Home
        TT24-7 User Manual
         Contents
1.      If the split function is configured  configuration e3  this  rear  valve has four  points of connection  i e  to the T piece connecting the two front valves  to the  tube desorption line  to the split effluent line and finally to a transfer line leading  to the capillary column     See section 2 5 for an overview flow schematic of this 3 heated valve  configuration     Figure 8 shows the schematic of this rear valve     QUI 1016 V2 2 Aug 06  22     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International       Figure 8  Schematic of third  rear  valve    Figure 9 shows the relative position of these three valves valve within the TT24 7   It also shows the location of the heated transfer line connection port into the  TT24 7  the twin trap box assembly  the pneumatics module  and the tube  desorption assembly    Rear heated valve Front heated valves Trap box assembly Pneumatics module    pnly on e2   e3  onfigurations    Tube desorption  assembly    Figure 9  Position of the principle system components of the TT24 7       2 4 2 Pneumatics module    The pneumatics module is a single sub assembly located at the front of the TT24 7  as seen in Figure 9  It is locked in position by a single M4 retaining screw   Removal of a trap from its location in the valve box requires the pneumatics  assembly to be removed  See section 3 4 for complete details of how to remove    replace the traps     The modu
2.      PC to GC Communications cable  eg LAN   RS232     Agilent  UTD 5098    Remote Start Cable P N    Thermo  UTD 5108       Figure 69  Direct serial connection  Figure 70 shows a similar schematic but in this case USB communication is used     QUI 1016 V2 2 Aug 06  95     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    USB Hub  USB to Serial cables          P N U USBSR    Standard  USB Cable    Serial Cables Remote Start   P N 2 0189     PC to GC Communications cable  eg LAN      Agilent  UTD 5038  Remote Start Cable P N       Thermo  UTD 5108    Figure 70  USB connection    It is possible to have a combined configuration in which one of the serial  communication cables is connected directly to the PC from the TT24 7  and the  other connects via USB     3 1 4 Power recycle box    Remote system control of the TT24 7 and associated GC or GCMS can be very  useful  Where access to the system is difficult e g  remote location  high security   worker exposure safety etc  the ability to run and monitor the respective  instruments remotely is very advantageous  This is particularly evident where the  continuous sampling mode is being used     Remote PC to PC control is achieved using commercially available software  packages  an example of this is VPN      A specific use of this is in the case of power failure  Power failures will close down  both the TT24 7 and the GC system  however the con
3.    11  Spherocarb      UniCarb     60   80 mesh only    12  Carbosieve SIII     60   80 mesh only    13  Carboxen 1000     range of mesh sizes    14  Molecular Sieve 5A   15  Molecular Sieve 13X    Retention Volumes and Safe Sampling Volumes    Table of Contents   Figures    TT24 7el1 basic system configured only for continuous sampling  TT24 7e2   standard system configured with the following additional features   TT24 7e3   Figure 1  TT24 7 User Interface   showing both sampling modes  Figure 2  Sample gas inlet on TT24 7    Figure 3  Schematic representation of TT24 7 sequence in continuous  operation    Figure 4  TT24 7e2 showing tube accessory fitted and sorbent tube  ready to be sealed into position    Figure 5  TT24 7e2 showing tube accessory fitted and sorbent tube  sealed into flow path    Figure 6  TT24 7e3 showing the tube desorb and split accessories fitted  Figure 7  Schematic showing the two TT24 7 trap valves   Figure 8  Schematic of third  rear  valve   Figure 9  Position of the principle system components of the TT24 7  Figure 10  Position of Trap A and Trap B within the trap housing    Figure 11  TT24 7 cold trap   note the restricted inlet   outlet end and the  quartz collar on the unrestricted end     Figure 12  Transfer line connection between TT24 7 and GC system    QUI 1016 V2 2 Aug 06  7     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    129  130  130  130  131  131  132  132  133  133  134  134  135   
4.    Desorption Tube    MFC  amp  Pump Sample in    Trap B       KEY           Flowof Carrier Gas  Carrier Gas  no flow       a    Flowof Sarple Gas  Sarrple Gas  no flow        Figure 24  TT24 7e2   Continuous sampling   trap A desorbing  trap B sampling    Trap A now fires  If this is the first run for A this run is treated as a clean up  run  however if this is a subsequent run then real sample will be transferred  across to the capillary column for analysis     To see how the time overlap of sampling  dry purging  trap fire and cooling  occurs relative to the GC oven cycle time please refer to section 2 3 1     QUI 1016 V2 2 Aug 06  41     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 2 2 Tube desorb mode  TT24 7e2   2 5 2 2 1 Standby    Desorption Tube       MFC  amp  Pump Sample in    Trap B       KEY          Flow of Carrier Gas  Carrier Gas  no flaw          gt     Flow of Sample Gas  Sample Gas  no flow        Figure 25  TT24 7e2   Tube desorb mode   standby    In this non sampling mode  carrier gas alone is continuously supplied to the  capillary column  No other gas flow occurs within the TT24 7 unit during this  time     QUI 1016 V2 2 Aug 06  42     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 2 2 2 Leak testing tube    MFC  amp  Pump Sample in             KEY          Flo
5.    Sorbent Strength  Medium Strong  Specific Surface Area  m2 g    560  Approximate analyte volatility range  n C3 4 to n Cg   Boiling point 50  C to 150  C  Example Analytes  Light hydrocarbons  Sorbent Maximum Temperature   gt 400  C   350  C to 400  C  350  C to 400  C    Notes  Hydrophobic    Recommended Conditioning Temperature   Recommended Desorption Temperature     Low artefacts   lt 0 1 ng   Some activity with labile compounds  Friable    6  Chromosorb 102     range of mesh sizes available    Sorbent Strength  Medium   Specific Surface Area  m2 g    350   Boiling point 50  C to 200  C    Alcohols  oxygenated compounds   haloforms less volatile than  methylene chloride    Approximate analyte volatility range   Example Analytes     Sorbent Maximum Temperature  250  C   Recommended Conditioning Temperature  225  C to 250  C   Recommended Desorption Temperature  No higher than 220  C   Notes  High artefacts   10 ng   Hydrophobic  Inert     suitable for labile compounds    For trace level analysis condition at  225  C  amp  desorb sample tubes no  higher than 200  C to reduce  background    7  Porapak Q  range of mesh sizes available     Sorbent Strength   Specific Surface Area  m2 g    Approximate analyte volatility range     QUI 1016 V2 2 Aug 06    Markes International Ltd  T   44  0 1443 230935    Medium   550  n Cs to n C12    F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Boiling point 50  C to 200  C    VOC s within volat
6.    TT24 7    Near Real Time  Monitoring    Operators Manual    AuGcust 2006  QUI 1016    VERSION 2 2    MARKES    ational Ltd  4  0 1443 230935 F   44  0 1443 231531 E  enquiries     markes       com    TT24 7 User Manual Markes International    This page left intentionally blank    QUI 1016 V2 2 Aug 06  1     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    TT24 7 User Manual   Table of Contents    Introduction    1 0 Quickstart Guide using the TT24 7    1 1 Running the TT24 7 system  1 1 1 Using an existing TT24 7 method    1 1 2 Creating a new TT24 7 method    2 0 The TT24 7 Thermal desorption System  2 1 An introduction to the TT24 7 system  2 2 TT24 7 Hardware configurations    2 3 Sampling modes  2 3 1 Continuous sampling  2 3 2 Single tube desorption    2 3 3 Split sampling    2 4 Principle system components   2 4 1 Valving assembly   2 4 2 Pneumatics module   2 4 3 Peltier cooled traps A and B   2 4 4 Heated transfer line   2 4 5 Tube desorption and sample split pneumatics    2 4 6 Electronic pneumatic control  EPC   2 4 6 1 Carrier gas connections  2 4 6 2 Connection to Agilent Technologies 6890GC   6850 GC  2 4 6 3 Connection to Thermo Electron Trace   FOCUS GC  2 4 6 4 Connection to other GC systems    QUI 1016 V2 2 Aug 06  1     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    11    13    13  13  14    15  15  16    16  16  18 
7.   07 500 ml mi    TrapB State  0 4 0 0   25 0   C Sample Gas  Air  Sample Time 0 0 min Transfer Line  0 0  120 0  C Carrier Gas  He  Dual Front HY  0 0   120 0  C       Figure 50  Instrument status bar  TT24 7e1     QUI 1016 V2 2 Aug 06  68     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    The instrument status bar  figure 50  is located at the bottom of the top level  screen  It is divided into four sections  i e  Instrument Status  Heated zones   GC and Flow Controller     2 9 1 3 1 Instrument status  This shows the status of the continuous sampling run which is operating     Instrument status  Indicates the current status of traps Aand B   Examples of status messages are     Sampling  Purging  Desorbing  Waiting for GC Ready  Equilibrating  Sample Time  How long the current trap has been sampling for  see  section 2 9 1 1 5    2 9 1 3 2 Heated zones  This section shows the heated zones monitored within the TT24 7   Trap A  Trap B  Transfer Line  Dual front HV    The status bar shows two adjacent numbers  The first represents the actual  value and the second the set point value  If these two numbers are equal they  appear in black font  if however the actual value is not at the set point it will  appear blue  If the TT24 7 is started when any one of these values is blue  then  the software will enter an equilibrating mode and the run will only commence  when the value becomes ready
8.   21     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 4 3 Peltier cooled cold traps  A and B   2 4 4 Heated sample transfer line  2 4 5 Optional tube desorption and   or sample split pneumatics    2 4 1 Valving assembly    The valving assembly within the TT24 7 consists of up to three heated valves   Each valve is manufactured from a solid block of PTFE  and machined to provide  the required flow paths for sampling and desorption of a sample gas stream     For hardware configuration e1 which only offers the continuous sampling  configuration  two valves are required  one for each trap  Each valve has three  points of connection i e  where the trap locates into the valve  where sample gas  enters the valve  and where the desorbed analytes leave the valve  Figure 7 is a  schematic which shows these connections  and the two opposing pins at the top  and bottom of the valve which control the active flow path within the valve i e   sampling  dry purging   backflush  desorption etc  These pins are either up or  down  and are actuated by an external gas supply        mo Two trap valves  HVA  amp  HVB    Figure 7  Schematic showing the two TT24 7 trap valves    When the additional configurations of tube desorption and split sampling are built  into the TT24 7  configurations e2 e3   then a third heated valve is required   which connects via a T connection to the two valves described above
9.   He 7   Sample Gas  air 7           Cancel         Figure 52  Options dialogue box   gas tab  This section defines the pressure units and the sample and carrier gas selection   Pressure units  The pressure unit selection is either psi  or kPa   Note 1 psi   101 325 kPa     However note that there are no displays of  pressure on the instrument status bar on a TT24   7e1 system    Carrier gas type  Choices of Helium  He   Nitrogen  N2   Hydrogen   H2   and diagnostic air are available     Sample gas type  Choices of Air  N2  He  Hz  safety considerations    may apply using this gas   and Carbon Dioxide   CO gt   are available  This selection configures the    flow calibration of the mass flow controller     QUI 1016 V2 2 Aug 06  72     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 9 1 5 2 Ports    Gas Ports   System   E Mail                Communications Port    Analyser Port  COMI    MFC Port  coms      Baud Rate  57600   Baud Rate  57600       r GC Interface Logic             f  Open   Start f Open   Ready    GCStart  out  GC Ready fin   i Closed   Start    Closed   Ready                               Figure 53  Options dialogue box   ports tab    This screen consists of two sections  the communications port  and the GC  interface logic  These two sections control communication between the TT24 7  and the associated PC system  and the ready   not ready and start signal logic  betw
10.   These values have to be equal   2  C tolerance  value  for the system to become ready     The transfer line and heated valves are controlled collectively by the flow path  temperature  see section 2 9 1 1 3      2 9 1 3 3 GC    The cable connecting the TT24 7 to the associated GC  see section 3 1    performs two functions  Firstly it monitors the GC ready status as specified in  the GC interface logic of the TT24 7 software  see section 2 9 1 5 2   If the GC  set point values are correct and equilibrated  and the software for the GC is in  a state to accept a start signal from the TT24 7  then the GC will be ready  This  ready status is relayed to the TT24 7 and observed in the GC status box  If the  GC is not ready  then a    Not Ready    comment appears     When the GC is ready the second function for the cable is to send the start  signal from the TT24 7 to the GC at the point of trap fire so it can begin its  analytical process  If the GC is not ready at the end of the sampling time  then  extended or fixed time sampling commences  see section 2 9 1 5 3      QUI 1016 V2 2 Aug 06  69     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 9 1 3 4 Flow controller    This component of the status box is not shown by default and requires  configuring within the TT24 7 software  see section 2 9 1 5 3   When  configured  following information is given     Flow rate  the sampling flow rate
11.   Where fine mesh size material is used  i e  80 100 mesh  or finer  the maximum flow rate will be reduced  Ultimately the fastest flow  rate achievable is determined by specifying the maximum flow rate of 1 L min  and observing the actual flow value on the status bar  see section 2 9 1 3  at  the bottom right side of the user interface     In certain applications where the analyte concentration is very low  i e  sub  ppb  and there is a time restriction on sampling  i e  NRT  there is a temptation  to set a very fast sampling flow rate to maximise the amount of analyte passing  into the trap in unit time  For example sampling at 1 L min for 10 minutes  results in a sampling volume of 10L  If the analyte concentration is very low  e g  in the pg L concentration  i e  ppt   then a 10L volume would contain  sufficient sample in theory for detection by selective GC detectors or by GCMS     However sensitivity is very dependant on the signal to noise  S N  ratio of the  resultant chromatographic peak  If this is sharp i e  peak widths  lt   5 seconds  then much better integration and detection is possible than for peak widths  gt  15  seconds     In practice it has been shown that very fast sampling flow rates   gt   800  mL min  with certain sorbent materials produces poor or broad peak shapes  and this can compromise the minimum detection limit  MDL  for that  compound  If the same compound is run at a lower flow i e  400 mL min as  opposed to 800 mL min  this can result in a m
12.   because the flow controller was unable to pull the required flow rate  over the sampling time  due to a failing cold trap or other problem      A warning message does not affect the operation of the TT24 7  system but gives information about something which may affect the  quantitation and the results     2 9 2 7 3 Error messages   ERROR  Trap A not heating   ERROR  Valve not heating   ERROR  Line not heating   ERROR  Trap A over temperature   ERROR  Valve over temperature   ERROR  Line over temperature    If one of the above errors occur then the TT24 7 system will shut down and  await operator   service engineer intervention     3 0 User Installed Components and Maintenance    3 1 System cabling TT24 7   PC   GC    This section describes the cabling connections required to correctly operate and  control the TT24 7 system  The TT24 7 has electrical connections with both the  associated GC  and the controlling PC  A power supply is also required  100    240V      There are four cables attached to the back of the TT24 7 system  These consist  of     i  and ii  Two  9 pin    D     serial  RS232  communication cables  P N Z 0189   iii  A single 25 pin remote  GC  start stop cable  which is instrument specific  Agilent GC P N UTD 5098  Thermo Electron GC P N UTD 5108  any other GC system uses UTD 5095  iv  A power cable  P N Z 0024      These four cables must be connected correctly for successful operation of the  TT24 7 system and remote starting of the associated GC     QUI 1
13.   described in section 1 1 above  The user interface for both sampling modes is  shown below  and varies slightly depending on the selection made  Figure 1      Method  default mth   modified Method  default mth   modified    Mode 14m n  A4 free hie     TrapB    Flow Path Temp   C Purge Flow Path Temp   C  Prepurge Time Prepurge Time      1 0 min H  and TL 120 0   1 0 min H   and TL 120 0  es    TT Sampling      Sample Flow Rate Trap Low Temp Desorb Flow Trap Low Temp      500 0 ml min   25 0 ie   500 0 ml min   25 0  e    Sample Time Trap High Temp Desorb Time Trap High Temp      10 0 min   100 0   C   3 0 a   100 0  c  Trap Hold Time Desorb Temp Trap Hold Time      10 0 min     250 0   C   10 0 fame                            Split   u Enable Trap Split    Save        Figure 1  TT24 7 User Interface   showing both sampling modes             Step 2  Enter set point values    After selecting the sampling mode  the desired set point values for each field  need to be specified so that the correct temperatures and flows are used for the  sampling and desorption phases  The Quickstart guide supplied with the  TT24 7  ITSO15  has a description of these fields on the reverse side  see  section 2 9 for further details      Note  the split functionality shown in the user interface above is only present  if the split option is configured within the TT24 7   Product Number TT24 7e3      QUI 1016 V2 2 Aug 06  14     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E
14.   enquiries markes com    TT24 7 User Manual Markes International    Without this  the system samples in a splitless mode directly into the capillary  column for maximum sensitivity     Step 3  Save the method    After all the set points are specified click the    Save    button shown on the  interface above  This allows the operator to specify the correct method file  name ending in a  mth suffix     To operate the TT24 7 simply follow the stepwise process described above in  section 1 1 1    2 0 The TT24 7 Thermal desorption System    This section covers the design philosophy of the TT24 7  the sampling protocols  available  and the hardware options that can be configured to achieve these  different sampling techniques     2 1 An introduction to the TT24 7 system    Thermal desorption as an analytical technique can be broadly divided into two  sampling categories namely off line  or on line  Off line sampling incorporates  both tube based  pumped or diffusive   and canister   bag  grab  samples   whereas for on line analysis the sample is a flowing gas stream passing directly  into the focusing trap for enrichment     Tube based sampling has the benefit of two stage pre concentration i e  from the  primary tube into the cold trap of the thermal desorption system  e g  UNITY   and  then from the cold trap into the capillary column  These two phases can typically  enrich the sample by a factor of 106     However for on line analysis a much faster analysis time is required  so
15.  11  12  12  12  14  17    18    19    19  21  22  23  23  24    25  26    TT24 7 User Manual    Markes International    Figure 13  Carrier gas supply connections on the back panel of the TT24 7 29    Figure 14  TT24 7 flow schematic  e3 configuration  showing the additional    gas lines associated with the GC EPC module    Figure 15  TT24 7e1   Standby   Figure 16  TT24 7e1   Trap A sampling  trap B purging   Figure 17  TT24 7e1   Trap A sampling  trap B desorbing   Figure 18  TT24 7e1   Trap A purging  trap B sampling   Figure 19  TT24 7e1   Trap A desorbing  trap B sampling   Figure 20  TT24 7e2   Continuous sampling   standby   Figure 21  TT24 7e2   Continuous sampling   trap A sampling  trap B  purging   Figure 22  TT24 7e2   Continuous sampling   trap A sampling  trap B  desorbing   Figure 23  TT24 7e2   Continuous sampling   trap A purging  trap B  sampling   Figure 24  TT24 7e2   Continuous sampling   trap A desorbing  trap B  sampling   Figure 25  TT24 7e2   Tube desorb mode   standby   Figure 26  TT24 7e2   Tube desorb mode   leak testing tube   Figure 27  TT24 7e2   Tube Desorb Mode   Leak testing trap A  Figure 28  TT24 7e2   Tube desorb mode   tube purge   Figure 29  TT24 7e2   Tube desorb mode   tube desorption   Figure 30  TT24 7e2   Tube desorb mode   trap  A  purge   Figure 31  TT24 7e2   Tube desorb mode   trap desorption   Figure 32  TT24 7e3   Continuous sampling   standby   Figure 33  TT24 7e3   Continuous sampling   trap A sampling  trap B  dry purging  
16.  20    21  22  23  24  25  27    27  28  30  31  31    TT24 7 User Manual    2 5 TT24 7 flow schematics    2 5 1 TT24 7e1 hardware configuration  2 5 1 1 Standby  2 5 1 2 Trap A sampling  trap B purging  2 5 1 3 Trap A sampling  trap B desorbing  2 5 1 4 Trap A purging  trap B sampling  2 5 1 5 Trap A desorbing  trap B sampling    2 5 2 TT24 7e2 hardware configuration    2 5 2 1 Continuous sampling mode  TT24 7e2   2 5 2 1 1 Standby  2 5 2 1 2 Trap A sampling  trap B purging  2 5 2 1 3 Trap A sampling  trap B desorbing  2 5 2 1 4 Trap A purging  trap B sampling  2 5 2 1 5 Trap A desorbing  trap B sampling    2 5 2 2 Tube desorb mode  TT24 7e2   2 5 2 2 1 Standby   2 5 2 2 2 Leak testing tube   2 5 2 2 3 Leak testing trap A  2 5 2 2 4 Tube purge   2 5 2 2 5 Tube desorption   2 5 2 2 6 Trap  A  purge   2 5 2 2 7 Trap  A  desorption    2 5 3 TT24 7e3 hardware configuration    2 5 3 1 Continuous sampling mode  TT24 7e3   2 5 3 1 1 Standby  2 5 3 1 2 Trap A sampling  trap B dry purging  2 5 3 1 3 Trap A sampling  trap B desorbing  2 5 3 1 4 Trap A purging  trap B sampling  2 5 3 1 5 Trap A desorbing  trap B sampling    2 5 3 2 Tube desorb mode  TT24 7e3   2 5 3 2 1 Standby  2 5 3 2 2 Leak testing tube    QUI 1016 V2 2 Aug 06  2     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531    Markes International    E  enquiries markes com    31    32  32  33  34  35  36    37    37  37  38  39  40  41  42  42  43  44  45  46  47  48    49    49  49  50  51  52  53  54 
17.  3 butadiene   350  C   400  C    300  C   350  C max  increase  temperature gradually from 100  C     To suit analyte    Significantly hydrophilic   do not  use in humid conditions     High artefacts   gt 10 ng     Easily  amp  irreversibly contaminated  by higher boiling components    F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Retention Volumes and Safe Sampling Volumes    The most accurate method of ensuring a particular sorbent tube will quantitatively  retain a particular analyte is to determine its retention volume for that analyte   Retention volumes are usually quoted in litres per gram of sorbent and must be  extrapolated to determine the retention volume on the mass of sorbent in a  standard tube  They are always quoted at a set temperature   usually 20  C     A comprehensive list of retention volumes for different analytes on several  sorbents packed into industry standard 3 5    x 1   4    OD tubes  is given in MDHS 72    Volatile Organic Compounds in Air   The MDHS  Methods for the Determination  of Hazardous Substances  series comprises validated methods produced by the  United Kingdom Health and Safety Executive for occupational hygiene work   Copies of these publications may be ordered from The Stationary Office  Holburn  Book Shop  59 60 Holburn Viaduct  London  EC1A 2FD  UK      It is possible to measure retention volumes in the laboratory by the following  procedure using a gas chromatograph configured with a
18.  54  55    TT24 7 User Manual    2 5 3 2 3 Leak testing trap A  2 5 3 2 4 Tube purge  2 5 3 2 5 Tube desorption  2 5 3 2 6 Trap  A  purge  2 5 3 2 7 Trap  A  desorption    2 6 Software control and configuration  2 7 Software installation  2 8 Firmware download    2 9 Software control   top level    2 9 1 User interface for TT24 7el1 hardware configurations    2 9 1 1 Method parameters  2 9 1 1 1 Mode   2 9 1 1 2 Prepurge time  2 9 1 1 3 Flow path temp  2 9 1 1 4 Sample flow rate  2 9 1 1 5 Sample time  2 9 1 1 6 Trap low temperature  2 9 1 1 7 Trap high temperature  2 9 1 1 8 Trap hold time  2 9 1 1 9 Save    2 9 1 2 Flow display    2 9 1 3 Instrument status bar   2 9 1 3 1 Instrument status   2 9 1 3 2 Heated zones   2 9 1 3 3 GC   2 9 1 3 4 Flow controller   2 9 1 4  Software icons and menu items  2 9 1 4 1 Software icons   2 9 1 4 2 Menu bar   2 9 1 5 Options   2 9 1 5 1 Gas    QUI 1016 V2 2 Aug 06  3     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E     Markes International    56  57  58  59  60    61    62    62    64    65    65  65  65  65  66  66  67  67  68  68    68    68  69  69  69  70    70  70  71  72  72    enquiries markes com    TT24 7 User Manual    2 9 1   2 9 1   2 9 1     5 2 Ports  5 3 System  5 4 eMail    2 9 1 6 User diagnostics    2 9 1 7 Reporting module    2 9 1 7 1 Information messages    2 9 1 7 2 Warning messages    2 9 1 7 3 Error messages    2 9 2 User interface for TT24 7e2 and TT24 7e3 hardware configurations    2 9 
19.  9 2 7 Reporting module  2 9 2 7 1 Information messages  2 9 2 7 2 Warning messages  2 9 2 7 3 Error messages    3 0 User Installed Components and Maintenance    3 1 System cabling TT24 7   PC   GC  3 1 1 TT24 7 power cable   ON OFF switch  3 1 2 Communications and Remote start cabling  3 1 3 Serial and   or USB Cable connections    3 1 4 Power recycle box    3 2 Gas requirements    3 2 1 User maintenance for gas requirements    3 3 Installation  and removal  of heated transfer line  3 3 1 Connecting the transfer line to the gas chromatograph  3 3 2 Installing the fused silica transfer line insert    3 3 3 Connecting the transfer line to the TT24 7  3 3 3 1 e1 systems  3 3 3 2 e2 e3 systems    3 3 4 Coupling the fused silica transfer line to the GC column  3 3 5 User maintenance of transfer line   3 4 Installation and removal of Cold Traps  3 4 1 Installing cold traps    3 4 2 Removing the cold traps    QUI 1016 V2 2 Aug 06  5     87  88  89    90    91  91  92  92    92    92  93  93  95  96    97  98    98  98  100    101  101  105    107  107  108    108  109    Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    3 4 3 Available Cold Traps   3 4 4 User maintenance of TT24 7 cold traps  3 4 4 1 Packing cold traps  3 4 4 2 Cold trap lifetime    3 5 Installation   removal of sampling tube  e2 e3     3 6 Installation   removal of split tube  e3     4 0 Troubleshooting  4 1 Leak testing the 
20.  A firing eventually   and the GC oven  After trap A fires both traps  will be conditioned and the GC column cleaned of any residual material     It is important to note that the GC cycle time must be less than the TT24 7  sampling time  or this will add to the overall cycle time of the method  If the GC  cycle time increases during the continuous sampling mode  e g  because the oven  cool down time increases  then the system will go into an extended sampling  mode for the trap which is currently sampling  This continues until the GC does  eventually become ready  at which point the trap fires  and the sequence  continues  See section 2 5 for further information on the TT24 7 flow schematics   Additionally the sampling time must be greater than the sum of purge  trap fire  and trap cooling times  typically 2 minutes      2 3 2 Single tube desorption  Available with hardware configurations e2 and e3     The TT24 7 may be configured with the capacity to desorb a single tube  For this  type of analysis    Tube Mode    must be selected as the sampling mode in the  TT24 7 software  figure 1      QUI 1016 V2 2 Aug 06  18     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    The tube to be analysed is inserted into a dedicated tube oven located on the left  hand side of the  front facing  instrument  The tube is retained and sealed by two  o rings  P N U COV10   one at either end of the oven ho
21.  By starting the TT24 7 software the continuous sampling process commences   The sample passes initially onto trap A  The date and time that the sampling  started  and the trap that is being sampled are reported as an information  message in the reporting module of the software  see section 2 9 2 7   e g     05 12 2005 10 49 31   INFO  Sampling Started on Trap A    At the same time trap B dry purges with carrier gas in the sampling  direction  This removes air and moisture from the trap prior to firing  The  dry purge flow rate through the trap  B  at this time is controlled by a needle  valve connected to SV6     The sample gas flow rate passing into the TT24 7 is controlled by the internal  mass flow controller  MFC      QUI 1016 V2 2 Aug 06  38     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 2 1 3 Trap A sampling  trap B desorbing    Desorption Tube       MFC  amp  Pump Sample in    Trap B             KEY          Flow of Carrier Gas  Carrier Gas  no flaw         a    Flow of Sample Gas  Sample Gas  no flow        Figure 22  TT24 7e2   Continuous sampling   trap A sampling  trap B desorbing    The first two runs of the continuous sampling mode  i e  trap A and trap B   would typically not be used for quantitative analysis of the sampled air but as  a clean up process for both traps  After both traps have been fired once then  the subsequent samples may be quantified     Afte
22.  Figure 34  TT24 7e3   Continuous sampling   trap A sampling  trap B  desorbing   Figure 35  TT24 7e3   Continuous sampling   trap A dry purging  trap B  sampling   Figure 36  TT24 7e3   Continuous sampling   trap A desorbing  trap B  sampling   Figure 37  TT24 7e3   Tube desorb mode   standby   Figure 38  TT24 7e3   Tube desorb mode   leak testing tube   Figure 39  T1T24 7e3   Tube desorb mode   leak testing trap A   Figure 40  TT24 7e3   Tube desorb mode   tube purge   QUI 1016 V2 2 Aug 06  8     Markes International Ltd     T   44  0 1443 230935 F   44  0 1443 231531    E  enquiries markes com    29  32  33  34  35  36  37    38    39    40    41  42  43  44  45  46  47  48  49    50    51    52    53  54  55  56  57    TT24 7 User Manual    Figure 41  TT24 7e3   Tube desorb mode   tube desorption   Figure 42  TT24 7e3   Tube desorb mode   trap  A  purge   Figure 43  TT24 7e3   Tube desorb mode   trap desorption   Figure 44  TT24 7 software   top level screen   Figure 45  Location of status   LEDs and front panel switch on TT24 7  Figure 46  Download activity timeout dialogue box   Figure 47  Ports tab in options configuration   Figure 48  TT24 7e1   User interface   Figure 49  Flow display showing flow direction through solenoid valves  Figure 50  Instrument status bar  TT24 7e1    Figure 51  Software icons and menu items for configurations e1  Figure 52  Options dialogue box   gas tab   Figure 53  Options dialogue box   ports tab   Figure 54  Options dialogue box   syste
23.  GC becomes ready  The extended  sampling time can be limited to a fixed time period by  entering a    Time to extend sampling    and ticking the     Enable Timeout    box     As a result of extended sampling the amount of sample  in the trap will be greater than that if extended sampling  had not occurred  The analytical result will therefore be  greater  and this may need to be compensated for in the  final quantitative data     The    No Back Valve    box should be checked   this  configures an e1 system     Gas   Ports   System E Mail      JV Enable E Mail Alerts    m E Mail Settings    Username    Password    Send To       Name  TT24elsystemt    EMailAddress  TT247etsysteml serverl com  Outgoing Server  92193101 tt t        lt       sSOsS   C       T1274 Location  Building 1330  Air Con System       IV Outgoing E mail server requires authentication    If you want to send an e mail to more than one recipient then separate each  e mail address using  e g  one example com  two example com               s ystem Operator    jtechnicalsupport company  com  service company com             Figure 55  Options dialogue box   email tab    This screen enables alert emails to be generated should the TT24 7 system    encounter a problem     This could be due to a set point value not being    controlled accurately due to a system fault  The nature of the fault  date and  time it occurred appear in the Report deviation screen of the top level software     The email alert facility has to
24.  UTD   5093   heated over its entire length by means of a distributed heater and at the  GC end by heat conduction from the GC oven     QUI 1016 V2 2 Aug 06  98     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    The line is 1 m long  which is sufficient to reach most gas chromatographs even  when a mass spectrometer is attached     Most GCs have access points built into the oven region by means of holes in the  side  top or back of the oven skin  with    knock out    sections in the outer casing   and one of these should ideally be utilised to install the transfer line     If all such access points are already in use it is possible to gain entry via an  unused injector or detector port with or without a heated zone     The general approach is illustrated in the three diagrams  figures 72  73 and 74     GC oven insulation    Silicone foam        PTFE protective sleeve tubing  0 25 mm I D  fused silica    End of line heater    Figure 72  Installation of transfer line through inner wall of GC oven    Note  In all cases the fused silica and PTFE sleeve tubing are fitted as the final  operation     Locate a hole in the inner oven wall with a corresponding hole leading to the  outside of the instrument  It is usually necessary to displace the oven insulation  material to enable the flexible metal line to be pushed against the outside of the  inner oven wall     If the GC oven wall insulatio
25.  be enabled by checking the    Enable eMail Alerts       box     Note  Some of the items in this section of the software may have to be    QUI 1016 V2 2 Aug 06    Markes International Ltd      75     T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    provided by the corporate IT department   eMail Settings    Name  This can be any descriptive name for the system in use   Where multiple TT24 7 systems are employed  a unique  identifying name per system is advised for accurate tracking     email Address  This is the email address of the PC system connected to  the TT24 7     Outgoing Server  The Internet Protocol  IP  address of the server  used by the PC must be specified here     T1247 Location  This is an important item  as it identifies the  exact location of the TT24 7 which has sent the email alert   It is recommended to make this section as clear and  descriptive as possible     Outgoing E mail server requires authentication  If the Outgoing E mail  server requires authentification  the box is checked as shown  above  However the alert email will not be sent unless a  Username and Password are typed  If the box is unchecked  i e  no authentification is required  then the alert email is  automatically generated and sent     Send to  This defines to whom the email is sent  This could be a single  person or consist of multiple email addresses  Each address  has to be separated by a semi colon         eMail alert
26.  followed by a plug of glass wool then the stronger  sorbent  Again this ensures that when air is drawn through the tube the mixture  reaches the weaker sorbent first where the higher boiling components are  trapped     Note that  if sampled tubes are to be stored for some time  for transportation or  other purposes then method 1 is preferred  After sampling the tube train can be  dismantled and each tube capped with long term storage caps  Long term storage  of tubes containing two or more sorbents is not recommended as higher boiling  components may migrate from the weaker to the stronger sorbent over time and  this can cause incomplete desorption when the tubes are eventually analysed  If  the multi sorbent tubes are to be stored for longer than one week then the  following procedure must be undertaken     take the samples in the normal way and cap with 1   4    brass storage caps  in the field as usual    place the tubes in an air tight container   e g  Tupperware box or clean tin  can with lid   and place in a clean refrigerator    when the tubes have reached the    cold    temperature  briefly remove them  from the refrigerator and check and re tighten the caps  due to the  difference in the thermal properties of brass and stainless steel the caps  may have become slightly loose     return the samples  in their container  to the refrigerator    immediately prior to analysis  remove the tubes from the refrigerator and  allow them to come to room temperature before analys
27.  from the menu bar  select File gt Open  and then the required method  The file suffix is    smth       Step 2  Preparing the GC GCMS to accept an external TT24 7 Start  Signal    The TT24 7 initiates the GC  GCMS and associated software by sending a  remote start signal at the point where the trap fires and releases its retained  sample  Prepare the GC  or GCMS  method or sequence  including any data file  name and storage location  so that on receipt of the start signal from the  TT24 7 the analytical system commences     Step 3  Check system ready status  Review the TT24 7 status bar at the base of the software to ensure all    temperatures and flows are stable and at their set point  The GC GCMS system  status should also be ready     QUI 1016 V2 2 Aug 06  13     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Step 4  Start run    Press the    Start Run    icon on the menu bar  this will initiate the TT24 7  continuous sampling mode  Initially this will be on to trap A during which time    trap B purges and trap fires in readiness for the first sample volume coming  from trap A     1 1 2 Creating a new TT24 7 method    Creating a new TT24 7 method follows a similar stepwise path which then leads  into the process described above for an existing method  section 1 1 1     Step 1  Choose the sampling mode   The first step to creating a new method is selection of the sampling mode as  
28.  lifted up    With the lever in the up position a tube can be positioned into the oven  It is  important that the tube is orientated correctly  the grooved end of the tube should  be towards the back of the instrument   With the tube in place the sealing  mechanism lever can be lowered which seals the tube into the flow path  When  sealing the tube into place take care that the tube is aligned with the sealing o   rings to prevent any o ring damage being incurred  Any damage to the o rings  could lead to a tube leak test failure     Removal of the tube is a very similar process  First  lift the sealing mechanism  lever to provide access to the tube and then pull the tube out by hand from the  sealing o ring  see figure 89   If the tube is difficult to extract from the sealing  o ring then the tube extractor can be used to provide extra grip        Figure 89  TT24 7e2 e3 system removing desorb tube    QUI 1016 V2 2 Aug 06  113     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    3 6 Installation   removal of split tube  e3     The split tube is located on the right hand side of the TT24 7 when looked at from  the front of the instrument  figure 93   The split tube is either a charcoal split  tube which traps any split volatiles and prevents them from being passed back  into the laboratory atmosphere  Or alternatively  when in Tube Desorb mode  the  split tube can be replaced by a conditi
29.  message  The email alert contains the following type of information     This is an automated message which has been sent as an error has occurred  with the TT24 7  Please see the details below     Problem  Transfer Line Over Temperature  Date  14 02 2005 10 45 41   Name  TT24 7el_system 1   Location  Building 1330   Air Con System 1    This provides sufficient information to identify the nature of the problem  the  system at fault  and its precise location     As the PC system connected to the TT24 7 could be part of a company    organisation internal network  then remote access software could be initiated  at this point  and the diagnostic section of the TT24 7 software accessed  This  could provide a clearer insight into the problem  before dispatching repair  personnel     2 9 1 6 User diagnostics    User diagnostics allows the user to manually manipulate the valves   both  heated valves and solenoid valves   in order to assist in tracking down a leak in  the system  see section 4 1   It is accessed from the    View    menu item   section 2 9 1 4 2 above      QUI 1016 V2 2 Aug 06  76     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Selecting    Diagnostics    opens a password Dialogue Box   enter the password     flowonly    and this will open the diagnostics screen   figure 56  Clicking on any  of the valves will cause the valve to move  e g  move UP from DOWN or OFF  from ON  
30.  packed column     Pack the chromatographic column with a known weight of the sorbent of  interest    Operate the column at a range of temperatures between 150  C and 250  C    Inject the analyte in question and note the retention time and column flow  at each temperature    From this information  calculate the specific retention volume in litres per  gram for each temperature    Plot the log of the specific retention volume against the reciprocal of the  absolute column temperature which gives a linear relationship    The slope of the graph may be extrapolated to give the log of the retention  volume at 20  C    The retention volume may then be calculated    In order to ensure that there is absolutely no chance of any breakthrough  occurring during a sampling procedure if  for example  the ambient temperature  were to rise  a Safe Sampling Volume  SSV  is usually quoted as being half the  retention volume  A sampling strategy which limits the volume of air sampled to  the SSV or less is therefore considered to be prudent     Note also that safe sampling volumes on some sorbents  particularly those with  some hydrophillicity  are significantly impacted by high atmospheric humidity  For  example SSVs on Spherocarb  UniCarb and Carbosieve SIII should typically be  reduced by a factor of 10 at 80  RH and above     Sampling in a mixed atmosphere    It is often the case that the list of analytes to be monitored requires more than  one sorbent  For example  if both toluene and meth
31.  pneumatic  control module  EPC  of the associated GC system  This demand will be based on  the carrier gas flow into the capillary column and any split flow or purge gas  requirements which are configured and specified during the process  This includes  the needle valve regulated split flow and dry purge flow values     QUI 1016 V2 2 Aug 06  31     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 1 TT24 7e1 hardware configuration  TT24 7el1 can only be operated in the continuous sampling mode     2 5 1 1 Standby       MFC  amp  Pump Sample in    KEY          Flow of Carrier Gas  Carrier Gas  no flaw        gt    _ Flow of Sample Gas      Sample Gas  no flaw        Figure 15  TT24 7e1   Standby    In this non sampling mode  carrier gas alone is continuously supplied to the  capillary column  No other gas flow occurs within the TT24 7 unit during this  time     QUI 1016 V2 2 Aug 06  32     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 1 2 Trap A sampling  trap B purging       KEY          Flow of Carrier Gas  Carrier Gas  no flaw        gt     Flow of Sample Gas      Sample Gas  no flaw        Figure 16  TT24 7e1   Trap A sampling  trap B purging    By starting the TT24 7 software the continuous sampling process commences   The date and time that the sampling started  and the trap that i
32.  restricted end        Figure 11  TT24 7 cold trap   note the restricted inlet   outlet end and the  quartz collar on the unrestricted end     Backflushing the trap allows the use of multiple sorbent beds in the trap  each bed  having a different sorbent strength  with the weaker sorbent at the start of the  trap  An example would be Tenax TA    followed by HayeSep D     or Chromosorb  106    etc  This sequence of sorbent materials allows for a much broader range of  analytes to be retained on the trap over that for a single bed  However greater  consideration is then required in terms of maximum operating temperature  as  many sorbents have quite different maximum upper temperature values   See  Appendix 3 for further information regarding sorbent selection and sorbent  maximum temperatures      Backflushing the traps allows each analyte to come off the retaining bed alone   without contact with a stronger sorbent material which could be irreversible or  detrimental to peak shape     Note  Installation and removal of the cold traps is a user operation  However  it  is necessary to follow the instructions carefully to avoid damaging the traps  Full  instructions are given in section 3 4  Collared traps are compatible with the  TT24 7 Trap Extraction Tool  TTD 5032     Note  You may have TT24 7 traps in stock which are of an earlier design with NO  collar on them   these traps may be used in your TT24 7 system but are not  compatible with the Trap Extraction Tool   extra care 
33.  so  sufficient time must be given for the tube oven to reach its desired set point  and stay at this value for at least 1 minute     2 9 2 1 8 Desorb temp   available in tube desorb mode    This is the upper tube set point temperature  Values ranging from 50  C to  350  C are possible  however the upper value should not exceed the maximum  temperature for the sorbents used  If a mixed bed sorbent tube is used the  upper value should not exceed the lowest of the maximum temperatures  See  Appendix 3 for details     Note  If you exceed the maximum sorbent temperature  the resulting  breakdown of the sorbent may severely contaminate the flow path of  your system  This may require the complete replacement of the flow  path and associated components by a fully qualified service engineer  and would not be covered by the instrument warranty     2 9 2 1 9 Trap A B   available in tube desorb mode    This selects which trap is used for sampling i e  onto which trap A or B the  desorbed or injected sample is passed     2 9 2 1 10 Trap low temperature   available in both sampling modes    This is the  lower  trap temperature value used to retain compounds on the  sorbent bed  Values ranging from 15 to 50  C are possible  Sub ambient values  are possible as a consequence of the Peltier cooling elements situated  immediately below each trap     QUI 1016 V2 2 Aug 06  81     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual M
34.  that  analytical data can be produced in the shortest time interval  This is facilitated by  sampling directly onto a cold trap for single stage enrichment  This ultimately  leads to near real time analysis  NRT  where all the sampling parameters are  running within the shortest time frame  and all the analytical conditions are  optimized for high speed chromatography  Time frames as short as 5 minutes are  possible with NRT     Off line analysis is therefore used for ultra low level sampling over extended time  periods  i e  8 hr time weighted average TWA   whereas on line analysis is used  for detecting potentially harmful levels at extremely low concentrations in the  shortest time frame     The main rate limiting step in a single trap TD system for on line analysis is the  cool down time of the trap  During this phase which can last for   5 minutes  no  sample can be adsorbed onto the trap  For most applications this is quite  adequate  and combined with high speed GC provides for excellent on line  analysis  However for certain compounds i e  chemical warfare agents  CWA  and  certain toxic industrial chemicals  TIC  this time frame is still too long  because of  their extreme toxicity to the human population  Ultimately a sampling regime is    QUI 1016 V2 2 Aug 06  15     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    required which is continuous  with no time blind spots  To a
35.  through the trap    Sample gas  the composition of the sample gas  as configured in the  options   section 2 9 1 5 1  is shown  typically air     Carrier gas  the composition of the carrier gas  as configured in the  options   section 2 9 1 5 1  is shown  typically Helium   2 9 1 4  Software icons and menu items  In the header section of the top level software there is a menu bar and a series  of icons as shown in figure 51   1124 7 Control    File wiew Instrument Help       D e W  gt    E    Figure 51  Software icons and menu items for configurations e1    2 9 1 4 1 Software icons    G  Creates a new method which starts with default parameters    Opens the method subdirectory so that an existing method  can be loaded into the TT24 7    4     H Requests a method file name to be saved into the methods    sub directory  No suffix is required as the     mth    suffix is  automatically added    a Starts a TT24 7 run with the loaded method conditions    Opens the Stop Run dialogue box which asks whether you  wish to    Stop Immediately    or    Continue running       QUI 1016 V2 2 Aug 06  70     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual    Markes International    2 9 1 4 2 Menu bar    File      1124 7 Control    File View Instrument Hel    L New    Se Open    Save Ctrl S  Save As  Ctri Alt 5    Simple Mode       Exit Ctrl x       New   Open     Save     Save As     Exit     View       1124 7 Control  File  
36.  top of the method window     QUI 1016 V2 2 Aug 06  82     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    When the desired method parameters have been specified they can be saved  directly into the active method by clicking on the SAVE button at the bottom of  the user interface     2 9 2 1 14 Split   available in both sampling modes    Note  The    Enable Trap Split    checkbox is only available with TT24 7e3  hardware configurations and when the split has been configured in Options   section 2 9 2 5      When enabled  this function opens the split solenoid valve SV7  see figure 37  in section 2 5 3 2 1  at the point of trap fire and splits the sample in proportion  to the split flow rate and column flow rate  see section 2 3 3      2 9 2 2 Flow Display    The flow display window  figure 59  is designed to help the user see the flow  direction process and the current run status of the TT24 7 system  The  diagram shows the flow direction through the solenoid valves SV1  2  3  4  5  and 6  the flow direction through the heated valves  HVA  HVB  HVC   the flow  through the traps A and B  and the flow through SV7 solenoid valve in a TT24   7e3 configuration  if the split is enabled   see 2 8 2 1 14      Desorption Tube       Figure 59  Flow Display showing flow direction through solenoid valves    2 9 2 3 Instrument status bar       Instrument Status Heated Zones GC Flow Controller   Tra
37.  tube  some analytes can diffuse into the side arm causing  measurable peak tailing  The 0 25 mm I D  fused silica transfer line must be  installed through the union and on into the narrow bore part of the inert coated  stainless steel tube     When connecting the transfer line to the TT24 7e1  pull about 20 cm of fused  silica from the PTFE line casing  Slide a   ic inch stainless steel Swagelok nut  and a   te inch x 0 4mm ferrule  Supplied with shipping kit  see Appendix 1  onto  the end of the fused silica and then cut off the first few mm of fused silica  Feed  the fused silica into the top of the union and slide the ferrule into position   Screw the nut onto the union and  with the fused silica still loose  position it so  that the fused silica is inserted around 20 mm into the union  Tighten the nut  to trap the fused silica and then tighten a further half turn using one of the 8  mm wrenches  spanners  provided in the shipping kit  Do not over tighten or  the ferrule will become distorted     Carefully bring the clamp plate  PTFE plate and shield tube down into position   shown in figures 79  amp  80  with the shield tube covering the union nut  The  shield tubing should be positioned such that the   isc inch side tubing projects  through one of the cutouts  As the transfer line is lowered into position the  clamp plate and PTFE plate should fit onto the two exposed threads     Use two M4 nuts  supplied with shipping kit  to lock the PTFE plate onto the  exposed threa
38.  view Instrument    O Options  Diagnostics       Options     Creates a new method which starts with default parameters    Opens the method subdirectory so that an existing method  can be loaded into the TT24 7    Saves the current TT24 7 parameters directly into the  loaded method    Requests a method file name to be saved into the methods  sub directory  No suffix is required as the  mth suffix is  automatically added    Immediately closes down the TT24 7 software    Accesses the 1124 7 configuration software  see section  2 9 1 5 for details     Diagnostics    Accesses two levels of diagnostic software for the TT24     Instrument    B 1124 7 Control  File view   Instrument    D    EP Run       Run   Stop     QUI 1016 V2 2 Aug 06    7    flow only  user  diagnostics and full  service engineer   diagnostics  see section 2 9 1 6      Starts the TT24 7 with the loaded method conditions    Opens the Stop Run dialogue box which asks whether you  wish to    Stop Immediately    or    Continue running        71     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 9 1 5 Options    Options is the principle configuration section for the TT24 7 system  It is  accessed from the    View    menu item  section 2 9 1 4 2 above      Within Options  there are three separate configuration screens   2 9 1 5 1 Gas    Gas   Ports   System   E Mail              PSI    Pressure Units  f kPa       Carrier Gas
39.  will display the current flow rate from the MFC in the status bar           Limit Sample Time   I Enable    Enabling this option will force the instrument to take a sample only for the specified sampling  time  No more gas will be sampled until the next sampling cycle               Sampling Extended Timeout    Time to extend sampling X IV Enable Timeout          Valves  IV No Back Valve                   Cancel      Figure 54  Options dialogue box   system tab       Display flow  If the enabling box is checked  the mass flow controller  flow rate and associated values are displayed in the top  level status bar  see section 2 9 1 3 4     Limit Sample Time  This option fixes the sampling time in continuous  sampling mode  No extended sampling will take place  if the GC is not ready when this is active  see 2 9 1 1 5   for further information regarding sample time      Sampling Extended Timeout  In continuous sampling mode  the TT24 7 checks  on the GC ready status at the completion of the  sampling time  If the GC is ready  then the system  continues into dry purge and then trap fire  However if  the GC system is not ready e g  the oven has not cooled  down to its set point value and equilibrated  then the  TT24 7 will automatically go into an extended sampling    QUI 1016 V2 2 Aug 06  74     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual    Valves     2 9 1 5 4 eMail    Markes International    mode until the
40.  within    QUI 1016 V2 2 Aug 06  27     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    the TT24 7  the column pressure remains at the defined value determined by the  GC method  Normally this is a constant pressure value  however in certain  circumstances pressure or flow programming may be required     By controlling the pressure at the head of the column  the retention time of  compounds eluting from the column will be fixed  Chromatographically this is  highly advantageous as it produces reproducible retention times for each  compound in all analyses  This ultimately leads to the generation of absolute  retention time databases for compound identification     Closed loop feedback is particularly important when the sample is being split  The  split flow passes through the rear heated valve and out via a needle valve for  control as discussed in section 2 3 3  This is immediately adjacent to the point  where the transfer line leading to the capillary column enters the TT24 7  and  without electronic control would directly affect the column head pressure i e  as  the split flow increased the head pressure would decrease and compound  retention times would vary accordingly     The pressure at this point is monitored by the pressure sensor within the GC flow  module  this sensor subsequently controls the total flow being delivered into the  TT24 7  This process is closed loop feedback  
41. 016 V2 2 Aug 06  92     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    3 1 1 TT24 7 power cable   ON OFF switch    The power cable is inserted into the power supply socket at the rear of the TT24 7  and the system is switched on using the adjacent ON OFF switch  The location of  this connection and switch is shown in figure 67  The power is supplied to the  system when the ON OFF switch is in the depressed state     To turn the TT24 7    OFF press the ON OFF switch so the depressed state is O     3 1 2 Communications and Remote start cabling    Figure 67 shows the rear of the TT24 7 illustrating the location of the sockets for  the two RS232 communication cables  the remote start cable and the power cable     1343990   PEPR rr      Remote start  cable connection    Serial cable 1   Sampler     Power ON OFF  switch    Serial cable   MFC     Unused  connection       Figure 67  Rear view of TT24 7 showing system cabling connections    Figure 68 shows a close up of the labelling on the back of the TT24 7   Note  that the label also shows the pneumatic connections into the TT24 7      Additional useful information such as the power requirements and instrument  serial number for the TT24 7 are also shown here     QUI 1016 V2 2 Aug 06  93     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International       Figure 6
42. 2 1 Method parameters    2 9 2     1 1 Mode    2 9 2 1 2 Prepurge time   available in both sampling modes    2 9 2   2 9 2   2 9 2   2 9 2   2 9 2   2 9 2   2 9 2   2 9 2   2 9 2   2 9 2   2 9 2   2 9 2     1 3 Flow path temp   available in both sampling modes   1 4 Sample flow rate   available in TT sampling mode   1 5 Sample time   available in TT sampling mode   1 6 Desorb flow   available in tube desorb mode   1 7 Desorb time   available in tube desorb mode   1 8 Desorb temp   available in tube desorb mode   1 9 Trap A B   available in tube desorb mode   1 10 Trap low temperature   available in both sampling modes  1 11 Trap high temperature   available in both sampling modes  1 12 Trap hold time   available in both sampling modes   1 13 Save   available in both sampling modes   1 14 Split   available in both sampling modes    2 9 2 3 Instrument status bar  2 9 2 3 1 Instrument Status  2 9 2 3 2 Heated Zones  2 9 2 3 3 GC   2 9 2 3 4 Flow controller    2 9 2 4 Software icons and menu items  2 9 2 4 1 Software icons  2 9 1 4 2 Menu bar    2 9 2 5 Options    QUI 1016 V2 2 Aug 06  4     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    Markes International    73  74  75    76    77  78  78  78    78    78  78    79  79  80  80  81  81  81  81  81  82  82  82  83    83  84  84  84  85    85  85  86    87    TT24 7 User Manual Markes International    2 9 2 5 1 Gas  2 9 2 5 2 Ports  2 9 2 5 3 System    2 9 2 6 User diagnostics    2
43. 4 7 Thermal Desorption System   where more detailed  knowledge and understanding is required    Section 3 User Installed Components and Maintenance  Section 4 Troubleshooting    There are three TT24 7 hardware configurations available from Markes  International  A visual inspection of the unit will ensure that you know which  configuration you have prior to reading this manual        TT24 7e1 basic system configured only for continuous sampling    QUI 1016 V2 2 Aug 06  11     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual    QUI 1016 V2 2 Aug 06    Markes International Ltd        T   44  0 1443 230935    Markes International    TT24 7e2    standard system configured with  the following additional features       tube desorption    pre purge to vent   dry purge    leak test    Note  the presence of the tube  desorption oven on the left hand  side of the trap box indicates that  this is an e2 system    TT24 7e3    standard system configured with  the following additional features       tube desorption    pre purge to vent   dry purge    split capability    Note  the presence of both the  tube desorption oven on the left  hand side and the split tube on  the right hand side  indicates that  this is an e3 system    F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    1 0 Quickstart Guide using the TT24 7    This section is in support of the    Quickstart Guide to Operating t
44. 44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Needle valve  to control  split flow    Split vent line       Figure 6  TT24 7e3 showing the tube desorb and split accessories fitted    The split ratio is then a function of the column flow relative to the total flow  entering the trap     Split Ratio   Column Flow       Column Flow   Split Flow    An additional benefit of split mode sampling is the increase in carrier gas flow rate  passing through the trap during the trap fire phase  This will be considerably  higher than in splitless mode  where the flow through the trap is equal to column  flow  Depending on the column   detector this ranges typically from  approximately 10 mL min down to 3 mL min  The increase in flow is directly  proportional to the split flow  and results in a faster movement of retained sample  from the sorbent bed into the column  This allows for higher initial operating  temperatures in the capillary column as the band width of sample entering the  column will be much sharper  In splitless mode the column needs to be cooler to  promote an efficient focusing effect on the stationary phase  This is a standard  procedure for splitless injection     The flow schematics for the split configuration  e3  are shown in section 2 5 3    2 4 Principle system components   The main functional components of the TT24 7  excluding electronics   consist of   2 4 1 Valving assembly  2 4 2 Pneumatics module    QUI 1016 V2 2 Aug 06
45. 6  84     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 9 2 3 4 Flow controller    This component of the status box is not shown by default and requires  configuring within the TT24 7 software  see section 2 9 2 5 3   When  configured  following information is given     Flow rate  the sample flow as follows  Trap A B TT Sampling mode  Tube desorb flow tube desorb mode    Sample gas  the composition of the sample gas  as configured in the  Options   section 2 9 2 5 3  is shown  typically air     Carrier gas  the composition of the carrier gas  as configured in the  Options   section 2 9 2 5 3  is shown  typically He   2 9 2 4 Software icons and menu items    In the header section of the top level software there is a menu bar and a series  of software icons  figure 61         1124 7 Control    File View Instrument Help    Ce HE   a    Figure 61  Software Icons and menu bar for configurations e2 e3       2 9 2 4 1 Software icons    i Creates a new method which starts with default parameters  a Opens the method subdirectory so that an existing method     can be loaded into the TT24 7   r Requests a method file name to be saved into the methods   mj sub directory  No suffix is required as the  mth suffix is    automatically added     p Starts a TT24 7 run with the loaded method conditions    Opens the Stop Run dialogue box which asks whether you  wish to    Stop Immediately    or    Con
46. 7 User Manual Markes International    2 9 1 User interface for TT24 7e1 hardware configurations    Figure 48 shows the user interface for the TT24 7e1 hardware configuration  Only  one sampling mode is available in this configuration   TT Sampling     Method  default mth   modified    Mode       Purge Flow Path Temp   C       4  Prepurge Time   HY and TL   120 0  TT Sampling    1 0  min  Sample Flow Rate Trap Low Temp      500 0 ml min   25 0   C    Sample Time Trap High Temp  10 0 min   100 0   C             Trap Hold Time  10 0 min                Figure 48  TT24 7e1   User interface    2 9 1 1 Method parameters    At the top of the method window is shown the active method name  e g   default mth  and its status i e  whether or not it has been modified from its  original set point values     The set point values used in the interface are described below   2 9 1 1 1 Mode    The sampling mode cannot be changed in the software   it is set to continuous  sampling  TT Sampling   as this is the only mode of operation     2 9 1 1 2 Prepurge time    This is the pre purge time when carrier gas is passed through the trap just prior  to trap fire  This purge  in the desorption direction  is to purge air   moisture  from the trap after sampling and the flow is equivalent to the column flow   which should be at least 3 mL min   The pre purge can be set between 0 and  99 9 minutes in increments of 0 1 minute and is typically 0 5 to 1 minute   although this can be reduced where the cycle t
47. 8  TT24 7 labelling    One of the 9 pin serial  RS232  cables connects into the Sampler port of the  TT24 7  This communicates with the analyzer part of the TT24 7 for instrument  control  This cable leads back to the PC  either directly into a serial  communication port  COM1  COM2 etc   in the PC or via a USB hub  The COM port  assignment is selected within the TT24 7 software  see section  2 9 1 5 2 2 9 2 5 2 Ports      The second serial cable connects into the MFC connector of the TT24 7  This is  the mass flow controller connection port and all flow related values within the  TT24 7 system are controlled and monitored via this cable connection to the PC   As for the Sampler connection described above this cable also leads back to the  PC  either directly into a serial communication port  COM2  COM1  in the PC or via  a USB hub     The COM port assignment is selected within the TT24 7 software  see section  2 9 1 5 2  e1  and 2 9 2 5 2  e2 e3    Note that the COM port selection for the  MFC and Sampler must be different     If the two serial cables become disconnected from the back of the TT24 7  ensure  that the correct serial cable is returned to its allocated socket as per the COM  settings  If the cables are crossed over the TT24 7 will not operate correctly     Connection of the 25 pin Remote start cable is made into the I O socket of the  TT24 7  This cable is also connected into the associated GC  and particularly into  the GC Remote start socket  Because GC manufa
48. Column flow   Split flow    If the split function is not enabled then the flow rate through the trap is equal  to the column flow rate alone and should be at least 3 mL min        QUI 1016 V2 2 Aug 06  60     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 6 Software control and configuration    This section reviews the software control and configuration for each of the  available TT24 7 hardware configurations  The user interface is discussed in  detail     The 1124 7 software runs as a Windows   application using Windows 98     onwards  however Windows 2000    or Windows XP    is the preferred version  If  Windows NT    is used no USB cable connection is possible as this is not supported  by NT     Ensure that you have one of these versions of Microsoft Windows installed on the  PC prior to installation of the software     The software consists of a top level screen containing a set point user interface   an active flow schematic  a reporting module and a status bar showing set point  and actual values  figure 44         1124 7 Control    Fie Yiew Instrument Help Function icons and  DEH pe eee bar    Method  default mth   modified    jode   lode   TT Sampling      Inlet   Desorption Tube _   Tube    Flow Path Temp C         pete    0 min HY and TL EEN       iw o   O o  o       05 12 2005 10 491  INFO  Samping Started on Trap A    User Active  interface Reporting flow  module schema
49. Desorption Tube    Needle   N Valve    Q    He Q     MFC  amp  Pump Sample in  N  on    Trap B          Needle  Valve    Split Tube    KEY       gt     Flow of Carrier Gas        Optional Carrier Gas Flow  Carrier Gas  no flow           Flow of Sample Gas  Sample Gas  no flow        Figure 42  TT24 7e3   Tube desorb mode   trap  A  purge    This is the dry purge phase prior to trap fire  to remove any residual air and  moisture coming from the tube  The flow is controlled by the needle valve  attached to SV6     The split line pnuematics through SV7 and its associated needle valve are also    purged during this operation  The flow through this line is controlled by this  SV7 needle valve     QUI 1016 V2 2 Aug 06  59     TT24 7 User Manual Markes International    2 5 3 2 7 Trap  A  desorption    MFC  amp  Pump Sample in    Trap B    Needle  Valve    Column            Flow of Carrier Gas        Optional Carrier Gas Flow  Carrier Gas  no flow            Flow of Sample Gas  Sample Gas  no flow        Figure 43  TT24 7e3   Tube desorb mode   trap desorption    The selected trap is now rapidly heated and backflushed with carrier gas to  remove retained analytes from the sorbent bed s  into the capillary column     If the split function is enabled in the method   see section 2 9 2 1 14  then this  will occur at trap fire  The flow rate through the trap is therefore equal to the  column flow plus the split flow through SV7  The split ratio is therefore     Split Ratio   Column Flow  
50. ED    Front panel switch       Figure 45  Location of status   LEDs and front panel switch on TT24 7  To download the firmware follow the stepwise procedure below     Step 1  The TT24 7 should be switched on at the back of the instrument  with  the front panel switch off  i e  not depressed   This should cause the top LED  to turn red  while the bottom LED remains OFF     If however the front panel switch is inadvertently ON  i e  depressed  at this    QUI 1016 V2 2 Aug 06  62     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    time both LED   s will appear red  Correct this by simply turning the front panel  switch OFF before continuing     Step 2  Start the TT24 7 software within the controlling PC  This is achieved    by clicking on the system icon  The software starts by attempting to  detect the sampler board  top LED  and if successful will begin downloading the  firmware  The top LED will turn amber at this time indicating the download is  in progress     If however the top LED remains red after approximately 10 seconds this  indicates the sampler board was not detected  To correct this  turn the main   rear  power switch off  wait 10 seconds and switch on again  The TT24 7  software will continue to look for the sampler board  and assuming there is not  a specific problem within the instrument  the firmware download will resume  and the top LED turns amber     Upon successful c
51. FC  amp  Pump    Trap B    KEY          Flow of Carrier Gas  Carrier Gas  no flaw        gt    _ Flow of Sample Gas      Sarple Gas  no flow        Figure 18  TT24 7e1   Trap A purging  trap B sampling    After the sampling time for trap A has completed trap A switches into its pre   purge mode to sweep out air and moisture prior to trap fire  Having cooled to  its starting temperature after the first clean up trap fire  trap B now goes into  its sampling phase  The date and time that the sampling started  and the trap  that is being sampled are reported as an information message in the reporting  module of the software  see section 2 9 1 7   e g     05 12 2005 10 59 17   INFO  Sampling Started on Trap B    QUI 1016 V2 2 Aug 06  35     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 1 5 Trap A desorbing  trap B sampling       MFC  amp  Pump Sample in    Trap B    KEY          Flow of Carrier Gas  Carrier Gas  no flaw        gt    _ Flow of Sample Gas      Sample Gas  no flow        Figure 19  TT24 7e1   Trap A desorbing  trap B sampling    Trap A now fires  If this is the first run for A this run is treated as a clean up  run  however if this is a subsequent run then real sample will be transferred  across to the capillary column for analysis     To see how the time overlap of sampling  purging  trap fire and cooling occurs  relative to the GC oven cycle time please refer to sectio
52. If the split flow is increased the  pressure at this point would normally drop  however the sensor compensates for  this by increasing the total flow into the TT24 7 to provide for the extra split flow  and maintain the column pressure     Most GC manufacturers have an LCD built into the instrument and information  relating to the GC inlet flows can normally be observed  One such parameter is  the    Total Flow    delivered by the inlet into the capillary injection port  When the  inlet flow module is configured with the TT24 7 this    Total Flow    parameter is now  the flow entering the thermal desorption system  This provides a mechanism to  observe the change in total flow as the split flow is varied  i e  as the split flow  increases this value will also increase in direct proportion     When the analysis is splitless  i e  no split flow   the flow rate through the trap is  equivalent to the column flow which is typically between 3 to 10 mL min  Once  again the flow is controlled by the column head pressure and the total flow  demand now is approximately equal to the column flow  The value may be slightly  higher than the exact column flow if a septum purge flow is running     2 4 6 1 Carrier gas connections    Manufacturers of GC systems which have the capability of electronic pneumatic  control typically have three pneumatic lines leading from the electronic module  into the capillary injection port  These consist of the carrier gas supply line  a  septum purge line an
53. Spherocarb   is a registered trademark of Phase Separations  UK   Tenax TA  Tenax GR   are registered trademarks of Buchem B V  The Netherlands    UniCarb   is a trademark of Markes International Ltd   UK    QUI 1016 V2 2 Aug 06  137     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    
54. UI 1016 V2 2 Aug 06  78     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Method  default mth   modified Method  default mth   modified    Mode jag i fen   Trapa    C TrapB    Purge Flow Path Temp   C   Purge  gt   gt  Flow Path Temp   C       Prepurge Time    1 0 min    TT Sampling TT Sampling  Sample Flow Rate Trap Low Temp Desorb Flow Trap Low Temp    500 0 mlmin   25 0   C   500 0 ml min   25 0 EE  Sample Time Trap High Temp Desorb Time Trap High Temp  10 0  min   100 0   C   3 0 min   100 0   C    Trap Hold Time Desorb Temp Trap Hold Time      10 0  fae   250 0   C   10 0 min    L    Prepurge Time    H and TL 120 0   1 0 min H   and TL 120 0                                  f Split    I Enable Trap Split    Save   Save      Figure 58  User interface for each sampling mode             TT Sampling  for continuous sampling from a stream of gas air  see section  2 3 1     Tube Desorb  allows the desorption of a sorbent tube  see section 2 3 2     2 9 2 1 2 Prepurge time   available in both sampling modes    This is the time that the trap is dry purged  carrier gas passed through the trap  in the sampling direction  just prior to trap fire  This dry purge removes air    moisture from the trap after sampling and the flow is controlled by a needle  valve SV6 as shown in figure 21  The flow should be typically set to a value of    50 mL min  The pre purge can be set between 0 and 99 9 m
55. a specified    time out     value  see section 2 8 1 5 3  is equalled at which point the system stops  The  GC becoming not ready is typically due to the oven not cooling down as fast as  normal and this affects the GC cycle time     The sampling time can be fixed to the specified value i e  preventing extended  sampling within the TT24 7 software  Under these conditions if the GC is not  ready then after the fixed sampling time the system then waits until the GC  becomes ready or again the time out value stops the process  See section  2 9 1 5 3 for extended sampling and fixed time sampling     2 9 2 1 6 Desorb flow   available in tube desorb mode    This is the flow rate of carrier gas which passes through the tube during its  primary desorption phase when the tube is heated to transfer the retained  sample into the selected trap   either Aor B  see 2 9 2 1 11   Flow values  ranging from 50 mL min to 1000 mL min are possible and typically flow rates  of  100 mL min are used     2 9 2 1 7 Desorb time   available in tube desorb mode    This is the primary desorption time interval  values ranging from 0 1 to 10  minutes are possible  Within this time frame sample from the tube must pass  into the selected trap and the tube must reach and maintain its upper  temperature value  section 2 9 2 1 8   Values of less than three minutes are  therefore not recommended unless the required desorb temperature is very  low  The heating rate for the tube is much slower than that of the trap
56. allowing different parts of the flow path to be leak checked with an  external device such as a helium leak detector  see Section 4 1      Before exiting the flow only diagnostic section  click the    standby    button  top  left hand side   this returns all valves to the standby position ready for system  operation        Diagnostics    Pneumatics         m Manual Valve Control    Standby   Out    To Column                Figure 56  Flow diagnostics screen    2 9 1 7 Reporting module    The reporting module  figure 57  displays a number of different Information   Warning or Error messages     All messages detailed below will appear for either Trap A or Trap B if relevant   Trap A is used as an example below  All error messages are date and time  stamped e g  05 12 2005 10 34 54    PR eports  05 12 2005 10 49 31  INFO  Samping Started on Trap A       Figure 57  Reporting module    QUI 1016 V2 2 Aug 06  77     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 9 1 7 1 Information messages   INFO  Sampling started on Trap A     date and time stamped when sampling started   INFO  Sampling has been extended on Trap A      reported when samping time exceeded the expected time set in the  method  this is typically due to the GC Ready signal being delayed    perhaps due to the oven taking longer than expected to cool down     Information messages do not affect the operation of the system in an
57. alyte retained on the trap     Valves  There are three tick boxes that may be checked    Enable Split option for method    For TT24 7e3 systems configured with the split option  this box should be checked to include this capacity in the  method    Purge Valve SV6    For TT24 7e2 e3 hardware configurations the SV6 valve  box should be checked to configure the purge valve and  allow dry purging of the traps    No Rear Valve    This box should be unchecked for e2 and e3  configurations        2 9 2 6 User diagnostics    User Diagnostics allows the user to manually manipulate the valves   both  heated valves and solenoid valves   in order to assist in tracking down a leak in  the system  see section 4 1   It is accessed from the    View    menu item   section 2 9 1 4 2 above      Selecting    Diagnostics    opens a password Dialogue Box   enter the password     flowonly    and this will open the diagnostics screen   figure 65  Clicking on any  of the valves will cause the valve to move  e g  move UP from DOWN or OFF  from ON  allowing different parts of the flow path to be leak checked with an  external device such as a helium leak detector  see Section 4 1      Before exiting the flow only diagnostic section  click the    standby    button  top  left hand side   this returns all valves to the standby position ready for system  operation     QUI 1016 V2 2 Aug 06  90     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Ma
58. and this can compromise the minimum detection limit  MDL  for that  compound  If the same compound is run at a lower flow i e  400 mL min as  opposed to 800 mL min  this can result in a much sharper peak and a  significantly better S N ratio even though the amount of analyte trapped is half     The science behind this effect is based on the depth the analyte passes into the  sorbent  and as a consequence how easily it comes off at trap fire  It is  therefore both compound and sorbent dependant     2 9 2 1 5 Sample time   available in TT sampling mode    This specifies the time interval for sample to pass into the trap  This parameter  is directly associated with the sampling flow rate as discussed above to  determine the total amount of sample transferred  The range for sample time  is from 0 1 to 99 9 minutes  however  values greater than 1 minute will typically  be used     The sampling time and therefore the sampling volume can be affected by the  ready   not ready status of the GC system  At the completion of the defined  set point time value and prior to the system going into the pre purge phase of  the trap  the TT24 7 monitors the GC ready status  If the GC is ready then the  pre purge process commences  If however the GC is not ready then the    QUI 1016 V2 2 Aug 06  80     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    sampling time is extended until the GC becomes ready  or 
59. anol are to be monitored  simultaneously two sorbents   one medium   weak for toluene and one strong for  methanol   should be used  For diffusive monitoring  this can only be achieved by  using two or more tubes in parallel and by stringently conditioning the tube  packed with the stronger sorbent before re use  Pumped monitoring requires    QUI 1016 V2 2 Aug 06  135     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    samples to be drawn through the two or more sorbents in series  Sampling onto  sorbents in series can be achieved in one of two ways     1  Linking together two tubes containing the sorbents required  2  Packing both sorbents as two separate beds in a single tube    In method 1 tubes are connected together using     1   4    brass couplings fitted with  PTFE combined ferrules  The tubes must be connected in series such that the tube  containing the weakest sorbent is at the front of the sampling train and all tubes  must be oriented such that the sample passes through from the sampling   grooved  end of each tube  This ensures that the higher boiling components in  the mixture are adsorbed by the weaker sorbent and eliminated from the sample  stream before reaching the strong sorbent     In method 2 small plugs of glass wool or sorbent retaining gauzes separate the 2  or three sorbents in a single tube  The weaker sorbent is packed at the front   sampling end of the tube
60. arkes International    The trap minimum value is sensitive to both the flow path temperature  described above  the flow rate of sample passing through the trap  and the  temperature of the air sample  For example if the flow path temperature is set  to 200  C  with a fast flow rate  i e   gt   700 mL min   then the trap minimum  value may be 25  C or higher at faster flows  If additionally the sample gas is  above ambient then this minimum value will be higher again     Using lower sampling flows will assist this value  and can indeed improve on  signal to noise for the resulting chromatographic peak  see section 2 9 1 1 5  above      Ultimately the minimum trap temperature achievable is determined  experimentally by reviewing the relationship between the set point and actual  values as shown in the status bar  and this will be a function of the flow path  temperature and the sampling conditions     2 9 2 1 11 Trap high temperature   available in both sampling modes    This is the upper temperature setting for the trap  Values ranging from 50  C  to 400  C are possible  The upper trap temperature should be set to the  optimum value which enables 100  recovery of analytes from the sorbent bed   but does not exceed the maximum permissible temperature for that  sorbent  See Appendix 3  Keeping the maximum trap temperature as low as  possible whilst still enabling 100  analyte recovery will extend trap lifetime     Note  If you exceed the maximum sorbent temperature  the resultin
61. aw           Flow of Sample Gas  Sample Gas  no flow        Figure 29  TT24 7e2   Tube desorb mode   tube desorption    The tube is now heated to the temperature set as the    Desorb Temp    within the  TT24 7 method  The tube    Desorb Flow    and    Desorb Time    parameters control  the desorption process  The set    Desorb Temp    value will depend on the sorbent  material and should not be set higher than that recommended for the  sorbent s  in the tube  See Appendix 3 for further details on sorbent selection  and maximum recommended temperatures     Sufficient flow and time should be maintained to completely desorb all analytes  from the tube onto the specified trap A or B     In the schematic above A is the chosen trap     QUI 1016 V2 2 Aug 06  46     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 2 2 6 Trap  A  purge    Desorption Tube    MFC  amp  Pump Sample in    Trap B       KEY          Flow of Carrier Gas     Carrier Gas  no flaw      a    Flow of Sample Gas  Sample Gas  no flow        Figure 30  TT24 7e2   Tube desorb mode   trap  A  purge    This is the dry purge phase prior to trap fire  to remove any residual air and  moisture coming from the tube  The flow is controlled by the needle valve  attached to SV6     QUI 1016 V2 2 Aug 06  47     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes In
62. can be sent to the TT24 7 Software   Command  Start   Parameters  None   Send a command to the TT24 7 to start run     Command  Stop  Parameters  None  Send a command to the TT24 7 to stop the current run     Command  LeakTest  Parameters  None    Send a command to the TT24 7 to start a leak test  Command will not work unless  instrument is in a standby state     Command  LoadMethod    QUI 1016 V2 2 Aug 06  126     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Parameters  Method Filename   Change the current loaded method on the TT24 7  If the instrument is in standby  the method will be changed immediately  If the instrument is running the method  will be changed at the end of sampling or at the end of the current run   Command  Split   Parameters  ON   OFF   Open or Close the split valve on the TT24 7    Command  Mode   Parameters  TUBEA   TUBEB   INJECTA   INJECTB   CONTINUOUS   Change the mode of operation of the TT24 7  Changing the mode will not take  effect if the instrument is in running mode    Command  StopFlow   Parameters  None   Sending command will set the sample flow rate to 0     Command  ResumeFlow  Parameters  None    Sending command will return the sample flow to the value specified in the method  file     QUI 1016 V2 2 Aug 06  127     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual    Markes Inter
63. chieve this  a system  containing two traps  A and B  is required  working in tandem  so that while one  is receiving and focusing sample  the other is desorbing into the analytical system   The process then alternates between each trap     The TT24 7 has therefore been designed as a twin trap system to facilitate NRT  analysis     2 2 TT24 7 Hardware configurations    There are three TT24 7 hardware configurations available  Ensure that you know  which configuration you have prior to reading this manual  see Introduction      TT24 7el1 standard system configured only for continuous sampling  TT24 7e2 standard system configured with the following additional features     tube desorption    pre purge to vent   dry purge    leak test  TT24 7e3 standard system configured with the following additional features     tube desorption    pre purge to vent   dry purge    leak test    split capability    2 3 Sampling modes    Depending on the hardware configuration of your TT24 7 there are two sampling  modes available via the User Interface     Continuous sampling   for continuous sampling from a stream of gas air   Tube Desorb   allows the desorption of a sorbent tube   Note  TT24 7e1 only allows the continuous sampling mode of operation  TT24 7e2 and TT24 7e3 allow both sampling modes of operation    2 3 1 Continuous sampling  Available with hardware configurations e1  e2 and e3    For continuous sampling  the    TT Sampling    mode of operation must be selected  in the TT24 7 softwar
64. cturers have different connectors  for a Remote Start cable  different cables are required for each GC     The N C socket is currently not used     QUI 1016 V2 2 Aug 06  94     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    3 1 3 Serial and   or USB Cable connections    The TT24 7 requires two serial connections with the controlling PC  Direct serial  connection therefore requires two spare COM ports in the PC  However many  modern PC systems do not have these serial ports available and may require  upgrading with an additional serial card     When using Microsoft Windows    98     2000    or XP     an extra serial port can be  generated using an available USB port and a USB to serial port conversion cable   P N U USBSR     Note  USB communication is NOT possible with Windows NT       The standard 9 pin RS232 serial cable is connected between the TT24 7 and the  USB   Serial conversion cable  The conversion cable may be connected directly to  a USB port on the PC or  if there are insufficient spare USB ports  into a USB hub   U USBHB  which can accommodate up to four connections  The hub is then  connected directly to one USB port on the PC     Figure 69 shows a cabling schematic using direct serial connection from the TT24   7 and the PC  and the remote start connection     Serial Cable to Sampler connection     P N Z 0189   Remote Start    Serial Cable to MFC connection   P N 2 0189
65. cycle time     The sampling time can be fixed to the specified value i e  preventing extended  sampling within the TT24 7 software  Under these conditions if the GC is not  ready then after the fixed sampling time the system then waits until the GC  becomes ready or again the time out value stops the process  See section  2 9 1 5 3 for extended sampling and fixed time sampling     2 9 1 1 6 Trap low temperature    This is the  lower  trap temperature value used to retain compounds on the  sorbent bed  Values ranging from 15 to 50  C are possible  Sub ambient values  are possible as a consequence of the Peltier cooling elements situated  immediately below each trap     The trap minimum value is sensitive to the flow path temperature described  above  the flow rate of sample passing through the trap and the temperature  of the air sample  For example if the flow path temperature is set to 200  C   with a fast flow rate  i e   gt   700 mL min   then the trap minimum value may  be 25  C or higher at faster flows  If additionally the sample gas is above  ambient then this minimum value will be higher again     Using lower sampling flows will assist this value  and can indeed improve on  signal to noise for the resulting chromatographic peak  See section 2 9 1 1 5  above      Ultimately the minimum trap temperature achievable is determined  experimentally by reviewing the relationship between the set point and actual  values as shown in the status bar  section 2 9 1 3  and this wil
66. d       valve       Figure 82  Precise positioning of end of fused silica transfer line    If the fused silica terminates in the space above the end of the inert coated  stainless steel tube  some analytes can diffuse into the side arm causing  measurable peak tailing  The 0 25 mm I D  fused silica transfer line must be  installed through the union and on into the narrow bore part of the inert coated  stainless steel tube     QUI 1016 V2 2 Aug 06  105     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    When connecting the transfer line to the TT24 7e2 e3  pull about 20 cm of fused  silica from the PTFE line casing  Slide a   is inch stainless steel Swagelok nut and  a    V ie inch x 0 4mm ferrule  supplied with shipping kit  see Appendix 1  onto the  end of the fused silica and then cut off the first few mm of fused silica  Feed the  fused silica into the top of the union and slide the ferrule into position  Screw the  nut onto the union and  with the fused silica still loose  position it so that the fused  silica is inserted 50 mm into the union  Tighten the nut to trap the fused silica  and then tighten a further half turn using one of the 8 mm wrenches  spanners   provided in the shipping kit  Do not over tighten or the ferrule will become  distorted     Carefully bring the clamp plate  PTFE plate and shield tube down into position   shown in figures 83  amp  84  with the shield tub
67. d a split vent line     When this module is connected to the TT24 7 system only the carrier supply  line and septum purge line are used  the split vent line is redundant  For  correct closed loop feedback control the pressure sensor for the inlet must be  connected into the septum purge line     The carrier  septum purge and split vent lines usually connect into the electronic  module by means of a single manifold  e g  Agilent technologies  or individual  connectors  e g  Thermo Electron   To protect the original gas lines  for future    QUI 1016 V2 2 Aug 06  28     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    re assembly of the capillary inlet  it is recommended that these lines are  replaced with additional gas lines rather than being cut     Figure 13 shows the back panel on the TT24 7 for connection of the carrier gas  supply  carrier inlet  and pressure monitoring line  pressure outlet      i a     TOLLO  E  T  TS  T  iiit   T  TT  TOLL  TO    Pressure outlet  Carrier inlet                     Figure 13  Carrier gas supply connections on the back panel of the TT24 7    A flow diagram showing the additional electronic flow module and gas lines to  the TT24 7 is shown in figure 14 and further details are given in section 2 5     Carrier Gas Supply Line  GC Pneumatics Module    From GC          Desorption Tube Pressure sensor    Needle    HY C i    H L  gt    Septum Purge  MFC  a
68. d traps    The quartz cold traps are fragile and packing them is a skilled task that should be  undertaken with care  Both traps in the TT24 7 should be identically packed and  should be flow tested prior to use to ensure that they have similar impedences   For these reasons we recommend the use of Markes pre packed TT24 7 cold traps   either one of the application specific traps  or cold traps which are custom packed  for your specific application     3 4 4 1 Packing cold traps    If you wish to pack your own traps Markes supply empty traps for this purpose   The traps should be packed from the wider bore end using the following  procedure     Insert a small plug of quartz or glass wool  depending on the application    using a suitable flexible tool     Pour in the required amount of sorbent s      If multiple sorbents are to be used  the weakest sorbent should be  inserted first  followed by the stronger sorbent  Sorbents should be  separated by small plugs   2 mm  of quartz   glass wool as above     Finish the trap with a 5 mm plug of quartz   glass wool     A 6 cm length of the trap  measured from the point of bore restriction  is  subjected to full heating   cooling power  All the trap packing   wool plugs   except the final wool plug  should be within this 6 cm length     Note  Care should be taken when using tools   funnels etc  to insert wool  and sorbent that the edge surface of the traps are not damaged or chipped  in any way as this will prevent them sealing correc
69. ded Conditioning Temperature   Recommended Desorption Temperature   Notes     Markes International    Boiling point 50  C to 200  C   VOC s within volatility range above   290  C   280  C to 290  C   No higher than 290  C    Non polar  requires extensive  conditioning  high artefact  background     10  Chromosorb 106     range of mesh sizes available     Sorbent Strength   Specific Surface Area  m2 g    Approximate analyte volatility range     Example Analytes     Sorbent Maximum Temperature   Recommended Conditioning Temperature   Recommended Desorption Temperature   Notes     Medium    750   n Cs to n Cy gt    Boiling point 50  C to 200  C    Hydrocarbons  benzene  volatile  oxygenated compounds    225  C to 250  C   225  C to 250   C   No higher than 200  C  High artefacts   10 ng     For trace level analysis condition at  225  C  desorb sample tubes no  higher than 200  C to reduce  background   Hydrophobic    Inert     suitable for labile compounds    11  Spherocarb      UniCarb     60   80 mesh only     Sorbent Strength   Specific Surface Area  m2 g      Approximate analyte volatility range     Example Analytes     QUI 1016 V2 2 Aug 06    Markes International Ltd  T   44  0 1443 230935     132     F   44  0 1443 231531    Strong     1200   also operates on molecular  sieve principle    C3 to n Cg  Boiling point  30  C to 150  C    Very volatile compounds e g  VCM   ethylene oxide  carbon disulphide   dichloromethane  chloromethane   Volatile polar compounds e g   methan
70. djacent numbers  The first represents the actual  value and the second the set point value  If these two numbers are equal they  appear in black font  if however the actual value is not at the set point it will  appear blue  If the TT24 7 is started when any one of these values is blue  then  the software will enter an equilibrating mode and the run will only commence  when the value becomes ready  These values have to be equal   2  C tolerance  value  for the system to become ready     The transfer line and heated valves are controlled collectively by the flow path  temperature  see section 2 9 2 1 3      2 9 2 3 3 GC    The cable connecting the TT24 7 to the associated GC  see section 3 1    performs two functions  Firstly it monitors the GC ready status as specified in  the GC interface logic of the TT24 7 software  see section 2 9 2 5 2   If the GC  set point values are correct and equilibrated  and the software for the GC is in  a state to accept a start signal from the TT24 7  then the GC will be ready  This  ready status is relayed to the TT24 7 and observed in the GC status box  If the  GC is not ready  then a    Not Ready    comment appears     When the GC is ready the second function for the cable is to send the start  signal from the TT24 7 to the GC at the point of trap fire so it can begin its  analytical process  If the GC is not ready at the end of the sampling time  then  extended or fixed time sampling commences  see section 2 9 2 5 3      QUI 1016 V2 2 Aug 0
71. ds     QUI 1016 V2 2 Aug 06  103     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International       Figure 79  Moving the clamp plate  PTFE plate and shield tube down into  position       Figure 80  The transfer line installed and the 8 way orange connector plugged  in    Plug the orange 8 way connector into the socket adjacent to the transfer line  connection  See figure 80 for the correct orientation of this connector     Replace the heated valve box and M4 nuts and re fix the back panel by  reversing the procedure above     QUI 1016 V2 2 Aug 06  104     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    3 3 3 2 e2 e3 systems    For e2   e3 systems the transfer line inlet is situated at the back of the third  heated valve and the transfer line is installed horizontally into this union     Figure 81 shows the     ic inch stainless steel union into which the fused silica  transfer has to be fitted  Figure 82 shows a partly sectioned view of this   16   inch union giving an indication of where to position of the end of the fused silica  tubing     CHCE   SA  Transfer a  installed here       Figure 81  Installation of fused silica tubing    Carrier bypass    Fused silica        terminates here  drilled out    1 16 inch inert  oated stainless  steel tubing  0 5  mm bore     Analytes from the    trap via the heate
72. e     Gas samples which are above atmospheric pressure  Max 50 psi  can be analysed  by the TT24 7 system  For samples where the pressure is at or below atmospheric  pressure a vacuum pump is required to pull the sample through the instrument   Sample flow is regulated for all pressures using an internal mass flow controller   MFC  capable of controlling gas flows from 50 mL min to 1 L min  Sample gas  is drawn into the TT24 7 through a 1   4    Swagelok fitting on the right hand side of  the instrument close to the valving assembly  This union could be extended with    QUI 1016 V2 2 Aug 06  16     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    additional plumbing to the exterior walls of the TT24 7  to facilitate connection of  a remote sample line or sampling bag  figure 2         Figure 2  Sample gas inlet on TT24 7    Continuous sampling starts by initiating the TT24 7 control software  see section  2 9   which in turn starts regulating the sample flow rate and time of the gas  stream to be analysed  This results in a total volume of gas sampled onto each  trap  After the sampling period the trap is purged for a defined period of time to  remove residual air   moisture  After purging the trap  the system enters the trap  fire phase in which sample is backflushed from the trap into the capillary column   Backflushing the trap with carrier gas provides an efficient mechanism to swee
73. e 87   Figure 88   Figure 89   Figure 90     QUI 1016 V2 2 Aug 06    Loosen the back panel screws  Remove the M4 nuts  Transfer line connection    Markes International    Precise positioning of end of fused silica transfer line    Moving the clamp plate  PTFE plate and shield tube down into    The transfer line installed and the 8 way orange connector    Installation of fused silica tubing    Precise positioning of end of fused silica transfer line    Aligning the clamp plate  PTFE plate and shield tube into position    Clamp plate  PTFE plate and shield tube in position    Plugging in the 8 way connector    TT24 7el1 with overlaid schematic showing trap positions and  user serviceable parts    TT24 7e2 e3 system with Desorb Tube option on LHS  TT24 7e2 e3 system   tube sealing mechanism lever lifted up    TT24 7e2 e3 system removing desorb tube  TT24 7e2 e3 system with Split tube option on RHS    Markes International Ltd  T   44  0 1443 230935     10     F   44  0 1443 231531    E  enquiries markes com    101  102  102  103    104    104  105  105  106  106  107    109  112  113  113  114    TT24 7 User Manual Markes International    TT24 7 Operators Manual    Introduction    This user guide has been designed so that the operator of a TT24 7 system can  start an analysis as quickly as possible after installation     There are four principle sections to the manual     Section 1 Quickstart Guide   for those users who simply want to start analysing  samples    Section 2 The TT2
74. e covering the union nut  The shield  tubing should be positioned such that the   is inch side tubing projects through one  of the cutouts  As the transfer line is lowered into position the clamp plate and  PTFE plate should fit onto the two exposed threads        Figure 84  Clamp plate  PTFE plate and shield tube in position    QUI 1016 V2 2 Aug 06  106     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Use two M4 nuts  supplied with shipping kit  to lock the PTFE plate onto the  exposed threads     Plug the orange 8 way connector into the socket adjacent to the transfer line  connection  See figure 85 for the correct orientation of this connector        Figure 85  Plugging in the 8 way connector    3 3 4 Coupling the fused silica transfer line to the GC column    Return to the GC and connect the column to the end of the transfer line using the  quick seal column connector provided with the shipping kit  Instructions are  included with the connectors  Alternatively you may use a       inch zero dead  volume connector with the appropriate ferrules     3 3 5 User maintenance of transfer line    The quick seal column connector should be replaced every time the seal between  the fused silica transfer line and the GC column is broken  If a zero dead volume  connector is being used the ferrules should be replaced as necessary when the  connection is broken     The fused silica transfer lin
75. e itself  and its accompanying PTFE sleeve  should be  replaced as part of a scheduled program of maintenance  Frequency of changing  the transfer line will depend entirely on the usage level of the system and the  nature of the compounds being analysed  In any event it is advised to replace the  fused silica transfer line at least annually     QUI 1016 V2 2 Aug 06  107     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    3 4 Installation and removal of Cold Traps  Note  Never turn on the power to TT24 7 without the cold traps installed    3 4 1 Installing cold traps  Refer to Figure 86       Loosen the locating screw on the trap pneumatics before sliding the  pneumatics forwards       Slide the narrow bore end of the cold traps into the trap box until they  touch the o rings in the heated valves  3 5 x 1 5 mm o ring  P N  U COV35        Switch on the instrument and download the firmware  see section 2 8      Allow the heated valves to reach their temperature set point     Push the cold traps the remaining 2 3 mm into the heated valve o rings     Fit the spacer o rings  P N U COV39  onto the end of the trap    e For collared traps use two o rings pushed up against the collar    e For old style uncollared traps use a sufficient number of o rings to fill  the space between the trap valve connector and the trap guides  see  figure 86       Replace the trap pneumatics at the front of the in
76. eak shape   peak splitting    Peak broadening  particularly of early eluting components  is often an early  indication that the cold trap packing needs changing  The trap sorbent is  subjected to rapid heating during the analysis of every sample and should  therefore be replaced regularly     Normal aging or the desorption of samples containing aggressive compounds can  produce activity in the transfer line or in the analytical column itself  This results  in peak broadening or tailing  If this occurs  the capillary column or the fused  silica insert inside the transfer line should be replaced     A poor connection between the transfer line and the analytical column will also  distort peak shapes  To avoid this  the connecting ends of both the column and  the transfer line should be cut cleanly using a fused silica column cutting tool  The  union or connector assembly should be a quartz quick seal connector or an inert   zero dead volume fitting recommended for butt connecting capillary tubing     Broad peaks can also result from the selection of too strong an adsorbent in the  cold trap or from low carrier gas flows through the trap during desorption  The  gas flow through the cold trap during secondary desorption  i e  the column flow   should be at least 3 ml min for optimum peak widths     If a cold trap is loaded with relatively large quantities   gt 1 mg  of water or solvent   flash vapourisation of the solvent or water may result in a temporary pressure  surge causing 
77. ection configures the flow calibration of the mass flow  controller  The selection made here is also shown in the flow  controller status box      87     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 9 2 5 2 Ports    Gas Ports   System   E Mail              Communications Port    Analyser Port cow  x  MFC Port   7   Baud Rate  57600 x  Baud Rate  57600          M GC Interface Logic    GCStart  out  GC Ready  in     Open   Start     Open   Ready     Closed   Start    Closed   Ready                            Figure 63  Options dialogue box   ports tab    This screen consists of two sections  the communications port  and the GC  interface logic  These two sections control communication between the TT24 7  and the associated PC system  and the ready   not ready and start signal logic  between the GC and the TT24 7     Communications port     The TT24 7 has two serial ports at the rear of the instrument  as shown  in figure 67 section 3 1  These are the serial interfaces for the mass  flow controller and the TT24 7 analyser     Each port requires its own dedicated communications  COM port   connection from the PC  This can be provided in two ways  i e  using two  separate serial port cables from the PC system  or if two serial ports are  not available using USB hub and USB to serial connection cables  N B   Not supported by Microsoft Windows NT   See section 3 1 for more  details     U
78. een the GC and the TT24 7     Communications port     The TT24 7 has two serial ports at the rear of the instrument  as shown  in figure 67 section 3 1  These are the serial interfaces for the mass  flow controller and the TT24 7 analyser     Each port requires its own dedicated communications  COM port   connection from the PC  This can be provided in two ways  i e  using two  separate serial port cables from the PC system  or if two serial ports are  not available using USB hub and USB to serial connection cables  N B   Not supported by Microsoft Windows NT   See section 3 1 for more  details     Under normal circumstances the default baud rate values of 57600  should be used  However if there are communications problems then a  lower baud rate be specified i e  38400     GC Interface logic  This section of the software affects the start   stop interaction between    QUI 1016 V2 2 Aug 06  73     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    the TT24 7 and the host GC   GCMSD system  The actual configuration  is dependent on the model of GC   GCMSD used  For Agilent  Technologies and Thermo Electron instrumentation the GC Start  out   and GC Ready  in  settings should be in the    closed    state     For other GC manufacturers please contact Markes International for details   2 9 1 5 3 System    Gas   Ports System   E Mail            m Display Flaw  JV Enable    Enabling this option
79. eer  diagnostics  see section 2 9 2 6      Starts the TT24 7 with the loaded method conditions    Opens the Stop Run dialogue box which asks whether you  wish to    Stop Immediately    or    Continue running        86     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual    2 9 2 5 Options    Markes International    Options is the principle configuration section for the TT24 7 system  It is  accessed from the    View    menu item  see 2 9 1 4 2      Within options  there are four separate configuration screens     2 9 2 5 1 Gas    Gas   Parts      System   E Mail           PSI  kPa    Pressure Units          Carrier Gas    Sample Gas        He      air 7                 Figure 62  Options dialogue box   gas tab    This section defines the pressure units and the sample and carrier gas selection     Pressure units     Carrier gas type     Sample gas type     QUI 1016 V2 2 Aug 06    The pressure unit selection is either psi  or kPa   Note 1 psi   101 325 kPa     In Tube Desorb mode the desorb pressure is displayed on the  instrument status bar  see section 2 9 2 3 1    Choices of Helium  He   Nitrogen  N gt    Hydrogen  H gt    and  diagnostic air are available  The selection made here is  shown in the flow controller status box as part of the  instrument status  See section 2 8 1 3 4   Choices of Air  No  He  Hz  safety considerations may apply  using this gas   and Carbon Dioxide  COV  are available  This    sel
80. er Manual Markes International    2 5 3 1 2 Trap A sampling  trap B dry purging    Desorption Tube    q Needle   N valve        n       D O  TA Y   MFC  amp  Pump Sample in  off      Trap B       Dr    Needle    valve Split Tube    on  _  aw      Ne    KEY          Flow of Carrier Gas      Optional Carrier Gas Flow  Carrier Gas  no flow             Flow of Sample Gas  Sample Gas  no flow        Figure 33  TT24 7e3   Continuous sampling   trap A sampling  trap B dry   purging   By starting the TT24 7 software the continuous sampling process commences    The sample passes initially onto trap A  The date and time that the sampling   started  and the trap that is being sampled are reported as an information   message in the reporting module of the software  see section 2 9 2 7   e g     05 12 2005 10 49 31   INFO  Sampling Started on Trap A  At the same time trap B dry purges with carrier gas in the sampling direction   This removes air and moisture from the trap prior to firing  The dry purge flow    rate through the trap  B  at this time is controlled by a needle valve connected  to SV6     Carrier gas also purges down the split line to purge any air from the line and to  prevent ingress into the system     The sample gas flow rate passing into the TT24 7 is controlled by the internal  mass flow controller  MFC      QUI 1016 V2 2 Aug 06  50     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes Internationa
81. eve  one or more of the  large washers supplied will be needed  If the heated zone block is particularly  deep both the M6 spacer nut and spacer tube will needed as shown     Large washer  Silicone foam rubber Heated zone block  insulation    Large washer    Spacer nut M6    Spacer tube    Spacer nut M6       Figure 74  Installation of transfer line through heated zone block    This part of the line derives its heat from the heated zone block which should be  set to run at a conveniently high temperature  preferably 50  C above the line  setting but not above 250  C as the silicone foam rubber insulation will be  damaged     The parts supplied can be used in other combinations to suit particular instrument  configurations     3 3 2 Installing the fused silica transfer line insert    Once the heated line has been fitted to the GC  the fused silica plus associated  PTFE sleeving  P N UTD 5093 see Appendix 1  are pushed from the GC end  along  the   s inch aluminium tube until they protrude from the other  TT24 7  end of the  transfer line     QUI 1016 V2 2 Aug 06  100     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    3 3 3 Connecting the transfer line to the TT24 7    Place the TT24 7 on the bench on the most convenient side of the GC  Ensure  that the transfer line will reach from the back of the TT24 7 to the selected entry  point into the GC oven     3 3 3 1 e1 systems    For e1 sys
82. g  breakdown of the sorbent may severely contaminate the flow path of  your system  This may require the complete replacement of the flow  path and associated components by a fully qualified service engineer  and would not be covered by the instrument warranty     When multi bed traps are being used  the maximum temperature that can be  set is that relating to the sorbent with the lowest allowable temperature   Consideration must then be taken of the recovery efficiency for analytes from  sorbents which go to higher values     2 9 2 1 12 Trap hold time   available in both sampling modes    This is the time interval when the trap is held at its maximum value  Time  intervals from 0 1 to 10 minutes are possible  however values in excess of    0 5 minutes are typically used  This value must be long enough to ensure  complete removal of analytes from the trap into the analytical column   Consideration of the trap flow conditions at trap fire are required  i e  the flow  through the trap at trap fire is equal to the column flow rate which could be as  little as 3mL min  In this case a time value less than 1 minute is not  recommended     Incorrect setting of this time can be the cause of poor sensitivity for certain  compounds  as insufficient time will reduce the amount of sample leaving the  trap     2 9 2 1 13 Save   available in both sampling modes    Once a method parameter has been changed then the word    modified    appears  alongside the method name in the blue bar at the
83. he TT24 7  Software     ITSO15   This is a laminated card  which is supplied separately with  the TT24 7  It is intended to help operators with a quick setup of the system  where a more detailed knowledge is not required     For the Quickstart approach to work successfully  it is assumed that the TT24 7  has been installed and tested by a qualified engineer  Quickstart uses a TT24 7  hardware configuration of e2 e3 to demonstrate the running of the system   however the instructions are also valid for TT24 7el1 systems     More detailed information about the instrumentation and software is supplied in  subsequent sections of this document   Note  Never turn on the power to the TT24 7 system  without the cold traps installed  see section 3 4     1 1 Running the TT24 7 system    The TT24 7  e2 and e3  can be operated in two sampling modes  continuous  sampling  and single tube desorption  The TT24 7e1 can only be operated in  continuous sampling mode     As the principle operating mode for the TT24 7 is continuous sampling this will be  included in the Quickstart setup  Both sampling modes are discussed in detail  later in the manual  Section 2 2      1 1 1 Using an existing TT24 7 method    Running the TT24 7 system in its continuous sampling mode using an existing  method is a simple four step process     Step 1  Loading the TT24 7 method    To load the required TT24 7 method either click on the menu bar icon at the top  of the software screen and select the desired method  or
84. hrough the trap     Sample flow rates from 50 mL min to 1 L min are possible  with 0 1 mL  increments  however the maximum flow rate is affected by the sorbent bed  depth and mesh size  Where fine mesh size material is used  i e  80 100 mesh  or finer  the maximum flow rate will be reduced  Ultimately the fastest flow  rate achievable is determined by specifying the maximum flow rate of 1  L minute and observing the actual flow value on the status bar  see section  2 9 2 3  at the bottom right side of the user interface     In certain applications where the analyte concentration is very low  i e  sub  ppb  and there is a time restriction on sampling  i e  NRT  there is a temptation  to set a very fast sampling flow rate to maximise the amount of analyte passing  into the trap in unit time  For example sampling at 1 L min for 10 minutes  results in a sampling volume of 10L  If the analyte concentration is very low  e g  in the pg L concentration  i e  ppt   then a 10L volume would contain  sufficient sample in theory for detection by selective GC detectors or by GCMS     However sensitivity is very dependant on the signal to noise  S N  ratio of the  resultant chromatographic peak  If this is sharp i e  peak widths  lt   5 seconds  then much better integration and detection is possible than for peak widths  gt  15  seconds     In practice it has been shown that very fast sampling flow rates   gt   800  mL min  with certain sorbent materials produces poor or broad peak shapes  
85. ility range above   oxygenated compounds    Example Analytes     Sorbent Maximum Temperature  250  C   Recommended Conditioning Temperature  225  C to 250  C  Recommended Desorption Temperature  No higher than 225  C  Notes  High artefacts   10 ng     For trace level analysis condition at  225  C desorb sample tubes no higher  than 200  C to reduce background levels    Low maximum temperature   repack  tubes after 50 thermal cycles    Inert    8  Porapak N  range of mesh sizes available   Sorbent Strength  Medium  Specific Surface Area  m2 g    300  Approximate analyte volatility range  n Cs to n Cg   Boiling point 50  C to 150  C    Volatile nitriles  e g  acrylonitrile   acetonitrile  propionitrile  Pyridine   volatile alcohols  ethanol  methyl  ethyl ketone    Example Analytes     Sorbent Maximum Temperature  190  C  Recommended Conditioning Temperature  180  C to 190  C  No higher than 180  C  Notes  Hydrophobic  High artefacts   10 ng     For trace level analysis condition at  180  C and desorb sample tubes no  higher than 160  C to reduce  background levels    Recommended Desorption Temperature     9  HayeSep D  range of mesh sizes available     Sorbent Strength   Specific Surface Area  m2 g    Approximate analyte volatility range     QUI 1016 V2 2 Aug 06    Markes International Ltd  T   44  0 1443 230935    Medium   795  n Cs5 to n Cy gt     F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual    Example Analytes    Sorbent Maximum Temperature   Recommen
86. ime of the TT24 7 system needs  to be kept to a minimum     2 9 1 1 3 Flow path temp    This is the set point temperature for the heated valves and the transfer line to  the GC  The temperature range extends from  100  C to 200  C  The  temperature should be set high enough to prevent any condensation of sample  within the valve   transfer line and values less than 120  C are rarely used   However this temperature value will also affect the lower trap temperature  value     QUI 1016 V2 2 Aug 06  65     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    If the flow path temperature is set to its maximum of 200  C then the trap  temperature cannot be controlled at values less than 25  C  If the flow path  temperature is set to 150  C then a minimum trap temperature of  15  C is  possible     2 9 1 1 4 Sample flow rate    This specifies the flow rate of sample vapour passing through the traps  The  sample may be at atmospheric pressure  in which case a vacuum pump is  required to pull sample into the system  The flow rate is then regulated by the  internal mass flow controller  MFC   If the sample is at positive pressure  Max  50 psi   then the vacuum pump is not required  and again the MFC controls the  flow through the trap     Sample flow rates from 50 mL min to 1 L min are possible  with 0 1 mL  increments  however the maximum flow rate is affected by the sorbent bed  depth and mesh size
87. ing   this is vital  to avoid any problems with condensation    Note also that if 2 or more sorbents are to be packed into a single tube  all the  sorbents must have similar maximum temperatures  If one or more of the  sorbents has a significant lower maximum temperature than the other s  in the  tube  it will not be possible to stringently condition the more stable sorbents  without exceeding the temperature limitations of the less stable material  For this  reason Chromosorb and Porapak sorbents are not recommended to be used in  mixed sorbent bed tubes     QUI 1016 V2 2 Aug 06  136     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Long term storage of clean and sampled tubes    Conditioned or sampled sorbent tubes should always be stored using 1   4    brass  Swagelok type screw caps fitted with combined PTFE ferrules  It is recommended  that these be tightened by hand plus a further quarter turn using conventional  spanners   wrenches or  ideally  a Markes International CapLok    tool  The Cap   Lok tool was invented by scientists at the French Environmental Centre   INERIS    and prevents over tightening and distortion of the PTFE ferrules     It is not necessary to store capped tubes in refrigerated conditions   except as  stated above  If refrigeration is to be used  caps must be retightened   approximately a quarter turn  using the CapLok tool once they have reached their  st
88. inutes in  increments of 0 1 minute and is typically 0 5 to 1 minute  although this can be  reduced where the cycle time of the TT24 7 system needs to be kept to a  minimum     2 9 2 1 3 Flow path temp   available in both sampling modes    This is the set point temperature for the heated valves and the transfer line to  the GC  The temperature range extends from  100  C to 200  C  The  temperature should be set high enough to prevent any condensation of sample  within the valve   transfer line and values less than 120  C are rarely used   However this temperature value will also affect the lower trap temperature  value     If the flow path temperature is set to its maximum of 200  C then the trap  temperature cannot be controlled at values less than 25  C  If the flow path  temperature is set to 150  C then a minimum trap temperature of  15  C is  possible     QUI 1016 V2 2 Aug 06  79     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 9 2 1 4 Sample flow rate   available in TT sampling mode    This specifies the flow rate of sample vapour passing through the traps  The  sample may be at atmospheric pressure  in which case a vacuum pump is  required to pull sample into the system  The flow rate is then regulated by the  internal mass flow controller  MFC   If the sample is at positive pressure  Max  50 psi   then the vacuum pump is not required  and again the MFC controls the  flow t
89. ional can supply a secondary pressure regulating accessory  P N  U GASO01   which regulates both the dry gas and the carrier gas pressure     The dry gas is attached at the rear of the instrument to the connector labelled     Purge Gas        The sample gas  typically air  will contain both oxygen and moisture  The air is  removed from the trap just prior to trap fire using the dry purge   pre purge gas  flow  Moisture can also be removed by prudent selection of sorbent material  e g   Tenax does not retain moisture  see Appendix 3   If a hydrophilic sorbent is used  the water will be transferred across to the analytical system  Split mode sampling   where available with e3 systems  will reduce this volume     3 2 1 User maintenance for gas requirements    Standard good practice should be observed for the maintenance of the gas lines  serving the TT24 7 and in particular the replacement   regeneration of any filters  on the system  de oxo filters  moisture traps  hydrocarbon traps etc    Attention  should be paid to the manufacturers instructions of these filters with respect to  frequency of regeneration   replacement     3 3 Installation  and removal  of heated transfer line    3 3 1 Connecting the transfer line to the gas chromatograph    The TT24 7 is supplied with a universal transfer line to deliver desorbed analytes  from TT24 7 to a gas chromatograph or other analytical system  The sample path  utilises a deactivated fused silica line  0 25mm I D  and 0 35mm O D   P N
90. l    2 5 3 1 3 Trap A sampling  trap B desorbing    Desorption Tube    Needle  Valve    MFC  amp  Pump Sample in       Neots oft  Split Tube    on  tO          Vent  G        KEY       gt     Flow of Carrier Gas      Optional Carrier Gas Flow  Carrier Gas  no flaw          _ Flowof Sample Gas  Sample Gas  no flow        Figure 34  TT24 7e3   Continuous sampling   trap A sampling  trap B desorbing    The first two runs of the continuous sampling mode  i e  trap A and trap B   would typically not be used for quantitative analysis of the sampled air but as  a clean up process for both traps  After both traps have been fired once then  the subsequent samples may be quantified     After purging the trap  B  for the first time  trap fire occurs and flow is directed   backflushed  across to the capillary column  This first  B  trap fire prepares  the trap for subsequent sampling     If the split function is enabled in the method   see section 2 9 2 1 14   then this  will occur at trap fire  Trap B Desorbing   The flow rate through the trap is  therefore equal to the column flow plus the split flow through SV7  The split  ratio is therefore     Split Ratio   Column Flow  Column flow   Split flow    If the split function is not enabled then the flow rate through the trap is equal  to the column flow rate alone and should be at least 3 mL min     QUI 1016 V2 2 Aug 06  51     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Man
91. l    Packing Lists   e2 specific parts    Description    QUI 1016 TT24 7 User Manual    Quick Start Guide for TT24 7 software    a       QUI 1016 V2 2 Aug 06  122     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Packing Lists   e3 specific parts    Description    pee eee A  7  Quick Start Guide for TT24 7 software    SERTTD 1044  e2 e3 Inlet support bracket  ITS014 TT24 7e3 Quick Reference Guide    a       QUI 1016 V2 2 Aug 06  123     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Routine Maintenance Spares   common parts  e1  e2 and e3 systems     U FV001 Pk 10 ferrule     16    graphitized vespel  0 4 mm id  hole  to connect fused silica transfer line to TT24 7   U FV002 Pk 10 ferrules    16    X     1e    graphitized vespel  use  with PEEK tubing i e  connection of carrier gas  lines      U FV003 Pk 10 ferrules    s    x   16    graphitized vespel    U FV005 Pk 10 ferrules    s    graphitized vespel  2 mm id  hole    transfer line insert with PTFE sleeve    TT24 7 cold trap optimised for high boiling point  reactive compounds such as the chemical  warfare Agent VX    T 2HD TT24 7 cold trap optimised for the chemical  warfare agent mustard  HD     T 3GBGE TT24 7 cold trap optimised for the more volatile  chemical warfare agents such as GB and GE  the G   analog of VX     T 6EMP TT24 7 emp
92. l be a function  of the flow path temperature and the sampling conditions     2 9 1 1 7 Trap high temperature    This is the upper temperature setting for the trap  Values ranging from 50  C  to 400  C are possible  The upper trap temperature should be set to the  optimum value which enables 100  recovery of analytes from the sorbent bed   but does not exceed the maximum permissible temperature for that  sorbent  See Appendix 3  Keeping the maximum trap temperature as low as  possible whilst still enabling 100  analyte recovery will extend trap lifetime     Note  If you exceed the maximum sorbent temperature  the resulting  breakdown of the sorbent may severely contaminate the flow path of  your system  This may require the complete replacement of the flow  path and associated components by a fully qualified service engineer  and would not be covered by the instrument warranty     When multi bed traps are being used  the maximum temperature that can be  set is that relating to the sorbent with the lowest allowable temperature     QUI 1016 V2 2 Aug 06  67     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Consideration must then be taken of the recovery efficiency for analytes from  sorbents which go to higher values     2 9 1 1 8 Trap hold time    This is the time interval when the trap is held at its maximum value  Time  intervals from 0 1 to 10 minutes are possible  however values i
93. le consists of three solenoid valves SV1  SV2  and SV3  See section 2 5  for full details of flow schematics     QUI 1016 V2 2 Aug 06  23     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    These three way  on   off  solenoid valves control the source and flow direction of  gas passing through the traps  These flows consist of     The sample gas for analysis  continuous or tube   The carrier gas to backflush the traps  The purge gas to flush the traps of air and moisture prior to trap fire    SV1 and SV3 are connected to the traps A and B  and SV2 connects sequentially  to both valves     When either SV1 or SV3 are ON they are connecting the associated trap   SVi trap A  SV3 trap B  to the mass flow controller for any of the trap sampling  modes discussed in section 2 3     When either SV1 or SV3 are OFF they are either receiving carrier gas from SV2 to  backflush the trap at trap fire into the capillary column  or they are receiving  dry purge gas from the traps to remove air and moisture in the trap     2 4 3 Peltier cooled traps A and B    The TT24 7 contains two identical quartz cold traps which are located within the  trap housing  figure 10   The two traps lie parallel to each other and connect at  one end into a heated valve and at the other end into the pneumatics assembly   The location of the trap box is shown in Figure 9     Trap A       Figure 10  Position of Trap A and Tra
94. lied to the  capillary column  No other gas flow occurs within the TT24 7 unit during this  time     QUI 1016 V2 2 Aug 06  54     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 3 2 2 Leak testing tube    MFC  amp  Pump Sample in    Trap B    off  iii       Needle    valve Split Tube    KEY       gt     Flow of Carrier Gas      Optional Carrier Gas Flow  Carrier Gas  no flow        gt    _ Flow of Sarnple Gas  Sample Gas  no flow        Figure 38  TT24 7e3   Tube desorb mode   leak testing tube    The first stage in tube desorption is a tube leak test  This is essential as the  tube has been inserted into the TT24 7 flow path and sealed with two retaining  o rings one at the front and back of the tube oven  See section 3 5 for tube  installation instructions     During this phase the tube desorption flow lines in the TT24 7 are initially  pressurised with carrier gas and then solenoid SV5 is switched OFF to seal the  system  The heated valves are configured so that there is no connection  beyond HVC  The pressure sensor P adjacent to SV5 is monitored to measure  the rate of pressure decay  If this is acceptable   lt  5  drop in pressure within  30 seconds  the software moves on to the next phase in the tube desorption  process     QUI 1016 V2 2 Aug 06  55     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes I
95. low controller  flow rate and associated values are displayed in the top  level status bar  see section 2 8 1 3 4     This option fixes the sampling time in continuous  sampling mode  No extended sampling will take place  if the GC is not ready when this is active  see 2 8 1 1 5   for further information regarding sample time      Sampling Extended Timeout  In continuous sampling mode  the TT24 7 checks    QUI 1016 V2 2 Aug 06    Markes International Ltd     on the GC ready status at the completion of the  sampling time  If the GC is ready  then the system  continues into dry purge and then trap fire  However if     89     T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    the GC system is not ready e g  the oven has not cooled   down to its set point value and equilibrated  then the   TT24 7 will automatically go into an extended sampling   mode until the GC becomes ready  The extended   sampling time can be limited to a fixed time period by   entering a    Time to extend sampling    and ticking the     Enable Timeout    box     As a result of extended sampling the amount of sample  in the trap will be greater than that if extended sampling  had not occurred  The analytical result will therefore be  greater  and this may need to be compensated for in the  final quantitative data     The total sampling volume is reported in the reports  deviation screen  which allows the re calculation of the  correct amount of an
96. lumns are available which have a built in length of uncoated tubing at the end  of the column  This length can be inserted through the PTFE sleeve described  above and connected directly into the TT24 7  The advantage of this method is  the exclusion of a connector within the transfer line  which may be the source of  problems with leaks  or activity  see Section 4   Troubleshooting for further  details   The disadvantage of this technique is that changing columns needs  disconnection at the TT24 7 fitting and removal from the transfer line     Where a single column is likely to be used consistently the latter approach is  recommended  where multiple columns are to be used in an interchangeable  manner the former is advised     2 4 5 Tube desorption and sample split pneumatics    These sampling options have been discussed in general in section 2 3 2 and 2 3 3   Additional information can be found in section 3 0   User Installed Components  and Maintenance and the sampling schematics can be found in section 2 5     2 4 6 Electronic pneumatic control  EPC     Two gases pass into the TT24 7 system during its operation  excluding the trap  box purge flow   namely carrier gas and sample gas     The sample gas is typically air however other gas types can be accommodated i e   nitrogen  hydrogen  helium and carbon dioxide  Control of the sample gas flow  rate is achieved using an internal mass flow controller with an operating range  from 50 mL min to 1 L min  The gas type must be s
97. m tab   Figure 55  Options dialogue box   email tab   Figure 56  Flow diagnostics screen   Figure 57  Reporting module   Figure 58  User interface for each sampling mode   Figure 59  Flow Display showing flow direction through solenoid valves  Figure 60  Instrument status bar   e2 e3 configurations   Figure 61  Software Icons and menu bar for configurations e2 e3  Figure 62  Options dialogue box   gas tab   Figure 63  Options dialogue box   ports tab   Figure 64  Options dialogue box   system tab   Figure 65  Flow diagnostics screen   Figure 66  Reporting module   Figure 67  Rear view of TT24 7 showing system cabling connections  Figure 68  TT24 7 labelling   Figure 69  Direct serial connection   Figure 70  USB connection   Figure 71  Cable connections including power recycle box   Figure 72  Installation of transfer line through inner wall of GC oven  Figure 73  Installation of transfer line through fan protection grill  Figure 74  Installation of transfer line through heated zone block  QUI 1016 V2 2 Aug 06  9     Markes International Ltd  T   44  0 1443 230935    F   44  0 1443 231531    E  enquiries markes com    Markes International    58  59  60  61  62  63  64  65  68  68  70  72  73  74  75  77  77  79  83  83  85  87  88  89  91  91  93  94  95  96  97  99  100  100    TT24 7 User Manual    Figure 75   Figure 76   Figure 77   Figure 78     Figure 79   position    Figure 80   plugged in    Figure 81   Figure 82   Figure 83   Figure 84   Figure 85   Figure 86     Figur
98. m the sorbent tubes  e2 e3 systems     If the contamination is shown  from the experiments described above  not to be  coming from the carrier gas or the sorbent in the traps  the next most likely  candidate is the sorbent in the tubes  e2 e3 systems      Sorbent tubes should be thoroughly conditioned prior to use  Ideally   and    QUI 1016 V2 2 Aug 06  116     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    certainly before use for the first time   this should be done off line to avoid  contaminating the TT24 7 flow path  This can be done using a tube conditioner  such as the TC 20 from Markes International  Sorbent tubes should be  conditioned at temperatures approximately 20   40   C higher than those used for  desorption of the samples   assuming that this does not exceed the sorbent  maximum temperature  See Appendix 3 for information regarding sorbent  conditioning   maximum sorbent temperatures etc     4 2 4 Other potential sources of contamination    Unsilanized glass or quartz wool should be used as standard in the cold traps and  should be conditioned at high temperatures before use  Silanized glass wool can  be used but is only recommended for the analysis of labile compounds     Note  NEVER heat silanized glass wool above 250  C  even during system  conditioning  as the silylating reagent will break down and may  irreversibly contaminate your sample flow path     4 3 Poor p
99. mon parts  e1  e2 and e3 systems     me p    SERUTD 1125  Autosystem clamp 1    poe ES Kit  TT24 7  consisting of  2 x 7 s    x 12    wrench  1 x  posidriver No 2   2 x 7 mm x 8 mm wrench   3   i   Nut t 16    St st Swagelok   Washer        x 1      Pk 10 Size 007 O Ring   Union brass   s    x 1       Pk 10 Ferrule   s x 4 16      Tube PEEK       OD x 0 03    bore   Pk 10 Ferrule   s    x 2 mm Graph Vesp    ffeil fle   EEEEE    u   gt             W  EE    QUI 1016 V2 2 Aug 06  120     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Packing Lists   GC Cable Options   Your TT24 7 system comes with one of the following cable options   SERUTD 5098 PCB GC Interface for Agilent 6890 6850 GC  or   SERUTD 5095 PCB GC Interface   general purpose    or   TGK 6000 Thermo GC Installation Kit including   SERUTD 5108 PCB GC Interface for Thermo GC  SERTTD 1109 Transfer line termination x1  SERZ 0119 Brass union t 16    t 16    x1  SERZ 0401 Plug cap   16    x1  U COV08 Pk 10 Size 008 O Ring X   10  SERUTD 1036 8 mm Retaining ring x2    Packing Lists   e1 system specific parts    Description    Ce  Rurale  T247 User Manual  Installation report back form  Transfer line 1  Mains cable 1   1   1    i  s p   Ran be pce wh sorbent ceria       QUI 1016 V2 2 Aug 06  121     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes Internationa
100. mp  Pump Sample in   5 down   Line from GC  Trap B               Flow of Carrier Gas         Optional Carrier Gas Flow  Carrier Gas  no flaw         gt     _ Flow of Sample Gas  Sample Gas  no flow        Figure 14  TT24 7 flow schematic  e3 configuration  showing the additional  gas lines associated with the GC EPC module    QUI 1016 V2 2 Aug 06  29     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    The pressure sensing Septum Purge line connecting to the TT24 7 may have  carrier gas flowing  In some GC systems this is variable  If maximum  sensitivity is required then this flow should be stopped or a small sample  splitting will occur  This can be achieved by either switching the flow off within  the instrument  or software   or simply blanking the exit line with a closed  Swagelok nut     2 4 6 2 Connection to Agilent Technologies 6890GC   6850 GC    For correct operation of the Agilent Technologies EPC system with the TT24 7   a 6890 GC currently requires firmware revision of A 03 08 or N 04 09  For the  6850 GC serial numbers upwards of US10243001 will be able to provide the  same functionality as the 6890 but may require a flashable firmware upgrade     Note  You should confirm these firmware requirements with your local Agilent  Technologies representative     The Injection Port Mode of the EPC module now has to be configured     Using either the GC keyboard  or hand held contr
101. must be taken when  removing these traps from the system     2 4 4 Heated transfer line    The analysis and detection of desorbed analytes from the cold traps is achieved  using conventional analytical instrumentation  This includes GC  GCMS and in    QUI 1016 V2 2 Aug 06  25     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    certain instances direct MS technology  For speciation of components  gas  chromatography is the preferred choice using capillary columns for maximum  separation efficiency  Typically columns with an internal diameter of 0 32 mm or  0 25 mm are used  with lengths ranging from 10 m to 60 m  All standard  stationary phases are suitable  For splitless injection  carrier flow rates equal to  or greater than 3 mL min are recommended for efficient transportation of analytes  from the sorbent bed into the column  In split mode sampling the flow rate is by  definition considerably greater than this anyway     Interfacing the TT24 7 to the GC capillary column can be achieved by a number  of mechanisms  The most common approach uses a length of uncoated   deactivated fused silica transfer line  which connects and seals into the TT24 7  using a   icth Swagelok nut and a 0 4 mm id graphitised vespel ferrule  Installation  of the transfer line is discussed in more detail in section 3 3     The transfer line is surrounded by a length of PTFE sleeving of slightly greater  inter
102. n 2 3 1     The first two runs of the continuous sampling mode  i e  trap A and trap B   would typically not be used for quantitative analysis of the sampled air but as  a clean up process for both traps  After both traps have been fired once then  the subsequent samples may be quantified     QUI 1016 V2 2 Aug 06  36     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 2 TT24 7e2 hardware configuration    TT24 7e2 may be operated in both sampling modes  continuous sampling and  tube desorb      2 5 2 1 Continuous sampling mode  TT24 7e2   2 5 2 1 1 Standby    Desorption Tube    MFC  amp  Pump Sample in       KEY          Flow of Carrier Gas  Carrier Gas  no flaw        a     Flow of Sample Gas    Sample Gas  no flow        Figure 20  TT24 7e2   Continuous sampling   standby    In this non sampling mode  carrier gas alone is continuously supplied to the  capillary column  No other gas flow occurs within the TT24 7 unit during this  time     QUI 1016 V2 2 Aug 06  37     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 2 1 2 Trap A sampling  trap B purging    Desorption Tube    q Needle  I valve       KEY       gt     Flow of Carrier Gas  Carrier Gas  no flaw           Flow of Sample Gas  Sample Gas  no flow        Figure 21  TT24 7e2   Continuous sampling   trap A sampling  trap B purging   
103. n excess of    0 5 minutes are typically used  This value must be long enough to ensure  complete removal of analytes from the trap into the analytical column   Consideration of the trap flow conditions at trap fire are required  i e  the flow  through the trap at trap fire is equal to the column flow rate which could be as  little as 3mL min  In this case a time value less than 1 minute is not  recommended     Incorrect setting of this time can be the cause of poor sensitivity for certain  compounds  as insufficient time will reduce the amount of sample leaving the  trap     2 9 1 1 9 Save    Once a method parameter has been changed then the word    modified    appears  alongside the method name in the blue bar at the top of the method window   When the desired method parameters have been specified they can be saved  directly into the active method by clicking on the SAVE button at the bottom of  the user interface     2 9 1 2 Flow display    The flow display window  figure 49  is designed to help the user see the flow  direction process and the current run status of the TT24 7 system  The  diagram shows the flow direction through the solenoid valves SV1  2  3 and 4   and the flow direction through the heated valves and traps     Flow Display    To Column       Figure 49  Flow display showing flow direction through solenoid valves    2 9 1 3 Instrument status bar    Instrument Status Heated Zones GC Flow Controller  Trap   State  0   0 0 2 25 0   C State  Not Ready Flow Rate
104. n is particularly thick it may be necessary to shorten  the silicone foam rubber insulation sleeve  which is intended to rest against the  outer wall of the GC oven     The M6 spacer nut  attached to a   4 inch spacer tube on the transfer line  secures  the line casing to the oven wall allowing the     s inch aluminium sleeve to protrude  into the oven  If the hole in the inner oven wall is larger than the end of the line   fit one of the large metal washers from the shipping kit at this point     In figure 73 the entry to the GC oven is through the fan protection grill  In this  situation the   4 inch spacer tube attached to the transfer line prior to the spacer  nut is used to extend the line and a special U shaped metal support bracket is  pushed through adjacent holes in the grill to press against the oven inner wall     Note  The line must not be secured with a nut against the fan grill as this could  be distorted causing it to hit the fan     QUI 1016 V2 2 Aug 06  99     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    me U shaped metal Spacer nut  M6  Silicone foam support bracket E ma  rubber insulati    Aluminium heat  conductor    Fan protection  Spacer tube grill       Figure 73  Installation of transfer line through fan protection grill    Figure 74 shows installation via a heated zone block  As the entry hole will  generally be larger than the diameter of the metal line sle
105. n molecular  sieve principle    permanent gases and light  hydrocarbons  C3  C3     Boiling point  60  C to 80  C  Ultra volatile hydrocarbons    E  enquiries markes com    TT24 7 User Manual    Sorbent Maximum Temperature     Recommended Conditioning Temperature   Recommended Desorption Temperature     Notes     14  Molecular Sieve 5A  Sorbent Strength    Approximate analyte volatility range   Example Analytes    Sorbent Maximum Temperature     Recommended Conditioning Temperature     Recommended Desorption Temperature     Notes     15  Molecular Sieve 13X  Sorbent Strength    Approximate analyte volatility range   Example Analytes    Sorbent Maximum Temperature     Recommended Conditioning Temperature     Recommended Desorption Temperature     Notes     QUI 1016 V2 2 Aug 06    Markes International Ltd  T   44  0 1443 230935    Markes International     gt 400  C  350  C   400  C  300  C   350  C    Significantly hydrophilic   do not use  in humid conditions     High artefacts   gt 10 ng      Easily  amp  irreversibly contaminated  by higher boiling components    Very strong for small molecules  Boiling point  60  C to 80  C  nitrous oxide   350  C   400  C    300  C   350  C max  increase  temperature gradually from 100  C     To suit analyte    Significantly hydrophilic   do not use  in humid conditions     High artefacts   gt 10 ng     Easily  amp  irreversibly contaminated  by higher boiling components    Very strong for small molecules  Boiling point  60  C to 80  C   1 
106. nal diameter than the outer diameter of the fused silica  The PTFE is used to  assist in the insertion and feeding of this length of tubing through the outer heated  jacket of the transfer line  It also maintains a constant temperature profile down  the length of the fused silica preventing hot spots  The outer layer of the transfer  line is insulation material which is used to maintain a consistent and controllable  temperature along its 1 2 m length  figure 12      The transfer line is maintained at the same temperature as the heated valves  i e   the valve temperature controls the transfer line temperature setting  This is  typically up to 200  C     Connection to the capillary column within the GC oven requires a suitable hole in  the GC covers to allow the outer insulation of the transfer line to pass as far as  possible to the surface of the oven without assimilating heat from the oven  For  certain GC systems the split   splitless injection port is used as a conduit after the  internal liner etc  is removed     The capillary column can then be connected to the transfer line using deactivated  glass unions  or deactivated steel connectors and ferrules     Figure 12  Transfer line connection between TT24 7 and GC system  As an alternative to the fused silica transfer line approach  certain capillary    QUI 1016 V2 2 Aug 06  26     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    co
107. national    Appendix 3  Sorbent specifications    1  Carbotrap C     20 40 mesh    Carbopack C     60 80 mesh     Carbograph 2TD  range of mesh sizes     Sorbent Strength   Specific Surface Area  m2 g    Approximate analyte volatility range     Example Analytes     Sorbent Maximum Temperature     Recommended Conditioning Temperature     Recommended Desorption Temperature     Notes     Very weak    12   n Cg to n Cz9   Alkyl benzenes  hydrocarbons to n C39   gt 400  C   350  C to 400  C   300  C to 350  C   Hydrophobic   Minimal   lt 0 1 ng  artefacts   Some activity with labile compounds  Friable    2  Tenax TA    or GR     range of mesh sizes     Sorbent Strength   Specific Surface Area  m2 g    Approximate analyte volatility range     Example Analytes     Sorbent Maximum Temperature     Recommended Conditioning Temperature     Recommended Desorption Temperature     Notes     QUI 1016 V2 2 Aug 06    Markes International Ltd     T   44  0 1443 230935     128     F   44  0 1443 231531    Weak    35   n C7 to n C39   Bpt  100  C to 450  C    Aromatics  except benzene   apolar  components bpt gt 100  C  polar  components bpt  gt 150  C  PAHs PCBs     350  C   325  C   Up to 300  C   Hydrophobic   Low inherent artefacts   lt ing    Inert     suitable for labile compounds  Graphitised form best for PAHs PCBs  Efficient desorption    Use 35 60 mesh to minimise fines  and eliminate    leakage    through  conventional sorbent retaining  gauzes    E  enquiries markes com    TT24 7 Use
108. nder normal circumstances the default baud rate values of 57600  should be used  However if there are communications problems then a  lower baud rate be specified i e  38400     QUI 1016 V2 2 Aug 06  88     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual    Markes International    GC Interface logic    This section of the software affects the start   stop interaction between  the TT24 7 and the host GC   GCMSD system  The actual configuration  is dependent on the model of GC   GCMSD used  For Agilent  Technologies and Thermo Electron instrumentation the GC Start  out   and GC Ready  in  settings should be in the    closed    state     For other GC manufacturers please contact Markes International for details     2 9 2 5 3 System    Gas   Ports System   E Mail             Display Flow       JV Enable    Enabling this option will display the current flow rate from the MFC in the status bar          Enable       m Limit Sample Time    Enabling this option will force the instrument to take a sample only for the specified sampling  time  No more gas will be sampled until the next sampling cycle        M Sampling Extended Timeout    Time to extend sampling   IV Enable Timeout          Valves          IV Enable Split option for method  V Purge Valve  S 6    No Back Valve                   Figure 64  Options dialogue box   system tab    Display flow     Limit Sample Time     If the enabling box is checked  the mass f
109. njection mode  TT24 7  if the column flow rate is 3mL   min and the septum purge approximately 3 mL min then the    Total Flow    will  show as approximately 6 mL minute on the 6890 keyboard     In split sampling mode  TT24 7  if the split flow is for example 50 mL min then  the    Total Flow    reading will appear as approximately 56 mL min  This value  can be seen to provide an additional electronic readout of the split flow used by  the TT24 7 at trap fire     QUI 1016 V2 2 Aug 06  30     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 4 6 3 Connection to Thermo Electron Trace   FOCUS GC    For correct operation of the Thermo Electron DPFC system with the TT24 7 the  GC requires an available inlet configured for back pressure regulation     The injection port mode of the DPFC module to be used has to be configured  for split operation but the split flow has to be set to zero  off      The DPFC module is physically connected to the TT24 7 using the gas lines on  the GC that are used for supplying the split   splitless inlet  The carrier supply  and septum purge lines should be cut and extended  using the   s inch unions  and green PEEK tubing supplied in the TT24 7 shipping kit  so that they can be  connected onto the back panel of the TT24 7  The carrier supply line should  be connected to the carrier inlet of the TT24 7 while the septum purge line  should be connected to the pres
110. nternational    2 5 3 2 3 Leak testing trap A    MFC  amp  Pump Sample in    Trap B    Needle  Valve    h off  Split  Vent    KEY   m Flow of Carrier Gas      Optional Carrier Gas Flow  Carrier Gas  no flow        gt    _ Flow of Sarnple Gas  Sample Gas  no flow        Figure 39  TT24 7e3   Tube desorb mode   leak testing trap A    After the tube leak test the selected trap also undergoes a further leak test  In  the example above trap A is to be used  The heated valve HVC now switches  to allow pressurisation  via SV5  of the line right through trap A  SV1 being    switched OFF  Once again the pressure sensor next to SV5 is monitored for  pressure decay     The split line pneumatics through to SV7 are also included within this leak test     QUI 1016 V2 2 Aug 06  56     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 3 2 4 Tube purge    MFC  amp  Pump       Trap B       Needle  Valve         gt     Flow of Carrier Gas      C   Optional Carrier Gas Flow  Carrier Gas  no flow            Flow of Sample Gas  Sample Gas  no flow        Figure 40  TT24 7e3   Tube desorb mode   tube purge    Prior to the tube desorption  heating  phase  the air and moisture which will be  present in the tube needs to be flushed out to preserve the integrity of the  sorbent material  Carrier gas purges through the tube for a default time of 30  seconds at the tube desorb flow rate and residual air   moistu
111. nual Markes International       Diagnostics    Pneumatics    r Manual Valve Control    stan  ou  HVA       To Column                Figure 65  Flow diagnostics screen    2 9 2 7 Reporting module    The reporting module  figure 66  displays a number of different Information   Warning or Error messages     All messages detailed below will appear for either Trap A or Trap B if relevant   Trap Ais used as an example below  All error messages are date and time  stamped e g  05 12 2005 10 34 54       Reports  05 12 2005 10 49 31  INFO  Samping Started on Trap A       Figure 66  Reporting module    2 9 2 7 1 Information messages   INFO  Sampling started on Trap A     date and time stamped when sampling started   INFO  Sampling has been extended on Trap A      reported when samping time exceeded the expected time set in the  method  this is typically due to the GC Ready signal being delayed    perhaps due to the oven taking longer than expected to cool down     Information messages do not affect the operation of the system in any way     QUI 1016 V2 2 Aug 06  91     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 9 2 7 2 Warning messages   WARNING  Sampling volume outside of expected range      this warning message is given when the total volume sampled is not  what would be expected from the method  For example this could be  because the sampling time was extended  see section 2 9 1 5 3  or
112. ol  ethanol  acetone    E  enquiries markes com    TT24 7 User Manual    Sorbent Maximum Temperature     Recommended Conditioning Temperature     Recommended Desorption Temperature   Notes     12  Carbosieve SIII     60   80 mesh    Sorbent Strength     Specific Surface Area   m2 g      Approximate analyte volatility range     Example Analytes   Sorbent Maximum Temperature     Recommended Conditioning Temperature     Recommended Desorption Temperature   Notes     Markes International     gt 400  C   350  C to 400  C   300  C to 350  C   Some hydrophillicity   Low artefacts   lt 0 1 ng    Inert   suitable for labile compounds  Excellent batch to batch reproducibility  Non friable    Easily contaminated by higher  boiling components   protect with  front bed of weaker sorbent    only   Very Strong     800   but primarily operates on  molecular sieve principle    Ethane to n Cs5   Boiling point  60  C to 80  C  Ultra volatile hydrocarbons   gt 400  C   350  C to 400  C   300  C to 350  C   Some hydrophillicity   Low artefacts   lt 0 1 ng     Easily and irreversibly contaminated  by higher boiling components    protect with front bed of weaker  sorbent    13  Carboxen 1000     range of mesh sizes     Sorbent Strength   Specific Surface Area  m2 g      Approximate analyte volatility range     Example Analytes     QUI 1016 V2 2 Aug 06    Markes International Ltd  T   44  0 1443 230935     133     F   44  0 1443 231531    Very Strong for small molecules     gt 1200   also operates o
113. oller  6850GC   or  Chemstation    software  access the injection port mode  This consists of Split   Splitless  Pulsed Split  or Pulsed Splitless  Select the    Splitless    mode of  injection and set the purge time to 999 99 minutes exactly  This is required to  prevent the EPC module resetting itself at the completion of each run which is  not required with the TD system     Note  This exact time value is very important for EPC control of the TT24 7     The EPC module is physically connected to the TT24 7 using either the original  gas lines  which must be cut  connecting to the split   splitless inlet  or  preferably by using a separate pair of gas lines connecting into their own  manifold  Agilent Part Number G2131 80500   which replaces the original three  gas line manifold  This type of connection enables the original split   splitless  inlet to be reconfigured very easily if required     If the capillary injection port is not required  then it can also provide an easy  access route for the heated transfer line to enter the GC oven for both the 6890  and 6850 models  See section 3 3 1 for further details     When using the 6890 GC  the total flow of carrier gas entering the TT24 7 can  be observed from the keyboard  by selecting the appropriate inlet  i e  front or  rear  and scrolling down to the line    Total Flow        The    Total Flow    consists of column flow  septum purge flow  if applicable  and  the split flow within the TT24 7     For example in splitless i
114. ompletion of the download process the top LED will turn from  amber to green     Step 3  If however the sampler board is still not detected  it may be due to  an incorrect COM port designation  and eventually download activity within the  software will time out  A dialogue box will then appear as shown in figure 46        Select Option    Unable to connect to the instrument  please select and option  from the list below          Edit Options       Continue Running                Figure 46  Download activity timeout dialogue box    From this dialogue box select the    Edit Options    configuration  this will launch  the instrument options software page  and from this selection of the    Ports    tab  enables the COM port settings to be reviewed  figure 47   Ensure that the  correct COM port settings are specified  See section 2 9 1 5 2 for further  details     QUI 1016 V2 2 Aug 06  63     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Parts   System   E Mail            r Communications Port    Analyser Port  COMI z  MFC Port  coms 7   Baud Rate  57600 7  Baud Rate  57600 7     m GC Interface Logic      GCStart  out        _ GC Ready  in             Open   Start   Open   Ready  C Closed   Start C Closed   Ready                                 Cancel      Figure 47  Ports tab in options configuration    With the correct COM port settings specified  return to Step 1 above and repeat  the 
115. oned sorbent tube and used to re collect the  split portion of the sample for method   data validation and repeat analysis   SecureTD Q      For further information regarding SecurelTD Q and sample re   collection please see the Markes International brochure    Validation of Thermal  Desorption featuring SecureTD Q   Quantitative sample re collection for thermal  desorption           Figure 90  TT24 7e2 e3 system with Split tube option on RHS    To access the split tube the sealing mechanism lever should be lifted and the tube  inserted   removed exactly as described for the sample tube above     4 0 Troubleshooting    4 1 Leak testing the system    In any gas flow path leaks may occur due to worn or ill fitting o rings  fittings etc   The TT24 7 flow path can be leak tested using the flow only diagnostics facility   see section 2 9 1 6   2 9 2 6   In order to leak test successfully  a leak detection  system such as a helium leak detector  Markes p n C HEL23   C HEL11  should  be used     QUI 1016 V2 2 Aug 06  114     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Note     wet    methods of leak testing  e g  with bubble solution  SNOOP etc    should be avoided at all costs     Using diagnostics  individual valves may be switched in order to isolate different  parts of the flow path and the helium leak detector applied to the appropriate  connection point to ascertain if there is a ga
116. orage temperatures     N B  There are other implications associated with storing tubes under refrigerated  conditions  Many laboratory fridge   freezer atmospheres are highly contaminated  with volatiles from other samples or from the refrigeration system itself     N B  If the tubes are to be transported in such a way as to be exposed to very  cold temperature i e  in an aircraft hold  by rail   road overnight during cold  weather  it is advisable to follow the above retightening procedure by cooling the  tubes  by placing outside  or in a clean refrigerator  prior to shipment     When monitoring trace level atmospheric components  conditioned and sampled  tubes can be wrapped in uncoated aluminium foil and   or placed in a sealed  non   outgassing container  such as an uncoated tin  during transportation and storage     For further information on minimising artifacts   see TDTS19    Trademarks and Registered Trademarks  UNITY     UttRA     TT24 7    are all trademarks of Markes International Ltd   UK     Carbotrap C  Carbopack C  Carbotrap  Carbopack B  Carbotrap X  Carbopack X  Carbosieve SIII   Carboxen 1000   are registered trademarks of Supelco Inc  USA    Chromosorb 102  Chromosorb 106   are registered trademarks of Manville Corporation  USA  Carbograph 1TD  Carbograph 2TD  Carbograph 5TD   are registered trademarks of LARA s r l   Italy  GCB1   is a registered trademark of SKC Inc  USA   Porapak N  Porapak Q   are registered trademarks of Waters Associates Inc  USA   
117. ound contamination which is obtained     If the contamination increases with the length of time that the instrument  is in standby then the carrier gas supply is likely to be contaminated and  individual components of the supply system should be checked     If the contamination is of more volatile components then you will need to sample  your carrier gas via the TT24 7 system itself using the following procedure     Disconnect the carrier gas supply from the rear of the TT24 7 and attach  a  clean  T piece and short length of tubing so that the carrier gas can now  be attached to both the rear of the TT24 7 and to the sample inlet of the  TT24 7     Set up a continuous sampling run on the TT24 7 to sample the carrier gas    QUI 1016 V2 2 Aug 06  115     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    for a specified time period  e g  10 minutes    Note the background contamination which is obtained    Repeat the above procedure but this time double the time period that the  system is left in standby  e g  20 minutes      Note the background contamination which is obtained     If the contamination increases with the length of time that the instrument  is in standby then the carrier gas supply is likely to be contaminated and  individual components of the supply system should be checked     To reduce the possibility of carrier gas contamination  ideally  each TT24 7 system  should have i
118. ow pressurisation  via SV5  of the line right through trap A  SV1 being  switched OFF  Once again the pressure sensor next to SV5 is monitored for  pressure decay     QUI 1016 V2 2 Aug 06  44     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 2 2 4 Tube purge    Desorption Tube    MFC  amp  Pump Sample in    Trap B          KEY          Flow of Carrier Gas     Carrier Gas  no flaw           Flow of Sample Gas  Sample Gas  no flow        Figure 28  TT24 7e2   Tube desorb mode   tube purge    Prior to the tube desorption  heating  phase  the air and moisture which will be  present in the tube needs to be flushed out to preserve the integrity of the  sorbent material  Carrier gas purges through the tube for a default time of 30  seconds at the tube desorb flow rate and residual air   moisture is then flushed  onto the chosen trap    The majority of the air will pass directly through the trap as it is unretained   however the trap is also dry purged in the next phase  section 2 5 2 2 6  to  remove any residual air and  depending on the sorbent phase  moisture as well     QUI 1016 V2 2 Aug 06  45     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 2 2 5 Tube desorption    Desorption Tube    MFC  amp  Pump Sample in    Trap B             KEY          Flow of Carrier Gas     Carrier Gas  no fl
119. p   State  Leak testing the Tube Trap A  0 0   25 0   C Tube Oven  0 0   250 0  C State  Not Ready Flow Rate  07 500 ml min  TrapB State  Not in use Trap B  0 0 7 25 0   C Rear HV  0 0 7 120 0  C Sample Gas  Air   Running Time 0 1 min Transfer Line  0 0  120 0  C Carrier Gas  He   Desorb Pressure  0 0 psi Dual Front HY  0 0   120 0  C    Figure 60  Instrument status bar   e2 e3 configurations    The instrument status bar  figure 60  is located at the bottom of the top level  screen  It is divided into four sections  i e  Instrument Status  Heated zones   GC and Flow Controller     QUI 1016 V2 2 Aug 06  83     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 9 2 3 1 Instrument Status  This shows the status of the sampling run which is operating   Instrument status     Indicates the current status of traps A and B e g  sampling  desorbing   leak testing tube etc     Sample Time   TT Sampling mode   How long the current trap has been sampling for  Running Time  Tube Desorb mode     The run time associated with the current operational phase i e   desorbing the tube  leak test  dry purge etc     Desorb Pressure  Tube Desorb mode   Equivalent to the column head pressure at the point of tube desorption   2 9 2 3 2 Heated Zones  This section shows the heated zones monitored within the TT24 7    Trap A   Trap B   Transfer Line   Dual Front HV   Tube Oven   Rear HV    The status bar shows two a
120. p  compounds off the trap in the reverse direction to sampling at the point of trap  fire  This allows the use of traps containing more than one sorbent bed in order  of adsorption strength thereby covering a much broader range of analyte that can  be analysed     This process can be entirely splitless for maximum sensitivity  or if the split option  is available  configuration TT24 7e3   the sample will be split at this point     See sections 2 5 1 and 2 5 2 1 to review the flow schematics for continuous  sampling     To assist in understanding the sequence of events occurring during continuous    QUI 1016 V2 2 Aug 06  17     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    sampling  and how this relates to the GC oven cycle  figure 3 represents  diagrammatically the operational status of both traps  A and B  relative to the GC  cycle  An arbitrary sampling time of 5 minutes is shown                 DP   Pre   Dry   Purge   TF   Trap fire   TC   Trap cooling   CD   GC oven cool down   R   Ready  awaiting sampling  TD   awaiting start run  GC      Figure 3  Schematic representation of TT24 7 sequence in continuous  operation    When the TT24 7 starts  trap A immediately goes into sampling mode  whereas  trap B goes through a purge  trap fire  and trap cool sequence of events  this also  instigates the first GC oven cycle  This maybe considered as a first run effect for  both traps  trap
121. p B within the trap housing    Each trap is inserted through a ceramic heater sleeve and at the point of trap fire  heats at rates approaching 100  C second  This extremely fast heating rate  removes the retained analytes from the sorbent bed s  very efficiently which  allows the traps to operate in splitless mode for maximum sensitivity     Each trap is positioned directly on to its own Peltier cooling element  This electro   thermo cooling enables each trap to cool to sub ambient temperatures during the    QUI 1016 V2 2 Aug 06  24     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    sampling phase which greatly enhances the trapping efficiency of the sorbent  material  The lower temperature is a function of the sample temperature  valve  temperature and sampling flow rate     Dry air  nitrogen or helium is used to purge the trap box to prevent the build up  of moisture and the potential for icing which would effect the temperature  measurement of the traps     The narrow bore traps are made of quartz  are restricted at the inlet   outlet end  and have a quartz collar on the unrestricted end  figure 11   They may be packed  with up to 60 mm of sorbent s  which are separated and retained by quartz wool   Sample enters the trap through the restricted inlet end and at the point of trap  fire the trap is backflushed  carrier gas flow reversed  and the analytes are  desorbed back out of the
122. peak splitting or discrimination as seen on conventional GC  injectors  In these cases  reduce the amount of water or solvent retained by the  cold trap  e g by raising the cold trap temperature or by using a longer pre purge  time      If the GC analytical column is overloaded this will cause band broadening  High  resolution capillary columns work at optimum with analyte masses in the order of  20 200 ng     QUI 1016 V2 2 Aug 06  117     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    4 4 Carryover of components of interest    Carryover is usually caused by incomplete desorption and is usually addressed  either by using more stringent desorption conditions  higher trap desorption  temperature where possible or longer trap hold time   or by selecting a weaker  sorbent for collecting the samples     The gas flow rate is also critical and should be at least 3 mL min     If the sorbent in the trap is over heated in error  this may have caused high boiling  materials to deposit in the internal flow path of the TT24 7     Note  If you exceed the maximum sorbent temperature  the resulting  breakdown of the sorbent may severely contaminate the flow path of  your system  This may require the complete replacement of the flow path  and associated components by a fully qualified service engineer and  would not be covered by the instrument warranty     4 5 Poor recovery   loss of sample    Recove
123. pecified for correct calibrated  control which is based on its density  See section 2 9 2 5 1 for further  information     The carrier gas should be chosen for optimum performance of the associated  chromatographic system and this is typically helium  although other choices are  available e g  nitrogen and hydrogen  The carrier gas flowing through the system  performs several functions including dry purging the traps prior to trap fire  back  flushing the traps at trap fire  transporting sample from the tube accessory or  capillary inlet into the selected trap etc     If split sampling is configured then this also places demands on the supply of  carrier gas  Without electronic pneumatic control this would affect the column  head pressure  resulting in large differences in the retention time of eluting  compounds run to run     Section 2 5 shows the schematics illustrating the flow protocol within the TT24 7  for each sampling mode within each hardware configuration     The carrier gas supply entering the TT24 7 originates from the associated GC  system and in particular from the electronic flow module normally supplying gas  to a split   splitless injection port  This gas supply is now redirected from the  capillary inlet into the TT24 7     An important aspect associated with the control of the carrier gas is called closed  loop feedback  This relates to controlling the capillary column head pressure so  that whatever flow demands are made by the various processes occurring
124. process     Step 4  The next step is downloading firmware into the mass flow controller  board  MFC  which is represented by the lower LED  To achieve this turn the  front panel switch On  depressed   at which point the lower LED turns red  The  software will automatically attempt to download the firmware into the MFC  board and this is indicated by the LED turning amber     As for the sampler board  if detection of the MFC board is unsuccessful  toggle  the front panel switch off  wait 10 seconds  and on  and the software will re   attempt to download the firmware again     On successful completion the lower LED will turn green     If the download is still unsuccessful then check the COM settings for the MFC  board as specified in Step 3 above     At this point both LEDs should appear green  and the user interface should be  loaded onto the PC display     2 9 Software control   top level  The top level software screen is shown in figure 44 above     Four separate windows are shown titled Method  Flow Display and Reports and  at  the base of the screen  the Instrument Status display  Above the screen are seven  graphical icons and a menu bar     The software control and user interface will vary according to the hardware  configuration installed  Ensure that you know which configuration you have prior  to reading this manual  see Introduction      QUI 1016 V2 2 Aug 06  64     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 
125. r Manual    Markes International    3  Carbotrap     20 40 mesh    Carbopack B     60 80 mesh     GCBi     range of mesh sizes    Carbograph i1TD  range of mesh    sizes    Sorbent Strength    Specific Surface Area  m2 g    Approximate analyte volatility range     Example Analytes     Sorbent Maximum Temperature   Recommended Conditioning Temperature   Recommended Desorption Temperature   Notes     Medium   Weak    100   n Cs7  6 to n Cy4   Ketones  alcohols  aldehydes and    apolar components within the above  volatility range     Perfluorocarbon tracer gases   gt 400  C   350  C to 400  C   300  C to 350  C   Hydrophobic   Low artefacts   lt 0 1 ng    Some activity with labile compounds  Friable    4  Carbopack X     40 60 and 60 80 mesh    Carbotrap X     20 40    mesh    Sorbent Strength    Specific Surface Area  m2 g    Approximate analyte volatility range     Example Analytes     Sorbent Maximum Temperature   Recommended Conditioning Temperature   Recommended Desorption Temperature   Notes     QUI 1016 V2 2 Aug 06  129     Markes International Ltd  T   44  0 1443 230935    Medium Strong    240   n C374 to n C6 7   Boiling point 50  C to 150  C    Light hydrocarbons  BTX for 2 week  diffusive exposure     gt 400  C   350  C to 400  C   350  C to 400  C   Hydrophobic   Low artefacts   lt 0 1 ng    Some activity with labile compounds  Friable    F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    5  Carbograph 5TD  various mesh sizes
126. r achieved using gas filters  Typically    QUI 1016 V2 2 Aug 06  97     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    this will include a Moisture trap  P N C MSTRP   Oxygen trap  P N C O2TRP  and  Hydrocarbon trap  P N C HCTRP      The external supply pressure of carrier gas needs to be   10 psi greater than the  capillary column head pressure requirements  If the column is operating in  constant pressure mode this is a simple addition  i e  for a head pressure of 30 psi  the supply pressure needs to be   40 psi etc  However if constant flow conditions  are being used then the supply pressure needs to be calculated at the oven  maximum value when the pressure demand by the column is at its greatest   Connection of the carrier gas to the TT24 7 is discussed in section 2 4 6 1     The trap box purge and valve actuator gas is typically dry air  nitrogen  or carrier  gas  He only   This needs to be supplied at a pressure between 50 and 70 psi and  must have a dew point lower than  35  C  The trap box purge gas is essential to  prevent moisture condensation and subsequent icing within the Peltier cooled trap  box  If this occurs  the lower  i e  sub ambient  temperatures will not be  achievable     The dry gas supply must always be on when the 1TT24 7 is being used and  secondary pressure regulation of this gas supply is required in addition to the  laboratory control     Markes Internat
127. r purging the trap  B  for the first time  trap fire occurs and flow is directed   backflushed  across to the capillary column  This first  B  trap fire prepares  the trap for subsequent sampling  The flow rate through the trap is equal to  the column flow rate and should be at least 3 mL min     QUI 1016 V2 2 Aug 06  39     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 2 1 4 Trap A purging  trap B sampling    Desorption Tube    q Needle   N valve    MFC  amp  Pump    Trap B       KEY          Flow of Carrier Gas  Carrier Gas  no flaw           Flow of Sample Gas  Sample Gas  no flow        Figure 23  TT24 7e2   Continuous sampling   trap A purging  trap B sampling    After the sampling time for trap A has completed trap A switches into its dry   purge mode to sweep out air and moisture prior to trap fire  Having cooled to  its starting temperature after the first clean up trap fire  trap B now goes into  its sampling phase  The date and time that the sampling started  and the trap  that is being sampled are reported as an information message in the reporting  module of the software  see section 2 9 2 7   e g     05 12 2005 10 49 31   INFO  Sampling Started on Trap B    QUI 1016 V2 2 Aug 06  40     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 2 1 5 Trap A desorbing  trap B sampling 
128. re is then flushed  onto the chosen trap     The majority of the air will pass directly through the trap as it is unretained   however the trap is also dry purged in the next phase  section 2 5 3 2 6  to  remove any residual air and  depending on the sorbent phase  moisture as well     QUI 1016 V2 2 Aug 06  57     TT24 7 User Manual Markes International    2 5 3 2 5 Tube desorption    Desorption Tube    MFC  amp  Pump Sample in    Trap B    Needle  Valve    5s off  Split  Vent    KEY   m Flow of Carrier Gas      Optional Carrier Gas Flow  Carrier Gas  no flow           Flow of Sample Gas  Sample Gas  no flow        Figure 41  TT24 7e3   Tube desorb mode   tube desorption    The tube is now heated to the temperature set as the    Desorb Temp    within the  TT24 7 method  The tube    Desorb Flow    and    Desorb Time    parameters control  the desorption process  The set    Desorb Temp    value will depend on the sorbent  material and should not be set higher than that recommended for the  sorbent s  in the tube  See Appendix 3 for further details on sorbent selection  and maximum recommended temperatures     Sufficient flow and time should be maintained to completely desorb all analytes  from the tube onto the specified trap A or B     In the schematic above A is the chosen trap     QUI 1016 V2 2 Aug 06  58     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 3 2 6 Trap  A  purge    
129. ry of labile components may often be improved by increasing the trap hold  time and column gas flow rate while reducing the desorption and flow path  temperatures  Many volatile labile analytes will pass successfully through TT24 7  with flow path temperatures as low as 50  C     For the analysis of extremely labile  relatively involatile components  bp gt  n C12    silanized glass or quartz wool alone should be used as the cold trap packing  material where possible     If a multibed cold trap is being used  ensure that the different sorbents are kept  in discrete beds separated by unsilanised glass wool   quartz wool plugs and  arranged in order of increasing sorbent strength   i e  weak to strong from the  sampling      QUI 1016 V2 2 Aug 06  118     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Appendix 1  Packing List and Routine Maintenance  Spares    The following items are included with your TT24 7 system  please check carefully  and inform your distributor if there are any shortages  Items marked with a   are  consumable items and may require changing at regular intervals   the commercial  re order number is given in the description and detailed in the routine  maintenance spares section     QUI 1016 V2 2 Aug 06  119     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Packing List   com
130. s  Sample Gas  no flow        Figure 36  TT24 7e3   Continuous sampling   trap A desorbing  trap B sampling    Trap A now fires  If this is the first run for A this run is treated as a clean up  run  however if this is a subsequent run then real sample will be transferred  across to the capillary column for analysis     If the split function is enabled in the method   see section 2 9 2 1 14 then this  will occur at trap fire  Trap A Desorbing   The flow rate through the trap is  therefore equal to the column flow plus the split flow through SV7  The split  ratio is therefore     Split Ratio   Column Flow       Column flow   Split flow    If the split function is not enabled then the flow rate through the trap is equal  to the column flow rate alone and should be at least 3 mL min     To see how the time overlap of sampling  dry purging  trap fire and cooling  occurs relative to the GC oven cycle time please refer to section 2 3 1     QUI 1016 V2 2 Aug 06  53     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 3 2 Tube desorb mode  TT24 7e3   2 5 3 2 1 Standby    Desorption Tube    MFC  amp  Pump Sample in    Trap B       KEY          Flow of Carrier Gas      Optional Carrier Gas Flow  Carrier Gas  no flow        gt    _ Flow of Sample Gas  Sample Gas  no flow        Figure 37  TT24 7e3   Tube desorb mode   standby    In this non sampling mode  carrier gas alone is continuously supp
131. s being sampled  are reported as an information message in the reporting module of the software   see section 2 9 1 7   e g     05 12 2005 10 49 31   INFO  Sampling Started on Trap A  The sample passes initially onto trap A  while trap B pre purges with carrier gas  in the desorption direction  This removes air and moisture from the trap prior    to firing  The pre purge flow is determined by the column flow and should be  at least 3 mL min     The sample gas flow rate passing into the TT24 7 is controlled by the internal  mass flow controller  MFC      QUI 1016 V2 2 Aug 06  33     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 1 3 Trap A sampling  trap B desorbing       MFC  amp  Pump    Trap B    KEY          Flow of Carrier Gas  Carrier Gas  no flaw        gt    _ Flow of Sample Gas      Sample Gas  no flow        Figure 17  TT24 7e1   Trap A sampling  trap B desorbing    After purging the trap  B  for the first time  trap fire occurs and flow is directed   backflushed  across to the capillary column  This first  B  trap fire prepares  the trap for subsequent sampling  The flow rate through the trap is equal to  the column flow rate and should be at least 3 mL min     QUI 1016 V2 2 Aug 06  34     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 1 4 Trap A purging  trap B sampling    M
132. s either from the tube or using  the continuous sampling mode     The tube desorption configuration can also be used for real tube sampling  In  certain applications  a confirmatory tube is positioned within an environment to  back up a continuous monitoring station  The sampling process could be pumped  or diffusive  If the continuous system alarms  then tubes could be analysed on  the TT24 7 to confirm this response     2 3 3 Split sampling  Available with hardware configuration e3 only     The 1124 7 may be configured with the capacity to split the sample at the point  of trap fire  This functionality can be turned on and off in the TT24 7 software   see section 2 9      When split sampling is not configured the TT24 7 will desorb sample from the trap  in a splitless manner directly into the analytical capillary column  For those  applications where maximum sensitivity is required this is the preferred mode of  sampling  However there are instances where a more flexible mechanism for  loading sample into the column is required  In this instance the ability to split the  sample at the point of trap fire is advantageous     The split flow is controlled by a small needle valve on the right hand side of the  TT24 7  Figure 6 shows the TT24 7 with the split accessory fitted  TT24 7e3    Measurement of the split flow is achieved using a digital flow measuring device  connected to this vent line     QUI 1016 V2 2 Aug 06  20     Markes International Ltd  T   44  0 1443 230935 F   
133. s leak     In addition to the flow diagnostics  the TT24 7 software can perform a manual leak  test of desorption tube  trap A and trap B by clicking on the appropriate icon in  the software  see section 2 9 2 4 1      4 2 Contamination   the presence of artifacts in the chromatogram    Artifacts are usually the result of either insufficient conditioning of sorbent from  the cold traps  or sorbent tube where applicable  e2 e3 systems    or  contamination from the carrier gas or carrier gas supply equipment     4 2 1 The carrier gas supply    The carrier gas supply is a common source of contamination in thermal  desorption  Contaminants may derive from the gas itself  cylinder head  regulators  gas lines or carrier gas filters     If the contamination is of high boiling compounds then use the following  procedure to establish whether the gas supply is at fault     Immediately after a GC run  cool the GC oven down to ambient  temperature and leave the TT24 7 in standby for a period of time  e g  10  minutes      Carrier gas will now be passing directly into the GC column and because  the column is at ambient temperature any high boiling contamination will  be focusing on the front of the column     After the selected time period  start the GC run by manually pressing RUN  on the GC system     Note the background contamination which is obtained     Repeat the experiment but this time double the time period that the  system is left in standby  e g  20 minutes      Note the backgr
134. strument and carefully  slide the pneumatics towards the traps until the trap valve connectors   P N TTD 5032  have sealed onto the traps     Note  It is easier to locate the trap tube links on the cold traps if they are  gently rotated as they are pushed onto the cold traps     QUI 1016 V2 2 Aug 06  108     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International       X    Figure 86  TT24 7e1 with overlaid schematic showing trap positions and user  serviceable parts    3 4 2 Removing the cold traps  Refer to Figure 86    Ensure that the instrument is in standby       Remove the trap pneumatics by loosening the locating screw and  carefully sliding the pneumatics forward     QUI 1016 V2 2 Aug 06  109     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Note  Take care not to rotate the trap pneumatics as they are moved to  avoid breaking the cold traps       Once clear of the cold traps lift the trap pneumatics clear of the traps  and move to the RHS of the instrument        Close the TT24 7 software     Switch the instrument off   Collared Traps      Place the trap extraction tool  TTD 5032  over the cold trap so that the  notch of the tool sits directly behind the cold trap collar       While holding the tool in place begin to unsdrew the knob at the front  of thre trap extraction tool which will extract 
135. sure outlet  see figure 13      If the capillary injection port is not required  then it can also provide an easy  access route for the heated transfer line to enter the GC oven  See section  3 3 1 for details     2 4 6 4 Connection to other GC systems    Electronic pneumatic control using Shimadzu GC systems is also possible  The  carrier gas and septum purge line are configured similarly to both the Agilent  and Thermo Electron systems  For more information please contact Markes  International Ltd     2 5 TT24 7 flow schematics    This section reviews the flow schematics for the TT24 7 for the three hardware  configurations available  TT24 7e1  TT24 7e2 and TT24 7e3   These will identify  the direction of flow of carrier gas  sampling gas  and dry purge gas for two  sampling modes Continuous Sampling and Tube Desorb     Each section shows the chronological order of events during the thermal  desorption process     The solenoid valves are labelled with a numerical SV designation  going from SV1  through to SV7  The status of each solenoid valve is either ON or OFF     SV1  SV2  SV3 and SV4 are three way valves whose ON   OFF status controls the  direction of carrier gas flow     SV5  SV6 and SV7  where fitted  are two way ON   OFF solenoids     The heated  PTFE  valves are labelled HVA  HVB and HVC  where fitted   Each  valve is in an UP or DOWN status depending on the two pin position of each valve     Total  carrier  gas flow into the TT24 7 is controlled by the electronic
136. system    4 2 Contamination   the presence of artifacts in the  chromatogram    4 2 1 The carrier gas supply  4 2 2 Contamination from the cold trap  4 2 3 Contamination from the sorbent tubes  e2 e3 systems   4 2 4 Other potential sources of contamination  4 3 Poor peak shape   peak splitting    4 4 Carryover of components of interest    4 5 Poor recovery   loss of sample    Appendix 1  Packing List and Routine Maintenance    Spares    Appendix 2  Dynamic Data Exchange  DDE  software    control    Appendix 3  Sorbent specifications    1  Carbotrap C     20 40 mesh    Carbopack C     60 80 mesh       Carbograph 2TD  range of mesh sizes   2  Tenax TA    or GR     range of mesh sizes   3  Carbotrap     20 40 mesh    Carbopack B     60 80 mesh       GCBi     range of mesh sizes    Carbograph 1TD  range of mesh sizes     QUI 1016 V2 2 Aug 06  6     110    111  111  112    112    114    114    114    115  115  116  116    117  117  118    118    119    125    128    128  128    129    Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International      Carbopack X     40 60 and 60 80 mesh    Carbotrap X     20 40 mesh     Carbograph 5TD  various mesh sizes     4  5  6  Chromosorb 102     range of mesh sizes available   7  Porapak Q  range of mesh sizes available    8  Porapak N  range of mesh sizes available    9  HayeSep D  range of mesh sizes available    10  Chromosorb 106     range of mesh sizes available
137. t  Trap A Not Heating   1 or O     lt TBH gt  Trap B Not Heating   1 or O     lt FVH gt  Front Valve Not Heating   1 or O     lt TLH gt  Transfer Line Not Heating   1 or O     lt TOH gt  Tube Oven Not Heating   1 or O     lt RVH gt  Rear Valve Not Heating   1 or O     lt TAT gt  Trap A Over Temperature   1 or O     lt TBI gt  Trap B Over Temperature   1 or O     lt FVT gt  Front Valve Over Temperature   1 or O     lt TLT gt  Transfer Line Over Temperature   1 or O     lt TOT gt  Tube Oven Over Temperature   1 or O    QUI 1016 V2 2 Aug 06  125     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International     lt RVT gt  Rear Valve Over Temperature   1 or O      lt split gt  Split Valve Open   1 or O      lt flow gt  Current Flow Rate    lt error gt  General Instrument Error occurred   1 or O    Temperatures    Returns  A comma separated list of values in the order shown below     lt TA gt  The Temperature of Trap A    lt TB gt  The Temperature of Trap B    lt TL gt  The Temperature of the Transfer Line    lt FV gt  The Temperature of the Front Heated Valve    lt RV gt  The Temperature of the Rear Heated Valve    lt TO gt  The Temperature of the Tube Oven  CurrentMethod    Returns  The name of the currently active method  FlowRate   Returns  The current Flow value   ActiveTrap   Returns  Which trap is currently active  SampleTime   Returns  The current sample time    Commands   The following commands 
138. t  ic    Tnstrument Status eated Zone  GC Flow Controller   TrapA State  0  rap A  0 0   25 0 TubeOven   00 2500C   State  Not Ready    low F 07 500 mimin  TrapB State  0  rap B  10   Rear HV  0 0  120 0  C e ir   Sample Time 00 min i xr       Figure 44  TT24 7 software   top level screen    QUI 1016 V2 2 Aug 06  61     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    At the top of the screen are function icons and a menu bar for instrument  configuration and control     2 7 Software installation    Remove the software CD from its packaging  and insert into the CD drive of the  PC     Access the CD drive from Windows and click on the Setup icon  Follow the on  screen instructions to complete the installation of the software     2 8 Firmware download    After the software has been loaded onto the PC the firmware has to be  downloaded into two separate boards within the TT24 7 system     Firstly ensure that all the electrical connections between the TT24 7  the PC and  the rest of the analytical system are in place  See section 3 1 for details     On the front panel of the TT24 7 there are two LED display lights and an  instrument activity switch  The LEDs refer to the status of the Sampler and MFC  boards within the TT24 7 and show different colours depending on the download  status of the firmware  figure 45             E              HHE  S352   E      SESz       s 2 Sampler LED  MFC L
139. tems the transfer line inlet is situated behind the front two heated  valves and the transfer line is installed vertically down into the union     To obtain access to the installation point  first remove the heated valve cover  as follows     Loosen the two screws holding the back panel in place  figure 75  and remove  the panel     Back panel screws       Figure 75  Loosen the back panel screws    Remove the four M4 nuts at the front of the valve box and lift the box off  figure  76     QUI 1016 V2 2 Aug 06  101     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International       Figure 76  Remove the M4 nuts    Figure 77shows the     ie inch stainless steel union into which the fused silica  transfer has to be fitted  Figure 78 shows a partly sectioned view of this   16   inch union giving an indication of where to position of the end of the fused silica    tubing        Figure 77  Transfer line connection    QUI 1016 V2 2 Aug 06  102     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Carrier bypass    Fused silica  terminates  ere  1 16 inch inert  oated stainless  steel tubing  0 5  mm bore     Analytes from the trap  via the heated ee       Figure 78  Precise positioning of end of fused silica transfer line    If the fused silica terminates in the space above the end of the inert coated  stainless steel
140. ternational    2 5 2 2 7 Trap  A  desorption    Desorption Tube  Purge  Vent     Needle  Vale    MFC  amp  Pump Sample in    Trap B       KEY          Flow of Carrier Gas    Carrier Gas  no flaw       a    Flow of Sample Gas  Sample Gas  no flow        Figure 31  TT24 7e2   Tube desorb mode   trap desorption    The selected trap is now rapidly heated and backflushed with carrier gas to  remove retained analytes from the sorbent bed s  into the capillary column     The flow rate through the trap is equal to the column flow and should be at  least 3 mL min     QUI 1016 V2 2 Aug 06  48     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 3 TT24 7e3 hardware configuration  TT24 7e3 includes the option to split sample and may be operated in both  sampling modes   continuous sampling and tube desorb    2 5 3 1 Continuous sampling mode  TT24 7e3    2 5 3 1 1 Standby    MFC  amp  Pump Sample in                  Flow of Carrier Gas       Optional Carrier Gas Flow  Carrier Gas  no flow            Flow of Sample Gas  Sample Gas  no flow        Figure 32  TT24 7e3   Continuous sampling   standby    In this non sampling mode  carrier gas alone is continuously supplied to the  capillary column  No other gas flow occurs within the TT24 7 unit during this  time     QUI 1016 V2 2 Aug 06  49     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 Us
141. the cold trap from the heated  valve o ring       Once fully unscrewed the knob can then be pulled forward to extract the  cold trap from its housing     Note  Cold traps are much easier to remove whilst the o rings in the  heated valves are still hot     Uncollared Traps     Carefully pull each of the cold traps out    Note  Using a latex glove will provide better grip on the traps to aid  removal     Note  Cold traps are much easier to remove whilst the o rings in the  heated valves are still hot      If the cold traps remain difficult to pull out then switch the instrument  back on and re download the firmware  allow the heated valves to attain  their temperature set point  switch the instrument off and repeat the  process     3 4 3 Available Cold Traps  A range of pre packed TT24 7 cold traps are available from Markes as follows     QUI 1016 V2 2 Aug 06  110     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Pareno   peseripton   T 1VX TT24 7 cold trap optimised for the sampling  of CW agents  specifically VX    T 2HD TT24 7 cold trap optimised for the sampling  of CW agents  specifically Mustard  HD      T 3GBGE TT24 7 cold trap optimised for the sampling  of CW agents  specifically Sarin  GB  and  the G analogue of VX  GE     T 6EMP TT24 7 empty cold trap for packing by user   T 7CUS 1124 7 cold trap custom packed to user  specification       3 4 4 User maintenance of TT24 7 col
142. then back ON  After this has occurred the firmware is  automatically downloaded into the TT24 7  The desired sampling method can  then be started again     The recycle box has two power  IN   OUT  supply sockets  and two  IN   OUT   serial ports  The power supply to the TT24 7 now passes through the recycle box  which in turn connects into the power socket of the TT24 7  The power cable  connecting the recycle box to the TT24 7 has a female socket at either end  P N  Z 0207      The serial port connection to the Sampler socket in the TT24 7 also passes  through the recycle box and it is this connection which initiates a power cycle  when the software is downloaded     Figure 71 shows a schematic of this configuration     USB Hub  USB to Serial MFC cable  P N U USBSR        Standard  USB Cable    Power Recycle Box  Serial Cable Serial Cable    Remote Start        P N 2 0189   PIN 2 0189     Power connector   P N seeccoad          PC to GC Comms cable  eg LAN RS232      Agilent  UTD 5098  Remote Start Cable P N  Thermo  UTD 5108       Figure 71  Cable connections including power recycle box    3 2 Gas requirements  The gases required by the TT24 7 system are  Carrier Gas  Sample gas  continuous sampling mode   Trap box purge gas and heated valve actuator pneumatics    The quality of both the carrier gas and the purge gas is an important  consideration     For the carrier gas the quality must be 5 0 grade  99 999   or better  This can  be achieved directly from the supply tank o
143. tinue running       t   Initiates a leak test procedure  The first function is to  perform a leak test of the tube  followed by a leak test of trap  A and then trap B    Allows manual  needle valve  control of the split flow through  z SV7  if enabled   e3 systems  and dry purge flow through  SV6  QUI 1016 V2 2 Aug 06  85     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual    Markes International    2 9 1 4 2 Menu bar    File  File View Instrument Hel  k New     Open  Save Ctrl 5  Save As     Ctrl Alt 5          Simple Mode       Exit Ctrl x       New   Open     Save     Save As     Exit     View    B 1124 7 Control  File   view Instrument    b Options  Diagnostics       Options     Creates a new method which starts with default parameters    Opens the method subdirectory so that an existing method  can be loaded into the TT24 7    Saves the current TT24 7 parameters directly into the loaded  method    Requests a method file name to be saved into the methods  sub directory  No suffix is required as the  mth suffix is  automatically added    Immediately closes down the TT24 7 software    Accesses the 1124 7 configuration software  see section  2 9 1 5 for details     Diagnostics    Accesses two levels of diagnostic software for the    Instrument    B 1124 7 Control  File View   Instrument    D    EP Run       Run   Stop     QUI 1016 V2 2 Aug 06    TT24 7    flow only  user  diagnostics and full  service  engin
144. tly and the TT24 7 will  not function     QUI 1016 V2 2 Aug 06  111     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    3 4 4 2 Cold trap lifetime  Cold trap lifetime is dependent upon a number of factors       the sorbent s  type used in the trap  porous polymers have a lower  lifetime than graphitised carbon sorbents for example       the maximum temperature that the trap is routinely heated to during  trap fire  and the length of time it is held at this temperature  operating  your traps close to the maximum temperature of the sorbents within them  will reduce trap lifetime       the nature of the compounds being analysed    When the performance of the trap starts to deteriorate it is time to replace both  traps     In any event it is recommended that cold traps are replaced annually     3 5 Installation   removal of sampling tube  e2 e3     The sampling tube is located on the left hand side of the TT24 7 when looked at  from the front of the instrument  figure 87         Figure 87  TT24 7e2 e3 system with Desorb Tube option on LHS    To access the tube oven which houses the sampling tube the sealing mechanism  lever should be lifted  see figure 88      QUI 1016 V2 2 Aug 06  112     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International       Figure 88  TT24 7e2 e3 system   tube sealing mechanism lever
145. trolling PC will typically have  a battery backup for a few hours     When the power returns the GC will switch on and reset to its original method  values  however the TT24 7 requires the firmware to be downloaded again  This  requires the TT24 7 software to be closed as it will still be active due to the battery  backup in the PC  If the power failure extends beyond the battery backup of the  PC  then the PC software will have to be rebooted from scratch  User name login  and passwords will be required to access the Microsoft Windows desktop  followed  by rebooting of both the TT24 7 and GC software     To successfully download the firmware into the TT24 7 it is recommended that the  instrument is switched OFF and then ON  at the back of the instrument   This will  turn both LED lights to red as discussed in section 2 8  The TT24 7 is now ready  to receive the firmware download and this is achieved by rebooting the software  as described in section 2 8     However where remote system control is active  the ON   OFF switch at the rear    QUI 1016 V2 2 Aug 06  96     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    of the instrument cannot be physically accessed  so an alternative mechanism is  required     This is achieved using a power recycle box  P N U RCYBX   The power recycle box  is activated when the TT24 7 software is initiated  this electronically switches the  system OFF and 
146. ts own independent carrier gas supply  separate from any other  conventional chromatographs in the laboratory     Note  TT24 7 is such a good concentrator of VOCs that normal laboratory gas  lines  which perform perfectly well for conventional GC analyses  can produce  artifacts on the system  It is recommended that the gas itself and gas line  components meet the requirements stated in section 3 2     4 2 2 Contamination from the cold trap    If the contamination is shown  from the experiment described above  not to be  coming from the carrier gas  the next most likely candidate is the sorbent in the  traps  or sorbent tube where applicable  e2 e3 systems       See Appendix 3 for information regarding sorbent conditioning   maximum  sorbent temperatures etc     To condition the traps set up a continuous sampling method with the following  parameters       an extended pre purge time  e g  5 minutes      a short sampling time  e g  30 seconds        a maximum trap temperature which is 10   20  C higher than that  normally used for analysis   assuming that this does not exceed the  sorbent maximum temperature  see Appendix 3   typically to extend trap  lifetime  your operating method should not require the trap to be used at  its maximum temperature        an extended trap hold time  e g  5 minutes      run this method through both traps       you may need to run the method through each trap more than once  depending on the contamination levels found     4 2 3 Contamination fro
147. ty cold trap  T 7CUS TT24 7 cold trap custom packed to your specification       QUI 1016 V2 2 Aug 06  124     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    Appendix 2  Dynamic Data Exchange  DDE  software  control    Dynamic data exchange  DDE  provides a mechanism to control and monitor the  TT24 7 externally to the standard system software  The section below supplies the  control and status commands to achieve this     DDE commands are typically embedded in macro programs within the  GC    GCMS  host software  and enable that software to take control of the TT24 7  This  is usually implemented to add some custom functionality to the system  An  example of this would be to modify a method parameter or entire method based  on an analytical result     The DDE functionality is designed for customers with a strong background in  software programming     TT24 7 DDE Interface     To create the connection to the DDE interface you will need to use the following  settings     Service Name  Topcat  Topic  DDEInterface    Information Requests  The following information can be returned from the TT24 7 software   The Request item names are case sensitive     State  Returns  A comma separated list of values in the order shown below   lt mode gt  Operating Mode   lt state gt  Standby  Running or LeakTest   lt TAL gt  Trap A Leak Error   1 or O     lt TBL gt  Trap B Leak Error   1 or O     lt TAH g
148. ual Markes International    2 5 3 1 4 Trap A purging  trap B sampling    Desorption Tube    q Needle   N valve    i n H  He i C  see   O H MFC  amp  Pump Sample in  off on    Trap B TA  a         Split Tube    KEY       gt     Flow of Carrier Gas      Optional Carrier Gas Flow  Carrier Gas  no flow           Flow of Sample Gas  Sample Gas  no flow        Figure 35  TT24 7e3   Continuous sampling   trap A dry purging  trap B  sampling    After the sampling time for trap A has completed trap A switches into its dry   purge mode to sweep out air and moisture prior to trap fire  Having cooled to  its starting temperature after the first clean up trap fire  trap B now goes into  its sampling phase  The date and time that the sampling started  and the trap  that is being sampled are reported as an information message in the reporting  module of the software  see section 2 9 2 7   e g     05 12 2005 10 49 31   INFO  Sampling Started on Trap A    Carrier gas also purges down the split line to purge any air from the line and to  prevent ingress into the system     QUI 1016 V2 2 Aug 06  52     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 3 1 5 Trap A desorbing  trap B sampling    MFC  amp  Pump Sample in       Needle    Valve off  Split Tube    me OOS    G  a     i    KEY       gt     Flow of Carrier Gas      Optional Carrier Gas Flow  Carrier Gas  no flow        gt    _ Flow of Sample Ga
149. uch sharper peak and a  significantly better S N ratio even though the amount of analyte trapped is half     The science behind this effect is based on the depth the analyte passes into the  sorbent  and as a consequence how easily it comes off at trap fire  It is  therefore both compound and sorbent dependant     2 9 1 1 5 Sample time    This specifies the time interval for sample to pass into the trap  This parameter  is directly associated with the sampling flow rate as discussed above to  determine the total amount of sample transferred  The range for sample time  is from 0 1 to 99 9 minutes  however  values greater than 1 minute will typically  be used     The sampling time and therefore the sampling volume can be affected by the    QUI 1016 V2 2 Aug 06  66     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    ready   not ready status of the GC system  At the completion of the defined  set point time value and prior to the system going into the pre purge phase of  the trap  the TT24 7 monitors the GC ready status  If the GC is ready then the  pre purge process commences  If however the GC is not ready then the  sampling time is extended until the GC becomes ready  or a specified    time out     value  see section 2 9 1 5 3  is equalled at which point the system stops  The  GC becoming not ready is typically due to the oven not cooling down as fast as  normal and this affects the GC 
150. using  Figures 4 and 5  below show the TT24 7 with the tube accessory fitted        Figure 4  TT24 7e2 showing tube accessory fitted and sorbent tube ready to be  sealed into position       Figure 5  TT24 7e2 showing tube accessory fitted and sorbent tube sealed into  flow path    The flow schematics for the tube desorption process can be seen in section 2 5 2    The ability to desorb a single tube using the TT24 7 considerably extends the  functionality of the system  beyond that of a single mode continuous sampler     QUI 1016 V2 2 Aug 06  19     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    An example of this flexibility is the capacity of the system to desorb the analytes  retained within the tube onto either trap A or trap B  This is a user selectable  function within the TT24 7    Tube Mode    sampling configuration in the software   See section 2 9   This enables equivalency of trap performance to be assessed by  desorbing a test sample initially into trap A and then into trap B  and comparing  the data  This could also be used as a method validation tool to check for system  bias for each trap     Single tube desorption also enables calibration of the detector within the  associated GC or GCMS system  If a tube is spiked with a standard of known  concentration and analysed then the detector can be easily calibrated  This will  enable quantitative analysis of subsequent sample
151. w of Carrier Gas     Carrier Gas  no flaw        gt     Flow of Sample Gas  Sample Gas  no flow        Figure 26  TT24 7e2   Tube desorb mode   leak testing tube    The first stage in tube desorption is a tube leak test  This is essential as the  tube has been inserted into the TT24 7 flow path and sealed with two retaining  o rings one at the front and back of the tube oven  See section 3 5 for tube  installation instructions     During this phase the tube desorption flow lines in the TT24 7 are initially  pressurised with carrier gas and then solenoid SV5 is switched OFF to seal the  system  The heated valves are configured so that there is no connection  beyond HVC  The pressure sensor P adjacent to SV5 is monitored to measure  the rate of pressure decay  If this is acceptable   lt  5  drop in pressure within  30 seconds  the software moves on to the next phase in the tube desorption  process     QUI 1016 V2 2 Aug 06  43     Markes International Ltd  T   44  0 1443 230935 F   44  0 1443 231531 E  enquiries markes com    TT24 7 User Manual Markes International    2 5 2 2 3 Leak testing trap A    MFC  amp  Pump Sample in    Trap B       KEY          Flow of Carrier Gas     Carrier Gas  no flaw       a    Flow of Sample Gas  Sample Gas  no flow        Figure 27  TT24 7e2   Tube Desorb Mode   Leak testing trap A    After the tube leak test the selected trap also undergoes a further leak test  In  the example above trap A is to be used  The heated valve HVC now switches  to all
152. y way   2 9 1 7 2 Warning messages   WARNING  Sampling volume outside of expected range      this warning message is given when the total volume sampled is not  what would be expected from the method  For example this could be  because the sampling time was extended  see section 2 9 1 5 3  or  because the flow controller was unable to pull the required flow rate  over the sampling time  due to a failing cold trap or other problem      A warning message does not affect the operation of the TT24 7  system but gives information about something which may affect the  quantitation and the results     2 9 1 7 3 Error messages   ERROR  Trap A not heating   ERROR  Valve not heating   ERROR  Line not heating   ERROR  Trap A over temperature   ERROR  Valve over temperature   ERROR  Line over temperature    If one of the above errors occur then the TT24 7 system will shut down and  await operator   service engineer intervention     2 9 2 User interface for TT24 7e2 and TT24 7e3 hardware configurations    2 9 2 1 Method parameters    At the top of the method window is shown the active method name  e g   default mth  and its status i e  whether or not it has been modified from its  original set point values     The set point values used in the interface are described below   2 9 2 1 1 Mode    Hardware configurations e2 and e3 allow two different sampling modes   each  with a slightly different user interface     Figure 58 shows the two user interfaces for the different sampling modes     Q
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Hinweise zur Bedienungsanleitung NV-GS1EG NV  Manual W64 Boretti aktuelle Version  Les couleurs du blues ! - Corbeil    VIAIR 90223 Use and Care Manual  MANUAL DEL OPERADOR  Guía del usuario Ulead Systems, Inc. Enero de 2006  SMARTOUCH DIGITAL USER MANUAL    Copyright © All rights reserved. 
   Failed to retrieve file