Home
        servicing procedures - Diagramasde.com
         Contents
1.                                                       Pre use Checks    Carry out the Pre use Checks listed in Section 6 of the A100  User Manual     NOTE  A  The A100SP is not fitted with a PEEP valve     B  Check the operation of the Bag Vent switch detector   see  section 3 5 14 in the AV900 V4 ventilator user manual     Ordering Information    Retrofit Kits    To retrofit an A100SP absorber to an existing Prima SP  not  applicable to Prima   order one of the following kits     58113 A100SP  UK specification  retrofit kit   Prima SP102  58114 A100SP  US specification  retrofit kit   Prima SP102  58115 A100SP  UK specification  retrofit kit   Prima SP101  58116 A100SP  US specification  retrofit kit   Prima SP101    Each kit consists of an A100SP absorber  complete with AV series  bellows assembly  and mounts     Accessories    Fresh gas hose 52590  Manometer 58112  Breathing circuit  black rubber  58436  Absorber detachables 52582  Note    1  Absorber detachables consists of     3 litre breathing bag   1 05 m  42 inch  breathing tube  Female mount   Connector mount   Facemask elbow   Y piece    2  All standard connectors are for use with 30 mm taper anti pollution    systems   Models compatible with 19 mm systems are available on request     155    C    0473    Cat  No  52793  Doc  No  PR SP 0204SM  March 2004    Copyright    Penlon Ltd 2004 All rights reserved     Penlon Limited  Abingdon Science Park  Barton Lane  Abingdon  OX14 3PH  UK     Service Tel  44  0  1235 54707
2.                  gt   lt  gt   lt  gt   lt  gt                O2 Air CO2 N20  Pneumatic pressure Gas cut off valve Flowmeter  source  normally open   Filter Vaporizer    Pressure gauge    Pressure regulator     l    Pressure relief valve    Audible alarm     lt  K Cy    Restrictor    ap  ae  y  i  i    Pneumatic switch   optional Air N20  interlock     Reservoir    Pneumatic on off  switch    Flow control valve   variable     Oxygen flush valve    Non return valve    Power take off point   or test point     Mechanical  connection    Visual indicator          DESCRIPTION    3 4 Gas Supply Safety  Devices    3 4 1 Gas Supply Cut off Device  A gas cut off device  triggered by low oxygen  supply pressure  cuts the supply of nitrous  oxide  and carbon dioxide  if fitted      The cut off operates when the oxygen  pressure falls to 186  14 kPa  27  2 psig      Gas supplies are reinstated only when the  oxygen supply pressure rises above 227   14 kPa  33  2 psig      3 4 2 Oxygen Supply Failure  Warning Whistle   A whistle gives an audible warning when   there is a reduction of oxygen supply   pressure    Operated solely by the remaining oxygen in   the machine system  the warning whistle is   prolonged by an oxygen reservoir built into   the gas circuit  allowing a minimum warning   whistle of 7 seconds duration     The whistle will start to sound when the  pressure falls to 200  21kPa  29  3 psig    and will continue to sound until the pressure  falls to approximately 70 kPa  10 ps
3.                 eX ObLPOL  03     30vdS  G3yvO4 SEN    hting System   Wiring Diagram    Ig    L    HOLIMS  JOULNOD  WALSAS    N    eel y          LX ZOZPOL  ONISNOH diyo  29540 AVM 9    WIE 9910 29510 AYME               TANVd 1V3lY 19313           3       142    iagram    D    Wiring    S13 1LNO XNV ON  NOLINOW ZO       14S  HOLINOW ZO ON    SLA LLNO XNv  Z WS    ZLLZOL          SLSTLNO XNV   HOLINOW ZO  L WS    YOSNIS       3SVO YOLINOW20  YOSNIS    Z 0 9L JYV YOLINOW ZO NO S 2 t SHOLOSNNOD OL SIJYIM      QALVLS ASIMYSHLO SSJINN   SIdIYLS MINY 9 TATIVeVd JYV SIJYIM HLYWS TIV       3N18  N    NMOS  71    Z 0 bZ SI ONIMIM SNIVW O3dS 93  8 AN        BLIHM   N    MOW1E   1    GaAOUddY VSO Smeg SI ONINIM SNIVW D3dS WSO       OYA ASSV JLVIYdOYddY 33S ONIMIM TANVd TWOINLO373 YOS         DNIIM          ASSY 89d TELON  YOLINOW  O vee SON    AWM  Y3ONIG           ZELPOL  OMd 3343  AVZ Y30ONIG                        1X 9ZZP0L  Ny    Sat    FO AVMS                teZE0L SGvaS pit ZX  766001 Y3C1IOH    SE0P0L HVS  EL SZL ISN    lt  lt                ezreot uwagi    anis  HLYv3                NNOO 92   YILT    dn  ONINIVLJY  QW31 931       HOLIMS  TOXLNOD   W3LSAS anis  Hiya    XO8 NIVW  ants    D    HLAVA   pX ezzeoL 3 TaNvd  ONIY Pa     4x E2ZEOL   died SiN ONY    SWYS SW             6zS008   VSO   829008  N   v1 SNIVW YOLINOW ZO EGO  S 3SN4 SZL    ere                  3 SoH LVZ xz       N    30vdS  91      wwe 9  30vdS b t  Y3SC10OH3SNS     60LP01  bZE0L  v66E0L    3OvdS b t JAYS  9b E
4.            PRE USE CHECKS   All models       5 6 Patient Breathing  System    5 6 1 Hose Connections    Check that all hoses  connected  as illustrated     are correctly    5 6 2 A100   A100SP Absorber    Always follow the pre use check  procedures given in the instruction  manual supplied with the absorber    The use of an oxygen monitor  and a carbon  dioxide analyser  is highly recommended  when using any partial rebreathing  anaesthetic system     5 6 3 Breathing System Hose   Reservoir Bag  Ventilator    Refer to illustrations on next page   Connectors for the Inspiratory hose  A  and  Expiratory hose  B   and the reservoir bag  connector  C  are 22 mm male    All connectors comply with ISO 5356 1     The ventilator connection point  D  is also 22  mm male     Hose and bag connections are fitted with  Penlon Safelock high security fittings     Check all connections for gas tightness     5 6 4 Fresh Gas Supply    Refer to illustrations on next page   The fresh gas hose assembly  E  supplied  with the machine has a Penlon connector at  the absorber inlet and a 22 mm Safelock  taper at the other end    This should be connected to the common  gas outlet  F  of the anaesthetic machine     Check all connections for gas tightness     38                   Prima SP  with A100SP  Absorber  For additional  information   refer to  Appendix 4                                                           A100SP   AV900 V4 connections    1     2     Inlet   from DRIVE GAS outlet on  AV900 
5.        Turn this way  to remove    SERVICING PROCEDURES    9 16 Oxygen Failure    Visual Indicator    CAUTION  Ensure that any residual gas in the machine is exhaust   ed before commencing the dismantling procedure     Removal    1     Remove all cylinders and disconnect all  pipeline supplies     2  Remove the front cover  see 9 6     3  Disconnect the tubing at the rear of the  visual indicator  1     4  Remove the cover at the back of the  machine   5  Remove the screws  2  to release the  switch mounting bracket  3     6  Unscrew the locking ring  4   and detach  the visual indicator   Refitting   8  Reverse the removal procedure     Fit the front cover   see section 9 6   Carry out a Function test  section 6     103                            r  To        PRN       O                   SERVICING PROCEDURES    9 17 CGO Block    Clippard Valve Replacement    CAUTION  Ensure that any residual gas in  the machine is exhausted before  commencing the dismantling  procedure     Removal   1  Remove all cylinders  and disconnect all  pipeline supplies    2  Depress the button  1   to allow access at  2   to the grubscrew that  retains the button    3  Slacken the grubscrew   using a M1 3 Allen key   and gently prise out the  button    4  Remove the two screws  3   and withdraw  the shield  4   valve  5  and O seal  6                  Refitting  5  Lightly coat the seal  6  with Fomblin   6  Reverse the removal sequence     Carefully refit the Clippard Valve  5  to  prevent damage 
6.     0293  O seal    0508 Filter             0226  O seal  462526  Hose   Backbar  Outlet to CGO  400262  Bobbin 01057  i es Washer  HE 36247  Bobbin                                     36247  Bobbin       400224  Bobbin    Cascade flowmeter system illustrated  On non   cascade models  a filter  0508  is fitted to each  flow tube     115    PARTS LIST       Cagemount Backbar    Preventive Maintenance Kit    Do NOT apply grease  72 Month Service or lubricant to    016  Bodok Seal  Part No  57489 043  O seal   Safelock    Kit Contents   Part No Qty  Component Description 043 O Seal   Safelock   462526 1 35m Hose   Backbar Outlet to CGO 011017 Tubing   Sample Block   0226 O Seal   Hose Assembly 045438 Secondary Regulators   Mech AHD  37946 Hose Support 90183 Clippard Valve   Oxygen Flush  37947 Seal Support 0314 O Ring   Oxygen Flush Clippard  37951 Split Ring Valve   37950 Off Line Hose 0691 6 O Ring   Cylinder Pipeline NRV   016 Bodok Seal Following not shown    400224 Flowmeter Bobbin   Metal with Filter 462544 3m 8 mm Hose   Oxygen   Internal  400262 Flowmeter Bobbin   Metal without 462543 2m 6 mm Hose   Nitrous Oxide   Internal  Filter 462542 2m 5 mm Hose   Air   Internal   0762 Filter   Yoke and Pipeline 462541 1m 4 mm Hose   Switch Feeds COz    0508 Filter   Flowmeter Internal   36247 Flowmeter Bobbin   Tube Seals 01089 1strip   Gas ID Label White   Oxygen  01057 Washer 011026 1strip   Gas ID Label Blue   Nitrous Oxide  36075 Shock Washer 011027 1strip Gas ID Label Black  
7.     30VdS  p b     NI ONG       1N 301S    SNOILOJNNO9  aSN4  SNIVW TWOIdAL     zS008   VSO     zS008  AN    ASSY GVA SNIVW  N                      ISO AVS ANo S10 s9910 AYME     6me gt  Lee00e   VSO   6me 91  9ze00      XN       WOOT Y3MOd    YOLINOW ZO   200 06zz sasna  OMHVEL wwz    5          TANVd  SSA990V  WOILYSA    TANVd 1V3lY 19313        j       143    Appendix    APPENDIX 4  Fixings Tightening Torques    Note     Only suitably skilled and trained persons should carry out service and  maintenance procedures on this anaesthesia equipment     Specific torque values are given where required     Where specific torque values are not provided  screws and fittings should be  tightened using a reasonable degree of tightness and skill  Where torque    guidance is required  the following values may be used     1 General tightening torques   Diameter Type  M3 Cap Head screws    Csk  amp  Button head screws    M4 Cap Head screws  Csk  amp  Button head screws  M5 Cap Head screws  Csk  amp  Button head screws  M6 Cap Head screws  Csk  amp  Button head screws  M8 Cap Head screws  Csk  amp  Button head screws  2 Specific torque values  Diameter Where used  M4 Cylinder bump stop  M6 Cylinder Yokes  M6 Yoke mounting plate into rivet bushes  M8 Top tray fixings  M8 Work surface to vertical extrusion  M16 Castor fixings  CGO Taper fitting  M8 Base to vertical extrusions  M6 Vertical extrusion T nuts  M6 Main control box to vertical extrusions  M3 AHD gear grub screws  M4 Front panel  M
8.     Battery Back up   Sensor Type     High Priority Alarm   Medium Priority Alarm   Low Priority Alarm   Alarm Mute     Low Alarm Set Range   High Alarm Set Range     Cable length     Sensor  Type   Life     0   100     1     2  of full scale  at constant temperature and pressure   90  of final value in approx  10 seconds  air to 100  O2      10 to 38  C  50 to 1000F     5  C to 50  C  23  F to 122  F     5  C to 50  C  23  F to 122  F    5  to 95   non condensing     See section 4 11  MOX 3 galvanic fuel cell    Flashing  5 audio pulses with 6 seconds repeat time   Flashing  3 audio pulses with 24 seconds repeat time  Static with single beep sound   30 seconds for high priority alarm   120 seconds for medium priority alarm   18  99       1     19  105       1      2 m  6 ft   fully extended    Galvanic fuel cell sensor  0  100      1 500 000 O2 percent hours at 20  C   One year minimum in typical applications     Interference Gases and Vapours  in 30  Oxygen  70  Nitrous Oxide     Interference    Nitrous Oxide  Carbon Dioxide  Halothane  Enflurane  Isoflurane  Sevoflurane    Humidity Effects    Volume   Dry Interference in 02     80   lt 1   5   lt 1   5   lt 1   5   lt 1   5   lt 1   5   lt 1     Sensor output is relatively unaffected by prolonged operation in either high or very low relative    humidity     If the sensor shows signs of being affected by condensation  dry the sensor with soft tissue     CAUTION    DO NOT use heat to dry the sensor     29    SPECIFICATION   
9.     Machines with Air sup   ply   NeO Air Interlock  switch correctly set     Remedy   Ensure correct gas  cylinder is attached   Check index pins not  loose or damaged   Replace gauge     Replace gauge     Open flowmeter and  empty system     Replace plunger    O     seal assembly     Test for leaks     Replace gauge   Replace as necessary   Replace as necessary     Contact hospital engi   neer     Replace as necessary     Contact hospital engi   neer     Check cylinder pipeline  connections     See instructions under  component group    See instructions under  component group     Tighten     Check position of  switch     FAULT FINDING       Component    Flowmeter units   Mechanical AHD only     Symptom    Float stays at bottom of  tube when control valve  opened NzO flowmeter     Restricted flow through  flowmeter  Control   valve opening does not  achieve expected flow     Incorrect O2 concentra   tion     62    Possible Fault    Gas delivery switch left  in off position     Gear linkage loose on  Oz or N20 primary  valve shaft     Gear linkage loose on  valve shaft     N20 flow limited by  N20 control valve     Secondary regulators  out of alignment     Gear linkage loose on  valve shaft     Remedy   Turn on gas delivery  switch   Reset gears and tight     en    Reset gears and tight   en    Open N20 control valve    Check and reset sec   ondary regulators   Replace if necessary    Reset gears and tight   en    FAULT FINDING       Component Symptom    Flowmeter unit Float
10.     O2 Monitor    5 12 6 Set Alarms Set Low Alarm  The low alarm value cannot be set lower  Set High Alarm than 18   or above 99      The high alarm value cannot be set below  19   or above 105      Note that in certain conditions of excess  pressure  the readout may show a value  above 100        1  Press the HIGH ALARM SET button     1  to enable high alarm set mode  2   The HIGH ALARM LED  2  will flash   and a single    beep    warning will 5    sound     2  Turn the adjacent    SET    knob  3  to  the desired setting     3  Press HIGH ALARM SET button  1   again   The unit will then exit the high alarm  set mode   The LED display  2  will now stop  flashing  and a single    beep    warning  will sound     48    1     Press the LOW ALARM SET button   4  to enable low alarm set mode   The LOW ALARM display  5  will  flash  and a single    beep    warning will  sound     Turn the adjacent SET CAL knob  6   to desired setting     Press LOW ALARM SET button  4   again    The unit will exit low alarm set mode   The LED display will now stop  flashing  and a single    beep    warning  will sound     6  FUNCTION TEST       6 1 Introduction    Every PrimaSP anaesthetic machine is  tested thoroughly before leaving the factory   but it is essential to undertake a complete  performance check as described in this  section before the machine is first used  Clinically    This can be done initially by a Penlon trained  engineer as part of the commissioning of a  new machine  Subsequentl
11.    Prima SP Anaesthetic  Machine Range    Service Manual                                                 Quality and Assurance in Anaesthesia    THE IMPORTANCE OF PATIENT MONITORING    WARNING    Anaesthetic systems have the capability  to deliver mixtures of gases and vapours  to the patient which could cause injury or  death unless controlled by a qualified  anaesthetist     There can be considerable variation in  the effect of anaesthetic drugs on indi   vidual patients so that the setting and  observation of control levels on the  anaesthesia systems does not in itself  ensure total patient safety     Anaesthesia system monitors and patient  monitors are very desirable aids for the  anaesthetist but are not true clinical moni   tors as the condition of the patient is   also dependent on his respiration and the  functioning of his cardio vascular system     IT IS ESSENTIAL THAT THESE ELEMENTS  ARE MONITORED FREQUENTLY AND  REGULARLY AND THAT ANY OBSERVA   TIONS ARE GIVEN PRECEDENCE OVER  MACHINE CONTROL PARAMETERS IN  JUDGING THE STATE OF A CLINICAL  PROCEDURE     IMPORTANT    Servicing and Repairs    In order to ensure the full operational life of  this anaesthetic machine  servicing by a  Penlon trained engineer should be  undertaken periodically     The machine must be serviced to the  schedule detailed in section 8    Details of these operations are given in this  Service Manual  available only for Penlon  trained engineers     For any enquiry regarding the servicing o
12.    Use soap solution to detect leak  repair  and repeat above tests  5  6  amp  7      Repeat the above tests  5  6  amp  7  with  all fitted vaporizers turned on  Test one  at a time  to check gas tightness of  each individual vaporizer     A leaking vaporizer must be withdrawn  from use    Return the unit to Penlon  UK  Customers  for repair  or arrange  repair by a Penlon Approved Dealer   Non UK Customers      58    6 10 Emergency Oxygen Flush    1    6 11    Control Test    Ensure all flowmeter control valves  are turned off     Oxygen Cylinder is turned on and  Oxygen Pipeline is unconnected from  the wall outlet     Depress the Emergency Oxygen  Flush  O2 Flush  button on the CGO  block     Check that the delivered flow rate is  between 35   75 L min     Ensure the spring action cuts off the  Oxygen flow when the button is  released     Check that the whistle does not sound  when the Oxygen flush is operated    If this occurs check that the Oxygen  Cylinder Valve is fully open    If the problem is not rectified  instruct  the service engineer to check    a  filters and Non return valves for  restriction    b  the pressure of the primary Oxygen  regulator     Connect the oxygen pipeline and  close the cylinder valve and repeat  tests 3   6     Auxiliary Outputs   Test    Check outputs for security and correct  function    With a suitable flowmeter measure the  output from each outlet to ensure that  greater than 100 litres min can flow to  free air    Ensure that the ox
13.    block   may be mounted on a universal back   bar manifold  built onto the Prima SP range   as an option     Single and two station manifolds are  available  with each station fitted with two  valve capsule assemblies for vaporizer  connector block attachment     When a vaporizer is installed on a station the  valves on that station open automatically to  allow gas flow into and out of the vaporizer   Removal of the vaporizer from the station  closes the valves on that station     Selectatec compatible vaporizer interlock  systems are described in the literature  supplied with the vaporizer     3 8 3 Cagemount Vaporizer  Vaporizers fitted with cagemount tapers  have the male taper  inlet port  on the left   and the female taper on the right  viewing  the front of the vaporizer      It is recommended that detachable  cagemount connectors are retained with a  safety clip  catalogue number 52275  to  prevent inadvertent disconnection     DESCRIPTION    3 9 Common Gas Outlet  CGO   Block    The CGO block is mounted on the rail on the front  of the machine  and can be moved along the rail   Slacken the securing screw  1  under the block and  carefully slide the block along the rail to the required  position    Tighten the screw to hold the block in place     The fresh gas outlet  2  is located on the front face  of the block  with 22 mm male taper and concentric  15 mm female taper  The male taper incorporates  the Penlon Safelock system designed to prevent  accidental disconnec
14.   0293  O seal             36247  Bobbin       400224  Bobbin       0508 Filter Cascade flowmeter system illustrated     On non cascade models  a filter  0508  is fitted  to each flow tube     120    PARTS LIST       Mechanical AHD    Selectatec Backbar  Preventive Maintenance Kit    72 Month Service  Part No  57490    Kit Contents  Part No Qty   462526 1 35 m  0226  041115  016  400224  400262  0762  0508  36247  01057  36075  0691  043  011017  045438  022533    Component Description   Hose   Backbar Outlet to CGO  O seal   Fresh Gas Sampling Adapter  O ring   Selectatec   Bodok Seal   Flowmeter Bobbin   Metal with Filter  Flowmeter Bobbin   Metal without Filter  Filter   Yoke and Pipeline   Filter   Flowmeter   Flowmeter Bobbin   Tube Seals  Washer   Shock Washer   O seal   Flowmeter   O seal   Safelock   Tubing   Sample Block   Secondary Regulator   Mech AHD  Dzus Clip    MMH H OMDRODOWAWWUAAA    37947    gio Seal Support    Split Ring    l Difi c    NN AK FFARR   RARASAANANA INANI  yaa             0314     O ring Seal  A   s  F  Ce      90183             d    We Y    Clippard Valve  CO       121    90183  0314    0691 6  37946 4  37947 4   37951 4  Following not shown   462544 3m  462543 2m    462542 2m  462541 1m    01089   011026  011027  011023    1strip  1strip  1strip  1strip  Dioxide       Do NOT apply grease or  lubricant to    016  Bodok Seal   043  O seal   Safelock  041115  O ring  Selectatec    Clippard Valve   Oxygen Flush  O ring   Oxygen Flush Clippard  Val
15.   7  Check that the flow of Air has stopped    8  Check that N20 is now delivered     Continue Pre use Checks at  section 5 4    34    PRE USE CHECKS   Machines with Mechanical AHD       5 3 Pre use Checks   Gas Supply   Machines with Mechanical AHD     5 3 1 Gas Pipeline Supplies   Machines with  Mechanical AHD   1  Connect the oxygen pipeline hose only    Check that the correct pressure gauge reading is  obtained    2  Turn on the Gas Delivery switch  A     Check that the warning whistle sounds briefly  and  that the correct basal flow of oxygen is delivered  see  section 4 8     3  Open both oxygen and nitrous oxide flowmeter  valves    NOTE  On machines with Air  and the optional  Air N2O interlock system  set the switch  B  to N2zO   see section 5 3 4     Check that flow is only shown in the oxygen  flowmeter    4  Close both valves  Turn off the Gas Delivery switch   Check that the warning whistle sounds briefly  and  that the oxygen basal flow is stopped    5  Connect the other pipeline hoses  in turn    Check the gauge reading for those gases   Check that gas flows when the relevant needle valve  is operated    6  On machines with Air  and the optional Air N2O  interlock system  set the switch  B  to Air  see section  5 3 4     NOTE   If the machine is not equipped for connection to a pipeline supply    carry out the above procedures using the cylinder supplies     5 3 2 Gas Cylinder Supplies    CAUTION Open the cylinder valves slowly to avoid damage to  the pressure 
16.   Air Selection switch is positioned for N2O     0  Fully open N20 Flow Control Valve  Check that NO flow of N2O occurs   11  Slowly operate O2 control valve throughout range and check that N2zO now flows  12  Operate N2O Air selector switch to Air position and check that N2O ceases to flow  and that Air is now available   13  Return switch to the N20 position and set the oxygen flow back to minimum      O NOOARYW    14  Refer to the table below  Check   O2 concentration at the oxygen flows specified   Adjust the O2 Flowmeter control only   Do NOT adjust the N20 flowmeter     Note The flow rates for N2O are given as a guide only   O2 Flow  02  in O2 N20 mixture  N20 Flow   0 25 L min 27   33   0 5   0 67   0 5 L min 27   33   1 0   1 35  0 8 L min 27   33   1 6   2 1  1 5 L min 27   33   3 0   4 1  3 0 L min 27   33   6 1   8 1      Cascade flowmeter models   March 2002 onwards    15  Repeat tests by reducing Oxygen flow back to Basal Flow conditions  referring to the  values in the above table   16  Check Gas Delivery Switch for correct operation     Oxygen Fail Alarms    1  Simulate an Oxygen fail by unplugging the Oxygen Pipeline from the outlet   Check that before the supply of oxygen is exhausted that the Oxygen Fail Whistle  sounds for a minimum of 7 seconds    Check that any Nitrous Oxide and Carbon Dioxide  if fitted  Gas Flows are cut off    2  Check that the Oxygen Supply Visual Indicator turns from fully green to fully red    3  Turn on Oxygen cylinder    Check Oxy
17.   Always follow the procedures and checklist  given in the instruction manual supplied with  the vaporizer  particularly when filling the  vaporizer with anaesthetic agent     WARNING   Vaporizers must always be mounted  never  used free standing    Free standing vaporizers may be accidentally  tipped resulting in excessive and uncalibrated  volumes of anaesthetic drug entering the  breathing system     Do not install or connect any vaporizer of any  description between the CGO and the  breathing system  unless it is specifically  designed for such use   This allows the  oxygen flush flow to pass through the  vaporizer  and severe overdosage may result      5 10 1 Selectatec Mounting System  Dependent on choice of backbar manifold  system  up to two Selectatec compatible  vaporizers may be fitted    To install the vaporizer  carefully offer the  vaporizer up to the manifold    Check that the gas connection ports on the  vaporizer are aligned with the valves on the  manifold    Carefully lower the vaporizer onto the  manifold and lock the vaporizer into position  by clockwise rotation of the locking lever    through 909    NOTE Do not use excessive force to lock the  vaporizer onto the manifold  Damage to the  locking fastener will result    CAUTION   To prevent damage to the locking shaft  ensure  that the gas connection ports are aligned with the  valves on the manifold  and are correctly  engaged  before tightening the locking lever     5 10 2 Selectatec Compatible    Vapor
18.   Reduced Pressure Gas Circuit Leakage  Emergency Oxygen Flush Control Test  Auxiliary Outlets    FAULT FINDING  Introduction  Fault Diagnosis Table    SERVICE SCHEDULE   Service Schedule   Six Month Service Checks   Twelve Month Service Checks  Twenty four Month Service Checks  Seventy two Month Service Checks    SERVICING PROCEDURES   Servicing   Ancillary Equipment   Cleaning and Sterilisation   Gas System Components   Internal Gas Pipework   Front Cover   Remove and Refit  Flowmeter Cover   Mechanical AHD System   Basal Flow Check   O2 Concentration Check   Gear Linkage Reset  Control Valve Capsule Removal Replacement  Oxygen Reservoir   Removal Replacement  Secondary Regulators   Gas Block   Remove and Refit   Selectatec Compatible Manifold Block   Gas Delivery Switch   Air N2O Interlock Switch   Oxygen Failure Visual Indicator   CGO   Clippard Valve Replacement  Flowmeter Tubes   Removal   Flowmeter Assembly   Removal  Flowmeter Lighting   Removal   Oxygen Monitor   Battery Replacement  Oxygen Monitor Sensor   A100 Circle System Absorber   removal    SPARE PARTS LIST    APPENDIX   Care of Back up Battery   Product Classification and Labelling Terminology  Wiring Diagram   Lighting   Wiring Diagram   General   Tightening Torques   A100SP Absorber User Information     iv     59  60    69  69  72  75  78    110  111    140  141  142  143  144  145    USER RESPONSIBILITY       This anaesthetic machine has been built to  conform with the specification and operating  procedur
19.   This machine is fitted with an hypoxia guard known as a Mechanical Anti Hypoxic Device     MAHD     This device prevents the delivery of hypoxic mixtures to the patient and must only be  adjusted by a Penlon trained Engineer or other persons certificated to do so by Penlon  Limited    Incorrect adjustment of the MAHD could result in the failure of this device and the loss of  hypoxia protection     The oxygen control is restricted to prevent the needle valve from fully closing  This ensures  a minimum flow of oxygen at all times while the machine is switched on     This small flow of oxygen is known as the Basal Flow    DO NOT attempt to fully close the oxygen flow to zero    DO NOT overtighten Flow Control knobs for the other gases as this will reduce the life of  the needle valves        1  Ensure that all flowmeter control valves are  shut  Disconnect all the pipeline gas supplies     2  Remove all gas cylinders     3  Check each cylinder yoke for security of  attachment  and free movement of the clamp  bar  A  and clamp screw  B    Check Pin Index system pins  C  are secure  and undamaged     4  Check that a seal  D  is in place and  undamaged  in each cylinder yoke     5  Attach each gas cylinder to its appropriate  yoke  ensuring full engagement of the  Pin  Index  system  Note  Use full bottles only     6  Turn on the Oxygen Cylinder   Check that the Cylinder pressure is indicated  on the Oxygen Cylinder pressure gauge in a  smooth sweeping movement   Check that all 
20.   To prevent dirt entering the gas system   cylinder yokes and pipeline inlets are fitted  with filters     Gas Inlet Block   Each individual gas supply  from a cylinder  or pipeline  is routed through a separate gas  block    Each gas block has an integral high    pressure gauge tapping for direct mounting  of a pressure gauge  and a non return valve  to prevent back flow of gas     In addition  gas blocks for cylinder supplies  have a diaphragm pressure regulator to  reduce the pressure of the compressed gas  supply  and a pressure relief valve  factory  set to prevent any pressure build up under  the diaphragm should any leakage develop  across the reducing valve seat     Secondary Pressure Regulator   For oxygen  nitrous oxide  and air  a second  stage regulator reduces the pressure  supplied to each flowmeter controls  see  section 4 8      The fitment of a secondary regulator for  oxygen and nitrous oxide enhances the  performance of the mechanical AHD system   if fitted      Secondary regulation of the air supply is  also utilised to allow connection to high  pressure  7 bar  air pipeline supplies     Carbon Dioxide Flow Restrictor  On machines with a carbon dioxide supply   an integral  factory set  flow valve is fitted to  restrict the flow of carbon dioxide to 500  ml min     3 3 Gas Circuit Schematics    Gas circuit schematics for    a  Non AHD machines   b  Mechanical AHD machines   are shown on the following pages  and both  show a four gas machine    All availab
21.   Z    A 054528     Stem adaptor     8 mm tube    054528       Stem adaptor  8 mm tube    139    11  Appendix    APPENDIX 1    Machines with Oxygen Monitor  Care of Back up Battery    A  Batteries installed in machines   The batteries must be charged before the machine is released  for use with a 24 hour charge    Subsequently the recharge periods for batteries on machines  in store are similar to those in B  below     Batteries in machines in normal use will be kept charged by the  internal power supply and will only require special charging care  following the discharge test carried out during function testing     B  Battery care storage requirements     During storage the batteries will require a periodic recharge  the  frequency of which is determined by the storage temperature   which must not exceed 122  F  50  C      Storage Recharge  temperature period    38 to 50  C  100 to 122  F  1 month  21 to 38  C  70 to 100  F  3 months    7 to 210C  45 to 70  F  6 months  0 to 70C  32 to 45  F  9 months   5 to 0  C  23 to 32  F  12 months    Duration   recharge until the charge current is less than 25 mA   typically overnight      It is recommended that at each charge an updated label is    affixed to each battery to indicate date of the last charge     C  Disposal of used batteries    Used batteries must be disposed of according to hospital   local  state and federal regulations     140    Appendix    APPENDIX 2    Product Classification and Labelling Terminology  The terms C
22.   must be equal to  or below the oxygen  concentration     O2 MONITOR INOP  inoperative    This alarm indicates when the oxygen  monitor is in malfunction condition    The alarm can be triggered by electronic  components failure or software malfunction   In this alarm condition  a red O2 MONITOR  INOP ALARM LED will flash at a 0 5 second  rate  accompanied by a high priority sound    If this mode occurs you can reset the system  by pressing the ALARM MUTE and LOW  ALARM SET buttons simultaneously for 3  seconds     O2 SENSOR FAULT   The alarm is triggered    a  when either the oxygen sensor is  disconnected or approaching the end of its  life    b  if the the oxygen concentration exceeds  110     In the alarm condition  a red O2 SENSOR  FAULT ALARM LED will flash at a 0 5  second rate  accompanied by a high priority  sound    To cancel this alarm  check the sensor  connection or replace the sensor     O2 SENSOR LOW   This alarm indicates the sensor  approached the end of its life    The yellow O2 SENSOR LOW LED will light  up  and a low priority sound will be triggered     has    20    The sensor must be replaced as the output  will fall very quickly to zero within two to  three weeks of normal usage    See section 6 for sensor replacement   NOTE   To maintain maximum sensor life  always  remove from breathing circuit after use     LOW BATTERY   This alarm indicates that the battery is  disconnected or the battery voltage has  dropped below acceptable limits    If the monitor
23.  01250  12  700647  13  041202  14  700638  15 71553  16 019067  17 041115  18 041214  19 041222  20 041223  21 800482  Not illustrated   800481  054540    Description    Screw  M4 x 25 socket head  Leak check label   Left hand end block  Manifold   Valve capsule   Screw  M5 x 12 socket head  TEC 3 location button   Clip retaining plate   Dzus spring   Screw  M2 5 x 6   Nut  M2 5   Tie rod   O ring   Right hand end block  Interlock peg   Screw  M4 x 12 socket head  O seal   O seal   O seal   O seal    Manifold Block   two station   one piece    Manifold Block   one station   one piece  Coupling  inlet and outlet       These components are not used on later  one piece manifolds    132                        PRR HRANHWNHKRANNNNANHNA     e     e    PARTS LIST             Ref Part No  Description Qty    1 045055 Gas delivery switch assembly 1  054945 Oxygen supply visual indicator assembly 1  045069 Gas interlock switch assembly 1    133    PARTS LIST    Gauge Cover                      Ref Part No  Description Qty    1 800157 Cover   Prima SP 102 1  800255 Cover   Prima SP 101 1   2 019033 Screw  M4 x 40 Csk Cross SS 2   3 019123 Screw  M4 x 10 Bik 2    134    PARTS LIST                                  Oxygen Monitor    Ref Part No  Description Qty   1 700823 Battery 1  2 103472 Cable 1  3 102714 Sensor assembly 1    includes flow diverter and O ring  4 800512 Tee piece 1  5 043 O ring  for Safelock connector  1    135    PARTS LIST                pie Be  D pez     Hoses    Pri
24.  10  High pressure regulator  11  Fresh gas pressure relief valve  12  Pipeline supply inlet block  13  Cut off valve capsules   14  Secondary regulator  O2   15  Secondary regulator  N2O     84    Secondary regulator  Air    a  Test points are mounted beneath  each secondary regulator    b  On pre 2002 machines  the regu   lators are mounted vertically    Hose from flowmeter   Fresh gas hose   O2 flush hose   Manifold block   Test point   primary regulator  Warning whistle   Air N2O interlock switch   Oz failure visual indicator   Gas delivery switch       SERVICING PROCEDURES    9 4 Gas System Components    The illustration on the preceding page shows  the components visible when the front cover  is removed     Machines with COz    An additional high pressure block and regu   lator  and an additional cut off valve capsule  are fitted     9 5 Internal Gas Pipework  9 5 1 Identification    Internal pipework is colour coded and diam   eter indexed  as follows     Colour codes    Oxygen White  or Green  on US  specification machines    Nitrous Oxide Blue   Medical Air Black white  or Yellow  on US specification  machines     Carbon Dioxide Grey    Tubing diameter    Oxygen 8mm  Nitrous Oxide 6mm  Medical Air 5mm  Carbon Dioxide 4mm  Mixed Gas 10 mm    85    SERVICING PROCEDURES    9 5 2 Tubing Connectors    Push in type fittings are used throughout the  low pressure circuit    To connect and disconnect the tubing from  the connectors  follow the procedures  detailed below     D
25.  3  Check O2 Emergency Flush flowrate is 35   75 L min    Auxiliary Outlets   1  Check outlets for security  damage and correct lock and release movement   2  Check dynamic flowrate and that Oxygen Fail Whistle does not operate     Oxygen Monitor  if fitted   Function and Calibration Tests   1  Connect a Test O2 Analyser into the patient circuit   2  Check that the machine O2 sensor is inserted into the absorber O2 sampling point   3  Test both O2 sensors in 100  Oxygen   4  Expose both sensors to air and check reading is 21    2   Adjust high and low O2 alarms and check alarms trigger when values are lower or  higher than reading on O2 analyser respectively   Return the alarm levels to original settings   5  Restore the sensor to correct location   Remove the test O2 analyser     Final Gas Concentration Tests   1  Check concentration by Gas   2  Check flowrate from CGO corresponds with flowmeter by concentration   3  Set 3 L min Oxygen and Nitrous Oxide Flows   Check  using O2 Monitor that the concentration from the CGO is 50    2    4  Set Oxygen to 0 5 L min  set N2O to 0 5 L min   Allow O2 monitor to stabilise then check concentration is 50    2    If fitted repeat above test substituting Carbon Dioxide for N2O   If fitted set Air to 3 L min  set Oxygen to 3 L min   Check Concentration reading on O2 Monitor is 60    2      5   6     Test Completion   Paperwork   Remove all test equipment from machine   Drain all Gas from machine    Turn OFF Gas Delivery Switch    Turn Off
26.  Air   0691 O Seal   Flowmeter 011023 1strip   Gas ID Label Grey   Carbon Dioxide    ADADWO WWOAAAAA    37947  37951  Split Ring Seal Support    SERES  S  piesa       37950  Offline          0226  O seal  Offline Hose       Bodok Seal Q    N     0314  O ring d 0762    K A  Filter   Yoke  tet  Z Or         90183           ay i  y Wr  _       Clippard Valve   x Q 7  a         U   i 4 N  oR             0762  Filter   Pipeline    116    PARTS LIST    045438  Secondary Regulators             Pe VT                       LT    0691    f K   p VAH A  A Hf ay f  a   KN kG  PRA  X oP  A P EX    GH    1     O ring    H    Is mn  jS GEM    oS    D    LA   V PG   ALNE  UPE    Se    he  of    VAN          2    X  a3                          ie                   Bobbin            a   5  9 O  Is  229 T T T T  lo O v    T O TOO ymo  88   AS B   a2 88   12 oO o0 oO oO    Cascade flowmeter system illustrated     0508 Filter    On non cascade models  a filter  0508  is    fitted to each flow tube     117    PARTS LIST    Selectatec Backbar  Preventive Maintenance Kit    12   36   60 Month Service  Part No  57476           Do NOT apply grease or  lubricant to   Component Description 016  Bodok Seal   Hose   Backbar Outlet to CGO 043  O seal   Safelock  Hose Support 041115  O ring  Selectatec  Seal Support   Split Ring   Dzus Clip   O ring   Selectatec   Bodok Seal   Filter   Yoke and Pipeline   O seal   Safelock   Tubing   Sample Block    Kit Contents    Part No  462526  37946  37947  37951  0225
27.  Compatible  Manifold Block          1  The valve capsule top O seal   arrowed on the illustration  must  be regularly changed to prevent  the possible occurrence of leaking  Check for splitting  perishing  and  expansion  and also check if the  seal is sticking to the vaporizer        To ensure satisfactory performance   a  ALWAYS perform a leak test  before using the manifold block or i    after changing the vaporizer  9 F  b  Replace these O seals regularly dd 3     maximum 6 months interval   D   DO NOT apply grease or lubricant S    Overhaul   CAUTION Ensure that any residual gas in  the machine is exhausted before  commencing the dismantling procedure     Removal    1  Remove all cylinders and    disconnect all pipeline supplies   2  Remove the front cover   see  section 9 6   to gain access to the       manifold block  e  3  Disconnect the inlet and outlet   3  fresh gas hoses  9    4  Remove the screws at the rear of  d  4  the machine  to allow removal of D    the manifold block   oe    Overhaul and Refit  5  Dismantle the block as illustrated     Note that later Prima SP models  are fitted with a one piece  manifold block assembly    Renew all O seals    To refit  reverse the dismantling  procedure    Function test the machine   see    NOO       One piece manifold block    section 6  i   fitted to later Prima SP models  4      100    SERVICING PROCEDURES    9 14 Gas Delivery Switch    CAUTION  Ensure that any residual gas in the machine is  exhausted before commenci
28.  DO NOT apply grease or lubricant  Replace flexible tubing from back bar outlet to CGO   1 35 m  Replace vaporizer locking spring   Check all pillar valves open and seal correctly    Check pillar valve inserts sit proud or flush with the pillar valve   A sunken pillar valve insert denotes failure of the internal spring   Check vaporizer interlock system for correct operation     Back Bar Assembly   Cagemount   Checks     1   2   3     Replace flexible hoses    Replace flexible hose from back bar to CGO   1 35 m   Check security of taper connectors   Use Fomblin or other Oxygen safe grease to lightly lubricate the metal surface where  male and female connector come into contact with each other     Fresh Gas Pressure Relief Valve  PRV  Test     1   2     Occlude CGO with a manometer   At 10 L min Flowrate of Oxygen the PRV relieves at 41 kPa   10      76    SERVICE SCHEDULE       Leak Test From Flowmeter to Common Gas Outlet   1  Fit a suitable Manometer Test gauge to the CGO    2  Set Oxygen Flow to maintain a pressure of 20 kPa  3 psi    Maximum permissible flow is 200 ml min    3  Perform test with and without Vaporizer s  fitted and with vaporizer s  turned both ON  and OFF    4 Set oxygen flow to minimum and remove test equipment     Common Gas Outlet Checks  CGO     1  Check security and freedom of movement of CGO    2  Replace Safelock O ring  DO NOT apply grease or lubricant  3  Check O2 Emergency Flush flow rate is 35   75 L min    Auxiliary Outlets   1  Check outlets 
29.  Part No     1  2  3  4  5  6  7  8    800230  15279  38320  019011  800227  462541 A R  800311  053207  054512  462542 A R  045057  01067  01007  01027  054516       a  32 ou nu un in Cc  ar oONM   Oo    045056       Description    Knob  CO2   Label  CO2   Needle Valve   Screw  M3 x 6   Flow Tube  CO2 Low   4 mm Tube   High Pressure Gas Block  CO2  Plug  G 1 8  with O ring  Elbow   4 mm Tube   5 mm Tube   Valve  CO2   M3 Plain Washer  Screw  M6 x 16  Grubscrew  M4 x 6  Tube Connector    128    045058    Qty       pA a       As required  1  2  1  As required  1            p a e     PARTS LIST                   CGO Block   Ref Part No  Description Qty   1 15045 O2 Flush Label 1  2 043 O Ring BS116 1  3 034331 Safelock Nut 1  4 800471 Flush Button 1  5 800472 Button Shield 1  6 800473 Screw  Flush Button 1  8 89247 Outlet Male Connector 1  10 90183 Clippard Valve 1  11 0251 O Ring 219 0 x 1 75 CSO 1  12 0314 O Ring 1  13 800257 CGO Block 1  14 0360 1 4  Ball 3  15 011107 Cap  M5    Skiffy    1  16 019164 Screw  M5 X 80 1  17 800258 CGO Rail Mount 1  18 054519 Coupling 1  19 800259 Wear Pad 1  20 0408 Dowty Seal G1 4 1  21 01130 Screw  M5 x 16 2  22 019073 Screw  M3 x 10 1  23 700650 Outlet Connector 1  24 0581 Dowty Seal G1 8 1  25 15135 Label O2 Flush  Green  1  26 800360 Adaptor  Orifice  1    129    PARTS LIST       High Pressure Gas Block    Ref         O ON OOK ON        oO    12  13  14  15  16  17  18  19    20    Part No     800308  800305  800309  800311  800310  80030
30.  Slowly operate O2 control valve throughout range  Check that N2O now flows  Operate N2O Air selector switch to Air position and check that N2O ceases to flow  and that Air is now available    Return switch to the N2O position and set the oxygen flow back to minimum   Refer to the table below  and check   Oxygen concentration at the oxygen flows  specified     Adjust the Oxygen Flowmeter control only   Do NOT adjust the N2O flowmeter     Note The flow rates for N2O are given as a guide only   O2 Flow  02  in O2 N2O mixture  N20 Flow   0 25 L min 27 33  0 5   0 67   0 5 L min 27 33  1 0   1 35  0 8 L min 27 33  1 6   2 1  1 5 L min 27 33  3 0   4 1  3 0 L min 7 33  6 1   8 1     Cascade flowmeter models   March 2002 onwards    Repeat tests by reducing Oxygen flow back to Basal Flow conditions  referring to the  values in the above table   Check gas delivery switch for correct operation    ee Fail Alarms     2   3    4     B  1   2   3   4  5    D    Simulate an Oxygen fail by unplugging the Oxygen Pipeline from the outlet   Check that before the supply of oxygen is exhausted that the Oxygen Fail Whistle  sounds for a minimum of 7 seconds    Check that any N2O and Carbon Dioxide  if fitted  Gas Flows are cut off    Check the Oxygen Supply Visual Indicator turns from fully green to fully red    Turn on Oxygen cylinder    Check Oxygen Basal flow reinstated and whistle sounds briefly    Turn off Oxygen cylinder and repeat above tests     ack Bar Assembly   Selectatec   Checks     
31.  Turn OFF Gas Delivery Switch   Turn Off all cylinders    Turn off all Flow Control Valves   Attach Service and Warning Labels     DPN    Penlon recommends that after servicing the Anaesthetic Machine should be given an   Acceptance Check  by an Anaesthetist before being returned to Operational Use     81    9  SERVICING PROCEDURES       9 1 Servicing   The Prima SP range of anaesthetic  machines must only be serviced by a  Penlon trained engineer to the following  service schedule     Every 6 months   Inspection and Function Check   Every 12 months   Annual service which includes routine  replacement of seals  etc  as preventive  maintenance   At 3 years interval   Pipeline hoses   replace   At 5 years interval   Primary regulators   replace   Every 2 and 6 years   Additional tests   component replacement    Details of these service operations are  given in this Prima SP Service Manual   available only to Penlon trained engineers     Mechanical AHD System   Additional servicing for the mechanical AHD  system only    Function Test    At 6 month intervals  see section 6 8  to 6 11    Additional Servicing    At 12 month intervals    Secondary Regulators   check pressure  setting   Gear linkage   inspect    Service Tools    015067  Regulator Spanner   Regulators manufactured  by    Gas Arc       800612  Flow setting tool    82    9 2 Ancillary Equipment  9 2 1 Sigma Delta Vaporizer    Follow the instructions given in section 8 of  the user instruction manual  These include   Eve
32.  all cylinders    Turn off all Flow Control Valves    Attach Service and Warning Labels     Ore iGO NDE    Penlon recommends that after servicing the Anaesthetic Machine should be given an   Acceptance Check  by an Anaesthetist before being returned to Operational Use     74    SERVICE SCHEDULE    8 4 Twenty Four Month Service Checks   To be performed at 24  and 48 Months of a 72 month cycle      Machine Frame     1  Check integrity of structure and security of all attachments   2  Check for damage    3  Check wheels for freedom of movement and braking    4  Check drawers  writing tablet move freely     Electrical Safety Test     1  Check mains lead and plug for damage   2  Perform Electrical Safety Tests as relevant for country of use   3  Check outlet sockets for correct function     Pipeline Hose Checks    1  Examine pipeline connectors  probes and hoses for cleanliness and damage    2  Perform anti confusion checks    3  Replace pipeline filters   4 Perform pressure Leak Checks of all hoses with cylinders removed and Gas Delivery  Switch in the OFF position     Cylinder Yoke Assemblies    1  Check for damage  cleanliness  all fittings are tight and clamping screws for freedom  of movement    2  Replace Bodok seals  DO NOT apply grease or lubricant   3  Replace yoke inlet filters     Pressure Gauge and regulator checks     1  Examine all pressure gauges for damage and clarity   2  Check movement of needles is smooth   3  Perform pressure Leak Test of all cylinder Gas Suppl
33.  all traces of Loctite from the grubscrews and apply new  Loctite 242     15  Check that each gear slides easily on its shaft   16  Set the Gas Delivery Switch to ON   17  Open the N20 valve to maximum  to check full flow     18  Close the valve gently until the N2O bobbin ceases to rotate   Do not over tighten the valve when closing   this will cause  irreparable damage to the valve seat     19  Fit the O2 gear  7  on the shaft   Fit the Setting Tool  9   Part No  800612        92       SERVICING PROCEDURES       20  Fit the O2 control knob and open the O2 control valve  to  check full flow   Cascade Flowmeter Tube Models   Close the valve until a flow of 50 75 ml min is obtained   Single Flowmeter Tube Models   Close the valve until a flow of 100 200 ml min is obtained     21 Distance the inner face of the O2 gear with a 0 1 mm feeler  gauge at    A      Tighten the grubscrews   Check the flow returns to the basal flow set in operation 20   above     22  Refit the N2O gear  8    be careful not to move either shaft   Tighten the grub screws     23  Apply a small amount of BG 87 to the gears     24  Ifthe machine is not fitted with an Oxygen monitor  connect  a stand alone analyser to the CGO outlet        25  Check Oxygen concentration     26  Turn the Nitrous Oxide flowmeter control to maximum   Refer to the table below  and check   Oxygen concentration  at the Oxygen flows specified    Adjust the Oxygen Flowmeter control only   Do NOT  adjust the N2O flowmeter   Note The f
34.  flash  and  a single    beep    warning will sound           ALARMS    HIGH Oz ALARM  LOW O2 ALARM  O2  MONITOR INOP  O2 SENSOR FAULT  0  SENSOR LOW  LOW BATTERY  MAINS FAILURE    K Penlon    GAS DELIVERY switch  must be  ON  to operate  O2 Monitor    O2 Monitor       6  Turn the SET CAL knob  7  until the  display shows 100      7  When value is set  press the CAL  button  6  again to accept the  calibration setting    The O2 concentration display LED  1   will now stop flashing  and a single     beep    warning will sound     The display will return to normal  operation     Sensor Low Indication   The unit has a self detect feature to indicate  when the sensor life is low    During calibration  if the O2 SENSOR LOW  alarm LED  8  illuminates  and a single     beep    warning sounds  this indicates that  the sensor must be replaced    Sensor output will fall very quickly to  zero over a period of two to three weeks  from the first time that the alarm is  activated     PRE USE CHECKS   Oxygen Monitor    HIGH ALARM SET    SET  LOW ALARM SET        a    SET CAL    5 12 5 Calibration Procedure    Using Room Air    NOTE Calibration in room air may not  provide as great an accuracy as calibration  carried out in 100  oxygen     1  Switch on the oxygen monitor using  the Gas Delivery switch on the  machine front panel    The LEDs  1  2  3  and alarm visual  indicators  4 and 5  will illuminate  and  the audible alarm will sound     2  Remove the sensor from the  anaesthetic machin
35.  flow of oxygen is indicated  as  shown above    If the basal flow is outside the range given  above    a  adjust the output of the secondary Oxygen  regulator   see section 9 11  to achieve the  correct basal flow    b  check the oxygen concentration  see section  9 8 2     Switch the Gas Delivery switch to OFF     If the correct basal flow can not be achieved   reset the gear linkage   see section 9 8 3     90                SERVICING PROCEDURES       9 8 2 Check Oxygen Concentration    NOTE  The machine must be on a flat surface     1  If the machine is not fitted with an Oxygen monitor  connect  a stand alone analyser to the CGO outlet     2  Set the Gas Delivery switch  1  to ON                                                                                                                                                                             3  Turn the Nitrous Oxide flowmeter control to maximum   4  Refer to the table below  and check   Oxygen concentration  at the oxygen flows specified    Adjust the Oxygen Flowmeter control only   Do NOT 3  adjust the N20 flowmeter  2 ka N  Note The flow rates for N20 are given as a guide only  HUN il  O2 Flow  02  in O2 N20 mixture  N20 Flow D ot     0 25Limin 27  33   0 5   0 67 a  ee les     i 5 y     a gt  z      D Ff     0 5 L min 27   33   1 0  1 35 a    gt  j    0 8 L min 27   33   1 6   2 1      1 5 L min 27   33   3 0   4 1 HSA eA O  3 0 L min 27   33   6 1   8 1 sA   Cascade flowmeter models   March 2002 onwards 4  If 
36.  integrated system with the AV900  V4 ventilator   this system is not MRI compatible     3  Nuffield 200 Ventilator  4   Delta Vaporizer                ep  A100 Absorber    Ayr    ion  ZA  A                          FIS  uy                      The following components are currently not  MRI compatible       Oxygen Monitor  Flowmeter lighting  Electrical power outlets    AV series ventilator    PONH    NOTE   a  MRI Compatible Plastic Laryngoscopes    see section 3 9 in the Penlon Price List    b  The IDP Pressure Failure Alarm must be   a  mounted securely and  b  fitted with appropriate batteries   see  section 1 5 in the Penlon Price List     4  SPECIFICATION    4 1 Physical Dimensions    Overall frame size   SP 101  SP 102    Work surface  Height    Size   Loading   Writing tablet   Top shelf     Loading     Base Drawer     Loading     Castors    Ventilator bellows post  Loading   Gas scavenging fixing    Loading    Common gas outlet     Height x Width x Depth  cm   139 x 45 x 70  139 x 71 x70    86 cm  SP 101  45cm x 25cm    SP 102  58cm x 25cm  30 kg  66 Ib    evenly distributed     30 x 22 cm  SP 101  58 5 cm x 35cm    SP 102  71 cm x 35 cm  30 kg  66 Ib    evenly distributed     Prima 101  16 x 37 x 35cm  Prima 102  12 x 54 5 x 35 cm    10 kg  22 Ib  evenly distributed    Front pair braked    Bushed to accept 25 4 mm  1 inch  or 22  mm  7 8 inch  poles   30 kg  66 Ib     Bracket on frame upright    30 kg  66 Ib     22 mm male taper with coaxial 15 mm  female taper con
37.  is 35   75 L min    Auxiliary Outlets   1  Check outlets for security  damage and correct lock and release movement   2  Check dynamic flowrate and that Oxygen Fail Whistle does not operate     Oxygen Monitor  if fitted   1 Replace backup battery     Oxygen Monitor   Function and Calibration Tests   Connect a Test O2 Analyser into the patient circuit   Check that the machine O2 sensor is inserted into the absorber O2 sampling point   Test both O2 sensors in 100  Oxygen   Expose both sensors to air and check reading is 21    2   Adjust high and low O2 alarms and check alarms trigger when values are lower or  higher than reading on O2 analyser respectively   Return the alarm levels to original settings   5  Restore the sensor to correct location   Remove the test O2 analyser     PON gt     Final Gas Concentration Tests   1  Check concentration by Gas   2  Check flowrate from CGO corresponds with flowmeter by concentration   3  Set 3 L min Oxygen and Nitrous Oxide Flows   Check  using O2 Monitor that the concentration from the CGO is 50    2    4  Set Oxygen to 0 5 L min  set N2O to 0 5 L min   Allow O2 monitor to stabilise then check concentration is 50    2    If fitted repeat above test substituting Carbon Dioxide for N20   If fitted set Air to 3 L min  set Oxygen to 3 L min   Check Concentration reading on O2 Monitor is 60    2      5   6     Test Completion   Paperwork  1  Remove all test equipment from machine     80    SERVICE SCHEDULE       Drain all Gas from machine   
38.  is in use under battery power   and the battery voltage is less than 11 5  volts  a low priority alarm is triggered  to  indicate that the battery has less than 20  minutes life left    If the battery voltage falls to less than 10 8  volts a flashing medium priority alarm is  triggered to indicate there is less than 5  minutes power left in the battery    To cancel this alarm  mains power must be  On     NOTE lf this condition persists  contact your  Penlon Service Centre  or Penlon Customer  Service Department in the UK     At the end of the final 5 minute warning  period  the oxygen monitor will shut down  to  prevent damage to the battery     MAINS FAILURE   This alarm indicates mains power failure or  cut off  The yellow MAINS FAILURE LED will  illuminate  and a low priority sound will be  triggered     3 13 5 Alarm Mute  In an alarm condition  pressing the ALARM    MUTE button will deactivate the alarm  sounder but the alarm LED will continue to  flash  The yellow MUTE ALARM LED will  illuminate  accompanied with a SINGLE     beep    sound     The alarm mute cannot be operated    a  until the mute time is over  or the alarm  condition has been rectified    b  when the oxygen concentration drops  below 18      In high priority and medium alarm conditions  the alarm mute deactivates the sounder for  30 seconds and 120 seconds respectively     DESCRIPTION   O2 Monitor       3 13 6 The MOX 3 Oxygen Sensor    The MOX 3 oxygen sensor offers quick  response  linear output over
39.  machine does not  exceed the values specified in any relevant  national standards that may apply in the country  where the machine is in use     b  Each socket is protected with two 5 A fuses     3 10 2 Auxiliary Power Supply  Sockets  if fitted     An optional mains electricity auxiliary panel  with three or four sockets can be specified   and fitted to the rear of the machine                                         The supply to the sockets is controlled by an  ON OFF switch  A   which also incorporates  a circuit breaker     3 10 3 Flowmeter Bank Lighting   Optional     The lighting system is controlled by the  main ON OFF switch  A      3 10 4 AV900 or AV800 Ventilator   if fitted  Power Supply    The mains lead for an AV series ventilator  can be plugged into one of the auxiliary  power sockets on the rear of the machine     AV900 V4 with interface link to Prima SP    See section 3 15    a  Turn the machine Gas Delivery Switch ON   The ventilator will power up    b  While the Prima SP power is ON  the  Ventilator can be turned OFF and ON  using  the ventilator On Off switch    c  Turn the Gas Delivery Switch to OFF    The ventilator will power down     Ventilator Back up Battery   If the power supply to the ventilator fails  the  ventilator back up battery will power the  ventilator for 60 minutes  if the battery has  been maintained in a fully charged condition   Refer also to the user instruction manual  supplied with the ventilator    Battery charging takes place aut
40.  machine to an approved  anaesthetic gas scavenging system  AGSS      WARNING   Do not contaminate pipeline connections   cylinder yokes  regulators or pressure gauges  with oil or grease  or any other flammable  lubricant or sealant     FUNCTION TEST    6 2 Machine Frame    1  Check the main structure of the  machine is firmly assembled  free  from obvious distortion and damage  and that all attachments are secure     2  Check that the castor wheels are  securely attached to the machine  frame and that the wheels run and  swivel freely  Check that the wheel  locking mechanism is working     3  Check the drawer unit s   where fitted   open and close smoothly     4  Check that the writing tablet  if fitted   opens and closes smoothly     6 3 Electrical Safety Tests    1  Disconnect the mains lead from the  electrical supply     2  Disconnect all equipment from the  auxiliary power outlet panel  A      3  Perform Electrical Safety Tests as  specified by the National Standards  applicable to your country     4  Check electrical outlet sockets for  correct output and earth continuity                    50    FUNCTION TEST    6 4 Pipeline Gas Supply and Non Return    10    11     12     Valve    Ensure that all reserve gas cylinders are turned off   and removed from the machine     Leave the Gas Delivery ON OFF switch in the  OFF   Position     Check ALL Pipeline Hose assemblies for correct  probes  colour coding  and attachment to the machine     Inspect ALL Pipeline hoses for cr
41.  regulators should be  set at 275 kPa  40 psi   but adjustment is  available from 261 to 289 kPa  85 to 41  psi     If the N20 flow is low the N2O pressure  should be increased to 289 kPa  41 psi    before the O2 pressure is reduced to 261  kPa  35 psi     Similarly  if the N2O flow is high  the O2  pressure should be increased  before the  N20 pressure is reduced    Carry out a full function test on the  machine  see section 6     95                                                                            Do   oOo oa    amp        dR Se     SE  oo   ahs 2 Gr GD  amp                           pas ee                                                                               oe   O VANNS O    7                Note   On machines built from March 2002   the regulators are mounted horizontally    SERVICING PROCEDURES    9 9 Control Valve Capsule  Removal Replacement    CAUTION  Needle valves are designed to seal with a    light torque only and may be damaged if    tightened excessively        Note   Early type capsule assemblies are  shown in the illustration             Note    There are no user serviceable components  within the flow control valve capsules  1  2   and 3     This procedure relates only to the exchange  of the complete  factory set  valves     1  Remove all gas cylinders and discon   nect all gas pipeline hoses     CAUTION   Ensure that all gas supplies are exhausted before  commencing these procedures  High pressure  gas can    fire    the valve out of the m
42.  rotate the retainer  A   to lock the canister in position     3  Reconnect the hoses  see 5 6 4    Slide the canister and support  plate under the machine work  surface     110      Lift the retaining  pin  C    Disconnect the  hoses   Support the  weight of the  valve block as it  is removed from    1        Note   For A100SP  Absorber  see  Appendix 4             Absorber Valve  Block Assembly  Removal         2  Disconnect the hoses  the machine   Rotate the retainer  A   and  support the canister as it is  removed from the machine   Refit    Align the buttons   D  on the top of  the canister block  with the slots  E        Slide the canister    into the support  plate       Reconnect the    hoses  see 5 6 4        Fit the retaining pin     C      10  PARTS LIST    Prima SP System   Preventive Maintenance Kits    Part No  Description   Prima SP Mechanical AHD and Selectatec  57455 6 month   57476 12 month   57479 24 month   57476 36 month   57479 48 month   57476 60 month   57490 72 month    Prima SP Mechanical AHD and    Cagemount Backbar    57455 6 month  58614 12 month  57480 24 month  58614 36 month  57480 48 month  58614 60 month  57489 72 month  A100 Absorber   58615 12 month  58609 60 month  AV900 Ventilator   57453 12 month  57453 24 month  57453 36 month  57453 48 month  57454 60 month  AV800 Ventilator   57633 12 month  57633 24 month  57633 36 month  57633 48 month  57635 60 month  Nuffield 200 Ventilator  57139 12 month  57139 24 month  57139 36 month  57139 4
43.  seal   Hose Assembly  011017 Tubing   Sample Block       37950     gt            Offline  Hose  0226  O seal  Offline           ery    A  a  y    SN    Qy    0762  i Filter   Pipeline    t          om  Je          5 l 37947    37951  Yi on Yj  Split Ring Seal Support  es i scorer eres   lt  lt  462526 tte yea    a  043 0226 Hose   s  O seal O seal Backbar    Safelock to CGO Hose Support       113    PARTS LIST    Cagemount Backbar  Preventive Maintenance Kit    24   48 Month Service  Part No  57480       Do NOT apply grease  Kit Contents or lubricant to     Part No Qty  Component Description 016  Bodok Seal  462526 1 35 m Hose   Backbar Outlet to CGO 043  O seal   Safelock  37950 4 Off line Hose   016 Bodok Seal   Flowmeter Bobbin   Metal with Filter   Flowmeter Bobbin   Metal with Filter   Filter   Yoke and Pipeline   Filter   Flowmeter   Flowmeter Bobbin   Tube Seals   Washer   Shock Washer   O seal   Flowmeter   O seal   Safelock   O seal   Hose Assembly   Hose Support   Seal Support   Split O ring   Tubing   Sample Block    400224  400262  0762  0508  36247    01057  36075  0691  043  0226  37946  37947  37951  011017    DODWOAWAWAWW UI    ak ee ek a  hO WM PW PP       37951 37947  Split Ring Seal Support    N SSS  ope  iy    i  y       mmaa  l    eooo  Nm  VBRYEIE PITTS       ae 0762       Offline  Hose    0226  O seal  Offline Hose Filter   Pipeline    114       PARTS LIST       043  O seal    Safelock    0293  O seal    0691  O seal    36075  Washer       0691  O seal
44.  sticks at above    zero reading when con     trol valve closed     Float level unstable  after start up     63    Possible Fault    Dirt in tube     Build up of electrostatic  charge     Flowmeter unit not ver   tical     Upper float retaining  peg not centralised or  moulding flash not  removed     Worn needle valve     Pressure reducing  valve defect     Flowmeter unit not ver   tical     Remedy    Remove tube and  clean     Check earthing con   tacts at bottom of  flowmeter tubes     Increase flow rate  sharply to break  charge  then close  valve     Ensure trolley is on  level surface     Remove flow tube and  retaining peg    Clean off moulding  flash and re position  retaining peg     Replace complete  valve and or  seat assembly     See instruction under  component group     Ensure machine is on  level surface     FAULT FINDING       Component    Flowmeter unit    Symptom    Restricted flow through  flowmeter  Control   valve opening does not  achieve expected flow      Float level falls after ini   tial setting     64    Possible Fault    Faulty cylinder pres   sure reducing valve   yoke or connections     Faulty pipeline supply  hoses or connections     Blocked flowmeter filter     Control knob loose on  shaft     Incorrect valve restric   tion shimming     Damaged valve seat    assembly     Reducing valve defect     Pipeline supply defect     Leak in    upstream    con   nection     Blockage in    down   stream    gas circuit      Not oxygen flowmeter   Gas cut off
45.  the entire O   100  oxygen range  and long service life     The MOX 3 is a self powered galvanic cell  that generates a current proportional to  oxygen concentration     The cell has a highly stable output over its  operating life  Significant output loss is only  shown at the very end of its life    Typical sensor drift rates are less than 1   per month when the sensor is exposed to  gas in typical applications    Sensor lifetime is governed by the mass of  lead available to react with the oxygen and  its rate of consumption  High oxygen partial  pressure and high temperature will increase  the sensor output current  thus shortening  the operation life    At the point where all lead has been  consumed  the output will fall very quickly to  zero over a period of two to three weeks     NOTE  To maintain maximum sensor life  always remove  from breathing circuit after use     21          Sensor Location  The sensor assembly consists of an  external probe and 2 m  6 ft  cable     1  Prima SP with A100SP Absorber    Please refer to the datasheet  Doc No A100SP  103DS U   supplied with the absorber     Sensor       2  Mounted on the dome of the  absorber inspiratory valve       Sensor          DESCRIPTION   O2 Monitor       3 13 7 Power Supply   O2 Monitor    Mains Power Supply   Power is fed to the machine via the mains lead to a  switching mode power supply    The oxygen monitor is powered by 14 2 V    The monitor is controlled by the machine gas system  master On Off switch  s
46.  the relevant user manual for further information     32    PRE USE CHECKS    Non AHD Machines        5 2 Pre use Checks  Gas Supply  Non  AHD     5 2 1 Gas Pipeline Supplies   1  Connect the oxygen pipeline hose  only    Check that the warning whistle  sounds briefly as the hose is  connected    Check that the correct pressure gauge  reading is obtained    2  Switch Gas Delivery Switch  A  to ON   On machines with Air  set the Air N2O  interlock switch  B  if fitted  to N2O   see 5 2 3     Open the oxygen and nitrous oxide  flowmeter needle valves    Check that flow is only shown in the  oxygen flowmeter     3  Close the flowmeter valves    4  Connect the other pipeline hoses in  turn  and check the gauge reading for  each gas     Check that gas flows when the  relevant needle valve is operated   For Air  set the interlock switch  B  if  fitted  to Air   NOTE  If the machine is not equipped for connection to a  pipeline supply  carry out the above procedures  using the cylinder supplies     5 2 2 Gas Cylinder Supplies  CAUTION   Open the cylinder valves slowly to avoid damage  to the pressure reducing valve and pressure  gauges  Ensure that valves are at least one full  turn open when in use     1  Fit the gas cylinders to their  respective yokes  open the cylinder  valves one at a time    Check the pressure on each cylinder  gauge     NOTE   A  When two cylinders are provided for a single  gas  test each separately  clearing pressure after  each test by opening the flowmete
47.  units in  operation or defective     Worn needle valve  and or seat     Remedy    See instruction under  component group     See instructions under  component group     Clean or replace lower  bobbin assembly     Tighten     Reset valve shims     Replace complete  valve     See instructions under  relevant component  group     See instructions under  component group     Test for leaks   Tighten leaking joints    DO NOT use exces   sive force      Trace site of blockage    and clear     See instructions under  component group     Replace complete  valve assembly     FAULT FINDING       Component    Vaporizer mounting  facility     Selectatec compatible  system     N20 and CO2 gas cut   off unit     Symptom    Leakage at back bar  manifold or vaporizer  interlock joints     Leakage from back bar     Leakage from vaporizer  connections  Vaporizer  difficult to remove     No nitrous oxide  third  or fourth gas     No N20  third or fourth  gas  or restricted flow  of these gases     Safety capsule valve  fails to operate correct     ly     Leaks at joint between  manifold and valve  seating     Valve cuts flow of  nitrous oxide as  required but does not  restore it     65    Possible Fault    Loose manifold or  vaporizer retaining  screws     Missing    O    seals    Damaged or worn gas  valves     Damaged or worn    O     seals or seal faces   Damaged or worn lock   ing system     Oxygen supply not  turned on     Mechanical AHD gas  delivery switch left in     Off position     F
48. 1  The taper cone joints must be engaged axially  and not sideways loaded    Use the shims provided so that the distance from  the back bar to the taper joint can be adjusted by  adding or removing shims from the vaporizer     2  The cone joints should then be lightly smeared  with an Oxygen compatible lubricant such as   Fomblin      The taper joints must be engaged by applying  axial pressure  and the fixing nuts tightened   Check all joints for gas tightness    3  Check flexible hoses for security of attachment  and integrity     57                   FUNCTION TEST    6 9 Reduced Pressure Gas    Circuit Leakage    Back Bar Leak Test and  Pressure Relief Test     Connect a suitable test pressure  gauge  Manometer  to the Common  Gas Outlet  CGO  using suitable tub   ing and connector     Ensure all vaporizers attached to the  machine are turned OFF     Using the Oxygen Flow Control Valve   raise the pressure in the circuit slowly  until a flow of 10 L min is set    Check that the Pressure Relief Valve   PRV  mounted under the gauge  cover  on the right hand side  begins to  leak in the range of 37 to 45 kPa     Do not exceed this pressure     Alter the pressure by reducing the  Oxygen Flow Rate until the displayed  reading on the test gauge is 20 kPa   150 mmHg      The maximum permissible flow to  maintain the above pressure is 200  ml min     If a higher flow rate is required to main   tain a pressure of 20 kPa  150 mmHg   there is a leak in the system that must  be fixed 
49. 10  3 5 Mechanical AHD 11  3 6 Pressure Gauges 12  3 7 Flowmeters and Controls 13  3 8 Vaporizers 14  3 9 Common Gas Outlet  CGO  Block 15  3 10 Electrical Power Supply 16  3 11 Third Fourth Gas Options 17  3 12 Auxiliary Gas Outlets 17  3 13  Oxygen Monitor 18  3 14 A100 and A100SP Absorber 23  3 15 Prima SP Anaesthetic System   MRI Compatibility 23  4  SPECIFICATION 24  4 1 Physical Dimensions 24  4 2 Gas Supplies 25  4 3 Flowmeters 25  4 4 Gas Pressures 26  4 5 Auxiliary Gas Outlets 27  4 6 Oxygen Failure Warning Devices 27  4 7 Oxygen Flush 27  4 8 Mechanical AHD Systems 27  4 9 Environmental 27  4 10 Third and Fourth Gas Options 28  4 11 Electrical Supply 28  4 12   Oxygen Monitor 29  5  PRE USE CHECKS 31  5 1 Pre use Check List 31  5 2 Pre use Checks  Non AHD Machines  33  5 3 Pre use Checks  Machines with Mechanical AHD  35  5 4 Leak Rate Check 37  5 5 Electrical Supply 37  5 6 Patient Breathing System 38  5 7 Oxygen Flush 42  5 8 Anaesthetic Gas Scavenge System  AGSS  42  5 9 Alarm System Testing 43  5 10 Vaporizers 44  5 11 Ventilator 44  5 12   Oxygen Monitor 45  6  FUNCTION TEST 49  6 1 Introduction 49  6 2 Machine Frame 50  6 3 Electrical Safety Tests 50  6 4 Pipeline Gas Supply and Non Return Valve 51    6 5 Cylinder Gas Supply and Pressure Reducing Valves 52   iii     CONTENTS       9 9  9 10  9 11  9 12  9 13  9 14  9 15  9 16  9 17  9 18  9 19  9 20  9 21  9 22  9 23    10   11     Flowmeter Unit   Gas Safety Devices   Vaporizers and Back Bar Manifold Assembly
50. 2 5 Flowmeter cover screws    144    Torque Nm  1 7   2 0  0 9   1 2  4 2  4 5  3 1  3 4  7 0  7 5  6 0   6 5  7 5   8 0  6 5   6 8  13   13 6  11 5  12    Torque Nm  Tighten to take up slack   and then 2 5 turns   7 8   6 7   10  12   12  14   20   22   38   42   14 16   5 6   3 4   0 5   1 0   1 0   2 0    0 27   0 32    Appendix       APPENDIX 5    A100SP Absorber    This information must be used in conjunction with the   E  A100 User Instruction Manual  Cat No 52689            A100SP Absorber Features             1  Interface with AV900 V4 Ventilator l  The A100SP is fitted with fitted with a sensor that  detects the position of the absorber bag vent lever  660  Note that the Bag Vent lever is fitted on the left  OO9    hand side of the manifold block  viewed from  lt a  the front of the machine    The sensor signal cabling is routed internally to a B  connector at the rear of the absorber valve block A                 Operation of the Bag Vent lever will trigger CS Zz  automatic Mode switching on the AV900 ventilator  Go   as follows  Se   a  If the Absorber Bag Vent Lever is moved from   Vent to Bag  the ventilator will change from Volume Lyx Z  Mode  or Pressure Mode  into Spontaneous Mode    b  Switching the absorber Bag Vent lever from Bag  to Vent will reset the ventilator from Spontaneous O  Mode to Standby Mode    If the ventilator is in any mode other than those   detailed above  operation of the absorber Bag Vent   lever will not affect the ventilator    NOTE   
51. 2 Cylinder ON    Set all gas flows  as required  to 5  litres min  except COz  set this to 500  ml min     Connect a manometer to each  Primary regulator test point in turn and  check regulator is set in accordance  with specification for relevant country   Connect a manometer to each  Secondary regulator test point in turn  and check regulator is set in  accordance with specification for  relevant gas   Reduce gas  settings     flows to minimum    FUNCTION TEST    6 6 Flowmeter Unit     10     11     Nitrous Oxide   Air Selection Switch   Oxygen Fail Cut off and Warning    Visually inspect the flowmeter tubes for damage   Check that the transparent shield is in place     Turn the Gas Delivery Switch  A  to ON   Check that a basal flow of oxygen is delivered  as follows     All models 100 200 ml min   up to March 2002     Cascade O2 flowmeter 50 75 ml min   March 2002 onwards     Single O2 flowmeter 100 200 ml min   March 2002 onwards     Open the Oxygen Flowmeter control slowly   Check that a full flow can be obtained     Set a flow rate of 6 L min  Check that the float is stable in its setting and spins freely   Repeat test at a flow rate of 500 ml min     Without rotating the control knob  check for wear by  gently pulling on the knob  and check that any fluctuation  in flow rate is within the range of  100 ml min     Close the Oxygen Flowmeter control valve gently until the  mechanical stop engages     Check Oxygen Basal Flow  see operation 2      Nitrous Oxide   Oxygen L
52. 33  041115  016  0762  043  011017    o a  gs  3         0762 KSX A  z Filter   Yoke CALL ao MY    016 043   462526  Bodok Seal 9 O seal 0226 Hose   Backbar   9 7   Safelock O seal Outlet to CGO  S  022533     gt  Q    Dzus Clip  041115 J    0762  Filter    Pipeline or       118    PARTS LIST    Selectatec Backbar  Preventive Maintenance Kit    24   48 Month Service  Part No  57479    z Do NOT apply grease or  Kit Contents lubricant Ae ae    Part No Component Description 016  Bodok Seal  462526 Hose   Backbar Outlet to CGO 043  O seal   Safelock  37946 Hose Support 041115  O ring  Selectatec  37947 Seal Support   37951 Split Ring   041115 O ring   Selectatec   016 Bodok Seal   400224 Flowmeter Bobbin   Metal with Filter   400262 Flowmeter Bobbin   Metal without Filter   0762 Filter   Yoke and Pipeline   0508 Filter   Flowmeter   36247 Flowmeter Bobbin   Tube Seals   01057 Washer   36075 Shock Washer   0691 O seal   Flowmeter   043 O seal   Safelock   011017 Tubing   Sample Block   022533 Dzus Clip    oo  Gad  3             0762    Filter    SS 0762 Dzus Clip        Filter   Yoke    016  Bodok Seal    041115  O ring    119    PARTS LIST    37947  37951  Split Ring Seal Support    ZXZ AA    of DD p ETES          SANONEET    Hose Support               0226   O seal  ae 462526   O seal    Safelock Hose   Backbar  Outlet to CGO  400262   0293 Bobbin diber  PRR oo Washer    SS    0691  O seal    oot    36247  a Bobbin    36075  Washer                                  0691  O seal  
53. 36247  400224 Bobbin  gers Bobbin  O seal  0508 Filter Cascade flowmeter system illustrated   On non cascade models  the filter  0508   is fitted to each flow tube              125    PARTS LIST       Air Cylinder   Pipeline Kit    Ref Part No   1 800230  2 38318  3 800224  4 800310  800306  5 462542 A R  6 053207  7 800205  800206  8 054519  9 462542  10 462542  11 462542 A R  12 054917  13 045068  14 0581  15 054543  16 020014  17 15064  18 0762  19 21240  20 700382  700389  21 01067  22 019011  23 054528  24 01007  25 054517  26 01027  27 462542  28 045069  MA7 Supply  58267  58268  064274    Description    Knob  Air  Needle Valve  Flow Tube  Air    HP Gas Block  Air UK  HP Gas Block  Air USA    5 mm Tube  Plug  G 1 8 with O ring    Low Pressure Gas Block  Air  UK  Low Pressure Gas Block  Air  USA    Coupling   5 mm Tube  5 mm Tube  5 mm Tube    Quick Release Coupling  auxiliary outlet  Air   Valve  Air  Dowty Seal G1 8    Elbow   5 mm Tube   Nut  M16 Lock   Label  Air   Filter   Spring   NIST Air Connector Body  DISS Air Connector Body    M3 Plain Washer  Screw  M3 x 6  Stem Adaptor  Screw  M6 x 16  Tube Connector  Grubscrew   M4 x 6  5 mm Tube  Actuator Switch    Pipeline hose assembly kit  comprising   Hose  Pressure reducing TU    126    Qty     As required  2    1  1    Poe      s required     i mi k      aa 4 4a p  O       a  34 3234 KR opp       E    KA  ZO    BANE                   X             KE             2D     127    PARTS LIST          CO2 Cylinder kit  Ref
54. 6    019129  011171  700447  016  700446  020406  32444  0762  32058  700440  15252  15051  15277  TBA  15064  15278  041122  045444  041121  90283  404194  700438  053207  049046  049047  049048  049049  049050  049051    0691    Description    High Pressure Gas Block Assembly   O2  UK   High Pressure Gas Block Assembly   O2  USA spec   High Pressure Gas Block Assembly   N20   High Pressure Gas Block Assembly   CO2   High Pressure Gas Block Assembly   Air  UK   High Pressure Gas Block Assembly   Air  USA spec     Screw  M6 x 40   Knob   Yoke Catch Plate   Bodok Seal   Yoke Spacer   Circlip   Yoke Insert   Filter   Pin   Gas Block  High Pressure  Label  O2  UK   Label  O2  USA   Label  N2O   Label  CO2   Label  Air  UK   Label  Air  USA   O ring   HP Regulator   O ring   Pressure Relief Valve  Spacer   Plunger   Plug  G 1 8  complete with O ring  Gauge   O2  UK  Cylinder  Gauge   O2  USA  Cylinder  Gauge   N20  Cylinder  Gauge   CO2  Cylinder  Gauge   Air  UK  Cylinder  Gauge   Air  USA  Cylinder    O ring 1 8  I D x 0 07 CS    130    E 0668 S69 oe ea oe oa oe oe on on on on eo on on on on on on pp ae oa ao py ae os 4 pp     e    Qty     a ee Cee ee Cee Ce    PARTS LIST          131    PARTS LIST       Selectatec Compatible Backbar    Note     Main illustration shows original split type    manifold     Inset illustration shows later  one piece    manifold        Ref Part No   17 01253  2 15464  3  700636  4  700631  5 045051  6 01056  7 71577  8 700634  9 022533  10 01059  11
55. 6 UK Sales Tel  01235 547036 International Sales Tel  44  0  1235 547001  Service Fax  44  0  1235 547062 UK Sales Fax  01235 547023 International Sales Fax  44  0  1235 547021  E mail  technicalsupport penlon co uk uksales penlon co uk export penlon co uk    
56. 8 month  57450 60 month    111    Circle System Module    57451  57451  57451  57451  57452    12 month  24 month  36 month  48 month  60 month    Delta Vaporizer Overhaul Kits    55146    55145    55144    55143    55142    55141    55140    55139    55147    5 year Halothane     Pour Filler   Halothane   Agent Specific   Key  Filler   10 year Enflurane     Pour Filler   10 year Enflurane     Agent Specific  Key  Filler  10 year Isoflurane     Pour Filler   10 year Isoflurane    Agent Specific  Key  Filler  10 year Sevoflurane    Pour Filler   10 year Sevoflurane    Agent Specific  Key  Filler  10 year Sevoflurane    Quik Fil Filler    PARTS LIST    Preventive Maintenance Kit  6 Month Service  use also for UK customer specified 3 month service     Part No  57455    Kit Contents    Part No Qty  Component Description  016    5 Bodok Seal  041115 6 O ring   Selectatec  043 1 O ring   Safelock Connector on CGO Block    WARNING Do NOT apply grease or lubricant           016  Bodok Seal    041115  O ring              112    PARTS LIST    Cagemount Backbar  Preventive Maintenance Kit    12   36   60 Month Service  Part No  58614      Do NOT apply grease  Kit Contents or lubricant to   Part No Qty  Component Description 016  Bodok Seal  462526 1 35 m Hose   Backbar Outlet to CGO 043  O seal   Safelock     not shown   37946 Hose Support  37947 12 Seal Support  37951 12 Split Ring    37950 Off line Hose   016 Bodok Seal   0762 Filter   Yoke and Pipeline  043 O seal   Safelock   0226 O
57. O2 Monitor       Oxygen Monitor   continued    Temperature Effects  The sensor has a built in temperature compensation circuit  and is relatively unaffected by  temperature changes within the operating temperature range given above     Pressure Effects   The sensor measures O2 partial pressure  and its output will rise and fall due to pressure  change  e g  changes in barometric pressure  or breathing system pressure     An increase in pressure of 10  at the sensor inlet will produce a 10  increase in sensor  output     30    5  PRE USE CHECKS       5 1 Pre use Checklist    A pre use checklist for the Prima SP range of  machines is printed on the next page    This checklist is also supplied with the  machine     Where necessary  subsequent sections in this  manual provide an explanation and procedure  for setting up the machine and ancillary  equipment and the various checks that must  be carried out before clinical use     In addition  checks specific to non AHD  machines  and mechanical AHD machines  are explained in separate sub sections     5 2 Non AHD machines  5 3 Machines with mechanical AHD    Details of checks common to ALL types of  machine  e g  Check correct connection and  functioning of vaporizers  are explained in  sections 5 4 onwards     WARNING   Pre use checks must be performed before each  period of clinical use    These checks must be supplemented by  periodic Function Testing  and full Service  Testing by a Penlon trained engineer to the  Service Schedul
58. ON    Penlon recommends that after servicing the Anaesthetic Machine should be given an   Acceptance Check  by an Anaesthetist before being returned to Operational Use     77    SERVICE SCHEDULE    8 5 Seventy Two Month Service Checks   To be performed at 72 Months of a 72 month cycle      Machine Frame     1  Check integrity of structure and security of all attachments   2  Check for damage    3  Check wheels for freedom of movement and braking    3  Check wheels for freedom of movement  check front braking   4  Check that drawers and writing tablet move freely    Electrical Safety Test    1  Check mains lead and plug for damage    2  Perform Electrical Safety Tests as relevant for country of use   3  Check outlet sockets for correct function     Pipeline Hose Checks    Examine pipeline connectors  probes and hoses for cleanliness and damage   Perform anti confusion checks    Replace pipeline filters   Replace non return valve o rings    Perform pressure Leak Checks of all hoses with cylinders removed and Gas Delivery  Switch in the OFF position     arwn  gt     Cylinder Yoke Assemblies    1  Check for damage  cleanliness  all fittings are tight and clamping screws for freedom  of movement    2  Replace Bodok seals  DO NOT apply grease or lubricant   3  Replace yoke inlet filters     4 Replace non return valve O rings     Pressure Gauge and regulator checks     1  Examine all pressure gauges for damage and clarity   2  Check movement of needles is smooth   3  Perform pressur
59. Regulator seat dam   aged     Loose connections   Tighten leaking joints    DO NOT use exces   sive force     Adjust replace regulator  as necessary     Cylinder empty     Bodok seal absent or  worn     Cylinder clamp screw  assembly damaged   Pressure gauge seal  damaged     Remedy    Fit replace seal     Replace as necessary     Ensure correct gas  cylinder is attached   Check index pins not  loose or damaged   Replace as necessary     Fit new relief valve     Adjust replace regulator  as necessary     Replace regulator     Test for leaks     Replace cylinder     Fit replace seal     Replace as necessary     Replace as necessary     FAULT FINDING       Component    Pipeline supply hoses  and inlet blocks    Flowmeter units    Symptom    Pressure gauge pointer  sticks at above zero  reading     Leaking high pressure  non return valve     Pressure gauge reads  zero when connection  to supply made     Insufficient free flow     Float stays at bottom of  tube when control valve  opened     61    Possible Fault  Inadequate  engagement of pin   index system     Defective gauge     Defective gauge     Gas trapped in system     Damaged    O    seal     Loose connections   Tighten leaking joints    DO NOT use  excessive force    Defective gauge   Faulty probe or hose     Blocked filter     Pipeline supply  defective     Blocked filter     Pipeline supply  defective     No gas supply     Pressure reduction  valve defect     Pipeline supply  defective     Control knob loose on  shaft 
60. Replace Selectatec O rings  DO NOT apply grease or lubricant  Replace flexible tubing from back bar outlet to CGO   1 35 m  Replace vaporizer locking spring    Check all pillar valves open and seal correctly    Check pillar valve inserts sit proud or flush with the pillar valve   A sunken pillar valve insert denotes failure of the internal spring   Check vaporizer interlock system for correct operation     Back Bar Assembly   Cagemount   Checks     1   2     Replace flexible hoses   Replace flexible hose from back bar to CGO   1 35 m    79    SERVICE SCHEDULE    3  Check security of taper connectors  Use Fomblin or other Oxygen safe grease to lightly lubricate the metal surface where  male and female connector come into contact with each other     Fresh Gas Pressure Relief Valve  PRV  Test   1  Occlude CGO with a manometer   2  At 10 L min Flowrate of Oxygen check the PRV relieves at 41 kPa   10      Leak Test From Flowmeter to Common Gas Outlet   T  Fit a suitable Manometer Test gauge to the CGO    2  Set Oxygen Flow to maintain a pressure of 20 kPa  3 psi    Maximum permissible flow is 200 ml min     3  Perform test with and without Vaporizer s  fitted and with vaporizer s  turned both ON  and OFF   4 Set oxygen flow to minimum and remove test equipment     Common Gas Outlet Checks  CGO      1  Check security and freedom of movement of CGO    2 Replace Safelock O ring  DO NOT apply grease or lubricant  3  Replace emergency flush valve    4 Check O2 Emergency Flush flow rate
61. Sterilisation  1  Check that all hoses and cables are  secure and correctly routed     2  Refit the covers     3  Refit the canisters   see page 7 of this  datasheet   Filling and Changing CO2  Absorbent     4  Refit the bag arm and bag     5  Refit the Manometer     Refitting the Absorber to the  machine    WARNING   The absorber assembly weighs  approximately 17 kg    Take care when lifting the absorber and  during installation     1  Apply the brake on the anaesthetic  machine     2  Align the mounting buttons  1  on the  top of the absorber manifold block  with the slots in the mounting plate  2   on the anaesthetic machine     3  Slide the absorber into position   Continue to push the unit inwards  until the right hand side mounting  buttons  1  drop into recesses in the  mounting plate  This indicates that  the assembly is correctly installed   Check that the assembly cannot be  moved to the left     4  Reconnect the hoses and cables at  the rear of the absorber manifold  block  3     Check for correct fitment with the  illustration on page 3 of this  datasheet     5  Refit the oxygen sensor  4  and  reconnect the cable  5      Ventilator Bellows Assembly   6  Reassemble the bellows  components   Refer to sections 7 2 and 5 2 in the  AV900 User Manual   Important   check that the O rings   6 and 7  are in position                          O                                                                               o Ho                                                 
62. Symptom    Valve does not operate     Low Oxygen flow  less  than 35 l min      Leaks around control  button     O2 flow into breathing  circuit or atmosphere   valve does not shut off  flow      67    Possible Fault    No oxygen supply     Defective valve     Low oxygen supply  pressure     Worn valve seals     Incorrect adjustment of  valve shaft     Incorrect control jet     Worn valve seals     Worn valve seals     Incorrect adjustment of  valve shaft     Defective valve     Remedy    See instructions under  component groups     Replace valve     See instructions under  components group     Fit new seals or  replace valves as nec   essary     Replace valve     Replace control jet    Fit new seals or  replace valve as neces   sary     Replace valve     Replace valve     Replace valve     FAULT FINDING       Component    Electrical power supply    Oxygen Monitor    Symptom    Machine    dead        Display blank when  powered up    O2 concentration dis   play shows    E         O2 MONITOR INOP  alarm is On    Battery Low LED and  Mains On LED illumi   nated     Battery In Use LED illu   minated  Mains supply  connected     68    Possible Fault    Mains supply  socket    Faulty on off switch    Mains supply faulty     Mains power off or  faulty     Battery not charged   Faulty battery     Battery missing     Sensor not connected     Faulty battery     Faulty circuitry     Mains supply faulty     Remedy    Check supply  plug  ON      Replace switch  front  panel      Reset ci
63. This function can be enabled disabled through the AV900   V4 on screen menus  Service Sub menu  see section   3 5 4 in the ventilator User Manual            X    2  The spirometer flow sensor heads are built into the  inspiratory and expiratory airways within the  absorber    The sensor signal cabling is routed internally to a  connector at the rear of the absorber valve block    3  The sensor for the oxygen monitor is built into the  inspiratory airway within the absorber     4  The breathing bag is mounted on a rotating arm     5  The canister release mechanism is operated by a  lever at the base of the canister assembly     6  Gas flow through the canisters is from top to  bottom     145    O09 TNE OD ON ee TO NS Se                   A100SP Circle System Absorber mounted on Prima SP Anaesthetic Machine  The absorber is fitted with optional on off  bypass  switch  and manometer    Adjustable pressure limiting valve  APL valve   Expiratory non return valve  NRV   Manometer  optional    Inspiratory non return valve  NRV    Absorber on off  bypass  switch  optional   Inspiratory hose connector   Bag ventilator switch    Expiratory hose connector    146    Condensate drain  Canister lever  Canister assembly  Reservoir bag   Bag swivel arm    Oxygen sensor                      A100SP   AV900 V4 connections       1  Inlet   from DRIVE GAS outlet on AV900  control unit     2  Exhaust outlet from APL Valve   connect  to Anaesthetic Gas Scavenge System          3  Inlet   fresh gas 
64. a smooth  sweeping movement   Check that the only Gauges indicating  pressure are those for Oxygen and  Nitrous Oxide   Check again that the only indicated  flow is the Oxygen Basal Flow    Turn the Nitrous Oxide Cylinder OFF   Observe the pressure gauge reading   a leak will be indicated by the needle  pointer falling towards zero   Maximum acceptable leak is 200 kPa   Turn the Nitrous Oxide Cylinder ON   Turn the Nitrous Oxide   Air Control  Switch  F  to select Air    Turn the Air Cylinder ON    Check that the Air pressure gauge  responds in a smooth sweeping  movement    Check that the only Gauges indicating  pressure are those for Oxygen   Nitrous Oxide and Air    Check again that the only indicated  flow is the Oxygen Basal Flow    Turn the Air Cylinder OFF    Observe the pressure gauge reading     53    25   26     27   28     29     31     32     33                 A leak will be indicated by the needle  pointer falling towards zero   Maximum acceptable leak is 500 kPa     Turn the Air Cylinder ON    If Carbon Dioxide  COz  Cylinder  supply is fitted   turn CO2 cylinder ON   Check that the CO2 pressure gauge  responds in a smooth sweeping  movement    Check that ALL cylinder pressure  gauges are indicating pressure   Check again that the only indicated  flow is the Oxygen Basal Flow    Turn the CO2 cylinder OFF    Observe the pressure gauge reading   a leak will be indicated by the needle  pointer falling towards zero   Maximum acceptable leak is 200 kPa   Turn the CO
65. above 1 L min  the high flow tube  reading indicates the rate of flow for that gas     3 7 3 Carbon Dioxide Flow  Restriction   The maximum flow of carbon dioxide  if   fitted  is restricted to 500 ml min     DESCRIPTION       3 8 Vaporizers    CAUTION  Read the instruction manual supplied with the    vaporizer before clinical use     3 8 1 Vaporizer Mounting    Systems  Vaporizers for the administration of volatile  anaesthetic agents can be fitted to  customer   s requirements as follows     Up to two Penlon Sigma Delta  Selectatec compatible vaporizers   mounted on a Selectatec compatible  universal backbar    One or two Penlon Sigma Delta  Cagemount vaporizers mounted on a  Modura rail  check that relevant  national standards for your country  allow fitment of more than one  cagemount type vaporizer       a     WARNING   Vaporizers must always be securely mounted   and never used free standing    Unmounted vaporizers may be accidentally  tipped resulting in uncalibrated and excessive  volumes of liquid anaesthetic drug entering  the breathing system     Vaporizers of any description must not be  installed or connected between the Common  Gas Outlet  CGO  and the Breathing  System  unless they are specifically  designed for such use   If this is done  the  oxygen flush flow will pass through the  vaporizer  and severe overdosage may  result      3 8 2 Selectatec Compatible  Vaporizers   Selectatec compatible vaporizers   e g  the   Sigma Delta with the Selectatec connector
66. achine     2  Remove the cover and flowmeter  knobs  as detailed in section 9 8   3  If the O2 valve capsule  1  is to be    removed  take off the O2 gear  4  by  loosening the grubscrews  5      96             Undo the two securing screws and    washers  6  and withdraw the  capsule  1     To replace the flow control valve  light   ly smear the O seals  7  with Fomblin  and insert the capsule into the mani   fold block    Align the two slots  8  with the secur   ing screw holes  and refit the screws  and washers  6     If necessary reset the gears  as  detailed in section 9 8  and check the  gas flows    Replace the flowmeter knobs and  cover    Perform a function test  section 6     SERVICING PROCEDURES    9 10 Oxygen Reservoir    Removal Replacement    1  Remove all oxygen cylinders and dis   connect the oxygen pipeline hose     CAUTION  Ensure that any residual gas in the machine is  exhausted before commencing the dismantling  procedure     2  Remove the cover at the back of the  machine to gain access to the rear of  the flowmeter assembly     3  Unscrew the nut  1  to release the  oxygen reservoir  2  from the mount   ing bracket    4  After refitting  perform a function test   section 6           Note   Early type reservoir is  shown in the illustration             97    SERVICING PROCEDURES    9 11 Secondary Regulators  Check and Reset Pressure  Output    1  Turn the Gas Delivery Switch to OFF     2  Disconnect all pipeline and cylinder  supplies    3  Remove the fro
67. acks  and check that  all clips and connections are secure     Connect the Oxygen Supply to the anaesthetic  machine using the pipeline hose     Check that the oxygen pipeline pressure gauge  indicates the correct pipeline pressure     Check the security of the Oxygen Hose connections  by tug testing the hose  A     Perform this test at both ends of the hose by tugging  gently whilst grasping the hose between thumb and  forefinger    Firm leak free joints should be maintained  without any  relative movement of the spigot  ferrule or hose   Note  Renew all faulty hose assemblies     Hoses   Test for leaks from each hose by brushing leak  detecting fluid generously about both end fittings     Pipeline Non return Valve  NRV    Test for leaks  using a pipeline non return valve or ball  valve assembly    The leak rate should be less than 10 kPa in one  minute    Alternatively  fit a tube securely to cover the hole in the  centre of the Oxygen Cylinder Yoke Seal  B     Brush leak detecting fluid over the other end of the  tube    Leaks will be indicated by bubbling of the fluid     Dry with paper towel  fix any leaks  replace hose NRV  if necessary      Connect the Nitrous Oxide Supply to the anaesthetic  machine using the pipeline hose   Repeat tests 5  to 10 for this gas     Connect the Air Supply to the anaesthetic machine    using the pipeline hose   Repeat tests 5 to 10 for this gas     51       FUNCTION TEST    6 5 Cylinder Gas Supply And  Pressure Reducing Valves     CAUTION
68. and Warning Labels     mo e NDE    Penlon recommends that after servicing the Anaesthetic Machine should be given an   Acceptance Check  by an Anaesthetist before being returned to Operational Use     71    SERVICE SCHEDULE       8 3 Twelve Month Service Checks   To be performed at 12  36  and 60 Months of a 72 month cycle      NOTE    AT 36 MONTHS ALL GAS CARRYING PIPELINE HOSES SHOULD BE REPLACED  These are not  included in the Preventive Maintenance Kits  please order separately    AT 60 MONTHS ALL CYLINDER GAS SUPPLY PRESSURE REDUCING VALVES SHOULD BE  REPLACED  These are not included in the Preventive Maintenance Kits  please order separately     Machine Frame     1  Check integrity of structure and security of all attachments   2  Check for damage    3  Check wheels for freedom of movement  check front braking   4 Check that drawers and writing tablet move freely     Electrical Safety Test     1  Check mains lead and plug for damage   2  Perform Electrical Safety Tests as relevant for country of use   3  Check outlet sockets for correct function     Pipeline Hose Checks    1  Examine pipeline connectors  probes and hoses for cleanliness and damage    2  Perform anti confusion checks    3  Replace pipeline filters   4  Perform pressure Leak Checks of all hoses with cylinders removed and Gas Delivery  Switch in the OFF position     Cylinder Yoke Assemblies    1  Check for damage  cleanliness  all fittings are tight and clamping screws for freedom  of movement    2  Replac
69. and neonatal  patients    The oxygen monitor is a module within an  anaesthesia system    The oxygen monitor is intended for use by  health care providers  i e  Physicians   Nurses and Technicians for use with patients  during general anaesthesia     3  DESCRIPTION    3 1 Framework and General  Construction    Frame   The machine has a cast aluminium base   extruded aluminium uprights  with aluminium  and plastic panels     Mobility   Trolley models have four castors  with a  brake on each of the front castors  The  castors are five inches diameter    A footrest is built into the front of the  machine    To aid manoeuvrability  two side handles are  provided     Mounting posts and brackets   A    T slot    mounting system is built into each  side upright  to allow the use of pole mount  brackets  V brackets  and ventilator  mounting brackets    The pole mount upright  see illustration  can  be used to mount a complete AV series  Ventilator  or bellows unit only    V brackets can be used to mount a gas  scavenging system  suction units  and  accessories     Draw units and work surfaces   The machine can be fitted with a base  drawer unit  as illustrated  plus two  additional smaller drawers    The work surface has raised edges to  prevent instruments  vials etc   from rolling  off     The Prima SP 102 illustrated on this page is  equipped with a standard full width top shelf  unit suitable for a large monitor  and optional  pull out writing tablet mounted under the  work surfa
70. anical AHD Label   Flowmeter Overlay   Flow Tube Cover   Knob   O2    Anti tamper    Block   Knob   N20  Air  CO2   Spur Gear  50 Teeth    M4 x 8 Grubscrew   M4 Nut   Flow Tube  O2   Low   Needle Valve  O2   up to March 2002  Needle Valve  O2   up to March 2002  Spur Gear  20 Teeth    M4 x 35 Screw  Cap Head St St   Flow Tube  O2   Cascade   Needle valve capsule  N2O   up to March 2002  Needle valve capsule  N2O   from March 2002  M3 x 6 Screw  Socket Head   Ball   0 25    Manifold Block   Cascade  UK spec   Needle Valve   M3 Plain washer   Spacer Plate   Screw  M4 x10  CSK St  St     Label   Prima SP   Mixing Chamber   Cascade  UK spec   Top Cover   Flowmeter Panel   Elbow   8 mm Tube   Luminescent Panel   Elbow   10 mm Tube   Screw  M4 x 12 SKT HD Cap   Flow Tube N20 Low   Flow Tube N20 Cascade   Screw  M3 x 6   Screw  M5 x12   6 mm Tube   Elbow   6mm Tube   Spring  not required on pre March 2002 models   Grubscrew  M4 x 6   Label O2 UK   Label N2O   Stop Pin   Grubscrew  M4 x 4   Selon Seal  2 BA   Coupling   Tube Connector   6 mm Tube   8 mm Tube   Product Label  including CE Mark    124    Qty     s required       ANMNDPHHHNMHAPNMNHHAWHAAHANMDHHAHAHANYNA oO KTH AWA HANA HHAnNNAH He Bees    As required  As required  1                re Hh CR       A O GE C                                                                    28    1   39 ARA    Aad   18    L  32N  AN  31  1  yy  ae    4   ae   0293 01057  O seal Washer  0691 36247  O seal Bobbin  36075  Washer  0691  O seal 
71. any residual gas in the machine is  exhausted before commencing the dismantling          procedure    Removal   1  Remove all cylinders and disconnect  all pipeline supplies    2  Disconnect the machine from the  mains electrical supply  if necessary     3  Remove the flowmeter cover   see  section 9 7     4  Remove the four screws  1  securing    the flowmeter assembly to the  machine frame     5  Disconnect the wiring connector  2     machines with flowbank lighting only   6  Disconnect the tubing  3  where nec     essary to allow the removal of the  flowmeter assembly                    Refitting   7  Reconnect the tubing at the back of  the flowmeter assembly   see section  9 5    Reverse the removal procedure   Carry out a machine Function Test    section 6        o     106    SERVICING PROCEDURES    9 20 Flowmeter Lighting    Removal    CAUTION   Ensure that any residual gas in the machine is  exhausted before commencing the dismantling  procedure     Removal   1  Remove all cylinders and disconnect  all pipeline supplies    2  Disconnect the machine from the  mains electrical supply    3  Remove the flowmeter cover   see  section 9 7     4  Remove the flowmeter tubes and    flowmeter assembly  see sections  9 18 and 9 19     Disconnect the tubing connector  1    Remove the three screws  2   and  detach the manifold  3     7  Remove the lighting assembly  4      ou                            Refitting   8  Reverse the removal procedure   Refer to section 9 5 for tubing 
72. as fall   en to the minimum safe level  the  oxygen monitor will automatically  shut down to avoid permanent  damage to the battery    If the internal battery is fully dis   charged  the oxygen monitor will  not function in the event of mains  power failures    The battery must be recharged  before the oxygen monitor is used  clinically  otherwise back up time  can not be guaranteed    See section 3 13     CAUTIONS    1     Do not sterilise the oxygen sensor or  control unit components    These components are not  compatible with sterilisation  techniques and damage may result     Do not autoclave or expose the  sensor to high temperatures     If the sensor shows signs of being  affected by condensation  dry the  sensor with soft tissue    Do not use heat to dry the sensor     NOTES    1     The O2 SENSOR FAULT alarm  indicates that one of the following  conditions has occurred    a  Internal electrical fault   b  Software electronics fault   c  Oxygen sensor fault     The concentration read out may  in  certain conditions of excess pressure   show a value above 100     To accommodate these conditions it is  possible to set the high alarm value up  to 105   see section 5      To maintain maximum sensor life   always remove the unit from the  breathing circuit after use     2  PURPOSE       The Prima SP anaesthesia workstation  range is intended to provide controlled  concentrations and flows of anaesthesia  gases into a patient breathing system  from  where the anaesthesia ve
73. at N2O now flows  Operate N2O Air selector switch to Air position and check that N2O ceases to flow  and that Air is now available    Return switch to the N2O position and set the oxygen flow back to minimum   Refer to the table below  and check   Oxygen concentration at the oxygen flows  specified     Adjust the Oxygen Flowmeter control only   Do NOT adjust the N2O flowmeter     Note The flow rates for N2O are given as a guide only   O2 Flow  02  in O2 N2O mixture  N20 Flow   0 25 L min 27   33   0 5   0 67   0 5 L min 27   33   1 0   1 35  0 8 L min 27   33   1 6   2 1  1 5 L min 27   33   3 0   4 1  3 0 L min 27   33   6 1   8 1     Cascade flowmeter models   March 2002 onwards    Repeat tests by reducing Oxygen flow back to Basal Flow conditions  referring to the  values in the above table   Check gas delivery switch for correct operation    niga Fail Alarms     2   3     4     B  1   2   3   4  5    D    Simulate an Oxygen fail by unplugging the Oxygen Pipeline from the outlet   Check that before the supply of oxygen is exhausted that the Oxygen Fail Whistle  sounds for a minimum of 7 seconds    Check that any Nitrous Oxide and Carbon Dioxide  if fitted  Gas Flows are cut off   Check that the Oxygen Supply Visual Indicator turns from fully green to fully red   Turn on Oxygen cylinder    Check Oxygen Basal flow reinstated and whistle sounds briefly    Turn off Oxygen cylinder and repeat above tests     ack Bar Assembly   Selectatec   Checks     Replace Selectatec O rings 
74. aulty safety capsule  valve     Valve shuttle or seals  defective     Loose valve retaining  screws    Worn or damaged    O     seals     Vent port blocked     Loose valve     Worn O seals     Valve shuttle or seals  defective     Remedy    Tighten     Replace    O    seals     Replace valve     Replace as necessary     Replace components  as necessary     Turn on supply     Turn on gas delivery  switch    Replace valve    Fit replacement cap   sule valve    Tighten    Replace seals    See instructions under  component group     Tighten retaining  screws    Replace seals    Fit replacement valve     FAULT FINDING       Component    Warning Whistle     Pressure relief valve   fresh gas circuit     Symptom    Leakage at whistle     Whistle continually  sounds when oxygen  on     Whistle fails to sound     Leakage through valve     Relief pressure too  high     Relief pressure too low     66    Possible Fault    Loose connection     Faulty whistle   Faulty whistle     Low pressure in circuit     Faulty whistle     Low pressure in circuit     High pressure in circuit     Faulty valve    Faulty valve    Remedy    Tighten   DO NOT use  excessive force      Replace whistle   Replace whistle     See instructions under  component group     Replace whistle     See instructions under  component group     Reduce pressure   Check for blockage in     downstream    circuit     Replace complete  valve    Replace complete  valve     FAULT FINDING       Component    Emergency oxygen  flush    
75. ce    Other options include additional high level  shelves and a CPU tray mounted on the  drawer unit                    Mme                                               5  aA  D    tin    S    Prima SP 102 with  standard top shelf  and base drawer unit             Pole mount system  fitted to frame upright    DESCRIPTION       3 2 Gas Circuit    Gas Supplies   For each size machine  a variety of cylinder  and pipeline combinations can be added to  the basic specification of oxygen and nitrous  oxide cylinder and pipeline supply     For example  the Prima SP102 can be sup   plied with two extra gas cylinders  choose  from one additional oxygen  one additional  nitrous oxide  one carbon dioxide  one air    and one extra pipeline supply   Air     Note a  Kits are available for fitment to  existing machines   see section  7  Ordering Information    b  Carbon dioxide is not available  on US specification machines   c  Helium and Xenon are available  to special order    Cylinder Yokes   The cylinder yokes are rear mounted and  conform with ISO standards for pin index  fitting    To ensure that only cylinders of the  appropriate gas may be installed the yokes  are designed so that the retaining latch  cannot be closed unless the index pins are  fully engaged     Pipeline Inlets   Machines can be fitted with up to three  pipeline gas inlets mounted on the rear of  the machine     Pipeline supply hoses are connected by  non interchangeable  screw threaded unions   NIST      Filters 
76. connec   tions    9  Carry out a machine Function Test     see section 6     107    SERVICING PROCEDURES    9 21 Oxygen Monitor    Battery Replacement    Removal   1  Disconnect the machine from the  mains electrical supply    2  Remove the four screws  1     3  Carefully detach the front cover    assembly  2   to allow access to the  battery  3     4  The battery is attached to the side  panel with an adhesive pad   Disconnect the battery wiring and  detach the battery and pad                    Refitting   5  Reverse the removal procedure   Carry out a machine function test    section 6        108    SERVICING PROCEDURES       9 22 Oxygen Monitor Sensor     Sensor   Cleaning and Disinfection    CAUTION   If you use ethylene oxide for sterilisation  use only a low  temperature ethylene oxide method  Do not immerse the  sensor in any cleaning solution    Do not autoclave or expose the sensor to high temperatures     Sensor Replacement    WARNING   The sensor contains    a  A small quantity of electrolyte  classified as a harmful  irritant which is potentially hazardous    b  Lead    Do not attempt to open a cell   ALWAYS check the integrity of the sensor assembly before    use   Once exhausted  the sensor must be disposed of according to  hospital  local  state and federal regulations     Sensor Expiry Date    The approximate expiry date is marked on the sensor label   using two boxes which represent the year and month    Thus  for a sensor marked as below  the approximate ex
77. control unit    Exhaust outlet from APL Valve    connect to Anaesthetic Gas Scavenge  System   Inlet   fresh gas hose from  anaesthetic machine Common Gas  Outlet   Outlet   sample line to Pressure  Monitor Port on ventilator   Interface cable   Bag Vent switch   connects internally to Prima SP  On Off Switch interface  then to  connector on AV900 rear panel    Spirometer signal cable   connected  to socket on AV900 V4 rear panel           In board A100 Absorber and AV900 Ventilator Bellows   AV900 Control Unit mounted on side bracket or shelf   For additional information  refer to the A100 user documentation           gt  AV900 Control Unit  Spirometer and pressure  monitor connections   Refer also to AV900 V4 user T  manual  EZ    su 1a    A  z               ELA CGO Block  rs                           amp   S     E    See QD      lt   2 a  unnn    CA  aro CEA    Canister Sen    itd          O   h  ion  Ss       In board A100 Absorber   AV900 Ventilator and Bellows Unit  mounted on side bracket or shelf   For A100SP Absorber  refer to the relevant  user documentation       Spirometer and pressure  monitor connections   Refer also to AV900 V4 user       Canister      oh    ONDA    o    10     PRE USE CHECKS   All models       ee pe    y     i     l  i     i  N                     Bellows   Control Unit   Outlets to Anaesthetic Gas  Scavenging System  AGSS   Bacterial Filter   Absorber valve block   Heat and moisture exchanger  Patient   CGO Block on anaesthetic  machine  Fres
78. d moisture exchanger 16  Connector   Reservoir Bag  SP on off switch  and A100SP  Patient 17  Inlet   Absorber   Fresh Gas Absorber Bag Vent lever position   CGO Block on anaesthetic Supply 26  Interface connections on Prima  machine  Fresh Gas Supply  18  Drive Gas Inlet   Ventilator SP and A100SP   Auxiliary Outlet on anaesthetic 19  Drive gas Outlet   ventilator 27  APL Valve   machine  Drive Gas Supply  control unit to bellows 28 Outlet from APL Valve to AGSS  Flow sensor   expiratory 20  Outlet   Exhaust Valve 29 Oxygen sensor    148    AV900 Ventilator   Oxygen  Sensor Calibration    Calibrate the AV900 oxygen monitor  system with the sensor in position in  the absorber     1     Pull the Absorber On   Off lever  1   down to its OFF position   The flow will bypass the absorbent     Remove the breathing circuit hoses  from the inspiratory and expiratory  connectors  2  on the absorber    This will give a free flow of oxygen  through the sensor     Switch on the ventilator and the  anaesthetic machine gas delivery  switch  3     The oxygen monitor automatically  switches ON when the ventilator is  switched on    Ensure that all vaporizers are OFF     Apply 100  oxygen only  at 5 L min   from the flowmeter     Allow the oxygen to flow until the  oxygen monitor readout stabilises     Calibrate the sensor  using the AV900  ventilator menu procedure     Press the menu switch and select the  Oz monitor sub menu     Scroll to CALIBRATION    If the menu shows 21   calibration   
79. de flow control  and check that the maximum flow rate  available is 600 ml min     18  Check that the float is stable and that it  spins within the glass flow tube     19  Close the Carbon Dioxide Flowmeter  control gently until no gas flows     20  Set the following flow rates   Oxygen  5 L min  N2O  5 L min  Air  5 L min     21  Check that the only flows indicated are  that of Oxygen and Air   Check that the Nitrous Oxide flow tube  indicates a flow of zero     22  Set the Nitrous Oxide   Air switch  B  to  select Nitrous Oxide     23  Observe that the Air supply is cut off and  that of Nitrous Oxide is reinstated   Check that flow tubes now indicate flows  of Oxygen and Nitrous Oxide  while the  Air float has dropped to zero     55    FUNCTION TEST    6 7 Gas Safety Devices   6   Mechanical AHD    1  Fully open the Nitrous Oxide  Flowmeter and check that there is no  flow of gas indicated on the N2O flow  tube  7     2  Fit an Oxygen Analyser to the output  of the Common Gas Outlet   Refer to the table below  and check    Oxygen concentration at the oxygen  flows specified    Adjust the Oxygen Flowmeter con   trol only   Do NOT adjust the N20  flowmeter   Note The flow rates for N2O are 8   given as a guide only     02 Flow  02 N20 Flow   L min   in O2 N20   mixture    0 25 27   33  0 5   0 67   0 5 27   33  1 0   1 35  0 8 27 33  1 6   2 1  1 5 27 33  3 0   4 1    3 0 27 33  61 81 10      Cascade flowmeter models   March 2002  onwards    3  Turn OFF the Gas Delivery Switc
80. de frame upright     26    SPECIFICATION    4 6 Oxygen Failure Warning Devices    1  Gas system whistle  2  Visual indicator  direct pressure operated    4 7 Oxygen Flush    Button on CGO block  The system supplies 35   75 L min when fully depressed     4 8 Mechanical AHD System    Minimum oxygen concentration 30   3   of total O2   N20 flow   Basal Flow   Cascade flow tubes   Oxygen basal flow 50 75 ml min    Single Flow tubes  Oxygen basal flow 100 200 ml min    All models   up to March 2002   Oxygen basal flow 100 200 ml min    Reduced pressure from secondary regulators   See section 4 4     4 9 Environmental    Operating Conditions    Temperature  10 to 38  C  50 to 1009F   Atmospheric Pressure range 70 kPa to 106 kPa   Altitude 2438 m  8000 ft  maximum  Humidity 10   95  R H  non condensing   Transport and storage temperature    Basic machine  5 to 600C  23 to 1409F   Oxygen monitor option  5 to 50  C  23 to 1220F    Cleaning Wipe external surfaces with dry or damp cloth     Use mild soap  or disinfectant solution if necessary     27    SPECIFICATION    MRI Compatibility    See additional notes in section 3 16     The following Prima SP system components are MRI compatible    Prima SP basic machine  includes any variant of Back Bar  Flowmeter Bank  Drawers   Monitor Shelves  Additional Gases    A100  In board and pole mounted    see section 3 16     A100SP Absorber  Caution   A100SP is normally fitted as part of an integrated system  comprising Prima SP and AV900 V4  ven
81. densate from the absorber    Refer to section 7 3 1 in the A100 User Manual   WARNING Condensation  which may collect in  the bottom of the absorber is caustic and care  must be taken not to spill it on the skin when  draining the condensate trap    After draining ensure that the drain valve  4  is  fully closed    Apply the brake on the anaesthetic machine    Lift the right hand side of the absorber and slowly  slide the assembly out from the side of the machine   The absorber assembly weighs approximately  17 kg   support the absorber at the front and  side as it is removed from the machine     Remove the covers and canisters    5     Press the retaining clip and detach the manometer   5   Do not autoclave   Unscrew the retaining nut  6  and remove the bag  arm  7     Remove the screws  8  securing the top cover  9    and detach the cover from the absorber    Do not autoclave    Turn the lever  10  at the base of the absorber anti   clockwise  and remove the top canister  11   sideways after disengaging the rubber seals by  vertical movement    Remove the bottom canister    Dispose of the soda lime granules  and wash the  canisters    Do not refit the canisters until the autoclave  procedure is completed    Remove the screws  12  securing the bottom cover   13   and detach the cover from the absorber    Do not autoclave     152                                  Bellows Assembly   10  Turn the bellows housing  12  anti clockwise   then lift it from the base   Remove the bello
82. e  gently move it  through the air to allow room air to  circulate for 20 seconds     3  Allow the O2 Concentration reading   1  to stabilise   This will take at least 30 seconds     4  Press the CAL button  6  to enable  sensor calibration mode   The O2 CONC LED  1  will flash  and  a single    beep    warning will sound     ALARMS    HIGH O2 ALARM  LOW Oz ALARM  O2 MONITOR INOP  O2 SENSOR FAULT  02 SENSOR LOW  LOW BATTERY  MAINS FAILURE    Q Ponion    GAS DELIVERY switch  must be  ON  to operate  O2 Monitor    O2 Monitor       5  Turn the SET CAL knob  7  until the  display shows 21      6  When value is set  press the CAL  button  6  again to accept calibration  setting    The display LED  1  will now stop  flashing and a single    beep    warning  will sound   The display will return to normal  operation     Sensor Low Indication   The unit has a self detect feature to indicate  when the sensor life is low    During calibration  if the O2 SENSOR LOW  alarm LED  8  illuminates  and a single     beep    warning sounds  this indicates that  the sensor must be replaced    Sensor output will fall very quickly to  zero over a period of two to three weeks  from the first time that the alarm is  activated     PRE USE CHECKS   Oxygen Monitor    HIGH ALARM SET    SET  LOW ALARM SET         m e    SET CAL       ALARMS    HIGH O2 ALARM  LOW O2 ALARM  O2 MONITOR INOP  O2 SENSOR FAULT  02 SENSOR LOW  LOW BATTERY  MAINS FAILURE    GAS DELIVERY switch  must be  ON  to operate  O2 Monitor
83. e Bodok seals  DO NOT apply grease or lubricant   3  Replace yoke inlet filters     Pressure Gauge and regulator checks     1  Examine all pressure gauges for damage and clarity   2  Check movement of needles is smooth   3  Perform pressure Leak Test of all cylinder Gas Supplies with all Pipelines    disconnected from supply    Perform this test with Gas Delivery Switch turned ON and Oxygen Basal Flow  blanked off    For Air   N20 repeat the test with the selector switch in the required position   On completion of tests reinstate basal flow by removing test blanking plug   Check output pressure of Primary Regulators    Check output pressure of Secondary Regulators     NOS    Flowmeter and Mechanical AHD Checks   1  Check Flow Control Valves for smooth operation and positive off position   2  Turn On Gas Delivery Switch  Check Basal Flow of O2 is delivered with all Flow Control Valves closed down     All models   up to March 2002 100 200 ml min  Cascade O2 flowmeter  March 2002 onwards 50 75 ml min  Single O2 flowmeter   March 2002 onwards 100 200 ml min    3  Check flow tubes     72    SERVICE SCHEDULE       14     15     Check Bobbin Floats move freely and rotate within the glass tube    Check maximum flow capability of all gases    If CO2 is fitted check that the maximum flow is restricted to 500 600 ml min flow rate   Connect O2 Monitor to CGO    Close all gas flow valves    Ensure N20   Air Selection switch is positioned for N2O    Fully open N2O Flow Control Valve   Chec
84. e Leak Test of all cylinder Gas Supplies with all Pipelines    disconnected from supply   Perform this test with Gas Delivery Switch turned ON and Oxygen Basal Flow  blanked off     4  For Air   N20 repeat the test with the selector switch in the required position   5  On completion of tests reinstate basal flow by removing test blanking plug   6  Check output pressure of Primary Regulators   7  Replace secondary regulators  8  Check output pressure of Secondary Regulators   Flowmeter and Mechanical AHD Checks   1  Turn off all cylinders  disconnect pipeline hoses  and drain all gases from the  machine   2  Replace flowmeter tube seals and filters   3  Check Flow Control Valves for smooth operation and positive off position   4 Turn On Gas Delivery Switch  Check Basal Flow of Oxygen is delivered with all Flow Control Valves closed   All models   up to March 2002 100 200 ml min  Cascade O2 flowmeter  March 2002 onwards 50 75 ml min  Single O2 flowmeter   March 2002 onwards 100 200 ml min    78    SERVICE SCHEDULE       OPNI    10     12     13   14     15     16     17     Check flow tubes    Check Bobbin Floats move freely and rotate within the glass tube    Check maximum flow capability of all gases    If CO2 is fitted check that the maximum flow is restricted to 500 600 ml min flow rate   Connect O2 Monitor to CGO    Close all gas flow valves    Ensure N20   Air Selection switch is positioned for N20    Fully open N2O Flow Control Valve   Check that NO flow of N2O occurs   
85. e given in the Prima SP Service  Manual     These checks will not in themselves ensure  the safe use of the apparatus  which remains  the responsibility of the qualified practitioner  in charge of it     31    PRE USE CHECKS    PRE USE CHECKLIST    The machine must be carefully inspected and checked as follows     An incorrectly functioning machine must be repaired by a suitably  qualified person before use     N    D OTA 2    0   1     Check for visible damage  machine stability  and condition of gas  supply hoses    Check for labelling which may indicate status of machine  including  faults or recent servicing    Check correct connection of electrical supply    Check correct connections of gas supplies    Check adequate pipeline supply and back up cylinder supply   Switch on gas delivery switch  and note special operating system   Check functioning of flowmeters    Check function of Mechanical AHD  if fitted     Check function of Air N2O interlock switch  if fitted     Check correct connection and functioning of the vaporizers   Check functioning of oxygen flush    Check leak rate of low pressure gas system    Check the integrity of the patient circuit    Test the alarm system     Refer to Section 5 in the User Manual for further information     12     14     Ancillary equipment   Check operation of the AGSS    Check functioning of ventilator  including disconnect alarm   Check that the oxygen analyser and other patient monitoring  equipment functions correctly     Refer to
86. e order  of  a licensed practitioner     Statements in this manual preceded by the  following words are of special significance     WARNING means there is a  possibility of injury to  yourself or others   CAUTION means there is a possibility  of damage to the apparatus    or other property     NOTE indicates points of  particular interest for more  efficient and convenient  operation     Always take particular notice of the  warnings  cautions and notes provided  throughout this manual     1  WARNINGS AND CAUTIONS       The following WARNINGS and  CAUTIONS must be read and  understood before using this  anaesthetic apparatus     WARNINGS    1  This apparatus is designed for use  only with non flammable anaesthetic  agents  It must not be used with or in  close proximity to flammable  anaesthetic agents  due to a possible  fire or explosion hazard     2  Exterior panels must not be removed  by unauthorised personnel and the  apparatus must not be operated with  such panels missing    On machines with an electrical power  supply  there is a possible electric  shock hazard     3  No oil  grease or other flammable  lubricant or sealant must be used on  any part of the machine in close  proximity to medical gas distribution  components    There is a risk of fire or explosion     Unless otherwise stated  O rings and  other elastomeric seals must only be  lubricated with an oxygen compatible  lubricant  Use sparingly     4  When attaching cylinders of medical  gases ensure that the 
87. ecks at section 5 4    36    PRE USE CHECKS   All models    5 4 Leak Rate Check   Low  Pressure Gas System    1  Attach a side branch connector to the fresh  gas outlet on the CGO block outlet   Connect the side branch tube to a  sphygmomanometer    2  Turn ona flow of 150 ml min of oxygen   Block the open port of the connector with a  finger    The pressure in the low pressure gas system  will rise and be displayed on the  sphygmomanometer    3  Check that the pressure rises to at least 100  mmHg    Release the finger seal immediately the  pressure is reached     CAUTION    Do not maintain closure of the open port longer than  necessary to perform the test     This test should be performed    a  With all vaporizers    off    and isolated    b  With each vaporizer in turn set to 1      If an optional CGO switch is fitted  see section 3 9   the  lever must be set to the vertical position     5 5 Electrical Supply  if fitted     1  Connect the machine power lead to a  suitable mains supply socket     2  Set the switch  A  to ON   Check for correct function of all electrical  equipment  including devices powered by the  auxiliary power outlets on the rear of the  machine     3  Machines with optional O2 Monitor   Check for correct fitment of the mains lead   B  into the rear of the monitor unit   Switch the Gas Delivery Switch  C  to ON   Check that the monitor control panel LEDs  activate     4  Machines with optional flowmeter lighting   Check for correct operation        37  
88. ee 3 13 2      Back up Battery   Should the electrical power supply to the machine  fail  the emergency battery supply for the unit comes  into action automatically    This is indicated by the illuminated yellow    MAINS  FAILURE    LED alarm  accompanied by a single  audible tone    The battery is maintained in a fully charged state  during normal use  i e  the machine connected to the  mains power supply     A fully charged battery will power the unit for a  minimum of 60 minutes     Low Priority Battery Low Alarm   When the battery is discharged  and the mains  power supply is not restored  the    BATTERY LOW     LED alarm will illuminate  accompanied with a low  priority alarm sound     Medium Priority Battery Low Alarm   When the battery is further discharged  and the  mains power supply is not restored  the    BATTERY  LOW    LED will flash at a two second rate and a  medium priority audible warning will be given when  the minimum safe level of voltage is reached     Low Battery Shut Down   WARNING   When the battery voltage has fallen to the minimum  safe level  the oxygen monitor will automatically shut  down to avoid permanent damage to the battery     Recharging the Battery   Charging of the back up battery takes place  automatically when the mains power supply is on   irrespective of the position of the machine gas  system On Off switch position    NOTE   The stated battery back up period will only be available if  the battery is kept fully charged  After the back 
89. ent Disposal   always follow the  instructions supplied with the filter or heat and  moisture exchanger    Always renew components at the recommended    interval    2  Follow the instructions in the relevant user  manual for connection to analysers and  monitors    3  Ventilator connections shown are for AV900 V4    with spirometry and oxygen monitor   For AV900 V 3 and AV800 ventilators  please  refer to section 5 in the relevant user manual     4  For A100SP  refer also to the user  documentation supplied with the absorber     5 6 6 Breathing System Pre use Test    Connect the CGO block outlet on the machine to the  fresh gas inlet of the breathing system    If an optional CGO switch is fitted  see section 3 9    the lever must be set to the vertical position     NOTE   This machine must be fitted with a breathing system  complying with approved design parameters  at the  selection of the qualified practitioner     The breathing system components do not constitute  part of the machine but connections between the  machine and breathing system should be verified as  follows     1  Occlude the adjustable pressure limiting   APL  valve  if fitted   and the patient  connection port     Press the oxygen flush valve button briefly   Check that the reservoir bag inflates    If the system includes a manometer  inflate  the bag to approximately 40 cmH20     2  Release the oxygen flush valve   Check that the pressure is maintained in the  system with less than 200 ml min fresh gas  de
90. ents must be sterilised to the manufactur   er   s recommended methods     Monitor  Refer to the monitor user instruction manual     Ventilator   AV series Ventilator Bellows Assembly  Cleaning   The user must follow the detailed instruc   tions included in section 7 5 of the ventilator  user instruction manual     Never use any hard object or abrasive agent  to clean any of the components  only a soft  cloth     If the valve seat is damaged  the diaphragm  valve will leak and this may cause serious  malfunction    CAUTION   Do not use any cleaning solution containing  alcohol or any harsh abrasive cleaning agent  on the bellows housing     Sterilisation   The user must follow the detailed instruc   tions included in section 7 of the ventilator  user instruction manual    WARNING   Do not autoclave the bellows housing  The  exhalation diaphragm valve assembly must  be disassembled prior to sterilisation   do not  sterilise the diaphragm     83                               4  I    NTS     y                                So        lt  lt   Vi  HG  Oy           lt  gt        Gi       Ny       gt     M  ZA         bed  i    JJ    DA  Feny   ja    ij                          Gas System Components   UK Specification 3 gas machine     1  CGO block   2  Auxiliary outlet block   3  Reservoir  O2   for whistle   4  Low pressure block  O2   5  High pressure block  O2   6  Low pressure block  Air   7  High pressure block  Air   8  Low pressure block  N20   9  High pressure block  N20  
91. es stated in this manual and or  accompanying labels and notices when  checked  assembled  operated  maintained  and serviced in accordance with these  instructions     To ensure the safety of this device it must be  checked and serviced to at least the  minimum standards laid out in this manual   A defective  or suspected defective  product  must not under any circumstances be used     The user must accept responsibility for any  malfunction which results from non   compliance with the servicing requirements  detailed in this manual     Additionally  the user must accept  responsibility for any malfunction which may  result from misuse of any kind or non   compliance with other requirements detailed  in this manual     Worn  broken  distorted  contaminated or  missing components must be replaced  immediately  Should such a repair become  necessary it is recommended that a request  for service advice be made to the nearest  Penlon accredited agent     This device and any of its constituent parts  must be repaired only in accordance with  written instructions issued by Penlon  Limited and must not be altered or modified  in any way without the written approval of  Penlon Limited  The user of this equipment  shall have the sole responsibility for any  malfunction which results from improper  use  maintenance  repair  damage or  alteration by anyone other than Penlon or its  appointed agents     USA and Canadian Federal Law restricts the  sale and use of this device to  or on th
92. et the Gas Delivery Switch to OFF   2  Disconnect all pipeline and cylinder supplies    MN ll  3  Remove the front cover   see section 9 6  D a 0  Retighten the screws securing the gas blocks  B l    a    e a   Reconnect the cylinder and pipeline supplies  aay P k a   gt   e  D Z O    4  Attach a pressure gauge to the oxygen regulator test point  1     i 2   Early type regulators shown in illustration  on    D     amp  ZY  SA   5  Set the Gas Delivery Switch to ON  L           0  6  Set a flow of 5 L min Oxygen  2  Reset the O2 secondary regulator to 172 kPa  25 0 psi   1  7  To increase output  turn the regulator cap   3  slowly clockwise     To decrease output  turn the regulator cap fully anti clockwise   then clockwise to the correct setting    8  Return the flow to minimum    9  Attach a pressure gauge to the N20 regulator test point  2      10  Seta flow of 5 L min nitrous oxide   Reset the N2O secondary regulator to 186 kPa  27 0 psi    To increase output  turn the regulator cap   3  slowly clockwise   To decrease output  turn the regulator cap fully anti clockwise   then clockwise to the correct setting   Return the flow to minimum     11  Set the Gas Delivery Switch to OFF   Gear Linkage    12  Remove the flowmeter control knobs  4   screws  5  and control  knob cover  6      13  Loosen the grubscrews and remove the O2 gear  7  and small  N20 gear  8               14    Check the condition of the gears and grubscrews  Replace with  new components if necessary   Clean
93. for security  damage and correct lock and release movement   2  Check dynamic flowrate and that Oxygen Fail Whistle does not operate     Oxygen Monitor  if fitted   Function and Calibration Tests   1  Connect a Test O2 Analyser into the patient circuit   2  Check that the machine O2 sensor is inserted into the absorber O2 sampling point   3  Test both O2 sensors in 100  Oxygen   4 Expose both sensors to air and check reading is 21    2   Adjust high and low O2 alarms and check alarms trigger when values are lower or  higher than reading on O2 analyser respectively   Return the alarm levels to original settings   5  Restore the sensor to correct location   Remove the test O2 analyser     Final Gas Concentration Tests   1  Check concentration by Gas   2  Check flowrate from CGO corresponds with flowmeter by concentration   3  Set 3 L min Oxygen and Nitrous Oxide Flows   Check  using O2 Monitor that the concentration from the CGO is 50    2    4  Set Oxygen to 0 5 L min  set N2O to 0 5 L min   Allow O2 monitor to stabilise then check concentration is 50    2    If fitted repeat above test substituting Carbon Dioxide for N2O   If fitted set Air to 3 L min  set Oxygen to 3 L min   Check Concentration reading on O2 Monitor is 60    2      5   6     Test Completion   Paperwork   Remove all test equipment from machine   Drain all Gas from machine    Turn OFF Gas Delivery Switch    Turn Off all cylinders    Turn off all Flow Control Valves    Attach Service and Warning Labels     So G
94. g screws  2  securing  the panel to the front of the machine    4  Slacken each screw  3  on the side of the  machine  to loosen the side clamp plates    5  Carefully pull the cover  4  away from the front of 1  the machine    Refit   6  Line up the cover with the gauges and switches     7  Check that the clamp plate on each side of the  machine is positioned to allow the sides of the  cover to fit between each plate and the frame    8  Carefully push the cover into place    9  Screw in the top fixings  2   Do NOT tighten    10  Tighten the clamp plate screws  3     11  Tighten the top fixings  2     12  Atthe rear of the machine  tighten the screws  1                                88    SERVICING PROCEDURES    9 7 Flowmeter Cover    Removal   1  Slacken the screws  1  securing the  filler strip  2     2  Remove the screws  3  and top  guard  4     3  Remove the cover  5     Refitting   4  Reverse the removal procedure                                                                 89    SERVICING PROCEDURES       9 8    9 8 1    Mechanical AHD System    Check and Adjust Basal Flow    All models   up to March 2002  Oxygen basal flow 100 200 ml min    Cascade flowmeter models   March 2002 onwards  Oxygen basal flow 50 75 ml min    Single flowmeter models   March 2002 onwards  Oxygen basal flow 100 200 ml min    NOTE  The machine must be on a flat surface     1     2     Check that all flowmeter controls are OFF   Switch the gas delivery switch  1  to ON     Check that a basal
95. gen Basal flow reinstated and whistle sounds briefly    4  Turn off Oxygen cylinder and repeat above tests     Back Bar Assembly   Selectatec   Checks   1  Replace Selectatec O rings  DO NOT apply grease or lubricant    2  Check all pillar valves open and seal correctly    3  Check pillar valve inserts sit proud or flush with the pillar valve   A sunken pillar valve insert denotes failure of the internal spring    4  Check vaporizer interlock system for correct operation     Back Bar Assembly   Cagemount   Checks    1  Check flexible hoses    2  Check security of taper connectors  Use Fomblin or other Oxygen safe grease to lightly lubricate the metal surface where  male and female connector come into contact with each other     70    SERVICE SCHEDULE       Fresh Gas Pressure Relief Valve  PRV  Test   1  Occlude CGO with a manometer   2  At 10 L min flow rate of Oxygen the PRV relieves at 41 kPa   10     Leak Test From Flowmeter to Common Gas Outlet   1  Fit a suitable Manometer Test gauge to the CGO    2  Set Oxygen Flow to maintain a pressure of 20 kPa  3 psi    Maximum permissible flow is 200 ml min     3  Perform test with and without Vaporizer s  fitted and with vaporizer s  turned both  ON and OFF   4  Set oxygen flow to minimum and remove test equipment     Common Gas Outlet Checks  CGO     1  Check security and freedom of movement of CGO    2  Check Safelock O ring  DO NOT apply grease or lubricant  3  Check Oz Emergency Flush flow rate is 35   75 L min    Auxiliary Ou
96. h   Check that both gas flows are  stopped     11     4  Turn ON the Gas Delivery Switch  Check that both flows are reinstated     5  Partially close the Nitrous Oxide  flowmeter to check that it limits the  Nitrous Oxide flow     Check that the reducing flow gives an 12     increased Oxygen concentration   Fully open the Nitrous Oxide flowme   ter     13     14     56    Close the Oxygen flowmeter control  valve gently until the mechanical stop  engages    Check that the same corresponding  values of Oxygen and Nitrous Oxide  flows are achieved as tabled above     Ensure that when the Oxygen flow is  at its basal level  all Nitrous Oxide  flows have ceased    Check that only the Oxygen Basal  Flow is indicated on the flow tubes   Close the Nitrous Oxide Flowmeter  and switch the Gas Delivery Switch to  OFF     Oxygen Fail Cut off   Set the following flow rates   Oxygen  5 L min   N2O  5 L min   COsz  if fitted    500 ml min     Turn Off the Oxygen Cylinder and  observe that the Oxygen Cylinder  pressure gauge indicates falling  Oxygen pressure     Check that the warning whistle   Oxygen Fail Alarm  sounds before  the supply of oxygen is exhausted  from the anaesthetic machine and that  its duration is a minimum of 7  seconds     Check that before the supply of  Oxygen is exhausted from the  anaesthetic machine  that Nitrous  Oxide and Carbon Dioxide flows are  cut off automatically by the Oxygen  Fail System     Oxygen Fail Warning   Check that before the supply of  Oxygen i
97. h Gas Supply   Auxiliary Outlet on anaesthetic  machine  Drive Gas Supply   Flow sensor   expiratory     rr S          11   12  13   14   15   16   17     18   19     20   21              Breathing System Connections    Note    Schematic shows AV900V4 with O2 monitor     26    On Prima SP with a built in O2 monitor  item  23 is located on the monitor rear panel  see  5 12 1     Prima SP interface cabling is shown    The absorber is fitted with a Bag Vent switch           Flow sensor   inspiratory  Sensor   pressure monitor  Expiratory Valve   Absorber  Inspiratory Valve   Absorber  Inlet   from Ventilator  Connector   Reservoir Bag  Inlet   Absorber   Fresh Gas  Supply   Drive Gas Inlet   Ventilator  Drive gas Outlet   control unit to  bellows   Outlet   Exhaust Valve   Inlet   Bellows Drive Gas    40    22  Outlet   to breathing system   23  Input socket   Oxygen monitor sen   sor   24  Input socket   spirometer   25  Input socket   Prima SP interface   SP on off switch  and A100SP  Absorber Bag Vent lever position    26  Interface connection on Prima SP   if fitted    27  Connector   pressure monitor   28  APL Valve   29  Oxygen sensor  for alternative    locations   see section 3 13 6     PRE USE CHECKS   All models       5 6 5 Breathing Circuit Schematic    Note   1  To protect the expiratory limb of the breathing  circuit  and the spirometer  use a breathing  circuit bacterial filter  4   or a heat and moisture  exchanger  6  at the patient Y piece     CAUTION   Replacem
98. he  needle valve from fully closing  This  ensures a minimum basal flow of  oxygen    DO NOT attempt to close the flow to  zero    Do not overtighten     Compressed gas supplies must be  clean and dry     When the auxiliary gas outlets are in use  ona machine with cylinder supply only  or  if the pipeline supply is not in use  check  flow rate requirements  and ensure that  adequate back up cylinders are available     NOTES    1     In the text  litres per minute is  abbreviated to L min to avoid confusion  that may occur due to the typeface used  in this manual     All tightening torques  unless specified  otherwise  are to the relevant Standard  for each fixing type     WARNINGS AND CAUTIONS  O2 Monitor       Oxygen Monitor    WARNINGS    1     We recommend calibration of the  oxygen monitor every time the  system is turned on  as a safety  precaution     Do not attempt to open the fuel  cell  The sensor contains a small  quantity of electrolyte  classified as  a harmful irritant which is  potentially hazardous     ALWAYS check the integrity of the  sensor assembly before use  See  section 3 4     Once exhausted  the sensor must  be disposed of according to  hospital  local  state and federal  regulations     The sensor measures oxygen  partial pressure  and its output will  rise and fall due to pressure  change    An increase in pressure of 10  at  he sensor inlet will produce a 10   increase in sensor output     Using the oxygen monitor    6     When the battery voltage h
99. hines     An integral flow restrictor is fitted to reduce  the flow of carbon dioxide to 500 ml min     Note that carbon dioxide is not available on  US specification machines     3 12 Auxiliary Gas Outlets  CAUTION   When the auxiliary gas outlets are in use ona  machine with cylinder supply only  or if the  pipeline supply is not in use  check flow rate  requirements  and ensure that adequate back up  cylinders are available     Oxygen  Auxiliary oxygen outlets are mounted on the  right hand side frame upright     Air   On machines fitted with an air  cylinder pipeline supply  an auxiliary air  outlet is fitted above the oxygen outlets     Supply pressure  See section 4 5                            DESCRIPTION   O2 Monitor       3 13 Oxygen Monitor  Optional                          The oxygen monitor continuously measures and   S  indicates the concentration of oxygen in the breathing  system  and triggers an alarm when the concentration Mann MUTE hace ae  varies from the set levels  pines   HIGH ALARM SET pa  rr    CERN T  CAUTION C  os se as  If your machine is fitted with an AV900 Ventilator with ai  a built in Oxygen Monitor  please refer to the AV900 NSE A  user manual for instructions on setting up and O aim Q Penton  operation  SEGAL   O  CONC   o GAS DELIVERY switch   3 13 1 System Description CEE                The Oxygen Monitor uses a fast responding  oxygen   specific  self powered sensor that achieves 90  of  final value in less than 10 seconds        O2 Monit
100. hose from anaesthetic  machine Common Gas Outlet          4  Outlet   sample line to Pressure Monitor  Port on ventilator  AV900   AV800 etc        5  Interface cable   Bag Vent switch   connects internally to Prima SP On Off  Switch interface  then to connector on  AV900 rear panel      6  Spirometer signal cable   connected to  socket on AV900 V4 rear panel                 7  Oxygen monitor sensor   connected to  socket on AV900 rear panel       Pressure Monitor Port    Spirometer connector        Input socket      Prima SP  A100SP Absorber Ob mohitorsengor    Interface connector    AV900 V4 Ventilator  Control Unit Rear Panel       Drive gas outlet                147    Co     ON OO    Schematic   Breathing System Connections    a   22            O                SE a  a E E OE T AET          hot    Breathing System Connections   Note   1  This Schematic shows an AV900 with  spirometry and oxygen monitor    2  Prima SP interface cabling is shown    3  The absorber is fitted with a Bag Vent switch           Bellows 11  Flow sensor   inspiratory 21  Inlet   Bellows Drive Gas  Ventilator Control Unit 12 Connectors   sensor   pressure 22  Outlet   to breathing system  Outlets to Anaesthetic Gas monitor 23  Input socket   Oxygen monitor  Scavenging System  AGSS  13  Expiratory Valve   Absorber sensor   Bacterial Filter 14  Inspiratory Valve   Absorber 24  Spirometer cable connectors  Absorber valve block 15  Inlet   from Ventilator 25  Input socket   Prima SP interface  Heat an
101. hysically distinguishable from the other flow  controls for identification by touch in  accordance with ISO standards     When fitted  air and carbon dioxide  flowmeters are always installed in the inner  positions on the flowmeter assembly  These  positions are blanked out if air or carbon  dioxide are not specified for the machine     NOTE   Machines with an Air supply option have the  additional option of an Air N2O Interlock switch   The machine will NOT deliver a mixture of Air  and nitrous oxide   see 3 4 4      On these machines  this switch must be in the  N20 position to allow a flow of Nitrous Oxide     The gas delivery switch  positioned to the  right of the flowmeter bank  controls the  supply of oxygen and must be in the ON  position for normal operation of the  machine     Flow control of each gas is achieved by a  needle valve comprising a polished stainless  steel needle mounted concentrically in a  common manifold block  To minimise wear  and material pick up the needle seat is  manufactured from silver  The flow control  knob is turned counter clockwise to increase  the gas flow     CAUTION   Needle valves are designed to seal with light  torque and may be damaged if tightened  excessively    DO NOT USE EXCESSIVE FORCE     3 7 2 Optional Dual Cascade  Flow Tubes   The flow of gas through dual cascade   system flow tubes always flows through the   low flow tube first  The high flow tube should   not show any flow until more than 1 L min is   set     At flows 
102. ia the ventilator  menu procedure    NOTE   Do NOT remove the spirometer heads   Press the menu switch on the front panel     Scroll down the main menu and select  SPIROMETRY     Scroll down the sub menu and select  CALIBRATION     Turn the wheel to switch display to  CALIBRATION  cal    A message will appear    Remove the spirometer head  Do NOT remove the spirometer heads  Press the wheel to initiate calibration   Repeat operations 8 to 12  twice more   Calibration is completed     Scroll to ESCAPE FROM MENUS     Press the wheel to confirm                             On Off Switch    Alarm Mute    Menu Switch       Navigator Wheel  and  Press Button       Main menu    O2 MONITOR   SPIROMETRY   FRESH GAS COMPENSATION   ON  GAS MIXTURE    VOLUME TYPE  WAVEFORM   SERVICE   ESCAPE FROM MENUS       Spirometry sub menu    SPIROMETRY   ON     gt  CALIBRATION   cal  ESCAPE FROM MENUS       150    Filling and Changing CO2  Absorbent    WARNING   If the absorbent is to be changed during  clinical use  the bypass switch  1  must be in  the    Absorber OFF    position  and adequate  fresh gas flow must be maintained to prevent  excessive build up of CO2        Removing the canisters    1  Turn the lever  2  at the base of the  absorber anti clockwise     This leaves space between the top  and bottom of the canister frame for  the top canister  3  to be removed  sideways after disengaging the rubber  seals by vertical movement             i gt      Eee  2  The absorbent in the top canis
103. ies    1  Check for damage  cleanliness  all fittings are tight and clamping screws for freedom  of movement    2  Replace Bodok seals  DO NOT apply grease or lubricant    Pressure Gauge and Regulator Checks     1  Examine all pressure gauges for damage and clarity   2  Check movement of needles is smooth   4  Perform pressure Leak Test of all cylinder Gas Supplies with all Pipelines    disconnected from supply    Perform this test with Gas Delivery Switch turned ON and Oxygen Basal Flow  blanked off    For Air   N20 repeat the test with the selector switch in the required position   On completion of tests reinstate basal flow by removing test blanking plug   Check output pressure of Primary Regulators    Check output pressure of Secondary Regulators     ONAN    Flowmeter and Mechanical AHD Checks   1  Check Flow Control Valves for smooth operation and positive off position   2  Turn On Gas Delivery Switch  Check Basal Flow of Oxygen is delivered with all Flow Control Valves closed down     All models   up to March 2002 100 200 ml min  Cascade O2 flowmeter  March 2002 onwards 50 75 ml min  Single O2 flowmeter   March 2002 onwards 100 200 ml min    69    SERVICE SCHEDULE       Check flow tubes    Check Bobbin Floats move freely and rotate within the glass tube    Check maximum flow capability of all gases    If CO2 is fitted check that the maximum flow is restricted to 500 600 ml min flow rate   Connect O2 Monitor to CGO     Check no other gas flows are indicated   Ensure N20 
104. ies with all Pipelines    disconnected from supply   Perform this test with Gas Delivery Switch turned ON and Oxygen Basal Flow  blanked off     4  For Air   N20 repeat the test with the selector switch in the required position   5  On completion of tests reinstate basal flow by removing test blanking plug   6  Check output pressure of Primary Regulators   7  Check output pressure of Secondary Regulators   Flowmeter and Mechanical AHD Checks   1  Turn off all cylinders disconnect pipeline hoses and drain all gases from the machine   2  Replace flowmeter tube seals and filters   3  Check Flow Control Valves for smooth operation and positive off position   4 Turn On Gas Delivery Switch  Check Basal Flow of Oxygen is delivered with all Flow Control Valves closed down   All models   up to March 2002 100 200 ml min  Cascade O2 flowmeter  March 2002 onwards 50 75 ml min  Single O2 flowmeter   March 2002 onwards 100 200 ml min    Check flow tubes    Check Bobbin Floats move freely and rotate within the glass tube    Check maximum flow capability of all gases    If CO2 is fitted check that the maximum flow is restricted to 500 600 ml min flow rate   Connect O2 Monitor to CGO    0  Close all gas flow valves      OPAONOO    75    SERVICE SCHEDULE       11   12     13   14     15     16     17     Ensure N20   Air Selection switch is positioned for N20    Fully open N2O Flow Control Valve   Check that NO flow of N2O occurs    Slowly operate O2 control valve throughout range and check th
105. ig      Oxygen consumption of the whistle is  approximately 2 L min when sounding and  nil at other times     3 4 3 Fresh Gas Pressure Relief  Valve   A pressure relief valve is mounted between   the vaporizer back bar and the common gas   outlet  CGO  on the inside face of the   machine right hand upright     It is designed to prevent fresh gas being  delivered to the breathing system at  pressures exceeding 41 kPa  6 psi     This valve also protects machine  components against excessive pressure in  the event of a total blockage of the CGO     3 4 4 Air N2O Interlock  optional     The user can switch  between Air and Nitrous    Oxide  A      NOTE   a  The machine will NOT  deliver a mixture of Air  and nitrous oxide    b  On machines with  Mechanical AHD  the  O2 N20 linkage continues  to operate                 3 4 5 Oxygen Supply Visual  Indicator   The indicator  B  is operated from the   oxygen supply and shows GREEN when the   supply is at working pressure  and RED if   the pressure falls     3 4 6 CO2 Flow Restriction   The maximum flow of carbon dioxide is  restricted to 500 ml min by a pre set flow  control valve    This valve is not user adjustable    Note that carbon dioxide is not available for  US specification machines     3 4 7 Mechanical AHD    A mechanical link between the oxygen  control valve and a needle valve in the  nitrous oxide flow ensures that the machine  delivers a fresh gas mixture with a minimum  27  oxygen  irrespective of the flow of  nitrous 
106. inkage  Switch the Nitrous Oxide   Air control  B  to select Nitrous  Oxide    Fully open the Nitrous Oxide Flow control and check that  no flow is indicated on the flow tube     Slowly open the Oxygen Flowmeter control and check  that the Nitrous Oxide flow tube float rises    Pause briefly at an indicated Nitrous Oxide flow of 4  L min    Check that the Nitrous Oxide flow tube float is stable in its  setting and spins freely    Increase the Oxygen flow until maximum flow of Nitrous  Oxide has been achieved     Close the Oxygen Flowmeter control valve gently   Pause briefly when Nitrous Oxide flow indicates 500  ml min   Check the stability of the Nitrous Oxide float and that it  spins freely   Decrease the Oxygen flow until the mechanical stop  engages    54                   FUNCTION TEST    Check that as the Oxygen flow rate is  reduced  so is the flow of Nitrous Oxide   Check again that the only flow indicated  is the Oxygen Basal Flow        12  Close the Nitrous Oxide Flowmeter W2    control     13  Nitrous Oxide   Air Selection Switch                 machines with Air supply fitted   NU   Set the Nitrous Oxide   Air switch  B  to an   select Air  ntok  14  Open the Air Flowmeter control slowly ree    and see that a full flow can be obtained     15  Set an Air flow of 5 L min and check that  the Air flow tube float is rotating and  stable     16  Close the Air Flowmeter control gently  until no gas flows     17  Carbon Dioxide Supply   if fitted   Open the Carbon Dioxi
107. ire  frequent cleaning and disinfection  they are not a permanent part of the  anaesthetic machine and therefore  cannot be directly under the control  of the anaesthetic machine  manufacturer    However  we strongly recommend  that only breathing systems which  have been approved and authorised  by Penlon for use with the Prima SP  range should be employed    This is particularly important when  mechanical ventilation is employed     When mechanical ventilation is  employed the patient breathing  system must be connected directly to  an over pressure relief valve to  prevent the possibility of  barotrauma     WARNINGS AND CAUTIONS       10     11     12     14     15     16     Always perform a pre use check of  the machine  including vaporizers   ventilator  circle absorber and  monitors before clinical use  Follow  the pre use checklist  see section 5   as a minimum requirement    Many clinical accidents occur  because of a failure to check for  correct function     The machine must not be used if any  of the alarm  monitoring or protection  system devices are not functioning  correctly     The machine must not be fitted with  more than four operator accessible  mains socket outlets  There is a risk  of an excessive leakage current     The use of antistatic or electrically  conductive breathing hoses is not  recommended when using high  frequency electrical surgery  equipment  e g    Diathermy     Burns may be caused     Before any electrically powered  machine is used cli
108. izers with Interlock  WARNING  Only vaporizers with the Selectatec  compatible interlock function will interlock if  installed on a two station manifold   The installation of non interlock vaporizers  allows the possible operation of more than one  vaporizer at the same time     Check that the interlock mechanisms of all  the vaporizers on the manifold are working    44    correctly  i e  check that only one vaporizer  at a time can be turned on     5 10 3 Cagemount Vaporizer  Some international standards demand that  this type of vaporizer should only be used on  machines with a single mounting station    In addition  use safety clip  Cat  No  52275   to retain the cagemount taper cones in posi   tion on the vaporizer     5 10 4 Pre use Checks   On ALL vaporizers  before use    1  Check all joints for gas tightness    2  Check vaporizer agent level    3  Check for correct agent delivery  concentrations   use an agent  analyser    Always follow the pre use check procedures   given in the vaporizer instruction manual     5 11 Ventilator  if fitted     Always follow the pre use check procedures  given in the ventilator instruction manual   Check all hose and tubing connections for  gas tightness    Check all wiring connections for correct  fitment and security     AV900 V4 Ventilator  Check for correct fitment of the interface  cable at each connector  A                                 Check the interface system function   a  Turn the Gas Delivery Switch ON   The ventilator wil
109. k that no flow of nitrous oxide occurs    Slowly operate O2 control valve throughout range and check that N2O now flows   Operate N2O Air selector switch to Air position and check that N2O ceases to flow and  that Air is now available    Return switch to the N2O position and set the oxygen flow back to minimum    Refer to the table below  and check   Oxygen concentration at the oxygen flows  specified    Adjust the Oxygen Flowmeter control only   Do NOT adjust the N2O flowmeter     Note The flow rates for N2O are given as a guide only   O2 Flow  02  in O2 N2O mixture  N20 Flow   0 25 L min 27   33   0 5   0 67   0 5 L min 27   33   1 0   1 35  0 8 L min 27   33   1 6   2 1  1 5 L min 27   33   3 0   4 1  3 0 L min 27   33   6 1  8 1     Cascade flowmeter models   March 2002 onwards    Repeat tests by reducing Oxygen flow back to Basal Flow conditions  referring to the  values in the above table   Check gas delivery switch for correct operation    ssa Fail Alarms     2   3     4     B  1   2   3   4  5    D    Simulate an Oxygen fail by unplugging the Oxygen Pipeline from the outlet   Check that before the supply of oxygen is exhausted that the Oxygen Fail Whistle  sounds for a minimum of 7 seconds    Check that any Nitrous Oxide and Carbon Dioxide  if fitted  Gas Flows are cut off   Check that the Oxygen Supply Visual Indicator turns from fully green to fully red   Turn on Oxygen cylinder    Check Oxygen Basal flow reinstated and whistle sounds briefly    Turn off Oxygen cyli
110. l power up   While the Prima SP power is ON  the AV900  V4 can be turned off and on  using the  ventilator On Off switch   c  Turn the Gas Delivery Switch to OFF   The ventilator will power down     b     PRE USE CHECKS   Oxygen Monitor       5 12 Oxygen Monitor   A100SP  Absorber    The pre use checks for the oxygen monitor  include a set up and calibration procedure    NOTE   Prima SP with A100SP Absorber    Please refer to Appendix 4   Doc No A100SP  103DS U   supplied with the absorber      5 12 1 Sensor Location   Check that the sensor  1  is correctly fitted  as  illustrated  depending on the system installed on  your anaesthetic machine    1  Prima SP with A100SP Absorber  2  Mounted on the dome of the absorber inspira   tory valve  3     T    piece adaptor  2  on CGO block  a  using a 22 mm taper and Safelock nut  or  b  using a 22 mm male female tee fitting of  suitable design     Check that the cables and leads  3  are correctly  fitted at the rear of the monitor control unit    For AV900 V4 ventilator with built in oxygen  monitor  please refer to the ventilator user manual     5 12 2 System Set up  Switch on the oxygen monitor using the Gas  Delivery switch  4  on the front panel     NOTE   The Gas Delivery switch must be in the On position for  gas delivery from the anaesthetic machine  and to  supply electrical power to the oxygen monitor        Check that the display LEDs  5  and the alarm  indicators  6  are lit for two seconds  and that the  audible alarm sou
111. lass 1 and Type B are defined in IEC 601 1  the  standard for electrical medical equipment   e    This symbol denotes  Type B equipment    Type B equipment calls for a particular degree of protection against  electric shock     Class 1 equipment has additional protection such that metal parts of  the unit that are accessible to the user  e g  the metal casing of the  oxygen monitor unit  cannot become live in the event of failure of the  basic insulation of the electrical components within the unit     This symbol denotes  Refer to the User Manual       141    APPENDIX 3       NMOHS SV G3LSIML 20 2 SHV SSYIM TWNOIS TV         GSLVLS SSIMYSHLO SSJINN   SAdINLS MIWNYS TATIVeVd JYV SSYIM HLS TIV          SN1G  N  NMOYG   1    ZOZ SI ONIMIM SNIVW O3dS 9318AN         BLIHM   N    MOVIE   1  03AOYddY WSO Smeg  SI ONINIM SNIVW O3dS WSO       OHO ASSV ALVINdOddV 8 ZEE008 DUC 33S ONIMIM TANVd TWOIMLOSTA HOS      027008    ASSY 99d    NILH91 YSALAWMO 1S    EX ZELPOL    STWNIWYEL wwz    1x 9ELPOL  SNISNOH wwz        ONIIM    TANVd SSS9OV WOILYSA    LX LpZEOL  03X    G30NOYHS    3avds  94       wwg 9  30YdS  b t  yaa 0H3asn4    60201  bPZeOL  v66E0     3AVdS b t JOYAS  94        30vdS  94          1NO 30IS 3OVdS    b t       NI ONS    SNOILOANNOO  asna  SNIVW WOIdAL     L ILON ONIXIM 33S          06  008   WSO   P8BE008 AN o   WOO7 SNIYW    NILH  N Y3LIWMOT4       yayvays  199 9 HOLIMS  YIMOd  XNV                3          noan    QN3    pzS008   VSO     zs008  AN    ASSY GV31 SNIVW     
112. le gas supply options are shown     Note that carbon dioxide is not available for  US specification machines     Prima SP  non AHD     2004 UK specification four gas  machine with dual cascade  oxygen and nitrous oxide  flowmeters    Air   Nitrous Oxide interlock    Note  US specification machines  are not fitted with CO2 facility    NOTE   The non return  valve  adjacent to  the O2 reservoir  is fitted on  machines built  from July 2003             O O          L  J  idg            gt   gt            Air CO2      Assembly    Flowmeter    Gas Tray       Pneumatic pressure  source  Filter    Pressure gauge    Pressure regulator         Pressure relief valve    Audible alarm      K Cy    Restrictor    Gas cut off valve   normally open     Pneumatic switch   optional Air N2O  interlock     Reservoir    Pneumatic on off  switch    Flow control valve   variable     Flowmeter    Vaporizer    Oxygen flush valve    Non return valve    Power take off point   or test point     Visual indicator          Prima SP with  Mechanical AHD    2004 UK specification four gas  machine with dual cascade  oxygen and nitrous oxide  flowmeters    Air   Nitrous Oxide interlock    Note  US specification machines  are not fitted with CO2 facility               Flowmeter  Assembly                       gt     es 4  NOTE  The non return  valve  adjacent to  the O2 reservoir  is fitted on  machines built  from July 2003      O Q Q Q           l            i  a K H  A q q q q  J J J  bili ii H ii  e  2     
113. livered into the breathing system  showing  that no leaks are present     41    PRE USE CHECKS       5 7 Oxygen Flush    Check for a high flow of oxygen through the  CGO outlet when the flush valve button is  pressed and that the flow ceases when the  button is released    NOTE   If an optional CGO switch is fitted  see section  3 9   the lever must be set to the vertical position     This test is most conveniently done after the  breathing system has been attached  using  the reservoir bag as an indicator of gas flow     5 8 Anaesthetic Gas  Scavenge System   AGSS     By inspection  check that all sources of  expired anaesthetic gases  e g  the absorber  APL valve  and the ventilator bellows patient  gas exhaust port  are connected to an  approved collection system leading to an  AGSS     WARNING   Vacuum systems must not be connected  directly to the APL valve on the absorber  A  receiving system with a positive and negative  pressure control function must be interposed   Systems must comply with standard ISO 8835  part 2     42    PRE USE CHECKS       5 9 Alarm System Test    WARNING  The anaesthetic machine must not be used if any alarm is  not functioning correctly     Primary Oxygen Failure Alarm   The machine is fitted with a warning whistle and a visual  indicator  A     These components act as oxygen supply failure devices  and constitute the primary alarm system  powered only by  the residual oxygen supply  as described in section 3     The system can be checked whe
114. low rates for N2O are given as a guide only     O2 Flow  02  in O2 N20 mixture  N20 Flow     0 25 L min 27   33   0 5   0 67   0 5 L min 27   33   1 0   1 35  0 8 L min 27   33   1 6 2 1  1 5 L min 27   33   3 0   4 1  3 0 L min 27   33   6 1   8 1     Cascade flowmeter models   March 2002 onwards       If the  02 concentration is outside the range given  above  at any of the set flows  adjust the output of  either or both of the secondary regulators   see 9 8 2        NOTE Maintain output pressures within the range   152   241 kPa  22   35 psi     27  Check the Oxygen basal flow   see 9 8 1   28  Recheck the Oxygen Concentration   29  Remove the O2 control knob and the Setting Tool  9    30  Carefully refit the control knob cover  6  and screws  5    ensuring that the basal flow does not change   Fit the O2 and N20 control knobs        93    SERVICING PROCEDURES    9 8 4 Gear Linkage   Reset   Models built up to Feb 2002    CAUTION   Ensure that any residual gas in the machine is  exhausted before commencing the dismantling pro   cedure        Removal   1  Remove all cylinders  and disconnect  Pipeline supply hoses    2  Remove the two screws  1  and control  knob cover  2     3  Check the condition of the gears and  grubscrews  and replace if necessary    4  To reset the gears  first remove the O2    gear  4  only    5  Turn the Gas Delivery switch  5  to On   Gently close the NzO primary valve so  that the N20 flowmeter bobbin reseats   NOTE  a  This valve is closed by tur
115. ma SP with in board A100 Absorber and stand alone AV900 Ventilator   Ref Part No  Description Qty    1 58459 Hose   1 5m 1   2 011292 Hose   300 mm 1   3 011291 Hose   600 mm As required    PARTS LIST       Pressure Relief Valve  PRV     Ref Part No  Description Qty   1 700650 Outlet connector 1   2 0408 Dowty seal 1   3 045435 Pressure relief valve  PRV  1   4 800238 Manifold block   PRV 1   5 054525 Stem adaptor 1  Note    Coupling    A    connects the PRV to the vaporizer manifold block  see page 132 for part  number details    137    PARTS LIST                                  Miscellaneous Components    Ref Part No  Description Qty    1 000083 Castor   locked  front  2   2 000084 Castor  rear  2  01252 Star washer  M6 As required  800644 Tube hook As required  025206 Spring washer  dia 16 mm 4   3 011163 Handle 2  01028 Screw  M6 x 30 4  01015 Washer  M6 4  800192 T nut 4   4 800483 Bump stop As required  800548 Screw  No 12 x 20 As required    Interface and Spirometry Cables  5 300414 1 Interface Cable   AV900 V4 to Prima SP  560 mm       300415 1 Interface Cable Assembly   A100SP Absorber to Prima SP    consists of socket assembly and two cables    138    PARTS LIST    Tubing Connectors          054514 ao 054521   Swivel elbow Straight   6 mm tube connector   8 mm tube  oS 054513  Equal elbow  5 mm tube    2054531 BY  Equal tee 054534    Stem adaptor  4 mm tube    5 mm tube    Swivel elbow  4 mm tube  bo e   raig  daplon OE      054529  Stud fitting    male 4 mm  tube 
116. machine yoke  and cylinder faces are dust free and  clean and that the sealing washer  provided is in position between the  cylinder valve and the yoke    Tighten the yoke securely before  opening the cylinder valve  Dust and  dirt presents a fire hazard in the  presence of high pressure gas   Leakage of high pressure gas can  cause serious injury     5  Anaesthesia apparatus must be  connected to an anaesthetic gas  Scavenging system  AGSS  to  dispose of waste gas and prevent  possible health hazards to operating  room staff     This requirement must be observed  during test procedures as well as  during use with a patient     Prima SP machines must only be  used with Sigma Delta vaporizers  or  other vaporizers approved by Penlon  Limited  installed by means of the  Cagemount or Selectatec system   Free standing vaporizers may be  accidentally tipped  resulting in  excessive and uncalibrated volumes  of anaesthetic drug entering the  breathing system    Do not install or connect any  vaporizers of any description  between the Common Gas Outlet   CGO  and the breathing system  unless they are specifically designed  for such use   If this is done  the  oxygen flush flow will pass through  the vaporizer and may result in gross  overdosage when the flush valve is  operated      The breathing system which  conveys gases from the anaesthetic  machine to the patient and disposes  of expired gases is a vital part of the  anaesthetic delivery system   Because breathing systems requ
117. nder and repeat above tests     ack Bar Assembly   Selectatec   Checks     Replace Selectatec O rings  DO NOT apply grease or lubricant  Replace flexible tubing from back bar outlet to CGO   1 35 m  Replace vaporizer locking spring   Check all pillar valves open and seal correctly    Check pillar valve inserts sit proud or flush with the pillar valve   A sunken pillar valve insert denotes failure of the internal spring   Check vaporizer interlock system for correct operation     Back Bar Assembly   Cagemount   Checks     1   2   3     Replace flexible hoses    Replace flexible hose from back bar to CGO   1 35 m   Check security of taper connectors   Use Fomblin or other Oxygen safe grease to lightly lubricate the metal surface where  male and female connector come into contact with each other     73    SERVICE SCHEDULE       Fresh Gas Pressure Relief Valve  PRV  Test   i Occlude CGO with a manometer   2  At 10 L min Flowrate of Oxygen  the PRV relieves at 41 kPa   10      Leak Test From Flowmeter to Common Gas Outlet   1  Fit a suitable Manometer Test gauge to the CGO    2  Set Oxygen Flow to maintain a pressure of 20 kPa  3 psi    Maximum permissible flow is 200 ml min     3  Perform test with and without Vaporizer s  fitted and with vaporizer s  turned both ON  and OFF   4 Set oxygen flow to minimum and remove test equipment     Common Gas Outlet Checks  CGO     1  Check security and freedom of movement of CGO    2  Replace Safelock O ring  DO NOT apply grease or lubricant 
118. nds     5 12 3 Calibration                                              A new unit must be calibrated before clinical use    Thereafter  we recommend calibration every                 time the system is switched on  as a safety   precaution    Calibration must also be performed  5 LOWALARN SET   A  when the sensor is replaced  or Q Penton   B  when point of use elevation changes by more ae   than 160 m  500 ft             gran  2 Monitor   We recommend calibration with a 100    oxygen standard source  at a pressure and O2 Monitor    flow similar to your application   Calibration at lower concentrations or with room  air is possible  but less desirable    45    PRE USE CHECKS   Oxygen Monitor       ALARM MUTE    HIGH ALARM SET pean      e J       SET  LOW ALARM SET        b IO    SET CAL          5 12 4 Calibration Procedure    Using 100  Oxygen    1  Switch on the oxygen monitor using  the Gas Delivery switch on the  machine front panel    The LEDs  1  2  3  and alarm visual  indicators  4 and 5  will illuminate  and  the audible alarm will sound     2  Ensure the oxygen sensor is mounted  correctly at the CGO hose connector    see section 5 12 1   All vaporizers must be OFF     3  Flush 100  oxygen through the CGO  and maintain the flow through the  total breathing circuit for  approximately 20 seconds     4  Allow the oxygen reading to stabilise   This will take at least 30 seconds     5  Press the CAL button  6  to enable  sensor calibration mode   The Oz CONC LED  1  will
119. nections  Safelock fitting    Weight  approximate  depending on specification      SP 101  SP 102    70 kg  154 Ib   75 kg  165  b     24    SPECIFICATION    4 2 Gas Supplies    Cylinders     SP 101  SP 102    Special order options     Pipeline   SP 101 102    Medical gas colour codes   Oxygen  Nitrous oxide  Medical air  Carbon dioxide    Helium  Xenon    Oxygen  nitrous oxide  air  and carbon dioxide  cylinder fittings can be specified to the maximum  numbers given below    All cylinder yokes are pin indexed    Maximum of three  Maximum of four    Helium  Xenon    Maximum of three  oxygen  nitrous oxide  air    All to relevant national standards     White    or Green   Blue   Black White  or Yellow  Grey   Brown   Green  bright        To comply with relevant national standards     Internal pipework is diameter indexed for each gas     Oxygen  8mm  Nitrous oxide 6mm  Air 5mm  Mixed gas 10 mm    4 3 Flowmeters    Flow ranges     Single flow tubes  Oxygen   Nitrous Oxide   Air  Carbon Dioxide  Helium    Cascade flow tubes  Oxygen  Air  Nitrous Oxide    Flowmeter Accuracy    Carbon Dioxide  4mm  Helium 4mm  Xenon 4mm   0 10 L min   0 10 L min   0 10 L min    0 500 ml min  flow limited to 500 ml min   0   10 L min     1  0  1000 ml min   2  0 10 L min    The accuracy of the flowmeter tubes is   2 5  of full scale reading     25    SPECIFICATION    Flowmeter construction and dimensions  Tubes and floats are matched  and must not be interchanged   Flowmeter tubes have antistatic c
120. never the low pressure  oxygen system is first pressurised by turning on a cylinder  or connecting a pipeline    a  The whistle will sound briefly as pressure increases   and    b  The visual indicator will turn from red to green     Whistle  Visual Indicator  and Gas Cut off   Device Test   A formal test  including the action of the internal gas cut    off device  is performed as follows    1  Connect both oxygen and nitrous oxide supplies    2  Set the Gas Delivery switch  B  to ON  and check  that the warning whistle sounds briefly    3  On machines with Air  and an optional Air N2O  interlock system  set the switch  C  to N2O     4  Set a flow of 2 L min on both flowmeters    include CO    if this gas is provided    5  Disconnect the oxygen supply at the wall socket or    close the oxygen cylinder valve and check    a  that as the oxygen flow slows down  the whistle   starts to sound and continues for at least 7 seconds    b  that the flow of nitrous oxide  and carbon dioxide   if fitted  is cut off completely before the oxygen   flowmeter shows zero flow    c  that the visual indicator  A  turns red before the   oxygen flow is entirely stopped    NOTE All gases must be included in the pre use check   6  Reinstate the oxygen supply    Check that all gas flows are reinstated  and that the   visual indicator turns green again     Oxygen Monitor  if fitted   See section 5 12     43       Req bay  Midic ates      tape   Ee                PRE USE CHECKS       5 10 Vaporizers  
121. ng the dismantling  procedure     1     Remove all cylinders and disconnect  all pipeline supplies     2  Remove the front cover  see 9 6     3  Remove the cover at the back of the  machine   4  Remove the screws  1  to release the  switch mounting bracket  2     5  Disconnect the tubing from the rear of  the switch assembly  3     6  Slacken the screws  4     7  Twist the front of the switch  as illus   trated  to unlock the two halves of the  switch assembly    Refitting   8  Reverse the removal procedure     Fit the front cover   see section 9 6   Carry out a Function test  section 6     101                                              SERVICING PROCEDURES    9 15 Air N2O Interlock Switch    CAUTION  Ensure that any residual gas in the machine is exhausted before  commencing the dismantling procedure     1     Remove all cylinders and disconnect all pipeline  supplies     2  Remove the front cover  see 9 6     3  Disconnect the tubing from the switch assembly  1     4  Remove the cover at the back of the machine   5  Remove the screws  2  to release the switch  mounting bracket  8     6  Turn the lever  4  clockwise to unlock the two  halves of the switch assembly    7  Unscrew the locking ring  5   and detach the toggle  switch  6     Refitting   8  Reverse the removal procedure        Align the toggle switch  6  as illustrated   Fit the front cover   see section 9 6     Carry out a Function Test  section 6           Align  as illustrated    102                           
122. nically for the  first time  check that the hospital  engineering department has carried  out an earth continuity test     Before using any additional electrical  equipment powered by the auxiliary  sockets on the machine  check that  the additional equipment is correctly  wired and is earthed through its plug   A missing or defective protective  earth conductor may increase earth  leakage currents to the patient to  values exceeding the allowable  limits  resulting in ventricular  fibrillation  or interference with the  pumping action of the heart     Additional equipment placed on the  top shelf must be securely attached    Take care when moving a fully loaded  machine  particularly when  negotiating ramps    Check that hoses or power leads are  not trailing on the floor     CAUTIONS    Flowmeter needle valves are designed  to seal with light torque and may be  damaged if tightened excessively    Take particular care with the carbon  dioxide flowmeter control  if fitted   do  not force the control knob past either the  fully open or fully closed positions     Open cylinder valves slowly to avoid  damage to pressure reducing valves   Ensure that cylinder valves are at least  one full turn open when in use     Under no circumstances should  anaesthetic agents be used for cleaning  purposes     After use  always disconnect the  machine from the piped gas supply  and or close the gas cylinder valves     Mechanical AHD system   The oxygen  flow control is restricted to prevent t
123. ning the   needle anti clockwise   b  On machines with Air  set the  Air N2O interlock switch  6  to N20                     6  Set the oxygen control valve so that a  basal flow of 150 ml min of oxygen is  supplied    T  Remove the N20 gear  7     Carefully replace the O2 gear so that the  gear stop  8  is positioned anti clock   wise  and just touching the mechanical  stop  9  in the front of the manifold block   see illustration     8  Distance the inner face of the O2 gear  from the mechanical stop with a 0 1 mm  feeler gauge    Tighten the grubscrews  10  to secure  the O2 gear    9  With the O2 gear against the stop   replace the N2O gear and push on so  that the front faces of the gears are flush   see X X as illustrated    Note that the rear face of the N2O gear is  closer to the manifold block than the O2  gear  m                                           7  Tighten the grubscrews  11  to secure 11    the N20 gear   Apply a small amount of Molycote BG87 a D   grease  evenly distributed  to the O2 and    N20 gear teeth  X    04      SERVICING PROCEDURES    Gas Flow    10     11     12     13     14     15     Check the gas flows  See Function Test   section 6 7    If the gas flows are inaccurate  small  adjustments can be made by adjusting the  oxygen basal flow within the range 150    200 ml  and then using the secondary reg   ulators  as follows    Attach a pressure gauge to the test point  on each regulator  12   oxygen  13    nitrous oxide     At 5 L min flow both
124. nt cover   see section  9 6   Retighten the screws securing the gas  blocks   Reconnect the cylinder and pipeline  supplies     4  Attach a pressure gauge to the oxygen  regulator test point  1     Early type regulators shown in  illustration     5  Turn the Gas Delivery Switch to ON     6  Set a flow of 5 L min Oxygen   Reset the Oxygen secondary regulator  to the figure given in the table     7  To increase output  turn the regulator  cap   2  slowly clockwise   To decrease output  turn the regulator  cap fully anti clockwise  then clockwise  to the correct setting    8  Return the flow to minimum     9  Attach a pressure gauge to the nitrous  oxide regulator test point  3      10  Seta flow of 5 L min nitrous oxide   Reset the N2O secondary regulator to  the figure given in the table    To increase output  turn the regulator  cap   4  slowly clockwise    To decrease output  turn the regulator  cap fully anti clockwise  then clockwise  to the correct setting     Return the flow to minimum   11  Turn the Gas Delivery Switch to OFF     12  Check the oxygen basal flow   see    section 9 8 1     13  Check   oxygen concentration    see    section 9 8 2                                                                             4    4 MM ill  P  0  B Se  OSE   Oa 2 A GF e                                           arane   3                                                            Note   Early type regulators are  shown in the illustration                         Secondary regula
125. ntilator and  breathing circuit will then deliver this fresh  gas mixture to the patient    Use the Prima SP in conjunction with  anaesthetic vaporizers  breathing hoses and  patient connection fittings which comply with  the relevant ISO standard or equivalent     Depending upon the patient circuit selected   the machines can be used in open  semi   open  semi closed or closed circuit  configurations     The range has been designed to give a wide  choice of configurations and accessories  as  follows     Machine size    Prima SP101 Induction machine with a  single vaporizer on the backbar manifold   and up to three gas cylinders     Prima SP102 Medium width machine with  capacity for two vaporizers on the backbar  manifold  and up to four gas cylinders     Gas supplies  SP101 Up to three gases    SP102 Up to four gases   Oxygen  nitrous oxide  carbon  dioxide  and air  other options  available   with pin index cylinder  yokes  and provision for up to  three pipeline supply inlets     Vaporizer mounting systems    Backbar manifold for Selectatec Compatible   or Cagemount type vaporizers     Anti hypoxic Device  AHD     Machines can be specified with a  mechanical AHD system  designed to  minimise the risk of a hypoxic mixture  reaching the patient  see section 3     Oxygen monitor  optional     The Oxygen Monitor is intended to  continuously measure and display the  concentration of oxygen in breathing gas  mixtures used in anaesthesia  and is  intended for adult  paediatric 
126. o not use excess force     Disconnecting tubing from connector    Push in the tube and the connector  end piece    A        Hold the end piece    A    in place     Pull the tube out to disconnect     Fitting tubing to connector    Push in the tube as far as possible  Do not use excessive force     The connector end piece    A    will also move  inwards     Pull the tube carefully outwards   The end piece    A    will be pulled outwards to  the    locked    position                             86    SERVICING PROCEDURES    9 5 3 Cascade Flowmeter  Assembly Tubing    The illustration shows the rear of a three gas  flowmeter assembly  with cascade oxygen  and nitrous oxide flowmeter assemblies  as  fitted to a UK specification machine with  mechanical AHD                 1  Oxygen supply    2  Oxygen   low flow tube to high flow  tube    3  Nitrous oxide   supply to primary  needle valve    4  Nitrous oxide   link to secondary  needle valve  mechanical AHD        5  Nitrous oxide   low flow tube to high  flow tube  6  Air   supply to flowmeter needle valve       7  Mixed gas supply to backbar  manifold     87          SERVICING PROCEDURES    9 6 Front Cover    Remove and Refit       CAUTION   Ensure that any residual gas in the machine is exhausted   before commencing the dismantling procedure    Removal   1  Remove all cylinders  and disconnect Pipeline  supply hoses    2  At the rear of the machine  slacken the screws   1  securing the gas blocks    3  Remove the two top fixin
127. oatings     Tubes are length indexed     Oxygen 260 mm  10 24 inch    Nitrous oxide 250 mm  9 84 inch    Other gases 240 mm  9 45 inch   see 3 12   Scale length 152 mm  6 in  minimum     all flow tubes except carbon dioxide     4 4 Gas Pressures    USA  UK  Canada Japan   Pipeline supplies  340 kPa 400 kPa    50 psig   58 psig   Cylinder supplies   Reduced pressure from regulator 310 kPa  15 kPa  35kPa 380 kPa  15 kPa  35 kPa   at 5 L min flow   45 psig  2 psig    5 psig   55 psig  2 psig  5 psig   Regulator diaphragm bursting pressure 2800 kPa 2800 kPa    406 psig   406 psig   Reduced pressure system safety valve 600 kPa 600 kPa    87 psig   87 psig     Reduced pressure from secondary regulators  at 5 L min flow     Oxygen and Nitrous Oxide 152   241 kPa  22   35 psi   Air 207   283 kPa  30   41 psi   Machines built before March 2002    All gases 207   283 kPa  30   41 psi   Safety valve 41 kPa  10   6 psi  10       to protect flowmeter  vaporizer etc      4 5 Auxiliary Gas Outlets    Pipeline supply  Gas is supplied at pipeline supply pressure  see above   Cylinder supply  Gas is supplied at reduced pressure from cylinder regulator  see above   Oxygen    Two self sealing connections on side frame upright   Total flow rate  not less than 100 L min to free air  80 L min against 243 kPa  36 psig  resistance  70 L min against 270 kPa  40 psig  resistance  50 L min against 297 kPa  44 psig  resistance   Air  on machines with Air supply option    One self sealing connection on si
128. omatically  when the ventilator mains lead is connected  to a    live    mains supply    The OFF indicator on the ventilator front  control panel will show a yellow light during  charging     NOTE   The stated battery back up period will only be  available if the battery is kept fully charged    If the battery has been allowed to discharge  below the LOW BATTERY condition  the  ventilator will not function correctly until the  voltage raises above the LOW BATTERY level   A fourteen hours recharge will be necessary to  bring the battery to full charge     3 10 5 Monitor and other    Accessories  if fitted   The mains lead  or adaptor  for a monitor  system or other accessories requiring an  electrical supply can be plugged into one of  the auxiliary sockets on the rear of the  machine   Penlon Oxygen Monitor   see section 3 13 7     DESCRIPTION       3 11 Third Fourth Gas  Options    3 11 1 Air   When air is specified as the third fourth gas   the machine specification is modified as  stated in 4 10 1    NOTE   Machines with an Air supply option can be  fitted with an optional Air N2O Interlock  switch  A    The machine will NOT deliver a mixture of  Air and nitrous oxide   see section 3 4 4   This switch must be in the NzO position to  allow a flow of Nitrous Oxide     3 11 2 Carbon Dioxide   When carbon dioxide is specified as the  third fourth gas  the machine specification is  modified as stated in 4 10 2     Gas supply cut off devices operate on  carbon dioxide in all mac
129. on control  5  Alarm mute key 10  High alarm set control    3 13 3 Displays    Oxygen Percentage Readout  The display provides direct readout of oxygen concentrations in the range of 0 100    If the oxygen concentration exceeds 100   the display will flash     Low Alarm Set Readout   The indicated value represents the oxygen percentage at which the low alarm will be activated   The low alarm set value is limited within 18 99     To set the low oxygen concentration alarm  see section 5 12     High Alarm Set Readout   The indicated value represents the oxygen percentage at which the high alarm will be activated    The high alarm set value is limited within 19 105   Note that in certain conditions of excess pressure   the readout may show a value above 100      To set the high oxygen concentration alarm  see section 5 12     19    DESCRIPTION   O2 Monitor       3 13 4 Alarm Conditions    HIGH O2 ALARM  The high O2 alarm is triggered when the    oxygen concentration is 1  above the  setting    In this alarm condition  a red HIGH O2  ALARM LED will flash at a 0 5 second rate   accompanied by a high priority sound    To cancel this alarm  the high alarm setting  must be equal to  or above the oxygen  concentration     LOW O2 ALARM  The low alarm is triggered when the oxygen    concentration is 1  below the setting    In this alarm condition  a red LOW O02  ALARM LED will flash at a 0 5 second rate   accompanied by a high priority sound    To cancel this alarm  the low alarm setting
130. or             Sensor life   approximately 1 500 000 O2 percent hours at  200C   minimum one year in most normal applications      An external probe is supplied with a 2m  6ft   extendable cable and diverter fitting for a standard 15  mm Tee adaptor     The system has user adjustable high level and low   level alarms with visual and audible indication of  alarm conditions     Easy to read  seven segment LED display for high   set  low set  and oxygen concentration readings     The monitor is controlled by the machine gas system  master On Off switch  A      A back up battery provides a minimum of 60 minutes  operation in the event of mains failure    The battery is charged when the machine is  connected to the mains supply     3 13 2 System On Off Switch   The switch  A  controls gas delivery from the  anaesthetic machine  and electrical power to the  oxygen monitor           The switch must be in the On position to use the  oxygen monitor and anaesthetic machine     When switched to On  the monitor will always default    to previous settings  re    DESCRIPTION   O2 Monitor    ALARMS    HIGH O2 ALARM  LOW Oz ALARM  O2 MONITOR INOP  O2 SENSOR FAULT  02 SENSOR LOW  LOW BATTERY  MAINS FAILURE    GAS DELIVERY switch  must be  ON  to operate  O2 Monitor    O2 Monitor       Oxygen monitor control panel    1  O2 Concentration display 6  High set key   2  Low alarm set calibration display 7  Low set key   3  High alarm set 8  Calibrate key   4  Alarm LEDs 9  Low alarm set   Calibrati
131. other gauges indicate zero  pressure   Check that there is no flow of gas indicated on All models 100 200 ml min  the Oxygen flow tube   up to March 2002        Basal Flow    Cascade O2 flowmeter 50 75 ml min  7  Turn on the Gas Delivery Switch  E    March 2002 onwards   Check that the warning whistle sounds briefly   Check that a basal flow of oxygen is delivered   as indicated in the table     Single O2 flowmeter 100 200 ml min   March 2002 onwards        52    FUNCTION TEST    Leak Tests  For each Leak Test  allow a minimum 2  minute period     8     10     11     12   13   14     15     16     17     18   19     20   21     22     23   24     Turn OFF Gas Delivery Switch  E    Turn the Oxygen Cylinder OFF    Drain down the oxygen system and  blank off the output of the O2  secondary regulator  Turn on the O2  cylinder then turn off again    Observe the pressure gauge reading   a leak will be indicated by the needle  pointer falling towards zero   Maximum acceptable leak is 500 kPa   Drain down the oxygen system and  remove the blank  Reconnect the  output of the secondary regulator   Turn the Oxygen Cylinder ON    Turn the Gas Delivery Switch  E  ON   warning whistle will sound briefly    Check that all flow control valves are  in the OFF position and that the only  flow indicated is the Oxygen Basal  Flow    Switch the Nitrous Oxide   Air Control   F  to select Nitrous Oxide    Turn the Nitrous Oxide Cylinder ON   Check that the Nitrous Oxide pressure  gauge responds in 
132. oxide set by the anaesthetist    With the nitrous oxide control valve fully  open  the oxygen and nitrous oxide flows are  then both controlled by the oxygen control  valve    See section 3 5 for a full description    3 4 8 Low Pressure Gas Tubing  Diameter indexed tubing is used for the low  pressure gas system   see section 4     DESCRIPTION       3 5 Mechanical AHD   Anti Hypoxic Device     3 5 1 Introduction   The Mechanical AHD is housed within the  flowmeter module and comprises a gear  linkage between the oxygen control valve  and a needle valve in the nitrous oxide flow     The system is designed to control the  relative flow rates of oxygen and nitrous  oxide    A predetermined minimum oxygen  concentration of 30   3  in the oxygen    nitrous oxide mixture is maintained over the  flow range to ensure that a hypoxic mixture  is not supplied from the anaesthetic  machine     3 5 2 Gas Delivery Switch   The Gas Delivery Switch  A  operates on the  oxygen supply and must be in the    On     position for normal operation of the  anaesthetic machine     The switch consequently controls the supply  of all gases provided with a gas cut off  triggered by a predetermined pressure level  within the oxygen supply  see section 3 4 1      A whistle  oxygen failure warning whistle   will sound briefly whenever the gas delivery  switch is turned on or off  Note that the  whistle functions continuously if the oxygen  supply fails  see section 3 4 2      NOTE   The switch also control
133. oxygen basal flow is continuously supplied   Single Flow Tubes  100   200 ml min  Dual Cascade System Flow Tubes    50   75 ml min  This basal flow can only be turned on and off  by using the Gas Delivery Switch     CAUTION   The oxygen control is restricted to prevent the  needle valve from fully closing  This ensures a  minimum oxygen basal flow    DO NOT attempt to close the flow to zero  Do not  overtighten the knob     DESCRIPTION    3 6 Pressure Gauges    Pressure gauges  50 mm diameter  are  located on the front panel below the  flowmeter bank     The gauges for the third and fourth gases  if  fitted  are positioned between oxygen and  nitrous oxide  Unused gauge positions are  blanked out     All pressure gauges are colour coded and    labelled for the gases whose pressures they  are indicating     Cylinder gauges are marked  CYLINDER   Pipeline gauges are marked  PIPELINE     The gauges are calibrated in kPa x 100     DESCRIPTION       3 7 Flowmeters and    Controls  3 7 1 All models  The flowmeters  mounted behind the    perspex cover on the left hand side of the  machine  are length indexed to prevent  inadvertent  incorrect installation   All floats indicate flow rate in line with the  upper surface as shown below     Read flow at  this level       Each flow control valve is positioned directly  underneath the flow tube assembly to which  it corresponds  and the control knob is  colour coded for the gas which it controls     The oxygen flow control knob is made  p
134. piry  date is end of December 2004     vRO123K56789    MHJ FMAMJJAS ON K    Sensor Unit   Remove and Refit    1  Pull out the cable connector  A  from the expired  sensor  B     Remove sensor from the absorber or Tee adaptor   Discard the expired sensor and flow diverter  C    Insert the cable connector into the new sensor  B    Screw the new flow diverter  C  onto the new sensor   and fit new O rings    Fit the assembly into the Tee adaptor or absorber   Reconnect the cable   Calibrate the new sensor  see section 5 12    Dispose of the used components according to hospital  regulations and relevant national legislation     ar wo Np    O OND    109       Oxygen Sensor Location    1  Prima SP with A100SP Absorber       2  Mounted on the dome of the  absorber inspiratory valve                         Replacement parts  102714 Sensor  B    includes flow diverter  C  and O rings     58779 Tee adaptor  CGO fitting              Absorber Canister    SERVICING PROCEDURES       9 23 A100 Circle System Absorber    Follow the user maintenance procedures and service  recommendations detailed in the A100 circle absorber user  instruction manual     Inboard Mounting System on Prima SP  access to the  Canister and Valve Block                            Remove  1  Slide the canister and sup   port plate out from the  machine           Refit    1  Align the buttons  B  on the top of  the canister block with the slots  in the support plate     2  Slide the canister into the support  plate  and
135. r  repair of this machine  contact the nearest  accredited Penlon agent     or communicate directly with     Technical Support Department  Penlon Limited   Abingdon Science Park   Barton Lane   Abingdon   OX14 3PH   UK    Tel   44  0  1235 547076  Fax   44  0  1235 547062  E mail  technicalsupport penlon co uk    Always give as much of the following  information as possible     Type of equipment   Product name   Serial number   Approximate date of purchase  Apparent fault    OBO Ns    FOREWORD       This manual has been produced to provide  authorised personnel with information on the  function  routine performance  maintenance  checks and repair procedures applicable to  the Prima SP anaesthetic machine range     Information contained in this manual is  correct at the date of publication    The policy of Penlon Limited is one of  continued improvement to its products   Because of this policy  Penlon Limited  reserves the right to make any changes  which may affect instructions in this manual   without giving prior notice     Personnel must make themselves familiar  with the contents of this manual and the  machine   s function before using the  apparatus     Copyright    Penlon Limited  2004   All rights reserved      ii     CONTENTS       Page No    USER RESPONSIBILITY 01   1  WARNINGS AND CAUTIONS 02  2  PURPOSE 05  3  DESCRIPTION 06  3 1 Framework and General Construction 06  3 2 Gas Circuit Non AHD MACHINES 08  3 3 Gas Circuit Schematic 09  3 4 Gas Supply Safety Devices 
136. r valve    B  Turn off the reserve cylinders during normal  use    C  N20   cylinder pressure does not indicate  cylinder content     33    3                    Open the cylinder valve of the third  and fourth gas cylinders  if fitted    Check the reading on the pressure  gauge for each gas    Ensure that all flowmeters are kept  closed until gas supplies are required     PRE USE CHECKS    Non AHD Machines     5 2 3 Flowmeters    1  Operate each flowmeter control knob  in turn   2  Check that full scale of flow can be    achieved  and that the floats in all  tubes move freely and rotate when at  a steady flow    3  On machines with Air and the optional  Air N2O Interlock switch  B   use the  switch to select air and nitrous oxide  in turn   4  Check the supply of each gas     5  Check that the flow can be turned off  by gentle rotation of the control knob   and that the float reseats on the  bottom stop    6  On machines with optional dual  cascade flow tubes  check that gas  flow is through the low flow tube  initially until full flow is achieved  then  through the high flow tube                 5 2 4 Air N2O Interlock    NOTE  This is an additional option on machines  with Air supply    1  Switch the Gas Delivery Switch  A  to    ON    2  Set the Air N 20 Interlock switch  B  to  Air    3  Open the Air flowmeter control and  check that Air is delivered    4  Open the N20 flowmeter control     5  Check that N2O is NOT delivered    6 Set the Air N2O Interlock switch to  N20  
137. rate the oxygen flowmeter needle valve     Check that full scale of flow of oxygen and nitrous  oxide can be achieved  and that the floats in both  tubes move freely and rotate when at a steady flow   5  Check that the nitrous oxide flow can be turned off  by gentle rotation of the oxygen knob   Check also that the nitrous oxide float reseats on  the bottom stop  and that the oxygen basal flow  continues to flow   6  Operate the other control knobs in turn to check     the full scale of flow can be obtained    the floats move freely and rotate at a steady flow   the flow can be turned off by gentle rotation of the  knob  and that   the floats reseat on the bottom stop              7  On machines with Air  and the optional Air N2O  interlock system  set the switch  B  to Air  to check  the supply for this gas    8  On machines with optional dual cascade flow  tubes  check that gas flow is through the low flow  tube initially until full flow is achieved  then through  the high flow tube     5 3 4 Air N20 Interlock   machines with Air supply option     1  Switch the Gas Delivery Switch  A  to ON    Set the Air N 20 Interlock switch  B  to Air    Open the Air flowmeter control and check that Air is  delivered    Turn on the oxygen supply at the flowmeter    Open the N20 flowmeter control    Check that N2O is NOT delivered    Set the Air N 20 Interlock switch to N20    Check that the flow of Air has stopped    Check that N20 is now delivered     oN    CONDOS    Continue Pre use ch
138. rcuit breaker   rear panel      Check supply   plug socket ON     Charge  Replace battery  Fit Battery    Check connections    Press and hold both  ALARM MUTE and  LOW ALARM SET but   ton    All the LEDs should  illuminate for 2 sec   onds    The high alarm set and  low alarm set will  default to 100 and 18  respectively     Replace batteries     Replace control PCB or  alarm PCB     A  Check supply   plug socket  ON     B  Reset circuit  breaker  rear  panel      8  SERVICE SCHEDULE       8 1 Service Schedule    Recommended Service Frequency  Six Month Service  Annual Service  Two Year Service  Six Year Service    NOTE  Preventive Maintenance Kit Part Numbers are listed in section 10     8 2 Six Month Service Checks   To be performed at 6  18  30  42  54   amp  66 months of a 72 month cycle      Machine Frame     1  Check integrity of structure and security of all attachments   2  Check for damage    3  Check wheels for freedom of movement  check front braking   4  Check that drawers and writing tablet move freely     Electrical Safety Test     1  Check mains lead and plug for damage   2  Perform Electrical Safety Tests as relevant for country of use   3  Check output sockets for correct function     Pipeline Hose Checks     1  Examine pipeline connectors  probes and hoses for cleanliness and damage   2  Perform anti confusion checks   3  Perform pressure Leak Checks of all hoses  with cylinders removed and Gas    Delivery Switch in the OFF position     Cylinder Yoke Assembl
139. reducing valve and pressure gauges  Ensure that  valves are at least one full turn open when in use     1  Fit the gas cylinders to their respective yokes  open  the cylinder valves one at a time and check the  pressure on each gauge    NOTE   A  When two cylinders are provided for a single gas  test each   separately  clearing pressure after each test by opening the   flowmeter valve    B  Turn off the reserve cylinders during normal use    C  N20   cylinder pressure does not indicate cylinder content     2  Check the third and fourth gas cylinders  if fitted    open the cylinder valve and check the contents on  the pressure gauge    3  Ensure that all flowmeters are kept closed until gas  supplies are required     35    CAUTION   On machines fitted with a mechanical  AHD  the oxygen flowmeter control is  restricted to prevent the needle valve  from fully closing    This ensures a minimum oxygen basal  flow    DO NOT attempt to close the flow to  zero    Do not overtighten the knob                    PRE USE CHECKS   Machines with Mechanical AHD    5 3 3 Flowmeter   Machines with  Mechanical AHD    1  Turn on the Gas Delivery switch  A  and check that  the warning whistle sounds briefly  and that the  correct basal flow of oxygen is delivered  see  section 4 8     2  On machines with Air  and the optional Air N2O  interlock system  set the switch  B  to N2O  see  section 5 3 4      3  Open the nitrous oxide needle valve and check that  there is no nitrous oxide flow   4  Ope
140. ry 6 months   Function test and calibration test   Every 10 years   Major overhaul  if unit to be used beyond  10 years     Note  Halothane vaporizers   major  overhaul required at 5 years    On vaporizers with interlock  test the  function of the interlock system during the  vaporizer calibration test     9 2 2 Monitor    Follow the service recommendations  detailed in the monitor user instruction  manual     9 2 3 AV800   AV900 Ventilator    If the anaesthetic machine was supplied with  an AV series ventilator  follow the  instructions given in section 7 of the user  instruction manual    These include     Every 6 months  Inspection and Function Check    Every 12 months  or 1200 hours   Replace bellows    Every 12 months  Replace O seals and oxygen inlet filter   etc   as preventive maintenance     Every 5 years  Replace batteries  Fit service exchange pneumatic assembly    9 2 4 Anaesthetic Gas  Scavenging System    Follow the service recommendations  detailed in the AGSS user instruction  manual     9 2 5 Circle System Absorber  Follow the service recommendations  detailed in the circle absorber user  instruction manual     SERVICING PROCEDURES       9 3 Cleaning and Sterilisation    Frequently wipe over the external surfaces  of the machine  the vaporizers  with a damp  cloth  Always allow the machine to dry off  thoroughly before using it      Mild antiseptic solutions may be used but  should be rinsed off with clean water      Breathing system hoses and other compo   n
141. s fully exhausted  that the  visual indicator  C  has turned fully  red     Turn OFF all Flowmeters and turn ON  the Oxygen cylinder so that only the  Oxygen Basal Flow is indicated     Check that the visual indicator  C  is  showing fully green     FUNCTION TEST    6 8 Vaporizers and Back Bar Manifold  Assembly    6 8 1 Selectatec Compatible Vaporizer    1  Check that all the vaporizers are securely  mounted on the manifold   If a vaporizer is not secure  check the condition of  the locking shaft  A      2  On interlock vaporizers check that the interlock  mechanism of each vaporizer is working correctly   i e only one vaporizer at a time can be turned on     3  Selectatec Manifold   check all pillar valves open  and seal correctly  Check pillar valve inserts sit  proud or flush with the pillar valve    A sunken pillar valve insert denotes failure of the  internal spring     6 8 2 Cagemount Vaporizer   Vaporizers fitted with the Cagemount tapers have the  male taper  Inlet Port   C  on the left and the female  taper on the right as viewed from the front of the  vaporizer    Two M6 studs with nuts  washers and a clamp plate  D   are provided to fix the vaporizer to the anaesthetic  machine     NOTE   Some international standards demand that this type of  vaporizer should only be used on machines with a single  mounting station     In addition  use a safety clip  Catalogue No 52275  to  retain the Cagemount taper cones in position on the  vaporizer     Fitting the Vaporizer   
142. s the electrical supply to  the optional flowmeter lighting unit  oxygen  monitor and integrated AV900 V4 ventilator     3 5 3 Gear Linkage and Nitrous  Oxide Control Valves   A gear linkage connects the oxygen control   knob on the flowmeter module and a needle   valve in the nitrous oxide flow  This linkage   limits the flow of nitrous oxide relative to the   flow of oxygen set by the user    Note that this needle valve acts as the   primary nitrous oxide valve  and is actuated   only by movement of the oxygen control    NOTE   Machines with Air option have the additional    11                opion of an Air N2O Interlock switch  B    he machine will NOT deliver a mixture of Air and    nitrous oxide   see section 3 4 4     This switch must be in the N2O position to allow a  flow of Nitrous Oxide     The nitrous oxide control knob on the  flowmeter module operates a secondary  needle valve in the nitrous oxide flow  It is  positioned downstream of the primary valve  and therefore is used only to restrict the flow  already set by the primary valve  which itself  has been determined by the position of the  oxygen control knob     Therefore  for any oxygen flow set by the  user  the mixture delivered will still contain a  minimum 30   3  oxygen even with the  nitrous oxide control knob fully open    As the nitrous oxide knob is progressively  closed  the oxygen content of the mixture  increases to 100      3 5 4 Oxygen Basal Flow  To allow the system to function correctly  an  
143. ter is S amp S SUSY    rrr AN  always exhausted first  so the EDN  USRS    canisters should be rotated as follows               a  liftthe top canister from the frame and          replace the absorbent  Place to one a   side  for refitment in the lower  position     a     b  the lower canister should be refitted in  the upper position   see below     Refitting the canisters    CAUTION  Efficient use of soda lime is only achieved if the  newly refilled canister is replaced in the lower    position    1  Refit the canisters into the frame  with  the newly filled canister in the lower  position    Check that the seals and canisters  align    2  Ensure that the top canister  3   engages correctly into the top seal  4   as you rotate the lever  2  clockwise to  the vertical  locked on position      3  Leak test the absorber     see section  6 2 in the A100 User Manual     151    Sterilisation    Caution  Do not autoclave     Oxygen Sensor Manometer  Top cover Bottom cover    Absorber Assembly   removal before  Sterilisation    WARNING  The absorber assembly weighs approximately 17 kg   Take care when the absorber is removed from the machine     Remove    1     Disconnect the cable  1  from the O2 sensor  2    Unscrew the sensor from the manifold block    Do not autoclave    At the rear of the absorber manifold block   disconnect the hoses  apart from the fresh gas hose    3  and cables    Disconnect the fresh gas hose from the CGO block  on the anaesthetic machine    Drain the con
144. the  02 concentration is outside the range given  NOTE     above  at any of the set flows  adjust the output of  either or both of the secondary regulators  oxygen   2    nitrous oxide   3   as follows     Early type regulators  shown in illustration    5  Switch the Gas Delivery switch to OFF   6  Remove the front cover   follow the procedure given in  section 9 6     Retighten the screws  5  securing the gas blocks   Reconnect the cylinder and pipeline supplies        7  Attach a pressure gauge to the regulator test point  4     8  Switch the Gas Delivery switch to ON    9  To increase output  turn the regulator cap   2 or 3  slowly  clockwise     To decrease output  turn the regulator cap fully anti clock   wise  then clockwise to the correct setting   Maintain the output pressure within the ranges     All models   up to March 2002  Secondary regulator pressure 207   283 kPa  30   41 psi     All models   March 2002 onwards  Secondary regulator pressure 152   241 kPa  22   35 psi     10  If the correct basal flow can not be achieved  reset the gear  linkage   see section 9 8 3     11  Recheck the Oxygen Concentration     91          SERVICING PROCEDURES    9 8 3 Gear Linkage   Reset    Cascade and Single Flowmeter Tube Systems  March 2002 Onwards    NOTE  The machine must be on a flat surface                                                                                                                                                     Set the secondary regulators 3  1  S
145. tilator   this system is not MRI compatible     Nuffield 200 Ventilator  Delta Vaporizer    IDP Pressure Failure Alarm   when secured to the machine  and used with appropriate batteries     The following components are currently not MRI compatible   Oxygen Monitor  Flowmeter lighting  Electrical power outlets  AV series ventilator    4 10 Third and Fourth Gas Options    4 10 1 Air   Air flowmeter range  1   10 L min   Cylinder yoke pin indexed for medical air   Pipeline inlet for air    Cylinder pressure gauge    Pipeline pressure gauge     Air pipework is colour coded at each junction     4 10 2 Carbon Dioxide   Carbon dioxide flowmeter range 20   500 ml min  flow restricted to 500 ml min    Cylinder yoke pin indexed for carbon dioxide    Cylinder pressure gauge    Carbon dioxide pipework is colour coded at each junction     4 11 Electrical Supply  if fitted   Standard  5 5 amp  220 240 V  50 Hz  Optional  6 amp  110 120 V  60 Hz    Permanently attached 3 metre lead   Stowage hooks for cable on rear   Auxiliary electrical power outlets  if fitted    SP102 4 outlets  SP101 3 outlets  Max  total current 5A    Battery back up  oxygen monitor    12 V  1 8 1 9 9 AH  rechargeable sealed lead acid battery  provides a minimum of 60 minutes  operation     28    SPECIFICATION   O2 Monitor    4 12 Oxygen Monitor    Measurement Range   Resolution    Accuracy and Linearity   Response Time     Operating Temperature   Storage Temperature   Transport Temperature     Relative Humidity Range 
146. tion of the breathing system                       A high mounting position for the CGO is available  as an option for all machines        Oxygen Flush   An emergency oxygen flush valve button  3  is  mounted on the top front of the CGO block and is  marked    Oz FLUSH       Depressing the button provides a delivery of  between 35 75 litres min of oxygen into the  common gas outlet  2     Releasing the button allows the spring loaded valve  to return to its normal position        Optional high mounting position    Optional two position Switch on CGO    An optional switch on the CGO outlet enables the A   user to choose to divert the fresh gas flow    a  directly to the absorber  or   b  through the CGO outlet to an open patient  circuit     Switch in vertical position  A    The fresh gas flow is directed to an open patient  circuit via the outlet  B  on the front of the common  gas outlet        Switch in horizontal position  C    The fresh gas flow is directed to the absorber via  tubing connected to the rear outlet of the common  gas outlet block     Optional CGO switch    DESCRIPTION       3 10 Electrical Power  Supply  if specified     3 10 1 Mains Power Supply    Power is fed to the machine via the mains  lead  to power an auxiliary output panel  and  optional flowmeter bank light    NOTE   a  It is the user   s responsibility to ensure that the  total sum of leakage currents from additional  equipment plugged into the auxiliary sockets plus  the leakage current from the
147. tlets   1  Check outlets for security  damage and correct lock and release movement   2  Check dynamic flow rate and that Oxygen Fail Whistle does not operate     Oxygen Monitor  if fitted   Function and Calibration Tests   1  Connect a Test O2 Analyser into the patient circuit   2  Check that the machine O2 sensor is inserted into the absorber O2 sampling point   3  Test both O2 sensors in 100  Oxygen   4  Expose both sensors to air and check reading is 21    2   Adjust high and low O2 alarms and check alarms trigger when values are lower or  higher than reading on O2 analyser respectively   Return the alarm levels to original settings   5  Restore the sensor to correct location   Remove the test O2 analyser     Final Gas Concentration Tests   1  Check concentration by Gas   2  Check flowrate from CGO corresponds with flowmeter by concentration   3  Set 3 L min Oxygen and Nitrous Oxide flows   Check  using O2 Monitor that the concentration from the CGO is 50    2    4  Set Oxygen to 0 5 L min  set N2O to 0 5 L min   Allow O2 monitor to stabilise then check concentration is 50    2    If fitted repeat above test substituting Carbon Dioxide for N20   If fitted set Air to 3 L min  set Oxygen to 3 L min   Check Concentration reading on O2 Monitor is 60    2      5   6     Test Completion   Paperwork   Remove all test equipment from machine   Drain all Gas from machine    Turn OFF Gas Delivery Switch    Turn Off all cylinders    Turn off all Flow Control Valves    Attach Service 
148. to the seals   Check that the countersunk screw and  locating screw  3  are fitted in the correct  orientation  as illustrated   7 Carry out a machine function check   section  6     104    SERVICING PROCEDURES    9 18 Flowmeter Tubes   Removal    CAUTION Ensure that any residual gas in the  machine is exhausted before commencing the dis   mantling procedure                 Removal   1  Remove all cylinders and disconnect all  pipeline supplies    2  Remove the flowmeter cover   see section  9 7     3  Hold the tube  as illustrated  and press  upwards  X  against the resistance of the  spring    4  Carefully detach the lower end of the tube  from the bobbin    5  Remove the tube  ensuring that the top  bobbin remains in place  Y     Overhaul   Bobbin Assemblies          6  Remove and refit the components  as illus   trated below  taking care not to damage the  O seals when refitting the bobbins        Refitting the Flowmeter  7  Reverse the removal procedure  noting   a  the scale must be centralised   b  the tube is touching the contact strip  1    8  Carry out a machine Function Test   sec   tion 6     400262    0293 Bobbin   01057  O seal E Washer    0691    EF 36247  O seal A   Bobbin    36075  Washer                                        36247  Bobbin    0508 Filter    Cascade flowmeter system illustrated   On non cascade models  a filter  0508   is fitted to each flow tube        105    SERVICING PROCEDURES    9 19 Flowmeter Assembly    Removal    CAUTION  Ensure that 
149. tors  Pressure output at 5 L min flow    Machines built from March 2002 onwards  Oxygen and   Nitrous Oxide 152   241 kPa  22   35 psi   Air 207   283 kPa  80   41 psi     Machines built before March 2002  All gases 207   283 kPa  30   41 psi              SERVICING PROCEDURES    9 12 High Pressure  Cylinder     Gas Block  Remove and Refit    CAUTION  Ensure that any residual gas in the machine is  exhausted before commencing the dismantling  procedure     1     2     Remove all cylinders and disconnect  all pipeline supplies    At the rear of the machine  remove the  two screws  1  at the rear of the block  to be removed    Refer to 9 6 2 if necessary    Remove the screws  2  at the rear of  the adjacent gas block    At the front of the machine  carefully  pull the blocks  8 and 4  from the  machine    Separate the blocks   detach the block   4  from the tubing connector  5     To refit  reverse the above procedure  and then carry out a full function test   see section 6      Yoke Filter   Remove and Refit  CAUTION   Ensure that any residual gas in the machine is  exhausted before commencing the dismantling  procedure     1     2   3   4    Remove all cylinders and disconnect  all pipeline supplies    remove the Bodok seal  1     Unscrew the yoke insert  2     Use a suitable rod to carefully remove  the filter  3     To refit  reverse the above procedure   Fit a new Bodok seal    Function test the machine  see sec   tion 6         99    SERVICING PROCEDURES    9 13 Selectatec
150. up power  supply has been run down  the oxygen monitor will not  function until the battery is recharged    An eight hours recharge will be necessary to bring the  battery to full charge     22    DESCRIPTION       3 14 A100  A100SP  Absorber    A100 Absorber In board Mounting   The canister and valve block assemblies are  mounted under the work surface and can be  detached separately for cleaning  section 6      A100SP Absorber  Please refer Appendix 4     3 15 AV900 V4 Ventilator    Interface to Prima SP   The interface cable links the socket  A  at  the rear of the ventilator control unit to  socket  B  on the rear panel of the  anaesthetic machine     AV900 V4 ON OFF function   a  Turn the machine Gas Delivery Switch  C   ON    The ventilator will power up    b  While the Prima SP power is ON  the  Ventilator can be turned OFF and ON  using  the ventilator On Off switch    c  Turn the Gas Delivery Switch to OFF    The ventilator will power down     3 16 Prima SP System    MRI Compatibility  CAUTION Field strengths vary at individual MRI    facilities   for additional information  contact  Penlon Ltd     The following system components are MRI  compatible     1  Prima SP basic  machine     A  basic  machine includes any variant of    Back Bar Flowmeter Bank  Drawers Monitor Shelves  Additional Gases    2  A100 and A100SP Absorber   Non magnetic inspiratory expiratory valve  discs must be fitted   Part No  2930 623    CAUTION A100SP Absorber is normally part of  an Prima SP
151. using air   press the navigator wheel  to switch to 100   calibration using   oxygen      A message will flash on the screen   O2 AT 100      Press the button to confirm  NOTE  The message    OXYGEN SENSOR LOW OUTPUT  will appear on screen if the user attempts  to calibrate at 21  in 100  oxygen     Repeat operations 8 and 9 twice  more    Scroll to ESCAPE FROM MENUS  and press the wheel to exit     Turn off the flow of oxygen        Git    Ce                      On Off Switch    Alarm Mute    Menu Switch    Navigator Wheel  and  Press Button          O2 Monitor sub menu   calibration    O2 MONITOR   ON   gt  CALIBRATION  100   HIGH ALARM SET    LOW ALARM SET  ESCAPE FROM MENUS       149    AV900 Ventilator   Spirometry  Calibration    The Spirometry heads must be calibrated  with zero flow going through them     1     Turn the Prima SP off at the Gas Delivery  switch  1     This will stop all gas flows  including the  AHD basal flow     This will also turn the AV900 off     Turn the AV900 on at the ventilator  Do not  use the SP Gas Delivery switch      Remove the breathing circuit hoses from  the inspiratory and expiratory connectors   2  on the absorber     Disconnect the fresh gas hose from the  CGO block on the anaesthetic machine     Disconnect the hose from the APL valve  outlet  3  at the rear of the manifold block    a  Ensure that the ventilator bellows is  empty  or    b  Remove the bag  and set the Bag Vent  lever  4  to Bag position    Calibrate the spirometer v
152. ve   O ring   Cylinder Pipeline NRV  Hose Support   Seal Support   Split Ring    8 mm Hose   Oxygen   Internal   6 mm Hose   Nitrous Oxide    Internal   5 mm Hose   Air   Internal   4 mm Hose   Switch Feeds COz    Internal   Gas ID Label White   Oxygen  Gas ID Label Blue   Nitrous Oxide  Gas ID Label Black   Air   Gas ID Label Grey   Carbon       041115  O ring            econdary Regulators        aS                         PARTS LIST          her S    Sy    SS   lt i    N   D  an se  Kon ys  S     lt  gt    lt p    Backbar Outlet  to CGO  SS  th   lt a    462526 Hose       PARTS LIST             400262   0293 BODDI 01057   O seal i E Washer   rah 36247  pee Bobbin    36075  Washer                               0691  O seal    0293  O seal             36247  Bobbin       400224  Bobbin  0508 Filter Cascade flowmeter system illustrated   On non cascade models  a filter  0508  is  fitted to each flow tube        123    PARTS LIST       Flowmeter Assembly   Two gas Cascade  UK specification     Ref        OONO A OO N       oO    ab  fee es pee  a kW PP    Part No     14978  800341  800196  800231  800269  800230  015043  019025  01023  800225  38315  800408  015044  01031  800223  38317  800409  019011  0360  700362  38316  01067  800270  019123  800239  700366  800285  800234  054515  402736  054524  019067  800226  800222  019133  01056  462543 A R  054514  031048  01027  15252  15277  38349  019083  0770  054519  054518  462543  A R  462544  A R  15023    Description    Mech
153. ws  13      11 Undo the three retaining screws  then  remove the exhalation valve assembly  14      Cleaning procedures before sterilisation    Absorber     Clean the frame and inspiratory and expiratory  valve assemblies   refer to section 8 4 in the A100  User Manual    The absorber assembly can then be autoclaved as  a single unit     Ventilator Bellows Assembly   Refer to section 7 2 in the AV900 V4 user manual       Sterilisation and Autoclave    CAUTION  To prevent possible damage to components  peak sterilisation temperatures must  not exceed      54  C  130  F  for gas  ethylene oxide  or    134  C  275  F  for steam autoclave    Low temperature autoclave must not exceed 120  C   Following sterilisation with ethylene oxide  components must be quarantined in a  well ventilated area to allow dissipation of any residual gases     Follow the recommendations given by the steriliser manufacturer for aeration  periods required     Sterilisation and Autoclave Treatment Table    ITEM METHOD  A100SP Absorber assembly Gas  liquid  autoclave  Absorber canisters Liquid  low temperature autoclave    Ventilator bellows assembly components     Bellows Gas  liquid  autoclave  20 cycles max     Hoses Gas  liquid  autoclave   Bellows base Gas  liquid  autoclave    Bellows canister Liquid  autoclave   Exhalation valve assembly Gas  liquid  pasteurise  low temperature autoclave  NOTE    Examples of suitable liquid agents are  Nu Cidex  Sporicidin  and Sonacide     153    After Cleaning   
154. y  this function  testing should be repeated at regular  intervals     Full function testing should be carried out  at  six month intervals  by a Penlon trained  engineer or other persons certificated by  Penlon Limited    Procedures for machine servicing are  described in subsequent sections in this  Service Manual     Read the procedure carefully before carrying  out each test  noting references to machine  specifications and equipment    Perform these tests in sequence     Mechanical AHD   Function tests specific to non AHD  and  Mechanical AHD systems are detailed  separately  These sections have special  page headings    Equipment Required  No complex test equipment is required for  the series of tests described below     Leak tests   Apply a Leak Test Spray or a diluted solution  of soap in water to the suspect connector    not on high pressure cylinders  and check  for a stream of bubbles at the joint     Flow rates  Check with a respirometer or other suitable  flow measurement device     Electrical circuits  Check with a multi meter     Pressure  Check with a manometer     49    Safety Precautions    WARNING   Before commencing any test work ensure that  all vaporizers are in the  Off  position    Note that the OFF position may be indicated  on the vaporizer as a 0  zero   Refer to the  vaporizer instruction manual     WARNING   During any test procedure the release of  Nitrous Oxide or anaesthetic vapours into the  working environment must be controlled   Connect the
155. ygen failure alarm  does not function during this test   Repeat the test for both pipeline and  cylinder supplies     7  FAULT FINDING       7 1 Introduction    Provided the Prima SP Anaesthetic Machine is  regularly serviced and correctly used  it is unlike   ly that serious component faults will occur     After several years    usage  some items may  require replacement and regular performance  checks   see section 8     To trace possible faults  refer to the Fault  Diagnosis Table  which will indicate the required  action    In many cases  apparent machine faults may be  attributable to causes other than machine mal   functions  these causes are also listed     59    FAULT FINDING       7 2 Fault Diagnosis Table    Component    Cylinder pressure  reducing valves   gauges and yokes     Pressure relief valve  leaks or out of adjust   ment     Symptom    Leaking high pressure  connection from cylin   der to yoke     Diaphragm or valve  disc worn  damaged     Outlet pressure too  high      may occur with leak   ing pressure relief valve   above      Outlet pressure too low     Diaphragm or valve  disc worn damaged or  regulator out of adjust   ment     Pressure gauge reads  zero when cylinder  valve opened     60    Possible Fault    Bodok seal absent or  worn     Cylinder clamp screw  assembly damaged     Inadequate engage   ment of pin index sys   tem     Adjust replace regulator  as necessary     Faulty relief valve     Diaphragm worn  dam   aged or out of adjust   ment    
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Talkswitch TS-200  QCD Expander User Manual v1.0    avez-vous une bonne faculte d`adaptation ? comment proceder  Adesso AKB-132UB  017877B5001A__W XM2-300HP TechMnl_PT.indd  Relazione Tecnica degli Interventi  WEB TIMETABLES USER GUIDE - University of St Andrews  Change the World. One Call at a Time.™  Lily Operating Instructions    Copyright © All rights reserved. 
   Failed to retrieve file