Home
        User Guide 420A
         Contents
1.             na 14       Cleaning Me Dual Wire WIV ORT OD ssa Edo Paus            chua ehem dran cimus ducc ductum daas ac psu SGH aab duca A AD ad cta su ERU REB P 14  4 2 Maintenance of the force transducer                    CUR RU ua Ri Reach cR kr ew ds QD Reine 15  LN                                            SU RETE T I                                     ee 18  2 2 2 Force transducer                                                                               nuc DEN                     Rune 16  4 3 Changing the Dual Wire Myograph window              2                               ena          aua Rua auk                                 ua sua E                      cidunt 17  4 4 Maintenance of the linear 511966                          RS XH URDU EEG Pica Eo PRICE EN ERR Rt ce iR e 17  Appendix 1  Buffer FeCIDOeS  cessa remm a Exe Cuni Ea Ex ES Ein        KEA ENEAN 18  Appendix 2   Normalization theory  ioskxccenkxuEEXETREEMUEN EUER SERYXRDEENEAKEESKNMENMUP  NGRKNKESKNUON EUR FEEKERNXEKBNRUREVEXEFEG MSN RUN REUNKODEEUANKFD YR nnmnnn 20  Appendix      Reading    millimetre                                        icri e pnr iex nob i c all c cR c rer n 22  NOTOS eR TP                                                                                       P       23    2 WIRE MYOGRAPH SYSTEM   420A   USER GUIDE    CHAPTER 1   DUAL WIRE MYOGRAPH OVERVIEW    Port for connection to Wire Interface Allen screws for fine alignment of the  e Myograph jaws  4 on each side     T
2.      0229 M      DAI    e Hener    ae    Micrometer            d    rio 06 EH    dI       Ln Bal    Myograph jaw connected            12  11      CJ              Myograph jaw connected to force transducer    UM re     49 Window at the bottom of the Myograph chamber     for imaging  under the wire jaws                Myograph chamber separator              Figure 1 1 Dual Wire Myograph with close up of the chamber    CHAPTER 1       3    CHAPTER 2   SETTING UP THE DUAL WIRE                     2 1 Adjustment of supports    A successful mounting of any kind of tubular tissue segment in the Dual Wire Myograph is to a high extent dependent on perfectly  matching supports  The supports are matched prior to the shipment but daily use of the Dual Wire Myograph and greasing of the  transducer pinhole will over time create a need for an adjustment of the supports     NOTE   THE TRANSDUCERS ARE FRAGILE AND SENSITIVE TO MECHANICAL STRAIN  BE VERY CAUTIOUS NOT TO PUT STRAIN ON THE  TRANSDUCER WHEN CHANGING OR ADJUSTING THE MOUNTING SUPPORTS  IN ADDITION  VERY LITTLE FORCE SHOULD BE  APPLIED TO THE SCREWS IN ORDER TO AVOID BREAKING THE THREADS     Adjustment of the supports is performed using the following step by step procedure  The procedure is illustrated in figure 2 1     1  Carefully loosen screw    D    on the top of the support connected to the force transducer  Align the horizontal Support and  carefully tighten the screw again     Loosen screw  E  on the top of the support connec
3.   Carefully disconnect the plug on the force transducer    5  Loosen and remove the two Allen screws          and washers as illustrated in figure 4 2 and carefully remove the old trans   ducer    6  Remove any remaining grease from the transducer pin left inside of the transducer compartment of the Dual Wire Myo   graph  Clean the hole leading from the transducer compartment to the Dual Wire Myograph chamber    f  Gently place the new force transducer into position  Place the Allen screws and washers in their positions ready to be tight   ened    8  Before tightening the Allen screws be sure that the transducer pin is placed directly in the center of the    chamber pinhole       when viewing from the side of the Dual Wire Myograph        Figure 4 2 Illustration of how to replace the force transducer    16 WIRE MYOGRAPH SYSTEM   420A   USER GUIDE    9  Tighten the two Allen screws and replace the bottom plate  Tighten the four screws on the bottom plate     10  Place some high vacuum grease around the transducer pin in the chamber  Make sure that the hole is completely sealed  so absolutely no buffer solution is able to enter the transducer compartment  which will damage the force transducer     11  Mount the support on the transducer pin and adjust the supports as described in chapter 2 1     IMPORTANT  BEFORE MAKING ANY EXPERIMENT ON THE DUAL WIRE MYOGRAPH  REMEMBER TO PERFORM A NEW FORCE CALIBRATION     4 3 Changing the Dual Wire Myograph window glass  The glass in the Dua
4. DMI    DANISH MYO TECHNOL Ost    DUAL WIRE MYOGRAPH SYSTEM   420A  USER GUIDE    User Guide    Version 1 0       CONTENTS    Chapter 1   Dual Wire Myograph overview                                                                                                                                                                                                                                                                      3  Chapter 2   Setting up the Dual Wire Myograph        ssssnsnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn nnmnnn nnmn 2  2 Adjustment or SUPPO                 E                         4  Zi OS AMANO MT TT eine escent nc A ain A sa a sce m                           ePP            5  Chapter 3   Experimental SOE UD  ooasieargrns cxi o ev nibo Rinceg QUON EDO RON K In xr c n Roe nO RETE RON OON BOE nonnisi 6  S MN Mounting Protocol            arte Emm 6  MSN TERMS ETONE ita cc cnc te cence erotic endear et wel elspa rt eee ram E E 6  Sd  2 MOURUS SIED cic atc N Fn cna                    ULCUS SNO ME rNu CER DM UUn ARR E ER DUNS 7  Bo  3 MOBURBUDS SLE D IRITE Sta    are gustu E E E E i 7  S L IACI SIC TOUT es                                                      8  iG MINNIE BITE Dmm 8  S MOURUDS SIOD CD                   cen are manne tr irre  een naar Dusche Xe Ra aa                  RCM D Tu ee en eee    UU GER          T          M DEER E UD PU  9  euo NOUBUDE SIBDSBVOD aided esum EAEE EE m depu ud oua a 
5. Figure 4 3 Greasing points on the linear slides     CHAPTER4 17    APPENDIX 1   BUFFER RECIPES    Physiological Saline Solution  PSS     1x PSS    Chemical Mol Wt mM g 0 5L gL g 2L g 4L           58 45  130 3 799 7 598 15 20 30 39  KCI  74 557  4 7 0 175 0 35 0 70 1 40  KH PO   136 09  1 18 0 08 0 16        0 64  MgSO  7H O  246 498  1 17 0 145 0 29 0 58 1 16  NaHCO   84 01  14 9 0 625 1 25 2 50 5 00  Glucose  180 16  5 5 0 5 1 00 2 00 4 00  EDTA  380  0 026 0 005 0 01 0 02 0 04  CaCl   110 99  1 16 O 8mL 1 6mL 3 2 ML 6 4mL     1 0 M solution     1  Make a 1 0M solution of CaCl   110 99  in double distilled H O  Filter sterilize the calcium solution through a 0 22 um filter   The sterilized solution can be stored in the refrigerator for up to 3 months     2  Dissolve all the chemicals except the CaCl  in approximately 80  of the desired final volume of double distilled H O while  being constantly stirred  For example  if 1 litre of PSS is to be made  then dissolve all the chemicals in 800mL of double  distilled H O     3  Add the appropriate volume of 1 0M CaCl  for the total volume of PSS being made  for example  1 6mL of 1 0M CaCl  for 1  litre of buffer   Continue to stir the PSS while the CaCl  is being added     4  Bring the solution up to the final volume with double distilled H O  Continue to stir the solution until the EDTA is fully  dissolved  This takes about 15 minutes at room temperature     5  Aerate the solution with carbogen  95        5         for about 20 mi
6. Myograph is a very delicate and sophisticated piece of research equipment  DMT recommends that the following  sections are read carefully and that the instructions are followed at all times     As a part of the general maintenance of the Dual Wire Myograph  DMT recommends that the Dual Wire Myograph is force cali   brated at least once a month  The Dual Wire Myograph should also be force calibrated every time the Wire Interface has been  moved  Although lab benches are all Supposedly perfectly horizontal  small differences in lab bench pitch can affect the calibra     tion of the system  The Dual Wire Myograph also should be calibrated if the system has been idle for longer than a month  A  step by step procedure is explained in chapter 3 7 1 in Wire Myograph System   User Manual     4 1 Cleaning the Dual Wire Myograph    DMT STRONGLY RECOMMENDS THAT THE DUAL WIRE MYOGRAPH AND SURROUNDINGS WILL BE CLEANED AFTER EACH  EXPERIMENT  PLEASE REMEMBER TO TAKE OUT THE CHAMBER DIVIDER     1  Fill the chamber to the edge with an 8  acetic acid solution and allow it to stand for a few minutes to dissolve calcium de   posits and other salt build up  Use a swab stick to mechanically clean all the surfaces of the Dual Wire Myograph     2  Remove the acetic acid and wash the chamber and supports several times with double distilled water         If any kind of hydrophobic reagent have been used  which might be difficult to remove using step 1  and 2  then try incubating  the chamber and supp
7. NT  DMT RECOMMENDS USE OF THE HIGH VACUUM GREASE ONCE A WEEK TO SEAL UP THE TRANSDUCER HOLE BY FRE   QUENTLY USE     DMT TAKES NO RESPONSIBILITIES FOR THE USE OF ANY OTHER KINDS OF HIGH VACUUM GREASE THAN THE ONE TO BE  PURCHASED FROM DMT     DMT TAKES NO RESPONSIBILITIES FOR ANY KIND OF DAMAGE APPLIED TO THE FORCE TRANSDUCER        Figure 4 1 Transducer pin hole to be sealed up with high vacuum grease    seen in the transducer house  left  and inside the chamber  right     4 2 1 Checking force transducer    The force transducer is a strain gauge connected in a Wheatstone bridge  The force transducers are placed in separate  compartment  The separate compartments provide some mechanical protection for the force transducer but the transducers  are still very vulnerable to applied forces exceeding 1 newton  100 g  or fluid running into the transducer compartment due to  insufficient greasing of the transducer pinhole     If the force reading on the Wire Interface appears unstable then first check that the Wire Interface and the Dual Wire Myograph  are properly connected through the 25 pin grey cable  If the force reading still appears unstable then perform a new force cali   bration of the force transducer  During the force calibration monitor the relative force reading values in the Calibration Menu on  the Wire Interface  The normal operating values for the force transducer during calibration should be between 3000 and 3500     e lf the value is O  a single digit  or a three 
8. Stimulus 1  amp  2       Wash out     KPSS   10 uM NA 4 x with PSS  Stimulate for 3 minutes Wait 5 minutes      Stimulus          Wash out     PSS   10 uM      4 x with PSS  Stimulate for 3 minutes Wait 5 minutes      Stimulus 4       Wash out     KPSS 4 x with PSS  Stimulate for 3 minutes Wait 5 minutes       Stimulus 5      Wash out    KPSS   10 uM NA 4 x with PSS  Stimulate for 3 minutes Ready for experiment       3 4 Endothelium function  The reasons for checking endothelium function may include     1  To check whether the relaxing function of the endothelium is intact  The procedure is performed to make sure that the en   dothelium is not damaged during the dissection or mounting procedure     2   f an experiment requires removal of the endothelium this procedure is useful to check whether the endothelial cells were  successfully removed     The procedure can be performed after the vessel segment has been heated  equilibrated and normalized  Preferably the proce   dure should be done after performing a standard start to make sure that the vessel segment is viable     The present procedure is for use with rat mesenteric arteries  Another procedure may be needed for other animal species and  tissue or vessel types     3 4 1 Principles of checking endothelium function    Stimulating a vessel segment with acetylcholine causes a release of nitric oxide  NO  also known as EDRF  from the endothelium  cells and subsequent relaxation of the vascular smooth muscle cells  If the en
9. apart     While controlling the movement of the wire with the forceps  use the other forceps to gently pull the vessel segment along  the wire until the area of interest is situated in the gap between the jaws  The near end of the vessel segment shall lie about  0 1 mm inside the jaw gap to insure no point of contact  figure 3 3 A     Still controlling the free wire end with the forceps  move the jaws together to clamp the wire and in one movement secure    the wire under the near fixing screw on the left hand jaw  Again in a clockwise direction so that tightening the screw also  tightens the wire  figure 3 3 B               Figure 3 3 A and    Mounting step 3       CHAPTERS 7    3 1 4 Mounting step four    e Using forceps  gently rub the vessel segment on the far side of the jaw to Separate any excess vessel Segment from the area  of interest clamped in the gap between the jaws  figure 3 4 A   Make sure that the vessel segment is Separated as close as  possible to the jaws  figure 3 4 B   The excessive vessel segment is finally dissected free and removed from the Dual Wire  Myograph chamber  figure 3 4 C      ba 94 94    Figure 3 4     B and    Mounting step 4    3 1 5 Mounting step five    e Move the jaws apart  figure 3 5 A   Take a second wire holding it about one third down from the far end using forceps  Align  the wire parallel with the vessel segment such that the wire can be passed into the far end of the lumen  Gently feed the  wire through the lumen of the vessel s
10. by     IC    IC           ot    1 0    2    APPENDIX2 21    APPENDIX 3   READING A MILLIMETRE MICROMETER    Sleeve scale Thimble scale       Figure A3 1 Overview of the micrometer parts  actual reading 20000 um   20 mm     Sleeve scale  The micrometer sleeve scale has a total length of 25 mm divided into 50 equal parts  Each part of a division above the horizontal  line represents 1 mm  where each 5th line is marked by a longer line and a number designating the length in mm  Each division  below the horizontal line is placed between each 1 mm mark  scale above the horizontal line  and represents 0 5 mm   Thimble scale  The thimble is divided into 5O equal parts  and one complete rotation of the thimble is indicated by the smallest division on the  sleeve  which equals 0 5 mm  Each division on the thimble scale is 10 um  If the thimble scale falls between two lines  then     number between O and 10 um must be approximated   Example 1   1  Note that the thimble has stopped at a point beyond    10    on the sleeve indicating 10000       10 mm      2  Note that there is no mark completely visible between the 10 mm mark and the thimble     3  Read the value on the thimble corresponding to the intersection with the horizontal line on the sleeve        A  Reading on sleeve  10000 um  B  Noadditional mark visible     um  C  Thimble reading  380 um  Total reading  10380       Figure A2 2 Example 1   reading   10380 um  Example 2    1  Note that the thimble has stopped at a point beyon
11. ciples of the normalization procedure    In practice the normalization is performed by distending the segment stepwise and measuring sets of micrometer and force read   ings  figure 3 8   These data are converted into values of internal circumference  um  and wall tension T  mN mm  respectively     Plotting wall tension against internal circumference reveals an exponential curve and by applying the isobar curve corresponding  to 100 mmHg           5 calculated from the point of intersection using the Laplace relation  figure 3 9   IC  is calculated from IC   by multiplying a factor giving an internal circumference at which the active force production as well as the sensitivity to agonists  of the segment is maximal  For rat mesenteric arteries the factor is O 9 but both this factor as well as the transmural pressure has  to be optimized for each particular segment  The normalized internal diameter is calculated by dividing IC  with        Appendix 2 contains a complete description of the mathematical rationale and calculations of the normalization procedure     Force  mN mm   Wal Tension     mN mm        Time  min  d a   lee PIER oed ub   T  Figure 3 8 lllustration of the stepwise Figure 3 9 Illustration of the exponential  normalization procedure curve fitting and determination of IC po    3 3 Standard start    The purpose of performing a standard start is to   1  Re activate the mechanical and functional properties of the vessel segment     2  Check that responses to differ
12. d    16    on the sleeve indicating 16000 um  16 mm    2  Note that this time a mark is visible between the 16 mm mark and the thimble indication 500 um     3  Read the value on the thimble corresponding to the intersection with the horizontal line on the sleeve     A  Reading on sleeve  16000 um  B  One additional mark visible  500 um  C  Thimble reading  280 um   Total reading  16780 ym       Figure A2 3 Example 2   reading   16780 um    22 WIRE MYOGRAPH SYSTEM   420A   USER GUIDE    NOTES    NOTES 23    DMT A S   Skejby Science Center  Skejbyparken 152  DK 8200 Aarhus N                    Tel    45 87 41 11 00  Fax   45 87 41 11 01    www dmt dk  sales dmt dk  support dmt dk    DMT Asia Ltd    Rm 2402B  Great Eagle Centre  23 Harbour Road   Wanchai  Hong Kong S A R   P R  China    Tel    852 6621 8337  Fax   852 3020 7554    www dmt asia com  sales dmt asia com  support dmt asia com    DMT Asia  China office    Rm 28C  No  8 Dong Fang Road  Lu Jia Zui Financial District  Shanghai 200120   P R  China          6021 5425 1330  Fax   86  0  21 5877 0063    www dmt asia com  sales dmt asia com  supportedmt asia com    DMT USA  Inc    525 Avis Drive   Suite 10   Ann Arbor  MI 48108  USA    Tel    1 770 612 8014  Fax   1 678 302 7013    www dmt usa com  sales dmt usa com  support dmt usa com       
13. d surfaces during the last 15 minutes of the incubation period   3  Wash the chamber and supports several times with double distilled water   4   ncubate the chamber with 9690 ethanol for 10 minutes while continuing the mechanical cleaning with a swab stick     5  Remove the ethanol solution and wash a few times with double distilled water  Incubate the chamber and supports with an  8  acetic acid solution for 10 minutes and continue the mechanical cleaning with a swab stick     6  Wash the chamber and supports several times with double distilled water   IMPORTANT    IN EXCEPTIONAL CASES IT MAY BE NECESSARY TO UNMOUNT THE SUPPORTS AND CLEAN THEM AND THE CHAMBER  SEPARATELY TO ENSURE THAT ALL SURFACES ARE CLEANED     14 WIRE MYOGRAPH SYSTEM   420A   USER GUIDE    4 2 Maintenance of      force transducer    The force transducer is the most delicate and fragile component of the Dual Wire Myograph  Careful handling is very important   Two of the jaws is connected to the transducer pins  on each side in the chamber   The transducer pins enters the chamber  through a pinhole in the chamber wall located below the surface level of the buffer  See figure 4 1   To prevent the buffer from  running into the transducer house the hole is filled with high vacuum grease  As a part of daily maintenance it is very important  to inspect the greasing of the transducer hole before starting any experiment  Insufficient greasing will cause damage and  malfunction of the force transducer     IMPORTA
14. digit number  the force transducer is broken and needs to be replaced    e    the value is less than 2000 or greater than 4500  the force transducer is broken and needs to be replaced    e    the force reading s  appear yellow  cannot be reset to zero  or the transducer cannot be recalibrated  the force transducer  is broken and needs to be replaced  If any other problems related to the force transducer are encountered  please contact  DMT for advice or further instructions     IMPORTANT  IF THE MESSAGE    OFF    IS DISPLAYED IN EITHER OF THE FORCE READING LINES IN THE MAIN MENU ON THE WIRE INTERFACE    THEN THE FORCE TRANSDUCER IS BROKEN AND NEEDS TO BE REPLACED  IN CASE OF ANY OTHER PROBLEMS RELATED TO  THE FORCE TRANSDUCER  PLEASE CONTACT DMT FOR FURTHER INSTRUCTION AND ADVICE     CHAPTER 4 15    4 2 2 Force transducer replacement    In case that the force transducer is broken and needs to be changed  please follow this step by step replacement procedure    carefully    1  Disconnect the Dual Wire Myograph from the Wire Interface  grey cable     2  Carefully remove the support connected to the transducer pin by loosening the screw    D    on top of the support as illustrated  in figure 2 1    3  Turn the Dual Wire Myograph up side down and remove the bottom plate by loosening the four screws  A  as illustrated in  figure 4 2   IMPORTANT  NOTICE HOW THE PLUG IS CONNECTED TO THE FORCE TRANSDUCER SO NO MISTAKE IS MADE WHEN CONNECTING  THE PLUG TO THE NEW FORCE TRANSDUCER    4
15. dothelium is undamaged by the dissection and  mounting procedures  then a substantial relaxation will occur  With complete removal or damaged endothelium  a partial relaxa   tion or no relaxation to acetylcholine is observed     Itis important to note that the amount of NO or EDRF in a vessel is often dependent upon its size  In certain vessels  endothelium   derived hyperpolarizing factor  EDHF  can contribute more or less than EDRF  and in other vessels the same stimulation with ACh  can promote release of endothelium derived contracting factor  EDCF   Therefore  it is important to check the existing literature  in order to determine the expected response in your particular vessel with the given concentration of agonist     CHAPTER    11    3 5 In vitro experiment 1  Noradrenaline contractile response    The purpose of the present protocol is to determine the sensitivity of rat mesenteric small arteries to the vasoconstrictor  noradrenaline norepinephrine with a cumulative concentration response curve     3 5 1 Background    Noradrenaline  norepinephrine  causes contraction of mesenteric small arteries through activation of o adrenoceptors whereas  noradrenaline activation of B adrenoceptors causes vasodilatation  As the purpose is to determine the contraction sensitivity to  noradrenaline  the vasodilatory effect of noradrenaline is eliminated throughout the experiment by the constant presence of the  B adrenoceptor antagonist  propranolol     Rat mesenteric arteries are d
16. e vessel on the far side of the jaws does not extend beyond the jaws  as even a small extension will  affect the normalisation procedure  In case of excess of vessel on the far side of the jaws then move the jaws together again  and remove excessive tissue using a forceps as described in mounting step four  A better method for the skilled operator is  to move the jaws slightly apart and use scissors to make a small slit in the vessel wall where the vessel is clamped        Figure 3 7 A and B Mounting step 7    3 2 Normalization  The importance of normalizing the preparation is three fold     1  Experiments with elastic preparations like vessels can only have meaning if they are performed under conditions where the  size is clearly defined     2  Clearly defined conditions are required in pharmacological experiments as the sensitivity of preparations to agonists and  antagonists is dependent on the amount of stretch         The active response of a preparation is dependent on the extent of stretch  which makes it important to set the preparation  to an internal circumference giving maximal response     CHAPTER3 9    The aim of the normalization procedure is to stretch the segment to a so called normalized internal circumference  IC   defined  as a set fraction of the internal circumference             that a fully relaxed segment would have at a specified transmural pressure   For small rat arteries the target transmural pressure is typically 100 mmHg   13 3 kPa     3 2 1 Prin
17. egment in one movement using the first mounted wire as a guide  figure 3 5 B C   Hold  the wire at a point at least 10 mm from the vessel to prevent the vessel being stretched during the manoeuvre  Be careful  not to touch the lumen of the vessel with the end of the wire and when pushing the wire end through the near end of the  lumen  Once the wire has successfully passed through the lumen of the vessel segment  place the wire in a position  which  ensures sufficient length for the wire to be secured both at the near and far fixing screws on the right hand jaw              Figure 3 5                  Mounting step 5       8 WIRE MYOGRAPH SYSTEM   420     USER GUIDE    3 1 6 Mounting step six    e Carefully move the jaws together while ensuring that the second mounted wire lies underneath the first one secured on the  left hand jaw  figure 3 6 A   The procedure clamps the second wire to prevent it from damaging the vessel segment when  securing the wire to the right hand jaw  connected to the transducer   Secure the near end of the wire in a clockwise direc   tion under the far fixing screw on the right hand jaw  figure 3 6 B         Figure 3 6 A and B Mounting step 6    3 1 7 Mounting step seven    e Secure the far end of the wire under the near fixing screw on the right hand jaw  Again the wire is passed clockwise around  the screw stretching the wire as the screw is tightened  figure 3 7 A B   Move the jaws apart to slightly stretch the vessel seg   ment  Make sure that th
18. ensely innervated by sympathetic nerves  which have a highly efficient reuptake mechanism that  removes noradrenaline from the neuromuscular junction  The reuptake mechanism will create a concentration gradient between  the solution around the vessel segment and the receptors on the smooth muscle  To correctly determine the sensitivity to  noradrenaline it is necessary to eliminate this concentration gradient by performing the experiment in the presence of cocaine  to block the noradrenaline reuptake     To determine the sensitivity to noradrenaline the vessel segment is exposed to increasing concentrations of noradrenaline   Each concentration is applied until a steady response has been reached and then the next concentration is applied  When the  vessel segment is fully contracted or does not response more upon increasing the noradrenaline concentration  the experiment  is ended     3 5 2 Protocol    Prepare the following stock solutions   Noradrenaline  105  103  10  M  Propranolol  10   M   Cocaine  10   M    1  Mount and normalize the vessels as described in chapter 3 1 and 3 2   2  Perform a standard start as described in chapter 3 3     3  Incubate the vessel segment in 1 uM propranolol  add 5 uL of 103 M to 5 mL PSS in chamber  and 3 uM cocaine         15  uL of 10   M to 5 mL PSS in chamber  for at least 10 minutes     4  Add increasing concentrations of noradrenaline into the bath  use the table below as a guideline   Wait for a stable contractile  response or a sta
19. ent types of stimuli are normal in appearance and thereby ensuring that the functionality of  the vessel segment has not been damaged during the dissection or mounting procedures     3  Ensure that the tension development gives an effective active pressure that is above the chosen accepted value  usually  13 3 kPa   100 mmHg      The standard start is performed after the vessel segment has been heated  equilibrated and normalized  The present procedure  is suitable for rat mesenteric arteries  Another procedure may be needed for other animal species and tissue or vessel types     3 3 1 Principles of the standard start procedure    The standard start procedure consists of a series of five stimuli and washout periods  The first two stimuli are performed using  a mixture of KPSS and 10 uM noradrenaline to give a maximum contractile response  The third stimulus is performed using a  mixture of PSS and 10 uM noradrenaline to give a maximum pure agonist mediated  o adrenoceptor  contraction  The fourth  stimulus is performed using KPSS to give a depolarising contractile response  this stimulus also includes a component from neu   rally released noradrenaline   The final stimulus is performed using a mixture of PSS and 10 uM noradrenaline  All solutions are  preheated to 37  C and aerated with a mixture of 9576 O  and 5  CO  before use  Instructions for making the necessary solutions  are described in appendix 1     10 WIRE MYOGRAPH SYSTEM   420A   USER GUIDE      Repeat 1 x        
20. fixing screw  so the wire is wound clockwise  tighten   ing the screw will then tighten the wire  This procedure should result in the wire being clamped between the jaws and with    near end of wire pointing towards operator  figure 3 1 B C      e Fill the Dual Wire Myograph chamber with PSS  at room temperature   See appendix 1 for example of buffer recipes        Figure 3 1 A  B and C Mounting step 1    6 WIRE MYOGRAPH SYSTEM   420A   USER GUIDE    3 1 2 Mounting step two    Using forceps to hold the handle segment  transfer excised vessel from petri dish to Dual Wire Myograph chamber  Hold the  vessel as close to the proximal end as possible and try to mount the vessel onto the wire     If the lumen is shut  try one of the following possibilities   1  Use the wire to gently push the lumen open  blood streaming out is a good sign      2  Hold excised vessel about 3 mm from the cut end with one set of forceps and use the other forceps to squeeze the blood  remaining in lumen out through the cut end     Pull the proximal end of the excised vessel segment along the wire such that the vessel segment acts as its own feeder to be  feed into the wire into the vessel  figure 3 2 A C   Be careful not to stretch the vessel segment if the end of the wire catches  the vessel wall        Figure 3 2 A  B and C Mounting step 2    3 1 3 Mounting step three    Once the vessel segment is threaded onto the wire  catch the free end of the wire  nearest you  with the forceps and move  the jaws 
21. h carbogen  9576 0    576 CO   for about 20 minutes     APPENDIX 1 19    APPENDIX 2   NORMALIZATION THEORY    The importance of making a normalization before initiating an experiment with any tubular tissue segment is described in chapter  3 2  In this appendix the mathematical rationale and calculations underlying the normalization procedure are described in detail     Mathematical calculations    Let  X  Y  be the pair of values representing the micrometer reading  see appendix 3  and force reading respectively character   izing each step in the normalization procedure  Y  is the force reading at the start position of the normalization procedure where  the wires are just separated and the force reading is approximately zero  Then  given that tension on the vessel is equal to force  divided by wall length  the wall tension at the i th micrometer reading is calculated by      Y   Y     T                 90            28 e  a           where 6 is the microscope eyepiece reticule calibration factor in mm per division and a  and a  are the vessel end points when  measuring the length of the mounted vessel segment     The internal circumference of the mounted vessel at the i th reading is calculated by   IC IO PEZ      e     where IC  is the internal circumference of the mounted vessel when the wires are just separated and is given by     IC    2 m  d  where d is the wire diameter  For 40      wires  IC    205 6 um     Using the Laplace relation  the effective pressure    is calc
22. l Wire Myograph chamber window is fixed in place and kept waterproof by a thin layer of high vacuum grease  on the circular edge between the glass and the chamber base  The following procedure describes how to change the window    glass     1  Screw the jaws as far apart as possible and carefully remove the jaw from the transducer pin side  it should not be neces   sary to remove the jaw on the micrometer side      2  Loosen the glass from the chamber by gently pushing up on the glass from below the window with a blunt tool    3  Remove the old grease and clean the area thoroughly with 96  ethanol    4  Carefully apply a small  continuous amount of high vacuum grease around the edge of the window  Using forceps  place  the new window glass in place  Push down gently around the edges to create a seal between the glass  the grease and the    chamber base     5  Check that the new window forms a tight seal by filling the chamber with distilled water  If there is a leak  repeat the replace   ment procedure     6  Finally  replace the jaw on the transducer pin and adjust the supports according to the instructions provided in chapter 2 1     4 4 Maintenance of the linear slides    Check the linear slides  under the black covers  for grease at least once a week  In case of insufficient lubrication  the micrometer  will not move as effortlessly as it Should  grease the slides with the original enclosed grease for linear slides at the points marked  by the arrows in figure 4 3        
23. ndard time such as 2 minutes between each application      NA  in chamber  uM   Volume of stock solution to add to chamber    ea              es             sO OM  sorte       12 WIRE MYOGRAPH SYSTEM   420A   USER GUIDE    3 6 In vitro experiment 2  Acetylcholine relaxation curve    The purpose of the present protocol is to determine the sensitivity of the endothelium dependent vasodilator acetylcholine in  noradrenaline pre contracted rat mesenteric small arteries     3 6 1 Background    Acetylcholine causes relaxation of rat mesenteric small arteries by activating of muscarinic M3 receptors at the endothelial cell  layer leading to release of endothelium derived relaxing factors     Rat mesenteric arteries do not show spontaneous tone in the wire myograph  which is why it is necessary to first induce a con   traction to be able to observe the relaxation to acetylcholine  In this protocol the contraction is induced by noradrenaline  The  required concentration of noradrenaline needs to be optimized since a too low concentration makes it impossible to evaluate the  relaxation  On the other hand it may be difficult to relax super maximally contracted arteries  which may lead to an underestima   tion of the sensitivity to acetylcholine  Therefore it is recommended to apply a concentration of noradrenaline inducing 60 70   of maximal contraction response  In practice this concentration is found by performing a noradrenaline concentration response  curve as described in the previ
24. nutes     25x Concentrated PSS     Chemical Mol Wt mM g 0 5L gL g 2L g 4L           58 45  3250 94 98 189 96 379 92 759 84  KCI  74 557  117 5 4 375 8 75 17 5 35 0  KH PO   136 09  29 5 2 0 4 0 8 0 16 0  MgSO  7H O  246 498  29 25 3 625 1 25 14 5 29 0  NaHCO   84 01  14 9 0 625 1 25 2 50 5 00  Glucose  180 16  5 0 0 5 1 00 2 00 4 00  EDTA  380  0 65 0 125 0 25 0 50 1 0  CaCl   110 99  40 20mL 40mL SOmL 160mL     1 0 M solution     18 WIRE MYOGRAPH SYSTEM   420A   USER GUIDE    Make    1 0M solution of CaCl   110 99  in double distilled H O  Filter sterilize the calcium solution through a 0 22 um filter   The sterilized solution can be stored in the refrigerator for up to 3 months     Dissolve all the chemicals except the CaCl  in approximately 80  of the desired final volume of double distilled H O while  being constantly stirred  For example  if 1 litre of PSS is to be made  then dissolve all the chemicals in 800mL of double  distilled H O    2    Add the appropriate volume of 1 0M CaCl  for the total volume of PSS being made  for example  1 6mL of 1 0M CaCl  for 1  litre of buffer   Continue to stir the PSS while the CaCl  is being added     Bring the solution up to the final volume with double distilled H O  Continue to stir the solution until the EDTA is fully dis   solved  This takes about 15 minutes at room temperature     Before use     Dilute the 25 x PSS stock solution 1 25 using double distilled H O     Add   091 g L Glucose  100 g L NaHCO     Aerate the solution wi
25. orts with 9690 ethanol or a weak detergent solution     4  Toremove more resistant or toxic chemicals incubate the chamber and supports with 1 M HCI for up to 1 hour  In exceptional  cases incubate the chamber and supports with an up to 3 M HNO3 solution for about 15 minutes     5  Wash the chamber and supports several times with double distilled water   To prevent the pipes from being blocked by buffer salt deposits after an experiment  remove the chamber cover from the myo   graph and turn on the vacuum pump and vacuum valve for about 10 seconds  Wait to turn off the oxygen supply until turning off  the vacuum pump  Wipe off any buffer remaining on the outside of the pipes using a piece of paper tissue   IMPORTANT NOTES  BE VERY CAREFUL USING STEP 3 AND 4 REPEATEDLY AS STRONG REAGENTS CAN CAUSE EXTREME DAMAGE TO THE DUAL  WIRE MYOGRAPH   BE VERY CAREFUL NOT TO EXERT ANY FORCE ON THE WIRE JAWS DURING THE CLEANING PROCEDURE     AFTER CLEANING  ALWAYS CHECK THAT THE GREASING AROUND THE TRANSDUCER PIN IS SUFFICIENT TO KEEP OUT THE  BUFFER SOLUTION FROM THE TRANSDUCER COMPARTMENT  SEE FIGURE 4 1      In cases of red or brown discolorations appearing on the chamber sides or on the supports  the following cleaning procedure will  work in most cases     1  Incubate the chamber and supports for 30 minutes with 20 ul of a 2 mM T 1210 Tetrakis   2 pyridylmethyl  ethylenediamine  solution dissolved in double distilled water     2  Use    swabestick to mechanically clean all the affecte
26. ous section     The vessel segment is exposed to the noradrenaline concentration and when the response has stabilised  increasing concentra   tions of acetylcholine are added to relax the vessel  Each concentration is applied until a steady response has been reached and  then the next concentration is applied  When the vessel segment is either fully relaxed or does not relax more upon increasing  the acetylcholine concentration  the experiment is ended     3 6 2 Protocol    Prepare the following stock solutions   Acetylcholine  10   103  10  M  Noradrenaline  10  M    1  Mount and normalize the vessels as described in chapter 3 1 and 3 2    2  Perform a standard start and check the vessel segment for endothelium function  as described in chapter 3 3 and 3 4    3  Add noradrenaline to obtain a response around 6096 of maximum  determined from the previous noradrenaline concen   tration response curve   When the contractile response is stable  add increasing concentrations of acetylcholine to the    chamber  using the table below as a guideline  Wait for a stable contractile response or a standard time such as two minutes  between each application      ea            iMWiM                       1    2 5 uL of 103 M  1 3 1 5 uL of 103 M    1 5 1 uL of 10   M  7 5 uL of 10   M  1 uL of 102 M  2 5 uL of 10  M   In calculating the  ACh  in the Dual Wire Myograph chamber  the applied volume of ACh is ignored        CHAPTER 3 13    CHAPTER 4   CLEANING AND MAINTENANCE    The Daul Wire 
27. quise A A L O A Ed oS das A E E EA E E E A ONE 9  ca NONN ZUON sanra oi o m                                                                   9  GN Re                  EIC V TETTE 10  3 3 1 Principles of the standard start procedure                  E n bx Date imei od nare Duce e 10  3 2 1 Principles of the normalization                       scs ous cui virt nca Riu ERES URSI ERR Ia EQ RARE YII A OA v FC Ck Ra ue dis 10  3 4 Endotheium CO DM mme                                       en 11  3 4 1 Principles or CHECKING endothelium function ofits                   dus                   Sada nb          un Eur ificnigue Dao                                                                                  11  3 5 In vitro experiment 1  Noradrenaline contractile reSPONSE             ccccssssscesececcsssseetececcccsseeeeenccccsseesencceceseeesenececesseerseececesseesesecees 12  uM      ON ENERO 12  DL FOO OT E O E N A E OE E EAO E E                      E QUEE EU  R VUE If EE 42  3 6 In vitro experiment 2  Acetylcholine relaxation CUrve          cccccssssecesececcsssseetececcecseseseeecccussseeeeeccccasseesenccecesseerenececesseerenececesssetenesees 15               Nc sere teeth                                                                                  15  OZ POLO C Ol RENE TNT IT A                                                m 15  Chapter 4   Cleaning and Maintenance                   randi OR Ra CR EN              A OK RUN                 RD ED ebd FH DN arr END       
28. rce calibration    DMT recommends that the Dual Wire Myograph is force calibrated at least once every month  DMT also recommends that the  Dual Wire Myograph is force calibrated every time the system has been moved or has not been used for a long period of time   See the force calibration procedure in chapter 3 7 1 in Wire Myograph Systems   User Manual     CHAPTER2 5    CHAPTER 3   EXPERIMENTAL SET UP    This chapter contains experimental set up for the Dual Wire Myograph  For dissection of a vessel  please see Procedures for  investigations of small vessels using a small vessel Myograph by M J  Mulvany     3 1 Mounting protocol for small arteries  The procedure involves attaching the mounting wires to jaws which are in turn mounted on the force transducer  This force trans   ducer is capable of measuring with a sensitivity of about 0 01 mN  1 mg   but can be damaged if the applied force exceeds about  1 N  100 g   Therefore care must be taken to avoid pressing the jaws too hard together  A movement of  20 um after they have  touched is sufficient to hold the wires clamped   3 1 1 Mounting step one   e Cut lengths of 40 um wire  2 2 cm long  Mount one wire on left hand jaw of the Dual Wire Myograph as follows     e Holding wire at far end  place centre of wire between jaws and screw jaws together so that the wire is clamped  figure 3 1 A      NOTE  DO NOT CLOSE THE JAWS TOO HARD AGAINST EACH OTHER     e Bend the far end of the wire towards the left  and wrap it around under 
29. ted to the linear slide micrometer  Align the horizontal support  matching the force transducer connected support as carefully as possible and gently tighten the screw again     2  Loosen screw    C    on the linear slide to roughly match the linear slide support to the force transducer support in the horizon   tal plane  Tighten the screw before proceeding with step 4     3  The plate          on which the linear slide is mounted is balanced on top of a small stainless steel ball making it possible to  finely adjust the linear slide support in all vertical and horizontal planes using the four Allen screws           Use the four Allen  screws to make the final horizontal  See arrow in figure 2 1  and vertical  See arrow in figure 2 2  adjustments to match the  linear slide support to the force transducer support  The correct matching of the supports is illustrated in figure 2 3     IMPORTANT   AVOID CONTINUOUSLY TIGHTENING THE ALLEN SCREWS DURING THE FINAL ADJUSTMENTS  LOOSEN THE ALLEN SCREW  PLACED DIAGONALLY TO THE ALLEN SCREW BEING TIGHTENED  OTHERWISE THERE 15 A HIGH RISK OF DAMAGING THE  DUAL WIRE MYOGRAPH FRAME                   gt O    gt        Figure 2 1 Illustration of screws for adjusting supports and horizontal adjustment    4 WIRE MYOGRAPH SYSTEM   420A   USER GUIDE       Figure 2 2 Illustration of vertical adjustment       Figure 2 3 lllustration of correctly aligned supports for small vessels  left  and  incorrectly aligned supports  middle and right     2 2 Fo
30. th carbogen  95760    5 CO   for at least 20 minutes  If necessary wait further for the pH of the buffer  to reach pH 7 4     High potassium Physiological Saline Solution      55     1x 60mM KPSS     Chemical Mol Wt mM g 0 5L g L g 2L g 4L           58 45  14 7 2 18 4 37 8 73 17 46  KCI  74 557  60 2 24 4 47 8 95 17 89             136 09  1 18 0 08 0 16 0 32 0 64  MgSO  7H O  246 498  1 17 0 145 0 29 0 58 1 16  NaHCO   84 01  14 9 0 625 1 00 2 00 5 00  Glucose  180 16  bib 0 5 1 00 2 00 4 00  EDTA  380  0 026 0 005 0 01  0 02 0 04  CaCl   110 99  1 6 0 8mL 1 6mL 32ml 6 4mL     1 0 M solution     T     Make a 1 0M solution of CaCl   110 99  in double distilled H O  Filter sterilize the calcium solution through a 0 22 um filter   The sterilized solution can be stored in the refrigerator for up to 3 months     Dissolve all the chemicals except the CaCl  in approximately 80  of the desired final volume of double distilled H O while  being constantly stirred  For example  if 1 litre of PSS is to be made  then dissolve all the chemicals in 800mL of double  distilled H O    2    Add the appropriate volume of 1 0M CaCl  for the total volume of PSS being made  for example  1 6mL of 1 0M CaCl  for 1  litre of buffer   Continue to stir the PSS while the CaCl  is being added     Bring the solution up to the final volume with double distilled H O  Continue to stir the solution until the EDTA is fully dis   solved  This takes about 15 minutes at room temperature     Aerate the solution wit
31. ulated for each pair of readings  The effective pressure is an estimate  of the internal pressure  which is necessary to extend the vessel to the measured internal circumference     T      amp     The stepwise distension is continued until the calculated effective pressure exceeds the target transmural pressure  The target  value needs to be optimized for the individual tissue preparation  optimal active force as determined by the length tension rela   tionship for that tissue   For rat mesenteric arteries the target transmural pressure is normally 100 mmHg  13 3 kPa      IC  T   100            7     100 mmHg T    An exponential curve is fitted to the internal circumference pressure data as illustrated in figure 3 9  Now the isobar correspond   ing to 100 mmHg is used to calculate the       value from the point of interception between the function of the exponential curve  and the function of the 100 mmHg isobar     The normalised internal circumference IC  is calculated by multiplying the internal circumference corresponding to LOO mmHg      op by a factor k  The factor is for rat mesenteric arteries 0 9  Again  this value should be optimized for the particular tissue  preparation being used by a length tension curve     IC   k    IC     20 WIRE MYOGRAPH SYSTEM   420A   USER GUIDE    The normalized internal  lumen  diameter is then calculated by      HE     1 TU    The micrometer reading X  at which the internal circumference of the normalized vessel is set to is calculated 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Heaven Fresh HF 380A air purifier  Tripod Adaptor VCT-55I    CAN スタータキット RX630 取扱説明書  Lavatrice Washing Machine  Manual de uso pHMETRO CRISON  User Manual WaveWriter (AWE) Arbitrary Waveform Editor 070  Roland VGA-3 Musical Instrument User Manual      Copyright © All rights reserved. 
   Failed to retrieve file