ITS-90 ISOTHERMAL TOWERS
Contents
1. Do not drill modify or otherwise change the shape of the apparatus Do not dismantle the apparatus Do not use the apparatus outside its recommended temperature range If cased do not return the apparatus to its carrying case until the unit has cooled There are no user serviceable parts inside Contact your nearest Isotech agent for repair Ensure materials especially flammable materials are kept away from hot parts of the apparatus to prevent fire risk Page 6 of 22 ITS 90 Isothermal Towers 155 07 05 14 _ISOT CH DO S AND DON TS DON T handle the accessories when they are very hot or very cold DON T rely on the controller to tell you the temperature of the Isotower Its job is only to provide an isothermal volume It is the calibrated working standard that is used to measure actual temperature DON T operate the equipment without the ceramic cap in place Page 7 of 22 ITS 90 Isothermal Towers Iss 07 05 14 _ISOT CH UNPACKING AND INITIAL INSPECTION Our Packing Department uses custom designed packaging to send out your unit but as accidents can still happen in transit you are advised after unpacking the unit to inspect it for any sign of shipping damage and confirm that your delivery is in accordance with the packing note If you find any damage or that part of the delivery is missing you must notify us or our agent and the carrier immediately If the unit is damaged you should keep the packing f
2. The top zone controller setting should be checked each time the Isotower is used and adjusted if necessary See the next section on how to perform this Monitoring and Adjusting the Isotower Profile Switch on and set a temperature 2 C below the cell s melt freeze temperature Monitor the cell with a stable thermometer Allow enough time for your lsotower to stabilise When the Isotower temperature stabilises record the temperature then raise the monitor 00mm 4 and allow the monitor to stabilise and record the temperature again ideally the raised temperature should be 0 1 C higher than the bottom measurement This takes only 10 minutes and checks for any drift on the Immersion Compensation Controller If the measurements indicate that the top is colder the temperature of the zone controller can be raised by increasing the setpoint value Please note that the relationship between the zone controller temperature and the temperature at the top of the cell is not linear and an increase of 0 1 C in the zone controllers value will not increase the temperature measurement at the top of the cell by 0 1 C DO NOT set the top temperature more than 2 C above the melt There is an over 4 temperature trip inside the ISOTower preset to 4 C above melt to protect the heat siphon Should you trip the unit you will need to contact Isotech for further instructions Page 11 of 22 ITS 90 Isothermal Towers 155 07 05 14 _ISOT CH
3. ISOTHERMAL TECHNOLOGY LTD PINE GROVE SOUTHPORT PR9 9AG ENGLAND TEL 44 0 1704 543830 54461 FAX 44 0 1704 544799 The company is always willing to give technical advice and assistance where appropriate Equally because of the programme of continual development and improvement we reserve the right to amend or alter characteristics and design without prior notice This publication is for information only Page 4 of 22 ITS 90 Isothermal Towers 155 07 05 14 _ISOT CH A CAUTIONARY NOTE ISOTECH PRODUCTS ARE INTENDED FOR USE BY TECHNICALLY TRAINED AND COMPETENT PERSONNEL FAMILIAR WITH GOOD MEASUREMENT PRACTICES IT IS EXPECTED THAT PERSONNEL USING THIS EQUIPMENT WILL BE COMPETENT WITH THE MANAGEMENT OF APPARATUS WHICH MAY BE POWERED OR UNDER EXTREMES OF TEMPERATURE AND ARE ABLE TO APPRECIATE THE HAZARDS WHICH MAY BE ASSOCIATED WITH AND THE PRECAUTIONS TO BE TAKEN WITH SUCH EQUIPMENT Page 5 of 22 ITS 90 Isothermal Towers 155 07 05 14 _ISOT CH HEALTH AND SAFETY INSTRUCTIONS Read this entire manual before use Wear appropriate protective clothing Operators of this equipment should be adequately trained in the handling of hot and cold items and liquids Do not use the apparatus for jobs other than those for which it was designed i e the calibration of thermometers Do not handle the apparatus when it is hot or cold unless wearing the appropriate protective clothing and having the necessary training
4. ITS 90 ISOTHERMAL TOWERS User Maintenance Manual Handbook Isothermal Technology Limited Pine Grove Southport PR9 9AG England Tel 44 0 1704 543830 Fax 44 0 1704 544799 Internet www isotech co uk E mail info isotech co uk The company is always willing to give technical advice and assistance where appropriate Equally because of the programme of continual development and improvement we reserve the right to amend or alter characteristics and design without prior notice This publication is for information only Page of 22 ITS 90 Isothermal Towers 155 07 05 14 _ISOT CH CONTENTS EMGINFORMA TION ee en eo Eu att d aut p Dur See nee oer 3 ELECTRICAL pcr 3 Environmental EIE D LO D Noo 3 ibe Salety IMIOM MACON e mc 3 Dl ited Re aera eae ree ree E er eter 4 CAUTIOINARY NOTE tut tei cue NE o bnt e ee ML oe sea ta A 5 HEALTEAND INSTRUG HONS silia deitate eite ba dba one neat tea bdo ad bo eds 6 DO SAND DON TS sik 7 UINPACKINGANBD INITIAL INSPECTION M 8 EEC CIS jsp Ae 8 SuemdeU Scope qe 9 COMMISSIONING TOUR ISO TOW ER epo E a euius E ec eva ca T 12 TEE IMMERSION COMPENSATOR Nee e ERR Lene ees bo deste ties Rei te Eb I ERE 3 Changing the Eeneth othe Melt Plateatis ste
5. COMMISSIONING YOUR ISOTOWER Switch on and set a temperature 5 C below the cell s melt freeze temperature Allow one or two hours for your lsotower to stabilise Monitor the cell with a stable thermometer such as lsotech s 670SQ 25 SPRT and microK bridge Once the system is stable raise the monitor 00mm 4 after a few minutes it will restabilise the new temperature reading should be the same or slightly above the fully inserted temperature If not increase the top controller offset in small steps until it is It is worth spending some time getting this right If the monitor thermometer is already calibrated it can be used to calibrate the main controller Once this is done set the controller 1 C above the melt temperature and monitor the complete melt When the melt is complete increase the temperature a further C so that the cell is 2 C above its melt temperature and allow one hour for the liquid metal to fully melt and mix Next set the controller 1 C below the freeze point and use the monitor to follow the cells temperature as it drops and then enters the super cool zone Carefully remove the monitor and insert a cold rod into the re entrant tube for 2 minutes Repeat with a second cold rod before reinserting the cool monitor SPRT Record the freeze of your lsotower You are now ready to calibrate thermometers either on the cells melt or its freeze Page 12 of 22 ITS 90 Isothermal Towers 155 07 05 14 _ISOT CH _
6. THE IMMERSION COMPENSATOR The offset value is provided as part of the calibration certificate you could regularly check its value and adjust it as described above Ensure the ceramic cap is in place before use see Figure for details Figure Always ensure that the ceramic is in place before powering the lsotower ZION ISOTHERMAL TOWER ITS 90 FIXED POINT TIN Page 13 of 22 ITS 90 Isothermal Towers 155 07 05 14 _ISOT CH Changing the Length of the Melt Plateau By setting the controller 0 1 C above the melt temperature you will obtain a melt up to 30 hours 0 2 C will give you a typical 15 hour melt etc Getting a Freeze Isotowers can be used as a melt or a freeze point apparatus The melt is arrived at more quickly and using a 24 hour timer and allows a melt to be available all day every day The freeze curve is slightly flatter in temperature than the melt and some of you may wish to use the lsotower during its freeze Firstly melt the cell and set the controller 2 C above the melt DO NOT set the controller more than 2 C above the melt There is an over temperature trip inside the lsotower preset to 4 C above the melt to protect the heat siphon Should you trip the unit you will need to contact Isotech for further instructions INEVER change the trip temperature Leave the Isotower for an hour before resetting the controller temperature 0 1 to 0 2 C below the freeze Monitor the temperature i
7. the scroll key until the lower displays shows UNITS Use the UP and DOWN keys to select the desired display unit Using the zone controller The zone controller is set to read the temperature difference between the heat siphon temperature and the immersion compensator Typically if there was no difference in the temperature then the controller would read 0 If the immersion compensator was cold by 1 C then the zone controller would be set to C etc To adjust the controller temperature simply use the up down buttons Page 19 of 22 ITS 90 Isothermal Towers 155 07 05 14 _ISOT CH USING THE PC INTERFACE The bath includes an RS422 PC interface and a special converter cable that allows use with the a standard RS232 port When using the bath with an RS232 port it is essential that this converter cable is used Replacement cables are available from Isotech part number ISO 232 432 A further lead is available as an option Part Number ISO 422 422 lead which permits up to 5 instruments to be daisy chained together The benefit of this approach is that a number of calibration baths may be connected together in a daisy chain configuration and then linked to a single RS232 see diagram RS232 RS422 RS422 Note The RS422 standard specifies a maximum lead length of 1200M 4000ft A true RS422 port will be required to realise such lead lengths The Isotech conversion leads are suitable for maximum combined lead l
8. user serviceable parts inside Contact your nearest lsotech agent for repair Voltage transients on the supply must not exceed 2 5kV Conductive pollution e g Carbon dust must be excluded from the apparatus EN61010 pollution degree 2 Environmental Ratings Operating Temperature 0 50 C Relative Humidity 5 95 non condensing Fire Safety Information Use only a Class D Metal X type fire extinguisher or Dry Soda Ash Na2CO3 for extinguishing these fires Protective equipment such as eye protection jackets gloves and respiratory equipment is recommended Please refer to the MSDS for further information Page 3 of 22 ITS 90 Isothermal Towers 155 07 05 14 _ISOT CH GUARANTEE This instrument has been manufactured to exacting standards and is guaranteed for twelve months against electrical break down or mechanical failure caused through defective material or workmanship provided the failure is not the result of misuse In the event of failure covered by this guarantee the instrument must be returned carriage paid to the supplier for examination and will be replaced or repaired at our option FRAGILE CERAMIC AND OR GLASS PARTS ARE NOT COVERED BY THIS GUARANTEE INTERFERENCE WITH OR FAILURE TO PROPERLY MAINTAIN THIS INSTRUMENT MAY INVALIDATE THIS GUARANTEE RECOMMENDATION The life of your ISOTECH Instrument will be prolonged if regular maintenance and cleaning to remove general dust and debris is carried out ISUTZCH
9. IS cions NR 20 Jsme the CSA CS be pan buts 20 CAL NOTEPAD aay ae 21 ae aAa IEAA 21 HOW TO INSTALL CAL NOTEPAD 22 PROTOCOL cdit ee aS e d 22 Page 2 of 22 ITS 90 Isothermal Towers 155 07 05 14 _ISOT CH EMC INFORMATION This product meets the requirements of the European Directive on Electromagnetic Compatibility EMC 89 336 EEC as amended by EC Directive 92 3 and the European Low Voltage Directive 73 25 EEC amended by 93 68 To ensure emission compliance please ensure that any serial communications connecting leads are fully screened The product meets the susceptibility requirements of EN 50082 1 criterion B Symbol Identification Publication Description N 1503864 Caution refer to manual AN IEC 417 Caution Hot Surface N ELECTRICAL SAFETY This equipment must be correctly earthed This equipment is a Class Appliance A protective earth is used to ensure the conductive parts cannot become live in the event of a failure of the insulation The protective conductor of the flexible mains cable which is coloured green yellow MUST be connected to a suitable earth The blue conductor should be connected to Neutral and the Brown conductor to Live Line Warning Internal mains voltage hazard Do not remove the panels There are no
10. Model Indicator Address Indicator Value Frozen PV 9 72 0 00 0 00 DEVELOPMENT Cal NotePad was developed by Isothermal Technology using LabVIEW from National Instruments The license details are shown on the download page and in the Cal Notepad manual Page 21 of 22 ITS 90 Isothermal Towers 155 07 05 14 _ISOT CH _ HOW TO INSTALL CAL NOTEPAD Download the ZIP from http www isotech co uk downloads 7 6 2 Extract the files to a temporary folder 3 Run setup exe m Isotech CNP Setup Oe x Welcome to Isotech CNP Installation Wizard is strongly recommended that vau exit all Windows programs before running this setup program Click Cancel to quit the setup program then close any programs pou have running Click Next to continue the installation WARNING This program is protected by copyright law and International treaties Unauthorized reproduction or distribution of this program any portion of it may result in severe civil and criminal penalties and will be prosecuted to the maximum extent possible under law Cancel 4 Follow the prompts which will install the application a user manual with setup information and the necessary LabVIEW run time support files S Should you ever need to uninstall the software then use the Add Remove Programs option from the Control Panel PROTOCOL The instruments use the Modbus Protocol If required e g for writing custom software
11. ards there will not be space for the expansion and the graphite crucible will rupture See illustration below 19 002 molten metal molten metal jueipeJe eunje1ee J OH Melting from the bottom upwards blocks Keeping the top at or above the temperature at room for expansion and can rupture cell the bottom allow for expansion without pressure It is vitally important that the top of the cell is at the same temperature or slightly hotter than the rest of the cell to stop any pressure building up within the cell The most common cause for the top of the cell to be colder is due to thermometers being placed in the Isotower conducting heat out of the cell Test thermometers with larger thermal mass such as Metal Clad Thermometers or a continuous throughput of thermometers being calibrated will cool the top of the cell and freeze the metal To avoid this it is important to pre warm the thermometers being inserted close to or ideally above the cell temperature before inserting it into the lsotower Page 10 of 22 ITS 90 Isothermal Towers 155 07 05 14 _ISOT CH To balance this loss a heated collar is included above the cell called an Immersion Compensator this collar has its own controller zone and follows the main controllers temperature The main controller and the top zone controller are factory set prior to despatch The settings can be found on the melt and freeze graphs that accompany the equipment
12. beu Mo Ug bite a 14 a R 14 eT 14 CONCERN crer A EA E A E 15 LEARNING MORE ABOUT YOUR ISO TOWER eu oe rt oti e ve eoe eoa Ep tert cated ee awe T l6 Note How the thermal profile along the cell can affect or even destroy it eese l6 Putting n mbers to the theory 17 USING TEX ONTROLBEER dto ite diu to tato ttt ato dile aba aede aiat one totes a luba de outta tessuti 18 Front Panella yOu sisino deiode alaaa a a a a aa 18 The Temperature essit eee 18 Alterne the dote oia E E tease pest tema ese dotum esM S US ie te pau 18 Advanced Controller Features 25 eoi epus tetera 18 RANI abe TEE 18 Instr rtie rt Addres Saisso e seeded PEE Eten tete e ene tette 19 Monitoring the Controler 19 WIGS E H 19 ISIS zone Controller s ERR 19 USING THE PC INTERFACE ton 20 CO
13. cover after a cold thermometer is inserted This is one hour for an Aluminium Slim Cell less for In Sn and Zn and longer for Ag Heavier UUT s such as Metal Clad Thermometers will freeze more metal and so will require longer for the cell to recover Preheating the UUT will reduce this time considerably because less metal is refrozen and is recommended in all cases Page 17 of 22 ITS 90 Isothermal Towers 155 07 05 14 _ISOT CH USING THE CONTROLLER Front Panel Layout Upper Readout lt Lower Readout Page Down Button Button Scroll Up Button Button The Temperature Controller The controller has a dual display the upper display indicates the nominal block temperature and the lower display indicates the desired temperature or setpoint Altering the Setpoint To change the setpoint of the controller simply use the UP and DOWN keys to raise and lower the setpoint to the required value The lower display changes to indicate the new setpoint Advanced Controller Features Setpoint Ramp Rate By default the bath is configured to heat and cool as quickly as possible There may be some calibration applications where it is advantageous to limit the heating or cooling rate An example might be when testing bimetallic thermostats by forcing the bath to heat at a controlled rate it is easier to determine the temperature at which the thermostat changes state The bath can have its heating rate limite
14. d with the Setpoint Ramp Rate feature This feature is accessed from the Scroll key Depress the key until the display shows OFF SP RAT The upper display will show the current value and is adjustable from OFF to 999 9 The units are C min and are adjustable via the UP DOWN keys When the SP RAT is active the lower setpoint display will now automatically update with the current value known as the working setpoint The setpoint can be seen by pressing either the UP and DOWN key Page 18 of 22 ITS 90 Isothermal Towers 155 07 05 14 _ISOT CH The Setpoint ramp rate operates when the bath is heating and cooling Instrument Address The controller has a configurable address which is used for PC communications Each instrument has an address this allows several instruments to be connected in parallel on the same communications bus The default value is This address would only need to be changed if more than one Dry Block is connected to the same PC port To check the Address value press the scroll key until the lower display indicates ADDR The upper display will show the current value that can be modified with the UP and DOWN keys Monitoring the Controller Status A row of beacons indicate the controllers status as follows Output opl Heater shows when the heater is active seen to pulse to display heater output REM This beacon indicates activity on the PC interface Units The units can be changed from C Press
15. engths of 10M that is adequate for most applications Connections For RS232 use simply connect the Isotech cable a 9 to 25 pin converter is included to suit PCs with a 25 pin serial converter RS422 Connections Pin Connection 4 5 8 9 Common Using the Interface The models are supplied with Cal NotePad as standard This easy to use package is compatible with MS Windows 9x A manual for Cal NotePad can be found on the first installation disk in Adobe PDF format If required a free Adobe PDF reader can be downloaded from www adobe com Page 20 of 22 ITS 90 Isothermal Towers 155 07 05 14 CAL NOTEPAD Cal Notepad can be used can be used to log and display values from the Dry Blocks and an optional temperature indicator such as the milliK or 10 The software requires Windows 9X XP a minimum of 5Mb of free hard drive space and free serial ports for the instruments to be connected LB Isotech CNP 2006 File Help Instrument Heat Source Chart Log ISUTZCH 1 700 0007 Setpoint Programming Active NEN Scroll Minutes Temperature A 1 20 00 4 600 000 500 000 400 000 0 30 00 210 00 300 000 4 10 00 200 000 100 000 Running 0 000 P Test Comms Measurements Start gm Source COM Port Controller Address Controller Value Controller Setpoint Z PLUS EiBISynch 4 an al J 0 00 0 00 J J Contacts Stop i SITE
16. evaporation and condensation creating a thermal conductance up to 200 times better than copper The ITS 90 cell inside the heat siphon therefore warms quickly and uniformly Due to thermal losses a thermometer under test will conduct heat out of the cell causing a thermal gradient To balance this loss a heated collar is included above the cell called an Immersion Compensator this collar has its own controller and follows the main controllers temperature The main controller and the top controller are factory set prior to despatch The settings can be found on the melt and freeze that accompanies the lsotower Safe Operation of an Isotower Always ensure that the ceramic cap is in place before connecting the power supply to the Isotower DON T operate the equipment without the ceramic cap in place To Prevent Irreparable Damage the top of the thermometer pocket MUST not be allowed to become cooler than the bottom of the thermometer pocket Inside the Isotower is a graphite crucible that contains metal Indium Tin Zinc or Aluminium These metals expand as they melt and contract as they freeze The crucible has sufficient volume to contain the metal when it is fully molten As the metal freezes it contracts therefore occupying a smaller area of the crucible Page 9 of 22 ITS 90 Isothermal Towers 155 07 05 14 If the top of the cell is colder than the rest of the cell then on melting then as the metal expands from the bottom upw
17. he heat siphon A user would notice that the cell was not melting properly An Aluminium lsotower heat siphon contains sodium under a partial vacuum A leak would allow a small amount of air to enter the heat siphon A user would notice that the cell was not melting properly In the unlikely event of a leak the user should contact lsotech The other remote possibility is that the heat siphon leaks into the cell while the cells metal is molten To Indium and Tin no reaction takes place with the water vapour With Zinc and Potassium neither is affected by the other Ref Hansen With Aluminium and Sodium neither is affected by the other Ref Hansen Again the cell would melt badly In this unlikely event please contact Isotech The lsotower wants to please you but to get the best from it you need to understand the controller temperature settings and the top heater offsets Page 15 of 22 ITS 90 Isothermal Towers 155 07 05 14 _ISOT CH LEARNING MORE ABOUT YOUR ISOTOWER Try adjusting the controller temperature between melts 0 1 0 2 C See the melts get longer and shorter Select the setting that gives you the length of plateau that suits you like about 15 hours so that can have a new plateau each day 2 Check your furnace profile each time you use your Isotower After the lsotower settles down 2 C below its melt raise the monitor 100mm 4 and check the temperature is the same or increases slightly This
18. n the cell and when it dips below the freeze temperature remove the monitor and in succession introduce two cold rods each for two minutes Replace the monitor which should quickly achieve the freeze temperature Calibration of the Units Under Test UUT can commence each UUT will freeze some metal and so shorten the freeze time and so if a number of thermometers are to be calibrated they should be pre warmed Certification The unique feature of the lsotower is that it is complete A certificate such as a UKAS certificate includes the uncertainties of not just the cell but also the apparatus and immersion characteristics of the UUT This reduces the overall uncertainties of your laboratory and simplifies the task of calculating uncertainties which in turn improves confidence in your measurements Page 14 of 22 ITS 90 Isothermal Towers 155 07 05 14 _ISOT CH A BASIC CONCERN Heat siphons contain fluids There is a basic concern that in the very unlikely event of a leak that there is a health and Safety issue Dealing with each fixed point in turn Both Indium and Tin use a heat siphon containing water as the fluid at either temperature the water is pressurised i e it is above bar If a leak occurred some moisture may leak out into the insulation A user would notice that the cell was not melting properly The Zinc Isotower heat siphon contains potassium under a partial vacuum A leak would allow a certain amount of air to enter t
19. or possible insurance assessment Electricity Supply Before connecting to the electricity supply please familiarise yourself with the parts of the manual relevant to your model Your unit s supply voltage requirement is specified on a plate on the instrument along with the serial number All instruments will work on an electricity supply frequency of 50Hz 60Hz The apparatus is provided with an approved power cord If the plug is not suitable for your location then the plug should be removed and replaced with an appropriate plug A Take care to ensure the old plug is disposed safely The cable is colour coded as follows COLOUR FUNCTION Green yellow Earth Ground Brown Live line Blue Neutral Please ensure that your unit is correctly connected to the electricity supply THE APPARATUS MUST BE CORRECTLY EARTHED GROUNDED The units on off switch is located on the power inlet Take care NOT to switch the unit off when it is hot allow to cool first Page 8 of 22 ITS 90 Isothermal Towers Iss 07 05 14 _ISOT CH USING YOUR ISOTOWER Congratulations you have just purchased the most advanced fixed point calibration cell and apparatus ever invented Before you power up your lsotower it is worth spending a little time to get to understand its operation Isotowers work by using the amazing thermal properties of heat siphons Once the fluid inside a heat siphon reaches its boiling temperature it transfers heat by
20. takes only 10 minutes and checks for any drift on the Immersion Compensation Controller 3 When calibrating leave the Unit Under Test UUT in the Ilsotower long enough for the system to fully restabilise see notes below We recommend pre warming the UUT in all cases Note How the thermal profile along the cell can affect or even destroy it The easiest cell to describe is the Water Triple Point because the container being made of glass is transparent so the effects can be seen As cold rods are inserted into the cell to create an ice mantle an ice bridge forms at the surface of the water in the cell If this is not immediately re melted then further cooling and ice formation expands the volume beneath the ice bridge and the cell will implode This is because ice is 8 less dense than water for example in Isotech 1 50 cell the water level increases 2mm as the ice increases from O to 5096 Although we cannot see inside other fixed point cells during melting a similar process is occurring For example the metal inside a Slim Aluminium Cell will expand 7 296 and the level of aluminium will increase by 2 6mm as the cell melts The cells are heated in furnaces which will be either 3 Zone or Heatpipe furnaces It should be obvious from the above that the cell should either have zero temperature gradient along its length or the top of the cell should be hotter than the rest of the cell If the cell melts from the bo
21. the technical details are available from our Document Library at http www isotech co uk Page 22 of 22 ITS 90 Isothermal Towers 155 07 05 14
22. ttom upwards the expanding liquid metal will rupture the graphite crucible Page 16 of 22 ITS 90 Isothermal Towers 155 07 05 14 _ISOT CH Note Putting numbers to the theory Traditionally the fixed points of ITS 90 have been used on their freezes This is because during freezing the metals shrink In by 2 3 Sn by 2 5 Zn by 4 6 Al by 7 2 and Silver by 4 5 During melting pressure can build up inside the cell as it melts and expands This is why the freeze is chosen In practice it is very convenient to use the melt plateau for calibrating thermometers and we along with other producers recommend it in all but the most exacting requirements However it is important to pre warm the UUT before inserting it into the melting cell to avoid refreezing it This note explores the worst case situation of a UUT being inserted into a melting cell from room temperature For Isotech model 670 thermometer the end 18cm weighs 10 5g The 670 will freeze 5 4cc s of In 4 4cc s of Sn 5 4cc s of Zn 7cc s of Al and 1 2cc s of Ag There are approximately 67cc s of metal in an Isotech Slim Cell and 200cc s an Optimal Cell A 670 will refreeze the following percentages of metal in a Slim Cell 7 8 of In 6 6 of Sn 7 8 of Zn 1296 of Al and 17 of Ag By setting the Slim Cell s apparatus a melt of eight hours can be achieved This means 12 of the metal is melted per hour So now we can calculate the time for the melting cell to re