ELISA PLATE SHAKER – INCUBATOR

image

Contents

1. Version 150511 PI number 1701329 Page 8 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 4 PACKING TRANSPORT AND STORAGE 4 1 General warnings Instrument has to be decontaminated before packing for transportation 4 2 Packing Packaging is needed whenever it is to be transported or shipped by courier or other means To pack the instrument follow the instructions as below described Decontaminate the instrument as explained in chapter No 13 Decontamination of this manual Place the instrument into the original packaging box Instrument has to be properly protected by plastic protective material Put copy of safety clearance certificate copy of Safety Clearance certificate is attached at the end of this manual Mark the package with address instrument identification and warning Labels 4 3 Instrument Transportation The transportation of the instrument in unpacked condition must be limited within the room where it is used to avoid damage 4 4 Storage of the Instrument Before storing the instrument for a long period pack it carefully as described above and store indoors Relative humidity can be up to 80 and temperature between 10 c and 35 c Version 150511 PI number 1701329 Page 9 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 5 INSTRUMENT DESCRIPTION Components of different views of the below pictured instrument 5 1 Perspective View A Front View Incubator Sh
2. 7 PRECAUTIONS Precautions Keep the place dry and clean Check all the grounding wires Use original packaging for transportation Version 150511 PI number 1701329 Page 12 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 8 GENERAL KEY AND OPERATION 8 1 KEYPAD 8 1 1 START YES KEY This option helps the user to START the process of Shaking Similarly YES key is used when the instrument asks certain questions as follows For eg CREATE TEST Y N OR START SHAKE Y N 8 1 2 STOP NO KEY This option helps the user to STOP the process of Shaking Whereas NO key is used to answer certain questions such as For eg CREATE TEST Y N OR START SHAKE Y N In such a case user must select either YES or NO options to proceed further Version 150511 PI number 1701329 Page 13 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 8 1 3 SHAKE KEY User can directly start the process of Shaking just by selecting SHAKE key present on the keypad SHAKE START SHAKE Y N YES SET RPM 400 Enter the RPM between 400 to 700 and press Enter key Its default value is 500 RPM PRESS ENTER KEY Finally Shaker starts at 400 RPM This process continues until user selects STOP button present on the keypad For Terminating the process select STOP
3. A solution of 0 5 Sodium Hypo Chlorite NaOCL solution Bleach is used Commercial bleach is 5 NaOCL household bleach is 3 NaOCL When using commercial bleach use a 10 1 mixture if using household bleach a 6 1 mixture is required This is a caustic solution It is important to wear gloves and eye protection when handling it Wipe down the carrier and all exposed surfaces of the unit with the bleach solution Remove the top shroud of the instrument and wipe down the top surface of the instrument base as well as the inside of the top shroud Reassemble the unit and discard the used gloves and towels PI number 1701329 Page 24 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 14 SAFETY CLEARANCE CERTIFICATE Please complete all information requests on this form prior to returning the instrument to the manufacturer or your local distributor for servicing repairs or return Thank you for your co operation Customer 2 Contact Address Position Dept Tel Country tA Fax Post Code Model No Serial No Accessories Returned Date of Purchase if known Complaint Has the equipment been exposed to any of the following delete as applicable a Blood body fluids pathological specimens YES NO If YES please specify b Other Biohazard YES NO If YES Please specify Version 150511 PI number 1701329 Page 25 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 15 SPARE PARTS 15 1 Ordering Spare Par
4. 9 4 3 INSTRUMENT TRANSPORTATION ccccccecsccessccessecesssccssscesssecsesecsssecesscesscesssesesseceasesesasessusessseesaresensess 9 4 4 STORAGE OF THE INSTRUMENT ss cccessssecesssececsssescevesseseccasseseseuseesevecseseecassusecenscesaseuseeeevessusescaseusevens 9 5 INSTRUMENT DESCRIPTION retient rhet ret S r a eek SP Praesta Mn nee 10 5 1 PERSPECTIVE VIEW Leite re ett ettet case bite e est beet teet cuss bite e LO A FRONT itte Retiro n reb Pee er na Er rego tron Presse sese ele eese LO B REAR VIEW 5 te tm HD t Un oH MG EU S 10 6 INSTALLATION PROCEDURE AND VERIFICATION CRITERIA eere 11 6 1 UNPACKING INSTRUCTIONS 5 5 sti N 6 2 PLACING THE INSTRUMENT sen ro treo aee peor eee er nent se LT 6 3 POWER SUPPLY REQUIREMENTS 11 6 4 PROTECTIVE GROUNDINGS d Ee nre d EE Ee 11 6 5 START UP INSTRUCTIONS 2 2 cscs decet eet o eter eh eet ba o Dua d eR ere aget 11 7 PRECAUTIONS ccir 12 8 GENERALE KEY AND OPERATI ON nn etit eati eoa ch bo Rr ve or ees ed vpn eos eco ese Une eo te 13 8 T4 KEYPAD 4 recie Ee te Rede Te ee oe ERR Rete ete e ene te ined 13 BAT START YES KEY e
5. ELISA PLATE SHAKER INCUBATOR 11 RECALLING THE TEST MENU ONCE ENTER TEST NO Enter the Test No and select ENTER key TEST RECALLED After Recalling the test it will flash following Test parameters and finally ask the user to Load the Plate and press Enter key TEMP 37C RPM 500 TIMER 1 10 TIMER 2 20 LOAD PLATE PRESS ENTER Load both the Plates and press ENTER key Since the Timer is set it will beep twice after reaching its time set for both the corresponding plates Note Before loading the plate it will attain its set temperature Version 150511 PI number 1701329 Page 22 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 12 ERROR MESSAGES MESSAGES CAUSE CORRECTIVE ACTION This message appears only when the test no is recalled Test Not Exist which does not exist Test Present When test is already present Invalid RPM This message appears when the entered RPM Value is out of the given range This message appears when the entered Timer Value is Invalid TM1 TM2 Val out of the given range This message appears when the entered Temperature Invalid Temp Value is out of the given range While deleting any test if user enters the test no which Invalid Test No does not exists then it will show the corresponding messa
6. button STOP RPM 400 STOP Y N Select option in order to continue the _ process of Shaking 207 c MOM UM On selecting YES option it STOPS the process of Shaking Version 150511 PI number 1701329 Page 14 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 8 1 4 FUNC KEY FUNC Press FUNC button present on the keypad CREATE TEST Y N YES SET RPM 400 Enter the RPM between 400 to 700 and press _ Enter key Its default value is 500 RPM SET TIMER 1 50 Set TIMER1 amp TIMER2 between 1 999 min and mem press Enter key Its default value is 30 min SET TIMER 2 70 SET TEMP 37C j Set TEMP between 37 C to 42 C and press Enter key Whereas its default value is 37 C SAVE TEST NO 5 j Enter the Test no in order to save the test parameters Remember user can save maximum 9 tests TEST SAVED i DELETE TEST Y N NO YES ENTERTESTNO __ CONFIRM YES TEST DELETED INITIAL SCREEN Version 150511 PI number 1701329 Page 15 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 8 1 5 MENU KEY This option helps the user to recall the saved test by entering its test no MENU ENTER TEST NO Enter the Test No and
7. 2 C Incubation Time Timer 1 30 min 1 to 999 min Timer 2 30 min Version 150511 PI number 1701329 Page 19 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 9 SAVING THE TEST Saving the Test itself means creating new test with new parameters FUNC Press FUNC button present on the keypad TEST YES Enter the RPM between 400 to 700 and press SELBEN EU 777 Enter key Its default value is 500 RPM 22 SET TIMER 1 50 Set TIMER1 amp TIMER2 between 1 999 min and 77 Enter key Its default value is 30 min a eed SET 2 70 SET TEMP 37C Set TEMP between 37 C to 42 C and press Enter _ key Its default value is 37 C Enter the Test no in order to save the test SAVE TEST NO 5 parameters Remember user can save maximum 9 tests TEST SAVED Version 150511 PI number 1701329 Page 20 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 10 DELETING THE TEST One can Delete the saved test by following the procedure mentioned below FUNC CREATE TEST wo DELETE TEST Y N YES ENTERTESTNO ___ CONFIRM Y N NO TEST DELETED NO gt INITIAL SCREEN Version 150511 PI number 1701329 Page 21 of 26 User Manual
8. PARE PARTS 15 1 ORDERING SPARE P ARTS 3 eb rote eee e ah cok AR Version 150511 PI number 1701329 Page 3 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 1 GENERAL INFORMATION 1 1 Warranty Information Each Instrument is completely tested and guaranteed for twelve months from delivery The warranty applies to all the mechanical and electrical parts It is valid only for proper installation use and maintenance in compliance with the instructions given in this manual DiAsource ImmunoAssays S A will at its discretion repair or replace parts which may be found defective in the warranty period The warranty does not include any responsibility for direct or indirect personal and or material damages caused by improper use or maintenance of the instrument Parts that are inherently subject to deterioration are excluded from the warranty In case of defects due to misuse of the instrument any incidental expenses like travel and man hour service charges will be charged extra 1 2 Technical Service DiAsource ImmunoAssays S A is always accessible to the customers for any kind of information about installation use maintenance etc While asking for service please refer to this manual and report the printed serial no on the identification label Only qualified technicians are entitled to fix the instrument the user as described in this manual should carry out ordinary maintenance 1 3 D
9. R 1 UMOR TIMER YES Version 150511 PI number 1701329 Page 17 of 26 User Manual ELISA PLATE SHAKER INCUBATOR e TIMER 2 Select NO option in order to continue the TIME 10 STOP Y N TIMER 8 1 7 CLEAR KEY This key is used to select the Language option and also to clear the test entry In other words user can abort the test by pressing CLEAR key CLEAR FRENCH LANGUAGE Y N Select Yes for French Language or otherwise No to continue in English Version 150511 PI number 1701329 Page 18 of 26 User Manual 8 1 8 TEMP KEY TEMP 4 ELISA PLATE SHAKER INCUBATOR Press TEMP button present on the keypad ROOM TEMP YIN INITIAL SCREEN vo 37 C SET TEMP Y N rene eter voie eise YES SET TEMP 37 C Incubator Heating OFF Room Temperature It will set to room temperature Set value and press ENTER Key XX 37 C to 42 C NOTE By Default condition incubator heating is off Instrument will show Room Temperature 8 1 9 ENTER KEY Enter button present on the keypad helps the user to set the value for the selected parameters ie on completion of data entry 8 2 DEFAULT SETTING OPERATING MODE PARAMETERS DEFAULT VALUES RANGES RPM 500 RPM 400 to 700 RPM Temperature range 37 37 C to 4
10. User Manual ELISA PLATE SHAKER INCUBATOR ELISA PLATE SHAKER INCUBATOR DIA4000 DIA ee DlAsource ImmunoAssays S A Rue du Bosquet 2 B 1348 Louvain La Neuve BELGIUM Phone 32 10 84 99 11 Fax 32 10 84 99 90 www diasource be IVDInstrumentation support 2 diasource be User Manual ELISA PLATE SHAKER INCUBATOR TABLE OF CONTENTS 1 GENERAL INFORMATION m 4 1 2 WARRANTY INFORMATION tette ec eee et diete eb nd nn 1 27 TECHNICAL SERVICE dar conoce c e dan ed ceret a do rt c at tco cca 1 3 DISPOSAL 505 en e teal la Ee b eet t ee dee tee 4 MEAS CONTACTS 2 GENERAL SAFETY WARNINGS cicer etienne inten nn torna tite n Gn vo aeo oss seesi sisirin 6 2 1 DANGER WARNINGS SYMBOLS ccsssccccsscsssscnssssecscesssssnnsseesssessssnnsssesescsscssansseesesessesansssesssssssenanees 6 2 2 USE OF THE INSTRUMENT 2 n od re a E d a eed rb re Pope ere el i 7 INTRODUCTION E EE 8 3 1 SPECIAL FEATURESS e ue e te dinates tnt ci ae oO 327 45 PEGIFIGATIONS 8 4 PACKING TRANSPORT AND 9 GENERAL WARNINGS 5 ena soos teas ser vr ede ia 4 2 APACKINGE 5 1 cte tot E bi mnt c ut E
11. aker Plate holder B Rear View On Off switch Cooling Fan 8 Input power socket Version 150511 PI number 1701329 Page 10 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 6 Installation procedure and verification criteria 6 1 Unpacking instructions Check accessories as per packing list CE Kindly store all packing materials so as to use it to repack and ship for maintenance or servicing 6 2 Placing the instrument The instrument has to be placed on a level bench Room temperature has to be between 10 and 35 C with a relative humidity below 85 Protect it from direct sunshine 6 3 Power supply requirements Once the instrument has been placed plug it into a power source by the locally available approved plug in cable Power cord should be CE CSA and UL marked 115 230 Volts 10V 60 50 Hz 6 4 Protective Grounding Warning Please make sure that electrical power source is properly grounded 6 5 Start up Instructions o Switch on the instrument The instrument will display the company name o The instrument initializes all the parameters internally and carries out a power on self test It then displays company name The temperature control of the plate starts The temperature of the plate will maintain at 37 0 5 o The instrument is now in IDLE mode and ready for use Version 150511 PI number 1701329 Page 11 of 26 User Manual ELISA PLATE SHAKER INCUBATOR
12. ge Version 150511 PI number 1701329 Page 23 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 13 DECONTAMINATION Check whether all the tubing is properly connected Switch on the instrument 13 1 13 2 Decontamination Procedure If the instrument is to be shipped after being exposed to potentially hazardous material it should be decontaminated The following procedure outlines the method to decontaminate the instrument before packaging and shipment Purpose of Decontamination Decontamination minimizes the risk to all who come in contact with the instrument during shipping handling and servicing 13 3 13 4 Version 150511 General Considerations Any laboratory instrument that has been used for clinical analysis is considered a bio hazard and should be decontaminated prior to handling Intact skin is generally considered an effective barrier against infectious organisms however small abrasions and cuts may not be always visible Prophylactic gloves must be worn when handling instruments that have not been decontaminated Gloved hands should be considered contaminated and must be kept away from eyes mouth and nose at all times Mucous membranes are considered as the prime entry routes for infectious agents Wear eye protection and a surgical mask when there is a possibility of aerosols Eating and or drinking while decontaminating instruments is not advisable Decontamination Procedure
13. he original spare parts gt Maintain room temperature and humidity as specified in the manual gt The instrument must be used as described in this manual only Usage in any other way will be regarded as improper gt Alterations to the instrument are strictly prohibited The user is liable and solely responsible for any improper modification to the instrument and for the consequences derived as a result Version 150511 PI number 1701329 Page 7 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 3 INTRODUCTION 3 1 Special Features Simultaneous shaking amp incubation operation Buzzer indication on completion of incubation Indication of Remaining time Current temperature of incubation on display on pressing TEMP Key A Separate Timer ON indication Keypad 3 2 Specifications Operating Modes Shaker Incubator Shaking amp Incubator Temperature Control Temperature Range 37 C to 42 C Resolution 1 C Incubation Time 1 to 999 min Shaker Frequency Amplitude 400 to 700 RPM 2mm Operating Position On horizontal flat rigid and vibration free surface Operating Conditions Temperature Relative Humidity From 18 C to 32 C Up to 80 Storage Conditions Temperature Relative Humidity From 10 C to 35 C Up to 80 Enclosure ABS Fire retardant Size cm 28 X 25 X 16 LX b X h Weight 3 Kgs Approx
14. isposal instruction In case of removal or disposal of instrument following instructions need to be followed e Do not dispose in municipal waste follow local regulations for instrument disposal Plastic parts Electronic PCBs and components can be recycled so return back the instrument to manufacturer Version 150511 PI number 1701329 Page 4 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 1 4 Contacts DlAsource ImmunoAssays S A Rue du Bosquet 2 BE 1348 Louvain La Neuve BELGIUM IVD Instrumentation Support Service Hotline phones availabilities Monday to Friday 08 30 to 17 00 Belgium time Please contact first Hotline 1 In case of no response please contact Hotline 2 Hotline 1 phone number 0032 0 10 849932 Hotline 2 phone number 0032 0 10 849976 Fax 0032 0 10 849990 Email IVDInstrumentation support diasource be Version 150511 PI number 1701329 Page 5 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 2 GENERAL SAFETY WARNINGS 2 1 Danger warnings symbols The following symbols are used to inform the user of the safety rules 94 BEE General Symbols Version 150511 This symbol indicates generic danger It means that serious damage can occur to the operator if described precautions are not observed This symbol indicates HIGH ELECTRIC VOLTAGE It is dangerous to touch any part having this label Only qualified operators can access these components after unplugging the instru
15. ment from the Supply This symbol indicates that the instrument involves the handling of samples which can be infected urine or human serum In this condition infection or contamination might occur Pay attention to the general safety warnings when in presence of such biological substances Use Protective clothes gloves and glasses This symbol in the user manual indicates that damages to the instrument or erroneous results could occur if the given warnings are not followed This symbol indicates a portion which is particularly important and should be studied carefully This symbol indicates a Protective Earth or Ground terminal Symbol for Manufacturer PI number 1701329 Page 6 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 2 2 Use of the Instrument The instrument has to be used for the designed purposes under specified conditions following proper procedures and safety rules by qualified personnel THIS MANUAL CONTAINS INSTRUCTIONS FOR OPERATION BY QUALIFIED PERSONNEL ONLY gt A qualified user has to make sure that the environmental condition is suitable the installation is correct the use and maintenance are proper according to the general safety rules as well as to the particular precautions described in the manual However the user is not entitled to repair the instrument gt A qualified technician is entitled to maintain and fix the instrument according to the instructions given using t
16. nse e Re N lo diit lc eee eel eet 13 8 1 2 STOP ZNO eee e e ee e e E e c e e Ms 13 STS SHAKE KEY s tuu utei NE 14 8 4 FUNC KEY eiiis onmi tin nieder quentin 15 81 5 MENU KEY it et temo eiie Mieten ie 16 8 1 6 TIMER KEY RR GP A 17 S7 CLEAR KEY iiis eee Ue nans 18 LONE Jud REY mE 19 STD ENTER KEV EE 19 8 2 DEFAULT SETTING OPERATING MODE LO 9 SAVING THE TEST iore eene trier esc cette ete aed erbe SAS 20 10 DELETING THE ee tree tee tette qe Ie PE Tee ttes on eee crane 21 11 RECALLING THE TEST 22 2 2 oor 23 13 24 150511 PI number 1701329 Page 2 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 13 1 DECONTAMINATION PROCEDURE eee eee nennen nennen eren 13 2 PURPOSE OF DECONTAMINATION eese nennen entren entren entrar scene scene ee irte dran eite 24 13 3 GENERAL CONSIDERATIONS 12 52 55 err re E t et E PEOR FEE Pa P t reete add 13 4 DECONTAMINATION PROCEDURE 24 14 SAFETY CLEARANCE CERTIFICATES 22 2 2 2 ne ettet eto eeu oai orae erento rreri soia ere 25 15 S
17. select ENTER key TEST RECALLED Enter the Valid Test No or else it will display following string INVALID TEST NO Since user can save maximum 9 tests one has to enter the test no in between 9 in order to recall the test If user enters the number which donot exist or in other words the test is not saved with that particular number then in such a case it will display following string TEST NOT EXIST Version 150511 PI number 1701329 Page 16 of 26 User Manual ELISA PLATE SHAKER INCUBATOR 8 1 6 TIMER KEY This option is used to set the timer for the respective plates TIMER TIMER 1 Y N YES START TIMER YES SET TIMER 1 30 Enter TIMER1 between 1 999 min and press Enter key Its default value is 30 min TIMER2 Y N YES START TIMER Y N YES SET TIMER 2 30 Enter TIMER2 between 1 999 min and press Enter key Its default value is 30 min NO After selecting Timer and Timer2 LED glows of the respective Timer On completion LED gets OFF and the buzzer beeps in order to indicate that the Timer has reached its set value Now select Enter key to stop the buzzer The TIMER continues until user selects STOP button present on the keypad For Terminating the process select STOP button STOP TIME
18. ts Parts subject to deterioration or defectives which need to be replaced have to be ordered by giving following details Ordering the spare parts the following data are to be mentioned Customer s purchase order No Name and version of the instrument Instrument code number Part code number Description of the part Requested quantity Name and company address for delivering the ordered goods While replacing the parts the use of ORIGINAL SPARE PARTS guarantees the efficiency and a lasting instrument life Version 150511 PI number 1701329 Page 26 of 26