Home
KERN FTC - KERN & SOHN GmbH
Contents
1. 5 2 1 TOLEDO Continuous commands The scale supports the following input commands in TOLEDO Continuous mode Command Meaning P lt CR gt lt LF gt Print out the current result T lt CR gt lt LF gt Tare the scale Z lt CR gt lt LF gt Zero the display C lt CR gt lt LF gt Clear the current value Tx xxx lt CR gt lt LF gt Define tare 5 2 2 Output format in TOLEDO Continuous mode Weight values are always sent in the following format in TOLEDO Continuous mode Status Field 1 Field 2 1 2 3 4 5 G 7 8 9 10 11 12 13 14 15 16 17 18 STX SWA SWB SWC MSD LSD 14150 LSD CR CHK Field 6 digits for the weight value that is sent without a decimal point and unit Field 2 6 digits for the tare weight that is sent without a decimal point and unit STX ASCII characters 02 hex characters for start of text SWA SWB SWC Status words A B C see below MSD Most significant digit LSD Least significant digit CR Carriage Return ASCII characters OD hex CHK Checksum 2 part complement of the binary sum of the 7 lower bits of all previously sent characters incl STX and CR User manual ME Nr 22017145 01 07 42 Interface description FTC BA e 0720 Status word A Status Bit Function Selection 6 5 2 1 0 Decimal X00 0 0 0 0 position XO 0 0 1 X 0
2. Command Meaning LEVEL 0 Reset the scale IO Inquiry of all available 5103 commands E Inquiry of SICS level and SICS versions I2 Inquiry of scale data I3 Inquiry of scale software version 14 Inquiry of serial number S Send stable weight value SI Send weight value immediately SIR Send weight value repeatedly Z Zero the scale Zl Zero immediately LEVEL 1 D Write text into display DW Weight display K Keyboard check SR Send and repeat stable weight value T Tare TA Tare value TAC Clear tare TI Tare immediately LEVEL 2 C2 Calibrate with the external calibration weight C3 Calibrate with the internal calibration weight 110 Inquire or set scale ID 111 Inquiry of scale type P100 Print out on the printer P101 Print out stable weight value P102 Print out current weight value immediately 39 User manual ME Nr 22017145 01 07 FTC BA e 0720 Interface description Command Meaning PWR Power On Off SIRU Send weight value in the current unit immediately and repeat SIU Send weight value in the current unit immediately SNR Send stable weight value and repeat after every weight change SNRU Send stable weight value in the current unit and repeat after every weight change SRU Send weight value in the current unit and repeat ST After pressing the Transfer key send the stable weig
3. 4 3 2 DISPLAY weighing unit and display accuracy UNIt1 Select weighing unit 1 g kg oz Ib t UNIt2 Select weighing unit 2 g kg oz Ib t rESOLU Select readability resolution model dependent UNt rOLL When UNT rOLL is switched on the weight value can be displayed in all available units and as pieces with 4 Notes e Oncertified scales the weighing units oz and Ib are displayed with the symbol Oncertified scales resolutions that deviate from the scale definition are displayed without a weighing unit and with the symbol e On dual range dual interval scales resolutions marked with I lt gt 1 21 are divided up into 2 weighing ranges intervals e g 2 x 3000 d 4 3 3 TARE tare function A tArE Switching on off automatic taring CHAIn tr Switching on off chain tare A CL tr Switching on off automatic taring with automatic clearing of the tare weight when the load is removed from scale 4 3 4 ZERO automatic zero update AZM On certified scales this menu item does not appear Switching on off automatic zero update and selecting zeroing range Possible settings OFF 0 5 d 1d 2d 5d 10 d User manual ME Nr 22017145 01 07 30 Settings in the menu FTC BA e 0720 4 3 5 RESTART automatic saving of zero point and tare value ON OFF When the Restart function is activated the last zero point and tare value
4. A B C D E F G H K L SM 335 265 100 240 200 46 276 208 216 LA 370 360 115 350 240 52 310 304 310 small platform model mm 2 large platform model mm 7 1 5 Net weights Model without battery with battery with internal calibration weight without battery small platform model 4 6 kg 5 3 kg large platform model 8 2 kg 8 9 kg User manual ME Nr 22017145 01 07 48 Technical data and accessories FTC BA e 0720 7 1 6 7 1 7 49 Interface connections The compact scale can be fitted with a maximum of 2 interfaces The following com binations are possible COMI COM2 RS232 RS232 Ethernet RS232 Analog second scale interface Assignment of the interface connections Pin RS232 Analog COM1 Interface COM2 Excitation 8 2 VDC 2 TxD1 2 Sense 3 RxD1 2 Shield 4 Sense b GND Excitation GND 6 7 Signal 8 Signal 9 VCC User manual ME Nr 22017145 01 07 FTC BA e 0720 Appendix 8 Appendix 8 1 8 1 1 Table of Geo Values For weighing instruments verified at the manufacturer s the geo value indicates the country or geographical zone for which the instrument is verified The geo value set in the instrument e g Geo 18 appears briefly after switch on or is specified on a label Table GEO VALUES 3000e shows the geo values
5. 1 3 4 Adjustment General According to the EU guideline 90 384 EEC balances must be verified officially if they are to be used as follows legally regulated area e For commercial transactions if the price of goods is determined by weighing e For the production of medines in pharmacies as well as for analyses in the medical and pharmaceutical laboratory e For official purposes e For the production of finished packages In case of doubt please contact your local office of weights and measures Verification Information An EU qualification approval is available for those balances marked as appropriate for verification in the technical data In the event that the balance is applied in an area subject to verification as described above it must be officially verified and re verified at regular intervals Re verification of a balance is carried out in compliance with the respective legal pro visions of the states The term of verification validity for balances in Germany for example is normally 2 years The legal provisions of the country of use are to be observed User manual ME Nr 22017145 01 07 Operation FTC BA e 0720 2 Operation 2 1 Switching on Switching off 2 2 Automatic 2 3 Switching on and off Press amp The scale conducts a display test When the weight display appears the scale is ready to weigh gt Press amp Before the display goes out OFF appears briefly
6. Zeroing Zero point correction Zeroing corrects the influence of slight changes on the load plate 1 Unload scale 2 Press Goel The zero display appears In the case of scales that cannot be certified the automatic zero point correction can be deactivated in the menu or the amount can be changed As standard the zero point of the scale is automatically corrected when the scale is unloaded Simple weighing 1 Place weighing sample on scale 2 Wait until the stability monitor O goes out 3 Read weighing result User manual ME Nr 22017145 01 07 FTC BA e 0720 Operation 2 4 2 4 1 2 4 2 2 4 3 2 4 4 Weighing with tare Taring Place the empty container on the scale and press rare The zero display and the symbol NET appear The tare weight remains saved until it is cleared Clearing the tare gt Unload scale and press fare The symbol NET goes out the zero display appears If A CL tr is activated in the menu the tare weight is automatically cleared as soon as the scale is unloaded Automatic taring Prerequisite A tArE is activated in the menu the symbol T flashes in the display Place the container or packaging material on the scale The packaging weight is automatically saved as the tare weight the zero display and the symbol NET appear Chain tare Prerequisite The tare function CHAIn tr is activated in the menu With this function it is possible to tare several
7. Only scales that have been aligned precisely horizontally provide accurate weighing results The certified scales have a spirit level to simplify alignment gt Turn the adjustable feet of the scale until the spirit level s air bubble is inside the inner circle The manufacturer adjusts each scale to the local gravity conditions GEO value In the event of major geographical location changes this setting must be adjusted by a service technician Certified scales must also be recertified observing the national cer tification regulations These steps are not necessary for scales with an internal cali bration weight User manual ME Nr 22017145 01 07 FTC BA e 0720 Introduction AN 1 3 2 Note 1 3 3 Connecting the power supply CAUTION Before connecting the scale to the mains check whether the voltage value printed on the rating plate corresponds with the local mains voltage A Never connect the device if the voltage value printed on the rating plate is different to the local mains voltage gt Plug the mains plug into the socket After connection the device performs a self test When the zero display appears the device is ready to weigh gt Calibrate the device in order to obtain the greatest possible precision see Section 4 3 1 Scales with a built in battery can work independently from the mains for approxi mately 30 hours in normal operation A prerequisite for this is that the background li
8. gt Check cable connection between E rr J scale the units gt Check interface settings r r e No calibration gt Unplug the mains plug then plug cere Li it back in switch unit off and then back on in battery mode gt Calibrate scale gt Contact your dealer or local rep resentative 41 e Average piece weight too low Counting is not possible on this E rrr scale with this average piece weight User manual ME Nr 22017145 01 07 44 Event and error messages FTC BA e 0720 Error Cause Remedy D e Unstable weight value when Ensure stable surroundings Err 9 referencing gt Ensure that the weighing pan is freely movable gt Adjust vibration adapter e Printout not yet ended End printout E rr 1 gt Repeat required action cre CO e Switching the weighing unit impermissible during dynamic weighing End dynamic weighing gt Switch weighing unit m LI LA e EAROM checksum error gt Unplug the mains plug then plug it back in switch unit off and then back on in battery mode gt Contact your dealer or local rep resentative Weight display unstable e Restless installation location e Draft e Restless weighing sample e Contact between weighing pan and or weighing sample and surroundings e Mains fault gt Adjust vibration adapter gt Avoid drafts Dynamic weighing gt Remedy contact Check mains Incorrect weight dis
9. or gt Remove the number of pieces displayed above the key and press ref The scale determines the average piece weight and then shows the number of pieces removed together with a minus sign 3 Remove more parts from the container until the required number of pieces is reached Counting with variable reference quantity If VAr SPL ON is set in the menu it is possible to select from 5 preset reference quantities via e gt Press as often as necessary until the display above the key has changed to the desired reference quantity The rest of the counting process is as described earlier Counting with minimum accuracy The item Min rEFW in the menu allows to preset a minimum accuracy of 97 5 99 0 or 99 5 96 On the basis of this the scale calculates the minimum reference weight necessary to reach the defined accuracy 1 Place the reference parts on the scale and press or er 2 If the average piece weight is not sufficient to ensure the desired accuracy Add x PCS appears 3 Add the displayed number of pieces The scale then automatically determines the average piece weight with the larger reference quantity The rest of the counting process is as described earlier User manual ME Nr 22017145 01 07 FTC BA e 0720 Counting 3 5 3 5 1 Note 3 6 3 7 Reference optimization The greater the reference quantity the more accurately the scale determines the number of pieces Autom
10. 7 0 fer 30 4t 33 9 11 12 10 11 9 10 18 9 7 8 33 9 5 3t 12 183 11 12 10 11 9 10 er 35 31 37 50 13 14 12 13 11 12 110 11 9 10 37 50 40 5 14 15 13 14 12 13 11 12 10 11 40 5 42 19 15 16 14 15 13 14 12 18 11 12 42 19 44 32 16 17 15 16 14 15 18 14 12 18 44 32 46 45 17 18 16 17 15 16 14 15 13 14 46 45 48 58 18 19 17 18 16 17 15 16 14 15 48 58 51 13 19 20 18 19 17 18 16 17 15 16 51 13 53 31 20 21 19 20 18 19 17 18 16 17 53 3t 55 52 21 22 20 21 19 20 18 19 17 18 55 52 58 17 22 28 21 22 20 21 119 20 18 19 58 17 60 49 23 24 22 23 21 22 20 21 19 720 60 49 63 30 24 25 23 24 22 23 21 22 20 21 63 30 66 24 25 26 24 25 23 24 X 22 28 21 22 66 24 69 35 26 27 25 26 24 25 X 128 24 X 22 28 _69 35 7 16 27 20 26 27 25 26 X 24 25 23 24 73 16 77 52 28 29 27 28 _ 26 27 X 25 26 24725 77 52 185 45 29 30 28 29 27 28 26 27 25 20
11. tings in the menu are described in Section 4 4 1 Counting parts into a container 1 Place the empty container on the scale and press rare The container is tared and the zero display appears 2 Place 10 reference parts on the scale and press Gen or Place the number of pieces displayed above the key on the scale and press E The scale determines the average piece weight and then shows the number of pieces 3 Add more parts to the container until the required number of pieces is reached The average piece weight remains saved in the factory setting until a new average piece weight is determined With it is possible to switch between the number of pieces and the weighing units preset When the number of pieces is displayed it is possible to display with the average piece weight i e the weight of a single reference part for 2 seconds If A CL APW ON is set in the menu the average piece weight is automatically cleared after each counting operation The average piece weight must be deter mined again for the next counting operation If ACCurCY ON is set in the menu the accuracy achieved is briefly shown after the number of pieces is determined User manual ME Nr 22017145 01 07 18 Counting FTC BA e 0720 3 2 3 3 3 4 Counting parts out of a container 1 Place the full container on the scale and press mac The container is tared and the zero display appears 2 Remove 10 reference parts and press e
12. 1 0 DA 0 1 0 0X 0 0 0 00X 0 0 000X 0 0 0000X Numerical X1 0 increment y 1 0 X5 Status word B Status word C Function value Bit Function value Bit Gross net Net 0 0 0 Sign Negative 1 0 1 Overload 1 2 0 2 Movement 3 Print request 3 Ib kg kg 1 4 Extended 1 4 1 5 1 Powerup 1 6 Manual taring only kg 1 6 43 User manual ME Nr 22017145 01 07 FTC BA e 0720 Event and error messages 6 Event and error messages Error Cause Remedy Display Dark e Back lighting set too dark gt Set back lighting b LIGHt brighter e No mains voltage gt Check mains e Unit switched off gt Switch on unit e Mains cable not plugged in gt Plug in mains plug e Brief fault gt Switch device off and back on again Insufficient load e Load plate not on the scale Place load plate on the scale L 4 e Weighing range not reached Set to zero Overload e Weighing range exceeded gt Unload scale nm m gt Reduce preload e Result not yet stable gt f necessary adjust vibration mm a mc m adapter or weigh dynamically e Function not permissible Unload scale and set to zero e Zeroing not possible with over gt Unload scale Fr n ce WE load or insufficient load Ronde u e Average piece weight too low Select and place larger number of E rr reference parts on the scale r e No valid value from the reference
13. 6 20 Counting with automatic reference determination 20 Counting with two scales s s couette eon ovt Rau eel shin esaet ques 20 23 Operating ING 23 Oval Via ois cerita eni UR nde a tp dh dev ee ede bete is ee 25 Scale Seftings CSOALE aan seinen 28 Application settings APPLICATION 32 Terminal settings TERMINAL ssssssssse n 34 Configuring interfaces COMMUNICATION 35 Diagnosis and printing out of the menu settings DIAGNOS 38 Interface description eeeeeee erre enne nnn 39 SICS interface COMMONS ana ante 39 TOLEDO Continuous mode nee Ri eat cle 42 Event and error messages eee 44 Technical data and accessories eeeeennee 46 E le Le ii etos banna snarl 46 3 Table of contents FTC BA e 0720 8 ET amm E T 50 8 1 Table of Geo Values i e emen 50 8 2 Sample Protocols was tran te ee e e e ka 53 9 IEN IS 54 User manual ME Nr 22017145 01 07 FTC BA e 0720 Introduction 1 Introduction 1 1 Safety instructions CAUTION Do not use the scale in hazardous areas Our product range includes special devices for hazardous areas DANGER Electric shock hazard A Always pull out the mains plug before any work on the device DANGER Electric shock hazard if the mains cable is damaged A Check the mains cable for damage regularly and replace
14. FTC BA e 0720 Introduction 1 2 3 Keypad Main functions Key Function in operating mode Function in the menu ON Switching device on off abort To the last menu item End OFF Setting scale to zero Scrolling back Taring scale Scrolling forward Transfer key PRINT Long key press Calling up menu ND m T Activating menu item Accepting selected setting Additional functions Function GA lt D t M m 4 Switching between gross and net weight displaying tare specification resolution Calling up additional information e g gross weight average piece weight higher Switching the scale d Lo 0 Switching between weight value and number of pieces Determining average piece weight from 10 pieces Determining average piece weight from any number of pieces User manual ME Nr 2201 7145 01 07 Introduction FTC BA e 0720 1 3 Major geographical location changes Putting into operation Selecting or changing the location The correct location is crucial to the accuracy of the weighing results gt Select a stable vibration free and if possible a horizontal location The ground must be able to safely bear the weight of the fully loaded scale Observe the following environmental conditions e No direct sunlight e No strong drafts e No excessive temperature fluctuations Aligning the scale
15. The scale determines the average piece weight and changes to the display in pieces PCS 3 Place the parts to be counted on the same scale The number of pieces is displayed User manual ME Nr 22017145 01 07 22 Settings in the menu FTC BA e 0720 4 Settings in the menu 4 1 4 1 1 Operator menu Supervisor menu Note 23 Settings can be changed and functions can be activated in the menu This enables adaptation to individual weighing requirements The menu consists of 6 main blocks containing various submenus on several levels Operating the menu Calling up the menu and entering the password The menu differentiates between 2 operating levels Operator and Supervisor The Supervisor level can be protected by a password When the device is delivered both levels are accessible without a password Press and keep it pressed until COdE appears 2 Press again The menu item tErMINL appears Only the submenu dEVICE is accessible Press and keep it pressed until COdE appears 2 Enter the password and confirm with mJ The first menu item SCALE appears No supervisor password has been defined when the device is first delivered Therefore respond to the password inquiry with when you call up the menu for the first time If a password has still not been entered after a few seconds the scale returns to weigh ing mode Emergency password for Supervisor access to the menu If a password has been issued fo
16. in various capacities and resolutions The power supply is carried out via a built in power supply device an internal rechargeable battery with an external mains adapter or an external battery One of the following options can also be ordered e Additional interface RS232 e Ethernet interface e Analog second scale interface User manual ME Nr 22017145 01 07 FTC BA e 0720 Introduction 1 2 1 Overview 1 Display 2 Scale specifications 3 Load plate 4 Adjustable foot 5 Keys 1 Power supply connection 2 Optional interface RS232 interface User manual ME Nr 22017145 01 07 7 Introduction FTC BA e 0720 1 22 Display oo oo P WN OO Aa Go Ma O a a O A O A dozne s kg Pes 1 234 5 E era ST D HH 8 11 10 9 7 segment display 7 digits with decimal point Active interface Symbol for displaying gross and net values Active scale Weighing range display Battery charge level only present on scales with a battery Weight units Selected reference quantity Symbols for optimizing the average piece weight Symbol for dynamic weighing Graphic display of the weighing range Stability monitor goes out when a stable weight value is reached Sign Identification for changed or calculated weight values e g higher resolution minimum weight not reached User manual ME Nr 22017145 01 07
17. is still available In the example approx 65 of the scale capacity is occupied Dynamic weighing With the dynamic weighing function it is possible to weigh restless weighing samples such as live animals If this function is activated the symbol amp appears in the dis play With dynamic weighing the scale calculates the mean value from 56 weighing oper ations within 4 seconds Prerequisite AVErAGE gt MAnuAL is selected in the menu The weighing sample must be heavier than 5 scale divisions 1 Place the weighing sample on the scale and wait until it has stabilized 2 Press to start dynamic weighing During dynamic weighing horizontal segments appear in the display and the dynamic result is then displayed with the symbol 3 Unload the scale to be able to start a new dynamic weighing operation User manual ME Nr 22017145 01 07 FTC BA e 0720 Operation With automatic start 2 9 2 10 Prerequisite AVErAGE gt Auto is selected in the menu The weighing sample must be heavier than 5 scale divisions 1 Place the weighing sample on the scale The scale starts the dynamic weighing automatically During dynamic weighing horizontal segments appear in the display and the dynamic result is then displayed with the symbol 2 Unload the scale to be able to perform a new dynamic weighing operation Printing results If a printer or computer is connected to the scale the weighing results can be pr
18. it immediately if it is dam aged A On the rear side of the device maintain a clearance of at least 3 cm in order to prevent the mains cable bending too much CAUTION On no account open the device The warranty is void if this stipulation is ignored The device may only be opened by authorized persons gt PP 9 User manual ME Nr 22017145 01 07 5 Introduction FTC BA e 0720 AN Note 1 2 CAUTION Handle the compact scale with care The scale is a precision instrument A When the weighing pan has been removed never clean the area under the load plate holder with a solid object A Do not put excessive loads on the scale A Avoid banging the weighing pan Disposal gt Observe the valid environmental regulations when disposing of the scale If the device has a rechargeable battery The battery contains heavy metals and therefore must not be disposed of with normal waste gt Observe the local regulations for disposing of environmentally hazardous materi als Use with foodstuffs Parts coming into contact with foodstuffs have smooth surfaces and are easy to clean The materials used do not splinter and are free of harmful substances With foodstuffs it is recommended to use the supplied protective cover gt Clean the protective cover regularly and carefully Replace damaged or very dirty protective cover immediately Description The compact scales are available in a small and large size
19. rEF Ct PCS StArLN CrLF F FEEd AIAGNOS tESt SC intErN ExtErN 38 KboArd AISPLAY SNr SNr2 List LiSt2 rESEt AL SUrE 27 User manual ME Nr 22017145 01 07 FTC BA e 0720 Settings in the menu 4 3 Scale settings SCALE 4 3 1 CAL calibration adjustment As the acceleration value due to gravity is not the same at every location on earth each balance must be coordinated in compliance with the underlying physical weighing principle to the existing acceleration due to gravity at its place of location only if the balance has not already been adjusted to the location in the factory This adjustment process must be carried out during the initial start up after change in location and variation of surrounding temperature It is also recommendable to adjust the balance periodically during weighing operation in order to obtain exact measured values Adjusting of non verifiable balances External For scales without an internal calibration weight 1 Unload scale 2 Activate menu item CAT with The scale determines the zero point 0 appears in the display The calibration weight to be placed on the scale then flashes in the display 3 If necessary change the weight value displayed with rare 4 Place the calibration weight on the scale and confirm with Pnr The scale calibrates with the calibration weight loaded After calibration is com pleted donE appears briefly in the display and the
20. reference quantities SPL qtY Reference quantity Sql 565 e Define 5 fixed reference quantities Min reFW Monitoring the minimum reference weight OFF e No monitoring of the minimum reference weight 97 5 99 0 e Monitoring the minimum reference weight so that a counting accuracy of 99 5 97 5 99 0 or 99 5 is achieved rEF OPt Optimizing the average piece weight OFF e No reference optimization AUtO e Automatic reference optimization User manual ME Nr 2201 7145 01 07 32 Settings in the menu FTC BA e 0720 A SMPL Automatic determination of the average piece weight ON e After taring the average piece weight is determined with the next weight placed on the scale and the displayed reference quantity OFF e No automatic determination of the average piece weight A CL APW Automatic clearing of the average piece weight ON e When the load is taken off the scale after a counting operation the average piece weight is automatically cleared The next counting operation begins with deter mining the average piece weight again OFF e The average piece weight is maintained until a new average piece weight is deter mined ACCurCY Displaying the counting accuracy ON e After the average piece weight is determined the counting accuracy that can be achieved is shown briefly in the display OFF e No counting accuracy display tOtAl Ct Counting on two scales bULK e Display number of p
21. scale automatically returns to weighing mode Adjusting of verifiable balances e Switch off the scale e Remove the scale bottom plate by undoing the Torx TX20 retaining screws Impor tant In order to remove the bottom plate it is necessary to break the certification seal affixed to it Once this seal has been destroyed the scale must be recertified User manual ME Nr 22017145 01 07 28 Settings in the menu FTC BA e 0720 29 by an accredited organization and a new certification seal must be affixed before the instrument may be used as a certified scale again e Hold down the service switch pushbutton on the analogprint indicated by an arrow in the drawings below while at the same time switching on the scale Keep ihe service switch pressed in until Service appears in the display Small platform model Large platform model 1 Display Scale Press the El key within 20s 2 Display Metrolo Press the 9 Aen 3 Display Scale 1 Press the E key 4 Display ramp Press the they 5 Display SNR Press the C key 6 Display SCAL bId Press the key 7 Display GEO adjustment by means of GEO value The GEO value can be set in this block and the balance also adapted to the local grav ity ratios without adjusting weights Case a You are familiar with the GEO values Once the o kay has been pressed the current GEO value will be displayed Press the oder Co ey in or
22. 9 KERN amp Sohn GmbH Ziegelei 1 Tel 0049 0 7433 9933 0 WAAGEN GEWICHTE BALANCES WEIGHTS D 72336 Balingen Frommern Fax 0049 0 7433 9933 149 Postfach 40 52 e mail info kern sohn com 72332 Balingen Web www kern sohn com Operating Instructions Compact scales KERN FTC Version 2 0 01 2007 FTC BA e 0720 ME Nr 22017145 FTC BA e 0720 Table of contents Table of contents L2 WN 2 1 2 2 2 3 2 4 2 5 2 6 2 7 2 8 2 9 2 10 2 11 3 1 3 2 3 3 3 4 3 5 3 6 3 7 4 1 4 2 4 3 4 4 4 5 4 6 4 7 5 1 5 2 7 1 User manual ME Nr 22017145 01 07 Page Lut D E 5 Safety 6 6501101000 ausser Ee Aim in 6 Putting into 6662 10 iip M 13 Switching ON dnd Off ic rere 13 Zeroing Zero point correction unse 13 Simple weighing rs ee Ong Tu Y ttk HEU DEA NACE EEE EEEE d 13 Weighing with tare 14 Calling up the gross weight and tare value 0000 15 Displaying weight values with a higher resolution ssssse 15 Displaying the capacity available sssssss 15 DYNGMIC WEIGHING nciht eee 15 Printing resus o en Usate Av be aa Radeon Fulco dura 16 SWIICHING Bel 16 GO aaa a A 17 COUNTING E 18 Counting parts into a container areas 18 Counting parts out of a container o 19 Counting with variable reference quantity 19 Counting with minimum accuracy RR 19 Reference
23. T AZM RESTART FILTER VIBRAT PROCESS StABILI Min WEiG S ONO 53 OF ON OF ON Et 0 ER 16 1 1 04 NO APPr 0000000 SINGLE R g 6100 00 20 01 e 19 g kg 0 01 g OFF 2 2 5 OFF MED UNIVEr StAnDr 000 g OFF APPLICATION VAr SPL Spl Qty sol SQ2 SQ3 504 SQ5 Min RefW REF OPT A SMPL A CL APW ACCurCy tOtAL Ct DYNAMIC Ge tun D G 0 1085 kg T 0 0145 kg N 0 0940 kg COMMUNICATION COM MODI PriNtEr tEmPLat ASCi Fmt LINE FMT1 MULTI E NGtH 1 24 ADD LF 1 0 PARAMET BAUD 1 2400 PAriTY 1 8 odd H SHAKE 1 XONXOFF ChECSUM 1 0FF Vcc L OFF COM 2 MODI PARAMET BAUD PAriTY H SHAKE ChECSUM Vcc OPTION EtH NEt USB ANALOG DiGitAL DEF PrN tEmPLt1 tEmPLt2 1 StdArd DIALOG 9600 8 nonE XONXOFF OFF OFF User manual ME Nr 22017145 01 07 N KERN amp Sohn GmbH P D 72322 Balingen Frommern Tel 0049 0 7433 9933 0 M OTE LALANCESWEIBHIS Postfach 4052 Fax 0049 017433 9933 149 E Mail info kern sohn de Internet www kern sohn de Konformit tserkl rungen Declaration of conformity for apparatus with CE mark Konformit tserkl rung f r Ger te mit CE Zeichen Declaration de conformit pour appareils portant la marque CE Declaraci n de conformidad para aparatos con marca CE Dichiarazione di conformit per apparec
24. United Kingdom 55 00 62 00 23 factory setting User manual ME Nr 22017145 01 07 FTC BA e 0720 Appendix 8 1 2 GEO VALUES 6000e 7500e OIML Class III Height Geografische Breite Geo Wert 00 00 12 44 05 46 17 10 6 12 44 20 45 7 17 10 23 54 8 20 45 26 45 9 23 54 29 25 10 26 45 31 56 11 29 25 34 21 12 31 56 36 41 13 34 21 38 58 14 36 41 41 12 15 38 58 43 26 16 41 12 45 38 17 43 26 47 51 18 45 38 50 06 19 47 51 52 22 20 50 06 54 41 21 52 22 57 04 22 54 41 59 32 23 57 04 62 09 24 59 32 64 55 25 62 09 67 57 26 64 55 71 21 27 67 57 75 24 28 71 21 80 56 29 75 24 90 00 30 User manual ME Nr 22017145 01 07 52 Appendix FTC BA e 0720 8 2 Sample protocols Weighing with tare Dynamic weighing G 0 1085 kg T 0 0145 kg N 0 0940 kg Dyn WT T 43 52 kg 3 78 kg Printout with header G Gross weight N Net weight T 2 Tare Dyn WT dynamically determined weight Protocol of the scale settings menu point List see page 38 SOFTWARE SNR Build SCAL TYP BAS UNIT SCL CAP RESOL GEO DISPLAY UNIt1 UNIt2 RESOLU UNt rOLL tArE A TArE CHAIn tr A CL tr PB TArE ZERO Z CAP
25. are saved After switching off on or after a power interruption the device continues to work with the saved zero point and tare value 4 3 6 FILTER adaptation to the ambient conditions and the weighing type VIbrAt Adaptation to the ambient conditions LOW e Very steady and stable environment The scale works very quickly but is very sensitive to external influences MEd e Normal environment The scale operates at medium speed HIGH e Restless environment The scale works more slowly but is insensitive to external influences PrOCESS Adaptation to the weighing process UNIVEr e Universal setting for all weighing samples and normal weighing goods dOSING e Dispensing liquid or powdery weighing samples StAbILI Adjusting the weighing speed FASt e The scale operates very fast StAndrd e The scale operates at medium speed PrECISE e The scale operates with the greatest possible reproducibility The slower the scale works the greater the reproducibility of the weighing results 4 3 7 X RESET resetting scale settings to factory settings SUrE Confirmation inquiry e Reset the scale settings to factory settings with e Do not reset scale settings with 31 User manual ME Nr 22017145 01 07 FTC BA e 0720 Settings in the menu 4 4 Application settings APPLICATION 4 4 1 COUNT settings for counting VAr SPL Adaptation of the reference quantity ON e The reference quantity can be changed in operating mode OFF e Counting only with defined
26. at SINGLE Add LF e Line feed O 9 35 User manual ME Nr 22017145 01 07 FTC BA e 0720 Settings in the menu 4 6 3 COM1 COM2 gt PARAMET communication parameter bAUd Selecting baud rate 300 600 1200 2400 4800 9600 19200 38400 baud PAritY Selecting parity 7 none 8 none 7 odd 8 odd 7 even 8 even H SHAKE Selecting Handshake NO XONXOFF NET Addr not documented ChECSuM Activating checksum byte appears only in TOLEDO Continuous mode Vcc not documented 4 6 4 COM1 COM2 gt RESET COMT RESET COM2 resetting serial interface to factory settings SUrE Confirmation inquiry e Reset interface settings to factory settings with e Do not reset the interface settings with 4 6 5 OPTION configuring options If no option is installed or is not yet configured N A appears in the display EtH NEt Configuration of the Ethernet interface IP AddrS Enter IP address SUBNEt e Enter Subnet address GACEWAY e Enter Gateway address USb not documented USb TEST diGitAL not documented IN 1 4 OFF ZErO tArE PriNt rEF 10 rEF n SCALE UNIt User manual ME Nr 22017145 01 07 36 OUT 1 4 OFF StAbLE bEL MIN AbV MIN UNdErLd OVErLd StAr ANALOG Configuration of the analog second scale interface Mode Operating mode of the second scale rEF e Second scale can only be used to determine the average piece weight bULK e Second scale can on
27. atic reference optimization rEF OPt gt AUtO must be set in the menu for this The symbol Auto Opt appears in the display 1 Place the reference parts on the scale and press or ern 2 Place additional reference parts max the same number as for the first reference determination on the scale The scale automatically optimises the average piece weight with the larger number of reference parts The rest of the counting process is as described earlier The reference optimization can be performed several times Counting with automatic reference determination Prerequisite A SMPL ON is set in the menu Place the number of pieces displayed above the key into the container The scale automatically determines the average piece weight and then shows the quantity The rest of the counting process is as described earlier Counting with two scales For piece counting it is possible to connect a second scale or weighing platform e g a floor scale for counting a large number of pieces via the optional analog second scale interface The necessary settings for the application and interface parameters are described in ihe Sections 4 4 1 4 6 1 and 4 6 4 User manual ME Nr 22017145 01 07 20 Counting FTC BA e 0720 21 3 7 1 Note 3 7 2 Note Counting with a reference scale Prerequisite The connected second scale is configured as reference scale l Place the reference parts on the reference scale an
28. ay ideally a 0 Og after which the scale changes to the next menu item KboArd o C N KboArd PUSH 1 10 Keyboard test e First press the large keys on the bottom row in order TARE PRINT e Then press the smaller keys in the top row a amp 8 4 If the key works the scale changes to the next key Note You cannot abort the keyboard test If you have selected the menu item KboArd you must press all keys dISPLAY Display test The scale displays all functioning segments SNr Display of the serial number SNr2 Display of the serial number of scale 2 This menu item only appears if an analog second scale is connected List Printout of a list of all menu settings List2 Printout of a list of all menu settings of scale 2 This menu item only appears if an analog second scale is connected rESEt AL SUrE Resetting all menu settings to the factory settings Confirmation inquiry e Reset all menu settings to the factory settings with e Do not reset the menu settings with Interface description FTC BA e 0720 5 Interface description 5 1 5105 interface commands The compact scales FTC support the command set MT SICS METTLER TOLEDO Standard Interface Command Set With SICS commands it is possible to configure query and operate the scales from a PC SICS commands are divided up into various levels 5 1 1 Available SICS commands
29. chi contrassegnati con la marcatura CE English We hereby declare that the product to which this declaration refers conforms with the following standards Deutsch Wir erkl ren hiermit da das Produkt auf das sich diese Erkl rung bezieht mit den nachstehenden Normen bereinstimmt Francais Nous d clarons avec cela responsabilit que le produit auquel se rapporte la pr sente d claration est conforme aux normes cit es ci apr s Espanol Manifestamos en la presente que el producto al que se refiere esta declaraci n est ade acuerdo con las normas siguientes Italiano Dichiariamo con ci che il prodotto al quale la presente dichiarazione si riferisce conforme alle norme di seguito citate Electronic Scale KERN FTC Mark applied EU Directive Standards Approval Test certificate N 73 23EEC Low EN61010 1 C Voltage 89 336EEC EMC EN55022 Emission KI B C EN61000 3 2 EN61000 3 3 EN50082 1 90 384EEC Non EN45501 1 T6179 1 C automatic year code M 1 weighing Instruments 1 1 applies only to certified balances gilt nur f r geeichte Waagen valable uniquement pour les balances v rifi es s lo aplicable a balanzas verficadas la dichiarazione vale solo per le bilance omologate Date 24 01 2007 Signature Gottl KERN amp Sohn GmbH Management Gottl KERN amp Sohn GmbH Ziegelei 1 D 72336 Balingen Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Notice Certifie
30. d balances and balances used for legal applications have the EU type approval The year of the initial verification is shown next to the CE mark Such balances are verified in the factory and carry the M mark on the actual balance and the packaging The year of initial verification is shown next to the CE mark The GEO value of verified balances explains for which location of use the balance has been verified This GEO value is shown on the balance itself and on the packing Further details see GEO value table Hinweise F r geeichte eichpflichtige Waagen liegt eine EU Bauartzulassung vor Das Jahr der ersten Eichung ist neben dem CE Zeichen aufgef hrt Solche Waagen sind ab Werk geeicht und tragen die Kennzeichnung M auf dem Ger t selbst und auf der Verpackung Der GEO Wert gibt bei vom Hersteller geeichten Waagen an f r welchen Aufstellungsort die Waage geeicht ist Dieser GEO Wert befindet sich auf der Waage sowie der Verpackung Genaueres ist der GEO Wert Tabelle zu entnehmen Remarques Les balances v rifi es admissibles la v rification font l objet d une approbation de mod le UE L ann e de la v rification primitive est indiqu s c t de la marque CE Ces balances sont v rifi es d origine et portent la marque M sur l appareil lui m me et sur l emballage Le valeur GEO indique le lieu d utilisation pour lequel la balance t v rifi e Ce valeur GEO se trouve sur la balance ainsi que sur l emballage Veui
31. d press or er The scale determines the average piece weight and changes to the display in pieces PCS Place the parts to be counted on the first scale The total quantity is displayed If tOtAL Ct gt bULK is set in the menu only the number of pieces on the bulk scale is displayed If tOtAL Ct gt botH is set in the menu the reference quantity is added to the bulk quantity Counting with a bulk scale Prerequisite The connected second scale is configured as bulk scale l Place the reference parts on the first scale and press or Rern The scale determines the average piece weight and changes to the display in pieces PCS Place the parts to be counted on the bulk scale The total quantity is displayed If tOtAL Ct gt bULK is set in the menu only the number of pieces on the bulk scale is displayed on the bulk scale If tOtAL Ct gt botH is set in the menu the reference quantity is added to the bulk quantity User manual ME Nr 22017145 01 07 FTC BA e 0720 Counting 3 7 3 Counting with an auxiliary scale Note This configuration allows counting of diverse parts for example very small parts on one scale and large parts on the other scale Prerequisite The connected second scale is configured as an auxiliary scale The scale doesn t change automatically but only after pressing the key 1 Activate the appropriate scale 2 Place the reference parts on this scale and press or er
32. der to alter the GEO value The next value is dis played every time the key is pressed adjustment range O 31 See the GEO value chart in chapter 7 1 1 for the appropriate value Confirm the selected GEO value using the key Please note The GEO value may not be readjusted following this adjustment by means of GEO value as this would cause the set adjusting values to become invalid Fall b Sie sind mit den GEO Werten NICHT vertraut Die Justierung muss in diesem Fall mit Justiergewicht erfolgen siehe Punkt 8 You are NOT familiar with the GEO values In this case adjustment must be made using adjusting weights see item 9 Taste dr cken User manual ME Nr 22017145 01 07 FTC BA e 0720 Settings in the menu 7 Display LIN CAL Press are key 8 Display CAL Press zc e Press the key The scale determines the zero point preload appears in the display The calibration weight to be placed on the scale then flashes in the display If necessary change the weight value displayed with Place the calibration weight on the scale and confirm with The scale calibrates with the calibration weight loaded After calibration is completed done appeats briefly in the display Return to weighing mode Press the key END will appear on the display Press the key order to save the alterations Save will appear on the display Press the key to confirm Following this the balance will return to weighing mode
33. for European countries Table GEO VALUES 6000e 7500e shows the geo values for different gravitation ZONES GEO VALUES 3000e OIML Class III European Countries Geographical latitude Geo value Country 46 22 49 01 18 Austria 49 30 51 30 21 Belgium 41 41 44 13 16 Bulgaria 42 24 46 32 18 Croatia 48 34 51 03 20 Czechia 54 34 57 45 23 Denmark 57 30 59 40 24 Estonia 59 48 64 00 20 Finland 64 00 70 05 26 41 20 45 00 17 France 45 00 51 00 19 47 00 55 00 20 Germany 34 48 41 45 15 Greece 45 45 48 35 19 Hungary 63 17 67 09 26 Iceland 51 05 55 05 22 Ireland 35 47 47 05 17 Italy 55 30 58 04 23 Latvia 47 03 47 14 18 Liechtenstein 53 54 56 24 22 Lithuiania 49 27 50 11 20 Luxemburg 50 46 53 32 2 Netherlands 57 57 64 00 24 Norway 64 00 71 11 26 User manual ME Nr 22017145 01 07 50 Appendix FTC BA e 0720 51 Geographical latitude Geo value Country 49 00 54 30 21 Poland 36 58 42 10 15 Portugal 43 37 48 15 18 Romania 47 44 49 46 19 Slovakia 45 26 46 35 18 Slovenia 36 00 43 47 15 Spain 55 20 62 00 24 Sweden 62 00 69 04 26 45 49 47 49 18 Switzerland 35 51 42 06 16 Turkey 49 00 55 00 21
34. ghting is switched off and that no peripheral devices are connected The device automatically switches to battery operation as soon as the mains supply is interrupted When the mains supply is restored the device automatically switches back to mains operation The battery symbol indicates the present charging level of the battery segment cor responds to approx 25 96 capacity When the symbol flashes the battery must be charged min 4 hours The charging period is extended if work is continued during charging The battery is protected against overcharging The battery s charging capacity can be reduced under continuous mains operation gt To maintain the charging capacity after a maximum of 4 weeks discharge the bat tery completely before recharging it Monitoring the test substances The metrology features of the balance and any possible available adjusting weight must be checked at regular intervals within the scope of quality assurance For this purpose the answerable user must define a suitable interval as well as the nature and scope of this check Information is available on KERN s home page www kern sohn com with regard to the monitoring of balance test substances and the test weights required for this Test weights and balances can be adjusted quickly and at a reasonable price in KERN s accredited DKD calibration laboratory return to national normal User manual ME Nr 22017145 01 07 11 Introduction FTC BA e 0720
35. ht value SU Send stable weight value in the current unit TST2 Start test function with external weight TST3 Start test function with internal weight LEVEL 3 PW Average piece weight LEVEL SPECIAL 13 Header for the printout ICP Send configuration of the printout LST Send menu settings MOI Weighing mode MO2 Stability setting MO3 Autozero function MI9 Send calibration weight M21 Inquire set weight unit P Print text P130 Weight value unit and price PCS Number of pieces PRN Print out at every printer interface REF Average piece weight RST Restart SFIR Send weight value immediately and repeat quickly SIH Send weight value immediately in high resolution SWU Switch weight unit SX Send stable data record SXI Send data record immediately SXIR Send data record immediately and repeat U Switch weight unit User manual ME Nr 22017145 01 07 40 Interface description FTC BA e 0720 41 Requirements for communication between scale and PC The scale must be connected to the RS232 or Ethernet interface of a PC with a suit able cable The interface of the scale must be set to Dialog mode see Section 4 6 1 A terminal progam must be available on the PC e g HyperTerminal The communication parameters baud rate and parity must be set in the terminal program and on the scale to the same values see Section 4 6 3 User manual ME Nr 22017145 01 07 FTC BA e 0720 Interface description 5 2 TOLEDO Continuous mode
36. ieces for the parts on the bulk scale only both e Display number of pieces for all parts on the bulk and the reference scale 4 4 2 AVERAGE determining the average weight for an unstable load OFF Calculating average weight switched off AUtO Calculating average weight with automatic start of the weighing cycle MAnuAL Calculating average weight with manual start of the weighing cycle via 4 4 3 RESET resetting application settings to factory settings SUrE Confirmation inquiry e Reset the application settings to factory settings with e Do not reset the application settings with 33 User manual ME Nr 22017145 01 07 FTC BA e 0720 Settings in the menu 4 5 4 5 1 Terminal settings TERMINAL DEVICE Sleep mode energy saving mode and display backlighting SLEEP This menu item only appears on devices in mains operation When SLEEP is activated the scale switches off display and backlighting after the time period set when not in use The display and backlighting are switched on again at the press of a key or if the weight changes Possible settings OFF 1 min 3 min 5 min PWr OFF This menu item only appears on devices in battery operation When Pwr OFF is activated the device switches itself off automatically after approx 3 minutes when not in use b LIGHt Switching the display backlighting on off On scales with a battery the background lighting switches itself off automatically if there has been no ac
37. inted out or sent to a computer Press The display contents are printed out and transferred to the computer See Section 8 2 for sample protocols Switching scales If a second scale or a weighing platform is connected e g via the optional analog second scale interface the currently active scale is shown in the display The second scale can be operated in exactly the same way as the first scale Press The display changes from one scale to the other User manual ME Nr 2201 7145 01 07 16 Operation FTC BA e 0720 AN AN 2 11 Cleaning CAUTION Electric shock hazard 4 Before cleaning with a damp cloth pull out the mains plug to disconnect the unit from the power supply CAUTION When the weighing pan has been removed never clean the area under the load plate holder with a solid object This could damage the weighing cell Other cleaning information Use damp cloths Do not use any acids alkalis or strong solvents Do not clean using a high pressure cleaning unit or under running water If very dirty remove the weighing pan protective cover if present and adjustable feet and clean these items separately Follow all the relevant instructions regarding cleaning intervals and permissible cleaning agents User manual ME Nr 22017145 01 07 FTC BA e 0720 Counting 3 Counting 3 1 Note The compact scales FTC have additional functions for piece counting The relevant set
38. llez trouver plus de d tails dans le tableau GEO Notas Las balanzas verificadas verificables cuentan con una aprobaci n de modelo UE El a o de la primera verificaci n est indicado al lado del distintivo CE Estas balanzas est n verificadas en f brica y llevan la designaci n M sobre el propio aparato y sobre el embalaje El valor GEO indica el lugar de ubicaci n por lo cual la balanza est verificado El valor se encuentra sobre la balanza as como sobre el embalaje Por favor toman demBas detalles de la tabla GEO Avvertenza Per le bilance approvate esiste un approvazione CE del tipo L anno della prima verifica 6 indicato a fianco della marcatura CE tipi marcati con un contrassegno M su sfondo verde peino possono essere impiegati da subito Il coefficiente GEO di bilance omolgate indca per quale luogo la bilancia 6 stata omologata Questo coefficiente GEO si trova sulla bilancia e sull imballo Ulteriroi informazioni vedi tabella coefficente GEO GEO WERT Tabelle GEO value table geographische H he ber Meer in Metern altitude Breite geo hical latitude 650 1300 1300 1950 1950 2600 2600 3250 or of 9552 4 5 3 4 X 2 3 X 7 2 J0O f 9 52 15 65 6 4 5 3 4 9X 2 3 12 15 ef 19 2 6 7 s e a5 X J 3 4 X 2 3 19 2 22 22 7 8 6 7 15 6 Jar ara 22 22 25 21 8 9 7 8 ei X s 6 ars 25 21 28s 6 9 10 8 9 amp 7 8 Ier J 5 6 28 6 4 10 11 9 10 8 9
39. ly be used as bulk scale AuXILIA e No difference between reference and bulk scale all functions available on the scale selected BYPASS e Second scale interface not assigned 4 6 6 DEF PRN configuring templates tEMPLt1 tEMPLt2 Selecting Template or Template 2 LINE 1 20 Select line NOt USEd e Line not used HEAdEr e Line as header The contents of the header must be defined via an interface com mand see Section 5 1 SCALE NO e Scale number GROSS e Gross weight tArE e Tare weight nEt e Net weight APW e Average piece weight rEF Ct e Reference quantity PCS e Pieces StARLN e Line with CrLF e Line feed blank line F FEEd e Page feed 4 7 Diagnosis and printing out of the menu settings DIAGNOS tESt SC Internal External Testing scale with internal calibration weight e Int CAL appears in the display during the test e After completion of the test ideally d 0 Og briefly appears in the display after which the scale changes to the next menu item KboArd Testing scale with external calibration weight 1 The scale checks the zero point 0 appears in the display The test weight flashes in the display If necessary change the weight value displayed with rare Put the calibration weight on the scale and confirm with rm The scale checks the calibration weight put on them After the test is completed the deviation from the last calibration briefly appears in the displ
40. mm high with back lighting Keypad e Pressure point membrane keypad e Scratch proof labeling Housing e Diecast aluminum housing chromium nickel steel weighing pan e Dimensions see Page 48 User manual ME Nr 22017145 01 07 46 Technical data and accessories FTC BA e 0720 FTC Protection Class IEC 529 DIN 40050 EN60529 IP43 not with Ethernet interface Resolution of the analog second scale interface 300000 points in noncertified configuration e 7000 points in certified configuration Supply of the weighing cell 8 2V 7 1 3 47 Weighing ranges and readability The compact scales FTC with strain gauge weighing cells are supplied in the config uration 2 x 3000 d Higher legibilities are available from the factory with the optional Premium weighing cells Capacity Configuration 2 x 3000 d standard 1x 6000 d with optional Premium weighing cells Weighing Readability Weighing Readability ranges certified range certified 3kg 1 5 kg 3 kg 0 5g lg 3 kg 0 5g 6 kg 3 kg 6 kg 1g 2g 6 kg lg 15 kg 6 kg 15 kg 2g 5g 15 kg 2g 35 kg 1bkg 35kg 5g 10g 35 kg 5g 60 kg 30 kg 60 kg 10g 20g 60 kg 10g User manual ME Nr 22017145 01 07 FTC BA e 0720 Technical data and accessories 7 1 4 Dimensions
41. on via MT SICS commands control of the scale via PC CONt OLd As per CONCINU see above but with 2 fixed blanks in front of the unit compatible with Spider 1 2 3 dIAL OLd As per dIALOG see above but with 2 fixed blanks in front of the unit compatible with Spider 1 2 3 dt b DigiTOL compatible format GROSS e Transfer of the gross weight identified with G tArE e Transfer of the tare weight nEt e Transfer of the net weight dt G As per dt b see above gross weight identified with G cont wt TOLEDO Continuous mode COnt Ct TOLEDO Continuous mode transfer of the number of pieces 2nd dISP For connecting a second display automatically activates the 5 V voltage supply at Pin 9 rEF Data transfer from the reference scale automatic switchover bULK Data transfer from the quantity scale automatic switchover AuXILIA Data transfer from the reference or quantity scale manual switchover 4 6 2 COM1 COM2 gt PRINTER settings for protocol printout This menu item only appears if the mode Print or A Print is selected tEmPLat Selecting protocol printout StdArd e Standard printout tEmPLt1 e Printout in accordance with Template 1 tEmPLt2 e Printout in accordance with Template 2 ASCi FmtT Selecting formats for the protocol printout LINE Fmt e Line format MULtI multi line or SINGLE single line LENGtH e Line length O 100 characters appears only with line format MULCI SEPArAt e Separator _ and space appears only with line form
42. play e Incorrect zeroing e Incorrect tare value e Contact between weighing pan and or weighing sample and surroundings e Scale tilted gt Unload scale set to zero and repeat weighing operation gt Clear tare gt Remedy contact gt Level scale 45 User manual ME Nr 22017145 01 07 FTC BA e 0720 Technical data and accessories 7 Technical data and accessories 7 1 Technical data 7 1 1 Type key The compact scales FTC are available with various capacities and platforms that can be seen from the complete type designation 7 1 2 General data FTC Applications e Weighing e Dynamic weighing e Counting with fixed or variable reference quantity e Counting with reference and bulk scale Settings e Resolution selectable e Weighing unit selectable g kg oz Ib t e Taring function manual automatic chain tare e Automatic zero point correction when the scale is switched on and during oper ation e Filter for adapting to the ambient conditions vibration adapter e Filter for adapting to the weighing type e g dispensing weighing process adapter e Switch off function sleep mode for mains operated devices energy saving mode for battery operation e Display lighting e Add mode for determining the piece weight when counting e Reference optimization e Graphic display of the weighing range Accuracy class OIML NTEP Ill Display e LCD liquid crystal display digits 16
43. r Supervisor access to the menu and you have for gotten it you can still enter the menu Press 3 times and confirm with ern User manual ME Nr 22017145 01 07 FTC BA e 0720 Settings in the menu 4 1 2 Selecting and setting parameters Scrolling on one level Scroll forward Press rare Scroll back Press Goe Activating menu items gt Press rnr accepting selection Exiting menu 1 Press The last menu item END appears 2 Press ferr The inquiry SAVE appears 3 Confirm inquiry with to save the settings and return to weighing mode or gt Press to discard changes and return to weighing mode User manual ME Nr 22017145 01 07 24 Settings in the menu FTC BA e 0720 4 2 Overview Level 1 Level 2 Level 3 Level 4 Level 5 Level 6 Page CAL 28 dISPLAY UNIt1 g kg oz lb t 30 UNIt2 g kg oz 1b t rESOLU UNt rOLL ON OFF tArE A tArE ON OFF 30 ChAIn tr ON OFF A CL tr ON OFF ZErO AZM OFF 0 5 d 1d 2d 5d 10 d 30 rEStArt ON OFF 3l FILtEr VibrAt LOW MEd HIGH 3l PrOCESS UNIVEr dOSING StABILI FASt StAndrd PrECISE rESEt SUrE 31 APPLIC COUNt VAr SPL ON OFF 32 SPL qtY Sql Sq5 Min reFW OFF 97 5 99 0 99 5 rEF OPt OFF AUtO A SMPL ON OFF A CL APW ON OFF ACCurCY ON OFF tOtAL Ct bULK bOth AVErAGE OFF AUtO MAnuAL 33 rESEt SU
44. rE 33 tERMINL dEVICE SLEEP OFF 1 min 3 min 5 min 34 PWr OFF YES NO b LIGHt ON OFF ACCESS SUPErVI 34 rESEt SUrE 34 25 User manual ME Nr 22017145 01 07 FTC BA e 0720 Settings in the menu Level 1 Level 2 Level 3 Level 4 Level 5 Level 6 Page COMMUNI COM 1 COM 2 MOdE Print 35 A Print CONtINU dIALOG CONt OLd dIAL OLd dt b GrOSS ON OFF tArE ON OFF nEt ON OFF dt G GrOSS ON OFF tArE ON OFF nEt ON OFF COnt Wt COnt Ct 2nd dISP rEF bULK AuXILIA PriNtEr tEmPLat StdArd tEMPLt1 35 tEMPLt2 ASCi Fmt LINE FMt MULtI SIN GLE LENGtH 14 100 SEPArAt KE Add LF 0 9 PArAMEt bAUd 300 38400 36 PAritY 7 nonE 8 nonE 7 odd 8 odd 7 EVEN 8 EVEN H SHAKE NO XONXOFF nEt 422 nEt 485 NEt Addr Oi ST ChECSuM ON OFF Vcc ON OFF rSt COMx SUrE 36 User manual ME Nr 22017145 01 07 26 Settings in the menu FTC BA e 0720 Level 1 Level 2 Level 3 Level 4 Level 5 Level 6 Page COMMUNI OPtION EtH NEt IP AddrS SUbNEt GAtEWAY 36 USb USb tESt 36 diGitAL IN 1 4 OFF ZErO tArE Print 36 rEF 10 rEF n SCALE Unit OUT 1 4 OFF StAbLE bEL Min AbV Min UndErLd OVErLd StAr ANALOG Mode rEF bULK AuXILIA 36 bYPASS dEF PrN tEMPLt1 LINE1 NOt USEd HEAdEr 37 tEMPLt2 LINE 20 SCALE NO GrOSS tArE nEt APW
45. times if for example cardboard is placed between individual layers in a container 1 Place the first container or packaging material on the scale and press rare The packaging weight is automatically saved as the tare weight the zero display and the symbol NET appear 2 Weigh the weighing sample and read print out the result 3 Place the second container or packaging material on the scale and press again The total weight on the scale is saved as the new tare weight The zero display appears 4 Weigh the weighing sample in the second container and read print the result b Repeat the last two steps for other containers User manual ME Nr 22017145 01 07 14 Operation FTC BA e 0720 2 5 2 6 2 7 l I En EN N 2 8 With manual start 100 Calling up the gross weight and tare value 1 Press 28 The gross weight is displayed for 2 seconds 2 Press again while the gross weight is displayed The tare value is displayed After a few seconds the scale changes back to the net weight display Displaying weight values with a higher resolution Press a The current weight value is displayed for 2 seconds in a higher resolution The scale then changes back to the normal resolution Displaying the capacity available The scale has a graphic display of the scale capacity available The bar indicates how many per cent of the scale capacity is already occupied and what capacity
46. tivity on the scale for 5 seconds Note This menu item is accessible without a Supervisor password 4 5 2 ACCESS password for Supervisor menu access SUPErVI ENtER C rEtYPE C Password entry for Supervisor menu access Request to enter password Enter the password and confirm with Request to repeat the password entry gt Enter the password again and confirm with Notes e The key may only be used in combination with another key e If you enter an impermissible code or make a typing error in the repetition e The password can consist of up to 4 characters e The key must not be part of the password It is required for confirming the password COdE Err appears in the display 4 5 3 RESET resetting terminal settings to the factory settings SUrE Confirmation inquiry e Reset terminal settings to the factory settings with e Do not reset the terminal settings with User manual ME Nr 22017145 01 07 34 Settings in the menu FTC BA e 0720 4 6 Configuring interfaces COMMUNICATION 4 6 1 COMI COM 2 gt MODE operating mode of the serial interface Print Manual data output to the printer with A Print Automatic output of stable results to the printer e g for series weighing operations CONtINU Ongoing output of all weight values via the interface dIALOG Bi directional communicati