        Agilent 1290 Infinity Thermostat - User Manual
         Contents
1.                                                                                     Waste tube in leak channel    Figure 10 Hydraulic Connections    1290 Infinity Thermostat User Manual    Installing the Thermostat 3    Stage 4  Installing the Sample Tray    Load the sample tray with sample vials as required     2 Slide the sample tray into the autosampler so that the rear of the  sample tray is seated firmly against the rear of the sample tray area     3 Press the front of the sample tray down to secure the tray in the  autosampler  If the tray pops out of its position the air channel adapter  is not inserted correctly        Figure 11 Installing the Sample Tray  Half Tray Combinations    In the autosampler only the 100 vial tray is supported for temperature control of the vials   Nevertheless the half trays of the standard autosampler  G1329A B  can be used in the  thermostatted autosampler as well  However when these trays are installed cooling or  heating of the vials in the tray will not work     Half trays can be installed in any combination enabling both 1 8 ml and  6 ml vials to be used simultaneously     1290 Infinity Thermostat User Manual 29    3 Installing the Thermostat    Numbering of Vial Positions    The standard 100 vial tray has vial positions 1 to 100  However  when  using two half trays  the numbering convention is slightly different  The  vial positions of the right hand half tray begin at position 101 as follows     Left  hand 40  position tray  1 40  Le
2.             p n 03396 61010 Pin 33XX Pin Agilent Signal Name Active  module  TTL          9 1   White Digital ground  NC 2   Brown Prepare run Low  8o    3 3   Gray Start Low  2 E NC 4   Blue Shut down Low  e  n NC 5   Pink Not  i connected  NC 6   Yellow Power on High  14 7   Red Ready High  4 8   Green Stop Low  NC 9   Black Start request Low  13  15 Not  connected       1290 Infinity Thermostat User Manual    73    8 Cable Identification    Agilent Module to Agilent 35900 A D Converters                                                                               p n 5061 3378 Pin 35900 Pin Agilent Signal Name Active  A D module  TTL   1   White 1   White Digital ground  2   Brown 2   Brown Prepare run Low  3   Gray 3   Gray Start Low  4         4   Blue Shut down Low  5   Pink 5   Pink Not  connected  6   Yellow 6   Yellow Power on High  7   Red 7   Red Ready High  8   Green 8   Green Stop Low  9   Black 9   Black Start request Low  Agilent Module to General Purpose  p n 01046 60201 Wire Color Pin Agilent Signal Name Active  module  TTL   White 1 Digital ground  A    1  Dn Brown 2 Prepare run Low  00  em OB KEY Gray 3 Start Low  oo  o0           Blue 4 Shut down Low  HE Pink 5 Not       connected  S BE 15     Yellow 6 Power on High  Red 7 Ready High  Green 8 Stop Low  Black 9 Start request Low       74 1290 Infinity Thermostat User Manual    BCD Cables                         Cable Identification 8    One end of these cables provides a 15  pin BCD connector to be connecte
3.    e Agilent Technologies       Notices       Agilent Technologies  Inc  2011 2014   2015    No part of this manual may be reproduced  in any form or by any means  including elec   tronic storage and retrieval or translation  into a foreign language  without prior agree   ment and written consent from Agilent  Technologies  Inc  as governed by United  States and international copyright laws     Manual Part Number  61330 90014 Rev  B    Edition  05 2015  Printed in Germany    Agilent Technologies  Hewlett Packard Strasse 8  76337 Waldbronn    Warranty    The material contained in this docu   ment is provided  as is   and is sub   ject to being changed  without notice   in future editions  Further  to the max   imum extent permitted by applicable  law  Agilent disclaims all warranties   either express or implied  with regard  to this manual and any information  contained herein  including but not  limited to the implied warranties of  merchantability and fitness for a par   ticular purpose  Agilent shall not be  liable for errors or for incidental or  consequential damages in connection  with the furnishing  use  or perfor   mance of this document or of any  information contained herein  Should  Agilent and the user have a separate  written agreement with warranty  terms covering the material in this  document that conflict with these  terms  the warranty terms in the sep   arate agreement shall control     Technology Licenses    The hardware and or software described in  this 
4.   CN  China    Alphabetic character A Z  used by manufacturing     Alpha numeric code 0 9  A Z  where each combination  unambiguously denotes a module  there can be more than one  code for the same module     Serial number    83    9 Hardware Information  Electrical Connections    Rear View of the Module           Relay contacts       CAN cable to next module    Vial number output       Therm Autosampler cable        Remote    RS232C  CAN bus          Thermostat    Autosampler cable                            Figure 17 Electrical Connections    84 1290 Infinity Thermostat User Manual    Hardware Information 9    Early Maintenance Feedback    Maintenance requires the exchange of components which are subject to  wear or stress  Ideally  the frequency at which components are exchanged  should be based on the intensity of usage of the module and the analytical  conditions  and not on a predefined time interval  The early maintenance  feedback  EMF  feature monitors the usage of specific components in the  instrument  and provides feedback when the user selectable limits have  been exceeded  The visual feedback in the user interface provides an  indication that maintenance procedures should be scheduled     EMF Counters    EMF counters increment with use and can be assigned a maximum limit  which provides visual feedback in the user interface when the limit is  exceeded  Some counters can be reset to zero after the required  maintenance procedure     Using the EMF Counters    The 
5.   Installing the Thermostat    Damaged Packaging 18  Delivery Checklist 18    Optimizing the Stack Configuration 19    Installing the Thermostat 21    Stage 1  Preparing the Thermostat and Autosampler  Fraction  Collector 22    Stage 2  Power Cable and Interface Cable Connection 25  Stage 3  Flow Connections 27   Stage 4  Installing the Sample Tray 29   Stage 5  Installing Tray Cover and Front Cover 31   Stage 6  Turning on the Thermostatted Autosampler  Fraction  collector 31    Transporting the Thermostatted Autosampler or Fraction Collector 32    RE Agilent Technologies    3 Installing the Thermostat    Damaged Packaging    CAUTION    If the delivery packaging shows signs of external damage  please call your  Agilent Technologies sales and service office immediately  Inform your  service representative that the instrument may have been damaged during  shipment      Defective on arrival  problems    If there are signs of damage  please do not attempt to install the module  Inspection by  Agilent is required to evaluate if the instrument is in good condition or damaged      gt  Notify your Agilent sales and service office about the damage      gt  An Agilent service representative will inspect the instrument at your site and  initiate appropriate actions        Delivery Checklist    18    Unpack the module  Ensure all parts and materials have been delivered   The delivery checklist is shown in Table 3 on page 18  Please report    missing or damaged parts to your local Agilen
6.   Sound Pressure Lp  lt  70 dB         At Operator Position    Normal Operation    e According to ISO 7779 1988 EN 27779 1991  Type Test     1290 Infinity Thermostat User Manual 93    10 Appendix    Agilent Technologies on Internet    For the latest information on products and services visit our worldwide  web site on the Internet at     http   www agilent com    94 1290 Infinity Thermostat User Manual    Index    A    Agilent Lab Advisor software 34  Agilent Lab Advisor 34   Agilent  oninternet 94   ALS thermostat parts and materials  accessory kit 63   main assemblies 62  63       ALS thermostat repairs   power supplyfuses 58  ambient non operating temperature 15  ambient operating temperature 15    analog  cable 70  Autosampler Thermostat Operation 8  auxiliary  cable 77  B  BCD  cable 75    benchspace 14    C   cable  analog 70  auxiliary 77  BCD 75  CAN 78  external contact 79        78  remote 72  RS 232 80   cables    1290 Infinity Thermostat User Manual    analog 68   BCD 68   CAN 69   external contact 69   LAN 69   overview 68   remote 68   RS 232 69  CAN   cable 78  cautions and warnings 54  cleaning 57  compensation sensor open 44  compensation sensor short 45  condensation 6  14  14    D    defect on arrival 18  delay volume 19  delivery checklist 18  18  dimensions 15    E    electrical connections  descriptions of 82  electronic waste 91  EMF  early maintenance feedback 85  environment 12  14  error messages  ALS thermostat 48  bad cooling   heating  performance
7.  41  Timeout 41    Shutdown 42   Remote Timeout 42   Lost CAN Partner 43   Leak Sensor Short 43   Leak Sensor Open 44  Compensation Sensor Open 44  Compensation Sensor Short 45  Fan Failed 45   Leak 46   Open Cover 46   Cover Violation 47    Thermostat Error Messages 48  Fan Failed 48  Temperature Control Failed  1 4  49  Temperature Sensor Failed  1 4  49  Lost Contact to Autosampler Thermostat 50  Power Fail for Autosampler Thermostat Module 50  Temperature out of Range 51  Bad Cooling   Heating Performance 52    This chapter describes the meaning of error messages  and provides  information on probable causes and suggested actions how to recover from  error conditions     oh Agilent Technologies 39    5    Error Information    What Are Error Messages    40    Error messages are displayed in the user interface when an electronic   mechanical  or hydraulic  flow path  failure occurs which requires  attention before the analysis can be continued  for example  repair  or  exchange of consumables is necessary   In the event of such a failure  the  red status indicator at the front of the module is switched on  and an  entry is written into the module logbook     If an error occurs outside a method run  other modules will not be  informed about this error  If it occurs within a method run  all connected  modules will get a notification  all LEDs get red and the run will be  stopped  Depending on the module type  this stop is implemented  differently  For example  for a pump the 
8.  52  compensation sensor open 44    Index    compensation sensor short 45  fan failed 45  48   ignition without cover 46  46  leak sensor open 44   leak sensor short 43   leak 46   lost CAN partner 43   lost contact to ALS thermostat 50  power fail ALS thermostat 50  remote timeout 42  shutdown 42   temperature control failed 49  temperature out of range 51  temperature sensor failed 49  timeout 41       external contact    cable 79  F  failure 35  fan failed 45  flow path 19  frequency range 15  fuses 12  58  G  general error messages 41  H  half tray combinations 29  humidity 15       injection          6    95    Index    installation    thermostatted autosampler 21    installation    benchspace 14    installing the thermostat    preparation 22       installing the thermostatted autosampler    flow connections 27   power cable and interface cable 25  sampletray 29   tray cover and front cover 31  turning on the instrument 31    internet 94  introduction to the thermostat 6    L  LAN   cable 78  leak sensor open 44    leak sensor short 43    le    ak 46    line frequency 15  line voltage 15  lost CAN partner 43    M    m    3       3    aintenance  definition of 56  feedback 85  essage   ignition without cover 46  46  remote timeout 42   issing parts 18    ulti draw option 6    Non operating altitude 15    non operating temperature 15    96    0    operating Altitude 15  operating temperature 15    optimizing the stack configuration 19     19    P    packaging   damaged 18 
9.  BNC Connector  p n 8120 1840 Pin BNC Pin Agilent Signal Name  module  Shield Shield Analog    el   Center Center Analog            1290 Infinity Thermostat User Manual    Agilent Module to General Purpose    Cable Identification                      p n 01046 60105 Pin Agilent Signal Name  module  Not connected  Black Analog    p  Red Analog        we                1290 Infinity Thermostat User Manual    71    Cable Identification    Remote Cables    72                One end of these cables provides a Agilent Technologies APG  Analytical  Products Group  remote connector to be connected to Agilent modules   The other end depends on the instrument to be connected to     Agilent Module to 3396A Integrators                                        p n 03394 60600 Pin 3396A Pin Agilent Signal Name Active  module  TTL             9 1  White Digital ground  NC 2   Brown Prepare run Low  8o  A 3 3   Gray Start Low  z    NC 4   Blue Shut down Low  e  Be NC 5   Pink Not  Ls connected  NC 6   Yellow Power on High  5 14 7   Red Ready High  1 8   Green Stop Low  NC 9   Black Start request Low  13  15 Not  connected       Agilent Module to 3396 Series II   3395A Integrators    Use the cable Agilent module to 3396A Series I integrators  03394 60600   and cut pin  5 on the integrator side  Otherwise the integrator prints    START  not ready     1290 Infinity Thermostat User Manual    Cable Identification    Agilent Module to 3396 Series      3395B Integrators                               
10.  BO  EEE   22 Primary fuses          Pull the fuse holder out of the socket    Remove the fuse from the fuse holder    Insert a new fuse in the fuse holder    Reinsert the fuse holder and fix with the screwdriver   Reinsert the power cables     Switch ON the power switch     1290 Infinity Thermostat User Manual    59    6 Maintenance  Exchanging the Power Supply Fuses    60 1290 Infinity Thermostat User Manual    1290 Infinity Thermostat User Manual    7  Parts for Maintenance    Main Assemblies  External Parts  62  Accessory Kit 63   HPLC System Tool Kit 64   Plastic Parts 65    RE Agilent Technologies    61    7  Partsfor Maintenance  Main Assemblies  External Parts     Main Assemblies  External Parts        Figure 15 Main Assemblies    ltem p n   1 2110 0015  2 2110 0029  3 5065 9982    G1330 81600    62    Description   Fuses T2 5 A 250V  CSA  UL listed    Fuse TCA   Board  T3 A 250 V  CSA  UL listed   Plastics kit  includes base  top  left and right sides     Cable  Autosampler  Fraction Collector to Thermostat    1290 Infinity Thermostat User Manual    Parts for Maintenance 7    Accessory Kit    ltem p n Description  1 G1330 67300 Waste Tube Assy  2 G1330 81600 Cable  Autosampler  Fraction Collector to Thermostat    1290 Infinity Thermostat User Manual 63    7  Partsfor Maintenance    HPLC System Tool Kit    HPLC System Tool Kit  G4203  68708  contains some accessories and tools  needed for installation and maintenance of the module     64 1290 Infinity Thermostat User 
11.  CAN cable to previous module V  GPIB   Sono aA irs   c       163             U  A  V  n  V  n  n       ees                           ccce  Ren P                      Thermostat to  Autosampler  Fraction  Collector Cable                                  Figure 9          Cable Connections     Stage 3  Flow Connections      WARNING   When opening capillary or tube fittings  solvents may leak out     The handling of toxic and hazardous solvents and reagents can carry health risks      gt  Observe appropriate safety procedures  for example  wear goggles  safety gloves    3    and protective clothing  as described in the material handling and safety data sheet    supplied by the solvent vendor  especially when toxic or hazardous solvents are    used        1290 Infinity Thermostat User Manual    27    3    28    Installing the Thermostat  Installing the Thermostat    To column compartment    Pump    From pump    To metering device    Plug  To waste    From needle seat  Column compartment    The following procedure describes the installation of the flow   connections of the G1329A B Autosampler  For information about other   autosampler or fraction collector modules see the  Installing  chapter in   the corresponding manuals    1 Connect the pump outlet capillary to port 1 of the injection valve    2 Connect column  compartment inlet capillary to port 6 of the injection  valve     3 Ensure that the waste tube is positioned inside the leak channel                                   
12.  Infinity Thermostat User Manual    1290 Infinity Thermostat User Manual    6  Maintenance    Warnings and Cautions 54  Introduction to Maintenance 56  Cleaning the Module 57   Exchanging the Power Supply Fuses 58    RE Agilent Technologies    53    6 Maintenance    Warnings and Cautions    CAUTION Damaged electronics    Disconnecting or reconnecting the autosampler  fraction collector to thermostat cable  when the power cords are connected to either of the two modules will damage the  electronics of the modules  In such a case  mainboards of both instruments must be  exchanged  otherwise they can damage the other instrument      gt  Make sure the power cords are unplugged before disconnecting or reconnecting the  autosampler  fraction collector to thermostat cable          WARNING   Module is partially energized when switched off  as long as the power cord is  plugged in     Risk of stroke and other personal injury  Repair work at the module can lead to  personal injuries  e  g  shock hazard  when the module cover is opened and the  instrument is connected to power      gt  Never perform any adjustment  maintenance or repair of the module with the top  cover removed and with the power cord plugged in      gt  The security lever at the power input socket prevents that the module cover is taken  off when line power is still connected  Never plug the power line back in when cover  is removed          WARNING   Sharp metal edges    Sharp edged parts of the equipment may cause in
13.  are in  the tray     In cooling mode condensation is generated on the cooled side of the  Peltier elements  This condensed water is safely guided into the leak  system     1290 Infinity Thermostat User Manual    Introduction  Introduction to the Thermostat        Heat sink fans  TCA board               Power supply    Heat sink    Heat exchanger        Heat exchanger fan    Condensation leak drain    Figure 1 Overview of the Thermostat    1290 Infinity Thermostat User Manual    1 Introduction  Thermostat Operation    Thermostat Operation         Heat exchanger       Special sample tray        Air channel in      autosampler sampling unit          Air outlet    Heating cooling module             Peltier elements    Ambient air in       Thermostatted air        Heated air out    Figure 2 Thermostat Principle    8 1290 Infinity Thermostat User Manual    Introduction 1    The thermostat is equipped with    cooling heating module which uses  Peltier elements for efficient air cooling  When turned on the front side of  the Peltier elements is heated cooled according to the temperature setting      fan draws air from the sample tray area and blows it through the  channels of the heating cooling module  The fan speed is determined  according to the environmental conditions  e g   ambient temperature   humidity   In the heating cooling module the air reaches the temperature  of the Peltier elements and this thermostatted air is blown underneath the  special sample tray where it is ev
14.  be impaired      gt  The operator of this instrument is advised to use the equipment in a manner as  specified in this manual        Safety Standards    This is a Safety Class I instrument  provided with terminal for protective  earthing  and has been manufactured and tested according to  international safety standards     Operation    Before applying power  comply with the installation section  Additionally  the following must be observed     Do not remove instrument covers when operating  Before the instrument is  switched on  all protective earth terminals  extension cords   auto transformers  and devices connected to it must be connected to a  protective earth via a ground socket  Any interruption of the protective  earth grounding will cause a potential shock hazard that could result in  serious personal injury  Whenever it is likely that the protection has been  impaired  the instrument must be made inoperative and be secured against  any intended operation     Make sure that only fuses with the required rated current and of the  specified type  normal blow  time delay  and so on  are used for    1290 Infinity Thermostat User Manual 89    10 Appendix    90    replacement  The use of repaired fuses and the short circuiting of fuse  holders must be avoided     Some adjustments described in the manual  are made with power supplied  to the instrument  and protective covers removed  Energy available at  many points may  if contacted  result in personal injury     Any adjust
15.  park transport assembly 32  parts and materials 18  performance specifications 16  physical specifications 15  power consideration 12  power consumption 15  powercords 13  power requirements 12  power supply indicator 36    R  radio interference 92  remote  cable 72  repairs    cautions and warnings 54  RS 232C  cable 80    S    safety class  89  safety  general information 89  standards 15  symbols 88  sample tray 6  sample trays  numbering of vial positions 30  serial number    information 83  83  shipping 32  shutdown 42  site requirements 12  12  power cords 13  sound emission 93  specification  physical 15  specifications 16  status indicators 36  status indicator 35  storage 14    T    temperature sensor 46  temperature 14  timeout 41  tool kit   hplc system 64  transport mechanism 6  transport 32  32  troubleshooting   error messages 40    V  vialracks 6  vial tray 32  vials 6    voltage range 15    Ww    warnings and cautions 54  waste    electrical and electronic  equipment 91    WEEE directive 91  weight 15    1290 Infinity Thermostat User Manual    Index    1290 Infinity Thermostat User Manual 97    www agilent com    In This Book    This manual contains information about the  Agilent 1290 Infinity Thermostat  This  manual describes the following       introduction      site requirements and specifications      installing the thermostatted autosampler      optimizing performance      troubleshooting and diagnostics      maintenance      parts and materials for mai
16.  representative   2 Leak sensor incorrectly routed  being Please contact your Agilent service  pinched by a metal component  representative     1290 Infinity Thermostat User Manual 43    5    44    Error Information    Leak Sensor Open    Error ID  0083  The leak sensor in the module has failed  open circuit      The current through the leak sensor is dependent on temperature  A leak  is detected when solvent cools the leak sensor  causing the leak  sensor  current to change within defined limits  If the current falls outside the  lower limit  the error message is generated     Probable cause Suggested actions  1 Leak sensor not connected to the main Please contact your Agilent service  board  representative   2 Defective leak sensor  Please contact your Agilent service  representative   3 Leak sensor incorrectly routed  being Please contact your Agilent service  pinched by a metal component  representative     Compensation Sensor Open    Error ID  0081    The ambient  compensation sensor  NTC  on the main board in the module  has failed  open circuit      The resistance across the temperature compensation sensor  NTC  on the  main board is dependent on ambient temperature  The change in  resistance is used by the leak circuit to compensate for ambient  temperature changes  If the resistance across the sensor increases above  the upper limit  the error message is generated     Probable cause Suggested actions  1 Defective main board  Please contact your Agilent service  repre
17. 2003  is  introducing producer responsibility on all electric and electronic  appliances starting with 13 August 2005     This product complies with the WEEE Directive  2002 96 EC  marking requirements  The  affixed label indicates that you must not discard this electrical electronic product in  domestic household waste     Product Category     With reference to the equipment types in the WEEE Directive Annex I  this product is  classed as a Monitoring and Control Instrumentation product     E    Do not dispose off in domestic household waste    To return unwanted products  contact your local Agilent office  or see www agilent com for  more information     1290 Infinity Thermostat User Manual 91    10 Appendix    Radio Interference    Cables supplied by Agilent Technologies are screened to provide optimized  protection against radio interference  All cables are in compliance with  safety or EMC regulations     Test and Measurement    If test and measurement equipment is operated with unscreened cables  or  used for measurements on open set ups  the user has to assure that under  operating conditions the radio interference limits are still met within the  premises     92 1290 Infinity Thermostat User Manual    Appendix 10    Sound Emission    Manufacturer s Declaration    This statement is provided to comply with the requirements of the  German Sound Emission Directive of 18 January 1991     This product has a sound pressure emission  at the operator position    lt  70 dB     
18. 6 integrators    Agilent module to general purpose    1290 Infinity Thermostat User Manual    CAN cables    p n  5181 1516  5181 1519    LAN cables    p n  5023 0203    5023 0202    Cable Identification 8    Description  CAN cable  Agilent module to module  0 5 m    CAN cable  Agilent module to module  1 m    Description    Cross over network cable  shielded  3 m  for point to point  connection     Twisted pair network cable  shielded  7 m  for point to point  connection     External Contact Cable    p n  G1103 61611    RS 232 cables    p n  G1530 60600  RS232 61601    5181 1561    1290 Infinity Thermostat User Manual    Description    External contact cable   Agilent module interface board to general  purposes    Description  RS 232 cable  2 m    RS 232 cable  2 5 m   Instrument to PC  9 to 9 pin  female   This cable has special pin out   and is not compatible with connecting printers and plotters  It s also  called  Null Modem Cable  with full handshaking where the wiring  is made between pins 1 1  2 3  3 2  4 6  5 5  6 4  7 8  8 7  9 9     RS 232 cable  8 m    69    Cable Identification    Analog Cables    70       One end of these cables provides a BNC connector to be connected to  Agilent modules  The other end depends on the instrument to which    connection is being made     Agilent Module to 3394 6 Integrators                   p n 35900 60750 Pin 3394 6 Pin Agilent Signal Name  module  1 Not connected  2 Shield Analog              3 Center Analog    Agilent Module to
19. Defective electronic board in autosampler or  autosampler thermostat    Suggested actions    Check cable connection between autosampler  and autosampler thermostat     Connect cable between autosampler and  autosampler thermostat       Exchange ASM board in the autosampler       Exchange TCA board in the thermostat       Contact the Agilent Technologies service  organization     Power Fail for Autosampler Thermostat Module    Error ID  4113    The  36    that is generated in the autosampler thermostat power supply  is checked by the autosampler electronics  If this voltage is missing the    error message will be generated     Probable cause    1 Autosampler thermostat module not  connected to line power    2 Autosampler thermostat module fuse s   defective    3 Autosampler thermostat power supply  defective    Suggested actions    Check correct power line connection     Check fuse s  and replace if defective     Contact the Agilent Technologies service  organization     1290 Infinity Thermostat User Manual    Error Information 5    Temperature out of Range    Error ID  4114    To protect the Peltier elements from damage their working range is limited  to  3      to 65       The error message is generated when at least one of  the four Peltier elements exceeds this limit     Probable cause Suggested actions    1 The four main heat exchanger fans are not       Clean the air filters and heat exchanger fins  able to blow enough air through the in the autosampler thermostat  Switc
20. Figure 4 on page 20 and Figure 9 on page 27   Refer  to the documentation of the Agilent 1200 Infinity Series Instant Pilot or  ChemStation for LC for more information     In an Agilent 1290 Infinity or 1260 Infinity system  the individual modules are connected by  a CAN cable  The Agilent 1200 Infinity Series Instant Pilot can be connected to the CAN bus  at any of the modules in the system  The control software can be connected to the system  by one GPIB cable at any of the modules  If an Agilent detector is part of the system  the   LAN connection must be at the detector  For more information about connecting the instant  pilot or control software refer to the respective user manual  For connecting the Agilent   1290 Infinity equipment to non Agilent 1290 Infinity equipment  see Autosampler manual      1290 Infinity Thermostat User Manual 25    3 Installing the Thermostat  Installing the Thermostat    7 Connect additional cables as required  see Figure 9 on page 27         0000000         Rutosampler  Fraction  ollector to  hermostat Cable       Power  connector             Safety lever    Figure 8 Power Connectors and Safety Levers at Rear of thermostatted  Autosampler Fraction Collector     26 1290 Infinity Thermostat User Manual    Relay contacts    Vial number output       Installing the Thermostat  Installing the Thermostat    CAN cable to next module                                a              C              E Tu 5     B   o     a RR      i  RS232C      CAN bus    
21. Manual    Parts for Maintenance 7  Plastic Parts    Plastic Parts       Figure 16 Plastic Parts    Item p n Description   1 G1330 68723 Cover kit  includes base  top  left and right    2 5065 9982 Plastics kit  includes base  top  left and right sides   5042 8907 Leak pan  solvent cabinet    1290 Infinity Thermostat User Manual 65    7  Partsfor Maintenance  Plastic Parts    66 1290 Infinity Thermostat User Manual    1290 Infinity Thermostat User Manual    8  Cable Identification    Cable Overview 68  Analog Cables 70  Remote Cables 72   BCD Cables 75   Auxiliary Cable 77  CAN LAN Cables 78  External Contact Cable     79  RS 232 Cable Kit 80    OE Agilent Technologies    67    Cable Identification    Cable Overview    68    Never use cables other than the ones supplied by Agilent Technologies to ensure proper  functionality and compliance with safety or EMC regulations     Analog cables    p n   35900 60750  35900 60750  01046 60105    Remote cables    p n  03394 60600    03396 61010  5061 3378  01046 60201    BCD cables    p n  03396 60560  G1351 81600    Description  Agilent module to 3394 6 integrators  Agilent 35900A A D converter    Analog cable  BNC to general purpose  spade lugs     Description  Agilent module to 3396A Series   integrators    3396 Series 11   3395A integrator  see details in section  Remote  Cables  on page 72    Agilent module to 3396 Series      3395B integrators  Remote Cable    Agilent module to general purpose    Description  Agilent module to 339
22. ater and mild detergent       WARNING   Liquid dripping into the electronic compartment of your module can cause shock  hazard and damage the module     gt  Do not use an excessively damp cloth during cleaning     Drain all solvent lines before opening any connections in the flow path     6       1290 Infinity Thermostat User Manual    57    6 Maintenance    Exchanging the Power Supply Fuses    When    Tools required    Parts required    CAUTION    The fuse holders are located on the rear panel of the thermostat     If wrong fuses are installed     Description    Flat head screwdriver    p n Description  2110 0015 Fuses T2 5 A 250V  CSA  UL listed     Damaged electronics    Disconnecting or reconnecting the autosampler  fraction collector to thermostat cable  when the power cords are connected to either of the two modules will damage the  electronics of the modules  In such a case  mainboards of both instruments must be  exchanged  otherwise they can damage the other instrument      gt  Make sure the power cords are unplugged before disconnecting or reconnecting the  autosampler  fraction collector to thermostat cable        58    1290 Infinity Thermostat User Manual    oo nN c Cc dA    Maintenance    Switch OFF the power switch at the front of the thermostatted  autosampler     Remove the power cable from the two modules     Insert the flat head screwdriver in the fuse holder  slightly press and  turn counter clockwise to release the fuse holder from the socket              
23. d  to the Agilent modules  The other end depends on the instrument to be    connected to    Agilent Module to General Purpose                         p n G1351 81600 Wire Color Pin Agilent Signal Name        Digit  module  Green 1 BCD 5 20  p Violet 2 BCD 7 80  Ree Blue 3 BCD 6 40  a   Yellow 4 BCD 4 10  E Black 5 BCD 0 1  MCN Orange 6 BCD 3 8  Red 7 BCD2 4  Brown 8 BCD 1 2  Gray 9 Digital ground Gray  Gray pink 10 BCD 11 800  Red blue 11 BCD 10 400  White green 12 BCD 9 200  Brown green 13 BCD 8 100  notconnected 14  notconnected 15  5V Low       1290 Infinity Thermostat User Manual    75    8 Cable Identification    Agilent Module to 3396 Integrators    p n 03396 60560 Pin 3396 Pin Agilent Signal Name        Digit                            module  p 1 1 BCD 5 20  2 2 BCD 7 80  Se 15        o 3 3 BCD 6 40      So 4 4 BCD 4 10  E           5 5 BCDO 1  fre e9  6 6 BCD 3 8  Ex 7 7 BCD2 4  8 8 BCD 1 2  9 9 Digital ground  NC 15  5V Low       76 1290 Infinity Thermostat User Manual    Auxiliary Cable    Cable Identification    One end of this cable provides a modular plug to be connected to the  Agilent vacuum degasser  The other end is for general purpose     Agilent Vacuum Degasser to general purposes       p n G1322 81600 Color Pin Agilent Signal Name  1100       White 1 Ground     Brown Pressure signal  Green 3  Yellow 4  Grey 5 DC 5VIN  Pink 6 Vent       1290 Infinity Thermostat User Manual    TI    Cable Identification    CAN LAN Cables    78       Both ends of this cable 
24. document are furnished under a license  and may be used or copied only in accor   dance with the terms of such license     Restricted Rights Legend    If software is for use in the performance of a  U S  Government prime contract or subcon   tract  Software is delivered and licensed as   Commercial computer software  as  defined in DFAR 252 227 7014  June 1995    or as a  commercial item  as defined in FAR  2 101 a  or as  Restricted computer soft   ware  as defined in FAR 52 227 19  June  1987  or any equivalent agency regulation  or contract clause  Use  duplication or dis   closure of Software is subject to Agilent  Technologies  standard commercial license  terms  and non DOD Departments and  Agencies of the U S  Government will    receive no greater than Restricted Rights as  defined in FAR 52 227 19 c  1 2   June  1987   U S  Government users will receive  no greater than Limited Rights as defined in  FAR 52 227 14  June 1987  or DFAR  252 227 7015  b  2   November 1995   as  applicable in any technical data     Safety Notices    CAUTION    A CAUTION notice denotes a  hazard  It calls attention to an  operating procedure  practice  or  the like that  if not correctly per   formed or adhered to  could  result in damage to the product  or loss of important data  Do not  proceed beyond a CAUTION  notice until the indicated condi   tions are fully understood and  met        A WARNING notice denotes a  hazard  It calls attention to an  operating procedure  practice   or the 
25. e  with full handshaking where the wiring  is made between pins 1 1  2 3  3 2  4 6  5 5  6 4  7 8  8 7  9 9     RS 232 cable  8 m    1290 Infinity Thermostat User Manual    1290 Infinity Thermostat User Manual    e Ove Hardware Information       amp  e  e e   amp  Electrical Connections 82     Information on Instrument Serial Number 83  Rear View of the Module 84  Early Maintenance Feedback 85  This chapter describes the module in more detail on hardware and electronics  NOTE  manuals     Details on electrical connections are described in the autosampler fraction collector             Agilent Technologies 81    9    Hardware Information    Electrical Connections    82    The CAN bus is a serial bus with high speed data transfer  The two  connectors for the CAN bus are used for internal module data transfer  and synchronization     The interface board slot is used for external contacts and BCD bottle  number output or LAN connections     The REMOTE connector may be used in combination with other  analytical instruments from Agilent Technologies if you want to use  features such as start  stop  common shut down  prepare  and so on     With the appropriate software  the RS 232C connector may be used to  control the module from a computer through    RS 232C connection   This connector is activated and can be configured with the  configuration switch     The Thermostat  Autosampler  Fraction collector connection is used for  control signal transfer and synchronization of the two m
26. e autosampler module   G1329A B  G1367A D  G1367E  G1377A  G2260A  G4226A  G5667A   G4303A  or fraction collector  G1364A C  G1364D  G5664A  and the  thermostat module  G1330B   Both modules have a separate power supply  and a power plug for the line connections  The two modules are connected  by a control cable and both are turned on by the autosampler or fraction  collector     The autosampler or fraction collector power supplies have automatic  voltage selectors  Consequently there are no voltage selectors in the rear  of the two autosampler or fraction collector modules  The autosampler or  fraction collector has no externally accessible fuses  because automatic  electronic fuses are implemented in its power supply  The thermostat  power supply has two externally accessible fuses     Hazard of electrical shock or damage of your instrumentation  can result  if the devices are connected to a line voltage higher than specified      gt  Connect your instrument to the specified line voltage only        CAUTION    Inaccessible power plug     In case of emergency it must be possible to disconnect the instrument from the power  line at any time      gt  Make sure the power connector of the instrument can be easily reached and  unplugged      gt  Provide sufficient space behind the power socket of the instrument to unplug the  cable        12    1290 Infinity Thermostat User Manual    Site Requirements and Specifications 2    Power Cords    Different power cords are offered as opt
27. e displayed in the user interface when an electronic   failure occurs with the autosampler thermostat module which requires   attention  In the event of such a failure  the red status indicator at the  front of the thermostatted autosampler is switched on  and an entry is  written into the instrument log book     This section describes the meaning of autosampler thermostat module  error messages  and provides information on probable causes and  suggested actions how to recover from error conditions     Fan Failed    Error ID  4109    Each fan in the thermostat is equipped with a speed sensor  that allows  monitoring and control the speed of the fans  If the sensor shows no  signal  when the fan is activated  it is very likely that the fan is defective   The number in brackets indicates the fan position  Position numbers of the  fans are as follows  seen from front of the thermostat       fanfailed 1   left side  fan in front position    fan failed 2   left side  fan      back position    fan failed 3   right side  fan in back position    fan failed 4   right side  fan in front position      fanfailed 5   small cooler fan on top of cooling   heating module    Probable cause Suggested actions    1 One of the 4 heat sink fans is defective  See Service Manual for Exchanging the  Heatsink Fans     2 The cooling heating module fan is defective      Service Manual for Exchanging Heat  Exchanger Fan     1290 Infinity Thermostat User Manual    Error Information 5    Temperature Contro
28. enly distributed and streams back into  the sample tray area  From there it is again drawn into the thermostat   This  recycle  mode assures a very efficient cooling heating of the sample  vials     In cooling mode the opposite side of the Peltier element will become very  hot and to maintain the performance of the elements they have to be  cooled down  This is done with large heat exchangers in the back of the  thermostat  Four fans blow air from left to right through the instrument to  remove the heated air  The fan speed is controlled according to the  temperature of the Peltier elements     During cooling condensation will appear in the heating cooling module   The condensed water will be guided out of the thermostat     1290 Infinity Thermostat User Manual 9    1 Introduction  Thermostat Operation    10 1290 Infinity Thermostat User Manual    1290 Infinity Thermostat User Manual    0        2     Qeu9  Site Requirements and Specifications     amp  e  5 o i Site Requirements 12    Power Consideration 12  Power Cords 13  Bench Space 14  Environment 14    Physical Specifications 15    Performance Specifications 16    This chapter provides information on environmental requirements  physical and  performance specifications     XE Agilent Technologies 11    2 Site Requirements and Specifications    Site Requirements       suitable site environment is important to ensure optimum performance  of the instrument     Power Consideration    The autosamplers comprises two modules  th
29. er should refer to  the instruction manual in order to protect risk of harm to the operator and  to protect the apparatus against damage     Indicates dangerous voltages     Indicates a protected ground terminal     CAN    Indicates eye damage may result from directly viewing the light produced  by the deuterium lamp used in this product     9       The apparatus is marked with this symbol when hot surfaces are available  and the user should not touch it when heated up     p       KITE UMS    alerts you to situations that could cause physical injury or death      gt  Do not proceed beyond a warning until you have fully understood and met the  indicated conditions        CAUTION A CAUTION    alerts you to situations that could cause loss of data  or damage of equipment      gt  Do not proceed beyond a caution until you have fully understood and met the  indicated conditions        88 1290 Infinity Thermostat User Manual    Appendix 10    General Safety Information    The following general safety precautions must be observed during all  phases of operation  service  and repair of this instrument  Failure to  comply with these precautions or with specific warnings elsewhere in this  manual violates safety standards of design  manufacture  and intended use  of the instrument  Agilent Technologies assumes no liability for the  customer s failure to comply with these requirements       WARNING   Ensure the proper usage of the equipment   The protection provided by the equipment may
30. flow will be stopped for safety  reasons  For a detector  the lamp will stay on in order to avoid  equilibration time  Depending on the error type  the next run can only be  started  if the error has been resolved  for example liquid from a leak has  been dried  Errors for presumably single time events can be recovered by  switching on the system in the user interface     Special handling is done in case of a leak  As a leak is a potential safety  issue and may have occurred at a different module from where it has been  observed  a leak always causes a shutdown of all modules  even outside a  method run     In all cases  error propagation is done via the CAN bus or via an APG  remote cable  see documentation for the APG interface      1290 Infinity Thermostat User Manual    General Error Messages    Error Information    General error messages are generic to all Agilent series HPLC modules  and may show up on other modules as well     Timeout    Error ID  0062    The timeout threshold was exceeded     Probable cause    1 The analysis was completed successfully   and the timeout function switched off the  module as requested     2 Anot ready condition was present during a    sequence or multiple injection run for a  period longer than the timeout threshold     1290 Infinity Thermostat User Manual    Suggested actions    Check the logbook for the occurrence and  source of a not ready condition  Restart the  analysis where required     Check the logbook for the occurrence and  so
31. ft    Non operating altitude Up to 4600 m  15091 ft  For storing the module    Safety standards  IEC  CSA  Installation category Il  Pollution degree 2 For indoor use only   UL    Using the module at high environmental temperatures may cause the rear panel to  become hot        Do not use the module at environmental temperatures higher than 50       122  F        1290 Infinity Thermostat User Manual 15    2    Site Requirements and Specifications    Performance Specifications    16    Table2 Performance specifications Agilent 1290 Infinity Thermostat       Type Specification  Temperature range Settable from 4   C to 40   C in 1   increments    Temperature accuracy at ambient    3      to 8   C at a setpoint of 4   C  temperatures  lt  25   C and humidity  lt  50      Temperature accuracy at ambient    3   C to 9   C at a setpoint of 4 ap  temperatures   30      and humidity   60           LER  Measurement conditions     G1329B   with 100 Vial Tray in vial location 2 10 92 and 100 vials filled with water  G1367A G1367B G1367C G1377A G1367E G5667A G2258A G4226A     For vials  Using the Thermostattable Tray  G1329 60011  or 100 Micro Vial   G4226 60021   both loaded with 100 vials  Temperature is measured in vial locations  1 10 23 25 45 75 91 and 100  filled with 1 mL of water     For well plates  Standard Tray  G2258 60011  for two Well Plates loaded with two  Agilent 96 Well Plate  5042 1386     1290 Infinity Thermostat User Manual    1290 Infinity Thermostat User Manual    3
32. ft  hand 15 position tray  1 15  Right hand 40 position tray  101 140    Right hand 15 position tray  101 115        Position 1    Position 101    Figure 12 Numbering of Tray Positions     30 1290 Infinity Thermostat User Manual    Installing the Thermostat 3    Stage 5  Installing Tray Cover and Front Cover    1 Fix the tray cover in the clips of the left autosampler cover side by  sliding it in position  Do not close the tray cover     2 Position the front cover in the top left corner of the autosampler and  turn it towards the instrument  Press the stop fastener to secure it in  the right side cover of the autosampler     3 Close the tray cover         Tray cover    pA        Autosampler front door       Figure 13 Installation of Tray Cover and Front Cover of the Autosampler    Stage 6  Turning on the Thermostatted Autosampler  Fraction  collector    1 Depress the power switch to turn on the two modules   The power switch stays depressed  1  and a green indicator lamp in the power switch is on    when the module is turned on  When the line power switch stands out  0  and the green  light is off  the module is turned off     1290 Infinity Thermostat User Manual 31       Installing the Thermostat    Transporting the Thermostatted Autosampler or Fraction    Collector    When moving the autosampler or fraction collector around the laboratory   make sure that any condensed water inside the thermostat is removed   Tilt the module to the front  so that the water inside the thermo
33. h off  autosampler thermostat the thermostat and wait 10 min for the   peltier elements to equilibrate       Make sure that there is enough space on  left and right side of the autosampler  thermostat and that the air channel inlets  and outlets are not blocked     2 Ambient temperature too high   low Make sure that the ambient air temperature is  within its specified limits  4   C to 55           1290 Infinity Thermostat User Manual 51    5 Error Information    Bad Cooling   Heating Performance    Error ID  4120    When turned on  the autosampler thermostat Peltier elements are  activated according to the given setpoint or to given setpoint change  The  electronics check whether the actual temperature is moving in the correct  direction  e g   decreasing when cooled down      Probable cause Suggested actions    1 Peltier element s  defective Contact the Agilent Technologies service  organization     2 Peltier element s  not connected or Check connections of the Peltier elements   incorrectly connected  e g   after repair or  maintenance     3 The four main heat exchanger fans are not       Clean the air filters and heat exchanger fins  able to blow enough air through the in the autosampler thermostat  Switch off  autosampler thermostat the thermostat and wait 10 min for the   peltier elements to equilibrate       Make sure that there is enough space on  left and right side of the autosampler  thermostat and that the air channel inlets  and outlets are not blocked     52 1290
34. icates a not ready condition  The module is in a  not ready state when it is waiting for a specific condition to be reached  or completed  for example  immediately after changing a set point   or  while a self test procedure is running     An error condition is indicated when the status indicator is red  An  error condition indicates the module has detected an internal problem  which affects correct operation of the module  Usually  an error  condition requires attention  e g  leak  defective internal components    An error condition always interrupts the analysis     If the error occurs during analysis  it is propagated within the LC  System  i e  a red LED may indicate a problem of a different module   Use the status display of your user interface for finding the root  cause module of the error     A blinking indicator indicates that the module is in resident mode  e g   during update of main firmware      A fast blinking indicator indicates that the module is in a low level  error mode  In such a case try to re boot the module or try a  cold start  then try a firmware update of the autosampler  fraction  collector  for details see the manual of your autosampler  fraction  collector      1290 Infinity Thermostat User Manual 37    4 Troubleshooting and Diagnostics  Status Indicators    38 1290 Infinity Thermostat User Manual    1290 Infinity Thermostat User Manual      eo  ee  5     909  Error Information         e  e e  e What Are Error Messages 40    General Error Messages
35. ions with the module  The female  end of all power cords is identical  It plugs into the power  input socket at  the rear  The male end of each power cord is different and designed to  match the wall socket of a particular country or region     Absence of ground connection or use of unspecified power cord    The absence of ground connection or the use of unspecified power cord can lead to  electric shock or short circuit      gt  Never operate your instrumentation from a power outlet that has no ground  connection      gt  Never use a power cord other than the Agilent Technologies power cord designed  for your region        Use of unsupplied cables    Using cables not supplied by Agilent Technologies can lead to damage of the  electronic components or personal injury      gt  Never use cables other than the ones supplied by Agilent Technologies to ensure  proper functionality and compliance with safety or EMC regulations        Unintended use of supplied power cords    Using power cords for unintended purposes can lead to personal injury or damage of  electronic equipment      gt  Never use the power cords that Agilent Technologies supplies with this instrument  for any other equipment        1290 Infinity Thermostat User Manual 13    2 Site Requirements and Specifications    Bench Space    The module dimensions and weight  see Table 1 on page 15  allow you to  place the module on almost any desk or laboratory bench  It needs an  additional 25 cm  10 inches  of space on eithe
36. juries      gt  To prevent personal injury  be careful when getting in contact with sharp metal  areas        54 1290 Infinity Thermostat User Manual    Maintenance 6      WARNING   Toxic  flammable and hazardous solvents  samples and reagents    The handling of solvents  samples and reagents can hold health and safety risks        When working with these substances observe appropriate safety procedures  for  example by wearing goggles  safety gloves and protective clothing  as described in  the material handling and safety data sheet supplied by the vendor  and follow good  laboratory practice      gt  The volume of substances should be reduced to the minimum required for the  analysis      gt  Do not operate the instrument in an explosive atmosphere        CAUTION Safety standards for external equipment     gt  If you connect external equipment to the instrument  make sure that you only use  accessory units tested and approved according to the safety standards appropriate  for the type of external equipment        1290 Infinity Thermostat User Manual 55    6 Maintenance    Introduction to Maintenance    The module is designed for easy maintenance  Maintenance can be done  from the front with module in place in the system stack     There are no serviceable parts inside     Do not open the module     56 1290 Infinity Thermostat User Manual    Maintenance    Cleaning the Module    To keep the module case clean  use a soft cloth slightly dampened with  water  or a solution of w
37. l Failed  1   4     Error ID  4110    There are four Peltier elements built into the autosampler thermostat for  efficient cooling   heating  The electronics monitor the current through the  Peltier elements  If the Peltier current is out of a specified limit the Peltier  element is defective     Probable cause Suggested actions    1 One of the four peltier elements is defective  Peltier elements cannot be replaced on site   Please contact Agilent Technologies service  organization to initiate instrument exchange  process     Temperature Sensor Failed  1   4   Error ID  4111    The autosampler thermostat is equipped with four sensors to monitor the  performance of the instrument  The sensors are connected to both sides of  the Peltier elements for control of the cooling   heating efficiency  As the  positioning of the sensor is critical to meet the performance requirements  of the autosampler thermostat the single sensors cannot be replaced    separately   Probable cause Suggested actions  1 Temperature sensor is defective See Service Manual for Exchanging the    Sensors     1290 Infinity Thermostat User Manual 49    5    50    Error Information    Lost Contact to Autosampler Thermostat    Error ID  4112    The autosampler control electronics continuously check whether the  autosampler thermostat is active or not     Probable cause    1 Badcable connection between autosampler  and autosampler thermostat    2 Cable between autosampler and thermostat  module not connected    3 
38. like that  if not correctly  performed or adhered to  could  result in personal injury or  death  Do not proceed beyond a  WARNING notice until the indi   cated conditions are fully under   stood and met        1290 Infinity Thermostat User Manual    Contents    Contents    1 Introduction 5    Introduction to the Thermostat 6  Thermostat Operation 8    2 Site Requirements and Specifications 11    Site Requirements 12  Physical Specifications 15  Performance Specifications 16    3 Installing the Thermostat 17    Damaged Packaging 18   Optimizing the Stack Configuration 19   Installing the Thermostat 21   Transporting the Thermostatted Autosampler or Fraction Collector 32    4 Troubleshooting and Diagnostics 33    Agilent Lab Advisor Software 34  Overview of the Thermostat s Indicators and Test Functions 35  Status Indicators 36    5 Error Information 39    What Are Error Messages 40  General Error Messages 41  Thermostat Error Messages 48    6 Maintenance 53    Warnings and Cautions 54  Introduction to Maintenance 56  Cleaning the Module 57   Exchanging the Power Supply Fuses 58    1290 Infinity Thermostat User Manual    Contents    7 Partsfor Maintenance 61    Main Assemblies  External Parts  62  Accessory Kit 63   HPLC System Tool Kit 64   Plastic Parts 65    Cable Identification 67    Cable Overview 68  Analog Cables 70  Remote Cables 72   BCD Cables 75   Auxiliary Cable 77  CAN LAN Cables 78  External Contact Cable 79  RS 232 Cable Kit 80    Hardware Information 81    Elec
39. ment  maintenance  and repair of the opened instrument under  voltage should be avoided whenever possible  When inevitable  this has to  be carried out by a skilled person who is aware of the hazard involved  Do  not attempt internal service or adjustment unless another person  capable  of rendering first aid and resuscitation  is present  Do not replace  components with power cable connected     Do not operate the instrument in the presence of flammable gases or  fumes  Operation of any electrical instrument in such an environment  constitutes a definite safety hazard     Do not install substitute parts or make any unauthorized modification to  the instrument     Capacitors inside the instrument may still be charged  even though the  instrument has been disconnected from its source of supply  Dangerous  voltages  capable of causing serious personal injury  are present in this  instrument  Use extreme caution when handling  testing and adjusting     When working with solvents  observe appropriate safety procedures  for  example  goggles  safety gloves and protective clothing  as described in the  material handling and safety data sheet by the solvent vendor  especially  when toxic or hazardous solvents are used     1290 Infinity Thermostat User Manual    Appendix 10    The Waste Electrical and Electronic Equipment  WEEE  Directive   2002 96 EC     Abstract    The Waste Electrical and Electronic Equipment  WEEE  Directive   2002 96 EC   adopted by EU Commission on 13 February 
40. ntenance      cable identification      configuring the thermostatted autosampler       appendix      O Agilent Technologies 2011 2014  2015    Printed in Germany    05 2015    G1330 90014  Rev  B                         Agilent Technologies    
41. odules  The  cable must be installed for operation of the thermostat  Make sure the  power cords are unplugged before disconnecting or reconnecting the  cable coupling between both modules     The power input socket accepts a line voltage of 100   240 VAC   10    with a line frequency of 50 or 60 Hz  Maximum power consumption  varies by module  There is no voltage selector on your module because  the power supply has wide ranging capability  There are no externally  accessible fuses  because automatic electronic fuses are implemented in  the power supply     Never use cables other than the ones supplied by Agilent Technologies to ensure proper  functionality and compliance with safety or EMC regulations     1290 Infinity Thermostat User Manual    Hardware Information 9    Information on Instrument Serial Number    Serial Number Information 1200 Series and 1290 Infinity    The serial number information on the instrument labels provide the    following information     CCYWWSSSSS  CC    YWW    SSSSS    Format    country of manufacturing    DE  Germany            Japan     CN  China    year and week of last major manufacturing change  e g  820  could be week 20 of 1998 or 2008    real serial number    Serial Number Information 1260 Infinity    The serial number information on the instrument labels provide the    following information     CCXZZ00000  CC    ZZ    00000    1290 Infinity Thermostat User Manual    Format    Country of manufacturing    DE  Germany              Japan   
42. on  Collector Cable  THU Aii                              Remote cable          Figure 4 Recommended Stack Configuration  Rear View     20 1290 Infinity Thermostat User Manual    Installing the Thermostat 3    Installing the Thermostat    Module is partially energized when switched off  as long as the power cord is  plugged in     Repair work at the module can lead to personal injuries  e g  shock hazard  when the  cover is opened and the module is connected to power      gt  Make sure that it is always possible to access the power plug    gt  Remove the power cable from the instrument before opening the cover      gt  Do not connect the power cable to the Instrument while the covers are removed        Personal injury  To avoid personal injury  keep fingers away from the needle area during autosampler  operation     Do not bend the safety flap away from its position  or attempt to remove the safety  cover      gt  Do not attempt to insert or remove a vial from the gripper when the gripper is  positioned below the needle        CAUTION     Defective on arrival  problems    If there are signs of damage  please do not attempt to install the module  Inspection by  Agilent is required to evaluate if the instrument is in good condition or damaged      gt  Notify your Agilent sales and service office about the damage     An Agilent service representative will inspect the instrument at your site and  initiate appropriate actions        1290 Infinity Thermostat User Manual 21    3 Ins
43. ostat 3    Installing the Thermostat    3 If the thermostat is located on top of another Agilent 1200 Infinity Series  module place the waste tube assembly into the top cover of the thermostat  and locate the other end in the waste funnel of the module beneath           Module locks    Waste tube  assembly        Condensation  leak tube    Figure 5 Preparation of the Thermostat    4 Connect the condensation leak tube to the main waste exit of the  thermostat and place into an appropriate vessel  It is possible to either let  the condensation leak tubing exit the module at the front or at the left side  of the module  Make sure that the leak tube is fully fixed on the outlet     Condensation  leak tube    Waste bottle    Figure 6 Condensation Leak outlet    1290 Infinity Thermostat User Manual 23    3 Installing the Thermostat  Installing the Thermostat    5 Install the front cover of the thermostat     6 Place the autosampler  fraction collector module on top of the  thermostat  Make sure that the autosampler  fraction collector is  correctly engaged in the thermostat locks     7 Remove the tray and place the air channel adapter into the  autosampler  fraction collector tray base  Make sure the adapter is fully  pressed down  This assures that the cold airstream from the thermostat  is correctly guided to the tray area of the autosampler or fraction  collector         Air channel  adapter    Figure 7 Air channel adapter    8 Install the diffusion plate on the air channel adap
44. perature sensors  leak sensor and   board  mounted temperature  compensation sensor  are used by the leak  algorithm to determine whether a leak is present  When a leak occurs  the  leak sensor is cooled by the solvent  This changes the resistance of the  leak sensor which is sensed by the leak sensor circuit on the main board     Probable cause Suggested actions   1 Loose fittings  Ensure all fittings are tight    2 Broken capillary  Exchange defective capillaries   Open Cover    Error ID  0205    The top foam has been removed     Probable cause Suggested actions   1 Foam not activating the sensor  Please contact your Agilent service  representative    2 Defective sensor or main board  Please contact your Agilent service  representative     46 1290 Infinity Thermostat User Manual    Error Information 5    Cover Violation    Error ID  7461  The top foam has been removed     The sensor on the main board detects when the top foam is in place  If  the foam is removed while the lamps are on  or if an attempt is made to  switch on for example the lamps with the foam removed   the lamps are  switched off  and the error message is generated     Probable cause Suggested actions  1 The top foam was removed during Please contact your Agilent service  operation  representative   2 Foam not activating the sensor  Please contact your Agilent service  representative     1290 Infinity Thermostat User Manual 47    5    Error Information    Thermostat Error Messages    48    Error messages ar
45. product that can be  used with or without data system  Agilent Lab Advisor software helps to  manage the lab for high quality chromatographic results and can monitor  in real time a single Agilent LC or all the Agilent GCs and LCs configured  on the lab intranet     Agilent Lab Advisor software provides diagnostic capabilities for all  Agilent 1200 Infinity Series modules  This includes diagnostic capabilities   calibration procedures and maintenance routines for all the maintenance  routines     The Agilent Lab Advisor software also allows users to monitor the status  of their LC instruments  The Early Maintenance Feedback  EMF  feature  helps to carry out preventive maintenance  In addition  users can generate  a status report for each individual LC instrument  The tests and diagnostic  features as provided by the Agilent Lab Advisor software may differ from  the descriptions in this manual  For details refer to the Agilent Lab  Advisor software help files     The Instrument Utilities is a basic version of the Lab Advisor with limited  functionality required for installation  use and maintenance  No advanced  repair  troubleshooting and monitoring functionality is included     1290 Infinity Thermostat User Manual    Troubleshooting and Diagnostics 4    Overview of the Thermostat s Indicators and Test Functions    Status Indicators    The thermostatted autosampler is provided with two status indicators  which indicate the operational state  prerun  run  and error states  of 
46. provide a modular plug to be connected to  Agilent modules CAN or LAN connectors     CAN Cables    p n  5181 1516  5181 1519    LAN Cables    p n  5023 0203    5023 0202    Description  CAN cable  Agilent module to module  0 5 m    CAN cable  Agilent module to module  1 m    Description    Cross over network cable  shielded  3 m  for point to point  connection     Twisted pair network cable  shielded  7 m  for point to point  connection     1290 Infinity Thermostat User Manual    External Contact Cable       10   6             One end of this cable provides a 15 pin plug to be connected to  Agilent modules interface board  The other end is for general purpose     Cable Identification    Agilent Module Interface Board to general purposes          p n G1103 61611 Color Pin Agilent Signal Name  module    White 1 EXT 1    Brown EXT 1  Ecce Green 3 EXT 2  UI Yellow 4 EXT2  Grey 5 EXT 3  Pink 6 EXT 3  Blue 7 EXT 4  Red 8 EXT 4  Black 9 Not connected  Violet 10 Not connected  Grey pink 11 Not connected  Red blue 12 Not connected  White green 13 Not connected  Brown green 14 Not connected  White yellow 15 Not connected       1290 Infinity Thermostat User Manual    79    Cable Identification    RS 232 Cable Kit    80    p n  G1530 60600  RS232 61601    5181 1561    Description  RS 232 cable  2 m    RS 232 cable  2 5 m   Instrument to PC  9 to 9 pin  female   This cable has special pin out   and is not compatible with connecting printers and plotters  It s also  called  Null Modem Cabl
47. r side and approximately  8 cm  3 1 inches  in the rear for air circulation and electric connections     If the bench shall carry a complete HPLC system  make sure that the  bench is designed to bear the weight of all modules     The module should be operated in a horizontal position     Environment    Your modules will work at ambient temperatures and relative humidity as  described in Table 1 on page 15     CAUTION Condensation within the module  Condensation will damage the system electronics      gt  Do not store  ship or use your module under conditions where temperature  fluctuations could cause condensation within the module      gt   f your module was shipped in cold weather  leave it in its box and allow it to warm  slowly to room temperature to avoid condensation        14 1290 Infinity Thermostat User Manual    Site Requirements and Specifications 2    Physical Specifications    Table 1 Physical Specifications          Type Specification Comments   Weight 20 7 kg  46 Ibs    Dimensions  height x 140 x 345 x 435 mm  5 5 x 13 5 x 17   width x depth  inches    Line voltage 100     240 VAC    10   Wide ranging  capability   Line frequency 50 or 60 Hz   5 96   Power consumption 260 VA   210 W   717 BTU Maximum   Ambient operating 4     40   C  39     104   F  See warning           temperature page 15   Ambient non operating  40     70   C   40     158   F    temperature   Humidity  lt  95   r h  at 40       104   F  Non condensing   Operating altitude Up to 2000 m  6562 
48. sentative     1290 Infinity Thermostat User Manual    Error Information 5    Compensation Sensor Short    Error ID  0080    The ambient  compensation sensor  NTC  on the main board in the module  has failed  short circuit      The resistance across the temperature compensation sensor  NTC  on the  main board is dependent on ambient temperature  The change in  resistance is used by the leak circuit to compensate for ambient  temperature changes  If the resistance across the sensor falls below the  lower limit  the error message is generated     Probable cause Suggested actions  1 Defective main board  Please contact your Agilent service  representative     Fan Failed    Error ID  0068  The cooling fan in the module has failed     The hall sensor on the fan shaft is used by the main board to monitor the  fan speed  If the fan speed falls below a certain limit for a certain length  of time  the error message is generated     Depending on the module  assemblies  e g  the lamp in the detector  are  turned off to assure that the module does not overheat inside     Probable cause Suggested actions   1 Fan cable disconnected  Please contact your Agilent service  representative    2 Defective fan  Please contact your Agilent service  representative    3 Defective main board  Please contact your Agilent service  representative     1290 Infinity Thermostat User Manual 45    5 Error Information    Leak    Error ID  0064     leak was detected in the module     The signals from the two tem
49. stat can  safely flow into the leak funnel  Otherwise no special precautions are  needed for the modules     Heavy weight   The module is heavy  20 7 kg  45 6 Ibs     Carry the module at least with 2 people     gt  Avoid back strain or injury by following all precautions for lifting heavy objects    gt  Ensure that the load is as close to your body as possible      gt  Ensure that you can cope with the weight of your load        CAUTION    Mechanical damage of the module    If the transport assembly is not parked  the module could be damaged due to excessive  shock of the shipping container during transport      gt  Always park the transport assembly before shipment        32    If the autosampler or fraction collector needs to be shipped to another  location via carrier  ensure       The two modules are shipped in separate boxes       The transport assembly of the autosampler or fraction collector is  parked  see  Park Arm  Park Gripper   in your respective Service  Manual for more information     The vial tray is secured     1290 Infinity Thermostat User Manual    1290 Infinity Thermostat User Manual    4  Troubleshooting and Diagnostics    Agilent Lab Advisor Software 34    Overview of the Thermostat s Indicators and Test Functions 35  Status Indicators 36   Power Supply Indicator 36   Module Status Indicator 37    RE Agilent Technologies    33    4    Troubleshooting and Diagnostics    Agilent Lab Advisor Software    34    The Agilent Lab Advisor software is a standalone 
50. t Technologies sales and  service office     Table3 Thermostat Checklist       Description Quantity Part Number  Thermostat 1   Power cable 1 as ordered  Accessory kit  see  Accessory              1 61330 68755    page 63 for more information     HPLC System Tool Kit  optional  see 1 G4203 68708   HPLC System Tool Kit  on page 64 for  more information        1290 Infinity Thermostat User Manual    Installing the Thermostat 3  Optimizing the Stack Configuration    Optimizing the Stack Configuration    If your autosampler fraction collector is part of a system  you can ensure  optimum performance by installing the autosampler fraction collector in  the stack in the position shown in Figure 3 on page 19 and Figure 4 on  page 20  This configuration optimizes the system flow path  ensuring  minimum delay volume  As the autosampler thermostat is very heavy  it  should always be installed at the bottom of a stack     Instant Pilot       Detector le         Column compartment TL  rims       IN                                      Solvent cabinet       Degasser  optional                                                  Autosampler  ALS  Fraction collector        Thermostat for the ALS Fraction collector  optional     Figure 3 Recommended Stack Configuration  Front View     1290 Infinity Thermostat User Manual 19    3 Installing the Thermostat  Optimizing the Stack Configuration    LAN to control software    RAN Bus cable    to Instant Pilot   xp    u E              Autosampler  Fracti
51. talling the Thermostat    CAUTION Damaged electronics    Disconnecting or reconnecting the autosampler  fraction collector to thermostat cable  when the power cords are connected to either of the two modules will damage the  electronics of the modules  In such a case  mainboards of both instruments must be  exchanged  otherwise they can damage the other instrument      gt  Make sure the power cords are unplugged before disconnecting or reconnecting the  autosampler  fraction collector to thermostat cable        Stage 1  Preparing the Thermostat and Autosampler  Fraction  Collector         Damage through condensation    If the condensation tube is located in liquid the condensed water cannot flow out of  the tube and the outlet is blocked  Any further condensation will then remain in the  instrument  This may damage the instruments electronics      gt  Make sure the condensation tube is always above the liquid level in the vessel    gt  Make sure the waste container is not sealed      gt  Ensure the drain tube has no loops and leads directly into the waste bottle        Even under average humidity conditions  a significant amount of condensed water gathers  every day  A suitable container must be provided and emptied regularly in order to avoid  overflow     1 Place the thermostat on the bench or in the stack     2 Remove the front cover  Press the two snap fasteners on the sides of  the cover and move it away     22 1290 Infinity Thermostat User Manual    Installing the Therm
52. ter        Diffusion plate    9 Re install the tray     24 1290 Infinity Thermostat User Manual    Installing the Thermostat 3    10 If there is no Agilent 1260 1290 Infinity Series module located beneath  the thermostat connect the waste tube to the central waste exit of the  autosampler  fraction collector and place in a waste vessel     Stage 2  Power Cable and Interface Cable Connection    CAUTION Damaged electronics    Disconnecting or reconnecting the autosampler  fraction collector to thermostat cable  when the power cords are connected to either of the two modules will damage the  electronics of the modules  In such a case  mainboards of both instruments must be  exchanged  otherwise they can damage the other instrument      gt  Make sure the power cords are unplugged before disconnecting or reconnecting the  autosampler  fraction collector to thermostat cable        1 Ensure the power switch on the front of the autosampler  fraction  collector is OFF and the power cables are disconnected     2 Connect the cable between the autosampler  fraction collector and the  thermostat  see Figure 8 on page 26     3 Move the safety lever at the rear of the two modules to the right  position  see Figure 8 on page 26     4 Connect the power cables to the power connectors     5 Connect the CAN interface cables to other modules in the system  see  Figure 4 on page 20 and Figure 9 on page 27      6 If required  connect additional interface and control cables to the  autosampler  see 
53. the  instrument  Both are located on the autosampler module  The status  indicators provide a quick visual check of the operation of the  thermostatted autosampler  see    Status Indicators    on page 36      Error Messages    In the event of an electronic  mechanical or hydraulic failure  the  instrument generates an error message in the user interface  For details  on error messages and error handling  please refer to the Agilent Lab  Advisor software     1290 Infinity Thermostat User Manual 35    4 Troubleshooting and Diagnostics    Status Indicators    Two status indicators are located on the front of the autosampler  The  lower left indicates the power supply status  the upper right indicates the  thermostatted autosampler status        Status indicator    Power supply indicator    Figure 14 Location of Status Indicators    Power Supply Indicator    The power supply indicator is integrated into the main power switch   When the indicator is illuminated  green  the power is ON     36 1290 Infinity Thermostat User Manual    Troubleshooting and Diagnostics 4    Module Status Indicator    The Thermostat itself has no status indicator  Any Error condition will only show up on the  Autosampler Fraction Collector status LED s     When the status indicator is OFF  and power switch light is on   the  module is in a prerun condition  and is ready to begin an analysis        green status indicator  indicates the module is performing an analysis   run mode      A yellow indicator ind
54. tly for analysis     2 Defective remote cable  Exchange the remote cable   3 Defective components in the instrument Check the instrument for defects  refer to the  showing the not ready condition  instrument s documentation      1290 Infinity Thermostat User Manual    Error Information 5    Lost CAN Partner    Error ID  0071    During an analysis  the internal synchronization or communication  between one or more of the modules in the system has failed     The system processors continually monitor the system configuration  If one  or more of the modules is no longer recognized as being connected to the  system  the error message is generated     Probable cause Suggested actions  1 CAN cable disconnected    Ensure all the CAN cables are connected  correctly       Ensure all CAN cables are installed    correctly   2 Defective CAN cable  Exchange the CAN cable   3 Defective main board in another module  Switch off the system  Restart the system  and    determine which module or modules are not  recognized by the system     Leak Sensor Short    Error ID  0082  The leak sensor in the module has failed  short circuit      The current through the leak sensor is dependent on temperature  A leak  is detected when solvent cools the leak sensor  causing the leak sensor  current to change within defined limits  If the current increases above the  upper limit  the error message is generated     Probable cause Suggested actions  1 Defective leak sensor  Please contact your Agilent service 
55. trical Connections 82  Early Maintenance Feedback 85    10 Appendix 87    General Safety Information 88   The Waste Electrical and Electronic Equipment  WEEE  Directive   2002 96 EC  91   Radio Interference 92   Sound Emission 93   Agilent Technologies      Internet 94    1290 Infinity Thermostat User Manual    1290 Infinity Thermostat User Manual    Introduction    Introduction to the Thermostat    6  Thermostat Operation    8    de Agilent Technologies    1    Introduction    Introduction to the Thermostat    The Agilent 1290 Infinity Thermostat is designed for use with other  modules of the Agilent 1200 Infinity Series or with other LC systems if  adequate remote control inputs and outputs are available  The thermostat  is controlled from the Agilent 1200 Infinity Series Instant Pilot or from  your Agilent control software for LC systems     The specially  designed thermostattable sample trays holds either  100 x 1 8 mL vials or two wellplates and 10 x 1 8 mL vials     The thermostat contains Peltier  controlled heat exchangers  A fan draws  air from the area above the sample vial tray of the autosampler and then  blows it through the fins of the cooling heating module  There it is cooled  or heated according to the temperature setting  The thermostatted air  enters the autosampler through a recess underneath the specially  designed  sample tray  The air is then distributed evenly through the sample tray  ensuring effective temperature control  regardless of how many vials
56. urce of a not ready condition  Restart the  analysis where required     41    5    5    42    Error Information    Shutdown    Error ID  0063         external instrument has generated a shutdown signal on the remote  line     The module continually monitors the remote input connectors for status    signals  A LOW signal input on pin 4 of the remote connector generates  the error message     Probable cause Suggested actions   1 Leak detected in another module with a Fix the leak in the external instrument before  CAN connection to the system  restarting the module    2 Leak detected in an external instrument Fix the leak in the external instrument before  with a remote connection to the system  restarting the module     3 Shut down in an external instrument with a Check external instruments for a shut down  remote connection to the system  condition     Remote Timeout    Error ID  0070    A not ready condition is still present on the remote input  When an  analysis is started  the system expects all not ready conditions  for  example  a not ready condition during detector balance  to switch to run  conditions within one minute of starting the analysis  If a not  ready  condition is still present on the remote line after one minute the error  message is generated     Probable cause Suggested actions  1 Not ready condition in one of the Ensure the instrument showing the not ready  instruments connected to the remote line  condition is installed correctly  and is set up    correc
57. user settable EMF limits for the EMF Counters enable the early  maintenance feedback to be adapted to specific user requirements  The  useful maintenance cycle is dependent on the requirements for use   Therefore  the definition of the maximum limits need to be determined  based on the specific operating conditions of the instrument     Setting the EMF Limits    The setting of the EMF limits must be optimized over one or two  maintenance cycles  Initially the default EMF limits should be set  When  instrument performance indicates maintenance is necessary  take note of  the values displayed by the EMF counters  Enter these values  or values  slightly less than the displayed values  as EMF limits  and then reset the  EMF counters to zero  The next time the EMF counters exceed the new EMF  limits  the EMF flag will be displayed  providing a reminder that  maintenance needs to be scheduled     1290 Infinity Thermostat User Manual 85    9 Hardware Information  Early Maintenance Feedback    86 1290 Infinity Thermostat User Manual    1290 Infinity Thermostat User Manual    10  Appendix    General Safety Information 88    The Waste Electrical and Electronic Equipment  WEEE  Directive   2002 96 EC  91    Radio Interference 92  Sound Emission 93    Agilent Technologies on Internet 94    RE Agilent Technologies    87    10 Appendix    General Safety Information    Safety Symbols    Table4 Safety Symbols       Symbol Description       The apparatus is marked with this symbol when the us
    