Home
Data Sheet - Test Equipment Depot
Contents
1. coded leads on spools red blue two 100 ft color coded leads hand tied green black one 30 ft lead green four T shaped auxiliary ground electrodes set of five spaded lugs one 100 ft AEMC tape measure DataView software ground tester workbook CD carrying bag for meter carrying bag for kit and user manual Accessories Optional 25Q Calibration Checker Cat 2130 59 Tape Measure 100 ft Cat 2130 60 Test Kit for 3 Point testing includes two 150 ft color coded leads on spools red blue one 30 ft lead green two 14 5 T shaped auxiliary ground electrodes set of five spade lugs 100 ft tape measure and soft carrying bag Cat 2135 35 Test Kit for 4 Point testing includes two 300 ft color coded leads on spools red blue two 100 ft color coded leads hand tied green black four 14 5 T shaped auxiliary ground electrodes set of five spade lugs 100 ft tape measure and soft carrying bag Cat 2135 36 Test Kit for 4 Point testing includes two 500 ft color coded leads on spools red blue two 100 ft color coded leads hand tied green black one 30 ft lead green four 14 5 T shaped auxiliary ground electrodes set of five spade lugs 100 ft tape measure and soft carrying bag Cat 2135 37 Ground Test Kit for 3 Point testing Supplemental for 4 Point testing includes two 100 ft color coded leads one 30 ft lead green two 14 5 T shaped auxiliary ground electrodes and soft carrying bag Cat 2135 38 Test Eq
2. Multi Function Ground Test Equipment 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Depot 1 800 517 8431 Resistance Soil Resistivity Tester Model 6470 The Digital Ground Resistance Tester Model 6470 performs grounding resistance soil resistivity bonding tests 2 and 4 wire Resistance and earth coupling measurements This direct reading tester measures from 0 01 to 99 99kQ and is auto ranging automatically seeking out the optimum measurement range test current and test frequency Easy to use simply connect the leads select the test mode Press to Measure and read the results Up to 512 test results can be stored in internal memory for later recall to the display or download to a PC for analysis and hard copy reports The large LCD is easy to read and indicates grounding electrode resistance test voltage current and frequency It also measures individual test electrode resistance battery status and more depending on the measurement The Model 6470 is Cat IV rated to 50V and is over voltage protected to greater than 250V against accidental connection to live circuits Live voltage if present is also displayed on screen In the event of a system fault the Model 6470 can withstand 250V between any two terminals Also included are manual and automatic test frequency selection from 40 to 513Hz user selectable Soil Resistivity test method Wenner or Schlumberger
3. ORMATION CATALOG NO Ground Resistance Tester Model 6470 3 Point 4 Point Bond Test Digital DataView software Cat 2135 01 Includes meter rechargeable NiMH batteries optical USB cable power adapter 110 240V with power cord 115V US DataView software ground tester workbook CD carrying bag for meter and user manual Ground Resistance Tester Model 6470 Kit 150 ft Cat 2135 02 Includes meter rechargeable NiMH batteries optical USB cable power adapter 110 240V with power cord 115V US two 150 ft color coded leads on spools red blue one 30 ft lead green two T shaped auxiliary ground electrodes set of five spaded lugs one 100 ft AEMC tape measure DataView software ground tester workbook CD carrying bag for meter carrying bag for kit and user manual Ground Resistance Tester Model 6470 Kit 300 ft Cat 2135 03 Includes meter rechargeable NiMH batteries optical USB cable power adapter 110 240V with power cord 115V US two 300 ft color coded leads on spools red blue two 100 ft color coded leads hand tied green black four T shaped auxiliary ground electrodes set of five spaded lugs one 100 ft AEMC tape measure DataView software ground tester workbook CD carrying bag for meter carrying bag for kit and user manual Ground Resistance Tester Model 6470 Kit 500 ft Cat 2135 04 Includes meter rechargeable NiMH batteries optical USB cable power adapter 110 240V with power cord 115V US two 500 ft color
4. age 16 or 32V user selectable Frequency From 41 to 128Hz selectable External Voltage Measurement Range Auto Ranging 0 1 to 65 0V DC to 450Hz Accuracy 2 of Reading 3cts Resistance Measurement Bond Testing Measurement Type 2 Point with lead resistance compensation or 4 Point Kelvin sensing user selectable Range Auto Ranging 2 Point 0 01 to 99 9kQ 4 Point 0 00102 to 99 99kQ Accuracy 2 of Reading 2cts Test Voltage 16Vpc or auto polarity Test Current Up to 250mA max MECHANICAL Dimensions 10 7 x 9 76 x 5 12 272 x 248 x 130mm Weight KG o Gase i PP ypropylene DISPLAY COMMUNICATION Software ENVIRONMENTAL 14 to 122 F 10 to 55 C 40 to 158 F 40 to 70 C SAFETY CE Mar o d S EJ AEMC INSTRUMENTS www aemc com 3of 9 Functional Displays 4 Point Bond Test 2000 The 4 Point Kelvin sensing Bond test shows lead connections bond resistance test results test voltage and current Frequency Selection The Frequency selection screen displays selected frequency and voltage for the test as well as lead connection Schlumberger Test 0 oe Se The Schlumberger test displays test lead connection soil resistivity test results p and electrode spacing EJ AEMC INSTRUMENTS 3 Point Fall of Potential Test The 3 Point Fall of Potential test displays test lead connection grounding rod resistance and test electrode res
5. an be easily disconnected e 4 Point tests or soil resistivity measurements Locating areas of lowest soil resistivity is essential for achieving an economical grounding installation e Touch and step potential testing measurements These tests are recommended when the ground cannot be disconnected where ground faults are highly likely to occur or when the footprint of grounded equipment the outline of the part of equipment in contact with the earth is comparable to the size of the ground to be tested e 2 or 4 Wire tests for continuity tests on bonding or on grounding systems This test is a DC resistive test using 250mA or more and is used to check bonding of all connection points on the ground system e Earth coupling measurement and display Used to estimate the influence of two earth resistance systems that are not connected to each other www aemc com 2of9 Specifications ELECTRICAL 3 Point Measurement Range Auto Ranging 0 019 to 99 99kQ Resolution 0 0102 to 100Q Test Voltage Nominal 16 or 32VRMS user selectable Resistance Measurement Frequency 41 to 513Hz user selectable or automatic selection Test Current Up to 250mA Accuracy 2 of Reading 1ct Earth Coupling Measurement Yes Soil Resistivity 4 Point Measurement Test Method Wenner or Schlumberger selectable with automatic calculation of test results in 2 meters Q cm or Q feet Range Auto Ranging 0 01 to 99 99KQ Resolution 0 01 to 10002 Test Volt
6. clear and easy to use tabbed dialog boxes all of Model 6470 functions can be configured and tests can be initiated Results can be displayed in real time and stored in your PC Reports may be printed along with the operator s comments and analysis DataView is included with the Model 6470 SIC R o s Cener AUTO Fall of Potential Plot Name of Operator Model Location Date Ground System Type Single Rod Rod Depth Serial Multiple Rods Grid Longest Diagonal Dimension Z Electrode Distance Voltage Electrode Y as distance Measured Test Conditions from Ground Resistance Rod under Temp Soil Dry Testi Soil Type Loam Sand amp Gravel Shale Limestone Sandstone Granite Slate Other 100 90 80 70 60 50 Resistance Q 40 30 0 10 20 30 40 50 60 70 80 90 100 Y Electrode Distance in from the Grounding Electrode to the Z Electrode www aemc com 5of 9 DataView Functional Displays lt Configuration Ee EJ E Configuration aani FV Ten Vollage Limi Bon
7. d test setup screen operator can choose two or four wire test method amp Configuration Soil resistivity test setup screen operator can select either the Wenner or Schlumberger test method select test voltage and frequency and program the spacing between electrodes in either feet or meters 3 Point Fall of Potential test setup screen operator can select test voltage manually select the test frequency or have the instrument automatically select the best test frequency Distance for Y electrode can also be entered manually and a Fall of Potential plot generated EJ AEMC INSTRUMENTS www aemc com 6of9 Construction AEMC Large electroluminescent backlit USB liquid crystal display communications port DC power input Z Electrode terminal Rotary Y Electrode dina terminal 2 ce wen switch Xv Electrode terminal Kir Oe Tans r Start stop test button X Electrode Se tity terminal l E DISTANCE ak START STOP DPLAYT aL el Ji eee SMPOTH GROUND TESTER B AEMC MODEL 6470 HETEL Six program function buttons Multi Function Ground Resistance Tester Model 6470 Kit 500 ft kit shown includes meter NiMH batteries optically isolated USB cable Data View Model 6470 includes meter NiMH battery software external battery charger AC power cord one 30 ft green wire one pack optically isolated USB cable each 500 ft red and blue wire one each 100 ft black and green wire set o
8. f five DataView software external battery spade lugs four 14 5 auxiliary electrodes tape measure 100 ft charger AC power cord and user manual carrying case and user manual Catalog 2135 01 Catalog 2135 04 Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 FAX 781 665 0780 TestEquipmentDepot com Accessories 4 Af p Test Kit for 3 Point testing includes two 150 ft color coded leads on spools red and blue one 30 ft lead green Test Kit for 4 Point testing includes two 300 ft color coded nT se leads on spools red and blue two 100 ft color coded leads green and black pwa 14a T paced aa OUNO es One Ser al four 14 5 T shaped auxiliary ground electrodes one set of five spade lugs five spade lugs 100 ft tape measurer and carrying bag 100 fit f Catalog 2135 35 ape measurer and carrying bag Catalog 2135 36 Test Kit for 4 Point testing includes two 500 ft color coded leads on spools Test Kit for 3 Point testing supplemental for 4 Point testing includes red and blue two 100 ft color coded leads green and black two 100 ft color coded leads one 30 ft lead green one 30 ft lead green four 14 5 T shaped auxiliary ground electrodes two 14 5 auxiliary electrodes and soft carrying bag for use with kit number one set of five spade lugs 100 ft tape measurer and carrying bag 2135 35 when 4 Point soil resistance testing is necessary Catalog 2135 37 Catalog 2135 38 ORDERING INF
9. ic frequency scan from 40 to 513Hz for optimum test accuracy in electrically noisy environments Selectable test voltage limit of 16 or 32V with up to 250mA of test current Automatic recognition of all electrode connections and measurement of their resistance value Auto off power management Optically isolated USB communication we a j EJ AEMC INSTRUMENTS Er Ery Asad r il F ae i r Sih PL ee a ee oe ppe E i ra 1 T brai ae e Eaa pT s Pa ah Tei kly checks ground rod resistance Connect Press Read approach Remote set up and operation of all measurements using DataView software supplied Automatic report generation Rechargeable NiMH batteries from wall charger or vehicle power Rugged dustproof and rainproof field case IP54 rated in closed position IEC 61557 parts 4 and 5 compliant Includes DataView software for data storage real time display anal ysis report generation and system configuration F i 5 E aS 3 A F iapa t i TE E j aw T or y ji F r mia pJ ala i n i F N a DF ar arte FT a n Fr Ep ee al fee Ley J ss Ss ma i i et Tl a v a 4 ae ry a Li E a a if i j La i pri Ce es HA ib r I aye Lay i Sa DAT pea Ra ie z le TIAE ETEF h j AES H Applications e 3 Point measurements of resistance to ground of ground rods and grids 3 Point measurements are generally used when the electrode or grid c
10. istances Test voltage current or frequency can also be displayed with a button press Data Storage Displays test results stored at a specific memory location as well as the test function Wenner Test The Wenner test displays test lead connection soil resistivity test results p electrode spacing and resistance www aemc com 4of9 DataView Software DataView Data Analysis and Reporting Software Configure all functions of the Model 6470 e Run tests and analyze real time data from your PC e Configure all test functions and parameters from your PC e Customize views templates and reports to your exact needs e Create and store a complete library of configurations that can be used with the Model 6470 as needed e Display Fall of Potential plots tabular listings of test results resistance vs frequency plots soil resistivity and bonding tests e Print reports using standard or custom templates you design e Free updates are available through our website www aemc com DataView Minimum System Requirements e Windows 2000 XP Vista 128MB of RAM for Windows 2000 256MB recommended 256MB of RAM for Windows XP 512MB of RAM for Windows Vista e 80MB of hard disk space 200MB recommended e CD ROM drive EJ AEMC INSTRUMENTS DataView software provides a convenient way to configure and control ground resistance tests from your computer Through the use of
11. uipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 FAX 781 665 0780 TestEquipmentDepot com
12. user selectable 2 Wire or 4 Wire Bond Resistance test and a Earth coupling test feature Additional features of the Model 6470 include a heavy duty field case sealed against dust and water when closed the test button is also sealed against the elements NiMH rechargeable batteries An external charger powered ceo from 120V 230V 50 60Hz is included The Model 6470 can also be powered and charged from a 12V vehicle battery Testing while recharging is allowed The Ground Resistance Tester Model 6470 is rugged easy to use and ideal for maintenance crews and contractors performing numerous tests It exceeds mechanical and safety specifications for shock vibration and drop tests called out in IEC standards The adjustable test frequency and filtering provide for rejection of high levels of interference allowing it to be used under difficult conditions such as the presence of high stray currents that normally affect accuracy and measured auxiliary rod resistance enables the user to optimize measurements in difficult soil environments Model 6470 quic Features 3 Point Fall of Potential measurement with manual or automatic frequency selection 4 Point soil resistivity measurement with automatic calculation of Rho p and user selection of the Wenner or Schlumberger test method 2 and 4 Wire DC resistance measurement Bond testing with automatic polarity reversal 3 Point earth coupling measurement Automat
Download Pdf Manuals
Related Search
Related Contents
Manhattan 177375 mice zelsius®C5-CMF Manuel de montage et d`instruction MANUEL D`UTILISATION Outil de coupe-herbe/débrouss West Penn Wire CN-FS11V coaxial connector 取扱説明書 警告 注意 User Manual - Scoota Mart 600TV Lines Super High Resolution Color Camera Installation Manual Altronix ALTV248300ULCB power extension INSTALLATION MANUAL Copyright © All rights reserved.