         Contents
1.    Proven    DryCal     TECHNOLOGY                ML 800  User Manual    Mesalabs          Mesalabs    Mesa Labs    Commitment    As a manufacturer of critical equipment  quality is the highest priority in  our manufacturing process  Mesa Labs commits to providing creative   innovative thinking and the engineering and scientific expertise needed to  produce gas flow related products and solutions that are universally  recognized for their superior performance  quality and value     Proven DryCal technology is the recognized leader in gas flow  measurement  providing the industry   s most reliable products  service and  solutions for professionals in environmental protection  workplace safety   industrial process control and laboratory calibration     We strive to provide the closest NIST traceable  legal defensibility of any  flow calibration equipment manufacturer  and we actively maintain our  NVLAP  NIST  ISO 17025 laboratory accreditation in order to support our  claims and continually improve our quality system and laboratory profi   ciency  Thank you for purchasing our products  From all of us at Mesa Labs   best wishes for many years of accurate  defensible primary flow  measurements     Mesa Laboratories Inc  10 Park Place Butler  NJ 07405 USA   973  492 8400 FAX  303  484 4992 www mesalabs com Symbol    MLAB    on the NASDAQ    Table of Contents    Introduction  Q nn near 2  Register  Your Product  u    nisse 2  About Your ME 800  2  een Gada aise ee 2  OP ralOMescse
2.   Parameters Factory Settings Optional Settings  No  of Readings in an Averaging Sequence 10 1 100  Atmospheric Pressure mm Hg mBar  kPa  PSI  Temperature oC OF   Standardized Temperature Setting 0   C 0 0 50 0   C  Date Format MM DT YR DT MM YR   Time Format AM PM 24 Hr       Connecting the ML 800 to a Gas Flow Source    The accuracy of the ML 800 is dependent upon its source being stable  An unstable flow source may produce  inconsistent readings  Flow direction is indicated on the cell  To use a pressure flow source  connect to the inlet  fitting  or to use a vacuum flow source  connect to the outlet fitting     The ML 800 is designed to be used at ambient pressures only  This is easily accomplished by leaving the outlet   of the flow cell open to atmosphere for pressure applications or the inlet open to atmosphere in vacuum scenarios    If tubing is needed to exhaust test gases to a fume hood insure that the exhaust tubing is of sufficient diameter such  that the pressure to the ML 800 calibrator does not exceed 5 inches of water column  WC  above ambient pressure     Taking Readings    The ML 800 default is preset for ten  10  readings in an averaging sequence  This parameter is user definable   see Setup Menu 2  Reading Type    in Average  amp  Min  Reading      Hain Meru  Fun  SeLur  Lesk Test    1  Press Enter  Read  or Auto to enter the  Run Menu           e Press the Read button to initiate a single reading   e Press the Auto button to initiate continuous hands free
3.  This is the unique piston tare value of a particular flow cell  It is factored into each flow reading for  leakage independent measurements  The piston tare value is factory set and non adjustable     Option 2     PTVM    Piston Tare Value Multiplier   This is the adjustable piston tare value multiplier used to adjust the piston tare value to enhance accuracy at lower  flows when using gasses other than nitrogen or air  The piston tare value multiplier   s factory setting is 1 000   although it may be changed to any value between 0 200 and 3 00 if gasses other then air or nitrogen are used     Using Sensor Factors    When calibrating an MFC or MFM using a surrogate or proxy gas  meaning the MFC or MFM was originally calibrated  by the manufacturer for a gas other than what you are currently using   the reading of the ML 800 can be scaled by  the Sensor Factor to allow for this     To enable your ML 800 to scale its actual flow measurements to match the adjusted flow from the MFC or MFM  input  the sensor factor into the ML 800 before calibrating the MFC or MFM  see SETUP     Readings      The Sensor Factor is displayed on the run screen of the ML 800 as SF X XXX  The default setting is 1 000  Because a  sensor factor other than your ML 800 s default value of 1 000 modifies the actual flow to a    reported    flow make sure  to change this setting back to 1 000 when a calibration involving the use of a sensor factor is completed     Leak Testing    To ensure proper function o
4.  can be used to measure the gas flow rates at atmospheric pressure    mmHg  Its Proven Dry   Cal Technology features a near frictionless piston and flow cell design that provides highly accurate  liquid free flow  measurements of unequaled speed and convenience  For more information about our patented Proven DryCal  Technology  please visit our website at drycal mesalabs com  The ML 800 consists of two primary components  The  base and the flow cell  there are five interchangeable flow cell models  each with a designated flow range   The flow  cell plugs into the base unit in order to create a complete  functional ML 800 system  Neither component can operate  independently  The base component houses the main computer and timing crystal  The flow cell is easily fitted into  the base unit   s 9 pin connector and two guide pins     The flow cell performs the actual flow measurements  For this purpose  it houses a  piston and glass assembly     comprised of a borosilicate glass tube housing a precision machined piston  The flow cell also contains an  integrated temperature sensor and barometric pressure transducer in the gas flow stream for instant conversion  of the volumetric readings into standardized flow  The ML 800 includes an RS 232 port  serial port  and cable     You can also utilize Mesa   s DryCal Pro software package available for download from http   drycal mesalabs com   drycal pro software      DryCal Pro software captures flow data from your DryCal   instrument fo
5.  readings   e Press the Stop button to stop current flow reading and open valve     e Press the Reset button to clear the display of current data     Setting User Preferences    The ML 800 offers enhanced electronics options to allow the user to define parameters specific to an application   There are six Setup Menus     General Menu Navigation    Use the Up Down arrows to navigate between different lines within a menu  Use the Enter button to select a field  to be modified and to lock in any changes  Use the keypad for data entry  such as the number of readings in an  averaging sequence  use the Up Down arrows to toggle between fields  such as Temperature or Pressure units   Use the Fwd and Back keys to advance to the following menu or return to the previous menu     Keypad Function   The keypad is used for numeric character entry  When pressed  each alphanumeric button will display its corresponding  number  To go back  use the back arrow  The ML 800 will overwrite previous entries  Use the Enter key to lock in any  selections     Hain Meru  Furi  Se bur  Leak Test    To enter    Setup Menu   1    from the    Main Menu     select Setup and press Enter     Setup Menu 1  Sensor Factor  Calibration Type       Setur Menu     1  Sensor Factor 1  BBB  Cal  Type Seo    Option 1     Sensor Factor     As applicable  change the Sensor Factor from its default value of 1 000 to the value provided by the MFC or MFM  manufacturer  see Using Sensor Factors      Option 2     Cal  Type     
6.  the Piston Tare Value  Multiplier  PTVM   When using the instrument with gas species other then air or nitrogen  the molecular behaviors   of these gases may degrade the Piston Tare Value  For highest accuracy  the instrument s Piston Tare Value  PTV    can be adjusted  Adjusting the Piston Tare Value is accomplished by entering a new Piston Tare Value Multiplier  PTVM    The Piston Tare Value Multiplier is multiplied to the Piston Tare Value and used to adjust the measurement  the de   fault value for air and nitrogen is 1 000  The Piston Tare Value Multiplier  PTVM  can be set to any value from 3 000 to  0 2000     For flows above 20 ccm  a new a Piston Tare Value Multiplier  PTVM  value can be calculated by using the viscosity of  the gas being measured and accurate results will be obtained  Calculate the PTVM by taking the ratio of the viscosity  of nitrogen to the viscosity of the gas under test  For example  to calibrate hydrogen consider the following  at 0  C   the viscosity of nitrogen is 165 31 microPoise  and the viscosity of hydrogen is 83 21 microPoise  Express these as  165 31 83 21  or 1 987  and enter 1 987 as the Piston Tare Value Multiplier  PTVM  for this cell     When measuring alternate gases at flows below 20 ccm  or for absolute best accuracy  it may be necessary to  perform a dynamic leak test using the gas under test  Contact Mesa for information on this test     Setur Menu     6  PTU B   188 ccm  FTUH 1  BBB    Option 1     PTV    Piston Tare Value 
7. Te aa aaea 12  LEak TESEING aerario a E E EE a E E asas 12  Annual Maintenanceiand CalibratiOn sissy isinna ana aiaa ua aasan a aiaaa 14  RECERCAT ON saree nu l SD ua E A E au A a E Sau yaaa asua s uy sua h a sa asha 14  Sending  YOUr ML 800    Mesa  ua Shaw a a RE EEE enden ee 15  SHIPPING Wa    anregen 15  SLO0rd08   s asas EHER RE HNIFEEENEENEEERGEESEELETEFERTEEHERERGERE FERE ER EEE IHRENEREERERSEEREDEERTEUTTELELTEREEREEEENPELFPEUEBEEELEELLEPELTEEFEEFENERERR 15  Met  Lab Series  ML 800 Specificationsi   442  u au a a ann aan ce da haha ann aha ran 16  EiitediWakrantyrua Qua ayy qatana asua aun Sua FELOETELHLOERTELENEUREEOPEDEUEFOHILEEFOREE STELLE Ea a aE a a 17      Accuracy    Reliability    Convenience    Introduction    Congratulations  You ve chosen the Met Lab   Series ML 800  featuring Proven DryCal   Technology  The ML 800 has  five interchangeable cells that offer    0 25  to   0 15  of reading standardized accuracy within a portable primary  standard that   s ideal for both volumetric and mass flow applications where a high degree of measurement accuracy   is required  Before you get started  we recommend taking a few moments to review this manual and familiarize  yourself with the ML 800  If at any time you have questions regarding its operation  please contact Mesa through  our web site  www drycal com   or call us at 973 492 8400 to speak with one of our knowledgeable customer support  representatives     Register Your Product    Before we begin  please registe
8. This option changes the sample type being displayed  Toggles between    std      standardized readings  and    vol       volumetric readings      Option 3     Low Bat Shutoff     When    On    the unit turns off when running on internal battery and the battery voltage is low     Setup Menu 2  Reading Type    in Average  amp  Minutes Reading       Setur Menu   2  Reading Ture Cort    in Avo  5   Min Reading GE    Option 1     Reading Type     Toggles between    Cont      continuous auto read  and    Burst     performs continuous auto read  then stops after the  quantity in averaging sequence is reached  as programmed in Setup Menu 2  Option 2    in Average       Option 2       in Average     Changes the quantity in an averaging sequence  consecutive readings  from 1 100     Option 3     Min  Reading     Specifies the time interval  in minutes  between flow readings     Setup Menu 3  Pres  Units  Temp  Units  Temp Corr        Setur Menu     3  Fres Units mma  Teme Units   C  Teme Corre et    Option 1     Pres  Units     This option allows you to toggle between mmHg  mBar  kPa and PSI     Option 2     Temp  Units     This option allows you to set the temperature units for    F or    C     Option 3     Temp  Corr      This option is used to set the standardization temperature for indicating standardized flow rate     Setup Menu 4  Date  Time  amp  Battery Voltage       Setur Menu     4  Ar 18 68 05 603 AM    Bat 5 90 Volts  This menu has an alternate navigation method     Within 
9. al mesalabs com drycal pro software    DryCal Pro  captures flow data from your ML 800 directly to a pre configured table  The data can be exported to selectable Microsoft office  environment     To run DryCal Pro  you must have Windows  XP or 7  Microsoft Excel   2003 and up  and a RS232 port  or if your PC does not have  an RS 232 port you will need a USB to RS 232 adapter      ML 800 3 Cell requires release 2 08 firmware or higher installed in base    ML 800 75 cell requires release 2 10 or higher installed in base      Limited Warranty    The DryCal Met Lab Series ML 800 is warranted to the original end user to be free from defects in materials and workmanship under normal use   and service for a period of one year from the date of purchase as shown on the purchaser s receipt  The ML 800 s battery is warranted for six   months from the original purchase date  If the unit was purchased from an authorized reseller  a copy of an invoice or packing slip showing the   date of purchase may be required to obtain warranty service    The obligation of Mesa Labs under this warranty shall be limited to repair or replacement  at our option   during the warranty period  of any part   that proves defective in material or workmanship under normal use and service  provided the product is returned to Mesa Labs  transportation   charges prepaid    Notwithstanding the foregoing  Mesa Labs shall have no liability to repair or replace any Mesa Labs product    1  That has been damaged followin
10. cision instruments and gauges  such as depth and laser micrometers  and a temperature bath  When completed  you receive NIST traceable  ISO 17025   ANSI Z 540  and NIST Handbook  150 backed calibration certificates     Recertification includes a 90 day service warranty should any related labor or parts replacements prove faulty     Due to the intensive nature of our ML 800 Recertification  the time spent within our service facility during this pro   cess is generally fourteen days  beginning from the day we receive your product     For a detailed explanation of our Recertification process  our laboratory and our accreditation  please visit our  website at www drycal com and visit the Calibration  amp  Repair page     Sending Your ML 800 to Mesa    Please contact Mesa for an RMA  return merchandise authorization  number before sending your ML 800 to our  factory for elective Recertification or other service  You can get an RMA number through our automated web based  system at http   drycal mesalabs com request an rma   by emailing csbutler mesalabs com  or  by calling us directly  at 973 492 8400     Sending your ML 800 to Mesa without an RMA number may result in return of the instrument without inspection or a  substantial delay in service turnaround time     When requesting your RMA number  provide your ML 800 model  serial number and revision level  Also  describe any  product issues you may be experiencing  Please keep in mind that Mesa will not begin evaluation and servic
11. display  The default setting for the backlight is  always on  If you wish to conserve battery power you may wish to turn the backlight off or enable the backlight only  when inputting information on the keypad     To turn the backlight off  from the    Main Menu    press the Light button one time  To enable the backlight only when  inputting information on the keypad press the Light button a second time  To return to the default setting  always  on   press the Light button a third time    Low Battery Indicator    A low battery condition is indicated by a  B    appearing in the upper right hand corner of the LCD  The low battery  indicator allows the user to connect to an external power source prior to the unit powering down  Time between low  battery indication and loss of power varies depending on the current application     Battery Disposal     PB          In compliance with the European Union CE directive  2006 66 EC the battery in your ML 800 or ML 500 base  should be removed for recycling prior to disposal of the  ML 800 or ML 500 base  The battery in the ML 800 500 is a  valve regulated sealed lead acid battery  Please note that  opening the ML 800 or ML 500 base may damage connections  so this procedure should only be used for battery  disposal                 PROCEDURE   Remove six black Phillips head screws on the back and  side of the ML 800 or ML 500 base  Lift off the top  control panel  remove the two screws holding the bracket  that retains the battery and disco
12. e of your  ML 800 until you have accepted and approved  in writing  our formal RMA quote for service  This protects both you  and Mesa during this process and ensures a fair and efficient service experience     If sending your ML 800 for repair or evaluation  rather than elective Recertification   please contact Mesa for technical  support or troubleshooting assistance prior to shipping the unit  We will first attempt to resolve the situation over the  phone or via email  If you ve provided us with a detailed description of your ML 800 issue and application details and  we re unable to resolve the situation by telephone or email  we ll issue you an RMA number for the prompt return of  your ML 800 for evaluation     Please note that Mesa will make every attempt to verify your issue  as we want you to get the most out of your ML 800   However  if we are unable to detect a product issue or if we determine that the issue is application related rather than  product related  we reserve the right to charge an evaluation fee     Shipping   When shipping your ML 800  please ensure that the packaging is adequate to protect the instrument  Whenever  possible  your ML 800 should be shipped in its original packaging or within a hard case  such as a Pelican carrying  case  available for purchase from Mesa   We highly recommend using a standard freight carrier  e g   FedEx  UPS   that supplies tracking numbers and insuring the product against damage in transit     Mesa is not responsible fo
13. f the ML 800 annual factory calibration is recommended  However  a quality assurance  self test feature is provided to verify proper integrity of the flow cell  It is recommended that the self check leakage  test be conducted periodically as part of an ongoing quality assurance program  Passing the leak test does not ensure  proper function of the ML 800  although it does ensure that total leakage is within the product s allowable limits  The  leak test may take as long as 6 hours or more to complete per fitting  Since the leak test may take several hours  it is  recommended to plug the charger into the base during the leak test to prevent unit from turning off     When performing the leak test procedure note the inlet and outlet leakage  If the inlet and outlet leakage do not  agree within 50  of each other  assure that the inlet and outlet Fittings are tight  then repeat the leak test procedure   If the problem persists  please contact Mesa     The maximum leakage limits are  200 ccm for the ML 800 10  1 000 ccm for the ML 800 24  and 3 000 ccm for the  ML 800 44  If the leakage exceeds the maximum limits  repeat the leak test procedure  If the problem persists   please contact Mesa for assistance  Due to the low leakage of the ML 800 3  the self leak test is not recommended     To Initiate the Leak Test  1  Place the ML 800 on a flat  vibration free surface     2  While in the    Main Menu    install the leak test cable  included with the ML 800 base unit   between the 9 p
14. g sale  including but not limited to damage resulting from improper electrical voltages or currents   defacement  misuse  abuse  neglect  accident  fire  flood  terrorism  act of God or use in violation of the instructions furnished by Mesa Labs     2  When the serial number has been altered or removed  or     3  That has been repaired  altered or maintained by any person or party other than Mesa Labs    own service facility or a   Mesa authorized service center  should one be established   This warranty is in lieu of all other warranties  and all other obligations or liabilities arising as a result of any defect or deficiency of the product   whether in contract or in tort or otherwise  All other warranties  expressed or implied  including any implied warranties of merchantability and  fitness for a particular purpose  are specifically excluded     Mesalabs    Mesa Labs  10 Park Place  Butler  NJ USA 07405    Phone  973 492 8400  Toll Free  800 663 4977  Fax  303 484 4992    Email  csbutler mesalabs com  Web  www mesalabs com       2015 Mesa Labs MKO01 33 Rev L      
15. in ML 800 Base Connector and the 9 pin Cell Connector        3  From the    Main Menu    remove the cell and select    Leak Test        4  After a leak test is initiated  the display will read        5  Place the leak test plug over either the inlet or the outlet port fitting on the ML 800 flow cell   6  Invert the ML 800 flow cell     7  Push any key to commence the leak test  At this point the ML 800 will time the descent of  the piston  The leak test may take up to 6 hours to complete  The display will read        8  If the test is completed successfully a leakage rate will be displayed  The display will appear similar to the following        9  Repeat the procedure with the leak test port cap over the opposite port fitting     Leak Test Setup        Inverted Cell    Leak Test Plug    N  a Swagelok Fitting    Leak Test Cable             Annual Maintenance and Calibration    Your ML 800 is engineered to provide years of reliable service  with appropriate care and maintenance  Mesa  recommends an annual calibration by our ISO 17025 accredited laboratory  to help ensure the best possible flow  measurements and to provide an audit trail for those applications subject to regulatory requirements  If you should  encounter any problems with your ML 800  immediately contact Customer Service and provide a detailed  description of your situation  including ML 800 model and serial number  information about the flow source   and the current calibration setup  environmental conditions d
16. ions  Cell Model 3   H x W x D  9 x 4 x 6 5 in   258 x 102 x 165 mm  Dimensions  Cell Models 10  24  44   H x W x D  10 15 x 4 x 6 4 in   258 x 152 x 102 mm  Dimensions  Cell Model 75   H x W x D  15    x 5      381mm X 5mm    Weight  Base 43 5 oz   1233 g  Gas Compatibility  Non corrosive  non condensing  non combustible gasses  Less than 70  humidity     Usage     Flow Mode  Suction and Pressure   Configuration  Base with modular  interchangeable flow cells  five    Temperature  amp  Pressure Sensors  Yes  in the flow stream of each interchangeable flow cell  Reading Modes  Single  Auto or User Specified Burst   AC Adapter Plug  12V DC   gt  500 ma  2 5 mm  center positive  North American standard  others available  Battery Type  6 volt lead acid    Inlet and Outlet Fittings  Swagelok      Ambient Temperature  15 30   C   Storage Temperature  0 70   C   Ambient Humidity  0 70   non condensing   Display  Backlit LCD   Operating Pressure  Absolute   15 PSI   Data Port  Serial   Warranty  1 year  battery 6 months    Mesa highly recommends annual Recertification  product maintenance  pre and post dynamic flow comparisons and  full dimensional calibration   This is elective and is not included as a warranty item     All specifications are subject to change  Please contact Mesa or visit our web site at http   drycal mesalabs com for  the most current product information     DryCal Pro Software        Visit Mesa   s website to download your copy of DryCal Pro software  http   dryc
17. l for more information regarding acceptable gas  flow ranges or visit our website at drycal mesalabs com for the most current product specifications     A For use with clean laboratory air or other inert  non corrosive gasses only     Operating Your ML 800    Getting Started    Although the ML 800   s battery comes charged from the factory  we recommend fully charging your new battery by  connecting the AC power adapter charger to the ML 800   s charging jack and then plugging it into a standard wall  outlet  Initial charging should take approximately eight hours     After initial charging  you may continue to charge your ML 800 indefinitely by leaving it connected to the power sup   ply  Otherwise  make sure to fully charge your ML 800 at least once every three months to help maintain battery life     On bases manufactured after January 2008 a small cooling fan was added for improved performance  The fan  operates when the base is connected to the charger    Turning the ML 800 On  amp  Off   Press On to start the ML 800  An opening screen will appear indicating the instrument   s firmware revision level  followed by the  Main Menu        The ML 800 has a battery saving automatic shut off system  After 65 minutes of inactivity  the ML 800 will shut off   Alternatively  the unit can be shut off manually by pressing the Reset button followed by the number 0 from the    Main  Menu        LCD Backlight    The ML 800   s LCD display includes a backlight function to illuminate the 
18. nnect the connectors  plugged into the battery Lift the battery from the case          ML 800 Keypad    Power button A   65 minutes auto shut off  3  Backlight ld    uP    1 2 3  i i  BACK Wu   wo Menu navigation  Use for data entry DOWN  35  ou cox a Reset Stop    Q    Product functions       Using the ML 800 Keypad    General Menu Navigation   Use the Up Down arrows to navigate between different lines within a menu  Use the Enter button to select a field  to be modified and to lock in any changes  Use the keypad for data entry  such as entering standard temperature  conditions  use the Up Down arrows to toggle between fields  such as Temperature or Pressure units  Use the  Fwd and Back keys to advance to the following menu or return to the previous menu     Keypad Function    The keypad is used for numeric character entry only  When pressed  each alphanumeric button will display its  corresponding number  To go back  use the back arrow  The ML 800 will overwrite previous entries  Use the Enter key    to lock in any selections     e Reset Button  Clears both the current reading and the group average and opens the valve  allowing the piston  to return to its resting state     e Stop Button  Stops current reading and opens the valve  allowing the piston to return to its resting state     e Auto Button  Initiates automatic readings  must be accessed from the    Run Menu      If the Reading type has been  set to    Cont    in    Setup Menu   2     the unit will run until the pre
19. r damage that occurs during shipment     Storage  If you need to store your ML 800 for an extended period  please follow these guidelines     Always store it in a clean  dry place   If possible  leave it attached to its AC power adapter charger while in storage   If your ML 800 cannot be attached to its AC power adapter charger while in storage     Fully charge it before extended storage  If the battery isn   t fully charged prior to storage   it might be permanently damaged     Fully charge it at least once every three months     When removing your ML 800 from storage  recharge the battery for at least 12 hours prior to use     Met Lab Series ML 800 Specifications    Interchangeable Flow Cell Models     Part Number ML 800 3  ML 800 10 ML 800 24 ML 800 44 ML 800 75     Model 50 Mililiters Per Min Cell 500 Mililiers Per Min 5 Liter Per Min Cell 50 Liter Per Min Cell 100 Liter Per Min Cell  Cell   Flow Range 0 5   50 mL min 5   500 mL min 50   5 000 mL min 500   50 000 mL min 1 100 L min   Accuracy   0 25  0 5 50 sccm  0 15   0 15   0 15   0 15    Standardized   0 002 sccm   Time Per Reading 1   60 seconds 3   135 seconds 3   90 seconds 1   35 seconds 1 50 seconds   Weight 80 0z 2300 grams 85 1 oz   2412 5 g 86 oz   2439 3 g 88 4 oz   2507 g 160 oz   4535 g   Inlet Fitting 1 8 1 4 1 4 1 4 1 2   Outlet Fitting 1 8 1 4 1 4 1 2 1 2    Specifications based on averaged readings  Accuracy is stated as a percent of reading  including standardization  if applicable    Basics     Dimens
20. r easy export into common software pack   ages     e Capture flow data from your DryCal   directly to a PC and export the data to a Microsoft environment    e Real time data monitoring   e Enter flow rates from pumps or other flow source and calibrate the flow source    e Enter flow data from a flow meter and compare the flow measurements from your DryCal   precision calibrator     e Upload the latest version of the firmware to your DryCal         L    Gas Inlet       Gas Outlet    Cell Connector         Guide Pins  Base Connector          Back and Side Front and Side       Unpacking Checklist   Your ML 800 has been packaged with care and includes all components necessary for complete operation  Please  take a moment to check that you have received the following items  If you believe you have not received a full  shipment or if you have any questions  please contact Mesa immediately     Your ML 800 Base with the following  Your ML 800 Flow Cell Includes  e ML 800 Electronic Base e ML 800 Flow Cell  e Battery Charger e Leak Test Plug  e Leak Test Cable e Certificate of Calibration  e RS 232 Cable e Inlet Filter  ML 800 3 only     e Certificate of Calibration    Warnings    N The ML 800 is not rated intrinsically safe and is not for use with explosive gasses or for use in  explosive environments       N The ML 800 is not designed for gas flows above the rated specifications of the flow cell in use  Please consult  the product specification on the inside front cover of the manua
21. r your product with Mesa Laboratories  Inc  To complete your registration  log on to our  website at http   drycal mesalabs com product registration  and complete the registration form  Registration of your new  DryCal product ensures your instruments warranty claim information is properly documented in Mesa   s database  and  provides you the peace of mind that your new calibration equipment is covered under our warranty service plan  Addi   tionally  this registration will allow you to better manage your service and calibration dates through the receipt of timely  reminders at 45 and 15 days prior to the recommended calibration date of your instrument     Mesa recommends annual service and calibration of your Proven DryCal primary gas flow standard as a periodic quality  assurance measure  as well as to provide you and your organization with a defensible audit trail of premier quality     About Your ML 800    The ML 800 is a positive displacement primary piston prover for gas flow measurements in either pressure or  vacuum applications  Using Proven DryCal Technology  it combines the accuracy of a primary standard with  unequaled speed and convenience  Volumetric or standardized flow readings are obtained with the push of a button   The ML 800 can be set to take flow readings manually   one reading at a time   or automatically in the continuous  auto read mode  The ML 800 can be programmed for up to 100 readings in an averaging sequence     Operation    The versatile ML 800
22. sdev NE ER sa ma yaqan usus Akasia aaoi a Eaa aaa aE aa BE aaea EINE EETLUTDERNEERRTE DDP ERASERTDEENUTRNEERESEREOEFETE FREUEN FERDR 2  Unpacking Checklist  ka ner a er m q imaqa LEERE e ua a ea tothe 3  Warnings a eu EEE RREESTEERCEREBLLLEDELTLELTLELTER DE E STH PLEPETEREREOELEUERITELEPEL e E EHEN A E 3  Operating Your ML 800 uuu a nee a ana aan pi daaa pidio En sas Naaa 4  Getting Started aseisiin runna inaani nn a En En nd ana HE nn nn ka was ayawqwa wasaka waka 4  Turning  the  ML 800 On     Off   uu ea ana a na a nn nn a a a 4  LED pacos m S sm aasan Su nis aS am ini A E a au 4   ow Battery Imdicatepu u L een ra a Ea aa an nn a a ah Tec bet ine ernennen 4  Using the MIS800 Keypadi gur anans ead tecavaaciiendsnntcnssedanaa Eaa aE aa aaa Eaa a aiaa Ead Epia ENa 5  General Menu NAVIGATION  kurus s s aae S a a vache Ea Sus S Si u ENARA Era CEEE aE wasa 5  Keypad  FUuMCtiOn sau unun SSS a a E a sa a ce iasnuamesatvoncacuads afoucecscaues 5  MI 800  Menu Trees    een 6  Error Messages sirgena o NEENEFREREERSTETHELTENESDLLSETLELFLETLLDRFERETETELELFEERELHESCE a E TEE E A Ty 7  Inst  lling 8 amp  Removing Flow Cells sorrisi riena enden 8  Factory Default Ste GS cassia usasapa E aaa aa a ATE usqay akapa EAE aa AR en 8  Connecting the  ML 800 toa Gas Flow SOURCE  uqa ana aaaea ia aiiidh 8  Taking ReadiNdgs saa unapuy a a aa aaa aaa a aa aa aaa aaa aaa aa ee 9  Setting User Preferences una aa naaa a Ae aE EaD a E aaa aa He a Da Naa aaa 9  Using Sensor FACTORS sisses yea aa Be HE aa 
23. se the Up Down arrows to navigate between different lines    e Use the Enter button to select a field to be modified depending on the function and to lock in any changes   Use either the keypad or the Up Down arrows to modify the desired field    e Use the Fwd and Back keys to advance to the following menu or return to the previous menu     Error Messages       Cell is not present or unit does not    see    cell  Piston did not return to bottom of cell properly   To exit  press Back  or wait to return to    Main Menu     To exit  press any key     Installing  amp  Removing Flow Cells    The ML 800 accepts interchangeable ML 800  only  flow cells for different flow ranges  If the user attempts to  enter the  Run Menu    prior to installing a flow cell  the unit indicates    No Cell    and returns to the    Main Menu    after  a 5 second delay     Installing Flow Cells  1  Position the selected flow cell into the base opening  its top label facing you     2  Turn the cell back and forth to locate the guide pins  when the guide pins are engaged  press down   3  When powered on  the ML 800 senses which cell is installed and displays the appropriate units for that cell     Removing Flow Cells  Grasp the flow cell firmly  hold the base in place  and lift upwards     Factory Default Settings    The ML 800 has a number of user definable features and settings  To return to factory default settings at any time   press Reset followed by Save from the    Main Menu                        
24. set average is reached and then restart the cycle   If the Reading type has been set to    Burst    in    Setup Menu   2     the unit will run until the preset average is  reached and then stop     e Read Button  Initiates a single reading  must be accessed from the    Run Menu         ML 800 Menu Tree    Main Menu ein  Run  Setur  Leak Test  4  I  I  Setur Menu     1  Sensor Factor  1  BBB  Cal  Type Sid   4  I  I  Setur Menu     2  Reading Tore Cort    in Bug if  Min Reading Be  4  I  I  Setur Menu       Pres Units mmHg  Teme Units  C  Teme Corre EDS   C  4  I  I  Setur Menu     4  Ar 18 68 65 03 AM    Bat 6 99 Volts  4  I  I  Setur Henu     5  Date   MT DT YR    Time     ANSP  Auto off Enable    4    Press Fwd arrow  after    Menu   6    to       I  I  I return to   Main Menu     I    Setur Menu     6  PTU 2 188 com  FTUH 1  BBB    e Use the Up Down arrows to navigate between different lines   e Use the Enter button to select a field to be modified depending on the function and to lock in any changes     e Use either the keypad or the Up Down arrows to modify the desired field      amp  m m m m m om m m m m m am am am am am am am a s s s                                             Forward navigation        gt    Backward navigation    e Use the Fwd and Back keys to advance to the following menu or return to the previous menu  Q        ML 800 Menu Tree           Forward navigation       p   Backward navigation       To exit at any time  press Reset       m m m m         e U
25. this menu  use the Fwd and Back arrows to select each option  such as Month or Day  To change a selected  option  use the Up Down arrows  Date and time formats are specified using Setup Menu 5  Date and Time Formats     Option 1  Date  This option allows you to set the date     Option 2  Time  This option allows you to set the time     Setup Menu 5  Date  amp  Time Formats       Setur Menu     5  Date   MT DT YR  Time     AMPH  Auto off Enable    Option 1     Date     Toggles between    MT DT YR    or    DT MT YR    date format     Option 2     Time     Toggles between    AM PM    or  24 Hr    time format     Option 3     Auto Off     Toggles between    Enable    or    Disable    65 minute auto shut off     Setup Menu 6  Piston Tare Value  PTV  and Piston Tare Value Multiplier  PTVM        PTV stands for Piston Tare Value  this is the amount of gas that passes around the piston during measurement  All Mesa  calibration equipment has a factory set Piston Tare Value that is stored in the memory of the Proven DryCal cell  PTV  stands for Piston Tare Value  this is the amount of gas that passes around the piston during measurement  The value is  typically very small  0 008 ccm for the ultra low flow cells  0 1 ccm for low flow cells  0 2 ccm for medium flow cells and  1 4 ccm for the high flow cells  We adjust for this leakage by adding the Piston Tare Value  PTV  to the measurements     On our highest accuracy instruments we allow for the adjustment of the Piston Tare Value with
26. uring the test  the flow point or points that you re  checking and an explanation of the issue you   re experiencing     Recertification    Your ML 800 primary piston prover is a precision measuring standard comprised of moving parts that are machined  to extremely close tolerances  Additionally  various environmental factors  product wear  drift of the temperature  sensors and pressure transducers  or inadvertent damage may adversely affect your ML 800   s measurement accuracy or  general performance     For these reasons  Mesa highly recommends having your ML 800 annually verified by our ISO 17025 accredited  laboratory in order to ensure its measurement integrity  For those applications subject to regulatory or ISO require   ments  verification by our accredited laboratory provides you with a defensible audit trail of the highest quality     As the ultimate quality assurance measure  as well as to keep your ML 800 in top condition and updated with the latest  hardware and firmware upgrades  as available   Mesa offers our elective Recertification program  Recertification is an  intensive service and calibration package that provides pre  and post dynamic flow comparisons against our Proven  DryCal Technology lab standard  complete product refurbishment   including cleaning of the piston and glass assembly   installation of any available upgrades and other routine and preventative maintenance items  and full dimensional cal   ibration  comprised of 20 or more tests using pre
    