HI 839800 COD REACTOR and Test Tube Heater 2008 Series

image

Contents

1. A Reactor Operation The Reactor incubates glass vials containing hazardous chemicals at high temperature Read the Material Safety Data Sheet MSDS before handling reagent vials Handling Reagent Vials Handle the vials very carefully and wear protective clothing After use dispose of the reagent vials according to local regulations Safety Equipment When operating with Reactor use protective clothing including goggles or face mask and gloves Use of the optional safety shield is highly recommended Reagent Spills Wipe up spilled reagents immediately If reagent contacts skin rinse the affected area thoroughly with water Avoid breathing released vapors Read the Material Safety Data Sheet MSDS for each reagent for complete chemical information Fire Hazard Avoid the presence of flammable liquids or paper near the operating Reactor A fire hazard could be created SPECIFICATIONS Temperature of Reaction 105 C or 150 C 221 F or 302 F Temperature Stability 0 5 C Temperature Range 10 C to 160 C Accuracy 2 C 25 C Capacity 25 vials dia 16x100 mm 1 receptacle for a stainless steel reference thermometer Warm up Time 10 15 minutes depending on selected temperature Operating Mode timed 0 to 180 minutes or infinity mode Environment 5 to 50 C 41 to 122 F Power Supply HI 839800 01 115 Vac 60 Hz 250 W fuse protected HI 839800 02 230 Vac 50 Hz 250 W Dimensions 190 x 300 x 95 mm 7 5 x 11
2. temperature During this period the heating indicator will remain ON continuously until block temperature is within 10 C of target temperature The heater and the heating indicator will cycle OFF and ON in order to maintain the desired temperature e When the desired temperature is reached and has stabilized a beep will signal that vials may be placed in the reactor At this time the vials can be inserted into the block and the timed digestion can be started Press START to start the timer WAIT TO WAIT te ri ina 3 IO Suc no 200 NAA si EL ne SAS The instrument displays the current temperature and altemately on the second row the set point value and the reaction time Notes HOLE FOR e Use of the optional safety shield is strongly recommended THERMOMETER to protect the operator from splattering reagents in case a vial should break e Do not place the reactor near a heating or cooling source Temperature stability could be affected e In order to allow the heat transmission from the block to the liquid inside the tubes the temperature of the metallic block is a few degrees higher than the set value During operation the block temperature will be slightly different for an empty block or one full of vials If cold tubes are placed in the block the temperature will decrease several degrees and a few minutes will be required to recover as the tubes warm An external stainless steel probe thermometer can be used
3. temperature well in the block A 3 hour timer is incorporated for applications where timed digestions are required When this feature is enabled a beep will be heard and the heating element will turn off at the end of the selected time period The reactor contains a thermal fuse that prevents overheating by turning off the heating element The temperature is continuously displayed on the LCD even when there is no active temperature program running Two models are available e HI 839800 01 runs with 115 Vac power supply e HI 839800 02 runs with 230 Vac power supply PRELIMINARY EXAMINATION Please examine this product carefully Make sure that the instrument is not damaged If any damage occurred during shipment please notify your Dealer Each HI 839800 Test Tube Heater is supplied complete with e Power Cable Instruction Manual Note Save all packing material until you are sure that the instrument works correctly Any defective item must be returned in its original packing ABBREVIATIONS C degree Celsius COD Chemical Oxygen Demand HEALTH 8 SAFETY Before unpacking setting up or using this reactor it is important to carefully read this manual Pay particular attention to all warnings cautions and notes Failure to do so could result in serious injury to the operator or damage to the equipment To ensure the safe use of this reactor it must not be installed or used in any manner other than that specified in this manual
4. to check the block temperature by placing it into the appropriate receptacle After inserting vials press START to start the timer The heating indicator will cycle ON and OFF maintaining the constant temperature After the time has elapsed the message DONE is displayed The heating element will turn OFF and the samples are ready to be removed from the reactor To begin another digestion press START To modify the temperature program use the 105 C 150 C key To change the timer value use A and W keys The reactor should be kept clean and any reagent spills should be wiped up immediately Do not touch the heater block unless you are certain it has cooled If some reagent accidentally spills into the heater block disconnect power and allow the reactor to cool then remove all tubes and clean the block Should overheating occur the reactor automatically shuts down and all LED indicators turn off Contact your dealer or your nearest Hanna Service Center for maintenance FUSE REPLACEMENT Replace fuse only with fuse of specified type and current rating e Turn the reactor off and disconnect the power supply e Unscrew the fuse holder by rotating it counterclockwise Replace the fuse with one 4A fuse for 01 or one 2A fuse for 02 model e Replace the fuse holder before turning the reactor on HI 151 00 Electronic thermometer C HI 151 01 Electronic thermometer F HI 740216 Test tube cooling rack 25 tube c
5. 8 x 3 7 Weight approx 4 8 kg 10 6 Ib FUNCTIONAL DESCRIPTION REACTOR DESCRIPTION 1 Heater block can hold up to 25 vials 2 POWER LED indicator lights green when the power switch on the back of the reactor is set in the ON position 3 HOT surface indicator flashes red if the block temperature is above 50 C 4 HEATING LED indicator lights yellow only when the heater is ON When the indicator cycles ON and OFF the heater is maintaining a stable temperature 5 A V keys are used to adjust the timer values when setting reaction time 6 105 C 150 C key is used to select the temperature program 105 C or 150 C 221 F or 302 F 7 START key is used to start the selected temperature program or to start the timer after placing the vials into the heater block 8 STOP key is used to stop the heating process or to stop the timer 9 Liquid Crystal Display LCD 10 Power switch turns the reactor ON and OFF When the power switch is set in the ON position the power indicator will light 11 Power supply 115 or 230 Vac depending on model 12 Fuse holder the fuse is placed on the rear of the case for easy maintenance 5 OPERATING MODES e Idle mode the default mode While in this mode the meter measures and displays the heater block temperature an IDLE message the target temperature and the reaction time e Heating mode activated by the START key First the reactor will heat to the des
6. Instruction Manual HI 839800 COD REACTOR and Test Tube Heater 2008 Series na ce dat 110 kini a a www hannains Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the correct use of the instrument Please read it carefully before using the meter If you need additional technical information do not hesitate to e mail us at tech hannainst com This instrument is in compliance with directives TABLE OF CONTENTS PRELIMINARY EXAMINATION ssesssssssesssssssssssssssssssssssssssesssssssessssseecessseeceeeeeeeceeceeeeeeeeeeeeseeeseee 3 GENERAL DESCRIPTION eo A 3 ABBREVIATIONS HEALTH amp SAFETY All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Hanna Instruments Inc Woonsocket Rhode Island 02895 USA GENERAL DESCRIPTION The HI 839800 thermo reactor is constructed of durable materials with a vial capacity to perform up to 25 digestions simultaneously The reactor is equipped with 2 predefined temperature programs dedicated temperature settings for COD and total phosphorous reactions at 150 C and also 105 C for total nitrogen digestion In addition HI 839800 features LED indicators for visual indication of power on POWER heating in progress HEATING and hot surface warning HOT The heater block temperature can be verified by placing a thermometer in the
7. apacity HI 740217 Laboratory bench safety shield 10 WARRANTY HI 839800 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to the instructions This warranty is limited to repair or replacement free of charge Damages due to accident misuse tampering or lack of prescribed maintenance are not covered If service is required contact your dealer If under warranty report the model number date of purchase serial number and the nature of the failure If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization Number from the Customer Service Department and then send it with shipment costs prepaid When shipping any instrument make sure it is properly packaged for complete protection To validate your warranty fill out and return the enclosed warranty card within 14 days ftom the date of purchase Recommendations for Users Before using these products make sure that they are entirely suitable for your specific application and for the environment in which they are used Operation of these instruments may cause unacceptable interferences to other electronic equipments this requiring the operator to take all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the in
8. ired temperature When the temperature is reached and it is stable a beep signals the user that the vials can be inserted into the heater block Pressing the START key initiates the countdown timer A beep tone is heard also at the end of the heating cycle the heater is off and the block starts cooling The heating mode will be stopped anytime if the STOP key is pressed ERRORS AND WARNINGS The instrument shows messages when special conditions occur E Temperature Low Appears during the heating mode if the measured lt HH temperature is lower than the set point Wait for warm up period x 1500 LIL LO TEMP ima Temperature high Appears during the heating mode if the heater block gc 1 temperature is over the set point Wait for cooling period x Cna uu HI TEMP Heating system malfunctions Appears when the heating mode is ON H E A for more than 5 minutes and the temperature has not increased FAI LEd HOT Hot surface Appears when the heating block temperature is over 50 C Prepare the reactor as follows e Place the reactor on a flat table Do not place the instrument under direct sun light e Connect the reactor to a power outlet with ground Verify the correct voltage and frequency are used See the label on the instrument e Turn the reactor ON using the power switch located on the rear panel Wait until unit displays the heater block temperature TEMPERATURE SELECTION The reactor has 2 predefined te
9. mperature profiles e 105 C 221 F e 150 C 302 F To choose from the available profiles e At start up the display briefly illuminates all message segments then displays the heater block tempera ture The current settings are displayed reaction time and temperature on the second line WAIT TO DONE Doo DI AC LIST AC LIST 65 05 CH es 6600 00 s0 gt SAMPLES HILO TEMP e Press the 105 C 150 C key to selec e Press the A or W key to increase or decrease the reaction time e An infinite mode for time may also be set if the reaction time is adjusted in the way described above until the display shows the message e Ifthe selected temperature program is not started within 10 seconds the temperature screen will be displayed DIGESTION PROCEDURE Turn the reactor on using the power switch located on the rear panel The green POWER LED will be illuminated The display briefly illuminates all message segments and then displays the heater block temperature Select the temperature according to the specific digestion procedure e Press the START button to start the heating mode The yellow HEATING LED will illuminate e The display will show the current temperature of the heating block and the set temperature LO TEMP e When the heating block is above 50 C the red HOT LED will flash The heater block warms for approximately 10 15 minutes depending on the selected
10. struments EMC performance To avoid damages or burns do not put the instrument in microwave oven Do not use or store the instrument in hazardous environments Hanna Instruments reserves the right to modify the design construction and appearance of its products without advance notice Eis pane a n n a In s 80 20 240086E8NVW