Home
        Rat Premixed Multi-Analyte Kit
         Contents
1.                               
2.   Protect the Streptavidin PE from light during handling and storage    1  Centrifuge the Streptavidin PE vial for 30 seconds at 1000 x g prior to removing the cap    2  Gently vortex the vial  taking precautions not to invert the vial     3  Dilute the Streptavidin PE Concentrate to a 1X concentration by adding 220 uL of  Streptavidin PE to 5 35 mL of Wash Buffer  This provides enough Streptavidin PE to assay one  96 well microplate  If assaying less than 96 wells  adjust these volumes accordingly     6 For research use only  Not for use in diagnostic procedures     INSTRUMENT SETTINGS  Luminex MAGPIX analyzer     a  Assign the microparticle region for each analyte being measured   see Certificate of Analysis for details    b  50 events bead   c  Sample size  50 uL    d  Collect Median Fluorescence Intensity  MFI     Luminex 100 200 and Bio Rad Bio Plex analyzers     Note  Calibrate the analyzer using the proper reagents for superparamagnetic microparticles  refer  to instrument manual      a  Assign the microparticle region for each analyte being measured  see Certificate of Analysis  for details    b  50 events bead   c  Minimum events  0   d  Flow rate  60 uL minute  fast    e  Sample size  50 uL   f  Doublet Discriminator gates at approximately 8000 and 16 500       g  Collect MFI    Note  The CAL2 setting for the Bio Rad Bio Plex analyzer should be set at the low RP1 target value     www RnDSystems com 7    ASSAY PROCEDURE    Bring all reagents and samples to room tempe
3.  e Deionized or distilled water    e Multi channel pipette  manifold dispenser  or automated dispensing unit    e 500 mL graduated cylinder    e Polypropylene test tubes for dilution of standards and samples     e Horizontal orbital microplate shaker  0 12  orbit  capable of maintaining a speed of  800   50 rpm     e Microcentrifuge     www RnDSystems com 3    SAMPLE COLLECTION  amp  STORAGE  The sample collection and storage conditions listed below are intended as general  guidelines  Sample stability has not been evaluated     Cell Culture Supernates   Remove particulates by centrifugation and assay immediately or  aliquot and store samples at  lt   20   C  Avoid repeated freeze thaw cycles     Serum   Allow blood samples to clot for 2 hours at room temperature before centrifuging for  20 minutes at 2000 x g  Remove serum and assay immediately or aliquot and store samples at   lt   20   C  Avoid repeated freeze thaw cycles     Plasma   Collect plasma using EDTA or heparin as an anticoagulant  Centrifuge for 20 minutes  at 2000 x g within 30 minutes of collection  Assay immediately or aliquot and store samples at   lt   20   C  Avoid repeated freeze thaw cycles     Note  Citrate plasma has not been validated for use in this assay     SAMPLE PREPARATION    To determine the appropriate dilution for each analyte  refer to the table located in the  following link http   www rndsystems com Products LXSARM     Serum and plasma samples require at least a 2 fold dilution  A suggeste
4. Magnetic Luminex    Screening Assay    Rat Premixed Multi Analyte Kit    Catalog Number LXSARM    For the simultaneous detection of multiple rat biomarkers in cell culture supernates  serum   and plasma     This package insert must be read in its entirety before using this product   For research use only  Not for use in diagnostic procedures     TABLE OF CONTENTS  SECTION    ATRODU CTION se sans ese GEE RE oe ee GEE ee ek eo ER ee ee EG  PRINGIELE OF EIE AS SA dees meisie sees ego sd gee ee see ee E ee ee Ve een ees ee ee  LIMITATIONS OF THE PROCE DU    cere enn ees ee ordes ee ee ese ee ee es ee  ae  ie  Ml  ER EE EE OE EA  PRECAUTION es is E ee ee OE E  MATERIALS PROVIDED  amp  STORAGE CONDITIONS esse se se seen se se se se een Ge Ge ee ee Ge ee  OTHER SUPPLIES REQUIRED sr eende Ge ee re err er ee ie ee ie de oe ee ee  SAMPLE COLLECTION  amp  STORAGE    sees esse se se see eene se eene Gee se ee gee ee ee ee Se Se eg ee ee eg  SAMPLE PREPARATION   ie messe eie eine ene oe oe e eien eek ese sees ese ee ee ede  KEAGENTPREPARATION es iedere etes de ese een eed er ee ee Pe ie ee oe seed ie  DILUTED MICROPARTICLE COCKTAIL PREPARATION    sesse se sesse se se see es se se se bee gee Gee ee  DILUTED BIOTIN ANTIBODY COCKTAIL PREPARATION       sesse sesse se se see ee se se se seges Ge se ee  SIREPTAVIDIN PE PREPARAIIO N    sesse eie segesse ee esse ee ende seksie see gewees be ees ee ees ed  INSTRUMENT SETTINGS si esse ie E N ee ie E De rrr  AS PROCEDURE is oie ee GE GR ed ee Ede E 
5. and detection platforms     Analyte specific antibodies are pre coated onto color coded magnetic microparticles   Microparticles  standards and samples are pipetted into wells and the immobilized antibodies  bind the analytes of interest  After washing away any unbound substances  a biotinylated  antibody cocktail specific to the analytes of interest is added to each well  Following a wash to  remove any unbound biotinylated antibody  streptavidin phycoerythrin conjugate   Streptavidin PE   which binds to the biotinylated antibody  is added to each well  A final wash  removes unbound Streptavidin PE  the microparticles are resuspended in buffer and read  using the Luminex MAGPIX Analyzer  A magnet in the analyzer captures and holds the  superparamagnetic microparticles in a monolayer  Two spectrally distinct Light Emitting  Diodes  LEDs  illuminate the microparticles  One LED identifies the analyte that is being  detected and the second LED determines the magnitude of the PE derived signal  which is in  direct proportion to the amount of analyte bound  Each well is imaged with a CCD camera  Kits  can also be used with Luminex 100 200 or a Bio Rad Bio Plex dual laser  flow based systems     www RnDSystems com    LIMITATIONS OF THE PROCEDURE      FOR RESEARCH USE ONLY  NOT FOR USE IN DIAGNOSTIC PROCEDURES   e The kit should not be used beyond the expiration date on the kit label   e Do not mix or substitute reagents with those from other lots or sources     e If samples generat
6. ator Diluent RD6 52 serves as the  blank  Refer to the Certificate of Analysis for values of Standard 1     100uL    100 uL 100uL   100 pL 100 uL    RA OR A OR RA       500 uL Std     D    STANDARD    Standard Cocktail Standard 1 Standard2 Standard3 Standard4  Standard5 Standard 6       www RnDSystems com 5    DILUTED MICROPARTICLE COCKTAIL PREPARATION    1  Centrifuge the Microparticle Cocktail vial for 30 seconds at 1000 x g prior to removing  the cap     2  Gently vortex the vial to resuspend the microparticles  taking precautions not to invert  the vial     3  Dilute the Microparticle Cocktail using Assay Diluent RD1W in the mixing bottle provided     Number of Wells Used Microparticle Cocktail   Assay Diluent RD1W  96 500 uL   5 00 mL  48 250 uL   2 50 mL  24 125 uL   1 25 mL    Note  Protect microparticles from light during handling  Prepare microparticles within 30 minutes  of use  Diluted microparticles cannot be stored     DILUTED BIOTIN ANTIBODY COCKTAIL PREPARATION    1  Centrifuge the Biotin Antibody Cocktail vial for 30 seconds at 1000 x g prior to removing  the cap     2  Gently vortex the vial  taking precautions not to invert the vial   3  Dilute the Biotin Antibody Cocktail in Assay Diluent RD1W  Mix gently     Number of Wells Used Biotin Antibody Cocktail   Assay Diluent RD1W  96 500 uL   5 00 mL  48 250 uL   2 50 mL  24 125 uL   1 25 mL    STREPTAVIDIN PE PREPARATION   Use a polypropylene amber bottle or a polypropylene test tube wrapped with aluminum  foil
7. d 2 fold dilution is  75 uL of sample   75 uL of Calibrator Diluent RD6 52  Mix thoroughly     High abundance biomarkers may require additional dilution such as 50  or 200 fold     A suggested 50 fold dilution is 10 uL of sample   490 uL of Calibrator Diluent RD6 52   Mix thoroughly     A suggested 200 fold dilution can be achieved by adding 10 uL of sample to 90 uL of  Calibrator Diluent RD6 52  Complete the 200 fold dilution by adding 10 uL of the diluted  sample to 190 uL Calibrator Diluent RD6 52     4 For research use only  Not for use in diagnostic procedures     REAGENT PREPARATION    Bring all reagents to room temperature before use     Wash Buffer   If crystals have formed in the concentrate  warm to room temperature and mix  gently until the crystals have completely dissolved  Add 20 mL of Wash Buffer Concentrate to  deionized or distilled water to prepare 500 mL of Wash Buffer     Standard   Reconstitute the Standard Cocktail with Calibrator Diluent RD6 52  Refer to the  Certificate of Analysis for the reconstitution volume  Allow the standard to sit for a minimum of  15 minutes with gentle agitation prior to making dilutions     Use polypropylene tubes  Pipette 500 uL of the reconstituted Standard Cocktail into a tube  labeled Standard 1  Pipette 200 uL of Calibrator Diluent RD6 52 into the remaining tubes  Use  Standard 1 to produce a 3 fold dilution series  below   Mix each tube thoroughly before the  next transfer  Standard 1 serves as the high standard  Calibr
8. e values higher than the highest standard  further dilute the samples with  Calibrator Diluent and repeat the assay       Any variation in standard diluent  operator  pipetting technique  washing technique   incubation time or temperature  and kit age can cause variation in binding     e Variations in sample collection  processing  and storage may cause sample value differences     e Discepancies may exist in values obtained for the same analyte utilizing different  technologies       Magnetic Luminex Screening Assays afford the user the benefit of multianalyte analysis of  biomarkers in a single sample  A multipurpose diluent is used to dilute samples  if necessary   and provide accurate estimates of natural analytes in cell culture supernates  serum  and  plasma       Only the analytes listed on the enclosed Certificate of Analysis can be measured with  this kit     TECHNICAL HINTS      When mixing or reconstituting protein solutions  always avoid foaming     e To avoid cross contamination  change pipette tips between additions of each standard level   between sample additions  and between reagent additions  Also  use separate reservoirs for  each reagent     e To ensure accurate results  proper adhesion of plate sealers during incubation steps is  necessary     e Protect microparticles and Streptavidin PE from light at all times to prevent photobleaching     PRECAUTIONS    Some components in this kit contain ProClin   which may cause an allergic skin reaction  Avoid  breat
9. ere oes  EALEULATIO N OF BESLIS ss ie Dee ei ee ee ee de ee sey Ge  EALIBRATION  osse ee ee ee Ge OEA Ge E NE  VEE iS  ER OE tre verre A re rrrrrer terre rrrtrer ir rerrrrerr frre trier ire errr erect    MANUFACTURED AND DISTRIBUTED BY     USA  amp  Canada   R amp D Systems  Inc    614 McKinley Place NE  Minneapolis  MN 55413  USA   TEL   800  343 7475  612  379 2956 FAX   612  656 4400  E MAIL  info RnDSystems com    DISTRIBUTED BY     UK  amp  Europe   R amp D Systems Europe  Ltd    19 Barton Lane  Abingdon Science Park  Abingdon OX14 3NB  UK  TEL   44  0 1235 529449 FAX   44  0 1235 533420   E MAIL  info RnDSystems co uk    China   R amp D Systems China Co   Ltd    24A1 Hua Min Empire Plaza  726 West Yan An Road  Shanghai PRC 200050  TEL   86  21  52380373 FAX   86  21  52371001   E MAIL  info RnDSystemsChina com cn    INTRODUCTION    This kit contains the components required to screen multiple rat biomarkers in cell culture  supernate  serum  and plasma samples in multiplexed sandwich ELISAs     Magnetic Luminex   Screening Assays can be used to assess the levels of biomarkers of your  choosing in a single sample  For ease of use  the microparticles are premixed in one vial as are  the biotinylated detection antibodies     PRINCIPLE OF THE ASSAY    Magnetic Luminex Performance Assay multiplex kits are designed for use with the Luminex  MAGPIX   CCD Imager  Alternatively  kits can be used with the Luminex 100 200    or Bio Rad    Bio Plex    dual laser  flow based sorting 
10. hing mist     Wear protective gloves  clothing  eye  and face protection  Wash hands thoroughly after  handling  Please refer to the MSDS on our website prior to use     2 For research use only  Not for use in diagnostic procedures     MATERIALS PROVIDED  amp  STORAGE CONDITIONS    Store the unopened kit at 2 8   C  Do not use past kit expiration date   This kit contains sufficient materials to run multiplex assays on one 96 well plate     STORAGE OF OPENED  DILUTED   PART PART   DESCRIPTION OR RECONSTITUTED MATERIAL    Rat Standard 893115   2 vials of recombinant rat biomarkers in a buffered Once reconstituted  any   Cocktail A protein base with preservatives  lyophilized  remaining standard must be  discarded  Use fresh a standard  for each assay     Microparticle Cocktail preservatives  1 month at 2 8   C   Rat Premixed 894783   0 6 mL ofa concentrated biotin antibody cocktail with      Streptavidin PE 893535   0 25 mL of a concentrated streptavidin phycoerythrin ae ES EN  Calibrator Diluent 895438   2 vials  21 mL vial  of a buffered protein base with  oe eean ayes rt  Concentrate surfactant with preservative  May turn yellow over time   volumes  and concentrations for the provided Standard s         Provided this is within the expiration date of the kit        OTHER SUPPLIES REQUIRED    e Luminex MAGPIX  Luminex 100 200  or Bio Rad Bio Plex analyzer with X Y platform     Hand held microplate magnet or platewasher with a magnetic platform    e Pipettes and pipette tips   
11. rature before use  It is recommended that  all samples and standards be assayed in duplicate     Note  Protect microparticles and Streptavidin PE from light at all times     a N    3     10       Prepare all reagents  standards  and samples as directed in the previous sections       Resuspend the diluted microparticle cocktail by inversion or vortexing  Add 50 uL of the    microparticle cocktail to each well of the microplate     Add 50 uL of Standard or sample  per well  Securely cover with a foil plate sealer  Incubate  for 2 hours at room temperature on a horizontal orbital microplate shaker  0 12    orbit  set  at 800   50 rpm  A plate layout is provided to record standards and samples assayed       Using a magnetic device designed to accommodate a microplate  wash by applying the    magnet to the bottom of the microplate  removing the liquid  filling each well with Wash  Buffer  100 uL  and removing the liquid again  Complete removal of liquid is essential for  good performance  Perform the wash procedure three times     Note  Refer to the magnetic device user manual for proper wash technique using a flat bottom  microplate       Add 50 uL of diluted Biotin Antibody Cocktail to each well  Securely cover with a foil plate    sealer and incubate for 1 hour at room temperature on the shaker set at 800   50 rpm       Repeat the wash as in step 4       Add 50 uL of diluted Streptavidin PE to each well  Securely cover with a foil plate sealer and    incubate for 30 minutes at 
12. room temperature on the shaker set at 800   50 rpm       Repeat the wash as in step 4       Resuspend the microparticles by adding 100 uL of Wash Buffer to each well  Incubate for    2 minutes on the shaker set at 800   50 rpm     Read within 90 minutes using a Luminex or Bio Rad analyzer   Note  Resuspend microparticles immediately prior to reading      Samples may require dilution  See Sample Preparation section     For research use only  Not for use in diagnostic procedures     CALCULATION OF RESULTS    Use the Standard concentrations on the Certificate of Analysis and calculate 3 fold dilutions for  the remaining levels  Average the duplicate readings for each standard and sample and  subtract the average blank Median Fluorescence Intensity  MFI      Create a standard curve for each analyte by reducing the data using computer software  capable of generating a five parameter logistic  5 PL  curve fit     If samples have been diluted  the concentration read from the standard curve must be  multiplied by the dilution factor     CALIBRATION    This assay is calibrated against highly purified recombinant rat biomarkers produced at  R amp D Systems     All trademarks and registered trademarks are the property of their respective owners     www RnDSystems com 9    PLATE LAYOUT    to record standards and samples assayed     ECX XXXXXX    eC OOOO O00   ECX XXXXXX    ECX XXXXXX    6 OOOOOOOO 7     DOOOOOOO      EESSESSSD     OOOOOOOO  7     OOOO O00   6 OOOOOOOO  OOOO OOOO 7       
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Untitled - Shopify  屋外用ドームハウジング    総合カタログ2015  Mode d`emploi ROTA 400  Instruction Sheet - Hair Dryer PNP-210 Register  manual do usuário  AVM FRITZ!WLAN Stick International  Agilent 83433A 光トランスミッタ ユーザーズ・ガイド  XL Vu VideoProbe® - GE Measurement & Control    Copyright © All rights reserved. 
   Failed to retrieve file