Home
        Human Factor X ELISA Kit
         Contents
1.  at 1 2 in MIX Diluent and assay   however  the user should determine the optimal dilution factor  The  undiluted samples can be stored at  20  C or below for up to 3 months   Avoid repeated freeze thaw cycles    e Urine  Collect urine using sample tube  Centrifuge samples at 800 x g for  10 minutes and assay  Store samples at  20  C or below for up to 3  months  Avoid repeated freeze thaw cycles    e Saliva  Collect saliva using sample tube  Centrifuge samples at 800 x g  for 10 minutes and assay  Store samples at  20  C or below for up to 3  months  Avoid repeated freeze thaw cycles    e CSF  Collect cerebrospinal fluid  CSF  using sample tube  Centrifuge  samples at 3000 x g for 10 minutes and assay  The undiluted samples can  be stored at  80  C for up to 3 months  Avoid repeated freeze thaw  cycles     Refer to Sample Dilution Guidelines below for further instruction     Guidelines for Dilutions of 1 100 or Greater   for reference only  please follow the insert for specific dilution suggested     1 100    1 10000       A     4 ul sample  396 ul buffer 100x     100 fold dilution    Assuming the needed volume is less than  or equal to 400 ul     4 ul sample   396 ul buffer  100x   4 ul of A   396 ul buffer  100x      10000 fold dilution  Assuming the needed volume is less than  or equal to 400 ul        1 1000    1 100000       4 ul sample   396 ul buffer  100x   24 ul of A  216 ul buffer  10x     1000 fold dilution    Assuming the needed volume is less than  or equal to 2
2.  determined by testing three plasma samples in    twenty assays     Intra Assay Precision Inter Assay Precision  Sample 1 2 3 1 2 3  n 20 20 20 20 20 20  CV      Average  CV                                            Spiking Recovery    e _ Recovery was determined by spiking two plasma samples with different  factor X concentrations     Unspiked  Sample   ng ml     Spike Recovery   ng ml         5 0 11 2 10 8 96   6 2 15 0 21 2 20 1 95   25 0 31 2 28 3 91   5 0 17 3 19 4 112   15 0 27 3 26 8 98   25 0 37 3 35 6 95     Average Recovery     98     Expected Observed                                              Linearity  e Plasma and serum samples were serially diluted to test for linearity     Average Percentage of Expected Value       Sample Dilution Plasma Serum  1 400 107  105   1 800 98  101   1 1600 101  96                          10    Cross Reactivity    Species    Cross Reactivity           Canine    None       Bovine    None       Monkey    None       Mouse    None       Rat    None       Swine    None       Rabbit    None       Proteins    Cross Reactivity           Human Factor X    100        Human Factor Xa       Troubleshooting    100                       Issue Causes Course of Action  Use of expired e Check the expiration date listed before use   components e Do not interchange components from different lots   e Check that the correct wash buffer is being used   e Check that all wells are dry after aspiration   Improper wash step e Check that the microplate was
3. 40 ul     Reagent Preparation       4 ul sample   396 ul buffer  100x   4 ul of A   396 ul buffer  100x   24 ul of B   216 ul buffer  10x      100000 fold dilution  Assuming the needed volume is less than  or equal to 240 ul        Freshly dilute all reagents and bring all reagents to room temperature    before use     MIX Diluent Concentrate  10x   If crystals have formed in the  concentrate  mix gently until the crystals have completely dissolved   Dilute the MIX Diluent Concentrate 1 10 with reagent grade water  Store    for up to 30 days at 2 8  C     Standard Curve  Reconstitute the 400 ng  48 mU ml  of Human Factor X  Standard with 2 ml of MIX Diluent to generate a 200 ng ml  24 mU ml   standard stock solution  Allow the standard to sit for 10 minutes with  gentle agitation prior to making dilutions  Prepare duplicate or triplicate  standard points by diluting the standard stock solution  200 ng ml  24  mU ml  1 2 with MIX Diluent to generate 100  50  25  12 5  6 25  3 125   and 1 563 ng ml solutions  MIX Diluent serves as the zero standard  0  ng ml   Any remaining solution should be frozen at  20  C and used    within 30 days     Standard wa  FX   Point Dilution  ng ml   1 part Standard  200 ng ml       PE 1 part MIX Diluent 100 0    1 part P1   1 part MIX Diluent 50 00   600    1 part P2   1 part MIX Diluent 25 00         Pa    1partP3 1 part MIX Diluent   1250   150      Ps    1partP5 1 part MIX Diluent   3 125   0375    ls   MiXDiluent_   0 000   0 000         e Bi
4. an active enzyme FXa  FXa  as the activator of prothrombin   occupies a central position linking the two blood coagulation pathways  1 4      Principle of the Assay    The AssayMax Human Factor X ELISA  Enzyme Linked Immunosorbent Assay   kit is designed for detection of human factor X in plasma  serum  milk  urine   saliva  CSF  and cell culture samples  This assay employs a quantitative  sandwich enzyme immunoassay technique that measures FX in less than 4  hours  A monoclonal antibody specific for FX has been pre coated onto a 96   well microplate with removable strips  FX in standards and samples is  sandwiched by the immobilized antibody and biotinylated polyclonal antibody  specific for FX  which is recognized by a streptavidin peroxidase conjugate  All  unbound material is washed away and a peroxidase enzyme substrate is  added  The color development is stopped and the intensity of the color is  measured     Caution and Warning    e This product is for Research Use Only and is Not For Use In Diagnostic  Procedures      Prepare all reagents  working diluent buffer  wash buffer  standard   biotinylated antibody  and SP conjugate  as instructed  prior to running  the assay    e Prepare all samples prior to running the assay  The dilution factors for  the samples are suggested in this insert  However  the user should  determine the optimal dilution factor    e Spin down the SP conjugate vial and the biotinylated antibody vial before  opening and using contents    e The Sto
5. e stored for up to 30 daysina  vacuum desiccator    e Diluent  1x  may be stored for up to 30 days at 2 8  C    e Store Standard at 2 8  C before reconstituting with Diluent and at  20  C  after reconstituting with Diluent     Other Supplies Required  e Microplate reader capable of measuring absorbance at 450 nm     e Pipettes  1 20 ul  20 200 ul  200 1000 ul  and multiple channel    e Deionized or distilled reagent grade water     Sample Collection  Preparation  and Storage    e Plasma  Collect plasma using one tenth of 0 1 M sodium citrate as an  anticoagulant  Centrifuge samples at 3000 x g for 10 minutes and use  supernatants  Dilute samples 1 800 into MIX Diluent and assay  The  undiluted samples can be stored at  20  C or below for up to 3 months   Avoid repeated freeze thaw cycles  EDTA or Heparin can also be used as  an anticoagulant     e Serum  Samples should be collected into a serum separator tube  After  clot formation  centrifuge samples at 3000 x g for 10 minutes  and  remove serum  Dilute samples 1 800 into MIX Diluent and assay  The  undiluted samples can be stored at  20  C or below for up to 3 months   Avoid repeated freeze thaw cycles    e Cell Culture Supernatants  Centrifuge cell culture media at 3000 x g for  10 minutes to remove debris  Collect supernatants and assay  Store  samples at  20  C or below  Avoid repeated freeze thaw cycles    e   Milk  Collect milk using sample tube  Centrifuge samples at 800 x g for  10 minutes  Milk dilution is suggested
6. her is dispensing properly   e If washing by pipette  check for proper pipetting    technique   o Splashing of reagents e Pipette properly in a controlled and careful manner   8 while loading wells  2 i e Pipette properly in a controlled and careful manner   a Inconsistent volumes y ee     e Check pipette calibration   3 loaded into wells i  o e Check pipette for proper performance            Insufficient mixing of  reagent dilutions    e Thoroughly agitate the lyophilized components after  reconstitution   e Thoroughly mix dilutions        Improperly sealed  microplate    e Check the microplate pouch for proper sealing    e Check that the microplate pouch has no punctures    e Check that three desiccants are inside the microplate  pouch prior to sealing           Unexpectedly Low or High    Signal Intensity    Microplate was left  unattended between  steps    e Each step of the procedure should be performed  uninterrupted        Omission of step    e Consult the provided procedure for complete list of steps        Steps performed in  incorrect order    e Consult the provided procedure for the correct order        Insufficient amount of  reagents added to  wells    e Check pipette calibration   e Check pipette for proper performance        Wash step was skipped    e Consult the provided procedure for all wash steps        Improper wash buffer    e Check that the correct wash buffer is being used           Improper reagent  preparation       e Consult reagent preparation secti
7. man Factor X ELISA Kit with Positive and Negative    Controls  Plasma  Serum  Urine  Milk  Saliva  Cell Culture  and CSF  samples        www assaypro com    e mail  support assaypro com  11    
8. on for the correct  dilutions of all reagents           Insufficient or  prolonged incubation  periods    e Consult the provided procedure for correct incubation  time        Deficient Standard Curve Fit    Non optimal sample  dilution    e Sandwich ELISA  If samples generate OD values higher  than the highest standard point  P1   dilute samples  further and repeat the assay    e Competitive ELISA  If samples generate OD values lower  than the highest standard point  P1   dilute samples  further and repeat the assay    e User should determine the optimal dilution factor for  samples        Contamination of    e A new tip must be used for each addition of different       reagents samples or reagents during the assay procedure   Contents of wells e Verify that the sealing film is firmly in place before placing  evaporate the assay in the incubator or at room temperature        Improper pipetting    e Pipette properly in a controlled and careful manner   e Check pipette calibration   e Check pipette for proper performance           Insufficient mixing of  reagent dilutions       e Thoroughly agitate the lyophilized components after  reconstitution   e Thoroughly mix dilutions        References     1  Ruf  W  and Edgington  T S   1994  FASEB J  8 385    2  Neuenschwander  P F  et al   1993  Thrombosis and Haemostasis 70 970   3  Messier  T L  et al   1991  Gene 99 291    4  DiScipio  R G  et al   1977  Biochemistry 16 5253    Version 5 3R  Related Products  e EF1010 7 AssayMax Hu
9. otinylated Human Factor X Antibody  60x   Spin down the antibody  briefly and dilute the desired amount of the antibody 1 60 with MIX  Diluent  Any remaining solution should be frozen at  20  C    e Wash Buffer Concentrate  20x   If crystals have formed in the  concentrate  mix gently until the crystals have completely dissolved   Dilute the Wash Buffer Concentrate 1 20 with reagent grade water    e SP Conjugate  100x   Spin down the SP Conjugate briefly and dilute the  desired amount of the conjugate 1 100 with MIX Diluent  Any remaining  solution should be frozen at  20  C     Assay Procedure    e Prepare all reagents  standard solutions  and samples as instructed  Bring  all reagents to room temperature before use  The assay is performed at  room temperature  20 25  C     e Remove excess microplate strips from the plate frame and return them  immediately to the foil pouch with desiccants inside  Reseal the pouch  securely to minimize exposure to water vapor and store in a vacuum  desiccator    e Add 50 ul of Human Factor X Standard or sample per well  Cover wells  and incubate for 2 hours  Start the timer after the last addition    e Wash five times with 200 ul of Wash Buffer manually  Invert the plate  each time and decant the contents  hit 4 5 times on absorbent material  to completely remove the liquid  If using a machine  wash six times with  300 ul of Wash Buffer and then invert the plate  decanting the contents   hit 4 5 times on absorbent material to completely remo
10. own sample concentration from the Standard Curve  and multiply the value by the dilution factor     Typical Data    The typical data is provided for reference only  Individual laboratory  means may vary from the values listed  Variations between laboratories  may be caused by technique differences     Standard Point Average OD  P1 100 0       P2 50 00       P3 25 00       P4 12 50       PS 6 250       P6 3 125       P7 1 563       P8 0 000             Sample  Pool Normal   Sodium Citrate Plasma  800x              Standard Curve    e The curve is used for illustration only  A standard curve should be  generated each time the assay is performed     Human FX Standard Curve    OD 450 nm          0 1 L  1 0 10 0 100 0     f X   ng ml     Human FX Standard Curve    OD 450 nm          0 1 tr 1   r  r  l po iil po iii   0 1 1 0 10 0 100 0     f X   mU ml   Reference Value  e Normal human factor X plasma levels range from 3 to 13 ug ml     e Human plasma and serum samples from healthy adults were tested   n 40   On average  factor X level was 8490 ng ml     Sample Average Value  ng ml        Human Pool Normal Plasma 7926       Human Normal Plasma 8124          Human Pool Normal Serum 9421          Performance Characteristics    e The minimum detectable dose of factor X as calculated by 25D from the  mean of a zero standard was established to be 0 4 ng ml   e Intra assay precision was determined by testing replicates of three    plasma samples in one assay   e inter assay precision was
11. p Solution is an acidic solution     e The kit should not be used beyond the expiration date   Reagents    e Human Factor X Microplate  A 96 well polystyrene microplate  12 strips  of 8 wells  coated with a monoclonal antibody against human FX    e Sealing Tapes  Each kit contains 3 precut  pressure sensitive sealing  tapes that can be cut to fit the format of the individual assay    e Human Factor X Standard  Human FX in a buffered protein base  400 ng   lyophilized     e Biotinylated Human Factor X Antibody  60x   A 60 fold biotinylated  polyclonal antibody against human FX  100 pl     e  Streptavidin Peroxidase Conjugate  SP Conjugate   A 100 fold  concentrate  80 ul     e _ MIX Diluent Concentrate  10x   A 10 fold concentrated buffered protein  base  30 ml     e Wash Buffer Concentrate  20x   A 20 fold concentrated buffered  surfactant  30 ml  2 bottles     e Chromogen Substrate  A ready to use stabilized peroxidase chromogen  substrate tetramethylbenzidine  8 ml     e Stop Solution  A 0 5 N hydrochloric acid to stop the chromogen  substrate reaction  12 ml      Storage Condition    e Upon arrival  immediately store components of the kit at recommended  temperatures up to the expiration date    e Store SP Conjugate and Biotinylated Antibody at  20  C    e Store Microplate  Diluent Concentrate  10x   Wash Buffer  Stop Solution   and Chromogen Substrate at 2 8  C    e     Unused microplate wells may be returned to the foil pouch with the  desiccant packs and resealed  May b
12. ve the liquid    e   Add 50 ul of Biotinylated Human Factor X Antibody to each well and  incubate for 1 hour    e Wash the microplate as described above    e Add 50 ul of Streptavidin Peroxidase Conjugate per well and incubate for  30 minutes  Turn on the microplate reader and set up the program in  advance     Wash the microplate as described above    Add 50 ul of Chromogen Substrate per well and incubate for 10 minutes  or till the optimal blue color density develop  Gently tap the plate to  ensure thorough mixing and break the bubbles in the well with pipette  tip    Add 50 ul of Stop Solution to each well  The color will change from blue  to yellow    Read the absorbance on a microplate reader at a wavelength of 450 nm  immediately  If wavelength correction is available  subtract readings at  570 nm from those at 450 nm to correct optical imperfections   Otherwise  read the plate at 450 nm only  Please note that some  unstable black particles may be generated at high concentration points  after stopping the reaction for about 10 minutes  which will reduce the  readings     Data Analysis    Calculate the mean value of the duplicate or triplicate readings for each  standard and sample    To generate a standard curve  plot the graph using the standard  concentrations on the x axis and the corresponding mean 450 nm  absorbance  OD  on the y axis  The best fit line can be determined by  regression analysis using log log or four parameter logistic curve fit   Determine the unkn
13. yssarpro    AssayMax      Human Factor X  ELISA Kit    Assaypro LLC  3400 Harry S Truman Blvd  St  Charles  MO 63301  T  636  447 9175  F  636  395 7419    WWW assaypro com    For any questions regarding troubleshooting or performing the assay  please contact our  support team at support assaypro com        Thank you for choosing Assaypro     Assay Summary    Step 1  Add 50 ul of Standard or Sample per well   Incubate 2 hours     Step 2  Wash  then add 50 ul of Biotinylated Antibody per well   Incubate 1 hour     Step 3  Wash  then add 50 ul of SP Conjugate per well   Incubate 30 minutes     Step 4  Wash  then add 50 ul of Chromogen Substrate per well   Incubate 10 minutes     Step 5  Add 50 ul of Stop Solution per well   Read at 450 nm immediately     Symbol Key    BE    Consult instructions for use     Assay Template       12       11       10                                                                   Human Factor X ELISA Kit    Catalog No  EF1010 1  Sample insert for reference use only       Introduction    Factor X  FX  is a plasma serine protease zymogen involved in the blood  coagulation cascade  FX is purified from plasma as a two chain protein  consisting of a 45 kDa heavy chain and a 17 kDa light chain  The FX heavy  chain is cleaved during coagulation by several different proteases including  the intrinsic Xase complex  the FX activating enzyme from Russell   s viper  venom  RVV  and trypsin  and also by extrinsic  tissue factor factor Vila   pathway to give 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
scarica il catalogo  Sencor SVC 900 GEMINO vacuum cleaner  Operating instructions for users WPM 2006 plus, WPM    Mode d`emploi Antiglisse Stabibain Acrylique  manuale d`installazione  Manuel d`installation, d`utilisation et d`entretien de la baie de  Philips Indirect light 70979/87/PU  TypeTool 3 for Macintosh User Manual    Copyright © All rights reserved. 
   Failed to retrieve file