Home

Moisture Target Series 5

image

Contents

1. Figure C 1 Main and User1 Menu Map Menu Maps Ca August 2004 NOTE press Ec at any time to go back one level is shown as er for better readability AIN ME U SEtUP E passcode 1 CONSULT FACTORY DISPLAY number number Screen Display Current Value ow Adjust Numeric Value Select Menu Option Must be pressed within 5 seconds passcode Figure C 1 USER2 ME See Figure C 1 dEFALT OFFSet Figure C 2 User2 Menu Map Menu Maps
2. 000 5 3 1 Quick Setup Guide B 3 Powering Up 0 0 eee eee eee 3 1 Userl scH 3 er en 3 5 B 1 PPMv Calculating A 14 Ser o oret ne Do Rr ERO se 3 6 B 2 PPMw Calculating A 16 Menu Map Probe Maine ha ra RE ELI EA B 1 Cable Length 2 22200 2 5 Quick Setup Guide B 3 Calibration Data Sheet 4 6 User iiie det oh aba os ass pair B 1 Calibration Interval 4 6 USer2 4 2 2 ae ae en B 2 Cleaning rye scenes net a A 9 Moisture Probe Connecting el ko een 2 5 Gas Flow Rates 2 22 222020 A 5 Entering Calibration Curve 3 22 Liquid Flow Rates A 5 Installing in Sample System 2 2 Materials of Construction A 12 Maintenance 00 00 eee A 8 MTS 5 M Series unsre a ERU REX 1 1 Configurations 0 5 1 Specifications 005 5 3 Dimensions 0 0000 ee 5 2 Keypad iiss tenebo ae taka oa 3 2 R Mounting ne PAs aes 2 1 Powering Down 4 3 1 Range Powering UP oo 3 1 Dew Point 0 0 20 eee eee eee 1 1 Troubleshooting 0005 4 2 Recorder User Program isses eee 4 4 Adjusting Lm 3 20 Connecting usie a eee es 2 7 O High Value 0 00000 3 18 Tow Value ur ni de es 3 17 Offset Dew Point 2 22 222000 3 12 Selecting
3. Output cont t AL Press s or t until t rcd appears on the display t rcd Press ENTER r out One of these displays appears Press s or t until r out appears Then press ENTER rADJ 0 20 One of these displays appears Press s or t until the desired output range type 0 20 wie mA 4 20 mA or 0 2 V is displayed Then 0 2 press ENTER X XH One of these displays appears Press s or t until the desired test point see table 3 5 E below is displayed Then press ENTER X XL Table 3 5 Recorder Output Test Points Output Range Test Point Desired Output H 10096 20 mA 0 20 mA t 5096 10 mA L 0 OmA H 100 20 mA 4 20 mA t 5096 12mA L 2 096 4 mA H 10096 2V 0 2 V t 5096 1V L 0 40V Operation amp Programming 3 19 August 2004 Testing the Recorder Output cont Adjusting the Recorder Output The MTS 5 produces the selected test signal at the recorder output terminals Use the digital multimeter to measure the test signal Note the difference if any between the measured recorder output reading and the desired recorder output reading X XH If desired press s or t to select another test point Then press ENTER When the testing is complete press ESC X XX When the range selection screen reappears press ESC r_out To exit the User Program press ESC until run appears Then press ENTER To continue programming the MTS 5 pr
4. ENTER moves to next submenu item confirms changes s scrolls up increases numerical values t scrolls down decreases numerical values ESC exits current submenu item cancels changes Note 7o make numerical values change more rapidly hold down the s or t key MOISTURE TARGET SERIES 5 DEW FROST POINT Figure 3 1 Front Panel Keypad 3 2 Operation amp Programming August 2004 Entering the User Program Exiting the User Program To enter the User Program while the MTS 5 is taking measurements you must press certain keys in the correct order and within a certain length of time If you do not complete the following instructions exactly the MTS 5 returns to taking measurements Follow the instructions below as summarized Table 3 1 below to enter the User Program from normal measurement mode For a complete programming overview see Appendix C Menu Maps XXXC Press ESC and wait until ESC appears on the display Note Perform the following two steps within 5 seconds of the appearance of ESC on the display ESC Press ENTER within 5 seconds Then press ESC within 5 seconds ALA You are now at the Main Menu Table 3 1 Entering the User Program Press These Keys Display Reads ESC ESC ENTER within 5 seconds ESC ESC within 5 seconds AL A you are in Main Menu Program and set up specific features of the MTS 5 acco
5. To turn on the MTS 5 do the following 1 Connect the power cord to an appropriate power source 2 Press the POWER key on the front panel The MTS 5 displays a sequence of informational screens while it initializes then displays a dew frost point measurement screen To turn off the MTS 5 do the following 1 If you are in the MTS 5 User Program exit the User Program see page 3 3 and wait until the unit displays a data screen 2 Press the POWER key on the front panel 3 When the display reads P OFF press the ENTER key on the front panel The MTS 5 display goes blank and the unit enters standby mode IMPORTANT When the unit is in standby mode it still is powered on Operation amp Programming 3 1 August 2004 Programming the The MTS 5 User Program enables the following procedures Instrument setup the display alarms and recorder output test the display alarms and recorder output adjust the recorder output enter calibration data dew frost point constant offset etc Note The MTS 5 stops taking measurements while you are in the User Program See the next section for a description of the front panel keypad or go directly to page 3 3 for instructions on entering the User Program Using the Keypad The MTS 5 front panel as shown in Figure 3 1 below contains five keys Use of the POWER key was described on page 3 1 and the other four keys are used for programming the meter
6. dP HI dP LO XX dP LO Press s or t until the dP display appears Press ENTER Press ENTER Use the s and t keys to set the desired high maximum dew point value Then press ENTER Press s or t until the dP LO display appears Press ENTER Use the s and t keys to set the desired low minimum dew point value Then press ENTER To exit the User Program press ESC until run appears Then press ENTER To continue programming the MTS 5 proceed to the appropriate section for instructions 3 10 Operation amp Programming August 2004 Setting the Backlight Timer Entering the Probe Serial Number If you set the backlight timer to a value greater than zero the display s backlight comes on whenever you press a key Then the backlight stays on for the number of minutes specified in the User Program up to 1 439 minutes To keep the backlight on all the time set the backlight timer to 1 440 minutes To set the backlight timer enter the User2 menu as described on page 3 8 Then proceed as follows unit Press s or t until bLitE appears on the display bLitE Press ENTER XX Use the s and t keys to set the desired backlight timer interval Then press ENTER bLitE To exit the User Program press ESC until run appears Then press ENTER To continue programming the MTS 5
7. 000005 5 2 Watchdog Function 0 3 13 Wiring Alarms iuntp eH eta p tears 2 9 DiaSram sss 032 sort cya Saeed Pep 2 6 Proben ace hires ae hdres a ea oars 2 5 Recorde socane tee ea ta bee See 2 7 Wiring Diagram 2 222222 B 4 Index DECLARATION Sensing OF CONFORMITY We Panametrics Limited Shannon Industrial Estate Shannon County Clare Ireland declare under our sole responsibility that the Moisture Target Series 5 Hygrometer to which this declaration relates are in conformity with the following standards e EN 61326 1998 Class A Annex A Continuous Unmonitored Operation e EN 61010 1 1993 A2 1995 Overvoltage Category II Pollution Degree 2 following the provisions of the 89 336 EEC EMC Directive and the 73 23 EEC Low Voltage Directive The units listed above and any sensors and ancillary sample handling systems supplied with them do not bear CE marking for the Pressure Equipment Directive as they are supplied in accordance with Article 3 Section 3 sound engineering practices and codes of good workmanship of the Pressure Equipment Directive 97 23 EC for DN 25 Shannon June 1 2002 pou Modos Mr James Gibson GENERAL MANAGER Cc LS ENISO9002 T V ESSEN Shannon ISO 9001 U S CERT DOC H2 August 2004 DECLARATION Sensing DE CONFORMITE Nous Panametrics Limited Shannon Industrial Estate Shannon County Clare Ireland d clarons sous notre propre responsabili
8. Application of the Hygrometer A 5 August 2004 Contaminants Non Conductive Particulates Industrial gases and liquids often contain fine particulate matter Particulates of the following types are commonly found in such process fluids carbon particles salts rust particles polymerized substances organic liquid droplets dust particles molecular sieve particles alumina dust For convenience the above particulates have been divided into three broad categories Refer to the appropriate section for a discussion of their affect on the GE Infrastructure Sensing moisture probe Note Molecular sieve particles organic liquid droplets and oil droplets are typical of this category In general the performance of the moisture probe will not be seriously hindered by the condensation of non conductive non corrosive liquids However a slower response to moisture changes will probably be observed because the contaminating liquid barrier will decrease the rate of transport of the water vapor to the sensor and reduce its response time Particulate matter with a high density and or a high flow rate may cause abrasion or pitting of the sensor surface This can drastically alter the calibration of the moisture probe and in extreme cases cause moisture probe failure A stainless steel shield is supplied with the moisture probe to minimize this effect but in severe cases it is advisable to install a Teflon or sta
9. few microns of mercury up to pressures of 5000 psig The calibration data supplied with the moisture probe is directly applicable over this entire pressure range without correction Note Although the moisture probe calibration data is supplied as meter reading vs dew point it is important to remember that the moisture probe responds only to water vapor pressure When a gas is compressed the partial pressures of all the gaseous components are proportionally increased Conversely when a gas expands the partial pressures of the gaseous components are proportionally decreased Therefore increasing the pressure on a closed aqueous system will increase the vapor pressure of the water and hence increase the dew point This is not just a mathematical artifact The dew point of a gas with 1000 ppm of water at 200 psig will be considerably higher than the dew point of a gas with 1000 ppm of water at 1 atm Gaseous water vapor will actually condense to form liquid water at a higher temperature at the 200 psig pressure than at the 1 atm pressure Thus if the moisture probe is exposed to pressure changes the measured dew point will be altered by the changed vapor pressure of the water It is generally advantageous to operate the hygrometer at the highest possible pressure especially at very low moisture concentrations This minimizes wall effects and results in higher dew point readings which increases the sensitivity of the instrument T
10. proceed to the appropriate section for instructions To enter the serial number of your probe enter the User menu as described on page 3 7 Then proceed as follows S No Press ENTER x Use the s and t keys to enter the serial number of the probe Then press ENTER S_No To exit the User Program press ESC until run appears Then press ENTER To continue programming the MTS 5 proceed to the appropriate section for instructions Operation amp Programming 3 11 August 2004 Applying a Dew Point To apply a constant offset to all dew point measurements enter the Offset User2 menu as described on page 3 8 Then proceed as follows unit Press s or t until OFFSEt appears on the display OFFSEt Press ENTER XXC Use the s and t keys to set the desired offset value Then press ENTER to accept new offset value or press ESC to keep the old offset value OFFSEt To exit the User Program press ESC until run appears Then press ENTER To continue programming the MTS 5 proceed to the appropriate section for instructions 3 12 Operation amp Programming August 2004 Using the Alarms Setting the Fault Alarm The standard MTS 5 is equipped with one fault alarm AL F and two high low alarms AL A and AL B To set up and or test each of these alarms proceed to the appropriate section If enabled the fault alar
11. 0 719 19 0 854 0 838 0 822 0 806 0 791 18 0 939 0 921 0 904 0 887 0 870 17 1 031 1 012 0 993 0 975 0 956 16 1 132 1 111 1 091 1 070 1 051 15 1 241 1 219 1 196 1 175 1 153 14 1 361 1 336 1 312 1 288 1 264 13 1 490 1 464 1 437 1 411 1 386 12 1 632 1 602 1 574 1 546 1 518 11 1 785 1 753 1 722 1 691 1 661 10 1 950 1 916 1 883 1 849 1 817 9 2 131 2 093 2 057 2 021 1 985 8 2 326 2 285 2 246 2 207 2 168 7 2 531 2 493 2 450 2 408 2 367 6 2 765 2 718 2 672 2 626 2 581 5 3 013 2 962 2 912 2 862 2 813 4 3 280 3 225 3 171 3 117 3 065 3 3 568 3 509 3 451 3 393 3 336 2 3 880 3 816 3 753 3 691 3 630 1 4 217 4 147 4 079 4 012 3 946 0 4 579 4 504 4 431 4 359 4 287 A 18 Application ofthe Hygrometer August 2004 D Table A 3 Vapor Pressure of Water Continued Aqueous Vapor Pressure Over Water Temp C 0 0 0 2 0 4 0 6 0 8 0 4 579 4 647 4 715 4 785 4 855 1 4 926 4 998 5 070 5 144 5 219 2 5 294 5 370 5 447 5 525 5 605 3 5 685 5 766 5 848 5 931 6 015 4 6 101 6 187 6 274 6 363 6 453 5 6 543 6 635 6 728 6 822 6 917 6 7 013 7 111 7 209 7 309 7 411 7 7 513 7 617 7 722 7 828 7 936 8 8 045 8 155 8 267 8 380 8 494 9 8 609 8 727 8 845 8 965 9 086 10 9 209 9 333 9 458 9 585 9 714 11 9 844 9 976 10 109 10 244 10 380 12 10 518 10 658 10 799 10 941 11 085 13 11 231 11 379 11 528 11 680 11 833 14 11 987 12 144 12 302 12 462 12 624 15 12 788 12 953 13 121 13 290 13 461 16 13 634 13 809 13 987 14 166 14 347 17 14 530 14 715
12. 3 10 To enter new high and low reference values enter the User2 menu as described on page 3 8 and proceed as follows unit Press s or t until rEF appears on the display rEF Press ENTER H rEF Press ENTER Operation amp Programming August 2004 Entering High and Low Reference Values cont Accessing Factory Setup X XXKK H rEF LrEF KXRKX LrEF Press s or t to enter a new high reference value Press ENTER Press s or t until L rEF appears on the display Press ENTER Press ENTER Press s or t to enter a new low reference value Press ENTER To exit the User Program press ESC until run appears Then press ENTER To continue programming the MTS 5 proceed to the appropriate section for instructions Users do not normally need to access the Factory Setup submenu and a special passcode is required to do so Never attempt to program this submenu without first contacting GE Infrastructure Sensing for assistance Operation amp Programming 3 25 Chapter 4 Service and Maintenance Introd ctlonit sos run ee eet te ees 4 1 Common Problems cc cece ccc ce e hen 4 2 Replacing the User Program cece cece eee ee eee eee 4 4 Replacing Recalibrating Moisture Probes Lsuuue 4 6 August 2004 Introduction The Moisture Target Se
13. 30 96 30 51 97 20 98 20 99 10 100 10 101 10 52 102 09 103 10 104 10 105 10 106 20 53 107 20 108 20 109 30 110 40 111 40 54 112 51 113 60 114 70 115 80 116 90 55 118 04 119 10 120 30 121 50 122 60 56 123 80 125 00 126 20 127 40 128 60 57 129 82 131 00 132 30 133 50 134 70 58 136 08 137 30 138 50 139 90 141 20 59 142 60 143 90 145 20 146 60 148 00 60 149 38 150 70 152 10 153 50 155 00 61 156 43 157 80 159 30 160 80 162 30 62 163 77 165 20 166 80 168 30 169 80 63 171 38 172 90 174 50 176 10 177 70 64 179 31 180 90 182 50 184 20 185 80 65 187 54 189 20 190 90 192 60 194 30 66 196 09 197 80 199 50 201 30 203 10 67 204 96 206 80 208 60 210 50 212 30 68 214 17 216 00 218 00 219 90 221 80 69 223 73 225 70 227 70 229 70 231 70 70 233 70 235 70 237 70 239 70 241 80 71 243 90 246 00 248 20 250 30 252 40 72 254 60 256 80 259 00 261 20 263 40 73 265 70 268 00 270 20 272 60 274 80 74 277 20 279 40 281 80 284 20 286 60 75 289 10 291 50 294 00 296 40 298 80 76 301 40 303 80 306 40 308 90 311 40 77 314 10 316 60 319 20 322 00 324 60 78 327 30 330 00 332 80 335 60 338 20 79 341 00 343 80 346 60 349 40 352 20 80 355 10 358 00 361 00 363 80 366 80 81 369 70 372 60 375 60 378 80 381 80 82 384 90 388 00 391 20 394 40 397 40 83 400 60 403 80 407 00 410 20 413 60 A 20 Application ofthe Hygrometer August 2004 D Table A 3 Vapor Pressure of Water Continued Aqueous Vapor Pressure Over Water cont Temp C 0 0 0 2 0 4 0 6 0 8
14. 4 4 Re installing the Main PC Board en ERR tle co eod en 4 5 Checking the New User ProgEatm c een RR Hd PIEEME 4 5 Replacing Recalibrating Moisture Probes 0 cece cece eet eee eene 4 6 Chapter 5 Specifications eC OGG ne eat RU cet p culvert et Maal adver SE Oe lots canes S EE 5 1 Moisture Megsurement ee nes rn 5 3 Appendix A Application of the Hygrometer Fg dave 5 leioa Eee A 1 Moist re MOBILOEEIWES za een A 2 Pressure cette toe a tact set br E i ec ae e ioter E eus AIME A 5 RESPONSE TIER ei ln era A 3 Temperal fe cscs stood ar era khaki See iubet Oe A 4 POW ROS a RER ee A 4 COMMON UMPCS 12 os pa Get due I ER RE A 6 NOR Corduetwe Partichlates Area mer A 6 FonduetvePartieulates ists seen len een A 7 TRANCA Partieulgtese E EE E ei ie A 7 Aluminum Oxide Probe M lntenance nee nee A 8 Cleaning the Moisture Probe um rs Be A 9 Corrosive Gases And LiquidS rn ee a ee nn A 11 Materials oF C nstruction som coo ea ee las en ee A 12 Calculations and Useful Formulas in Gas Applications 0 0 0 cece ccc ene eee ee A 13 NomencdtU TT ne SE ee NE E A 13 Parts per Milion by VoOlUITie uan orb an oe Hure A 14 Parts ber MIMO t WC IQ ts ct crews ee NE Reel A 16 Relative kim ae ae os ee a E EE RO eed Shela hi A 16 Weight of Water per Unit Volume of Carrier Gas 0 6 cece ccc e eee neces A 16 Weight of Water per Unit Weight of Carrier GS 2 ccc seen A 17 Comparison or PPIMMVCOlCUIGEONS 4 0 vesnys re ern A 22 Bre
15. Passcode aaa dew neue nenn ie ac 3 8 Seting Upthe MTS geom S ecu st EM ne Me Ar VEM acs el wok 3 9 Selecting the Display DIES staat e tele aes Peden Mola a hd etr re 3 9 Setting the High and Low Dew Paints Seesen 3 10 Setting the Backight Tiner Aeee ee es res 3 11 Entering the Probe Serial Number 6k cece nent e ee 3 11 ADBIU NG Dew Point Offset re ee ee er ee 3 12 Using TO IGEITIS o ooo oti Messias eher 3 13 Setting the Fault AAR es re tego ANS 3 13 Setting the High Low ATQERIS eere a recie EN te Oe eee teehee 3 14 Testing the Alarmi Reas ertragen 3 15 Using the Recorder Outpt Teer 3 16 Setting Up the Recorder OUT OU ce acts ea 3 16 Testing and Adjusting the Recorder Output 2 0 2 cece cece eee rrna 3 18 Entering New Probe Calibration Date 25H te a 3 22 Additional Setup Procedures u u a a 3 23 Testing the Display unse ee se nes SERRE Ode aca 3 23 Restoring the Defgult SertiltiS use arire ae e n ted Ania Ne 3 24 Entering High and Low Reference Values au ri 3 24 Accessing Factor Setup rn ste nn ee irte tee E 3 25 November 2004 is ee Table of Contents cont Chapter 4 Service and Maintenance OLEO HEURE aust saris yes Oe etes LM a entices LEM a M LUE 4 1 COMMON PEODISITISS ner cel repe EE etw uta cork EE RC E E eee S 4 2 R placing the User Progra SS od karten uo dare e bv etlam 4 4 Removing the Main PC BOO cos rct tuto rp EP UE D b Ep MEE 4 4 Replacing the PROM CI ot acetate ee iter M UH teer es are
16. backlight timer page 3 11 enter the unit s serial number page 3 11 apply a constant dew point measurement offset page 3 12 Proceed to the indicated page to perform the desired task Selecting the Display Units The MTS 5 can display measurements in C F or MH sensor signal To select the display units enter the User2 menu as described on page 3 8 Then proceed as follows unit dEGC dEG F unit Press ENTER Press s or t until the desired units dEG C dEG F or H appears on the display Then press ENTER To exit the User Program press ESC until run appears Then press ENTER To continue programming the MTS 5 proceed to the appropriate section for instructions Operation amp Programming 3 9 August 2004 Setting the High and Low Set the minimum low and maximum high dew frost points Dew Points measured by your unit to suit your specific application needs To set the dew point measurement range enter the User menu as described on page 3 7 Then proceed as follows Note Find the high and low dew points on the data sheet that came with your unit Usually the MTS 5 is set at the factory for low dew point dP LO of 110 C and high dew point dP HI of 20 C You can choose high and low dew point values from 110 to 60 C in ten degree increments S_No dP dP HI XX
17. follows M ppm ppm x 77 1 2 T Relative humidity is defined as the ratio of the actual water vapor pressure to the saturation water vapor pressure at the prevailing ambient temperature expressed as a percentage P RH x 100 1 3 Ps Typical Problem Find the relative humidity in a system if the measured dew point is 0 C and the ambient temperature is 20 C Solution From Table A 3 on page A 18 the water vapor pressure at a dew point of 0 C is 4 579 mm of Hg and the saturation water vapor pressure at an ambient temperature of 20 C is 17 535 mm of Hg Therefore the relative humidity of the system is 100 x 4 579 17 535 26 1 Three units of measure are commonly used in the gas industry to express the weight of water per unit volume of carrier gas They all represent a vapor density and are derivable from the vapor pressure of water and the Perfect Gas Laws Referenced to a temperature of 60 F and a pressure of 14 7 psia the following equations may be used to calculate these units P mg of water _ ag x _W 1 4 liter of gas Tx P Ib of water _ 9 9304 x T 1 5 ft of gas R 10 Ib of water _ _ PPM ox Py 1 6 MMSCF of gas 21 1 21 1 x Py Note MMSCF is an abbreviation for a million standard cubic feet of carrier gas Application of the Hygrometer August 2004 Weight of Water per Unit Weight of Carrier Gas Occasionally the moisture content of a gas is expressed in terms of t
18. of the hygrometer dew point readings for solutions of various known water concentrations must be performed Such a calibration can be conducted in either of two ways perform a Karl Fischer analysis on several unknown test samples of different water content prepare a series of known test samples via the addition of water to a quantity of dry liquid In the latter case it is important to be sure that the solutions have reached equilibrium before proceeding with the dew point measurements Note Karl Fisher analysis is a method for measuring trace quantities of water by titrating the test sample against a special Karl Fischer reagent until a color change from yellow to brown or a change in potential indicates that the end point has been reached Either of the empirical calibration techniques described above can be conducted using an apparatus equivalent to that shown in Figure A 4 on page A 31 The apparatus pictured can be used for both the Karl Fischer titrations of unknown test samples see page A 32 and the preparation of test samples with known moisture content see page A 33 A 30 Application of the Hygrometer August 2004 M2 Probe Rubber Septum 3 4 26 THD Female soft soldered to cover Exhaust A Stainless Steel Tubing Soft Solder soft soldered to cover Metal Cover with Teflon Washer Glass Bottle Magnetic Stirrer Bar Magnetic Stirrer Figure A 4 Moisture Content Test Apparatus App
19. of the owner s approval to proceed the instrument will be repaired and returned lii November 2004 EI oe eO 9 a r1 Table of Contents Chapter 1 Features and Capabilities IEEE IUGUM EEE ae ELECT M uM Se eer UE UE 1 1 EISCHOMIES UNIT EI 1 1 fiel m asien E 1 1 Chapter 2 Installation abEOSDIC HOD oS mcer toe oa uS reci std stir mafia cn lh ES 2 1 Mounting the Electronics Unit a ee i el o Be Mele Le cmt et 2 1 Mounting the Sample System ouch nett ie he Ie t CE t ees 2 2 Installing He PRODI ooa cias bee nee ale 2 2 Making Wiring Connections nee rer 2 4 installing the Power Cable ssa cic uid u RR a ial a pee ee rg 2 4 Conmecting KREA RO o sn ee Vener D MI A hee Caen ee fd 2 5 Connecting the Recorder Outpt s yi nail es vo bw etn an len Tace So ME 2 7 GHATS CHG the Alde oases we eec dotted et E 2 9 Chapter 3 Operation amp Programming Powerttig HE TE a eei dn Meu Cate ate ea cte Me Dogg Feat Ed 3 1 POWERING DOWN fn etre to Rh ent ce ea en ae TA AA ea eee aw Dalya 3 1 Programming thelnstr menki sehe aus hint 3 2 Using the Neg D Od iioi ea a een Pere re Et ee ER 3 2 Entering the User Program 2 22 24 a ee ee Bee tet 3 3 Exiting tie WSer Proc nants ei ee el 3 3 Navigating Through the MENUS sense ka in Hain ds 3 4 The User and User MORBUS rue 3 6 EAbet ine the USOC Merl zn Deere ern tell thee tuts 3 7 Eh ndingthellserliPasseoden an neuerer 3 7 Entering the User Menu ee ee see Idea bebe 3 8 Changing the User
20. recorder output enter the User menu as described on page 3 7 Then complete the following steps in the order indicated S No Press s or t until rcrd appears on the display rcrd Press ENTER Selecting the Units r unit One of these displays appears Press s or t until r unit appears Then press Eu ENTER rcd LO rcd HI dEG C Press s or t until the desired units dEG C dEG F or H appears Then press ENTER dEG F H r unit Press t until r out appears 3 16 Operation amp Programming August 2004 Selecting the Output Type Range r_out Press ENTER Press s or t until the desired output range type 0 20 mA 4 20 mA or 0 2 V appears Then press ENTER IMPORTANT Be sure switch S1 on the main PC board is set to the correct output type current or voltage See Connecting the Recorder Output on page 2 11 r_out Press t until rcrd LO appears Setting the Output Low Value rcrd LO Press ENTER The current recorder output low value in the units specified on page 3 16 is displayed X X C Use s and t to set the desired recorder output low value Then press ENTER rcrd LO Press t until rcrd HI appears Operation amp Programming August 2004 Setting the Output High Value Testing and Adjusting the Re
21. the sample For best results the hydrocarbon liquid used above should be pre dried to a moisture content that is insignificant compared to the moisture content of the sample Note Since the addition of the solid may significantly change the saturation value for the solvent published values should not be used Instead an empirical calibration as discussed in the previous section should be used A dew point of 110 C which can correspond to a moisture content of 10 ppm or less represents the lower limit of sensor sensitivity The maximum measurable moisture content depends to a great extent on the liquid itself Generally the sensor becomes insensitive to moisture contents in excess of 1 by weight OR An alternative technique involves driving the moisture from the solids sample by heating The evaporated moisture is directed into a chamber of known volume which contains a calibrated moisture probe Convert the measured dew point of the chamber into a water vapor pressure as discussed earlier in this appendix From the known volume of the chamber and the measured vapor pressure dew point of the water the number of moles of water in the chamber can be calculated and related to the percent by weight of water in the test sample Although this technique is somewhat tedious it can be used successfully An empirical calibration of the procedure may be performed by using hydrated solids of known moisture conte
22. 14 903 15 092 15 284 18 15 477 15 673 15 871 16 071 16 272 19 16 477 16 685 16 894 17 105 17 319 20 17 535 17 753 17 974 18 197 18 422 21 18 650 18 880 19 113 19 349 19 587 22 19 827 20 070 20 316 20 565 20 815 23 21 068 21 324 21 583 21 845 22 110 24 22 377 22 648 22 922 23 198 23 476 25 23 756 24 039 24 326 24 617 24 912 26 25 209 25 509 25 812 26 117 26 426 27 26 739 27 055 27 374 27 696 28 021 28 28 349 28 680 29 015 29 354 29 697 29 30 043 30 392 30 745 31 102 31 461 30 31 824 32 191 32 561 32 934 33 312 31 33 695 34 082 34 471 34 864 35 261 32 35 663 36 068 36 477 36 891 37 308 33 37 729 38 155 38 584 39 018 39 457 34 39 898 40 344 40 796 41 251 41 710 35 42 175 42 644 43 117 43 595 44 078 36 44 563 45 054 45 549 46 050 46 556 37 47 067 47 582 48 102 48 627 49 157 38 49 692 50 231 50 774 51 323 51 879 39 52 442 53 009 53 580 54 156 54 737 40 55 324 55 910 56 510 57 110 57 720 41 58 340 58 960 59 580 60 220 60 860 Application of the Hygrometer A 19 August 2004 es Table A 3 Vapor Pressure of Water Continued Aqueous Vapor Pressure Over Water cont Temp C 0 0 0 2 0 4 0 6 0 8 42 61 500 62 140 62 800 63 460 64 120 43 64 800 65 480 66 160 66 860 67 560 44 68 260 68 970 69 690 70 410 71 140 45 71 880 72 620 73 360 74 120 74 880 46 75 650 76 430 77 210 78 000 78 800 47 79 600 80 410 81 230 82 050 82 870 48 83 710 84 560 85 420 86 280 87 140 49 88 020 88 900 89 790 90 690 91 590 50 92 51 93 50 94 40 95
23. 36 EG Die oben aufgef hrten Ger te und zugeh rige mitgelieferte Sensoren und Handhabungssysteme tragen keine CE Kennzeichnung gem der Druckger te Richtlinie da sie in bereinstimmung mit Artikel 3 Absatz 3 gute Ingenieurpraxis der Druckger te Richtlinie 97 23 EG f r DN lt 25 geliefert werden Shannon June 1 2002 pu Modos Mr James Gibson GENERALDIREKTOR NSAI LS ENISO9002 TUV ESSEN Shannon ISO 9001 U S CERT DOC H2 August 2004 USA 1100 Technology Park Drive Billerica MA 01821 4111 Web www gesensing com Ireland Shannon Industrial Estate Shannon County Clare Ireland
24. 70 ppm saturation concentration with the 30 C ambient temperature and read the Henry s Law Constant of 27 4 on the center scale c Using the straightedge connect the above K value of 27 4 with the measured dew point of 0 C and read the correct moisture content of 125 ppm where the straightedge crosses the moisture content scale Find the moisture content in heptane at an ambient temperature of 50 C if a dew point of 3 C is measured with the GE Infrastructure Sensing hygrometer a From the literature it is found that Cg for heptane at a temperature of 50 C is 480 ppm b Using a straightedge on Figure A 3 on page A 26 connect the 480 ppm saturation concentration with the 50 C ambient temperature and read the Henry s Law Constant of 5 2 on the center scale c Using the straightedge connect the above K value of 5 2 with the measured dew point of 3 C and read the correct moisture content of 29 ppm where the straightedge crosses the moisture content scale Application of the Hygrometer A 27 August 2004 Special Case cont Note Jf the saturation concentration at the desired ambient temperature cannot be found for any of these special case hydrocarbons the value at any other temperature may be used because K is constant over a large temperature range 3 Find the moisture content in hexane at an ambient temperature of 10 C if a dew point of 0 C is measured with the GE Infrastructu
25. 84 416 80 420 20 423 60 426 80 430 20 85 433 60 437 00 440 40 444 00 447 50 86 450 90 454 40 458 00 461 60 465 20 87 468 70 472 40 476 00 479 80 483 40 88 487 10 491 00 494 70 498 50 502 20 89 506 10 510 00 513 90 517 80 521 80 90 525 76 529 77 533 80 537 86 541 95 91 546 05 550 18 554 35 558 53 562 75 92 566 99 571 26 575 55 579 87 584 22 93 588 60 593 00 597 43 601 89 606 38 94 610 90 615 44 620 01 624 61 629 24 95 633 90 638 59 643 30 648 05 652 82 96 657 62 662 45 667 31 672 20 677 12 97 682 07 687 04 692 05 697 10 702 17 98 707 27 712 40 717 56 722 75 727 98 99 733 24 738 53 743 85 749 20 754 58 100 760 00 765 45 770 93 776 44 782 00 101 787 57 793 18 798 82 804 50 810 21 Application of the Hygrometer A 21 August 2004 Comparison of ppm Calculations There are three basic methods for determining the moisture content of a gas in ppm the calculations described in this appendix calculations performed with the slide rule device that is provided with each GE Infrastructure Sensing hygrometer values determined from tabulated vapor pressures For comparison purposes examples of all three procedures are listed in Table A 4 below Table A 4 Comparative ppm Values Calculation Method Dew Pressure Slide Appendix Vapor Point C psig Rule A Pressure 0 0 5 0 55 0 526 100 0 065 N A 0 0675 20 800 0 009 NA 0 0095 1500 0 005 N A 0 0051 0 37 40 38 88 100 4 8 5 2 4
26. 98 300 0 65 0 8 0 7016 1500 0 36 0 35 0 3773 0 N A 20 000 23 072 36 100 3000 3000 2956 9 id 800 420 400 416 3105 1500 220 200 223 0 A 22 Application of the Hygrometer August 2004 Liquid Applications Theory of Operation Moisture Content Measurement in Organic Liquids Henry s Law Type Analysis The direct measurement of water vapor pressure in organic liquids is accomplished easily and effectively with GE Infrastructure Sensing Aluminum Oxide Moisture Probes Since the moisture probe pore openings are small in relation to the size of most organic molecules admission into the sensor cavity is limited to much smaller molecules such as water Thus the surface of the aluminum oxide sensor which acts as a semi permeable membrane permits the measurement of water vapor pressure in organic liquids just as easily as it does in gaseous media In fact an accurate sensor electrical output will be registered whether the sensor is directly immersed in the organic liquid or it is placed in the gas space above the liquid surface As with gases the electrical output of the aluminum oxide probe is a function of the measured water vapor pressure When using the aluminum oxide sensor in non polar liquids having water concentrations lt 1 by weight Henry s Law is generally applicable Henry s Law states that at constant temperature the mass of a gas dissolved in a given volume of liquid is proportional to th
27. ALARM B Connection Connection Connection Normally Closed NC Normally Closed NC Normally Closed NC Armature Contact A Armature Contact A Armature Contact A Normally Open NO Normally Open NO Normally Open NO Figure 2 5 Back Panel Wiring Connections 2 6 Installation November 2004 Connecting the Recorder Output The MTS 5 has one isolated analog recorder output Connect your recorder to terminal block TB2 on the back of the MTS 5 as shown in Figure 2 5 on page 2 6 WARNING Never connect line voltage or any other power input to the recorder output terminals This output provides either a current or voltage signal which is set using switch S1 on the main PC board The MTS 5 is configured at the factory but you should check the switch position before making connections Follow these steps to check or reset switch S1 1 Make sure the MTS 5 is turned off and unplugged IWARNING You must disconnect line power before opening the MTS 5 enclosure 2 Remove the back cover of the enclosure by removing four screws IMPORTANT When you remove the back cover and main PC board do not dislodge or disconnect the flexible jumper that connects the display to the main PC board Remove the board from the enclosure only far enough to reach switch S1 3 Slowly slide the main PC board from the enclosure by pulling straight back being careful not
28. Alarm A and Alarm B relays through the TB2 terminal block on the back of the MTS 5 as shown in Figure 2 5 on page 2 6 Installation 2 9 Chapter 3 Operation amp Programming Powering Up u au 3 1 Powering DOWN ar een 3 1 Programming the Instrument 2 2 eee cece eee eee ee eens 3 2 Using the Ke pad rss een ee 3 2 Entering the User Program cece eee e cette eee 3 3 Exiting the User Program cece cece cece eee e 3 3 Navigating Through the Menus cece eee ee eee eens 3 4 The User1 and User2 Menus eee ee eee cece nena 3 6 Setting Up the MTS 5s ceo br e Eee do te c e Ales 3 9 Using the Alarms aa art 3 13 Using the Recorder Output eee cece e eee eee eee 3 16 Entering New Probe Calibration Data ccc cece eee eee 3 22 Additional Setup Procedures eee eee eee ee ee eee 3 23 August 2004 Powering Up Powering Down Whenever the power cord is connected to a power source the MTS 5 is actually powered up If the display is blank the unit is in standby mode which means it is powered up but inactive IMPORTANT For compliance with the European Union s Low Voltage Directive 73 23 EEC this unit requires an external power disconnect device such as a switch or circuit breaker The disconnect device must be marked as such clearly visible directly accessible and located within 1 8 m 6 ft of the MTS 5
29. ENTER To continue programming the MTS 5 proceed to the appropriate section for instructions Operation amp Programming 3 7 August 2004 ss Entering the User2 Menu Enter the MTS 5 User Program as described on page 3 3 ALA Press s or t until USEr2 appears on the display USEr2 Press ENTER Note The MTS 5 comes from the factory with the User2 passcode set to 1 If desired change the passcode as instructed below 1 Use the s and t keys to enter the passcode Then press ENTER unit You are now in the User2 menu Changing the User2 To change the User2 passcode continue as follows Passcode unit Press s or t until uSEr2 appears on the display uSEr2 Press ENTER XX Use the s and t keys to enter a new passcode Then press ENTER to save the new passcode or press ESC to keep the old passcode uSEr2 To exit the User Program press ESC until run appears Then press ENTER To continue programming the MTS 5 proceed to the appropriate section for instructions 3 8 Operation amp Programming August 2004 Setting Up the MTS 5 When you turn on the MTS 5 it begins to display measurements To program the meter and adapt it to your needs you can perform the following tasks select the units in which measurements are displayed page 3 9 enter the dew point measurement range page 3 10 set the
30. GE Infrastructure Sensing Moisture Target Series 5 Hygrometer User s Manual GE Infrastructure Sensing Moisture Target Series 5 Hygrometer o User s Manual 910 151E November 2004 Moisture Target Series 5 is a GE Panametrics product GE Panametrics has joined other GE high technology sensing businesses under a new name GE Infrastructure Sensing November 2004 Warranty Return Policy Each instrument manufactured by GE Infrastructure Sensing Inc is warranted to be free from defects in material and workmanship Liability under this warranty is limited to restoring the instrument to normal operation or replacing the instrument at the sole discretion of GE Infrastructure Sensing Inc Fuses and batteries are specifically excluded from any liability This warranty is effective from the date of delivery to the original purchaser If GE Infrastructure Sensing Inc determines that the equipment was defective the warranty period is one year for general electronic failures of the instrument one year for mechanical failures of the sensor If GE Infrastructure Sensing Inc determines that the equipment was damaged by misuse improper installation the use of unauthorized replacement parts or operating conditions outside the guidelines specified by GE Infrastructure Sensing Inc the repairs are not covered under this warranty The warranties set forth herein are exclusive and are in lieu of all oth
31. NTER t AL Press s or t until dSPtSt appears on the display tSPtst Press ENTER 8 8 8 8 8 8 All LCD segments are lit To exit the User Program press ESC until run appears Then press ENTER If one or more of the LCD segments is defective contact GE Infrastructure Sensing for assistance To continue programming the MTS 5 proceed to the appropriate section for instructions Operation amp Programming 3 23 August 2004 Restoring the Default Settings Entering High and Low Reference Values To restore all settings and data to their factory default values enter the User2 menu as described on page 3 8 and proceed as follows unit Press s or t until dEFALT appears on the display dEFALT Press ENTER SurE Press ENTER to reset all default values or ESC to keep the existing values done To exit the User Program press ESC until run appears Then press ENTER To continue programming the MTS 5 proceed to the appropriate section for instructions The MTS 5 is factory programmed with high and low reference values However you may have to enter new reference values if you replace the User Program or make other changes to the meter Consult the factory for more information IMPORTANT The high and low reference values relate to the moisture measurement circuitry of the MTS 5 They are not the same thing as the high and low dew points see page
32. O NC NO NC NO e FAULT ALARM A ALARM INTERNAL LINE L NEUTRAL WARNING WHEN SERVICING DO NOT FOLD OR CREASE KEYPAD DISPLAY CABLE THIS CAN CAUSE UNIT TO FAIL AC POWER Figure B 4 Interconnection Diagram 702 212 ALARM B INTERNAL TB1 26 30 DC POWER Outline and Installation Drawings B 4 Appendix C Menu Maps Main and User1 Menu Map User2 Menu Map Quick Setup Guide August 2004 to go back one level for better readability NOTE press at any time D is shown as DISPLAY number Screen Display Current Value ara V Adjust Numeric Value C T Select Menu Option v ia E passcode passcode 1 1 CONSULT FACTORY See Figure C 2 Must be pressed within 5 seconds USER1 ME dP ED DP value
33. Type Range 3 17 Selecting Units 04 3 16 Setting Up eese 3 16 Specifications lees 5 1 Switch Location 2 04 2 8 ESTING ro ones tar Ea 3 18 Reference Values Changing 3 24 Relative Humidity Calculating A 16 2 Index November 2004 Index cont S Sample System Mounting ess erases nuanses eee eee eee 2 2 Probe Installation 2 2 Schematic Wiring 00 2 6 Serial Number Entering 3 11 Solids Applications A 35 Specifications Alarms 2o xe LEAST Re ENS 5 1 Display se pies ee bile eee 5 1 Electronic 22 6 omissus 5 1 Moisture Probe 4 A 12 Probe 2 nta easet quor i3 5 3 R cord r re 2 eisen 5 1 Standby Mode 00 3 1 T Temperature Range 2 5 2 Testing Alarms oei 4220 sen eed bbe deeds 3 15 Display ft ike eee ve a ee 3 23 Recorder 2 beta ee bok 3 18 Troubleshooting Guide 4 2 U User Program Entering u a ren 3 3 EXIUnDg o vient reor poris NS 3 3 Navigating 000000 3 4 PROM iur asien s 4 4 Replacing oon sis corroe oirne ee eee 4 4 Restoring Defaults 3 24 V Versions Bench Mount 04 B 2 Board Mount 005 B 3 Panel Mount 05 B 1 Voltage Input Options 1 1 5 2 W Warm Up Time
34. aightedge connect the dew point as measured with the GE Infrastructure Sensing Hygrometer with the known system pressure Read the moisture content in ppm where the straightedge crosses the moisture content scale Typical Problems 1 Find the water content in a nitrogen gas stream if a dew point of 20 C is measured and the pressure is 60 psig Solution In Figure A 2 on page A 15 connect 60 psig on the Pressure scale with 20 C on the Dew Frost Point scale Read 200 ppm on the Moisture Content scale 2 Find the expected dew frost point for a helium gas stream having a measured moisture content of 1000 ppm and a system pressure of 0 52 atm Solution In Figure A 2 on page A 15 connect 1000 ppm on the Moisture Content scale with 0 52 atm on the Pressure scale Read the expected frost point of 27 C on the Dew Frost Point scale A 14 Application ofthe Hygrometer August 2004 Te DEW FROST POINT F DEW FROST POINT C PRESSURE PSIG PRESSURE ATMOSPHERES MOISTURE CONTENT ppm by volume Figure A 2 Moisture Content Nomograph for Gases Application ofthe Hygrometer A 15 August 2004 Parts per Million by Weight Relative Humidity Weight of Water per Unit Volume of Carrier Gas The water concentration in the gas phase of a system in parts per million by weight can be calculated directly from the ppm and the ratio of the molecular weight of water to that of the carrier gas as
35. al moisture content of 47 ppm on the left scale A 28 Application of the Hygrometer August 2004 Special Case cont Note The saturation value at 50 C for this liquid could also have been determined by connecting the K value of 5 1 with the ambient temperature of 50 C and reading a value of 475 ppm on the right scale For many applications a knowledge of the absolute moisture content of the liquid is not required Either the dew point of the liquid or its percent saturation is the only value needed For such applications the saturation value for the liquid need not be known The GE Infrastructure Sensing hygrometer can be used directly to determine the dew point and then the percent saturation can be calculated from the vapor pressures of water at the measured dew point and at the ambient temperature of the liquid P Saturation Eos x 100 Wy 100 1 12 Cs Ps Application of the Hygrometer A 29 August 2004 Empirical Calibrations For those liquids in which a Henry s Law type analysis is not applicable the absolute moisture content is best determined by empirical calibration A Henry s Law type analysis is generally not applicable for the following classes of liquids liquids with a high saturation value 2 by weight of water or greater liquids such as dioxane that are completely miscible with water liquids such as isopropyl alcohol that are conductive For such liquids measurements
36. also provide the sample system in an enclosure if requested If a sample system was ordered outline and dimension drawings are included with the shipment Follow the steps below to mount the sample system 1 Fasten the sample system plate or enclosure with a bolt in each of the four corners 2 Connect the sample system inlet to the process and the outlet to the return using appropriate stainless steel fittings and tubing Caution Do not start a flow through the system until the probe has been properly installed Installing the Probe GE Infrastructure Sensing probes are usually installed in a sample system The sample system protects the probes from any damaging elements in the process The probes are inserted into a cylindrical shaped container called the sample cell which is included as part of your sample system M2 probes have 3 4 16 straight threads sealed with an o ring to secure the probe either into the sample system or directly into the process line Other fittings are available for special applications Caution If the probe is to be mounted in the process line consult GE Infrastructure Sensing for proper installation instructions and precautions 2 2 Installation August 2004 Installing the Probe Refer to Figure 2 2 below and follow these steps to install the probe cont into the sample cell 1 Insert the probe into the sample cell so it is perpendicular to the sample inlet 2 Screw the probe into the
37. ature If the moisture probe temperature is expected to vary the test should be performed at more than one temperature Application of the Hygrometer A 25 August 2004 3 E o 2 a uw za O O Ww a Fr o Oo DEW FROST POINT AT TEMPERATURE T C Figure A 3 TEMPERATURE T C SATURATION VALUE AT TEMPERATURE T ppmw HENRY S LAW CONSTANT K Moisture Content Nomograph for Liquids A 26 Application of the Hygrometer August 2004 Special Case As mentioned earlier saturated straight chain hydrocarbons represent a special case where the Henry s Law constant does not vary appreciably with temperature In such cases use the nomograph for liquids in Figure A 3 on page A 26 to complete the analysis Determination of moisture content if the Henry s Law constant K is known 1 Using a straightedge connect the known K value on the center scale with the dew frost point as measured with the GE Infrastructure Sensing hygrometer Read moisture content ppm where the straightedge crosses the scale on the left Typical Problems 1 Find the moisture content in benzene at an ambient temperature of 30 C if a dew point of 0 C is measured with the GE Infrastructure Sensing hygrometer a From the literature it is found that Cs for benzene at a temperature of 30 C is 870 ppm b Using a straightedge on Figure A 3 on page A 26 connect the 8
38. be at least 10 C higher than the measured dew point up to a maximum of 70 C Because an ambient temperature increase may cause water vapor to be desorbed from the walls of the sample system it is possible to observe a diurnal change in moisture concentration for a system exposed to varying ambient conditions In the heat of the day the sample system walls will be warmed by the ambient air and an off gassing of moisture into the process fluid with a corresponding increase in measured moisture content will occur The converse will happen during the cooler evening hours GE Infrastructure Sensing hygrometers are unaffected by the fluid flow rate The moisture probe is not a mass sensor but responds only to water vapor pressure The moisture probe will operate accurately under both static and dynamic fluid flow conditions In fact the specified maximum fluid linear velocities see Table A 1 and Table A 2 on page A 5 for the M Series Aluminum Oxide Moisture Probe indicate a mechanical stability limitation rather than a sensitivity to the fluid flow rate If the measured dew point of a system changes with the fluid flow rate then it can be assumed that off gassing or a leak in the sample system is causing the variation If secondary moisture is entering the process fluid either from an ambient air leak or the release of previously absorbed moisture from the sample system walls an increase in the flow rate of the process fluid will dilute the s
39. below 50 C 58 F Note Standard cable assemblies including connectors can be ordered from GE Infrastructure Sensing in any length up to 600 meters 2000 feet To connect the probe cable complete these steps 1 Connect the cable to the probe by inserting the bayonet type connector onto the probe and twisting the shell clockwise until it snaps into a locked position approximately 1 8 turn 2 Connect the probe cable see Figure 2 4 below to terminal block TB3 on the back of the MTS 5 see Figure 2 5 on page 2 6 IMPORTANT To maintain good contact at each terminal block and to avoid damaging the pins on the connector pull the connector straight off not at an angle make cable connections while the connector is away from the unit and push the connector straight on not at an angle when the wiring is complete Shield Green Figure 2 4 Probe Cable Two Wire Shielded Installation 2 5 November 2004 Connecting the Probe cont PROBE RECORDER Connection Connection Shield Gnd Signal Green Return Red LINE POWER internal connections AC DC Line Input Neutral Return Ground Ground sion Faur ararma_JararmB N NIN NIN N c lolc lo c o e lsisisie e opmjujo2 Never Connect Any Power Inputs to TB2 or TB3 FAULT ALARM ALARM A
40. c2 USErl 1 ENT Y passcode S 4 v E E wn Z ja o ENT Y serial no om Y v ALF 4 Y ALF di zi ALFEn Must be pressed within 5 seconds dPHI Figure C 3 Quick Setup Guide 9 D 4 August 2004 XX X C a NOTE press at any time BE to go back one level Y ESC is shown as ent for better readabilitu ENT ESC ENT Y Y ALA ESC RUN ENT Y dEG C dEG F A H E ALA di ENT Y AL LO a ALHI ProbE ENT Y setpoint DP range ENT Y MH value 4 v DISPLAY Screen Display 4 9 Adjust Numeric Value ENT number Current Value Select Menu Option y Menu Maps C 3 November 2004 Index A E Alarms Electronic Specifications 5 1 Connecting s secsi opacs ania dyan 2 9 Empirical Calibrations A 30 Faults earo 2 ee es 2 9 3 13 High IEoW onere temi 2 9 3 14 F Programming 204 5 3 13 Specifications 0200 20 5 1 Factory Setup 6 sees eeees 3 25 iq A fos dk Cie et ee 3 15 Fault Alarm 0 0 0 eee eee eee 2 9 Tap Point na 3 14 Flow Rates Applications Ga
41. center scale 3 Using a straightedge connect above K value with the dew frost point as measured with the GE Infrastructure Sensing hygrometer 4 Read the moisture content ppm where the straight edge crosses the moisture content scale Empirical Determination of K and Cs If the values of K and Cg are not known the GE Infrastructure Sensing hygrometer can be used to determine these values In fact only one of the values is required to determine ppm from the nomograph in Figure A 3 on page A 26 To perform such an analysis proceed as follows 1 Obtain a sample of the test solution with a known water content or perform a Karl Fischer titration on a sample of the test stream to determine the ppm of water Note The Karl Fischer analysis involves titrating the test sample against a special Karl Fischer reagent until an endpoint is reached 2 Measure the dew point of the known sample with the GE Infrastructure Sensing hygrometer 3 Measure the temperature C of the test solution 4 Using a straightedge connect the moisture content ppm with the measured dew point and read the K value on the center scale 5 Using a straightedge connect the above K value with the measured temperature C of the test solution and read the saturation concentration ppm IMPORTANT Since the values of K and Cs vary with temperature the hygrometer measurement and the test sample analysis must be done at the same temper
42. corder Output Testing the Recorder Output rerd HI Press ENTER The current recorder output high value in the units specified on page 3 16 is displayed XXC Use s and t to set the desired recorder output high value Then press ENTER rerd HI To exit the User Program press ESC until run appears Then press ENTER To continue programming the MTS 5 proceed to the appropriate section for instructions After you have set up the recorder output as described in the previous section the output may be tested and or adjusted as necessary IMPORTANT Be sure switch SI on the main PC board is set to the correct output type current or voltage See Connecting the Recorder Output on page 2 11 To measure the recorder output signal use a digital multimeter that can measure 0 20 mA at a resolution of 0 01 mA or 0 2 V ata resolution of 0 0001 V Connect the meter in parallel with the two recorder connections for a voltage measurement or in series with either of the recorder connections for a current measurement To test the recorder output enter the User2 menu as described on page 3 8 Then proceed as follows unit Press s or t until tESt appears on the display tESt Press ENTER 3 18 Operation amp Programming August 2004 Testing the Recorder
43. corresponding to the specified dew point Press ENTER XXX Press s or t to enter the next dew point value Press ENTER Note Dew point temperatures range from the minimum to the maximum settings in 10 degree increments Continue entering dew point MH pairs until you have completed entering all the calibration curve points 3 22 Operation amp Programming August 2004 Entering New Probe Calibration Data cont XXX To exit the User Program press ESC until run appears Then press ENTER To continue programming the MTS 5 proceed to the appropriate section for instructions Additional Setup In addition to those procedures already discussed the following Procedures programming steps may be performed testing the liquid crystal display page 3 23 restoring the factory default settings page 3 24 entering new high and low reference values page 3 24 changing factory setup information page 3 25 Note Itis not normally necessary or recommended that the factory setup information be changed Before attempting this contact GE Infrastructure Sensing for assistance Testing the Display The liquid crystal display LCD may be tested to ensure that all of its segments are working To perform this task enter the User2 menu as described on page 3 8 and proceed as follows unit Press s or t until tESt appears on the display tESt Press E
44. cs of the fluid flow Response times and measurement values will be affected by the degree of equilibrium reached within system Factors such as gas pressure flow rate materials of construction length and diameter of piping etc will greatly influence the measured moisture levels and the response times Assuming that all secondary sources of moisture have been eliminated and the sample system has been allowed to come to equilibrium then the measured dew point will equal the actual dew point of the process fluid Some of the most frequently encountered problems associated with moisture monitoring sample systems include the moisture content value changes as the total gas pressure changes the measurement response time is very slow the dew point changes as the fluid temperature changes the dew point changes as the fluid flow rate changes A 2 Application of the Hygrometer August 2004 Moisture Monitor Hints cont Pressure Response Time GE Infrastructure Sensing hygrometers measure only water vapor pressure In addition the instrument has a very rapid response time and it is not affected by changes in fluid flow rate If any of the above situations occur then they are almost always caused by a defect in the sample system The moisture sensor itself cannot lead to such problems GE Infrastructure Sensing hygrometers can accurately measure dew points under pressure conditions ranging from vacuums as low as a
45. dio warea leva idiolg CER ERREUR TC ERAS TRU TU ee eT SE OER aa A 23 Theory or Operationerna ea telat ee re A 23 Moisture Content Measurement in Organic LiquidS 0 cece ccc eee nn A 23 ERDpIFISOECOHIDIOUO HS ics abre b cena er ee RE ar ni ctor terc Lt p ee d A 30 Instructions Tor Karl FisehieAndltsissz ar due ee te ee Al etit A 32 Instructions for Preparing Known Samples ut ae ae a A 33 Additional Notes for Liquid Applications 2 0 0 cece cece ccc nee e A 34 SONGS AD DICOONS en een alliierten qmd A 35 November 2004 a Li a O aa Table of Contents cont Appendix B Outline and Installation Drawings Panel Mount Enclosure 7 12 2042 os o xc BR a EX Ru Rus B 1 Bench Mount Enclosure 712 1043 nase B 2 Board Mount Version 7 12 1045 see are B 3 Interc nnection Diagrami 7022 12 nero air nota B 4 Appendix C Menu Maps Mamsand UserkMenu Map A er hehe hh A Mee es C 1 User Ment MOp ete etu er aa eue ad C 2 Q ick Set p GIO 2000228 nee Ido o ToS UO asien Am bb bre basti veros C 3 vi Chapter 1 Features and Capabilities Introd ctloncu ss s hae Bese ee Bae ehe Dees Electronics Units oes sees aed ocean er sn Nee Probes a nn ern er a August 2004 Introduction Electronics Unit Probes The Moisture Target Series 5 is a microprocessor based single channel hygrometer that measures moisture content in gases It is intended for Original Equipment Manufacturer OEM application
46. e ALT or too dry Sensor temperature is greater than 70 C Reads too dry Return the probe to factory for evaluation 158 F Stream particles causing abrasion Reads too wet or too dry Return the probe to factory for evaluation Service and Maintenance August 2004 Symptom A blinking E on the left side of the display Screen always reads the wettest highest programmed moisture calibration value while displaying dew frost point Possible Cause Probe is saturated Liquid water is present on the sensor surface and or across electrical connections Response Table 4 1 Troubleshooting Guide for Common Problems cont Clean the probe as described in Appendix A page A 8 Then reinstall the sensor There is a shorted circuit in the sensor Run dry gas over the sensor surface If high reading persists the probe is probably shorted and should be returned to the factory for evaluation Sensor is contami nated with conduc tive particles see Appendix A page A 7 Clean the probe as described in Appendix A page A 8 Then reinstall the sensor Improper cable connection Check the cable connections to both the probe and the hygrometer A blinking E on the left side of the display Screen always reads the driest lowest programmed moisture calibration value while displaying dew frost point Open circuit in sensor Return the p
47. e at a specified dew point In addition all definitions involving humidity can then be expressed in terms of the water vapor pressure Nomenclature The following symbols and units are used in the equations that are presented in the next few sections RH relative humidity Ty temperature K C 273 Tp temperature R F 460 ppm parts per million by volume ppm parts per million by weight e My molecular weight of water 18 e Mr molecular weight of carrier gas Ps saturation vapor pressure of water at the prevailing temperature mm of Hg Pw water vapor pressure at the measured dew point mm of Hg e P total system pressure mm of Hg Application of the Hygrometer A 15 August 2004 EEE SE EE Parts per Million by The water concentration in a system in parts per million by volume Volume is proportional to the ratio of the water vapor partial pressure to the total system pressure P ppm n x 10 A 1 T In a closed system increasing the total pressure of the gas will proportionally increase the partial pressures of the various components The relationship between dew point total pressure and ppm is provided in nomographic form in Figure A 2 on page A 15 Note The nomograph shown in Figure A 2 is applicable only to gases Do not apply it to liquids To compute the moisture content for any ideal gas at a given pressure refer to Figure A 2 on page A 15 Using a str
48. e exposed sensor carefully reinstall the protective shield over the sensor 9 Connect the probe cable to the cleaned probe and measure the dew point of the same ambient air recorded in Step 1 10 If the probe is determined to be in proper calibration 2 C 3 6 F it has been successfully cleaned and may be reinstalled in the sample cell If the probe is not in proper calibration proceed to Step 11 11 Repeat Steps 1 10 using soaking time intervals of 5 times the previous cleaning sequence until two consecutive cleanings produce identical probe responses to the ambient dew point Application of the Hygrometer A 9 August 2004 S4938uu IUJ SAYDU ui 310 SUOISUBUIG v uo1 A DIG ZE EX Gl e ezigbu O i 88 ST S290 ij AijISOLOg uoJollA OOT PISIUS Y JOSUaS di7 8 HT Ld uuo xipueg 0 JUa DAINDZ S869 0S 2 OT TT 2EW0 OHLIT W E P d S27 28TO 4 S 9 0S2 0 m I ST4S 08E 2 0822 2260 m S6081 28TE U 85 82 SCTT 4 1 GE Infrastructure Sensing M Series Moisture Probe Figure A Application of the Hygrometer A 10 August 2004 Corrosive Gases And Liquids GE Infrastructure Sensing M Series Aluminum Oxide Moisture Probes have been designed to minimize the affect of corrosive gases and liquids As indicated in Materials of Construction on page A 12 no copper solder or epoxy is used in the construction of these
49. e partial pressure of the gas in the system Stated in terms pertinent to this discussion it can be said that the ppm of water in hydrocarbon liquids is equal to the partial pressure of water vapor in the system times a constant As discussed above a GE Infrastructure Sensing aluminum oxide probe can be directly immersed in a hydrocarbon liquid to measure the equivalent dew point Since the dew point is functionally related to the vapor pressure of the water a determination of the dew point will allow one to calculate the ppm of water in the liquid by a Henry s Law type analysis A specific example of such an analysis is shown below For liquids in which a Henry s Law type analysis is applicable the parts per million by weight of water in the organic liquid is equal to the partial pressure of water vapor times a constant ppm Kx Pw 1 9 where K is the Henry s Law constant in the appropriate units and the other variables are as defined on page A 13 Application of the Hygrometer A 23 August 2004 ss Henry s Law Type Analysis Also the value of K is determined from the known water saturation cont concentration of the organic liquid at the measurement temperature t Saturation ppm Kar 1 10 Ps For a mixture of organic liquids an average saturation value can be calculated from the weight fractions and saturation values of the pure components as follows Ave Cy X Xi C 1 11 izl where X i
50. econdary moisture source As a result the vapor pressure will be lowered and a lower dew point will be measured Note Refer to the Specifications chapter in the Hygrometer User s Manual for the maximum allowable flow rate for the instrument A 4 Application of the Hygrometer Flow Rate cont August 2004 Hle Table A 1 Maximum Gas Flow Rates Based on the physical characteristics of air at a temperature of 77 F and a pressure of I atm the following flow rates will produce the maximum allowable gas stream linear velocity of 10 000 cm sec in the corresponding pipe sizes Inside Pipe Diameter in Gas Flow Rate cfm 0 25 7 0 50 27 0 75 60 1 0 107 2 0 429 3 0 966 4 0 1 718 5 0 2 684 6 0 3 865 7 0 5 261 8 0 6 871 9 0 8 697 10 0 10 737 11 0 12 991 12 0 15 461 Table A 2 Maximum Liquid Flow Rates Based on the physical characteristics of benzene at a temperature of 77 F the following flow rates will produce the maximum allowable fluid linear velocity of 10 cm sec in the corresponding pipe sizes Inside Pipe Diameter Flow Rate in gal hr I hr 0 25 3 11 0 50 12 46 0 75 27 103 1 0 48 182 2 0 193 730 3 0 434 1 642 4 0 771 2 919 5 0 1 205 4 561 6 0 1 735 6 567 7 0 2 361 8 939 8 0 3 084 11 675 9 0 3 903 14 776 10 0 4 819 18 243 11 0 5 831 22 074 12 0 6 939 26 269
51. ecorder output cannot be properly adjusted contact GE Infrastructure Sensing for assistance rADJ To exit the User Program press ESC until run appears Then press ENTER To continue programming the MTS 5 proceed to the appropriate section for instructions Operation amp Programming 3 21 August 2004 Entering New Probe Calibration Data Whenever a new or recalibrated probe is installed in the system the calibration curve for that probe as supplied with the probe must be entered into the MTS 5 User Program IMPORTANT Enter a new probe calibration curve only when necessary Consult GE Infrastructure Sensing for guidance A probe calibration curve consists of 2 to 18 pairs of dew point temperature MH values The MH value is the sensor signal at the associated dew point IMPORTANT Before entering the calibration curve data make sure the high and low dew points programmed into the MTS 5 match those on the calibration curve see page 3 10 for instructions To enter the new probe calibration data enter the User menu as described on page 3 7 Then complete the following steps S_No Press s or t until ProbE appears on the display ProbE Press ENTER The current minimum dew point reading see page 3 10 is displayed XXX Press ENTER The current MH sensor signal value at that dew point is displayed X XXXX Press s or t to enter the new MH value
52. er enter page 3 22 probe calibration XXX select dew point value X XXXX corresponding curve data MH value dEG C SC AL b dEG F F Low Alarm Base ia SEDE A H MH High Alarm ALB di disable Alarm B C r_unit select output units F MH rerd set up recorder page 3 16 output ont select output mode and range rcdLO enter output minimum rcd HI enter output maximum current passcode enter new passcode displayed change Userl passcode Operation amp Programming 3 5 August 2004 unit page 3 9 select display units Table 3 4 The User2 Menu User2 Menu Item Submenu Item Sub Submenu Item C F MH OFFSEt page 3 12 set dew point offset C N A dEFALT page 3 24 reset to factory default settings press ENTER to default N A tESt see submenu page references test alarms recorder output display test alarm relays ALA On ALA OF ALb On ALb OF ALF On ALF OF press t to turn alarms on and off test and adjust recorder output r_out page 3 19 test output select output mode and range select H 100 t 50 L 0 rAdJ page 3 20 adjust output dSPtSt page 3 23 test display rEF page 3 24 calibration references HrEF LrEF enter high and low reference values bLitE page 3 11 set backlight timer uSErl page 3 7 change Userl passcode c
53. er warranties whether statutory express or implied including warranties of merchantability and fitness for a particular purpose and warranties arising from course of dealing or usage or trade If a GE Infrastructure Sensing Inc instrument malfunctions within the warranty period the following procedure must be completed 1 Notify GE Infrastructure Sensing Inc giving full details of the problem and provide the model number and serial number of the instrument If the nature of the problem indicates the need for factory service GE Infrastructure Sensing Inc will issue a RETURN AUTHORIZATION number RA and shipping instructions for the return of the instrument to a service center will be provided 2 If GE Infrastructure Sensing Inc instructs you to send your instrument to a service center it must be shipped prepaid to the authorized repair station indicated in the shipping instructions 3 Upon receipt GE Infrastructure Sensing Inc will evaluate the instrument to determine the cause of the malfunction Then one of the following courses of action will then be taken fthe damage is covered under the terms of the warranty the instrument will be repaired at no cost to the owner and returned f GE Infrastructure Sensing Inc determines that the damage is not covered under the terms of the warranty or if the warranty has expired an estimate for the cost of the repairs at standard rates will be provided Upon receipt
54. he response time of the GE Infrastructure Sensing standard M Series Aluminum Oxide Moisture Probe is very rapid a step change of 63 in moisture concentration will be observed in approximately 5 seconds Thus the observed response time to moisture changes is in general limited by the response time of the sample system as a whole Water vapor is absorbed tenaciously by many materials and a large complex processing system can take several days to dry down from atmospheric moisture levels to dew points of less than 60 C Even simple systems consisting of a few feet of stainless steel tubing and a small chamber can take an hour or more to dry down from dew points of 5 C to 70 C The rate at which the system reaches equilibrium will depend on flow rate temperature materials of construction and system pressure Generally speaking an increase Application of the Hygrometer A 3 August 2004 Response Time cont Temperature Flow Rate in flow rate and or temperature will decrease the response time of the sample system To minimize any adverse affects on response time the preferred materials of construction for moisture monitoring sample systems are stainless steel Teflon and glass Materials to be avoided include rubber elastomers and related compounds The GE Infrastructure Sensing hygrometer is largely unaffected by ambient temperature However for best results it is recommended that the ambient temperature
55. he weight of water per unit weight of carrier gas In such a case the unit of measure defined by the following equation is the most commonly used grains of water _ 7000 x My x Pw 1 7 lb of gas M x Pr For ambient air at 1 atm of pressure the above equation reduces to the following grains of water _ 572xP 1 8 Ib of gas ae W l Application of the Hygrometer A 17 August 2004 Table A 3 Vapor Pressure of Water Note f the dew frost point is known the table will yield the partial water vapor pressure Pw in mm of Hg If the ambient or actual gas temperature is known the table will yield the saturated water vapor pressure Ps in mm of Hg Water Vapor Pressure Over Ice Temp C 0 2 4 6 8 90 0 000070 0 000048 0 000033 0 000022 0 000015 80 0 000400 0 000290 0 000200 0 000140 0 000100 70 0 001940 0 001430 0 001050 0 000770 0 000560 60 0 008080 0 006140 0 004640 0 003490 0 002610 50 0 029550 0 023000 0 017800 0 013800 0 010600 40 0 096600 0 076800 0 060900 0 048100 0 037800 30 0 285900 0 231800 0 187300 0 150700 0 120900 Temp C 0 0 0 2 0 4 0 6 0 8 29 0 317 0 311 0 304 0 298 0 292 28 0 351 0 344 0 337 0 330 0 324 27 0 389 0 381 0 374 0 366 0 359 26 0 430 0 422 0 414 0 405 0 397 25 0 476 0 467 0 457 0 448 0 439 24 0 526 0 515 0 505 0 495 0 486 23 0 580 0 569 0 558 0 547 0 536 22 0 640 0 627 0 615 0 603 0 592 21 0 705 0 691 0 678 0 665 0 652 20 0 776 0 761 0 747 0 733
56. his type of particulate with water will cause pitting or severe corrosion of the sensor element In such instances the sensor cannot be cleaned or repaired and the probe must be replaced Obviously the standard moisture probe cannot be used in such applications unless the complete removal of such particulates by adequate filtration is assured Application of the Hygrometer A 7 August 2004 Aluminum Oxide Probe As part of a routine preventive maintenance program the moisture Maintenance probe should be removed and returned to the factory for recalibration once a year Between these calibrations if the aluminum oxide moisture probe becomes contaminated with an electrically conductive liquid the moisture measurements will be erroneously high In sucha situation the probe should be removed from the sample system and cleaned as described in Cleaning the Moisture Probe on page A 9 WARNING The probe cleaning procedure should be performed only by a qualified technician or chemist IMPORTANT Moisture probes must be handled carefully and cannot be cleaned in any fluid which will attack its components The probe s materials of construction are Al Al O3 nichrome gold stainless steel glass and Viton A Also the sensor s aluminum sheet is very fragile and can be easily bent or distorted Do not permit anything to touch it WARNING Make sure you replace the moisture probe before restarting the system A 8 Applicati
57. ics unit mounting the sample system installing the probe into the sample system wiring the input power wiring the probe recorder output and alarm connections IWARNING To ensure safe operation the MTS 5 must be installed and operated as described in this manual Also be sure to follow all applicable local safety codes and regulations for installing electrical equipment Mounting the Electronics The standard MTS 5 panel mount unit can be installed in a Unit rectangular cutout in a panel up to an inch thick see Appendix B Outline and Installation Drawings IMPORTANT For NEMA 4 and IP66 installation the MTS 5 must be mounted in a rigid flat panel using the panel gasket provided and all four mounting brackets see Figure 2 1 below Brackets ge i io Gasket Figure 2 1 Mounting the MTS 5 in a Panel Installation 2 1 August 2004 ss Mounting the Electronics To mount the MTS 5 complete the following steps Unit cont 1 Remove the four clip on mounting brackets from the unit see Figure 2 1 on page 2 1 2 Make sure the gasket is installed then slide the meter into the front of the panel cutout 3 Reinstall the four clip on mounting brackets from behind the panel 4 Secure the unit to the panel by tightening the four mounting bracket screws Mounting the Sample The sample system is normally fastened to a metal plate that has four Sy stem mounting holes GE Infrastructure Sensing can
58. inless steel filter in the fluid stream On rare occasions non conductive particulate material may become lodged under the contact arm of the sensor creating an open circuit If this condition is suspected refer to Aluminum Oxide Probe Maintenance on page A 8 and Cleaning the Moisture Probe on page A 9 for the recommended cleaning procedure A 6 Application of the Hygrometer August 2004 Conductive Particulates Corrosive Particulates Note Metallic particles carbon particles and conductive liquid droplets are typical of this category Since the hygrometer reading is inversely proportional to the impedance of the sensor a decrease in sensor impedance will cause an increase in the meter reading Thus trapped conductive particles across the sensor leads or on the sensor surface which will decrease the sensor impedance will cause an erroneously high dew point reading The most common particulates of this type are carbon from furnaces iron scale from pipe walls and glycol droplets from glycol based dehydrators If the system contains conductive particulates it is advisable to install a Teflon or stainless steel filter in the fluid stream Note Sodium chloride and sodium hydroxide particulates are typical of this category Since the active sensor element is constructed of aluminum any material that corrodes aluminum will deleteriously affect the operation of the moisture probe Furthermore a combination of t
59. larm relay is a single pole double throw contact set that contains the following contacts Normally Open NO Armature Contact Normally Closed NC The fault alarm if enabled trips when there is one or more of the following faults power failure range error system reset by the watchdog function Note The watchdog function is a supervisory circuit that automatically resets the user program in the event of a system program error The fault alarm operates in a fail safe manner Pins 4 and 5 provide a normally closed contact When the MTS 5 is operating in a non fault state the fault alarm relay is energized to keep the contact between pins 4 and 5 open When a fault occurs the fault alarm relay is de energized so the contact between pins 4 and 5 closes The contact between pins 5 and 6 normally open works in the opposite way closed during ordinary operation open when there is a fault Make connections to the fault alarm through terminal block TB2 on the back of the MTS 5 as shown in Figure 2 5 on page 2 6 IMPORTANT To maintain good contact at each terminal block and to avoid damaging the pins on the connector pull the connector straight off not at an angle make cable connections while the connector is away from the unit and push the connector straight on not at an angle when the wiring is complete Each of these alarms can be set to trip on a high or low condition Make the connections to
60. lication of the Hygrometer A 31 August 2004 Instructions for Karl Fischer Analysis To perform a Karl Fisher analysis use the apparatus in Figure A 4 on page A 31 and complete the following steps 1 2 Fill the glass bottle completely with the sample liquid Close both valves and turn on the magnetic stirrer Permit sufficient time for the entire test apparatus and the sample liquid to reach equilibrium with the ambient temperature Turn on the hygrometer and monitor the dew point reading When a stable dew point reading indicates that equilibrium has been reached record the reading Insert a syringe through the rubber septum and withdraw a fluid sample for Karl Fischer analysis Record the actual moisture content of the sample Open the exhaust valve Open the inlet valve and increase the moisture content of the sample by bubbling wet N gt through the liquid or decrease the moisture content by bubbling dry N through the liquid When the hygrometer reading indicates the approximate moisture content expected close both valves Repeat steps 3 8 until samples with several different moisture contents have been analyzed A 32 Application of the Hygrometer August 2004 Instructions for Preparing Note This procedure is only for liquids that are highly miscible with Known Samples water Excessive equilibrium times would be required with less miscible liquids To prepare samples
61. m is not at equilibrium or there is a leak Allow sufficient time for sample system to equilibrate and moisture reading to become steady Check for leaks Dew point at the sampling point is different from the dew point of the main stream Reads too wet or too dry Readings may be correct if the sampling point and main stream do not run under the same process conditions The different process conditions cause readings to vary Refer to Appendix A page A 3 for more information If sampling point and main stream conditions are the same check sample system pipes and any pipe between the sample system and main stream for leaks Also check sample system for adsorbing water surfaces such as rubber or plastic tubing paper type filters or condensed water traps Remove or replace the contaminating parts with stainless steel parts Sensor or sensor ss ae Reads too wet Clean the sensor and the sensor shield as YP or too dry described in Appendix A page A 8 contaminants see Then reinstall the sensor Appendix A pages A 6 and A 7 Pacha P ith Clean the sensor and the sensor shield as ee Reads high described in Appendix A page A 8 Then conductive f f ticles dew point reinstall the sensor Also install a proper filter par i e sintered or coalescing element see Appendix A page A 7 Sensor is corroded see Appendix A Reads too wet Return the probe to factory for evaluation pac
62. m oxide moisture probes have been designed to reliably measure the moisture content of both gases and liquids The measured dew point will be the real dew point of the system at the measurement location and at the time of measurement However no moisture sensor can determine the origin of the measured moisture content In addition to the moisture content of the fluid to be analyzed the water vapor pressure at the measurement location may include components from sources such as moisture from the inner walls of the piping external moisture through leaks in the piping system and trapped moisture from fittings valves filters etc Although these sources may cause the measured dew point to be higher than expected it is the actual dew point of the system at the time of measurement One of the major advantages of the GE Infrastructure Sensing hygrometer is that it can be used for in situ measurements i e the sensor element is designed for installation directly within the region to be measured As a result the need for complex sample systems that include extensive piping manifolds gas flow regulators and pressure regulators is eliminated or greatly reduced Instead a simple sample system to reduce the fluid temperature filter contaminants and facilitate sensor removal is all that is needed Whether the sensor is installed in situ or in a remote sampling system the accuracy and speed of measurement depend on the piping system and the dynami
63. m triggers when one of the following events occurs e a power failure arange error a reading outside the programmed dew point range reset by the watchdog function Note The watchdog function is a supervisory circuit that automatically resets the User Program whenever a system error occurs The fault alarm operates in fail safe mode That is when the meter is in a non fault state the fault alarm relay is energized to keep the contact between the normally closed pins TB2 4 and TB2 5 open When a fault occurs the fault alarm relay is de energized so that the contact between pins 4 and 5 closes The contact between the normally open pins TB2 5 and TB2 6 works in the opposite way closed during ordinary operation open when there is a fault To enable or disable the fault alarm enter the User Program as described on page 3 3 and proceed as follows ALA Press s or t until AL F appears on the display ALF Press ENTER ALF En Press s or t until the enabled or disabled ALF En or ALF di option appears on the ALF di display Press ENTER ALF To exit the User Program press ESC until run appears Then press ENTER To continue programming the MTS 5 proceed to the appropriate section for instructions Operation amp Programming August 2004 ss Setting the High Low Alarm A is set up in the Main menu see page 3 3 and Alarm B is set Ala
64. mum Fuse 100 120 VAC units 5 mm x 20 mm Type T 0 125 A 250 V 230 240 VAC units 5 mm x 20 mm Type T 0 125 A 250 V 24 VDC units 5 mm x 20 mm Type T 0 4 A 250 V Temperature operating 0 to 60 C storage 30 to 70 C Warm Up Time meets specified accuracy within three minutes Dimensions panel mount 2 83 x 5 67 x 4 71 in H x W x D 71 9 x 144 x 119 6 mm cutout required 2 65 x 5 4 in H x W 67 3 x 137 2 mm bench mount 4 11 x 6 42 x 4 71 in H x W x D 104 4 x 144 x 119 6 mm board mount 8 00 x 5 50 x 2 53 in H x W x D 203 2 x 139 7 x 64 2 mm European Compliance complies with EMC Directive 89 336 EEC and 73 23 EEC Low Voltage Directive Installation Category II Pollution Degree II Note The mains cord installation must comply with European Standard EN61010 to maintain Low Voltage Directive compliance 5 2 Specifications August 2004 Moisture Measurement Sensor Type thin film aluminum oxide moisture sensor probe Moisture Probe Compatibility compatible with all GE Infrastructure Sensing M Series aluminum oxide moisture probes Probe Cable Length 2 000 ft 600 m maximum Moisture Probe Pressure Rating M1 5 microns Hg to 75 psig M2 5 microns Hg to 5 000 psig Dew Frost Point Temperature Overall Calibration Range 110 to 60 C Available Calibration Range Options Standard 80 to 20 C with data to 110 C Extended High 80 to 60 C with data
65. nt for test samples A 36 Application of the Hygrometer Appendix B Outline and Installation Drawings Panel Mount Enclosure 712 1042 0 cece cece cee eee B 1 Bench Mount Enclosure 2 712 1043 eee cece eee B 2 Board Mount Version 712 1045 cee eee cece eee B 3 Interconnection Diagram 702 212 ce cece eee eee eee B 4 August 2004 5 35 135 9 5 40 137 2 PANEL CUTOUT Notes 1 Dimensions are in inches millimeters 32 8 1 2 Weight 1 5 Ib 68 kg 5 67 144 0 MOISTURE TARGET SERIES 5 Figure B 1 Panel Mount Enclosure 712 1042 Outline and Installation Drawings B 1 August 2004 5 35 135 9 4 71 1193 6 Notes 1 Dimensions are in inches millimeters 2 Weight 1 5 Ib 68 kg 32 8 1 MOISTURE TARGET SERIES 5 DEW FROST POINT 5 67 144 0 6 42 163 1 Figure B 2 Bench Mount Enclosure 712 1043 Outline and Installation Drawings B 2 August 2004 Notes 5 50 139 7 1 Dimensions are in inches millimeters 5 00 127 0 2 Weight 1 22 Ib 554 kg JAAA DWAOeT dp 23 rg e mS o 2B 7 50 8 00 190 5 203 2 DEW FROST POINT 2 53 64 2 Figure B 3 Board Mount Version 44712 1045 Outline and Installation Drawings B 3 November 2004 OUT RTN N
66. oceed to the appropriate section for instructions If the recorder output testing performed in the previous section indicates the need for an adjustment continue the programming from the screen above as follows r_out Press s or t until rADJ appears Then press ENTER Note Only the high end reading of the recorder output range can be adjusted the low end of the range is not adjustable KA The current adjustment value is displayed Press s or t enter an adjustment value from 100 to 100 Then press ENTER To determine the adjustment value needed to make the 100 test reading match the desired high reading see the sample calculation on page 3 21 3 20 Operation amp Programming August 2004 Adjusting the Recorder Output cont Each recorder output adjustment unit equals 0 005 mA or 0 0005 V As an example assume that the 100 test output produced a reading of 20 2 mA for a recorder output range of 4 20 mA Then the required output adjustment would be calculated as follows Adjustment zu mar 202 IDA I AO Units 3 1 0 005 mA Unit Enter the calculated adjustment at the previous display page 3 20 rADJ Press s or t until r out appears Then press ENTER r_out Re test the recorder output as described on page 3 18 Continue to test and adjust the recorder output until the output is within the required tolerance Then proceed as follows Note Ifthe r
67. of known moisture content use the apparatus in Figure A 4 on page A 31 and complete the following steps 1 Weigh the dry empty apparatus 2 Fill the glass bottle with the sample liquid 3 Open both valves and turn on the magnetic stirrer 4 While monitoring the dew point reading with the hygrometer bubble dry N through the liquid until the dew point stabilizes at some minimum value 5 Turn off the N supply and close both valves 6 Weigh the apparatus including the liquid and calculate the sample weight by subtracting the step 1 weight from this weight 7 Insert a syringe through the rubber septum and add a known weight of H O to the sample Continue stirring until the water is completely blended with the liquid 8 Record the dew point indicated by the hygrometer and calculate the moisture content as follows _ Weight of water 10 1 13 total weight of liquid d l ppm 9 Repeat steps 6 8 until samples with several different moisture contents have been analyzed Note The accuracy of this technique can be checked at any point by withdrawing a sample and performing a Karl Fischer titration Be aware that this will change the total liquid weight in calculating the next point Application of the Hygrometer A 33 August 2004 Additional Notes for Liquid Applications In addition to the topics already discussed the following general application notes pertain to the use of GE Infrastructure Sensing m
68. oisture input optional Analog Output single isolated recorder output for dew point internally optically isolated 10 bit 0 1 resolution 0 2 V 10 kQ minimum load resistance 0 20 mA 400 maximum series resistance 4 20 mA 400 maximum series resistance Outputs are user programmable within the range of the instrument and the corresponding probe Alarm Relays 1 fault alarm and 2 programmable high low alarms Form C SPDT Standard Hermetically Sealed Relays 8A 9 250VAC 0 3A 115VAC 8A 30VDC 2A 28VDC Standard and hermetically sealed designs are available for the high low alarms set to trip at any level within the range of the instrument programmable from the front panel The fault alarm is the same type as the high low alarms Note 70 maintain Low Voltage Directive Compliance EN Standard EN61010 the following rating applies 2A 28VDC Alarm Setpoint Repeatability 0 1 C dew point Configurations panel mount bench mount PC board Display 1 line 6 digit Liquid Crystal Display LCD with programmable backlight Front Panel weatherproof membrane front panel display keypad meets NEMA 4 and IP66 requirements panel mount version only Specifications 5 1 August 2004 ss Electronics cont Display Functions dew point temperature C F MH Input Power 100 120 230 240 VAC 50 60 Hz 10 24 VDC Power Dissipation 100 120 230 240 VAC units 10 VA maximum 24 VDC units 10 W maxi
69. oisture probes in liquid applications 1 All M Series Aluminum Oxide Moisture Probes can be used in either the gas phase or the liquid phase However for the detection of trace amounts of water in conductive liquids for which an empirical calibration is required the M2 Probe is recommended Since a background signal is caused by the conductivity of the liquid between the sensor lead wires use of the M2 Probe which has the shortest lead wires will result in the best sensitivity The calibration data supplied with GE Infrastructure Sensing Moisture Probes is applicable to both liquid phase for those liquids in which a Henry s Law analysis is applicable and gas phase applications As indicated in Table A 2 on page A 5 the flow rate of the liquid is limited to a maximum of 10 cm sec Possible probe malfunctions and their remedies are discussed in previous sections of this appendix A 34 Application of the Hygrometer August 2004 Solids Applications In Line Measurements GE Infrastructure Sensing moisture probes may be installed in line to continuously monitor the drying process of a solid Install one probe at the process system inlet to monitor the moisture content of the drying gas and install a second probe at the process system outlet to monitor the moisture content of the discharged gas When the two probes read the same or close to the same dew point the drying process is complete For example a
70. on of the Hygrometer August 2004 Cleaning the Moisture Probe To clean the moisture probe the following items are required approximately 600 ml of reagent grade hexane or toluene divided into two batches of 300 ml each approximately 300 ml of distilled NOT deionized water three glass NOT metal containers to hold the above liquids To clean the aluminum oxide moisture probe complete the following steps 1 Record the dew point of the ambient air IMPORTANT To avoid damaging the sensor during the following steps do not allow the sensor to contact the walls or the bottom of the containers 2 Carefully remove the protective shield covering the sensor without touching the exposed sensor see Figure A 1 on page A 10 3 Soak the sensor in one of the containers of hexane or toluene for 10 minutes 4 Remove the sensor from the hexane or toluene and soak it in the container of distilled water for 10 minutes 5 Remove the sensor from the distilled water and soak it in the second container of clean hexane or toluene for 10 minutes 6 Remove the sensor from the hexane or toluene and place it sensor side up in an oven set at 50 C 2 C 122 F 3 6 F for 24 hours 7 Repeat Steps 3 6 to clean the protective shield To ensure the removal of any contaminants that may have become embedded in the porous walls of the shield swirl the shield in the solvents during the soaking procedure 8 Without touching th
71. probes The moisture content of corrosive gases such as H2S SO cyanide containing gases acetic acid vapors etc can be measured directly IMPORTANT Since the active sensor is aluminum any fluid which corrodes aluminum will affect the sensor s performance By observing the following precautions the moisture probe may be used successfully and reliably 1 The moisture content of the corrosive fluid must be 10 ppm or less at 1 atmosphere or the concentration of the corrosive fluid must be 10 ppm or less at 1 atmosphere 2 The sample system must be pre dried with a dry inert gas such as nitrogen or argon prior to introduction of the fluid stream Any adsorbed atmospheric moisture on the sensor will react with the corrosive fluid to cause pitting or corrosion of the sensor 3 The sample system must be purged with a dry inert gas such as nitrogen or argon prior to removal of the moisture probe Any adsorbed corrosive fluid on the sensor will react with ambient moisture to cause pitting or corrosion of the sensor 4 Operate the sample system at the lowest possible gas pressure Using the precautions listed above the moisture probe may be used to successfully measure the moisture content in such fluids as hydrochloric acid sulfur dioxide chlorine and bromine Application of the Hygrometer A 11 August 2004 Materials of Construction M1 and M2 Probes Sensor Element Back Wire Contact Wire Fron
72. rding to the instructions in the appropriate sections of this chapter To exit the User Program at any time 1 Press ESC repeatedly until the display reads run 2 Press ENTER The MTS 5 then returns to taking measurements Operation amp Programming 3 3 August 2004 Navigating Through the Menus ALA page 3 14 ALF page 3 13 User 1 page 3 7 Main menu Userl menu User2 menu The MTS 5 User Program is divided into three sections The contents of these menus is summarized in Table 3 2 below Table 3 3 on page 3 5 and Table 3 4 on page 3 6 Also there is a complete map of the User Program in Appendix C Menu Maps Table 3 2 The Main Menu Main Menu Item Submenu Item Sub Submenu Item set up Alarm A ALF En set up Fault Alarm ALF di enter Userl Menu C F MH disable Alarm A enable Fault Alarm disable Fault Alarm enter passcode Low Alarm High Alarm N A see Table 3 3 on page 3 5 User2 page 3 8 enter User2 Menu enter passcode see Table 3 4 on page 3 6 SEtUP page 3 25 Factory Setup enter passcode 3 4 Operation amp Programming August 2004 Table 3 3 The User1 Menu User1 Menu Item Submenu Item Sub Submenu Item S No page 3 11 enter serial number N A N A A N A dP enter dew point dP LO enter lowest dew point page 3 10 range dPHI enter highest dew point N A ProbE ent
73. re Sensing hygrometer d From the literature it is found that Cs for hexane at a temperature of 20 C is 101 ppm Using a straightedge on Figure A 3 on page A 26 connect the 101 ppm saturation concentration with the 20 C ambient temperature and read the Henry s Law Constant of 5 75 on the center scale Using the straightedge connect the above K value of 5 75 with the measured dew point of 0 C and read the correct moisture content of 26 ppm where the straightedge crosses the moisture content scale 4 Find the moisture content in an unknown organic liquid at an ambient temperature of 50 C if a dew point of 10 C is measured with the GE Infrastructure Sensing hygrometer d Either perform a Karl Fischer analysis on a sample of the liquid or obtain a dry sample of the liquid Either use the ppm determined by the Karl Fischer analysis or add a known amount of water i e 10 ppm to the dry sample Measure the dew point of the known test sample with the GE Infrastructure Sensing hygrometer For purposes of this example assume the measured dew point to be 10 C Using a straightedge on the nomograph in Figure A 3 on page A 26 connect the known 10 ppm moisture content with the measured dew point of 10 C and read a K value of 5 1 on the center scale Using the straightedge connect the above K value of 5 1 with the measured 10 C dew point of the original liquid and read the actu
74. ries 5 is designed to be maintenance and trouble free However because of severe process conditions and other factors minor problems may occur from time to time Some of the most common problems and recommended maintenance procedures are discussed in this chapter If you can not find the information you need in this chapter please consult GE Infrastructure Sensing for help Caution Do not attempt to troubleshoot the MTS 5 beyond the instructions in this chapter If you do you may damage the unit and or void the warranty This chapter covers the following topics common problems page 4 2 replacing the User Program page 4 4 replacing recalibrating moisture probes page 4 6 Proceed to the appropriate section to perform any of the above tasks Service and Maintenance 4 1 August 2004 Common Problems If the MTS 5 measurements read too wet or too dry or if they do not make sense there may be a problem with either the probe or a process component Use the descriptions of common problems in Table 4 1 below to troubleshoot and solve such problems Table 4 1 Troubleshooting Guide for Common Problems Possible Cause Symptom The accuracy of the moisture sensor is questioned There is insufficient time for the system to equilibrate Response Reads too wet during dry down conditions or too dry in wet up conditions Change the flow rate A change in dew point indicates the sample syste
75. rms up in the User menu see page 3 7 To set up either of the high low alarms proceed as follows ALA Alarm A Enter the Main menu and press ENTER at the AL A screen ane Alarm B Enter the User menu Press s or t until the AL b screen appears Then press ENTER dEG C Press s or t until the desired setting dEG 2 C dEG F H or ALX di appears on the ERE display NOTE X A for Alarm A or X H B for Alarm B Then press ENTER ALX di If you disabled the alarm by selecting ALX di above go directly to the final screen on this page Otherwise continue as follows Note A low LO alarm trips when the specified parameter drops below the trip value a high HI alarm trips when the specified parameter goes above the trip value AL LO Press s or t until the desire alarm type LO or HI appears on the display Then AER press ENTER XXX Use the s and t keys to set the desired trip point Then press ENTER ALA You are returned to one of these displays To exit the User Program press ESC until BER run appears Then press ENTER To continue programming the MTS 5 proceed to the appropriate section for instructions 3 14 Operation amp Programming August 2004 Testing the Alarm Relays To test the alarm relays enter the User2 menu as described on page 3 8 Then proceed as follows unit Pre
76. robe to the factory for evaluation Non conductive material is trapped under contact arm of sensor Clean the probe as described in Appendix A page A 8 Then reinstall the sensor If the low reading persists return the probe to the factory for evaluation Improper cable connection Check the cable connections to both the probe and the hygrometer Response is slow Slow outgassing of system Replace the system components with stainless steel or electro polished stainless steel The sensor is contaminated with non conductive particles see Appendix A page A 6 Clean the probe as described in Appendix A page A 8 Then reinstall the sensor Service and Maintenance 4 3 August 2004 Replacing the User Program Removing the Main PC Board Replacing the PROM Chip The User Program is stored on a PROM Programmable Read Only Memory chip This chip is installed in a socket on the main PC printed circuit board located inside the MTS 5 electronics enclosure To replace the PROM chip complete the following steps r Turn the power off and disconnect the main power source to the instrument IWARNING You must disconnect the line power before opening the MTS 5 enclosure Discharge any static electricity from your body by touching a grounded metal object Caution PROM chips can be damaged by static electricity Open the MTS 5 enclosure by removing the four scre
77. s and is suitable for a wide range of process conditions that require real time moisture measurement The MTS 5 measures dew points from 80 to 20 C 112 to 68 F with data to 110 C 166 F It comes equipped with two standard alarm relays one fault alarm relay and a single analog output The MTS 5 displays measurement data on a one line 6 digit liquid crystal display LCD You can program your unit and enter probe information using the keypad on the front panel see Figure 1 1 below The MTS 5 accepts line voltages of 100 120 230 and 240 VAC or 24 VDC SERIES 5 DEW FROST POINT Figure 1 1 Front Panel The moisture probe is the part of the system that comes in contact with the process The MTS 5 uses any M Series probe to measure dew point temperature in C or F The sensor assembly is secured to the probe mount and protected with a sintered stainless steel shield see Figure 1 2 below Other types of shields are available Shield Sensor Figure 1 2 M Series Probe Features and Capabilities 1 1 Chapter 2 Installation Introduction aus ea 2 1 Mounting the Electronics Unit ccc cece eee een eee eee 2 1 Mounting the Sample System cc cece cc ccc cece eee eeee 2 2 Installing the ProDE u ner rn een 2 2 Making Wiring Connections 0 cece cece eee cee ee ee eens 2 4 August 2004 Introduction Installing the MTS 5 includes mounting the electron
78. s the weight fraction of the jh component Cs is the saturation concentration ppm of the jn component and n is the total number of components In conclusion the Henry s Law constant K is a constant of proportionality between the saturation concentration Cs and the saturation vapor pressure Ps of water at the measurement temperature In the General Case the Henry s Law constant varies with the measurement temperature but there is a Special Case in which the Henry s Law constant does not vary appreciably with the measurement temperature This special case applies to saturated straight chain hydrocarbons such as pentane hexane heptane etc General Case Determination of Moisture Content if Cs is Known The nomograph for liquids in Figure A 3 on page A 26 can be used to determine the moisture content in an organic liquid if the following values are known the temperature of the liquid at the time of measurement the saturation water concentration at the measurement temperature the dew point as measured with the GE Infrastructure Sensing hygrometer A 24 Application of the Hygrometer August 2004 General Case cont Complete the following steps to determine the moisture content from the nomograph 1 Using a straightedge on the two scales on the right of the figure connect the known saturation concentration ppm with the measurement temperature C 2 Read the Henry s Law constant K on the
79. sample cell fitting making sure not to cross the threads 3 Tighten the probe securely Figure 2 2 below shows a typical probe installed in a sample cell Caution For maximum protection of the aluminum oxide sensor the stainless steel end cap should always be left in place Sample Cell Outlet Figure 2 2 Probe Installed in Sample Cell Installation 2 3 November 2004 Making Wiring Wiring the MTS 5 includes the following procedures Connections installing the power cable connecting the probe connecting the recorder output connecting the alarms IWARNING To ensure safe operation the MTS 5 must be installed and operated as described in this manual In addition be sure to follow all applicable local safety codes and regulations for installing electrical equipment Installing the Power Cable To install the power cable included with the MTS 5 simply plug the female connector end into the male connector on the rear panel see Figure 2 3 below Figure 2 3 MTS 5 Rear Panel 2 4 Installation November 2004 Connecting the Probe The probe must be connected to the MTS 5 with a continuous run of GE Infrastructure Sensing two wire shielded cable see Figure 2 4 below When connecting the probe be sure to follow these guidelines for cable use protect the cables from excessive strain bending pulling etc do not subject the cables to temperatures above 65 C 149 F or
80. ses ee te ee abes Sona VR A 5 Gases nut aan lan a ha dr A 13 Liquids a ae lee A 5 e E ee A23 FuseRating 0 0 eee eee 5 2 SOUdS es seee a a A 35 G B Gases Backlight Display 3 11 Flow Rates cene A 5 Bench Mount Enclosure B 2 Board Mount Version B 3 H High Low Alarms 004 2 9 C Cable Proben 2 5 Calculations 0 0 00 eee eee Als Interconnection Diagram 0000 B 4 Calibration Data Sheet o zotr e RESTER 4 6 K Empirical eol REDDERE A 30 Entering Curve 0055 3 02 Keypad unse er 3 2 Probe ee eset aoa nee ra 4 6 Certification NEMA A IP66 2 1 5 1 L D es ee Display Default Settings Restoring 3 24 Liquids Dew Point Applications 0 00008 A 23 Offset m heehee dak ees 3 12 Flow Rates 0 00 00 0000 A 5 R ngen 1 1 5 3 Low Voltage Directive 3 1 5 2 Dimensions 0 020s 5 2 Display Selecting Units 0000 3 9 Setting Backlight 3 11 Specifications 04 1 1 5 1 POSTING aussen led 3 23 Index November 2004 Index cont M P Menu Panel Mount Enclosure B 1 Entering Userl 22222222020 3 7 Passcodes nass es ens 3 7 3 8 Entering User2 22222202 3 8 Power Main ee 3 4 B 1 Rating is 222 ea 5 2 P ssc des u 2 4202 3 7 3 8 Powering Down
81. ss s or t until tESt appears on the display tESt Press ENTER t AL Press ENTER Note At the following display alarms A and B are shown as being tested even if they have been disabled ALA On Alarm A tripped Press t ALA OF Alarm A reset Press t abon Alarm B tripped Press t ALb OF Alarm B reset Press t ALF On Alarm F tripped Press t ALF OF Alarm F reset Press t At the above screen the test modes are cycled through continuously in the order indicated ALX To terminate the testing loop at any of the above screens press ESC t AL To exit the User Program press ESC until run appears Then press ENTER To continue programming the MTS 5 proceed to the appropriate section for instructions Operation amp Programming 3 15 August 2004 ss Using the Recorder The MTS 5 is equipped with one recorder output This output may be Output configured for 0 20 mA 4 20 mA or 0 2 V In addition the recorder output may be tested and adjusted Proceed to the appropriate section for specific instructions Setting Up the Recorder Before the recorder output can be used the following parameters Output must be programmed units to be recorded type current or voltage and range of output low end minimum of the recorder output range high end maximum of the recorder output range To set up the
82. system of this type has been used successfully to monitor the drying of photographic film If one wishes to measure the absolute moisture content of the solid at any time during such a process then an empirical calibration is required 1 Ata particular set of operating conditions i e flow rate temperature and pressure the hygrometer dew point reading can be calibrated against solids samples with known moisture contents 2 Assuming the operating conditions are relatively constant the hygrometer dew point reading can be noted and a solids sample withdrawn for laboratory analysis 3 Repeat this procedure until a calibration curve over the desired moisture content range has been developed Once such a curve has been developed the hygrometer can then be used to continuously monitor the moisture content of the solid as long as operating conditions are relatively constant Application of the Hygrometer A 35 August 2004 Laboratory Procedures If in line measurements are not practical then there are two possible laboratory procedures The unique ability of the GE Infrastructure Sensing probe to determine the moisture content of a liquid can be used as follows Using the apparatus shown in Figure A 4 on page A 31 dissolve a known amount of the solids sample in a suitable hydrocarbon liquid The measured increase in the moisture content of the hydrocarbon liquid can then be used to calculate the moisture content of
83. t que les Moisture Target Series 5 Hygrometer r latif cette d claration sont en conformit avec les documents suivants e EN 61326 1998 Class A Annex A Continuous Unmonitored Operation EN 61010 1 1993 A2 1995 Overvoltage Category II Pollution Degree 2 suivant les r gles de la Directive de Compatibilit Electromagn tique 89 336 EEC et de la Directive Basse Tension 73 23 EEC Les mat riels list s ci dessus ainsi que les capteurs et les systemes d chantillonnages pouvant tre livr s avec ne portent pas le marquage CE de la directive des quipements sous pression car ils sont fournis en accord avec la directive 97 23 EC des quipements sous pression pour les DN 25 Article 3 section 3 qui concerne les pratiques et les codes de bonne fabrication pour l ing nierie du son Shannon June 1 2002 p Fu ND Mr James Gibson DIRECTEUR G N RAL NSAI LS ENISO9002 T V ESSEN Shannon ISO 9001 U S CERT DOC H2 August 2004 Sensing KONFORMITATS ERKLARUNG Wir Panametrics Limited Shannon Industrial Estate Shannon County Clare Ireland erkl ren in alleiniger Verantwortung da die Produkte Moisture Target Series 5 Hygrometer folgende Normen erf llen e EN 61326 1998 Class A Annex A Continuous Unmonitored Operation e EN 61010 1 1993 A2 1995 Overvoltage Category II Pollution Degree 2 gem den Europ ischen Richtlinien Niederspannungsrichtlinie Nr 73 23 EG und EMV Richtlinie Nr 89 3
84. t Wire Support Shell Shield O Ring Electrical Connector Pins Glass O Ring 99 99 aluminum aluminum oxide gold Nichrome A6 316 stainless steel gold 304 stainless steel 316 stainless steel Corning 9010 glass 303 stainless steel 304 stainless steel Vitron 302 Stainless Steel Corning 9010 silicone rubber Application of the Hygrometer August 2004 Calculations and Useful A knowledge of the dew point of a system enables one to calculate all Formulas in Gas other moisture measurement parameters The most important fact to Applications recognize is that for a particular dew point there is one and only one equivalent vapor pressure IMPORTANT The calibration of GE Infrastructure Sensing moisture probes is based on the vapor pressure of liquid water above 0 C and frost below 0 C GE Infrastructure Sensing moisture probes are never calibrated with supercooled water Caution is advised when comparing dew points measured with a GE Infrastructure Sensing hygrometer to those measured with a mirror type hygrometer since such instruments may provide the dew points of supercooled water As stated above the dew frost point of a system defines a unique partial pressure of water vapor in the gas Table 3 on page 18 which lists water vapor pressure as a function of dew point can be used to find either the saturation water vapor pressure at a known temperature or the water vapor pressur
85. tion and Orientation Service and Maintenance 4 5 August 2004 Replacing Recalibrating Moisture Probes For maximum accuracy moisture probes should be returned to the factory for recalibration every 6 12 months depending on the application Under very severe conditions more frequent calibrations are recommended under very mild conditions less frequent calibrations are necessary Contact a GE Infrastructure Sensing applications engineer for a specific recommended calibration frequency All new or recalibrated moisture probes must be installed in accordance with the instructions presented in Chapter 2 Installation IMPORTANT To maintain good contact at each terminal block and to avoid damaging the pins on the connector pull the connector straight off not at an angle make cable connections while the connector is away from the unit and push the connector straight on not at an angle when the wiring is complete After the probe has been installed and wired enter the probe calibration curve data as described on page 3 22 Each probe is shipped with its own Calibration Data Sheet which includes the serial number for that probe 4 6 Service and Maintenance Chapter 5 Specifications Electronics Moisture Measurement August 2004 ee ____ Electronics Input moisture signal from GE Infrastructure Sensing thin film aluminum oxide moisture sensor Intrinsic Safety external safety barrier for m
86. to 110 C Accuracy 2 C from 65 to 60 C 3 C from 110 to 66 C Repeatability 0 5 C from 65 to 60 C 1 0 C from 110 to 66 C Specifications 5 3 Appendix A Application of the Hygrometer Introduction su a au Do e i la Dd A 1 Moisture Monitor Hints 0 0 cece ccc eee eee nennen A 2 Contaminants s sua ot trate en eae MI Per A 6 Aluminum Oxide Probe Maintenance ceee eee e eens A 8 Corrosive Gases And Liquids eee eee eee eee eens A 11 Materials of Construction a Zu A 12 Calculations and Useful Formulas in Gas Applications A 13 Liquid Applications ea A 23 Empirical Calibrations se nen ee ee A 30 Solids Applications 2 0 02 0 ee wees A 35 August 2004 Introduction This appendix contains general information about moisture monitoring techniques System contaminants moisture probe maintenance process applications and other considerations for ensuring accurate moisture measurements are discussed The following specific topics are covered Moisture Monitor Hints Contaminants Aluminum Oxide Probe Maintenance Corrosive Gases and Liquids Materials of Construction Calculations and Useful Formulas in Gas Applications Liquid Applications Empirical Calibrations e Solids Applications Application of the Hygrometer A 1 August 2004 Moisture Monitor Hints GE Infrastructure Sensing hygrometers using aluminu
87. to dislodge the flexible jumper 4 Locate switch block S1 shown in Figure 2 6 on page 2 8 5 Set switch S1 in the appropriate position I for current or V for voltage 6 After you set the switch slide the main PC board back into the enclosure and fasten the screws on the back cover Connect the recorder output to terminal block TB2 on the back of the MTS 5 as shown in Figure 2 5 on page 2 6 IMPORTANT To maintain good contact at each terminal block and to avoid damaging the pins on the connector pull the connector straight off not at an angle make cable connections while the connector is away from the unit and push the connector straight on not at an angle when the wiring is complete Installation 2 7 November 2004 Connecting the Recorder Output cont uad TO E e C20 R22 C52 m mem j Oes eo Oe O Switch S1 oO oooOO00O03O29000900 Figure 2 6 Switch S1 on the Main PC Board 2 8 Installation November 2004 Connecting the Alarms Connecting the Fault Alarm Connecting High Low Alarms A and B The MTS 5 has one fault alarm and two high low alarms Hermetically sealed alarms are available Each a
88. urrent passcode displayed enter new passcode uSEr2 page 3 8 The User1 and User2 change User2 passcode current passcode displayed enter new passcode The instructions that follow show how to enter the User menu and Menus the User2 menu To prevent tampering with critical parts of the User Program a passcode security feature protects the User and User2 menus factory set default passcodes are 0 for User1 and 1 for User2 For greatest security set new passcodes as described in the following sections 3 6 Operation amp Programming August 2004 ee ___ Entering the User1 Menu Enter the MTS 5 User Program as described on page 3 3 ALA Press s or t until USEr1 appears on the display USEr1 Press ENTER Note The MTS 5 comes from the factory with the Userl passcode set to 0 If desired change the passcode as instructed below 0 Use the s and t keys to enter the passcode Then press ENTER S No You are now in the User menu Changing the User1 To change the User passcode continue as follows Passcode S No Press s or t until uSEr1 appears on the display uSEr1 Press ENTER XX Use the s and t keys to enter a new passcode Then press ENTER to save the new passcode or press ESC to keep the old passcode uSEr1 To exit the User Program press ESC until run appears Then press
89. ws on the back of the enclosure IMPORTANT Do not dislodge or disconnect the flexible jumper that connects the display to the main PC board Remove the board from the enclosure only far enough to replace the PROM chip Slowly slide the main PC board from the enclosure by pulling straight back being careful not to dislodge the flexible jumper Refer to Figure 4 1 on page 4 5 to locate the PROM chip on the main PC board The PROM chip socket is labeled as U19 Usea chip puller to remove the PROM chip from its socket Caution PROM chips can be damaged by static electricity Observe anti static precautions before proceeding Place the new PROM chip in the socket labeled U19 making sure that the beveled corner on the PROM chip matches the beveled corner on the socket see Figure 4 1 on page 4 5 Gently seat the new PROM chip completely into the socket 4 4 Service and Maintenance August 2004 Re installing the Main PC 1 Carefully slide the main PC board back into the enclosure Board 2 Replace the back panel on the enclosure and reinstall the four screws Do not overtighten the screws Checking the New User 1 Power up the MTS 5 see page 3 1 Program 2 Check to make sure the calibration and reference data are not corrupted see Chapter 3 Operation amp Programming If any data is corrupted re enter the data as described in Chapter 3 c pes w 703 1251 Figure 4 1 PROM Loca

Download Pdf Manuals

image

Related Search

Related Contents

Manual de instrucciones - HBSO  取扱説明書  GlassedPad  460MCs-R - フェローズジャパン  Puritron Technology Inc      Makita AC310H Use and Care Manual  

Copyright © All rights reserved.
Failed to retrieve file