Home
        HG-DOx User Manual
         Contents
1.         Flow Flow  Control Outputs witch Meter   internal   External  Flow  Switch    3 2 2 Input Switches    Flow input switch terminal blocks on the I O module allow for three optional  input switches to be connected to the system to ensure that there is flow in  the system and or to the analyzer     Two flow switches and one flow meter may be connected     e Flow Switch  internal   Flow switch connected to flow cell   o If no flow switch is installed  a jumper must be installed for the  analyzer to operate properly   e External Flow Switch   external off    Connection for external 2 wire  flow switch  If an external switch is not connected  a jumper must be  installed for the analyzer to operate properly     e Flow Meter  Connection for 2 or 3 wire flow meter where   o White   VCC  o Green  IN  o Brown   Ground    Note  Electrical connections in this section are ONLY recommendations  All    electrical connections should comply with National Electrical Code  NEC  and  all local regulations     8    3 3 Relays    Relay 1  J2  will activate a blower or other air system based on the DO set   point     Relay 2  J6  will active the automatic cleaning cycle     3 4 Installing Probes  Probe 3 4 1 Dissolved Oxygen Probe    Opening    The Dissolved Oxygen  DO  probe is supplied    S from the factory with the following components   MALENA e DO Probe with 7m  21 ft  of cable  ia Flow Assembly  optional     e  e Replacement Membranes  e Abrasive Paper       Cage         Outlet  lt     
2.      2     aew    oND    9     Remove the probe or sensor  clean with a dry cloth and place in the  standard solution   Place the probe or sensor in the standard solution and wait for the  reading to stabilize    Press Menu until    ___ Calibrated to    appears in the LCD display    Press OK    Enter the password  Press the up arrow or down arrow until the  password is reached    Press OK    Press OK again    Press the up arrow or down arrow until the value is the same as the  standard solution    Press OK to save the new calibration or Esc to abort without saving     10  Press Esc to return to the main display     i  pH Probe     gt              Grounding Probe     amp     pH Buffer  Solution       17    O Blue I  Water Technologies    6 Maintenance    6 1 DO Probe Maintenance    Probe maintenance will be required periodically and will depend primarily on  water conditions  Before performing any maintenance  turn OFF the power  and water to the probe     6 1 1 Cleaning the sensor    Always clean the sensor before calibration and whenever there is slow  response time to changes or errors in measurement     Depending on the soiling of the probe  proceed as follows   Salt Deposits   Soak in 1 5  HCL solution for several minutes   Rinse with plenty of water   Dirt on Sensor Body   Clean with water and brush or cloth   Dirt on Membrane   Clean with water and soft sponge    6 1 2 Regenerating the Sensor    Cleaning the Gold Cathode    The Gold Cathode needs to be cleaned when visi
3.    2   73   811    6   43   1242     2   81   7     7    amp    1211  _     70   8   4   1181     2   84   769    g   48   1153     30   8   755    10   50   1125     3   88   742    11   52   1099     32   90   730    12    54   1075     3   gi   718    3   5   1051     34   93   706    mh ae ee ae  15    69   1006     3   97   68    16   61   98     3   99   672    17   63   964     38   100   661    18   64   945     39   102   651    19   6   926     40   104   641    20   68   908                       6  Press Menu until    Oxygen in Air Calibrated to    appears in the LCD  display     The top line will display    Oxygen in Air Calibrated to    and a number  The  number displayed is the last value someone entered for the calibration  The  bottom line will display    Sensor was    and a number  This number is the  sensor reading without any calibration at the time of the last calibration  If  there is a large discrepancy between these two numbers  the sensor was  calibrated improperly or the probe may require maintenance  The value  displayed normally on the main screen and the value the analyzer uses to  determine dosing rates is the calibrated value     7  Press OK    8  Enter the password  Press the up arrow or down arrow until the  password is reached    9  Press OK    10 Press OK again     The display will now show    Cal  Oxygen in Air to    on the top line and    Sensor  Reading    on the bottom line  The    Sensor Reading    is the current reading of  the sensor 
4.   4 2 Configuring Internal 4 20mA Outputs                cceeeeeeeeeeeeeeeeeeetneeeeeeees 13   D AO TATIOME SR nn 14  51 DO Calibrator ee io n aaae aaa aaa aa aa a e a aaa i 14  5 1 1 Air AMBP ADO  setcstesicentesecencnerieneneeesenepaceinereedeexesacdeannemededeeueadeecetions 14  5 1 2 Reference Calibration                                                 16   5 2 Calibrating other Sensors and Meters            ssssssssneneeeeseessrrnnnreeseerenne 17   0  Maintenant ieee dete jcbacunedemeceesdu caste usenecteiden actescececteidevedtesdasececte  18  6 1 DO Probe Maintenance sind Mutereverend Mess 18  6 1 1 Cleaning the sensor              cceeeeeeeeeeceeeeeeeeeeeeeeeaeeeeeeeeeeseeeeeeneeees 18  6 1 2 Regenerating the Sensor                                                            18  6 1 3 Replacing El  CtoMte   sn dnhen  18   2         TROUBNG SO OUING nest 19   3    O    Blue I    Water Technologies       1 General Safety Precautions    This section presents important information intended to ensure  safe and effective use of this product     Read the following carefully before handling the product  These  warnings and cautions must be followed carefully to avoid injury to  yourself or damage to equipment     Warning  Only properly trained and licensed  electricians should attempt to wire or service the  electronic components of the analyzer controller     There is an Electrical Shock Hazard when servicing  this system     Always verify that all electrical power source s  a
5.   To enter the Technician Menu    1  Press Menu to enter the operator menu then press UP   DOWN together  until the display changes to menu   51    2  Navigate the menus exactly the same as the operator menus  but the  technician password is the only password accepted     4 2 Configuring Internal 4 20mA Outputs    The menu for internal 4 20mA settings is found in the technician menu     Setting the 4 20mA outputs    1  Enter the technician menu and scroll until    4 20 Settings    appears in the  LCD   2  Enter the technician password and press OK    3  Select the Output Channel  1 to 2 Built In or 1 to 4 on NTU 4 20 card    4  Select the output Parameter  DO  pH  etc     5  Select the Value for 4mA Output   6  Select the Value for 20mA Output   7  Test the output  with multi meter set on mA using the test output settings  will help adjusting the external unit  PLC     8  Repeat the above steps for the ieee tan Channel   Variable Min Value   Max Value  outputs  The table on the right outlines 1  some example settings  Note that a    2    single variable may be the output of more 3  than one channel  a   Temp   0   500         oO        Set the 4 20mA alarm output  a  2mA  4mA  20mA  or hold last value    13 Q Blue     Water Technologies    The 4 20 alarm output is the output value that will be sent in case of a  problem with the HYDROGUARD that does not have flow or cannot  perform a test  e g  stuck piston or unclean cell   In case of low or high  level  i e  low chlorine   
6.   est le bon fusible qui est en place     Each relay connection is limited to 4 amps  to prevent  overheating  The relays may show a higher rating but  do not connect equipment exceeding 4 amps     All electrical connections should comply with National  Electrical Code  NEC  and all local regulations     Caution  Do not use chemicals that reduce the  surface tension  When using hydrochloric acid   observe all safety regulations     Electrodes    Warning  Do not swallow the electrolyte  Avoid  electrolyte contact with skin or eyes  In case of  accidental contact  wash with a lot of cold water  In  case of eye inflammation  contact a doctor  immediately  Wear safety glasses and gloves when  working with the electrolyte solution     Caution  Do not touch or damage the electrodes  The  electrolyte is sensitive to oxidation  Always keep the  electrolyte bottle closed after use  Do not transfer the  electrolyte to other containers  The electrolyte should  not be stored for more than one year and should be  clear  not yellow  in appearance  for use by date  see  label   Avoid forming air bubbles when pouring the  electrolyte into the measuring chamber     Caution  HYDROGUARD s control board unit should  not be opened except for initial installation and  troubleshooting  and should only be opened by a  trained and approved technician     Chaque connexion relais est limit  e a 4 Amp  afin  d   viter toute surchauffe  M  me si les relais affichent    ventuellement une valeur sup  rieur
7.  bowel     Installation Using Flow Cell  optional     1  Mount the DO flow assembly such that the probe opening is facing  between horizontal and vertical up  never facing down   see diagram   2  Thread the Probe into the opening of the flow cell   3  Connect the Inlet and Outlet Flow to the Flow Assembly  a  Connect the Inlet to the Bottom      fitting  b  Connect the Outlet to the Top      fitting  c  Max Pressure   10 bar  150 psi   d  Min Flow Rate   30 L hr  0 13 GPM        Installation without Flow Cell  immersion in water     1  Mount the piping and auto cleaning assembly as describe in separated  manual   immersion assembly manual     2  Secure the immersion assembly with mounting bracket  2 in figure A   station    9  OBluel       3  Submerge the probe into the sample ambient and secure the probe or  probe cable to keep the probe in the desired location with the probe  facing down  Make sure the cleaning assembly as describe below  photo   4  Immersion in water should be 40 100 cm below surface to ensure  correct reading    El                Figure A    e Connect the auto cleaning assembly to clean water supply  1 in figure  A    e Use an electric valve to activate the auto cleaning  Command for  cleaning is used by relay 2  J6 on the I O board    e Set in the cleaning interval in the main menu       10 OBluel                   Probe Wiring  7  Route the DO cable through an open s  gland in the bottom of the electronics i i  enclosure  FO    8  The cable may be exten
8.  dependent power supply  interlocked power supply      3 2 1 Connecting the Main Electrical Power    The Main Power Supply may be connected to either 90 120 or 190 240VAC  50 60Hz  Switching between voltages is accomplished by changing two  2   jumpers located above the main power connection  to the left of the  transformer  For 90 120VAC  a 1amp fuse should be use  for 190 240VAC  a  0 5amp fuse should be used  These changes must be completed prior to  wiring    Caution  Before making a connection to a power source  confirm that both  the J21 and J22 jumpers are located on the correct voltage and that the  appropriate fuse is in place  1 0A for 110V and 0 5A for 220V      1         Verify that the power switch or circuit breaker to the non dependent power  source is off    Connect the line  live  wire to the I O board connector marked Line   Connect the neutral wire to the I O board connector marked Neutral   Connect the earth wire to the I O Module connector marked Ground   Continue with the other electrical connections    Turn on electrical power only after all electrical connections have been  completed     RC RAR    O    Blue I    Water Technologies             jumpers for voltage Chipset I  25   selection  J21   J22                                                           A p      i  J16 J2 J6 J11 J17 J7 J3 J4 J5 J8    BLW   CLN   PH    NTU   Alarm   Temp   12 3 TAA  C NC NO  C NC N C NC NO   C NC NO  C NC NOC NC NO S    amp    S   2     BLW LN NTU Alarm Temp  Piy   bd
9. OBluel    HYDROGUARD HG DOx  Water Quality Analyzer       Per RS    Lei    7       User Manual    Version 5 3    No part of this publication may be reproduced  transmitted  transcribed  stored in a retrieval  system  or translated into any language or any computer language  in any form or by any third  party  without the prior written permission of Blue   Water Technologies Ltd     Disclaimer    Blue   Water Technologies Ltd  does not accept any responsibility for any damage caused to  its products by unauthorized personnel  Use of non Blue   Water Technologies    reagents  and or replacement parts will void all warranties     Trademark Acknowledgements    HYDROGUARD is the registered trademark of Blue   Water Technologies Ltd     Copyright    2014 by Blue   Water Technologies Ltd     O    Blue I    Water Technologies    Table of Contents    1 General Safety Precautions samsung 4  2  Preface n aae a a a a a aaa ia 6  2 1 Measurements and Features    6  2 2 System Components ss 6     NS ANA en idad nid iaaiiai aa 7  3 1 Working Environment 2 528 inner 7  3 2 Electrical Requirements and Installation    7  3 2 1 Connecting the Main Electrical Power    7  3 2 2 Input SMIC  ea een est een ren ent 8  BIOS nn eme 9  34  Installing Probes sisis entre EEE 9  3 4 1 Dissolved Oxygen Probe    9  Probe Wiring ns en eme ben naine 11   DR tra nas re ne 11   4 First Time Operation and Calibration                c   ceceeeeeeeeeeeeeeeeeeeeeeeeeeeeeees 12  4 1 Menus and Settings sn nn do ee 12
10. bly soiled or if there is a  coating of silver     Unscrew the membrane cap from the membrane body  Carefully clean the gold cathode in 2 steps using the supplied abrasive paper   Use the Green paper first then use the yellow paper until the soil or silver  coating is removed  Clean the electrode with tap or distilled water  Fill the membrane cap with new electrolyte solution and re attach to the  membrane body    6 1 3 Replacing Electrolyte    Electrolyte is consumed during normal operation and requires replacement  periodically     Remove the membrane cap  1  Replace the electrolyte  and if damaged the membrane cap    18  OBluel       2  Fill the complete contents of an electrolyte ampoule into the membrane  cap   3  Remove air bubbles by tapping the side of the membrane cap   4  Hold the sensor body at_an angle and carefully screw the membrane cap  down to the stop     5  Place the membrane cap on the body and screw the cap to the stop     CAUTION  The electrolyte solution is strongly alkaline and proper care and  personal protective equipment should be used when handling the  electrolyte     7 Troubleshooting    The table below outlines very basic troubleshooting  Refer to the latest  documents and technical notes  available for download at    www blueitechnologies com  For more information or to contact your Blue    Technologies    representative     Before troubleshooting a problem    1  Perform a System Reset  last menu of operator menu    2  Check that all flat cable 
11. connections between electronic cards are secure  3  Check that all chipsets on electronic cards are secure and no pins are bent    Problem   Symptoms Potential Cause Solution   Suggestion    Connection between boards is loose Check all connections between boards       Check wiringing    compare to wiring  diagram    Check that no pins on the chipset are  Chipset is loose or pin bent bent and it is in completely and  correctl    Check and Replace fuse above main  Fuse is blown  Power Suppl    Check for burn marks on O board   VO board has been damaged due to improper Recheck for proper wiring    confirm all  electrical installation neutrals and grounds in contact are from  the sa    me power suppl  Sensor Calibrated too Low Recalibrate Sensor  Flow too Low Increase Flow  Coating on Membrane Clean or Replace Membrane    Electroylte Exhausted Replace Electoltye    Display not functioning or Improper wiring or bad connection  is displaying odd numbers    Analyzer won t power up    DO Measurement is Low          Sesnor Calibrated too High Recalibrate Sensor    Damaged Membrane Replace Membrane  Gold Cathod Dirty Clean Cathode      5  Anode Coating Dissolved Anode Silver oe Brown  Replace  Unstable pH or ORP   Check  PT100  I O board ground and CI  A Poor Grounding i    Readings and pH wire grounding    DO Measurement is High       a Q Blue        No part of this publication may be reproduced  transmitted  transcribed  stored  in a retrieval system  or translated into any language 
12. ded up to 100m       330 ft  with appropriate cables and i    connectors  contact your Blue   KOIN  Technologies representative for more me pH  Rx  Do  amp  Tempr  Card  details   9  Connect the Probe Cable to the DO ana  electronics card  me eo  e GreenWire  CL A j  E     e White  thin  Wire   CL K i  e Gray  thick  Wire   CL GND ue  e Brown Wire   NTC GND    ee  e Red Wire   NTC IN  pH  The pH sensor is emerged with the same immersion assembly  Probe wiring  1  Route the pH cable through an open gland in the bottom of the  electronics enclosure   2  Connect the Probe Cable to the pH electronics card at pH    and pH     which is the shield  OBluel       4 First Time Operation and Calibration    4 1 Menus and Settings    HYDROGUARD has two menu levels  Operator and Technician  The  Operator menu includes settings that may be controlled by on site operators   The Technician menu includes settings and calibrations that should be  restricted to specially trained HYDROGUARD maintenance technicians  Each  menu has a separate password  The technician level password may be used  whenever a password is required  however the operator password will only be  accepted in the operator menu     Measurement 3 54 DO 7 62 pH  Display  0 23 NTU 498 uS  Menus and  Alarm Display    Buttons for  Menu  Navigation       Note  The default Operator Password is  123 and the default Technician  Password is  456     Caution  DO NOT Forget your password  There is no way to reset the  technician password 
13. e  ils ne se  connecteront pas    un   l  ment d  passant 4 Amp     Tous les branchements   lectriques doivent   tre  conformes au Code Electrique National  NEC      National Electrical Code  ainsi qu      toutes les  consignes locales     Attention  N utilisez pas de produits chimiques  susceptibles de r  duire la tension superficielle  Lors  de l   utilisation d acide chlorhydrique  appliquez  scrupuleusement toutes les consignes pertinentes     Les   lectrodes    Attention  N   avalez pas de substance   lectrolyte   Evitez tout contact de I     lectrolyte avec la peau ou les  yeux  En cas de contact accidentel avec cette  substance  rincez abondamment    l   eau froide  En  cas d   inflammation oculaire  consultez imm  diatement  un m  decin  Portez des lunettes et des gants de  protection lors de la manipulation de la solution    lectrolyte     Attention  Ne touchez pas ni n   abimez les   lectrodes   L     lectrolyte est sensible    l oxydation  Maintenez la  bouteille contenant l   lectrolyte toujours ferm  e apr  s  utilisation  Ne transvasez pas l     lectrolyte dans  d   autres r  cipients  L     lectrolyte ne doit pas   tre  conserv   plus d   un an et doit garder une apparence  claire  pas jaun  tre   pour la p  riode d utilisation  voir  l   tiquette   Evitez la formation de bulles d   air en  versant la solution   lectrolyte dans le compartiment  de dosage     Attention  Le tableau de commandes de  l HYDROGUARD ne doit en aucun cas   tre ouvert si  ce n   est lor
14. lectrique est ou sont bien  d  connect  e s  avant d ouvrir l unit   ou  d entreprendre toute op  ration de service technique  et tout branchement des composants   lectroniques     Attention  Il y a lieu d   agir avec une extr  me prudence  lors de l installation  de la mise en   uvre et de la  maintenance l HYDROGUARD    Seuls des  techniciens d  ment form  s    cet effet sont autoris  s     effectuer l installation et la maintenance de  l   analyseur du contr  leur     Seuls des op  rateurs qualifi  s ayant re  u la formation  ad  quate sont habilit  s    modifier les dosages des  produits chimiques utilis  s     Conformez vous sans exception aux consignes  locales de sant   et de s  curit   lorsque vous effectuez  toute op  ration technique sur l   analyseur le  contr  leur  ou lorsque vous modifiez les param  tres  de dosages chimiques     L alimentation g  n  rale peut   tre branch  e sur 110   120 ou sur 220 240VAC 50 60Hz  Pour basculer  d une tension    l   autre  il suffit de changer les deux  2   cavaliers situ  s au dessus de la principale connexion    lectrique     gauche du transformateur  Une tension     110 120VAC requiert un fusible de 1 Amp    une  tension    220 240VAC requiert un fusible de 0 5 Amp   Ces modifications doivent   tre accomplies avant le  branchement   lectrique     Pr  cautions  Avant de relier l appareil    une  quelconque alimentation   lectrique  v  rifiez que les  deux cavaliers sont situ  s sur les valeurs correctes de  tension et que c 
15. no 4 20mA alarm will be activated     5 Calibration  5 1 DO Calibration    The oxygen calibration has 2 options  air calibration and reference calibration   The air calibration is a direct calibration of the DO sensor and the reference  calibration is a comparative calibration method     5 1 1           Air Calibration      Remove the sensor from the flow cell or water sample    Clean the sensor body with a damp cloth and dry the membrane with a  soft dry cloth  If the membrane was under pressure  greater than atmospheric  pressure   you must equilibrate the pressure before calibrating    a  Remove the membrane cap   b  Refill the electrolyte solution and replace the membrane cap   c  Wait for the probe to re polarize   lt  60 minutes       Wait approximately 40 minutes for the probe to adjust to the ambient    air temperature and the measured value to stabilize   make sure to  keep the probe away from direct sunlight       Determine the Calibrate to Value using the following Formula     Calibrate _to _ Value   Sx K x Lx M   where     S   Saturation Value  see Table    K   Altitude Factor  see Table    L   Relative Air Pressure at Calibration in hPA   use 1013 hPa if unknown    M   1 02 in air or 1 00 in air saturated water    14    Temp r Temp S factor Altitude K factor  ec   F    mg        C     F    mg      o   32   1464     21   70   8     a a aa ee   ara    2        38    3   73   857    3   37   1345     24   75   841    4   3   1309    23   77   8235    5   41   1275  
16. or any computer  language  in any form or by any third party  without the prior written  permission of Blue   Water Technologies Ltd     Trademarks and Patents  HYDROGUARD is a registered trademark of Blue   Water Technologies Ltd   Patents issued and pending at the time of this printing     Disclaimer  Blue   Water Technologies Ltd  does not accept any responsibility for any  damage caused to its products by unauthorized personnel  Use of non Blue    Water Technologies    reagents and or replacement parts will void all  warranties              18 Ha melacha St   Rosh Ha ayin 4809148  Israel   Tel  972 9 7680004   Fax 972 9 7652331  www blueitechnologies com    20  OBluel       
17. re off  before opening the analyzer controller unit or  attempting to service electronic components or wiring     Caution  Extreme caution should be used when  installing  operating  and maintaining the  HYDROGUARD   Analyzer  Only properly trained  technicians are authorized to install and maintain the  analyzer controller     Only properly trained and licensed operators should  attempt to make any changes to chemical dosing  levels     Always follow local health and safety regulations  when performing any service on the  analyzer controller unit or when changing chemical  dosing settings     The main power supply may be connected to either  110 120 or 220 240VAC 50 60Hz  Switching between  voltages is accomplished by changing two  2  jumpers  located above the main power connection  to the left  of the transformer  For 110 120VAC  a 1amp fuse  should be use  for 220 240VAC  a 0 5amp fuse  should be used  These changes must be completed  prior to wiring     Caution  Before connecting to a power source   confirm that both jumpers are located on the correct  voltage and that the appropriate fuse is in place     Attention  Seuls des   lectriciens qualifi  s ayant re  u  la formation ad  quate peuvent entreprendre le  branchement  l entretien ou la r  paration des  composants   lectroniques de l   analyseur du  contr  leur     Il existe un risque de choc   lectrique lors de  l entretien de ce syst  me     Ayez soin de toujours v  rifier que la ou les source s   d alimentation   
18. s de l installation initiale et en cas de  d  pannage     auquel cas son ouverture ne doit   tre  effectu  e que par un technicien ayant re  u la  formation ad  quate et d  ment habilit       2 Preface    2 1 Measurements and Features    The HYDROGUARD HG DOx can be configured to measure any combination  of the following water quality parameters     Dox  Dox  Dox  Dox    Dissolved oxygen  Dissolved oxygen  Dissolved oxygen  Dissolved oxygen      pH    ORP    pH   ORP    mam     YS as Ss n    Several communication options are also available   e Internal 4 to 20 mA Outputs  up to 6 channels programmable   e HydroSoft   Direct Connection Software  e WaterGuard OL     Wireless Communication Package    2 2 System Components    The unit includes HG 602 main platform  Electrode and cable  All the additional partS in the drawing are optional and could be purchase  separately       6 O Blue I  Water Technologies       3 Installation    3 1 Working Environment    Pollution Degree  2   Installation Category  2   Altitude  2 000 m   Humidity  1 to 90  non condensing   Electrical Supply  100 115Vac  1 0A or 200 230Vac  0 5A  50 60Hz  Temperature  5  C to 45  C    3 2 Electrical Requirements and Installation    HYDROGUARD requires a 90 120 or 190 240 VAC  50 60 Hz electrical power  source  The main HYDROGUARD power supply should be connected to a  non dependent power supply  so that the unit remains powered constantly   Any relays used to directly activate equipment should be powered by a 
19. with no calibration  The    Cal  Oxygen in Air to    value is the new  value which you want to set     11 Press the up arrow or down arrow until the value is the same as the  value given by the digital photometer    12 Press OK to save the new calibration or Esc to abort without saving    13 Press Esc to return to the main display     15 OBluel    5 1 2 Reference Calibration    1  Test the water sample using an external digital Dissolved Oxygen tester   2  Press Menu until    Oxy  Reference Calibrate to    appears in the LCD  display    Press OK    Enter the password  Press the up arrow or down arrow until the password  is reached    Press OK    Press OK again    Press the up arrow or down arrow until the value is the same as the value  given by the external tester   8  Press OK to save the new calibration or Esc to abort without saving    9  Press Esc to return to the main display     3  4     gt  no    5  6  7    x n n    O    Blue I    Water Technologies       5 2 Calibrating other Sensors and Meters    Calibration of other sensors and meters can be done using reliable external  testing device or standard solutions  When using an external testing device   make sure to take the sample water from the same supply to that probe or  sensor     Using External Testing Devices    Follow Test Method for Reference Calibration    Using Standard Solutions    Note  for pH  or ORP  calibration  the temperature probe must also be placed  in the standard for the reading to stabilize     1
20. without a complete reprogramming of the  HYDROGUARD System     Each of the parameters in the operator menu is configured in the same way   The following procedure describes how to configure a typical setting     1  Locate the desired parameter in the menu   a  Press Menu until the desired parameter name appears in the LCD  display   2  Press OK  Enter Password 100 appears in the LCD display   3  Enter the Operator password  or technician password  both are accepted   4  Press the up arrow or down arrow until the password number is reached     Holding Menu why pressing up or down will advance the first digit  Holding up    or down for an extended period of time will proceed through the numbers  more quickly     12    O    Blue I    Water Technologies       5  Press OK to accept the password  The parameter name and current  setting appear in the LCD display    6  Press OK  again  The LCD display shows the parameter and the current  setting    7  Enter the new parameter setting    b  Press the up arrow or down arrow until the desired value is reached    c  The second row of the menu display  below the value that is being  changed  shows the current value    Press OK to save the new setting or Esc to abort without saving the new   setting     2    To change the settings of additional parameters  press Menu  until the desired parameter appears in the LCD display and  repeat steps 6 8 above to set the new parameter  The table  on the right outlines an example of menu settings      
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Samsung ProXpress C2620DW  Barvni      Copyright © All rights reserved. 
   Failed to retrieve file