1. YES Sets save 0 0 Not saved 1 Save new settings 9 Adam Equipment Company 2012 4 0 CALIBRATION e Press the Zero key and power up the scale using the on off button e Release the Zero key and the display will show SPAN Make sure the hook is empty unless you intend to use any chains slings etc to hold the calibration mass Add any items such as these to the hook before continuing e Press the Zero button and the display will show SET then 00000 when the reading at zero is stable This completes the zero calibration e Following zero calibration the value of the mass being used to calibrate the scale should be entered Use the Hold button to increment the right hand digit and the Unit button to move this digit to the left Proceed by entering a number for the next digit if required until the value of your chosen calibration weight is displayed correctly e Load this mass onto the hook and when the scale is stable press the Zero key The display shows HOLD e After a few seconds the calibration is complete and the scale returns to normal weighing showing the value of the calibration mass as used To calibrate in Lbs make sure that the scale is Lbs mode before commencing to calibrate in Kg make sure the scale is in Kg mode before commencing 10 Adam Equipment Company 2012 5 0 PARTS LIST LHS LH
2. H ADAM Adam Equipment LHS LHSa SERIES P N 7 00 6 6 0200 Revision B November 2012 Software version L1 1 LA1 1 Adam Equipment Company 2012 1 0 CONTENTS TO CONTENTS dou tnnt hib asl tee eS a a cet sa ste eaa 1 210 INTRODUC TION c 2 21 FEATURES n ne ee ener ee a EN E Ree ere eee 2 2 2 TECHNICAL SPECIFICATIONS eMe 3 2 9 SAFETY NOTIGE cds OR ett a deber a ttes 4 2 4 KEY AND PANEL DESCRIPTION see eee 5 2 5 DISPLAY INDICATIONS WITH 6 30 OPERATIONS d od bit tb c t bi tee um ettet 6 3 1 SAFETY INSTRUCTIONS sese 6 3 2 BATTERY OPERA TON 555 a gag ct dace abun te BE Be aa ag Ea 7 3 3 BASIC OPERATIONS idt cat Ee aaa aa UE eet ERE R setts 8 34 REMOTE CONTROL UNIT dcn eite eR 9 9 5 USER PARAMETERS sng ecu aaa ag a a A e na Bg aa baa 9 2 0 GALIBRATION anaa aaa a a a neat niea Ev oa aa a a ue vam a NG Ta s 10 50 PAE SHES Tana aaa ana a uh ak aa a a a a a aa a Et 11 WARRANTY INFORMATION eeeen mnn 13 1 Adam Equipment Company 2012 2 0 INTRODUCTION 2 1 FEATURES The LHS LHSa Series are a range of heavy duty lower capacity crane scales The main features include e Easy Operation amp Rugged construction e 25mm bright red LED display enabling a good viewing angle e 360 degree rotary safety hook and support ring design e Overload safety up to 150 of model capacity e Selectable weighing units of kg Los and Newtons Newtons not ava
3. power switch is located on the rear housing Use this key to ON OFF p S i turn the scale ON and OFF Zero Press the Zero key to return the reading to zero Press the Unit key to change the weighing unit being displayed Unit between Kg Lbs and Newtons Newtons not available for a models Use this key to activate standard Hold When pressed the Hold LED will light at the top left side of the display and the reading on the display will lock and flash Press the Hold key again to cancel the Hold function When Peak hold is selected from the parameter section the Hold press of the Hold key will activate Peak Hold The Hold LED will flash and the display will also flash and lock showing the maximum weight registered of a single weighing process Press the Hold key again to cancel the Hold function 5 Adam Equipment Company 2012 2 5 DISPLAY INDICATIONS WITH LED The LED will light up in the weighing unit that is being used to Kg Ib N on PE indicate when the weight is stable Hold Hold Indicator 3 0 OPERATION 3 1 SAFETY INSTRUCTIONS Read this manual carefully before you operate the SHS Crane Scale Unpack the Crane Scale carefully from the packaging Attach it to your hoist or equivalent existing equipment securely at the ground level to the support ring in the upright position Make sure that the equipment to which the scale is attached is in proper condition The operator should be traine
4. publication may be reprinted or translated in any form or by any means without the prior permission of Adam Equipment Adam Equipment reserves the right to make changes to the technology features specifications and design of the equipment without notice All information contained within this publication is to the best of our knowledge timely complete and accurate when issued However we are not responsible for misinterpretations which may result from the reading of this material The latest version of this publication can be found on our Website www adamequipment com Adam Equipment Company 2012
5. Sa Part number Description Item No 7 00 1 2 0005 LHS COMPLETE HOUSING 500kg 1000lb 1500kg 3000lb 1 2 3 7 00 1 2 0012 LHS COMPLETE HOUSING 2000kg 4000lIb 1 2 3 7 00 4 0 0083 LHS ON OFF SWITCH 4 7 00 4 0 0086 LHS POWER CONNECTOR SOCKET 5 7 00 4 0 0084 LHS LEAD ACID BATTERY 6Vdc 3 2ah 6 7 00 1 0 0279 LHS RING SUPPORT 7 7 00 1 0 0273 LHS LOAD CELL 500kg 1000Ib 8 7 00 1 0 0274 LHS LOAD CELL 1500kg 3000Ib 8 7 00 1 0 0275 LHS LOAD CELL 2000kg 4000Ilb 8 7 00 1 0 0276 LHS HOOK 9 7 00 4 0 0085 LHS MAIN PCB 10 7 00 5 6 0045 LHS FASCIA PANEL 500kg 11 7 00 5 6 0046 LHS FASCIA PANEL 1500kg 11 7 00 5 6 0047 LHS FASCIA PANEL 2000kg 11 7 00 2 0 0081 LHSa FASCIA PANEL 1000 11 7 00 2 0 0082 LHSa FASCIA PANEL 3000lb 11 7 00 2 0 0083 LHSa FASCIA PANEL 4000lb 11 7 00 4 0 0094 LHS KEYBOARD PUSHBUTTON 7 00 3 0 0025 LHS COMPLETE PACKAGING SET 7 00 4 0 0109 LHS eVDC 600ma POWER CHARGER UNIT USA 7 00 4 0 0110 LHS 6VDC 600ma POWER CHARGER UNIT AUSTRALIA 7 00 4 0 0111 LHS 6VDC 600ma POWER CHARGER UNIT SA 7 00 4 0 0087 LHS 6VDC 600ma POWER CHARGER UNIT UK 7 00 4 0 0088 LHS 6VDC 600ma POWER CHARGER UNIT EU 7 00 4 0 0090 LHS REMOTE CONTROL UNIT 11 Adam Equipment Company 2012 WARRANTY INFORMATION Adam Equipment offers Limited Warranty Parts and Labour for the components failed due to defects in materials or workmanship Warranty starts from the date of delivery During the warranty period shou
6. d for its safe operation Carry out regular visual inspection for any damage or wear and tear on parts The split pins in particular at the hook and eyebolts as well as the inner wall of the hook and all joints should be checked regularly If any damage or wear has taken place stop using the scale and replace the parts Do all inspection and maintenance work when the scale is not in use Do not expose the scale to extreme temperatures The operating temperature range is O C to 40 C Secure the sample to be weighed in the hook of the scale Ensure that the sprung Safety Lever returns to its resting position which prevents the sample from slipping out of the hook Do not exceed the maximum capacity of the Scale Clean and protect your scale after use 6 Adam Equipment Company 2012 Overload Warning Please do not add any item that exceeds the maximum capacity of the scale When the display shows remove the item to avoid damage to the load cell and scale 3 2 BATTERY OPERATION e The rechargeable battery will operate continuously for between 20 and 30 hours before requiring recharging e To charge the battery plug the charger unit into the socket as marked on the rear of the scale Plug the charger into the mains power supply and wait until the LED on the charger shows green The battery is now fully charged e Charging time for a completely discharged battery is approximately 16 hours A full battery pack is recomm
7. ended to keep the scale continuously operational e Only use the charger as supplied with the unit e The scale will automatically power down after 30 minutes of non use to further protect the battery e To replace the battery turn the locking screw on the battery cover 90 degrees clockwise and lift open the cover Disconnect and remove the old battery and replace with a new one e Contact your dealer for a replacement battery 7 Adam Equipment Company 2012 3 3 BASIC OPERATIONS Power on and off Press the ON OFF button to turn the scale on or off The display on the scale will show the model capacity then countdown from 9 to before displaying zero weight Zero the scale Press the Zero key so that the display shows zero Change the weighing Unit Press the Unit key to change the weighing unit Standard Hold setting 1 When the reading may not be perfectly stable or you cannot view a reading easily you can press the Hold key to lock it The reading will be locked and also flash at the point you press the Hold button enabling you to view it when you can The Hold LED will also light up and once you have noted the reading a press of the Hold key will return the scale to the normal weighing mode Peak Hold setting Peak Hold is the highest weight reading registered by the scale when a weighing takes place Press the Hold key when the display is at zero and the Hold LED will flash which denotes Peak Hold
8. has been activated When weight force is applied to the scale the display will also flash and lock the highest recorded weight of that weighing process This value and the Hold LED will remain flashing on the display until the Hold key is pressed again at which point the scale will return to normal weighing mode displaying the current weight on the scale Press Hold again to perform another Peak Hold weighing at any time 8 Adam Equipment Company 2012 3 4 REMOTE CONTROL UNIT The remote control works via infra red and can be used to operate the SHS from long distances It operates using 2 x AAA batteries and these can be exchanged by unclipping a cover on the rear of the remote control unit 3 5 USER PARAMETERS Five USER parameters are on the SHS to access these power up the scale and whilst the display is counting down press and then release the Unit button The display will show 00000 Use the Hold and Unit buttons to enter the code 11111 and then press Zero bEE 1 will be displayed Press the Hold button to change the setting and the Zero button to move to the next parameter PARAMETER SETTING TABLE FUNCTION DESCRIPTION DEFAULT BEE Sets buzzer 1 0 Disable 1 Enable Hop Sets the Hold function 1 0 Peak 1 Standard STB Sets the Stable function 1 0 Disable 1 Enable Sets weighing unit on power up CHT 0 kg 1 Lb s 0 2 Newtons Newtons not available for a models
9. ilable for a models e Standard and Peak Hold feature e Internal rechargeable battery fitted as standard e Auto power off after 30 minutes of non use e 6Vdc 600ma battery charger e Strong Die cast Alluminium housing e Infrared remote control 2 Adam Equipment Company 2012 2 2 TECHNICAL SPECIFICATIONS Model No Capacity Division Model No Capacity Division LHS 500 500kg 0 1kg LHS 1000a 1000Ib 500kg 0 21b 0 1kg LHS 1500 1500kg 0 2kg LHS 3000a 3000lb 1500kg 0 5Ib 0 2kg LHS 2000 2000kg 0 5kg LHS 4000a 40001b 2000kg 1Ib 0 5kg PT TC Nw EV DL 2 E Adam Equipment Company 2012 2 3 SAFETY NOTICE The limit stated below is the Ultimate Overload Limits of the scales Adam Equipment Co Ltd Its directors owners and affiliated companies and organisations cannot be held liable for any resulting damage or injury sustained as a result of the machines being used in excess of their stated capacities or ultimate overload limits Model No LHS 500 1000a LHS 1500 3000a LHS 2000 4000a Max capacity 500kg 1000Ib 1500kg 3000lb 2000kg 4000lb Overload capacity 750kg 1500Ib 2000kg 4000Ib 2500kg 5000Ib NOTICE Do not use the scales if any parts are bent or loose or show signs of damage or wear Refer to the user manual 4 Adam Equipment Company 2012 2 4 KEY AND PANEL DESCRIPTION The
10. ld any repairs be necessary the purchaser must inform its supplier or Adam Equipment Company The company or its authorised Technician reserves the right to repair or replace the components at any of its workshops depending on the severity of the problems However any freight involved in sending the faulty units or parts to the service centre should be borne by the purchaser The warranty will cease to operate if the equipment is not returned in the original packaging and with correct documentation for a claim to be processed All claims are at the sole discretion of Adam Equipment This warranty does not cover equipment where defects or poor performance is due to misuse accidental damage exposure to radioactive or corrosive materials negligence faulty installation unauthorised modifications or attempted repair or failure to observe the requirements and recommendations as given in this User Manual Additionally rechargeable batteries where supplied are not covered under warranty Repairs carried out under the warranty does not extend the warranty period Components removed during the warranty repairs become the company property The statutory right of the purchaser is not affected by this warranty The terms of this warranty is governed by the UK law For complete details on Warranty Information see the terms and conditions of sale available on our web site Adam Equipment Company 2012 Manufacturer s Declara
11. ntended to be incorporated into EEE devices as defined by European Directive 2002 95 EEC must be recycled or disposed using techniques that do not introduce hazardous substances harmful to our health or the environment as listed in Directive 2002 95 EC or amending legislation Battery disposal in Landfill Sites is more regulated since July 2002 by regulation 9 of the Landfill England and Wales Regulations 2002 and Hazardous Waste Regulations 2005 Battery recycling has become topical and the Waste Electrical and Electronic Equipment WEEE Regulations are set to impose targets for recycling Adam Equipment Company 2012 ADAM EQUIPMENT is an ISO 9001 2008 certified global company with more than 35 years experience in the production and sale of electronic weighing equipment Adam products are predominantly designed for the Laboratory Educational Health and Fitness retail and Industrial Segments The product range can be described as follows Analytical and Precision Balances Compact and Portable Balances High Capacity Balances Moisture analysers balances Mechanical Scales Counting Scales Digital Weighing Check weighing Scales High performance Platform Scales Crane scales Health and Fitness Scales Retail Scales for Price computing For a complete listing of all Adam products visit our website at www adamequipment com Copyright by Adam Equipment Co Ltd All rights reserved No part of this
12. tion of Conformity This product has been manufactured in accordance with the harmonised European standards following the provisions of the below stated directives Electro Magnetic Compatibility Directive 2004 108 EC Low Voltage Directive 2006 95 EC Adam Equipment Co Ltd Bond Avenue Denbigh East Milton Keynes MK1 1SW United Kingdom FCC COMPLIANCE This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment The equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense Shielded interconnect cables must be employed with this equipment to insure compliance with the pertinent RF emission limits governing this device Changes or modifications not expressly approved by Adam Equipment could void the user s authority to operate the equipment WEEE COMPLIANCE Sealed Lead Acid Battery Must be recycled Properly Any Electrical or Electronic Equipment EEE component or assembly of parts i

Download Pdf Manuals

image

Related Search

Related Contents

Fujifilm FinePix J50 User's Manual  MRS Web Assistant  Flexible architecture for an embedded interrupt controller    Manual - Network thermometer Querx PT100  A propos d`Amadeus  Vantec ICEBERQ 4 Premium VGA Cooling Kit  Philips 28PW6108/01 Flat Panel Television User Manual  Whirlpool MW7500XW User's Manual  