Leica MATS Type TL - Leica Microsystems

image

Contents

1. A A 170 134 98 Y Y 140 ne p 190 gt IN 220 Dimensions in mm 8 Technical Data Leica MATS Thermocontrol System Power supply Operating environment Transport and storage Power cable Disposal 100 240 V 10 alternating current 50 60 Hz 0 68 A protection class only to be used in enclosed rooms Temperature 5 C to 40 C maximum relative humidity 35 to 80 non condensing Elevation up to max 2000 m Ambient conditions installation category Il according to IEC 664 Pollution degree 2 Transport and storage at 25 70 C and max 80 humidity non con densing with power supply from 100 up to 120 V only use the following power cables 3 wire grounded instrument connection cable SVT No 18 AWG UL listed not permanently connected to the instrument nominal data at least 125 V 7A When using the extension cable use only power cables with protective ground With power supply from 220 up to 240 V Use only a three pin power cable plug and socket according to EU EN standards within the EU When using Class accessories use a connection with protective ground When using the extension cable use only power cables with protective ground Once the product has reached the end of its service life please contact Leica Service or Sales about disposal Please observe and comply with the national and federal laws and regulations that are equiva
2. User Manual Living up to Life Seica MICROSYSTEMS Copyright General Notes Copyright All rights to this documentation are held by Leica Microsystems CMS GmbH Reproduction of text or illustrations in whole or in part by print photocopy microfilm or other method including electronic systems is not allowed without ex press written permission from Leica Microsys tems CMS GmbH The instructions contained in the following docu mentation reflect state of the art technology We have compiled the texts and illustrations as ac curately as possible Still we are always grateful for comments and suggestions regarding poten tial mistakes within this documentation The information included in this manual may be changed without prior notice Revision 1 0 published May 07 2013 by Leica Microsystems CMS GmbH Ernst Leitz Strae 17 37 D 35578 Wetzlar Germany http www leica microsystems com Responsible for contents Marketing CMS Contents o N m m wm Description e 4 Safety Concept ss 6 Safety Instructions 7 Design me 10 Assembly sace 11 12 Size and Weights 13 Technical 14 1 Description MATS for TL bases Microscope stage Auto matic Thermo
3. Repairs service work Repairs may only be carried out by Leica Mi crosystems trained service technicians Only original Leica Microsystems spare parts may be used Disconnect power supply during maintenance and repair Avoid contact with powered elec trical circuits which can lead to injury Power cable Only use the power cables specified on p 14 Regularly check if the power cable is intact Replace defective power cables immediately because they can cause the instrument and other equipment to become electrically live and injure people Ensure that the cable is routed carefully Pre vent people from getting caught in cables or hurting themselves while tripping The instru ment could tip over and be damaged Avoid pulling the power cable Liquids Exercise particular care when handling liq uids Spilled liquids electrical instru ments can cause the instrument and oth er equipment to become electrically live People could be injured or the instruments could be damaged 3 Safety Instructions Cleaning care Carefully handle your Leica MATS Thermocontrol System for TL bases Unplug the power cable before cleaning Do not open or disassemble the control unit under any circumstances Observe warning messages when handling liq uids Do not immerse the instrument in water or sol vent Never use volatile materials such as gasoline or thinners for wiping The color of the surface could ch
4. ange or peel off Imprinted letters could also be removed Avoid anything that could scratch the thermo plate N wae If contaminated carefully clean the control unit and thermoplate with a soft cloth that has been immersed in a water diluted cleaning agent 4 Design 4 Design Components 1 Control unit 2 Thermoplate type 10450620 for Leica stereo microscopes with TL3000 ST TL4000 BF DF TL4000 RC I or TL5000ergo transmitted light base Heating stage 10 450 620 5 Assembly 9 Assembly Thermoplate Control units e To reach the correct temperature remove the e Connect the thermoplate to the control unit glass stage plate from the base of the stereo Plug the power cable into the input socket of microscope the heating unit Position the thermoplate according to the fig ure on the specimen stage of the stereomicro For compatible power cables see p 14 scope Heating stage 10 450 620 for Leica stereomicroscopes with TL3000 ST TL4000 BF DF TL4000 RC I or TL5000ergo transmitted light base TL TL RC and TL RCI 11 6 Operation Controls 1 Power switch 2 Temperature control 3 Adjusting temperature up down buttons 4 PV display shows the process value and differ ent characters 5 SV display shows different configured values oF Before activating the Leica MATS Thermo control Sy
5. control System for transmitted light bases Intended use The Leica MATS Thermocontrol System for TL bases is a heating system for stereomicroscopes and enables the observation of sensitive micro scopic specimens under accurate temperature conditions This includes examining specimens taken from the human body for the purpose of gaining information relating to physiological or pathological states or inborn anomalies to examine safety and compatibility with potential recipients or to monitor therapeu tic measures The Leica MATS TL Heating System complies with Council Directive 98 79 EC concerning in vi tro diagnostics This safety class 1 instrument was built and tested in accordance with the harmonized stan dards EN 61010 1 2001 Safety requirements for electrical equipment for measurement control and laboratory use and EN 61010 2 101 2002 Safety requirements for electrical equipment for measurement control and laboratory use part 2 Particular requirements In vitro diagnostic IVD medical equipment This IVD instrument is not intended for use in a patient environment as defined by DIN VDE 0100 710 It is also not intended to be combined with medical devices as defined by EN 60601 1 If a microscope is electrically connected to a medi cal instrument in accordance with EN 60601 1 the requirements defined in EN 60601 1 1 shall apply 1 Description User Manual N wx This User Manual describes the as
6. copes transmitted light stands The thermoplate and control unit are matched AS we Use the thermoplate and control unit from the same package never from different packages the serial numbers of both units must be identi cal Non intended use Using the Leica MATS TL Thermocontrol System in any way other than that described in this User Manual could result in personal injury or damage to property Never Install other plugs or cables Change or alter the Leica MATS Thermocontrol System for TL bases Take apart parts if not specifically instructed to in this User Manual Have the Leica MATS Thermocontrol System for TL bases opened or maintained by unau thorized persons Reuse a damaged thermoplate 3 Safety Instructions Place of use The Leica MATS Thermocontrol System for TL bases may only be used within closed rooms The following areas of use are prohibited Locations in which flammable gases corrosive gas oil vapors and substances that can impair electrical insulation develop or arise in large quantities Locations in which strong vibrations or im pacts occur or are transmitted Locations where high voltage lines are in the area or in which induction interference can af fect the operation of the Leica MATS Thermo control System for TL bases Locations in which condensate or humidity de velops or direct sunlight can occur Temperature regulation The process value shown o
7. lent to EC directives such as WEEE Like all electronic devices neither the product nor its accessories may be disposed of as general household waste 14 8 Technical Data Temperature control Temperature indicator Duration Method the temperature control is based on the PID control unit with solid state relay Achieved accuracy T actual temperature in the center of the heating stage is within a range of t gt T gt t 1 C where t stands for the set tem perature under the condition that the set value is t 37 C Increment 0 1 C Method of adjustment with up down button Range of adjustment room temperature up to 50 C Attainable temperature accuracy 0 3 C with indicator temperature Sensor thermocouple Connection to heating stage 4 pole plug cable length 1000 mm Method of display digital display with 7 segments and individually illumi nated indicator Increment 0 1 C Accuracy of display 0 5 50 C within 10 minutes 15 6102 1 uoisinoy y e319 JO 51 1 1 1 soDo e3197 pue v2I31 6682 62 1099 6t XeJ 0 62 1099 Gy BUOY 102 1e 219 B GGE 95Je I1S 219 19U13 e HW SIN suiejsAso421 A 29197 1 MICROSYSTEMS www leica microsystems com
8. n the control unit cor responds to the temperature of the stage sur face If the temperature of the specimen in a pe tri dish or on a specimen slide is lower than the displayed process value the programmed value must be modified in order to obtain a suitable specimen temperature Otherwise the specimen could be damaged A malfunction of the sensor can cause the surface of the thermoplate to become very hot Protective measures of the manufacturer Dielectric strength 2 sec at 1350 V alternating current between power supply and grounding jack Insulation resistance at 500 V more than 100 MO minimum between power supply and grounding jack Fire protection temperature cannot be set over 50 C Fuse T1 6A 250VAC According to Directive 2004 108 EC the Leica MATS Thermocontrol System for TL bases is electromagnetically compatible and can be used with other electric instruments 3 Safety Instructions Responsibilities of person in charge of instru ment Ensure that The Leica MATS TL Thermocontrol System is operated maintained and repaired only by au thorized and trained personnel All operators have read understood and ob serve this stereomicroscope s User Manual and particularly the safety regulations The heating plate is intact before every use If the plate is defective stop using it immedi ately Otherwise an uncontrolled temperature increase and damage to the specimen can re sult
9. sembly and handling of the Leica MATS Thermocontrol System for TL bases With your instrument you receive a printed user manual in your local lan guage Further language versions and_ information can be found on the interactive CD ROM that you received with your Leica stereomicro scope or macroscope User manuals and up dates can be downloaded on our website www stereomicroscopy com For detailed information relating to the micro scope stereomicroscope or macroscope please see the separate user manual for your instru ment 2 Safety Concept Before assembling the Leica MATS Thermo control System for TL bases and putting it into operation please read this user manual and observe the safety instructions Legal regulations Adhere to general and local regulations relating to accident prevention and environmental pro tection EC Declaration of Conformity The Leica MATS Thermocontrol System for TL bases is constructed based on the state of the art of technology and is provided with an EC Decla ration of Conformity Intended use The Leica MATS Thermocontrol System for TL bases is a heating system for stereomicroscopes and is intended for the observation of sensitive microscopic specimens under accurate temper ature conditions The thermoplate heating stage can be heated up to 50 C The Leica MATS Thermocontrol System for TL bases consists of a control unit a power cable and a thermoplate for Leica stereomicros
10. stem for TL bases please read the safety instructions on p 6 9 Press the power switch 1 After completing work switch the instrument off at the power switch and disconnect the power plug 12 Temperature control The thermoplate heating stage can be heated up to 50 The process value shown on the control unit corresponds to the temperature of the stage surface If the temperature of the specimen in a petri dish or on a specimen slide is lower than the displayed process value the programmed value must be modified in order to obtain a suit able specimen temperature Otherwise the spec imen could be damaged N ws Make sure that when using the stereomicro scope the glass stage plate is removed before disconnecting the heating stage from the base Otherwise the correct temperature is not guar anteed Adjusting the temperature To adjust the temperature press the up down but tons If the button is held for one second or lon ger the displayed temperature changes continu ously Two seconds after completed adjustment the temperature control is activated to reach the defined temperature Set the temperature control using the buttons 2 Wait 10 minutes until the thermoplate s tem perature has stabilized 7 Size and Weights Control unit Weight approx 1 3 kg Dimensions width 113 mm length 210 mm height 128 74 mm Thermoplate 10 450 620 Weight approx 170 g