Rigel 1550 User Manual

image

Contents

1. TeTechS Rigel 1550 Terahertz Spectrometer User Manual Te Techs Contents System Description Overview Part Names Operating Conditions Start up Procedure Safety Laser Safety Electrical hazards CE compliance Installation How to Install Sensors heads and connect to Spectrometer How to Restore Terahertz Signal Data Acquisition for Fast Scan Installing Drivers for the DAQ and Spectrometer Software Spectrometer Operation Running Rigel 1550 Spectrometer Software Comparing Spectra Running FAST SCAN Calculating Absorption Coefficient Zero Padding Option Accessories Exiting and Shutdown of Spectrometer Rigel 1550 Terahertz Spectrometer User Manual info tetechs com 10 10 10 11 13 15 16 16 23 26 30 31 31 32 Te TechS info tetechs com System Description Overview The Rigel 1550 Spectrometer is a versatile instrument operating in the terahertz range Rigel 1550 spectrometer is a portable modular compact and reconfigurable terahertz time domain spectrometer system capable of testing solid powder thin film gas and liquid samples for material sensing and characterization applications The fiber coupled movable transmitter and receiver heads can be mounted around the sample or process under test for the cases where bringing the sample or process under test inside the spectrometer housing 1s not feasible The terahertz transmitter and detectors are driven by a Femtosecond fib
2. The new position should be close to the peak position Repeat process by doing Steps 6 and 7 for finer alignment and should get a value for THz signal around 6 Rigel 1550 Terahertz Spectrometer User Manual 12 Te TechS info tetechs com Data Acquisition for Fast Scan The Rigel 1550 terahertz THz Spectrometer has fast scan option for conducting fast scan with setting of 1 Hz or one waveform per second The waveform scan length is 50 ps and step size is 50 fs By selecting the Fast Scan button and clicking START SCAN button in Rigel 1550 spectrometer software followed by turning on the FAST SCAN switch on the cabinet for the electronic housing one can do fast scan CAUTION Before selecting the Fast Scan option make sure the lock in setting in the Rigel 1550 Spectrometer software is set to sensitivity of High Dynamic mode and with 12 dB Oct filter setting and time constant of 3 ms The display range scaling should be the value of the sensitivity for example 1E 3 for I mV gain setting For continuous real time fast scan and data acquisition DAQ after doing one fast scan in main window of Rigel 1550 spectrometer software and switching on the FAST SCAN switch on the cabinet of the electronic housing run the DAQ FAST Real time program The DAQ FAST Real time use a data acquisition of the analog THz signal from lock in amplifier with a sampling rate of 20 kSamples second on each channel and is used in acquiring one waveform per second and disp
3. 34 05 09 2013 12 29 05 09 2013 12 54 05 09 2013 1 39 PM 05 09 2013 2 57 PM 05 09 2013 3 20 PM 06 09 2013 10 34 Type CSV File CSV File CSV File CSV File CSV File CSV File CSV File CSV File CSV File CSV File CSV File CSV File g g g g g g a A 4i A A j Save as type Al Files Cancel For example here we compare the scan of ABS plastic amie 1 After putting the sample in between the terahertz sensors conduct a scan over the same range from 20 mm to 25 mm with a speed of 0 1 mm s The resulting time domain signal and spectrum are shown below Rigel 1550 Terahertz Spectrometer User Manual 23 Te TechS info tetechs com Rigel Spectrometer from TeTech Inc ae File Edit Operate Tools Window Help I I I I I 25 25 2775 30 325 Scan Options Spectrum Options Br Normalized Reference Spectrum Y axis GO START POSITION Start Position mm Peak Position mm FWHM A ps dB Linear Y 20 23 075 1 49925 Zero Padded gt aan 0 FIND CURRENT POSITION Scan Width mm Signal Reading 5 000 FWHM C ps Clear Comparison Graphs Bandwidth C THz 7 0 0566682 0 Motor Speed mm s 0 rm 0 1 mm s FWHM D ps Bandwidth A THz Bandwidth D THz F 11 ee a oo r y g mEn Beso Z 9 F uA 7T Scapa g u It is shown that the pulse is broader from 1 3 ps in freespace to 1 5 ps with the ABS sample and the peak position shifted to 23
4. a _ K OO Figure 4 Rigel 1550 Terahertz Spectrometer Sensor Heads 2 Conduct a long THz scan using Rigel 1550 Software by setting motor speed to 0 5 mm s and scan range from 0 to 20mm Follow the instructions on Running Rigel 1550 Spectrometer Software to conduct slow THz scan If you do not find the THz peak and signal in this range change the delay in increments of 20 mm from 20 mm to 40 mm until you find the right delay for the THz peak Rigel 1550 Terahertz Spectrometer User Manual 11 Te TechS info tetechs com 3 Once the THz signal is obtained on the time domain graph the software automatically finds the THz peak and goes to the optical delay position for that THz peak The next step involves the maximizing the THz peak signal In order to do that launch Show Peak THz Signal application that shows the signal level in real time for the current delay position To improve the alignment the Tx and Rx sensor heads should be tilted along y axis z axis being optical axis one by one to maximize signal ensuring the polarization is the same The direction of the enclosure of the Tx and Rx antenna must be changed along the plane made by the marker for the polarization and the optical axis Then change the delay after adjustment and changing the direction of the antennas to maximize the THz signal reading The delay can be changed by entering the position in mm for start position and going there by clicking GO START POSITION
5. noise floor is shown too ia Rigel Spectrometer from TeTech Inc gt File Edit Operate Tools Window Help gt amp 3 Scan Options Spectrum Options o Normalized Reference Spectrum Y axis GO START POSITION f Start Position mm Peak Position mm FWHM A ps dB Linear g 20 23 075 1 49925 Zero Padded gt GO PEAK POSITION stop Position mm Current Position mm FWHM B ps u FIND CURRENT POSITION AH Scan Width mm Signal Reading lt j 5 000 a en Bandwidth C THz h 0 0566682 0 Motor Speed mm s 0 o sartsean 7 0 1 mm s FWHM D ps Bandwidth A THz Bandwidth D THz Start Scan l 0 0 66 0 Running Fast Scan CC FWHM E ps Bandwidth B THz Bandwidth E THz 0 0 929535 0 em POP a aa Famam Rigel 1550 Terahertz Spectrometer User Manual 25 Te TechS info tetechs com Rigel Spectrometer from TeTech Inc in So a uO O File Edit Operate Tools Window Help PA Te TechS Terahertz Signal Spectrum Bu ime ps Scan Options 2 Culley SOE ee ol Ne cer There is a clear difference in the spectrum and delay in the pulse in the time domain Here again the ABS sample is shown in white and the reference air measurement shown in red If you click clear comparison graphs it will clear the loaded signal and spectra to compare and shows the last scan which is in white line Note For convenience one can scan t
6. or changing the start position of the scan with the same scan width to measure and store the time domain signal and spectrum for post processing and comparison Note The value of 6V THz signal corresponds to 6V 10V x 300u V x 1 uA V Gain 180 pA THz current on Rx antenna Rigel 1550 Terahertz Spectrometer User Manual 22 Te TechS Comparing Spectra info tetechs com After finding the delay and obtaining the THz pulse using procedure of initializing and running Rigel 1550 put the sample in between the transmitter and receiver To compare the spectra first put sample at focus of the transmission module or if using the collimated beam between Tx and Rx put sample in the middle o amp O f ex Fee Name i Date modified _ Scan Spectrum _13 09 05 12 25 Referencel _ Scan Spectrum 13 09 05 12 52 Reference _ Scan Spectrum _135 09 05_ 15 31 References _ Scan Spectrum 13 09 05 14 39 Referenced _ Scan Spectrum _13 09 05 15 14 References _ Scan Spectrurmn _13 09 06_10 30_ Reference _ Scan_13 09 05 12 25 Referencel csv _ Scan_13 09 05 12 52 Reference2 csv _ Scan_13 09 05 13 31 _Reference3 csv _ Scan_15 09 05 14 39 Referencet csy _ Scan_15 09 05 15 14 ReferenceS csv _ Scan_13 09 06_10 30 Referencel csv 4 File name Scan_1309 06_11 23_ABS csv 05 09 2013 12 29 05 09 2013 12 54 05 09 2013 1 39 PM 05 09 2013 2 57 PM 05 09 2013 3 20 PM 06 09 2013 10
7. 08 mm The is also attenuation due to the sample 2 To compare the spectra with other measurements or for this example the reference click on COMPARE Select the file to compare which is a time domain scan csv file Do not load the spectrum as the only time domain signal csv file is used to compare the signal and spectrum Rigel 1550 Terahertz Spectrometer User Manual 24 Te TechS info tetechs com e Fee Date modified Type __ Scan Spectrum _15 09 05_15 31 References 05 09 2013 1 39 PM CSV File __ Scan Spectrum _ 13 09 05 14 39 Reference 05 09 2013 2 57PM CSV File __ Scan Spectrum _15 09 05_15 14 References 05 09 2013 320 PM CSV File __ Scan Spectrum _13 09 06 10 30 Referencel 06 09 201310 34 CSV File _ Scan Spectrum _13 09 06_11 25 ABS csv 06 09 2013 11 26 CSV File _ Scan_13 09 05_12 25 Referencel csv 05 09 2013 12 29 CSV File _ Scan_135 09 05_12 52 Reference csv 05 09 2013 12 54 CSV File _ Scan_13 09 05_13 31_Reference3 csv 05 09 2013 1 59PM CSV File _ Scan_13 09 05_14 39 Referencet csy 05 09 2013 2 57 PM CSV File _ Scan_13 09 05_15 14 ReferenceS csv 05 09 2013 3 20 PM CSV File can_13 09 06_10 30_ Referencel csy 06 09 2013 10 34 CSV File _ Scan_13 09 06 11 23 ABS csv 06 09 2013 11 26 CSV File Fe Files Cancel The Reference is loaded in red and the ABS sample measurement is in white shown below The change in FWHM of pulses and 20 dB bandwidth above
8. Graphs Linear Stage Settings VISA resource name for Stage Stage Device Number D COM4 0 1 Responses from Stage T Max Acceleration A J 10 01 STOP MOTOR Bandwidth Accuracy Settings Bandwidth Accuracy dB rh ih Bandwidth Accuracy Linear rh 7 0 01 Zero pad accuracy A ey 1024 7 Rigel 1550 Terahertz Spectrometer User Manual Lock in Settings Overload Unlocked Choose Time Constant 7 12 dB Oct 100 ms Choose Sensitivity High Dynamic Mode 300 uV 300 nA 0 Display Range Scaling r Y 1E 3 Signal to Maximize Fast Scan Settings Length of scan s is connected to uP y gt Delta t ps a 7 0 05 eeo 20 TeTech5 info tetechs com The setting for the lock in Time constant is 100 ms and the Sensitivity is 300 uV For the fast scan and other normal scans the Delta t or time resolution of the scan is 50 fs or 0 05 ps 4 Click on GO HOME 5 Under Scan Options after scrolling up enter the Start Position in mm Scan Width in mm and Motor speed in mm s for the normal scan The Stop position is automatically calculated so it does not have to be entered Rigel Spectrometer from TeTechS Inc File Edit Operate Tools Window Help i gt 2 1 f I I I I I I I 1 l I I I I e 25 5 ya nR 325 35 U 2 25 5 25 9702375 Time ps Frequency THz Ha Scan Options Spectrum Options Pr Normalized Reference Spectrum Y axis GO START PO
9. SITION Sant Position mm Peak Position mm FWHM A ps ao y 20 22 3025 1 29935 Zero Padded gt GO PEAK POSITION Stop Position mm Current Position mm FWHM B ps i pam 2 0 FIND CURRENT POSITION Scan Width mm Signal Reading ee s Y 5 000 M C ps Clear Comparison Graphs Bandwidth C THz 0 0566682 0 Motor Speed mm s 0 0 1 mm s FWHM D ps Bandwidth A THz Bandwidth D THz Start Scan a 0 93 0 Fast Scan 0 Running C D FWHM E ps Bandwidth B THz Bandwidth E THz _ r R Zu p Error Codes Settings Data Readings Graphs OENE a n WD E O For example for the position of where the Tx and Rx are located We do a scan for the delay starting at 20 mm and for 5 mm scan with a motor speed of 0 1 mm s This corresponds to a 33 3 ps scan Rigel 1550 spectrometer returns the time domain signal spectrum peak position in mm Full width half maximum FWHM of the pulse and the THz power spectrum bandwidth which is 20 dB above the noise floor The pulse width of our measurement in free space for a large aperture transmitter and 5 um dipole as receiver is 1 3 ps and bandwidth of 0 93 THz Rigel 1550 goes to the peak position automatically after each scan Note When clicking Start Scan for the first time the Shaker controller and motor for it 1s turned on and comes to center position affecting delay by 5 mm The is connected to uP LED lights up when servo motor is on and goes to c
10. age Settings Lock in Settings VISA resource name for Stage Stage Device Number T COM4 1 0 Responses from Stage n 12 Jo o lt 0 0 mh 3 6 2 16 4 50 18 g Choose Time Constant A 2 gt g 10 01 q Choose itivi nal to Maximi STOP MOTOR 2 TE BIO EEE y High Dynamic Mode 300 uV 300 nA 5 94344 Display Range Scaling y 1E 3 Bandwidth Accuracy Settings Bandwidth Accuracy dB Fast Scan Settings J 1 f Bandwidth Accuracy Linear Length of scan s is connected to uP wy y 7 0 01 g 5 Zero pad accuracy g 1024 Delta t ps A J 0 05 B ALICA Tele 2 Turn the FAST SCAN manual switch at the side of spectrometer ON 3 Run RigelRealTime exe NOTE The Main Rigel exe application program must be running with the lock n settings mentioned above for the real time signal and spectrum to show in the Rigel 1550 Real Time application Rigel 1550 Terahertz Spectrometer User Manual 27 Te TechS info tetechs com DAQ FAST Realtime2 vi SE ile Edit Operate Tools Window Help DAQ FAST 2 Real Terahertz Wave Signal 15 12 5 Ham Spectrum Y axis dB Linear gt Gl oo ee re 3 9 28a e Er 22 The Figure above shows the real time signal and its spectrum after running RigelRealTime application If a sample is placed between the transmitter and receiver the change would be shown in real time or in a matter of ms DAQ_FAST_Realtime2 vi ile Edit Ope
11. e Time domain csv file of the material that was scanned in step 1 and show the absorption coefficient once the files is loaded in the Spectrum window 5 To clear the graphs showing the absorption coefficient click Clear Comparison Graphs Rigel 1550 Terahertz Spectrometer User Manual 30 Te TechS info tetechs com Zero Padding Option Before a scan one can select Zero Padded checkbox to zero pad the time domain signal with the number of zeros equal to the Zero pad accuracy or default setting of 1024 The user can change the zero pad accuracy or number of zeros to add to the signal for zero padding before computing the spectrum for higher resolution by changing the Zero pad accuracy setting under the Bandwidth Accuracy Settings Accessories The Rigel 1550 terahertz Spectrometer comes with a transmission mode measurement module for conducting spectroscopy of power samples compressed into a 13 mm diameter pellet or to put any sample at focus of terahertz beam The terahertz transmission mode measurement module uses two focusing lenses to focus the collimated terahertz beam from the transmitter module unto the sample with a beam spot size of 3 mm or less The second focusing lens collects the transmitted beam through the sample and collimates the beam that is collected by the receiver sensor module Unlike the case of putting a sample such as a large plastic in between the transmitter and receiver without the transmission mode measurement
12. enter position for fast scan if FAST SCAN is turned on the servo motor for the Shaker will oscillate w r t the center position Rigel 1550 Terahertz Spectrometer User Manual 21 TeTechs info tetechs com 6 Click GO START POSITION 7 Make sure the Emission light on the Laser is ON before doing a scan Once the motor finishes going to the Start Position Click START SCAN then click Create New Scan Would you like to create a new scan file or open an existing imcomplete scan Open Existing Scan 8 Enter Filename for the Time domain signal to store and Spectrum to store as csv comma delimited files The filename contains the data and time of scan Once you click OK the shaker is turned on for the first time scan and then the Running Button will turn on green At the end of the scan we get the signal and spectrum shown below Rigel Spectrometer from TeTechS Inc a LO x Eile Edit Operate Tools Window Help Riigel i gt Te TechS Terahertz Signal Spectrum 575 HARI Scan Options Spectrum Options me Normalized Reference Spectrum Y axis _ m aa dB Linear z 20 22 3025 1 29935 Zero Padded Ta a p y gt i al tll DES ro are N See En a 9 One can do repeated measurements by putting a sample in transmission mode between the transmitter and receiver and doing the scan over the same range to see delay of the pulse and effect on the spectrum
13. er laser at 1550 nm Part Names Figure shows the main components of the front and back of Rigel 1550 terahertz spectrometer Optical Delay module Onefive ORIGAMI Femtosecond Laser Shaker electronic controller Laser driver Linear stage for slow scan Lock in amplifier NI DAQ for data acquisition Square wave generator USB Hub 10 Shaker power supply CHONDA RWN Rigel 1550 Terahertz Spectrometer User Manual 3 Te TechS info tetechs com Computer heads switch USB Fast scan a b Figure 1 Main features of Rigel 1550 Terahertz spectrometer a Front panel and b Back panel Rigel 1550 Terahertz Spectrometer User Manual 4 Te TechS info tetechs com _ p a M Sample Holder THz Receiver THz Transnutter Module Module e THz ee gt har Transmission THz Collimating mode weil Lens Figure 2 Rigel 1550 Spectrometer Head for Transmission mode THz Time domain Spectroscopy TDS Rigel 1550 spectrometer has optical light delivery system that includes an optical delay box for slow and fast terahertz scan The slow scan can be used to find the THz signal peak and scan over longer ranges longer distance between transmitter and receiver with less noise or high signal to noise ratio The fast scan option allows for almost real time measurement of changes when a sample is placed in the THz transmission mode unit or the user put
14. he sample first and store the data and do a reference measurement in air next to show the reference in white line as last scan The data for the samples are then loaded using COMPARE from the stored data and shown in color such as red Running FAST SCAN The fast scan option uses a sine wave generator with position voltage ranging from 1 5 V to 1 5 V and frequency of 1 Hz The Fast scan acquires a waveform every 500 ms using this setting The sine wave generator that drives the shaker servo motor can be increased to 20 Hz Rigel 1550 Terahertz Spectrometer User Manual 26 TeTech5 info tetechs com To run Fast scan mode first load Rigel exe for normal slow scan mode and get the signal using the Rigel 1550 spectrometer software as outlined above Rigel 1550 spectrometer automatically goes to the peak position Then follow the following steps 1 Change the Time constant of the lock in amplifier under Lock in settings to 12 dB Oct 3 ms as shown below The unlocked led light turns on if the signal is unlocked if reference 1s disconnected The overload turns on if the gain is high But the Signal to Maximize should show a signal between range of 10 V to 10 V with display range scaling of le 3 If the signal is within this range the DAQ is working and the lockin is not overloaded Rigel Spectrometer from TechS Inc File Edit Operate Tools Window Help SEID Error Codes Settings Data Readings Graphs Linear St
15. iber output to the right of the Rx Fiber output 3 Connect the Rx fiber labeled as Rx with one end already connected to the sensor to the collimator output labeled Rx on the chassis after passing the fiber through the hole on the top of spectrometer 4 Connect the MMCX to SMA connector on Transmitter sensor to SMA cable from spectrometer electrical connections label Tx The Tx and Rx output SMA cables pass through the hole at the top of the spectrometer 5 Connect the MMCX to SMA connector on Receiver sensor to SMA cable from spectrometer electrical connections label Rx output Rigel 1550 Terahertz Spectrometer User Manual 10 Te TechS info tetechs com How to Restore Terahertz Signal CAUTION The Transmitter and Receiver can be placed around any process under test or on optical breadboard but the shortest or minimum distance between the Tx sensor lens to Rx sensor lens is 17 cm One can increase the distance between the sensor heads up to a distance of 47 cm Ensure the polarization of the Tx and Rx sensor is correct by lining up the markers on the end caps of the sensors 1 Place the transmitter and receiver sensor heads in optical view of each other The Tx and Rx should on the same level vertically and horizontally The user has the option of placing the transmitter and receiver sensor heads around process or material under test but a breadboard is also provided for mounting the Tx and Rx on post i En A ee e ie
16. ified Type di data 05 09 2013 1 21 PM File folder _ demo csw 05 09 2013 3 43PM CSV File _ demod2 csv 05 09 2013 3 34 PM CSV File _ demo3 csv 05 09 2013 3 37 PM CSV File _ dermo csy 05 09 2015 3 45PM CAV File _ RigelRealtime aliases 04 09 2015 1 18 PM ALIASES F RigelRealtime 04 09 2013 1 18 PM Applicatic 55 RigelRealtime 04 09 2015 1 18 PM Confiqura IT j File name Rigel _Fastscan csv E Save as type All Files Cancel Rigel 1550 program will ask you to save the THz signal when you clicked STOP into a CSV file Rigel 1550 Terahertz Spectrometer User Manual 29 Te TechS info tetechs com Calculating Absorption Coefficient 1 In order to calculate absorption coefficient of a material first place material in transmission module or in between the transmitter and receiver and conduct a scan within the same scan range that you would use for reference scan in air 2 Check the checkbox on the right that is named Reference and run another scan removing the material and conducting an air scan within the same range NOTE Make sure the pulse for the reference is in the same scan range window as the material under test 3 While the Reference checkbox is clicked the last scan you did without a sample is the reference scan and will be used to compute the absorption coefficient by clicking COMPARE button 4 Similar to the comparison of spectra the software will ask the user to select the file nam
17. ion Always Take This Action 2 Right click on the NI 9125 BNC that is highlighted and run Self Test and the program returns a message saying The device passed the self test The click on Test Panels at the top of the window and you should see the screen below Rigel 1550 Terahertz Spectrometer User Manual 16 File Edit View Tools Help info tetechs com p amp My System b E Data Neighborhood 4 g Devices and Interfaces 4 Sp NI cDAQ 9174 cDAQ1 amp 1 NI9215 BNC cDAQIMod1 a Network Devices b PXI PXI System Unidentified b J Serial amp Parallel gt 4 Scales D 6 Software gt il M Drivers gt amp Remote Systems BE Properties X Delete Ep Self Test YA Test Panels 2 Reset Device Create Task Name Serial Number Slot Number Value 0x18657F5 1 Attributes Device Routes _ Calibration Back v A NI DAQmx Device Basics What do you want to do gt Run the NI DAQmx Test Panels Remove the device gt View or change device properties Rigel 1550 Terahertz Spectrometer User Manual 17 Te TechS info tetechs com Test Panels 1 NI 9215 BNC cDAQIMod1 Analog Input Channel Name Amplitude vs Samples Chart Auto scale chart W DAQ iMod 1 ai0 1000 3 Mode On Demand Measurement Type Voltage Max Input Limit Min Input Limit Units 10 10 V Terminal Co
18. laying the terahertz signal and spectrum The DAQ has capability to go up to 100 kS s per channel The spectrometer comes with 4 slots for use of DAQ and is extendable for an array of terahertz transmitter and receiver modules This is advantageous for industrial applications where an array of fiber coupled terahertz transmitter and receiver sensor pairs are used to inspect an industrial process and the data is acquired in parallel The DAQ FAST Real time allows for any change in the path of the terahertz beam between transmitter and receiver such as putting a plastic or paper sample or any material to be observable in the delay of the terahertz signal change in the amplitude of the THz signal and spectrum of the THz signal Rigel 1550 Terahertz Spectrometer User Manual 13 Te Techs info tetechs com 140 66 mm 120 68 mm 5 538 in 4 751 in 138 30mm 118 26 mm 5 445 in 4 656 in U National Instruments DAQ that can be used by Rigel 1550 Spectrometer software Rigel 1550 Terahertz Spectrometer User Manual 14 Te TechS info tetechs com Installing Drivers for the DAQ and Rigel 1550 Spectrometer Software For Rigel 1550 Spectrometer Application Rigel exe to work properly the user has to install the Labview Run Time 8 5 and NI DAQmx Device Driver 9 3 on your computer or Laptop To install the Labview Run Time 8 5 1 Download the installation from the National Instruments NI website at http www ni com dow
19. nfiguration Differential A 3 Click Start to run the DAQ and make sure there is a signal on the screen then click on Stop then Close Now DAQ is ready to use in Rigel 1550 Note Make sure the name of the Channel is as shown on the screen image above Rigel 1550 Terahertz Spectrometer User Manual 18 TeTech5 info tetechs com Test Panels 1 NI 9215 BNC cDAQIMod1 Analog Input Amplitude vs Samples Chart Auto scale chart 7 Stop m Launch Rigel l Run Rigel exe from the Rigel 1550 Spectrometer folder 2 Adjust the screen to fit window of your Laptop or Computer Rigel 1550 Terahertz Spectrometer User Manual 19 Te TechS info tetechs com Rigel Spectrometer from TeTechS Inc File Edit Operate Tools Window Help i gt Te TechS Terahertz Signal Spectrum 50 55 60 65 70 m 75 85 Scan Options GO START POSITION Position mm Peak Position mm FWHM A ps 0 0 Vi 0 TOD i nar wg BE ro 29 Spectrum Options _ Normalized _ Reference Spectrum Y axis dB Linear Zero Padded nn Sudan 3 Scroll down to screen with Tabs and select Settings Tab The Lock in Settings Linear Stage settings and Fast Scan settings are as shown below Rigel Spectrometer from TeTech Inc File Edit Operate Tools SEIT Window Help Error Codes Settings Data Readings
20. nloads 2 Run the application and follow the instructions To install the NI DAQ driver version 9 3 1 If you have other versions of the NI DAQ on your computer remove it and install the NI DAQ driver by running the NIDAQ930f2 exe that comes with Rigel 1550 Spectrometer and Rigel 1550 Real Time terahertz signal acquisition tool 2 Follow the instructions for the installation of the driver for the DAQ Rigel 1550 Terahertz Spectrometer User Manual 15 Te TechS info tetechs com Spectrometer Operation Running Rigel 1550 Spectrometer Software After installing the drivers connect the USB cable to the Laptop or computer and the operating system will detect the 4 devices connected using the USB cable The cDAQ 9174 should be detected and the NI 9125 connected to it The other three devices are the lock in amplifier Linear stage for the slow scan and shaker controller for the Fast scan 1 Before running Rigel exe to launch Rigel 1550 spectrometer software make sure the DAQ is connected and run a test using Measurement amp Automation Explorer NI MAX to make sure the DAQ is connected Select Configure and Test This Device below New Data Acquisition Device MI DAGiins has detected a new data acquisition device W cDAQ 3174 What would you like to do begin an Application with This Device Using Ml Labview 64 bit Gy Configure and Test This Device ws Using Nl Measurement amp Automation Explorer Take No Act
21. rate Tools Window Help ErrorCodes Data Graphs Waveforms to Average ai Deita_t J 0 05 Ji 0 01 THz_Signal num of samples per waveform A 0 lt 7 10000 Jo o 000715 Counter Terahertz Time Terahertz Signal Amplitude Spectrum t3 i j 1 Jo Jaan Jo fna Ao ro rn Num of Waveforms 7 0 22 2805 J 0 0 244958 0 22 2727 0 238125 me cm Stack full 22 2805 0 167009 E TaN Reference Loaded 22 2805 0 320084 3 22 2805 0 249925 eC ms Eu 22 2701 0 248011 Zr J m n mere Goa 32 8 Ep Po 2m p The Data tab shows the setting for Delta t which is 50 fs the number of waveforms to average is set to 1 for fast scan and up to 5 to average out the noise Since a 3 ms time constant averages out Rigel 1550 Terahertz Spectrometer User Manual 28 TeTechS info tetechs com the noise at that speed there might not be a need to average the waveforms for this setting but if the Fast scan is increased to 20 Hz this function is used to average the waveforms say up to 11 waveforms before showing the THz signal and spectrum The X scale is a scale for the amplitude of the THz signal and it is set to 0 01 This scale is adjusted to view the THz signal to values of between 15 to 15 for sensitivity of 300 uV on the lock in 4 Click the STOP to stop the continuous data acquisition after doing measurements EB Choose CSV file to save Terahertz signal J Real Time Terahertz Sign 5 e Eal Name Date mod
22. rical Hazards The Rigel 1550 spectrometer contain electrical circuits devices and components operating at high voltages Contact with these circuits devices and components can cause death serious injury or painful electrical shock Good grounding 1s essential to avoid a potentially serious electric shock hazard Application of the wrong supply voltage can create a fire hazard and potentially serious shock hazard and could seriously damage the Rigel 1550 spectrometer system and any attached equipment CAUTION Do NOT use power cords with faulty or frayed insulation Rigel 1550 Terahertz Spectrometer User Manual 9 Te TechS info tetechs com CE compliance The Rigel 1550 spectrometer and sensor heads have been designed to comply with the requirements of the Electromagnetic Compatibility EMC Directive and the Low Voltage Directive or LVD of the European Union Installation How to Install Sensors heads and connect to Spectrometer WARNING Do not disconnect the fiber and MMCX cable from the Rx and Tx sensor heads 1 The Transmitter and Receiver sensor heads come disconnected from the chassis of the Spectrometer Both the optical fiber and electrical SMA connector must be connected during installation process 2 Connect the Tx fiber labeled as Tx with one end already connected to the sensor to the collimator output labeled Tx on the chassis after passing the fiber through the hole on the top of spectrometer The Tx is the middle F
23. s in the THz transmitter and receiver modules in the path of a process outside the spectrometer head Rigel 1550 Terahertz Spectrometer User Manual 5 Te TechS info tetechs com b Figure 3 Rigel 1550 Spectrometer Head for a Reflection mode and b ATR THz Time domain Spectroscopy TDS Rigel 1550 Terahertz Spectrometer User Manual 6 Te TechS info tetechs com Operating Conditions Eectrical requirements 100 240 V 50 60 Hz lt 5 A Operating temperature 15 35 C Storage temperature 15 65 C Humidity lt 50 Warm up time 10 minutes for laser emission Start up Procedure CAUTION Do not turn the power to the instrument on and off quickly as this may cause damage to the units and power supply 1 Open the sample compartment spectrometer head and make sure there is no sample in sample holder and place sample holder in the center 2 Make sure the THz Beam path between transmitter and receiver modules is free from obstructed 3 Connect the power supply the fiber optic cable for the Transmitter Blue and the other to the receiver Red one SMA connection labeled Tx to the Transmitter and the other SMA connection to the output from the receiver module 4 Connect the USB from your Laptop or computer to the USB connection on side of the spectrometer cabinet for electrical components 5 Turn the FAST SCAN OFF and POWER ON and green light comes on and the spectrometer is ready for use Rigel 1550 Teraher
24. tz Spectrometer User Manual 7 TeTechS info tetechs com 6 Turn the key for the laser to On position and wait till the status LED stops blinking and has stable green light Then press Emission button until it turns red light 7 Wait until Windows recognizes the instruments and initialization is complete 8 Run the Rigel 1550 Spectrometer Software to do a scan and find the THz peak signal before fast scan or simply conduct a slow THz scan using the spectrometer Rigel 1550 Terahertz Spectrometer User Manual 8 Te TechS info tetechs com Safety Laser Safety Rigel 1550 spectrometer transmitter and receiver are driven by Fiber coupled femtosecond laser from Onefive Onefive s ORIGAMI is a Class 3B laser If user opens optical delay module direct and mirror reflected high peak power invisible light pulses can be hazardous to eyes If fiber is disconnected or the optical delay module open the laser power emitted by the ORIGAMI may cause retinal or corneal damage The skin of the human body can severely be damaged when exposed to the radiation Therefore extreme care should be exercised during laser operation User does not need to eye wear during normal operation of the Rigel 1550 THz spectrometer Direct exposure of the eye to the invisible laser beam must be avoided DANGER AVOID EYE AND SKIN EXPOSURE TO LASER RADIATION RADIATION OF THE ORIGAMI IS INVISIBLE AND IN FIBER AND DO NOT DISCONNECT THE TX AND RX FIBERS Elect
25. unit where the collimated beam interacts with the sample using the transmission mode measurement unit focuses the beam to a sample spot size so the beam interacts with the sample over the small region such as the case for a 13 mm diameter pellet The transmission mode measurement unit also has two other positions to put the sample holder with the sample to have the sample interact with a slightly de focused THz beam Although the spot size at these locations is not at focus the beam spot sizes are smaller than the case for the collimated THz beam Rigel 1550 Terahertz Spectrometer User Manual 31 Te TechS info tetechs com Exiting and Shutdown of Spectrometer Turn the Emission red button on the laser off Press Quit button in Rigel 1550 Terahertz Spectrometer software Unplug the USB connection to the Spectrometer Turn the key on the laser to switch OFF the laser Turn the Green button for the Spectrometer OFF An a Rigel 1550 Terahertz Spectrometer User Manual 32 Te TechS Contacts TeTechS Inc 295 Hagey Blvd 1 Floor Waterloo Ontario N2L 6R5 Canada Toll Free 1 855 574 1764 Tel 1 519 584 0791 Fax 1 519 513 2421 www tetechs com Email info tetechs com TeTechS Inc 2014 Rigel 1550 Terahertz Spectrometer User Manual info tetechs com 33