        Ettan™ DALTsix and Ettan DALTtwelve Systems
         Contents
1.     Programming Ettan DALTsix    Turn on the power supply control unit according to the user documentation provided  with the equipment     Refer also to the user documentation for programming instructions   For recommended electrophoresis conditions  see Electrophoresis conditions with  precast gels  on page 34     Programming Ettan DALTtwelve    If not already turned on  turn on the power supply control unit by pressing the Power  switch to the I position     When the unit is turned on  the default settings of constant power  continuous run  auto  pump  and 25   C are shown on the display  see Figure 4 2     The control unit has four programmable parameters  Run Type  Timer  Pump  and  Temperature  see Figure 4 2     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 31    4 Operation    4 2    32    Preparations before start    1 Const Watt 6W  2 Time 1 1 00hrs  3 Pump Auto  4 Temperature 22  C  5  6 6        7 8          Description Item Description  1 Run type 5 HV ON indicator  2 Timer 6 Arrow buttons  3 Pump 7 set read button  4 Temperature 8 start stop button                      Figure 4 2  Controller interface in programming mode     The set read button toggles the controller between the set and read modes  The start   stop button starts and stops the execution of the programmed electrophoresis run  The   lt  and P gt  buttons move the cursor between run parameters  and the A and W buttons  change the settings of the parameters     Below
2.   4 Operation  Preparations before start    4 2    Note  The gel face of the strip must not touch the opposite glass plate        Melt an aliquot of agarose sealing solution  see Appendix B 4  in a heating block or  boiling water bath for each Immobiline DryStrips  Allow the agarose to cool slightly  and slowly pipette the molten agarose solution  along the upper surface of the gel   up to the top of the glass plate  Take care not to introduce bubbles  Do not allow  the agarose to solidify     Once the agarose solution has completely set the gel should be run in the second  dimension as soon as practically possible     Inserting gels into Ettan DALT  electrophoresis units    30          WARNING  Disconnect the unit from the power supply when inserting gels           Inserting gels into Ettan DALTsix    When the electrophoresis buffer has reached the desired temperature  insert the  loaded Gel Cassettes with the Immobiline DryStrip Gels in place     1    Wet the UBC sealings with cathode buffer or 0 1  SDS  immerse the sealings in  solution or spray the sealings of the UBC using a plant sprayer  and carefully slide  the UBC over the gel cassettes     Note  Do not move the UBC repeatedly up and down as this will reduce the sealing  effect     In a separate container  add 2 x 125 ml  2 bottles  of cathode buffer  Rinse the  bottles and fill up with distilled or deionized water to 1 2       Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Operati
3.   full range of proteins        Vertical protein streaks     IPG strip not properly placed  on gel surface     Make sure IPG strip uniformly  contacts the gel surface along  itsentire length  Avoid gouging the  surface of the separating gel        Spots skewed or  distorted     Gels run too fast   uneven  migration     Run at a lower power setting  Use a  two step program  Start at a low  power setting until the proteins enter  the gel  then increase the power for  the remainder of the run        Uneven gel surface     Overlay the running gel with water   saturated butanol before  polymerization begins to avoid  forming an uneven gel surface     Uneven gel polymerization or  gradient formation        Heavy background after  silver staining              Use reagents of the highest purity   preferably electrophoresis grade     Use deionized  double distilled water           48    Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Symptom    Distortion in the 2 D  pattern     Possible cause    Bubbles or liquid between the  gel and the glass plate     Troubleshooting 6  Stained gels 6 3    Corrective action    Use the roller to remove any bubbles  or excess liquid between the gel and  the glass plate  Ensure that no visible  bubbles remain and that the gel  adheres firmly to the glass and  resists movement        Interfering substances in the  first dimension     Contaminants in the sample can  cause distortions or swollen regions  in the IPG stri
4.  DALTtwelve Systems are intended for research use only  and  shall not be used in any clinical procedures  or for diagnostic purposes     Safety notices    These Operating Instructions contain WARNINGS  CAUTIONS and NOTICES concerning  the use of the product  with meanings as defined below        WARNING   WARNING indicates a hazardous situation which  if not avoided  could  result in death or serious injury  It is important not to proceed until all  stated conditions are met and clearly understood           CAUTION   CAUTION indicates a hazardous situation which  if not avoided  could  result in minor or moderate injury  It is important not to proceed until all  stated conditions are met and clearly understood     NOTICE     NOTICE indicates instructions that must be followed to avoid damage to       the product or other equipment              6 Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Introduction 1  Regulatory information 1 2    Notes and tips    Note  A Note is used to indicate information that is important for trouble free and  optimal use of the product     Tip  A tip contains useful information that can improve or optimize your procedures     Typographical conventions    Software texts and commands are identified by bold italic text  A colon is used to  separate menu levels  e g  File  Open refers to the Open option in the File menu      1 2 Regulatory information    This section lists the directives and standards that are fulf
5.  XXXXXXXX Voltage    Serial No XXXXXXX Frequency    Mfg Year  2009 Power max   Fuse     N12406    A A D  C        TNT    751 84 Uppsala Sweden    2895233600       Made in Sweden GE Healthcare Bio Sciences AB      Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 19    2 Safety instructions  2 2 Labels    Symbols used in safety labels    Label       Meaning    Warning  Read the user documentation before using the system  Do  not open any covers or replace parts unless specifically stated in the  user documentation        Indicates that hazardous voltages are generated inside the  instrument        The system complies with the requirements for electromagnetic  compliance  EMC  in Australia and New Zealand        CE          The system complies with applicable European directives        Labels concerning hazardous    substances    Label    Meaning             This symbol indicates that the waste of electrical and electronic  equipment must not be disposed as unsorted municipal waste and  must be collected separately  Please contact an authorized  representative of the manufacturer for information concerning the  decommissioning of equipment           20    Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Safety instructions 2  Emergency procedures 2 3    Meaning    This symbol indicates that the product contains hazardous materials  in excess of the limits established by the Chinese standard SJ     T11363 2006 Requi
6.  a homogeneous 12 5  polyacrylamide gel  It is intended to be used in  Ettan DALTsix or Ettan DALTtwelve systems together with the DALT Buffer Kit  see 2 D  Electrophoresis Principles and Methods     DALT Pre cast Gel Cassettes    The DALT Gel 12 5 pre cast gels are inserted into a specially designed reusable cassette   see below  and run vertically in the Ettan DALTsix and Ettan DALTtwelve Systems              Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA       Introduction 1  Instrument 1 3    Gel casting equipment    Gel Caster  The Ettan DALTsix and Ettan DALTtwelve Gel Casters can hold a different number of gel  cassettes with separator sheets  for casting homogenous or gradient gels         Cassette capacity  Cassette capacity     System  1 0 mm gels 1 5 mm gels    Ettan DALTsix Gel Caster    14 13                   Fewer gels can be cast at one time by using blank cassette inserts to occupy unneeded  volume  For more information on gel casting  see Ettan DALTsix and Ettan DALTtwelve    User Manuals     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 13    1  13    14    Introduction  Instrument    Gel Casting Cassettes    The Ettan DALTsix and Ettan DALTtwelve use the same cassettes  The gel casting  cassettes are pre assembled  Two glass plates are joined along one edge by a hinge  strip of silicone rubber  and the vinyl side spacers  1 0 mm or 1 5 mm thick  are glued in  place  To complete assembly  cl
7.  alone unit  or  e connected to other CE marked instruments  or    e connected to other products recommended or described in the user  documentation  and    e used in the same state as it was delivered from GE Healthcare  except for  alterations described in the user documentation or explicitly authorized by  GE Healthcare     Regulatory compliance of  connected equipment    Any equipment connected to Ettan DALTsix and Ettan DALTtwelve Systems should meet  the safety requirements of EN 61010 1 IEC61010 1 or relevant harmonized standards   Within the European Union  connected equipment must be CE marked     8 Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Introduction 1  Instrument 1 3    1 3 Instrument    This chapter contains a general description of Ettan DALTsix and Ettan DALTtwelve  Systems     General description    In 2 D electrophoresis  proteins are separated according to isoelectric point by  isoelectric focusing  most reliably on immobilized pH gradient  IPG  strips using the  IPGphor    IEF System  The second dimension electrophoresis separates the proteins on  the basis of their molecular mass using sodium dodecyl sulphate polyacrylamide gel  electrophoresis  SDS PAGE   The Ettan DALTsix and Ettan DALTtwelve Systems are  designed to handle multiple large second dimension gels in a simple  efficient  and  reproducible manner     A sample  focused in an IPG strip  is placed on the cathodic  upper  surface of a slab gel  and sealed
8.  be heard as air is drawn into the lower chamber  For runs of six or fewer gels   alternating Gel Cassettes with blank cassettes also eases the removal of the Gel  Cassettes at the end of a run  Leave enough of the electrophoresis buffer in the  tank to act as a lubricant between the glass cassettes and the buffer seal     There are two methods for removing the first cassettes from the unit  using the  cassette removal tool or the hands     e Using the cassette removal tool  Carefully insert the cassette removal tool  between the cassette and the buffer seal  with the folded tip facing the  cassette  until the tip is beneath the bottom edge of the cassette  Verify that  the tool is caught on the bottom edge of the cassette  then lift it out slowly with  the tool  Figure 4 3         Figure 4 3  Using the cassette removal tools     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Operation 4  Procedures afterarun 4 4    e Using the hands  By hand  apply upward pressure alternately to each side of  the cassette  gradually shifting it up until you can grasp it and remove it  Figure  4 4         Figure 4 4  Removing cassettes by hand     2 Open the cassettes using a Wonder Wedge and carefully transfer the gels to a  staining tray  for example  Take care to ensure that the gel does not adhere to the  spacers     Note  Vinyl gloves are less sticky than latex gloves and make it easier to handle  large gels     Detection    Scan the gels as soon as po
9.  bowing should not  affect the performance of the seal     To remove the anode plate or any of the other internal components  follow the  directions below     1 Insert the buffer seal removal tools through the fifth or sixth slot  with the wider end  underneath the tubes  Turn the tools 90   and place as close to the end blocks as    42 Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Maintenance 5  Component maintenance 5 3    possible  Carefully pull upward on the buffer seal until it is removed from the unit   see Fig 5 1         Figure 5 1  Using the buffer seal removal tools    2 Slide the two baffle plates upward and out of the unit     3 Remove the two flow guides by pulling out the retaining pins and lift the blocks out     4 Remove the anode plate by unscrewing the flathead nylon screw and lifting the  plate out     5 To reassemble the unit  replace the anode plate and screw  making sure that the  sealing sleeve is in place  Spread a small amount of GelSeal compound on the plug  with a swab applicator to prevent corrosion     6 Replace the flow guides and baffle plates     7 Puta light film of GelSeal compound on the gasket of the buffer seal and reinsert it  into the unit using even pressure  Make sure that it is fully seated before using the  unit     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 43    5 Maintenance  5 3  Component maintenance    44 Ettan DALTsix and Ettan DALTtwelve Systems Operat
10.  can cause fatal electrical shock if the safety features  are disabled  The safety lid must be securely closed before starting a  protocol     WARNING   The instrument covers must not be opened by the user  It contains  electrical circuits which can give a lethal electric shock  Service and  planned maintenance should be performed by personnel authorized by  GE Healthcare        WARNING   Disconnect all units from mains power supply before removing the  safety lid  unloading the gels  draining and cleaning the electrophoresis  unit      gt   gt   PIP Pe             16 Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Safety instructions 2  Safety precautions 2 1    Using flammable liquids    When using flammable liquids with the Ettan DALTsix and Ettan DALTtwelve Systems  system  follow these precautions to avoid any risk of fire or explosion        WARNING  Fire Hazard  Before starting the system  make sure that there is no  leakage           WARNING  A fume hood or similar ventilation system shall be installed when  flammable or noxious substances are used      gt   gt              Personal protection       WARNING   Always use appropriate personal protective equipment during  operation and maintenance of Ettan DALTsix and Ettan DALTtwelve  Systems           WARNING   Hazardous substances  When using hazardous chemical and  biological agents  take all suitable protective measures  such as  wearing protective glasses and gloves resistant to t
11.  in place with agarose  The Gel Cassette is then inserted into the  electrophoresis unit through the buffer seal slots flanked by rubber gaskets  Any unused  slots are filled with the blank cassette inserts  The buffer seal is an effective current and  liquid barrier  For more information  see Chapter 4     System components  The following components are included in Ettan DALTsix and Ettan DALTtwelve Systems     e Electrophoresis unit   e Power Supply Control Unit  e Gel Caster   e Gel Casting Cassettes    e Gradient maker    Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 9    1 Introduction  1 3 Instrument    The Electrophoresis units    Description of Ettan DALTsix Electrophoresis unit                     1  2  3  Part Description Part Description  1 Safety interlock 5 Upper buffer chamber  2 Anode assembly cassette 6 Lower buffer chamber  carrier   3 Buffer circulation pump 7 Ceramic heat exchanger  4 Lid with leads                10 Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA       Introduction 1  Instrument 1 3    Description of Ettan DALTtwelve Electrophoresis unit             5  1   a  2  3   q     7  4             Part Description Part Description  1 Lid with leads  cathode  5 Safety interlock  2 Upper buffer chamber 6 Rubber gasket  3 Lower buffer chamber 7 Ceramic heat exchanger  4 Buffer circulation pump                      Electrophoresis unit data    The Ettan DALTsix and Ettan DALTtwelve can a
12.  is a description of the programmable parameters  see also Table 4 2     e Run Type determines the method of power regulation  The options for run type  are constant power  constant current  constant voltage  and crossover mode  In  crossover mode the voltage and current limits for the run are set manually   instead of using the instrument defaults of 600 V and 1 000 mA  As the run  progresses  the power supply operates in the mode that is limiting     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Operation 4  Preparations before start 4 2    Timer controls the duration of the electrophoresis run  The options are continuous  run  timed step  timed and hold  and volt hours  Up to three timed steps of up to  100 h each can be programmed using the timed step mode  With timed step  mode  all the steps must have the same run type     Pump controls buffer circulation through the separation unit  The options for the  pump are On  Off  and Auto  Auto mode activates the pump only when power is  applied     Temperature controls the cooling or heating of the buffer in the separation unit   The temperature range is from 10  C to 50  C  On the display  cooling is indicated  by    and heating is indicated by 4      The pump must be on to properly cool or  heat buffer in the tank  The lower temperature limit is a function of the ambient  temperature and the power reading  To reach the lower temperature limit  10  C   or in laboratory environments where t
13.  respective owners   Bj  rkgatan 30    2009 General Electric Company   All rights reserved   751 84 Uppsala First published Nov  2009   Swede n All goods and services are sold subject to the terms and conditions of sale    of the company within GE Healthcare which supplies them  A copy of these  terms and conditions is available on request  Contact your local GE  Healthcare representative for the most current information        www gelifesciences com 2DE    GE Healthcare UK Ltd  Amersham Place  Little Chalfont  Buckinghamshire  HP7 9NA  UK    GE Healthcare Bio Sciences Corp  800 Centennial Avenue  P O  Box 1327  Piscataway  NJ 08855 1327  USA    GE Healthcare Europe GmbH  Munzinger Strasse 5  D 79111 Freiburg  Germany    GE Healthcare Japan Corporation  Sanken Bldg  3 25 1  Hyakunincho  Shinjuku ku  Tokyo 169 0073  Japan    imagination at work    28 9640 96 AA 11 2009    
14.  the separation tank with distilled or deionized water is  sufficient  If desired  the internal components of the separation unit can be removed for  a more thorough cleaning  see below   The unit can also be periodically cleaned with a  dilute solution of a mild detergent     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 41    5 Maintenance  5 3  Component maintenance    Clean the gel casting cassettes and pre cast gel cassettes with a dilute solution of a  laboratory cleanser such as RBS 35  from Pierce Chemical Company  Rinse the  cassettes thoroughly with distilled or deionized water        NOTICE  Do not autoclave or heat any part above 50   C           NOTICE  Do not expose the unit or its parts to organic solvents  including  gt 20   ethanol           NOTICE  If using radioactive reagents  decontaminate the unit with a cleaning  agent such as CONTRAD 70 or Decon    90 from Decon Laboratories  Inc        0 0 0          5 3 Component maintenance       WARNING  Electrical Shock Hazard  Do not use the equipment if any part is  cracked or broken      gt              If any of the components become cracked or broken it should be replaced  The upper  buffer chamber  UBC  should be replaced if any of the rubber flaps become cut or torn     To prevent damage to the flaps  ensure that all sharp edges on glass plates are  smoothed and exercise care when using the cassette removal tool     With normal use the UBC ribs may become bowed upward  this
15.  unit is 400 mA  100 W  and  600 V  For overnight runs we recommend to set the temperature to 15  C  Alternatively   set the temperature to 22  C     For further run conditions  see the DIGE Gels and DIGE Buffer Kit Short Instructions in the  DIGE Buffer Kit     Run conditions  Day run      Set the MultiTemp temperature to 22  C        Step mA gel Voltage  V  W gel Time   hours mins    1 10 80 1 1 0   2 50 500 17   4 00   5 00           The maximum rated input power for the electrophoresis unit is 100 W     Continue the electrophoresis until the bromophenol blue reaches the end of the gel     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Operation 4  Preparations before start 4 2    Run Conditions  Overnight    Set the MultiTemp temperature to 15  C     Step mA gel Voltage  V  W gel Time   hours mins   1 12 150 15 15 00   17 00          Continue the electrophoresis until the bromophenol blue reaches the end of the gel     Recommended running conditions for DIGE Gels   Ettan DALTtwelve  Program the Ettan DALTtwelve electrophoresis unit to deliver the following protocol     Constant power    Temperature  22  C       Step Power W gel Duration  1 1 1 hour  2 17  max 180   Until the bromophenol dye front reaches the    bottom of the gel  approximately 4 5 hours for a  full set of 12 gels           For overnight runs  the power is set to 1 W gel and the recommended temperature is 15  C   Alternatively  set the temperature to 22  C  Run for 1 hou
16. 115115112112     6 2 Pre cast GEIS  eeeeeessessssssssssteesssessssscenceeeeececeeeeceeeeeeestteeet  6 3   Stained Jels srriicseipe nesia  7 Reference information             csecceceeeeeees  TI    SPOGCIIGATIONS sssscccscersasscxscensccsccascacszssszescattassascsauscncsteece  TZ  LALOR Essin  7 3 Ordering INFOMATION      eeesssssssseeeseseststtttneeenenneeseeee    Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA                Table of Contents    Appendix A_ Electrical and communication connections      53    A 1 Front of Ettan DALTtwelve Power Supply Control unit uu    53  A 2 Rear of Ettan DALTtwelve Power Supply Control Unit wes 54  Appendix B RecipeS          ssssesssssssesssssressssssressssssesseresssessssseeesesse 55  B 1 SDS equilibration buffer solution           sss11s11ssisss1ssrssrrrrssrisnrsssrssrrenrns 55    B 2 4x resolving gel buffer solution  B 3 SDS electrophoresis buffer oo ceeccccccsssssssssssseesssessccssssssssssseesessesscsessseess  B 4 Agarose SCALING SOLUTION oo  cecssssssssssssssssssssseesssesscssssssssssssesessessesessseess       4 Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Introduction 1  Important user information 1 1    1 Introduction    Purpose of the Operating  Instructions    The Operating Instructions provide you with the instructions needed to handle Ettan  DALTsix and Ettan DALTtwelve Systems in a safe way     Prerequisites    In order to operate Ettan DALTsix and Ettan D
17. 2  C for a day run and to 15  C for an overnight run  Equilibrate the  buffer to 15  C before starting an overnight run  Refer to the MultiTemp Ill User  Manual for more information     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Operation 4  Preparations before start 4 2    4 Set aside upper chamber  The upper chamber is prepared once the gel has been inserted into Ettan DALTsix   See Inserting gels into Ettan DALT electrophoresis units  on page 30     For more detailed instructions  refer to Ettan DALTsix Electrophoresis System User  Manual     Preparing Ettan DALTtwelve    1 Prepare cathode buffer  In a separate container dilute the concentrated cathode buffer included in the  DIGE Buffer Kit to working strength by adding four bottles of DIGE cathode buffer   total volume 500 ml  and fill up with distilled or deionized water to a total volume  of 2 25       2 Prepare anode buffer  Ensure that the valve on the separation unit is set to    circulate     Add the entire  contents of two bottles of DIGE anode buffer stock solution included in the DIGE  Buffer Kit into the tank  Rinse the bottles with distilled or deionized water and pour  it into the tank  Fill the tank to the 7 5   fill line with distilled or deionized water  in  this way washing the DIGE anode buffer from the buffer seal     Note  Avoid pouring the DIGE buffer on the tubing by spreading the tubing slightly  with one hand while pouring the solution with the other  Figure 
18. 2 kPa  12 PSI            WARNING  Be careful when lifting and moving the electrophoresis unit  It is best to  move the unit when empty      gt  PP PPE             System operation       WARNING   Hazardous biological agents during run  When using hazardous  biological agents  flush the entire system tubing with bacteriostatic  solution followed by a neutral buffer and finally distilled water  before  service and maintenance      gt              18 Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Safety instructions 2  Labels 2 2       WARNING   Avoid spilling liquids on the body of the instrument  If large volumes of  liquid have penetrated the casing of the instrument and come into  contact with the electrical components  immediately switch off the  instrument and contact an authorized service technician              Maintenance       WARNING  Do not remove the main cover  There are no userserviceable  components inside  and you can be exposed to high voltage              2 2 Labels    This section describes the safety labels and labels concerning hazardous substances  that are attached to the Ettan DALTsix and Ettan DALTtwelve Systems  For information  about marking of the computer equipment  refer to the manufacturer   s instructions     Labels on the instrument    The illustration below shows an example of the identification label that is attached to  the Ettan DALTsix and Ettan DALTtwelve Systems     XX XXXK XX Ettan DALTxxxx    Code No 
19. 4 1         Figure 4 1  Spreading the tubing elements apart with one hand while pouring the solution  with the other  to avoid pouring the DIGE buffer onto the tubing      3 Turnon the electrophoresis unit  Turn on the electrophoresis unit by pressing the Power switch to the I position     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 27    4 Operation    4 2 Preparations before start    4 Start the pump  Start the pump to mix the buffers and set the separation unit to desired  temperature  A temperature of 22  C is recommended for day runs  and 15  C for  overnight runs     Note  A temperature of 25  C is recommended for electrophoresis     For more detailed instructions  refer to Ettan DALTtwelve System User Manual     Equilibrating Immobiline  DryStrip Gels    28    Equilibration is carried out in a two step process to reduce vertical streaking in the  second dimension  The first step saturates the IPG strip with the SDS system and the  second step blocks the protein thiol groups  The equilibration solution contains buffer   urea  glycerol  reductant  SDS and dye  Tubes and volume of equilibration solution to  use are specified in Table 4 1     Prepare equilibration solution    1 Prepare SDS equilibration buffer  see Appendix B 1   This is a stock solution    2 Just prior to use  add 50 mg DTT per 10 ml SDS equilibration buffer  0 5   w v    Equilibration   1 Place the IPG strips in individual tubes with the support film toward the tube wa
20. 96 AA 25    4 Operation    4 1 Operation overview    4 1 Operation overview    The typical workflow in Ettan DALT system  when using precast gels  can be divided into  the following steps     Step Action Page                      1 Preparing Ettan DALT system for electrophoresis using precast gels   26  3 Equilibrating Immobiline    DryStrip Gels 28  4 Applying equilibrated Immobiline DryStrip Gels 29  5 Inserting gels into Ettan DALT electrophoresis units 30  6 Programming the power supply control unit 31  7 Starting the electrophoresis 37  8 Unloading the gels from the Ettan DALT electrophoresis units 37                Note  Precast DIGE Gels have been used as an example throughout this chapter     4 2 Preparations before start    Preparing Ettan DALT system  for electrophoresis using    precast gels    Preparing Ettan DALTsix    1    26    Prepare anode assembly   Insert the anode assembly cassette carrier into the tank  The anode assembly is  molded so thot it can only be inserted in one orientation  The side edge of the  assembly should fit into the slot in the side of the tank     Fill with anode buffer   Add one bottle  125 ml  of DIGE Anode Buffer stock solution included in the DIGE  Buffer Kit into the tank  Fill the electrophoresis unit to the 4 5   fill line with distilled  or deionized water  Connect the pump to mains power supply     Turn on the temperature controller   Turn on the MultiTemp IIl temperature controller  or similar  and adjust the  temperature to 2
21. ALTtwelve Systems safely and according  to the intended purpose the following prerequisites must be met     e You should be acquainted with the use of general laboratory equipment and with  handling of biological materials     e You should understand the concepts of electrophoresis   e You must read the Safety Instructions in Chapter 2 of these Operating Instructions     e The system should be installed according to the instructions in Chapter 3 of these  Operating Instructions     In this chapter    This chapter contains important user information and a general description of Ettan  DALTsix and Ettan DALTtwelve Systems and their intended use     1 1 Important user information    Read this before using Ettan  DALTsix and Ettan DALTtwelve    Systems    All users must read the Safety Instructions in Chapter 2 of these Operating Instructions  before installing  using or maintaining the instrument     Do not operate Ettan DALTsix and Ettan DALTtwelve Systems in any other way than  described in the user documentation  If you do  you may be exposed to hazards that can  lead to personal injury and you may cause damage to the equipment     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 5    1 Introduction  1 1 Important user information    Intended use    The Ettan DALTsix and Ettan DALTtwelve Systems are designed to handle multiple large  second dimension electrophoresis gels in a simple  efficient  and reproducible manner     Ettan DALTsix and Ettan
22. GE Healthcare    Ettan    DALTsix and  Ettan DALTtwelve Systems    Operating Instructions    Original instructions       Ettan    Table of Contents    T REO CUCU ascites  1 1  Important user Information      eseeesesssssssseceeeeeeeee  1 2 Regulatory information  LS WASUMUMS WG sscessssadeensztcctencpestassictecncastiarencssetectaeonncestints  2 Safety instructionS           seessseessessssessssesesses   2 1 Safety precautions  eeeeeeesssssssssssseeseesesesettttteetenesseesee  Z2  LADES nenn ek    2 3 Emergency procedures       24 Recycling INFOrmation oe eeeesssssssssceeeecesesesettsttteneeneeeeee       Table of Contents       3 MASCOT ON nnana 23    3 1 Site requirements    3 2 Unpacking     3 3 Placement       34 TRONS POM  se csacracseressss sasncsscessasesiateetaniessmeencesensioanvacectaveie  BBs  ASSEDI resen S  3 6 CONNECTIONS      eeeeecssssscssssseeeee    3 7 Spare parts and accessories    4  Operati  M cisdiseissascsaiscessiasessstiasctectisssasesssess  4 1 Operation Overview aoeeeescccssssssssesseccccsssssneeeseceesssnneees    4 2 Preparations before start        4 3 Performing a run             GA Procedures OFF g rUn  arassssiiiiriresiiiis  GS MaimtenaNCE sissies  5 1 User maintenance schedule  52 CIC CIAUIAG secasscassadsssasecsvtsnccsonssceonesccadecsedeatonscsavessanconnsbnnetot  5 3  Component MAINTENANCE     reeeeeeessssessssseeseeeeeeeeeceeee  6 Troubleshooting siisisccsisiccsescsscssssiaissosccardisencs  6 1 Electrical and mechanical       s ss ss15115115
23. Second dimension  separation proceeds    slowly with high current     All of the slots in the buffer seal  are not occupied by either gel  cassettes or blank cassettes     Ensure that all 12 slots in the buffer  seal are occupied  Do not pour more  than the suggested volume  7 5    into  the lower reservoir        Anodic buffer has mixed with  cathodic buffer from overfilling  of either the cathodic or the  anodic reservoir     Ensure that the level of the anode   lower  buffer does not come above  buffer seal when the separation unit  is fully loaded     Remove any excess anode buffer  from the upper reservoir  Ensure that  the level of cathode buffer is not  above the air vents in the upper  reservoir        Dye front is irregular     The top surface of the gel has  been damaged during  application of the IPG strip     Take care during application of the  IPG strip that neither gel is damaged        Bubbles or liquid between the  gel and the glass plate     Use the roller to remove any bubbles  or excess liquid between the gel and  the glass plate  Ensure that no visible  bubbles remain and that the gel  adheres firmly to the glass and  resists movement        Interfering substances in the  first dimension     Contaminants in the sample can  cause distortions or swollen regions  in the IPG strip following IEF  Modify  sample preparation to limit these  contaminants  Refer to 2 D  Electrophoresis  Principles and  Methods        Pronounced downward  curving of the dye front  
24. ccommodate a different number of gels   either 1 mm or 1 5 mm thick  in a common tank under identical conditions           Electrophoresis system Gel size Gel capacity  Ettan DALTsix 25 5 x 20 5 cm 6  Ettan DALTtwelve 25 5 x 20 5 cm 12    Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 11    1  13    Introduction    Instrument    Power supply control unit    Electrophoresis system Power supply unit Requirements    Ettan DALTsix EPS 601 Power Supply or 600 V  400 mA  or 100 W  any other power supply  with similar specifications       Ettan DALTtwelve Ettan DALTtwelve Power 600 V  1A  or 200 W  Supply Control Unit                Pre cast Gels and Gel cassettes    12    Pre cast gels    If fluorescent staining labeling techniques will be used  do not run gels cast on plastic  backing  as it can pose a problem of high background with some dyes during  subsequent analysis  Instead  a DIGE Gel can be used  A DIGE Gel is a 12 5  precast  polyacrylamide gel in a low fluorescent glass cassette for second dimension 2 D  electrophoresis and specially developed for 2 D DIGE analysis  It is intended to be used  in Ettan DALTsix or Ettan DALTtwelve systems together with the DIGE Buffer Kit  For more  information on DIGE Gels  see Ettan DIGE System User Manual     DALT Gel 12 5 is a precast polyacrylamide gel  25 5 x 19 6 cm  1 mm thick  for the second  dimension of 2 D electrophoresis  The gel is provided already cast onto a plastic support  film  The gel is
25. cription  1 High voltage connector 2 Signal and power cable  connector       Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 53    Appendix A Electrical and communication connections  A 2 Rear of Ettan DALTtwelve Power Supply Control unit    A 2 Rear of Ettan DALTtwelve Power Supply Control unit    Part Description Part Description       1 Power switch 2 Power inlet       54 Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Recipes Appendix B    SDS equilibration buffer solution B 1    Appendix B Recipes    B 1 SDS equilibration buffer solution     6 M urea  50 mM Tris HCl pH 8 8  30  glycerol  2  SDS  0 001  bromophenol blue        200 ml     Final concentration Amount  Urea  FW 60 06  6M 72 07 9  Tris HCl  pH 8 8  see Appendix B 2  50 mM 6 7 ml  Glycerol  87  w w  30   v v  69 ml  84 2 g   SDS  FW 288 38  2   w v  40g  1  Bromophenol blue stock solution 0 001   w v  2 mg  Double distilled water     to 200 ml       1 This is a stock solution  Just prior to use  add DTT or iodoacetamide  for first or second  equilibration  respectively  as described in the in Equilibrating Immobiline DryStrip Gels  on    page 28     Store in 40 ml aliquots at  20   C     B 2 4x resolving gel buffer solution     1 5 M Tris base  pH 8 8  1          Final concentration Amount  Tris base  FW 121 1  1 5 M 181 7 g  Double distilled water     750 ml    HCigg        Double distilled water        adjust to pH 8 8    to1l       Ettan DALT
26. e  placed on a stable support such as a bench or sturdy table     A length of rubber or vinyl tubing sufficiently long to reach a sink should be attached to  the drain port on the back of the unit before use  The unit should not be placed on bench  paper or any other material that might be pulled in by the air intake fans  as any  hindrance to air circulation will reduce the cooling capacity     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 23    3 Installation  3 4 Transport    For Ettan DALTsix  The tubing leading to and from the heat exchanger should be  connected to a circulating water bath such as the MultiTemp III  the heat exchanger  should not be connected to a water tap or any other coolant supply that lacks pressure  regulation     The Power Supply should be placed conveniently near the electrophoresis unit in a dry  place without risk for liquid exposure    3 4 Transport  Before moving the system     e remove liquids and gels from the separation unit  e disconnect all cables    e lift the instrument by the base unit    3 5 Assembly    The following parts must be added to the Ettan DALTsix and Ettan DALTtwelve  instruments before they can be used     e IPG strips  e Gel Cassettes containing gels    e Buffer liquids    3 6 Connections    Communication    Connect Ettan DALTtwelve according to the electrical drawings in Appendix A  Connect  Ettan DALTsix according to the user documentation supplied with the Power Supply     Electrical 
27. etting        Problems with first dimension     Refer to troubleshooting guides for  IPGphor    or Multiphor    units  or 2 D  Electrophoresis  Principles and  Methods        Smeared or comet  shaped spots  Ettan  DALTsix only            Buffer problems        Check pH of cathode buffer  Should  be between 8 3 and 8 8     Make sure that 2x Laemmli buffer is  used in the upper  cathode  chamber     Buffer or SDS depleted           Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    47    6 Troubleshooting  6 3 Stained gels        Symptom                Protein spots are near the  buffer front        Protein spots have not  entered the gel when  buffer front has reached  the bottom of the gel           Possible cause    Incorrect pore size     Corrective action          Pore size of the gel is too large   Increase the  T        Degraded proteins     during sample preparation     Proteins degraded during sample  preparation  Add protease inhibitors          Incorrect pH     Incorrect pore size     Incorrect pH     The gel pore size is too small     Check the pH of the 4x gel buffer  It  should be pH 8 8  Proteins will  migrate faster below pH 8 8            Decrease the  T     Check the pH of the 4x gel buffer  It  should be pH 8 8  Proteins will  migrate more slowly above pH 8 8        Protein spots are at both  extremes but not in  center     The molecular weight range of the    sample requires an acrylamide  concentration gradient to resolve the
28. he ambient temperature is elevated  it may  be necessary to place the unit in a cold box or cold room     Table 4 2  Parameter options                                Parameter Option Display Range   Run Type Constant power Const Watt 1 to 200 W  Constant current Const Amp 1 to 1000 mA  Constant voltage   Const Volt 1 to 600 V  Crossover mode V  amp  mA 1 to 600 V  1 to 1000 mA   Timer Timed step Time 1   Time 3 00 01 to 99 59 hrs  Time and hold T Hold 00 01 to 99 59 hrs  Volt hours Stop at 1 to 9999 Vhr  Continous run Continous run      Pump On On    off off    Auto Auto     Temperature     Temperature 10 to 50  C                      Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    33    4 Operation    4 2 Preparations before start    See Table 4 3 for a sample program  Twelve gels are electrophoresed at 180 W   constant power  and 25   C with a 5 W gel entry phase  The second phase is extended  longer than required to ensure that the dye front runs off the gel        Step Parameter Value   1 Const Watt 60 W  Time 1 00 45 hrs  Pump Auto  Temperature 25  C   2 Const Watt 180 W  Time 2 04 00 hrs  Pump Auto  Temperature 25  C       Table 4 3  Sample program    For recommended electrophoresis conditions  see Electrophoresis conditions with  precast gels  on page 34     Electrophoresis conditions with    precast gels    34    Recommended running conditions for DIGE gels   Ettan DALTsix    The maximum rated electrical input for the electrophoresis
29. he substances used   Follow local and or national regulations for safe operation and  maintenance of the system s           WARNING   Spread of biological agents  The operator has to take all necessary  actions to avoid spreading hazardous biological agents in the vicinity of  the instrument  The facility should comply with the national code of  practice for biosafety      gt  PP             Installing and moving the  instrument       WARNING   Supply voltage  Make sure that the supply voltage at the wall outlet  corresponds to the marking on the instrument  before connecting the  power cord              Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 17    2 Safety instructions  2 1 Safety precautions       WARNING  Protective ground  The Ettan DALTsix and Ettan DALTtwelve Systems  must always be connected to a grounded power outlet           WARNING  Power cord  Only use power cords delivered or approved by  GE Healthcare        WARNING   Access to power switch and power cord  Do not block the rear and  side panel of the instrument  The Power switch must always be easy to  access  The power cord must always be easy to disconnect        WARNING  Use EPS 601 as power supply for Ettan DALTsix           WARNING   Only connect MultiTemp Ill  or a circulating water bath with earthed  water pipe and low pressure water circulation  Do not connect the heat  exchanger to a water tap or any other water source with unregulated  pressure that may exceed 8
30. ifications 7 1    Parameter Value    Ingression protection    IP20       Supply voltage    100 to 240 V    50 to 60 Hz       Power consumption     Ettan DALTsix  Ettan DALTtwelve    Fuse specifications    100 W  650 W    T2 5 A 250V       Dimensions  H x W x D     160 x 278 x 471 mm             Weight  8 3 kg   Ettan DALTsix 11 1 kg  Ettan DALTtwelve 24 5 kg  Ambient temperature  4 to  40   C  Relative humidity 0 to 90   tolerance       Atmospheric pressure    Acoustic noise level          84 to 106 kPa  840 to 1060 mbar    lt 70dBA          Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    51    7 Reference information  7 2 Literature    7 2 Literature    For further information regarding the Ettan DALTsix and Ettan DALTtwelve Systems  refer  to the following     e Ettan DALTsix Electrophoresis System User Manual  e Ettan DALTtwelve System User Manual   e Ettan DIGE System User Manual   e 2 D Electrophoresis  Principles and Methods   e DIGE Gels and DIGE Buffer Kit Short Instructions    e   MultiTemp Ill User Manual    7 3 Ordering information    For ordering information visit www gelifesciences com 2DE     52 Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Electrical and communication connections Appendix A  Front of Ettan DALTtwelve Power Supply Control unit A 1    Appendix A Electrical and  communication connections    A 1 Front of Ettan DALTtwelve Power Supply Control unit       Part Description Part Des
31. illed by Ettan DALTsix and  Ettan DALTtwelve Systems     Manufacturing information          Requirement Content   Name and address of manufacturer GE Healthcare Bio Sciences AB   Bj  rkgatan 30  SE 751 84 Uppsala  Sweden   Name and ID of notified body INTERTEK SEMKO AB  NB 0413   Place and date of declaration Uppsala  Sweden  May 2009   Identity of person authorized to sign DoC   See EC Declaration of Conformity                CE Conformity    Directive Title             2006 42 EC Machinery Directive  MD   2006 95 EC Low Voltage Directive  LVD   2004 108 EC ElectroMagnetic Compatibility  EMC  Directive             Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 7    1 Introduction  1 2 Regulatory information    International standards             Standard Description Notes   EN 61010 1  Safety requirements for electrical   IEC 61010 1  equipment for measurement  control   CAN CSA C22 2 and laboratory use   No  61010 1   EN 61326 1 EMC emissions and immunity Harmonized with  requirements for electrical equipment 2004 108 EC  measurement  control and laboratory  use   EN ISO 12100 1  Safety of machinery   Basic concepts  Harmonized with   12100 2 general principles and design 2006 42 EC   EN ISO 14121 1  Safety of machinery   Principles of risk Harmonized with   14121 2 assessment 2006 42 EC                   CE marking    The CE marking and the corresponding Declaration of Conformity is valid for the  instrument when it is     e used as a stand
32. ing Instructions 28 9640 96 AA    6 Troubleshooting    6 1 Electrical and mechanical    Symptom    No current at start of run     Possible cause    Insufficient volume of buffer in  upper reservoir     Troubleshooting 6  Electrical and mechanical 6 1    Corrective action    Ensure that the unit contains enough  buffer to contact the upper electrode        Buffer not circulating     Pump is not primed     Turn circulation valve to Drain to fill  with buffer then back to circulate        Pump is off or set to Auto     Set unit to ON on control unit        Pump is broken     Contact GE Healthcare Service        Display on the control unit  blank  Ettan DALTtwelve  only      Unit is not turned on     Turn unit on with power switch        Unit is not plugged in     Display is broken     Plug in unit     Contact GE Healthcare Service        Control unit display  malfunction     Turn the AC power switch off for a  few seconds  then on again  If the  problem persists  call GE Healthcare  Service        Open circuit warning   Ettan DALTtwelve only            Safety interlocks not engaged        Make sure that the lid is completely  closed    Make sure that the pump valve is  turned completely to circulate    If this does not correct the error   contact GE Healthcare Service           Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    45    6 Troubleshooting  6 2  Pre cast gels    6 2 Pre cast gels    Symptom    Possible cause    Corrective action       
33. irculate mode  empty again  and repeat if necessary     Remove the lid from the unit by sliding it to the left and rinse it with distilled or  deionized water  Slide the lid back on its hinges before using the unit again     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Maintenance 5  User maintenance schedule 5 1    5 Maintenance    WARNING  Disconnect power  Always disconnect power from the instrument  before performing any maintenance task              WARNING   Hazardous substances  When using hazardous chemical and  biological agents  take all suitable protective measures  such as  wearing protective glasses and gloves resistant to the substances used   Follow local and or national regulations for safe operation and  maintenance of the system           5 1 User maintenance schedule    Before each run check the following     All parts are clean    All parts are intact and free of cracks and other damage  If a part is damaged  replace the part or entire instrument before use     The power cord and high voltage cable is intact and free of cracks and other  damage  If the cable is damaged replace it before use     There is no leakage from parts containing buffer or cooling water  Moist or liquid  must be removed from the equipment before use     The circulation pump spins and that there is circulation of liquid     5 2 Cleaning    For day to day operation of the unit  the cleaning procedure outlined in unit  operation   thoroughly rinsing
34. ll   2 Add 10 ml DTT containing solution to each tube    3 Place the tubes on a rocker and equilibrate for 15 min     Note  When using CyDye    DIGE saturation dyes  Labeling Kit for scarce samples    repeat the first equilibration with DTT containing SDS equilibration solution  for another 15 min     Second equilibration  Asecond equilibration is performed with an iodoacetamide solution  instead of DTT      1 Prepare a solution of 450 mg iodoacetamide per 10 ml of SDS equilibration buffer   4 5   w v      2 Decant the first equilibration solution and add the same volume of iodoacetamide  containing equilibration solution to each tube     3 Place the tubes on a rocker and equilibrate for an additional 15 min     Note  The subsequent steps of electrophoresis unit preparation  insertion of gels  and melting of the Sealing Solution can be performed while the IPG strips are  equilibrating     Note  Be consistent with the timing of the equilibration steps     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Operation 4  Preparations before start 4 2    Table 4 1  Suggested containers and volumes of equilibration solution     Strip length Container Equilibration   cm  solution  ml   7 Disposable  15 ml conical tubes 2 5 5   11 25 x 200 mm screw cap culture tubes 5 10   13 25 x 200 mm screw cap culture tubes 5 10   18 25 x 200 mm screw cap culture tubes  Equilibration 10 15    24    tubes available from GE Healthcare  or Petri dish    Equilibration 
35. ll the upper buffer chamber upward using the handles at each end     As the upper buffer chamber is pulled upward make sure that the cassettes remain  in the anode assembly     Take out the anode assembly   Remove the Gel Cassettes from the anode assembly     Open the Gel Cassettes using a Wonder Wedge to separate the plates on the side  opposite the hinge     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 37    4 Operation    4 4 Procedures after a run    38    For lab cast gels  run an edge of the Wonder Wedge down each side of the  cassette along the spacer and carefully lift the gel out of the cassette  For precast  gels  lift the gel out by grasping the GelBond backing     Tilt the entire electrophoresis unit on the side of the unit  do NOT tilt to front or back   to empty the buffer from the lower buffer chamber into a sink or a waste container   See Section 3 3 for instructions of how to place the unit for safe use     Rinse all the components with distilled or deionized water  Flush the pump with  distilled or deionized water by filling the unit and turning the pump on     Unloading the gels in Ettan DALTtwelve    1    After the run has been completed  remove one or more of the Blank Cassette  Inserts or gels and drain enough of the electrophoresis buffer from the tank to  expose 2 3 cm of the Gel Cassettes  This will ease removing the remaining  cassettes  When the first cassette  either blank or gel  is removed  a sucking sound  will
36. on 4  Preparations before start 4 2    3 Fill the UBC with 1 2 liters of diluted cathode buffer and use a funnel to adjust the  buffer level in the lower buffer chamber to the same height as in the UBC  by adding  water or diluted anode buffer     4 Attach and close the lid  Connect the power leads to the power supply     Inserting gels into Ettan DALTtwelve  When the electrophoresis buffer has reached the desired temperature  insert the  loaded Gel Cassettes with the Immobiline DryStrip Gels in place     Note  Gel Cassettes and Blank Cassette Inserts slide much more easily into the unit if  they are wet  Wetting the cassette with some cathode buffer using a soaked  kleenex or alternatively distilled or deionized water from a squirt bottle can be  used to wet the cassettes and Blank Cassette Inserts as they are being loaded into  the unit     1 Fit Blank Cassette Inserts into any unoccupied slots   2 Load the unit from back to front     3 Whenall 12 slots are occupied  the buffer level should be slightly below the level of  the gaskets  If the buffer level is too low  add distilled or deionized water to the  lower buffer chamber  If excess anode buffer is in the upper reservoir  remove it  with a pipette  Pour the diluted cathode buffer into the tank to the fill line  some of  this buffer may drip through the gasket and mix with the anode buffer during the  run  but this will not affect performance or results      4 Close the lid     Programming the power  supply control unit
37. on one side of the gel        There is an unfilled gap  between the gel and one of the  spacers        When sealing the IPG strip into place   ensure that some of the agarose  sealing solution flows down any gap  that may exist between the gel and  the spacer        46    Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA       6 3 Stained gels    Symptom    Protein spots are diffuse  or broader than usual     Possible cause    Poor quality reagents     Troubleshooting 6  Stained gels 6 3    Corrective action    Use only the highest quality  reagents        Incomplete polymerization     Make sure that polymerization is  complete        Incorrect equilibration time     Check equilibration time of IPG strips   Too long can lead to diffusion  and  too short can lead to incomplete  equilibration        Gel surface damaged by IPG  strip     Problems with first dimension     Make sure the IPG strip rests on the  slab gel surface without damaging     Refer to troubleshooting guides for  IPGphor    or Multiphor    units  or 2 D  Electrophoresis  Principles and  Methods        Protein spots are poorly  resolved     Incomplete polymerization     Make sure that polymerization is  complete        Diffusion of proteins     Begin electrophoresis as soon as the  IPG strips are loaded  to prevent  diffusion of low molecular  weight  proteins        Running too fast     Reduce the power  current  or  voltage        Too high temperature     Reduce the temperature s
38. ose the two plates like a book and press the plates  together over the length of the spacer  Gels are removed by opening the book after the  run and carefully lifting out the gel slab     Gradient maker    A Gradient maker is used to make gradient gels using two gel solutions of different  acrylamide concentrations and densities     a light solution and a heavy solution  Below  is a description of Ettan DALTtwelve Gradient maker                                                                       6  T  1 g Pl  IG Pl 5   1  2  in ee  m   4  3  Part Description Part Description  1 Feed tube 4 Gradient maker  2 Gel caster 5 Heavy chamber  3 Pump 6 Light chamber                   For more information on how to cast gradient gels  see Ettan DALTsix and Ettan  DALTtwelve User Manuals  For information on how to use the pump  refer to the user  documentation supplied with the pump     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Safety instructions 2  Safety precautions 2 1    2 Safety instructions    This chapter describes safety compliance  safety labels  general safety precautions   emergency procedures  power failure and recycling of Ettan DALTsix and Ettan  DALTtwelve Systems     2 1 Safety precautions    Introduction    The Ettan DALTsix and Ettan DALTtwelve Systems are powered by mains voltage and  handles liquids that may be hazardous  Before installing  operating or maintaining the  system  you must be aware of the hazards described in 
39. p following IEF  These  distortions can result in disturbances  in the seconddimension        Vertical gap in the 2 D  pattern     Vertical streaking     Spots are vertically  doubled  or    twinned        Bubble between IPG strip and  top surface of second  dimension gel     Incorrectly prepared  equilibration solution     Poor transfer of protein from  IPG strip to second dimension    gel     Ensure that no bubbles are trapped  between the IPG strip and the top  surface of the second dimension gel     Prepare equilibration solution  according to instructions     Use low power for sample entry  phase  Extend entry phase if  necessary        Insufficient equilibration     IPG strip is not placed properly     Prolong equilibration time     Ensure that the plastic backing of the  IPG strip is against the glass plate of  the second dimension cassette        Poor representation of  higher molecular weight  proteins        Incorrectly prepared  equilibration solution     Prepare equilibration solution  according to instructions           Poor transfer of protein from  IPG strip to second dimension    gel        Use low power for sample entry  phase  Extend entry phase if  necessary           Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    49    6 Troubleshooting  6 3 Stained gels    50 Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    7 1 Specifications    Reference information 7    7 Reference information    Spec
40. power  Connect the power cord to a grounded power outlet specified in Section 3 1     3 7 Spare parts and accessories    For correct up to date information on spare parts and accessories visit   www gelifesciences com 2DE    24 Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Operation 4    4 Operation    WARNING   Check electrical cables for damage and the electrophoresis unit for  damage or leakage before use  If damage and or leakage is present  do  not use the equipment  There is a risk for high voltage injuries           WARNING   The high voltage power supply must always be disconnected when the  safety lid of the electrophoresis unit is taken off  The high voltage power  supply must never be switched on unless the safety lid is on the  electrophoresis unit           WARNING   Do not overfill the tank  Spillage can cause short circuiting of internal  circuits and the breakdown of electrical isolation which increases the  risk of electrical shock  Always make sure the equipment is  disconnected from power supply when filling the tank           WARNING  Any liquid on the equipment must be dried off before connecting the  power supply           WARNING   Always wear gloves  protective clothing  and eye protection when  handling DIGE Gels  DIGE Buffer  IPG strips  gel cassettes or any other  equipment these items will come into contact with     PP Pe  gt              Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 
41. r and then increase to 1 5 W gel     Recommended running conditions for DALT Precast Gels and 1 mm Laemmli gels    Ettan DALTsix    Run conditions  Day run      Applicable to 1 mm thick 12  PAA gel and Laemmli Buffer system  Set the MultiTemp  temperature to 25  C        Step mA gel Voltage  V  W gel Time   hours mins    1 10   80 1 1 00   2 40   500 13 4 30   6 00           For 1 5 mm gels  increase the current by 50    Continue the electrophoresis until the bromophenol blue reaches the end of the gel     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 35    4 Operation    4 2 Preparations before start    36    Run Conditions  Overnight    Set the MultiTemp temperature to 30  C        Step mA gel Voltage  V  W gel Time  hours   1 10 80 1 1  2 12   150 2 15 17        1 For 1 5 mm gels  increase the current by 50      Continue the electrophoresis until the bromophenol blue reaches the end of the gel     Recommended running conditions for DALT Precast Gels and 1 mm Laemmli gels    Ettan DALTtwelve    Electrophoresis is performed at constant power in two steps  Stop electrophoresis when  the dye front is approximately 1 mm from the bottom of the gel     Temperature control improves gel to gel reproducibility  especially if the ambient  temperature of the laboratory fluctuates significantly     For best results  gels should be run at 25  C  For overnight runs we recommend to set the  temperature to 30  C        Step Power Approximate run   W gel  d
42. rements for Concentration Limits for Certain  Hazardous Substances in Electronics        2 3 Emergency procedures    This section describes how to do an emergency shutdown of Ettan DALTsix and Ettan  DALTtwelve  The section also describes the result in the event of power failure     Emergency procedures    In an emergency situation  do as follows to stop the run     Step Action       1 Switch off power to the instrument by pressing the Power switch on the  external power supply to the 0 position  If required  disconnect the mains  power cord  The run is interrupted immediately     2 For Ettan DALTsix  If required  disconnect the Ettan DALTsix mains power  cord     Power failure  The result of a power failure depends on which unit that is affected     Power failure to    will result in          Ettan DALTsix and Ettan   The run is interrupted immediately  in an undefined  DALTtwelve state       Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 21    2 Safety instructions  2 4 Recycling information    2 4 Recycling information    The equipment shall be decontaminated before decommissioning and all local  regulations shall be followed with regard to scrapping of the equipment     Disposal  general instructions    When taking Ettan DALTsix and Ettan DALTtwelve out of service  the different materials  must be separated and recycled according to national and local environmental  regulations     Recycling of hazardous  substances    Ettan DALTsix and Et
43. six and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    55    Appendix B Recipes  B 3 SDS electrophoresis buffer    B 3 SDS electrophoresis buffer     25 mM Tris base  192 mM glycine  0 2  SDS  20 I      Final concentration Amount  Tris base  MW 121 14  25 mM 60 5g  Glycine  MW 75 07  192 mM 288 0 g  SDS  MW 288 38  0 2   w v  40 0g  Double distilled water     to 101       1 The pH of this solution should not be adjusted     Store at room temperature     B 4 Agarose sealing solution     25 mM Tris base  192 mM glycine  0 1  SDS  0 5  agarose  0 002  bromophenol blue     100 ml    Final concentration Amount  Laemmli SDS electrophoresis buffer 100 ml   see Appendix B 3   Agarose  NA or M  0 5  0 5g  1  Bromophenol blue stock solution 0 002   w v  200 ul       Add all ingredients into a 500 ml Erlenmeyer flask  Swirl to disperse  Heat in a  microwave oven on low or on a heating stirrer until the agarose is completely dissolved   Do not allow the solution to boil over  Dispense 1 5 ml aliquots into screw cap tubes and  store at room temperature     56 Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    For local office contact info rmation  visit GE  imagination at work and GE monogram are trademarks of General    fi   Electric Company   www gelifesciences com contact 7  CyDye  Ettan  Immobiline and Typhoon are trademarks of GE Healthcare  companies   GE Healthcare Bio Sciences AB All third party trademarks are the property of their
44. ssible after the second dimension SDS PAGE is finished in  order to minimize protein diffusion  Typhoon    Variable Mode Imager or Ettan DIGE  Imager are recommended for scanning DIGE second dimension SDS PAGE gels  Store  the gels in a refrigerator in a closed container and keep the gels moist  Allow the gels to  reach room temperature before scanning  Keep the gels in the glass cassettes  throughout scanning     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 39    4 Operation    4 4 Procedures after a run    Cleaning the Ettan DALT units    after a run    40    Cleaning Ettan DALTsix    1    When all of the gels and Blank Cassette Inserts have been removed  tilt the entire  electrophoresis unit on the side of the unit  do not tilt to front or back  to empty the  buffer from the lower buffer chamber into a sink or a waste container  See  Section 3 3 for instructions of how to place the unit for safe use     Rinse all the components with distilled or deionized water  Flush the pump with  distilled or deionized water by filling the unit and turning the pump on     Cleaning Ettan DALTtwelve    1    When all of the gels and Blank Cassette Inserts have been removed  drain the  buffer by turning the pump valve to drain with the pump on  Emptying will take  about 1 min     After the buffer has been removed  pour 3 4   of distilled or deionized water into the  unit and allow it to drain  Rinse the unit with 5 7   of distilled or deionized water in  c
45. tan DALTtwelve contain hazardous substances  Detailed  information is available from your GE Healthcare representative     Disposal of electrical  components    Waste of electrical and electronic equipment must not be disposed as unsorted  municipal waste and must be collected separately  Please contact an authorized  representative of the manufacturer for information concerning the decommissioning of  equipment     22 Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Installation 3  Site requirements 3 1    3 Installation    Ettan DALTsix and Ettan DALTtwelve are delivered in protective packing material and  shall be unpacked with great care     Any equipment connected to Ettan DALTsix and Ettan DALTtwelve must fulfill applicable  standards and local regulations     For detailed information on Installation  see Ettan DALTsix and Ettan DALTtwelve User  Manuals     3 1 Site requirements    Parameter Requirement                      Electrical power 220 240 V  50 60 Hz  Ambient temperature 4  C to 40  C   Placement Stable laboratory bench  Humidity 0 to 90   non condensing    3 2 Unpacking    e Check the equipment for damage before starting assembly and installation   e Document any damage and contact your local GE Healthcare representative     Remove straps and packing material and stand equipment upright before starting  installation     3 3 Placement    Position the electrophoresis unit close to a sink for easy rinsing and draining  It must b
46. this manual  Follow the  instructions provided to avoid personal injury or damage to the equipment     The safety precautions in this section are grouped into the following categories     General precautions   Using flammable liquids   Personal protection   Installing and moving the instrument  System operation    Maintenance    General precautions       WARNING   Do not operate the Ettan DALTsix and Ettan DALTtwelve Systems in any  other way than described in the Ettan DALTsix and Ettan DALTtwelve  Systems manuals              WARNING   Operation and user maintenance of the Ettan DALTsix and Ettan  DALTtwelve Systems should be performed by properly trained personnel  only           Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 15    2 Safety instructions  2 1 Safety precautions       WARNING  Do not use any accessories not supplied or recommended by  GE Healthcare           WARNING  Do not use the Ettan DALTsix and Ettan DALTtwelve Systems if they are  not working properly  nor if it have suffered any damage  for example     e damage to the power cord or its plug  e damage caused by dropping the equipment    e damage caused by splashing liquid onto items not supplied or  recommended by GE Healthcare        WARNING   Operation and user maintenance of the Ettan DALTsix and Ettan  DALTtwelve Systems should be performed by properly trained personnel  only     WARNING   Ettan DALTsix and Ettan DALTtwelve Systems are high voltage  instruments that
47. tubes available from GE Healthcare 10 15  or Petri dish       Note  The subsequent steps of gel assembly  preparation of electrophoresis unit     Applying equilib    insertion of the gel into the Precast Gel Cassette  and melting of the sealing  solution can be performed as the Immobiline DryStrip Gels are equilibrating  as  long as the time frames above are adhered to     rated    Immobiline DryStrip Gels          CAUTION  Handle the gels with care  The edges of the glass plates of the gels are  sharp           Place the gels in the Ettan DALT cassette rack     Briefly rinse the Immobiline DryStrips by submerging them in a measuring cylinder  containing SDS electrophoresis buffer  see Appendix B 3  for Ettan DALT     Holding one end of the Immobiline DryStrip with forceps  carefully place the  Immobiline DryStrip in between the two glass plates of the gel  Using a thin plastic  spacer  push against the plastic backing of the Immobiline DryStrip  not the gel  itself  and slide the strip between the two glass plates until it comes into contact  with the surface of the gel     Note  The strip should just rest on the surface of the gel  Avoid trapping air bubbles  between strip and the gel and avoid piercing the second dimension gel with  the strip     Note  The acidic end of the Immobiline DryStrip should be on the right side of the  gel when the shorter of the two plates is facing the user     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA 29  
48. uration  h min    Day runs    set temperature to 25   C   1 mm thick gels 1 2 0 45   lab cast and precast  2 17  max 180  4 00  1 5 mm thick gels 1 5 0 30   2 17  max 180  6 00       Overnight runs     set temperature to 30   C  power supply   for continuous run    1 0 mm thick gels run overnight    1 18 00    1 5 mm thick gels run overnight    15 17 00       1 For the best possible resolution  faster separation times should be used  Use the faster      lt  6 h  protocols instead     Ettan DALTsix and Ettan DALTtwelve Systems Operating Instructions 28 9640 96 AA    Operation 4  Performing arun 4 3    4 3 Performing a run    For Ettan DALTsix    For instructions on how to start and stop the electrophoresis program  refer to the user  documentation for the power supply control unit used     For Ettan DALTtwelve    1    Press the start stop button on the power supply control unit to start the  electrophoresis program     To stop the electrophoresis program before it is finished  press the start stop  button     4 4 Procedures after a run    Unloading the gels from the  Ettan DALT electrophoresis    units          WARNING   Disconnect all units from mains power supply before removing the  safety lid  unloading the gels  draining and cleaning the electrophoresis  unit           Unloading the gels in Ettan DALTsix          CAUTION  Ettan DALTsix electrophoresis unit is heavy  especially when filled with  buffer  Handle the unit with care to avoid personal injury        Carefully pu
    