         Contents
1.   Totalizer Totalizer  FlowAlarm   GCEZIEGL    oo a l Out  Flow V Out  Flow       _set_     set     Pulse Out Setup    _Return      Process  munit       1  Tap    SETUP    2  Tap    Output Config    3  Tap    Pulse Out Setup    4  Select    Flow Alarm     5  Enter Set point Value  6  Select    Pulse Transition       This page is subject to proprietary rights statement on last page  19 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC    Appendix B  List of Configuration Software Commands    The following is a summary list of Configuration Software Commands     Process     Displays Process data  no sub menus     Unit ID     Display Instrument manufacturing information  no sub menus     Setup     Instrument Setup functions  up to 3 levels of sub menus   Units  KFactor  Line Size  Temp   Flow  Totalizer  Output Cal  LCD  Output Config  Pulse Out Setup    Diagnostics     Troubleshooting Functions  no sub menus   Reset STS0 ST75  Display Parameter Memory  Calibration Values  Factory Restore  Process  amp  System Faults    This page is subject to proprietary rights statement on last page     20    ST50 ST75 PDA SOFTWARE    O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC ST50 ST75 PDA SOFTWARE    NOTES    This page is subject to proprietary rights statement on last page  21 O6EN003372 Rev  A    ORIENTATION FLAT FlexMASSter    Y NOMENCLATURE OUTUNE DIMENSION       COMPRESSION AITING       ao  FCI   s Complete Customer Commitment  Worldwide  ISO 9001 2000 and AS9100 Cert
2.  E O E E E E EN 6  Reading and Setting the CMIN and CMAX values for Flow and Temperature                  cccccccccceeseseeeeeeeeeeeeeeeeeseeeeeseeeeeeeeeeeeeeeeeees 6  Readme and se thine the KIaetOr Serer a E AE AE E ey Aes EO A AE E E AE EARNE 7    Breer De By Boy play oe mu   meena emer pT ee 1  Contiourino and Setting up the A malo OUPS iriiri nnan EEEE E EAEE EE AAEE AE EAEE E A AE AEEA AE  i  Contouring and settine the Pulse OutChanne Sercio A A NE E A A AAE ENA A E EE 8  Chapter 3  Using the Configuration Software to Read the ST50 ST75 Process Information                          ccccccccceeeeeeeeeeees 9  Genine Proces Dal iarsna ne a a E RN 9  Chapter 4  Using the Configuration Software to Read the ST50 ST75 ID Information        0     0 cc ceccessseeeeeeeeeeeeeeeeeaes 10  Getting ST50 ST75 Instrument Identification Information                  ccccceeeeseseseeeeeesessessssscssesseaesssaaaaaaeesasaaaaassaaassassaeaaaaaaaaaaaaaaaaas 10  Chapter 5  Using the Configuration Software to Access the STS50 ST75 UTILITIES Functions               eee 11  Using the Configuration Software to Reset the ST50 ST75 Flowmeter             cccccccccccccccccecceeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeees 11  Using the Configuration Software to Display the ST50 ST75 Configuration and Calibration Parameters                 ccssssssceeeeeeeees 11  Using the Configuration Software to Display and Set the Calibration Values of the ST50 ST75             cccccccccceeeeeeeeeeeeeeeeee
3.  LLC ST50 ST75 PDA SOFTWARE    INTENTIONALLY LEFT BLANK    This page is subject to proprietary rights statement on last page  6 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC ST50 ST75 PDA SOFTWARE    Chapter 2  Using the Configuration Software to Setup the ST50 ST75    The Configuration Software provides access to the 8 Instruments setup functions of the ST50 ST75  This chapter will show  you how to use these 8 setup functions of the ST50 ST75 instrument  The functions are     e Process Units selection       e Pipe Type  amp  Size selection    e Totalizer  amp  setup selection    e Local LCD setup  Line Size     Temp   Flow    e K Factor Setup    Output Cal e Temp  amp  Flow min  amp  max  o LOD   Output Config e Analog Output Channels Configuration      Return   e Pulse Output Channels Configuration    e Tap on the    Setup    button of the main screen  to open  the Setup selection screen        Setting up and Selecting Process Units    Setting up the Line Type and Size       Tap on the    Units    button to open the Process  Units selection screen     Tap on the    Get    button to read the  current setup of the ST50 ST75     Select the new Process unit to use   remember to change the Units Sticker  on front of the Instrument     Tap on the    Set    button to down load  the new Units into the ST50 ST75     From the Setup Menu tap on the    Line Size    button to  open the Line Size selection screen     Tap On the    Get All    button to read the current setup  of
4.  a diagnostic data window  tap on the    Get     button next to it     e To exit Diagnostic mode tap on the    Return    button        This page is subject to proprietary rights statement on last page  17 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC ST50 ST75 PDA SOFTWARE    INTENTIONALLY LEFT BLANK    This page is subject to proprietary rights statement on last page  18 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC ST50 ST75 PDA SOFTWARE    Appendix A  Using The Configuration Software Quick Users Guide    IR Operation    Set the JP5 jumper on the STS0 ST75 in the    IR    position  This jumper is located under the bezel on the LCD display board  The     IR    jumper position is located closest to the LCD display     Selecting the Pipe Size       ae   FF FLUID COMPONENTS  FP ho el  INTERNATIONAL LLC    Units   __KFactor__  Line Size    _ Temp  Flow      Rectangular    Output Cal    PROCESS    1D Unit SETUP  LCD     Output Config    DIAGNOSTICS     UTILITIES   Return      1  Tap    SETUP    2  Tap    Line Size    3  Select Plenum Type  4  Tap    Get All     5  Enter Dimension  6  Tap    Set     Tap    Return             Changing Units       Lime Size  MA Temp  Flow    Output Cal    PROCESS    1D Unit SETUP    LCD     Output Config    DIAGNOSTICS   UTILITIES   Return      1  Tap    SETUP    2  Tap    Units    3  Tap    Get     4  Select desired Units  amp   Tap    Set     Tap    Return       C A  2 j          Setting Pulse Output Channels as Alarm Outputs  
5.  familiar with the basic functionality of the Configuration Software  you can use the rest of this manual as a  reference guide for less common tasks     This page is subject to proprietary rights statement on last page  2 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC ST50 ST75 PDA SOFTWARE    Chapter 1  Setting Up the Configuration Software for Use With Your Handheld PDA    The Configuration Software enables you to communicate with the STS50 ST75 Flowmeter through the IrDA wireless link or the  use of a RS232 cable  The Configuration Software loads into a Palm OS based PDA     The IR connection provides the convenience of not having to open the ST50 ST75 and connect a cable to access its information   The ST50 ST75 data and parameters are organized in a way that make it easy to find the information that you need     The Configuration Software with the Palm OS PDA offers a new level of ease in using FCI   s ST50 ST75 Flowmeters     PDA Requirements  The Configuration Software requires a PDA with the Palm OS 4 1 or newer  it requires a PDA with a IrDA wireless link  The  Configuration Software requires about 200K of memory to run in the PDA     The Configuration Software has been tested with the Palm Tungsten e  the Palm Zire    71  and the Palm m5185     For downloading the Configuration Software into the PDA  the PDA will need the serial cradle cable connected to the PC with  the Palm Quick Install software installed     Installing the Configuration Software into y
6.  the ST50 ST75     Select the Pipe type  Rectangular Duct or Round Pipe  by taping on the corresponding button     Enter the Dimensions of the Plenum selected     Tap on the    Set    button to down load to the  ST50 ST75 the configuration data        e From the Setup Menu tap on the    Temp   Flow    button to open the setup screen for  CMAX and CMIN for Flow and Temperature   e Tap On the    Get All    button to read the current setup of the ST50 ST75     e Select the fields you want to change  enter the new data     e Tap on the    Set    button to down load to the ST50 ST75 the configuration data     This page is subject to proprietary rights statement on last page  7 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC ST50 ST75 PDA SOFTWARE    Reading and Setting the KFactors    From the Setup Menu tap on the    Kfactor     button to open the setup screen for the K factors     Tap On the    Get All    button to read the current  setup of the STS0 ST75     Select the fields you want to change  enter the  new values     Tap on the    Set    button to down load the  Parameter Values to the STS0 ST75        Local Display Setup       From the Setup Menu tap on the    LCD    button to open the setup screen for the ST50 ST75 Local Display     e Tap On the    Get All    button to read the current setup of the STS0 ST75     e Choose the new multiplier by taping on the desired selection         e Ifyou want to change the display to   of full scale  tap on the check box for    
7.  with a check on the check    box   O Below Min Flow e Process related Faults   O Above Max Flow Below Min Flow  O Below CMin Flow Above Max Flow  O Above CMax Flow  O Below CMin Temp Below CMin Flow  O Above CMax Temp Above CMax Flow  O Board Temp Out of Limit Below CMin Temp  O TC DeltaR Below DeltaR Min Above CMax Temp    O TC DeltaR Above DeltaR Max    TC DeltaR Below DeltaR Min  TC DeltaR Above DeltaR Max       e Instrument related Fault  Board Temp Out of Limit     This page is subject to proprietary rights statement on last page  16 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC ST50 ST75 PDA SOFTWARE    Chapter 6  Using the Configuration Software to Access the ST50 ST75 Diagnostic Functions    The Configuration Software gives access to three ST50 ST75 Diagnostic Functions  These functions help in troubleshooting the  ST50 ST75 operation  when a problem is suspected     The ST50 ST75 gives the technician a window from which to analyze the processing of the flow signal information  This window  opens into 3 areas of the signal processing sequence  The first area is the signal as it comes out of the A D in raw counts  The  second is as the signal is converted into ohms  And the third area is after the signal has been temperature corrected     e Tap on the    Diagnostics    radio button of the main screen   to open the ST50 ST75 Diagnostics Troubleshooting  selection screen     e Only one of the 3 areas of the diagnostics can be viewed  at a time     e To look at
8. FLUID COMPONENTS  INTERNATIONAL LLC    FL        PDA Software User   s Manual  For Use with FCI   s ST50 and ST75 Mass Flowmeter       Fluid Components International LLC  FCI  All Rights Reserved Doc  O6EN003372 Rev  A    Fluid Components International LLC  FCI  All Rights Reserved Doc  O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC ST50 ST75 PDA SOFTWARE    IMPORTANT  Please read the End User Software License Agreement with this product before using the accompanying  software program s   Using any part of the software indicates that you accept the terms of the End User Software License  Agreement     Contents  About ETS Mantal reee cdawardeasieseseebae is adeosee densest teed das bs eas dense aae esti dati aa adi ad eateeseeaaa 2  Chapter 1  Setting Up the Configuration Software for Use With Your PDA ou               ccc cccccccccccceceee eee eeeseeeeeeeeeeeeeeeeaaaas 3  PDA  Reguireme nt sirnioone e E ETE a eee 3  Installing the Configuration Software into your PDA            cccccceeesseessseeesesessseeesessssssessssssssseeceeeeeeeeeeeeeeseaseasasasaaasasaaaeaeaaaaaaaaaeaaaaas 3  Setting up the Configuration Software  Preferentes coson a a A ouees 4  Connecting 10 Hess 1 50 5 T Ine a N N E doe ieens duane duastoueadtomeaean aoe ieeaucaseus 5  Chapter 2  Using the Configuration Software to Setup the ST50 ST75 0 00 0    ccccccccsssseeeeceeeeeeeeeeaaaaseeeeeseeeeeeceeeeeeeeeaaaas 6  Pctine Up ald  Selecting Process UniS aiar A N N 6  Seneu he Lime Lype Iid SIZE ee me
9. as  been selected it will continue to update about once a second  and will stop when the Configuration Software is moved away from  the ST50 ST75 IR illumination area  or the    Return    button is tap     There are 2 possible screens  depending on the parameters selected during setup  the basic screen will always show Process Flow  and Temperature  And the alternative screen will show Process Flow  Temperature and Totalizer values   Getting Process Data    e Tap on the    Process    button of the main screen  to open the Process information screen     Basic Screen          Return         e Tap On the    Get All    button to read the current setup of the ST50 ST75     Basic Screen with Totalizer active       Return        This page is subject to proprietary rights statement on last page  11 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC ST50 ST75 PDA SOFTWARE    INTENTIONALLY LEFT BLANK    This page is subject to proprietary rights statement on last page  12 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC ST50 ST75 PDA SOFTWARE    Chapter 4  Using the Configuration Software to Read the ST50 ST75 ID Information    The Configuration Software provides access to ST50 ST75 manufacturing information  such as Unit Serial Number  ST50 ST75  Firmware Version  User Name and Shop Order number  called Shop Name     This information helps to identify the flow meter  and can be used by technical support personnel to retrieve factory records  and  help resolve filed problem
10. e    019819 01 exe    to start the installation     e The installation software will load the    Configuration Software prc       file into the    Palm Add on subdirectory    e Press the    HOTSYNC    button on the PDA or cradle to load the Configuration Software into the PDA  Congratulations you  are ready to use the Configuration software     This page is subject to proprietary rights statement on last page  3 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC ST50 ST75 PDA SOFTWARE    Step2  Setting up the Configuration Software Preferences  e Ifyou are going to be communicating with the STS0 ST75 through the IrDA link  you need do nothing  The instructions in  this step are to be use only if you are going to use other than the default setting     e From the applications menu in the PDA tap the i  FCI ST50 Icon and the Configuration Software f    opening screen will display        HotSync    9    Phone Pad       F FLUID COMPONENTS  PPL  A  INTERNATIONAL LLC       To Do List       e Tap Start to open up to the    Top Level Functions    screen of the Configuration Software     e Tap on the    FCI    ST50 75    to open the options tab tap on    Preferences                   a FLUID COMPONENTS rrererences   OMPONENTS  ye INTERNATIONAL LLC About ST50__sJATIONAL LLC  Exit to Palm OS        A Jj Aland lfa A Ty   g   i 2    DIAGNOSTICS   UTILITIES    The    Preferences    screen sets up 3 items        1  The ability to select the RS232 Serial port by  taping on the check box 
11. e assigned to DAC 1 or    V OUTPUT        e Tap on the    Set    button to down load the configuration data to the    ST50 ST75   Pulse Out Setup _ Return      l Out  Flow V Out  Flow  l Out  Temp V Out  Temp    _Get_     set       This page is subject to proprietary rights statement on last page  8 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC    Configuring and Setting up the Pulse Output Channels    ST50 ST75 PDA SOFTWARE    The Pulse Output Channels are used to Provide Totalizer and Process Alarm information  Totalizer information is provided  as a voltage pulse  and Alarm information is provided as a Voltage    State Transition     Low To    High or High to Low   There are 2 channels  Channel 1 is a Source Channel  and Channel 2 is a Sink Channel   See ST50 ST75 Installation    Instructions for information on how to connect to these 2 channels      From the Setup Menu  tap on the    Output Config    button to open the setup screen for the ST50 ST75 Analog Channel    selection  and the Pulse Output channel setup     Tap on the    Pulse Out Setup    button to open the Pulse Output  Channels Configuration screen     Tap On the    Get All    button to read the current setup of the  ST50 ST75     Chose the signal to be assigned to the Source Channel   Choose the signal to be assigned to the Sink Channel     Select and enter the new values for    Pulse Factor     and    Sample  Period     In most cases these values do not do change from the  factory settings     If t
12. eeees 12  Using the Configuration Software to Restore the ST50 ST75 Factory Set Parameters               ccccccccceceeeseeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeees 12  Using the Configuration Software to View ST50 ST75 Process and Instrument Faults                cccccccccceeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeees 12  Chapter 6  Using the Configuration Software to Access the ST50 ST75 DIAGNOSTIC Functions            0000000000    13  Using the Configuration Software to Troubleshoot the ST50 ST75 With the Diagnostic Functions                 ccccceceeeeeeeeeeeeeeeeeeees 13  Appendix A  The Configuration Software Quick Users Guide                 cccccccccccssssssssseseececeeeeeeeeeeeaasseeeeeseeeeeeeeeeeeeeeeasaaaaeeeeees 14  Appendix B  List of Configuration Software Commands                     cccccccccc cece cease eeeeeeeseeeeeeeeeeeeeeeeeaaaeeeeeeeeeeeeeeeeeeeeeeeaaaaaaeeesses 15    This page is subject to proprietary rights statement on last page  1 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC S1T50 ST75 PDA SOFTWARE    About This Manual  This manual is designed to quickly familiarize you with the use of the Configuration Software  It describes all you need to know    to effectively use the configurator during the commissioning  operational  and maintenance phases   e Viewing Process data  Flow  Temperature  and Totalizer   e Configuring the ST50 ST75 i e  Pipe size  Type  Process units  etc    e Troubleshooting    e Viewing the ST50 ST75 Configuration    After you become
13. for    Use Serial Comm  Port     You will need a serial cable kit FCI P N  014108 02 connected to the ST50 ST75   The  ST50 ST75 must be set for serial wire  communication  and a PDA specific serial cable  may be required     2     Timeout    sets the time the configurator waits or  response from the ST50 ST75 before indicating a  problem        3     Password    is needed for some of the utility  functions  More of this in chapter 5     NOTE  When using the Configuration Software and the ST50 ST75 in serial cable mode  remember to change the jumper   JP5  in the ST50 ST75 to RS232 communication position     This page is subject to proprietary rights statement on last page  4 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC ST50 ST75 PDA SOFTWARE    Step3  Connecting to the STS0 ST75  The recommended method for connecting to the STS50 ST75 with the Configuration Software is by means of the IR Link     IR interface  For best results on establishing a good IR link between the ST50 ST75 and the PDA  place the PDA in front of the  ST50 ST75 as shown below  at a distance of no more than 5 feet        The ST50 S75 can also be connected to the  Configuration Software by means of an RS322 Serial  cable     The serial cable attaches on the back of the ST50 ST75                          _ gt   with the other side connecting to the PDA serial  connection        This page is subject to proprietary rights statement on last page  5 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL
14. he    Flow Alarm    has been selected  then also select the pulse  transition format Low to High or High to Low     Select the Flow Alarm set point  It is selectable in SFPS      Some Definitions       Sey  TAS  re    Totalizer  Flow Alarm Flow Alarm    Geta    set   Return    Sample Period  This is the    Pulse Output    channel update rate  Pulse output is not a continuous outputting signal  but a new    value is updated based on the Sample Period parameter     Pulse Factor  This is the equivalency number to the corresponding instrument volumetric flow units   1 e  1 pulse is equal  to 1 NCMH for a factor of 1  if NCMH are the volumetric units selected     This page is subject to proprietary rights statement on last page     O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC ST50 ST75 PDA SOFTWARE    INTENTIONALLY LEFT BLANK    This page is subject to proprietary rights statement on last page  10 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC ST50 ST75 PDA SOFTWARE    Chapter 3  Using the Configuration Software to Read the ST50 ST75 Process Information    The Configuration Software Provides access to 3 Process parameters monitored by the ST50 ST75 Flow meter  these parameters  are Process Flow  Process temperature and Process totalization of flow  This chapter will show you how to use the Configuration  Software to read these parameters     Once the link is established between the STS50 ST75 and the Configuration Software  and the Process information reading h
15. ified    NOTICE OF PROPRIETARY RIGHTS    This document contains confidential technical data  including trade secrets and proprietary information which is the property of Fluid Components  International LLC  FCI   Disclosure of this data to you is expressly conditioned upon your assent that its use is limited to use within your company    only  and does not include manufacture or processing uses   Any other use is strictly prohibited without the prior written consent of FCI     Visit FCI on the Worldwide Web  www fluidcomponents com  1755 La Costa Meadows Drive  San Marcos  California 92078 USA   760 744 6950   800 854 1993   Fax 760 736 6250       Fluid Components International LLC  FCI  All Rights Reserved O6EN003372 Rev  A    
16. our PDA  To Install the Configuration Software into the PDA a PC with Microsoft Windows is required     The minimum requirements for Windows computers are as follows   A Pentium class PC running one of the following operating systems     e Windows 98   e Windows 2000 Pro  requires administrator rights to run the Configuration Software Installation software    e Windows XP Home or Pro  requires administrator rights to run the Configuration Software Installation software   e Less than 1 megabyte  MB  of available hard disk space   e VGA monitor or better   e CD ROM or DVD ROM drive   e Mouse    e One available USB port or serial port for the PDA   s serial cradle cable   Step1  Run the Configuration Software Installation  P N 019819 01 exe     WARNING  If you are using Windows 2000 or Windows XP you need to be in administrator mode  or you need to have your  IT administrator install it     e Confirm that the PDA is connected to the PC through the Cradle cable   e The installation software is located on the FCI Provided CD  load the CD into the CD ROM drive     e During the Installation you will be asked for a pass word  there is label attached to the back of CD case with the password   the password is case sensitive  be sure to enter the information as it appears in the label  Have the information ready before  you Start the installation     e The installation software will run automatically  if it doesn   t  open the CD directory with windows explorer  select and click  on fil
17. pervision of an  FCI trained technical personnel     e Tap on the    YES     RESET    button  At this time the ST50 ST75 will break the  communication link and the Configuration Software will respond with an error  message  Communication has to be re established with the STS0 ST75     FCIST50 75    Unexpected data  received   Please try again  O    Return      e After restoring communication tap on the    Return    button to return to the Utilities Screen       OK_        Using the Configuration Software to Display the ST50 ST75 Configuration and Calibration Parameters    e From the Utilities Menu tap on the    Display Parameter  Memory    button to open the Configuration and        Coefficient 2 2   366 724  lt   Calibration Parameters screen  Coefficient 2 3  aoe  e Tap On the    Get All    button to read the current setup of Coefficient 2 4   1145938 0  the ST50 ST75  Coefficient 2 5  162140 093    Scroll up and down to view the Configuration and Break Point     calibration parameters     Delta R Min   Delta R Max     e This screen allows viewing only  Parameters cannot be  changed   e Tap on the    Return    button to return to the Utilities Menu        This page is subject to proprietary rights statement on last page  15 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC ST50 ST75 PDA SOFTWARE    Using the Configuration Software to Display and Set the Calibration values of the ST50 ST75    e From the Utilities Menu tap on the    Calibration Values    button to open 
18. s and issues     Getting the ST50 ST75 Instrument Information     e Tap on the    Unit ID    button of the main screen   to open the ST50 ST75 Manufacturing  Information screen  ST50 75 Info      e Tap On the    Get All    button to read the current  setup of the STS0 ST75        This page is subject to proprietary rights statement on last page  13 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC ST50 ST75 PDA SOFTWARE    INTENTIONALLY LEFT BLANK    This page is subject to proprietary rights statement on last page  14 O6EN003372 Rev  A    FLUID COMPONENTS INTERNATIONAL LLC ST50 ST75 PDA SOFTWARE    Chapter 5  Using the Configuration Software to Access the ST50 ST75 Utilities Functions    The Configuration Software gives access to the five ST50 ST75 Utility Functions  These functions help to Check the calibration  and configuration parameters  help restore factory setting  and display any process or instrument alarms that may have been set     e Tap on the    Utilities    button of the main screen     Reset ST50  ST75    to open the ST50 ST75 Utilities selection screen  Display Parameter Memory  Calibration Values    Factory Restore  Process  amp  System Faults       Using the Configuration Software to Reset the STS50 ST75 Flowmeter    From the Utilities Menu tap on the    Reset ST50 ST75    button to open the Soft Reset screen of the ST50 ST75  This command  will force a re initialization of the STS0 ST75 software  This command should only be use with the advice and su
19. the  calibration Parameters screen     Coefficient 1 1  lt    Coefficient 1 2 e Tap On the    Get All    button to read the current setup of the ST50 ST75   Coefficient 1 3  Coefficient 1 4  Coefficient 1 5    e To Change the value of a Parameter  scroll up or down until you find the  parameter  tap on it to select it  then enter the new value in the value field     Coefficient 2 1 e Repeat the previous step for as many parameters as need to be changed   Coefficient 2 2  Coefficient 2 3 e When you are ready to down load the new values to the ST50 ST75 tap on the    Coefficient 2 4    Set All    button        Using the Configuration Software to Restore the STS50 ST75 Factory set Parameters    e From the Utilities Menu tap on the    Restore    button to  open the Factory Restore screen  By performing a restore   the parameters that have changed in the field  will be lost   and the ST50 ST75 Factory calibration setting will be  reloaded into the current active memory of the instrument     e Tap on the    Yes     Restore    button to initiate the Factory  Restore     e Tap on the    Return    button to return to the Utilities  Menu  YES   RESTORE       Using the Configuration Software to View the ST50 ST75 Process and System Faults  The ST50 ST75 has nine possible alarm conditions it can detect  8 are process related and 1 is instrument related     e Tap On the    Get All    button to read the current setup of the ST50 ST75  At  this time any fault conditions that exist will show
20. x 1000    Display as   of full scale        x1 Million    e Remember any changes that you make here need to be followed by a change  7 on the display stickers  to reflect the new display mode        Get all _ acta  Return  e Tap on the    Set    button to down load the configuration data to the ST50 ST75   ko    Configuring and Setting up the Analog Outputs  Zero and Span Analog Trim Setting    The Zero and Span setting for Flow and Temperature are set at the factory  and in most cases there is no need to adjust  them in the field     From the Setup Menu  tap on the    Output Cal    button to  open the setup screen for the Zero and Span settings of the  two analog output channels  Each of the 2 channels gets 2  sets of Zero and Span values that correspond to Flow and  Temperature     Tap On the    Get All    button to read the current setup of the  ST50 ST75     Select the fields you want to change  enter the new values     Tap on the    Set    button to down load to the ST50 ST75 the  Parameters Values        Assigning a Signal to the 2 DAC   s    Typically the Factory Sets DAC 0  I OUTPUT  to Flow and DAC 1  V OUTPUT  to Temperature        e From the Setup Menu tap on the    Output Config    button to open the  setup screen for the ST50 ST75 Analog Channel selection  and the  Pulse Output channel setup     e Tap On the    Get All    button to read the current setup of the  ST50 ST75     e Choose the signal to be assigned to DAC 0 or    I OUTPUT      e Choose the signal to b
    