Home

Technical Specification FLUXUS® F608**-A2

image

Contents

1. technical type PT12N PT12N PT12F PT12F order code 670415 1 670414 1 670415 2 670414 2 design short response time type Pt100 2x Pt100 matched Pt100 2x Pt100 matched according to EN 1434 1 according to EN 1434 1 connection 4 wire 4 wire measuring range C 30 250 50 250 accuracy T 0 15 C 2 10 T C class A 0 15 C 2 10 T C class A accuracy AT lt 0 1K 3K lt AT lt 6K lt 0 1K 3K lt AT lt 6kK more corresponding to more corresponding to EN 1434 1 EN 1434 1 response time S 50 8 housing aluminum PEEK stainless steel 304 1 4301 copper degree of protection IP66 IP66 according to IEC EN 60529 weight without con kg 0 25 0 5 0 32 0 64 nector fixation clamp on clamp on accessories thermal conductivity X x paste 200 C thermal conductivity x x foil 250 C plastic protection plate x insulation foam dimensions length mm 15 14 width b mm 15 30 height h mm 20 27 dimensional drawing A B 5 AUI r c1 El extension cable h dE i i T Connection Temperature Probe red blue red white blue white Connector pin cable of temperature probe extension cable 1 white blue blue 2 red blue gr
2. material FEP outer diameter mm 5 1 color black shield x BECEL A em lc ee Corre Pt DWR1NZ7 Perret TSFLUXUS_F608xx A2V1 5 1EN_Leu 2014 03 20 21 FLUXUS F608 A2 Technical Specification FLEXIM FLEXIM GmbH Wolfener Str 36 12681 Berlin Germany Tel 49 30 93 66 76 60 Fax 49 30 93 66 76 80 internet www flexim com e mail info flexim com Subject to change without notification Errors excepted FLUXUS is a registered trademark of FLEXIM GmbH TSFLUXUS_F608xx A2V1 5 1EN_Leu 2014 03 20
3. 4 TSFLUXUS_F608xx A2V1 5 1EN_Leu 2014 03 20 Technical Specification FLUXUS F608 A2 Typical Measurement Setup ATEX zone 2 power adapter outside of explosive atmosphere transmitter transducers temperature probes output adapter o outputs ATEX zone 2 ATEX zone 1 outside of explosive atmosphere ATEX zone 1 outputs transmitter transducers TSFLUXUS_F608xx A2V1 5 1EN_Leu 2014 03 20 FLUXUS F608 A2 Flow Transmitter Technical Data FLUXUS F608 A2 design portable ATEX zone 2 measurement measurement principle transit time difference correlation principle automatic NoiseTrek selection for measurements with high gaseous or solid content flow velocity 0 01 25 m s repeatability 0 15 of reading 0 01 m s medium all acoustically conductive liquids with lt 10 gaseous or solid content in volume transit time difference principle temperature compensation corresponding to the recommendations in ANSI ASME MFC 5 1 2011 accuracy with standard calibration with advanced calibration optional with field calibration 1 6 of reading 0 01 m s 1 2 of reading 0 01 m
4. FLEXIM Technical Specification FLUXUS F608 A2 Portable Ultrasonic Flow Measurement of Liquids in Hazardous Areas Portable instrument for non intrusive quick ultrasonic flow measurement with clamp on technology for all types of piping Features Precise bi directional and highly dynamic flow measure ment with the non intrusive clamp on technology High precision at fast and slow flow rates high tem perature and zero point stability Portable easy to use flow transmitter with 2 flow chan nels multiple inputs outputs an integrated data logger with a serial interface Extremely resistant carbon fiber housing Covered by ATEX IECEx zone 2 certification Compact and very lightweight allowing the measuring system to be easily carried as personal luggage e g for offshore visits Water tight resistant against oil many liquids and dirt Li lon battery provides up to 14 hours of measurement operation Automatic loading of calibration data and transducer detection for a fast and easy set up less than 5 min providing precise and long term stable results User friendly design Transducers available for a wide range of inner pipe diameters 10 6500 mm and fluid temperatures 40 200 C Rugged transducers ATEX IECEx zone 1 and 2 resis tant to rough environments dust and humidity Robust water tight IP67 transport case with compre hensive accessories HybridTrek automatically switches between transit time
5. 130 mineral grease paste type N coupling compound 990739 2 30 200 silicone paste type E coupling compound 990739 3 30 250 fluoropolymer paste type H TSFLUXUS_F608xx A2V1 5 1EN_Leu 2014 03 20 Technical Specification FLUXUS F608 A2 Connection Systems connection system NL l X y get rm mo e 5 mE c E UJ transducer frequency G H K M P Q S 3d character of transducer order code x y x y x y l x y l N cable length m 2 3 2 2 lt 1 2 1 1 1 lt 10 L cable length option LC m 2 7 ES 7 2 lt 10 8 1 ES X y transducer cable length max length of extension cable Transducer Cable Technical Data transducer cable extension cable type 1699 6111 1750 standard length m see table above 5 10 ambient temperature C 55 200 100 225 lt 80 sheath material stainless steel 304 1 4301 stainless steel 304 1 4301 stainless steel 304 1 4301 outer diameter mm 8 8 9 cable jacket material PTFE PFA PE outer diameter mm 2 9 2 7 6 thickness mm 0 3 0 5 0 5 color brown white black shield x x x TSFLUXUS F608xx A2V1 5 1EN Leu 2014 03 20 FLUXUS F608 A2 Technical Specification Clamp on Temperature Probe optional Technical Data
6. 5 S S9 E E Zo 152 x a 5 5E g va o 3 10 Ilo FS set of ultrasonic flow transducers for liquids measurement shear wave K 0 5 MHz 1 MHz Q 4 MHz normal temperature range E extended temperature range shear wave transducers with transducer frequency M Q A1 ATEX zone 1 IECEx zone 1 A2 ATEX zone 2 IECEx zone 2 NL with Lemo connector XXX cable length in m for max length of extension cable see page 19 connector outside of ATEX zone 1 IECEx zone 1 LC long transducer cable ATEX zone 2 IECEx zone 2 only FSK and ATEX zone 1 IECEx zone 1 example FS M N A2 NL 010 shear wave transducer 1 MHz normal temperature range ATEX zone 2 IECEx zone 2 connection system NL with Lemo connec tor and extension cable 10 m 12 TSFLUXUS_F608xx A2V1 5 1EN_Leu 2014 03 20 Technical Specification FLUXUS F608 A2 Technical Data Shear Wave Transducers zone 1 technical type CDK1NW1 CLK1NW1 CDM2NW1 CLM2NW1 CDQ2NW1 CLQ2NW1 order code FSK NA1NL FSK NA1NL LC FSM NA1NL FSM NA1NL LC FSQ NA1NL FSQ NA1NL LC transducer frequency MHz 0 5 1 4 inner pipe diameter d min extended mm 100 50 10 min recommended mm 200 100 25 max recommended mm 3600 2000 150 max extended mm 6500 3400 400 pipe wall thickness min mm max mm material housing PEEK with stainless steel cap and PEEK with stainless stee
7. IP65 IP65 IP65 according to IEC EN 60529 transducer cable type 1699 1699 1699 1699 length m 5 9 4 3 dimensions length mm 136 5 136 5 84 70 width b mm 59 59 40 30 height h mm 90 5 90 5 59 47 5 dimensional drawing E Sn lt c r c B T I o qi o lo o lo E o la E 35 8 lel ambient temperature min C 40 40 40 40 max C 130 130 130 130 temperature x x x x compensation explosion protection category gas 3G dust 2D gas 3G dust 2D gas 3G dust 2D gas 3G dust 2D EPL Gc Db Gc Db Gc Db Gc Db zone 2 21 2 21 2 21 2 21 explosion protection temperature pipe surface A min C Ts 55 55 55 max C 190 190 190 190 x marking C 0637 0 C 0637 I8 C 0637 jac C 0637 IISG Ex nA IIC T6 T2 Gc X ExnAIIC T6 T2 Gc X Ex nA IIC T6 T2 Gc X ExnA IIC T6 T2 Gc X E Ex tb IIIC TX Db Ex tb IIIC TX Db Ex tb IIIC TX Db Ex tb IIIC TX Db c certification ATEX IBExU10ATEX1163 X IBExU10ATEX1163 X IBExU10ATEX1163 X IBExU1OATEX1163 X E certification IECEx IECEx IBE 12 0005X IECEx IBE 12 0005X IECEx IBE 12 0005X IECEx IBE 12 0005X x type of protection gas non sparking gas non sparking gas non sparking gas non sparking dust protection by dust protection by dust protection by dust protection by enclosure enclosure enclosure enclosure TSFLUXUS F608xx A2V1 5 1EN Leu 2014 03 20 FLUXUS F608 A2 Technical Specification Shear Wave Transducers zone 2 extende
8. and NoiseTrek mode of measurement when high partic ulate flows are encountered QuickFix for fast mounting of the flow transmitter in diffi cult conditions Applications Designed for the following industries Upstream on and offshore Midstream and downstream pipelines and refineries Chemical industry Energy sector e g HVAC geothermal power plants TSFLUXUS_F608xx A2V1 5 1EN_Leu 2014 03 20 FLUXUS F608 supported by handle Measurement with transducers mounted by the portable Variofix VP Measurement with the flow transmitter fixed to the pipe by the QuickFix pipe mounting fixture FLUXUS F608 A2 Technical Specification Table of Contents Eunctioniv ond eo celle onda P LM cido sd oa tinted ri AM tn 3 Measurement Principles A A di 3 Calculation of Volumetric Flow Rate a a non nnn ona nn conc ana nn ases lends sas sadness nn 3 Number of Sovind Paths iaa tee da e do de Lo ios 4 Typical Measurement Setup tete aee rete A te edd rece nes eleif todas 5 Elow Tra nsmlitler 2 eee eb ea ied e ae aia dedu M a 6 Technical Pata wicks hice seit scele oa ae 6 DIMENSIONS 5 A aha ert ae deine aed a he ne ees ie be pet ee 8 Standard Scope Of SUPDIY cita Ui Ee RUNE RE te mea aces MIRI 8 Adapters Optional pte 9 Transducers isc ioicoihischeninuee ea eee ae ded sed eet aa No ht ee eed So pis a a 11 Transducer SCISCU ON 22 3 2 12 5 0 tieniti Tet saetetes sith iemie dubicee
9. q nA IIC T6 T2 Gb Gc Ex q nA IIC T6 T2 Gb Gc E Ex tb IIIA TX Db Ex tb IIIA TX Db C certification ATEX IBExU10ATEX1162 X IBExU10ATEX1162 X E certification IECEx IECEx IBE 12 0004X IECEx IBE 12 0004X X type of protection gas powder filling non sparking gas powder filling non sparking dust protection by enclosure dust protection by enclosure TSFLUXUS_F608xx A2V1 5 1EN_Leu 2014 03 20 Technical Specification FLUXUS F608 A2 Shear Wave Transducers zone 2 cap and transducer shoe 304 1 4301 cap and transducer shoe 304 1 4301 technical type CDK1NH1 CLK1NH1 CDM2NH1 CDQ2NH1 order code FSK NA2NL FSK NA2NL LC FSM NA2NL FSQ NA2NL transducer frequency MHz 0 5 0 5 1 4 inner pipe diameter d min extended mm 100 100 50 10 min recommended mm 200 200 100 25 max recommended mm 3600 3600 2000 150 max extended mm 6500 6500 3400 400 pipe wall thickness min mm max mm material housing PEEK with stainless steel PEEK with stainless steel PEEK with stainless steel PEEK with stainless steel cap and transducer shoe 304 1 4301 cap and transducer shoe 304 1 4301 contact surface PEEK PEEK PEEK PEEK degree of protection IP65
10. s 10 5 of reading 0 01 m s flow transmitter power supply 100 240 V 50 60 Hz power supply unit outside of explosive atmosphere 10 5 15 V DC socket at transmitter with power adapter optional integrated battery battery power consumption Li lon 7 2 V 4 5 Ah operating time without outputs inputs and backlight gt 14 h lt 6W number of flow measuring channels 2 signal attenuation 0 100 s adjustable measuring cycle 1 channel 100 1000 Hz response time 1 s 1 channel option 70 ms housing material PA TPS PC Polyester stainless steel degree of protection accord ing to IEC EN 60529 IP65 dimensions see dimensional drawing weight 1 9 kg fixation QuickFix pipe mounting fixture ambient temperature 10 60 C display 2 x 16 characters dot matrix backlight menu language English German French Dutch Spanish explosion protection category gas 3G dust 2D EPL Gc Db zone 2 21 marking 113G CE 0637 II2D E Ex nA nC ic IIC T6 T4 Gc X Ex tb IIIC T 100 C Db I Ta 10 50 60 C certification ATEX IBExU10ATEX1067 E certification IECEx IECEx IBE 12 0006 C type of protection gas non sparking E dust protection by enclosure x temperature inputs intrinsic safety intrinsic safety parameters Um 16 V DC intrinsically safe inputs U 22 V ly 6 mA P 33 mW C 450
11. time difference At is measured and allows the flowmeter to determine the average flow velocity along the propagation path of the ultrasonic signals A flow profile correction is then performed in order to ob tain the area averaged flow velocity which is proportional to the volumetric flow rate Two integrated microprocessors control the entire measuring process This allows the flowmeter to remove disturbance signals and to check each received ultrasonic wave for its validity which reduces noise HybridTrek If the gaseous or solid content in the medium increases occasionally during measurement a measurement with the transit time difference principle is no longer possible NoiseTrek mode will then be selected by the flowmeter This measurement method allows the flowmeter to achieve a stable measurement even with high gaseous or solid content The transmitter can switch automatically between transit time and NoiseTrek mode without any changes to the measurement setup to t t gt Path of the ultrasonic signal Transit time difference At Calculation of Volumetric Flow Rate U k A ka At 2 t where V volumetric flow rate kRe fluid mechanics calibration factor cross sectional pipe area Ka acoustical calibration factor At transit time difference tr B transit time in the medium TSFLUXUS F608xx A2V1 5 1EN Leu 2014 03 20 3 FLUXUS F608 A2 Technical Specification Number of Sound Paths The number
12. 08ST A22 3N GG FLUXUS F608ST A22 3N GG 2D II NN NN 2D II NN NN 2D II GG NN application flow measurement on liquids 2 independent measuring channels temperature compensated calculation of mass flow rate integrated heat flow computer for monitoring of energy flows simultaneous monitoring of flow simultaneous monitoring of 2 and energy flow energy flows e g heating Systems heat exchangers outputs passive current output 2 2 2 binary output 2 2 2 inputs temperature input 2 4 accessories transport case x x x power supply unit mains cable x x x battery x x x power adapter output adapter input adapter 2 QuickFix pipe mounting fixture x x x for transmitter serial data kit x x x measuring tape x x x user manual x x x safety instructions Quick Start Guide connector board at the upper side of the transmitter q GR Gx60 S 66 1 if required to be ordered separately TSFLUXUS F608xx A2V1 5 1EN Leu 2014 03 20 Technical Specification FLUXUS F608 A2 Adapters optional output adapter power adapter input adapter transducers transducers measuring channel A measuring channel B RS232 inputs outputs power supply unit battery charging unit Technical Data output adapter power adapter technical type OA608A2 PA608A2 connection voltage 10 5 15 V DC dimensions see dimensio
13. ay 3 4 5 not connected 6 red red 7 white white 8 not connected Cable cable of temperature probe extension cable type 4 x 0 25 mm black or white LIYCY 8 x 0 14 mm gray standard length m 3 5 10 25 max length m 100 cable jacket PTFE PVC 20 TSFLUXUS F608xx A2V1 5 1EN Leu 2014 03 20 Technical Specification FLUXUS F608 A2 Wall Thickness Measurement optional The pipe wall thickness is an important pipe parameter which has to be determined exactly for a good mea surement However the pipe wall thickness often is unknown The wall thickness probe can be connected to the transmitter instead of the flow transducers and the wall thickness measurement mode is activated automatically Acoustic coupling compound is applied to the wall thickness probe which then is placed firmly on the pipe The wall thickness is displayed and can be stored directly in the transmitter Technical Data technical type DWR1NZ7 measuring range mm 1 250 resolution mm 0 01 accuracy 1 96 0 1 mm medium temperature C 20 200 short time peak max 500 explosion protection cable type 2616 length m 1 5 The measuring range depends on the attenuation of the ultrasonic signal in the pipe For strongly attenuating plastics e g PFA PTFE PP the measuring range is smaller Cable type 2616 ambient temperature C lt 200 cable jacket
14. ction by dust protection by enclosure enclosure TSFLUXUS_F608xx A2V1 5 1EN_Leu 2014 03 20 Technical Specification FLUXUS F608 A2 Transducer Mounting Fixture Order Code 1 2 3 4 5 6 7 9 no of character o description 5 ES SE o gt 9 EL o vP v oE 25 35 2 59 c as Bes m oc o 2 53 lS 8 lg 8 gS EE E ES 2 35 VP portable Variofix A all transducers D reflection arrangement or diagonal arrangement R reflection arrangement M medium C chains N without fixation 055 10 550 mm example VP A D M C 055 portable Variofix and chains material stainless steel 304 1 4301 301 1 4310 303 1 4305 dimensions 414 x 94 x 76 mm chain length 2 m TSFLUXUS_F608xx A2V1 5 1EN_Leu 2014 03 20 FLUXUS F608 A2 Technical Specification Coupling Materials for Transducers normal temperature range 4th character of transducer order code N extended temperature range 4th character of transducer order code E lt 100 C lt 170 C lt 150 C lt 200 C coupling compound type N coupling compound type E coupling compound type E coupling compound type E or H Technical Data type order code ambient temperature material C coupling compound 990739 1 30
15. d temperature range technical type CDM2EH5 CDQ2EH5 order code FSM EA2NL FSQ EA2NL transducer frequency MHz 1 4 inner pipe diameter d min extended mm 50 10 min recommended mm 100 25 max recommended mm 2000 150 max extended mm 3400 400 pipe wall thickness min mm max mm material housing PI with stainless steel cap PI with stainless steel cap and transducer shoe 304 and transducer shoe 304 1 4301 1 4301 contact surface PI PI degree of protection IP56 IP56 according to IEC EN 60529 transducer cable type 6111 6111 length m 4 3 dimensions length mm 84 70 width b mm 40 30 height h mm 59 47 5 dimensional drawing Er lt o Ic I RS o9 la E gt o O lal ambient temperature min C 30 30 max C 200 200 temperature x x compensation explosion protection category gas 3G dust 2D gas 3G dust 2D EPL Gc Db Gc Db zone 2 21 2 21 A explosion protection temperature pipe surface T min RE is 45 F max ec 14235 235 markin 113G 113G i d CE 0637 i55 CE 0637 i55 E Ex nA IIC T6 T2 Gc X Ex nA IIC T6 T2 Gc X c Ex tb IIIA TX Db Ex tb IIIA TX Db g certification ATEX IBExU10ATEX1163 X IBExU10ATEX1163 X x certification IECEx IECEx IBE 12 0005X IECEx IBE 12 0005X type of protection gas non sparking gas non sparking dust prote
16. l cap and PEEK with stainless steel cap and transducer shoe 304 1 4301 transducer shoe 304 1 4301 transducer shoe 304 1 4301 contact surface PEEK PEEK PEEK degree of protection IP65 IP65 IP65 according to IEC EN 60529 transducer cable type 1699 1699 1699 1699 1699 1699 length m 5 9 4 9 3 9 dimensions length mm 136 5 84 70 width b mm 59 40 30 height h mm 90 5 59 47 5 dimensional drawing E e NE c lt e LE LI a LL E e xw 56 is 9m o la Sl ambient temperature min C 40 40 40 max C 130 130 130 temperature x x x compensation explosion protection category gas 2 3G dust 2D gas 2 3G dust 2D gas 2 3G dust 2D EPL Gb Gc Db Gb Gc Db Gb Gc Db zone 1 2 21 1 2 21 1 2 21 A explosion protection temperature pipe surface T min C 55 55 E max C 180 180 180 X marking 112 3G 112 3G 112 3G l CE 0637 amp 112D CE 0637 112D CE 0637 amp 112D l Ex q nA IIC T6 T2 Gb Gc Ex q nA IIC T6 T2 Gb Gc Ex q nA IIC T6 T2 Gb Gc E Ex tb IIIC TX Db Ex tb IIIC TX Db Ex tb IIIC TX Db C certification ATEX IBExU10ATEX1162 X IBExU10ATEX1162 X IBExU10ATEX1162 X E certification IECEx IECEx IBE 12 0004X IECEx IBE 12 0004X IECEx IBE 12 0004X X type of protection gas powder filling non sparking gas powder filling non sparking gas powder filling non sparking dust protection by enclosure dust protection by enclosure dust protection by enclo
17. n timi dreds d iilen fenis n cer st A 11 Transducer Order Gode rccte o tete eie ctc e Mb aee teta ubl t eec alode cabe tote Me coss Cte dae cad 12 Technical Data O oe fet il 13 Transducer Mounting Fixture sssssssssssssesee eene eene nennen nennen ener rn en nennen nennen nnns 17 Coupling Materials for TranSducerS ooooooooocononccononcooconococcnononcnnnnnnn non nn nncnnnnnnnn enne nnn nn nr rra enne 18 Connection Systems en ne ee eee r rene nenr sedere nnne nennen nnn 19 TRANSGUGER GADIe es E E EEEE EETA EE EE AE AEE E E A EAE AEE AE AE E E hans 19 Clamp on Temperature Probe optional ccccccceeeeeeeeeeeeeeecaeceeeeeeeeeeeeeeceaaaeaeeeeeeeeeeeseesesseeniaeeees 20 Wall Thickness Measurement optional ssssssseeenn emere nenne 21 2 TSFLUXUS F608xx A2V1 5 1EN Leu 2014 03 20 Technical Specification FLUXUS F608 A2 Function Measurement Principle Transit Time Difference Principle In order to measure the flow of a medium in a pipe ultrasonic signals are used employing the transit time dif ference principle Ultrasonic signals are emitted by a transducer installed on the pipe and received by a se cond transducer These signals are emitted alternately in the flow direction and against it As the medium in which the signals propagate is flowing the transit time of the ultrasonic signals in the flow direction is shorter than against the flow direction The transit
18. nF Lo 10 mH C 1 8 nF Li 10 uH 1 for transit time difference principle reference conditions and v 0 15 m s reference uncertainty 0 2 96 TSFLUXUS F608xx A2V1 5 1EN Leu 2014 03 20 Technical Specification Technical Specification FLUXUS F608 A2 FLUXUS F608 A2 measuring functions physical quantities volumetric flow rate mass flow rate flow velocity heat flow if temperature inputs are installed totalizer volume mass optional heat quantity calculation functions average difference sum diagnostic functions sound speed signal amplitude SNR SCNR standard deviation of amplitudes and transit times data logger loggable values all physical quantities totalized values and diagnostic values capacity gt 100 000 measured values communication interface RS232 USB serial data kit software all Windows ver sions FluxData download of measurement data graphical presentation conversion to other formats e g for Excel TM FluxKoef creating medium data sets FluxSubstanceLoader upload of medium data sets cable RS232 adapter RS232 USB transport case dimensions 500 x 400 x 190 mm outputs The outputs are galvanically isolated from the transmitter number analog outputs binary outputs max 4 0 2 or 4 active current outputs or passive current outputs or frequency outputs or 2 acti
19. nal drawing weight kg 0 36 0 29 material housing polyester gasket silicone degree of protection IP66 according to IEC EN 60529 ambient temperature min C 20 max C 90 explosion protection zone 2 z marking CE E 113G Ex nA IIC T6 Gc X Ta 10 60 C type of protection non sparking TSFLUXUS_F608xx A2V1 5 1EN_Leu 2014 03 20 FLUXUS F608 A2 Technical Specification Dimensions output adapter power adapter thread 3x M20 x 1 5 thread M20 x 1 5 56 80 56 EL e e E e e e in mm Terminal Assignment output adapter power adapter 10 TSFLUXUS_F608xx A2V1 5 1EN_Leu 2014 03 20 Technical Specification FLUXUS F608 A2 Transducers Transducer Selection transducer order code FSK FSM 1000 5000 inner pipe diameter mm Xi recommended possible TSFLUXUS F608xx A2V1 5 1EN Leu 2014 03 20 11 FLUXUS F608 A2 Technical Specification Transducer Order Code 1 2 3 4 5 6 7 8 9 11 12 13 no of character description 2 Q o G gt 5 ct o c S 32 ES SS Be z B 23 amp 9 29
20. of sound paths is the number of transits of the ultrasonic signal through the medium in the pipe Depending on the number of sound paths the following methods of installation exist reflection arrangement The number of sound paths is even Both of the transducers are mounted on the same side of the pipe Correct positioning of the transducers is easier diagonal arrangement The number of sound paths is odd Both of the transducers are mounted on opposite sides of the pipe In the case of a high signal attenuation by the medium pipe and coatings diagonal arrangement with 1 sound path will be used The preferred method of installation depends on the application While increasing the number of sound paths increases the accuracy of the measurement signal attenuation increases as well The optimum number of sound paths for the parameters of the application will be determined automatically by the transmitter As the transducers can be mounted with the transducer mounting fixture in reflection arrangement or diagonal arrangement the number of sound paths can be adjusted optimally for the application a M E HD A DM a a transducer distance Reflection arrangement number of sound paths 2 Diagonal arrangement number of sound paths 3 a gt 0 am H ES Diagonal arrangement number of sound paths 1 Diagonal arrangement number of sound paths 1 negative transducer distance a
21. sure TSFLUXUS_F608xx A2V1 5 1EN_Leu 2014 03 20 FLUXUS F608 A2 Shear Wave Transducers zone 1 extended temperature range Technical Specification technical type CDM2EW5 CLM2EW5 CDQ2bEW5 CLQ2EW5 order code FSM EA1NL FSM EA1NL LC FSQ EA1NL FSQ EA1NL LC transducer frequency MHz 1 4 inner pipe diameter d min extended mm 50 10 min recommended mm 100 25 max recommended mm 2000 150 max extended mm 3400 400 pipe wall thickness min mm max mm material housing PI with stainless steel cap and PI with stainless steel cap and transducer shoe 304 1 4301 transducer shoe 304 1 4301 contact surface Pl Pl degree of protection IP56 IP56 according to IEC EN 60529 transducer cable type 6111 6111 6111 6111 length m 4 9 3 9 dimensions length mm 84 70 width b mm 40 30 height h mm 59 47 5 dimensional drawing er c oY lc a 78 O la al o lal ambient temperature min C 30 30 max C 200 200 temperature x x compensation explosion protection category gas 2 3G dust 2D gas 2 3G dust 2D EPL Gb Gc Db Gb Gc Db zone 1 2 21 1 2 21 A explosion protection temperature pipe surface T min C 45 E max C 225 225 X markin 112 3G 112 3G I j CE 0637 112D C 0637 amp 112D l Ex
22. ve current outputs and 2 passive current outputs or 2 active current outputs and 2 frequency outputs or 2 passive current outputs and 2 frequency outputs max 4 accessories output adapter necessary option current output range 0 4 20 mA accuracy 0 1 of reading 15 pA active output Rext lt 200 Q passive output Uext 4 9 V depending on Ret Rext lt 200 Q frequency output range 0 5 kHz open collector 24 V 4 mA binary output optorelay 26 V 100 mA binary output as alarm output functions limit change of flow direction or error binary output as pulse output pulse value 0 01 1000 units pulse width 1 1000 ms inputs The inputs are galvanically isolated from the transmitter number max 4 accessories input adapter if number of inputs gt 2 temperature input intrinsic safety type Pt100 Pt1000 connection 4 wire range 150 560 C resolution 0 01 K accuracy 0 01 96 of reading 0 03 K TSFLUXUS F608xx A2V1 5 1EN Leu 2014 03 20 7 FLUXUS F608 A2 Technical Specification Dimensions 59 SEHEM O CHANNEL A CHANNEL BO 213 226 in mm Standard Scope of Supply F608 Standard F608 Energy F608 Double Energy order code FLUXUS F608ST A22 3N NN FLUXUS F6