1. 10 With all four 4 mounting holes drilled re align the back shell with the band clamps and install the mounting hardware screws washers outer plates through the band clamps and through the new mounting holes as indicated below 7 n Back Shell Installation le aoe Outer Plate DL yD Lol me E ee 5 0 Mx1SuperLite Mount Installation 11 Working from the inside of the back pan secure the mounting hardware screws into place via the threaded backing plates provided 8 12 Fine tune height adjustments can be made prior to fully tightening the mounting screws or by loosening the mounting screws and sliding the cane clamps up or down as needed along the back canes 9 fe aaa Screws are secured into the threaded backing plates f Backing Plate Rear View NOTE Prior to tightening hardware minor height adjustments can be made if necessary by adjusting the cane clamps up or down on the back canes Front View NOTE Height adjustments made post installation may be subject to limitations depending on the location of the rigidizer bar relative to the band clamps 13 Verify that all mounting hardware is fully tightened and re attach the outer cushion cover onto the back shell If you have any questions or concerns when installing your Invacare Matrx Mx1 SuperLite Mount Back please contact our Technical Sevice Department for assistance Canada 800 680 4191 USA 888 433 6
2. 2 Back Cushion 4 Installation User Manual DEALER This manual contains important safety Installation and Set up information specific to the Invacare Matrx Mx1 SuperLite Mount Back This Manual MUST be given to the user of the product BEFORE using this product please read through the instructions carefully and save for future reference TRD0317 00 This product complies with Directive 93 42 EEC concerning medical devices The launch date of this product is stated in the CE declaration of conformity 2012 Motion Concepts All rights reserved Republication duplication or modification in whole or in part is prohibit ed without prior written permission from Motion Concepts Trademarks are identified by and All trademarks are owned by or licensed to Motion Concepts or its subsidiaries unless otherwise noted Table of Contents s WON A606 General Symbols Overview Limited Warranty Safety General Guidelines Intended Use Installation Information Operating Information Weight Limitations and Model Numbers Installation Overview Tools Required Mx1 SuperLite Mount Installation Installation Instructions Options Installing Adjusting the Lumbar Support Plug Inserts Adhesive Pads Installing Optional Privacy Flap Maintenance Cleaning Inspection Re Use matr Yes you can 1 0 General 1 1 Symbols Installing your Invacre Matrx Seating Series product safely dep
3. 16 in 41 cm IMX11609 SL IMX11612 SL IMX11616 SL 250 Ibs 113 kg 17 in 43 cm IMX11709 SL IMX11712 SL IMX11716 SL 18 in 46 cm IMX11809 SL IMX11812 SL IMX11816 SL 3 0 Tools Required Amm hex key provided Tape measure ruler Quick Grip Clamps x2 or similar Grease Pen Marker Power Drill Drill Bits 1 8 3 0mm 1 4 or 17 64 or 6 5mm Fine Grit Sandpaper 4 0 Installation Overview i Important Before beginning the installation process please fully read through all the instructions to understand the steps and adjustments involved If you have any questions or concerns during the installation process please contact our Technical Service Department for assistance Band Clamp Assembly e a Mx1 Back m Shell 8 5 0 Mx1SuperLite Mount Installation Please read through all installation instructions before proceeding with the Mx1 SuperLite Back installation IMPORTANT The mounting hardware band clamps for Mx1 SuperLite Back are designed for installation onto 1 26mm back canes only WARNING DO NOT install the Invacare Matrx MX1 assembly onto back canes with an outside diameter greater than or less than 1 26mm Otherwise injury or damage may occur Installation of the Mx1 SuperLite Back requires holes to be drilled into the back shell We strongly recommend that the installation be completed by Qualified Technician The Mx1 Back is designed for a one time set up on
4. you can Concepts Motion Invacare USA Elyria Ohio 44036 2125 Tel 440 329 6000 Tel 800 333 6900 Motion Concepts USA 700 Ensminger Rd Suite 112 Tonawanda NY 14150 Tel 888 433 6818 www invacare com Canada 570 Matheson Blvd E Unit 8 Mississauga Ontario L4Z 4G4 Tel 800 668 5324 www motionconcepts com Canada 84 Citation Drive Concord Ontario L4K 3C1 Tel 866 748 7943 Tel 905 695 0134 CE TRD0317 00
5. Laundering 1 Release the hook and loop fasteners securing the back cushion cover The hook and loop fasteners are located on the top and bottom edge and inside of the back shell 2 Remove the foam cushion Remove the lumbar support Refer to Installing Adjusting the Lumbar Support on page 28 Remove the cover from the back shell Follow the washing instructions on the cover 3 4 5 6 Allow the cover to fully dry 7 Insert the foam cushion into the cover 8 Ensure the foam cushion is fully inserted and the perimeter of the foam is aligned with the cover seams 9 Install the cover onto the back shell 10 Install the lumbar support Refer to Installing Adjusting the Lumbar Support on section 6 1 7 2 Inspection Visually inspect parts including hardware brackets upholstery materials foams if accessible and plastics for deformation corrosion breakage wear or compression WARNING Check ALL fasteners weekly to ensure that mechanical connections and attaching hardware are tightened securely otherwise injury or damage may occur DO NOT continue to use this product if problems are discovered Corrective maintenance can be performed at or arranged through your equipment supplier 7 3 Reuse WARNING Risk of Injury Always have a dealer inspect the product for damage before transferring the product to a different user If any damage is found DO NOT use the product Otherwise injury may occur 454 Yes
6. information for the safe operation and use of this product di IMPORTANT Please refer to the Wheelchair Owners Operators Manual for additional safety information warnings and cautions related to the operation of your wheelchair 2 1 General Guidelines WARNING DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner s manuals service manuals or instruction sheets supplied with this product or optional equipment If you are unable to understand the warnings cautions or instructions contact a healthcare professional dealer or technical personnel before attempting to use this equipment otherwise injury or damage may occur ACCESSORIES WARNING I Invacare products are specifically designed and manufactured for use in conjunction with Invacare acces sories Accessories designed by other manufacturers have not been tested by Invacare and are not recom mended for use with Invacare products i NOTICE THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE Check all parts for shipping damage before using In case of damage DO NOT use the equipment Contact the Dealer for further instructions 2 2 Intended Use The Invacare Matrx MX1 SuperLite back is designed to provide precise orientation within the wheelchair for optimal postural support 2 3 Installation Inform
7. sold separately for use with all standard sized MX1 backs All standard sized backs come with a pre sewn loop strip across the bottom rear of the cushion cover B for installation of the privacy flap Refer to the chart for proper sizing of the privacy flap for the MX1 back WIDTH PRIVACY FLAP 14 in 36 cm PFLAP XS 10 x 16 26 cm x 41 cm 15 in 38 cm PFLAP S 16 in 41 cm 13 x 16 33 cm x 41 cm I7 in 43 cm PFLAP M 18 in 46 cm 15 x 16 38 cm x 41 cm If installed remove the seat cushion Ensure the back is adjusted to the desired height Refer to section 5 1 Installation Instuctions Position the wheelchair in front of you so the back is facing you Align the hook strip A on the privacy flap with the loop strip sewn on to the bottom rear of the back cover B Pull the opposite end of the privacy flap through the gap between the back and the wheelchair seat C O 1 AaWN a Secure the loop strip D on the opposite end of the privacy flap to the fastening strip on the wheelchair seat NOTE The fastening strip pre installed on the wheelchair seat pan or sling upholstery may be either a hook strip or a loop strip It may be necessary to install the the double sided hook strip E onto the privacy flap to secure it to the seat ea ey REAR VIEW 14 7 0 Maintenance 7 1 Cleaning General Cleaning 1 Wipe the outer cover with a damp cloth when necessary
8. 818 00 6 0 Options The optional Lumbar Support is a pre fabricated foam insert that provides lower back support for additional comfort The position of the lumbar insert can be adjusted inside the Mx1 cushion cover or it may be removed if no lumbar support is desired 6 1 Lumbar Support Installation Adjustment 1 To install or access the lumbar support open the outer cover at the bottom rear of the back shell A Unfasten velcro to install access the lumbar support Lumbar Support FRONT Foam Cushion 2 Install adjust the foam lumbar support inside the Mx1 cover between the foam back cushion and the back shell See image B and C below 3 The lumbar support can be manually adjusted to any desired position or it can be removed completely if no lumbar support is desired Adjust lumbar foam as needed or remove completely 6 2 Plug Inserts Optional plug inserts Model DP 250 1 may be used to fill unused mounting holes 2 on the Mx1 back shell a Plug Inserts 13 6 0 Options 6 3 Adhesive Pads Adhesive pads 1 are available to protect the Mx1 back shell from scratches 1 Peelthe protective paper off of the back of the adhesive pad 2 Install the adhesive pad between the Mx1 back shell and the back cane 2 or other wheelchair component that may rub against the back shell Adhesive Pads O 6 4 Installing Optional Privacy Flap Privacy flaps 1 are available
9. M NORMAL WEAR OR FAILURE TO ADHERE TO THE ENCLOSED INSTRUCTIONS IN ADDITION THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED PRODUCTS SUBJECTED TO NEGLIGENCE ACCIDENT IMPROPER OPERATION MAINTENANCE OR STORAGE OR PRODUCTS MODIFIED WITHOUT INVACARE S MOTION CONCEPTS EXPRESS WRITTEN CONSENT INCLUDING BUT NOT LIMITED TO MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE MOTION CONCEPTS PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND INVACARE S MOTION CONCEPTS CONTROL PRODUCTS REPAIRED BY ANYONE OTHER THAN AN INVACARE MOTION CONCEPTS DEALER SUCH EVALUATION SHALL BE SOLELY DETERMINED BY INVACARE MOTION CONCEPTS THE FOREGOING WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER EXPRESS WARRANTIES IF ANY INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE IT SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESSED WARRANTY PROVIDED HEREIN AND THE REMEDY FOR VIOLATIONS OF ANY IMPLIED WARRANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN INVACARE MOTION CONCEPTS SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE PROVINCIAL LAWS AND REQUIREMENTS 2 0 Safety The Safety section contains important
10. at a cushion choice matches your individual needs As the needs of the individual become more complex cushion evaluation becomes more important Invacare Matrx MX1 The Invacare Matrx MX1 back is contoured to provide precise orientation within the wheelchair for optimal postural support The foam is oversized to provide extra comfort and protection The Invacare Matrx MX1 back includes a movable foam lumbar pad that can be installed inserted behind the existing foam cushion to provide additional positioning capability Refer to Section 6 1 for instructions on how to insert and adjust your lumbar support Outer Cover The outer cover is made of a mesh material that is moisture resistant and breathable Regular cleaning and inspection of the outer cover isrecommended Refer to Section 7 1 for Cleaning Instructions 1 0 General 1 3 Limited Warranty North America i PLEASE NOTE THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4 1975 This warranty is extended only to the original purchaser user of our products This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state Invacare Motion Concepts warrants this product to be free from defects in materials and workmanship for two years of use by original purchaser This warranty does not apply to punctures tears or burns nor to the removable cushion cover If within s
11. ation WARNING 2 The procedures in this manual should be performed by a qualified technician After any adjustments repair or service and before use make sure that all attaching component parts are secure DO NOT install the Invacare Matrx MX1 assembly onto back canes with an outside diameter greater than or less than 1 inch Otherwise injury or damage may occur The mounting position of the Invacare Matrx MX1 SuperLite back is directly related to the chair s stability Use extreme caution when using a new seating position Ensure the Invacare Matrx MX1 SuperLite Back is properly secured to the wheelchair before using Otherwise injury or damage may occur 2 4 Operating Information WARNING Skin condition should be checked very frequently after the installation of any new seating system Your therapist and physician should be consulted if you have any questions regarding individual limitations and needs Working with your therapist physician and equipment supplier is the best way to assure that a seating choice matches your individual needs As the needs of the individual become more complex the seating evaluation becomes more important 2 0 Safety 2 5 Weight Limitation and Model Numbers Refer to the chart for the weight limitation and stock model numbers SL SuperLite HEIGHT WIDTH Sin 23cm Tree i6in 4lamn WEIGHT CAPACITY 14 in 36 cm IMX11409 SL IMX11412 SL IMX11416 SL 15 in 38 cm IMX11509 SL IMX11512 SL IMX11516 SL
12. ends upon your diligence in following the warnings cautions and information provided in this Installation Manual Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage See the information below for definitions of the signal words WARNING Warning indicates a potentially hazardous situation which if not avoided could result in death or serious injury CAUTION Caution indicates a potentially hazardous situation which if not avoided may result in property dam age or minor injury or both IMPORTANT Important information and or instructions to remember when installing and or adjusting your Invacare Matrx Seating Series product NOTE e Notes indicate useful tips recommendations and information for efficient trouble free use Requirements e This symbol identifies a list of various tools components and items which you will need in order to carry out certain work READ WELL BEFORE OPERATION e This symbol advises you to read information carefully 1 2 Overview Important Information The best way to avoid problems related to pressure sores is to understand their causes and your role ina skin management program Your therapist and physician should be consulted if you have questions regarding individual limitations and needs All cushions should be selected carefully Working with your therapist and physician is the best way to assure th
13. ll in place amp mark all 4 mounting holes on the back shell Le Important Ensure the back shell is centered and horizontal prior to marking and drilling 10 5 0 Mx1SuperLite Mount Installation 7 With the mounting holes clearly marked remove the back shell from the wheelchair and carefully secure stabilize the back shell so that the mounting holes can be safely drilled 8 Using a power drill and a 1 8 3 0mm drill bit carefully drill a pilot hole at each of the 4 mounting locations marked on the MX1 back shell All holes should be drilled perpendicular to the surface of the back shell 9 Using a larger drill bit in the range of 1 4 up to 17 64 or 6 5mm drill out each of the 4 pilot holes to produce the final mounting locations 6 AN CAUTION After drilling use fine sandpaper or file to remove any rough edges or fibres around the holes NOTE the final mounting holes are intentionally oversized from the installation hardware screws in order to allow a small amount of tolerance around the drilled hole locations IMPORTANT To ensure a Drill Bit more accurate hole position use 1 4 up to 17 64 drill a pilot hole first using a or 6 5mm smaller 1 8 3 0mm drill bit prior to drillling out the final mounting hole FT cis wn e e a WARNING Ensure the back shell is properly stabilized and keep hands and fingers clear of the drill bit when drilling to prevent injury
14. ly Take care to ensure the accuracy of the hole locations prior to drilling NEVER use install a back that has had the holes modified or has been re drilled i e has more than 4 holes 5 1 Installation Instructions 1 Remove the existing backrest and any existing mounting hardware from the wheelchair 2 Install the band clamps on the wheelchair back canes 1 by snapping them onto the back cane tube as indicated in the image below 3 Adjust slide each band clamp along the back cane to the desired approximate mounting height and temporarily secure the band clamps into place using a set of quick grip clamps or similar 2 IMPORTANT Ensure that both band clamps are installed at the same height on the back canes Use a tape measure ruler to measure the height from a fixed point on the wheelchair Quick Grip nile Snap each band clamp onto its back cane and adjust to desired height Clamp both band Rigidizer Bar clamps into place ensuring they are adjusted to the same height i NOTE Rubber or padded quick grip clamps are recommended to prevent accidental scratching of the band clamps or the back Shell during the installation process 1 Back Cane 9 5 0 Mx1SuperLite Mount Installation 4 Remove the outer cushion cover from the back shell by carefully disengaging the velcro strips along the bottom and top of the back shell 5 Carefully place the back shell up against the mounting cla
15. mps and adjust the back shell to the desired approximate height on the wheelchair O IMPORTANT The back height should be set as accurately as possible during the initial hardware installation The overall height of the back rest can be re adjusted following installation however limitations may occur depending on the location of the rigidizer bar relative to the band clamps WARNING Prior to marking the back shell ensure that all mounting holes are located a minimum of 1 26mm from the outer edge of the back shell and that the mounting holes are located on the bent face of the back as indicated by the shaded area identified below 3 Hardware Mounting A TA Drilling Zones NS Hardware Mounting Mx1 Back Shell Drilling Zones OVERHEAD VIEW Wee PA i WARNING drill hole location must be a minimum of 1 26mm REAR VIEW from the edge of the back shell SELE di e _ 6 Make certain that the back shell is properly centered between the back canes and level horizontal 4 Use the band clamps as a template and accurately mark the mounting hole locations x4 on the back shell using a grease pen or marker 5 NOTE Quick grip clamps may also be used to hold the back shell in position prior to marking the mounting hole locations Alternatively a second person may be needed to ensure the back shell does not move slip when marking the hole locations Hold secure the back she
16. uch warranty period any such product shall be proven to be defective such product shall be repaired or replaced at Invacare s Motion Concepts option with refurbished or new parts This warranty does not include any labor or shipping charges incurred in replacement part installa tion or repair of any such product Product repairs shall not extend this warranty coverage for repaired product shall end when this limited warranty terminates Invacare s Motion Concepts sole obligation and your exclusive remedy under this warranty shall be limited to such repair and or replacement For warranty service please contact the dealer from whom you purchased your Invacare Motion Concepts product In the event you do not receive satisfactory warranty service please write directly to Invacare Motion Concepts at the address on the back cover Provide dealer s name address model num ber the date of purchase indicate nature of the defect and if the product is serialized indicate the serial number Invacare Corporation Motion Concepts will issue a return authorization The defective unit or parts must be returned for warranty inspection using the serial number when applicable as identification within thirty 30 days of return authorization date DO NOT return products to our factory without our prior consent C O D shipments will be refused please prepay shipping charges LIMITATIONS AND EXCLUSIONS THE WARRANTY SHALL NOT APPLY TO PROBLEMS ARISING FRO
Download Pdf Manuals
Related Search
Related Contents
View PDF Kobo Arc User Guide DE Arzneimittel-Kompendium 2005: Wichtige Hinweise Guia de Recuperação, Backup e Solução de Problemas Manual de Instrucciones ASUS CROSSHAIR_IV_FORMULA Owner's Manual Griffin PowerBlock Reserve 5 – Wiedergeben Design House 741033 Installation Guide Kicker ZX550.3 Owner's Manual Copyright © All rights reserved.