1. 11 2012 JK Fax 1 601 355 1831 sales thermoprobe net www thermoprobe net
2. RIZED REPAIR It is recommended that service beyond the scope of this manual be performed by ThermoProbe Inc or one of its authorized distributors 11 2012 JK USER INTERFACE EN Function Button Power Button Pressing the Power button once will turn on the device The instrument will shut off automatically within 20 minutes Pressing the Power button twice within 1 second will save whatever reading is on the display Pressing and holding the Power button will and shut the instrument off and clear all logged and averaged readings Pomer buno Operation Display One quick push Power On AIOE PRESSES Saves current reading LoggEd when display teads OFF a OFP pay Clears Readings Function f button Press and Hold the Function Button f to display options in a menu format When the desired function is displayed release the button Up to 10 readings can be saved and averaged Hold and Release when the Function PF Example on Display display reads List ae aca ull ETO EN rEAd 1 78 2 F rEAd 2 ree y 8 74 4 F etc At displays the average of all saved readings At 76 3 F C F Changes units 76 3 F or 24 6 C dEC Changes units resolution 0 1 or 0 01 76 3 F or 76 36 C Backlight When the instrument is operating in low light conditions a photocell will detect this situation and permit t
3. USER INSTRUCTIONS TP7C amp TP9 INTRODUCTION This manual describes the basic function use and safety instructions for a model TP7C or TP9 portable digital thermometer instrument These ThermoProbe instruments are intended for use in both hazardous flammable and non hazardous areas under dry conditions at ambient temperatures between 20 to 40 C The instruments are not intended for use in permanent outdoor installations and are not intended or tested for icing conditions Additional means of protection should be used where the equipment may be exposed to excessive external stresses e g vibration heat impact etc SAFETY INFORMATION BEFORE USE ThermoProbe thermometers are designed for safe operation in hazardous locations Potentially Flammable or Explosive The user must have a working knowledge of appropriate safety instructions a The inspector must have a thorough knowledge of the products to be measured and must know of the safety precautions to be taken when working with the material to be measured b The instrument shall be checked concerning severe defects check that instrument is complete including grounding bonding cable has good batteries etc If necessary check measurement accuracy If any defects are found the instrument should not be used until repairs have been made c The instrument especially cable and probe should be clean both for safety reasons and ease of use d The physical measurement location shoul
4. d be evaluated for primary and secondary risks e Power source must be removed before performing any maintenance f Exchange of components other than the batteries may compromise ATEX IECEXx certification and shall only be undertaken by Thermoprobe or one of its qualified service providers See also Authorized Repair section g To reduce the risk of fire or explosion this device must be bonded to the tank according to clause 6 3 2 e IEC EN 60079 14 before and during introduction into the tank and shall remain bonded until complete withdrawal from the tank h The device must remain bonded to ground earth using the provided connection whenever a hazardous atmosphere could be present as well as during situations where electrostatic charging can occur such as the unwinding winding of the thermometer cable or filling or emptying of the tank CAUTION In the event that any part of the instrument should become electrostatically charged in a potentially hazardous location follow company policies for testing and clearing the area of any hazardous gases before attempting to bond the instrument to earth ground If this is not possible allow sufficient time for the instrument to naturally dissipate any charges before attempting to bond to earth ground Given the atmosphere this could take several hours GUIDANCE NOTE Problems with aggressive substances and environments Be aware of aggressive substances and that extra protection may be needed Caus
5. he backlight to function Battery Check When the voltage of the batteries is low the device will indicate Lo bAtt on the display before resuming normal display functions When the batteries are low the backlight will not operate in order to conserve power while the user completes his operations Replace batteries as soon as possible in a safe location after Lo bAtt is noticed this will ensure backlight operation and avoid possible malfunctioning Do not attempt to calibrate the instrument if the Lo bAtt has been displayed since the new calibration values may not be properly stored to memory 11 2012 JK Error Codes ErrHI indicates the sensor is operating above its temperature limit the Probe Assembly is open circuited from a cut or broken section or the cable is not properly inserted at the circuit board terminal The most common cause is a damaged cable ErrLO indicates the sensor is operating below its temperature limit the Probe Assembly is short circuited due to a smashed or cut section or the cable wire polarity is reversed at the circuit board terminal The most common cause is a damaged cable nO rEAd the user accessed List or At before saving temperatures REPLACING THE PROBE ASSEMBLY NOTES 1 Replacement of the probe assembly requires re calibration of the device Replacement should only be done by experienced personnel and if calibration equipment is available 2 Please refer to IEC EN 60079 19 when maki
6. he highest temperature point the display will flash rEAdY and the new calibration settings will be in effect The buttons will now resume their normal operating functions The calibration settings are saved to flash memory when the device is turned off The unit will not turn off automatically manually turning the unit off saves the calibration data Ensure that your calibration efforts are saved by shutting the instrument off now Calibration Error Codes 1 nO CAL This is seen after the instrument is turned on when no calibration data has been saved yet The device must be calibrated before using 2 CAL Err2 This is seen after the instrument is turned on when there is a flash memory error or the calibrations data is corrupted When this error is displayed the device will inaccurately display temperatures without using any calibration data This error probably indicates a device error The user should contact the distributor or ThermoProbe Inc 3 CAL Err3 This is seen after the instrument is turned on when calibrations data reads OK but is invalid This could be an error in calibrating the device When this error is displayed the device will read temperatures without using any calibrations data This error could occur as a result of not saving the low or mid temperature before moving the device to the next bath during calibration The device should be recalibrated ThermoProbe Inc 112A JETPORT DR PEARL MS 39208 Tel 1 601 939 1831
7. icated by black marking 11 2012 JK Figure 1 Probe Assembly Lead Attachment CALIBRATION PROCEDURE e The calibration mode should only be accessed by qualified personnel with proper equipment otherwise calibration integrity may be compromised Read the following instructions carefully e A 2 Point 3 Point or a 4 point calibration can be performed A third or fourth point is only necessary when high accuracy is required at temperatures of 300 F and higher You must have the proper equipment for every point of calibration e Do not attempt to calibrate the instrument if the Lo bAtt has been displayed since the new calibration values may not be properly stored to memory e Refer to API 7 2 or another recognized standard for routine calibration verification recommendations e Calibration must not be performed in any environment considered to be hazardous Equipment needed e Ice Bath or other low temperature bath with certified reference thermometer e Warm to hot fluid bath between 20 C approx 68 F or higher up to 90 C approx 194 F with certified reference thermometer see Note e Optional high temperature oil bath at about 150 C 300 F and certified reference thermometer Note for limited calibration If entire range capability of instrument is not required the 2 point high adjustment can be made at a temperature relatively close to the common temperature of the liquid measured and accuracy will be maintained withi
8. le at each temperature calibration point the arrows will flash 3 times when the temperature reading stabilizes Calibration can be performed to hundredths of a degree The On Button increases the display reading the f Button decreases the display reading Holding the On Button or the f Button adjusts 0 1 degrees increments Momentary Presses of the On Button or the f Button for less than 0 5 seconds adjusts 0 01 degrees for every press When you leave calibration mode the display remains in hundredths for that session only to allow you to re check the temperatures Once you have turned the device off the display will only show in tenths 3 Once the temperature has stabilized using a certified reference device check the actual temperature in the bath and use the On Button or the f Button to adjust the device to the actual temperature The On button will decrement the readings and the f Button will increment the reading Once the device temperature matches the actual temperature press the Calibration Button to save the setting The display will scroll SAUE or CAnCEL If the SAUE option is chosen the blinking letter will change to represent the next temperature level A gt b If the CAnCEL option is chosen then the calibration procedure is exited and the prior calibration values are re activated 4 Move the probe to the next bath and repeat step 3 After you save t
9. lift battery up from the holder and then remove the remaining battery d Install each new battery making sure the end of the battery is aligned with symbol embossed in the battery case e Replace the retaining device and reinstall the cover CERTIFIED Batteries for the TP9 and TP7C are as follows Manufacturer Type Part Number Duracell AA LR6 Alkaline MN1500 Panasonic AA LR6 Alkaline LROXWA GP Gold Peak AA LR6 Alkaline GPI5A MEASURING PROCEDURES See www thermoprobe net for video on proper use of this instrument Contact your distributor or ThermoProbe Inc for names of training facilities Refer to American Petroleum Institute measurement standard Chapter 7 2 When the temperature has stabilized both Up and Down arrows will be displayed If you wish to log the temperature for averaging purposes rapidly press the On button twice An acknowledgment of a saved reading will occur with a display of LoggEd If the display shows then reading was not logged due to a non stabilized temperature The real time temperature will then be displayed again indicating the next reading is ready to be taken If you have completed all the readings needed for your operation the f Button may be used to LiSt all the readings logged and At average of the logged temperatures Document your readings and average temperature before powering off the device minutes after the last button push AUTHO
10. n the limited range For example If liquid product to be measured is commonly less than 38 C approx 100 F then a high point calibration can be made near that temperature Temperature accuracy above this calibration point cannot be assured Calibration Button Power Button Function Button 11 2012 JK To calibrate proceed with the following steps 1 A hidden Calibration Button is located on the front overlay underneath the ThermoProbe logo see diagram above While the instrument is on first press and hold the f Button and then press and hold the hidden Calibration Button until the display scrolls through the options 2Pt CAL 3PtCAL 4PtCAL CAnCEL When the desired option is displayed release the buttons The Calibration Mode can be exited in two ways If the user is not ready to enter the calibration mode the CAnCEL option can be chosen If a user needs to end a calibration before completing it then STEP 3 below provides an option to safely exit the operation and revert to previously stored calibration values 2 The device is now in calibration mode The last character on the right side of the LCD will now be blinking A representing the lowest temperature calibration point b the next higher temperature point C the next higher temperature point only used in 3 point calibration or 4 point calibration mode d the highest temperature point only used in 4 point calibration mode Whi
11. ng the repair 3 Only use replacement probe assemblies obtained from ThermoProbe Inc or one of its authorized distributors with part specifications as follows PART FORMAT AS FOLLOWS PG XXX XX XX Designates cable length in meters L Designates population of brass or feet not to exceed 50 meters incremental markings over cable length Designates Sensor Weight as follows RC Rail Car SW Standard Weight EW Extra Weight AW Asphalt Weight a First follow REPLACING BATTERY instructions a through c to remove batteries b On the circuit board push the terminals clamps down and remove the wires noting the wire lead color code arrangement White positive sensor wire Black negative sensor wire and sheild c Set the cover and circuit board aside and remove the strain relief knot in cable assembly d Unwrap the cable from the assembly and pull the cable free of the rubber grommet e Insert the new cable wire through the rubber grommet and then pull several inches of cable slack past the grommet f Tie a simple overhand knot in the cable at the grommet for strain relief and pull any slack through the grommet g On the circuit board push the terminal clamps down and insert the new wire leads making sure the black wire lead is installed in the terminal indicated with black paint See Figure 1 h Reinstall the batteries and cover and re spool the cable assembly 1 Perform a calibration Negative Terminal ind
12. rinsically safe system assumes the fuel and oxygen is present in the atmosphere but the system is designed so the electrical energy or thermal energy of a particular instrument loop can never be great enough to cause ignition T PIC T P9 Hazardous odone Location application Zone 0 ia Hazardous Area Zone ib TP Instrument Body Temperature Probe Probe Cable Ground Connection Zone 0 Where ignitable concentrations of flammable gasses vapors or liquids can exist all of the time or for long periods of time under normal operating conditions Zone 1 Where ignitable concentrations of flammable gasses vapors or liquids can exist some of the time under normal operating conditions 11 2012 JK REPLACING BATTERY WARNING e Batteries must be changed in Non hazardous area e Batteries must be of correct approved type e Batteries must be installed with correct polarity making sure the end of the battery is aligned with symbol embossed in the battery case e New batteries must not be mixed with old batteries Batteries must not be mixed with batteries of other manufacturers e Batteries must not be installed with polarity reversed where one cell could charge another cell a Ensure the instrument is in a non hazardous area b Remove the 3 screws holding the front cover on the TP7C or the 2 screws holding the front cover on the TP9 c Remove battery retaining device push one battery towards the spring contact and
13. tic soda highly basic and acidic substances will erode aluminum and copper ground clip and wire The Sensor Cable assembly has external surfaces of stainless steel and fluoropolymer material Exposure to Excessive heat can melt the plastic components of the instrument 11 2012 JK Safety Approvals for TP7C and TP9 II 2 1 G Ex ib ia IIB T4 ThermoProbe Inc 112A JETPORT DR PEAN BUSA t DI www thermoprobe net 0359 Model TP7C Year 20 SN 70 II 2 1 G Ex ib ia IIB T4 112A JET r ThermoProbe Inc Ex DEMKO 11 ATEX 1104891X RL N MER EATE IECEx ULD 11 0008X 39208 USA D I2 1 G Exib fia IIB T4 ne A ne 1 601 939 1831 DEMKO 11 ATEX 1104891X e lesion ected ry Randi ese IECEx ULD 11 0008X Applicable Standards are Agency or Safety Designation IEC 60079 0 2007 Ed 5 TECEx IEC 60079 11 2006 Ed 5 TECEx EN 60079 0 2009 Europe ATEX EN 60079 11 2007 Europe ATEX INTRINSIC SAFETY Intrinsically safe equipment is defined as equipment and wiring which is incapable of releasing sufficient electrical or thermal energy under normal or abnormal conditions to cause ignition of a specific hazardous atmospheric mixture in its most easily ignited concentration ISA RP12 6 This is achieved by limiting the amount of power available to the electrical equipment in the hazardous area to a level below that which will ignite the gases In order to have a fire or explosion fuel oxygen and a source of ignition must be present An int