Home

Laurel Laureate Series Panel Meters Manual PDF

image

Contents

1. Input Display Scaling Fite Relay Alarms Communicatior Analog Out Lockouts InputsDisplay Scaling Filter Relay Alarms Communication Analog Out Lockouts Alarm 1 4 2 BASIC Sig Cond Option Relay Option Comm Option Analog Option 4 Setpoint 1 Seipoint 2 ion EXTENDED RIS emend Board Board v 010000 020000 r Deviation 1 r Deviation 2 r Signal Input 000000 000000 r Mode Range p Custom Curve Rate r 50 60 Hz Line Freq r Alarm 1 Source gt Alarm 2 Source gt E 0 2 Disabled Disabled hd 60 Hz Filtered Filtered z rDisplay Alarm 1 State Alarm 2 State r Resolution Display Filter r Decimal Point Peak Button Action Auto Tare High Active High Active v 4 5 Digits ha em ddddd Disabled kal Relayl Alrm State r Relay2 Alrm State Active On z Active On r Deviation 1 Type 5 r Deviation 2 Type Band Deviat n Y Band Deviat n Y p Control Inputs rpAlrmlTye Alarm2 Control Input Control Input B Both Ctrl A B Non Latching v Non Latching v leter Rese r Alarms 1 2 No Rdgs to Alarm 1 Reading No Deviation in Menu S
2. Logic 0 resets all meter functions Function Reset npa toe Darse iba eer e rU dote 2s Logic 0 resets peak values and alarms Decimal Point input Overrides internal DP selections and controls DP position Display Blank InpUbs unde et occ ud uiid etude i qued qnae Logic 0 shuts off the display Power Requirements Input Voltage rating standard 85 264 Vac or 90 300 Vdc DC range not ETL approved Input Voltage rating low voltage 12 32 Vac or 10 48 Vdc Power Line Frequency qnin Fea andati qute Det aaa qued A DC and 47 63 Hz Power Consumption Max tea ius 5 Watts Excitation Outputs Voltage amp Current Levels jumper selectable 5 Vdc 5 100 mA max 10 5 120 mA max 24 5 40 mA max Excitation Output Ripple lp RT 100 mVp max Isolation from power and outputs 250 Vac Insulation dielectric strength to power and outputs 3 5 kV ac for 5 sec 2 3 kV ac for 1 min Isolation to signal COMMOoOn lsd lel a mln Ud ro iu 50 Vdc Maximum Applied Voltage Signals Voltage input ranges of 2V and above AC or 600 Vrms Voltage input ranges of 200 mV AC or DC load cell thermocouple RTD 100 Vdc Input Types Range
3. conditioner boards The PC first recognizes Fm ni EHE the type of board then prompts you to apply specific jumpers and calibration signals Press Ready to take a reading Press Repeat to take more readings When you have decided on which reading to accept press on the number 1 through 10 of that reading Additional cali E bration software is available online 213 5 InPut d C METER SETUP WITH AN UNCONNECTED PC SEtuP 0 0 Instrument Setup software is also of benefit when the is not connected to a meter Upon launching the 00 1 1656 software click on None for Communications then on ine Sees DPM and Series 2 Click on File gt Default Setup to 00010 retrieve a default setup file from disk or on File gt Open Setup to retrieve a previously saved setup file from disk s 2 To enter new setup information click on View gt Setup Loc 1 00000 then make your screen selections as if you were Loc2 0010 connected to meter Tabs will be grayed out if you Loc3 0000 have not selected the required hardware under the Input Display tab When done press on Main Menu then on View Menu The selections made under Setup will now be shown in the form of the required front panel programming sequence where each row corresponds to a menu item selected by the k key and
4. 0 20 000 kohms BOX2 BOX1 plus IPC 5 0 200 00 kohms Input DC Volts DCV1 200 00 mV DGV2 se 2 0000 V 20 000 V DOCVA 200 00 V DGV5 ee 600 0 V 6 300 0 V R6 0 2 Mohms fixed range Process Signals 4 20 mA 0 10V etc Peces 4 20 mA 0 100 00 Custom Scaling Specify min signal amp reading max signal amp reading Range not ETL certified 2 TABLE OF CONTENTS T 2 2 TABEE OF CONTENTS ae e eite Et um MA RM aaa cae as 3 3 PRODUCT INTRODUCTION cscs csscne nee eRe Renee eee 4 4 RECEIVING amp UNPACKING 5 5 SAFETY GONSIDERA TIONS eee Fer eee cete re eed cer e Eh 5 6 CONNECTOR WIRING 6 7 MECHANICAL ASSEMBLY 8 8 5 KEYS tnit tuta Octa 10 9 ENABLING amp LOCKING OUT MENU ITEMS seem 12 10 READING COORDINATES OF 2 POINTS SCALING 13 11 DC VOLTS AMPS PROCESS 14 12 LOAD CELL S MICROVOL T INBU Ta 20 13 AC RMS VOLTS amp AMPS INPUT reser epeio reget pee reu pets 25 14 THERMOGOUDPEE INPUT end
5. oo 2775 2750 256 2725 2700 0 0 1 2 3 4 5 6 7 8 4 10 2725 2750 2775 2800 2825 2850 2875 Print Save Pause Continue New Scale End Print Save Pause Continue New Scale End Plot Graph 50 e The Readings pull down menu provides three formats to display DPM data on the PC monitor Use the Pause and Continue buttons to control the timing of data collection then press Print for a hardcopy using your PC printer List presents the latest readings in a 20 row by 10 column table Press Pause at any time to freeze the display Press Print for a hardcopy List can capture peak readings Plot generates a plot of readings vs time in seconds It effectively turns the DPM PC combination into a printing digital oscilloscope Graph generates a histogram where the horizontal axis is the reading and the vertical axis is the number of occurrences of readings The display continually resizes itself as the number of readings increases The Jumpers pull down menu provides prz jumper positions for the various meter boards duplicating information in this manual er Calibration The Calibration pull down menu allows eee ae easy calibration of voltage and current ranges Desired Current ux Input Reading Reading Calib Calib Reading for the DC load cell and AC RMS signal vee 5 ws ere
6. To prevent electrical or fire hazard do not expose the instrument to excessive moisture e Do not operate the instrument in the presence of flammable gases or fumes such an environment constitutes a definite safety hazard This meter is designed to be mounted in a metal panel or a bench or wall mount style case The spacing around the meter and the ventilation must be sufficient to maintain the ambient temperature at less than 55 C e Verify the panel cutout dimensions and mount according to instructions Symbols used N Caution refer to accompanying documents a Earth ground terminal AN Caution risk of electric shock gt Both direct and alternating current Equipment protected throughout by double insulation or reinforced insulation Operating environment The meter is Class double insulated equipment designed for use in Pollution degree 2 6 CONNECTOR WIRING INFORMATION CONNECTORS Connectors for signal and power are UL rated screw clamp terminal blocks that plug into mating jacks on the printed circuit board Communication connectors are a single RJ11 plug for RS232 dual RJ11 plugs for RS485 dual RJ45 plugs for RS485 Modbus or USB P1 POWER AND DIGITAL CONTROLS ACHI DCHI 1 000 AC NEUTRAL DC RET 2115 EARTH GROUND TH CONTROL INPUT 2 5V OUT CONTROL INPUT 1 3 4 5 000 DIGITALGROUND 6 Warning Hazardous voltages be present on pins 4 5 amp 6 of P1 since
7. 5 000 A C427C Cu10 202 to 850 C C427F Cu10 331 to 1562 F Power 85 264 Vac or 90 300 Vdc 10 48 Vdc or 12 32 Vac Load Cells 6 wire ratio WM1 99 999 to 99 999 Thermocouples Setpoint Output 5 Specify min input signal amp dis A p p d played reading and max input ee Two contact relays 2A 11972 E discale noate RE Two solid state relays KF 408 to 2501 F nV 10 Vde excitation NM Four 8A contact relays TG 257 to 400 C Y Four solid state relays Options amp Accessories Analog Output Goss 400 to 1830 BL Blank lens no button n NCMO OR MERE None 245 to 1300 C pads 4 4 20 mA 0 10V 10 410V 410 to 2370 F CBLO1 RJ11 to DB9 RS232 FE SC b es 46 to 1768 C cable Connects me Digital Interface i MR 51 to 3214 F ter to PC com port 45 1768 CBLO2 USB to DB9 adapter eee Rs232 49 to 3213 F For use with CBLO1 PAPE et RS485 CBLO5 USB cable Type A d mal RS485 Modbus ee male to Type B male 0 20 000 ohms IPC NEMA 4 panel cover 6 NACE USB to RS485 converter R2 Channa wig E Ne 0 200 00 ohms BOX1 NEMA4 wall mount Ethernet 0 2 0000 kohms enclosure 8 Ethernet to RS485 converter R4
8. Adaptive filter 0 Low adaptive filter threshold level threshold High adaptive filter threshold level 00000 Input signal filtering Can be applied to dis play setpoint analog output data output Autofilter Batch average 16 readings Moving average 0 08 sec Moving average 0 15 sec Moving average 0 3 sec Moving average 0 6 sec Moving average 1 2 sec Moving average 2 4 sec Moving average 4 8 sec Moving average 9 6 sec Unfiltered dEc Pt d dddd SINUS Dec point selection Decimal point flashes Press A to shift the decimal point gt 17 Press Menu 82 Press Digit Press Value Select Select Key Select Key 7 Key Scaling method Scale Offset if selected under eee ONY Select E thru for flashing first digit thru 2 0 WV Wu for other flashing digits Select decimal point 0 0000 location when decimal point is flashing Select digit to flash Select E thru 8 for flashing first digit 0 thru E for other flashing digits Decimal point location is selected by 0 000020 0000 for other flashing digits Decimal point is set Select digit to flash by input range chosen Select Ej thru E for flashing first digit thru B Desired reading at for other flashing digits Decimal point is set
9. Both control inputs 1 amp 2 set to 1 for selec tions 3 7 5 9 Meter Reset Operation asa Not rate of change amp Rate x 10 Meter Configuration rate of change meter J Rate x 0 1 A Rate x 1000 Extended meter only Rate x 1 Rate x 10000 00000 Peak Display Also selects Peak Operation of front Peak or Valley at connector above panel PEAK button Valley Display Also selects Valley in and rear connector for Peak or Valley at connector above Peak or Valley Display Peak 1st push Valley 2nd push Front panel Tare 00000 0 Auto tare selection fl Auto tare No auto tare 00000 RTD Ohms RTD or Ohms wiring 3 or 4 wire 3 or 4 wire Not allowed Custom curve IT 2 wire read 2 wire read 11 2 wire short 2 wire short FiLt 00000 Unfiltered output Filtering Alarm filtering fl Filtered output 00000 0 Unfiltered Peak amp Valley Peak amp Valley filtering 1 Filtered Peak amp Valley 00000 0 Display batch average every 16 readings Display filtering fl Display filtered signal GR Adaptive filter a Low adaptive filter threshold level threshold High adaptive filter threshold level 37 Press Menu 344 Press Digit Press Value Select mud Select Key Select Key y Key FiLt 00000 f Autofilter Filtering Input signal filtering fl Batch average 16 readings continued be applied to dis Moving average 0 08 sec play setpoint
10. High signal input Lo rd Desired reading at Lo In Desired reading at Hi In Apply a low reference Press A to store the low signal input in the signal to the meter meter Apply a high reference Press A to store the high signal input in the signal to the meter meter ON Spp enc SE Se Select E thru E for flashing first digit thru 2 Select digit to flash for other flashing digits Decimal point is set by Eae an ON Spp Sep GATT SE Se Select E thru E for flashing first digit i thru 2 Select digit to flash for other flashing digits Decimal point is set by 23 Option board dependent menu items 2 1 2 Menu items related to alarm setup These will only appear if a relay board is detected If so please see Section16 AnSEt An Hi Menu items related to analog output setup These will only appear if an analog output board is detected If so see Section 17 SEr 1 JSEr 2 3 SER Addr Menu items related to serial communications These will only appear if an RS232 RS485 or USB I O board is detected If so see Section 18 Menu lockout items Menu items used to enable or lock out hide other menu items Loc menu items may in turn be locked out by a hardware jumper Please see Section 9 24 13 AC TRUE RMS VOLTS amp AMPS INPUT AC voltage or current measurement utilizes the True RMS signal conditioner board w
11. 11111 2 Alarm setup 3 3 Alarm setpoint value Press to display the lockout status consisting of 1 s and 0 s The left digit will flash Press gt again to step to the next digit which will flash programming 4 Analog output scaling 5 Serial interface setup 00000 Loc 3 12345 2 View peak value Press A to set the flashing digit to 0 to enable the 3 View alarm setpoints menu item or to 1 to disable Press MENU to enter 4 Reset peak amp latched alarms See the table to the right for list of menu items that 9 Reset meter reset can be enabled or disabled 12 10 READING COORDINATES OF 2 POINTS SCALING METHOD When the reading coordinates of 2 points scaling method has been selected under the four menu items below will appear ahead of all other menu items when the MENU or k is first pressed from the run mode This scaling method applies a straight line fit between two points which are determined from actual transducer signals and the desired corresponding meter readings A low signal such as the output of a pressure transducer at zero pressure and high signal such as the output of the same transducer at a known high pressure are applied to the meter The desired corresponding low and high readings are then entered from the front panel The meter then applies straight line fit between the high and low calibration points This scaling method has the advantage of calibrati
12. Excitation Signal input Signal input 4 To eliminate lead wire resistance follow this procedure 1 Set the InPut menu item and jumpers for the desired range 2 Set COnFG to Wald 3 Short the two leads at the sensor end 4 When the display has settled push the key The meter will then store a value propor tional to lead resistance automatically change COnFG to 80 and then reset The same result is obtained by grounding a Control Input that has been selected for Peak or Valley SEtuP where Y 21 2 4 5 Remove the short and connect the sensor The meter now processes the sensor resistance only If the range and associated jumpers are subsequently changed the above procedure must be repeated This procedure is also available through Instrument Setup software Revision 2 7 0 or later 35 KEYSTROKES FOR SETUP If the MENU Ww key does not work see Section 9 Enabling amp Locking Out Menu Items Press Menu Press Digit Press Value Select mud Select Key Select Key y Key EEA Pt100 RTD DIN alpha 00385 F Selection of signal Pt100 RTD DIN alpha 00385 C mpu ppe stange Pt100 RTD ANSI alpha 003902 F Pt100 RTD ANSI alpha 003902 C Ni120 RTD alpha 00672 F Ni120 RTD alpha 00672 C Cu10 RTD alpha 00427 F Cu10 RTD alpha 00427 C OHnnS BET 0 to 20 000 ohms Ohmmeter to 200 00 ohms Set decimal point 0 to
13. SCALE amp OFFSET SETUP For DC voltmeters amp ammeters a scale factor of 1 and an offset of 0 are used for direct readings in milli volts or milliamperes Decimal point selection does not affect the displayed digits For example 0 20V or 0 20 mA signals can both be displayed as 0 20000 A full scale of 20000 may be displayed as 20 000 mA or 20000 pA Use with a current shunt will require a scale factor to be set For example for a 500 100 500A 100 mV shunt divide 5000 the desired full scale display with 0 1A resolution by 10000 displayed value with 100 mV when the scale factor is 1 0 for a scale factor of 0 5 For process amp strain meters scaling is normally set up from the front panel using the and A keys but can also be set up via RS232 485 using special PC compatible setup software available at no charge The meter allows three scaling methods to be selected 1 Scale and offset 2 Coordinates of 2 points and 3 Reading coordinates of 2 points Only menu items applicable to the selected method will be presented DC RATIO OPERATION In the DC ratio or potentiometer follower mode the meter tracks a ratio of the applied excita tion voltage and is unaffected by changes in the excitation voltage This capability is used for resistive bridge sensors and voltage dividers such as potentiometers which track wiper posi tion To implement set to next page For bridges and potentiometers make P5 Signal Input connec
14. plugs PANEL MOUNTING Ensure that the panel mounted gasket is in place against the back of the bezel Turn the two mounting screws counterclockwise until the space between the mounting pawl and the rear of the gasket is greater than the panel thickness Insert the meter in the panel cutout Turn the mounting screws clockwise until the meter is securely mounted in the panel Do not overtighten FRONT VIEW REAR VIEW Mme P2 P3 P4 P5 3 78in omm 04971 V TOP VIEW SIDE VIEW if E 2 Mounting Pawl 102mm 4 02in PANELCUTOUT 45 0 6 0 0mm 1 772 024 0 0in Y 92 0 8 0 0mm 3 622 032 0 0in Dimensioned case drawings 8 FRONT PANEL SETUP KEYS Alarm status indicators 1 through 4 MENU key ALARMS key reverse MENU RESET key value select PEAK key digit select Meter Front Panel There are four front panel keys which change function for the Run Mode and Menu Mode effectively becoming eight keys The keys are labeled with alphanumeric captions MENU PEAK RESET ALARMS for the Run Mode and with symbols right arrow gt right triangle up triangle left arrow for the Menu Mode FRONT PANEL LOCKOUT The Menu Mode will not work with most meters as received from the factory since all menu items have been disabled in software and a lockout jumper is in place That jumper needs to be removed for the Menu Mode to work and menu items under
15. 200 kQ ranges and above measurements are susceptible to signal noise To minimize noise pickup the input signal wiring should utilize a shielded twisted pair and the shield should be connected to signal low and earth ground at the meter as illustrated Use 2 wire hookup 34 RTD amp RESISTANCE CONNECTION With the appropriate jumper settings RTD and resistance measurements allow 2 3 or 4 4 wire RTD or resistance Excitation wire RTD hookup to the J5 connector as illustrated The meter applies a fixed excita tion current cater 5 Tf Signal input Signal input 4 3 wire RTD or resistance In 4 wire hookup lead resistance is not a factor since different pairs of wires are used for excitation and sensing Excitation Excitation Signal input 3 Signal input 4 In 3 wire hookup the meter automatically compensates for lead resistance by measur ing the voltage drop in one current carrying lead and assuming that the voltage drop in the other current carrying lead is the same In 2 wire hookup the meter senses the voltage drop across the load and both lead wires The effect of the lead wires can be measured and subtracted by shorting out the load during meter setup The short should be as close as possible to the load Ambient temperature changes will still cause some error in the readings the higher the lead resistance the greater the error 2 wire RTD or resistance Excitation
16. Scaling method Scale and Offset if selected under SCALE Scale factor Offset value eee A Select E thru Y for flashing first digit thru 2 WV Wu for other flashing digits Select decimal point 0 0000 location when decimal point is flashing Select digit to flash OMe Select thru for flashing first digit thru B 0 000080 000 for other flashing digits Decimal point Select digit to flash location is selected by Scaling method Coordinates of 2 points if selected under Low signal input Desired reading at Lo In High signal input Desired reading at Hi In Sm WT Select Ej thru 9 for flashing first digit I thru B 0 000020 0000 for other flashing digits Decimal point is set Select digit to flash by input range chosen SETS SN Select E thru E for flashing first digit J thru 2 0 000020 0000 for other flashing digits Decimal point is set Select digit to flash by 15 eye SET Select Y for flashing first digit thru 2 0 00000 0000 for other flashing digits Decimal point is set Select digit to flash by input range chosen OS Select Ky thru for flashing first digit thru g 0 000080 0000 for other flashing digits Decimal point is set Select digit to flash by Scaling method Reading coordinates of 2 points if selected under Low signal input
17. amp 4 if detected 00000 Relay alarm state when alarm is active 00000 Alarm latching or non latching auto reset 00000 Alarm operates at and above setpoint active high or at and below setpoint active low 00000 Hysteresis mode or band deviation mode 00000 Number of consecutive readings in alarm zone to cause an alarm 00000 Relay alarm state when alarm is active 00000 Alarm latching or non latching auto reset 0 1 2 3 4 5 6 7 8 EANA 40 Relay 1 active on Relay 1 active off Relay 1 active on Relay 1 active off AL1 non latching AL1 latching AL1 non latching AL1 latching AL1 active high AL1 active low AL1 disabled AL1 active high AL1 active low AL1 disabled AL1 active high AL1 active low AL1 disabled AL1 band deviation AL1 split hysteresis AL1 band deviation AL1 split hysteresis Press Digit Press Value Select Select Key A Key Relay 2 active on Relay 2 active on Relay 2 active off Relay 2 active off AL2 non latching AL2 non latching AL2 latching AL2 latching AL2 active high AL2 active high AL2 active high AL2 active low AL2 active low AL2 active low AL2 disabled AL2 disabled AL2 disabled AL2 band deviation AL2 band deviation AL2 split hysteresis AL2 split hysteresis No deviation or hysteresis in menu AL1 span hysteresis AL2 band deviation AL1 span hysteresis AL2 split h
18. analog Moving average 0 15 sec output data output Moving average 0 3 sec E Moving average 0 6 sec Moving average 1 2 sec 4 Moving average 2 4 sec Moving average 4 8 sec Moving average 9 6 sec IN Unfiltered HAE d dddd ecimal point Decimal point flashes selection if ohms are selected under SCALE 11 Select E thru for flashing first digit I thru B Scale factor WV Wu for other flashing digits Select decimal point 0 0000 location when decimal point is flashing Select digit to flash OFFst UNS epp WT Select E thru Y for flashing first digit thru 2 Offset value WW voee for other flashing digits Use offset for display Select digit to flash in Rankine or Kelvin Decimal point location is selected by 101 Option board dependent menu items STET STURT STSETI PISIS Menu items related to alarm setup These will only appear if a relay board is detected If so please see Section16 AnSEt An Hi Menu items related to analog output setup These will only appear if an analog output board is detected If so see Section 17 Menu items related to serial communications These will only appear if an RS232 RS485 or USB I O board is detected If so see Section 18 Menu lockout items Menu items used to enable or lock out hide other menu items Loc menu items may in turn be locke
19. capabilities namely linearization of nonlinear inputs and display of rate of change from successive readings Full Scale The maximum input signal range for which the meter has been configured For example the most sensitive full scale for the load cell meter is 20 mV signal range from 20 mV to 20 mV Function Reset A rear panel control input which resets Peak Valley and any latched alarms when the input is tied to logic ground by a switch or OV is applied logic level true To reset the value again the input must be open or 5V applied logic level false and then set low Ground Loop A closed conductive path in external ground wiring that allows stray currents to flow in ground wiring creating ground noise The meters in this manual minimize ground loop problems by mutually isolating the grounds associated with meter power signal input and all output and communication options Jumper A push on component which provides a short between two adjacent posts on a circuit board Jumpers are used to configure signal conditioner boards for specific signal types and full scale ranges and to configure power supply and communications boards for various modes of operation Unused jumpers are stored by pushing one side over an unused post Hysteresis Split Relay operation is specified symmetrically Heater Heater Heater around setpoint The relay closes OFF OFF 4 OFF AE re opens when the reading rises above the setpo
20. low input desired reading at low input high input and desired reading at high input The meter then applies a straight line fit The decimal point is set by the separate dEC Pt menu item 62 Scaling Scale and Offset Method A scaling method where scale and offset are entered manually Scaling Reading Coordinates of 2 Points Method Setpoint Span Tare A scaling method where the low and high input values are determined from actual signals A known low signal is first applied to the meter That signal is captured as the low input value and the desired low reading is entered A known high signal is then applied That signal is captured as the high input value and the desired high reading is entered The meter then applies straight line fit This scaling method has the advantage of calibrating the transducer and meter as a system The actual voltage or current at either point does not need to be known The decimal point is set by the separate dEC Pt menu item A value compared to the reading to determine the state of a relay Term often used interchangeably with alarm setpoint The relay action can by latching or non latching utilize a hysteresis band or utilize a deviation band Hysteresis bands and deviation bands are specified by two symmetrical limits around the setpoint The number of counts corresponding to a given signal range A rear panel input which causes the display to be set to zero when the input is
21. 2000 0 ohms under 4 NT 0 to 20000 ohms 0 to 200 00 kohm 00 00 0 12 RTD or 4 1 2 digits for ohms Meter Setup Display selection with f 5 digit remote display 199 999 scale factor of 1 0 01 RTD 4 1 2 digit ohms count by 10 1 RTD or 3 1 2 digits for ohms 00 00 Power line frequency 00 00 Scale and offset method RTD amp ohms Scaling method Coordinates of 2 points 1 Reading coordinates of 2 points ohms Reset 2 Meter Hold Function Reset 2 Peak or Valley Hold 2 Peak or Valley Display Hold 2 Tare Peak or Valley Display 2 Tare Tare 2 Reset decimal point 1 XXXXX decimal point 1 XXXX X decimal point 1 XXX XX decimal point 1 2 XX XXX decimal point 2 2 XXXX X decimal point 2 XXX XX decimal point 2 2 XX XXX decimal point 2 X XXXX 00 00 Control inputs 1 amp 2 True logic 1 OV or tied to digital ground False logic 0 5V or open 0 1 0 1 0 1 0 1 lt 00 00 36 Wits Press Menu Press Digit icc Press Value Select Select Key Select Key y Key 00 00 1 Function Reset 2 Display Blank Meter Setup Control inputs 1 amp 2 1 Hold 2 Display Blank continued continued IN 1 Peak or Valley 2 Display Blank 1 Tare 2 Display Blank 1 Valley Display 2 Peak Display 1 2 Tare Reset Both control inputs 1 amp 2 set to 1 for selec tions M Function Reset
22. 9 14 O71 G0 NO 1 After 1 reading After 2 readings After 4 readings After 8 readings AL4 active high AL4 active high AL4 active high AL4 active low AL4 active low AL4 active low AL4 disabled AL4 disabled AL4 disabled AL4 band deviation AL4 band deviation AL4 split hysteresis AL4 hysteresis After 16 readings A After 32 readings After 64 readings After 128 reading Select E thru for flashing first digit thru 2 for other flashing digits Alarms will activate above the setpoint by the value entered and deactivate below the setpoint by the value entered Select E thru for flashing first digit thru 2 for other flashing digits Alarms will activate above and below the setpoint by the value entered and will deactivate in the middle of the band Select E thru for flashing first digit I thru 2 for other flashing digits Alarms will activate above the setpoint by the value entered and deactivate below the setpoint by the value entered Select E thru flashing first digit I thru 2 for other flashing digits Alarms will activate above and below the setpoint by the value entered and will deactivate in the middle of the band 17 ANALOG OUTPUT OPTION An optional analog board may be installed in the meter at rear panel jack position J4 adjacent to the signal conditioner board Once installed this board is recognized by the meter which will bring up the appropriate menu
23. Board Full Duplex Operation b amp e Bias jumpers should be installed on 1 board a amp d Installed on last meter in long cable run RS485 Modbus Board Half Duplex Operation b amp e bias jumpers installed on 1 board c amp f installed for half duplex operation a installed on last meter in line with long cable runs Note Board is shipped with no jumpers installed Ethernet to RS485 Converter Board amp USB to RS485 Converter Board Full Duplex Operation No jumpers for short cable runs Add b amp d for long cable runs Half Duplex Operation a amp c Installed for half duplex operation d Installed on last meter in line with long cable runs 44 RS 232 seufugs mam ndn mdnmCm nm m ufu mam Modbus Ethernet to RS485 mam USB RS485 ade SERIAL CONNECTION EXAMPLES 999999 SPORE Meter with Ethernet USB or RS232 serial interface board Host computer 5485 hub via RJ11 connectors Ethernet or USB cable and 6 wire data cables 999999 999999 999 999 MINE Mri Host computer Meter with RS485 server Meters with RS485 I O boards board Ethernet to serial each with two RJ11 connectors or USB to serial Ethernet cable 999999 Ethernet cable Router or switch Meter with Ethernet board with RJ45 9 9 99 9 9 Meter with Ethernet HANARA to serial server board RJ45 t
24. Boat Select digit to flash by eye SET Select Y for flashing first digit thru 2 High signal input 0 000080 0000 for other flashing digits Decimal point is set Select digit to flash by input range chosen eee Opp Select E thru Y for flashing first digit thru 2 Desired reading at for other flashing digits Decimal point is set pa Select digit to flash by Scaling method Reading coordinates of 2 points if selected under Low signal input Apply a low reference Press A to store the low signal input in the signal to the meter meter High signal input Apply a high reference Press A to store the high signal input in the signal to the meter meter Lo rd 0 000020 00000 00002 0 0000 Desired reading at Select E thru E for flashing first digit thru 2 Lo In Select digit to flash for other flashing digits Decimal point is set by Hi rd S Su Desired reading at NA Select E thru E for flashing first digit i thru 2 Hi In Select digit to flash for other flashing digits Decimal point is set by 18 Option board dependent menu items 28102 STHETI Menu items related to alarm setup These will only appear if a relay board is detected If so please see Section16 AnSEt An Hi Menu items related to analog output setup These will only appear if
25. Peak Operation of front Peak or Valley at connector above panel PEAK button 1 Valley Display Also selects Valley in and rear connector for Peak or Valley at connector above Peak or Valley Display Peak 1st push Valley 2nd push B Front 5 000 0 Meter comes up in normal run mode Auto tare Meter comes up in auto tare mode 000 0 Linear input Nonlinear input scaling Custom curve linearization Extended meter only FiLt 00000 1 Unfiltered output Filtering Alarm filtering 1 Filtered output 00000 0 Unfiltered Peak amp Valley Peak amp Valley filtering f Filtered Peak amp Valley 00000 0 Display batch average every 16 readings Display filtering Display filtered signal Adaptive filter 0 Low adaptive filter threshold level threshold High adaptive filter threshold level 00000 Input signal filtering Can be applied to dis play setpoint analog output data output Autofilter Batch average 16 readings Moving average 0 08 sec Moving average 0 15 sec Moving average 0 3 sec Moving average 0 6 sec Moving average 1 2 sec Moving average 2 4 sec Moving average 4 8 sec Moving average 9 6 sec Unfiltered dEc Pt d dddd TE ER Dec point selection Decimal point flashes 22 00 gt Press Menu md Select Key 4444 Press Digit M3 Press Value Select Select Key Key
26. and offset method Scaling method Coordinates of 2 points method Reading coordinates of 2 points method Reset 2 Meter Hold Function Reset 2 Pk or Valley Disp Meter Hold 2 Pk or Valley Disp Meter Hold 2 Tare Peak or Valley 2 Tare Ta 2 Reset 0 decimal point 1 0 decimal point 1 XXXX X 1 decimal point 1 XXX XX 1 decimal point 1 0 0 1 00 00 Rear connector control inputs 1 amp 2 True logic 1 OV or tied to digital ground False logic 0 5V or open s 0 2 1 2 0 2 1 2 0 2 1 2 0 2 1 2 Fun N decimal point 2 XXXX X decimal point 2 XXX XX decimal point 2 XX XXX 1 decimal point 2 X XXXX ction Reset 2 Display Blank Hold 2 Display Blank Peak or Valley 2 Display Blank Tare 2 Display Blank 1 Valley Display 2 Peak Display 1 Tare 2 Tare Reset Both control inputs 1 amp 2 set to 1 for selec tions A M I Function Reset Both control inputs 1 amp 2 set to 1 for selec tions 3 7 amp 9 Meter Reset 1 1 1 1 1 1 1 1 1 1 1 73 1 1 1 1 1 1 O O 00 gt KO Oo 21 Wits Press Menu Press Digit vd Press Value Select mud Select Key Select Key 7 Key Not rate of change Meter Configuration E on asarateof Rate x0 1 change meter A 1 Extended meter only Rate x 10 Rate x 100 Rate 1000 Q Rate x 10000 000 0 Peak Display Also selects
27. digital ground is tied to pin 3 of P5 Signal Input Keep pin 3 close to earth ground to minimize common mode voltage or shock hazard at pins 4 5 amp 6 of P1 P5 SIGNAL INPUT DC amp PROCESS EXCITATION EXCITATION SIGNAL INPUT SIGNAL INPUT Signal Source 24V EXCITATION 1 24V EXCITATION 2 SIGNAL INPUT SIGNAL INPUT 4L DC RATIO EXCITATION 1 EXCITATION 2 SIGNAL SIGNAL 4L AC TRUE RMS VOLTAGE SIGNAL IN 1 2 20 200 600V NC 2 SIGNAL RETURN 3 AC NEUTRAL CURRENT OR LOW 4 VOLTAGE SIGNAL IN 6 Bridge Potentiometer 2 20 200 mA 5A 0 2V Note Control inputs 1 amp 2 of P1 are menu selectable LOAD CELL METER 10 V EXCITATION 1 SENSE 2 SIGNAL 3 SIGNAL 4 10 V EXCITATION 5 SENSE 6 RTD amp OHMS 2 WIRE EXCITATION 1 EXCITATION 2 SIGNAL INPUT 3 SIGNAL INPUT 4 RTD amp OHMS 3 WIRE EXCITATION 1 EXCITATION 2 SIGNAL INPUT 3 SIGNAL INPUT 4 RTD amp OHMS 4 WIRE EXCITATION 1 EXCITATION 2 SIGNAL INPUT 3 SIGNAL INPUT 4 THERMOCOUPLE 1 2 SIGNAL INPUT gt SIGNAL INPUT 4 P2 SETPOINT CONTROLLER DUAL MECHANICAL RELAY OUTPUTS N O CONTACT COMMON N C CONTACT ALARM 1 1 2 3 N OCONTACT 4 5 6 ALARM 1 ALARM 1 ALARM 2 ALARM 2 ALARM 2 COMMON N C CONTACT QUAD MECHANICAL RELAY OUTPUTS ALARM1 N OCONTACT 1 ALARM1 amp 2 COMMON 2 ALARM2 N OCONTACT 3 4 5 ALARM3 N OCONTACT
28. file 52 4 Equation entry mode where the coefficients of a polynomial Y K1X P1 K2X P2 K3 X P3 are entered Up to 20 terms are allowed An offset can be built into X You will be asked if your DPM has a revision of DPMAL or later You will normally select 2 yes since revision DPMA4L started to ship in August 2000 You will be asked to supply the following LOW X COORDINATE VALUE gt LOW INPUT MEASUREMENT VALUE HIGH X COORDINATE VALUE HIGH INPUT MEASUREMENT VALUE gt This informs the computer of your signal conditioner jumper settings Enter 0 and 0 for the two LOW values For HIGH X enter your signal conditioner jumper range in the same units of measure that you will be using in your RAW data input file Enter 20 for 20 mV or 20V Enter 200 for 200 mV or 200V Enter 5 for 5A AC or DC For HIGH INPUT MEASUREMENT VALUE enter 20000 except for 5A DC where you should enter 5000 Position of the decimal point from 6 52 XX XXX 42XXX XXX 32XXXX XX 2 XXXXX X 1 XXXXXX for DPMs the leading X is a blank Specify the same position that you specified in the JAJA decimal point menu selection Follow the steps on the screens to finish generating the custom curve When prompted to download the file to the meter select Y When prompted to set the meter to custom curve mode also select Y KEYPAD CONTROL You can take a meter in and out of custom curve linearization using the meter keypad From the
29. see Section 1 Ordering Guide 5 Curve exe software downloadable from the web at no charge RJ11 to DB9 RS 232 cable GETTING STARTED Download curve exe into the same directory that will contain your data files such as c curves Set the meter baud rate to 9600 To do so press the key to get to then set the entry to Set the meter address to 1 To do so press the lm key to get to Fay then set the entry to To execute the program from Windows simply double click on curve exe which is an executable file Follow the steps on computer screens which will prompt you and provide extensive information Pressing R Enter returns to the main menu You will be given the choice to enter your data in one of four modes 1 Text file entry mode with an X value in one column and a Y value in another There can be additional columns which are ignored The file must have a DOS name of up to 8 characters and the extension RAW There can be from 5 to 180 rows X is the input value and should be in the unit of measure for which the meter was set up such as mV V mA or A Y is the desired corresponding reading and can range from 99999 to 99999 with any decimal point 2 2 coordinate keyboard entry mode where an actual X input signal is applied and the desired Y reading is entered from the keyboard 3 2 coordinate file entry mode where an actual X input signal is applied and the desired Y reading is provided from a
30. to 60 per second 50 for 50 Hz operation are made possible by Concurrent Slope Conversion Pat 5 262 780 which integrates the signal over an AC power line cycle for maximum noise rejection High read rates provide accurate peak and valley capture and quick response for control applications An adaptive digital filter supplies a time constant for the encountered signal noise level yet responds rapidly to changes that exceed a selected threshold Self calibration occurs automatically after every 17th reading The standard power supply is a high efficiency switching unit that operates from AC or DC and allows the meters to be powered from worldwide AC without changes A low voltage supply is optional for power from 10 48V batteries or from 12 30 Vac Both supplies provide an isolated 5 10 or 24Vdc transducer excitation output The meter case conforms to the 1 8 DIN size standard It is made of high impact 94 0 UL rated plastic and is watertight to NEMA 4 IP65 when panel mounted Mounting is from the front of the panel and requires less than 110 mm behind the panel Power and signal wiring is via removable plugs conforming to UL61010C safety standards All output options are isolated from meter and power ground to 250 Vac Extended meter versions can linearize nonlinear inputs Up to 180 data points may be linearized by a computer program that stores setup parameters in nonvolatile memory Extended meters can also display rate of change for e
31. 25000 500 mV 50000 Notes 1 See Section 19 to select 10V excitation 2 Jumpers are 2 5 mm 0 1 in SCALE amp OFFSET SETUP For DC microvolt meters a scale factor of 1 and an offset of 0 are used for direct readings in microvolts or millivolts Decimal point selection does not affect the displayed digits For example 20 mV can be displayed as 20 000 mV or 20000 uV The decimal point is set separately For load cell applications scaling is normally set up from the front panel using the and keys but can also be set up via RS232 485 using special PC compatible setup software available at no charge The meter allows three scaling methods to be selected 1 Manual scale and offset 2 Coordinates of 2 points and 3 Reading coordinates of 2 points Please see the Glossary for an explanation of each method 20 KEYSTROKES FOR SETUP If the MENU Ww key does not work see Section 9 Enabling amp Locking Out Menu Items Press Menu Press Digit accu Press Value Select mud Select Key Select Key Key InPut Strn Selection of signal Strain ratiometric 20 50 100 250 500 mV FS i tt input type amp range Em DC millivolts EX 50 IET 250 ERES mV E T Eu 0 4 1 2 digit meter counts by 1 Meter Setup mE type fl 5 digit remote display 499 999 4 1 2 digit meter counts by 10 g 3 1 2 al meter 00 00 Power line frequency 00 00 Scale
32. 459 7 will change the display to Rankin The addition of a relay output board turns the thermocouple meter from a temperature indicator into an on off temperature controller Please see further manual sections for setup of the following features relay output Section 16 analog output 17 and communications 18 SIGNAL CONDITIONER BOARD SETUP VIA PNP transistor JUMPERS J5 X J K E N none T R S j Sig J5 Y Sig E3 E4 i Open Indication E3 Jumper Upscale h Downscale i Letters indicate jumper position Jumpers are installed on pins adjacent to letters Use 2 5 mm 0 1 jumpers Store spare jumpers on an unused jumper post 31 KEYSTROKES FOR SETUP If the MENU Ww key does not work see Section 9 Enabling amp Locking Out Menu Items Press Menu Press Digit Press Value Select mud Select Key Select Key Key Type J F or C Selection of signal Thermocouple Type K F or C input type amp range Type N F or C Type E F or C Type T F or C Type S F or C Type R F or C 00 00 0 1 degree resolution Meter Setup Display selection Remote display 99 999 0 01 degree resolution 1 degree resolution 00 00 Noise minimized for 60 Hz Power line frequency 1 Noise minimized for 50 Hz 00 00 Offset only for thermocouple input Scaling method 00 00 ontrol inputs 1 amp 2 True logic 1 OV or tied to digital ground
33. ALARM3 amp 4 COMMON ALARM4 N OCONTACT P3 SERIAL COMMUNICATIONS RS232 INTERFACE Computer RS485 INTERFACE FULL DUPLEX ISO GND BRX ARX ATX BTX ISO GND A RXDO B RXD1 TXD1 A TXDO ISO GND DUAL SOLID STATE RELAY OUTPUTS DCORAC NOTUSED DCORAC DCORAC NOTUSED DCORAC QUAD SOLID STATE RELAY OUTPUTS ALARM 1 ALARM 1 ALARM 1 ALARM 2 ALARM 2 ALARM 2 Ooh WN ALARM1 N OCONTACT ALARM 1 amp 2 COMMON ALARM2 N OCONTACT 1 2 3 ALARM3 N OCONTACT 4 ALARM3 amp 4 COMMON i ALARM4 N OCONTACT 6 P4 ANALOG OUTPUT UNIPOLAR CONNECTIONS 4 20 mA or 0 20 mA OUTPUT 4 mmm 0 10V OUTPUT 5 ISOLATED GROUND 6 BIPOLAR CONNECTIONS REFERENCE or RETURN 4 xm 10V to 10V OUTPUT 5 6 RS485 INTERFACE HALF DUPLEX ISO GND ATX ARX BTX BRX ISO GND B TX RXD1 A TX RXDO ISO GND 7 7 MECHANICAL ASSEMBLY REMOVING THE REAR PANEL First remove any connectors Use one hand to press in the two sides of the rear of the case and the other hand to press down the two protruding tab releases at the top of the rear panel see figure below This will unhook the rear panel from the case Retaining tab with tab release fosse Retaining tab with tab release Retaining tab 1 oe Retaining tab Rear Panel REMOVING THE ELECTRONICS With the rear panel removed grasp the power supply board to the left
34. CII command only 0000 Select gf thru d for addresses 1 thru 15 Meter address with Custom Select thru 8 with decimal point for ASCII protocol addresses 16 ty 31 00000 Full duplex Serial Setup 3 Half or full duplex Half duplex 00000 1 Standard continuous mode Special start amp stop char 1 Special start amp stop characters 00000 Normal RTS RTS mode i Single transmission 00000 0 Only at end of all items Termination characters At end of each item 00000 Serial Setup 3 Data sent in continuous continued mode Reading Peak Valley Reading peak Reading valley Reading peak valley 107 gt 00 NO S Serial Setup 4 Modbus ASCII gap timeout Custom ASCII Serial protocol Modbus RTU Modbus ASCII None 2 or more stop bits Parity Odd 1 or more stop bits Even 1 or more stop bits ELS Modbus Address Select digit to flash Select through for flashing digit Appears only if the Address range is 1 to 247 Modbus protocol is selected 47 19 EXCITATION OUTPUT amp POWER SUPPLY Three isolated transducer excitation output levels are available from the power supply board These are selectable via jumpers b c d e f in the upper right of the board as illustrated In addition the board provides three jumper positions for special features The same jumper locations apply to the universal power supply 85 264 Vac and to the low voltage power
35. EMS For security reasons and ease of meter operation any and all menu items may be disabled or locked out so that they are no longer directly accessible from the front panel Each function to be disabled is set to 1 in menu items Loc 1 Loc 2 or Loc 3 and each function to be enabled is set to 0 The top menu items Loc 1 Loc 2 and Loc 3 can in turn be locked out by installing an internal hardware jumper With the jumper installed the operator only has access only to enabled menu items With the jumper removed the operator also has access to menu items Loc 1 Loc 2 and Loc 3 SETTING HARDWARE LOCKOUT JUMPER To access the lockout jumper remove the rear panel per Section 9 and locate jumper a in the lower portion of the power supply board next to the input connectors see figure at right Jumpera SETTING SOFTWARE LOCKOUTS Enabled or Disabled When setting up the meter it may be necessary to Menu Items enable specific menu items by setting the corres ponding lockout digit to 0 Be sure to reset the lockout digit to 1 if you do not want the menu item to be 1 Input type selection changed by an operator 2 Meter setup configuration amp decimal pt 3 Filter selection Press the le MENU key until Loc 1 Loc 2 or Loc 3 4 Scale or Lo Hi input is displayed as desired Note the hardware lockout 5 Offset or Lo Hi reading jumper must be removed see above Loc 2
36. ETUP OF CONNECTED METER A setup file can be retrieved from the meter DPM gt Get Setup be edited View gt Setup be saved to disk File gt Save Setup be retrieved from disk File gt Open Setup and be downloaded into one or multiple meters DPM Put Setup Downloading of setup files from a PC can be a major time saving when multiple meters have to be set up in the same way You will find that nstrument Setup software is very user friendly with separate tab selectable windows for nput Display Scaling Filter Relay Alarms Communications Analog Out and Lockouts If the required hardware such as the analog output board is not sensed the corresponding tab will be grayed out ADDITIONAL FEATURES The Commands pull down menu allows you to execute certain meter functions by using your computer mouse You can reset individual meter functions display current or peak readings and enter numbers to be displayed remotely by the DPM The first position of a transmitted number must be a blank sign or sign Five digits and a decimal point must be transmitted Leading O s serve as blanks The Commands pull down menu will be grayed out unless a Get Setup has been executed 2900 1280 2875 1024 ew 2850 2825 768 E _______________ 2800 512
37. False logic 0 5V or open Reset 2 Meter Hold Function Reset 2 Peak or Valley Hold 2 Peak or Valley Display Hold 2 Tare CEACE CINA 0 decimal point 1 XXXXX 0 decimal point 1 XXXX X 1 decimal point 1 XXX XX 1 decimal point 1 XX XXX 0 decimal point 2 XXXX X 0 1 IU decimal point 2 2 XXX XX decimal point 2 2 XX XXX 1 decimal point 2 2 X XXXX unction Reset 2 Display Blank 2 7 Display Blank eak or Valley 2 Display Blank 2 Display Blank Valley Display 2 Peak Display 1 Tare 2 Tare Reset Both control inputs 1 amp 2 set to 1 for selec tions Function Reset Both control inputs 1 amp 2 set to 1 for selec tions 0 0 9 9 3 7 5 9 Meter Reset 2 2 2 2 2 2 2 2 LA LA LA 9 LA LA mm IA ICA so UIO W gt KO 32 Press Menu Press Digit Masai Press Value Select Select Key 7 Key md Select Key Merer Connguration Peak Display Also selects Peak in Operation of front Peak or Valley at rear connector panel PEAK button Valley Display Also selects Valley in and rear connector for Peak or Valley at rear connector Peak or Valley Display Peak 1st push Valley 2nd push Front panel Tare FiLt 0000 A Unfiltered output Filtering Alarm filtering 1 Filtered output 0 0 Unfiltered Peak amp Valley eak amp Valley filtering 1 Filtered Peak amp Valley 000 1 Display batch average
38. LAUREATE SERIES 2 DIGITAL PANEL METER OWNERS MANUAL wn 1 PEAK RESET ALARMS j 4 n 5 Now with Ethernet c Intertek 4006497 L LAUREL Electronics Inc 3183 G Airway Ave Costa Mesa CA 92626 USA Tel 714 434 6131 Fax 714 434 3766 Website www laurels com 1 1 ORDERING GUIDE Configure a model number in this format L20201DCV1 CBL01 ss Digital panel meter with DC Amperes RMS Volts screw terminal connectors DGA 2 0000 mA 1 200 00 mV Display Color 2 20 000 mA 72 2 0000 V 200 00 mA 3 20 000 V with green 5 000 200 00 V V RT Cn DPM with red LED 5 600 0 V 3 Extended DPM green LED RTDs 300 0 V C Extended DPM red LED P385C Pt100 202 to 850 C Note Extended adds rate of P385F Pt100 331 to 1562 F RMS Amperes change amp linearization of non P392C Pt100 202 to 850 C RMA 2 0000 mA linear inputs Not applicable to P392F Pt100 331 to 1562 F 2 20 000 mA thermocouple or RTD inputs N672C Ni120 80 to 260 C 200 00 mA N672F Ni120 112 to 500 F RMAA
39. Loc 1 Loc 2 and Loc 3 then need to be set to 0 via the front panel for these menu items to be unlocked See Section 9 The paragraphs below assume that all menu items have been unlocked MENU MODE KEY ACTION In the Menu Mode pressing a key momentarily advances to the next menu item Holding down a key automatically advances through multiple menu items for fast menu navigation KEYS IN RUN MODE MENU Key Pressing MENU from the Run Mode enters the Menu Mode Pressing MENU repeatedly will step the meter through the various menu items if these have not been locked out and then back to the Run Mode PEAK Key Pressing PEAK normally causes the peak value of the input signal to be EI displayed The peak display then blinks to differentiate it from the normal present value display Pressing PEAK again returns the display to the present value The PEAK key can also be programmed to display Valley alternating Peak or Valley or to Tare the reading to zero When Peak or Valley is selected periodic horizontals bars at the top of the display indicate Peak and periodic horizontals bars at the bottom indicate Valley RESET Key Pressing RESET with PEAK resets peak and valley values Pressing jc RESET with ALARMS resets latched alarms Pressing RESET with MENU performs meter reset same as power on Meter reset can also be applied via a rear panel connect or a serial ASCII command ww ALARMS Key Pressing ALARMS once displays the setp
40. Menu mode press the amp key to get to 99 then set the fifth digit to either normal linear operation or to custom curve operation This fifth digit will only be displayed with an Extended meter FILES USED OR CREATED BY CURVE EXE 1 RAW is the raw input file generated by all four data entry methods 2 DVD adds three columns from which the smoothness of the input data and obvious input errors can be judged The more data points and the smoother the data the better the curve fit 3 NUM lists Y readings prior to custom curve linearization and addition of the decimal point 4 CCF is an internal file used by the software 5 SIM lists simulated linearized meter readings and calculated corresponding errors 6 PRM contains the final hex data that is downloaded into the meter 53 22 METER CALIBRATION All analog input and analog output ranges of the meter have been digitally calibrated at the factory prior to shipment using calibration equipment certified to NIST standards Calibration constants are stored digitally in non volatile memory in EEPROM on the signal conditioner board and analog output board This allows the boards to be changed without requiring meter recalibration Digital calibration also eliminates much of circuitry that would be associated with analog calibration providing superior long term accuracy and stability If periodic recalibration is required the meter may be returned to the factory or to an
41. S485 digital interface board with dual 6 pin RJ11 jacks not our RS485 digital interface with dual 8 pin RJ45 jacks The dual 6 pin RJ11 jacks on the RS485 board are wired in parallel to allow multiple meters to be daisy chained using 6 wire data cables with no need for hand wiring or an RS485 hub The outer two wires are used for ground Use 6 wire straight through data cables not 4 wire telephone cables or crossover cables all the way from the device server to the last device on the RS485 bus Connect ATX to ATX BTX to BTX etc with no crossover as you go from device to device To connect a meter with a USB board to a Windows PC use a USB cable with Type A and Type B connectors Upon first connection your computer may display Found new Hardware and automatically download and install the required USB driver from the Internet If installation is not automatic download the driver file with a name like CDM v2 10 00 WHQL Certified zip from http www ftdichip com Drivers VCP htm Unzip it into its own directory and point to that directory as the location of the driver You will need to use Device Manager accessible from Control Panel to determine the com port Go down the device list and click on Ports COM amp LPT and USB serial port com Note the com port for use with communications to your meter then exit Control Panel If you later need to change the Com port right click on USB serial port com then on Properties Por
42. UNPACKING Your meter was carefully tested and inspected prior to shipment Should the meter be damaged in shipment notify the freight carrier immediately In the event the meter is not configured as ordered or the unit is inoperable return it to the place of purchase for repair or replacement Please include a detailed description of the problem 5 SAFETY CONSIDERATIONS A Warning Use of this equipment in a manner other than specified may impair the pro tection of the device and subject the user to a hazard Visually inspect the unit for signs of damage If the unit is damaged do not attempt to operate Caution e The unit must be connected to a Disconnect switch or a branch circuit breaker which must be in a suitable location e This unit must be powered by 85 264 Vac with the high voltage power supply option or 10 48 Vdc 12 32 Vac with the low voltage power supply option Verify that the proper power option is installed for the power to be used This meter has no On Off switch It will be in operation as soon as power is connected e The 85 264 Vac power connector P1 Pins 1 3 is colored Green to differentiate it from other input and output connectors The 12 32 Vac 10 48 Vdc power connector is colored Black e Do not make signal wiring changes or connections when power is applied to the instrument Make signal connections before power is applied If reconnection is required disconnect power before such wiring is attempted
43. an analog output board is detected If so see Section 17 SEr 1 JSEr 2 5 4 Menu items related to serial communications These will only appear if an RS232 RS485 or USB I O board is detected If so see Section 18 Menu lockout items Menu items used to enable or lock out hide other menu items Loc menu items may in turn be locked out by a hardware jumper Please see Section 9 Scaling method 2 Reading Coordinates of 2 Points Scaling Method will appear before all other Menu items including Decimal point is set by 19 12 LOAD CELL amp MICROVOLT INPUT Load Cell Strain Gauge and Microvolt meters utilize the load cell signal conditioner board which offers sensitivity to 20 mV full scale and 4 or 6 wire load cell connection This board needs to be configured via jumpers for the desired voltage range All signal ranges are factory calibrated with calibration factors stored in EEPROM The meter software recognizes the board and will bring up the appropriate menu items for it however it does not recognize the jumper settings Please see further manual sections for setup of the following features relay output 16 analog output 17 communications 18 and transducer excitation output 19 RANGE SELECTION VIA JUMPERS Ranges amp Display with Scale Factor 1 j Full scale Input Jumpers 20 mV 20000 X 50 mV 50000 100 mV 10000 250 mV
44. ance Measurement 0 Hz and 10 Hz to 10 kHz crest factor 3 0 2 0000 mA 0 1 pA 1000 0 200 000 20 000 mA 1 uA 10 0 0 2000 00 200 00 10 10 0 20000 0 200 00 5 000 A 0 019 20mA 9 20000 MO 600V ranges are only ETL certified to 300V 56 Load Cell Input 20 000 mV 50 000 mV 100 00 mV 160 9999910 0to 299 999 01 of FS 250 00 mV 500 00 mV Thermocouple Accuracy 0 003 of reading C 0 0015 of reading C for load cell meter cM cou A AET 0 2 uV C Reference Junction 1 C 10 C to 40 C Dual amp Quad Relay Options Power to Relay Option c nncnesesnnceuneeeseseeenneeeeaanase Powered by meter SUDO ML SUID oii Via front panel pushbuttons or RS232 485 Updat Rate 2 ETE 56 s at 60 Hz 47 s at 50 Hz Response to input signal Display update rate Input Signal selectable eeeeeeeseeee eese Filtered or unfiltered input signal Actuation Modes selectable Above or below setpoint latching or non latching disabled Output Time Delay selectable ce 1 to 128 readings Front Panel Enable Lockout Modes selectable 1 Display and change setpoints 2 Display but do not change set
45. and signal condi tioner board to the right and carefully slide the electronic assembly out through the rear of the case see figure below INSTALLING NEW OPTION BOARDS Options boards plug into the main board at the front of the meter These are plug and play and may be installed in the field They will be recognized by the software which will provide access to the menu items associated with that board If necessary remove rear panel knockouts for new boards Boards plug into connectors as follows Option Board Main Board Plug Rear Panel Jack Power supply J1 Relay board J2 Serial interface board J3 Analog output board J4 Signal conditioner board J5 Note Corresponding main board and option board connectors have the same number of electrical lines When an option board is correctly installed the top and bottom edges of the main board and option board are aligned REASSEMBLING YOUR METER Slide the electronics assembly into the case until the display board is seated flush against the front overlay Insert the bottom tabs of the rear panel into the case and then carefully align the board connectors with the openings in the rear panel If necessary remove any rear panel knockouts for new option boards that may have been installed Ensure that all option boards are properly aligned with the molded board retaining pins on the inside of the rear panel Once the rear panel is in place reinstall the input output screw clamp terminal
46. auses the meter to reinitialize and take tare reading when set up for auto tare Achieved powering up the meter by pressing the RESET and MENU keys simultaneously stepping through all top level menu choices grounding a rear panel connector or supplying an ASCII command rESEt is displayed briefly RS485 Half Duplex Serial communications implemented with two wires allowing data transmission in both directions but not simultaneously RS485 Full Duplex Run Mode Scale Scaling Serial communications implemented with four wires allowing data transmission in two directions simultaneously The normal operating mode of the meter where readings are taken as opposed to the menu mode A constant multiplier used to go from A D converter counts to displayed counts This is the slope term m in the straight line formula y mx b where y is the displayed reading in counts m is the scale factor x is the measured reading in counts and b is the offset For direct readout in milli volts or milli bamps scale is 1 High input high reading Low input low reading Offset b gt o B E o Ee o gt E a n ea o Full span 99 999 to 39 999 The process of setting scale and offset so 0 Counts x from A D converter that the meter reads properly in engineering Full span 32 768 to 32 768 units such as psi Scaling Coordinates of 2 Points Method A scaling method where four numbers are entered manually
47. average filter signal High threshold adaptive moving avg filter 00000 20070 signal filtering Can be applied to dis play setpoint analog output data output Autofilter Batch average 16 readings Moving average 0 08 sec Moving average 0 15 sec Moving average 0 3 sec Moving average 0 6 sec Moving average 1 2 sec Moving average 2 4 sec Moving average 4 8 sec Moving average 9 6 sec Unfiltered Dec point selection Decimal point flashes m IN gt 28 Maes Press Menu md Select Key 4444 Press Digit a4 Press Value Select Select Key Key Scaling method Scale and Offset if selected under SCALE Scale factor Offset value Select K thru for flashing first digit J thru E 0 melee WO for other flashing digits Select decimal point Select digit to flash location when decimal point is flashing Select E thru 8 for flashing first digit f thru 9 0 000020 0000 for other flashing digits Decimal point Select digit to flash location is selected by Scaling method Coordinates of 2 points if selected under Low signal input Desired reading at Lo In High signal input Desired reading at Hi In SW Select E thru 8 for flashing first digit thru 9 0 000080 0000 for other flashing digits Decimal point is set Select digit to f
48. d out by a hardware jumper Please see Section 9 38 16 DUAL OR QUAD RELAY OUTPUT OPTION An optional relay board may be installed in the meter main board at plug position P2 adjacent to the power supply board Four board versions are available 2 or 4 relays mechanical or solid state Once installed the relay board is recognized by the meter software or PC based Instrument Setup software which will bring up the appropriate menu items for the type of board These menu items will not be brought up if no relay board is detected Menu selections for relays 3 and 4 will not be brought up if the dual relay board is detected All relay boards offer a choice of operating modes normally off or on latched or non latched hysteresis band deviation band alarm based on the filtered or unfiltered signal and selectable number of readings in alarm zone to cause an alarm KEYSTROKES FOR VIEWING amp CHANGING SETPOINTS The Alarms key be used to step through and view setpoints while the meter conti nues to make conversions and performs setpoint control If the Peak key is pressed while a setpoint is displayed conversion stops and the setpoint can be changed After pressing ell you have 30 seconds or the meter reverts to the normal display To view setpoints menu item Loc3 digit 2 must have been set to 0 To change setpoints menu item Loc2 digit 2 must have been set to 0 Press Alarms Press Digit jc Press Val
49. e 0 20 mA 12V 0 600 Ohm load Current Compliance 0 10V 10 to 10V Output 2 mA 5 kOhm or higher load PFERDE IEEE Meter input accuracy 0 02 of full scale analog output MD EE EE 16 bit 1 part in 65 536 Fesponse TIC a ede ERE 50 60Hz update rate Scaling of Reading for Zero 99 999 to 99 099 Scaling of Reading for Full Scale 99 999 to 99 999 Isolation rating between signal common and analog 250V ac Insulation dielectric strength between signal common and analog output DI UE SHE AU IER RU M ts 3 5 kV ac for 5 sec 2 3 kV ac for 1 min Serial Interface Option USB RS232 RS485 RS485 Modbus boards Output TYPOS s RS232 RS485 RS485 Modbus USB MTM USB to RS485 converter Ethernet Ethernet to RS485 converter Power to Interface E eR C eC dees Powered by meter RS485 d ade aI Half or full duplex Rates seo eu en ead 300 600 1200 2400 4800 9600 19200 Serial Protocols treten ten Custom ASCII Modbus RTU Modbus ASCII selectable Signal Eevelsc Rot on Betttoutedtendicetis Meet RS232 RS485 USB Ethernet standards Isolatio
50. e Section16 AnSEt An Hi Menu items related to analog output setup These will only appear if an analog output board is detected If so see Section 17 SEr 1 JSEr 2 3 SER Addr Menu items related to serial communications These will only appear if an RS232 RS485 or USB I O board is detected If so see Section 18 Menu lockout items Menu items used to enable or lock out hide other menu items Loc menu items may in turn be locked out by a hardware jumper Please see Section 9 Scaling method 2 Reading Coordinates of 2 Points Scaling Method will appear before all other Menu items including Decimal point is set by 30 14 THERMOCOUPLE INPUT The thermocouple signal conditioner board used for temperature measurement be confi gured via jumpers for 7 thermocouple types each in a single range J K T E N S R The meter software recognizes the board and will bring up the appropriate menu items for it however it does not recognize the jumper settings Display in C or F and resolution of 1 0 1 or 0 01 are user programmable High resolution should only be used for relative readings not absolute readings Although available 0 01 resolution is not recommended for thermo couples Offset adjustment is available for thermocouples and is normally set to 0000 0 If C is selected entering an offset of 0273 2 will change the display to Kelvin If F is selected entering an offset of 0
51. eak Display Tare 2 Tare Reset Both control inputs 1 amp 2 set to 1 for selec tions A M I Function Reset Both control inputs 1 amp 2 set to 1 for selec tions 0 0 9 5 3 7 5 9 Meter Reset g 1 2 4 5 0 0 0 1 1 0 0 1 1 0 1 0 1 N DX aX SE GEE S a OEE LAOLIX IX a OIX IX LX O O 0 gt KO Oo 27 Press Menu 2314 Press Digit ucc Press Value Select Select Key Select Key y Key 000 0 Not rate of change Meter Configuration Operation as a rate of Rate x 0 1 change meter Rate x 1 Extended meter only Rate x 10 Rate x 100 Rate x 1000 Rate x 10000 Peak Display Also selects Peak in Operation of front Peak or Valley at connector above panel PEAK button Valley Display Also selects Valley in and rear connector for Peak or Valley at connector above Peak or Valley Display Peak 1st push Valley 2nd push Front E Tare Meter comes up in normal run mode Auto tare fl Meter comes up in auto tare mode 000 0 0 Linear input Nonlinear input scaling Custom curve linearization Extended meter only 00000 Unfiltered output 0 Filtering Alarm filtering 1 Filtered output 00000 0 Unfiltered Peak amp Valley Peak amp Valley filtering 1 Filtered Peak amp Valley 00000 0 Display batch average every readings Display filtering 1 Display filtered signal Filtering for AC Non adaptive moving
52. every 16 readings Display filtering 1 Display filtered signal MP Adaptive filter 1 Low adaptive filter threshold level threshold 1 High adaptive filter threshold level 00000 Autofilter Input signal filtering Batch average 16 readings Can be applied to dis Moving average 0 08 sec play setpoint analog Moving average 0 15 sec output data output Moving average 0 3 sec Moving average 0 6 sec Moving average 1 2 sec Moving average 2 4 sec Moving average 4 8 sec Moving average 9 6 sec JN Unfiltered ONS SET MWe Select E thru Y for flashing first digit thru 2 Offset value 0 000080 0000 for other flashing digits Use offset for display Select digit to flash in Rankine or Kelvin AIN gt Option board dependent menu items STSETI Menu items related to alarm setup if a relay board is detected Please see Section16 AnSEt JAn Lo Hi_ Menu items related to analog output if detected Please see Section 17 1 5 2 5 EET Menu items related to serial communications if detected Please see Section 18 Menu lockout items Menu items used to enable or lock out hide menu items Please see Section 9 33 15 RTD amp RESISTANCE INPUT The standard RTD and resistance signal conditioner board can be configured via jumpers for four RTD types DIN 1000 platinum ANSI 1000 platin
53. he latest reading when the accumulated difference between individual readings and the filtered reading exceeds that threshold Adaptive moving average filtering allows a meter to respond rapidly to actual changes in signal while filtering out normal noise The accumulated difference is also reset to zero when the latest reading has a different polarity than the filtered reading A low adaptive filter threshold is normally selected A high filter threshold should be selected if the signal has large transients Alarm Latched An alarm which stays actuated until reset Latched alarms can shut down machinery or a process when an operating limit has been exceeded or maintain an alarm condition until acknowledged by an operator LATCHED ALARM 5 na gt Alarm Non latched An alarm which changes state automatically when the reading rises above a specified limit and changes back automatically when the reading falls below a limit Setpoint Pi Autofilter A selectable digital filter mode which auto matically selects an appropriate moving aver age filter time constant from 0 08 sec to 9 6 OFF OFF sec for the encountered noise condition Meter Reading 8 Auto tare A selectable meter operating mode where NON LATCHED ALARM the first reading following power on or meter reset is used to zero the display Further readings are then relative to this new zero Batch Average Filte
54. hich uses precision circuitry to compute the root mean square of complex waveforms from 10 Hz to 10 kHz Accurate measure ments are obtained with spikes up to 3 times the maximum of each range The input can be AC coupled to read only the AC component such as ripple on a power supply or DC coupled to read AC plus DC The board needs to be configured via jumpers for the desired voltage or current range and for AC or DC coupling All signal ranges are factory calibrated with calibration factors stored in EEPROM The meter software recognizes the board and will bring up the appropriate menu items for it however it does not recognize the jumper settings These need to be set manually Please see further manual sections for setup of the following features relay output 16 analog output 17 and communications 18 Voltage Ranges AC or DC Coupling RANGE SELECTION VIA JUMPERS Coupling Type 1 Letters indicate jumper position Jumpers DC coupling for AC DC f are installed on pins adjacent to letters AC coupling for AC only none 2 Use 2 5 mm 0 1 jumpers 3 Store spare jumpers on unused jumper post METER SCALING Refer to the above tables for the full scale counts or displayed digits produced by the available full scale input ranges with a scale factor of 1 and an offset of 0 The decimal point 25 can be set f
55. igh digits amp 3 LED indicators CON 0 E n Red or green Rage ean ass ett 99999 to 99999 99990 to 99990 A to D Conversion Technique 5 262 780 Concurrent Slope Read 60 5 for 60 Hz NMR 50 5 for 50 Hz NMR Output Update Rate ne nelenee 56 s at 60 Hz 47 s at 50 Hz 5 nien eio 3 5 s at 60 Hz 3 s at 50 Hz Noise Rejection dd Gate sire Withstand 250Vac 3 5 kV ac for 5 sec 2 3 kV ac for 1 min CMR TOM DG TOU HZ et EE a 130 dB NMR aEDOU IDUTIZS 90 dB with minimum digital filtering Control Inputs CMOS TTL levels logic 0 tied to digital ground logic 1 open Logic 0 holds display and outputs Peak Valley In pub x cene Logic 0 displays peak valley value APSE Logic 0 offsets input value to zero Tare eae Logic 0 resets Tare value to zero 7
56. int plus one hysteresis limit and opens or closes when the reading falls below the setpoint less one hysteresis limit A narrow Setpoint 1000 hysteresis band can be used to minimize Hysteresis limit 50 relay chatter A wide hysteresis band can be Hysteresis band 100 used for control 60 Hysteresis Span Menu Mode Meter Hold Modbus Same relay operation as for split hysteresis Heater Heater Heater but specified differently Here the setpoint is OFF OFF OFF the upper control limit and the lower control 7 limit is the setpoint less the hysteresis band Hee The meter programming mode used for input 3 7 i and range selection meter setup and meter Setpoint 1050 configuration Entered into from the Run Hysteresis band 100 mode by pressing the MENU key The Menu mode can be locked out completely by a jumper A rear panel input which freezes the meter display and all meter outputs while that input is tied to logic ground by a switch or is held at OV resume operation when the input is allowed to float or is held at 5V logic level false An industry standard serial communications protocol which allows devices by different manufacturers to be digitally addressed by a PC on the same commu nication line with up to 247 digital addresses More complex than the Custom ASCII protocol which is also supported by these meters Moving Average Filter Offset Peak Dis
57. isplayed requires E 10to 10V voltage output tput angie eu pease NJ Analog output unfiltered Analog output filtering fl Analog output filtered NV V Om Select E thru for flashing first Low displayed value 0 000080 0000 digit thru E for other flashing for 0 mA 4 mA OV or Select digit to flash digits Decimal point location is 10V output fixed by e z eda selection An Hi Select E thru for flashing first High displayed value digit thru E for other flashing for 20 mA or 10V out Select digit to flash digits Decimal point location is put fixed by selection 18 SERIAL COMMUNICATION OPTIONS A serial communications board may be connected to the meter main board at plug position P13 middle position Available boards are RS232 RS485 with dual RJ11 connectors RS485 Modbus with dual RJ45 connectors USB USB to RS485 converter Ethernet and Ethernet to RS485 converter The dual connectors of RS485 boards are wired in parallel to allow daisy chaining of addressable meters without use of a hub Three serial communication protocols are selectable for all serial boards Custom ASCII Modbus RTU and Modbus ASCII A USB to RS485 converter board or an Ethernet to RS485 converter board allows a meter to be interfaced to a computer and be the device server for a network of up to 31 other meters on an RS485 bus while itself retaining all capabilities of a meter The remote meters need to be equipped with our R
58. items for it These will not be brought up if an analog output board is not installed The analog output can be a 0 20 mA 4 20 mA or 0 10V unipolar signal with respect to isolated ground or a bipolar 10V to 10V voltage signal with respect to a reference return line Unipolar or bipolar operation is selected by a jumper A unipolar current or voltage output is selected at the connector Unipolar 4 20 mA or 0 20 mA current is selected in software UNIPOLAR CONNECTIONS 4 20 mA 0 20 mA OUTPUT 1 0 10V OUTPUT 2 ISOLATED GROUND BIPOLAR CONNECTIONS REFERENCE or RETURN 1 1 10V to 10V OUTPUT 2 31 Unipolar current of voltage Jumper Bipolar 10 to 10 voltage Jumper b The low analog output 0 mA 4 mA OV or 10V may be set to correspond to any low displayed reading An Lo The high analog output 20 mA OV or 10V may be set to corres pond to any high displayed reading An Hi The meter will then apply a straight line fit between these two end points to provide an analog output scaled to the meter reading KEYSTROKES FOR SETUP If the MENU Ww key does not work see Section 9 Enabling amp Locking Out Menu Items Ui Press Menu Press Digit Select ju Press Value Select Select Key Key y 3 Key AnSEt EDT 0 20 mA current output Analog Output Setup Analog output signal fl 0 10V voltage output Press k until AnSEt selection 4 20 mA current output is d
59. lash by input range chosen A Select E thru 8 for flashing first digit 0 thru 8 0 00000 0000 for other flashing digits Decimal point is set Select digit to flash by 15 Select E thru for flashing first digit I thru E 950 for other flashing digits Decimal point is set Select digit to flash by input range chosen A SE Select thru flashing first digit J thru 9 0 000080 0000 for other flashing digits Decimal point is set Select digit to flash by Scaling method Reading coordinates of 2 points if selected under Low signal input High signal input Lo rd Desired reading at Lo In Desired reading at Hi In Apply a low reference Press A to store the low signal input in the signal to the meter meter Apply a high reference Press A to store the high signal input in the signal to the meter meter eee Vp epp S Select E thru E for flashing first digit thru 2 Select digit to flash for other flashing digits Decimal point is set by epp Sy Select E thru E for flashing first digit thru 2 Select digit to flash for other flashing digits Decimal point is set by 29 Option board dependent menu items 2 1 2 LST Menu items related to alarm setup These will only appear if a relay board is detected If so please se
60. lley 2 Display Blank Tare 2 Display Blank Valley Display 2 Peak Display Tare 2 Reset Tare to Zero Both inputs 1 and 2 set to logic 1 for selec tions 8 Function Reset Both inputs 1 and 2 set to logic 1 for selec tions 1 I 8 3 3 9 Meter Reset 0 0 0 1 0 F 10 C 00 gt 303 CLA L A IA IA LA LA LA LA IA LA IA LA IA lA lA lA 16 Press Menu 82 Press Digit Press Value Select aud Select Key Select Key y Key 000 0 Not rate of change Meter Configuration Operation as a rate of I Rate x 0 1 Rate x 1 change meter 8 Rate x 10 Rate x 100 Extended meter only Ratex1000 f Rate x 10000 000 0 Peak Display Also selects Peak in Operation of front Peak or Valley at connector above panel PEAK button Valley Display Also selects Valley in and rear connector for Peak or Valley at connector above Peak or Valley Display Peak 1st push Valley 2nd push Front H Tare 000 0 Meter comes up in normal run mode Auto tare fl Meter comes up auto tare mode 000 0 Linear input Nonlinear input scaling Custom curve linearization Extended meter only FiLt 00000 Unfiltered output Filtering Alarm filtering fl Filtered output 00000 0 Unfiltered Peak amp Valley Peak amp Valley filtering Filtered Peak amp Valley 00000 0 Display batch average every 16 readings Display filtering Display filtered signal
61. momentarily tied to logic ground by a switch or is held at OV logic level true When the input is allowed to float or is held at 5V logic level false the meter displays readings relative to this new zero A common application is in weighing where an external Tare button is pressed to read the weight of an empty scale tare and tare is then automatically subtracted as a constant from gross weight for display of net weight Tare can also be used for other appli cations where a reading relative to starting point is desired Valley Display Zero The minimum or most negative reading since that minimum was last reset Reset can be via the meter front panel an external input or a software command The displayed value can reflect the filtered or unfiltered readings When used with process meters zero is an adjustment so that a given low transducer output reads zero on the meter Zero is adjusted by programming offset 63 25 WARRANTY Laurel Electronics Inc warrants its products against defects in materials or workmanship for a period of one year from the date of purchase In the event of a defect during the warranty period the unit should be returned freight prepaid and all duties and taxes by the Buyer to the authorized Laurel distributor where the unit was purchased The distributor at its option will repair or replace the defective unit The unit will be returned to the buyer with freight charges prepaid by the distribu
62. n rating between signal common serial 250V ac Insulation dielectric strength between signal common and serial I O LI 3 5 kV ac for 5 sec 2 3 kV ac for 1 min Option Board Connectors Etui idi inu Single RJ11 jack nier GS p Two RJ11 jacks for daisy chaining with 6 wire data cables RS485 Modbus Two RJ45 jacks for daisy chaining with 8 wire data cables Bose ci Lum RA cns xtA A eri USB type B plug USB to RS485 converter USB type B plug plus RJ11 jack to RS485 bus iplo eM Single RJ45 to Ethernet Ethernet to RS485 converter RJ45 jack to Ethernet plus RJ11 jack to RS485 bus Environmental Operating Temperature caca dix eet etn ete RERO In DA uicta ql 0 C to 55 C Storage Temperature dnd ls estin d imd ded 40 C to 85 C Relative Humidity cce ienie 95 from 0 C to 40 C non condensing e r NEMA 4X IP65 from front when panel mounted not verified for UL SHOCK face cie Mh M 10 G at 1 kHz applied in X Y Z axes VIDIAUOMN osten des 15 Hz to 150 Hz 1 mm to 2 mm amplitude 20 G max 58 24 GLOSSARY OF TERMS Adaptive Filter Threshold A threshold which causes an adaptive moving average filter to be reset to t
63. ner board which needs to be configured via jumpers for the desired voltage or current range All signal ranges are factory calibrated with calibration factors stored in EEPROM The meter software recognizes the board and will bring up the appropriate menu items for it however it does not recognize the jumper settings Please see further manual sections for setup of the following relay output 16 analog output 17 communications 18 and excitation output 19 Board Revision N Voltage Ranges Jumpers FS Input 200 00 mV 2 0000 V 20 000 V 200 00 V 300V UL 600V not UL Current Ranges FS Input 2 0000 mA 20 000 mA 200 00 mA 5 000 A Board Revision P shipped between 15 Jan 2012 and 1 May 2012 Voltage Ranges Jumpers 200 00 mV 2 0000 V 20 000 V 200 00 V 300V UL 600V not UL Current Ranges Jumpers 2 0000 mA 20 000 mA 200 00 mA 5 000 A 14 Board Revisions Q and R Voltage Ranges Jumpers 200 00 mV 2 0000 V 20 000 V 200 00 V 300V UL 600V not UL Current Ranges Jumpers ws ee pe 2 0000 mA 20 000 mA 200 00 mA 15 000 A Letters indicate jumper position Jumpers are installed on pins adjacent to letters Use 5 mm 0 2 jumpers for locations designated by a capital letter Use 2 5 mm 0 1 jumpers for locations designated by a lower case letter Store spare jumpers on an unused jumper post not associated a capital letter Boo NS
64. ng the transducer and meter as a system The actual voltage or current at either point does not need to be known This method is ideal for process and load cell meters which require zero and span adjustment It is also available for DC or AC meters It is not available with thermocouple or RTD meters The programming example below is for a process meter used with a 4 20 mA pressure transducer for O to 100 psi Decimal points are set separately using the dEC Pt menu Press Menu Select Press Digit Select accu Press Value Select mud Key Key Key Apply low signal input to display Press A to store low e g transducer output for O reading at low signal input reading psi e g 4 021 mA Apply high signal input Press A to store high e g transducer output for WEL Press gt to display reading known 100 00 psi source reading at high signal input e g 20 094 mA Lord 000 008000 00000 00 Select E thru Mode to enter desired low SW NW Select digit flashing first digit 0 thru reading e g 0 00 to flash other flashing digits 0010010 0 0300 UR Select 9 thru g for Mode to enter desired high WW Select digit flashing first digit thru reading e g 100 00 to flash other flashing digits 13 11 DC VOLTS AMPS PROCESS DC RATIO The DC Volts Amps Process and Strain meters utilize the DC signal conditio
65. ngs Process Signal A signal whose display requires setup of scale and offset settings for display in engineering units A classical process signal is 4 20 mA where the 4 mA and 20 mA end points can each correspond to a desired meter reading Rate of Change Meter A configuration mode of the Extended meter which allows the display of rate based on successive readings The conversion to engineering units is achieved with the combination of a multiplier from 0 1 to 10 000 and a scale factor 61 Reading The value displayed by the meter Taking a reading is the action of the meter to make an analog to digital conversion Readings are taken at 60 sec with 60 Hz power or 50 sec with 50 Hz power and are displayed with an update rate of 3 5 sec with 60 Hz power or 3 0 sec with 50 Hz power Remote Display Reset A display mode which allows the meter to serve as a remote display to another meter when connected to it by a 4 wire phone cord Also allows the meter to transmit raw measurement data to a computer and then display processed data from the computer A serial communications option board is required in the meter If such a board is not installed or no serial data is received the meter displays rESEt There are three types of Reset Peak and Valley Reset Achieved by simultaneously pressing the RESET and PEAK keys Latched Alarm Reset Achieved by simultaneously pressing the RESET and ALARMS keys Meter Reset C
66. o Ethernet RJ11 to RS485 bus Daisy chained 999999 6 wire dat EE EA 4 Mos SHAE Meters with RS485 I O boards each with RS485 EEEEEEB for daisy chained RS485 Host computer 45 Ethernet Internet amp modems Router Ethernet to serial server Router Host computer Transmitters KEYSTROKES FOR SETUP If the MENU k key does not work see Section 9 Enabling amp Locking Out Menu Items Press Menu Press Digit Select Maa Press Value Select mud Select Key Key y Key 000 Send unfiltered signal Fixed Parameters Output filtering fl Send filtered signal No parity 8 data bits __ 000 300 baud 1 stop bit Baud rate 600 baud 1200 baud 2400 baud 4800 baud 9600 baud 19200 baud 60 Hz 50 Hz Line frequency Line frequency 0 28 sec 0 34 sec 0 57 sec 0 68 sec 1 1 sec 1 4 sec 2 3 2 7 4 5 5 4 sec 9 1 sec 10 9 sec 18 1 sec 21 8 sec 36 6 sec 43 5 sec 72 5 sec 97 sec g 1 2 4 5 Output update rate NO gt 46 Wet Press Menu 4 Press Digit Select p Press Value Select mud Select Key Key Key 0000 0 No line feed after carriage return Serial Setup 2 ru feed Line feed after carriage return 0000 1 No alarm data Alarm data with readings 1 Alarm data with reading 0000 1 Continuous data output E of data output fl Data output on AS
67. octet pe epe ep aep eet p tete 31 15 RTD amp RESISTANCE INPUIT 34 16 DUAL amp QUAD RELAY OUTPUT OPTIONS eene eene 39 172 ANALOG OUTPUT OP TON de Er 42 18 SERIAL COMMUNICATION OPTIONS esee nnm enne 43 19 EXCITATION OUTPUT amp POWER SUPPLY seen rennen 48 20 WNSTISUMENT SETUP VIA PO s nuda cdd cdd de ta 49 21 CUSTOM CURVE emm 52 22 METER GALIBRATION tco te dette ter oto oce dept iecore 54 23 SSPEGCIFIOATIONS 55 24 GLOSSARY OF TERMS 59 295 WARRANTY dem cC 64 3 PRODUCT INTRODUCTION Our digital panel meters are versatile cost effective solutions to a wide variety of monitoring and control applications Depending on the choice of signal conditioner they are easily set up for an accurate display of temperature pressure flow weight voltage or current all in appropriate engineering units and with zero and span adjustment when needed Setup can be via front panel pushbuttons or the meter s serial interface Selective security lockout of the front panel keys protects against accidental changes to meter setup High read rates up
68. oint for Alarm 1 Pressing it meal again displays the setpoint for Alarm 2 Pressing it again returns to the present value 10 KEYS IN MENU MODE Right Arrow Key MENU Pressing steps the meter through all menu items that have been enabled and then back to the Run Mode With the DC signal conditioner board and no option boards available menu items are 1159 PIX NNI ni ETE ETE ETE if a change has been made to a menu item that change is saved to non volatile memory when the key is pressed next and is displayed briefly gt Right Triangle Key Digit Select Pressing from the nPut menu brings up all meter functions available with the meter s signal conditioner For the DC signal conditioner these are and Pressing from the SEtuP ConfFG FiLtr SCALE OFFSt Loc 1 Loc 2 or Loc 3 menus items sequentially selects digit positions 1 5 as indicated by a flashing digit 00000 00000 00000 00000 00000 Pressing gt from the dEC Pt menu item sequentially selects decimal point positions which will flash Up Triangle Key Value Select Pressing A for flashing item digit position or decimal point position will increment that item Pressing MENU will save any changes Left Arrow Key Reverse Menu Pressing has the same effect as the MENU key hau except that menu items are brought up in reverse order 11 9 ENABLING amp LOCKING OUT MENU IT
69. om the meter to a PC RS232 com port ora USB cable toa PC USB 11 40 DB9 RS232 cable with rear port see Section 1 Ordering Guide 5 Instrument Setup software view of DB9 connector to PC ESTABLISHING COMMUNICATIONS Connect the meter and PC Apply power to the meter Be sure that the meter is in Run Mode not Setup Mode To start the software from Windows click on Start gt Programs gt IS2 gt 152 Click on RS232 gt Establish The program will temporarily set the selected Com port to the required baud rate parity data bits and stop bit Once communications have been established click on Main Menu The software will sense the type of meter and installed boards but it cannot sense jumpers positions nor set jumpers for you If the computer is not connected to a meter select DPM and Series 2 ini x r Com Ports Baud Rate Protocol Select Instrument and Series Com1 9 300 Custom ASCI Panel Meter Transmitter Com2 600 C Modbus RTU dom C 1200 C Modbus ASCII DPM C Analog Input O Com C 2400 Help i C Com 5 C 4800 Weight Meter arity C Com 6 9600 Counter Timer Counter Timer C Com 7 C 19200 None Back C Com 8 C Com 9 3 Series _ Series 1 Other Com Port Series 2 Wait while attempting to establish communications Meter Type 49 1 x 101
70. or direct readout in milli volts or milli amperes Decimal point selection does not affect the counts For example a 20V input may be displayed as 20 000V or 20000 mV The 5A range designed for use with a 5A current transformer CT is scaled to produce 5000 counts with a scale factor of 1 and an offset of 0 Use with a specific CT will require the scale factor to be set For example for an 800A input 5A output CT set a scale factor of 1 6 This is the desired 8000 count display at 5A divided by the default 5000 count display at 5A Then set the decimal point to display to 800 0 at 5A All scaling methods applicable to DC process strain and load cell meters are available with AC RMS meters INTERNAL SHIELD To reduce noise pickup inside the meter or transmitter the RMS board is fitted with a flexible plug on shield If necessary This shield may be removed for jumper setting but must be reinstalled before closing the instrument SIGNAL SHIELDING Sig High RMS Signal Source SigLow Board Shield around twisted pair Earth Ground Shielding for noise reduction AC RMS measurements are susceptible to signal noise This is especially true when the instru ment has a wide bandwidth To minimize noise pickup the input signal wiring should utilize a shielded twisted pair and the shield should be connected to signal low at the meter as illustrated If signal low is close to earth ground such as within 2V signal low can fur
71. play A digital filter mode which displays a weighting moving average of readings Readings are taken at 60 sec with 60 Hz power and 50 sec with 50 Hz power Display update rates remain 3 5 sec with 60 Hz power and 3 0 sec with 50 Hz power There are eight moving average modes Old average x 1 2 new reading x 1 2 equivalent to 0 08 sec RC time constant Old average x 3 4 new reading x 1 4 equivalent to 0 15 sec RC time constant Old average x 7 8 new reading x 1 8 equivalent to 0 3 sec RC time constant Old average x 15 16 new reading x 1 16 equivalent to 0 6 sec RC time constant Old average x 31 32 new reading x 1 32 equivalent to 1 2 sec RC time constant Old average x 63 64 new reading x 1 64 equivalent to 2 4 sec RC time constant Old avg x 127 128 new reading x 1 128 equivalent to 4 8 sec RC time constant Old avg x 255 256 new reading x 1 256 equivalent to 9 6 sec RC time constant A constant adder used for the displayed reading This is the term b in the straight line formula y mx b where y is the displayed reading in counts m is the scale factor x is the measured reading in counts and b is the offset For direct readout in milli volts or milli bamps offset is 0 The maximum or most positive reading since that maximum was last reset Reset can be via the meter front panel an external input or a software command The displayed value can reflect the filtered or unfiltered readi
72. points 3 Neither display nor change setpoints Alarm Status IndICallOET 2001029042 25 0202240 008601 2 or 4 red LED lamps Status Indication Setup selectable Lit when output is ON or OFF or disabled Form C SPDT Relay Output AG IB scenes ot cec cutus cute uS cU LL Le 8A 240 Vac BDC Rating E FM NN 8A 24 Vdc Isolation rating between signal common and contacts 250 Vac Insulation dielectric strength between signal common and contacts Rel Stes Od Fes eae sd st A 3 5 kV ac for 5 sec 2 3 kV ac for 1 min Form A SPST Solid State Relay Output 120 mA 140 Vac EE E arene 120 mA 180 Vdc Isolation rating between signal common and contacts 250V ac Insulation dielectric strength between signal common and contacts Rect dine EE 3 5 kV ac for 5 sec 2 3 kV ac for 1 min Analog Output Option Power to Analog Output Powered by meter Output Levels 0 20 mA 4 20 mA 0 10V 10 to 10V Voltage Complianc
73. r A digital filter mode which averages 16 readings and then displays the average Readings are taken at 60 sec with 60 Hz power and 50 sec with 50 Hz power Counts The reading displayed on the panel meter ignoring the decimal point Custom ASCII Protocol A simplified short protocol for use with these panel meters It allows 31 digital addresses Not an industry standard protocol like the more complex Modbus protocol which is also offered with the meters Set oint Pass 2 p Fail Deviation Band Setpoint 1000 band in counts which controls relay action Deviation limit 50 symmetrically around a setpoint The relay Deviation band 100 actuates when the reading falls within the 59 deviation band and de actuates when the reading falls outside A limit e g 50 counts is set up around both sides of the setpoint to create a deviation band e g 100 counts Setting up a passband around a setpoint is often used for component testing Deviation limits are programmed by entering menu item dEU1b for Alarm 1 and dEU2b for Alarm 2 The deviation band equals two limits Display Blank A rear panel input which blanks the display when the input is tied to logic ground by a switch or when OV is applied logic level true The meter display will light when the input is open or is held at 5V logic level false Extended Meter A digital panel meter with an enhanced firmware that provides added
74. s Resolution Error at 25 C 55 DC Volts Thermocouple 0 1 1 resolution 200 00 mV 2 0000 V 20 000 V J 210 to 760 C 01 FS 0 09 C 347to 1400 F 01 FS 0 16 F K 244 to 1372 C 01 FS 0 10 C 408to 2501 F 01 FS 0 17 F 0 to 400 C 01 FS 0 03 C 257 to0 C 01 FS 0 20 C 32 to 752 F 01 FS 0 05 F 430 to 32 F 01 FS 0 36 F 200 00 V 300 0 V 0 1V 10 MQ 0 4 V 600 0 V 0 1V 10 MQ 0 4 V DC Amps 2 0000 mA 0 1 pA 100 Q 20 000 mA 1 pA 100 200 00 mA 10 pA Process DC Ratio Potentiometer Follower E 240 to 1000 C 01 FS 0 18 C 400 to 1830 F 01 FS 0 32 F N 245 to 1300 C 01 FS 0 10 C 410 to 2370 F 01 FS 0 17 F S 46 to 1768 C 01 FS 0 12 C 51 to 3213 F 01 FS 0 22 F 45 to 1768 C 01 FS 0 17 C 49 to 3214 F 01 FS 0 31 F RTD 0 01 0 1 1 resolution True RMS Volts 0 to 100 of Full Scale f 0 Hz and 10 Hz to 10 kHz crest factor 3 0 C OL ES 20 09 00385 331 to 1562 F 01 FS 0 05 F R ist 003925 331 to 1168 F 01 FS 0 07 F 200 00 mV 2 0000 V 20 000 V Ni 1200 80 to 260 C 01 FS 0 05 C 00672 112 to 500 F 01 FS 0 09 F Cu 100 97 to 260 C 01 FS 0 05 C 00427 143 to 500 F 01 FS 0 09 F 200 00 V 300 0V 100 mV 1 MQ 0 8 V 600 0 V 100 mV 1 MQ 0 8 V True RMS Amps 0 to 100 of Full Scale Resist
75. supply 12 32 Vac or 10 48 Vdc bc e d f Excitation output Jumper locations 5 5 100 mA max 10 Vdc 5 120 mA max 24 5 50 mA max SELECTION OF OTHER JUMPERS Jumper a Front panel menu lockout locked when installed See Section 9 Jumper g Provides 5V power output at P1 4 when installed Jumper h Connects Control Input 2 to P1 4 when installed 48 20 INSTRUMENT SETUP VIA PC Instrument Setup software is a PC program which is much easier to learn than front panel programming It is of benefit whether or not the meter is connected to a PC With the meter connected to a PC it allows uploading editing and downloading of setup data execution of commands under computer control listing plotting and graphing of data and computer prompted calibration With the meter unconnected to a PC it provides quick selection of jumper locations and a printable display of menu selections for front panel setup SOFTWARE INSTALLATION Download S2 exe onto your PC from the web or the distribution CD Double click on the downloaded file to unzip it into a special directory such as c temp Within that directory double click on setup exe which will install the software on your PC PREREQUISITES FOR CONNECTED USE 1 PC with an available RS232 or USB port 2 Meter to be set up 3 RS232 or USB board in the meter This board can be removed following meter setup 4 RJ11 to DB9 cable fr
76. t settings and Advanced Change port to the desired number click OK then exit Control Panel To connect a meter with an Ethernet board to a computer see our separate Ethernet Manual which covers our Node Manager Software This Windows based application runs on a host computer and is used to configure our Ethernet Nodes It automatically discovers all Nodes on a LAN or WAN plus any devices connected to Server Nodes via an RS485 bus It is used to configure each Node such as setting communication parameters naming the Node and associated devices entering email addresses for alarm notification and data requests selecting the Node s time zone for time stamping of emails and streaming data and upgrading firmware Once configuration data has been stored in flash memory of all Nodes Node Manager Software can be closed Node resident web server software is also provided BOARD SETUP VIA JUMPERS USB Board No jumpers required Basic Ethernet Board No jumpers needed RS232 Board e Pull up resistor on RTS line f Slave display to RS232 from another meter g Normal operation Note Board is shipped with jumpers e and g installed RS485 Board Full Duplex Operation b amp d Installed on last meter in long cable run RS485 Board Half Duplex Operation a amp c Installed for half duplex operation d Installed on last meter in line with long cable runs Note Board is shipped with no jumpers installed RS485 Modbus
77. the seven data columns correspond to values entered via the gt and A keys Click on any step in the sequence to bring up a detailed help window Click on Print for a hardcopy which you can then use as an instruction sheet to program your meter via its front panel Click on Main Menu gt File gt Save Setup As to save your setup to disk and have an electronic record 51 21 CUSTOM CURVE LINEARIZATION Curve exe is a DOS based executable PC program used to set up an Extended meter so that the readings have a user defined non linear relationship with the input signal The calculated linearizing parameters are downloaded into non volatile memory of the meter For example it allows a meter to correct for transducer nonlinearity or to display volume of an irregularly shaped tank based on liquid level The curve fitting algorithm uses quadratic segments of varying length and curvature and provides diagnostics to estimate curve fitting errors The program is self prompting avoiding the need for a detailed printed manual This manual section is only intended as intro duction and get started guide Tank Volume from Pressure at the Bottom of the Tank PREREQUISITES A520 TRA 1 PC compatible computer with an available RS232 or USB port 2 Extended meter 3 An RS232 or USB board in the meter This board can be removed following meter setup 4 An RJ11 to DB9 RS232 cable or a USB cable to connect the meter to the PC
78. ther be connected to earth ground at the meter as illustrated 26 KEYSTROKES FOR SETUP If the MENU Ww key does not work see Section 9 Enabling amp Locking Out Menu Items Press Menu Press Digit accu Press Value Select mud Select Key Select Key y Key InPut BA Selection of signal AC Volts 0 2 2 20 200 660V FS input type amp range FAC Al HID ES 2 20 200 mA 5A FS 0 4 1 2 digits 420 000 Meter Setup em selection with f Remote display 499 999 scale factor of 1 2 4 1 2 digits counts by 10 20 000 3 1 2 digits 2 000 00 00 Noise minimized for 60 Hz Power line frequency Noise minimized for 50 Hz 00 00 Scale and offset method Scaling method fl Coordinates of 2 points method Reading coordinates of 2 points method 00 00 Rear connector inputs 1 amp 2 True logic 1 OV or tied to digital ground False logic 0 5V or open Reset 2 Meter Hold Function Reset 2 Pk or Valley Disp Meter Hold 2 Pk or Valley Disp Meter Hold 2 Tare Pe or Valley 2 Tare Ta 2 Reset 0 decimal point 1 XXXXX decimal point 1 XXXX X decimal point 1 XXX XX decimal point 1 XX XXX decimal point 2 XXXX X decimal point 2 XXX XX 1 decimal point 2 XX XXX 2 1 decimal point 2 X XXXX Fu netion Reset 2 Display Blank Hold 2 Display Blank eak or Valley 2 Display Blank 2 Display Blank Valley Display 2 P
79. tions as shown under Connector Wiring Information page 6 15 KEYSTROKES FOR SETUP If the MENU Ww key does not work see Section 9 Enabling amp Locking Out Menu Items Press Menu Press Digit p Press Value Select mud Select Key Select Key EX dc U IBEX 2 Selection of signal DC Volts 2 2 20 200 660V FS input type amp range ET E END BE DC Amps 0 2 20 200 mA 5A FS Strain amp ratio NE 2 2 FS 0 4 1 2 digits 20 000 Meter Setup emm selection with f Remote display 499 999 scale factor of 1 4 1 2 digits counts by 10 20 000 E 3 1 2 digits 42 000 00 00 1 Noise minimized for 60 Hz line frequency 1 Noise minimized for 50 Hz Scale and offset method m method fl Coordinates of 2 points method A Reading coordinates of 2 points method 00 00 Control inputs 1 amp 2 True logic 1 OV or tied to digital ground False logic 0 5V or open Reset 2 Meter Hold Function Reset 2 Peak or Valley Hold 2 Peak or Valley Display Hold 2 Tare Peak or Valley Display 2 Tare Tare 2 Reset 1 decimal point 1 decimal point XXXX X 0 decimal point XXX XX 0 decimal point XX XXX 1 decimal point XXXX X 1 0 1O gt O71 G0 NO 1 1 1 decimal point XXX XX decimal point XX XXX 0 decimal point tion Reset 2 Display Blank Hold 2 Display Blank Pe ak or Va
80. tor LIMITATION OF WARRANTY The foregoing warranty shall not apply to defects resulting from 1 Improper or inadequate maintenance by Buyer 2 Unauthorized modification or misuse 3 Operation outside the environmental specifications of the product 4 Mishandling or abuse The warranty set forth above is exclusive and no other warranty whether written or oral is expressed or implied Laurel specifically disclaims the implied warranties of merchantability and fitness for a particular purpose EXCLUSIVE REMEDIES The remedies provided herein are Buyer s sole and exclusive remedies In no event shall Laurel be liable for direct indirect incidental or consequential damages including loss of profits whether based on contract tort or any other legal theory Copyright 1993 2015 Laurel Electronics Inc 15 June 2015 64
81. ue Select Lad Key Select Key y Key 300 24 200 00 295 00 Press Alarms Current setpoint 1 value blinks and To change setpoint 1 value to display Alarm 1 Alarm 1 LED indicator lights Press press A to change selected setpoint to select a digit which will blink blinking digits 395 00 395 00 305 00 Press Alarms Current setpoint 2 value blinks and To change setpoint 2 value to display Alarm 2 Alarm 2 LED indicator lights Press press to change selected setpoint to select a digit which will blink blinking digits 395 00 395 00 305 00 Press Alarms Current setpoint 3 value blinks and To change setpoint 3 value to display Alarm 3 Alarm LED indicator lights Press gt press to change selected setpoint to select a digit which will blink blinking digits 395 00 395 00 305 00 Press Alarms Current setpoint 4 value blinks and To change setpoint 4 value to display Alarm 4 Alarm 4 LED indicator lights Press gt press to change selected setpoint to select a digit which will blink blinking digits STET Press Alarms again Meter will reset and display current reading 39 KEYSTROKES FOR SETPOINT SETUP If the MENU Ww key does not work see Section 9 Enabling amp Locking Out Menu Items Press Menu mud Select Key ALSEt Alarm Setup for relays 1 amp 2 if detected Press lum until ALSEt is displayed ALS34 Alarm Setup for relays 3
82. um 1200 nickel 100 copper or five resistance ranges from 20 0000 to 200 00 A fixed 2 MQ resistance range R6 ordering option is provided by a factory modified signal conditioner board with component changes All ranges are factory calibrated with calibration factors stored in EEPROM on the signal conditioner board The meter software recognizes the board and will bring up the appropriate menu items for it however it does not recognize the jumper settings With RTDs display in C or F and resolution of 1 0 1 or 0 01 are user programmable SIGNAL CONDITIONER BOARD SETUP VIA JUMPERS Standard Board Modified Board E1 Jumper Pt100 Ni120 N A Cu10 20 0000 N A 0 200 000 N A 0 2 0000 kO N A 0 20 000 kO N A 0 200 00 kQ 0 2 0000 MO 2 or 4 wire none 3 wire g 1 Program decimal point selection under menu item 2 Letters indicate jumper position Jumpers are installed on pins adjacent to letters 3 Store spare jumpers on an unused jumper post SCALE amp OFFSET SETUP Scale is normally set to 1 0000 Scale can be used as an RTD correction when actual resistance is other than nominal as stated on the RTD calibration sheet For a Pt100 RTD divide 100 by the stated resistance at 0 C For example for a 99 04 ohm RTD scale should be set to 100 99 04 1 0097 SIGNAL SHIELDING Resistance m Sig High Ohms T Sigtiow Board Shield around twisted pair Shielding for noise reduction
83. xample to display flow rate based on changing tank level Alarm or setpoint control is provided by an optional relay board with two or four Form C 8A mechanical relays or two or four Form A 120 mA solid state relays The setpoints may be latching or non latching be energized above or below the setpoint or operate in a fail safe mode The relays can operate from the filtered signal to reduce relay chatter or from the unfiltered signal for fastest response Snubber circuits and a programmable relay switching time delay extend relay contact life An isolated analog output of 4 20 mA 0 20 mA 0 10V or 10 to 10V can be provided by an optional analog output board The output is linearized to the display and can operate from the filtered or unfiltered signal input It can be scaled via front panel pushbuttons or the meter s serial interface Ethernet USB RS232 or RS485 2 wire half duplex or 4 wire full duplex serial communi cations options are available with the Modbus protocol or a simpler custom ASCII protocol Modbus operation includes RTU or ASCII modes up to 247 digital addresses and up to 32 devices per RS485 line without a repeater Ethernet to RS485 and USB to RS485 converter boards allow a meter to be interfaced to a PC and to multiple meters on an RS485 network Meter programming can be via the meter s front panel or a PC running Windows based Instrument Setup Software serial interface option required 4 4 RECEIVING amp
84. y authorized distributor A modest fee will apply which also covers a Calibration Certificate DC load cell AC RMS and thermocouple signal conditioner boards can be calibrated using Instrument Setup Software running on a host PC Section 20 An RS232 RS485 USB or Ethernet interface board must be installed in the meter for communication to the PC This board may be removed following calibration A certified calibrated signal source is also required The Calibration screen of Instrument Setup Software is accessed by clicking on Calibration at the top of the DPM Main Menu screen The PC first recognizes the type of board then prompts you to apply specific jumpers and specific known signals for each range Press Repeat to take more readings When you have decided on which reading to accept press on the number 1 through 10 of that reading The RTD Ohms signal conditioner board cannot be calibrated using the Calibration screen of Instrument Setup Software Instead use the Scaling tab under Setup Here you can enter values for Scale and Offset for a specific range These corrections apply to resistance not to RTD temperature To calibrate RTD temperature refer to the published resistance table for your RTD type and calibrate resistance For example if your measured resistances are 0 1 low apply a scale factor of 1 01 54 23 SPECIFICATIONS Meter Display a 5 LED 7 segment 14 2mm 56 h
85. ysteresis AL1 span hysteresis AL2 span hysteresis After 1 reading After 2 readings After 4 readings After 8 readings Relay 3 active on Relay 3 active off Relay 3 active on Relay 3 active off AL3 non latching AL3 latching AL3 non latching AL3 latching After 16 readings EF After 32 readings After 64 readings After 128 reading Relay 4 active on Relay 4 active on Relay 4 active off Relay 4 active off AL4 non latching AL4 non latching AL4 latching AL4 latching Press Menu md Select Key ALS34 Alarm Setup for relays 3 amp 4 continued Alarm 1 hysteresis Alarm 2 hysteresis DEU1b Alarm 1 band deviation Alarm 2 band deviation Alarm 3 hysteresis Alarm 4 hysteresis Alarm 3 band deviation DEU4b Alarm 4 band deviation Press Digit Select Key 00000 Alarm operates at and above setpoint active high or at and below setpoint active low 00000 Hysteresis mode or band deviation mode see Glossary 00000 Number of consecutive readings in alarm zone to cause an alarm 0 0000 0 0000 0 0000 0 0000 Select digit to flash 0 0000 0 0000 0 0000 00000 0000 Select digit to flash elect digit to flash M Press Value Select Key AL3 active high AL3 active low AL3 disabled AL3 active high AL3 active low AL3 disabled AL3 active high AL3 active low AL3 disabled AL3 band deviation AL3 hysteresis AL3 band deviation AL3 hysteresis

Download Pdf Manuals

image

Related Search

Related Contents

Manual de instrucciones Balanza portátil  Tetenal Spectra Jet Wallpaper 170g  GPC 323    User Manual - Direct Healthcare Services  Block 1 User`s Manual.book  取扱説明書 ご使用前に、「取扱説明書」  Multilift MOG, MDG  Livret d`instructions  CPI 2000 - CB Gazette  

Copyright © All rights reserved.
Failed to retrieve file