PowerPac™ JR II Power Supply Instruction Manual - Bio-Rad
Contents
1. 5 mm OD x 20 mm length fuse Reinsert the fuse holder into its original position and screw back on Problem What to check Resolution Run times are Verify that the Run times will vary longer or gel and running with different gel shorter than buffers are made types and buffer expected properly systems Prepare fresh materials and rerun if necessary Output voltage Contact Bio Rad may be outside Laboratories for the specification suggestions on verifying output voltage Red LED Over current load Verify that the gel and blinks greater than 250 mA buffer are made properly If problem persists contact Bio Rad Technical Support Note PowerPac JR II has no repairable or replaceable parts other than the fuse Section 4 Specifications Input power Input power connector Output voltage Output current Output power Terminals Operating temperature Humidity Dimensions W x D x L Regulatory compliance Section 5 90 264 VAC 47 63 Hz 60 W max IEC 320 appliance inlet 100 or 200 V DC 200 mA max 40 W max 2 pair recessed female banana jacks 4 40 C 0 98 noncondensing 13 x 10x 7 cm EN61010 1 safety see Declaration of Conformity for details Product Information Catalog Description 165 5048 PowerPac JR II Power Supply 100 120 220 240 V includes power cord 1of7 Bio Rad Laboratories Inc Life Science Web site www bio rad com USA 800 4BIORAD A
2. V output One or two cells can be connected in parallel to the power supply Note Do not stack PowerPac JR II power supplies To insure adequate cooling of the power supply be sure that there is at least 10 cm clearance around the power supply Output voltage selector switch F777 PowerPac JR ul 100 V 200 V N Output jacks An Off C O f Overload A Fig 1 Front panel 1 2 Unpacking When you receive the power supply carefully inspect the container for any damage that may have occurred during shipping Severe damage to the container may indicate damage to the power supply itself If you suspect damage to the unit immediately file a claim with the carrier in accordance with their instructions before contacting Bio Rad Laboratories Unpack the power supply The PowerPac JR II power supply is shipped with the following e Power supply unit e AC power cord e User manual If any part is missing or damaged contact Bio Rad Laboratories immediately Section 2 Setup and Operation Step Procedure Description 1 Connect the power cord to an appropriate outlet and the electrophoresis cell s to the power supply 2 Select the output voltage 3 Turn on the power 4 Turn off the power The power leads are color coordinated with the power supply output terminals A indicates high voltages may be present Move the selector switch on the front panel to either 100 V or 200 V position P
3. 010 1 safety standards when used in accordance with the instruction manual and when used with EN61010 1 compliant electrophoresis cells and accessories 1 Always connect the power supply to an appropriate grounded AC outlet using the 3 prong AC power cord provided with the power supply 2 Bio Rad electrophoresis cells have molded two prong plugs which are inserted into the power supply s high voltage output jacks These plugs have been IEC 1010 certified for safety com pliance for use with PowerPac power supplies Use of other plugs or banana jacks including the PowerPac adaptor is done at the user s own risk and is not recommended by Bio Rad When inserting and removing the molded two prong plug always grasp the plug by the molded support at the rear of the plug Do not grasp the individual prong ends Table of Contents Section 1 Introduction BPP Bn 1751 01 1 2 Unpacking Section 2 Setup and Operation Section 3 Maintenance and Troubleshooting 3 1 Maintenance cccecccecceeceeeeeeeeteeeeeeeaeeeeeeeeeeeeeeee 3 2 Troubleshooting Section 4 Specifications ccccccccsessseeeecseeessessesseeeeeeeees 4 Section 5 Product Information cccccccecsssesseseeeeees 4 Power switch Section 1 Introduction 1 1 Overview The PowerPac JR II power supply provides constant voltage to instruments used in electrophoresis applications It is capable of supplying either 100 V 10 V or 200 V 20
4. PowerPac JR Il Power Supply Instruction Manual Catalog Number 165 5048 For Technical Service Call Your Local Bio Rad Office or in the U S Call 1 800 4BIORAD 1 800 424 6723 Safety Caution Warning PowerPac power supplies use high output voltages that are electrically grounded to minimize the risk of electrical shock to the user The following guidelines should be followed when using the power supply PowerPac JR II is intended for laboratory use only This product conforms to the Class A standards for electromagnetic emissions intended for laboratory equipment applications Emissions from this product may interfere with some sensitive appliances when placed nearby or in the same circuit as those appliances The user should be aware of this potential and take appropriate measures to avoid interference PowerPac power supplies have passed tests for operation at temperatures between 4 and 40 C with relative humidity between 0 and 98 noncondensing Operating the power supply outside these conditions is not recommended by Bio Rad and will void the warranty This instrument should not be modified altered or serviced by the user in any way User modification alteration or service of this instrument will void the manufacturer s warranty and compliance with EN61010 1 and create a potential safety hazard for the user Bio Rad is not responsible for any consequent injury or damages PowerPac JR II complies with EN61
5. ress the power switch on the front of the unit to the on position The LED indicator light will be lit green At the end of the run press the power supply switch to the off position Note For 220 V outlets select 220 V option on back panel of power supply before connecting power supply Section 3 Maintenance and Troubleshooting 3 1 Maintenance The PowerPac JR II power supply requires very little maintenance to ensure reliable operation To clean the case first unplug the power supply Use a damp cloth to wipe down the outer case Avoid wetting the connectors on the front panel and on the rear of the unit 3 2 Troubleshooting The front panel LED indicates the presence of voltage at the output jacks The following chart will aid with troubleshooting Problem What to check Resolution Green LED Verify that power If the electrophoresis blinks or is off supply is connected cell is connected to appropriate outlet disconnect the cell If that the power switch the LED stays on the is ON cell or cable is causing excessive current draw Verify that the gel and buffers are made properly and that the cables are in good condition If the electrophoresis cell is not connected the unit is defective and should be replaced Fuse may be blown Disconnect the power cord from the electrical outlet Unscrew fuse container located on back of power supply Remove fuse from the fuse holder Replace with 250 mA 250 V
6. ustralia 02 9914 2800 Austria 01 877 89 01 Belgium 09 385 55 11 Brazil 55 21 2527 3454 Canada 905 712 2771 China 86 21 63052255 Czech Republic 420 2 41 43 05 32 Denmark 44 52 10 00 Finland 09 804 22 00 Group France 01 47 95 69 65 Germany 089 318 84 0 Hong Kong 852 2789 3300 Hungary 36 1 455 8800 India 91 124 6398112 113 114 6450092 93 Israel 03 951 4127 Italy 39 02 216091 Japan 03 5811 6270 Korea 82 2 3473 4460 Latin America 305 894 5950 Mexico 55 52 00 05 20 The Netherlands 0318 540666 New Zealand 64 9 415 2280 Norway 23 38 41 30 Poland 48 22 331 99 99 Portugal 351 21 472 7700 Russia 7 095 721 1404 Singapore 65 6415 3188 South Africa 00 27 11 4428508 Spain 34 91 590 5200 Sweden 08 555 12700 Switzerland 061 717 9555 Taiwan 8862 2578 7189 2578 7241 United Kingdom 020 8328 2000 Sig 1103 4006135 Rev D