Checklist - Agilent Technologies
Contents
1. Windows XP Professional Service Pack 2 or 3 CD CDR CDRW or DVD DVD R DVD RW drive Peripherals Windows compatible keyboard and mouse Agilent Technologies does not guarantee the function of the Chemstation or MS Workstation software on computer hardware or operating systems that do not meet the specified requirements Contact your Sales Representative for a list of qualified equipment or for more information Gas Requirements Nitrogen A nitrogen gas supply that can provide up to a maximum of 12 L min of gas regulated at 80 psi is sufficient for ESI or APCI operation for one LC MS If there is more than one LC MS use a nitrogen supply that can deliver a maximum of 18 L min The nitrogen must be at least 99 pure with less than 0 1 ppm hydrocarbons and less than 1 oxygen It must be clean and dry with a 40 C dew point Use a nitrogen generator or a liquid nitrogen boil off to supply the nitrogen gas All nitrogen generators require regular maintenance The customer is responsible for setting up the preventive maintenance procedures and schedules contact the nitrogen generator manufacturer Any MS problems caused by poor gas quality are not covered by the Warranty It is the responsibility of the customer to conform to all regulations regarding the installation and operation of the gas system The nitrogen supply uses 1 4 in connectors Check the checklist box Nitrogen gas at least 99 pure regulator and gas lines
2. has information about the power cords and plugs Figure 3 shows the plug types Table 5 Power Cord 500 lon Trap LC MS Foreline Pump MS40 Length 1 each 2 5 m 8 ft 1 each 2 5 m 8 ft Power Plug Type 1 each CEE 7 7 plugs 1 each CEE 7 7 plugs or or 1 each NEMA 6 15P 1 each NEMA 6 15P The power cables for the CPU monitor and printer are approximately 2 13 m 7 ft long They have NEMA 5 15P plugs Figure 3 shows the conformation of the plugs and outlets Replacing or substituting power cords or plugs must be done with strict CAUTION compliance with all regulations including electrical codes power cord color coding and appropriate regulatory agency certification marks Outlets Plugs B CEE 7 7 t ae NEMA 6 15P NEMA 5 15P Figure 3 Power plugs and outlets Installation Site Power The power supply for the MS and the foreline pump must be 200 240V ac The power supply for other modules such as the LC printer and autosampler can be 100 120V ac or 200 240V ac The power supply must be stable free of fluctuations due to slow changes in the average voltage or to changes resulting from surges sags or transients 14 Agilent 500 lon Trap LC MS System MS Site Preparation Guide Site Preparation A measured ground to neutral potential of greater than 3V ac or de indicates grounding problems that may need correction before connecting an instrument module to the power source Evaluate any power source suspected o
3. pump and LC are heavy To prevent personal injury use CAUTION i H appropriate moving and lifting techniques Bench Use the following information to plan the layout of the modules of your LC MS system The Agilent Representative will unpack the cartons and puts the modules on the bench The system requires a clean flat bench The bench must have enough space for modules the mouse solvent bottles and such The bench must be strong enough to support the weight of the modules If the bench is against a wall drill a hole large enough for the vacuum tubing of your pump through the rear of the bench The area under the bench must be large enough for the height of the foreline pump If the bench is higher than 92 cm 86 in place a vibration isolation bench under the foreline pump The vibration isolation bench must be capable of supporting the foreline pump gilent 500 lon Trap LC MS System MS Site Preparation Guide 7 Site Preparation The Agilent 500 Ion Trap LC MS system is designed for indoor use at altitudes of up to 2 000 meters Table 1 lists dimension and weight information for the 500 Ion Trap LC MS system and related components The installation site must have enough bench space for the 500 Ion Trap LC MS system LC system data system and accessories In addition there must be sufficient space around the instrumentation for air flow and maintenance access At least 30 cm 12 in to the left of the 500 Ion Trap LC MS and at
4. ships with the system contains plenty of tubing for the installation of the MS LC pumps and LC autosampler If additional valves or modules are to be installed then additional tubing PEEK or stainless steel may be required Please contact your sales representative if you have any questions Check the checklist box Materials and solvents of specified grade are on site 18 Agilent 500 lon Trap LC MS System MS Site Preparation Guide Site Preparation 3 When the LC MS Arrives Inspecting the Shipping Cartons Do not open any shipping cartons The Agilent Representative will open them during installation Move the shipping cartons to a warm dry and secure area near the installation site After the instrument arrives carefully inspect the exterior of the shipping cartons for evidence of any damage that could have possibly occurred during shipment Inspect the cartons for the following e Waiter stains e Cuts punctures or deep indentations e Crushed corners or excessively abraded edges e Arrow point on the Tip N Tell indicator is blue see Table 9 e ShockWatch indicator tube is red see Table 9 Table 9 Tip N Tell One indicator is on the exterior Read and follow the instructions on the label If the Tip N Tell arrow point is blue the carton was on its side or tipped in transit and instrument damage may have occurred MONITORED SHIPMENT IF TIP N TELL ARROW POINT IS BLUE THIS PACKAGE HAS BEEN ON ITS SIDE OR TI
5. short term heat dissipation of 8900 Btu hr during startup Check the checklist box Temperature maintained between 16 and 30 C 61 86 F Humidity The relative humidity of the operating environment must be between 20 and 80 with no condensation Operating the LC MS at a very low humidity may result in the accumulation and discharge of static electricity shortening the life of electronic components Operating the system at high humidity may produce condensation and result in short circuits Use a combination temperature and humidity monitor at the installation site Check the checklist box Relative humidity maintained between 20 and 80 Particulate Matter The installation site must not have excessive dust smoke or other particulate matter Particulate matter may block airflow vents and cause the electronics to overheat Check the checklist box Installation site is free of excessive particulate matter gilent 500 lon Trap LC MS System MS Site Preparation Guide 11 Site Preparation Electrical Requirements The customer is responsible for providing appropriate electrical power and power outlets for all of the components and accessories used with the 500 Ion Trap LC MS system Power considerations include e Voltage ranges of major components e Power requirements of major components e Single phase power configurations e Power plugs receptacles and cords 12 Agilent 500 lon Trap LC MS System MS Site Preparati
6. Agilent 500 lon Trap LC MS System Site Preparation Guide re Agilent Technologies Checklist NOTE Do not unpack the shipping cartons Check off each checklist box after satisfying each requirement as described in the instructions All requirements must be met before requesting installation NOTE If the installation site is not ready for installation when the Agilent Representative arrives Agilent Technologies reserves the right to invoice for the Representative s time Requirements Installation site is in compliance with all relevant safety regulations User Representative will be available during the installation and certification period Entrance to the installation site is at least 92 cm 36 in wide Bench space is sufficient Bench can support system weight at least 114 kg 250 Ib Bench does not vibrate Exhaust system Is suitable Temperature maintained between 16 and 30 C 61 86 F Relative humidity maintained between 20 and 80 Installation site is free of excessive particulate matter Specified electrical supply and power outlets are installed Nitrogen gas at least 99 pure regulator and gas lines are installed Air less than 0 1 ppm hydrocarbons regulator and gas lines are installed Helium at least 99 999 pure regulator and gas lines are installed Materials and solvents of specified grade are on site O0 0 00 0000 0 0 00 0j ojojo Shipping cartons examined for damage If there w
7. PPED OVER IN TRANSIT MAKE NOTE ON BILL OF LADING AND CHECK FOR DAMAGE ANY CLAIMS FOR TIPPING DEPEND ON THIS NOTATION TELL PATENT NO 3 926 144 PRINTED IN U S A ShockWatch One indicator is on the exterior Read and follow the instructions on the label If the tube on the ShockWatch indicator is red the carton was dropped in transit and instrument damage may have occurred The instrument label may be yellow purple or red depending on the sensitivity of the instrument to impact that exceeds a specified G level force gilent 500 lon Trap LC MS System MS Site Preparation Guide 19 Site Preparation 20 MONITORED SHIPMENT BERWACHUNG CONTR LE TRANSPORT CARGA MONITOREADA IF SHOCKWATCH INDICATOR IS RED S W amp WENN DER INDIKATOR ROT IST HOCK ATCH SI LE CHOC LABEL EST ROUGE SI EL INDICADOR ESTA ROJO Do aot refuse shipment a Ne pas reluser le colis Note on delivery receipt and inspect tor damage s WARNING Indiquer sur le bordereau que le label est d ctench Veriliez les d g ts ventuets m 3 H damaged leave in original comtainer and packaging then request immediate inspection from carrier within 3 days of delivery N En cas de d g ts mettez des r serves circonstanci es par lettre recommand e aupr s du transporteur www shockwatch com If no external damage is apparent write Received but not inspected on the receiving documents to indicate that the cartons were not o
8. are installed Air 16 Agilent 500 lon Trap LC MS System MS Site Preparation Guide Site Preparation A compressed air gas supply capable of providing greater than or equal to 2 L min of gas regulated to 80 psi with a two stage regulator is required as a nebulizing gas for negative ESI The customer is responsible to conform to all regulations for the installation and operation of the gas system The air must be clean and dry with less than 0 1 ppm total hydrocarbons including methane and have a 40 C dew point The air supply uses 4 in connectors Check the checklist box Air less than 0 1 ppm hydrocarbons regulator and gas lines are installed Helium Use helium as the damping gas with a flow rate of 1 mL min Helium must be at least 99 999 pure and regulated to 40 psi 2750 bar The minimum requirement is one 257 ft3 tank with an Alltech regulator Part Number AL8111 or equivalent tank and regulator If you are concerned about the cleanliness of the helium use a Gas Clean filter Part Number CP17973 Check the checklist box Helium at least 99 999 pure regulator and gas lines are installed Materials and Solvents The Agilent Representative prepares solutions to tune and evaluate the LC MS Please supply the materials on Table 7 Error Reference source not found and the solvents on Table 8 Table 7 Quantity Item each 100 mL clean and new volumetric flasks 10 mL clean and new volumetric flasks Micr
9. as any damage the conditions were reported Requesting Installation After preparing your site contact the Customer Service office in your region to schedule installation Agilent 500 lon Trap LC MS System MS Site Preparation Guide Checklist Requesting Installation General Information Safety 2 Before the Installation Customer Representative Agilent Responsibility Activities Not Covered as Part of Installation Procedure Space and Weight Requirements Entrance Bench MS40 Foreline Pump Table or Supporting Surfaces Vibration Requirements Exhaust System Temperature Humidity Particulate Matter Electrical Requirements 1200 Infinity LC Series 212 LC 460 LC 500 lon Trap LC MS Power Requirements Summary Power Plugs and Cords Installation Site Power Qualified Computer Equipment Gas Requirements gilent 500 lon Trap LC MS System MS Site Preparation Guide Sg dd O OQ OQA O Materials and Solvents 17 3 When the LC MS Arrives 19 Inspecting the Shipping Cartons 19 Unpacking and Installation 20 Spare Parts and Consumables 22 System Verification 22 Preventive Maintenance 22 Trademark Acknowledgment 22 Agilent 500 lon Trap LC MS System MS Site Preparation Guide Site Preparation 1 General Information The site conditions must meet the minimum specifications before the Agilent Field Support Engineer FSE can proceed with the system installation The Site Preparation Guide prepares you for each r
10. elivery are accepted The process requires at least three days gilent 500 lon Trap LC MS System MS Site Preparation Guide 21 Site Preparation Spare Parts and Consumables Please refer to the 500 MS Ion Trap LC MS Mass Spectrometer Hardware Operation Manual part number 395407600 which has a list of the part numbers for spare parts and consumables System Verification The final step in the installation process is system verification Your Agilent service representative will test the system against Agilent standards as documented in the Agilent 500 Ion Trap LC MS System Installation Manual or in the installation guides for the for the 500 Ion Trap LC MS accessories you have purchased The Agilent service representative will not test your system using your standards or samples Furthermore the Agilent Customer Engineer will not set up your laboratory procedures Assistance with laboratory procedures can be obtained from your local Agilent Applications Engineer AE on a consulting basis at an additional cost Preventive Maintenance The customer is responsible for performing routine and preventive maintenance of the LC autosampler MS and data system If using a nitrogen generator perform the preventive maintenance to ensure that the nitrogen supply is clean and dry Any instrument problems resulting from a contaminated gas supply are billable and not included in the Warranty Perform regular preventive maintenance to increase t
11. equirement of the checklist Follow these instructions to ensure that the installation requires no more than the usual three days Have the completed checklist available to schedule the installation After the Agilent Representative completes the installation and it is accepted samples can be analyzed After meeting all of these requirements contact the Customer Service office in your region to schedule installation The LC MS operates reliably under carefully controlled environmental conditions You must provide suitable power sources operating environment materials and solvents Failures may occur if you use or maintain the system outside of the power and operating environment ranges and limits described in these instructions All phases of the installation site preparation must conform to local safety ical ildi Th k CAUTION electrical and building codes These codes take precedence over any recommendations in these instructions Compliance is the responsibility of the customer Safety Safety is the most important consideration for instrument use Determine if the installation site complies with all relevant safety regulations Check the checklist box Installation site is in compliance with all relevant safety regulations gilent 500 lon Trap LC MS System MS Site Preparation Guide 5 Site Preparation 2 Before the Installation Customer Representative Schedule the installation when the Customer Representative wi
12. f having noise problems evaluated with a recording type power line monitor before operating the system Contact your Sales Representative for a list of qualified equipment or for more information Table 6 MS Configuration Maximum Power Consumption 500 lon Trap LC MS without foreline 1200 VA pump 500 lon Trap LC MS with foreline pump 3400 VA Specified electrical supply and power outlets are installed SS WARNING Do not use extension cords with the Agilent 500 lon Trap LC MS Extension cords can be a safety hazard Qualified Computer Equipment If MS Workstation or Chemstation software is to be installed ona computer not purchased from Agilent the customer is responsible to ensure that the computer is adequately equipped and compatible with the operation of the data system and its communication interfaces Chemstation ask your Sales Representative for information MS Workstation Please check the list of requirements in the following table gilent 500 lon Trap LC MS System MS Site Preparation Guide 15 Site Preparation MS Workstation Computer Requirements MS Workstation Version 6 X Pentium 4 or higher Processor 3 GHz minimum a drive with a minimum of 20 GB ss is One full size PCI slot for each DSP or GPIB board required Expansion Slots One PCI slot for each PCI Ethernet board required ce i 1280 x 1024 x 256 color minimum video Display 1280 x 1024 x 65536 color recommended Operating Systems
13. f the instrument i i Check the checklist box Bench does not vibrate Exhaust System Do not put the foreline pump in an enclosed space Foreline pumps exhaust most compounds introduced into the MS and small amounts of oil vapor Provide an adequate fume exhaust system for the outlet of the foreline pump The foreline pump must be vented at least 2 L min The API spray chamber must be vented at least 20 L min to a separate vent line Ensure that the exhaust system does not pull a vacuum on the API chamber Consult local regulations for the proper method of exhausting the fumes lef Check the checklist box Exhaust system is suitable 10 Agilent 500 lon Trap LC MS System MS Site Preparation Guide Site Preparation Temperature The optimal operating range is between 16 and 30 C 61 86 F for best mass stability the temperature should not vary by more than 3 C 5 F during instrument operation As the installation site temperature increases system reliability decreases due to heat generated by electronic components during operation This heat must dissipate to the surrounding air for reliable operation The airflow around the system must be adequate The air conditioning system must maintain a constant temperature within operational limits around the system Do not place the system near air ducts windows or heating and cooling systems The average steady state heat load for the MS alone is 5200 Btu hr with a possible
14. he life of the system to maximize system uptime and optimize system performance Please refer to the 500 MS Ion Trap LC MS Mass Spectrometer Hardware Operation Manual part number 395407600 for details The Agilent Representative will describe and demonstrate these procedures during the installation Trademark Acknowledgment Microsoft and Windows are registered trademarks of Microsoft Corporation Swagelok is a registered trademark of Swagelok Company Solon Ohio PEEK is a registered trademark of the Victrex PLC Company Lancashire England Other brand and product names are trademarks or registered trademarks of their respective holders 22 Agilent 500 lon Trap LC MS System MS Site Preparation Guide hr Agilent Technologies www agilent com In This Book This book contains information to prepare your site for the Agilent 500 Ion Trap LC MS system components Agilent Technologies 2010 December 2010 Revision 9 395407500
15. ht of the Data system depends on the components included Reserve at least 100 cm 39 in of bench space for the data system A typical data system weight is 23 kg 50 Ib Agilent 500 lon Trap LC MS System MS Site Preparation Guide Site Preparation Check the checklist box Bench space is sufficient NW Check the checklist box Bench can support system weight at least 114 kg 250 Ib Please refer to Figures 1 and 2 for suggested system configurations Figure 1 Agilent 1200 Infinity Series and the 500 lon Trap LC MS Figure 2 Varian 212 LC pumps and 460 LC autosampler with the Agilent 500 lon Trap LC MS gilent 500 lon Trap LC MS System MS Site Preparation Guide 9 Site Preparation MS40 Foreline Pump Table 2 Item Description Diameter of hole in bench if needed 5 8 cm 2 3 in Width 29 7 cm 11 8 in Depth 41 8 cm 16 5 in Height 22 8 cm 9 in Vacuum hose length 1 8 m 6 ft Power cord length 2 5 m 8 ft Serial cable length 2 7 m 9 ft Dedicated circuits 1 circuit for the MS40 pump which can provide 200 240 V ac 50 60 Hz 10A Power cord plug CEE 7 7 or NEMA 6 15P Table or Supporting Surfaces Vibration Requirements For all 500 Ion Trap LC MS Systems the bench or table where the instrument will be installed must be vibration free Do not place a HPLC modules or the CTC Autosampler on top of the CAUTION instrument These modules can cause vibrations which can affect the resolution o
16. least 15 cm approximately 6 inches behind the 500 Ion Trap LC MS must kept clear WARNING Do not stack the LC modules on top of the 500 lon Trap LC MS system This arrangement is potentially unstable and dangerous The foreline pump must be located under the laboratory bench or on the floor It must be within the 200 cm 79 in length of the vacuum hose The hose exits only the back of the 500 Ion Trap LC MS The hose is metal and cannot be bent sharply Do not put the foreline pump on your laboratory bench with the 500 lon Trap CAUTION LC MS due to the vibration that the pump creates This vibration can lead to a loss of mass accuracy and resolution hould be k CAUTION The supporting surface for the 500 lon Trap LC MS system should be kept relatively vibration free Please refer to Table 1 for further information regarding the dimensions and weights of the 500 Ion Trap LC MS foreline pump and other modules Product dimensions Table 1 Product Width Weight Dimensions cm in kg Ibs 500 lon Trap LC MS 54 21 62 135 1200 Infinity Series LC 34 5 13 5 48 104 212 LC 26 10 5 14 5 32 460 LC without cooling 330 12 19 42 460 LC with cooling 330 12 21 46 Agilent 1200 Infinity Size and weight of the LC system depends on the Series LC number and type of modules included Most Agilent 1200 Infinity Series LC modules are approximately 35 cm 14 in wide and 45 cm 18 in deep Data System Size and weig
17. ll be available One of the important duties of the Agilent Technologies service representative is to familiarize the Customer Representative with the basic functions of the LC MS Check the checklist box User Representative will be available during the installation and certification period Agilent Responsibility An Agilent Technologies service representative will install the 500 Ion Trap LC MS and verify its performance The service representative s responsibilities are limited to 1 Unpacking the 500 Ion Trap LC MS system and verifying that all components are present and undamaged 2 Connecting the carrier gas line to the instrument from the tank regulators and lines previously installed by the customer 3 Installing connecting and turning on 500 Ion Trap LC MS system components 4 Verifying that the system meets Agilent Technologies performance standards 5 Providing basic user familiarization for system hardware and software 6 Signing the customer up through the Response Centers for instrument and software support Activities Not Covered as Part of Installation Procedure Agilent Technologies is not responsible for e Any task not listed in the Agilent 500 Ion Trap System Installation Manual or in the installation guides for the LC data system and other accessories e Installing a nitrogen gas generator and or air compressor unless additional installation time is purchased e Connecting or verifying the perfo
18. on Guide Site Preparation 1200 Infinity LC Series See the 1200 Infinity LC Site Preparation Guide for information about the electrical requirements 212 LC The 212 LC requires 85 264V ac 47 63 Hz 60 Watts 100 VA 460 LC The 460 LC requires 95 240V act 10 50 60 Hz 200 VA 500 lon Trap LC MS The 500 Ion Trap LC MS requires a separate circuit and the outlet must have adequate amperage capacity and a reliable ground The power source must be clean and provide 200 240V ac 50 60 Hz 3 Hz 1 0 kW 10A 1200 VA steady state Power Requirements Summary Table 3 Component Circuit Power Plug and Outlet Type 500 lon Trap LC MS 1 each 220V US 1 each NEMA 6 15P EU 1 each CEE 7 7 Foreline Pump 1 each 220V US 1 each NEMA 6 15P MS40 EU 1 each CEE 7 7 Computer 110V or 220V see appropriate user manual LC modules 110V or 220V see appropriate user manual Use a separate dedicated power source for the HPLC modules additional instruments and equipment Never plug the mass spectrometer and the foreline pump into the same power source or the power source may overload Never use the free outlet on any of the power sources for equipment that draws more than 2A Table 3 and Table 4 provide information about power requirement Table 4 Modules Max Current Draw Amps 500 lon Trap LC MS 6A Foreline Pump MS40 10A gilent 500 lon Trap LC MS System MS Site Preparation Guide 13 Site Preparation Power Plugs and Cords Table 5
19. opipette capable of dispensing 1 0 mL and tips Micropipette capable of dispensing 200 uL and tips It is recommended that the highest grade solvents and chemicals available be used with the 500 lon Trap LC MS Lower purity grades will result in higher background noise in the 500 lon Trap LC MS Table 8 Quantity Solvent each 1 4 liter bottle LC MS grade water new and unopened 2 1 liter bottle LC MS grade acetonitrile new and unopened 1 1 liter bottle LC MS grade methanol new and unopened gilent 500 lon Trap LC MS System MS Site Preparation Guide 17 Site Preparation Figures 4 and 5 illustrate how critical LCMS grade solvents are for optimum 500 Ion Trap LC MS performance The 500 Ion Trap LC MS is sensitive enough to detect impurities in HPLC solvents which cause a very noise baseline ap ol all RA AIG gt Bol au kCount pesimsms003 XMS TIC Filtered 609 3 gt 80 460 1 92V i i 20 S N PP 11 S 14225 N 1245 minutes Seg 1 Time 0 00 12 46 reserpine Scan Functions 1 1225 1832 2442 De GE eS RE Figure 4 Signal to Noise Ratio with HPLC Grade Solvents lol ll Rx Rol AIS l Bel lal kCount pesimsng 5fg XMS TIC Filtered 7 609 3 gt 380 460 1 42V SIN PP 3398 S 3398 N 1 Seg 1 Time 0 00 3 69 reserpine Scan Functions 1 141 211 261 351 421 491 TL brs wep Scans Figure 5 Signal to Noise Ratio with LC MS Grade Solvents The supply kit that
20. pened Systems are shipped either FOB Agilent or FOB Destination The manner of shipment determines who is responsible for filing a claim against the carrier if the system was damaged in transit Most systems are shipped FOB Agilent so any damages incurred in shipment are the responsibility of the purchaser and the carrier Contact your local office for assistance with filing claims and billing repairs If the system ships FOB Destination contact the your local office Agilent will not accept liability for damage if obviously received damaged materials were received without noting the damage on the receiving documents Check the checklist box Shipping cartons inspected for damage If there was any damage the conditions were reported Unpacking and Installation The Agilent Representative will review the Checklist with the customer to ensure that the site requirements were met The Agilent Representative will unpack and install the instrument and demonstrate the fundamental operation and maintenance procedures The User Representative must be present during the installation The Agilent Representative will demonstrate that the system meets the performance specifications which are on the data sheet and any additional criteria explicitly written into the sales contract Agilent 500 lon Trap LC MS System MS Site Preparation Guide Site Preparation Plan to analyze samples only after the installation is complete and the conditions of d
21. rmance of hardware and software not provided by Agilent Technologies Items that do not include installation by Agilent Technologies must be installed by the customer 6 Agilent 500 lon Trap LC MS System MS Site Preparation Guide Site Preparation e Testing the 500 Ion Trap LC MS system using customer standards or samples e Detailed instruction in the operation of the computer operating system and Workstation software Contact Agilent Technologies for information concerning training classes Setting up laboratory procedures Assistance with laboratory procedures can be obtained from an Agilent Applications Engineer AE on a consulting basis at additional cost Contact Agilent Technologies for information concerning assistance with laboratory procedures application development or chemical analysis consulting e Operation of the 500 Ion Trap LC MS system following installation Space and Weight Requirements Entrance Before arranging for delivery of the LC MS determine that there is sufficient clearance to move the shipping cartons to the installation site The largest shipping carton is 72 cm 28 in wide by 92 cm 36 in long If the cartons are to be moved with the pallet at least 92 cm 36 in clearance is needed Allow additional room for maneuvering the shipping containers around corners or through doors Check the checklist box Entrance to the installation site is at least 92 cm 36 in wide The MS foreline