Home

ST 3000 Smart Transmitter Series 900 Remote Diaphragm Seals

image

Contents

1. 4 Plastic Plugs TEMPORARY ONLY to protect threads and MUST be REMOVED before installation 8 If Table III Bolt Nut option is chosen Seal bolts will ship as same material and MAWP may change Note Remote seal system pressure rating is body rating or seal rating whichever is less ST 3000 Smart Pressure Transmitter 35 34 ST 16U 34 Issue 47 Page 7 of 11 Availability STR9xx Customer clamp rating or 600 psi whichever is less i ae eee ea mum sess seess na el Sanitary Seal Non Wetted Material No Selection Sides s No S lection PRESE Seal Pressure Rating Diameter Bolt Pattern S Bolts SS Bolts lt for 3 Pipe gt 8 Bolt 9 di 1 500 psi 1 500 psi Design RGK aa T e 6 93 ue gt 1 250 psi 1 250 psi Design ROK Seem Seen 316L SS Carbon Steel 316L SS 316L SS Wetted Material Hastelloy C 276 316L SS Saddle Seal Hastelloy C 276 Hastelloy C 276 316L SS N A Body Only 19 Hastelloy C 276 N A Body Only Klinger C 4401 non asbestos Gasket Grafoil Teflon Gylon 3510 8 If Table Ill Bolt Nut option is chosen Seal bolts will ship as same material and MAWP may change All sanitary seals have dairy grade 3A approval 10 Bolts are not included with body only selection Note Remote seal system pressure rati
2. Rating Figure 34 ST 03 ANSI 150 ANSI 300 ANSI 150 ANSI 300 ANSI 150 ANSI 300 ANSI 150 ANSI 300 ANSI 150 ANSI 150 23 ANSI 300 23 ANSI 150 22 ANSI 300 22 ANSI 150 22 ANSI 300 22 ANSI 150 ANSI 150 23 ANSI 300 23 ANSI 150 23 ANSI 300 23 ANSI 150 23 ANSI 300 ANSI 150 ANSI 300 1 Standard facing 125 250 AARH RF raised face serrated surface finish 4 Plastic Plugs are TEMPORARY ONLY to protect threads and MUST be REMOVED before installation Note Remote seal system pressure rating is body rating or seal rating whichever is less 31 34 ST 16U 34 Issue 47 Page 3 of 11 Availability ol STR9xx lt I o o oO Const See Spec Table Il continued below STR9xx CCA CCC CGA CGC CDA CDC DCA DCC DGC DDA Table Il continued next page 32 ST 3000 Smart Pressure Transmitter 34 ST 16U 34 Issue 47 Page 4 of 11 Availability STR9xx L saam tower Samen 30 46 316L SS 316L SS Hastelloy C 276 316L SS Hastelloy C 276 Hastelloy C 276 Monel 4009 Monel 4009 Tantalum 316L SS Tantalum Hastelloy C 276 Tantalum Tantalum Clad Upper Upperiset A on ee Carbon Steel 316L SS Flush Flanged Seawih Fishing TN KENNEN EE EIER Lower Connections One 1 4 with plastic plug continued 4 EA and Plugs
3. QF 2 VS Special High Vacuum 50 ANNE required 40 D T Consu ens ae a ES 40 2 25 30 175 350 i 450 200 300 Process Temperature Degrees F Maximum Pressure Limited Rating of Seal Pressure NEOBEE 20 30 to 400 F 10 40 5 25 30 175 350 450 550 0 10 200 400 Process Temperature Degrees F Figure 13 ST 3000 Remote Seals operable limits for pressure versus temperature 1Am D psig 9 psig ST 3000 Smart Pressure Transmitter 1200 Loop Resistance ohms 800 650 450 250 Note A minimum of 250 ohms of loop resistance is necessary to support communications Loop resistance equals barrier resistance plus wire resistance plus receiver resistance 4 Operating Area Figure 14 Supply voltage and loop resistance chart 10 8 16 28 20 63 25 28 3 37 0 Operating Voltage Vdc 42 4 Performance Under Rated Conditions Model STR93D 0 to 100 psi 7 bar Parameter Description Upper Range Limit psi 100 bar 7 Minimum Span psi 0 9 bar 0 063 Turndown Ratio 110 to 1 Zero Elevation and Suppression No limit except minimum span within 100 URL Accuracy Reference Includes combined effects of linearity hysteresis and repeatability Accuracy includes residual error after averaging successive readings For FouNDATION Fieldbus use Digital Mode specifications For HART use Analog Mode specifi
4. 34 ST 16U 34 Issue 47 Page 5 of 11 Availability STR9xx TABLE SEALS continued Diaphragm Flange Pressure Rating Dependent on Flange Size 1 Selection Diameter Customer Flange 316L SS 316L SS Hastelloy C 276 316L SS Wetted Material poe i Hastelloy C 276 Hastelloy C 276 Monel 4009 Monel 4009 Tantalum Tantalum Non Wetted Materials No Selection tac mir No Selection No Selection Calibration Rings None 316L SS Hastelloy C 276 Monel 400 Pancake Seal Flushing Connections One 1 4 with plastic plug and Plugs One 1 4 with metal plug Metal plug material Two 1 4 with plastic plugs will be the same as Two 1 4 with metal plugs Lower material if One 1 2 with plastic plug metal plug is chosen One 1 2 with metal plug SS Plug for CS Lower Two 1 2 with plastic plugs and Tantalum Clad Two 1 2 with metal plugs Table continued below Availability STR9xx TABLE SEALS continued i B 1 Wistted 316L SS 316L SS a tena Hastelloy C 276 316L SS Hastelloy C 276 Hastelloy C 276 Chemical Tee Taylor Wedge Non Wetted Material No Selection SCENE GENES No Selection REESE Styles No Selection No Selection No Selection Xeno Table Il continued next page 1 Standard facing 125 250 AARH RF raised face serrated surface finish 4 Plastic Plug
5. C the type of protection used on the Not FISCO FNICO Ci 4 2nF 14 Ta 50 to 40 C equipment certification nameplate Li 0 Once a type of protection has been Pi 1 2W checked on the nameplate subsequently the equipment shall not be reinstalled using any of the other certification types Li 0 except Li 150uH when Option ME Analog Meter is selected 14 ST 3000 Smart Pressure Transmitter Type of Protection Comm Option Field Parameters Temp Codes Flameproof Zone 1 All All T5 Ta 50 to 93 C BR Ex d IIC T6 Ta 50 to 78 C Enclosure IP 66 67 Ui 30V li 100mA T4 Ta 50 to 93 C 4 20 mA DE Ci 4 2nF T5 Ta 50 to 85 C INMETRO Li T6 Ta 50 to 70 C CERTUSP Pi 1 2W Brazil Intrinsically Safe Zone 0 1 Ui 30V li 100mA T4 50 to 93 C BR Ex ia IIC 4 20 mA HART Ci 4 2nF T5 Ta 50 to 63 C c Ies Li T6 Ta 50 to 48 C Enclosure IP 66 67 Pi 1 2W Ui 24V Fieldbus Pere T3 Ta 50 to 93 C Not FISCO Liz 0 4 Ta 50 to 40 C Pi 1 2W Li 0 except Li 150uH when Option ME Analog Meter is selected ST 3000 Pressure Transmitter Marine Certificate MT Option applications This certificate defines the certifications covered for the ST 3000 Pressure Transmitter family of products including the SMV 3000 Smart Multivariable Transmitter It represents the compilation of the five certificates Ho
6. diaphragm gasket surface wetted Configuration HS Figure A Flange Non Wetted x D 555 JL 00066 Aa g J D effective Body diaphragm Diaphragm extends over diaphragm the entire Raised face gasket surface wetted non wetted Configuration IS Figure C ST 3000 Smart Pressure Transmitter 155 Instr Conn F rea Z i diaphragm ang Raised face Gasket surface wetted Configuration HT Figure B Instr Conn A Flange Ring Non Wetted we diaphragm Body Raised face Gasket wetted surface wetted Configuration IT Figure D ST 3000 Smart Pressure Transmitter Dimensions and Drawings Flush Flanged Seal with Lower A BO B1 B2 EJE B2 ES B1 B2 A SE B1 B2 A BO B1 B2 A BO B1 B2 BO B1 Class 300 82 BO ESI d B2 A B1 B2 BO B1 B2 BO pe BR B2 A EI B1 B2 A B1 B2 B1 B Dimension with 1 4 NPT Flushing Connection B2 B dimension with 1 2 NPT Flushing Connection Na mw w ro iro ro m o GIN NN Wu 8 23 24 AY Uppermsert f Instr Conn Flange Ring Mon Wetted Diaphregrm Wetted Wetted Flush Flanged Seal with Lower ST 3000 Smart Pressure Transmitter Upper Mon lWetted 0 90 Flush Conn Opti
7. 100 linear scale The Smart Meter option SM provides an LCD display for both analog and digital output and can be configured to display pressure in selected engineering units HART Output Protocol Options HC and H6 Optional electronic modules for the ST 3000 provide HART Protocol compatibility in either HART 5 x or 6 x formats Transmitters with a HART Option are compatible with any HART enabled System that provides 5 x or 6 x format support Foundation Fieldbus Output Option FF Equips transmitter with FF protocol for use in 31 25 kbit s FF networks See document 34 ST 03 72 for additional information on ST 3000 Fieldbus transmitters SIL2 SIL3 Certification Option SL This ST 3000 product is available for use with safety systems With the SL option we are fully certified to SIL 2 capability for single transmitters and SIL 3 capability for multiple transmitter use through T V Nord Sys Tec GmbH amp Co KG We are in compliance with the following SIL standards IEC 61508 1 1998 IEC 61508 2 2000 IEC 61508 3 1998 ST 3000 Smart Pressure Transmitter Lightning Protection Option LP A terminal block is available with circuitry that protects the transmitter from transient surges induced by nearby lightning strikes NAMUR NE43 Compliance Option NE This option provides software the meets the NAMUR NE43 requirements for failsafe software Transmitter failure information is generated when the measuring information
8. 78 C Ui 30V li 100mA T4 Ta 50 to 93 C 4 20 mA DE Ci 4 2nF T5 Ta 50 to 85 C Li T6 Ta 50 to 70 C Pi 21 2W Intrinsically Safe Zone 0 1 Ui 30V li 100mA 4 Ta 50 to 93 C 9II 1 G Ex ia 11C 4 20 mA HART Ci 4 2nF T5 Ta 50 to 63 C Li T6 Ta 50 to 48 C Enclosure IP 66 67 Pi 1 2W Ui 24V Fieldbus T3 Ta 50 to 93 Not FISCO ir 0 T4 Ta 50 to 40 C Pi 1 2W Ui 30V li 100mA 4 Ta 50 to 93 C 4 20 mA DE Ci 4 2nF T5 Ta 50 to 85 C Li T6 Ta 50 to 70 C Pi 1 2W Non Sparking Zone 2 Ui 30V li 100mA TA Ta 50 to 93 C G Ex na 4 20 mA HART Ci 4 2nF T5 Ta 50 to 63 C Honeywell Li T6 Ta 50 to 48 Enclosure IP 66 67 Ui 24V Fieldbus LE T3 Ta 50 to 93 C Not FNICO Liz 0 4 Ta 50 to 40 C Pi 1 2W Multiple Marking Ui 30V li 100mA 4 Ta 50 to 93 C Flameproof Zone 1 4 20 mA DE l unen Hs B ES buie mE 2 2 G Ex d IIC Pi 1 2W Intrinsically Safe Zone 0 1 Ui 30V Ex ia IIC li 100mA 4 Ta 50 to 93 C 1 G 4 20 mA HART Ci 4 2nF T5 Ta 50 to 63 C Li T6 50 to 48 C Non Sparking Zone 2 Pi 1 2W G Ex nA lic NOTE The user must determine the type of protection required for Ui 24V installation of the equipment The user p shall then check the box adjacent to Fieldbus li 250mA T3 Ta 50 to 93
9. 93 C Not FISCO Liz 0 4 Ta 50 to 40 C Pi 1 2W Multiple Marking Ui 30V li 100mA 4 Ta 50 to 93 C Flameproof Zone 1 4 20 mA DE Ci 4 2nF T5 Ta 50 to 85 C Ex d IIC Enclosure IP 66 67 dr 1 2W Intrinsically Safe Zone 0 1 Ui 30V ia IIC 66 67 l 100A T4 T8 7 901095 6 4 20 mA HART Ci 4 2nF T5 Ta 50 to 63 C 1 user must determine the type of B 1 2W T6 Ta 0 to 48 protection required for installation of the Us equipment The user shall then check the box V adjacent to the type of protection Ui 24V used on the equipment certification li 250mA nameplate Once a type of protection has Fieldbus ES 4 2nF T3 Ta 50 to 93 C been checked on the nameplate Not FISCO 1 0 H T4 Ta 50 to 40 C subsequently the equipment shall not be Bi 1 2W reinstalled using any of the other certification types Li 0 except Li 150uH when Option ME Analog Meter is selected ST 3000 Smart Pressure Transmitter 13 Type of Protection Comm Option Field Parameters Temp Codes ATEX LCIE Flameproof Zone 0 A20 IP6X T95 C Ta 93 C 91I 1 D Ex tD All All or T80 C Ta 78 C Enclosure IP 66 67 Flameproof Zone 1 T5 Ta 50 to 93 C 1260 Ex d IIC Ex tD All All Tolas sore G i A21 IP6X T95 C Ta 93 C Enclosure IP 66 67 or T80 C Ta
10. Pressure Leak Test with Certificate F3392 Calibration Test Report and Certificate of Conformance F3399 Certificate of Conformance F3391 Certificate of Origin 0195 SIL Certificate SIL 2 3 FC33337 NACE Certificate Process Wetted amp Non Process Wetted FC33339 NACE Certificate for welded meter bodies only F0198 NACE Certificate Process Wetted FC33338 Material Traceability Certification per EN 10204 3 1 FC33341 Marine Type Approvals DNV ABS BV KR amp LR Table Ill continued next page 5 Supplied as 316 SS Grade CF8M the casting equivalent of 316 SS ST 3000 Smart Pressure Transmitter TABLE Ill OPTIONS continued Warranty Options 37 34 ST 16U 34 Issue 47 Page 9 of 11 STR9xx Selection Y Y Additional Warranty 1 year Additional Warranty 2 years Additional Warranty 3 years Additional Warranty 4 years Approval Body No hazardous location approvals Explosion Proof Approval Type Location or Classification Class I Div 1 Groups A B C D FM Dust lgnitionproof Approvals Non Incendive Intrinsically Safe Intrinsically Safe Zone 0 Intrinsically Safe Zone 1 Dust tight Enclosure Zone 0 Flameproof and Dust tight Enclosure Zone 1 Class Il III Div 1 Groups E F G Class I Div 2 Groups A B C D Class 1 Il III Div 1 Groups A B C D E F G Ex ia IIC T4 Ta 50 C to 93 C T5 Ta 50 C to 85
11. T6 Ta 50 C to 70 C Enclosure IP 66 67 Ex ia IIC T4 Ta 50 C to 93 C T5 Ta 50 C to 85 T6 Ta 50 C to 70 C Enclosure IP 66 67 Ex tD A20 IP6X T95 C at Ta 93 C or T80 C at Ta 78 C Enclosure IP 66 67 Ex d IIC T5 Ta 40 C to 93 C T6 Ta 40 C to 78 C Supply 11 42Vdc Ex tD A21 IP6X T95 C at Ta 93 C or T80 C at Ta 78 C Enclosure IP 66 67 Selection Non Sparking 10 eee Zone 2 LCIE Multiple Marking Int Safe Zone 0 1 and Dust tight Enclosure or Flameproof Zone 1 and and Dust tight Enclosure or Non Sparking Zone 2 Ex nA IIC T5 40 C to 93 C T6 Ta 40 C to 78 C Zone 2 Supply lt 42Vdc 23mA Ex tD A22 IP6X T95 C at Ta 93 C or T80 C at Ta 78 C Honeywell Enclosure IP 66 67 Ex ia IIC 4 Ta 50 C to 93 C T5 Ta 50 C to 85 T6 Ta 50 C to 70 C Ui 30V li 100mA Ex tD A20 IP6X T95 C at Ta 93 C or T80 C at Ta 78 C Ex d IIC T5 Ta 40 C to 93 T6 Ta 40 C to 78 C Supply 11 42Vdc Ex tD A21 IP6X T95 C at Ta 93 C or T80 C at Ta 78 C 1Il 3 GD Ex nA IIC T5 Ta 40 C to 93 C T6 Ta 40 C to 78 C Zone 2 Supply lt 42Vdc 23mA Ex tD A22 IP6X T95 C at Ta 93 C or T80 C at Ta 78 C Honeywell Enclosure IP 66 67 Table Ill Approvals c
12. is no longer valid Transmitter failure values are lt 3 6 mA and 2 21 0 mA The normal ST 3000 ranges are lt 3 8 mA and gt 20 8 mA Write Protection Options WP and WX Provides the capability to hardwire write protect installed transmitter configurations Stainless Steel Tagging Option TG Up to 30 characters can be added on the stainless steel nameplate mounted on the transmitter s electronics housing at no extra cost A stainless steel wired on tag with additional data of up to 4 lines of 28 characters is also available The number of characters for tagging includes spaces Transmitter Configuration Options TC and FC With Option TC the factory can configure the analog DE or HART transmitter s linear square root extraction damping time LRV URV and mode analog digital and enter an ID tag of up to eight characters and scratchpad information as specified With Option FC the Device ID Transmitter Tag Unit Level Node Address Output Mode and Damping Time Constants can be specified Custom Calibration and ID in Memory Option CC The factory can calibrate any range within the scope of the transmitter s range and enter an ID tag of up to eight characters in the transmitter s memory Indicator Configuration Option CI Provides custom configuration of Smart Meters Lifetime Warranty Option WL Extends limited 1 year warranty policy to 15 years for ST 3000 8100 pressure transmitters See Honeywell Terms and
13. Ambient Temperature 2541 7742 251070 13to158 40t085 40to 185 55 to 125 67 to 257 Process Interface 25 1 7742 See Figure 13 55 to 125 67 to 257 Temperature Humidity RH 10 to 55 0 to 100 0 to 100 0 to 100 Vacuum Region Minimum See Figure 13 Pressure mmHg absolute Atmospheric inH O absolute Atmospheric Supply Voltage Current and Load Resistance Voltage Range 10 8 to 42 4 Vdc at terminals Current Range 3 0 to 21 8 mA Load Resistance 0 to 1 440 ohms as shown in Figure 14 Maximum Allowable Working Pressure MAWP ST 3000 products are rated to Maximum Allowable Working Pressure MAWP depends on Approval Agency and transmitter materials of construction MAWP is minimum of Body Rating or Seal Rating See Model Selection Guide for Seal MAWP Body MAWP STR93D 750 psig 52 bar STR94G 500 psig 35 bar 1 For CTFE fill fluid the rating is 15 to 110 C 5 to 230 F Short term equals 2 hours at 70 C 158 F 3 MAWP applies for temperature range 40 to 125 C However Static Pressure Limit is de rated to 3 000 psi from 26 C to 40 C Use of graphite o rings de rates transmitter to 3 625 psi Use of adaptor with graphite o rings de rates transmitter to 3 000 psi Consult factory for MAWP of ST 3000 transmitters with CSA approval ST 3000 Smart Pressure Transmitter Maximum Pressure Limited Rating of Seal 10000 4000 2 Silicone DC200 E 600 555 gt
14. Conditions ST 3000 Smart Pressure Transmitter 29 Model Selection Guides are subject to change and are inserted into the specifications as guidance only Prior to specifying or ordering a model check for the latest revision Model Selection Guides which are published at http hpsweb honeywell com Cultures en U S Products Instrumentation ProductModelSelectionGuides default htm Model Selection Guide Honeywell ST 3000 Smart Transmitter DP amp GP Remote Seals Series 900 Instructions Select the desired Key Number The arrow to the right marks the selection available e Make selections from each Table I Il and IV using the column below the proper arrow e Select as many Table III options as desired plus a communications option selection e A denotes unrestricted availability A letter denotes restricted availability e Restrictions follow Table IV Key Number 1 II III Optional KEY NUMBER 34 ST 16U 34 Issue 47 Page 1 of 11 Model Selection Guide A e Selection Ava 0 25 to 0 2 700 0 62 2 to 0 7 000 mbar Body Rating 750 psi 51 7 bar Compound Characterized 0 5 to 0 500 psig 0 0 35 to 0 35 bar Body Rating 500 psi 35 bar STR93D Remote seal system pressure rating is body rating or seal rating whichever is less Important Note communication option from Table Ill AN DE HC H6 or FF TABLE I METER BODY Description 1 Remote Seal High
15. DN80 PN40 Diaphragm Upper Insert weis T 316L SS 316L SS 316L SS Wetted Material Hastelloy mae Hastelloy C 276 Hastelloy C 276 Monel 400 Monel 400 Tantalum 2 316L SS MEE CS Nickel Plated ange Materia 316L SS Seal Capillary Center Seal Connection Side Seal 2 Calibration Rings None 316L SS Hastelloy C 276 D Monel 400 Secondary Fill A B C D E F G H J K L asas Table Il continued next page 1 Standard facinq 125 250 AARH RF raised face serrated surface finish 2 Limited vacuum availabilitv 3 Minimum static pressure requirement No vacuum allowed See Specifications Fiqure 15 5 Tantalum Upper insert has Tantalum wetted parts and 316 SS or CS non wetted parts Note Remote seal system pressure rating is body rating or seal rating whichever is less ST 3000 Smart Pressure Transmitter TABLE II SEALS continued Flush Flanged with Lower 2 3 1 2 e 1 Metal plug material will be the same as Lower material if metal plug is chosen SS Plug for CS Lower and Tantalum Clad 4 2 1 1 2 1 2 One 1 4 with plastic plug One 1 4 with metal plug Two 1 4 with plastic plugs Two 1 4 with metal plugs One 1 2 with plastic plug One 1 2 with metal plug Two 1 2 with plastic plugs Two 1 2 with metal plugs Flange Pressure
16. IL 2 related Safety Systems single use and for redundant multiple use in SIL 3 Safety Systems through TUV Nord Sys Tec GmbH amp Co KG under the following standards IEC61508 1 1998 IEC 61508 2 2000 IEC61508 3 1998 Physical and Approval Bodies Parameter Description Process Interface See Model Selection Guide for Material Options for desired Seal Type Seal Barrier Diaphragm 316L Stainless Steel Monel Hastelloy C Tantalum Seal Gasket Materials Klinger C 4401 non asbestos Grafoil Teflon Gylon 3510 Mounting Bracket Carbon Steel zinc plated or Stainless Steel angle bracket or Carbon Steel flat bracket available Neg Silicone DC 200 S G 25 C 77 F 0 94 Fill Fluid Meter Body CTFE Chlorotrifluoroethylene S G 25 C 77 F 1 89 Silicone DC 200 S G 25 C 77 F 0 94 CTFE Chlorotrifluoroethylene S G 25 C 77 F 1 89 Fill Fluid Secondary Silicone Oil 704 S G 25 C 77 F 1 07 Syltherm 800 S G 25 C 77 F 0 90 NEOBEE M 20 S G 25 C 77 F 0 93 Electronic Housing Epoxy Polyester hybrid paint Low copper aluminum alloy Meets NEMA 4X watertight and NEMA 7 explosion proof Capillary Tubing Armored Stainless Steel or PVC Coated Armored Stainless Steel Length 5 10 15 20 25 and 35 feet 1 5 3 4 6 6 1 7 5 and 10 7m A 2 51 millimeter S S close coupled nipple is also available
17. One 1 4 with metal plug Metal plug material Two 1 4 with plastic plugs will be the same as Two 1 4 with metal plugs Lower material if One 1 2 with plastic plug metal plug is chosen One 1 2 with metal plug SS Plug for CS Lower Two 1 2 with plastic plugs and Tantalum Clad Two 1 2 with metal plugs Klinger C 4401 non asbestos Grafoil Teflon Gylon 3510 Diaphragm ANSI Class 150 OD ANSI Class 300 E extension DIN DN80 PN40 EFM ANSI Class 150 FGA 3 70 OD ANSI Class 300 FGC extension DIN DN100 PN40 FGP Flange Seal Diaphragm Selection with Extended 316L SS 316L SS EA Diaphragm Wetted Material Hastelloy C 276 316L SS E Hastelloy C 276 Hastelloy C 276 CS Nickel Plated e SHH No Selection No Selection s s Table Il continued next page TABLE II SEALS continued Wetted Material 1 Standard facing 125 250 AARH RF raised face serrated surface finish Plastic Plugs are TEMPORARY ONLY to protect threads and MUST be REMOVED before installation Bolt material will be same as Upper Material However if Table III bolt nut option is chosen seal bolt material will be the same Note Remote seal system pressure rating is body rating or seal rating whichever is less ST 3000 Smart Pressure Transmitter 33
18. Saddle Seal 2 2 2 4 Diaph Instr Conn ose E EM j i cM 5 Saddle B 280 a Seal A 3 50 4 or larger 5 3 04 Note Specify B or 8 bolt pattem by User c Lower Housing ad Wetted Gasket Diaphragm Wetted Wetted 3 Saddle Seal Instr Conn Upper Body Bolt Non wetted gt Lower Housing A Wetted Gasket Diaphragm Wetted Vetted 4 Saddle Seal Calibration Rings Type Size Rating Dimension 1 4NPT 12 NPT 5 00 5 00 LZ 2 gone 3 1504 8008 1 00 1 50 D Z B i 3 00 3 00 C B Ll mm Calibration Ring 28 Options High Accuracy Option HA Extends applicable S100 models to 0 025 analog reference accuracy Angle Mounting Bracket Options MB MX SB SX FB The angle mounting bracket is available in either zinc plated carbon steel or stainless steel and is suitable for horizontal or vertical mounting on a two inch 50 millimeter pipe as well as wall mounting An optional flat mounting bracket is also available in carbon steel for two inch 50 millimeter pipe mounting An option also exists for Marine approved mounting brackets used with Marine certification options Indicating Meter Options ME and SM Two integral meter options are available An analog meter option ME is available with a dual 0 to 10 square root and 0 to
19. See Model Selection Guide Wiring Accepts up to 16 AWG 1 5 mm diameter Mounting See Figure 16 Dimensions See Figures 19 and 20 for transmitter dimensions See Model Selection Guide for Seal dimensions Net Weight Transmitter 4 1 Kg 9 Ibs Total weight is dependent on seal type and capillary length See Figure 13 for Fill Fluid temperature limits 2 inch Sanitary Seals are limited to 15 ft 4 6 m capillary length NOTE Pressure transmitters that are part of safety equipment for the protection of piping systems or vessel s from exceeding allowable pressure limits equipment with safety functions in accordance with Pressure Equipment Directive 97 23 EC article 1 2 1 3 require separate examination 1 Vent Drains are sealed with Teflon or PTFE 2 Hastelloy C 276 or UNS N10276 3 Monel 400 or UNS N04400 a Supplied as 316 SS or as Grade CF8M the casting equivalent of 316 SS 5 Carbon Steel heads are zinc plated and not recommended for water service due to hydrogen migration For that service use 316 stainless steel wetted Process Heads i Hastelloy C 276 or UNS N10276 Supplied as indicated or as Grade CW12MW the casting equivalent of Hastelloy C 276 7 Monel 400 or UNS N04400 Supplied as indicated or as Grade M30C the casting equivalent of Monel 400 10 Certifications ST 3000 Smart Pressure Transmitter Type of Protection Comm Op
20. Side 2 Remote Seals 1 Remote Seal Low Side Value Added Model VAM unit Number of Seals Base STR models no longer include a default communications option All units now require the selection of a DC 200 Silicone CTFE Non Wetted Material 316 SS 316 SS for Close Couple Fill Fluid Meter Body Construction In Line Gauge 316 SS Heads 316 SS Heads for Close Couple connection 316 SS with all welded meter body Dual Head DP See Specification Sheet 34 ST 03 57 for figures on construction In Line Gauge Dual Head DP All Welded 30 ST 3000 Smart Pressure Transmitter 34 ST 16U 34 Issue 47 Page 2 of 11 TABLE SEALS Format for Seal Selection Specify 12 characters Comm n Requifed Seal Availability Note The first 3 characters are common to all seals When selecting required seal you must specify only the 9 selections within the required seal No Fill Fluid Silicone DC 200 CTFE Silicone Oil 704 Neobee M20 Syltherm 800 No Capilary No Nipple 5 feet 10 feet 15 feet 20 feet SS Armor 25 feet Capillary 35 feet Length 5 feet Connection of 10 feet Remote Seal 15 feet PVC Coated to Meter Body 20 feet SS Armor 25 feet 35 feet 2 inch long SS nipple close coupled No Selection pex s No Seal Attached to Core Transmitter 000000000 3 3 E Diameter EE Rating ANSI Class 150 ANSI Class 300 DIN
21. TR93D DP Transmitter with one Remote Seal Diaphragm Direct Capillary Capillary Length Size Mount 5 10 15 20 30 35 20 25psi 30psi 40psi 29 spi opsi 10 ipsi 12psi 14psi 15psi 41 o9ps 0 9 ps 2ps 3ps ps 5ps Minimum span is the higher of the value from the table above or the value defined under Performance Conditions for the range transmitter Figure 15 Maximum capillary length and diaphragm size chart ST 3000 Smart Pressure Transmitter Performance Under Rated Conditions General for all Models Parameter Description Output two wire Analog 4 to 20 mA or DE digital communications mode Options available for Foundation Fieldbus and HART protocols Supply Voltage Effect 0 005 span per volt Damping Time Constant Adjustable from 0 to 32 seconds digital damping CE Conformity Europe 89 336 EEC Electromagnetic Compatibility EMC Directive NAMUR NE 43 Compliance Option NE Transmitter failure information is generated when the measuring information is invalid or no longer present Failure information is transmitted as a current signal but outside the normal 4 20 mA measurement signal level Transmitter failure values are lt 3 6 mA and gt 21 0 mA The normal signal range is 2 3 8 mA and s 20 5 mA SIL 2 3 Compliance SIL certified to IEC 61508 for non redundant use in S
22. Technical Information ST 3000 Smart Transmitter Series 900 Remote Diaphragm Seals Mo Specifications 34 ST 03 57 March 2013 Introduction In 1983 Honeywell introduced the first Smart Pressure Transmitter the ST 3000 In 1989 Honeywell launched the first all digital bi directional protocol for smart field devices Today its ST 3000 Series 900 Remote Seal Transmitters continue to bring proven smart technology to a wide spectrum of pressure measurement applications For applications in which the transmitter must be mounted remotely from the process Honeywell offers the remote seal line of gauge absolute and differential pressure transmitters Typical applications include level measurement in pressurized vessels in the chemical and hydrocarbon processing industries A second application is flow measurement for slurries and high viscosity fluids in the chemical industry Honeywell remote seal transmitters are available with secondary fill fluids for corrosive or high temperature process fluids STR93D 0 to 500 psig 0 to 35 bar All ST 3000 transmitters can be ordered to provide one of STR94G the following output communication options Communications options 4 20 mA Honeywell Digitally Enhanced DE HART versions 5 x or 6 x FOUNDATION Fieldbus The devices provide comprehensive self diagnostics to help users maintain high uptime meet regulatory requirements and attain high quality standards S900 t
23. ass Ill Division 2 Pi 1 2W Class I Zone 2 Group IIC i Vmax 32V _ Enclosure Type 4X IP 66 67 FNICO Ci 4 2nF T4 Ta E 40 qu T3 Ta 93 C Li 0 except Li 150uH when Option ME Analog Meter is selected FM Approvals is a service mark of FM Global ST 3000 Smart Pressure Transmitter 11 Type of Protection Comm Option Field Parameters Temp Codes Explosion Proof Class Division 1 Groups B C D locations Dust Ignition Proof Class Il III v Division 1 Groups E F G a Al locations Enclosure Type 4X Vmax 42V Imax 225mA 4 20 mA DE Ci 4 2nF T4 Ta 93 C Li Pi 1 2W Intrinsically Safe Vmax 42V Imax 225mA Class Il 111 Division 1 Groups 4 20 mA HART Ci 4 2nF T4 Ta 93 C B C D E F G locations Li Enclosure Type 4X Pi 1 2W Vmax 24V Canadian Fieldbus Entity eee T4 Ta 40 C Standards Not FISCO PEN T3 Ta 93 C Association Pi 1 2W CSA Vmax 42 4V Imax 225mA 4 20 mA DE Ci 4 2nF T4 Ta 93 C Li Pi 1 2W i Vmax 30V Nonincendive nA E i 0 Class Division 2 Groups 420 oe D locations Enclosure Type 4X Pi 1 2W Vmax 24V Fieldbus Entity T4 40 Not Liz 0 T3 Ta 93 C Pi 1 2W Canadian Registration Number CRN All ST 3000 models except STG19L STG99L STG170 and STG180 have been regist
24. ax 65 6445 3033 South Korea Honeywell Korea Co Ltd Phone 822 799 6114 Fax 822 792 9015 Specifications are subject to change without notice For More Information Learn more about how Honeywell s ST 3000 Series 900 Remote Seal Transmitters can increase performance reduce downtime and decrease configuration costs visit our website www honeywellprocess com or contact your Honeywell account manager Honeywell Honeywell Process Solutions 1860 West Rose Garden Lane 34 ST 03 57 Phoenix Arizona 85027 March 2013 1 800 423 9883 or 1 800 343 0228 2013 Honeywell International Inc www honeywellprocess com
25. cations In Analog Mode 0 20 of calibrated span or upper range value URV whichever is greater terminal based For URV below reference point 50 inH2O accuracy equals 50 inH O 125 mbar S 0 10 0 10 or 0 10 0 10 in of span span inH span mbar In Digital Mode 0 175 of calibrated span or upper range value URV whichever is greater terminal based For URV below reference point 50 inH2O accuracy equals 50 inH O 125 mbar ES or 0 075 0 10 in of span span inH 0 span mbar 0 075 Combined Zero and Span Temperature Effect per 28 C 50 F In Analog Mode 1 5 of span For URV below reference point 200 inH2O effect equals 200 inH O 500 mbar 5 0 30 1 2 or 0 30 1 2 in of span span inH O span mbar In Digital Mode 1 475 of span For URV below reference point 200 inH O effect equals 200 inH O 500 mbar 0 275 1 2 or 0 275 1 2 in of span span inH span mbar Performance specifications are based on reference conditions of 25 C 77 F zero 0 static pressure 10 to 55 RH and 316L Stainless Steel barrier diaphragm Specification applies to transmitters with 2 seals only Apply 1 5 times factor to temperature effect for capillary lengths greater than 10 feet or for 2 inch sanitary seals 8 ST 3000 Smart Pressur
26. duit Plug ATEX CSA amp IECEx M20 316 SS Certified Conduit Plug ATEX CSA amp IECEx 1 2 NPT Non certified Conduit plug Zinc plated carbon steel general use NAMUR Failsafe Software SIL 2 T V Certified transmitter requires HC or H6 and WP options Lightning Protection Custom Calibration and I D in Memory Transmitter Configuration non Fieldbus Transmitter Configuration Fieldbus Write Protection Delivered in the enabled position Write Protection Delivered in the disabled position Stainless Steel Customer Wired On Tag 4 lines 26 characters per line customer supplied information Stainless Steel Customer Wired On Tag blank Meter Body Options Carbon Steel standard A286 SS NACE Bolts and 304 SS NACE Nuts for Heads 316 SS Bolts and 316 SS Nuts for Process Heads B7M Bolts and Nuts for Process Heads Remote Seal Options Gold Plated Seal Diaphragm 1 Seal Gold Plated Seal Diaphragm 2 Seals Teflon Coated Seal Diaphragm 1 Seal only for anti sticking Teflon Coated Seal Diaphragms 2 Seals only for anti sticking Angle Mounting Bracket Carbon Steel Marine Approved Angle Mounting Bracket Carbon Steel Angle Mounting Bracket 304 SS Marine Approved Angle Mounting Bracket 304 SS Flat Mounting Bracket Carbon Steel ervices Certificates Options Users Manual Paper Copy Standard HC H6 or FF ships accordingly Clean Transmitter amp Seals for Oxygen or Chlorine Service with Certificate Over
27. e Transmitter Performance Under Rated Conditions Models STR94G 0 to 500 psi 35 bar Parameter Description Upper Range Limit psi 500 bar 35 Minimum Span psi 20 bar 1 4 Turndown Ratio 100 to 1 Zero Elevation and No limit except minimum span from absolute 0 zero to 100 URL Suppression Accuracy Reference Includes In Analog Mode 0 10 of calibrated span or upper range value whichever is combined effects of linearity greater terminal based hysteresis and repeatability Accuracy includes residual error after averaging successive readings For FOUNDATION Fieldbus use Digital Mode specifications For HART use Analog Mode specifications In Digital Mode 0 075 of calibrated span or upper range value URV whichever is greater terminal based Performance specifications are based on reference conditions of 25 C 77 F zero 0 static pressure 10 to 55 RH and 316L Stainless Steel barrier diaphragm Transmitter Minimum Span and Maximum Capillary Length Minimum recommended span for STR93D DP Transmitter with two Remote Seals Diaphragm Capillary Capillary Length Size 5 10 15 20 25 3 35 maximum 20 15psig 20psig 25psig 15 29 50 75 100iwc 125iwc 150iwc 175iwc 200iwc 41 25iwc 25 25 25 25iwc 27 30 Minimum recommended span for STR94G or S
28. ends verifying the calibration of these devices once every four years Calibration Approved Honeywell Process Solutions York PA USA Manufacturing Honeywell Tianjin Limited Tianjin P R China Locations Honeywell Automation India Ltd Pune 411013 India Foundation Fieldbus is a trademark of the Fieldbus Foundation HART is a registered trademark of HART Communications Foundation Hastelloy C 276 is a registered trademark of Haynes International Monel 400 is a registered trademark of Special Metals Corporation Viton is a registered trademark of DuPont Teflon is a registered trademark of DuPont DC 200 is a registered trademark of Dow Corning FM Approvals is a service mark of FM Global ST 3000 and Experion are registered trademarks of Honeywell International Inc 16 ST 3000 Smart Pressure Transmitter Mounting LP Side H2 Fixed ref leg H1 HP Side NOTE Lower flange seal should not be mounted over 22 feet below the transmitter for silicone fill fluid 11 feet for CTFE fill fluid with tank at one atmosphere The combination of tank vacuum and high pressure capillary head effect should not exceed 9 psi vacuum 300 mmHg absolute Consult Honeywell for installation of STR93D Figure 16 Typical mounting arrangement for ST 3000 Transmitter with Remote Diaphragm Seals ST 3000 Smart Pressure Transmitter Application Data Liquid Level Closed Tank Determine the minimum and maxim
29. ered in all provinces and territories in Canada and are marked CRN 0F8914 5C 12 ST 3000 Smart Pressure Transmitter Type of Protection Comm Option Field Parameters Temp Codes IECEx International Electrotechnical Commission LCIE Flameproof Zone 1 T5 Ta 50 to 93 C All All T6 Ta 50 to 78 C Ex d IIC Enclosure IP 66 67 Ui 30V li 100mA 4 Ta 50 to 93 C 4 20 mA DE Ci 4 2nF T5 Ta 50 to 85 C Li T6 Ta 50 to 70 C Pi 1 2W Ui 30V Intrinsically Safe Zone 0 1 li 100mA TA Ta 50 to 93 C 4 20 mA HART Ci 4 2nF T5 Ta 50 to 63 C Ex ia IIC Enclosure IP 66 67 Li T6 Ta 50 to 48 C Pi 1 2W Ui 24V Fieldbus T3 Ta 50 to 93 Not FISCO Heo 4 Ta 50 to 40 C Pi 1 2W Li 0 except Li 150uH when Option ME Analog Meter is selected Type of Protection Comm Option Field Parameters Temp Codes SAEx South Africa Flameproof Zone 1 T5 Ta 50 to 93 C Ex d IIC Enclosure IP 66 67 Al T6 Ta 50 to 78 C Ui 30V li 100mA T4 Ta 50 to 93 C 4 20 mA DE Ci 4 2nF T5 Ta 50 to 85 C Li T6 Ta 50 to 70 C Pi 1 2W Ui 30V TE li 100mA 4 Ta 50 to 93 C ETE iab add 4 20 mA HART 42nF T5 Ta 50 to 63 C x ia Enclosure Li T6 Ta 50 to 48 C Pi 1 2W Ui 24V Fieldbus EE T3 Ta 50 to
30. g transmitter configuration ST 3000 Smart Pressure Transmitter 19 Reference Dimensions millimeters inches With Smart 829 94 9 Removal 36 meter 3 26 3 74 Clearance 0 14 Plug 53 1 65 1 for All Caps 457 2 09 2 56 with meter 135 TR Without Without 1 8 meter meter 55 3 2 18 nn T 23 5 925 Optional Optional Meters external Rotational LP seal omitted for clarity 67 2 64 29 6 3 69 STR93D DP I Remote Seal with Horizontal Pipe Mount With Smart 82 9 94 9 Removal 3 6 Plug meter 3 26 3 74 Clearance 0 14 53 1 65 1 for All Caps 135 45 7 2 09 2 56 with meter 5 32 18 Without Without meter meter D Seal Optional omitted for external clarity ground Optional Meters Rotational Lock STR94G LGP I Remote Seal with Horizontal Pipe Mount 24272 Figure 19 Approximate horizontal mounting dimensions for Remote Seal Transmitter Dimensions are for reference only and may be subject to change For specific information contact Honeywell customer service per the contact information located with in this document 20 Reference Dimensions ST 3000 Smart Pressure Transmitter millimeters inc
31. he welded ST 3000 meter body is an important part of an All Welded Remote Seal Solution which is commonly used in Vacuum applications Description The ST 3000 transmitter can replace any 4 to 20 mA output transmitter in use today and operates over a standard two wire system The measuring means is a piezoresistive sensor which actually contains three sensors in one It uses a differential pressure sensor a temperature sensor and a static pressure sensor in delivering the most comprehensive compensated output signal available today Microprocessor based electronics provide higher span turndown ratio improved temperature and pressure compensation and improved accuracy The transmitter s meter body and electronics housing resist shock vibration corrosion and moisture The electronics housing contains a compartment for the single board electronics which is isolated from an integral junction box The single board electronics is replaceable and interchangeable with any other ST 3000 Series 100 or Series 900 model transmitters Features e Choice of linear or square root output conformity is a simple configuration selection e Direct digital integration with Experion PKS and other control systems provides local measurement accuracy to the system level without adding typical A D and D A converter inaccuracies e Unique piezoresistive sensor automatically compensates input for real world temperature and static pressure variati
32. hes 82 9 94 9 _ Removal 3 26 3 74 Clearance 3 6 Plu 53 1 65 1 for Caps 0 14 g 2 09 256 with meter 135 45 7 Without Without 5 32 1 8 meter meter r 7 Pc TE E 55 3 og Seal 2 48 44 omitted for _ E p p 23 5 clarity li Rotational ES SEM s Ila Lock ptional omitted for Optional Meters clarity d ground Cn 93 6 STR93D DP l Remote Seal with Vertical Pipe Mount 3 69 g With 82 9 94 9 Removal Smart meter 39 Pug 135 326 374 53 1 65 1 for All Caps Pit with meter 4 2 09 2 56 55 5 45 7 Without Without AG 1 meter meter 2 18 EE ck d 1 Rotational 19 Lock Optional Lar 1 lt CS Meters I 1 E sl d 57 Optional external ground Seal o Ch omitted for clarity b Co N STR94G LGP I Remote Seal with Vertical Pipe Mount 24273 Figure 20 Approximate vertical mounting dimensions for Remote Seal Transmitter Dimensions are for reference only and may be subject to change For specific information contact Honeywell customer service per the contact information located with in this document ST 3000 Smart Pressure Transmitter Dimensions and Drawings Flush Flanged Seal 21 22 Instr Conn Flange Non wetted f effective diaphragm 0 Diaphragm extends over the entire Raised face
33. n be used with differential pressure transmitters and are available with Taylor Wedge 5 O D process connection Figure 6 Seals with Threaded Process Connections Seals with Threaded Process Connections can be used with differential gauge and absolute pressure transmitters and are available with 12 34 and 1 NPT Female process connections ST 3000 Smart Pressure Transmitter Figure 7 Sanitary Seals Sanitary Seals can be used with differential gauge and absolute pressure transmitters and are available with 3 and 4 Tri Clover Tri Clamp process connections Figure 8 Saddle Seals Saddle Seals can be used with differential gauge and absolute pressure transmitters and are available with 3 and 4 6 bolt or 8 bolt designs process connections Figure 9 Calibration Rings Calibration Rings are available with Flush Flange Seals and Pancake Seals Flushing ports 1 4 or gt are available with calibration rings Figure 10 Stainless Steel Armor and PVC Coated Stainless Steel Armor Capillaries Stainless Steel Armor and PVC Coated Stainless Steel Armor Capillaries are available with Honeywell Remote Seal Solutions Figure 11 2 Stainless Steel Nipples 2 Stainless Steel Nipples are available for Close Coupled remote seal solutions Figure 12 Welded Meter Body for All Welded Remote Seal Solution Welded Meter Body for All Welded Remote Seal Solution T
34. n the nameplate subsequently the equipment shall not be reinstalled using any of the other certification types TABLE IV Selection Factory Identification XXX X RESTRICTIONS Restriction Available Only With Not Available With Letter Table Selection Selection b Select only one option from this group BF BG c cU Lee TC ME or FISCO FNICO compliance not available with 3C 3N 33 3S 2J CA Z2 ZD ZA 6D amp 6S FISCO FNICO compliance available only FF ME l og I I og fel cc 51055555 L 4 did ME MM ME q Restrictions continued next page ST 3000 Smart Pressure Transmitter 39 34 ST 16U 34 Issue 47 Page 11 of 11 RESTRICTIONS continued Restriction Available Only With Not Available With Letter Table Selection Tabe Selection 2 This approval code requires the selection of a certified conduit plug A5 A6 or A7 FM Approvals is a service mark of FM Global Hastelloy is a registered trademark of Haynes International Monel 400 is a registered trademark of Special Metals Corporation HART is a registered trademark of HART Communication Foundation FOUNDATION V Fieldbus is a registered trademark of Fieldbus Foundation Teflon is a registered trademark of DuPont Neobee is a registered trademark of Ste
35. neywell currently has covering the certification of these products into marine For ST 3000 Smart Pressure Transmitter and SMV 3000 Smart Multivarible Transmitter American Bureau of Shipping ABS 2009 Steel Vessel Rules 1 1 4 3 7 4 6 2 5 15 4 8 3 13 amp 13 5 4 8 4 27 5 1 4 9 7 13 Certificate number 04 HS417416 PDA Bureau Veritas BV Product Code 389 1H Certificate number 12660 B0 BV Det Norske Veritas DNV Location Classes Temperature D Humidity B Vibration A EMC B Enclosure C For salt spray exposure enclosure of 316 SST or 2 part epoxy protection with 316 SST bolts to be applied Certificate number A 11476 Korean Register of Shipping KR Certificate number LOX17743 AE001 Lloyd s Register LR Certificate number 02 60001 E1 amp E2 ST 3000 Smart Pressure Transmitter 15 The ST 3000 Smart Pressure Transmitters are in conformity with the essential requirements of the Pressure Equipment Directive Honeywell ST 3000 Smart Pressure Transmitters are designed and manufactured in accordance with the applicable portions of Annex I Essential Safety Requirements and sound engineering practices These transmitters have no pressurized internal volume or have a pressurized internal volume rated less than 200 bar 2 900 psig and or have a maximum volume of less than 0 1 liter Article 3 1 1 a first indent Group 1 fluids Therefore these transmitters are not subject to the essen
36. ng is body rating or seal rating whichever is less TABLE Il SEALS 36 34 ST 16U 34 Issue 47 Page 8 of 11 TABLE III OPTIONS Analog only can be configured using appropriate Honeywell DE tool DE Protocol communications HART 5 x Protocol compatible electronics HART 6 x Protocol compatible electronics FOUNDATION Fieldbus Communications Indicating Meter Options Analog Meter 0 100 Even 0 10 Square Root Smart Meter Custom Configuration of Smart Meter Local Zero Local Zero and Span ST 3000 Smart Pressure Transmitter Selection Availability STR9xx ae Transmitter Housing amp Electronics Options No housing conduit plugs or adaptors come standard with the ST 3000 For certain approval codes you must select a certified conduit plug from below and it will come packaged in the box with your transmitter 316 SS Electronics Housing with M20 Conduit Connections 316 SS Electronics Housing with M20 to 1 2 NPT 316 SS Conduit Adapter for use with FM and CSA Approval codes 1 2 NPT Male to M20 Female 316 SS Certified Conduit Adapter ATEX CSA amp IECEx 1 2 NPT Male to 3 4 NPT Female 316 SS Certified Conduit Adapter ATEX CSA amp IECEx M20 Male to 1 2 NPT Female 316 SS Certified Conduit Adaptor ATEX CSA amp IECEx 1 2 NPT Zinc plated Certified Conduit Plug ATEX CSA amp IECEx 1 2 NPT 316 SS Certified Con
37. on Requires Sanitary Connections e Process Application Subjects the Measuring Instrument to Hydrogen Permeation e Tank Level Applications with Maintenance Intensive Wet Legs e Tank Application with Density or Interface Measurements e Measuring Instrument Requires Remote Mounting The following diaphragm seals are standard from Honeywell please call your local salesperson if you do not see the product you need for your application Figure 2 Flush Flange Seals Flush Flange Seals can be used with differential gauge and absolute pressure transmitters and are available with 3 ANSI Class 150 ANSI Class 300 and DIN DN80 PN40 process connections Flush flange seals can also be provided with Lowers Lowers are essentially calibration rings which allow flushing connections if needed ST 3000 Smart Pressure Transmitter Figure 3 Flange Seal with Extended Diaphragm Flange Seal with Extended Diaphragm can be used with differential gauge and absolute pressure transmitters and are available with 3 and 4 ANSI Class 150 ANSI Class 300 DIN DN80 PN40 and DIN DN100 PN40 process connections 2 4 and 6 extension lengths are available n Figure 4 Pancake Seals Pancake Seals can be used with differential gauge and absolute pressure transmitters and are available with 3 ANSI Class 150 300 and 600 process connections Figure 5 Chemical Tee Taylor Wedge Chemical Tee Taylor Wedge seals ca
38. onal Gasket Lower ree Wetted Wetted Weed Stud Flush Flanged Seal with Lower Note 0 90 dimension is 0 70 for 4 1 Dia Diaphragm ST 3000 Smart Pressure Transmitter 25 Dimensions and Drawings Flanged Seal with Extended Diaphragm gg J Instr Conn 7 Fill Screw A 7 50 B 0 84 2 80 Flange A 8 25 Non Wetted i C 2 80 DINDNso B PN40 B oa Seal with C 2 80 Extended A 2 B aised Face pec C Wetted Extension Length Ext i T C Wette DIN DNB0 PN40 m Diaphragm Wetted Designed to meet with schedule 40 pipe Flange Extended Seal Pancake Seal 388 24NF X 0 38 Hin Connection Pancake Flange Seal Chemical Tee Taylor Wedge Seal A Fill Instr Conn Diaphragm Wetted Wetted Chemical Tee Taylor Wedge Seal 26 Seal with Threaded Process Connection Without Flush B Dimension with 1 4 NPT Flushing Connection B dimension with 1 2 NPT Flushing Connection Sanitary Seal 1 42 B 8 s B ERIS 8 BB ST 3000 Smart Pressure Transmitter Upper Body Non Wetted Instr Conn Bolt Fill Screw 4 Flush Conn Optional Lower Threaded Process Connection Seal Instr Conn Body Wetted Fill Screw Diaphragm y A Wetted Sanitary Seal ST 3000 Smart Pressure Transmitter 27
39. ons e Added smart features include configuring lower and upper range values simulating accurate analog output and selecting preprogrammed engineering units for display e Smart transmitter capabilities with local or remote interfacing means significant manpower efficiency improvements in commissioning start up and ongoing maintenance functions e ST 3000 transmitters feature full Dual Seal certification based on ANSI NFPA 70 202 and ANSI ISA 12 27 01 requirements without the use of additional seal protection elements e ST 3000 transmitters are available fully compliant to SIL 2 3 requirements as a standard option ST 3000 Smart Pressure Transmitter Advanced Diagnostics ST 3000 is now available for both HART 6 and Foundation Fieldbus with advanced diagnostics that minimize unplanned plant outages minimize maintenance costs and by providing the industry s most reliable transmitter e Provide advanced warning of possible failure events and avoid costly shutdowns e Three levels of failure reporting e Comprehensive list of on board diagnostics Ref ST 3000 User manual with HART 6 34 ST 25 17 Rev June 09 and Foundation Fieldbus option manual 34 ST 25 15 Rev June 09 ST 3000 Smart Pressure Transmitter Operating Conditions All Models Reference Rated Condition Operative Limits Transportation and Parameter Condition at Storage zero static C oF C oF eC F C F
40. ontinued next page 38 ST 3000 Smart Pressure Transmitter 34 ST 16U 34 Issue 47 Page 10 of 11 STR9xx mc TABLE III Approvals Options continued RSEN cu i s Canadian Explosion Proof Class I Div 1 Groups B C D Standards Dust lgnitionproof Class Il Ill Div 1 Groups E F G 2 26 26 Association T Intrinsically Safe Class 1 II Ill Div 1 Groups A B C D E F G CSA Flameproof Exec 2T T5 Ta 40 to 93 C T6 Ta 40 to 78 C CA 26 26 Ex ia IIC T3 T4 T5 T6 See Spec for detailed temperature codes by b Communications option Intrinsically Safe Zone 0 1 Intrinsically Safe Zone 0 1 ee Flameproof Zone 1 Ex dlIC T5 T6 Enclosure IP 66 67 zD ele Africa Multiple Marking r Int Safe Zone 0 1 or Ex IIC 4 1 ZA e e Flameproof Zone 1 Ex d IIC T5 T6 Enclosure IP 66 67 cERTUSP Flameproof Zone 1 BR Ex d IIC T5 T6 6D ele INMETRO intrinsically Safe BR Ex ia IIC T4 T5 T6 gs zs Brazil Zone 0 1 CERTUSP certificate for detailed temperature codes by ommunications option 4 T5 T6 22 e o 10 See ATEX installation requirements in the ST 3000 User s Manual 11 The user must determine the type of protection required for installation of the equipment The user shall then check the box N adjacent to the type of protection used on the equipment certification nameplate Once a type of protection has been checked o
41. pan Company Syltherm 800 is a Trademark of Dow Corning Corporation Klingersil C 4401 is a registered trademark of THERMOSEAL INC GRAFOIL is a registered trademarks of GrafTech International Holdings Inc Gylon 3510 is registered trademark of Garlock Sealing Technologies Tri Clover Tri Clamp is a registered trademark of Alfa Laval DC 200 and DC 704 are registered trademarks of Dow Corning 40 ST 3000 Smart Pressure Transmitter Sales and Service For application assistance current specifications pricing or name of the nearest Authorized Distributor contact one of the offices below ASIA PACIFIC EMEA NORTH AMERICA SOUTH AMERICA TAC hfs tac Phone 80012026455 or 44 Honeywell Process Solutions Honeywell do Brazil amp Cia support honeywell com 0 1202645583 Phone 1 800 423 9883 Phone 55 11 7266 1900 FAX 44 0 1344 655554 1 800 343 0228 FAX 55 11 7266 1905 Email Sales sc cp apps Email Sales ask Email Sales ask oneywen salespa62 honeywell com ssc phoneywell com or ssc honeywell com or Phone 61 7 3846 1255 Or FAX 61 7 3840 6481 hfs tac s tac support honeywell com support honeywell com Toll Free 1300 36 39 36 Toll Free 1300 36 04 70 China PRC Shanghai Honeywell China Inc Phone 86 21 5257 4568 Fax 86 21 6237 2826 Singapore Honeywell Pte Ltd Phone 65 6580 3278 F
42. ransmitters allow smart performance at analog prices Accurate reliable and stable Series 900 transmitters offer greater turndown ratio than conventional transmitters Honeywell del Figure 1 Series 900 Remote Seal Pressure Transmitters feature proven piezoresistive sensors and advanced seal technology with standard weld connections All ST 3000 transmitters can provide a 4 20 mA output Honeywell Digitally Enhanced DE output HART output or FOUNDATION Fieldbus output When digitally integrated with Honeywell s Process Knowledge System EXPERION PKS ST 3000 instruments provide a more accurate process variable as well as advanced diagnostics Honeywell s cost effective ST3000 S900 transmitters lead the industry in reliability and stability e Stability e Reliability ST3000 S900 Transmitter Benefits Stability 0 01 per year Reliability 470 years MTBF Description of Diaphragm Seals Diaphragm seals are traditionally used when a standard pressure transmitter should not be exposed to the process pressure directly Diaphragm seals typically protect the pressure transmitter from one or more damaging aspects of the process media Consideration for using a diaphragm seal should be made in the following circumstances e High Process Temperature e Process Media is Viscous or Contains Suspended Solids e Process Media is Subject to Solidifying e Process Media is Corrosive e Process Applicati
43. s are TEMPORARY ONLY to protect threads and MUST be REMOVED before installation 7 Tantalum Body has Tantalum wetted parts and 316 SS non wetted parts Note Remote seal system pressure rating is body rating or seal rating whichever is less 34 ST 3000 Smart Pressure Transmitter 34 ST 16U 34 Issue 47 Page 6 of 11 Availability STR9xx TABLE SEALS continued eal Pressure Rating Diaphragm Threaded Process Connection Diameter i NPT Female Bolts 1 2 NPT 3 4 NPT 2 500 psi 1 250 psi 1 NPT 1 2 NPT 3 4 NPT 2 500 psi 1 250 psi 1 NPT 1 2 NPT 3 4 NPT 1 500 psi 750 psi 1 NPT ppp NN NN NE NN i 316L SS Carbon Steel 316L SS 316L SS Wetted Material Hastelloy ere 91 ipid g Hastelloy C 276 Hastelloy C 276 Monel 400 Monel 400 Tantalum 316L SS Seal with Tantalum Hastelloy C 276 Threaded Non W etted Material CS Nickel Plated um wmm ume DR Connection Carbon Steel G Flushing Connections a 1 4 with plastic plug and Plugs One 1 4 with metal plug Metal plug material Two 1 4 with plastic plugs will be the same as Two 1 4 with metal plugs Lower material if One 1 2 with plastic plug metal plug is chosen One 1 2 with metal plug SS Plug for CS Lower Two 1 2 with plastic plugs and Tantalum Clad Two 1 2 with metal plugs Klinger C 4401 non asbestos Grafoil Teflon Gylon 3510
44. tial requirements of the directive 97 23 EC PED Annex l and shall not have the CE mark applied European Pressure D SEDE Equipment For transmitters rated gt 200 bar 2 900 psig lt 1 000 bar 14 500 psig Honeywell maintains a technical file in accordance with Annex III Module A internal production control when the CE mark is required Transmitter Directive PED Attachments Diaphragm Seals Process Flanges and Manifolds comply with Sound Engineering Practice 97 23 EC NOTE Pressure transmitters that are part of safety equipment for the protection of piping systems or vessel s from exceeding allowable pressure limits equipment with safety functions in accordance with Pressure Equipment Directive 97 23 EC article 1 2 1 3 require separate examination A formal statement from TUV Industry Service Group of T V America Inc a division of T V S ddeutschland a Notified Body regarding the Pressure Equipment Directive can be found at www honeywell com A hard copy may be obtained by contacting a Honeywell representative Electro Magnetic Compatibility EMC 2004 108 EC CE Mark All Models EN 50081 1 1992 EN 50082 2 1995 EN 61326 1 1997 A1 A2 and A3 Industrial Locations Dual Seal Dual Seal Certification based on ANSI NFPA 70 202 and ANSI ISA 12 27 01 requirements without the use of Certification additional seal protection elements Recommended Frequency of Honeywell recomm
45. tion Field Parameters Explosionproof Class I Division 1 Groups A B C D locations o Dust Ignition Proof Al d TEACH Class Il Division 1 Groups F G locations Enclosure Type 4X Vmax 42 4V Imax 225mA RUE 4 20 mA DE Ci 4 2nF 4 Ta 93 C Intrinsically Safe Li 2 Pi 1 2W Class 1 Il Ill Division 1 Groups A B C D E F G locations Enclosure Vmax 30V Tvpe 4X 225 yp 4 20 mA HART Ci 4 2nF T4 Ta 93 C Li 1 2W Vmax 32V Fieldbus Entity T4 40 Not FISCO Liz 0 T3 Ta 93 C Intrinsically Safe Pi 0 84W Vmax 24V Class Il 111 Division 1 Groups B C DEF Gc w Fieldbus Entity 2 T4 40 Riga Not FISCO Liz 0 T3 Ta 93 C FM sw Class 1 Zone 0 AEx ia Group IIC Pi 1 2W Approvals Enclosure Type 4X IP 66 67 Vmax 17 5V Imax 380mA eal T4 Ta 40 C FISCO oo T3 Ta 93 C Pi 5 32W Vmax 42 4V Imax 225mA 4 20 mA DE Ci 4 2nF T4 Ta 93 C Nonincendive Li Pi 1 2W Class Division 2 Groups A B C D Vmax 30V locations Enclosure Type 4X Imax 225mA 4 20 mA HART Ci 4 2nF T4 Ta 93 C Li Pi 1 2W Vmax 32V Nonincendive Fieldbus Entity ae T4 Ta 40 C Not FNICO Liz 0 T3 Ta 93 C Class I Division 2 Groups A B C Pi 0 84W D Vmax 24V Suitable for Fieldbus Entity ni us TA Ta 40 C Class II Division 2 Groups F amp G Not FNICO 1 0 T3 Ta 93 C Cl
46. um pressure differentials to be measured Figure 17 PMin SGp x a SGf x d LRV when HP at bottom of tank URV when LP at bottom of tank PMax SGp x b SGf x d URV when HP at bottom of tank LRV when LP at bottom of tank Where minimum level 4mA maximum level 20 mA a distance between bottom tap and minimum level b distance between bottom tap and maximum level d distance between taps SGf Specific Gravity of capillary fill fluid See Page 9for values SGp Specific Gravity of process fluid 17 Max Level Min Level X Transmitter above datum Max Level Min Level X Transmitter at datum Max Level Min Level X Transmitter below datum 24253 Figure 17 Closed tank liquid level measurement distance 18 ST 3000 Smart Pressure Transmitter Density or Interface Calculate the minimum and maximum pressure differentials to be measured Figure 18 Pmin SGmin SGf x d minimum density 4mA output Pmax SGmax SOf x d maximum density 20mA output Where d distance between the taps 24257 SGmax maximum Specific Gravity SGmin minimum Specific Gravity SGf Specific Gravity of capillary fill fluid See Page 9 for values Figure 18 Density direct actin