Physics 235 Multichannel Analyzer (MCA) & Gamma

image

Contents

1. peaks all peaks that are visible of the spectrum Which of the peaks is most affected by the leads Explain why Exercise 2 Analysis of the Energy Spectrum of Known and Unknown Gamma Sources Delete all the ROIs and clear the spectrum of Na22 you collected but do not clear the energy calibration or close the Maestro program Closing the window will eliminate your energy calibration Do not change the geometry and amplifier settings Take the Na22 source and put it away from the detector put it back in the sources storage 6 TS2011 MF2001 RA98 a Cs 137 Co 57 and Co 60 Spectra Run the measurement for the energy spectrum for Cs 137 Co 57 and Co 60 Click on the prominent peak s and read the corresponding energy at the bottom of the window with the corresponding channel number and counts Go to the Calculation menu then Peak Info Take a note of the information displayed in a box above the peaks Go to the isotope library and confirm the corresponding energy of the peak s for each of the three isotopes Make a table of the peak values keV for the Gamma ray Compton edge and backscatter and compare to expected theoretical values as well as percentage error please research and try to understand what the Backscatter and Compton peaks are and their source Save your data in your directory Use WinPlots to print the results Clear the window but do not close it b Unknown Source s Obtain the mystery gamma source enclosed in black t
2. the arrangement shown in Fig 4 to the right photo tube base Important safety note Be sure to use one of the special high voltage cables with a longer than usual barrel for the high voltage connection Please ask someone if you can t find one of these special cables Using a regular BNC connector is dangerous The bi polar output of the amplifier should be used if it has one with the leading edge positive There are two parameters that ultimately determine the overall gain of the system the high voltage that is furnished to the photo Figure 4 Schematic for y Spectrometer multiplier tube and the gain of the linear amplifier The gain of the phototube is largely dependent upon its high voltage The high voltage value depends on the photo tube being used You will be using the Harshaw Nal TI1 detector along with a Fluke high voltage supply Set the latter for about 700 volts never exceed 1000 volts and adjust the amplifier gain to the middle of its range Be sure to record the gain both coarse and fine of the amplifier and high voltage settings in your lab notebook Having these numbers on hand will allow you to continue the experiment on another day without having to do a new calibration curve High Voltage Power Supply Amplifier Oscilloscop MCA card Computer G a Use a Na 22 gamma ray source for the energy calibration as in the MCA exercise experiment Put the source in the smaller cradle in th
3. F2001 RA98
4. Physics 235 Multichannel Analyzer MCA amp Gamma ray Spectroscopy Purpose Say you have a radiation detector that puts out an electrical pulse whenever a gamma ray passes through it Furthermore the voltage of that pulse is proportional to the energy of the gamma ray You want to keep track of how many gammas are detected as a function of gamma energy How do you do it One way is to use a Multichannel Analyzer An MCA is a device that bins voltage signals from a detector and produces a histogram of countrate vs energy In this lab you will learn the basics of using an MCA to analyze detector signals Important Preliminary Note on Background Material The type of radiation detector we will be using in this course is a phototube that is optically coupled to a scintillating crystal There is a great deal of physics involved in these detectors The MCA spectrum that we will be examining contains many features that are due to processes that occur in the scintillating crystal To fully understand the spectra you ll be studying you ll need to know the meaning of such terms as photoelectric absorption Compton scattering Compton shoulder back scatter peak pair production double escape peak single escape peak and how all these are related to the working of a phototube detector Many texts on Modern Physics cover these topics See for example Modern Physics for Scientists and Engineers by S K Kim and E N Strait pages 385 391 check it in the l
5. al scale to Log and go to Acquire then Start Depending upon the resulting spectrum adjust the amplifier gain up or down so that the entire spectrum occupies most of the 512 or 1024 channels which are available When you are satisfied with your setup clear the spectrum and reset the Live Time for a sufficiently long period so as to accurately determine the centroids of the peaks Then start a normal collection Once a good spectrum has been accumulated stop acquiring and save the data in your home directory and do an energy calibration following the procedure that you learned while doing the MCA exercise Save your data with file extension Chn in your directory You can also print it by opening it in WinPlots see section 7 1 of the manual about WinPlots To start WINPLOTS click on Start on the Windows Taskbar then Programs MAESTRO 32 and WinPlots The spectrum files are associated with WINPLOTS by the installation program so double clicking on a spectrum filename within Windows Explorer will start WINPLOTS and display that spectrum too You can smooth your spectrum once you have saved it in the buffer but not necessary for now Finally check the effect of putting lead bricks or lead sheets around the source detector Clear the spectrum and surround the source detector with lead sheets Collect the data again on Na 22 for about the same duration and same geometry as when there were no lead sheets What effect does this have on the
6. ape do not remove the tape from the radioactive source storage area and use it to replace the last source you used which should be moved well away from the detector The geometry should be the same as in the previous exercise Accumulate a spectrum for a period long enough to accurately locate all of the peaks Backscatter and Compton edge After the collection has stopped note the energy of all peaks Compare these energies with the isotope library energies to deduce what the mystery source is Remember that the sources you are looking at has been in our lab for at least 10 years Thus it does not contain any isotopes with very short half lives Include a copy of the spectrum in your notebook and be sure to identify all the features Compton edges Photopeaks Backscatter peaks etc by hand or by using Word After the isotope s has have been correctly identified record the energies of the gamma rays their relative strengths and the lifetime s Also include a sketch of the decay scheme do research on this for the isotope in your notebook Exercise 3 Mass Absorption Coefficients What is an absorption coefficient Make sure you know the answer to this question before continuing See Melissinos or any other modern physics textbook for a discussion of the absorption of radiation by matter The total mass absorption coefficient can be measured with a gamma ray spectrometer In this experiment you will measure the number of gamma rays t
7. ay to a 255 min particular level in an adjacent isotope followed by one or more gamma emissions as the residual nucleus de excites to its ground state Any het et gt electrons emitted by the source will be absorbed by the aluminum light 661 6 keV shield surrounding the detector s scintillator material and therefore will iin my 5 not be counted at all The gamma rays however are quite penetrating and will easily pass through the shield In this experiment you will set up and calibrate the spectrometer using Na 22 following which you will Figure 3 Cs Decay attempt to identify an unknown gamma ray source Afterwards you will investigate the penetration power of gamma rays in lead absorbers and measure the dead time of the spectrometer Before you begin take a look in the Table of Isotopes available in the lab and find the Cs decay shown above Careful study will reveal that there is another pathway through which Cs can decay to S7Ba What is that pathway Find the decay scheme for Na Is there more than one pathway for it Also look on the Chart of Nuclides found on the wall in the lab and find 137Cs and Na and see what information is contained there about their decays With the introduction above you are now in a position to carry out your own experiment on different isotopes However first you need to re do the energy calibration of your spectrometer set up using Na22 isotope To do this set up the electronics according to
8. e holder which has been provided Position the holder so 5 TS2011 MF2001 RA98 that the other end butts up against the front of the phototube the source should then be about 5 cm in from the detector Observe the amplified pulses from the detector on the oscilloscope The vast majority of them should be 2 6 volts high and clipping of pulses should not be a frequent event a few are OK Do you know what is meant by clipping Adjust the amplifier gain to achieve a reasonable distribution of pulse heights note the analyzer will accept up to 10 volt pulses Click on the Maestro shortcut on the desk top to open it if you have already closed the window from the MCA exercise Since the Nal crystal peaks are quite broad you do not need many channels 512 or 1024 will do You can increase or decrease the number of channels by clicking on the zoom out or zoom in in the menu bar To adjust the amplifier gain so that the peaks occupy most channels but are still on scale you will need to take some short sample spectra Select what seems like a reasonable short collection time by going to the Acquire menu then MCB properties and choosing Presets select Live Time as the Preset Type and type in a Preset Value of 60 seconds The Pulse Height Analysis PHA window is on the upper right corner of the window to monitor and register the Real Time Live Time and percentage Dead Time Record them and explain what they represent Set the vertic
9. f by default Go to ROI and choose Mark This sets the ROI status to be active for marking Next click at the left bottom of the annihilation peak and using the right arrow on the keyboard move the cursor to the right bottom of the annihilation peak You can equivalently do this by placing the marker at the center of the peak and choosing Insert from the keyboard You will see the peak painted If you want to clear the painting go to ROI then Clear Clear all for more than one peak painted An alternate method to paint the peaks is by using the menu bar Calculate Peak Search The software automatically paints peaks that fulfill the sensitivity criterion If unwanted peaks are painted then click on the center of the peak to locate the marker then go to ROI and clear This option sometimes omits peaks that you are after if they do not meet the sensitivity criterion so manual painting or using the Insert key are the better options Now locate the marker centered on the top of the annihilation peak and select ROI from the main menu Then paint the area under the peak using the Insert key observe that the peak is painted red If you feel the painted area is wider than the width of the peak you are intending to paint then go to ROI mark and paint it manually by clicking at the left side bottom of the peak to locate the marker and then use the right arrow key to paint until reaching the right hand side bottom of the peak For more fac
10. hat are absorbed when lead sheets are placed between the source and the detector By varying the thickness of the absorber it is possible to measure the mass absorption coefficient Since Na 22 has two well defined peaks in its spectrum one can determine this coefficient for two different energies at the same time 1 Return the mystery source to the radioactive storage area and place the Na 22 source in front of the Nal Tl detector with the same geometry as used previously Go to Acquire MCB properties and set the Live Time to 60 seconds and start a collection Paint the ROIs for the photo and annihilation peaks If the combined counts in these peaks are not adequate then increase the Live Time and perform another collection Once a satisfactory count level is attained record the absorber thickness and NET counts in each ROI in a table constructed for purpose of 7 TS2011 MF2001 RA98 this experiment Also record the Live Time that you used to collect the data so that counts per second can be calculated later 2 Clear the spectrum and place the thinnest lead absorber from the absorber set A E in the adjustable slot immediately in front of the source you may need to use some putty to hold the slider in place You should be careful to maintain the position of the source just as it was in the previous measurement Accumulate the spectrum for the same period as in step 1 above Record the absorber thickness and NET counts in both peaks in
11. ibrary Please make brief notes on the above topics before starting this lab You ll be lost without this background information Introduction A single channel analyzer SCA will generate a 5 V logic pulse each time an input pulse is greater than E but less than E AE A multi channel analyzer is a succession of single channel analyzers whose windows are all the same width and are arranged sequentially in order of increasing energy By plotting the count rate of each SCA versus its mean energy window setting one obtains a spectrum of count rate versus energy In an actual MCA the single channel analyzers are replaced by a single device called an analog to digital converter ADC The ADC measures the height of each pulse as it comes in and adds a count to a memory bin or channel corresponding to that voltage The voltage range of 0 5V is typically divided into 512 or more channels As well as the counts in each channel the MCA also records the collection time often called real or live time so that count rates can be determined The exact correspondence between pulse height and gamma ray energy depends on many factors for example the type of phototube being used what voltage it is operated at how much its output signal is amplified the geometry of the scintillating crystal etc Consequently the MCA must be calibrated by using an input with known pulse heights For example if a gamma ray spectrum with two peaks whose energies are kn
12. ilities and different options to be used read more of the MCA Emulator User s Manual Exercise 1 Energy Calibration and Peak Identification To accurately determine the energy of these peaks go to the Service menu then Library Select Peak This will open a window containing a list of nuclide their energy peaks percentage of disintegration and half life You can sort the library list by nuclide energy percent or life time Search for the Na 22 isotope and record any gamma ray energies you find Also write down the percent a number between 0 and 100 that indicates relative rate of the gamma decays and the half life When you are done minimize this window and maximize the MCA window that contains the Na 22 data sample0 Chn or sample01 Chn If the marker is not already centered on the annihilation peak locate the marker at the center of the peak and press Insert to paint it or paint it manually as explained above Go to Calculate Calibration The program will ask for the energy of the peak in keV not MeV Enter the 3 TS2011 MF2001 RA98 Energy and press lt Enter gt or click OK Move the marker to the next gamma peak and repeat the process After the second peak value is entered the program will ask for the units click OK The system is now calibrated and the marker location is shown at the bottom of the window in both channels and input keV units If you are not happy with your calibration process and want a complete
13. new recalibration you can click on the Destroy Calibration button by going to Calculate Calibrate then Destroy Calibration If the spectrum is calibrated the channel centroid rates ani a FWHM FW1 xM in channels and calibration he Ace 600850 2873 units e g energy library best match energy and activity gross area net area and net area uncertainty are displayed for the ROI peak marked by the marker This information is displayed beneath the Spectrum Window on the Supplementary Information Line and or in coos Marker 4 074 4 074 00 keV 64 255 C a text box inside the Expanded Spectrum View Peak 4073 43 4073 43 keV No close library match at the top of the peak Both displays are illustrated Gross Area 634163 Net Area 600668 2873 in Fig 2 To remove the text press lt Esc gt The program subtracts the calculated background channel by channel and attempts a least squares fit of a Gaussian function to the remaining data If unsuccessful it displays Could Not Properly Fit Peak If successful the centroid is based on the fitted function The reported widths are linearly interpolated between the background subtracted channels Fig 2 Peak Info Beneath Marker and at Top of Peak To learn about energy calibration techniques go to the menu bar and select Calculate This menu gives Sensitivity factor This is a criterion for the peak to be acceptable based on the second difference compared to weigh
14. own is analyzed by the MCA then a correspondence between the energies and the associated peak channel numbers can be made as E shown in Fig 1 This calibration assumes that all the equipment is strictly linear For a more precise calibration three or more known peaks up to 96 peaks Figure 1 Energy Calibration 1 TS2011 MF2001 RA98 can be used then a least squares fit can be made to the calibration points using higher order polynomials If only one calibration point is available then the MCA automatically assumes a linear fit that passes through the origin The MCA provides for the manipulation of the spectrum display with expand contract and linear logarithmic sqrt modes and various ways of displaying the data points In addition there are powerful data processing capabilities available in the MCA The main features are location of multiple regions of interest in the spectrum determination of centroid positions background subtraction energy calibration and peak identification While most of these operations can be performed automatically it is possible to do some of them under the control of the user Using the MCA software Start the computer and open the Maestro for Windows program on the desktop If not deleted you will see the spectrum previously collected Go to Acquire on the main menu and clear the window The MCA hardware is on a card installed in the computer Help from the user s manual All multi channel analy
15. ted error The sensitivity factor can be set between 1 and 5 an integer 1 being the most sensitive finds the most peaks i e it includes most small peaks Please read section 4 3 of the manual on this Close or clear the window and you are ready for the gamma ray spectroscopy If you wish you can save your work in your home directory with the file extension Chn You cannot save on the K drive Gamma ray Spectroscopy We will now begin collecting and analyzing the spectra of actual gamma ray sources One of the goals of the lab will be to identify an unknown source by measuring its gamma ray spectrum and comparing this with data in published isotope tables We will also investigate the absorption of gamma rays by matter A word of caution It is absolutely imperative to understand the physics involved in a phototube scintillator detector before trying to interpret spectra Please review the reference given on the first page before proceeding with this lab Another good reference for this lab is your textbook Experiments in Modern Physics by Melissinos All of Chapter 8 is devoted to a discussion of various types of radiation detectors Section 8 4 page 333 344 deals specifically with scintillation counters 4 TS2011 MF2001 RA98 gost Most isotopes that are used for gamma ray measurements have 300 yr associated beta particles in their decay schemes A typical decay B scheme for an isotope as shown in Fig 3 will include beta dec
16. the two annihilate to produce two photons traveling in opposite directions to conserve linear momentum The first and largest peak is due to annihilation photons What is the energy of one of those photons Notice that the channel position of the marker is indicated along with the corresponding number of counts at the bottom of the window Try moving the marker by using the mouse and or the arrow keys on the keypad There are many options available for viewing the spectrum and we will begin by looking at a few of them First put the marker near the top of the photo peak by positioning the mouse pointer there and then clicking the left mouse button At times one wants to see more detail either for more accurate positioning of the marker or when there are several peaks close together The 2 TS2011 MF2001 RA98 small inserted window helps you to zoom in or out The colors of the spectrum and data points can be selected by using the Display References and Spectrum Colors ROIs Regions of Interest The goal in any spectrum analysis is to identify the elements responsible for the radiation and to determine the activity of each of them Consequently the peaks in the spectrum become regions of interest ROTD that need to be delineated Each peak in the spectrum will have one ROI for closely spaced peaks the ROIs may overlap To learn about painting and editing a ROI go to Section 4 5 of the MCA Emulator User sManual The ROI status is of
17. your table e Clear the spectrum and insert the next thickest piece of lead in the holder Once again determine the NET counts for the same time Repeat in this fashion until you have used all the lead absorbers in the set In your notebook give a qualitative description of the effect of the lead absorbers upon the energies and number of gamma rays reaching the detector To make it quantitative normalize the count rate R for thickness x to the gross count rate Ro for no absorber Plot the natural log of R Ro vs absorber thickness in cm for both energies and determine the resulting slopes Also make an estimate of the error in these slopes The slopes correspond to the linear absorption coefficient Ux A mass absorption coefficient may also be defined for a material by Un Ux p where p is the material s density What are the advantages of reporting absorption in terms of Um instead of ux Using your slopes calculate the mass absorption coefficient in cm mg for gamma rays at these two energies along with their errors Finally compare your results with those plotted in Fig 5 14 Cs Zn sb Na t p m Oo ht o to 2 P Linear absorption coefficient cm rp a Fed o K 01 02 030405 07 10 2 345 7 10 Energy mev Figure 5 Linear absorption coefficients of several elements as a function of gamma ray energy taken from Introductory Nuclear Physics by David Halliday page 143 1955 8 TS2011 M
18. zers have numerous features in common learning about one model will help you operate any of them To get more help on using this program got to the Physics Manuals website at http www physics wwu edu manuals Then under the list of categories go to Ortec and open MAESTRO 32 MCA Emulator User s Manual You ll need to refer to this manual whenever you need help Viewing the spectrum In order to learn more about the MCA open the file called sample0 Chn or sample01 Chn which should be located in the directory K Physics P235 MCA To do this go to File then Recall and open the above file s One of the files is stored in 512 channels while the other one is stored in 1024 channels You can Recall these files one by one and view them The window will be divided into two to accommodate the two files spectra You can close one of the windows and expand the other to work with The vertical scale is logarithmic Log The other possible scale is linear A Click and see these scale indicators and observe the spectrum Select one the Log scale is recommended to see possible peaks clearly above the background The displayed spectrum was taken in this lab using a Na 22 source and a Nal T1 detector Na 22 decays by emitting a positron and then the resulting isotope very quickly emits a gamma ray which is evidenced by the second prominent photo peak to the right The positron does not travel far before it meets an electron and