900141 O2/CO2 Analyzer Operating Manual
Contents
1. ZERO buttons Press CAL first and hold it down then press ZERO on the analyzer front panel for 1 second When CAL appears on displays the analyzer is in the calibration mode The gas to be calibrated will show numbers on its display while the other display will show CAL 7
2. The Sample and Hold measurement mode is available to facilitate data recording when testing multiple packages or sites In this mode the analyzer draws gas for measurement over the user specified sample period 900141 product is shipped with the factory default setting of 15 seconds then turns the pump off and freezes the display The displayed values are held until the next measurement cycle is executed The UP ARROW button puts the analyzer in the Sample and Hold Measurement Mode and initiates the first test cycle Pressing this button will cause the analyzer to count down the sample time selected then turn the pump off and freeze the displays 3 4 3 When the button is pressed for the first time the analyzer will indicate execution of a sample period by flashing the display every second while taking measurements At the end of the test interval it will turn off the pump and freeze the displayed readings as described above Consecutive measurements may be taken by pushing the UP ARROW button again This process continues indefinitely until the analyzer is either powered off or set back into Continuous Measure Mode done by pressing the DOWN ARROW button Changing the Sample time from the factory default The sample time used in the sample measurement mode may be easily 6 6 adjusted by the operator from the factory default value This is done by first putting the analyzer in Continuous Measurement M
3. Port 9600 baud 11 7 2 RESET the CO2 andO2 Channel Calibrations to Factory Settings Use the UP DOWN Arrows If you suspect that the analyzer is operating incorrectly and has been previously mis calibrated you may RESET previous calibrations and revert to the as shipped factory calibration constants Use this procedure if calibration values have been saved but you want to restore default starting calibration values To do this simple enter the Calibration Mode as described above and then press the DOWN and UP arrows simultaneously for 1 second When CAL disappears from the display the Calibration Mode has been terminated and the analyzer calibration constants have been reset to the original as shipped factory values After the calibration constants have been reset the analyzer may be operated with factory calibration constants or may be re calibrated as described above The previous calibration values are permanently erased The factory default calibration values are used and stored in power safe memory for later use when powered down Consumables Spare Parts and Accessories The following spare parts consumables and accessories can be purchased from Bridge Analyzers Consumables P N Part Name Life 001015 1 16 Thk Sticky Nickel Septum N A 108126 Sample Filter 6 pack 6 12 months 001140 Oxygen Sensor 12 24 months Spare Parts P N Part Name Life 000617 Sample Needle N A 001370 Sample Line w
4. switch with a white protective rubber sealing boot The first activation turns power ON the second activation turns power OFF The analyzer will perform a self test indicated by temporarily lighting all display elements with 8 s and running the pump At the end of the self test the pump will be off and the display will show dashes indicating successful completion of self test and awaiting operation Operation begins when the operator initiates the first Zero A note about Zeroing the Analyzer The most important thing to remember is that the accuracy of the analyzer is assured by periodically Zero ing the analyzer on ambient air When the analyzer is zeroed the CO2 measurement channel is set to zero and the oxygen channel is calibrated to ambient O2 20 6 The automatic Zero function may be performed at any time by simply pressing the ZERO button for 1 2 second 3 1 3 2 Note For residual oxygen applications the sample needle does not have to be removed from the praduct package during the Zero procedure The analyzer automatically switches to a separate Zero port at the top left side of the case for drawing in ambient air for the room air Zero process For High Oxygen applications room air preconditioning may be required before the Zero is activated First Zero after Power Up Initiate the first Zero after power up by pressing the ZERO button on the display panel for one second The pump will start the d
5. Do Not Print This Page e Load 1 sheet of cardstock paper e Print Page 2 the cover e For the body of the text load 3 sheets of 281b matte finish paper into the printer e Print pages 3 5 7 Sheet 1 2 3 e Load page 7 The third text body sheet printed in the printer face up and print page 8 e Load page 5 The second text body sheet printed in the printer face up and print page 6 e Load page 3 The first text body sheet printed in the printer face up and print page 4 Organize the pages in order so that the cover is on the bottom facing down with pages 6 and 7 facing up Reverse the pile and staple together in the center REV 080801 Changed factory default sample time to 15 seconds from 10 seconds Added line filter needle assy to page 1 Changed O2 sensor P N to 001140 Changed title to simplify product REV 090129 Cleaned up printing instructions to put them in the right order for individual page backside printing Re arranged spare parts and accessory items to change septa to 1 16 thick version include 001370 sample line and 001364 high capacity battery pack option and change charger to 001363 Added comment that the analyzer is intended to operate off its internal battery pack and that a high capacity battery was available REV 090821 Changed Power supply accessory to 001400 Changed title of 001246 Added line in specifications table to note 16 hour battery option Removed Wash Down Note due to so
6. air to confirm that it measures both the ambient air and the calibration gas with stability 8 5 2 To charge the battery simply plug the yellow cable from wall mounted 12V supply charger into the Power Connector at the upper left corner of the analyzer front panel Plug tke wall supply into any 110 VAC power NOTE The 12V supply provides two functions operating the analyzer from a line AC power source and charging the battery When the power switch is in the ON position the analyzer will draw the power it needs from the 12V supply and any power remaining charges the battery When the analyzer is powered OFF all of the wall supply power charges the battery With the analyzer off the battery is 80 charged in approximately 4 hours with a full charge requiring 6 hours You may leave the battery charger in the wall outlet and connected to the analyzer and continue to charge the battery while using the analyzer NOTE DANGER OF FIRE OR EXPLOSION Only use the Bridge Analyzers Inc wall power pack provided with the Model 900141 analyzer to charge the battery pack If a different wall charger is used the battery pack may overheat resulting in a danger of fire or explosion Daily Inspect the sample filter and needle Your Model 900141 Gas Analyzer comes with a 25 mm diameter sample filter which should last approximately 3 months in normal use The sample filter and the needle should be visually inspected before the a
7. ample Time External Gas Contaminate Protection Internal Gas Contaminate Protection Start up Warm up Operating Temperature Power Line Power Internal Battery Optional Extended Battery Pack Battery Charger External Supply Weight Unit Alone In Soft Case Size Unit Alone In Soft Case PC Interface Carbon Dioxide CO2 and Oxygen O2 CO2 NDIR Non Dispersive InfraRed Oxygen O2 Electro Chemical Sensor CO2 0 00 to 100 0 02 0 000 to 100 0 CO2 0 01 02 0 01 above 2 00 O2 0 001 below 2 000 4 relative to gas reading 1 relative to gas reading 8 seconds to 95 Final Reading 15 Seconds to 99 5 Final Reading 4 ml sec nominal Adjustable in 1 second intervals Factory Default 15 Seconds External water blocking and particle filter customer accessible protects pneumatics and optical components from test gas borne contamination Internal water blocking and particle filter customer inaccessible protects pneumatics and optical components from test gas borne contamination 10 Relative 30 seconds from power up Full accuracy in 5 minutes 0 to 40 degrees C 110V 220 VAC Universal Wall Supply NiMH battery pack 6 8 hours per charge Optional extended battery pack provides 16 hours continuous operation on an overnight charge 110 VAC 60 Hz 220 VAC S0 Hz Regulated E Star compliant 2 5 Lbs 3 5 Lbs 6 1 4L x 3W x 3 2H 8 1 2L x5W x 5H Fully Sealed Serial
8. bilized adjust the analyzer to read the High O2 value contained in the calibration gas using the UP and DOWN arrows NOTE After Hi O2 calibration subsequent ZERO processes will result in a value different than 20 60 being displayed after subsequent room air Zeros NOTE Low O2 Offset and High O2 Span do not interact or effect each other and one may be performed independently of the other ACCEPT the Gas Channel Calibrations use CAL ZERO To save any calibration values for later use press both the CAL and ZERO buttons on the analyzer front panel again for 1 second the same procedure used to enter CAL mode When numbers appear on all the gas displays the new calibration values have been accepted and stored for future use in the analyzer and the analyzer is returned to normal mode Once the analyzer has been calibrated the new calibration constants are retained in power safe memory even during power off conditions and are activated and used whenever the analyzer is powered up IF THINGS GO WRONG OPTIONS FOR ABORTING OR RESETTING CALIBRATION ABORT the CO2 andO2 Channel Calibrations Use the ZERO Button If for any reason you feel that the calibration procedure was not performed correctly you may press the Zero button on the analyzer front panel to abort the calibration procedure before changes are saved This allows you to ABORT the current calibration process if you want to reset previous calibrations see
9. cket in front panel analyzers inc Bridge Analyzers 1805 B Clement Avenue Alameda CA 94501 USA 510 337 1605 voice 510 337 0388 FAX www bridgeanalyzers com P N 001283 Rev 090821 Mode 9001741 User Manual d BRIDGE analyzers inc Technical Support Warranty and Service Technical Support Unlimited no charge technical support is available from Bridge Analyzers Inc at the telephone numbers and email address below Please feel free to make use of this service in case you have any questions Warranty Service Policy Your BRIDGE GAS ANALYZER is covered by a one year parts workmanship warranty In order to return the analyzer for service please contact Bridge Analyzers for detailed instructions at Bridge Analyzers Inc 1805 B Clement Ave Bldg 28 Alameda CA 94501 USA 510 337 1605 voice 510 337 0388 FAX email support bridgeanalyzers com For product returned under warranty the customer pays freight charges for product return Bridge will repair or replace the analyzer and return the analyzer freight prepaid to the customer by the same shipping class used to ship the analyzer to Bridge There will be no charge for parts labor or outbound freight for products serviced under warranty Non Warranty Service Policy For product no longer under warranty or for malfunctions not warrantable the customer pays freight charges for product return Bridge will evaluate the u
10. isplay will flash dashes and the Zero will begin The CO2 Zero O2 Cal is fully automatic and takes approximately 30 seconds to complete At the end of the Zero sequence the analyzer will show gas readings and the pump will remain on the analyzer is in continuous sample and measure mode The operator may note the CO2 display counting down to 0 00 as an indication of the readiness of the analyzer Periodic Zero Requests 5 minutes and 30 minute intervals At 5 minutes from power on and every 30 minutes thereafter the analyzer will alternate gas readings with dashes This is the Zero Request indication The operator should Zero the analyzer when convenient after the dashes appear The analyzer continues to measure gas during the Zero Requests Oxygen Sensor Test The electro chemical Oxygen sensor gradually degrades loses sensitivity over 12 18 months As part of the analyzer Zero routine the analyzer measures the oxygen detector voltage output when calibrating it with ambient air If a low output cell is detected the analyzer will display Lo on the oxygen display as an indication that the O2 sensor should be replaced soon This warning may be cleared by pushing any button on the front panel Model 900141 CO2 O2 Map Headspace Gas Analyzer Specifications Gases Measured Measurement Method Measuring Ranges Resolution Accuracy Repeatability Response time Sampling Rate S
11. ith MLL Fittings N A Accessories Options P N Part Name Life 000888 Needle Cleaning Tool N A 001246 MAP Analyzer Gas Test Stand N A 001400 110 220 VAC Wall Charger Supply N A 001364 High Capacity Battery Pack Option N A 10 4 2 The voltage test limit is set high enough to ensure that the analyzer will typically remain functional for 15 to 30 days after low oxygen detector sensitivity is detected but the operator should make plans to replace the oxygen detector at some time during this period The oxygen sensor can be easily replaced in the field New oxygen detectors may be ordered directly from Bridge Analyzers Taking Gas Measurements Continuous Measurement Mode After the first Zero from power on the analyzer will automatically be in continuous measure mode The pump will be on and gas sample will be continuously drawn through the sample line measured and displayed on the analyzer The analyzer will remain in this mode until the operator selects Sample Measurement Mode below NOTE While the continuous sampling mode is the initial measuring mode for the MAP analyzer and is normally used to verify analyzer function and for calibration where continuous gas flow is valuable the typical operational use of the analyzer is to measure consecutive product packages for headspace gas content and the sample and hold test method is more appropriate for this purpose Initiating the Sample and Hold Measurement Mode
12. n gas usually does not contain any O2 The analyzer Low O2 calibration is used to set the zero reading of the O2 channel and is actually an O2 offset adjustment and is only active with O2 levels less than 2 000 When the indicated O2 is less than 2 000 the Low O2 calibration mode may be used to adjust the O2 reading to agree with the level of O2 in the calibration gas which is generally 0 000 This can be done on a gas mix containing CO2 but the presence of CO2 will not have an effect on low O2 measurement so N2 or some other oxygen free gas mix may be used If a small amount of O2 is specified to be present in the calibration gas the Low O2 calibration mode can be used to adjust the analyzer to correctly read the trace O2 value Conversely because the O2 sensor is automatically span calibrated to ambient air every time an analyzer Zero is performed Hi Span calibration of the O2 sensor is not generally necessary unless the customer wishes to have High O2 substantially above 20 60 process gas read correctly The High O2 calibration mode is activated only at O2 levels above 2 000 When the non active O2 calibration mode is selected during calibration it flashes indicating that it is not usable NOTE Hi Calibrating the O2 channel will cause the analyzer to read other than 20 60 when it is Zeroed on ambient air 6 1 Customer Span Cal and Cal Verification Procedure Power up the analyzer and pe
13. nalyzer is used Excessive filter discoloration or indication of restricted flow either through the filter or needle is an indication that the sample filter should be replaced or the needle serviced or replaced NOTE Use only genuine Bridge filters and needles as others may seriously degrade analyzer performance and risk contamination of the analyzer 5 5 4 5 3 1 As Required Verify Accuracy with Cal Gas The Bridge CO2 CO2 gas analyzer does not need frequent gas calibration for the analyzer to operate within the accuracy specification as the gas accuracy are generally maintained by the Zero procedure as described above However particularly if there are any questions about the accuracy of the analyzer during use the user may verify and readjust the analyzer accuracy with calibration gas containing known quantities of CO2 and O2 at any time NOTE This procedure should be performed only by trained and qualified technical personnel as it will permanently change the analyzer calibration Calibration Overview CO2 and Hi Lo O2 The calibration function is generally used to set the CO2 and Low O2 accuracy of the analyzer to a certified calibration gas mixture containing gas values similar to those used in the packaging process CO2 is span calibrated to read the calibration gas accurately while O2 is offset calibrated Also if the process is intended to replace ambient air with a process gas containing no O2 the calibratio
14. ng to creep down to zero when moving from room air or the equivalent be patient to provide enough time for the system to reach a stable zero gas level Use the DOWN or UP arrows to make the reading as close to 0 000 as possible NOTE The analyzer is capable of displaying negative gas values to assist the operator in making Low O2 trim adjustments High O2 Mode Adjusting the High O2 Span on Calibration Gas containing High levels of O2 The O2 channel is automatically calibrated on ambient O2 during the Zero process The factory default value shown at the end of the Zero process is 20 60 O2 to accurately correct for the anticipated 1 5 water vapor dilution in ambient air High values of O2 are used for some headspace gases and the analyzer should be recalibrated at high O2 gas levels to increase the high O2 accuracy in this application NOTE This High O2 calibration function is automatically disabled at O2 readings below 2 000 indicated by flashing the High O2 indication When calibration gas containing greater than 2 00 O2 is delivered to the analyzer the High O2 CAL function is enabled and the O2 reading can be adjusted to the correct value by using the UP and DOWN arrows NOTE The high O2 reading is stabilized by occasionally Zero ing the analyzer on room air especially when first going to high O2 levels from very low levels of O2 Allow the analyzer to stabilize at the new O2 reading then Zero the analyzer on room
15. nit and respond to the shipper with a service estimate for repairs recommended upgrades and additionally requested parts Once the service estimate is reviewed and the work authorized by the customer Bridge will proceed with the service authorized Customers who do not have a 30 day terms account established with Bridge may choose to have service charges paid by COD or Major Credit Card or establish a 30 day terms account which requires credit verification and takes approximately 10 working days to approve Bridge will return the analyzer to the customer by the same shipping method used to return the analyzer to Bridge unless advised of a different shipping class by the customer Shipments from Bridge Analyzers are by UPS Ground Service normally but other UPS shipping methods may be selected by the customer The customer will be charged for service and upgrade parts labor and outbound freight for products not under warranty 12 CO2 O2 MAP Gas Analyzer Model 900141 Operating Instructions 1 Initial set up Li Install the 25 MM Dia white sample filter and sample needle to one end of the black sample line Install the other end to the gas input port Luer Lock fitting on the intake manifold at the upper right side of the analyzer case indicated as INPUT 2 Power On Self Test Depress the white Power Switch located on the top right side of the analyzer to turn power on to the unit This is a 2 Action push button
16. ode To put the analyzer in this mode press the DOWN button instead of the UP button when the display is frozen Then pushing the DOWN and UP buttons together will cause the analyzer to display SEC in the upper CO2 display and the current sample time in seconds in the lower O2 display The sample time may be then increased or decreased in one second increments by using the UP and DOWN arrows Once the desired sample time is displayed press the 7 CAL button and the new value will be stored and used The new sample time value is saved in power safe memory so it will be used subsequently even if the analyzer is powered down 71 Periodic Maintenance Daily The battery provided with the analyzer provides 8 hours of standalone operation under fully charged conditions Battery life is maximized if it is kept fully charged by connecting the wall charger into the unit when it is not being used Recharge the Battery The analyzer is intended to be operated in the industrial environment in cordless mode without the charger plugged in and powered from its own internal battery pack A high capacity battery pack option P N 001364 is available from Bridge Analyzers to provide 16 hours cordless operation for this gas analyzer This is a factory installed option and installation requres return of the analyzer to Bridge Contact Bridge Analzyers Inc for pricing and availability information Once the readings have sta
17. rform scheduled zeros Allow it to operate normally with the sample pump continuously on for at least 30 minutes Note The gas analyzer should be in the continuous measurement mode during the calibration process so that gas measurement stability may be confirmed Use calibration gas that can be regulated to produce a low flow sample around 400 ml min that can be drawn by the gas analyzer at ambient pressure Do not directly connect the calibration gas tank to the analyzer Instead allow the analyzer to draw a sample of the gas through a Tee fitting or the equivalent and through a clean needle and filter The analyzer draws a sample at approximately 4 ml sec flow so very little gas flow is required to ensure correct delivery of calibration gas NOTE Bridge Analyzers Inc has a MAP Analyzer calibration stand accessory P N 001246 which has been especially designed to provide the correct flow rate and pressure and interfaces between the certified calibration gas regulator and the gas analyzer to ease the calibration process Refer to Bridge P N 001246 for more information Verify that the calibration gas measurement is stable before attempting a gas calibration Check for leaks sample dilution etc The service technician should confirm that calibration gas is being delivered to the analyzer correctly before attempting to re calibrate the analyzer ENTER Calibration Mode use CAL ZERO Press both the CAL and
18. the RESET Calibration procedure When CAL disappears from the display the calibration has been aborted and the analyzer has been returned to normal operating mode Any changes to calibration settings made during an aborted calibration mode will not be saved The analyzer will use the previous calibration constants for present and future operation 6 2 6 3 6 4 6 5 Pressing CAL Selects the Gas to be Calibrated The analyzer automatically selects CO2 as the gas to be calibrated when you enter CAL mode Pressing the CAL button attain selects High O2 then Low O2 and then back to CO2 and the sequence repeats Calibrate the CO2 Gas Readings Using of the UP and DOWN arrows correct the displayed value to equal the calibration gas cylinder tag value The next gas can be selected by simply pushing the CAL button when the previous gas is calibrated Low O2 Mode Adjusting the O2 Offset using calibration gas containing low or no O2 Often the low oxygen reading is the most important to the customer as it is used to test for residual air in the package headspace The oxygen gas channel is very stable close to zero but over time it may show a small positive or negative measurement error The Low O2 Calibration feature allows the user to fine adjust the analyzer close to 0 000 O2 Use the CAL button Select LO O2 CAL for oxygen and allow the display readings to stabilize It may take a few minutes for the O2 readi