Home

9320 TEDS Plug & Play Ready Hand Held

image

Contents

1. PEAK Button When pressed this button puts the 9320 into peak mode This will display the highest display reading and hold it on the display until it is reset or a higher value is reached To reset the peak display press the peak and valley buttons simultaneously In peak mode it is possible to capture peaks at a rate of up to 25Hz To turn off the peak mode press the peak button VALLEY Button When pressed this button puts the 9320 into valley mode This will display the lowest display reading and hold it on the display until it is reset or a lower value is reached To reset the valley display press the peak and valley buttons simultaneously In valley mode it is possible to capture valleys at a rate of up to 25Hz To turn off the valley mode press the valley button 12 Interface 9320 TEDS Ready Indicator Manual Issue 2 4 Configuration Menu Parameters vy Ce The parameter is meant to be accessible to the user It allows the removal of fixed display offset values from the display so that the GROSS and NET features can operate from a zero point This may also be considered as a manual tare facility To zero the display simply enter the value that you wish to subtract from the display in the ee parameter i e if the display reads 000 103 and you wish it to read 000 000 then enter 000 103 in the SEt parameter Set Zero may also be achieve by pressing Gross Net and Shunt Cal button simultaneously Different values can be
2. When the value has been set press A you do not know this simply enter all zeros You will be prompted with dISP LO Press to enter the display value required for the low input figure entered The value can be entered by using the KI and O button to select a digit and the a and M buttons to change the digit When the value has been set press You will be prompted with InPut HI press Now enter the sensitivity figure supplied for the sensor by using the KI and O button to select a digit and the a and v buttons to change the digit When the value has been set press PA You then be prompted with dISP HI Press to enter the display value required for the high input figure entered The value can be entered by using the KI and O button to select a digit and the a and M buttons to change the digit When the value has been set press You should now see donE displayed This means the calibration was successful press to the 9320 to normal operation mode with the new calibration data stored If you see FaiLEd then you will need to repeat the calibration checking that you have completed the procedure in the correct order and that the sensor is connected correctly 16 Interface 9320 TEDS Ready Indicator Manual Issue 2 4 Internal Calibration Menu Internal calibration of the 9320 is important when the indicator is used in the TEDS mode or when it is scaled by the method In either of these scenarios the indicator either reads the transducer E
3. E E A AA ad 21 2 interface 9320 TEDS Ready Indicator Manual Issue 2 4 What is TEDS Plug and play sensor hardware and software make configuring a smart TEDS sensor as easy as plugging a mouse into a PC The technology has greatly improved efficiency and productivity by completely eliminating manual sensor configuration Basic concept TEDS is at the heart of the new universally accepted IEEE 1451 4 standard for delivering Plug and Play capabilities to analogue measurement and test instruments In essence information in a Transducer Electronic Data Sheet provides interfacing devices with the critical sensor calibration information in order to perform accurate and precise measurements every time TEDS works in a similar way in which USB computer peripherals immediately work as they are connected TEDS enabled equipment may be swapped and changed without recalibration saving time and money TEDS holds information such as a sensor manufacturer model and serial numbers and more importantly all the calibration settings determined by the manufacturer Analog Signal TRANSDUCER MIXED MO DE INTERFACE ANALOG UE AND DIGITAL How it works Plug and play is a data acquisition technology that can simplify the configuration of automated measuring systems by making a sensor s unique identification data available electronically As implemented according to IEEE 1451 4 data in the form of a transducer electronic data sheet T
4. factory offset value can be changed to a measured value Once the derived value has been entered Press to confirm Interface 9320 TEDS Ready Indicator Manual Issue 2 4 Internal Calibration Procedure 1 Read and note the existing settings for gain and offset using the Internal Calibration Menu If desired the settings for gain and offset may be reset to 1 and 0 respectively for convenience 2 With the 9320 in normal operating mode connect a mV V calibration source to the sensor connector A precision simulator such as the Interface CX series is recommended For the 5 mV V range a source of 3 or 4 mV V is ideal If a simulator is not available another mV source may be used along with a precision DVM to measure both the mV signal and excitation voltage in which case the input in units of mV V is calculated as the ratio However the simulator provides the most reliable and trouble free method 3 Take 9320 0 mV V Reading and the 9320 4 mV V Reading using the preferred 4 mV V as the example Find the net reading by subtracting the former from the later 4 Calculate a new gain value per the formula 5 Set the New Gain value into the 9320 using the Internal Calibration Menu 6 If desired the offset may be adjusted The offset value is not critical to the TEDS or LIVE or FABLE methods of calibration To adjust the offset parameter for a true zero reading at zero input find the New Offset value per the formula 7 Set the
5. set for each RANGE Parameter The value sets the display update rate The options available are 25Hz 10Hz 3Hz 1Hz and 0 5Hz Different updates rates can be set for each RANGE The 25Hz rate only updates at this rate when in the PEAK or VALLEY mode When in normal display mode it has been limited to a 3Hz update as the digit fluctuations are impossible to view with the human eye The 10Hz 3Hz 1Hz and 0 5Hz rates update the display every 100mS 300mS 1000mS and 2000mS respectively The 9320 when it leaves the factory is set at 3Hz meter The parameter allows the user to set a visual alarm The value that is entered is the display value that you want the alarm to activate at When the alarm is activated the word OVErLOad appears on the screen To remove the alarm the display value must be reduced to a value that is lower than that set in the parameter This can be very useful as a safety feature or simply as a quick indication of when a preset level has been reached This value entered can be anywhere over the entire display range so there are no limitations Different values and settings are available for each RANGE SEt OPEr Parameter The 9320 has a special power saving mode which can be enabled or disabled within this parameter pressing when asked whether you wish to select P SAVE will put the 9320 into power save mode for the RANGE selected Pressing will de activate the power save facility When the power save facility is
6. 32 output from the 9320 by pressing when prompted by EnAbLEd On the display pressing will disable the RS232 The output format is ASCII The display value is passed to the RS232 port each time the display updates with a carriage return at the end of each data string The string information is as follows Baud Rate 9600 baud Stop bits 1 Parity None Data bits 8 14 Interface 9320 TEDS Ready Indicator Manual Issue 2 4 Calibration Menu Parameters SENS 5 0 Parameter The 9320 is factory set to enable calibration with sensors generating an input signal of 5mV V or less In the majority of cases it will not be necessary to read higher signal levels If however a higher sensitivity sensor is used with the 9320 it will be necessary to gain access to the internal PCB you must turn the 9320 off to move link LK1 to JP1 see picture below to allow the 9320 to accept sensitivities of up to 50mV V TEDS should only be used with 5mV V as 50mV V is not factory calibrated Once this link has been moved you will need to go back into the CALIBRATION MENU When re entering the menu you will notice that the parameter has changed to to change the sensitivity to 50mV V press PA the 9320 will now check the position of the link and change the sensitivity It will now be necessary to re calibrate any sensors that you may previously have calibrated to this instrument Sensitivity link should be in Be I O RIIIARAN gt this position for use w
7. EDS is burned on an electrically erasable programmable read only memory EEPROM chip located on the sensor so when a properly adapted signal conditioner interrogates the sensor it can interpret the self identification data This technology provides a great benefit by eliminating the need for paper calibration sheets In addition it can simplify labeling and cabling problems as well as inventory control issues by letting you burn location data onto the chip when installing a sensor And because all sensors produced according to the standard will carry the same basic identically formatted self identification information you will be able to mix and match sensors and applicable signal conditioners across manufacturers Interface 9320 TEDS Ready Indicator Manual Issue 2 4 3 Advantages Plug and play sensors are revolutionizing measurement and automation With Transducer Electronic Data Sheets TEDS your data acquisition system can detect and automatically configure sensors This technology provides e Reduced configuration time by eliminating manual data entry e Better sensor tracking by storing data sheets electronically e Improved accuracy by providing detailed calibration information e Simplified asset management by eliminating paper data sheets e Reliable sensor location by identifying individual sensors electronically Introduction The 9320 Portable Strain Display Load Cell Force transducer readout is a microprocessor based portable in
8. EPROM to get the transducer sensitivity in mV V or obtains it from manual key entry The 9320 must therefore have an internal mV V reference This section describes a procedure for periodic calibration of the internal reference If the 9320 is not used in the TEDS mode and is scaled only by the LIVE method not the FABLE method internal calibration is unnecessary To enter the Internal Calibration Menu press and hold ES and for 10 seconds Parameter Set up Information Press El To skip to next menu item Press To change the 5mV V gain ra 1 00000 a 1 00000 ra 1 00000 r 0 00000 Here the factory gain calibration can be changed to a measured value Once the derived value has been entered Press to confirm Press E to skip to next menu item Press to change the 5mV V offset Once the derived value has been entered Press to confirm Press El To skip to next menu item Press To change the 50mV V gain THIS CAN ONLY BE SET WHEN USING 50mV V RANGE Here the factory gain calibration can be changed to a measured value Once the derived value has been entered Press to confirm Press El to skip to next menu item Press to change the 5mV V offset THIS CAN ONLY BE SET WHEN USING 50mV V RANGE 5 0 OFFS Here the factory offset value can be changed to a measured value Here the
9. LEd a ff O Note Only when TEDS is disabled Configuration Menu To enter the CONFIGURATION MENU press and hold Parameter Se buttons for 3 seconds Set up Information Press EJ To skip to next menu item Press To set a new system zero This allows the user to introduce a fixed offset to the display value The GROSS and NET values are then displayed with this offset taken into account Values between 9999999 and 9999999 can be entered using the KI and O arrows to select a digit and the a and v arrows to increment or decrement the digits Press to accept the value and move onto the next parameter Set Zero may also be set by pressing and Tree at the same time Press To skip to next menu item Press To change the update rate This allows the user to set the display update rate the options available is the update rate of the display in Hz Please note that the 25Hz update is only available in PEAK or VALLEY mode When you choose to change the update rate you will be prompted whether you want to select 25Hz if you do not press EJ you will then be prompted to select any of the other values which in order are 10Hz 3Hz 1Hz 0 5Hz to set the update rate for the value you want press g Press To skip to next menu item Press To set overload alarm 8 interface 9320 TEDS Ready Indicator Manual Issue 2 4 This allows the setting of a visual overload The value entered is the di
10. New Offset value into the 9320 using the Internal Calibration Menu Internal Calibration Gain Example 1 Existing Gain 1 007920 Simulator CX 0404 4 mV V setting 3 99992 9320 0 mV V Reading 0 0003 9320 4 mV V Reading 3 9996 New Gain 1 007920 3 99992 3 9996 0 0003 1 008076 Internal Calibration Gain Example 2 Existing Gain 1 008220 mV source DVM reading 40 008 mV Excitation DVM reading 10 055 V Calculated input 40 008 10 055 3 9789 mV V 9320 0 mV V Reading 0 0003 9320 4 mV V Reading 3 9861 New Gain 1 0082200 3 9789 3 9861 0 0003 1 006475 18 Interface 9320 TEDS Ready Indicator Manual Issue 2 4 Specifications Performance Input Type Strain Gage Full Bridge Sensors Input Range Up 5mV V 50mV V can be supplied with factory set option Nonlinearity 0 005 FS Thermal Drift lt 25 ppm C Excitation Voltage 5Vdc 4 59mA maximum current nimium Prida 850 4 each 3500 sensors in parallel Resistance Internal Battery 2 each AA size alkaline access via sealed rear compartment Battery Life 45 hours Typical 450 hours in low power mode with 3500 sensor Update Rate 0 5 Hz to 25 Hz can be set in configuration menu Indication Display Type 7 digit LCD display 8 8mm high digits Display Resolution 1 part in 250 000 at 1Hz update rate 1 part in 65 000 at 10Hz update rate Control Variables Annunciators Front Panel User Keys Sett
11. a a 7 pin 723 series Binder connector The wiring for this is as detailed below 2 4 PIN 1 Tx PIN 2 Rx 3 1 PIN 3 Gnd 6 7 Internal Connections It may be necessary from time to time to know what the internal connections are For example if you disturb some of the connections while trying to insert the range legends or if you need to change the internal shunt calibration resistor These are shown below for reference only Sensor Connections Shunt Calibration Resistor Sensor Connections Interface 9320 TEDS Ready Indicator Manual Issue 2 4 5 There are six push buttons on the front panel of the 9320 which are available for use in normal operation Each of these is described below Front Panel Button Function of Button in Normal Operation Mode To switch the 9320 ON or OFF press and hold the button The RANGE button allows the user to toggle between two independent scales The range that has been selected is highlighted by an annunciator The HOLD button allows you to hold freeze the current display value when the button is pressed Pressing the HOLD button again releases the display The HOLD annunciator is illuminated when in the HOLD mode and the display will flash to alarm further that the user is not viewing instantaneous display values The GROSS NET button when pressed allows the user to toggle between displaying the Gross or Net display values This can be useful in many applications where i
12. a period of 1 one year from the date of dispatch If the Interface Inc product you purchase appears to have a defect in material or workmanship or fails during normal use within the period please contact your Distributor who will assist you in resolving the problem If it is necessary to return the product request an RMA and include a note stating name company address phone number and a detailed description of the problem Also please indicate if it is a warranty repair The sender is responsible for shipping charges freight insurance and proper packaging to prevent breakage in transit The warranty does not apply to defects resulting from action of the buyer such as mishandling improper interfacing operation outside of design limits improper repair or unauthorised modification No other warranties are expressed or implied Interface Inc specifically disclaims any implied warranties of merchantability or fitness for a specific purpose The remedies outlined above are the buyer s only remedies Interface Inc will not be liable for direct indirect special incidental or consequential damages whether based on the contract tort or other legal theory In the interests of continued product development Interface Inc reserves the right to alter product specifications without prior notice Doc No 517 173 Issue 2 2 Dated 07 09 06 20 Interface 9320 TEDS Ready Indicator Manual Issue 2 4
13. able Parameters Low Battery warning peak valley hold net shunt cal range Tactile Keys with recessed rims for ON OFF Switches 9320 power on off RANGE Selects between two ranges HOLD Hold the current display value press again to release GROSS NET Zero s display 100 range SHUNT CAL Generates simulated input for indicator testing PEAK Enables peak hold VALLEY Enables valley hold Tare Zero value display resolution decimal point position display update rate low power mode auto power off Mechanical Electrical Connection 5 pin Binder socket mating plug supplied Physical Size See drawing below mm Weight 260 grams Legends Insert legends for engineering unit identification supplied Environmental Operating Temperature 10 C to 50 C Environmental Rating IP65 when mating plug fitted Enclosure Type ABS dark grey Leather Carry Case Optional Safety Low Voltage 73 23 EEC amended by 93 68 EEC To IEC 1010 1 1990 EN Directive 61010 1 1993 89 336 EEC EMC Directive EN 50 081 1 1992 Light Industrial Emissions EN 50 081 2 1992 Heavy Industrial pr EN 50 093 1991 EMC Emissions EN 50 082 1 1992 Light Industrial EN 50 082 2 1992 Heavy Industrial Mechanical Dimensions 90 34 Interface 9320 TEDS Ready Indicator Manual Issue 2 4 19 Warranty The 9320 is warranted against defective material and workmanship for
14. activated battery life is conserved by pulsing on the excitation voltage to the sensor As a result the accuracy is reduced as is the update rate When in this mode the quickest update rate is 3Hz and the accuracy of the display is reduced to 1 digit in 10 000 It is important to note these limitations when deciding whether to use the power save facility However it is also possible to set one RANGE with power save activated and the other without The benefit is that the battery life based on a 3500 sensor bridge being connected increases from 45 hours to 450 hours It is also important to remember that when the 9320 is re calibrated with a sensor the power save facility will be automatically turned off The power save facility will therefore need to be re activated after calibration has been completed Interface 9320 TEDS Ready Indicator Manual Issue 2 4 13 A The parameter is another power saving feature It allows for the setting of a time period in minutes between 05 and 99 00 de activates AUtO OFF i e if this was set to 25 then if the 9320 detects no keyboard activity for a continuous 25 minute period then the 9320 will power down to conserve power If keyboard activity is detected at any time during the 25 minute period then the time period is restarted This can be a useful feature in a site environment should the 9320 be left unintentionally powered on 15232 Parameter This parameter allows the user to enable the RS2
15. cation it is to be used in The display can display the instantaneous peak or valley values It is also possible to hold the display value this only operates when not in peak or valley mode The display update rate decimal point position and resolution can be set to suit The 9320 has two independent ranges All values set in one range are totally independent from the other Switching the 9320 On Off The 9320 is switched ON or OFF by pressing and holding down the FG button for 3 seconds It is also possible to set an Auto off value in the configuration menu so that the 9320 automatically switches itself off after a preset time if there is no keyboard activity RANGE Button The range feature allows for the setting of two totally independent setup ranges to be selected if required To switch between ranges simply press the range button If TEDS has been enabled then only 1 range is permissible When you enter either the calibration menu or configuration menu the parameters you will be setting are those for the range you have selected An annunciator is lit to identify which range has been selected The 9320 is supplied with engineering unit legends these can be slid into a window located on the inside of the front panel These labels then help to further identify the units being displayed for each range Please refer to the photo below Legend labels are ma inserted both sides e rororo Interface 9320 TEDS Ready Indi
16. cator Manual Issue 2 4 11 HOLD Button The hold button allows the user to freeze the display when it is pressed When pressed again the display returns to it s normal operating mode When in hold mode the display will flash and the hold annunciator will be lit to ensure that this feature is not accidentally turned on without the user noticing The hold feature cannot be used when the 9320 is in either peak or valley hold mode GROSS NET Button The gross net button when pressed toggles between the gross and net display values This enables the user to zero the display by putting the 9320 into net mode and displaying the change in display value from that point This is useful for certain weighing applications where a tare weight exists which can be removed by putting the 9320 into net mode SHUNT CAL Button The shunt calibration button when pressed puts an internal 100kQ resistor across the ve excitation and ve signal of the sensor generating a simulated output from the sensor therefore giving a simulated display value This can be pressed immediately after the sensor has been calibrated with the 9320 and noted down for later reference The value noted can be used to get an idea of the calibration accuracy at a later date or for checking the integrity of the sensor and sensor cabling The shunt calibration resistor can be changed to suit specific requirements It is suggested that a 15ppm 0 1 tolerance resistor is used
17. ggested that the RS232 output is disabled when it is not required To enable press To disable presk Interface 9320 TEDS Ready Indicator Manual Issue 2 4 Calibration Menu To enter the Calibration Menu press and hold and buttons for 5 seconds Parameter Set up Information Press Ea To skip to next menu item Press To change sensor input sensitivity This allows the calibration engineer to change the sensitivity range of the 9320 when connecting to sensors with a sensitivity of greater than 5mV V The 9320 is factory set for SEnS 5 0 5mV V To ensure the unit is set to 5mV V press To select 50mV V you need to power down the unit and access the internal circuit board Move link LK1 and place it onto JP1 Power on the 9320 and return to this point of the calibration menu You will notice that the menu parameter has changed to SENS 50 0 press to change the sensitivity to 50mV V and move on to the next parameter Press Ea to skip to next menu item Press to the set the display resolution This parameter sets the decimal point position for the display and the resolution i e a value of 000 005 would display the reading to 3 decimal places and the readings will change in steps of 0 005 SEt rES The decimal point position is moved one place to the right each time you press the PS and v together Any value can be entered for the resolution using the KI and OD arrows to select a digit and the a a
18. incerface ADVANCED FORCE MEASUREMENT 9320 Load Cell Hand Held Digital Indicator TEDS Plug amp Play Ready IEEE 1451 4 Compliant User Manual 13 69 Rev 2 4 Interface Inc e 7401 E Butherus Dr e Scottsdale AZ 85260 e 800 947 5598 www interfaceforce com 9320TEDSman2R2_4 doc Contents VINOS LEDS ade 3 Basic CONCE ina A A AA AAA AAA 3 POW IU WO a 3 IVANA A A A A CETNA AA c 4 a A O cia NANINIRA 4 a o RR 4 Electrical Connection Information ccccceeeeneeeeeeeeeeccccceeeeeeetesseseecesecceeeeeseeeeee 5 Sensor CONNECTIONS sr See ii EEE E AA il 5 RS232 Port CONNECTIONS iii iaa lr e eee ad 5 Internal CONNECHONS essas a roda vers 5 MENU STFUCLUTO iii is cin chy Dd ia as 7 Confis ration MEU A A A Ad 8 Calibration MU A A oa dee aa 10 Operation Features 2000 A ad 11 Normal Display Operation ii A A E A a 11 Switching the 9320 On Off id a Dl das 11 RANGE BUON eae A AN AAA AA AAA AAA A 11 HOLD BUON a A E aos 13 GROSS NEF BUON sii ad iaa 13 SHUN TF CAL BUCOM rt a a a a S 13 PEAK BUON da eds 13 VALLEY BUE ao 13 Configuration Menu Parameters ni A A A AR cde ARA AAA 14 Calibration Menu Parameters cccccccceccccececccccececceeeeeeeseteeeesesceeeeeeeeetetesseeeeess 16 Calibration Procedures a as 17 Internal Calibrations ac A A buvnae d E Ea 18 SPECIFICALIONS a ANI UE E EA A ante See 20 Mechanical DiMeNnsiOns cccecccccccccccccccceeceeeeeeeeccseeeeeeeceeeeeeeseeeeeeeeeseeeeenseseeees 20 NARA
19. ith A sensors with sensitivities lt 5mV V 510 0189 issue 2 Sensitivity link should be in this position for use with sensors with sensitivities gt 5mV V Peeneee SEt FES Parameter This parameter enables the setting of two features on the 9320 It allows you to set the decimal point position of the display by pressing the a and v together to move the point position each press moves the decimal point position one place to the right It also allows for the setting of the display resolution or the number of display counts the display changes with an input change To change the resolution use the KI and OD arrows to select a digit you want to change and the a and M arrows to increment or decrement the digits Press to accept the value CALibrAt Parameter disabled when TEDS is enabled This parameter is used to calibrate and scale the 9320 with a sensor There are two basic methods of calibration available These are LIVE and ABLE There is also a third parameter which can be used for maintenance and recording purposes This parameter is CAL VAL The CAL VAL value can be viewed after a calibration has been completed and will show the offset and gain figures from any stored calibration If these figures are noted they can be used to re enter at a later date if calibration data is lost for any reason or if the calibration data from a sensor needs to be transferred to another 9320 tedS Parameter Thi
20. le of approx 3 seconds then press You then be prompted with dISP LO Press to enter the display value required with the low stimulus applied to the sensor The value can be entered by using the KI and O button to select a digit and the a and v buttons to change the digit When the value has been set press You will then be prompted with APPLY Hi unless you chose to USE SC in which case jump to the next stage At this point ensure that the high calibration stimulus is applied to the sensor and allow to settle of approx 3 seconds then press You then be prompted with dISP HI Press to enter the display value required with the high stimulus applied to the sensor The value can be entered by using the KI and OD button to select a digit and the a and v buttons to change the digit When the value has been set press You should now see donE displayed This means the calibration was successful press to the 9320 to normal operation mode with the new calibration data stored If you see FaiLEd then you will need to repeat the calibration checking that you have completed the procedure in the correct order and that the sensor is connected correctly tAbLE Calibration Procedure When is displayed press will now be displayed press will now be displayed press You will be prompted with InPut LO press Now enter the zero offset sensitivity of the sensor by using the KI and O button to select a digit and the a and v buttons to change the digit
21. n 6 Interface 9320 TEDS Ready Indicator Manual Issue 2 4 Menu Structure The 9320 has two menus details of which are outlined below A CONFIGURATION MENU which enables the user to tailor the operation to meet a specific application requirement The values selected in the CONFIGURATION MENU are completely independent for each range SEtZEro A 0000000 SEtrAte EA 2527 ES 10 132 1 E 0 5 El SEt OUEr PA 0000000 SEtOPEr PA Psave PA E AutO OFF Zo PA rs232 PA EnabLea BA A Calibration Menu which is used to calibrate each of the two ranges with independent scales as well as setting the display resolution for each range Interface 9320 TEDS Ready Indicator Manual Issue 2 4 7 SENS 5 0 KA use sc Ed APPLY Lo dISP LO 0000000 APPLY HI dISP HI 0000000 donE APPLY LO BA asp Lo 0000000 WEA dise Hi 0000000 donE InPut LO 0000000 diSP LO 0000000 InPut HI 0000000 dISP HI 0000000 donE El SEtrES 000 000 a CALibrAt LIVE E E tAbLE E CAL VAL SEt9Ain Z 0000000 EA SEt OFFS 0000000 donE tedS EnAb
22. nd M arrows to increment or decrement the digits Press to accept the value and move onto the next parameter To save the settings and move on to the next parameter press THIS MENU IS DISABLED WHEN TEDS IS ENABLED Press E to skip to next menu item Press to the enter the calibration routine If you have chosen to enter the calibration routine you will be prompted whether you want CALibrAt to select Live if you do not press ig otherwise press PA You will then be prompted to select either of the other calibration methods which in order are FABLE and CAL VAL to select any of the calibration methods press A Otherwise press El For more detailed calibration information please refer to the calibration section of the manual 10 Interface 9320 TEDS Ready Indicator Manual Issue 2 4 ENABLING TEDS DISABLES CALIBRATE MENU Press E to skip this parameter and exit menu Press to enable or disable TEDS If you have chosen to enter the TEDS calibration EnAbLEd appears tedS If you have chosen to enter the TEDS you will be prompted whether you want to select if you do not press iq otherwise press PA If you have selected enabled two flashing indicators will appear For more detailed TEDS calibration information please refer to the TEDS section of the manual Operation Features Normal Display Operation The 9320 has a full 7 digit display which can be scaled using the calibration menu to suit the appli
23. s parameter automatically calibrates the 9320 with the data from the TEDS chip The two annunciators appear when active connection with a TEDS peripheral has been made When there is a loss of connection these annunciators flash When changing a sensor the 9320 should be power cycled as this is when the TEDS data is read Calibration Procedures are not available when TEDS is enabled Interface 9320 TEDS Ready Indicator Manual Issue 2 4 15 Calibration Procedures The best method of calibration if it is possible to do so is the LiVE calibration as this reads in the sensor signal at two calibration points and scales the 9320 automatically If this is not possible then the sensitivity figure in mV V from the sensor calibration certificate can be used to scale the 9320 by using the tAbLE calibration This may be the only option available if you are unable to apply a known stimulus to the sensor which quite often is the case LIVE Calibration Procedure When is displayed press will now be displayed press You will be prompted USE SC this can be selected if you wish to use the shunt calibration figure from a sensor calibration certificate care should be taken that the shunt calibration resistor used originally with the sensor is the same as is fitted in the 9320 If you wish to use this press otherwise press You will then be prompted APPLY LO At this point ensure that the low calibration stimulus is applied to the sensor and allow to sett
24. splay value at which the 9320 displays Values between 9999999 and 9999999 can be entered using the KI and O arrows to select a digit and the a and M arrows to increment or decrement the digits Press to accept the value and move onto the next parameter Parameter Set up Information Press E To skip to next menu item Press To select operation mode This allows the enabling or disabling of the power save mode which updates at 1 update per second and pulses the sensor excitation This results in a lower accuracy 1 part in 20 000 To enable press To disable press eal Press To skip to next menu item Press To set auto power off This enables the setting of an auto power off value The value entered is in minutes If no front panel buttons are pressed for the time set here then the indicator will automatically power off to conserve battery life Values between 05 and 99 can be entered 00 leaves the 9320 permanently powered using the KI and OD arrows to select a digit and the a and v arrows to increment or decrement the digits Press to accept the value and move onto the next parameter Press EJ To skip this parameter and exit menu Press To enable the RS232 output This feature enables you to enable or disable the RS232 output Further details of the RS232 format are provided further into this manual The RS232 output is an option that has to be ordered with the 9320 To conserve battery life it is su
25. strument designed to interface with any full bridge sensor with an output sensitivity of up to 50mV V Bridge resistances from 85Q upwards can be used with the 9320 Configuration and calibration of the 9320 is achieved using the front panel push buttons to navigate through a very simple menu structure User functions available on the 9320 include Range Selection Display Hold Freeze Gross Net indication selection Peak Hold selection Valley Hold selection Shunt Cal check The 9320 is powered by two internal non rechargeable AA alkaline batteries There is an option to have rechargeable alkaline batteries which can be charged without the need for removing them from the 9320 User Operation Full 7 digit LCD display Operation Annunciators Unit Labels Push Buttons used for normal operation and for configuration 4 Interface 9320 TEDS Ready Indicator Manual Issue 2 4 Electrical Connection Information Sensor Connections The standard sensor connection is a 5 pin 723 series Binder connector The sensor connector is also used as a charging socket when the re chargeable alkaline battery option has been ordered the wiring for this is detailed below PIN 1 ve Excitation PIN 2 ve Excitation ve Charge Input amp TEDS Common PIN 3 ve Signal PIN 4 ve Signal PIN 5 ve Charge Input or TEDS RS232 Port Connections If the 9320 has been ordered with the optional RS232 output then this will be available vi
26. t is necessary to display the change in display value from a certain part of the measurement range When in NET mode the NET annunciator is lit When in GROSS mode the NET annunciator is not lit The SHUNT CAL button allows the user to press this at any point in time CA The standard unit shunts a 100kG resistor across the negative excitation and negative signal connections If this is performed at the end of the calibration procedure then a figure can be noted so the user can check calibration accuracy or connection integrity The button has to be held down to operate When held down the SHUNT CAL annunciator is lit and the display will flash to alarm further that the user is not viewing instantaneous display values Ey When the PEAK button is pressed the display will show the last Peak reading To reset the Peak readings press the PEAK and VALLEY buttons simultaneously When in PEAK mode the PEAK annunciator will be lit and the display will flash to alarm further that the user is not viewing instantaneous display values To turn off Peak mode press the PEAK button When the VALLEY button is pressed the display will show the last Valley reading To reset the Valley readings press the VALLEY and PEAK buttons simultaneously When in VALLEY mode the VALLEY annunciator will be lit and the display will flash to alarm further that the user is not viewing instantaneous display values To turn off the Valley mode press the VALLEY butto

Download Pdf Manuals

image

Related Search

Related Contents

一括ダウンロード (PDF 5.1MB)  User Manual - ECOsens Fuel Report - Fueldata Information Systems  A Feather-weight Virtual Machine for Windows Applications      Module E3: Documents de formation - Bundesamt für Justiz  Skuttle Indoor Air Quality Products 2100 Humidifier User Manual  Virtual Keyboard (VKB) for Windows  < ニッテル携帯レンタル >サービス案内 返却 支払 宅配 申込確認 お  

Copyright © All rights reserved.
Failed to retrieve file