1. Figure 1 View of a MDCC20 chip lying on top of a Petri dish right allowing multiplexed culturing compared with conventional culturing of a single strain on agar left 2 Sterility The MDCC is usually supplied sterile but in some cases the user might wish to perform their own sterilization procedure The MDCC can be sterilized by washing with 95 v v ethanol in the supplied tube Excessive ethanol can be poured off The remaining ethanol can be evaporated by placing the open tube ina laminar flow cabinet overnight Do not remove ethanol by flaming www microdish nl 1 MDCC User Manual and Protocols July 2012 Alternatively chips can be sterilized using ultraviolet light plasma cleaning or gamma radiation Normal power and time settings as used for sterilizing dry glassware or plastics are recommended Sterilization by autoclaving and dry heat are not appropriate 3 Chip Handling Orientation shiny side is UP this side is where microorganisms can be inoculated The dull grey surface is the underside of the chip that should contact the nutrient source such as an agar plate To place the MDCC on an agar plate open the tube under sterile conditions and gently tap the tube until one end of the chip contacts the agar Then gently withdraw the tube so the chip lies flat with the shiny side upwards Several ways are possible to move a MDCC from one agar plate to another Such an operation might be required when moving a ch
2. M 2007 A highly subdivided microbial growth chip constructed on a porous ceramic support PNAS 46 18217 18222 GREAT THE FUNNY LOOKING COCCI 2 0330 FROM THE LEFT LOOK NEW ft Aa Ds Zoe N Contact information MicroDish B V Kruytbuilding Padualaan 8 3584 CH Utrecht The Netherlands T 31 0 30 25 38 298 E info microdish nl www microdish nl This is proprietary information of MicroDish BV and is intended as guideline for usage No guarantee for a particular application can be given CAUTION Not intended for human or animal diagnostic or therapeutic use The General Terms and Conditions of MicroDish B V apply ioe www microdish nl
3. MDCC User Manual and Protocols July 2012 MicroDish Culture Chip MDCC 1 Introduction to the MDCC The MDCC is a disposable item for culturing microorganisms It is basically a large number of miniaturized Petri Dishes on a chip The MDCC has an 8x36 mm footprint although other configurations are possible The base of the MDCC is made out of porous aluminium oxide PAO with a pore size of about 0 2 um This structure allows nutrients from beneath the chip to diffuse through the pores to the surface and consequently supplying microorganisms with nutrients that are growing on top of the surface as microcolonies Microorganisms cannot penetrate the porous support The upper surface of the MDCC is divided into individual growth compartments microwells MicroDish BV supplies MDCCs in various designs depending on the envisaged assay For example the MDCC180 contains thousands of round compartments that are 180 um in diameter The MDCC180 is suitable for the isolation of previously uncultured organisms from complex samples and other analyses where differential growth rates and or a high degree of segregation is required The MDCC20 has about 180 000 square wells that are 20 um across This chip is applicable for high throughput screening HTS and viable counting The MDCCs is versatile allowing the rapid exchange of culture medium working with difficult organisms and is highly suitable for imaging and recovery from target microcolonies
4. chip starts to curve the plasma treatment has been too extensive D Remove the chip The chip can be briefly treated with ethanol UV or gamma radiation for sterilization purposes The necessity of this step will depend on the quality of the air supply and has to be considered individually www microdish nl 6 MDCC User Manual and Protocols July 2012 11 Troubleshooting Inoculation takes a while to dry Patience is advised Sometimes the surface chemistry does cause wetting to be delayed in time The agar plates should be solidified before the MDCC is deployed Washing with methanol and or plasma cleaning can be used to make the MDCC more hydrophilic if necessary Section 10 Plasma Cleaning MDCC Overgrowth of microcolonies The individual compartments on the MDCC will eventually be overgrown This will occur faster with the MDCC20 as the wells are small and the walls are only 10 um high It is recommended to use the MDCC180 for culturing complex samples e g in which some strains grow much faster than others The chip breaks Shear forces should be avoided It is recommended to slide a small piece of laboratory parafilm or a very thin spatula under one corner of the MDCC until the chip can be lifted and transferred to a different location Note Broken pieces can still be used 12 Frequently Asked Questions Can you custom build chips Yes Various MDCC designs are available Please contact MicroDish B V for more infor
5. dure might be applied for identification purposes by sequencing 16S rRNA encoding genes Alternatively cells can be inoculated for multiple displacement amplification MDA reactions for whole genome sequencing Recovered microcolonies can be transferred onto poly L lysine glass slides for FISH analysis immunofluorescence or negative staining Inoculation of recovered cells into specific growth media for further cultivation 10 Plasma Cleaning MDCC Plasma cleaning is a method for decontamination and for increasing the hydrophilicity of surfaces Plasma is a partially ionized gas and includes free electrons that vaporizes hydrocarbons from a surface which are immediately and completely removed This cleaning method is often superior to solvent cleaning methods Plasma cleaning does not prevent the re adsorption of material onto the cleaned surface storage under vacuum is therefore recommended Plasma treatment can be used to make the MDCC chip more hydrophilic and to fine tune the surface chemistry Plasma Cleaning Protocol A Wash the MDCC chip briefly with methanol air dry and place it on a clean microscope slide compartments up inside the plasma chamber Use a sterile microscope slide if necessary B Create the vacuum in the plasma chamber according to the normal protocols for the used device C Treat chips with plasma starting with 15 sec at lowest power for platinum treated chips and 10 sec if uncoated If the
6. ferred to a second LMP slide without dye for de staining purposes Figure 3 A section of a MDCC20 chip using fluorescence microscopy equipped with a 10x Fluorotar objective lens Escherichia coli strain XL2Blue expressing GFP was grown on the chip The picture was taken with a black and white CCD camera Compartments filled with GFP expressing cells appear white whereas empty ones are darker with a brighter rim Dissection Microscope A relatively low powered microscope total magnification 4x to 20x is also suitable for imaging Light is most effective from above ideally at a 45 angle This is an effective method when monitoring a chip over a period of time for example as part of a procedure to recover growing microcolonies USB microscopes may also be suitable Note Not all cultures within the MDCC may be visible by this method www microdish nl 4 MDCC User Manual and Protocols July 2012 Figure 4 Section of a MDCC160 using a dissecting microscope 10x Microcolonies can be seen in some of the compartments which appear white over the entire well surface Here illumination was from above at a 30 angle Scanning Electron Microscopy SEM It is recommended to fixate microcolonies prior SEM imaging SEM fixing Protocol A o0 It Prepare an agar plate containing the SEM fixative Other components e g growth medium cofactors may be added to the agarose Note Fixatives might be harmful In such a ca
7. ip from an agar plate to a staining slide It is left to the individual user which technique to apply Possible procedures are 1 Chips can be moved and transferred to a different agar plate by gently lifting them with a piece of laboratory parafilm To do so gently slide a piece of sterile parafilm between the MDCC and the agar then lift the MDCC to the new location 2 Avery thin spatula 5 10 mm width can be used to lift up the MDCC For this carefully place the tip of the spatula at the short end of the MDCC lift it up by 1 2 mm and then slowly slide it under the MDCC While lifting the chip up transfer it to the new location The transfer of the MDCC should occur as quickly as possible to prevent the cells from drying out Dry chips can be manipulated with tweezers but are brittle so some care is required 4 Inoculation Inoculation is done by pipetting or gently spreading microorganisms on the upper surface of the chip In HTS it is recommended to plate organisms with a probability of growth in any given compartment of lt 5 to ensure good separation in other words not every compartment has to contain a microcolony for an effective screening it is often better to ensure that most compartments that support growth are a monoculture The MDCC180 can be inoculated effectively with 5 ul droplets of the sample It is advisable to use a dilution series in order to find the optimum dilution Higher dilutions increase the possibili
8. mation Can I cut chips This is not advisable Can I reuse chips The MDCCs are designed as disposable items see faint rings around the edges of compartments when using fluorescence microscopy why This is autofluorescence of the material and a result of chip manufacturing It can be a useful feature enabling identification of compartments 13 References Staining and SEM protocols suitable for MDCC e Ingham CJ van den Ende M Pijnenburg PC Wever PC and Schneeberger PM 2005 Growth of microorganisms in a multiplexed format on a highly porous inorganic membrane Anopore App Environ Micro 71 8978 8981 e Ingham CJ van den Ende M Wever PC and Schneeberger PM 2006 Rapid antibiotic sensitivity testing and trimethoprim mediated filamentation of clinical isolates of the Enterobacteriaceae assayed on a novel porous culture support Journal Medical Microbiology 55 1511 1519 www microdish nl 7 MDCC User Manual and Protocols July 2012 e Hesselman MC Ingham CJ et al A multi platform flow device for microbial co cultivation and microscopic analysis PLoS ONE 2012 7 5 e36982 e Ingham CJ Sprenkels A Bomer JG van den Berg A and van Hycklama Vlieg JET 2010 High resolution contact printing and transfer of massive arrays of microorganisms on porous aluminium oxide Lab on a Chip 10 1410 1416 The MDCC20 chip e Ingham CJ Sprenkels A Bomer JG Molenaar D van den Berg A van Hycklama Vlieg JET and de Vos W
9. s are low in auto fluorescence Very intense stains can help to detect a higher amount of microcolonies on a MDCC It is also possible to identify specific microorganisms in HTS depending on the applied dye It is recommended to use Fluorotar lenses or other objectives suitable for fluorescence microscopy Microorganisms grown on the MDCC can be stained using solidified agarose containing the dye as described below www microdish nl 3 MDCC User Manual and Protocols July 2012 Fluorescence Staining Protocol A Pour 1 ml of barely molten low melting point agarose containing the dye on a standard microscope slide Typically use the concentration recommended by the manufacturer for staining in liquid culture or slightly higher Other components e g growth medium cofactors may be added to the low melting point LMP agarose If the dye is robust at high temperatures conventional agarose can be used B Allow the LMP agarose to solidify in the dark to project the applied dye from photobleaching at room temperature typically for 1 h C Transfer the MDCC onto the dye slide and incubate for a specific period As guideline an incubation time of 30 minutes at room temperature should be sufficient for nucleic acid stains Note The LMP should be gelled and not too wet prior staining as this might result in redistribution of microcolonies on the MDCC D The MDCC containing stained microorganisms is ready for imaging or might be trans
10. se plates are best poured using barely molten agar in a fume hood that protects the user from the fixative Allow the agarose to solidify at room temperature typically for 1 h Transfer the MDCC onto the agar plate and incubate Note The agarose should be gelled and not too wet prior fixation as this might result in redistribution of microcolonies on the MDCC chip The MDCC chip containing fixed microorganisms is ready for imaging Recovery of Samples is possible to recover microcolonies within the individual MDCC wells A sharp sterilized toothpick can be used to remove microorganisms from the compartments Microcapillaries or fragments of fibre optic cables or laser capture dissection microscopy may also be suitable The critical aspect of recovery is the imaging method we recommend dissection or transmission light microscopy using lenses with a working distance of about 1 cm Recovery of microcolonies stained by fluorogenic dyes is possible for example as part of HTS al though viability might be reduced depending on the toxicity of the applied dye In general it is advisable to use an automated X Y stage allowing the user to return to a specific position on the MDCC The use of a micromanipulator might additionally help when recovering a sample www microdish nl 5 MDCC User Manual and Protocols July 2012 9 Possible Procedures after Recovery The recovered material can be used for PCR amplification Such a proce
11. ty that the wells contain microcolonies derived from a single species This is an effective strategy when isolating previously uncultured organisms 5 MDCC Chemical Stability and Incubation Conditions Incubation can be performed at temperatures between 4 and 60 C and a pH between 3 to 9 Incubation at elevated temperatures up to 90 C is possible for a limited time max two weeks The MDCC is limited resistant to ethanol methanol dimethylformamide and DMSO but not to acetone www microdish nl 2 MDCC User Manual and Protocols July 2012 6 Staining Staining can be performed by introducing a fluorogenic dye from underneath the MDCC without disturbing the microcolonies on the chip surface as described in Section 7 Imaging Fluorescence Staining Protocol 7 Imaging Imaging is best performed using microscope objective lenses that do not touch the MDCC chip except using a cover slip and or immersion oil Depending on the magnification areas from a few mm to 10 um can be imaged in a single field of view Barrier Figure 2 Section of a MDCC160 chip on an agar plate using transmitted light microscopy A and B are compartments that contain microcolonies Microorganisms can overgrow individual wells as depicted in C which is not an issue at this point Targeted strain recovery is even eased in such a case Fluorescence Light Microscopy Intrinsic fluorescence e g GFP and fluorogenic strains can be used as the MDCC