1. Battery Installation This scale is powered by six 6 standard disposable C cell alkaline batteries Scales will arrive with batteries installed Use the following procedure when installing a new set of batteries 1 Gently remove the cradle by pulling up on the cradle to unlock it from the platform mounting pins 2 Remove the scale top cover by removing the six 6 screws Fig 11 3 Locate the battery tray in the rear center of the scale platform It has a white cover that is held down with two screws Remove the screws and cover 4 The batteries may be held in place with tape This was to add further protection during initial shipping and is no longer required Remove old tape and batteries 5 Insert new batteries Note the orientation that is shown on the bottom of the battery tray Press ON ZERO to make sure the display powers up properly 6 Replace the battery tray cover and screws scale cover and cradle Figure 11 Removing scale cover Calibration Your scale has been properly calibrated at the factory and should not require any further adjustments The following procedure should be performed only to correct any error in calibration Calibration is protected by a security code to avoid inadvertent access For maximum accuracy calibration must be done in grams If you only have certified weights in pounds convert them to grams i e 10 pounds 4536 grams 1 2 Depress the ON ZERO button on the con
2. Note that some buttons perform multiple functions ON ZERO TARE LB GM REWEIGH RECALL MENU Depress to turn scale on or reset to zero Depress to zero out the weight of an object such as a blanket on the scale Place the baby on the cradle and the console will display the net weight by automatically deducting the weight of the blanket Depress to switch between pounds and ounces mode and grams mode or between pounds and ounces mode and kilogram mode see User Setting Options section on how to switch from grams to kg or how to disable button Secondary function is the up arrow and is used for menu selection Depress while baby is on cradle to reweigh The last weight measurement can be recalled after scale has turned itself off and display is blank Secondary function is the down arrow and is used for menu selection Depress for advanced functions such as programming the length of the automatic time off Secondary function is the ENTER button and is used for menu selection 8 D B 9 Q Health o meter fer WT ProPlus PRIOR WT HEIGHT kg lb oz gram BMI AVERAGE WT A Vv ENTER REWEIGH TARE LB GM RECALL Figure 2 Readout Console Keypad not to scale Special Features The 2210KL scale comes standard with these special features 1 Auto Zero The scale will automatically display 0 when turned on If the display reads between 20 gram and 20 gram at any ti
3. to achieve a true measurement in line with the listed accuracy specification However the average weight deviation from the true weight of the baby will be minimal based on the severity of the motion during the weight reading Auto Off As a battery saving feature the scale will automatically turn off when not being used The length of the Auto Off timeout is programmable by the user The factory setting is one minute Using the Tare Function When using this scale the weight of any object such as a blanket accompanying the baby being weighed can be subtracted from the total weight to determine the net weight of the baby alone The Tare function automatically performs this subtraction avoiding the need for manual calculations TARE Figure 3 Tare button n08 gram Figure 4 Net weight display 1 Turn the scale on and wait for the display to read 0 2 Place the object i e blanket on the cradle The weight of the object will be displayed 3 Depress the TARE button NET WT and 0 0 will be displayed If the object is removed from the platform its weight will be displayed as a negative value 4 Place the baby on the cradle with the object still on it The readout will now display the baby s net weight 5 Depress ON ZERO to return the scale back to normal weighing operation Using the Weightlock Function When using this 2210KL scale the weight of any object over 512 gr
4. Health o meter Professional 2210KL Neonatal Pediatic Scale 1 gram accuracy USER INSTRUCTIONS Rev 20120516 THIS PAGE INTENTIONALLY LEFT BLANK Table of Contents Introduction ariii ia 4 Cautions amp Warnings 4 Assembly Instructions Before Assembly mere 5 Assembling the 2210KL 5 Operation Instructions Quick Start Instructions 0 5 Console Keypad cece 6 Special Features 0 0 ee 7 Using the Tare Function Using the Weightlock Function Using the Reweigh Function Using the Recall Function 9 Programming Auto Off Time 9 Motion Detection eee 10 Battery Life Indicators 10 User Setting Options 08 11 Battery Installation ceea 11 Calibration 0 0 0 eects 12 Care and Maintenance Cleaning Moving the Scale Troubleshooting Specifications Model 2210KL Specifications Warranty Model 2210KL Warranty How to Obtain Warranty Service Serial Number 13 13 13 14 15 15 Introduction Health o meter Professional scales by Pelstar LLC use the most sophisticated microprocessor technology available Each precision instrument is designed to provide accurate reliable and repeatable weight measurements as well as many other features that make the weighing process simple fast and convenient This opera
5. IREMENTS Batteries Battery Life approx With 1 minute Auto Off OPERATING CONDITIONS STORAGE amp TRANSPORTATION SCALE DESCRIPTION Type Capacity Accuracy O 11300 g 0 25 Ib gt 11300 g gt 25 lb Resolution O 11300 g 0 25 Ib gt 11300 g gt 25 lb Degree of protection against electrical shock Type BF This scale is a Latex free product 24 inches 61 cm 15 inches 38 cm 3 inches 8 cm 26 inches 66 cm 16 5 inches 42 cm 7 inches 18 cm 26 Ibs 11 8 kg 6 C cell Alkaline 9 V 3 years 80 000 weight readings 68 to 85 degrees F 20 to 30 degrees C 0 to 85 relative humidity 700 to 1030 hPA atmospheric pressure Avoid exposure to high pressure water or steam 14 to 122 degrees F 10 to 50 degrees C 0 to 85 relative humidity 500 to 1060 hPA atmospheric pressure Remove batteries if storing longer than 3 months Strain Gauge 20000 grams 45 pounds 19 0 1 oz 5g 0 1 oz 1g 0 1 oz 5g 0 1 oz 2210KL 3 Year Limited Warranty This Health o meter Professional scale is warranted from date of purchase against defects of materials or in workmanship for a period of three 3 years If product fails to function properly return the product freight prepaid and properly packed to Pelstar LLC see How To Get Warranty Service below for instructions If the manufacturer determines that a defect of material or in workmanship ex
6. L The stored value will now display 3 With the weight measurement in the display the scale will be returned to the unlocked mode ready for weighing by either of the following a RECALL is depressed again b ON ZERO is depressed Programming Auto Off Time The 2210KL has been preset at the factory to shut down after one minute of no use This can be programmed from 12 seconds to 25 minutes to suit your needs 1 Depress ON ZERO and wait for the display to show 0 0 Do not place anything on cradle 2 Hold MENU button down and depress the RECALL button Set 0 will be displayed along with the up and down arrows the 0 will be flashing 3 Depress the A button repeatedly until Set 3 appears depress ENTER 4 The Auto Off time currently programmed in the scale will display as minutes and tenths of minutes i e 03 2 the far left digit will be flashing Use the Aand W buttons to set the desired value press ENTER to advance to the next digit For example setting the time to 00 5 will turn the scale off after 30 seconds of no use 02 0 will turn the scale off after 2 minutes 25 0 will turn the scale off after 25 minutes 5 Press ENTER when complete to save and return to normal weighing operation ENTER MENU Figure 7 Menu button Motion Detection Motion detection uses a proprietary technology that automatically provides accurate weight measurement when there is e
7. ams will be locked on the display even after the baby has been removed from the cradle allowing the care provider to tend to the needs of the baby without having to remember the weight measurement This value is stored and can be recalled even after the scale has turned itself off When the baby s weight is locked PRIOR WT is displayed Fig 5 The weight will remain displayed even after the baby is removed from the cradle No further weighings can take place until Weightlock is deactivated Weightlock is deactivated by depressing ON ZERO or REWEIGH RECALL Fig 6 08 gram Figure 5 Weightlock display hA REWEIGH RECALL Figure 6 Reweigh Recall button Using the Reweigh Function The Reweigh function is used to reweigh the baby while he or she is still on the cradle To reweigh the baby should remain on the cradle and a weight measurement should be on the display Simply press REWEIGH RECALL Using the Recall Function The Recall function is used to recall the last weight measurement after the scale has shut down and the display is blank The scale automatically stores the last weight measurement This value is stored until a new measurement is taken or it is intentionally cleared To recall a weight value 1 The baby s weight has been measured the baby has been removed from the cradle and the display is blank scale has shut down 2 Depress ON ZERO and then REWEIGH RECAL
8. ct features or specifications without notice Pelstar LLC 2012 Rev 5
9. ists the customer s sole remedy will be repair or replacement of the scale at no charge Replacement will be made with a new or remanufactured product or component If the product is no longer available replacement may be made with a similar product of equal or greater value All parts including repaired and replaced parts are covered only for the original warranty period Who is Covered The original purchaser of the product must have proof of purchase to receive warranty service Please save your invoice or receipt Authorized dealers selling Pelstar products do not have the right to alter or modify or any way change the terms and conditions of this warranty Your warranty does not cover normal wear of parts or damage resulting from any of the following negligent use or misuse of the product use on improper voltage or current use contrary to the operating instructions abuse including tampering damage in transit or unauthorized repair or alternations Further the warranty does not cover Acts of God such as fire flood hurricanes and tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction How to Get Warranty Service Serial Number To Obtain Warranty Service make sure you Your model and serial numbers are located on the keep your sales receipt or document showing back of the scale Take time to reco
10. me regardless of what is on the cradle the scale will automatically reset to 0 after 8 seconds English Metric Readout With normal operation you may switch the display units between grams or kilograms and pounds at any time before or after a weighing When turned on the scale will display the last used unit of measure Tare Option An object weighing more than 20 grams can be subtracted from the total weight to determine the net weight Weightlock Any weight over 512 grams will lock in so that no further weighings can take place until Weightlock is deactivated The measurement will display even after the baby is removed from the cradle so that the caregiver can attend to the baby rather than trying to remember the weight measurement The value is retained even after the scale turns itself off Motion Detection In most cases the scale will display a true accurate weight based on a series of consistent and stable measurements However if a baby or object is moving on the platform the scale may not achieve consistent readings that are required to give a stable weight This condition is indicated by Weightlock not activating on the display If the movement on the platform continues for seven seconds the scale will automatically activate the motion detection feature and AVERAGE WT will be displayed on the console Because of the movement on the platform average weight indicates that the scale was not able
11. olutions or sanitizing wipes to clean the display surface e Do not use abrasive cleaners e Check overall appearance of the scale for any obvious damage or wear and tear Moving the Scale The scale should be lifted by the two black handles only For your added convenience an optional wheeled cart is available NOTE The scale cradle should never be used for transporting a baby or objects The cradle surface must be clear of any objects during transport Troubleshooting Scale will not turn on e Check that batteries are seated properly in battery tray as they may have shifted in shipping See BATTERY INSTALLATION section on how to access the batteries and check their positioning e Check that fresh batteries are properly inserted into the battery tray Incorrect weight readings e Cradle is touching wall or other object and must be moving freely To test remove all objects from cradle Turn scale on Display should read 0 Gently push down on the cradle with your finger and release The display should show some weight value and then quickly return to 0 e Cradle is not level Weight reading consistently too high or too low e Check scale calibration and calibrate if necessary e Possible broken wire in the connector Call the service department for further 2210KL Scale Specifications DIMENSIONS Cradle Width Depth Height Scale Assembly Width Depth Height Weight POWER REQU
12. rd the serial proof of purchase Call 1 800 638 3722 to number for future reference when communicating receive a return authorization RA number with the service department which must be included on the return label Attach your proof of purchase to your defective product along with your name address daytime telephone number and description of the Model Number 2210KL problem Carefully package the product and send with shipping and insurance prepaid to Pelstar LLC Attention R A Repair Department 11800 South Austin Avenue Unit B Alsip IL 60803 Date Purchased If your scale is not covered by warranty or has been damaged an estimate of repair costs or replacement costs will be provided to you for approval prior to servicing or replacing Serial Number 15 Health o meter This scale model is CSA listed te Koy a XS S SN fo inh e Medical Electrical Equipment for U S A and Canada Pelstar LLC 11800 South Austin Avenue Unit B CSA Symbol Legend Alsip IL 60803 Degree of protection against F electrical shock Type BF Phone 708 377 0600 Toll free 800 815 6615 i f Fax 708 377 0601 Caution or attention required h Risk of potential danger Ne MaCales Com ini Health o meter is a registered trademark of qe i Lo t oa of banery Sunbeam Products Inc used under license i We reserve the right to improve enhance or __ Direct electrical current modify Health o meter Professional produ
13. s RECALL button Release both buttons Set 0 will be displayed and the 0 will be flashing Depress the A button repeatedly until Set 6 appears Depress the ENTER button Aweight value will display representing the current calculation of the known weight The hundreds of pounds digit will be flashing Use the and W keys to input the proper value Press ENTER to advance to the next digit After entering the tenths of pounds digit press ENTER and CAL will display while the scale calibrates itself The readout will then quickly show the old weight then re weigh to show the new calibration weight 10 Remove weight and depress the ON ZERO button Wait for 0 0 to display and verify calibration by placing the same known weight back on the platform Calibration Error Code Err 4 Scale is below minimum calibration weight Use at least 1000 grams to calibrate the scale CARE AND MAINTENANCE 2210KL Cleaning Procedures Proper care and cleaning of the scale is essential to ensure a long life of accurate and effective operation In addition to routine cleaning include a regular check of all structural apparatus All external surfaces can be wiped clean with a damp cloth or tissue A mild soap and water solution may be used as well Wipe unit dry with a clean soft towel e DO NOT immerse scale platform into cleaning or other liquid solution e DO NOT use isopropyl alcohol other such s
14. s from the shipping containers Your scale has been shipped in one 1 carton and contains the following components e Scale Platform e Acrylic Cradle Six C cell alkaline batteries are pre installed in the scale platform Length rule Figure 1 Install Cradle Assembling the 2210KL 1 Place the scale platform on a table top with the display buttons facing you 2 The acrylic cradle has a built in length rule Align the cradle so that the rule is positioned over the back end of the scale platform Fig 1 3 Carefully lower the cradle onto the four mounting pins on the scale platform When the cradle is properly aligned with the pins gently push down on the cradle over each pin until it snaps into its locked position When properly positioned there will be about a 3 8 space between the platform top and the bottom of the cradle Always lift the scale by the black handles only never by the cradle OPERATION INSTRUCTIONS Quick Start Instructions 1 If desired place blanket on cradle 2 Press blue ON ZERO button Make sure 0 appears on display 3 Place baby on cradle 4 Allow scale to lock onto the weight Prior Weight will show on the display 5 Remove baby from cradle Measurement value remains on the display 6 Press blue ON ZERO button to ready scale for next weighing OPERATION INSTRUCTIONS Console Keypad The scale control console has five buttons
15. settings that you may customize for your specific needs Tone on off A tone sounds whenever a weight is locked in or a button is pressed Factory setting is on LB GM lock The LB GM button can be disabled so that the scale will weigh in only the selected unit of measure Factory setting is unlocked Gram kg display The metric unit of measure may be displayed in grams i e 2541 gram or kilograms 2 541 kg Factory setting is grams To change any or all of these settings 1 Depress ON ZERO button and wait for the scale to display 0 0 2 Hold MENU button down and depress RECALL button Set 0 will be displayed and the 0 will be flashing 3 Depress the button repeatedly until Set 5 appears Depress the ENTER button 4 11011 will display with the left most 1 flashing This sets the Tone 1 on 0 off Use the A and W buttons to input your desired setting Press ENTER 5 The second position will now be flashing This disables the LB GM button 1 unlocked 0 locked out 6 The third and fourth positions are not currently used Press ENTER twice 7 The fifth right most position will now be flashing This switches the display from gram to kilogram 1 gram 0 kg Press ENTER Example If you desire to leave the tone on and disable the keypad so it only displays kilograms the user settings would be 10010
16. sole and wait for the display to show 0 0 Place an accurately known sure weight 3 000 to 5 000 grams recommended on the platform If the display reads other than the known weight value by 1 gram then the scale is out of calibration IMPORTANT NOTE A sure weight is a weight that has been accurately calibrated and its license is traceable to NBS weights amp measures Weight room free weights are not normally suitable to use as check weights as they may vary from their stated value If you do not have access to sure weights use a weight that has been weighed on a scale you are sure is accurate Remove weight from platform The calibration security code must be entered before attempting to calibrate the scale Hold the MENU button down and press RECALL button Release both buttons Set 0 will be displayed and the 0 will be flashing Depress the A button repeatedly until Set 2 appears Depress the ENTER button Use the A and keys to input 125 as the calibration security code Press ENTER to advance to the next digit After entering the third digit press ENTER and the scale will return to normal weighing mode The security code needs to be entered only once unless the scale is powered off Access to calibration is denied if the proper code is not entered With the known weight on the scale press REWEIGH Quickly before WeightLock activates hold the MENU button down and pres
17. tions manual contains instructions for use of this scale In order to achieve satisfactory results the operator must read this manual thoroughly before using this scale If any questions arise regarding this scale or its use please contact Pelstar LLC Cautions amp Warnings Failure to comply with the following may void the warranty on the scale To ensure proper operation of this electronic scale this operator s manual should be reviewed carefully before operation Keep this manual available for future reference and also for use in the orientation of new personnel If the scale becomes damaged it should not be used until properly serviced All repairs should only be performed by Pelstar LLC Do not gas sterilize or autoclave the scale Do not exceed the weight limit of 45 Ibs 20 000 g Scales contain delicate sensors Do not bang into doors walls etc as damage may result No scale should be used for patient transfer The manufacturer assumes no responsibility for patient injury or scale damage should this warning be ignored Do not leave patients unattended on the scale ASSEMBLY INSTRUCTIONS Before Assembly Carefully inspect the carton for shipping damage before unpacking If significant damage is found keep all packaging and contact your shipper or Pelstar immediately Claims must be filed with the shipper as soon as possible after receipt of damaged goods Carefully unpack the component
18. xcessive motion on the cradle If there was little to no motion during the most recent weight measurement STABLE will be displayed along with PRIOR WT Fig 8 If there was considerable motion during the recent weight measurement motion detection will activate after seven seconds and AVERAGE WT will be displayed along with PRIOR WT Fig 9 B WT Figure 8 Stable prior weight display mn tu gram mn tu gram 9 WT AVERAGE WT Figure 9 Motion detected during weighing Battery Life Indicators This scale has excellent battery life with most users getting more than two years from a single set of batteries Your battery longevity will vary with the frequency of use and the length of the Auto Off timer To help ensure you are never without the use of your scale it comes with a battery life indicator e Battery Pictorial Located in the bottom left corner of the LCD display is a facsimile of a battery With new batteries all three segments will be dark As the batteries become depleted the segments will turn off e Flashing Battery When the battery pictorial is empty and the frame is flashing replace batteries within a couple weeks Fig 10 NOTE Properly dispose of old batteries Install six 6 new alkaline C cell batteries Do not mix new and used batteries Figure 10 Battery indicators User Setting Options Your scale has three additional
Download Pdf Manuals
Related Search
Related Contents
BOB21M / BOB21ME BOB30M / BOB30ME Model MB130 Baby Scale Modelo MB130 Báscula para bebés Apocalypse Maya DD 110-D Manual de instrucciones Manual de instruções - Online Categorisation of Defects Betriebs- Instruction Notice anleitung handbook d'utilisation MARE001 palourdiere 6 gratins poisson sauce creme_a 1.2MB - 日立住友重機械建機クレーン Copyright © All rights reserved.