Instructions for manually
Contents
1. Press the SET button to save setting Use the arrow buttons to select another programming mode or press MODE to exit the programming mode 5 42280 V4 0 10 2007 Programming Mode P2 Datalogger Start Modes PON gt Press the SET button to access the setup mode Use the arrow buttons to select parameter P2 Press the SET button again Select one of the four Start modes using the up down arrow buttons see list and diagram below a rEp Repeat Start mode Logger starts once readings are downloaded to PC hEy Keypad Start Datalogging begins after the START STOP button is held for 2 seconds imm Immediate Start Recording starts immediately after exiting the setup mode SCh Scheduled Start Logging begins at a preset date and time Press the SET button to save the setting Use the arrow buttons to select another programming mode or press MODE to exit the programming mode Stn Pe SEn rE p Sta hE y Stn SCx 42280 V4 0 10 2007 Programming Mode P3 Datalog Start Time Note that parameter P3 will only be visible if the Start MODE is set to SHCEDULE Parameter P2 set to SCh 1 10 11 12 13 14 15 16 17 Press the SET button to access the setup mode Use the arrow buttons to select parameter P3 Press the SET button again Use the up down arrow buttons to set the year Press the SET button again Use the
2. 2 Press and hold the MODE for at least 2 seconds This will change the top display to Dew Point temperature The DP icon will appear on the LCD when in Dew Point mode Press and hold the MODE button for at least 2 seconds to return the top display to the normal temperature display mode 3 The middle line displays Relative Humidity in The bottom line automatically alternates between Time and Date displays 5 Ifthe ALM status LED illuminates and the display flashes one of the Alarm limits has been reached Program the High and Low Temperature Humidity limits in Programming parameter P5 detailed later in this user guide Datalogger Basics The Model 42280 has a built in Datalogger that can store up to 8000 temperature and 8000 humidity readings Readings can be stored in the meter and transferred at a later time to a PC The meter can also display real time readings on a connected PC in other words the PC can display readings at the instant the readings are taken The meter must be configured before datalogging can be attempted Parameters such as Real Time Clock and Datalogger Sample Rate must be programmed before datalogging can begin It is strongly recommended that the user take advantage of the supplied software to configure the meter Instructions for manually configuring the meter are provided below however configuring the meter via software is very convenient Basic instructions for connecting the meter to a PC and ge
3. s HELP utility provides in depth software and downloading instructions which is beyond the scope of this user guide please refer to the HELP utility in the supplied software program for detailed guidelines and instructions 11 42280 V4 0 10 2007 Humidity Calibration 1 Turn on the meter and plug the humidity sensor probe into the 33 Calibration Reference bottle Simultaneously press and hold the START STOP and the SET buttons for at least 2 seconds to enter the 33 Calibration mode The display 32 8 will begin to flash on the display Approximately 30 minutes later the 32 8 display will stop flashing indicating that the 33 calibration is complete Plug the humidity sensor probe into the 75 Calibration Reference bottle Press and hold the START STOP button for at least 2 seconds to enter the 75 Calibration mode The display 75 3 will begin to flash on the display Approximately 30 minutes later the display will stop flashing indicating that the entire calibration is complete The meter will automatically return to the normal operating mode CAL cA 328 n53 Calibration Notes e The user can exit the calibration without finishing by pressing the START STOP button for 2 seconds before reaching step 6 above For best accuracy perform the calibration at room temperature e Calibration Reference bottles are available optionally through Extech Instruments distributors
4. Datalogging is indicated by the flashing REC light as described in the other Start Modes above To stop recording press and hold the START STOP button for at least 2 seconds The REC status light will stop blinking 10 42280 V4 0 10 2007 Connecting the Meter to a PC A USB cable is provided with the meter One end of the cable connects to the meter s USB jack and the other end connects to a USB port on the PC Running the supplied PC software The supplied PC software is included on the supplied CD ROM Insert the CD into the computer s CD ROM drive and follow the on screen prompts and instructions In the software COM PORT SETTING field select the COM port to where the meter is connected Software Datalogging Parameters From the main software screen select LOGGER SET and set the sample rate sample points and start mode These are described in some detail in the separate software manual and the software HELP utility on the supplied CD ROM Remember to press OK after changing a setting to ensure that the software processes the changes The meter is now ready to record Transferring readings to PC Connect and run the supplied software as previously described 2 Press the DOWNOAD START button on the main software screen to transfer the readings 3 The separate software instructions supplied in the software s HELP utility explain how to view graph save export and print the downloaded data NOTE The Software
5. 12 42280 V4 0 10 2007 Specifications General Specifications Display Measurement ranges Resolution Accuracy Datalogger memory Over range indication Low battery indication Power supply Multi function LCD with status LED indicators Temperature 4 to 144 F 20 to 70 C Humidity 0 to 100 Dew Point temperature 4 to 144 F 20 to 70 C 0 1 C F Temp 1 0 F 0 6 C from 32 to 122 F 0 to 50 C 2 0 F 1 2 C outside of this range Humidity 3 from 10 to 90 5 outside of this range 16 000 total 8000 temperature and 8000 humidity readings Error code appears on the LCD see Troubleshooting section below for list of codes Battery symbol appears on the LCD Four 4 1 5v AA batteries 13 42280 V4 0 10 2007 Troubleshooting Power is ON but there s no display Check that the batteries or the AC adaptor is installed and connected properly Ensure that the START STOP button is pressed firmly and for at least 100ms Calibration Failure Check that the batteries or the AC adaptor is installed and connected properly Ensure that the calibration bottles are tightly sealed around the meter sensor Ensure that calibration takes place at room temperature Flashing LED Indicators REC flashes when datalogging takes place ALM light flashes when alarm limit has been exceeded Error Codes E02 Under flow error E03 Over flow error E04 E02 and E03 combined error E11 CAL error
6. the SET button again oO Use the up down arrow buttons to set the hours i Press the SET button again N Use the up down arrow buttons to set the minutes wo Press the SET button again x A Use the up down arrow buttons to set the seconds oa Press SET again to stores the settings e gt Press the MODE button to back out to the top level of this menu N Use the arrow buttons to select another programming mode or press MODE to exit the programming mode 9 42280 V4 0 10 2007 Datalogging Starting a Datalogging session The 42280 can automatically store up to 16 000 readings 8000 temperature and 8000 humidity for later transfer to PC Program the controller as discussed earlier before Datalogging If the Datalogger has not been configured at this point configure as necessary If programming is complete continue with the next section Starting the Datalogger by push button Key Mode 1 on Programming mode P2 Datalogging Start Modes must be set to KEY hEy mode for this method When P2 is set to hEy press and hold the START STOP button for 2 seconds The REC front panel status light will blink indicating that readings are currently recording at the rate programmed in parameter P4 sample rate To stop recording press and hold the START STOP button for at least 2 seconds The REC status light will stop blinking Starting the Datalogger in
7. Department at 781 890 7440 ext 210 for authorization or visit our website www extech com for contact information A Return Authorization RA number must be issued before any product is returned to Extech The sender is responsible for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenance or repair or unauthorized modification Extech specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct indirect incidental or consequential damages Extech s total liability is limited to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is expressed or implied 15 42280 V4 0 10 2007
8. Humidity Sensor internal Temperature sensor Audible alarm with AC adaptor power only Visual Alarm LED Relative Humidity display Date and Time display Temperature display RECORD status LED USB port AC Adaptor jack Datalogger START STOP button MODE button SET button Up Arrow button Down Arrow button Tripod mount jack 0 7A o YM 07 06 21 Temperature amp Humidity Datalogger 42280 C 0 15 10 11 12 13 14 Notes The battery compartment is located on rear of instrument Desktop stand footers are not shown in diagram LCD Display Description 1 oa fF wo DN Temperature Display Dew Point icon Datalogger icon Date and Time Humidity Display Low battery icon 3 AOSE 08 8 8 4s 8888 88 3 42280 V4 0 10 2007 Operation Important Note before starting This meter relies on firm button presses to operate correctly Be sure to press firmly and hold the button for as long as directed in the instructions Meter Power The meter is powered by four 4 AA 1 5V batteries or the 117V AC adaptor 2 When the batteries are installed or the adaptor is connected to an AC power source momentarily press the START STOP button to turn the meter ON 3 Press the START STOP button again to turn the meter OFF Viewing the displayed readings 1 Normally the top line indicates temperature in degrees C or F Use the MODE button to switch between degrees C and degrees F
9. Immediate Mode 1 Programming mode P2 Datalogging Start Modes must be set to Imm mode for this method When the programming mode is exited Datalogger automatically begins The REC front panel status light will blink indicating that readings are currently recording To stop recording press and hold the START STOP button for at least 2 seconds The REC status light will stop blinking Starting the Datalogger in Schedule Mode 1 In Schedule Mode the meter begins recording at a specific date and time as programmed by the user Programming mode P2 Datalogging Start Modes must be set to SCh mode for this method After setting P2 to SCh set the desired time and date in Programming Mode P3 P3 only appears in the menu when SCh is selected in P2 Datalogging will start at the programmed date and time Note that the Real Time Clock Programming Mode P6 must first be set to the current date and time When datalogging begins the REC front panel status light will blink indicating that readings are currently recording at the rate programmed in parameter P4 sample rate To stop recording press and hold the START STOP button for at least 2 seconds The REC status light will stop blinking Starting the Datalogger in Repeat Mode In Repeat Mode the meter simply begins datalogging again after readings are downloaded to a PC This allows the user to maintain the same programming selections time after time
10. User Guide EXTECH INSTRUMENTS Temperature and Humidity Datalogger Model 42280 16 be ir aE 1 ye 05 21 Temperature amp Humidity Datalogger 42280 START STOP Introduction Congratulations on your purchase of the Extech 42280 Thermometer and Relative Humidity Datalogger The 42280 is a wall mount tripod mount or desktop indoor air quality monitor with an internal memory capable of storing 16 000 readings 8000 temperature readings and 8000 relative humidity readings that can later be transferred to PC using the USB interface Real time meter to PC logging is also supported This meter is shipped fully tested and calibrated and with proper use will provide years of reliable service Material list Supplied Materials Meter User Guide USB cable Program CD Four 4 batteries 1 5V AA Optional accessories 9V adaptor Calibration bottles 33 and 75 for Relative Humidity Calibration Features Large LCD display shows temperature humidity and date time clock simultaneously 16 000 internal memory capacity 8000 temperature 8000 RH Visual and Audible with 117V AC adaptor only Alarms Easily configure datalogger through the keypad or via PC interface USB PC interface with Windows software Wall mount tripod mount or for desktop use 2 42280 V4 0 10 2007 Description Meter Description 1 oon oa PF Oo ND as 2 11 12 13 14 15
11. er programming mode or press MODE to exit the programming mode Programming Mode P5 High and Low Alarms 1 Press the SET button to access the setup mode 2 Use the arrow buttons to select parameter P5 AL PS 3 Press the SET button again t H 4 Use the up down arrow buttons Pad 8 g to set the High Temperature 7 Alarm limit H Lo Press the SET button again 9 Co 8 Use the up down arrow buttons to set the Low Temperature Alarm limit 7 Press the SET button again 8 Use the up down arrow buttons to set the High Humidity Limit 9 Press the SET button again 10 Use the up down arrow buttons to set the Low Humidity Limit 11 Press the SET button again to store the settings 12 Press the MODE button to back out to the top level of this menu 13 Use the arrow buttons to select another programming mode or press MODE to exit the programming mode Note The AC Adaptor must be used for the audible alarm to sound 8 42280 V4 0 10 2007 Programming Mode P6 Real Time Clock rec rec n PS II NA ums d 1 08 25 Press the SET button to access the setup mode Use the arrow buttons to select parameter P6 Press the SET button again Use the up down arrow buttons to set the year Press the SET button again Use the up down arrow buttons to set month Press the SET button again 00 DO JOY a O Use the up down arrow buttons to set the day Press
12. recalibration required E32 IC read write error Return unit for repair E33 Circuit Error Return unit for repair 14 42280 V4 0 10 2007 Calibration and Repair Services Extech offers repair and calibration services for the products we sell Extech also provides NIST certification for most products Call the Customer Care Department for information on calibration services available for this product Extech recommends that annual calibrations be performed to verify meter performance and accuracy Saw amp Support line 781 890 7440 Technical Support Extension 200 E mail support extech com Repair amp Returns Extension 210 E mail repair extech com Product specifications subject to change without notice For the latest version of this User Guide Software updates and other up to the minute product information visit our website www extech com Extech Instruments Corporation 285 Bear Hill Road Waltham MA 02451 Copyright 2007 Extech Instruments Corporation All rights reserved including the right of reproduction in whole or in part in any form Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment a six month limited warranty applies to sensors and cables If it should become necessary to return the instrument for service during or beyond the warranty period contact the Customer Service
13. tting started with the supplied software are provided later in this guide Detailed instructions regarding the supplied software are provided in a separate manual included on the supplied CD ROM and the HELP utility in the software program 4 42280 V4 0 10 2007 Manual Programming The Model 42280 has several programming modes P1 P2 P3 P4 P5 and P6 These parameters must be programmed before the Datalogger or the Alarm can be used List of Programming Modes P1 Number of samples to record 1 2 4 8 12 or 16 x 1000 1 1000 2 2000 etc P2 Datalogging Start Methods Immediate Keypad Scheduled and Repeat modes P3 Start Time set the time and date when Datalogging will automatically begin Note that P3 will not appear in the menu if SCh is not selected in P2 P4 Sample Rate set the datalogging interval from 1 second up to 12 hours P5 Alarm Mode set High and Low Temperature Humidity limits P6 Real Time Clock set the meter s internal date and time clock Programming Mode P1 Number of readings sample points to record a O P Press the SET button to access the setup mode Use the arrow buttons to select parameter P1 see sample display below Press the SET button again Use the up down arrow buttons to pL 5 pL 5 select the desired number of samples p 1 5 in thousands to record for a l h datalogging session 1 2 4 8 12 or 16 where 1 1000 2 2000 4 4000 and so on
14. up down arrow buttons to set month Press the SET button again Use the up down arrow buttons to set the day Press the SET button again Use up down buttons to set the hours Press the SET button again Use up down buttons to set the minutes Press the SET button again Ste St P3 YE 010208 Ste Non Y M D 0702 08 Ske df B08 St Kor nms i 0825 St Nia ums f 1 08 26 Ste SEc ums f 1 08 25 Use the up down arrow buttons to set the seconds Press the SET button again to store the settings Press the MODE button to back out to the top level of this menu Use the arrow buttons to select another programming mode or press MODE to exit the programming mode 42280 V4 0 10 2007 Programming Mode P4 Datalogger Sample Rate 1 Press the SET button to access the setup mode 2 Use the arrow buttons to select parameter Se l Sr 3 Press the SET button again p 4 Hf Or 4 Use the up down arrow buttons to set hours ius 0826 5 Press the SET button again 6 Use the up down arrow buttons to set the Sr Sr minutes Press the SET button again E 1n S E E aus 480825 nms f 08 26 Use the up down arrow buttons to set the seconds 9 Press SET again to store the settings 10 Press the MODE button to back out to the top level of this menu 11 Use the arrow buttons to select anoth