in a symmetrical manner such that the right edge stops against the 100mm 50mm and finally 25mm stops Always break each glass piece exactly in half u
14. s Methods Enzymol 96 435 450 16 SECTION 9 ORDERING INFORMATION Ultramicrotomy Glass Strips Part No 315626 12mm x 1 x 8 25 4 x 203mm package of 16 pieces Part No 316995 6mm x 1 x 16 25 4 x 406mm package of 30 pieces Part No 316996 8mm x 1 x 16 25 4 x 406mm package of 24 pieces 17
15. sing the appropriate stop so that all the breaks are balanced Note that the last square produced from the left end of the original strip will probably not be an exact usable square unless the original strip was exactly 400mm long 16 inch long glass strips are 406mm long and the last square will be an unusable rectangle Section 4 Makina Knives From THE SQUARES The glass squares are scored nearly exactly diagonally to produce two glass knives from each square The glass squares are positioned diagonally in the knifemaker A knife edge should be produced just to one side of the top corner and just to the other side of the bottom corner If the score were exactly diagonal the break would occur right through the top and bottom corners of the square and no useful cutting edges would be produced Since the glass knife maker scores and breaks symmetrically it may be adjusted to form knife edges in either the upper left and the lower right corners for example or the upper right and lower left Just be consistent in order to simplify adjustments l Be sure the scoring shaft is pushed all the way in This places the scoring wheel behind the glass Rotate the score selector to number 2 Examine the glass square and determine which sides were freshly broken and which were broken by the manufacturer Most microtomists prefer to make the knife edges on the edges which were produced by the manufacturer since these edges are v
Download Pdf Manuals
Related Search
Related Contents
Bedienungsanleitung - Landhouse equipment Visualizza - Service, Support Documento PDF - AMS Tesi di Laurea d`informations NB! Skal ha ASK forside. Pass på riktig ombrekking ーnstructi。n FQーA ] FQーV Mention : Français Langue Étrangère : Parcours 取付・取扱説明書 - Dynamic DO!.jp S4D95LE-3 DIESEL ENGINE Bases Chroma Dupont - automotriz servi 124, ca Copyright © All rights reserved.