SP300 “Just Infusion” Programmable Syringe Pump
Contents
1. 6 4 Changing the Pumping Rate While Pumping The pumping rate can be changed while the pump is operating To change the pumping rate display the current pumping rate by pressing the Rate key if needed With the pumping rate displayed press the up arrow keys to change the rate The rate units will blink while the rate is being changed Rate units can not be changed while pumping The new rate is stored after a 2 second delay or by pressing a non arrow key If the new rate is within the operating range of the pump the display will blink and the new rate will be stored in memory and the pump will begin pumping at the new rate If the new rate is out of the operating range of the pump the display will show Pressing any key clears the message and restores the previous rate 6 5 Volume Dispensed The accumulated Volume Dispensed can be displayed at any time by pressing the Volume key The display will show the total accumulated volume pumped with the ml or ul LED lit and the Dispensed LED lit Volume is computed based upon the syringe inside diameter setting The Volume Dispensed accumulation is reset to 0 when A With the pump stopped pressing and holding any up arrow key while displaying the volume B The syringe diameter is changed C The accumulated Volume Dispensed rolls over from 9999 to 0 D The pump is powered on Publication 1200 01 8 03 23 10 NEXT gt ADVANCE www nextadvance com
2. 6 5 11 SPECIAL POWER UP FUNCTIONS eee ee eese eene eese sese eese eese eese ees sese sese ee eese eee eese sese sese 7 5 11 1 FIRMWARE VERSION DISPLAY eeeeeeeeer eene here rer eter eterne een 7 5 11 2 RESET PUMPING PARAMETERS c ccccsssssssssssssssssssssscssssssssssesssesessessesesssesseeseessesees 7 5 12 ERROR AND STATUS MESSAGES cccssssssssssscscssssssscscscssccscscscssssssssscssscscscscssscssssscess 7 6 ORERA TION ME mmMe 4M Mmete 7 6 1 SYRINGE INSIDE DIAMETER 6 osse ense espe Pe Ge es Se Po c s po oO Ue pa pe Do UC ED US DO Oe s ne oO Pe US e eO EC ES Ve DO PCEUER 7 6 1 1 VOLUME UNITS DEFAULT AND HOW TO CHANGE eee eene eren nns 7 6 2 OPERATING THE PUMP sccsssissscccessssecvcsssscecesstceseseoceseseoassvssesessoecsssdecascees cdsesssseseseces cess Rau eos 8 63 du telnfci c G M 8 6 4 CHANGING THE PUMPING RATE WHILE PUMPING ee eeeeeese sense etes sensns see tees een 8 6 5 VOLUME DISPENSED ee ies o cono ui ke oen ue eo oou Vi Fe Le en ado o uo ua e ee neg ooo Ru V 0e oen ndo Co vo Va FE Vena noe PEE 8 7 SETUP CONFIGURATION SET POWER FAILURE MODE 9 8 REAR OF PUM 2 0 c 0 25 00G 0500GGc eB Ge cate E E T 9 9 PY oL St 04 9 91 ACCESSORIES iere resietor sees vv uo edes
3. The last decimal point position to the right of the right most digit is not displayed indicating whole numbers with no decimal point To change the position of the decimal point use the left most up arrow key decimal point key AN Press and hold this key for at least 1 second and wait until the left most digit scrolls from 9 to 0 While continuing to hold this key the decimal point will shift 1 position to the right After the right most decimal point position the decimal point will shift to the first decimal point position Release the key when the decimal point is in the required position 5 5 Diameter and Setup Key The Diameter key allows the syringe inside diameter to be viewed and set While displaying the diameter the mm LED is lit With the Pump stopped momentarily pressing this key will display the current diameter setting Pressing the arrow keys will change the current diameter see sec 5 4 Arrow and Decimal Point Key The mm LED will blink while the diameter is being changed If the Diameter key is pressed and held Setup mode will be entered see sec 5 10 Setup When the Pump is operating pressing this key will display the current syringe diameter for review When the key is released the display returns to its previous display 5 6 Rate Key The Rate key allows the pumping rate to be viewed or changed while pumping or stopped While the pump is
4. Next Advance Inc before returning a unit for repair Warranty covered repairs will not be performed without a return authorization number At the option of Next Advance Inc a defective unit will be either repaired or replaced This warranty does not cover damage by any cause including but not limited to any malfunction defect or failure caused by or resulting from unauthorized service or parts improper maintenance operation contrary to furnished instructions shipping or transit accidents modifications or repair by the user harsh environments misuse neglect abuse accident incorrect line voltage fire flood other natural disasters or normal wear and tear Changes or modifications not approved by Next Advance Inc could void the warranty The foregoing is in lieu of all other expressed warranties and Next Advance Inc does not assume or authorize any party to assume for it any other obligation or liability 2 Overview The SP300 is a general purpose single syringe pump capable of infusing at digitally set pumping rates It is controlled from a microcontroller based system which drives a step motor allowing a wide range of pumping rates configured to the inside diameter of the loaded syringe The syringe is driven from a drive screw and drive nut mechanism Features Infusion pumping of syringes up to 140 cc Non volatile memory of all operating ters Pumping rates from 0 73 ul hr with a 1 cc ae to 1500 ml hr with a
5. SP300 Programmable Syringe Pum 7 Setup Configuration Set Power Failure Mode To change or view the setup configuration the Pump must be stopped Press and hold the Diameter Setup key until the parameter PF is displayed After 2 seconds or when any non arrow key is pressed the next parameter will be displayed See sec 5 10 Setup Key Pressing an arrow key under a value will change the parameter The following will be displayed Power Failure Mode Where n is the current setting 0 Disabled 1 Enabled When enabled if the Pump was pumping when power to the pump was disrupted the Pump will automatically start pumping when power is reconnected to the pump The accumulated volume will be reset to 0 Pressing any key on the keypad while powering up the pump will stop the Pump from starting in Power Failure Mode e Low Noise Mode Where n is the current setting 0 Disabled 1 Enabled A side affect of the SP300 s high precision micro stepped motor driver is a high frequency resonance sound at very low pumping speeds This mode minimizes this sound by reducing the micro stepping increasing pulsations 8 Rear of Pump VOLTS Hz 1 VDC AMPERAGE SERIAL NO 201001 MADE IN U S A C A m Figure 2 Rear of Pump 9 Appendix 9 1 Accessories 9 1 1 Syringe
6. eevv saos voa ooe sb us Iu ee voa ple ees oo ENS UP sanos issiro oses sopies sasoien bae Ue dE Ue 9 9 1 1 SYRINGE HEATER iss T 9 9 1 2 FIRMWARE AND HARDWARE UPGRADE eee eme eene ene ener eres esses nsns esee 9 9 2 TROUBLESHOOTING AND MAINTENANCE eeee esee ene teet estie etes senses een anna 10 9 3 SPECIFICATIONS e ereces ceceseeeesuoctes eee eve uuu ek o FVa ee e NU URUNEn nb ea ee eva SUN o Eu nE Vl e Ue Ud pun EK EVE e E EEUU RU 10 9 4 CUSTOM APPLICATIONS coc ec sese oso oes vo sa sess bp oan rene sceccessassesbanncec SEES USE USE RES URN SURE PEU SRo RUE OUE 10 9 5 SYRINGE DIAMETERS AND RATE LIMITS ee ee eee eee eene teet eese neenon as 11 ii 03 23 10 NEXT gt ADVANCE www nextadvance com SP300 Programmable Syringe Pump 1 General Information and Packing List Thank you for purchasing the Model SP300 Just Infusion TM Syringe Pump With the SP300 syringe pump you will be able digitally set an infusion rate for steady pumping Please familiarize yourself with the SP300 s operation by reading this user s manual For future reference record the serial number located on the rear of the pump and the date of purchase Next Advance Inc can be contacted at Phone USA 800 738 1681 International 518 674 3510 FAX 518 674 0189 Email info nextadvance com www nextadvance com This Operating Manual and the SP300 s hardware electronics and firmwa
7. gt ADVANCE www nextadvance com SP300 Programmable Syringe Pum mm ul i EE Om Omin i Millilit ipse er ub OG d m A pr wei Ww diac e UT Withdraw Pumping Direction ml min Lit Withdraw Purge Pumping ul min Ou A Jhr 2 Displayed volume is dispensed volume Lit Motor is operating Om min Not lit The Pump is stopped um Ou hr_ 5 4 Arrow and Decimal Point Keys Each of the four digits in the display is associated with the up arrow key directly below it When applicable the arrow key is used to increment the value of that digit or advance to the next setting Each press of an up arrow key will increase the digit by 1 up to 9 and then back to 0 The arrow keys may also be held down for continuous incrementing of numbers Some parameters have a fixed range of values such as some setup parameters that are either turned on or off In these cases the arrow key will only scroll up to the maximum value for that parameter then back to the minimum value When changing a value s units each press of any arrow key will change the units LEDs to the next units selection When the display blinks the new value is stored and takes affect This will occur when a non arrow key is pressed or after a 2 second delay since the last key press Decimal Point Key There are 4 decimal point positions on the LCD display Each decimal point position is to the right of a digit in the display
8. stopped momentary presses of this key will switch between the Rate display and the select rate units mode To change the pumping rate displayed while pumping or stopped use the up arrow keys see sec 5 4 Arrow and Decimal Point Key The rate units will blink while the rate is being changed The new pumping rate is entered and takes affect immediately when the display blinks after a 2 second delay or Publication 1200 01 5 03 23 10 NEXT gt ADVANCE www nextadvance com SP300 Programmable Syringe Pump when a non arrow key is pressed The new pumping rate is stored in memory See section 9 5 Syringe Diameters and Rate Limits for a list of minimum and maximum pumping rates A pumping rate of 0 0 will stop the pump When the pumping rate is changed if it is out of range of the pumping rate limits the display will show Pressing any key clears the message and returns to the previous pumping rate 5 6 1 Pumping Rate Units The pumping rate units can only be changed when the Pump is stopped A momentary press of the Rate key will enter Rate Units Change mode The 2 LEDs representing the units will blink and the display will show int Each press of any up arrow key selects the next rate units as indicated by the blinking units LEDs When the required rate units are blinking press any non arrow key or wait 2 seconds The display will blink indicating the rate units are stored in memory 5 7 Volume Key Pressing t
9. 60 cc Power Failure Mode Re starts pumping i after a power interruption Display of infused volume p p 2 1 Glossary of Terminology and Concepts Omm QOUispaensed Oml Omn Du Dhr OWifidrow D Pumping m C E aa Prabos Funston When a device has as many features as the NE 1000 family understanding its operation could be a daunting task at first By understanding the key concepts and terminology used in this manual the 2 03 23 10 NEXT gt ADVANCE www nextadvance com SP300 Programmable Syringe Pump operation of the SP300 will become quite intuitive Every effort has been made to design the SP300 with a consistent and intuitive user interface To facilitate and enhance your understanding of the SP300 s operation please take the time to familiarize yourself with the basic concepts below Parts of the Pump 1 Pusher Block 5 Drive Screw 8 Syringe Clamp 2 End Plate 6 Syringe Holder Block 9 Keypad User Interface 3 Power On Off Switch 7 V Slot on Syringe 10 Guide Rod 2 guide Ay Dive Nat Button Holder Block rods Parts of a Syringe Plunger Flange Plunger Barrel Flange Terminology Momentary Press A quick press less then 1 second then release of a key on the keypad Display Blink A momentary blanking of the LCD display This indicates that the new data entered by the user is valid and has taken affect 3 Setup Place the pump on a stable surface Plug the round connector end o
10. Heater Part SYRINGE HEATER see www SyringeHeater com for details Flexible heating pad that wraps around the syringe Digital PID controller will heat a syringe to a set temperature up to 100 C 9 1 2 Firmware and Hardware Upgrade Contact your dealer about upgrading your pump with other features from the NE 1000 Programmable Syringe Pump family See www SyringePump com Publication 1200 01 9 03 23 10 NEXT gt gt ADVANCE www nextadvance com 9 2 Troubleshooting and Maintenance Maintenance Periodically a small amount of all purpose oil should be applied to the guide rods The mechanism should be kept clean to prevent impeded operation No other special maintenance or calibrations are needed Pusher block makes a snap or click sound when the pump is started This is a normal condition When the pusher block is manually moved the drive nut may not have been fully engaged on the drive screw The sound heard is the drive nut engaging on the drive screw 9 3 Specifications SP300 Programmable Syringe Pump Syringe sizes Number of syringes Motor type Motor steps per revolution Microstepping Advance per step Motor to drive screw ratio Drive screw pitch DC connector Voltage at DC connector Amperage Power supply type Power supply output rating Dimensions Weight Maximum speed Minimum speed Maximum pumping rate Minimum pumping rate Maximum force Syringe inside diameter range 9 4 Cust
11. NEXT gt ADVANCE www nextadvance com SP300 Just Infusion Programmable Syringe Pump WARNING NOT FOR CLINICAL USE ON HUMANS Doc 1200 02 Rev 11 V3 7 03 23 10 NEXT gt ADVANCE www nextadvance com SP300 Programmable Syringe Pump Quick Start Instructions e Plugin the pump e Press the power switch to turn on power e Press any key to stop the display from blinking Setup Pumping Parameters To Change Numbers e Use the arrow keys to increment individual digits e To move the decimal point Press and hold the left most arrow key for at least second When the digit increments from 9 to 0 the decimal point will begin to shift Release the key when the decimal point is correct e Press any non arrow key or wait 2 seconds to enter the new setting The display will blink when a new value is entered and stored in memory Set the Syringe Inside Diameter e Momentarily press the Diameter key Set the inside diameter of the syringe in millimeters mm Set the Pumping Rate e Momentarily press the Rate key e Tochange the pumping rate units Momentarily press the Rate key again The display will show Press any arrow key to select the next available rate units Press any non arrow key or wait 2 seconds to set the rate units e Setthe pumping rate If the pumping rate is out of range the display will show rene the Syringe Press in the white drive nut button to move the pu
12. arrel over the syringe holder and the syringe plunger towards the middle of the pump Place the barrel on the syringe holder in the V slot with the barrel flange to the left of the syringe holder block 4 Liftthe syringe holder to slightly above the height of the syringe barrel and turn the syringe holder 1 4 turn back to its original position and then lower it onto the syringe barrel 3 03 23 10 NEXT gt ADVANCE www nextadvance com SP300 Programmable Syringe Pump 5 Then press the white drive nut button and slide the pusher block against the syringe plunger Release the white drive nut button To unload the syringe reverse the instructions for syringe loading 5 User Interface Units Volume Dispensed Reverse Purge Motor Indicators Indicator Indi i ndicator Operating i A O mm Dispensed EUELOMELEE EET 3 ml i d Lit lel IL laii Oo i O Withdraw O Pumping e e e CREE T r 44 D 4 t Rate Volume Diameter p zl H Setup Unused Functions Reverse Purge Key Figure 1 Front Panel 5 1 Entering Values When applicable values can be changed by displaying the current value then using the arrow keys The new value will be stored in the pump s non volatile memory meaning that the new value will not be lost the next time that power is applied to the pump A displayed value can be changed by pressing the arrow keys below each digit If the value to be chang
13. d the Dispensed LED will be lit Pressing any up arrow key will change the display to and the current volume units will blink Then press any up arrow key to switch the volume units between ml and ul Press any non arrow key or wait 2 seconds to enter the new volume units The display will blink when entered The selected volume units will remain in affect and override the default volume units Changing the diameter will no longer change the volume units Performming a system reset will cancel the override and allow the volume units to change to the default volume units when setting the syringe diameter 6 2 Operating the Pump When the Start Stop key is pressed the Pump will begin pumping and the Pumping LED will be lit The pump will pump continuously until stopped Either the pumping rate Volume Dispensed or syringe diameter can be displayed while pumping Press the relevant key to change the display While displaying the pumping rate press the up arrow keys to change the pumping rate 6 3 Purging To purge the syringe with the Pump stopped press and hold the Start Stop key The Pump will start pumping then after one second purge will begin The pump will pump at its top speed Purging will continue until the Start Stop key is released and then the pump will stop While purging the display will show Fir E Press and hold the direction button to perfrom a reverse purge The display will show
14. ed is not currently displayed when applicable press the key associated with the required value The display will show the setting s current value and its units if any While the current value is being changed the units LEDs associated with the value if any will blink Except where noted the new value is stored and or the selected operation takes affect when either 1 A non arrow key is pressed or 2 After a 2 second delay since the last arrow key was pressed If the new value is valid and different from the original value the display will blink indicating that the new value was stored Otherwise if the value was invalid an error message will be displayed Pressing any key clears the error message and restores the original value In general if a parameter has 2 values off and on they are represented by the numbers 0 and 1 respectfully 5 2 LCD Display The display consists of a 4 digit reflective LCD display This is the general purpose user display device for displaying floating point values functions and parameters The colon is used for separating function abbreviations from their parameter values 5 3 LEDs To the right of the LCD are 8 red round LED indicators The first 2 columns display the units of the displayed values Units are expressed using 1 or 2 LEDs For instance ml hr is expressed by lighting the ml and the hr LEDs Publication 1200 01 4 03 23 10 NEXT
15. f the supplied power supply adapter into the power plug located on the lower right of the pump s rear See section 8 Rear of Pump for a diagram of the rear of the pump Plug the other end of the power supply adapter into an appropriate electrical outlet The pump will be powered when the bottom of the power switch located on the upper right of the rear of the pump labeled 1 is pressed The red indicator on the switch is visible when the power switch is in the on position After power is applied to the pump the pump s display will flash Next the Pumping Parameters can be entered Before the pump can be started the pump needs the measurement of the inside diameter in millimeters of the syringe that will be loaded The syringe diameter can be entered using the keypad on the front panel of the pump Finally the syringe can be loaded and the pump started 4 Loading Syringes The syringe is loaded by securing the barrel and the pusher flange as follows 1 Press in fully the white drive nut button on the pusher block releasing the block Taking care not to drag the drive nut on the drive screw slide the block away from the syringe holder providing sufficient space for the loaded syringe Then release the white button 2 Lift the syringe holder above the syringe holder block Turn it 1 4 turn and then lower it onto the syringe holder block The syringe holder should be out of the V slot 3 Load the syringe with the b
16. his key will display the Volume Dispensed as indicted by the Dispensed LED The units of the volume are set according to the syringe diameter but can be changed 5 7 1 Clearing Volume Dispensed With the pump stopped and displaying the Volume Dispensed pressing and holding any up arrow key for one second will reset the dispensed volume to 0 Immediately after entering a new syringe diameter and before starting the pump pressing any up arrow key while displaying the Volume Dispensed will enter the Set Volume Units mode 5 8 Reverse Purge Key The direction key is used to un jam the pusher block after an over infusion making it difficult to release the nut block Press and hold this button for at least 1 second to reverse purge the pump until the pusher block is released Release this key to stop the pump While this key is held the display will show E The Withdraw and Pumping LED s will be lit during the reverse purge 5 9 Start Stop Key The Start Stop key starts or stops the Pump s operation Pressing this key switches between Pumping and the Pump stopped The Pumping LED will indicate that the Pump is pumping Pressing and holding this key while starting the Pump will start the purge mode Purge will begin after the key is held for one second and continue until the key is released The pump stops after the key is released 5 10 Setup Key The secondary func
17. lay or similar Pressing any key will clear the 5 11 2 Reset Pumping Parameters To clear out the current Pump setups press the right most up arrow key 4S while turning on power to the pump The display will show EE Pressing any key will clear the display With pumps having as many complex features as the NE 1000 family it is easy for a novice user experimenting with the pump s setup to get the pump into a weird state Performing this reset function will bring the pump out of a weird state 5 12 Error and Status Messages Value entered is Out Of Range of the pump s operational limits Check the pumping rate and or the syringe inside diameter setting is correct Key pressed is not currently applicable Indicates pumping rate or volume units change mode The units LED s will also be blinking Press any up arrow key to change the units Indicates that the pump is purging Displayed while holding down the Start Stop key Displayed while holding down the direction key and performing a reverse purge Before the pump can be operated the pumping data must be setup The syringe inside diameter and a non zero pumping rate needs to be set The operation of the pump can then be started by pressing the Start Stop key 6 1 Syringe Inside Diameter The syringe inside diameter can only be set while the Pump is stopped Use the up arrow keys to set the diameter value While the diameter value is being set the m
18. m LED will blink The new diameter value is stored after pressing any non arrow key or after a 2 second delay Valid syringe diameters are from 0 1 mm to 50 0 mm If the diameter is out of this range the display will show oor Pressing any key restores the diameter display to its previous value Changing the syringe diameter will not zero the pumping rate Section 9 5 Syringe Diameters and Rate Limits is a representative list for reference of syringe diameters for various syringe manufacturers and syringe sizes 6 1 1 Volume Units Default and How to Change The units of the accumulated infusion volume is set according to the syringe diameter setting If the default volume units are changed see next section the selected volume units will remain in affect until a reset function is performed From 0 1 to 14 0 mm Syringes smaller than 10 ml Volume units are pl From 14 01 to 50 0 mm Syringes greater than or equal to 10 ml Volume units are ml Publication 1200 01 7 03 23 10 NEXT gt ADVANCE www nextadvance com SP300 Programmable Syringe Pump Changing Volume Units The volume units used for the accumulated volume can be changed to either ml or ul NOTE Volume units can only be changed immediately after setting the syringe diameter and before the pump is started After the syringe diameter is entered display the Volume Dispensed by pressing the Volume key The current volume units an
19. ncssncsecsccssssasssuvstcccsversoisensevsssssesestecuscecdaccssstecsusccesssesvessssesccasusesesseeseess 3 4 LOADING SYRINGE G cccccccosssssssscsccssscsssvssscccsssssssssnsccccsssessecsceccscsssesscsccscesesssssseees 3 5 USER INTERFACE 2 2 5 52 2 50 50 0500 2 50855055560 0 555000505080 5 G 0s0B0SGB0BepEGoDEBODUSOSSEe 4 5 1 ENTERING VALUES weccssccsccvscescecesscsssconsscscbssseasccosseccececestecconssesesssscsseossuoetecessassseobeoseccsecees 4 52 LCD DISPLAY ueeieccceveenveeu eco veo vk es vo eaae ova cuui av vu enun svescsesvecsecceevesesessuususeseveessessuessscseveses 4 P NEM LEDS qe 4 5 4 ARROW AND DECIMAL POINT KEYS eese eese sese nn nena nasa se sese se sese sese sese sese sese sese senes 5 55 DIAMETER AND SETUP REY vi scsecssssvccscesscsoressessocteeccoossseessceceescsoseseessocseesdoassssesececeeoces 5 HOMME CU Vus QDPNETERE EEE 5 5 6 1 PUMPING RATE UNITS eo 5 ao eese rte toe eccesso rte eo epeee couper tete Eee veo A 6 5 7 VOLUME KEY qr 6 5 7 1 CLEARING VOLUME DISPENSED cecececesecececececececececeeecececeeseseseceseseseeeneeeeceeeeeee 6 5 38 REVERSE PURGE KEY scicisccsssicsccccscscectiscccscstacoteccsssccessescoscccsessdscscocosccsbecscesseseasccossesssestes 6 MEE OV ud E Mi vcccctesscosstecesscsvesscossoct cscs ccesivesosccscedscossccscoctesescsosines co ecssesessssceseesesessss 6 5 10 SETUP d
20. om Applications Publication 1200 01 Mechanical amp Electrical Up to 140 cc 1 Step motor 400 1 8 to 1 2 depending on motor speed 0 2126 uM to 0 8504 uM depending on motor speed 15 28 20 revolutions 2 1 mm center positive 12V DC at full load 750 mA at full load Unregulated linear external wall adapter country and power source specific 12V DC 800 mA 8 3 4 x 5 3 4 x 4 1 2 High 22 86 cm x 14 605 cm x 11 43 cm 3 6 Ibs 1 63 kg Operational 3 7742 cm min 0 004205 cm hr 1257 ml hr with a B D 60 cc syringe 0 73 ul hr with a B D 1 cc syringe 35 Ibs at minimum speed 18 Ibs at maximum speed 0 100 to 50 00 mm For specialized and OEM applications contact your dealer or Next Advance Inc Custom modifications can be made to the mechanics or the firmware 10 03 23 10 NEXT gt ADVANCE www nextadvance com SP300 Programmable Syringe Pump 9 5 Syringe Diameters and Rate Limits Next Advance www nextadvance com Model SP300 syringe pump Maximum and minimum flow rates Inside Syringe Diameter Rate Manufacturer mm mL hr HSW Norm Ject Monoject Terumo Air Tite Inside Minimum Syringe Diameter Rate Rate uL mm ul hr u1 hr Hamilton Microliter Publication 1200 01 11 03 23 10 O m iojfo nm m o
21. re are copyrighted Copyright 1999 2009 all rights reserved Packing List Included with the SP300 Syringe Pump are the following items e One of the following external unregulated power supply adapters Input One of 120V AC 60 Hz 220V AC 50 Hz 240V AC 50 HZ Output 12V DC 800 mA or compatible regulated power supply e This Operating Manual 1 1 Warnings and Cautions CO D Read the user s manual A Use only with the supplied power supply connected to a power source as specified on No user serviceable parts are inside ihe power supply label A Disconnect power from the pump when connecting or disconnecting cables Do not push objects of any kind int the chassis openings except for appropriate Do not immerse the pump in liquid cables and connectors A Install on a stable surface Ar the pump becomes damaged do not use in Keep hands and loose clothing away from unless certified safe by a qualified the pump s moving parts technician Damage includes but is not excluded to frayed cords and deterioration The pump can automatically start when the in performance Pumping Parameters is operating or when attached to an external control device D Discharge static from control cables before A Prevent liquids from entering openings in connecting by touching the cable to ground the rear of the pump Before touching the pump discharge static by touching ground 1 2 Disclaimer and Warranty Disclaimer Next Ad
22. sher block e Liftand turn the syringe clamp away from the syringe holder block e Position the syringe on the pump with the flange to the left of the syringe holder block e Lift and turn the syringe clamp onto the syringe body e Move the pusher block next to the syringe plunger Purge Press and hold the Start Stop key for one second Release to stop Press and hold the direction button to reverse the pump and release the pusher block if jammed Start the Pump Press and release the Start Stop key to start or stop the pump When Pumping e The pumping rate can be changed PUMP RESET Press and hold the right most up arrow key while turning on power to the pump Need more pumping features Visit www nextadvance com to see our full featured syringe pumps i 03 23 10 NEXT gt ADVANCE www nextadvance com SP300 Programmable Syringe Pump Table of Contents 1 GENERAL INFORMATION AND PACKING LIST 0sssssssssssssssssssssssssssseseees 1 1 1 WARNINGS A AND CAUTIONS D 1 1 2 DISCLAIMER AND WARRANTY eeeeeee eese eese eese eese sese sese sese eese sese eese eese sese sese see sese sese sese see 1 2 OVERVIEW inssisscccescisssccscscesccstissseresocdsceccssvsntevecsscsucssessletecsscesessvovscsecs secedssessbececsssevessseseass 2 2 1 GLOSSARY OF TERMINOLOGY AND CONCEDPTS ccccccssssssssssssssscsssesssesssssssssssessseeces 2 3 SEV UDP ies svssvecssnsscscccvccc
23. tion of the Diameter key is Setup While the Pump is stopped press and hold the Diameter key until the setup configuration parameter Power Failure Mode is displayed Fre n The display will consecutively display for about 2 seconds each Setup Configuration parameter and its current setting Pressing any non arrow key will immediately advance to the next Setup Configuration parameter if any To change a Setup Configuration parameter press an arrow key under the parameter s value To store the new value press any non arrow key or wait 2 seconds If the parameter value differs from its previous value the display will blink The new parameter value will be stored and the next parameter will be displayed if any See section 7 Setup Configuration for a complete description of the Setup Configurations After the last configuration parameter is displayed the display reverts back to displaying the syringe Publication 1200 01 6 03 23 10 NEXT gt ADVANCE www nextadvance com SP300 Programmable Syringe Pump diameter Any new parameter value will take affect immediately upon being stored 5 11 Special Power Up Functions The following special functions are accessed by pressing the relevant key while turning on power to the pump 5 11 1 Firmware Version Display To display the pump s firmware version press the left most up arrow key AS while turning on power to the pump The display will show H d u disp
24. vance Inc makes no representations or warranties expressed statutory or implied regarding the fitness or merchantability of this product for any particular purpose Further Next Advance Inc is not liable for any damages including but not limited to lost profits lost savings or other incidental or consequential damages arising from ownership or use of this product or for any delay in the performance of its obligations under the warranty due to causes beyond its control Next Advance Inc also reserves the right to make any improvements or modifications to the product described in this manual at any time without notice of these changes Next Advance Inc products are not designed intended or authorized for use in applications or as system components intended to support or sustain human life as a clinical medical device for humans or for any application in which the failure of the product could create a situation where personal injury or death may occur All brand and product names used in this manual are the trademarks of their respective owners 1 03 23 10 NEXT gt ADVANCE www nextadvance com SP300 Programmable Syringe Pump Warranty Next Advance Inc warranties this product and accessories for a period of two year parts and labor from the date of purchase The repaired unit will be covered for the period of the remainder of the original warranty or 90 days whichever is greater A return authorization number must be obtained from