        UNICORN version 2.01 - GE Healthcare Life Sciences
         Contents
1.      Also  when running a method containing too many calls to blocks set to Restore mode  the run stops  after a while and the message  Error in method  system stopped  appears  The number of calls  allowed varies with the method used     Action  Do not use Restore mode in blocks   The parameter Restore was removed in V2 20     SYSTEM CONTROL    Messages from printers    If a message from the printer appears at the end of a scouting run in which the result should be  printed automatically  the method might stop or the computer may lock up         56 3029 23    Edition AD  19 May 1998  56 3029 23  Edition AD  UNICORN vers  2 01  Known bugs and advice on operation    Amersham  e Biosciences    Action  To avoid the most common messages  make sure that your printer is ready and has enough  paper loaded     Another possible action  Do not allow any notify messages from the printer  Error messages can  usually not be turned off  How to turn off notify messages depends on the printer  Refer to the printer  User Manual     For network printers  turn off notify messages by first double clicking on the Novell icon on the OS 2  desktop  then on NetWare Tools  Select Tools Printer ports and double click on the printer in  the list  De select  Notify when job is printed   Click on Capture  This change will not automatically  apply the next time the computer is switched on  To make the change permanently  please contact  your network administrator  How to do it depends on the login routine
2.   File Report    If a report format is selected in File Report and more than one chromatogram is available on the  evaluation desktop  only the first one can be selected in the list of chromatograms     Action  Select Preview  then Exit  and Exit again in the Generate report box  When File Report is  entered again  all chromatograms can be selected     DEFAULT report format    If a DEFAULT report format has been defined and File Report is selected  chromatogram layout  settings will be same as on screen and not as specified in the DEFAULT report format     Action  If you want to use the DEFAULT report format chromatogram layout  you have to manually  select the DEFAULT format from the list of available formats     Column volume as X axis base    Column volume can not be selected as x axis base     Integration    Peak labels and start end markers shown in the curve window are always calculated from the latest  integration  If an earlier calculated baseline or peaktable is selected  the peak information in the curve  window is not updated  Displayed peak labels and markers and the peaktable will then not correspond     Action  Always work with the latest calculated baseline and peaktable     Edit peak table    If an integration has been performed and a new baseline is then calculated but without performing a  new integration  the peak table can not be edited     Action  Perform a new integration     Integration with a non calculated baseline    If a non calculated baseline  e g
3.   a curve from a blank run  is used as baseline in an integration  peak  labels will not be shown in the curve window     Action  Use a calculated baseline     56 3029 23  Edition AD  UNICORN vers  2 01  Known bugs and advice on operation 5 7     3  User Manual    START UP    Login on a local computer in a network    Page 3 1  When UNICORN is installed in a network and you want to control a system from the local  computer without logging in to the network  press Enter when the startup cmd window appears  Do  not just minimize or close the startup cmd window     METHOD EDITOR    Report formats in procedures    Page 5 36  Report formats in evaluation procedures which are saved in methods can not be edited or  viewed  A report format in a method will not be updated when a report format with the same name is  changed in the Evaluation  To edit the report formats  define a new report format in the Evaluation  and save it in a procedure  Import this new procedure into the method     SYSTEM CONTROL    Leaving and locking the system    Page 6 17  It is not possible to leave or lock a system when sequence or scouting runs are performed     EVALUATION    Printer setup    Page 9 38  The printer selected in File Printer setup is only valid for print outs from the evaluation  module  Printing from other parts of UNICORN is performed via the default OS 2 printer  Make sure  that a correct default OS 2 printer is selected on the OS 2 desktop     Report    Page 9 39  If  Current chromatogram
4.   is selected  the layout in the printout is not automatically the  same as on the screen  A layout for the printout must always be selected by clicking on the  Chromatogram Layout button in the File Report box     Changing x axis base  Page 10 2  After an integration  the x axis base can not be changed  If the result after integration is  not saved  it is possible to change base when the result file is opened again     Edit baseline and Edit Peak table    Page 10 13 and 10 14  Only the latest calculated baseline and peak table can be edited with Edit  Baseline and Edit Peaktable respectively     Evaluation procedures    Page 10 28  Personal procedures are stored in the home directory of each user     INSTALLATION    Pipe server name    Page 12 6  Two computers in a network must not be given the same pipe server name  Pay special  attention to this if you change the computer to your system     56 3029 23  Edition AD  UNICORN vers  2 01  Known bugs and advice on operation 6 7     3 or 4 systems connected to one computer    Page 13 8  If 3 or 4 systems are connected to a computer  either physically or through a network   the  line THREADS 256 in the Config sys file must have been changed to THREADS 512     Chapter 11  12  and 13    At the bottom of each page in chapters 11  12  and 13 it is stated the Edition is 0  That is wrong  The  edition is AA  the same as for all other chapters     At least 30 Mb free space    Page A 1  At lease 30 Mb free space should be available on the
5.  hard disk before each run  The swap  file requires this         Amersham  e Biosciences    56 3029 23  Edition AD  UNICORN vers  2 01  Known bugs and advice on operation 7 7     
6. Peak Label in the Chromatogram Layout box  the text is still  on  some printers  printed horizontal  Furthermore  in File Report Preview the text is not shown at all     Action  No action possible     56 3029 23  Edition AD  UNICORN vers  2 01  Known bugs and advice on operation 3 7     2  Advice on operation    START UP and SHUTDOWN    Login and Screen Saver    If a Screen Saver appears before you have had time to log in when UNICORN has been started  the  login box disappears     Action  Reboot the computer  Always login immediately when the login box appears     Shutdown    If UNICORN is minimised and Shutdown is selected in OS 2  a message asking to Quit UNICORN  appears  When entering UNICORN  the System control window is already closed down     Action  Never select Shutdown in OS 2 when UNICORN is running     METHOD EDITOR    Too many Watch instructions    If a Watch instruction calls a block  e g  named  Peak   which contains a new Watch instruction  that also calls the block  Peak   the message  Too many Watches  appears in the text instruction  window and a very large number of blocks will be shown     Action  The method can still be saved and run     Template    Empty     only for BioProcess System     When the template    Empty    has been selected for a BioProcess System  the unit for the column  volume in the Base instruction in the text window is incorrectly shown as ml  The real unit is liter     Action  Mark the Base instruction and click on Replace in the ins
7. UNICORN   version 2 01 Known bugs and  advice on operation    The list below contains the known and verified bugs in UNICORN version 2 01  Some advice on  operations is also given and some information missing in the User Manual is listed     1  Known bugs    START UP and SHUTDOWN    Untitled window    On some Compaq computers  OS 2 will not start up properly  An Untitled window appears and the  computer locks     Action  Install the latest available display driver for your computer  e g   Compaq CL5434 Display  Driver Support for OS 2 2 1X and 3 0  version 1 23 or  Compagq QVision 1024 1280 Display Driver  Support for OS 2 2 1X and 3 0  version 2 30     MAIN MENU    System run hours    The accumulated run time for a system  also called System run hours  that is displayed when you press  the Run hours button in the Administration Audit trail dialog may show an incorrect value  such  as  47000 hours      Action  No action possible   Corrected in V2 10     Directory names    If you create a user home directory called METHOD from within UNICORN and then try to create a  Sequence  the computer will lock up     Action  Do not use the name METHOD for a home directory   Corrected in V2 20     METHOD EDITOR    Blocks in Restore mode    Calling blocks which are set to Restore mode  instead of Normal mode   may cause malfunction of  the Wash  SyncPump or StepValve instructions in the called block  Even if they appear in the  logbook as executed  the system does not perform these actions
8. le will crash as soon as the  Documentation is opened    The cause is that the calibration file for the system is corrupt  If this happens  it is necessary to delete  the calibration file   otherwise  all subsequent result files will be corrupt too    Once the calibration file is deleted  see Action  below  it is either replaced by the default file delivered  with the strategy if such a file exists  or a new file is created  However  all calibration values and  system settings are lost and a new calibration needs to be done afterwards  To make this easier  print  calibration data and system settings values in the Documention of a valid result file    Action  Corrupt result files can be viewed in UNICORN V2 30    Perform following steps to delete the calibration file     o Quit UNICORN     56 3029 23  Edition AD  UNICORN vers  2 01  Known bugs and advice on operation 2 7     e Delete the file C  UNICORN LOCAL STRATEGY S YSTEMa strategyname SET  where n is  0    zero  for the first installed system   1  for the second  etc  and strategyname is the name of the  strategy     e Shutdown and reboot the computer     Preview in File Report    When Preview has been selected in File Report and you accidentally click outside the Preview  window  the window becomes hidden  but still open  behind the UNICORN desktop  This can cause  the computer to lock up at later stages     Action  Do not click outside the Preview window when open     Peak labels    If Horizontal text is de selected in 
9. s in your network     Calling Pause from a Watch instruction    If Pause is used in a Watch instruction and the Pause event is triggered  the system goes in Pause  state as expected but about 2 minutes later the method is erronously continued     Action  This error can be avoided if a Call to a block  that contains a Pause instruction  is made  instead of directly executing Pause     Column as scouting variable    If Column in the Base instruction is defined as a scouting variable and a new column is selected for  some of the runs after the method has been started  the column volume may be set to zero for the run  with the new column and that run will finish immediately     Action  Do not change columns during runs  Select all columns to be included in the scouting scheme  in the method   Corrected in V2 10     Run notes  If the Run notes are open at the end of the run the text may not be saved in the result file     Action  Do not let the Run notes be open at the end of runs   Corrected in V2 10     Fraction marks   only for BioPilot System and FPLC System     Some fraction marks are occasionally omitted when the fraction collector has been started with the  instruction Frac Start     Action  For version 2 0 of a standard strategy  use instruction Collect instead     EVALUATION    Corrupt result files    It might happen that a result file becomes corrupt in such a way that the Documentation information  cannot be opened or printed out  In some cases the whole Evaluation modu
10. truction dialog box to change the  unit to liter     Evaluation Procedures as variables    Defining a variable Procedure parameter in an Evaluate instruction can give method compilation  error when you try to run the method     Action  Do not define variable Procedure parameters in any Evaluate instructions     Collect    If the Collect instruction is set for a second time in a method after the CollectEnd instruction   fractions will sometimes not be collected     Action  Change to EPROM version 1 13 in the control unit to avoid this problem     Pressure limit    When the pressure limit set in UNICORN is exceeded  it may take about a second before a run is  paused     Action  Set the pressure limit on the pumps as well for faster response to pressure changes     Superloop    only for Multipurpose strategies     If the flow rate is changed during execution of the instruction Superloop  the time for filling or  emptying Superloop is not adjusted to the new flow rate     Action  Do not change the flow rate during filling or emptying of Superloop     56 3029 23  Edition AD  UNICORN vers  2 01  Known bugs and advice on operation 4 7     EVALUATION    DEFAULT layout    When a result file is opened and a DEFAULT chromatogram layout has been defined  all curves will  initially be shown even if only a few of them were defined in the DEFAULT layout     Action  Open the Chromatogram layout box  select the DEFAULT layout and click on OK  Only the  pre defined curves will then be shown   
    